327.3 Usermanual CewePrometerA0172e-1a
327.3 Usermanual CewePrometerA0172e-1a
327.3 Usermanual CewePrometerA0172e-1a
User Manual
Ver. 1.0
Contents
Introduction........................................................................................................................5
About this user manual.................................................................................................5
Contacting us................................................................................................................5
Product description ...........................................................................................................6
Sealing of CewePrometer-W........................................................................................9
Sealing of CewePrometer-R.......................................................................................10
Connections ................................................................................................................10
Mechanical design......................................................................................................13
Function modules .......................................................................................................16
Measuring principles ..................................................................................................17
Configuration, reading and maintenance ......................................................................18
Connecting to CewePrometer.....................................................................................19
Basic configuration.....................................................................................................20
Overview of functions ................................................................................................22
Changing configuration..............................................................................................24
Working with configurations......................................................................................25
Reading.......................................................................................................................25
Information about the meter .......................................................................................26
Versions and version conflicts ...................................................................................27
Updating firmware .....................................................................................................27
Language ....................................................................................................................28
Resetting registers and logs........................................................................................28
Functions..........................................................................................................................29
Meter clock.................................................................................................................29
Energy registers..........................................................................................................30
Instant values..............................................................................................................31
Digital inputs and outputs...........................................................................................33
Display sequences ......................................................................................................36
Communications and security ....................................................................................37
Loggers.......................................................................................................................39
Alarms ........................................................................................................................41
Event log.....................................................................................................................43
Maximum demand......................................................................................................44
Historical registers......................................................................................................45
Time of use.................................................................................................................46
Transformer compensation.........................................................................................48
3
Contents
Power quality..............................................................................................................50
Miscellaneous.............................................................................................................52
Using the display .............................................................................................................53
Voltage order display layout ......................................................................................53
Navigating in display sequences ................................................................................54
Display layouts with active choices ...........................................................................54
Using the display for communication diagnostics .....................................................56
Appendix A Display Layouts .......................................................................................57
Appendix B - Events ........................................................................................................61
Appendix C Serial Communication Port RS232/422 ..................................................63
Appendix D - Module block diagram ..............................................................................65
Appendix E Frequently Asked Questions...................................................................66
Appendix F Calculation Principles ..............................................................................67
Appendix G Material Declaration.................................................................................72
Appendix H Connection Data ......................................................................................73
Introduction
Introduction
Thank you for choosing the CewePrometer.
The CewesPrometer is a multi-functional, electronic, electricity energy meter. It provides very high
accuracy in measuring electrical energy and also in measuring instantaneous quantities such as
current, voltage, power, frequency etc. CewePrometers extensive configurable functional features
together with the high accuracy enable application areas more numerous than for traditional electric
meters. Besides having well-designed traditional features such as meter registers, data logging and
tariff capability, there are also extensive functions for monitoring alarm conditions, harmonics,
measuring sags and swells and other power quality quantities.
Contacting us
For more information and technical support, please contact Cewe Instrument.
Internet
www.ceweinstrument.se
support@ceweinstrument.se
marknad@ceweinstrument.se
Telephone
Address
Cewe Instrument AB
Box 1006
SE-611 29 Nykping
Sweden
Product description
Product description
15
14
13
12
11
10
3
Figure 1 CewePrometer-W
10
15
14
13
12
16
Figure 2 CewePrometer-R
11
Product description
1
Display
Alarm LED
The alarm LED flashes when an alarm state has occurred. For alarms, such as
for high or low voltage, phase balance or similar external errors, the LED stops
flashing when the alarm state has passed. If the meter has had an internal error,
the alarm must be acknowledged and manually reset.
Terminal cover
Meter cover
10
11
12
Sealed button
The button is hidden under the front panel window and provides access to a fifth
configurable display sequence. Holding down the button for about two seconds
opens a system menu.
Product description
13
Step backward
The button is used to step backward to a previous display layout in the selected
display sequence, or to step downwards in a menu.
14
Menu/Apply
The Menu/Apply button is used to open the menu where one of four
configurable display sequences can be chosen. The button confirms choices
made or initiates other activities depending on the current display layout.
15
Step forward
The button is used to step forward to the following display layout in the selected
display sequence, or to step upwards in a menu.
16
Product description
Sealing of CewePrometer-W
See Figure 1 CewePrometer-W for location of the sealing points. The front panel window can be
sealed with lead seals (or comparable seals) and sealing wire through the sealing points. The seal
prevents access to the fourth sealed button and the optional backup battery for the real-time clock.
Battery cover
Sealed button
Figure 3 Under the front panel window is a fourth button and backup battery for the real-time clock.
The terminal cover can be sealed with lead seals (or comparable seals) and sealing wire through the
sealing points.
The meter cover can be sealed with lead seals (or comparable seals) and sealing wire through the
two sealing points under the terminal cover.
Sealing point
Figure 4 The meter cover can be sealed at the sealing rings on both sides under the terminal cover.
Product description
Sealing of CewePrometer-R
See Figure 2 CewePrometer-R for location of sealing points. The enclosure permits the meter to be
sealed at three points. The top and bottom covers are normally each sealed with a lead seal and with
sealing wire through a sealing screw and a fixed hole on the cover at the rear of the meter. The
exterior window on the meter front can be sealed on a sealing point with a lead seal and sealing
wire. In this way, the meter can be sealed so that it cannot be opened without breaking all the seals.
One of the buttons on the front is protected from access behind the sealed exterior window.
Battery
Sealed button
Under the front panel window is a fourth button and backup battery for the real-time clock.
Connections
Connections to the CewePrometer-W are made on the meter terminal under the terminal cover.
Connections to the CewePrometer-R are made on the rear panel of the meter subrack. The
connections are: measuring voltages, measuring currents, relay outputs, optically isolated digital
inputs, separate auxiliary power and connections to communication modules, if any.
Digital outputs
Digital inputs
Output #8
Output #1
I'L1 UL1
IL1
I'L2 UL2
IL2
I'L3 UL3
Input #1
IL3
Input #4
N
Separate aux. power
(Uaux)
10
Product description
Connections
UL1
UL2
UL3
N
B1
B2
B3
B0
Uaux1
Uaux2
B5
B6
IL1
I'L1
IL2
I'L2
IL3
I'L3
A01
A1
A02
A2
A03
A3
IN1
IN2
IN3
IN4
C4, C9
C3, C8
D4, D9
D3, D8
IN1
IN2
C4, C9
C3, C8
12 outputs, 0 inputs
OUT1 C0, C5
OUT2 C1, C6
OUT3 C2, C7
OUT4 D0, D5
OUT5 D1, D6
OUT6 D2, D7
OUT7 D3, D8
OUT8 D4, D9
OUT9 B7, B8
OUT10 B4, B9
OUT11 C3, C8
OUT12 C4, C9
A standardised Entrelec Essailec plug-in connector is used, of the same type as for other meters
according to standard DIN 43862. The cable connection is made in the subrack. The meter is
connected via the contact pins after insertion. The connector has a short-circuit protection feature
for electrical circuits, permitting the meter to be withdrawn and inserted into the subrack during
operation.
