Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Node Commander User Manual (8500-0038)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 121

LORD USERMANUAL

Node Commander
Wireless Sensor Network Software Suite

Node Commander 2.10.0


Live Connect 2.1.3

2014 LORDCorporation
MicroStrain Sensing Systems
459 Hurricane Lane
Suite 102
Williston, VT 05495
United States of America

Phone: 802-862-6629
Toll Free: 800-449-3878
Fax: 802-863-4093

http://www.microstrain.com
sensing_support@LORD.com
sensing_sales@LORD.com

Copyright 2014 LORD Corporation


IEPE-Link, Torque-Link, 3DM-RQ1, Strain Wizard , DEMOD-DC , DVRT , DVRT-Link, WSDA , HSLink , TC-Link , G-Link , V-Link , SG-Link , ENV-Link, Watt-Link, Shock-Link, LXRS , Node
Commander , SensorCloud, Live Connect, MathEngine , EH-Link , 3DM , FAS-A , 3DM-GX1 , 3DMGX3 , 3DM-GX4, 3DM-DH , 3DM-DH3, EmbedSense , MicroStrain , and Little Sensors, Big Ideas. are
trademarks of LORD Corporation.

Document 8500-0038 Revision A


Subject to change without notice.

Node Commander Software User Manual

Table of Contents
1.

Wireless Sensor Network Overview

2.

Software Overview

3.

4.

2.1 Navigation and Views

2.2 Software Installation

11

2.3 Software Preferences

11

Quick Start Setup and Operation


3.1 System Connections

13

3.2 Gateway USB Communication

14

3.3 Automatic Node Discovery

15

3.4 Channel Configuration

17

3.5 Sampling Settings

19

3.6 Data Acquisition

20

3.7 Viewing Sensor Data

22

Gateway Communication

23

4.1 Gateway Serial Communication

24

4.2 Gateway USB Communication

26

4.3 Gateway Ethernet Communications

27

4.3.1 Connecting to a DHCP Network with Live Connect

5.

12

27

4.4 Verify Gateway Communication

29

4.5 Removing a Gateway

29

Gateway Settings

30

5.1 Gateway Information

30

5.2 Transmit Frequency

31

Node Commander Software User Manual

6.

7.

8.

5.3 Transmit Power

32

5.4 Setting the Serial Baud Rate

33

5.5 Analog Gateway Configuration

34

Node Communication

35

6.1 Automatic Node Discovery

37

6.2 Manually Adding Nodes

39

6.3 Scanning for Nodes by Address

40

6.4 Verifying Node Communication

41

6.5 Reestablishing Communication

41

6.6 Network Broadcast

42

6.7 Removing Nodes

43

Node Settings

44

7.1 Node Information

44

7.2 Node Power Management

45

7.3 Transmit Frequency

47

7.4 Transmit Power

48

7.4.1 Range Test

49

7.5 Channel Configuration

50

7.6 Node Boot Modes

52

7.7 Applying Node Configuration

54

7.8 Saving Node Settings

55

7.9 Node EEPROM Settings

56

Sensor Settings

57

8.1 Measurement Units

59

8.2 Conversion Values

60

Node Commander Software User Manual

8.2.1 Calculating a Linear Slope

63

8.2.2 Differential Input Gain and Offset

65

Example Gain and Offset Calculations

66

8.3 Sensor Calibration

9.

67

8.3.1 EXAMPLE: Lab or Field Calibration

69

8.3.2 EXAMPLE: Internal Shunt Calibration

73

8.3.3 EXAMPLE: Manufacturer Calibration

78

Sampling Modes and Settings

80

9.1 Streaming

82

9.2 Synchronized Sampling

83

9.2.1 Using the Beacon

84

9.3 Low Duty Cycle

85

9.4 Datalogging

86

9.5 Event Driven Sampling

87

9.6 Bandwidth and Sample Rate

88

9.7 Time Stamping

88

10.

Data Acquisition

89

10.1 Streaming Data

90

10.2 Synchronized Sampling

91

10.3 Low Duty Cycle Data Sampling

95

10.4 Armed Datalogging

98

10.5 Ending Sampling Sessions

102

10.6 Monitor Mode

104

10.7 Viewing Sensor Data

105

11.

Support

106

Node Commander Software User Manual

11.1 Troubleshooting Guide

106

11.2 Communications Ports inWindows

111

11.3 Live Connect Interface

113

11.4 Software Specifications

113

11.5 Technical Support

114

11.6 Warranty Information

115

11.7 Sales Support

116

12.

References

117

12.1 Related Documents

117

12.2 Glossary

118

Node Commander Software User Manual

1.

System Overview

Wireless Sensor Network Overview

The LORD MicroStrain Wireless Sensor Network is a high speed, scalable, sensor data
acquisition and sensor networking system. The system consists of wireless sensor interface
nodes, a data collection gateway, and full featured user software platforms based on the LORD
MicroStrain Lossless Extended Range Synchronized (LXRS ) data communications protocol.
Bidirectional wireless communication between the node and gateway enables sensor data
collection and configuration from up to two kilometers away. Gateways can be connected locally to
a host computer or remotely on local and mobile networks. Some gateways also features analog
outputs for porting sensor data directly to standalone data acquisition equipment.

The selection of available nodes allows interface with many types of sensors, including
accelerometers, strain gauges, pressure transducers, load cells, torque and vibration sensors,
magnetometers, 4 to 20mA sensors, thermocouples, RTD sensors, soil moisture and humidity
sensors, inclinometers, orientation, and displacement sensors. Some nodes come with integrated
sensing devices, such as accelerometers. System sampling capabilities are IEEE 802.15.4
compliant and include lossless synchronized sampling, event and burst sampling, streaming, and
data logging. One gateway can coordinate many nodes of any type, and multiple gateways can be
managed from one computer with the Node Commander and SensorCloud software
platforms. Integration to customer systems can be accomplished using OEM versions of the
sensor nodes and leveraging the LORD MicroStrain data communications protocol.
Common applications of the LORD MicroStrain Wireless Sensor Networks are wireless strain
sensor measurement, wireless accelerometer platforms, wireless vibration monitoring, wireless
energy monitoring, wireless environmental monitoring, and wireless temperature monitoring.

Node Commander Software User Manual

2.

Software Overview

Software Overview

Node Commander is a full featured data acquisition tool designed for use with the LORD
MicroStrain Wireless Sensor Network. Node Commander is included with all gateway devices
and is used for gateway and node configuration, and data collection tasks. It includes settings for
node and gateway communications, power management, sensor type and output parameters,
sensor sampling rates and modes, and data handling and storage. There are also interactive
features that aid in field installation, such as node discovery, transmission range tests, sensor
calibration routines, and near real time sensor data display. Node Commander manages
configuration and data collection of multiple gateways and nodes simultaneously, including
synchronized sampling modes.
Node Commander is designed for use with all LORD MicroStrain gateways and nodes, and
runs on most Windows platforms. Included with all Node Commander distributions is an
auxiliary program called Live Connect . Live Connect is used to translate between the Node
Commander serial protocol and the TCP/IP protocol used for Ethernet gateways.

Figure 1 - Node Commander Software

Node Commander Software User Manual

2.1

Software Overview

Navigation and Views

The Node Commander user interface includes a status bar, an icon toolbar, a header row of
drop down menus, device menus, and three main information windows, as shown in Figure 2 User Interface . When the software is first launched, only the Controller and Action Log
windows are shown. The Sampling windows appear only when sampling is initiated and vary
depending on the type of sampling being performed. The device menus are accessible only
when the devices, such as gateways and nodes, are connected and communicating with Node
Commander . They are accessed by right clicking on the device name in the Controller
window.

Figure 2 - User Interface


Windows can be repositioned by selecting the window header bar and dragging and dropping it
in the desired location. Outline guides will appear when the item is dragged to a new position.
Windows can also be closed or minimized with the icons in the upper right corners or with the
header row View menu. The header row can be repositioned in the same way but cannot be
closed.
In all menus only the commands that are currently available are selectable. They are shown in
color and will appear highlighted when hovered over. Some icons and selections will display
feature or device details as well. Features that are not available will be gray. Available

Node Commander Software User Manual

Software Overview

commands are dependent on what devices are connected and what tasks the software and
system are currently executing.
l

10

CONTROLLER INFORMATION WINDOW The Controller information window


displays a list of all attached, or previously attached, devices. The devices are
organized in a hierarchical tree and display the gateways and corresponding node
names and addresses. Additional node information can be viewed by selecting the
+ symbol next to the node name. Inactive devices can be removed from the list
through the device menu.
HEADER ROW MENUS - includes the program level commands in Node
Commander , such as file management and view settings.
DEVICE MENUS such as the gateway and node menus are accessed by right
clicking on the device name in the Controller window. All actions available for the
particular device can be found in its menu and will vary between device types.
Settings will only be applied to the selected device.
ICON TOOLBAR provides quick access to the most common commands.
ACTION LOG displays command execution information, including errors. For
example, the action log will display confirmation of device verification and an error if it
cannot be verified when requested. Information about configuration, testing, and
sampling tasks are also displayed after commands are executed. Refer to the log to
determine the current status of the system.
STATUS BAR displays the Node Commander program status. This menu can be
hidden using the header row View menu.
SAMPLING WINDOW displays the data being acquired from actively sampling
node. The sampling window is only displayed when sampling has been activated, and
the display varies depending on the sampling mode. Some modes include a graph
view and a data view, while others include only one.

Node Commander Software User Manual

2.2

Software Overview

Software Installation

To install Node Commander on the host computer, complete the following steps.
1. Insert the CD into the host computer.
2. Follow the on-screen prompts.

NOTE
The Node Commander software includes hardware drivers required for use
with USB gateways. Once installed, the software will automatically detect and
configure any USB gateways that are plugged into the host computer.

2.3

Software Preferences

The Preferences menu is used to configure basic software operational features such as the
default data directory, disk space allocation for data, and device and graphing preferences.
To access the Preferences menu, click on the Edit header and select Preferences.

Figure 3 - Software Preferences

11

Node Commander Software User Manual

3.

Quick Start Setup and Operation

Quick Start Setup and Operation

To acquire sensor data nodes are used with any LORD MicroStrain data gateway, such as the
WSDA -Base -10x -LXRS or WSDA -1000 - LXRS, and a software interface.
LORD MicroStrain has two software programs available for the Wireless Sensor Network:
SensorCloud and Node Commander . SensorCloud is a web based data collection,
visualization, analytical, and remote management platform based on cloud computing technology.
SensorCloud provides the most complete functionality. The Node Commander software
package is also fully featured and includes all functions needed for sensor configuration and data
acquisition. Node Commander is included with all data gateways and is sufficient for the basic
operations explained in this section.
In this section hardware and software setup is described, including an overview of the Node
Commander software menus required to configure a sensor connected to the node and begin
data acquisition. It is intended only as a quick start guide and is not a complete demonstration of all
system or software features and capabilities.

12

Node Commander Software User Manual

3.1

Quick Start Setup and Operation

System Connections

To acquire sensor data the following components are needed 1) external sensors, as
applicable 2) wireless sensor node 3) data gateway 4) local or networked host computer with
access to the data acquisition software, such as Node Commander or SensorCloud. The
selection of sensors, nodes, gateways, and software are application dependent, but the basic
interfaces are the same. See Figure 4 - System Connections for a connections overview.
Nodes will communicate with any gateway. Gateways with analog outputs can be connected
directly to standalone data acquisition devices for data collection, however, system
configuration will still occur through a digital communication interface.
Communication protocols between the gateway and host computer vary depending on which
model gateway is used. For example, the WSDA -Base -10x -LXRS gateway utilizes local
serial connections to the host computer, such as RS232 and USB, and interface with the Node
Commander software. The WSDA - 1000 - LXRS gateway utilizes Ethernet
communications and can be used with either Node Commander or SensorCloud. Users
can also write their own programs by utilizing the LORD MicroStrain Wireless Sensors
Network Software Development Kit.

Figure 4 - System Connections

13

Node Commander Software User Manual

3.2

Quick Start Setup and Operation

Gateway USB Communication

For USB gateways, drivers need to be installed on the host computer. These drivers are
included with the Node Commander software. After the software is installed, the USB
gateway will be detected automatically when the gateway is plugged in.
1. Open the Node Commander software

2. See "System Connections" on page 13. Make all hardware connections. Power is
applied to the gateway through the USB connection. Verify the gateway status
indicator is illuminated.
3. Open Node Commander.
4. When connected, the gateway should appear in the Controller window automatically
with a communication port assignment. If it is not automatically discovered, verify the
port is active.

Figure 5 - USB Gateway Communication

14

Node Commander Software User Manual

3.3

Quick Start Setup and Operation

Automatic Node Discovery


NOTE
Automatic node discovery only works in some boot up modes, such as the
normal boot mode. When the node is set for normal boot mode and powered
on, the device status indicator on the node will flash rapidly and then pulse in
one second intervals thereafter. If any other indicator behavior is observed, the
node may be configured for a different mode. If the node is in another boot
mode, it can be bypassed by toggling the node ON/OFF switch rapidly three
times, and then leaving it in the ON position for normal power up.

1. After establishing communication with the gateway, right click on the gateway name
in the Controller window and select Add Node, Node Discovery.

Figure 6 - Adding a Node in Node Commander


2. Turn the node ON with the ON/OFF switch.During power up the node will transmit a
message with its operating frequency. This will occur within a few seconds.
3. When the device status indicator on the node ends the rapid flash sequence and
begins pulsing at one second intervals, it has completed the normal boot up sequence
and is running in idle mode. At this point the node should be listed in the Controller
window and scanning can be stopped by selecting the Stop button in the Node

15

Node Commander Software User Manual

Quick Start Setup and Operation

Discovery window. Additional node information can be viewed by selecting the +


symbol next to the node name.

