Let us see some of the types of column chromatography. But before knowing that let us check
it out definition of chromatography. As we all know that chromatography is a technique applied
for the separation of the components of mixture by continuous distribution of the components
between two phases, of which one is mobile phase and another one is stationary phase. There
are four different types of column chromatography which works on four different mechanisms.
1. Adsorption chromatography in which the components of mixture are selectively adsorbed on the
surface of packing column material i.e. adsorbant.
2. Partition chromatography in which component is partitioned between the mobile phase and their
stationary phase held stationary on inert solid support.
3. Ion exchange chromatography in which the constituent of sample is selectively retained by
exchange resin by replacing ion/s on packing material.
4. Gel chromatography is the method in which column is packed with a permeable gel which brings
out separation by seiving or molecular filtration action.
These are some of the main types of column chromatography which are widely used.
There are other types of chromatography such as High performance liquid chromatography , Gas
chromatography and much more please check other articles on this blog.
It is typical for normal-phase chromatography on silica that the mobile phase is 100%
organic; no water is used. Reversed-Phase HPLC. The term reversed-phase describes
the chromatography mode that is just the opposite of normal phase, namely the use of a
polar mobile phase and a non-polar [hydrophobic] stationary phase ...
Answer: d
Explanation: Gas-solid chromatography is used for the separation of low
molecular weight gaseous species. Its application is limited because of semi-
permanent retention of the analyte.
Answer: c
Explanation: It should be highly pure. Further, it should be readily available and
non-inflammable.
Answer: a
Explanation: Hydrogen is dangerous to use. It has better thermal conductivity and
lower density.
Answer: b
Explanation: Helium is expensive. Its advantages are that it has low density and it
allows greater flow rates.
Answer: c
Explanation: Nitrogen has reduced sensitivity. It is still one of the commonly used
carrier gas in gas chromatography.
6. Slow injection of large samples leads to band broadening and loss of resolution.
a) True
b) False
View Answer
Answer: Slow injection of large samples leads to band broadening and loss of
resolution. Hence, for desired column efficiency, samples should not be too large.
7. In which of the following methods are liquid samples injected into the column in
gas chromatography?
a) Gas tight syringe
b) Micro-syringe
c) Rotary sample valve
d) Solid injection syringes
View Answer
Answer: b
Explanation: Liquid samples injected into the column in gas chromatography using
micro-syringe. Syringes of various capacities are available.
8. What must be done to the solid samples for it to be introduced into the column
without using solid injection syringes in gas chromatography?
a) Introduced in hot-zone of the column
b) Dissolved in volatile liquids
c) Introduced using rotary sample valve
d) Introduced using sampling loops
View Answer
Answer: b
Explanation: Solid samples must be dissolved in volatile liquids for introducing it
into the column. They can be introduced directly using solid injection syringes.
9. Which of the following is the commonly used support material for the packed
column in gas chromatography?
a) Glass
b) Metal
c) Diatomaceous earth
d) Stainless steel
View Answer
Answer: c
Explanation: Diatomaceous earth is the commonly used support material for the
packed column in gas chromatography. The columns could be made of glass or
metal.
Answer: c
Explanation: The advantage of straight column is that it can be repacked easily. It
is not compact in size.
11. Which of the following is the disadvantage of coiled or helical shaped packed
chromatographic column?
a) It cannot be packed uniformly
b) It cannot be repacked easily
c) It is not compact
d) It is not easy to heat it evenly
View Answer
Answer: b
Explanation: The disadvantage of coiled or helical shaped packed
chromatographic column is that it cannot be repacked easily. It is compact in size
and can easily be heated in an even manner.
12. Capillary columns are open tubular columns constructed from which of the
following materials?
a) Glass
b) Metal
c) Stainless steel
d) Fused silica
View Answer
Answer: d
Explanation: Capillary columns are constructed using fused silica. It is a very high
purity glass.
Answer: a
Explanation: Sample injection port must be maintained at a temperature at which
rapid vapourisation occurs but thermal degradation does not occur. The column is
maintained at a different temperature.
14. Which of the following is not a desirable feature of the ovens used in gas
chromatography?
a) It must have a fast rate of heating
b) Power consumption should be kept low
c) It must have maximum thermal gradients
d) It should have proper insulation
View Answer
Answer: c
Explanation: The ovens used in gas chromatography must have maximum thermal
gradients. The temperature must be uniform over the whole column.
15. Given below is the block diagram of gas chromatography. Identify the
unmarked component.
a) Pumping system
b) Pressure regulator
c) Flow regulator
d) Sample injection system
View Answer
Answer: d
Explanation: The unmarked component is syringe. Hence, the answer is sample
injection system. It is for the introduction of sample into the flowing gas stream.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Operation of Chromatography.