Auxiliary power
The CewePrometer is supplied with separate auxiliary power. Auxiliary power can be supplied both
with alternating current and polarity-independent direct current within a specified range. There are
two versions that have either a high or a low auxiliary voltage range.
Digital outputs
The CewePrometer's outputs are solid-state MOS-FET bipolar semiconductor relays, with normally
open contact function. Internal current limit protects the relay from being damaged by excessively
high current.
11
Product description
Digital inputs
The optically isolated digital inputs consist of opto couplers, where the LED is powered from an
external voltage signal through series resistance to limit the current. This means that an external
voltage must be connected between the passive pulse and level transmitter and the opto coupler's
LED, or the transmitter must deliver an active voltage signal to send pulses that are detected by the
optical input. The optically isolated inputs are unaffected by reverse polarity. Both DC and 50/60
Hz AC can be used.
Serial communication ports
The CewePrometer is always equipped with one optical port and is available in versions with or
without additional communication ports. A meter can be equipped with up to two extra ports, either
with the RS232 or RS422 interface.
The RS422 serial communication port is intended to be used to connect the CewePrometer
to a network for multi-drop communications. The CewePrometer is connected to the
communication equipment (RS422 converter) with two twisted-pair wires connected to a
5-pole connector.
For information on pin configuration, see Appendix C Serial Communication Port RS232/422
(pg. 63).
Communication port #2
Communication port #1
Communication port #1
Communication port #2
12
Product description
Mechanical design
Alternative
mounting
13
Product description
14
Product description
The subrack is also available in alternative connector versions that will fit other brands of meters
with the same type of connector. Contact Cewe Instrument for more information.
Enclosure and protective earth
CewePrometer-W
The enclosure consists of a meter base, meter cover, terminal cover and a plastic exterior window.
The meter is not connected with protective earth. There is also an optional 19 inch rack/panel
mounting kit available for to CewPrometer-W to be mounted side by side.
CewePrometer-R
Enclosure fits for installation of two meters in a 19" subrack in compliance with DIN438862. The
enclosure consists of a box made of surface-finished steel sheet and a front with a plastic window.
The meter is designed for protective earthing through the 19" subrack.
Isolation and personal safety
The electronic circuits are isolated from all connections to the supply mains with transformers.
Digital I/O connections are isolated with opto couplers and solid state relays. The communication
ports are also isolated but for a lower voltage.
15
Product description
Function modules
Modulisation
The CewePrometer is made up of function modules. Modulisation is primarily used to make it easy
to produce different versions with the same basic design, and to enable development of new
modules for the basic design to add or replace functions. Modulisation is not primarily intended for
being able to switch plug-in modules in the field the meter must be opened and the seals broken to
switch modules.
Modules
The CewePrometer consists of the following basic modules, which are necessary for a functioning
meter in its simplest version:
Measurement module
Power module
Moreover, the meter is normally equipped with the following modules for additional functionality:
Communication module
There is also space for additional function modules for expansion of function properties to meet
future market needs.
For more information, see Appendix D - Module block diagram (pg. 65).
16
Product description
Measuring principles
The measuring circuit in CewePrometer consists of current and voltage transformers that provide
signals to six parallel AD converters (analogue to digital converters) that are synchronised by a
common clock signal. The digital signals are thereafter processed by a DSP. Using voltage
transformers makes the electronics in the meter galvanically isolated from the measurement voltage,
which provides good personal safety and protection for connected equipment, such as modems.
Calculation flow
All values are calculated in the CewePrometer based on calibrated current and voltage values.
Current and voltage amplitudes and phase angles are fully compensated in regards to accuracy,
harmonics, frequency and temperature. Adjustments for accuracy are made throughout the dynamic
range for both voltage and current. Based on these individually compensated current and voltage
signals, power, energy, power factor and all other quantities that the CewePrometer can present, are
subsequently calculated. This means that accuracy for instant values is good and that active and
reactive power are correctly calculated, including harmonic power. For more information on
calculation principles and the formulas used, see Appendix F Calculation Principles (pg. 67).
3-element meter
On the 3-element CewePrometer, phase voltages and neutral wires are connected to the meter. The
voltages measured are phase voltages. Power and energy are calculated from three phase voltages
and three currents. Harmonic measurement, alarm monitoring and transformer corrections are made
on phase voltages. The phase to phase voltage is calculated from the phase voltages.
2-element meter
On the 2-element CewePrometer, the neutral conductor is not connected to the meter and the three
voltages are internally D-connected. The voltages measured are subsequently phase to phase
voltages. Power and energy are calculated based on two phase to phase voltages (U12 and U23) and
two currents (I1 and I3) according to the 2-watt meter method. The 2-element meter is primarily
used for D-connected systems (3-wire). Harmonic measurement, alarm monitoring and transformer
corrections are made on phase to phase voltages.
17
Configure
Configuring means that parameters affecting meter function can be set. Examples of
parameters that can be configured are: transformer ratios, logging interval and limits for
alarms.
Reading
Examples of information that can be read are: registers, logged values, alarms and
harmonics. The information can be printed out or saved to a file.
Maintenance
Examples of maintenance tasks are: resetting the event log and updating the firmware in the
meter.
Menu
Structure tree
18
Toolbar
Status bar
Configuration form
If the meter cannot be connected, an error message is displayed. Depending on the reason, the
message can suggest corrective actions, such as changing the port or port baud rate.
Tip: When communication problems occur, the trace feature on the meter's display can be used to
see if that which has been sent has been registered by the meter. See the section Using the display
for communication diagnostics (pg. 56).
Basic configuration
Some basic settings may be required before the CewePrometer will be able to measure and operate
correctly in a system.
Note: Settings are only necessary if they have not been made at the factory prior to delivery.
Meters supplied without backup batteries retain the time setting for a maximum of three days. For
time-dependent functions such as logging and event logging to work, the clock may need to be
adjusted.
To change the configuration for a meter, you must be connected to it. Click the Configuration
folder in the structure tree to the left in CeweConfig to display the various functions that can be
configured. For more information see the section Changing configuration (pg. 24).
Tip: You can save a configuration from a meter to a file. A summary of the configuration can also
be printed out. You can also create a configuration without being connected to a meter. For more
information, see Working with configurations (pg. 25).
20
Meter clock: To configure the meter clock, choose the node Meter Configuration Clock in the structure tree. Find out how the meter clock works and how it can be
configured in the section Meter clock (pg. 29).
Transformer ratios: For the meter to measure accurately, the ratios must be correct. To
configure the transformer ratio in CeweConfig, choose the node Meter Configuration
Measuring in the structure tree. Click the General tab in the window that opens. Fill in
the primary and secondary values for current and voltage.
Note: The values you choose as primary and secondary values will be considered as the
nominal values. These nominal values are used by several functions in the meter: alarms,
voltage monitoring and transformer compensation.
Presentation format for energy registers: To configure the presentation format for
energy registers, choose the node Meter Configuration Measuring in the structure
tree. Choose the prefix and number of decimals for energy registers. Based on the nominal
values, a capacity is calculated for how long the meter can measure without registers
resetting. Read about energy registers in the section Energy registers (pg. 30).