Figure 7 - Node Discovery

16

Node Commander Software User Manual

3.4

Quick Start Setup and Operation

Channel Configuration

Node channels are configured for the sensor that is connected to it. These setting are stored in
the node memory for that channel. Only the channels and configuration options that are
available on the type of node being used will be available in the configuration menus.
1. To enter the configuration menu, right click on the node name, and select Configure,
Configure Node. The Channels tab displays channel options available for the current
node.
a. Channel Enabled: indicates the channel number, and the check box is
used to enable the channel and select it for sampling. The icon next to the
check box describes the channel type, which is intrinsic to the node being
used. In the following example; a1) analog differential channel icon, a2)
analog single ended channel icon, a3) temperature channel icon.
b. Current Channel Configuration: The data output, units, input and label
describe how the channel is currently configured.
c. Configure: Select the channel Configure button to change the channel
parameters.The channel must be enabled first by selecting its adjacent
check box.

Figure 8 - Node Channels Menu

17

Node Commander Software User Manual

Quick Start Setup and Operation

2. To enter the channel configuration menu select the channel Configure button as
shown in Figure 8 - Node Channels Menu.The channel must be enabled first by
selecting its adjacent check box. The channel configuration menu options change
depending on the sensor type selected. Below is a description of possible settings.
a. Channel Label: names the channel.
b. Channel Diagram: describes the channel electronics and data flow.
c. Conversion Coefficients: defines the type and units of the
measurement being made.
d. PGA Settings: These settings determine what gain is applied to the
sensor measurement, and sets the position of the no load baseline
measurement for the sensor signal. Only available for differential input
channels with gain amplifiers.
e. Calibration Values:Includes the slope, offset, scale and formula used to
convert the sensor reading to engineering units. The slope and offset can
be determined from the sensor manufacturer calibration data, or through a
calibration process.

Figure 9 - Channel Setup

18

Node Commander Software User Manual

3.5

Quick Start Setup and Operation

Sampling Settings

Depending on the type of node being used, Node Commander offers up to four sampling
modes; Streaming, Synchronized Sampling, Low Duty Cycle, and Datalogging. Some modes
have user configurable settings for sample rate, sample duration, event based sampling,
beaconing, and data logging schemes. Other settings are automatic depending on the number
of active channels and other variables.
In general, when determining what sample mode and rate is most suitable for the application,
consider the following;
l

Increasing the sample rate reduces the system bandwidth and therefore the number
of nodes that can be reporting simultaneously.
Increasing the sample rate increases the duration the node is on over time, and
therefore reduces battery life.
Some sampling modes, such as Streaming (which is continuously recording data),
will use more bandwidth and battery power than fixed interval or event based polling
modes such as Low Duty Cycle and Event Triggered.

Sampling settings are accessed through the Configure Node menu. There is a tab for each
sampling mode.

Figure 10 - Sample Settings Menu

19

Node Commander Software User Manual

3.6

Quick Start Setup and Operation

Data Acquisition
NOTE
Touching connected sensors and test boards or charging the node battery
while acquiring sensor data may induce noise on sensitive sensor signals and
is not recommended.

When data acquisition is started, each of the four sampling modes has different menu
options and views. Some have a settings menu before data acquisition begins and may
include a data list view and/or a graph view.
The following is an example of Synchronized Sampling:
1. Right click on the node name and then Sample, Synchronized Sampling.
2. In the Synchronized Sampling window select the destination folder for the data in the
Save Location field.
3. Select Apply Network Configuration, and then select Start Sampling.

Figure 11 - Node Sampling Settings

20

Node Commander Software User Manual

Quick Start Setup and Operation

4. Close the Syncronized Sampling window by clicking the window X in the upper right.
The Sampled Data window is behind it. The default view is the Data Grid view. Use
the + symbol next to the node heading to view the data statistics.

Figure 12 - Sampling Data Grid View


5. Select Graph from the View Mode field, and then click on the Node name to view a
graphical representation of the data.
6. To end sampling, close the Sampled Data window by clicking the red and white X on
the window tab, and select Exit, Stop Nodes.

Figure 13 - Sampling Data Graph View

21

Node Commander Software User Manual

3.7

Quick Start Setup and Operation

Viewing Sensor Data

Acquired data is stored in CSV format and can be viewed with Microsoft Excel, Quattro Pro,
Open Office and other CSV editors and spreadsheet programs. The files can be found in the
default directory or the location specified at the beginning of sampling (as applicable).
The default directory is: C:\ProgramData\Microstrain\NodeCommander\SampledData
Use Windows Explorer or the Node Commander File menu to access the files. Different
sampling modes will output different file types, and they will be categorized in separate folders
by sampling mode and then further categorized by date, session, and/or node serial number.
Synchronized sampling and low duty cycle files are found in the Sampled Data folder.
Datalogging files need to be downloaded from the node before they are available for viewing
and can be accessed through datalogging menus as well as the File menu. They are stored, by
default, in the Downloaded Triggers folder.
Streaming data is stored in the Streaming folder.

Figure 14 - Exploring Data

22

Node Commander Software User Manual

4.

Gateway Communication

Gateway Communication

When a gateway is communicating with the host computer it will appear on a list in the Controller
window with a name that includes the communications port it is configured for and an icon
symbolizing the type of communication interface.
Node Commander remembers gateways and nodes that have been communicated with
previously. They will appear on the list when the software is started, however the name and icon
will be grayed out if the device is not currently communicating. To remove a gateway that is no
longer used, select Remove Base Station from the gateway menu.
Right clicking on the gateway name will produce a list of menu options available for the gateway .
Figure 15 - Gateway Communication shows a USB gateway that is communicating on Com Port
3, with the corresponding menu, and a serial gateway that is not connected, with the
corresponding menu.

Figure 15 - Gateway Communication


Some gateways will reconnect automatically when plugged into the host computer, while others
must be prompted to look for attached devices. To re-establish communication with a gateway on
the list, select Verify Base Station from the gateway menu. The Action Log window will display a
report on the success of the command.
To establish communication with a new device, or if the Verify Base Station command is
unsuccessful, refer to the following section that corresponds with the gateway communication
appropriate for the model of gateway being used.

23

Node Commander Software User Manual

4.1

Gateway Communication

Gateway Serial Communication


1. See See "System Connections" on page 13. Make all hardware connections and
power on the gateway. Verify the gateway status indicator shows that it is on.

NOTE
If using a serial gateway and the computer does not have a serial port, a USB to
RS232 converter may be used. Install the converter drivers, as directed by the
manufacturer, and set the converter to match the gateway port settings.

1. Open the Node Commander software

2. Right click anywhere in the Controller window, and select either Scan or Add Com
Port from the menu.
1. Scan Com Ports will attempt to find the connected device
automatically, and is useful if the port number is not known. Refer to
the Action Log window to monitor the scan progress.
2. Add Com Ports can be used to assign the port number the gateway is
attached to. For information on communication port numbers and
settings in Windows, see "Communications Ports inWindows" on
page 111. If the name and icon are grayed out, the gateway has been
added but is not yet communicating.

Figure 16 - Scan or Add Com Port

24

Node Commander Software User Manual

Gateway Communication

3. If using the Add Com Ports option, select the port that the gateway is connected to,
and then click OK.

Figure 17 - Select Com Port

4. The gateway will now appear in the Controller window.

Figure 18 - Serial Gateway Communication

25

Node Commander Software User Manual

4.2

Gateway Communication

Gateway USB Communication

For USB gateways, drivers need to be installed on the host computer. These drivers are
included with the Node Commander software. After the software is installed, the USB
gateway will be detected automatically when the gateway is plugged in.
1. Open the Node Commander software

2. See "System Connections" on page 13. Make all hardware connections. Power is
applied to the gateway through the USB connection. Verify the gateway status
indicator is illuminated.
3. Open Node Commander.
4. When connected, the gateway should appear in the Controller window automatically
with a communication port assignment. If it is not automatically discovered, verify the
port is active.

Figure 19 - USB Gateway Communication

26

Node Commander Software User Manual

4.3

Gateway Communication

Gateway Ethernet Communications

The default setting of the Ethernet gateway is for DHCP enabled network communications. In
order to configure the gateway for other network schemes, communication must be established
in this way first. The following are the primary ways the Ethernet gateway can be configured for
Ethernet communication:
l

Automatically through a DHCPenabled network.


By assigning the gateway a static IPaddress and connecting it directly to a host
computer.
By assigning the gateway a static IPaddress and connecting it to a similarly configured
network .
By connecting remotely through the SensorCloud web interface.

All options require connecting to the gateway through the Live Connect software interface.
Live Connect is essentially a TCP/IPto serial conversion tool that translates the
communications between Node Commander and the Ethernet gateway. The Live Connect
interface is described in the following section.

4.3.1

Connecting to a DHCP Network with Live Connect

1. Connect the host computer and gateway to the DHCP network and apply power to
the gateway. Verify the gateway status indicator shows that it is on and has
completed the boot up process.
2. Open Live Connect . The gateway will be detected automatically but depending on
the network, it may take 2 to 3 minutes. Once detected, the gateway will appear on
the list of discovered devices and can be identified by its serial number in the Device
Name column (Figure 20 - Live Connect Interface).
3. When the gateway appears on the list, highlight it and click the Connect button. The
Status column will indicate when communication has been established, and if so, the
Port column will display the communications port. All active gateways on the network
will be displayed, and can be connected to in this manner.

27

Node Commander Software User Manual

Gateway Communication

NOTE
Once communication has been established, the gateway settings can be
adjusted through the gateway Control Panel. Access the Control Panel by
highlighting the gateway and selecting the View button in Live Connect.

Figure 20 - Live Connect Interface


4. Open Node Commander.
5. The gateway will appear in the Controller window automatically with a
communication port assignment.

Figure 21 - Ethernet Gateway Communication

28

Node Commander Software User Manual

4.4

Gateway Communication

Verify Gateway Communication

The Verify and Reverify Base Station commands can be used to check the communication
status of the gateway to the host computer. The Verify command is available in the gateway
menu when no communication has been previously established. The Reverify command is in
the gateway menu if communication has already be established, and is used to check the
current status. Right click on the gateway name and select the command. The Action Log will
report the results of the gateway query.

Figure 22 - Verify Gateway Communication


4.5

Removing a Gateway

To remove a gateway that is no longer used, right click on the gateway name in the Controller
window and select Remove Base Station from the gateway menu .

Figure 23 - Removing Gateways

29

Node Commander Software User Manual

5.

Gateway Settings

Gateway Settings
5.1

Gateway Information

Gateway information, such as the model type, serial number, radio type, operating frequency,
current communications port, and firmware revision, is found in the gateway Information
window. The window is accessed by right clicking on the gateway name and selecting
Configure, Configure Base Station.
There is also a Label field to give the gateway a unique name, which is useful when there are
multiple gateways in the system.
Gateway transmit power can also be selected in this window. See "Transmit Power" on page
32 for additional information.

Figure 24 - Gateway Information and Name

30

Node Commander Software User Manual

5.2

Gateway Settings

Transmit Frequency
NOTE
l

The gateway can automatically manage nodes operating on different


frequencies by using the Node Discovery feature in Node Commander. In
this routine, the gateway listens for node broadcasts on the frequency
channel it is set to. If the node is in normal boot up mode, it will provide the
broadcast when it is initially powered on and broadcast on all channels. As
long as the node is powered on after the Node Discovery feature is
activated, the gateway will link to it and remember the channel setting for
future node queries.
Manually setting the node and gateway frequency channels to match is
required in some applications. For example, when sending broadcast
messages from the gateway to multiple nodes (including the synchronized
sampling beacon), all nodes must be on the same channel as the gateway to
receive the broadcast. Assigning channels is also a good idea when multiple
gateways are attached to one host computer, or when other wireless
equipment is nearby, and frequency or transmission interference may occur.

The frequency setting for the gateway is found in the gateway Configure menu.

Figure 25 - Gateway Frequency

31

Node Commander Software User Manual

5.3

Gateway Settings

Transmit Power

The transmit power level may require adjustment if power consumption is a concern or in
regions where there are transmit power restrictions. Lowering the power output reduces power
consumption, but it also reduces the wireless communication range between the gateways and
nodes.

NOTE
Actual range is highly dependent of how the nodes and gateways are installed
and the surrounding environment. Refer to the device user manuals for
installation recommendations.

Setting

Power Output

Maximum
Range

Extended
Standard
Low

16 dBm (39mW)
10dBm (10mW)
0dBm (1mW)

2 km
1 km
70 m

Table 1 - Transmit Power Settings


The transmit power setting for the gateway is found in the gateway Configure Base Station
menu.

Figure 26 - Transmit Power Setting

32

Node Commander Software User Manual

5.4

Gateway Settings

Setting the Serial Baud Rate

Changing the RS232 gateway baud rate may make it inoperable.


l

The RS232 gateway is set to communicate at a 115,200 baud rate when it is


manufactured. For most computer serial ports, 115,200 baud is the
maximum supported rate. On these computers, if the gateway is set to
921,600 baud it will no longer be able to communicate with the computer. To
restore, the gateway will need to be connected to a (non-standard) high
speed port or the returned to the factory for reconfiguration.
The RS232 gateway does not support Synchronized Sampling unless it is
set to 921,600 baud and used with a high speed serial port.

