Application Note 97 January 2006
Application Note 97 January 2006
Application Note 97 January 2006
January 2006
Accurate Measurement of LT5514 Third Order
Intermodulation Products
Dorin Seremeta
INTRODUCTION
Accurate measurement of the third order intercept point
for low distortion IC products such as the LT5514 requires
certain precautions to be observed in the test setup and
testing procedure. The LT5514 linearity performance is
high enough to push the test equipment and test set-up to
their limits. A method for accurate measurement of third
order intermodulation products, IM3, with standard test
equipment is outlined below.
It is also important to correctly interpret the LT5514
specification with respect to ROUT, and the impact of
demo-board transmission-line termination loss when
evaluating the linearity performance, as explained in the
LT5514 Datasheet and in Note 1 of this document.
EVALUATION OF LT5514 LINEARITY PERFORMANCE.
The difficulty in providing an adequate low distortion, twotone test signal (Note 2) and failure to adjust the spectrum
analyzer for maximum resolution (Note 3) are the most
common issues when evaluating the LT5514. Below is a
description of a test set-up and a method for evaluating
high linearity components using standard test equipment.
This method makes it easy to identify which section of the
set-up limits the measurement, and to optimize the set-up
SIGNAL
GENERATOR #1
10dB PAD
SPECTRUM
ANALYZER
2-WAY
COMBINER
AN97 F01
SIGNAL
GENERATOR #2
10dB PAD
Figure 1. Basic Set-Up for Two-Tone Signal Source IM3 Performance Verification Measurements
an97f
AN97-1
Application Note 97
spectrum analyzer should be inspected for:
a) Intermodulation products. Adjust the spectrum analyzer to obtain the maximum possible resolution
(intermodulation product amplitude in dBc relative to
tone power).
b) Second and third harmonic content (which will determine if a lowpass filter is required at the combined
signal generator output).
If the observed intermodulation products from the test
setup are of comparable size to those expected from the
LT5514, then an accurate characterization of the LT5514
is not possible. Likewise, if the observed harmonic content
is significant, then it must be reduced in order to avoid
generating spurious responses during the LT5514 characterization. Depending on the available test equipment,
an adjustment of the spectrum analyzer settings and the
signal combiner may be required in order to obtain satisfactory results. The intermodulation product level may be
limited by the two-tone signal source or by the spectrum
analyzer resolution. The insertion of a small attenuation
pad (as described in Note 4) can be used to find which
section of the setup limits the intermodulation product
level. No results better than this initial measurement can
be expected from the LT5514 demo board.
2. Evaluate the LT5514 Demo Board Performance
Remove matching resistor R7 (255) from the LT5514
demo board when no transmission lines are expected in
the LT5514 application (Note 1)
Program the LT5514 for maximum gain
Measure the LT5514 demo board gain
SIGNAL
GENERATOR #1
10dB PAD
SPECTRUM
ANALYZER
LT5514
DEMO-BOARD
2-WAY
COMBINER
ATTEN
DUT
AN97 F02
SIGNAL
GENERATOR #2
10dB PAD
AN97-2
Application Note 97
NOTE 1. The Demo-Board Output Linearity and the
LT5514 Device Linearity
VCC
C2
0.1F
IF
IN
ENA
1
2
3
4
5
6
7
8
9
10
T1
1:2
TC2-1T
C1
0.47F
ENA
ENB
VCC1
VCC2
GND
GND
LT5514
GND
GND
IN+
OUT
IN
OUT+
GND
GND
GND
GND
PGA0
PGA3
PGA1
PGA2
PGA0 PGA1
VOSUP
C4
0.1F
C3
4.7F
IF
OUT
T2
4:1
ROUT
100
R7
255
TC4-1W
RLOAD
50
J2
0
PGA2 PGA3
AN97 F03
J1
0
ENB
20
19
18
17
16
15
14
13
12
11
49
46
43
40
37
34
25
50
AN97-3
Application Note 97
transformed impedance gives an approximate ROUT =
100.
Thick lines (BOARD in legend) are the OIP3 figures for the
board output power as measured with a spectrum analyzer
for both conditions.
an97f
AN97-4
Application Note 97
increase to about 10dBm. In this case, methods (1) and (2)
fail to provide an adequately clean (low distortion) signal.
3. The best isolation between the two signal generators
can be achieved by buffering the individual signal generator outputs with high linearity amplifiers, prior to driving
a resistive combiner. This method was used to generate
the two-tone test signals for the LT5514 during datasheet
characterization.