1. Which of the following steps takes place after injection of feed in Column chromatography?
a) Detection of components
b) Separation in the column
c) Elution from the column
d) Collection of eluted component
View Answer
Answer: b
Explanation: The operation that occurs after the injection of feed is separation in the column.
After that, elution from the column and detection of components takes place.
2. Components with strong attraction to the support move more slowly than components with
weak attraction.
a) True
b) False
View Answer
Answer: a
Explanation: Different components will adsorb to the stationary phase in varying degrees.
Components with strong attraction to the support move more slowly than components with weak
attraction.
3. What happens during the elution from the column phase in chromatography?
a) Components with greatest affinity elute first
b) Components with least affinity elute first
c) Components elute in a random manner
d) Components elute according to their concentration in the mixture
View Answer
Answer: b
Explanation: During the elution phase, different components elute at different times. Components
with least affinity elute first.
4. In chromatogram, the position of peaks on the time axis can be used to determine which of the
following?
a) Components of the sample
b) Amount of component in the sample
c) Column efficiency
d) Column resolution
View Answer
Answer: a
Explanation: Chromatogram is a detector that responds to concentration solute and is placed at
the end of the column. The position of peaks on the time axis can be used to determine
components of the sample.
5. In chromatogram, the area under the peak can be used to determine which of the following?
a) Components of the sample
b) Amount of component in the sample
c) Column efficiency
d) Column resolution
View Answer
Answer: b
Explanation: Chromatogram is a detector that responds to concentration solute and is placed at
the end of the column. The area under the peak can be used to determine amount of component
in the sample.
6. The stationary phase could be a viscous liquid coated over a surface of solid particles.
a) True
b) False
View Answer
Answer: a
Explanation: The stationary phase could be a viscous liquid coated over a surface of solid
particles. The solid particles can also be the stationary phase.
9. In older analytical methods, which of the following methods were used to allow movement of
mobile phase?
a) Pumps
b) Pressure
c) Gravity
d) Blowing air into the column
View Answer
Answer: c
Explanation: In older analytical methods, gravity facilitated the movement of mobile phase. The
effect of capillary action was also employed to allow movement.
10. Given below is the diagram of Process of chromatographic column. Identify the unmarked
component in the process.
a) Reservoir
b) Collection tank
c) Microprocessor
d) Detector
View Answer
Answer: d
Explanation: The unmarked component is detector. It is placed at the bottom of the column. It
responds to solute concentration.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Chromatographic Behaviour of Solutes.
1. How is molar concentration of solute in stationary phase related to molar concentration of
solute in mobile phase?
a) Directly proportional
b) Inversely proportional
c) Equal
d) Not related
View Answer
Answer: a
Explanation: In chromatography, molar concentration of solute in stationary phase is directly
proportional to molar concentration of solute in mobile phase.
Cs Cm
Cs = k Cm
k is the distribution constant.
2. If the value of the distribution constant k is one, then what could be inferred about the
distribution of solute?
a) Its distribution in stationary phase is greater
b) Its distribution in mobile phase is greater
c) It is equally distributed in stationary and mobile phase
d) It is distributed in a random manner
View Answer
Answer: c
Explanation:
Cs Cm
Cs = k Cm
If k=1, it denotes that the solute is equally distributed in mobile and stationary phase.
3. The time taken by the analyte after sample injection to reach the detector is called _________
a) Dead time
b) Solute migration rate
c) Adjusted retention time
d) Retention time
View Answer
Answer: d
Explanation: The time taken by the analyte after sample injection to reach the detector is called
retention time. The retention of a solute in the system can be used to identify the solute.
4. The time required for a molecule of the mobile phase to pass through the column is called
___________
a) Dead time
b) Solute migration rate
c) Adjusted retention time
d) Retention time
View Answer
Answer: a
Explanation: The time required for a molecule of the mobile phase to pass through the column is
called dead time. The effectiveness of the system depends on dead time.
5. Adjusted retention time is the remaining retention time after subtracting __________ from
___________
a) Solute migration rate and retention time
b) Retention time and solute migration rate
c) Dead time and retention time
d) Retention time and dead time
View Answer
Answer: c
Explanation: Adjusted retention time is the remaining retention time after subtracting dead time
from retention time. It affects the effectiveness of the system.
6. Which of the following is the volume of mobile phase required to make a solute band move
from the point of injection through the column to the detector?
a) Dead volume
b) Retention volume
c) Void volume
d) Adjusted retention volume
View Answer
Answer: b
Explanation: Retention volume is the volume of mobile phase required to make a solute band
move from the point of injection through the column to the detector.
7. Adjusted retention volume is the remaining retention volume after subtracting ___________
from _____________
a) Solute migration rate and retention volume
b) Retention volume and solute migration rate
c) Dead volume and retention volume
d) Retention volume and dead volume
View Answer
Answer: c
Explanation: Adjusted retention volume is the remaining retention volume after subtracting dead
volume from retention volume.