1. Open Node Commander and establish communication with the gateway ( see
"Gateway Serial Communication" on page 24).
2. Right click on the gateway name and select Configure, Configure Base Station, and
adjust the Baud Rate as needed.

Figure 27 - Setting the Baud Rate

33

Node Commander Software User Manual

5.5

Gateway Settings

Analog Gateway Configuration

In addition to normal digital communications, such as serial and Ethernet, analog gateways
have eight channels of configurable outputs that recreate selected sensor input channels of
connected nodes. The analog gateway channels are mapped to a node sensor channel
through the Node Commander Configure Node menu.
This menu window is accessed by right clicking on the node heading in the Controller window
and selecting Configure, Configure Node. The Analog Pairing menu is used to assign node
channels to analog gateway outputs by entering the node address and channel number.
Channels can be assigned from multiple nodes. Further configuration options, such as enabling
a no signal timeout and selecting signal processing modes, are available in this menu. Refer to
the gateway user manual for more information.
Analog gateways have two programmable buttons located on the front of the device. The
function of these buttons can be assigned or disabled in the Buttons menu, and include power
and sampling options.

Figure 28 - Analog Gateway Configuration

34

Node Commander Software User Manual

6.

Node Communication

Node Communication

The node has four communication states; Boot up, Idle, Sleep and Sample. Node configuration
is only possible when the node is in the idle state. After boot up, the node will enter idle state if it is
configured for a normal boot mode. The node can also be configured to enter the sample or sleep
state after boot up in a few different ways. For example, the node may be configured to boot up
and then start sampling continuously. In that scenario the node will never go into the idle state so
configuration of the node is not possible until a stop command is executed to end the sampling. In
other boot modes, the node will boot up, sample, and then go into idle state, at which point
configuration can occur. The sample duration on these boot up sequences is determined by the
sampling settings.
Newly manufactured nodes are configured for normal boot mode. See "Node Boot Modes" on
page 52 for more information about setting and overriding boot modes.

NOTE
Nodes and gateways have status indicator lights that flash in different patterns
to describe what mode they are currently in or activity that is occurring. Refer to
the device user manual for details.

Depending on the boot mode selected, the node is also set to go into another state after the user
timeout interval has occurred. For example, when the node is in the normal boot mode, it will go
into the sleep state after the timeout interval has passed. This is useful for power conservation,
however, to put it back in idle state for configuration, the node stop or wake command must be
executed.

NOTE
If the node has been inactive for longer than the user timeout period, or if the
node has entered sleep or sample modes for any reason, it may be necessary
to execute a stop node command in order to perform configuration tasks.

35

Node Commander Software User Manual

Node Communication

When a node is communicating with the gateway and able to be configured, it will appear in the
Controller window on a list under the host gateway. The listing will include the node name and
serial number. Additional node information can be viewed by selecting the + symbol.
Node Commander remembers nodes and gateways that have been communicated with
previously. They will appear on the list when the software is started, however may not actually be
ready for configuration. If the + symbol next to the node name is selected, an error will appear if
the node is not connected. To reconnect with a previously connected node see "Reestablishing
Communication" on page 41.

Figure 29 - Node Communication


There are several methods to establish communication with a new device, depending on how
complicated the wireless sensor network is and what is known about the device;
1. Automatically with the Node Discovery feature
2. Manually by entering the node address
3. By scanning a range of node addresses
In Normal boot mode, the node will power on, send a status message, and then enter idle state
to wait for a command. Automatic node discovery only occurs in the status message phase of
the boot up. If the node is already in idle state, or in any other state, node discovery will not
work. Using other methods of establishing communication with the node, such as adding the
node manually, works only after boot up is complete and the node is in the idle state.

36

Node Commander Software User Manual

6.1

Node Communication

Automatic Node Discovery


NOTE
Automatic node discovery only works in some boot up modes, such as the
normal boot mode. When the node is set for normal boot mode and powered
on, the device status indicator on the node will flash rapidly and then pulse in
one second intervals thereafter. If any other indicator behavior is observed, the
node may be configured for a different mode. If the node is in another boot
mode, it can be bypassed by toggling the node ON/OFF switch rapidly three
times, and then leaving it in the ON position for normal power up.

1. After establishing communication with the gateway, right click on the gateway name
in the Controller window and select Add Node, Node Discovery.

Figure 30 - Adding a Node in Node Commander


2. Turn the node ON with the ON/OFF switch.During power up the node will transmit a
message with its operating frequency. This will occur within a few seconds.
3. When the device status indicator on the node ends the rapid flash sequence and
begins pulsing at one second intervals, it has completed the normal boot up sequence
and is running in idle mode. At this point the node should be listed in the Controller
window and scanning can be stopped by selecting the Stop button in the Node

37

Node Commander Software User Manual

Node Communication

Discovery window. Additional node information can be viewed by selecting the +


symbol next to the node name.

Figure 31 - Node Discovery

38

Node Commander Software User Manual

6.2

Node Communication

Manually Adding Nodes

Nodes can be manually added if the node address is known. The address is indicated on the
functional test or calibration document included with the node when it is purchased. The
gateway will search for the node on the frequency it is currently set to. If the node is on a
different frequency, the gateway frequency must be changed in order for the manual addition of
the node to be successful. See " Transmit Frequency" on page 31 for instructions on how to set
the gateway frequency.
1. Make the hardware connections and establish communications with the gateway.
2. Power on the node.
3. Right click on the gateway name in the Controller window and select Add Single
Node.
4. Enter the node address and select OK. If the node is not found, an error message will
appear and provide the option to scan for the node on other frequencies.
5. If the node is found, verify communication by expanding the node information list with
the "+" symbol next to the node name. If the information list appears, communication
has been established.

Figure 32 - Adding a Node by Address

39

Node Commander Software User Manual

6.3

Node Communication

Scanning for Nodes by Address

Nodes can found by scanning an address range. The gateway will search for nodes on the
frequency it is currently set to. If the node is on a different frequency, the gateway frequency
must be changed in order for the scan to be successful. See " Transmit Frequency" on page 31
for instructions on how to set the gateway frequency.
1. Make the hardware connections, and establish communications with the gateway.
2. Power on the node.
3. Right click on the gateway name in the Controller window, and select Scan Network.
4. Enter the node address range. The wider the range, the longer the scan will take.
5. Select OK.
6. If the node is found, verify communication by expanding the node information list with
the "+" symbol beside the node heading. If the information list appears,
communication has been established.

Figure 33 - Scanning Addresses

40

Node Commander Software User Manual

6.4

Node Communication

Verifying Node Communication

Use the Ping command to verify communication with a node at any time. Right click on the node
name in the Controller window and select Communicate, Ping from the node menu. The Action
Log window will report if the Ping was successful or not. A successful ping means the node is
communicating with the gateway.

Figure 34 - Test Node Communication


6.5

Reestablishing Communication

Node configuration cannot occur when the node is sampling or sleeping. The Stop Node
command is used to end whatever activity the node is currently performing, put the node into
idle mode, and allow configuration. It can be use to exit sampling modes, such as sampling on
boot up, and to interrupt other routines, such when the node is in sleep mode. To execute a
Stop Node command, right click on the node name in the Controller window, and select
Communicate, Stop Node from the node menu. After the node has been stopped, node
menus, such as Configure, will become accessible.

Figure 35 - End Node Activity to Reestablish Communication

41

Node Commander Software User Manual

6.6

Node Communication

Network Broadcast

The network broadcast feature is used to send the same command at the same time to multiple
nodes connected to the same gateway. Nodes must be on the same transmission frequency
(channel) as the gateway to receive broadcast commands. Network broadcast functions
include communications, sampling, and configuration commands.
Network broadcast commands are found in the gateway menu. Right click on the gateway
name, and select Network Broadcast. Then select the desired category and command.

Sample
l

Synchronized Sampling Start

Low Duty Cycle Start

Armed Datalogging

Trigger Datalogging Session

Erase

Communicate
l

Ping

Stop Node

Wake Node

Sleep

Cycle Power

Configure
l

42

Read/Write EEPROM

Figure 36 - Network Broadcast Commands

Node Commander Software User Manual

6.7

Node Communication

Removing Nodes

To remove a node that is no longer used, right click on the node name in the Controller window
and select Remove Node from the node menu.

Figure 37 - Removing a Node

43

Node Commander Software User Manual

7.

Node Settings

Node Settings
7.1

Node Information

Node information, such as the model type, serial


number, radio type, operating frequency, node
address, and memory size, is found below the
node name in the Controller window or more
extensively in the node Info window.
Clicking in the "+" symbol next to the node name
will produce a list of some of the node
information (Figure 38 - Node Information). The
same thing can also be accomplished through
the Communicate menu by right clicking on the
node name and selecting Communicate, Load
Node Information.

Figure 38 - Node Information

If the node information will not load, the node is not communicating. Try executing the Stop
Node command to ensure the node is not in a sampling or sleep mode, and then load
nodeinformation.
More node information can be found in the node Info window. This window is accessed by right
clicking on the node name and selecting Configure, Configure Node and the Info tab. In the
node Info tab there is a field to give the node a unique name, which is useful when there are
multiple nodes (Figure 39 - Additional Node Information and Name). Node settings can be
saved and loaded as well. See "Saving Node Settings" on page 55.

Figure 39 - Additional Node Information and Name

44

Node Commander Software User Manual

7.2

Node Settings

Node Power Management

The node power management features can be used to conserve power when the node is not
actively sampling. Power management features include an ultra low power sleep mode, node
boot up options, radio check interval settings, and an inactivity timeout.
Node power management features are available in two menus. For direct control over the
current state of the node, right click on the node name and select, Communicate, Power
Management. This menu is used to set the mode into sleep mode. The node will stay in sleep
mode until the Wake command is executed from the same menu. The Action Log window will
display the status of the command. There is also a command to cycle power to the node. This
software reboot feature is used in special cases, such as when using the Read/Write EEPROM
command to program the node or for troubleshooting if the node is unresponsive.

Figure 40 - Node Power Basic Commands


Additional power settings are found by right clicking on the node name and selecting Configure,
Configure Node. Select the Power tab.

NOTE
If the inactivity timeout is set to a very short duration, it may be difficult to
execute commands before the timeout occurs. In this case, use the Stop Node
command to re- establish communication and extend the timeout before it
occurs again.

45

Node Commander Software User Manual

Radio Check Intervals is how frequently the node turns on the radio to listen for
communications while it is sleeping. Longer wait intervals will result in less power use but
alos means there may be a delay between the time when the node is sent a Wake command
and when it actually wakes up.
User Inactivity Timeout is the amount of time that has to pass with no commands before
the activity related to node boot mode occurs. For Normal Boot mode, this is the time of
inactivity that has to pass before the node goes to sleep. In some boot modes, such as
Synchronized Sampling, it will begin sampling after that period of time .
Boot Mode determines what the node will do when it is turned on. Most of the boot options
are related to what type of sampling is desired and not directly to do with power
consumption, however some sampling modes use significantly less power than others. For
example, Low Duty Cycle uses a lot less than streaming because the node is only on and
transmitting periodically. See "Node Boot Modes" on page 52.

Figure 41 - Advanced Power Settings

46

Node Settings

Node Commander Software User Manual

7.3

Node Settings

Transmit Frequency
NOTE
l

The gateway can automatically manage nodes operating on different


frequencies by using the Node Discovery feature in Node Commander. In
this routine, the gateway listens for node broadcasts on the frequency
channel it is set to. If the node is in normal boot up mode, it will provide the
broadcast when it is initially powered on and broadcast on all channels. As
long as the node is powered on after the Node Discovery feature is
activated, the gateway will link to it and remember the channel setting for
future node queries.
Manually setting the node and gateway frequency channels to match is
required in some applications. For example, when sending broadcast
messages from the gateway to multiple nodes (including the synchronized
sampling beacon), all nodes must be on the same channel as the gateway to
receive the broadcast. Assigning channels is also a good idea when multiple
gateways are attached to one host computer, or when other wireless
equipment is nearby, and frequency or transmission interference may occur.

The frequency setting for the node is found in the node Configure menu.

Figure 42 - Node Frequency

47

Node Commander Software User Manual

7.4

Node Settings

Transmit Power

The transmit power level may require adjustment if power consumption is a concern or in
regions where there are transmit power restrictions. Lowering the power output reduces power
consumption, but it also reduces the wireless communication range between the gateways and
nodes.

NOTE
Actual range is highly dependent of how the nodes and gateways are installed
and the surrounding environment. Refer to the device user manuals for
installation recommendations.

Setting

Power Output

Maximum
Range

Extended
Standard
Low

16 dBm (39mW)
10dBm (10mW)
0dBm (1mW)

2 km
1 km
70 m

Table 2 - Transmit Power Settings


The transmit power setting for the node is found in the Configure Node menu.

Figure 43 - Transmit Power Setting

48

Node Commander Software User Manual

7.4.1

Node Settings

Range Test

After establishing communication between node and gateway, use the range test in Node
Commander to monitor the signal strength and best position the node, gateway, and
antennas for installation. Achievable range is determined by the transmit power setting and
is highly dependent on the physical environment surrounding the devices.
1. Right click on the node header, and select Communicate, Range Test.

Figure 44 - Range Test Menu


2. The total RSSI range for the node and gateway is -90 to 0dBm. The higher the value
(closer to zero), the better, but reliable communication can be achieved between 0
and -75dBm. The devices may still communicate between -75 and -90dBm, but it
could be intermittent or result in data loss. Position the node where the best RSSI
value is observed.