Pulse constants for pulse outputs (Digital I/O): To configure pulse constants for pulse
outputs in CeweConfig, choose the node Meter Configuration Digital I/O in the
structure tree. Find out how Digital I/O works and how it can be configured in the section
Digital inputs and outputs (pg. 33).
21
22
Configuration location in
CeweConfig
Meter Configuration
Communication
Meter Configuration
Communication
Meter Configuration
Maximum Demand
Meter Configuration
Historical Registers
Meter Configuration
Measuring, Transformer
Compensation tab
Meter Configuration
Measuring, Power Quality tab
Meter Configuration
Measuring, General tab
23
To open a configuration form, click the folder Configuration in the structure tree and then click
one of the nodes: Display Sequences, Measuring, Clock, etc. Configuration changes can be made
in all configuration form. In the lower right corner, there is an Apply button. If a meter is connected
and you click Apply, changes to the configuration will be immediately transferred to the meter. If
the changes affects registers or measurement, a warning dialog will be displayed. You can choose
not to send a portion of the configuration by clearing checkboxes in the dialog.
Warning dialog that sometimes is displayed when configurations are sent to the meter.
24
Reading
If you are connected to a CewePrometer with an authorisation level between 1 and 5, several
alternatives will be available in the structure tree under the node Reading. For the alternatives
25
Information about the connected meter can be obtained by choosing View Information about the
meter. Some of the information displayed is the same as what appears on the meter plate.
Additionally, there is information on when the meter was manufactured and when the most recent
configuration change was made.
Under the Modules heading, the firmware currently used in the meter is listed. See the section
Updating firmware (pg. 27).
26
Updating firmware
CewePrometer is designed with a number of modules that each have their own processors and their
own firmware. The module firmware is distributed in a package file (.pkg) which is transferred to
the meter using CeweConfig. Newly developed and improved functions can thus be added in a
meter that lacked these functions when delivered.
Note: Be sure to update CeweConfig to the latest version before updating the meter. There is
otherwise the risk that CeweConfig will no longer be version-compliant after firmware updating.
On the Tools menu, there is an Update firmware command when the meter is connected with
authorisation level 4 or higher. Begin by choosing the file that contains the update. The file name
and version number will then be displayed, and sometimes a message. Click the Advanced button
to view CeweConfig's version number by module for both the file and the meter. Additionally, you
can force modules to be updated even if the file has the same version number. Click Update to
begin updating.
27
Language
CeweConfig can be set to different languages. The available languages can be seen under
Language on the View menu.
28
Functions
Functions
Meter clock
The CewePrometer has an integrated real-time clock for time-dependent functions. Information
about time adjustment and daylight saving time status is recorded with tags on the logged values;
see Loggers (pg.39). Time adjustments are also noted as an event in the event log; see Appendix B Events (pg. 61).
Instantaneous adjustment
The meters date and time can be set to an absolute point in time. Instantaneous adjustment of the
meters time can influence logged values. To avoid this, the meter's log memory is reset when doing
instantaneous adjustment. For this reason, instantaneous adjustment of the meter clock is primarily
intended for use at initial configuration of the meter.
Sliding adjustment
The meters time can be adjusted successively. Instead of the meter adjusting the clock
instantaneously, the total adjustment is spread out over a longer period. A speed of adjustment can
be chosen in the 140% range. With 40%-adjustment, the clock is adjusted 24 seconds for each
minute until the total adjustment has been made.
Daylight savings time
CewePrometer offers the alternative of letting the meter clock follow daylight savings time. At a
specified date, the meter clock is adjusted forward, and at another, adjusted backward.
Example: On 28 March the clock is to be adjusted forward, from 02:00 to 03:00. The adjustment
back to standard time is to occur on 31 October at 3:00 (daylight savings time) when the clock is to
be set back to 02:00. The following is set in the meters: Begin March, 28, 02:00. End October, 31,
03:00 and the standard time is to be adjusted by 60 minutes.
External synchronisation
The meter time can be adjusted by a pulse on one of the meters digital inputs. When a pulse is
registred, the clock is adjusted to the closest multiple of a specified synchronisation interval. If the
synchronisation interval is, for example, one hour and the time is 13:29, a pulse will adjust the clock
to 13:00. If the time had instead been 13:31, the clock would have been adjusted to 14:00.
Available synchronisation intervals are:
10, 15, 20 and 30 minutes
1 hour
Times 12:00 and 00:00
Time 00:00
Besides specifying that the meter clock will synchronise via external pulses, a digital input must
also be configured for clock synchronisation; see the section Digital inputs and outputs (pg. 33).
29
Functions
Energy registers
Energy registers in the CewePrometer are electronic counters that accumulate energy. Prefixes and
number of decimals are configured which apply to all energy registers. Energy registers have a
width of nine digits, regardless of the total number of decimals and the prefix configured. When an
energy register has reached its maximum figure (for example, 999999.999 kWh), it resets to zero.
Fewer decimals and larger prefixes give the energy registers space for more energy without
resetting. Read more about how energy is calculated and defined by quadrants in Appendix F
Calculation Principles (pg. 67).
Overview
Energy direction
Import, export
Total values
Yes
Reactive energy
Import, export,
quadrants I-IV,
capacitive, inductive
Import, export
Apparent energy
Reactive power
import (+)
Total number
8
Yes
Yes
No
II
I
S
Capacitive
Q
P
Inductive
Reactive power
export (-)
III
IV
Quadrants
I
Phase angle
0 to 90
II
90 to 180
Lagging
III
-180 to -90
Leading
IV
-180 to 0
Leading
All registers are saved in non-volatile memory once per second. Loss of auxiliary voltage therefore
results in no more than one second of lost historical energy measurement data. For optimal
safekeeping, all registers are saved in three alternating memory areas.
30
Functions
Instant values
Besides energy, the CewePrometer can also measure instant values. Instant values are constantly
changing values such as current, voltage, power and harmonics. The formulas and definitions used
to calculate the values are presented in Appendix F Calculation Principles (pg. 67).
Overview
This table provides an overview of the instant values that can be read on the meter. Readings can be
viewed with CeweConfig, on the display and with other software that has implemented
CewePrometers communication protocol. Most instant values can be logged as average values; for
more information, see section Loggers (pg. 39).
Instant value
Frequency
Power factor total
Power factor per phase
Active power total
Active power per phase
Reactive power total
Reactive power per phase
Apparent power total
Apparent power per phase
Phase to phase voltage L12, L23, L31
Phase voltage
THD voltage
Amplitude voltage harmonics (231)4
Phase symmetry voltage
Current
THD current
Amplitude current harmonics (231)4
Phase symmetry current
Phase angle total
Phase angle per phase
Available on 3-element
meter
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes (calculated) 1
Yes
Yes (phase volt.)
Yes (phase volt.)
Yes (phase volt.)
Yes (L1, L2, L3)
Yes (L1, L2, L3)
Yes (L1, L2, L3)
Yes
Yes
Yes
Available on 2-element
meter
Yes
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes (phase to phase volt.) 2
Yes (phase to phase volt.)2
Yes (phase to phase volt.)
Yes (L1, L3) 3
Yes (L1, L3)
Yes (L1, L3)
Yes 4
Yes
No
Update frequency
The update frequency for instant values is proportional to the frequency of the measuring voltage.