Vadjusted retention = Vretention Vdead
8. Which of the following is defined as the ratio of moles of solute in stationary phase to the
moles of solute in mobile phase?
a) Distribution constant
b) Volumetric phase ratio
c) Retention factor
d) Total porosity
View Answer
Answer: c
Explanation: Retention factor is defined as the ratio of moles of solute in stationary phase to the
moles of solute in mobile phase. It is used for determining the migration rates of solutes in the
column.
9. Which of the following is the ratio of interstitial volume of packing to the volume of its total
mass?
a) Distribution constant
b) Volumetric phase ratio
c) Retention factor
d) Total porosity
View Answer
Answer: d
Explanation: Total porosity is the ratio of interstitial volume of packing to the volume of its total
mass. In capillary column, total porosity is 1.
10. Which of the following is the ratio of length of column packing to dead time?
a) Average linear rate of solute migration
b) Average linear rate of mobile migration
c) Relative migration rate
d) Selectivity factor
View Answer
Answer: b
Explanation: Average linear rate of mobile migration is the ratio of length of column packing to
dead time. It influences the effectiveness of the column in separating solutes.
11. Which of the following is the ratio of length of column packing to retention time?
a) Average linear rate of solute migration
b) Average linear rate of mobile migration
c) Relative migration rate
d) Selectivity factor
View Answer
Answer: a
Explanation: Average linear rate of solute migration is the ratio of length of column packing to
retention time. It influences the effectiveness of the column in separating solutes.
12. Retention distance is the distance between point of injection and minimum peak in the
recorder or computer generated chart.
a) True
b) False
View Answer
Answer: b
Explanation: Retention distance is the distance between point of injection and maximum peak in
the recorder. The chart is drawn between time and concentration of the solute.
13. Retention volume can be obtained by finding the product of which of the following
parameters?
a) Dead time and total porosity
b) Retention time and volumetric flow rate
c) Adjusted retention time and volumetric flow rate
d) Retention time and total porosity.
View Answer
Answer: b
Explanation: Retention volume can be obtained by finding the product of retention time and
volumetric flow rate.
Flow rate = Cross section of empty column Average linear velocity Total porosity
14. Retention factor is also known as capacitance factor.
a) True
b) False
View Answer
Answer: a
Explanation: Retention factor is also known as capacitance factor. It can be denoted as k.
15. What must be the value of selectivity factor?
a) Equal to 1
b) Less than 1
c) Greater than 1
d) Greater than 0
View Answer
Answer: Selectivity factor/ Relative retention must always be greater than 1. It is also known as
Relative migration rate.
This set of Analytical Instrumentation Questions and Answers for Freshers focuses on Column
Efficiency and Column Resolution.
1. Which of the following is the distance that the solute moves while undergoing one partition?
a) Retention distance
b) Distribution constant
c) Plate height
d) Column packing length
View Answer
Answer: c
Explanation: Plate height is the distance that the solute moves while undergoing one partition. It
is expressed in length.
2. Which of the following conditions will cause the efficiency of the column to increase?
a) Plate number becomes greater, plate height becomes smaller
b) Plate number becomes smaller, plate height becomes smaller
c) Plate number becomes greater, plate height becomes larger
d) Plate number becomes greater, plate height becomes larger
View Answer
Answer: a
Explanation: The efficiency of the column increases when the plate number becomes greater and
the plate height becomes smaller. Column efficiency greatly depends on plate height and plate
number.
3. Which of the following is the expression for Eddy diffusion in the column, if represents
obstruction factor, dp represents particle diameter, DM represents Solute diffusion co-efficient
and represents function for packing uniformity?
a) dM
b) 2 DM
c) DM
d) /dM
View Answer
Answer: a
Explanation: Eddy diffusion is given as, A = dp. It is also called multipath term.
4. Which of the following is the expression for longitudinal diffusion in the column, if represents
obstruction factor, dM represents particle diameter, DM represents Solute diffusion co-efficient
and represents function for packing uniformity?
a) dM
b) 2 DM
c) DM
d) /dM
View Answer
Answer: b
Explanation: Longitudinal diffusion is given as, B = 2 DM.It is also called axial diffusion.
5. In gas-liquid chromatography, when films are used in the interior of capillary column, then what
is the value of Eddy diffusion?
a) Greater than 1
b) Less than 1
c) Zero
d) Less than zero
View Answer
Answer: c
Explanation: In gas-liquid chromatography, when films are used in the interior of capillary column
the value of Eddy diffusion (A) is zero. To minimize A, diameter of packing should be small and
uniform.