Figure 45 - Range Test Statistics

49

Node Commander Software User Manual

7.5

Node Settings

Channel Configuration

Node channels are configured for the sensor that is connected to it. These setting are stored in
the node memory for that channel. Only the channels and configuration options that are
available on the type of node being used will be available in the configuration menus.
1. To enter the configuration menu, right click on the node name, and select Configure,
Configure Node. The Channels tab displays channel options available for the current
node.
a. Channel Enabled: indicates the channel number, and the check box is
used to enable the channel and select it for sampling. The icon next to the
check box describes the channel type, which is intrinsic to the node being
used. In the following example; a1) analog differential channel icon, a2)
analog single ended channel icon, a3) temperature channel icon.
b. Current Channel Configuration: The data output, units, input and label
describe how the channel is currently configured.
c. Configure: Select the channel Configure button to change the channel
parameters.The channel must be enabled first by selecting its adjacent
check box.

Figure 46 - Node Channels Menu

50

Node Commander Software User Manual

Node Settings

2. To enter the channel configuration menu select the channel Configure button as
shown in Figure 46 - Node Channels Menu.The channel must be enabled first by
selecting its adjacent check box. The channel configuration menu options change
depending on the sensor type selected. Below is a description of possible settings.
a. Channel Label: names the channel.
b. Channel Diagram: describes the channel electronics and data flow.
c. Conversion Coefficients: defines the type and units of the
measurement being made.
d. PGA Settings: These settings determine what gain is applied to the
sensor measurement, and sets the position of the no load baseline
measurement for the sensor signal. Only available for differential input
channels with gain amplifiers.
e. Calibration Values:Includes the slope, offset, scale and formula used to
convert the sensor reading to engineering units. The slope and offset can
be determined from the sensor manufacturer calibration data, or through a
calibration process.

Figure 47 - Channel Setup

51

Node Commander Software User Manual

7.6

Node Settings

Node Boot Modes

Nodes can be configured to boot up in several different ways, depending on the application
requirements ( Table 3 - Node Boot Modes ). For sample on boot up modes, the node will
sample according to the sampling settings configured for the node. This means it may either
boot up and sample continuously or boot up, sample for a fixed interval, and then go into idle
state.
Setting the node up for different modes will effect the way communication can be established
with the node. See "Node Communication" on page 35.
l

If the node is sampling or sleeping on boot up, node configuration cannot occur until the
sampling is completed or the Stop Node command is executed.
Node boot modes can be bypassed by toggling the node ON/OFF switch rapidly three
times and leaving it in the ON position. This will result in a normal boot up.

Boot Mode

Description

Node Device
Indicator Sequence

Normal

Node turns on, sends out a


status message, and then waits,
in idle, for a command.

Several rapid flashes, then


pulses in one second intervals.

Low Duty
Sampling
Synchronized
Sampling
Stream

Datalog

Sleep

Node turns on and takes a


sample at set parameters.
If set for a fixed sample interval,
on completion, the node sends
out a status message and then
waits, in idle, for a command.
If set for continuous sampling,
the node turns on and samples
until a stop command is sent.

Node turns on, and then goes


into sleep mode.

Pulses at the set sample interval when sampling. Normal


boot indicator sequence on
completion
On solid while sampling.
Normal boot indicator
sequence on completion
On dimly while sampling,
normal boot indicator sequence
on completion
One flash when sampling, then
on dimly when datalogging.
Normal boot indicator
sequence on completion
Off, with occasional flash at set
radio check interval.

Table 3 - Node Boot Modes

52

Node Commander Software User Manual

Node Settings

1. To change the node boot mode, right click on the node name and select Configure, Configure Node from the node menu.
2. Select the Power tab.
3. Select the desired boot mode from the drop down menu.
4. Select Apply or OK. The next time the node is restarted it will boot up in the new mode. It
must be a full power reset; the Cycle Power command in Node Commander will not work
for this.

Figure 48 - Boot Mode Setting

53

Node Commander Software User Manual

7.7

Node Settings

Applying Node Configuration

Node values set in any of the tabs in the node Configure menu are saved to the node using
either the OK or Apply buttons. The OK button saves and exits the menu, while the Apply
button writes the values, but does not exit the menu, allowing additional settings tobe changed.
Changes must be applied before selecting another tab, or they will not be saved.

Figure 49 - Applying Node Settings

54

Node Commander Software User Manual

7.8

Node Settings

Saving Node Settings

The Save Node State feature creates a file to backup all of the node settings, including factory
and user entered calibration values, channel configurations, sampling settings, and
communication settings. The settings file contains the unique node address and any applicable
factory calibration values and is therefore unique to each node . It is important that a saved file
only be loaded on the node it was created on, or the node address and factory calibration
values will be overwritten. If external sensors have been replaced or disconnected since the
settings file was created, recalibration of the sensors may be necessary.
The Save Node State and Load Node State features are found in the node Info window. The
Info window is accessed by right clicking on the node name in the Controller window and
selecting Configure, Configure Node. Name the node settings file with a name that indicates
which node it is, for example using the serial number.
Use the Load Node State button to reload previously saved node settings files.

Figure 50 - Node Settings File

55

Node Commander Software User Manual

7.9

Node Settings

Node EEPROM Settings

Advanced settings can be configured by using coded commands to write directly to the node
EEPROM. The EEPROM addresses and settings are available from LORD MicroStrain
Technical Support Engineers and are only required in special circumstances.
The EEPROM read/write menu is available though the node menu. Right click on the node
name and select Configure, Read/Write EEPROM. Enter the EEPROM command address
and value and then click Read or Write, as applicable.

Figure 51 - Read/Write Node EEPROM

56

Node Commander Software User Manual

8.

Sensor Settings

Sensor Settings

LORD MicroStrain sensor nodes are designed to accept many sensor types, and some have
integrated sensors. Some nodes have multiple input channels and channel types. Available sensor
settings vary depending on the type of node being used and what channel the sensor is connected
to. The sensor channel configuration menus change accordingly and can includes settings for
measurement units, conversion values and gain settings. See "Channel Configuration" on page 50
for menu interface information.
Sensor readings can be displayed and recorded in volts, A/D value (bits), or engineering units.
There are preset measurement units, as well as a user defined field. Because the wireless sensor
system is digital, the analog voltage readings from the sensors are converted into a digital
equivalent value based on the volt-to-bit scale of the internal analog to digital voltage converter
(A/D converter). A/D value can be read directly or further converted to engineering units by
applying conversion values and a conversion formula.
Some sensors also require calibration. Calibration incorporates coefficients that normalize the
sensor output to a known reference device in order guarantee accuracy of the sensor readings.
See "Sensor Calibration " on page 67 for more information.
Internal sensors are assigned to a specific channel at the factory and the channel number cannot
be changed. Some internal sensors are also calibrated at the factory. The calibration values can
be changed. Table 4 - Example Internal Sensor Types, describes some internal sensor types,
units and calibration options.

example
internal sensors

accelerometer

units
G Force

factory calibration

A/D value

user entry from lab or field


calibration

custom
temperature
temperature

calibration options

A/D value
custom

calibration not required


factory calibration
user entry from lab or field
calibration

Table 4 - Example Internal Sensor Types

57

Node Commander Software User Manual

Sensor Settings

External sensors can be attached to any channel that is applicable to the configuration of the
sensor. Table 5 - Example External Sensor Types , describes example sensors, units, and
calibration options.

channel type

example
external sensors
strain gauges in full, half,
quarter and custom
Wheatstone Bridge
configurations

units

calibration
options

strain

calibration wizard

volts

user entry from


manufacturer
data, lab or field
calibration

A/D value
custom

other Wheatstone Bridge


sensors such as:
analog
differential input

some pressure sensors


some force sensors
some mass sensors
some displacement
sensors
some accelerometer
some temperature
sensors

G force
A/D value
volts
custom
English and metric
measurements for;
mass, pressure,
force, distance, and
temperature.

user entry from


manufacturer data,
lab or field
calibration

4-20mA sensors
analog
single ended
input

sensors with voltage


outputs referenced to the
system ground.

volts
A/D value
custom
temperature

thermocouple

thermocouples

A/D value
custom

Table 5 - Example External Sensor Types

58

user entry from


manufacturer data,
lab or field
calibration
user entry from
manufacturer data,
lab or field
calibration

Node Commander Software User Manual

8.1

Sensor Settings

Measurement Units

Sensor measurement units are set in the Channel Configuration menu.


1. To enter the configuration menu, right click on the Node heading and select Configure, Configure Node. The Channels tab displays channel options available for the current node.

Figure 52 - Channel Configuration Menu


2. Select the type of measurement from the Class menu, and then select Units.

Figure 53 - Select Sensor Units

59

Node Commander Software User Manual

8.2

Sensor Settings

Conversion Values

The conversion values include the slope, offset, gain, scale, and formula used to convert the
sensor A/D value to engineering units. The A/D value (bits) are the digital representation of the
sensor voltage output. The type of sensor, channel, and desired engineering units determine
what conversion values are available. The values are entered through the Node Commander
software and saved in the node memory for the applicable node channel.

NOTE
In order to report accurate readings, many sensors require calibration.
Calibration coefficients normalize the sensor output to a known
reference device and are often expressed in the measurement unit
conversion values. Calibration accounts for the slight variations between
sensor elements, wiring, system electronics, sensor mounting, and
environmental conditions.

The conversion values can be determined mathematically from the sensor sensitivity
specifications, from the sensor manufacturer calibration data, or through a calibration process.
Calibration incorporates coefficients that normalize the sensor output to a known reference
device in order guarantee accuracy of the sensor readings, especially when making small or
precise measurements. See "Sensor Calibration " on page 67 for more information. Not all
sensors require calibration.
Conversion Formula:The default formula assumes a linear relationship between the A/D
value (bits) and engineering units and is expressed mathematically as y=mx+b, where the y is
the engineering units at a given point (measurement), m is the slope of the line that represents
the linear ratio, x is the A/D value at a given point, and b is the fixed zero load offset of the
sensor.
Slope: the ratio of A/D value (bits) to engineering units (EU) that is used to convert the sensor
measurements, or bits per EU. The slope conversion value will vary depending on the
engineering units desired. For example, if the units are a measurement of force in pounds, the
desire slope conversion would describe how many bits equal one pound (bits/pound).
Mathematically, the slope is m in the formula y = mx +b.

60

Node Commander Software User Manual

Sensor Settings

Offset: The offset value is the starting output value of the sensor with no load applied.
Mathematically, the offset is b in the formula y = mx +b.
Effective Range: Once the formula, slope, and offset values are applied, the sensor
measurement range, in engineering units, is calculated. This is the effective range and is based
on the resolution of the node. For example, a 16-bit node will have 65536 A/D values and a 12bit node will have 4096 A/D values. The number of A/D values per EU (slope) multiplied by the
total number of A/D values, minus the offset, determines the effective EUrange of the sensor.
The effective range does not take gain or offset scale into consideration.
Input Range (Gain): sets the amplification of the signal within the node. This feature is only
available for channels with differential inputs and gain amplifiers.
Offset Scale (with Auto Balance):This feature is only available for channels with differential
inputs, and assigns the position and value of the no load measurement of the sensor. The
offset scale level adjusts the operating window of the sensor measurements in reference to the
entire range. For example, in mid scale, the sensor no load measurement will be placed in the
middle of the range, providing it 50% of the range for positive readings and 50% of the range for
negative readings. Once the scale level is selected, the Auto Balance procedure is used to
assign the actual sensor no load measurement to the designated scale.
l

Low is for positive going signals (zero at 25% of total range)

High is for negative going signals (zero at 75% of total range)

Midscale is for and positive and negative going signals(zero at 50% of range)

Figure 54 - Offset Scale Setting

61

Node Commander Software User Manual

Sensor Settings

The conversion values can be entered in two menus. The Channel Configuration menu has
more options than the Calibration Coefficients menu, but both are acceptable ways to enter the
values and formulas.

Figure 55 - Abbreviated Conversion Values Menu

Figure 56 - Advanced Conversion Values Menu

62

Node Commander Software User Manual

8.2.1

Sensor Settings

Calculating a Linear Slope

A data analysis tool, such as Microsoft Excel, can be used determine the slope of a linear
relationship between sensor output A/D value (bits) and engineering units. This is not a
calibration unless a calibrated reference device is used to measure the applied loads. See
"Sensor Calibration " on page 67 for information and examples for determining calibrations
coefficients.
Here is an example, using Excel:
1. Open a blank spreadsheet.
2. Enter the A/D value (bits) measurements and applied load in the desired engineering
units in two columns. Enter A/D value is in the left column ( x-axis value) and the
applied load in the right (y-axis value).
3. From the Insert menu, select Chart, Scatter. Select the preferred format.

Figure 57 - Generate a Scatter Chart

63

Node Commander Software User Manual

Sensor Settings

4. Right click on the graphed line, and select Trendline.


5. Designate the line as Linear, and check the option to display the formula on the chart.

Figure 58 - Plot Trendline


6. The formula of the line is y=mx+b, where the y is the engineering units at a given point
(measurement), m is the slope of the line that represents the linear ratio, x is the A/D
value at a given point, and b is the fixed zero load offset of the sensor. Enter the slope
and offset as the conversion values for the sensor channel under the applicable
engineering units. In this example, enter 0.1338 for the slope and -282.36 for the
offset for the pounds units conversion values on the measured channel.

