EM300-50 (LectroCount Remote Display) PDF
EM300-50 (LectroCount Remote Display) PDF
Installation: EM100-13
Table of Contents
! WARNING
Model Numbers
Two LectroCount Remote Display Models are available
for interface to Liquid Controls LectroCount Electronic
Registers. Ensure that the model number of the
LectroCount Remote Display is compatible with the
LectroCount Electronic Register prior to installation. If
the device does not match the model number, the system
will not work and may be damaged when power is applied. Figure 1. Remote Display Identification Label
2
Specifications
The LectroCount Remote Display is a 6-digit, backlit display with 2" high digits, viewable to 100 ft, for flow metering
applications.
LectroCount Remote Display models are available for LectroCount LCR, LectroCount LCR-II and LectroCount
Electronic Registers.
Features include:
6-digit, backlit display with 2" high display characters.
30 ft, 4-wire, shielded cable.
Class I, Division 2 Group C & D weatherproof NEMA 4X enclosure.
Weights & Measures Approvals in USA and Canada when used with LectroCount Electronic Registers.
Viewable up to 100 ft.
Wiring: Connections from the LectroCount register to the LectroCount Remote Display should be made using 22
gage, 4-wire, shielded cable or larger. Thirty feet of cable is provided with the display. The maximum length of cable
allowable is 100 ft. The 4-wire cable includes conductors for power, ground, and two signals.
SPECIFICATIONS
Environmental Signal Input
Design for: Class I, Div. 2, Group C & D Input Signal Levels: Hi 2.50 VDC
NEMA 4X Lo 2.0 VDC
*At lower temperatures, the liquid crystal display may have a delayed
response rate.
3
Installation
! CAUTION
The LectroCount Remote Display and accessories (whether supplied by Liquid Controls or other) must be installed
and operated in accordance with all applicable federal, state, and local construction, electrical, environmental
and safety codes. Failure to do so could result in serious injury or death.
Installation Requirements
Read this manual prior to start of installation. If you The LectroCount Remote Display is supplied with a
have any questions, consult with your full service 30 foot cable, which should be adequate for most
distributor or call the Service Department at Liquid installations. If additional cable is required, cable
Controls. kits are available from Liquid Controls. It is
recommended that a 4-wire, shielded cable with 22
This display should always be securely mounted to gage wire or larger be used. The maximum length
a platform or supportive member. Ensure that the of cable allowable is 100 feet.
display does not experience excessive vibration or
shock. The LectroCount Remote Display is supplied ready for
final cable termination. Should it be necessary to replace
The LectroCount Remote Display is designed to be the cable between the signal source and the remote
compatible with a specific LectroCount Electronic display, be sure to secure the cable and tighten the cover
Register as specified by the customer and as so as to maintain the vapor seal.
indicated on the Serial Number Tag. No attempt
should be made to use the display with an input
different from the one originally specified.
4
Installation
Mounting Dimensions
The LectroCount Remote Display should be mounted The diagram below shows the overall dimensions and
using the four holes provided in the rear cover. The hole the locations of the mounting holes on the LectroCount
size is designed for use with " screws. It is important Remote Display.
to select a mounting location which will position the
display as close to the optimal viewing angle as possible
(see Page 6).
5
Installation
Optimum Viewing Angle
LCD displays have an optimum viewing angle. Displays to accommodate as wide a viewing area as possible.
lose contrast and become more difficult to read at some Optimum viewing with the greatest contrast will occur
viewing angles and are easier to read at other angles. when the display is viewed between 20 above and 40
Because the viewing angle is limited, a bias is designed below horizontal. If the display is viewed outside this
into the module at the time it is manufactured. The result range, the display contrast will reduce to a point where it
is a nominal viewing angle which is offset from the reaches an unacceptable viewing level. The horizontal,
perpendicular by a specified amount. The Liquid Controls side to side viewing angle is 40 left and right from center.