6. For an unpacked coated capillary column, the obstruction factor takes which of the following
values?
a) 0
b) 0.6
c) 1
d) 1.6
View Answer
Answer: c
Explanation: For an unpacked coated capillary column, the value of obstruction factor is 1. For a
packed column, the value is 0.6.
7. Base line resolution is achieved when degree of resolution/separation is ______
a) 1
b) 0
c) 0.5
d) 1.5
View Answer
Answer: d
Explanation: Base line resolution is achieved when degree of resolution/separation is 1.5. The
baseline bandwidths of adjacent bands are almost constant.
8. Which of the following equations give the expression for plate number, N when tR is the
adjusted retention time and W b is the width at the base of the peak which is equal to 4 standard
deviations.
a) 16 tR2/W b
b) 4 tR2/W b
c) (4tR/W b)2
d) 4 (tR/W b)2
View Answer
Answer: c
Explanation: The expression for plate number, N is given as N = (4tR/W b)2. It is a dimensionless
quantity.
9. It is more difficult to measure the width at half peak height than the base width in the detector
output.
a) True
b) False
View Answer
Answer: b
Explanation: The width at half peak height is easier to measure than the base width. Hence, the
plate number is calculated at peak height.
10. The solute, while moving in the column is always in equilibrium with both the stationary phase
and the mobile phase.
a) True
b) False
View Answer
Answer: a
Explanation: The solute, while moving in the column is always in equilibrium with both the
stationary phase and the mobile phase. But, the two phases will never be in equilibrium with
each other.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Gas Chromatography.
1. For the separation of which of the following substances, Gas-solid chromatography is being
used?
a) Thermally stable organic components
b) Volatile organic components
c) Thermally stable inorganic components
d) Low molecular weight gaseous species
View Answer
Answer: d
Explanation: Gas-solid chromatography is used for the separation of low molecular weight
gaseous species. Its application is limited because of semi-permanent retention of the analyte.
2. Which of the following is not a feature of carrier gas used in gas chromatography?
a) It must be chemically inert
b) It should be suitable for the detector employed
c) It should not be completely pure
d) It should be cheap
View Answer
Answer: c
Explanation: It should be highly pure. Further, it should be readily available and non-inflammable.
3. Which of the following is the disadvantage of hydrogen, which can be used as carrier gas in
gas chromatography?
a) Dangerous to use
b) Expensive
c) Reduced sensitivity
d) High density
View Answer
Answer: a
Explanation: Hydrogen is dangerous to use. It has better thermal conductivity and lower density.
4. Which of the following is the disadvantage of helium, which can be used as carrier gas in gas
chromatography?
a) Dangerous to use
b) Expensive
c) Reduced sensitivity
d) High density
View Answer
Answer: b
Explanation: Helium is expensive. Its advantages are that it has low density and it allows greater
flow rates.
5. Which of the following is the disadvantage of nitrogen, which can be used as carrier gas in gas
chromatography?
a) Dangerous to use
b) Expensive
c) Reduced sensitivity
d) High density
View Answer
Answer: c
Explanation: Nitrogen has reduced sensitivity. It is still one of the commonly used carrier gas in
gas chromatography.
6. Slow injection of large samples leads to band broadening and loss of resolution.
a) True
b) False
View Answer
Answer: Slow injection of large samples leads to band broadening and loss of resolution. Hence,
for desired column efficiency, samples should not be too large.
7. In which of the following methods are liquid samples injected into the column in gas
chromatography?
a) Gas tight syringe
b) Micro-syringe
c) Rotary sample valve
d) Solid injection syringes
View Answer
Answer: b
Explanation: Liquid samples injected into the column in gas chromatography using micro-syringe.
Syringes of various capacities are available.
8. What must be done to the solid samples for it to be introduced into the column without using
solid injection syringes in gas chromatography?
a) Introduced in hot-zone of the column
b) Dissolved in volatile liquids
c) Introduced using rotary sample valve
d) Introduced using sampling loops
View Answer
Answer: b
Explanation: Solid samples must be dissolved in volatile liquids for introducing it into the column.
They can be introduced directly using solid injection syringes.
9. Which of the following is the commonly used support material for the packed column in gas
chromatography?
a) Glass
b) Metal
c) Diatomaceous earth
d) Stainless steel
View Answer
Answer: c
Explanation: Diatomaceous earth is the commonly used support material for the packed column
in gas chromatography. The columns could be made of glass or metal.
10. Which of the following is the advantage of straight packed column?
a) It can be packed uniformly
b) It can be repacked easily
c) It is compact
d) It is easier to heat it evenly
View Answer
Answer: c
Explanation: The advantage of straight column is that it can be repacked easily. It is not compact
in size.
11. Which of the following is the disadvantage of coiled or helical shaped packed
chromatographic column?
a) It cannot be packed uniformly
b) It cannot be repacked easily
c) It is not compact
d) It is not easy to heat it evenly
View Answer
Answer: b
Explanation: The disadvantage of coiled or helical shaped packed chromatographic column is
that it cannot be repacked easily. It is compact in size and can easily be heated in an even
manner.