Figure 59 - Slope and Offset Values

64

Node Commander Software User Manual

8.2.2

Sensor Settings

Differential Input Gain and Offset

The combination of the gain, offset, and sensor signal cannot exceed the 0 to 3VDC input of
the analog to digital converter within the node (See "Example Gain and Offset Calculations"
on page 66).
l

Resolution: Applying gain to the sensor signal can be used to maximize the
measurement resolution. The more of the range that is used, the more digital counts
are available to measure the signal, which typically means higher resolution
measurements. Limitations to the gain adjustment are the sensor's measurement
capabilities and the 0 to 3V input range of the node. The signal produced after gain is
applied to the sensor at full scale must not exceed the input range of the node.
Offset Scale:The scale setting positions the no load measurement of the connected
sensor within the 0 to 3V range of the node input. The range of A/D counts that
corresponds with the 0 to 3V node input depends on the resolution of the node. For
example, a 12-bit node will have a full scale bit range of 4096 and a 16-bit node will
have a full scale bit range of 65535. A mid range setting positions the baseline offset
in the middle of the range (1.5V or full scale bits*1/2) and is used for sensors with
negative and positive going signals. The low range setting positions the baseline
offset in the bottom quarter range (75mV or full scale bits*1/4) and is used for sensors
with mostly positive going signals. The high range setting positions the baseline offset
in the top quarter of the range (2.25V or full scale bits *3/4) and is used for mostly
negative going signals.

Figure 60 - Differential Input Resolution and Offset (16-bit Node)

65

Node Commander Software User Manual

Sensor Settings

Example Gain and Offset Calculations


EXAMPLE 1:
Sensor signal range: 0 to 50mV
Gain setting: 21
Baseline offset range setting: Mid range
Calculations:
50mV * 21 = 1.05V (maximum voltage of sensor signal with gain)
1.05V + 1.5V = 2.55V (maximum input voltage to node with gain and offset)
Calculated node input over sensor range: 1.5 to 2.55V
This is a good setting because the node input voltage is within the 0 to 3V range.

EXAMPLE 2:
Sensor signal range: 0 to 50mV
Gain setting:30
Baseline offset range setting: Low range
Calculations:
50mV * 30 = 1.5V (maximum voltage of sensor signal with gain)
1.5V + 75mV = 2.25V (maximum input voltage to node with gain and offset)
Calculated node input over sensor range: 75mV to 2.25V
This may be a better setting than in Example 1 because the gain is higher, which could increase the
resolution of the measurement. The node input voltage is still within the 0 to 3V range.

EXAMPLE 3:
Sensor signal range: 0 to 50mV
Gain setting:75
Baseline offset range setting: Low range
Calculations:
50mV * 75 = 3.75V (maximum voltage of sensor signal with gain)
3.75V + 75mV = 4.5V (maximum input voltage to node with gain and offset)
Calculated node input over sensor range: 75mV to 4.5V
This setting will not work because the node input voltage is outside of the 0 to 3V range.

66

Node Commander Software User Manual

8.3

Sensor Settings

Sensor Calibration

Many sensors require calibration coefficients to accurately report measurements. Methods for
determining the calibration coefficients depend on the type of sensor measurement and
application. The Node Commander software facilitates multiple calibration methods.
Calibration calculators for some applications are also available by contacting LORD
MicroStrain Technical Support. See "Technical Support" on page 114.
l

Sensor manufacturers specifications or calibration: The slope and offset


values, or the data to derive them, are provided with the sensor by the manufacturer
to prove its accuracy and describe expected voltage output. Some sensors are
calibrated individually, while others are manufactured to a standard sensitivity value
(plus or minus some tolerance), which is provided in the device specifications.
Sensor lab calibration: If the manufacturers calibration is not available, or
outdated, calibration of the sensor can be performed with calibrated equipment in a
controlled environment. The calibration equipment and process will typically be
traceable to an industry standard, such as NIST or ASTM in the United States. Fixed
loads are applied to the sensor while the sensor output is recorded.The load is
applied or measured by a calibrated reference device. The known load value from the
calibrated device is then plotted against the measured output of the sensor to
determine the calibration slope and offset.In Node Commander this can be
accomplished by taking sensor readings while applying the known loads.

Sensor wiring, tolerances in system electronics, and differences in mounting techniques are
examples of systemic variables that can influence the sensor readings. Sensors that are
making small measurements or are otherwise sensitive to these slight differences may benefit
from a system calibration. The following techniques are system calibrations.
l

67

System shunt calibration (internal and external): This option is only available for
Wheatstone bridge type sensors (such as strain gauges) and utilizes a calibration
wizard in Node Commander. In the shunt calibration process, an internal or external
precision resistor is used to load part of the sensor bridge while the sensor remains
unloaded. The bridge output is measured and used as a loaded calibration point for
the sensor. In addition to the no load value it can be used to derive the calibration
slope and offset. The internal shunt resistor is suitable for most applications, however
an external shunt may be beneficial in high gain scenarios.

Node Commander Software User Manual

68

Sensor Settings

System field calibration: The field calibration is a similar methodology to the


sensor lab calibration. Known loads are applied to the sensor while the sensor output
is recorded. The load is applied or measured by a reference device. In this scenario,
the sensor may be installed in final field configuration, and the load may be applied
with the actual stimulus that the sensor will be monitoring. The known load value from
the reference device is then plotted against the measured output of the sensor to
determine the calibration slope and offset.In Node Commander this can be
accomplished by taking sensors readings while applying the known loads.

Node Commander Software User Manual

8.3.1

Sensor Settings

EXAMPLE: Lab or Field Calibration

The lab and field calibrations use similar methodology. See "Sensor Calibration " on page
67. The primary difference is the traceability and calibration environment. Lab calibrations
are performed in controlled environments with traceable equipment and procedures. Field
calibrations are more improvised, although calibrated equipment can still be used to improve
accuracy.
NODE:V-Link -LXRS ,16 bit (65536 A/D values)
CHANNELTYPE: differential analog input, 0 to 3VDC input range
SENSOR TYPE: load cell
SENSOR PARAMETERS: application voltage range: +/-20mV
This is the expected output voltage of the sensor based on the range of force being
measured in the application and the sensitivity of the sensor (V/engineering units)
DESIREDOUTPUT: engineering units (EU), force (lbs)
PROCEDURE:
1. Enter the node Channel Configuration menu by right clicking on the node name in
Node Commander and selecting the Configure option and Channels tab. Select the
check box for the channel the sensor is installed on, and then select the Configure
button for that channel.

Figure 61 - Node Configuration Menu

69

Node Commander Software User Manual

Sensor Settings

2. In this example, use the following settings (Figure 62 - Channel Settings):


a. Coefficients, Class: A/D value
b. Conversion Coefficients, Units: Bits
c. PGA Settings, Input Range: +/-20mV (expected sensor range)
d. PGA Setting: Mid scale (for positive and negative going signals)
3. Select the Auto Balance button to tare the no load value of the strain gauge. Click OK
to apply the node settings and exit configuration.

Figure 62 - Channel Settings


4. Right click on the Node heading, and then Sample, Stream, Start.

Figure 63 - Start Node Streaming

70

Node Commander Software User Manual

Sensor Settings

5. The streaming graph shows the bit output of the channel.


6. Using a calibrated tool, or some other way of applying and measuring a known load,
apply loads to the sensor at a number of intervals over the expected range of use. At
each of the calibration intervals, record the applied force and the corresponding
sweep value on the Yaxis of the graph (the A/D value output of the sensor).
a. Zoom in and out on the graph by un-checking the Auto Y-Axis Zoom box,
and then right clicking on the graph and selecting Zoom In. Draw a box
around the area to zoom in on.
b. Adjust the X-Axis Width from the field next to the Y-Axis Zoom.
c. End sampling by clicking the red X box on the Streamin Graph tab.

Figure 64 - Node Sampling

71

Node Commander Software User Manual

Sensor Settings

7. After making all measurements, calculate a slope from the data using the formula y =
mx+b in a data analysis program, such as Microsoft Excel. See "Calculating a Linear
Slope" on page 63.
8. Return to the Node Configuration screen for the sensor channel, and select the
desired Conversion Class and Units, and enter the slope and offset values derived in
the data analysis program.

Figure 65 - Enter Calibration Values


9. Save the values and exit configuration. This is the end of a lab calibration.
10. For field calibrations, begin node data streaming again with no load on the sensor.
11. Observe the value in the stream graph. If the stream is not at zero, return to the
channel configuration menu and adjust the Offset by increasing or decreasing the
value.
12. Once the offset has been zeroed, verify the calibration by applying known loads to the
sensor throughout the range and observing output in engineering units.

72

Node Commander Software User Manual

8.3.2

Sensor Settings

EXAMPLE: Internal Shunt Calibration

NODE: V-Link -LXRS ,16 bit (65536 A/D values)


CHANNELTYPE:differential analog input, 0 to 3VDC input range
SENSOR TYPE: strain gauge, Wheatstone Bridge, full bridge configuration
SENSOR PARAMETERS: application voltage range: +/-2mV
This is the expected output voltage of the strain gauge based on the range of strain being
measured in the application and the sensitivity of the gauge (volts/strain).
DESIREDOUTPUT: engineering units, microstrain
PROCEDURE:
1. Enter the node channel configuration menu by right clicking on the node name in
Node Commander and selecting Configure, Configure Node, and then the
Channels tab. Select the check box for the channel the strain gauge is installed on,
and then select the Configure button for that channel.

Figure 66 - Node Configuration Menu

73

Node Commander Software User Manual


2. In this example, use the following settings;
a. Coefficients, Class: Strain
b. Conversion Coefficients, Units: uStrain
c. PGA Settings, Input Range: +/-2.5mV (expected sensor range)
d. PGA Setting: Mid scale (for positive and negative going signals)
3. Select the Auto Balance button to tare the no load value of the strain gauge.

Figure 67 - Channel Settings

74

Sensor Settings

Node Commander Software User Manual

Sensor Settings

4. Select the Strain Wizard.


5. Select Full Bridge for the Bridge Type. ClickNext. Select Use the Strain
Measurement Wizard, and then click Next.
6. Set the following for this example;
a. Number of Active Gauges: 4 (for a full bridge strain gauge)
b. Gauge Factor: (ratio of mechancial strain to electrical output, a gauge
specification)
c. Gauge Resistance: (enter strain gauge ohm value, a gauge
specification)
d. Shunt Resistance: 499000 ohms

Figure 68 - Strain Wizard Settings


NOTE
Touching connected sensors and test boards or charging the node battery
while acquiring sensor data may induce noise on sensitive sensor signals and
is not recommended.

75

Node Commander Software User Manual

Sensor Settings

7. Select Calibrate.
8. Verify the calibration looks as shown in Figure 69 - Strain Gauge Calibration. The
green line represents the output of the strain gauge. With no load applied it should sit
near the Auto Balance baseline value and is represented by the red dashed line.
During calibration, a shunt resistance (selecting on the Parameters page) is applied
across the strain bridge, shown by the square pulse on the output. The Offset value,
shown with the dashed blue line, is the average output value of the pulse and should
sit across the top of the pulse. If the gauge has not had to time to equilibrate before
sampling or if varying environmental factors exist, spikes in the gauge output may
occur and affect the Baseline and Offset values. If this occurs, the Offset and
Baseline values can be adjusted to clip the spikes in the output values. Adjust them
as needed and select Accept when completed.

Figure 69 - Strain Gauge Calibration

76

Node Commander Software User Manual

Sensor Settings

9. Select Finish to end the Strain Wizard. Note the slope and offset values have been
calculated automatically.

Figure 70 - Completed Strain Wizard


10. Select OK to exit the Channel Configuration window, calibration is complete.

77

Node Commander Software User Manual

8.3.3

Sensor Settings

EXAMPLE: Manufacturer Calibration

NODE: V-Link -LXRS ,16 bit (65536 A/D values)


CHANNELTYPE: differential analog input, 0 to 3VDC input range
SENSOR TYPE: pressure transducer, voltage output, positive going
SENSOR PARAMETERS:
From the manufacturer calibration sheet included with the sensor:
sensor range: 0-250 psi
sensor zero load output: 0.0032 VDC
sensor full scale output (FSO) with 10V excitation: 86.07mV
Application parameter:
sensor excitation in application: 3V supply from the node
DESIREDOUTPUT: engineering units (EU), psi
CALCULATIONS:
Because the sensor will be powered from the node with 3V, and the sensor manufacturer
calibrated it a 10V, the manufacturer full scale output (FSO) value needs to be scaled to 3V.
(3V/10V) * 86.07mV = 25.82mV
Select a gain and offset scale value appropriate for the sensor. Because the signal is
positive going in this example application, the low offset scale will provide the largest range.
With the low offset selected, the effective input range of the node is 75mV to 3V (see
"Differential Input Gain and Offset" on page 65 ), or 2.25V. Calculate the highest gain
possible by dividing the actual input range by the sensor FSO.
2.25V/25.82mV = 87
The closest gain setting below optimal gain for a V-Link -LXRS is 75 (+/-20mV). Using a
higher gain value would exceed the input voltage capacity of the node when the sensor is at
higher pressures. This selection makes sense because the approximate input range
designation for a gain of 75 is +/-20mV (a 40mV delta minus 10mV for the low offset), which
is close to the FSO range of the sensor.