With a 50 Hz measuring voltage, updating occurs 12.5 times per second, and at 60 Hz, 15 times per
second. Instant values can be read up to twice per second via the serial port when it is set to 19200
baud.
31
Functions
Accuracy
The table shows typical accuracy for a CewePrometer with accuracy class 0.2 for a selection of
instant values.
Instant value
Voltage amplitude
Current amplitude
Current amplitude
Frequency
Power
Range
Umin - Umax
5 % of Imax - Imax
15% of Imax
4763 Hz
1 % of Imax - Imax
Power factor
Harmonics measurement
Harmonics numbers 2 to 31 are measured for all currents and voltages. At a fundamental frequency
of 50 Hz, the second harmonic is 100 Hz, the third harmonic is 150 Hz, etc. Both the harmonics
amplitude and phase angle are measured and included in the calculation of power and energy, and
can be read via the meters communication protocols. In CeweConfig, harmonic amplitudes are
presented with a diagram.
THD
THD stands for Total Harmonics Distortion and is a measurement of the amount of harmonics
present in a signal. Voltages and currents THD can be read via CeweConfig and on the display.
Harmonics on 2-element meter
The harmonics measurements of voltage on the 2-element meters should be seen from the
standpoint of the voltages being D-connected internally in the meter. The harmonics measured are
the vector sums of the phases harmonics.
32
Functions
Digital inputs and outputs
The CewePrometer has several inputs and outputs that can be configured to perform various tasks.
Both inputs and outputs are protected against overvoltages by varistors. They also have an isolated
interface between the electronics and the surroundings to ensure personal safety. For electrical data
on the meters inputs and outputs, see Appendix H Connection Data (pg. 73).
Inputs
Not used
The input is not used.
Pulse input
To register pulses from pulse-producing units, such as energy meters or water meters, pulse
inputs are used. Incoming pulses are accumulated in registers called external registers. There
is an external register connected to each input on the meter. For external registers, a factor is
configured by which the number of incoming pulses is multiplied. Prefixes and the number
of decimals can also be configured for the registers. Moreover, the registers can be
configured with descriptive texts.
Time synchronisation
When incoming pulses are received, the meters clock is synchronised at a specific interval
For available synchronisation intervals and more detailed information on time
synchronisation, see the section Meter clock (pg. 29).
Registration of pulses
The meter registers pulses on positive or negative flanks, depending on if the input is set to inverted
or not. A pulse must be at least 16 ms long to be guaranteed of being detected by the meter. The
maximum pulse width that the meter can handle is 60 s. By setting limits for maximum and
minimum pulse lengths, the meter can be limited as to what it detects as a valid pulse. Pulses with
lengths beyond the established limits are ignored. The occurrence of to long or to short pulses can
also be configured to be recorded in the event log.
VCC
GND
Pulse
length
Pulse length
(inverted input)
The figure shows pulse lengths when an input is inverted or non-inverted, respectively.
33
Functions
Outputs
Not used
The output is not used.
Pulse output
The output is used to pulse an energy type that the meter is measuring. A pulse constant is
specified for the output as pulses/unit, primary or secondary, and the pulse length is
specified for all pulse outputs. The shortest possible pulse length is 40 ms. Pulses are not
allowed not come too often, and because of this, there is a relationship between the pulse
length and the specified pulse constant that maximises the pulse frequency to 1000/(pulse
length(ms) * 2).
Pulse
Gap
Maximum pulse frequency at outputs limited so that the gap is at least as long as the pulse
length.
Remote control
With this function, the output can be made active or inactive by sending commands to the
meter via the IEC62056-21/IEC1107 protocol. This functions could be used control
anything that can be controlled with a digital relay output.
Alarm output
When an output is set to functions as an alarm output, one or more of the user-defined
alarms can be chosen to indicate at the output. When an alarm occurs, the output switches to
active, and when the alarm state ceases, the output returns to inactive. In the section Alarms
(pg. 41), user-defined alarms are described and how they can be configured.
In contrast to the other selectable functions for outputs, alarm outputs can be inverted. Note
that outputs are inverted via firmware. If the meter loses its auxiliary power, the relay will
open, regardless of it is inverted or not.
Output states
An active output means a closed relay when the output is not inverted. When the output is inverted,
the active relay is open. Only when the output is used as an alarm output can it be inverted. The
CewePrometer outputs are of the solid-state type and when the meter is turned off, they are open.
34
Functions
Meter variants
Meter
CewePrometer-R
CewePrometer-R
CewePrometer-R
CewePrometer-R
CewePrometer-W
CewePrometer-W
No of inputs
4
2
0
0
4
0
No of outputs
6
10
12
0
8
0
35
Functions
Display sequences
The following is a description of the function properties of CewePrometers display sequences. The
section Using the display (pg. 53) describes how to navigate in the display sequences via the
meters buttons, and Appendix A Display Layouts (pg. 57) lists all of the CewePrometers
available display layouts.
Menu for choosing one of the four display sequences. The names of the sequences can be configured.
The CewePrometer is equipped with a graphic display that can show the information available in
the meter. To organise the information, there can be up to five display sequences with a maximum
of 40 selectable display layouts in each. Each display sequence can be given a name that will be
presented on the display.
Display sequence five is only accessible via the sealed button under the front panel window. The
front panel window can be sealed to provide additional security. For this reason, it is appropriate
that display layouts with sensitive information and sensitive functions be placed in display sequence
five.
Automatic display cycling
Automatic display cycling means that the display layouts in one or more display sequences are
automatically cycled by the meter. If, for example, display sequences 1 and 3 are configured for
automatic display layout cycling, the meter will first cycle the display layouts in sequence 1 from
the first to last layout, and then continue with sequence 3. After the last layout in sequence 3,
cycling restarts with the first layout in sequence 1. The display layouts cycle with a configurable
delay of 1 to 30 seconds. If a user manually navigates in the display sequences via the meter's
buttons, cycling automatically stops and continues again after one minute.
Passwords and security
The display sequences can be configured to be password protected to limit access for various users.
When a user chooses a password-protected display sequence via the meters buttons, a password
entry field is displayed. Only a correctly entered password gives access to the display sequence and
to the display layouts that are included in it. A password can be up to six characters long and
include the characters A-Z, 0-9 and :;<>=?@. Passwords are not case-sensitive.
Note: If a password-protected display sequence is configured for automatic display layout cycling,
password protection is deactivated.
Automatic return to first display layout
One minute after a user has stopped navigating in the display sequences, the CewePrometer returns
to the first display layout in display sequence 1 if automatic display layout cycling is not activated
for any display sequence.
36
Functions
Communications and security
All CewePrometers are equipped with an optical port for communication. The meter can be
optionally equipped with one or two additional communication ports with RS232 or RS422
interfaces. The communication protocol used is IEC62056-21, or as the older version of the
standard is called, IEC1107. For more information on how the protocol is implemented, see the
document CewePrometer IEC1107/IEC62056-21 meter reading.
Communication speed
The meters optical port always starts with a baud rate of 300 bps, regardless of what is configured,
before shifting over to the specified communication speed. This means that software (for example,
CeweConfig) that communicates with the meter via the optical port does not need to know the
speed that the meters optical port is set to. The optional ports designated as communication ports
#1 and #2 differ in this respect. They start at the specified baud rate from the beginning, which
means that connected software must be aware of the speed to be able to communicate.