12. Capillary columns are open tubular columns constructed from which of the following
materials?
a) Glass
b) Metal
c) Stainless steel
d) Fused silica
View Answer
Answer: d
Explanation: Capillary columns are constructed using fused silica. It is a very high purity glass.
13. Sample injection port must be maintained at a temperature at which rapid vapourisation
occurs but thermal degradation does not occur.
a) True
b) False
View Answer
Answer: a
Explanation: Sample injection port must be maintained at a temperature at which rapid
vapourisation occurs but thermal degradation does not occur. The column is maintained at a
different temperature.
14. Which of the following is not a desirable feature of the ovens used in gas chromatography?
a) It must have a fast rate of heating
b) Power consumption should be kept low
c) It must have maximum thermal gradients
d) It should have proper insulation
View Answer
Answer: c
Explanation: The ovens used in gas chromatography must have maximum thermal gradients.
The temperature must be uniform over the whole column.
15. Given below is the block diagram of gas chromatography. Identify the unmarked component.
a) Pumping system
b) Pressure regulator
c) Flow regulator
d) Sample injection system
View Answer
Answer: d
Explanation: The unmarked component is syringe. Hence, the answer is sample injection
system. It is for the introduction of sample into the flowing gas stream.
This set of Analytical Instrumentation Interview Questions and Answers for freshers focuses on
Detection System of Gas Chromatography.
1. Which of the following is not an ideal characteristic of a detector used in gas chromatography?
a) Linear response to the solutes
b) Short response time
c) High reliability
d) Sensitive to the changes in the flow rate of carrier gas
View Answer
Answer: d
Explanation: The detector used in gas chromatography must be insensitive to the changes in
flow rate of carrier gas. There are many detectors used in gas chromatography.
2. Which of the following is not a type of detector used in gas chromatography?
a) Argon ionisation detector
b) Thermal conductivity detector
c) UV visible spectrometric detector
d) Electron capture detector
View Answer
Answer: c
Explanation: UV visible spectrometric detector is not used in gas chromatography. It is used in
liquid chromatography.
3. Which of the following detectors have high sensitivity to all organic compounds?
a) Sulphur chemiluminescense detector
b) Thermionic emission detector
c) Flame ionization detector
d) Argon ionisation detector
View Answer
Answer: c
Explanation: Flame ionization detector has high sensitivity to all organic compounds. It is the
commonly used detector for gas chromatography.
4. Which of the following is not the advantage of thermal conductivity detector used in gas
chromatography?
a) Simple in construction
b) High sensitivity
c) Large linear dynamic range
d) Non-destructive character
View Answer
Answer: b
Explanation: Thermal conductivity detector has relatively low density when compared to other
detectors used in gas chromatography. It is based on the principle that all gases conduct heat in
varying degrees.
5. Which of the following detectors is widely used to detect environmental samples like
chlorinated pesticides and polychlorinated biphenyls?
a) Flame ionization detector
b) Thermal conductivity detector
c) Argon ionisation detector
d) Electron capture detector
View Answer
Answer: d
Explanation: Electron capture detector is used to detect environmental samples like
polychlorinated biphenyls and chlorinated pesticides. It is highly sensitive to molecules containing
functional groups such as halogen and phosphorous.
6. In which of the following detector is the eluent mixed with hydrogen and burned and then
mixed with ozone and its intensity is measured?
a) Sulphur chemiluminescense detector
b) Thermal conductivity detector
c) Flame ionization detector
d) Electron capture detector
View Answer
Answer: a
Explanation: In Sulphur chemiluminescense detector, the eluent is mixed with hydrogen and
burned and then mixed with ozone and its intensity is measured. The resultant is a measure of
sulphur compounds present.
7. Filter photometer detector is primarily responsive to which of the following
compounds/elements?
a) Volatile sulphur or phosphorous compounds
b) Nitrogen
c) Halogen
d) Potassium
View Answer
Answer: a
Explanation: Flame photometric detector is primarily responsive to volatile sulphur or
phosphorous compounds. It is also responsive to tin and nitrogen.
8. Which of the following detector uses ultraviolet radiation from lamps to produce ionisation of
solute molecules?
a) Sulphur chemiluminescense detector
b) Thermal conductivity detector
c) Photo ionization detector
d) Electron capture detector
View Answer
Answer: c
Explanation: Photo ionization detector uses ultraviolet radiation from lamps to produce ionisation
of solute molecules. The current produced is measured and recorded.
9. Flame ionisation detector is also known as Katharometer.
a) True
b) False
View Answer
Answer: b
Explanation: Thermal conductivity detector is known as Katharometer. It uses heated filament as
sensing element and it is placed in the emerging gas stream.