78

Node Commander Software User Manual

Sensor Settings

Multiply the sensor FSO by the gain setting to get the sensor voltage after amplification.
75 * 25.82mV = 1.9365V
Scale the (gained) sensor input voltage/EU ratio to the node input voltage/EU ratio to
determine the equivalent node FSO value (x).
1.9365V/250psi=3V/x
(250psi * 3V)/1.9365V = x = 387. 3psi
The node converts voltage inputs to A/D values. For a 16-bit node, there are 65536 A/D
values over the 3V input range. Divide the node EU FSO by the A/D value to get the ratio, or
slope, of EU to A/D value.
387.3 psi/65536 bits = 0.00591 = slope
Once the slope is entered, the sensor offset value can be measured in a data sampling
session, such as streaming. Sample the sensor channel with no load applied, and read the
EUvalue. Enter this as a negative value for the offset in order to have it subtracted from
readings.

79

Node Commander Software User Manual

9.

Sampling Modes and Settings

Sampling Modes and Settings

Depending on the type of node being used, Node Commander offers up to four sampling modes.
Some modes have user- configurable settings for sample rate, sample duration, event based
sampling, beaconing, and data logging schemes. Other settings are automatic depending on the
number of active channels and other variables. Table 6 - Sampling Mode Settings Summary
describes the features and settings available in each mode.
Streaming: a general purpose sampling mode with very little latency between the sampling
and the data display. Sampling rates are automatic, although duration is configurable.
Streaming mode does not support time stamping.
Synchronized Sampling: the only mode that uses the lossless data protocol, and has the
most accurate time stamping scheme. Includes burst and event based sampling features,
datalogging and adjustable sample rates, and duration. It also coordinates node networks to
eliminate data collisions and guarantees node sampling within 32 microseconds of each other.
There is some latency between sampling and data viewing.
Low Duty Cycle: used for low frequency sampling over long durations. Features included
configurable sample rate and duration, event based sampling, datalogging, and time stamping.
Datalogging: saves data to internal node memory for user retrieval. Features include
adjustable sample rates, duration, multiple node triggering, and time stamping.
Sampling settings are accessed through the Configure Node menu. There is a tab for each
sampling mode.

Figure 71 - Sample Settings Menu

80

Node Commander Software User Manual

Sampling Mode

Feature
sample rate, each active
channel

Streaming
sampling duration
standard sample rate,
each active channel
sampling duration
burst sampling rate, each
active channel
Synchronized
Sampling

Sampling Modes and Settings

burst duration

time between bursts

Setting
l

node dependent, user selectable or


continuous
node dependent, user selectable
node and sample rate dependent, user
selectable
node dependent, user selectable
node and sample rate dependent, user
selectable
node and burst duration dependent, user
selectable

event based sampling

on/off, trigger levels, pre/post event logging

logging scheme

log only, transmit only, log and transmit

beaconning

system wide clock synchronization

sample rate, each active


channel

node dependent, user selectable

sampling duration

Low Duty Cycle

Datalogging

automatic

node dependent, user selectable or


unlimited

event based sampling

on/off, trigger levels, post/pre event logging

logging scheme

log only, transmit only, log and transmit

sample rate, each active


channel

node dependent, user selectable

overall sampling duration


multiple node trigger

node dependent, user selectable or


continuous
starts node sampling simultaneously

Table 6 - Sampling Mode Settings Summary

81

Node Commander Software User Manual

9.1

Sampling Modes and Settings

Streaming

In streaming mode, sample rates are automatic, based on the number of active channels. Table
7 - Streaming Sample Rates describes the automatic settings. Sampling can be set to repeat
for a fixed interval of time after the sampling has been started or continuously until a stop
command is sent. The data is transmitted continuously during sampling, and none is saved to
the node memory. It is saved on the host computer in a data file. No time stamp is used in
streaming mode.

Active
Channels

Sample Rate
(samples/sec)

1
2
3
4
5
6
7
8

736
679
617
565
520
485
452
424

Table 7 - Streaming Sample Rates


To adjust streaming sample settings, open the node configuration menu by right clicking on the
node name and then selecting Configure, Configure Node. Select the Streaming tab.

Figure 72 - Streaming Settings

82

Node Commander Software User Manual

9.2

Sampling Modes and Settings

Synchronized Sampling

Synchronized sampling is the most capable and accurate sampling method. Sample rates are
set by the user and there is a burst mode for high speed, short duration sampling. Event based
sampling is also available with pre and post event data saving options.
In standard synchronized sampling, the sampling can be set to repeat for a fixed interval of time
after the sampling has been started or continuously until a stop command is sent. In burst
sampling, the duration of the burst and time between bursts is configurable. The sampling
repeat interval is limited to one day. In event based sampling, the data will be continuously
sampled but only recorded when it meets the event criteria. The node can also be configured to
store the data in the node memory, to store the data and transmit it to the gateway, or to
transmit only. A time stamp is obtained from the host computer when sampling is started. Each
sample is then incrementally stamped based on that starting point and the node clock.
To adjust synchronized sampling settings, open the node configuration menu by right clicking
on the node name and then selecting Configure, Configure Node. Select the Synchronized
Sampling tab. Key menu options include the burst and event based sampling features.

Figure 73 - Synchronized Sampling Settings

83

Node Commander Software User Manual

9.2.1

Sampling Modes and Settings

Using the Beacon

In synchronized sampling, the beacon feature is used to coordinate sampling and


transmission timing between multiple nodes. The primary purpose is to avoid data collisions
and guarantee timing between samples from different nodes. The beacon cannot be used in
any another sampling mode.
When the beacon is enabled, the gateway broadcasts a data message once every second
containing a Coordinated Universal Time (UTC) time stamp, and is used to update each
nodes real time clock (RTC). This continuous synchronization beacon, coupled with the
accuracy of the gateway and node clocks (+/- 3 ppm drift), provides a node to node
synchronization of +/-32 microseconds. The gateway receives the UTC from the host
computer, so the host must be connected to a UTC time server to achieve that accuracy.
There is some lag inherent in the Microsoft Windows distribution of the UTC. This dictates
the overall accuracy of the time stamp, but the synchronization times between the gateway
and nodes will be consistent because of the accuracy of the embedded RTC clocks.
Nodes must be on the same transmission frequency (channel) as the gateway to receive the
beacon broadcast command. It is recommended that the beacon be disabled when not
performing synchronized sampling, or when not needed, to minimize radio interference.
To enable and disable the beacon, right click on the gateway name and select the
appropriate command. When the beacon is active, the gateway device status indicator will
be blue and pulse once a second.

Figure 74 - Synchronized Sampling Beacon

84

Node Commander Software User Manual

9.3

Sampling Modes and Settings

Low Duty Cycle

Low duty cycle sampling is used for sampling up to 500 times per second. Sample rates are
user adjustable, but are eventually limited by the number of nodes that are sampling and
transmitting simultaneously (see "Bandwidth and Sample Rate" on page 88). Sampling can be
set to repeat for a fixed interval of time after the sampling has been started or continuously until
a stop command is sent. In low duty cycle, the node can be configured to store the data in the
node memory, to store the data and transmit it to the gateway, or to transmit only. Event driven
sampling is also available in low duty cycle. See "Event Driven Sampling" on page 87. For
sample rates over 1Hz, the time stamp in low duty cycle is set by the host computer when the
sampling is started. For sample rates under 1Hz, samples are stamped when the data arrives
at the host computer after sampling and transmission.
To adjust low duty sample settings, open the node configuration menu by right clicking on the
node name and then selecting Configure, Configure Node. Select the Low Duty Cycle tab.

Figure 75 - Low Duty Cycle Settings

85

Node Commander Software User Manual

9.4

Sampling Modes and Settings

Datalogging

In datalogging mode, sampled readings are stored in the node memory instead of being
transmitted to the gateway during sampling. The sample rate is adjustable, and sampling can
be set to repeat for a fixed interval of time after the sampling has been started or continuously
until a stop command is sent. If the node is set to sample continuously, it will do so until the node
memory is full. The data is not transmitted to the gateway until the download command is
initiated from the gateway menu. In datalogging mode, the time stamp in is set by the host
computer when the sampling is started. Multiple nodes can be triggered to start data logging at
the same time. See "Armed Datalogging" on page 98.
To adjust datalogging settings, open the node configuration menu by right clicking on the node
name and then selecting Configure, Configure Node. Select the Datalogging tab.

Figure 76 - Datalogging Settings

86

Node Commander Software User Manual

9.5

Sampling Modes and Settings

Event Driven Sampling

Event driven sampling is available in both the low duty cycle and synchronized sampling
modes. In either mode the node is still taking continuous measurements based on the sample
rate and duration settings, however it only reports the measurement that meet the event
criteria. The node can be configured to also transmit the data just before and just after the
event. The pre and post event duration settings define the window of time before and after the
event that will be reported.
The event triggers are set for each channel in the form of a voltage threshold. The trigger
occurs when the channel reading is either above or below the threshold, depending on the
trigger operation settling.
The event driven sampling menus are accessed through the node configuration menu. Right
click on the node name and then select Configure, Configure Node. Select the desired
sampling mode (synchronized sampling or low duty cycle) and then select the Event Driven
Sampling check box, or the Configure button if the check box is already selected. The
Configure Event Driven Sampling menu will appear.

Figure 77 - Event Based Sampling Settings

87

Node Commander Software User Manual

9.6

Sampling Modes and Settings

Bandwidth and Sample Rate

In general, when determining what sample mode and rate is most suitable for the application,
consider the following;
l

9.7

Increasing the sample rate reduces the system bandwidth and therefore the number
of nodes that can be reporting simultaneously.
Increasing the sample rate increases the duration the node is on over time and
therefore reduces battery life.
Some sampling modes, such as Streaming (which is continuously recording data),
will use more bandwidth and battery power than fixed interval or event driven modes
such as Low Duty Cycle and Event Driven Sampling.
Synchronized Sampling includes a bandwidth calculator in the node configuration
menu and in the Synchronized Sampling menu that updates depending on what
options and nodes are selected.

Time Stamping

A time stamp may be attached to the sensor data packets to indicate what time a sample was
taken. Each sampling mode handles time stamping differently.
Streaming: No time stamp is used in streaming mode.
Synchronized Sampling: Synchronized sampling has the most accurate sample timing. The
time stamp is obtained from the host computer when sampling is started. Each sample is then
incrementally stamped based the precision real time clock in the node. Timing accuracy is
guaranteed to be within +/-32 microseconds. Absolute time is dependent on the accuracy of the
host computer in the initial stamp.
Low Duty Cycle: For sample rates under 32Hz, samples are stamped when the data arrives
at the host computer after sampling and transmission. For sample rates over 32Hz, a time
stamp is obtained from the host computer when sampling is started. The time stamp for each
subsequent sample is then calculated from the sample rate. Absolute time is dependent on the
accuracy of the host computer in the initial stamp.
Datalogging: A time stamp is obtained from the host computer when sampling is started. The
time stamp for each subsequent sample is then calculated from the sample rate. Absolute time
is dependent on the accuracy of the host computer in the initial stamp.

88

Node Commander Software User Manual

10.

Data Acquisition

Data Acquisition

During data acquisition, one sampling mode is chosen for each node. Nodes connected to the
same gateway can sample in different modes, however each node can only run one sampling
mode at a time. The sampling settings for the node will be applied to all channels. See "Sampling
Modes and Settings" on page 80 for information about the sampling settings.
Nodes can continue to sample even without Node Commander running or actively displaying
data. See "Ending Sampling Sessions" on page 102.
All sampling modes automatically save data to the host computer. See "Viewing Sensor Data" on
page 105.

NOTE
Touching connected sensors and test boards or charging the node battery
while acquiring sensor data may induce noise on sensitive sensor signals and
is not recommended.

89

Node Commander Software User Manual

10.1

Data Acquisition

Streaming Data

Streaming may only be initiated on one node at a time. To start a streaming data session, right
click on the node name in the Controller window and select Sample,Stream, Start. The
Resync menu option is used to resume viewing a streaming session that is already running.

Figure 78 - Streaming Data Graph


a. Node information:
communications port.

displays

the

node

serial

number,

frequency,and

b. Axis range: select the X-axis width and Y-axis zoom percentage, or used the Auto
check box for automatic scaling.
c. Graph: the node graph shows the sampled data. Each active channel is displayed
as a different color. The X-axis is time in seconds and the Y-axis in the A/D value
(bits). Right click on the graph for additional menu options such as show key, pan,
zoom, and stop streaming.
d. End sampling: the red "X" is used to exit the sampling window and/or end
sampling. See "Ending Sampling Sessions" on page 102.

90

Node Commander Software User Manual

10.2

Data Acquisition

Synchronized Sampling

To start a sampling session, nodes can be selected individually or as a group. When selected
as a group, they will all be set to the same sampling mode. Right click on the node or group of
node names and select Sample. The menus are different, depending on which method is
selected.

Figure 79 - Starting a Sampling Session


When a synchronized sampling session is started, the sampling menu appears and includes
settings to enable optional sampling features, configure nodes, and specify where the data is
saved. The built in bandwidth calculator displays the total bandwidth used by the nodes
selected for synchronized sampling (Figure 80 - Synchronized Sampling Menu).
a. Save Location: indicates where the data file will be saved on the host computer.
Use the Browse button to select a preferred location.
b. Node configuration: includes the node serial number, sampling settings,
bandwidth calculation, and current status. Highlight any node or group of nodes,
and the Remove, Configure, and Refresh buttons become active. The Configure
button opens the node configuration menus to adjust settings as needed, and
recalculates the node bandwidth. Multiple nodes can be configured together by
using the Shift or Ctrl key to select them.