Communication port #1 can be set at a speed of between 300 bps and 19200 bps, and
communication port #2 can be set at a speed of between 1200 and 19200 bps. The optical port can
be set at a speed of between 300 and 9600 bps.
Modem initiation
If a modem is connected to one of the optional ports, a modem initiation string can be configured
that the meter sends via the serial port. The initiation string can be used to set the correct
communication settings for the modem and can consist of AT-Commands. The meter sends the
initiation string when the meter is turned on and thereafter every tenth minute. During ongoing
communications, the meter waits to send the initiation string until communications has stopped.
Security
The meter has five authorisation levels that can limit access to the meter during communication via
any of the meters communication ports. Authorisation levels 1 and 2 are password-protected.
Authorisation levels 3 to 5 may be protected by passwords or by a circuit board jumper.
Authorisation levels
1 Provides access to reading.
2 Provides access to everything in level 1 plus
access to set the clock and finish historical
periods (also resets maximum demand
values).
3 Provides access to everything in level 2 plus
access to configure the meter.
4 Provides access to everything in level 3 plus
access to transfer new firmware to the meter.
5 Provides access to everything in level 4 plus
access to calibrate the meter and reset
registers.
37
Functions
Limitation of total access attempts
The meter limits the total number of access attempts to six when incorrect passwords are entered.
At the seventh attempt, the meter blocks access whether the password is correct or not. The block is
in effect until the next hour shift. After that, new password attempts may be made.
Passwords
A password consists of up to 12 case insensitive alfa-numerical characters. The authorisation check
may be deactivated for a level by deleting the password. When connecting to the meter, access is
granted to the highest level that is lacking password regardless of the password given by the user.
Access restriction via circuit board jumper
Inside the CewePrometer, there is space for a circuit board jumper that provides access to
authorisation level 5 when fitted. If the meter is connected with authorisation at level 5, the function
of the jumper can be modified. Security can be heightened by requiring the jumper even for lower
authorisation levels down to level 3. Security can also be relaxed by making authorisation level 5
accessible via a password. Cewe Instrument can provide a technical fact sheet, tfcr0004, that
describes the procedure for opening the meter and fitting the jumper.
38
Functions
Loggers
The CewePrometer has two identical, parallel and individually configurable loggers. That which is
described in this section applies both to logger 1 and logger 2.
Overview
A logger in a CewePrometer can log average values for instant quantities, energy registers and
external registers. Some quantities can be logged both by phase and as total values for all three
phases, others only as total values. The table provides an overview of quantities that can be logged.
Certain instant values in the table are not available in 2-element meters and thus cannot be logged;
see the section Instant values (pg. 31).
Quantity
Active energy import
Active energy export
Reactive energy import
Reactive energy export
Reactive energy inductive
Reactive energy capacitive
Reactive energy QI
Reactive energy QII
Reactive energy QIII
Reactive energy QIV
Apparent energy import
Apparent energy export
Phase voltage
Phase to phase voltage
Current
Active power
Reactive power
Apparent power
Frequency
Phase angle
Power factor
THD voltage
THD current
External registers 18
By phase
Yes
Yes
No
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Not applicable
Yes
Yes
Yes
Yes
Not applicable
Total
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Energy is logged as register values, i.e. at each logging occasion the current registers values is
logged. Based on these register values, the reading software can present the energy as register
values, periodic energy or as average power. CeweConfig provides all three alternatives.
Logging interval and total channels
A logger can store data in 1 to 10 channels. Common for all channels in a logger is that that the
logging interval that can be configured from one minute up to one hour. A loggers capacity is
dependent on number of channels and logging interval. When the logger is full, the oldest values
will be written over. The table shows the capacity in number of days before the oldest value is
written over.
39
Functions
Capacity in days
Number of logging channels
Logging interval
1
2
3
4
(min)
1 21
14
10
8.6
2 43
28
21
17
10
7.2
6.2
5.4
4.8
4.3
3.9
14
12
10
9.6
8.6
7.9
5 108
72
54
43
36
31
27
24
21
19
10 217
144
108
86
72
62
54
48
43
39
15 326
217
163
130
108
93
81
72
65
59
30 652
434
326
260
217
186
163
144
130
118
60 1304
869
652
521
434
372
326
289
260
237
Logged values are saved with time stamps, tariff information and flags that indicate events that have
occurred during the logging interval.
The time stamp indicates the end-time. If the logging interval is configured to one hour, a
value with the time stamp 15:00 refers to the period 14:00 to 15:00.
Tariff information indicates active rates for energy and power during the past interval.
To indicate events or states during an interval, a logged value can be stored with one or
more flags.
Event or state
Time adjusted
Name of flag
T
Disturbed
Alarm
Parameter changed
Incorrect energy
direction
Daylight savings
time
Voltage
loss/missing
Invalid value
Explanation
During the past interval, the meter clock has been adjusted either instantaneous or a sliding
adjustment is in progress.
The past interval is incomplete. For example, an interval shortened by the meter being
without auxiliary power or if the logging memory has been reset. The first value after the
logging memory having been configured will thus always be indicated with "Faulty value
(the logging memory is reset in conjunction with reconfiguration).
In conjunction with user-defined alarms being configured, it may be specified that an alarm
will also be indicated with logged values. When a user-defined alarm has triggered during
the past interval, this is indicated with the flag Alarm.
The CewePrometers configuration, calibration or initiation has changed. Which of these
three the flag refers to can be seen in the event log.
The CewePrometer can be configured for a normal energy direction, import or export. If the
meters energy direction deviates from this, this is indicated with the flag Incorrect energy
direction.
Daylight saving time has been in effect during the past interval.
During the past interval all measuring voltages have been lost or missing.
To be able to guarantee that a logged value is correct, it is logged with a control sum. If the
check sum is incorrect when the value is read, this is indicated with Invalid value. This is a
rare event, but could occur if the meter is subjected to powerful, external electromagnetic
interference.
For several of the flags, additional information can be viewed in the event log. A more exact
time for events is specified in the log.
Note: When power is calculated from logged energy values, the resulting values will be somewhat
more precise than when power is logged directly. This is because power is logged as an average
value of instant values. The instant values are read twice per second, while energy is accumulated
continuously.
40
Functions
Alarms
The CewePrometer is equipped with alarms to be able to indicate when measured quantities are
over or under a configurable limit value. The meter enters the alarm state when the limit value is
reached. An alarm is generated only after the alarm state has continued for a configurable time
(delay). Alarms are configurable by the user and are therefore called user-defined alarms. For the
CewePrometer 1.2.0 or earlier, the maximum delay is 64 seconds.
Note: There are alarms and events that cannot be configured, but instead, are always active.
Examples of such alarms are indication that the clock has been changed or that an auxiliary power
loss has occurred. For more information, see the section Event log (pg. 43).
Overview
For most user-defined alarms, the limit value is specified as a percentage of the nominal value,
which is the configured, nominal primary value (current, voltage or power). For 3-element meters,
the limit value corresponds to phase voltage, and for 2-element meters, phase to phase voltage. The
following table provides an overview of available alarms.