10. Thermionic emission detector used in gas chromatography is most sensitive to which of the
following elements?
a) Nitrogen
b) Phosphorous
c) Halogen
d) Carbon
View Answer
Answer: b
Explanation: Thermionic emission detector used in gas chromatography is most sensitive to
phosphorous. It is 500 times more sensitive to phosphorous than Flame ionization detector.
11. Which of the following detectors has a non-volatile bead of rubidium silicate placed above the
flame tip?
a) Argon ionisation detector
b) Thermionic emission detector
c) Flame ionization detector
d) Electron capture detector
View Answer
Answer: b
Explanation: Thermionic emission detector has a non-volatile bead of rubidium silicate placed
above the flame tip. It is maintained at about 180V with respect to the collector.
12. In which of the following detectors, the carrier gas is excited by a radioactive source and the
atoms of carrier gas are excited to metastable state?
a) Argon ionisation detector
b) Thermionic emission detector
c) Flame ionization detector
d) Electron capture detector
View Answer
Answer: a
Explanation: In Argon ionisation detector, the carrier gas is excited by a radioactive source and
the atoms of carrier gas are excited to metastable state. It uses argon as carrier gas.
13. Which of the following is not used as a heating element in Thermal conductivity detector?
a) Platinum
b) Gold
c) Graphite
d) Tungsten wire
View Answer
Answer: c
Explanation: Graphite is not used as heating element in Thermal conductivity detector. Platinum,
gold and tungsten wire are used as heating elements.
14. Electron capture detector is much less susceptible to contamination when nickel is used
instead of tritium.
a) True
b) False
View Answer
Answer: a
Explanation: Electron capture detector is much less susceptible to contamination when nickel is
used instead of tritium. The sensitivity of nickel is less than that of tritium.
15. Given below is a diagram of electron capture detector. Identify the unmarked component in
the diagram.
a) Glass shield
b) Electrode
c) Quartz shield
d) Radioactive - emitter
View Answer
Answer: d
Explanation: The unmarked component is Radioactive - emitter. Nitrogen and hydrogen are the
best carrier gases for these detectors.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Gas Solid Chromatography.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Liquid Chromatography.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
High Pressure Liquid Chromatography.
1. Which of the following is not true about High pressure liquid chromatography (HPLC)?
a) It requires high pressure for the separation of the specious
b) There is no need to vaporise the samples
c) It is performed in columns
d) It has high sensitivity
View Answer
Answer: b
Explanation: In High pressure liquid chromatography (HPLC), samples need to be vaporised. It
has high sensitivity.
2. High pressure liquid chromatography can be performed only in columns.
a) True
b) False
View Answer
Answer: a
Explanation: High pressure liquid chromatography can be performed only in columns. This is
because application of high pressure in open bed will not be effective.
3. Which of the following is not an advantage of Syringe type pumps used in High pressure liquid
chromatography?
a) Independent of viscosity
b) Pulse-less flow
c) High pressure capability
d) Unlimited solvent capacity
View Answer
Answer: d
Explanation: The limitation of Syringe type pump is that it has limited solvent capacity and is
inconvenient when solvents are to be changed.
4. Which of the following is not true about solvent programming which is done in high
performance liquid chromatography?
a) It provides unequal bandwidths
b) It provides fast overall separation
c) It provides maximum resolution
d) It provides maximum sensitivity
View Answer
Answer: a
Explanation: Solvent programming is done to provide equal bandwidths. It provides maximum
sensitivity for every solute in the sample.
5. Which of the following pulse damper takes up some amount of the pulsation energy which is
released to provide smooth pressure without pulsations?
a) Flexible bellows or compressible gas passed through tee columns
b) Flexible inert diaphragm
c) Electronic pulse damper
d) Electrical pulse damper
View Answer
Answer: a
Explanation: Flexible bellows or compressible gas passed through tee columns take up some of
the pulsation energy which is released to provide smooth pressure without pulsations. Its main
purpose is to avoid pulses.
6. Which of the following is not a characteristic of the syringe pump used in high pressure liquid
chromatography?
a) Pressure capability is high
b) Maintenance is frequent
c) Limited reservoir capability
d) Slight change of flow rate when extremely high pressure compresses the solvent
View Answer
Answer: b
Explanation: In syringe pumps, maintenance is infrequent. However, it has high pressure
capability.
7. Syringe pumps used in High pressure liquid chromatography are most suitable for which of the
following columns?
a) Capillary columns
b) Guard columns
c) Short-fast columns
d) Small bore columns
View Answer
Answer: d
Explanation: Syringe pumps used in High pressure liquid chromatography are most suitable for
small bore columns. It is the most commonly used piston type pump.
