Professional Documents
Culture Documents
Belec CompPort GB 042010
Belec CompPort GB 042010
Optics
double spectrometer in Paschen-Runge mounting
Rowland circle diameter 300 mm
usable wavelengths 190-410 nm, 220-430 nm
groove 3600 lines/mm
reciprocal dispersion 0,9 nm/mm (1st order)
shock resistance
temperature stabilised photoelectric cells
Probes
argon-flushed sparking probe for exact analysis,
including carbon
argon flow 0,1 l / min in standby and 2,5 l / min
when analysing
argon control on cable socket
air probe for quick mix-up checking
adapters for wires, pipes and small parts available
for all probes
lightweight shockproof plastic probe housing
start and clear buttons easily hand-operated
signal on mix-up identification: visual display for
repeat and reject, start button is blocked until
confirm button is pressed
multi fibre quartz optics, standard lengths 3m or 5m
low-wear tungsten electrode
silver electrode for air probe
probe connector system
Source
sparking generator with maximum 400 Hz frequency
unipolar discharge
separate parameter for pre-sparking and integration
selectable via software
ignition frequency program specifically selectable
via software
discharge power program specifically selectable via
software
arc source for air probe, optional
ignition voltage 20 kV
Electronics
stabilised high voltage
zero-stabilised analogue amplifier
6-decade dynamic A-D converter for each channel
36 channels with digital integration, configurable for
several bases
Dimensions
width 14.2 in. (360 mm)
height 7.9 in. (200 mm)
depth 16.5 in. (420 mm)
Weight
analysis unit 37.5 lbs. (17,00 kg)
special probe 2.9 lbs. ( 1,30 kg)
argon probe 2.1 lbs. ( 0,95 kg)
air probe
1.8 lbs. ( 0,80 kg)
measuring stand 20.0 lbs. ( 9,50 kg)
Power Supply
230V/50Hz or 110V/60Hz
100W in stand-by mode
600W during analysis
Computer Hardware
system-integrated industrial computer system
Intel Celeron M 1.0 GHz with real time clock
1 GB RAM minimum
2.5 hard disk 60 GB minimum
10.4 TFT colour display
touch screen
special dust and moisture protected keyboard
integrated in hinged lid
USB 2.0 ports
Ethernet interface RJ45*
Specifications
easy handling
mobility by trolley
easy transportation
user friendly Software Belec Win 21
analysis of P and S with Special Sparking Probe
quick mix-up checking with Air Sparking Probe
high accuracy grade (C < 0,01%)
connectable measuring stand
adapters for the most different sample shapes
installation and instruction by qualified personnel
Unique Option:
In addition to other elements, the carbon content can be analyzed less than
0,01%. The Argon Special Probe allows the analysis of short wavelengths of
Phosphor and Sulphur by a small spectrometer which is integrated in the probe.
The air probe comes into operation for quick mix-up checking and achieves
inspection cycles of less than three seconds.