Alarm
Low voltage
High voltage
Low power factor
Voltage unbalance
Current unbalance
Internal error
Comments
41
Functions
Indication
Alarms are always stored in the event log. Alarms can also be configured to indicate in one or more
of the following ways:
The alarm LED stops flashing and the digital output returns to inactive low after the alarm state
passes.
42
Functions
Event log
The CewePrometer has a number of defined events and alarms that are stored in the meters event
log when they occur. The event log can contain a maximum of 40 events. Thereafter, the oldest will
be removed to make room for the new. The meter has a counter for the total number of events that
have occurred since the latest reset.
An event is represented by a time stamp and a code that indicates what has occurred. See Appendix
B - Events (pg. 61) for a complete list of all events. For certain events, data is also stored for the
events, for example, which phase the event concerns.
Events are divided into three categories:
User-defined alarms
The meter has a number of alarms that when they occur, generate an event in the event log.
The alarms each have an individual configuration that determines the conditions for when an
alarm is to be activated. See the section Alarms (pg. 41).
Meter events
The meter is also equipped with a number of alarms that are not configurable by the user,
so-called meter events. These have set conditions and are stored only in the event log.
Examples of meter events are auxiliary power loss and clock resetting. A complete list of
meter events is in Appendix B - Events (pg. 61).
Internal errors
The contents of the internal data structures, such as configuration and energy registers, are
checked by the meter. If an error is detected in one of the data structures, it is designated as
an internal error. If an internal error is indicated, the event log should be checked for related
information, and suitable actions should be taken. Internal errors are listed in Appendix B Events (pg. 61).
Display of events
The event log can be viewed on the meters display if the display layout Events is configured to be
included in one of the meters display sequences. Only the time stamp and event code are shown on
the display. A more detailed description of the meter's events can be viewed in CeweConfig, which
converts the code into plain text and displays any related data. Via CeweConfig, the event log can
also be printed out or saved to a file.
43
Functions
Maximum demand
Maximum demand is the highest average value of, for example, active power during a historical
period. The average value is calculated for an interval and compared with a previously stored value.
If the new average value exceeds the previously stored value, it becomes the new value for
maximum demand. When a historical period ends, the current maximum demand values are copied
to historical registers and then reset. The CewePrometer always stores the three highest average
values from three different days.
Overview
Up to eight quantities can be saved in the maximum demand registers, which makes a total of 24
values with time stamps for when they occurred. The table shows the types of energy that can
produce maximum demand values.
Active energy import
Active energy export
Reactive energy import
Reactive energy export
Reactive energy inductive
Reactive energy capacitive
Reactive energy QI
Reactive energy QII
Reactive energy QIII
Reactive energy QIV
Apparent energy import
Apparent energy export
The calculation interval for average values determines the period during which average values are
calculated. At the end of the interval, the new average value is compared with the previous and
possibly stored as a new maximum demand value. The calculation interval can be chosen between 1
and 60 minutes in the following predefined steps: 1, 2, 3, 4, 5, 10, 15, 30 and 60 minutes.
Example: A historical period has just been finished and all maximum demand values have been
reset. The meter is configured with a calculation interval of 30 minutes and active power as the
quantity for which to calculate the average value. The table is read from top to bottom and
illustrates what will be stored in the maximum demand registers at the various times.
Point in time
Average
power
Highest value
Second highest
Third highest
12:59, 4 July
13:00, 4 July
13:30
14:00
90 kW
89 kW
91 kW
90 kW 07-04 13:00
90 kW 07-04 13:00
91 kW 07-04 14:00
00:00, 5 July
00:30
75 kW
95 kW
91 kW 07-04 14:00
95 kW 07-05 00:30
75 kW 07-05 00:00
91 kW 07-04 14:00
00:00, 6 July
93 kW
95 kW 07-05 00:30
93 kW 07-06 00:00
91 kW 07-04 14:00
16:00, 7 July
94 kW
95 kW 07-05 00:30
94 kW 07-07 16:00
93 kW 07-06 00:00
Note: There is never more than one maximum demand value for one day.
44
Functions
Historical registers
Historical registers are used by the CewePrometer to store current register values at defined points
so as to be able to read them later. Stored in historical registers are all maximum demand values,
external registers, TOU registers and energy registers, with the exception of energy registers by
phase. The historical registers are time stamped to indicate when storage occurred. The
CewePrometer can store up to 14 historical registers.
Date and time
Energy registers
Maximum demand
values
External registers
Historical register includes a number of stored registers, and dates/times when they were stored.
Finish historical periods
By finishing a historical period, the current registers values are stored in historical registers and the
maximum demand values are reset. When a period is finished, an event is stored in the meters
event log. Periods can be finished in various ways:
Via CeweConfig
The period is finished when the meter clock reaches the end
of the month.
Note: The display layout Finish period will only work if the meter is configured to permit ending a
period via a button.
45
Functions
Time of use
Time of use is a function that enables energy to be divided up into various registers depending on
the rate that applied when the energy was measured. In the CewePrometer, tariff structures can be
stored that switch rates at predetermined times according to a configurable pattern. A tariff structure
consists of seasons, day types and special days. The maximum number of rates is eight.
Seasons specify the day types that apply during the days of the week, Monday to Sunday.
Special days specify the day type that applies on a certain date.
Tariff structures
In the CewePrometer, there are two separate tariff structures. Both tariff structures have a set of day
types, seasons and special days. By setting a starting date for a tariff structure, it can be configured
before it goes into force.
Day type
In the CewePrometer, there are 16 day types (AP). A day type specifies witch rate, from a
maximum of eight, should apply when during the day. 16 changing points per day may be specified.
A day type can be connected to a day of the week in a season or to a special day.
Season
A season refers to a period. During this period, the season defines the day types that will apply
during the days of the week. In the CewePrometer, there are 16 seasons per tariff structure. The
seasons are arranged in a sequence where one season replaces the previous at a predetermined date.
Special days
Days that change rates according to a pattern that is not covered by linking day types to seasons
(such as holidays) are called special days. A special day specifies the day type that applies on a
certain date. Special days can be configured to apply every year on the same date or for a single
year. In the CewePrometer, up to 30 special days per tariff structure can be configured.
TOU registers
An energy register or external register that is chosen to be divided into the rates is a TOU register.
There are eight TOU registers and each TOU register has separate registers for eight rates.
Maximum demand values and tariffs
Maximum demand values are stored together with information on which rate applied when
registered. For one and the same day, there can be separate day types for energy and maximum
demand values.
46
Functions
Example: The conditions are Monday to Saturday, rate 2 applies from 7:00 in the morning until
17:00 in the afternoon. For the rest of the day, rate 1 applies. On Sundays, rate 1 applies around
the clock. During the summer months, 1June to 31 August, rate 1 applies around the clock. On
Christmas Eve, 24 December, rate 1 applies around the clock. For maximum demand values, rate 3
always applies. It is active energy import that is divided in to different rates.
Day types
Three day types must be configured. Day type A to apply to energy during the autumn, winter and
spring. Day type B to apply during the summer months, and day type C for maximum demand
values throughout the year.