8. Gravity feed method for solvent delivery is not used with narrow bore columns packed with fine
mesh particles.
a) True
b) False
View Answer
Answer: a
Explanation: Gravity feed method for solvent delivery is not used with narrow bore columns
packed with fine mesh particles. This is because they cannot deliver solvent at high pressure.
9. Which of the following cannot be done to reduce ripple in High pressure liquid
chromatography?
a) Using bellows
b) Using restrictors
c) Using long nylon tube between pump and column
d) Avoiding the use of solvent pump
View Answer
Answer: d
Explanation: Solvent pumps have to be used to reduce pulses in solvent flow. When there is a
pulsed flow, the efficiency of column decreases.
10. Which of the following is not true about Hydraulic capacitance flow control system used in
HPLC?
a) It can be used only for liquids with low viscosity
b) It is irrespective of solvent compressibility
c) It maintains constant flow
d) It smoothens high pressure pump pulsations
View Answer
Answer: a
Explanation: Hydraulic capacitance flow control system used in HPLC is irrespective of the
solvent viscosity. It is also irrespective of the solvent compressibility.
1. Bulk property detectors used in liquid chromatography does not respond to which of the
following properties?
a) Refractive index
b) Density
c) Properties of solutes
d) Dielectric constant
View Answer
Answer: c
Explanation: Bulk property detectors respond only to bulk properties of mobile phase. It does not
respond to properties of solutes.
2. Which of the following is not a property of a good detector used in liquid chromatography?
a) Good sensitivity
b) Ability to function in the presence of large background signal
c) Short response time
d) Volume of detector must be large
View Answer
Answer: d
Explanation: The volume of detector must be small. If the volume of the detector is large it may
lead to band broadening.
3. Detector selectivity is more important in Liquid chromatography than in Gas chromatography.
a) True
b) False
View Answer
Answer: a
Explanation: Detector selectivity is more important in Liquid chromatography than in Gas
chromatography. This is because in liquid chromatography, chances of overlapping are higher.
4. Which of the following UV absorbance detectors provide a real time spectrum of the
component of interest?
a) Continuous wavelength detector
b) Variable wavelength detector
c) Scanning wavelength detector
d) Fixed wavelength detector
View Answer
Answer: c
Explanation: Scanning wavelength detector provides a real time spectrum of the component of
interest. UV visible detector is the most widely used detection system.
5. Fluorescence detection is less selective than absorption detection.
a) True
b) False
View Answer
Answer: b
Explanation: Fluorescence detection is more selective than absorption detection. This is because
of its high sensitivity.
6. Which of the following detectors depend on Snells law at the interface between the cell wall
and the flowing liquid to deflect the light beam?
a) Electrochemical detectors
b) Fluorescence detectors
c) Refractive index detectors
d) Thermal conductivity detectors
View Answer
Answer: c
Explanation: Refractive index detectors depend on Snells law at the interface between the cell
wall and the flowing liquid to deflect the light beam. Changes in refractive index are monitored by
a position sensor.
7. Refractive index detectors used in liquid chromatography are not based on which of the
following processes?
a) Interference
b) Refraction
c) Reflection
d) Absorption
View Answer
Answer: d
Explanation: Refractive index detectors used in liquid chromatography are not dependent on
absorption. It depends on reflection, refraction and interference.
8. Which of the following detectors can be used for detection of amino acids in protein
hydrolyzates by introducing the reagent dansylchloride in the sample?
a) Electrochemical detectors
b) Fluorescence detectors
c) Refractive index detectors
d) Thermal conductivity detectors
View Answer
Answer: b
Explanation: Fluorescence detectors can be used for detection of amino acids in protein
hydrolyzates by introducing the reagent dansylchloride in the sample. Other compounds that can
be detected are petroleum products and natural products.
9. The reference cell is packed with which of the following in the Adsorption detector used in
liquid chromatography?
a) Inactive glass beads
b) Porous glass beads
c) Alumina
d) Silica
View Answer
Answer: a
Explanation: In Adsorption detector used in liquid chromatography, the reference cell is packed
with inactive glass beads. The other detector cavity is packed with silica, alumina or porous glass
beads.
10. Which of the following is true about Laser detectors used in liquid chromatography?
a) Causes thermal distortion, has decreased sensitivity
b) Causes thermal distortion, has increased sensitivity
c) Does not cause thermal distortion, has decreased sensitivity
d) Does not cause thermal distortion, has increased sensitivity
View Answer
Answer: a
Explanation: Laser detectors used in liquid chromatography cause thermal distortion and have
decreased sensitivity. This is because of the high energies of laser.
11. Which of the following detectors used in liquid chromatography is also called micro-
adsorption detector?
a) Electrochemical detectors
b) Fluorescence detectors
c) Refractive index detectors
d) Thermal detectors
View Answer
Answer: d
Explanation: Another name for thermal detector is micro-adsorption detector. The operation
depends upon temperature changes taking place due to the heat of adsorption.