91

Node Commander Software User Manual

Data Acquisition

Figure 80 - Synchronized Sampling Menu


c. Lossless: enables the lossless data protocol . The protocol enables buffering and
retransmission of data in order to provide 100% data collection success. Using this
feature may increase data display latency.
d. High Capacity: reduces the transmit rates in order to optimize bandwidth and
power savings among nodes with slower sample rates.
e. Network Bandwidth: the total bandwidth used by all the nodes. See "Bandwidth
and Sample Rate" on page 88 for more information.
f. Enable Beacon on Start: This setting is used in conjunction with (or as an
alternate to) enabling the beacon in the gateway menu (see "Using the Beacon" on
page 84). Enabling the beacon on start will begin sampling on all nodes immediately
instead of waiting for the beacon from the gateway, which is on a fixed interval and
may be delayed. Enabling the beacon on start can also be used instead of the
gateway enable for scenarios in which a continuous beacon is not desired, such as
when the beacon is being received from another source.
g. Apply settings and start sampling before acquisition can begin, use the Apply
Network Settings to save the session settings to the node. When completed, select

92

Node Commander Software User Manual

Data Acquisition

Start Sampling to begin.


h. Close sampling window with the red "X" to exit sampling, or once the sampling
has been started to view the data window behind it.
Syncronized sampling features two data views, a grid view and a graph view. Once sampling is
started, the data grid view is the default view.

Figure 81 - Synchronized Sampling Data View


a. Device status: Node sampling mode and gateway status are displayed in
parenthesis next to the device name.
b. Node information: includes node serial number and sampling statistics. Right click
on the node name for more menu options such as Stop Nodes.
c. Data: display of the sampled data with each channel in a column
d. Radiostrength: indicates how good the communication is between the gateway
and node. See "Range Test" on page 49.
e. Data file: the location and size of the data file, as data is added. View the data in
CSV format with the Open File button.
f. View menu: select between Data Grid and Graph views.
g. End sampling: the red "X" is used to exit the sampling window and/or end
sampling. See "Ending Sampling Sessions" on page 102.

93

Node Commander Software User Manual

Data Acquisition

Use the view menu to select the Graph view of the data. Click on the node to view the graph for
that node. Click again to hide it.

Figure 82 - Synchronized Sampling Graph View


a. Available Nodes: Click on the node to display the graph for that node. Click again
to hide it. Right click on the node name for more menu options such as stop nodes
and save stream.
b. Axis range: Select the X-axis width and Y-axis zoom percentage, or used the Auto
check box for automatic scaling.
c. Graph: the node graph shows the sampled data. Each active channel is displayed
as a different color. The X-axis is time in seconds and the Y-axis in the A/D value
(bits). Right click on the graph for additional menu options such as View Graph Key,
Pan, Zoom, Pause and Remove Graph.
d. View menu: Select between Data Grid and Graph views.
e. Data file: The location and size of the data file as data is added. View the data in
CSV format with the Open File button.
f. End sampling: The red "X" is used to exit the sampling window and/or end
sampling. See "Ending Sampling Sessions" on page 102.

94

Node Commander Software User Manual

10.3

Data Acquisition

Low Duty Cycle Data Sampling

To start a sampling session, nodes can be selected individually or as a group. When selected
as a group, they will all be set to the same sampling mode. Right click on the node or group of
node names and select Sample. The menus are different, depending on which method is
selected.

Figure 83 - Starting a Sampling Session

Low Duty Cycle sampling features two data views a grid view and a graph view. The data grid
view is the default view when sampling is started. Figure 84 - Low Duty Cycle Data Grid View,
shows the indicators and options in the grid view.

95

Node Commander Software User Manual

Data Acquisition

Figure 84 - Low Duty Cycle Data Grid View


a. Device status: Node sampling mode and gateway status are displayed in
parentheses next to the device name.
b. Node information: includes node serial number and sampling statistics. Right click
on the node name for more menu options such as Stop Nodes.
c. Data: display of the sampled data with each channel in a column
d. Radiostrength: indicates how good the communication is between the gateway
and node. See "Range Test" on page 49.
e. Data file: The location and size of the data file as data is added. View the data in
CSV format with the Open File button.
f. View menu: Select between Data Grid and Graph views.
g. End sampling: The red "X" is used to exit the sampling window and/or end
sampling. See "Ending Sampling Sessions" on page 102

96

Node Commander Software User Manual

Data Acquisition

Use the view menu to select the Graph view of the data. Click on the node to view the graph for
that node. Click again to hide it.

Figure 85 - Low Duty Cycle Graph View


a. Available nodes: Click on the node to display the graph for that node. Click again
to hide it. Right click on the node name for more menu options such as stop nodes
and save stream.
b. Axis range: Select the X-axis width and Y-axis zoom percentage, or used the Auto
check box for automatic scaling.
c. Graph: The node graph shows the sampled data. Each active channel is displayed
as a different color. The X-axis is time in seconds and the Y-axis in the A/D value
(bits). Right click on the graph for additional menu options such as View Graph Key,
Pan, Zoom, Pause and Remove Graph.
d. View menu: Select between Data Grid and Graph views.
e. Data file: The location and size of the data file as data is added. View the data in
CSV format with the Open File button.
f. End sampling: The red "X" is used to exit the sampling window and/or end
sampling. See "Ending Sampling Sessions" on page 102.

97

Node Commander Software User Manual

10.4

Data Acquisition

Armed Datalogging

Datalogging is started by opening a trigger session in Node Commander . Datalogging can be


initiated for one or multiple nodes at once. Datalogging settings are selected in the node
configuration menu and can vary between nodes.
1. Right click on one node or multiple node selections and select Sample, Datalogging,
Trigger Session or Trigger Datalogging, depending if there is one or multiple nodes
selected.
2. The selected nodes will appear in the Armed Datalogging window. The gateway will
send a Ping command to each node. The Status column will indicate if the ping was
successful, meaning the node is communicating with the gateway and ready for data
acquisition.
3. Enter a note for the data file, if desired.
4. Select Trigger All to begin data acquisition.

Figure 86 - Datalogging Menu

98

Node Commander Software User Manual

Data Acquisition

5. The Progress column will indicate the percentage of sampling that has occurred.
When it reaches 100%, data acquisition is complete. At any time during the data
acquisition, one or all nodes can be stopped by selecting the Cancel Selected or
Broadcast Stop All button. This stops the data acquisition process and sets the node
back in idle. Alternatively, the Close button can be used to close the datalogging
window without ending the data acquisition. Data acquisition will continue until done
or until a stop command is sent to the node.

Figure 87 - Datalogging

NOTE
Depending on datalogging settings for each node, data acquisition may take
longer to complete on some nodes than others. The Close button can be used
to close the datalogging window without stopping the data acquisition. The
Close Selected or Broadcast Stop All buttons are used to stop data acquisition
on one, or all nodes.

99

Node Commander Software User Manual

Data Acquisition

6. Once the datalogging menu is closed, the nodes that have acquired data will
automatically display the node information in the Controller window. The Memory
Percent Full indicates the percentage of memory that has been used for datalogging.
Number of Triggers indicates how many datalogging files have been created.

Figure 88 - Node Memory Statistics


7. To download the datalogging files off of the node onto the host computer, right click
on a node name, or group of nodes, and select either Sample, Datalogging,
Download or Sample, Download Triggers.
8. In the Download Triggers menu, select the node or nodes, and the Start Download
button.

Figure 89 - Datalogging File Download

100

Node Commander Software User Manual

Data Acquisition

The data that is stored in the node memory can be erased from the download window or
from the node menu. The erase selection is only available in the node menu when selecting
just one node at a time. Right click on the node name, and select Sample, Datalogging,
Erase.

Figure 90 - Erase Node Memory


Data can be viewed from the download window, or from the node menu. The view triggers
selection is only available in the node menu when selecting just one node at a time. Right
click on the node name, and select Sample, Datalogging, Browse Triggers. Browse to the
desired files to open. The data files are in CSV format.

Figure 91 - View Downloaded Data

101

Node Commander Software User Manual

10.5

Data Acquisition

Ending Sampling Sessions

There are multiple ways to end sampling sessions. Some end the session and stop the
sampling, while other methods just end the session and allow the node to continue sampling.
1. Exit the data view window: Click the red "X" in the upper right of the data sampling window. This option is used to continue sampling after Node Commander is closed, or to
end sampling.

Figure 92 - Ending a Sampling Session


When exiting the sampling session in this way, data acquisition can be stopped or continue
after the data view is closed.

Figure 93 - Stop Nodes or Continue Sampling

102

Node Commander Software User Manual

Data Acquisition

2. Stop the nodes during sampling: In modes with data grid and graph views, right click in
the data sampling window or on the node name to display a menu with the Stop All Nodes or
Stop Streaming command. In datalogging mode there is a Stop Broadcast button. This ends
all data acquisition.

Figure 94 - Stop All Nodes


3. Use the Stop Node command: If the node is sampling without an active view of the data
acquisition, right click on the node name in the Controller window and select Stop Node to
end sampling. This ends data acquisition for the selected node.

Figure 95 - Stop Single Node

103

Node Commander Software User Manual

10.6

Data Acquisition

Monitor Mode

Monitor Mode is used to observe any sampling activity that is occurring on nodes connected to
a particular gateway. Monitoring detects low duty cycle and synchronized sampling and
displays the sampling window. This is useful if the node was configured to start sampling on
boot up or if the data acquisition screen was closed without stopping the node. It is a way to
enter the data and graph view windows after data acquisition has already begun. The gateway
will only monitor nodes that are on the same frequency (channel) that it is on.
To monitor a node, right click on the gateway name and select Monitor Mode.

Figure 96 - Monitor Mode

104

Node Commander Software User Manual

10.7

Data Acquisition

Viewing Sensor Data

Acquired data is stored in CSV format and can be viewed with Microsoft Excel, Quattro Pro,
Open Office and other CSV editors and spreadsheet programs. The files can be found in the
default directory or the location specified at the beginning of sampling (as applicable).
The default directory is: C:\ProgramData\Microstrain\NodeCommander\SampledData
Use Windows Explorer or the Node Commander File menu to access the files. Different
sampling modes will output different file types, and they will be categorized in separate folders
by sampling mode and then further categorized by date, session, and/or node serial number.
Synchronized sampling and low duty cycle files are found in the Sampled Data folder.
Datalogging files need to be downloaded from the node before they are available for viewing
and can be accessed through datalogging menus as well as the File menu. They are stored, by
default, in the Downloaded Triggers folder.
Streaming data is stored in the Streaming folder.

Figure 97 - Exploring Data

105

Node Commander Software User Manual

11.

Support

11.1

106

Troubleshooting Guide

Support

Node Commander Software User Manual


Problem

Possible cause and recommended solution


1.1 power switch is off

1. POWER
gateway or node does
not turn on, or node
does not charge

The status indicator LED on the device will be off. Turn the
device on and the status indicator LED should illuminate.
1.2 external power is off or miswired
Verify the device power source is connected correctly and
turned on.
1.3 wrong power supply
Using a power supply other than the one specified the device or
an external supply that is outside of the device operating range
could result in permanent damage to the device. Or it may just
not work.
1.4 node internal source select switch is set incorrectly
When the node is manufactured, it is set to internal battery
operation, but it can be configured to accept an external source.
When set to an external source, the battery cannot be charged.
1.5 node battery is dead
If the node power source selector is set to internal and the node
will not power on, attempt to charge the node battery. If the
battery needs to replaced, contact Technical Support ( See
"Technical Support" on page 114).
1.6 node battery fault
If the battery charge indicator on the node is only dimly
illuminated when charging is attempted, a battery fault condition
has occurred. Unplug power and then plug it back in. The
indicator should turn on brightly, indicating charging.
1.7 sensors are drawing too much current
The node battery can only supply a limited amount of power to
the connected sensors. If an over-current condition occurs, the
node will shut down. Consider powering the node or sensors
with an external source.

107

Support

Node Commander Software User Manual


Problem

Possible cause and recommended solution


1.8 node or gateway is damaged
If all power settings and connections have been verified, and the
node is still unresponsive, contact Technical Support. ( See
"Technical Support" on page 114).
2.1 node or gateway has no power

2. COMMUNICATION
no communication to
the gateway or node

Verify the node and gateway have power applied and applicable
power switches are on. Power is indicated on both devices by a
status indicator LED.
2.2 gateway has no communication with the computer
Verify gateway communication in the software. Check, remove,
and reconnect communications and power cables as
applicable.
l

For serial gateways, verify the COM port setting.


For USBgateways, verify the drivers are installed on the
computer (included with Node Commander) and that the
software has had sufficient time to detect it.
For Ethernet gateways, use Live Connect to verify
communications on a DHCPnetwork. Check that the
extended timeouts are enabled in the Node Commander
Edit, Preferences menu, under Devices. Once
communication has been established the network
configuration can be changed.

2.3 node cannot be configured


Observe the node status indicator LED to determine what state
it is in: boot, idle, sample or sleep. If the node is sampling or
sleeping, it cannot be configured. In Node Commander ,
execute the Stop Node command to put the node in idle state
and allow configuration to occur.
If the user inactivity timeout is set very low, the configuration
menu will have to be entered quickly, before the timeout occurs
and puts the node back in a sample or sleep state.