A lowpass filter may be inserted between the combiner
output and the LT5514 board input in order to remove
harmonic content from the signal generator output. This
precaution is particularly recommended when signal generators with a high second harmonic output are used.
When the signal generators harmonic level is below
40dBc, the LT5514 intermodulation products can be
tested without this lowpass filter.
NOTE 3. Dynamic Range of the Spectrum Analyzer
Spectrum analyzers produce significant intermodulation
products if operated incorrectly.
As a rule, spectrum analyzers are designed to operate at
their best with about 30dBm to 40dBm at the analyzer
input. Normally the only device in front of the analyzer is
the instruments input attenuator. Therefore, with tone
levels of 10dBm, the (total) input attenuation should be set
to about 40dB to 50dB. As the spectrum analyzer attenuation level is increased, the instruments noise floor rises
and potentially hides any intermodulation products.
Attempting to display both the high side and low side
intermodulation products simultaneously is the most common problem in making use of the full resolution potential
of the spectrum analyzer. For example, a spectrum analyzer span setting of 4MHz is required to observe the
intermodulation products with 1MHz tone spacing. This
results in unnecessary elevation of the displayed noise
floor and reduced resolution.
The spectrum analyzer measurement span and resolution
bandwidth must be reduced in order to lower the displayed
noise floor, and to increase resolution. The distortion
products of the LT5514 should be measured by setting the
spectrum analyzer center frequency to that of each individual intermodulation product and tone frequency, one at
a time. The spectrum analyzer span is set to a minimum
value (100Hz) and resolution bandwidth is set to 10Hz. For
AN97-5
Application Note 97
spectrum analyzer filter. This method uses a second,
low distortion reference signal path that is placed in
parallel with the tested LT5514 device path. The outputs
of these two paths are subtracted, one from the other,
at the input to the spectrum analyzer. The net gain and
electrical delay of the two paths are tuned such that the
main tone level presented to the spectrum analyzer is
reduced by about 40dB typically, while the tested LT5514
output signal intermodulation products are not altered.
This method is most accurate, but requires a large
amount of hardware. The calibration procedure for this
method is also cumbersome, particularly when all gain
settings are to be tested.
As a quick check, if the intermodulation products displayed on the spectrum analyzer are real, and not artifacts
of the test setup, then their amplitude relative to the main
tone level will not be affected by varying the input attenuation on the analyzer (typically in 10dB steps). External
attenuator pads (3dB or 6dB) can be used for smaller level
variation for this test.
AN97-6
Application Note 97
For the previous example, the OIP3 at demo board output
when LT5514 is set at Max-Gain is:
OIP3 = 23.3 + 30.1 + (77.4)/2 = 45.5dBm
Note that the higher (worst) sidetone measurement is
used for this calculation.
The OIP3 measurement for the LT5514 Demo Board at
21dB Attenuation Step follows a similar sequence.
The set-up changes for step 1 only (source qualification,
Figure1) are:
Insert about 21dB pad at the combiner output
Increase by 21dB the signal generator output levels
Fill in the SOURCE column in Table 2
The set-up changes for step 2 only (Demo Board measurement, Figure 2) are:
Set LT5514 PGA0 high and PGA1 = PGA2 = PGA3
inputs low
Remove the 21dB pad from the combiner output in
step 1
PARAMETER
UNIT
Column Suffix:
SOURCE
(STEP1)
LT5514
(STEP2)
_src
_sa
99.7
Pwr_IMD3
dBm
116.5
101.3
99.9
Pwr_tone
dBm
23
23.3
100.1
Pwr_tone
dBm
23
23.3
100.3
Pwr_IMD3+
dBm
116.0
100.7
IMD3
dBc
93.5
78
IMD3+
dBc
93
77.4
UNIT
SOURCE
(STEP1)
LT5514
(STEP2)
IMD3LT5514 IMDSOURCE
DIFFERENCE
15.2dB
15.3dB
PARAMETER
_src
_sa
99.7
Column Suffix:
Pwr_IMD3
dBm
116.5
98.5
99.9
Pwr_tone
dBm
23
23.1
100.1
Pwr_tone
dBm
23
23.1
100.3
Pwr_IMD3+
dBm
114.5
97.8
IMD3
dBc
93.5
75.4
IMD3+
dBc
91.5
74.7
IMD3LT5514 IMD3SOURCE
DIFFERENCE
18dB
16.7dB
an97f
AN97-7
Application Note 97
12
40dB
20dB
10dB
6dB
3dB
10
8
6
4
2
0
6
8
4
10 12
IMD3 DIFFERENCE (dBc)
14
16
AN97 F05
an97f
AN97-8