M9290A CXA-m
PXIe X-Series Signal Analyzer
10 Hz to 3.0, 7.5, 13.6 or 26.5 GHz
Data Sheet
Table of Contents
Technical Specifications....................................................................3
Definitions and conditions.........................................................3
Recommended best practices in use.......................................3
Frequency and Time Specifications..........................................4
Amplitude Accuracy and Range Specifications.......................6
Dynamic Range Specifications.................................................9
PowerSuite Measurement Specifications............................... 12
General Specifications............................................................ 13
Inputs and Outputs................................................................. 13
I/Q Analyzer............................................................................. 14
System Requirements............................................................. 14
Software.......................................................................................... 15
Related Literature............................................................................ 16
Deploy a smaller
microwave footprint
In test system development, one
of your crucial requirements is
doing more in less spacebut this
often means tradeoffs between
footprint and precision in signal
analysis. That is, until now: weve
expanded the Keysight X-Series
with the CXA-m, a PXIe signal
analyzer that offers fully specified
performance up to 26.5 GHz. It
lets you handle RF and microwave
signals in four slots, and you
can leverage your existing code.
Choose the CXA-m and deploy a
smaller microwave footprint.
Technical Specifications
Definitions and conditions
Temperatures referred to in this document are defined as follows:
Full temperature range = Individual module temperature of 5 to 68 C, as reported by
the module, and environment temperature of 0 to 55 C.
Controlled temperature range = Individual module temperature of 25 to 40 C, as
reported by the module, and environment temperature of 20 to 30 C.
Specifications describe the warranted performance of calibrated instruments. Data
represented in this document are specifications under the following conditions unless
otherwise noted.
It is within its calibration cycle
Under auto couple control, except when Auto Sweep Time Rules = Accy
The analyzer has been stored at an ambient temperature within the allowed
operating range for at least two hours before being turned on; if it had previously
been stored at a temperature range inside the allowed storage range, but outside
the allowed operating range
The analyzer has been turned on at least 30 minutes with Auto Align set to normal,
or, if Auto Align is set to off or partial, alignments must have been run recently
enough to prevent an Alert message; if the Alert condition is changed from Time and
Temperature to one of the disabled duration choices, the analyzer may fail to meet
specifications without informing the user
95th percentile values indicate the breadth of the population (approx. 2 ) of
performance tolerances expected to be met in 95 percent of the cases with a 95 percent
confidence, for any ambient temperature in the range of 20 to 30 C. In addition to the
statistical observations of a sample of instruments, these values include the effects of
the uncertainties of external calibration references. These values are not warranted.
These values are updated occasionally if a significant change in the statistically observed
behavior of production instruments is observed.
Typical describes additional product performance information that is not covered by
the product warranty. It is performance beyond specifications that 80 percent of the
units exhibit with a 95 percent confidence level over the temperature range 20 to 30 C.
Typical performance does not include measurement uncertainty.
Nominal values indicate expected performance, or describe product performance that is
useful in the application of the product, but are not covered by the product warranty.
Temperature stability
20 to 30 C
Full temperature range
Achievable initial calibration accuracy
DC coupled
10 Hz to 3.0 GHz
10 Hz to 7.5 GHz
10 Hz to 13.6 GHz
10 Hz to 26.5 GHz
LO multiple (N)
1
2
2
2
2
4
4
4
4
4
4
4
AC coupled
10 MHz to 3.0 GHz
10 MHz to 7.5 GHz
10 MHz to 13.6 GHz
10 MHz to 26.5 GHz
10 Hz to 3.08 GHz
2.95 to 7.575 GHz
7.45 to 9.55 GHz
9.45 to 12.60 GHz
12.50 to 13.05 GHz
12.95 to 13.80 GHz
13.40 to 15.55 GHz
15.45 to 19.35 GHz
19.25 to 21.05 GHz
20.95 to 22.85 GHz
22.75 to 24.25 GHz
24.15 to 26.55 GHz
[(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Option PFR
Standard
1 x 10-7 / year
1 x 10-6 / year
-7
1.5 x 10 / 2 years
Option PFR
Standard
1.5 x 10-8
2 x 10-6
-8
5 x 10
2 x 10-6
Option PFR
Standard
4 x 10-8
1.4 x 10-6
= (1 x 1 x 10-7 + 5 x 10-8 + 4 x 10-8)
Trigger
Trigger delay
Time gating
Gate methods
Gate length range (except method = FFT)
Gate delay range
Gate delay jitter
Sweep (trace) point range
All spans
Resolution bandwidth (RBW)
Range (3.01 dB bandwidth)
Bandwidth accuracy (power)
Span = 0 Hz
Span 10 Hz
Span 10 Hz, swept
Span 10 Hz, FFT
Span = 0 Hz
Free run, video, external, RF burst, periodic timer
Span = 0 Hz or FFT
Span 10 Hz, swept
Resolution
1 s to 6000 s
1 ms to 4000 s
0.01 % nominal
40 % nominal
1 % nominal
150 to +500 ms
1 s to 500 ms
0.1 s
4.1:1 nominal
200 Hz, 9 kHz, 120 kHz, 1 MHz
10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz
(Option EMC)
(Option EMC)
Option B25
Standard
25 MHz
10 MHz
1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or
processing in the time, frequency, or modulation domain.