LectroCount Remote Display is designed with a bias set
6
Installation
Remote Display Shield Accessory (PN 81879)
There are certain applications where sun or bright light
may produce glare on the display. Use the optional shield
to reduce the effect of glare. This shield fits in place over
the LectroCount Remote Display as shown in Figure 6.
The holes in the shield are slotted for ease of installation.
7
Installation
Decimal Point Jumper Setting
NOTE: Model E1610 automatically displays the identical Once the enclosure is opened, the jumpers may be moved
decimal configuration appearing on the LectroCount to configure the display according to the LectroCount
LCR-II display. There is no need to adjust the jumpers Electronic Register for which it is designed. See the table
on Model E1610. below for jumper settings.
The LectroCount Remote Display may be configured to NOTE: The jumpers need to be in position. Do not remove
show the decimal point in one of three positions: and discard them. If Whole Units is desired, then J3 and
Whole Units (No Decimal Point) J4 must BOTH have jumpers in the OFF position.
Tenth Units
Hundredth Units NOTE: If J3 and J4 are BOTH in the ON position, then 2
decimal points will appear on the display. This is not a
This is accomplished by moving the J3 and J4 jumpers valid jumper configuration.
on the Remote Display PCB. The PCB is accessed by
removing the 14 screws on the back of the enclosure.
The jumpers are located on the bottom left side of the
PCB, to the left of the J2 connector (Figure 9).
J3 J4 J2
Figure 9. Jumper Locations
LectroCount LCR,
E1611 OFF OFF ON OFF OFF ON
LectroCount
8
Wiring Diagrams
Model No. E1610
For use with LectroCount LCR-II
Once the display is mounted, the factory installed
cable should be routed to the LectroCount LCR-
II. Choose an unused port on the LCR-II and
connect the cord grip provided. Route the cable
through the cord grip and connect the cable to
the J12 Connector on the LCR-II PCB as shown
in Figure 10.
J12 J1
Pin 39 Pin 21 (WHITE)
Pin 40 Pin 22 (GREEN)
Pin 41 Pin 23 (BLACK)
Pin 45* Pin 20 (RED)
*If the power supplied to the LCR-II ever exceeds Figure 10. LCR-II Wiring Diagram
+24VDC, then the remote display should be powered
from Pin 32 on J8, +5VDC.
RED to Pin 35 When a delivery is active, pressing the Select Switch will
BLACK to Pin 38. cause the display to show the current flow rate. After 5
seconds, the display returns to current flow total.
9
Wiring Diagrams
Model No. E1611
For use with LectroCount LCR
Once the display is mounted, the cable should
be routed to the LCR. Choose an unused port
on the LCR and connect the cord grip provided.
Route the cable through the cord grip and
connect the cable to the J12 Connector on the
LCR PCB as shown in Figure 12.
J12 J1
Pin 39 Pin 21 (WHITE)
Pin 40 Pin 22 (GREEN)
Pin 41 Pin 23 (BLACK)
Pin 45* Pin 20 (RED)
When functional, the remote display will match the LC Figure 13. LC Wiring Diagram
display. No external reset switch is required for operation.
10
Dual Display Kit
Dual Display Kit (PN 82594)
(For use with Models E1610 & E1611) 5. Route the cable from the unopened Remote Display
The Dual Display Kit is designed for use with two (2) through a cord grip attached to the junction box.
LectroCount Remote Displays and a single LectroCount
Electronic Register. This kit contains: 6. Route the removed cable through a second cord grip
4-Port Junction Box in the junction box.
Cord Grip (2)
Nipple & Seal Washer 7. Connect both cables to J1 on the PCB.
3. Remove the cord grip and cable from the unit. 9. Connect the loose end of the cable to the LectroCount
Electronic Register as shown on Pages 9-10.
4. Install the 4-Port Junction Box in the space where
the cord grip was just removed.
Figure 14. Dual Display Kit - 2 Remote Displays with 1 LectroCount Electronic Register
11
Illustrated Parts Breakdown