108

Support

Node Commander Software User Manual


Problem

Possible cause and recommended solution


2.4 node is out of range
Perform a bench test with the node and gateway in close
proximity to verify they are operational. See "Range Test" on
page 49 for range test and installation recommendations. The
system has been tested to operate with the node and gateway
up to 2km apart, with clear line of sight.
2.5 node is not in normal start-up mode
If the node status indicator shows the node booting in a mode
other than the normal boot mode, it can be bypassed by toggling
the node ON/OFF switch rapidly three times, then leaving it in
the ON position for normal power up. In normal boot mode the
conmmunication can be established with automatic node
discovery or manually once the boot process is complete and
the node is in the idle state. Start up mode can then be changed
in the software.
2.6 node is sampling
Observe the node status indicator LED to determine what state
it is in: boot, idle, sample or sleep. If the node is sampling, it
cannot be configured. In Node Commander, execute the Stop
Node command to put the node in idle state, and allow
configuration to occur.
2.7 node is sleeping
Observe the node status indicator LED to determine what state
it is: boot, idle, sample or sleep. If the node is sleeping, it cannot
be configured. In Node Commander , execute the Stop Node
command to put the node in idle state, and allow configuration to
occur.
2.8 gateway or node is damaged
Verify all connections, power, and settings. If available, try
installing alternate nodes and gateways one at a time to see if
the faulty device can be identified. If no conclusion can be
determined, or to send a device in for repair, contact Technical
Support (See "Technical Support" on page 114).

109

Support

Node Commander Software User Manual


Problem

Possible cause and recommended solution


3.1 no communication to node or gateway

3. DATAACQUISITION
sensor data is missing
or incorrect

Verify connections and power to the node and gateway. Verify


they are powered on and communicating in the software. Enter
a configuration menu, to verify that the node can be accessed.
3.2 sampling settings are incorrect
If the sampling mode, rate, or duration are not performing as
expected, enter the node configuration menu, and verify the
sampling settings.
3.3 sampling has not started
If sampling is occurring, the mode will be displayed next to the
node name in Node Commander . The node device status
indicator will also be flashing the sampling mode code. If the
node is not sampling, activate it in the software or with a sample
on start up boot sequence.
3.4 sensor is not connected correctly
Verify sensors connections and wiring. For non standard
connections contact Technical Support ( See "Technical
Support" on page 114).
3.5 sensor channel not configured correctly
Verify that the sensor is configured on the correct channel and
has been enabled for data acquisition.
3.6 sensor calibration is invalid
If possible, perform a field verification of the sensors by applying
known loads and comparing the measured values. In the
channel configuration settings, verify that the sensor channel
units are selected correctly. Verify that the calibration
calculations are correct. Verify that the gain and offset range
are correct and that the baseline offset is set with Auto Balance,
as applicable. Verify that the calibration values (slope and
offset) are correct. Verify that all parameters were written to the
node channel. Recalibrate as needed.

110

Support

Node Commander Software User Manual

11.2

Support

Communications Ports inWindows

The Windows Device Manager can be used to determine what communication port a RS232
or USB gateway is connected to.
1. Click on the Windows Start icon and select Control Panel.
2. Navigate to the System menu and select Device Manger. The menus may appear
different depending on the version of Windows and the View settings.

Figure 98 - Windows System Menu

111

Node Commander Software User Manual

Support

3. In Device Manager, expand the view for Ports (COM and LPT). Active COM ports will
appear on this list with the COM port number. A USB gateway will be displayed as
USB to UART Bridge (COM X ). A serial gateway will be attached to the
Communications Port (COMX). If no port is listed, the port is not recognized by the
computer, and no gateway communication can be established.

Figure 99 - Windows Device Manger Menu


4. To view the port settings, select the device, right click on it, and select Properties.
Table 8 - Serial Port Settings, describes the settings for serial communication. The
baud rate for the RS232 gateway can be adjusted in Node Commander only.
Changing it in the Windows port settings menu will not work. See "Setting the Serial
Baud Rate" on page 33. USB gateways run only at 921,600 baud.

Port Setting
Baud Rate
Parity
Data Bits
Stop Bits

Value
115,200 (RS232 gateways) or
921,600*
none
8
1

* for RS232 gateways, see "Setting the Serial Baud Rate" on page 33.

Table 8 - Serial Port Settings

112

Node Commander Software User Manual

11.3

Support

Live Connect Interface

Live Connect is a bridge between Node Commander , Ethernet connectivity, and web
based applications.
See "Gateway Ethernet Communications" on page 27 for information on connecting to the
Ethernet gateway in Node Commander.
Live Connect also provides an interface to the LORD MicroStrain web based data
aggregation and data visualization software, SensorCloud. Data files created in Node
Commander can be uploaded to SensorCloud through the Node Commander File menu.

11.4

Software Specifications
Parameter

Software Name
Operating Systems
Computer Requirements
For Use With

113

Specification
Node Commander and Live Connect
Microsoft Windows XP Pro, Vista, and 7, 32/64-bit
1 GHz processor, 256MB memory, 50MB free hard drive space for the application,
32 MB video card, 800 x 600 video resolution
All LORD MicroStrain LXRS , mXRSand legacy 2.4Ghz wireless sensors nodes
and gateways.

Node Commander Software User Manual

11.5

Support

Technical Support

There are many resources for product support found on the LORD MicroStrain website
including technical notes, FAQs, and product manuals.
http://www.microstrain.com/support_overview.aspx

For further assistance our technical support engineers are available to help with technical and
applications questions.

Technical Support
sensing_support@LORD.com
Phone: 802-862-6629
Toll Free: 800-449-3878
Fax: 802-863-4093
SKYPE: microstrain.wireless.support
Live Chat is available from the website during business hours:
9:00AM to 5:00PM (Eastern Time US & Canada)

114

Node Commander Software User Manual

11.6

Warranty Information
Warranty
LORD MicroStrain warrants its products to be free from defective material
and workmanship for a period ofone (1) yearfrom the original date of
purchase. LORD MicroStrain agrees to repair or replace, at its sole
discretion, a defective product if returned to LORD MicroStrain within the
warranty period and accompanied by proof of purchase. This warranty does
not extend to any LORD MicroStrain products which have been subject to
misuse, alteration, neglect, accident, incorrect wiring, mis-programming or to
use in violation of operating instructions furnished by us, nor extend to any
units altered or repaired for warranty defect by anyone other than LORD
MicroStrain . This warranty does not cover any incidental or consequential
damages and is in lieu of all other warranties expressed or implied and no
representative or person is authorized to assume for us any other liability in
connection with the sale of our products. Some states do not allow limitations
on how long an implied warranty lasts, and/or the exclusion or limitation of
incidental or consequential damages so the above limitations and exclusions
may not apply to the original customer.
Terms and Conditions of Sale
Please refer to the LORD MicroStrain website Support page for
completeTerms and Conditions of product sales.
Terms and Conditions of Service
Please refer to the LORD MicroStrain website Support page for
completeTerms and Conditions of product service.
Trial System
To enable customers to try our products risk free, LORD MicroStrain offers a
30-day return policy on the purchase of a starter kit. In order to take advantage

115

Support

Node Commander Software User Manual

Support

of this offer, a purchase order or payment for the starter kit is required when
the order is placed. If the product is not suited to the application, the product
may be returned within 30 days from the date of receipt for a full refund
(excluding shipping and handling), as long as the product is unaltered and
undamaged. Items can only be returned after LORD MicroStrain has issued
an Return Material Authorization (RMA). Items must be packed to withstand
shipping and returned freight pre-paid. LORD MicroStrain will inspect the
items returned and issue a refund or credit once the items have been
examined and are deemed to be unaltered or undamaged.Non-standard or
custom products may only be returned with approval and a re-stocking penalty
may be assessed. A 30 Day Return must be initiated by receiving a
RMAfromLORD MicroStrain.

11.7

Sales Support

Products can be ordered directly from the LORD MicroStrain website by navigating to the
product page and using the Buy feature.
http://www.microstrain.com/wireless
For further assistance, our sales team is availalbe to help with product selection, ordering
options, and questions.

Sales Support
sensing_sales@LORD.com
Phone: 802-862-6629
Toll Free: 800-449-3878
Fax: 802-863-4093
9:00AM to 5:00PM (Eastern Time US & Canada)

116

Node Commander Software User Manual

12.

References

References

12.1

Related Documents

Many references are available on the LORD MicroStrain website including product user
manuals, technical notes, and quick start guides. These documents are continuously updated
and new applications are added. They may provide more accurate information than printed or
file copies.

Document

Where to find it

Product Datasheets
Product Manuals and Technical Notes
Product Application Notes
NIST Calibration Procedures

http://www.microstrain.com/wireless/sensors
http://www.microstrain.com/support/docs
http://www.microstrain.com/applications
http://www.nist.gov/calibrations/
http://www.astm.org/Standard/standardsand-publications.html

ASTM Testing Procedures

Table 9 - Related Documents

117

Node Commander Software User Manual

12.2

References

Glossary

These terms are in common use throughout the manual:


A/D Value: the digital representation of the analog voltages. Analog to digital (A/D)
conversion. The accuracy of the conversion is dependent on the resolution of the system
electronics. Higher resolution produces a more accurate conversion.
ASTM: Association of Standards and Testing. A nationally accepted organization for testing
and calibration of technological devices.
Base Station: The base station is the transceiver that attaches to the host computer and
provides communication between the software and the node(s). It is also referred to as a
gateway.
Bits: the digital equivalent of voltage on the node.
Calibration: to standardize a measurement by determining the deviation standard and
applying a correction, or calibration, factor.
Configuration: a general term applied to the node indicating how it is set up for data
acquisition. It includes settings such as sampling mode and rate, number of active channels,
channel measurement settings, offsets, hardware gain, and calibration values.
Coordinated Universal Time (UTC): the primary time standard for world clocks and time. It
is similar to Greenwich Mean Time (GMT).
Cycle Power: a command transmitted to the node to re- boot the node, either through a
hardware or software switch.
Data Acquisition: the process of collecting data from sensors and other devices.
Data Logging: the process of saving acquired data to the system memory, either locally on
the node, or remotely on the host computer.
DHCP (network): Dynamic Host Configuration Protocol is the standardized networking
protocol used on Internet Protocol (IP) networks, which automatically configures devices that
are attached to it by assigning and configuring the device IP address.
Differential (signal): is a method of transmitting electrical signals in which they are paired
together as a differential pair and measured with reference to each other only. This method
makes the pair less susceptible to electrical noise.

118

Node Commander Software User Manual

References

EMI: Electromagnetic Interference is inductive or radiated disturbance that can create signal
degradation on electrical signals, including loss of data.
Firmware: the code that is programmed onto a microcontroller or similar device in an
embedded system that includes device operation commands, conditions, memory allocation
and many other tasks.
Gateway: The gateway is a transceiver that attaches to the host computer and provides
communication between the software and the node(s). It is also known as a base station.
Host (computer): The host computer is the computer that orchestrates command and control
of attached devices or networks.
LED: Light Emitting Diode is an indicator light that is used in electronic equipment.
LOS: Line of Sight is an acronym used in radio communications that describes the ideal
condition between transmitting and receiving antennas in a radio network. As stated, it means
they are in view of each other with no obstructions.
LXRS : Lossless Extended Range Synchronized is the LORD MicroStrain data
communications protocol used in the wireless sensor network.
NIST: National Institute of Standards and Testing is a nationally accepted organization for
testing and calibration of technological devices.
Node: The node is the wireless transceiver that the sensor (s) are connected to, providing
communication with the gateway. The G-Link -LXRS , V-Link -LXRS , and SG-Link LXRS are all nodes made by LORD MicroStrain.
Node Tester board: The Node Tester board is a device designed by LORD MicroStrain that
can be plugged into nodes to test functionality.
Offset: When describing a mathematically linear relationship, the offset is the value where the
line that represents the relationship in a graph crosses the y-axis. The equation of a straight
line is: y = mx+b, where x is the x-axis coordinate, y is the y-axis coordinate, m is the slope and
b is the offset.
Ping: A byte is transmitted by the gateway to the node and the node responds by echoing the
byte, indicating communication exists between them.
PGA: A Programmable Gain Amplifier is an electronic device used to amplify small electrical
signals.

119

Node Commander Software User Manual

References

Range Test: a continuous string of pings used to validate communication between the
gateway and the node over distance and obstruction
Read/Write EEPROM: commands transmitted to the node to read or write parameters stored
in the nodes operating system
Real Time Clock (RTC): a computer clock that keeps track of the current time.
Resolution:in digital systems, the resolution is the number of bits or values available to
represent analog values or information. For example, a 12- bit system has 4096 bits of
resolution and a 16-bit system has 65536 bits.
RFI: Radio Frequency Interference
RSSI: Received Signal Strength Indication is a measurement of the transmission power in a
radio signal. The units are decibels per meter (dBm).
RS232: a serial data communications protocol
Sensor: a device that physically or chemically reacts to environmental forces and conditions
and produces a predictable electrical signal as a result
Sleep: a command transmitted to the node to set a node into sleep configuration
Sampling: the process of taking measurements from a sensor or device
Sampling Mode: the type of sampling that is being utilized, such as event triggered,
continuous, or periodic. The nodes have several sampling modes that employ these types of
sampling
Sampling Rate: the frequency of sampling
Single Ended: electrical signals that are measured with reference to a system ground
Slope: When describing a mathematically linear relationship, the slope is the steepness of the
line that represents that relationship on a graph. The equation of a straight line is: y = mx+b,
where x is the x axis coordinate, y is the y axis coordinate, m is the slope and b is the offset.

120

Node Commander Software User Manual

References

Streaming: Streaming is a sampling mode in which all active channels (and the sensors
attached to them) are measured, and the data acquired is transmitted to the gateway and
software. The data is not written to non- volatile memory during streaming. Streaming can
either be finite (have a user defined start and end time) or continuous (continue until the power
is cycled on the node).
USB: Universal Serial Bus, a serial data communications protocol
Wheatstone Bridge: an electrical circuit used to measure unknown electrical resistances
WSN: Wireless Sensor Network

121

You might also like