Day type A
00:00
Rate 1
07:00
Rate 2
17:00
Rate 1
Day type B
00:00
Rate 1
Day type C
00:00
Rate 3
Seasons
Two seasons must be configured. The first season will apply during the summer period and the
second the rest of the year. The season that applies from 1 September will also apply from 1
January to 31 May.
06-01
Energy
MD
A
C
A
C
A
C
A
C
A
C
A
C
B
C
09-01
Energy
MD
B
C
B
C
B
C
B
C
B
C
B
C
B
C
Special day
For rate 1 to apply all of Christmas Eve, a special day must be configured. Because no year is
specified, the special day will apply every year.
12-24
Energy
MD
C
C
TOU register
A TOU register is configured to accumulate active energy in several registers for the different
rates.
47
Functions
Transformer compensation
Transformer compensation is a function for compensating for measurement errors in instrument
transformers and for losses in power transformers. The function enables the CewePrometer to
present measurement values for which errors and losses have been compensated. The formulas used
in the meter are presented in Appendix F Calculation Principles (pg. 67).
Overview
Value to entered
Active loss as percent of nominal power
Reactive loss as percent of nominal power
Active loss as percent of nominal power
Reactive loss as percent of nominal power
To compensate for errors in instrument transformers, their amplitude error in percent and phase
angle error in minutes are configured in the meter. One minute is equal to the angle 1/60. The
errors can be specified separately for all voltages and currents. When instrument transformer
compensations are used, current and voltage are affected, as well as all quantities that arise from
these: power, energy, etc.
2-element meter
When voltage errors are compensated on a 2-element meter, this is done on phase to phase voltages
L12, L23 and L31 instead of on phase voltages. Only L12 and L23 are included in the calculation of
power and energy. Compensation of L31 has no significance in this respect. In compensation of
current errors for 2-element meters, only L1 and L3 can be compensated for because they are the
only currents measured.
Power transformer losses
Power transformer losses consist of copper and iron losses. They are expressed as percentages of
nominal power. One value is specified for active loss and one for reactive. When compensation of
losses is configured, power, energy, power factor, etc. are affected but not current and voltage.
48
Functions
Calculating loss values
Based on the nominal total power and the measured loss value in watts, a loss value can be
calculated as a percentage of nominal power. It is the loss value that is configured in the meter.
Nominal power is calculated with configured nominal current and voltage.
Nominal power: Powernom = Current nom Line voltagenom 3
Loss value: Loss value = Loss / Powernom 100
49
Functions
Power quality
Power quality encompasses voltage monitoring and harmonics measurement. Harmonics
measurement is described in the section Instant values (pg. 31).
Voltage monitoring
Voltage monitoring monitors three states: swells (overvoltage), sags (undervoltage) and interrupts.
Monitoring is activated by configuring the limit values for sags and swells, and is deactivated by
setting the limits to zero. The limits are expressed as percentages of configured nominal voltage.
Example: For an upper limit of 110% and a lower limit of 90%, and the configured nominal
primary voltage of 10 kV, the limits attained are 9 kV and 11 kV primary.
For 3-element meters, the average value for phase voltage is monitored, and for 2-element meters,
the average value for phase to phase voltage is monitored.
Sags and swells
Sags and swells shorter than one second are registered by accumulating registers. If the state lasts
longer than one second, it is instead registered in the event log with a time stamp and duration. The
accumulating counters and the event log can be read in CeweConfig.
Interrupts
When the average value for voltage falls to under 10% of the configured nominal voltage, this is
registered as an interrupt in the event log with a time stamp and duration. Duration for an interrupt
of less than three seconds, and down to 10 ms, is presented as "< 3 s".
State
Sags and swells
Duration
160 ms 5 1 s
Registered in
Accumulating counter
Duration presented as
Duration not specified
> 1s
10 ms 6 3 s
Event log
Event log
Duration of state
<3s
Interrupt
>3s
Event log
Duration of state
5
6
50
Functions
Hysteresis
Hysteresis means that the limit for a state and the limit for when it is restored are different. This to
avoid several states being registered when the voltage level varies around a limit. The hysteresis
limit is always midway between the low or high limit and nominal voltage.
Primary voltage (example)
Percent
11 kV
110%
10,5 kV
105%
10 kV
100%
9,5 kV
95%
9 kV
90%
Duration sag
1 kV
10%
Duration
interrupt
51
Functions
Miscellaneous
General information texts
General information texts are used in the CewePrometer for all types of information, such as the
name of the station where the meter is installed. The information texts can be shown on the meter's
display and read with CeweConfig. There are four information texts that consist of a label and an
associated text. Both the label and text can be up to 17 characters long. To present information texts
on the display, display layouts for this can be entered in a display sequence.
Example of general information texts where the label is "Owner" and the text "Cewe Instrument".
Language on display
The CewePrometers display can present texts in a number of languages. The following languages
are available in meter version 1.2.0.
52
Swedish
English
German
French
Italian
Unit
Quadrant I
Phase
Quadrant II
Quadrant III
Quadrant IV
Display layout
53
To choose between rates, press the Menu/Apply button for about two seconds. Use the step buttons
to change rates and Menu/Apply to confirm your choice. The rate will continue to apply for other
display layouts that show TOU registers.
Finish historical periods
Layout for finishing historical periods. #5 indicates total number of previously finished periods.
Finish a historical period means that all registers are saved and maximum demand values are reset.
The meter may be configured so that periods may be finished by pressing a button on the meter. For
more information, see the section Historical registers (pg. 45). Additionally, the display layout
shown above must be included in a display sequence. To finish a period, press the Menu/Apply
button for about two seconds.
54
There are display layouts for viewing historical energy registers, external registers and maximum
demand registers. The title line with the date indicates that the display layout is showing a historical
register. To step between periods, the Menu/Apply button must first be pressed for about two
seconds. The step buttons are then used to change periods. Press Menu/Apply to return to stepping
between display layouts. The period will continue to apply for other display layouts. See Historical
registers (pg. 45).
Historical TOU registers
Switching rates
Historical registers also contain TOU registers, and for these, rates also can be changed. Choose
historical period and then press Menu/Apply to be able to change rate with the step buttons. Press
Menu/Apply to return to stepping between display layouts.
Events
No of events.
The meter registers different types of events; see the section Event log (pg. 43). This display layout
shows how many events are saved in the meter. Press Menu/Apply for about two seconds to view a
list of events in chronological order with ID numbers. Explanations of these IDs are in Appendix B
- Events (pg. 61).
Confirming alarms
By pressing Apply for about two seconds, alarms of the type internal error in the meter can be
confirmed/acknowledged. For more information on internal errors, see section Event log (pg. 43).
55
Check the display after the call to the meter: /?sssssss!CRLFAK0Z1CRLF. The sssssss
string represents the meters serial number and it is only included when the meter must be
addressed. Z can be a digit between 0 and 7.
If you do not attain these results, try one or more of the following:
56
Check the cabling and connections between the meter and reading unit.
Check that the meter and the reading unit have the same baud rate.
Instant values
57
Energy registers
58
Graphs
Historical registers
Time of Use
TOU register 2 of 8.
Miscellaneous
Total events.