2. Sweep points = 101.
50 Vdc
0.2 Vdc
0.1 to 1 dB/division in 0.1 dB steps
95th percentile ( 2)
0.4 dB
0.5 dB
0.8 dB
1.0 dB
1.3 dB
1.3 dB
0.5 dB
1.0 dB
1.2 dB
1.0 dB
1.2 dB
1.8 dB
2.4 dB
Additional information
0.15 dB typical
0.30 dB nominal
0.50 dB nominal
0.70 dB nominal
1.0 dB
Response
70 dBc (80 dBc typical)
70 dBc (80 dBc typical)
70 dBc (80 dBc typical)
70 dBc (80 dBc typical)
70 dBc (80 dBc typical)
70 dBc (80 dBc typical)
64 dBc typical
Preamp on
(Two 45 dBm tones at the preamp input,
spaced by 100 kHz, 0 dB attenuation,
20 to 30 C)
10 MHz to 2 GHz
2 to 3 GHz
3 to 7.5 GHz
7.5 to 13.6 GHz
13.6 to 26.5 GHz
10 MHz to 26.5 GHz
Specification
Typical
102 dBc/Hz
106 dBc/Hz
108 dBc/Hz
130 dBc/Hz
90 dBc/Hz nominal
105 dBc/Hz
110 dBc/Hz
110 dBc/Hz
132 dBc/Hz
145 dBc/Hz nominal
[span/1000] nominal
Adjacent
Alternate
0.21 dB
0.89 dB
0.25 dB
0.67 dB
68 dB
73 dB
72 dB
78 dB
General Specifications
Temperature range
Operating
0 to 55 C
Storage
40 to 70 C
EMC
Complies with European EMC Directive 2004/108/EC
IEC/EN 61326-1
CISPR Pub 11 Group 1, class A
AS/NZS CISPR 11
ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme la norme NMB-001 du Canada
Environmental stress
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration,
altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.
Power requirement
Power drawn from chassis
65 W
Weight
Net
1.9 kg (4.2 lbs)
Shipping
4.2 kg (9.3 lbs)
Dimensions
Height
132 mm (5.2 in)
Width
82 mm (3.2 in)
Length
202 mm (8.0 in)
Warranty
The CXA-m PXIe signal analyzer is supplied with a three-year warranty
Calibration cycle
The recommended calibration cycle is one year; calibration services are available through Keysight service centers
I/Q Analyzer
Frequency
Frequency span
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
Resolution bandwidth (spectrum measurement)
Range
Overall
100 MHz to 3 MHz
Span = 1 MHz
50 Hz to 1 MHz
Span = 10 kHz
1 Hz to 10 kHz
Span = 100 Hz
100 MHz to 100 Hz
Window shapes
Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB)
Analysis bandwidth
Standard instrument
10 Hz to 10 MHz
Option B25
10 Hz to 25 MHz
IF frequency response (standard 10 MHz IF path)
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
3.0
10
0.40 dB
0.03 dB
3.0 < f 26.5
10
0.10 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
Span (MHz)
Peak-to-peak
RMS
3.0
10
0.5
0.2
3.0 < f 7.5
10
0.5
0.4
Data acquisition (standard 10 MHz IF path)
Time record length
4,000,000 IQ sample pairs
Sample rate
90 MSa/s
ADC resolution
16 Bits
Option B25 25 MHz analysis bandwidth
IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 C)
Center frequency (GHz)
Span (MHz)
Max. error
RMS (nominal)
3.0
10 to 25
0.45 dB
0.03 dB
3.0 < f 26.5
10 to 25
0.20 dB
IF phase linearity (deviation from mean phase linearity, nominal)
Center frequency (GHz)
0.02 f < 3.0
3.0 < f 7.5
Data acquisition (B25 IF path)
Time record length
IQ analyzer
Sample rate
ADC resolution
Span (MHz)
10 to 25
10 to 25
Peak-to-peak
1
1
RMS
0.3
0.5
System Requirements
Operating system
Processor speed
Available memory
Available disk space
Video
Browser
Module management
Keysight connection expert
Programming
Driver
lVI-COM
IVI-C
LabVIEW
MATLAB
Programming assistance
Command expert
Development environments
Visual Studio (VB .NET, C#, C/C++), VEE, LabVIEW,
LabWindows/CVI, MATLAB
Related Literature
Literature
M9290A CXA-m PXIe Signal Analyzer - Product Fact Sheet
M9290A CXA-m PXIe Signal Analyzer - Configuration Guide
M9018A PXIe 18 slot Chassis - Data Sheet
M9037A PXIe High Performance Embedded Controller - Data Sheet
M9036A PXIe Embedded Controller - Data Sheet
Web
For more information or literature resources please visit the web:
Product page: www.keysight.com/find/M9290A
X-Series measurement applications: www.keysight.com/find/X-Series_Apps
X-Series signal analyzers: www.keysight.com/find/X-Series
Pub number
5992-0044EN
5992-0193EN
5990-6583EN
5991-3661EN
5990-8465EN
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysights commitment to superior product quality and lower total cost
of ownership. The only test and measurement company with three-year
warranty standard on all instruments, worldwide.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to five years of protection and no budgetary surprises to ensure your
instruments are operating to specification so you can rely on accurate
measurements.
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2008
Quality Management System
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysights measurement expertise and product
breadth, combined with channel partner convenience.
www.keysight.com/find/CXA-m
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637