60
Appendix B - Events
This appendix lists the events that can appear in the CewePrometer event log.
ID indicates the code that is displayed for an event on the meter's display. Category indicates one of
the following:
H
A
I
Meter event
User-defined alarm
Internal error
Category ID
H
1
H
H
H
H
H
H
3
4
5
6
7
14
H
H
A
A
A
A
A
A
A
A
A
A
15
16
17
18
19
20
21
22
23
24
25
26
61
62
value.
%1 indicates harmonics order
number.
%2 indicates the phase on which
the harmonic exceeded the limit
value.
%1 indicates inputs order
number.
%1 indicates inputs order
number.
%1 indicates one more missing
voltage phases.
None
None
None
%1 indicates one or more phases
where the energy direction is not
the normal.
None
None
None
27
28
29
30
H
H
H
H
42
43
44
45
H
I
I
46
1000
1001
None
1002
None
1003
None
1004
None
1005
None
1006
None
1007
None
1008
None
1009
RD
TD
Signal GND
6
7
8
9
RTS
CTS
-
RX+
RXTXTX+
Signal GND
63
64
65
Meter clock
What happens if a new successive adjustment is made when one is already underway?
The current adjustment will be stopped and the new begun.
What happens if the meter time is set within an hour before the change from daylight savings
time to standard time?
If the meter time is set to daylight savings time before the change, it will be set back to standard
time automatically at the approaching time change. If the meter time is set to standard time before
the change, it will not be affected at the approaching time change.
What happens if more than one input is used for clock synchronisation?
Several inputs can be used simultaneously to receive synchronising pulses even if this is unlikely.
How are logged values affected by the daylight savings time change?
According to the example:
Change from standard to daylight savings time
2005-03-28 01:30 3467.4 kWh
2005-03-28 03:00 3523.2 kWh
2005-03-28 03:30 3602.5 kWh
Change from daylight savings to standard time
2005-10-31 02:30 3345.4 kWh
2005-10-31 02:00 3756.7 kWh
2005-10-31 02:30 4132.8 kWh
2005-10-31 03:00 4583.1 kWh
Logged values during daylight savings time are labelled with the flag S.
What are the possible consequences of adjusting the meter clock immediately?
Because CeweConfig does not permit the clock to be adjusted immediately without simultaneously
resetting the logging memory, this is purely hypothetical. If it were to be allowed, then if the clock is
adjusted forward over one or more interval limits, there will be empty spaces in the logging
memory and values will be lacking for one or more intervals.
It the clock is adjusted backward over one or more interval limits, one or more values will exist
with the same time stamps. In the worst event, this can result in data read from the meter not
matching the expected period.
66
I RMS =
Phase to phase voltage is calculated from the phase voltages fundamental vectors. This calculation
is only made in the 3-element meter; in the 2-element meter, the phase to phase voltage is
measured.
In the 2-element meter, I2 is not measured but is calculated for monitoring purposes. It is calculated
from the currents fundamental vectors.
I 21
Power
Harmonic component power
The calculations below are for active power, the calculations for reactive are identical except for
that cos-functions are replaced with sin-functions.
P1n
Pn
3-element meter:
P1n = U 1n I1n cos( n )
2-element meter:
For 2-element meters, only the total power is calculated in each harmonic component.
67
1n
2n
Active and reactive power is calculated as the sum of harmonic component power up to 31st
harmonic. The calculation is made with plus and minus signs, where negative phase angle
represents export direction and positive represents import direction.
P
P1
Q
Q1
P = P1 + P 2 + P3
Q = Q1 + Q 2 + Q3
For 2-element meters, two elements are added instead of three.
Apparent power
S
S1
S = P2 + Q2
S1 = P12 + Q12
Energy
Energy is calculated by integrating power (P, Q and S) over time.
Definition of quadrants
The term phase angle is described under its own heading below.
Active energy is calculated for import and export. The direction is controlled by the sign for active
power (+ import, export).
Active energy import: quadrant I and IV
Active energy export: quadrant II and III
68
Reactive energy
Reactive energy is calculated for four quadrants. The quadrant is controlled by the sign for active
and reactive power (e.g., active power >= 0 and reactive power >= 0 corresponding to quadrant I).
Reactive energy import: quadrant I and II
Active energy export: quadrant III and IV
Reactive energy inductive: quadrant I and III
Reactive energy capacitive: quadrant II and IV
Apparent energy
Apparent energy is calculated for import and export. The direction is controlled by the sign for
active power; apparent energy is registered for the direction that the active energy has during the
same period.
Apparent energy import: quadrant I and IV
Apparent energy export: quadrant II and III
Power factor
pf ( L1) = P1 / S1
pf (Tot ) = P / S
The power factor is calculated without signs and is thus always positive.
Phase angle
( L1) = U 1 p I1 p
Phase angle for an element is calculated from the fundamental phase angles.
THD
Total harmonic distortion
THDeur =
I 22 + I 32 + ...I n2
I 12 + I 22 + ...I n21
100%
Where I 1 I n are the currents harmonic components. The calculation is made in the same ways
for current and voltage.
69
Transformer compensations
Instrument transformer compensations
Instrument transformer compensations are specified as error in percent for amplitude, and as error
in minutes (one-sixtieth of a degree) for phase angle. These errors can be specified separately for all
voltages and currents.
AmpErr
PhaseErr
u meas
u
Amplitude compensation
up = up meas PhaseErr
In 3-element meters, phase voltages are compensated and in 2-element meters, phase to phase
voltages. The same calculation is used for all currents and voltages.
Power transformer losses
Power transformer losses consist of copper and iron losses. They are expressed as percentages of
nominal power. One value is specified for active loss and one for reactive.
Iron loss
FeLoss
P1meas
P1
Nom
70
U 1p = 0
U 2 p = 120
U 3 p = 120
The same system is represented graphically below. The vectors rotate in an anti-clockwise
direction.
71
Plastic parts
Inner window
Outer window
Side pieces
Battery cover
Polycarbonate V0
Polycarbonate V0
Polycarbonate/ABS V0
Polycarbonate/ABS V0
CewePrometer-W
Enclosure
Bottom plate
Meter cover
Inner window
Outer window
Battery cover
Terminal cover
Polycarbonate/ABS V0
Polycarbonate/ABS V0
Polycarbonate V0
Polycarbonate V0
Polycarbonate/ABS V0
Polycarbonate/ABS V0
72
Current circuit
Measurement current (I in)
Burden
1(2) A
5(10) A
1(6) A
<0.4 VA/phase
Voltage circuit
Measurement range (Un)
3-wire system:
4-wire system:
Frequency range
Burden
Max. overvoltage
3x57120 V
3x120-240 V
3x57/99120/208 V
3x120/208240/415 V
Auxiliary power
Separate auxiliary voltage Uaux
Power Consumption
Digital outputs
Type
Relay performance
Digital inputs
Type
Voltage (AC or DC)
Incoming resistance
Optical switch
48220 V
20 kohm
73
Notes
Notes
Cewe Instrument AB
Box 1006 SE-611 29 Nykping SWEDEN
Tel: +46 155 775 00 Fax: +46 155 775 97
e-mail: info@ceweinstrument.se www.ceweinstrument.com
A0172e-1a