12. In UV-Visible detectors, the cells are not made of which of the following materials?
a) Quartz
b) Teflon
c) Silica
d) KELF
View Answer
Answer: c
Explanation: The cells in UV-Visible detectors are made of quartz, KELF and Teflon. They are
not made of silica.
13. Which of the following is not a feature of refractive index type detectors?
a) Higher potential sensitivity
b) Low cost
c) High volume flow rates
d) Easy cell accessibility
View Answer
Answer: c
Explanation: Refractive index type detectors have low volume flow rates. It is one of the major
advantages of the refractive index type of detectors.
14. Which of the following detectors can detect almost any component?
a) Combining two UV detectors
b) Combining RI and UV detectors
c) Mass detector
d) Laser-based detectors
View Answer
Answer: b
Explanation: Combining RI and UV detectors almost any component can be detected. In some
cases, single detection system may give incorrect results.
15. Given below is the diagram of Ultraviolet detector used in liquid chromatography. Identify the
unmarked component.
a) Collimator
b) Lens
c) Monochromator
d) Quartz window
View Answer
Answer: d
Explanation: The part that is not marked is quartz window. This detector is used to detect
components that absorb any wavelength in the UV-Visible region.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Amino-Acid Analysers.
1. What is the drawback that occurs in using ion exchange chromatography on sulphonated
polystyrene resin and colourimetry for amino-acid analysis?
a) Less accuracy
b) Low resolution
c) Inconvenient to handle many individual samples
d) Slow in operation
View Answer
Answer: c
Explanation: When ion exchange chromatography is used for amino-acid analysis, high accuracy
is obtained. It is inconvenient to handle many individual samples and cleaning of glassware is
laborious.
2. Which one of the following methods is the most suitable for amino-acid analysis?
a) Gas chromatography
b) Ion exchange chromatography
c) Paper electrophoresis
d) Resin column chromatography
View Answer
Answer: d
Explanation: Resin column chromatography is the most suitable for amino-acid analysis. Here,
ion exchange resin chromatography is followed.
3. Which of the following colour reagents are used in Resin column chromatography?
a) Marquis reagent
b) Benedict reagent
c) Ninhydrin
d) Nesslers reagent
View Answer
Answer: d
Explanation: Ninhydrin is the colour reagent used in resin column chromatography. The colour
reaction occurs during a known period and controlled temperature.
4. Which of the following amino-acids is measured at a wavelength of 440nm using photometric
systems?
a) Proline
b) Alanine
c) Glutamine
d) Valine
View Answer
Answer: a
Explanation: Proline and hydroxyproline are measured at 440nm. All the other amino-acids are
measured at 570nm.
5. In Automatic amino-acid analyzer, the sample containing ___________ of each amino
compound is introduced at the top of the ion exchange column.
a) 1 to 10 moles
b) 1 to 10 moles
c) 0.05 to 2 moles
d) 0.05 to 2 moles
View Answer
Answer: d
Explanation: In Automatic amino-acid analyzer, the sample having 0.05 to 2 moles of each
amino compound is introduced at the top of the ion exchange column. Buffer is then supplied to
the column.
6. Which of the following cannot be analysed using resin column chromatography?
a) Peptides
b) Amines
c) Amino compounds
d) Components which are ninhydrin negative
View Answer
Answer: d
Explanation: Components which are ninhydrin negative cannot be analysed using resin column
chromatography. Only the components that are nindydrin positive are analysed using this
method.
7. Trapped air in pumps of Automatic amino-acid analyzer is eliminated using which of the
following ways?
a) Suction pump
b) De-aerator
c) Overcome naturally by pump pressure
d) Using pulse damper
View Answer
Answer: b
Explanation: De-aerator is used to eliminate trapper air in pumps of automatic amino-acid
analyzer. Here, bubble-trap type de-aerator is used.
8. Amino-acids are distinguishable from each other only by one or two atoms in their structure.
a) True
b) False
View Answer
Answer: a
Explanation: Amino-acids are distinguishable from each other only by one or two atoms in their
structure. They are components all proteins and are essential for growth of organisms.
9. What is the drawback that occurs in using gas chromatography for amino-acid analysis?
a) Costly
b) Cannot be used for proteins or polypeptides
c) Slow process
d) Complex in operation
View Answer
Answer: b
Explanation: Gas chromatography cannot be used for proteins or polypeptides. Further, accuracy
is lost while converting amino-acids into gaseous phase.
10. Resin column chromatography does not involve breaking of bonds in amino-acids.
a) True
b) False
View Answer
Answer: b
Explanation: Resin column chromatography involves breaking of bonds in amino-acids. This
causes a delay as acid is eluted through the column by a buffer.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs)
focuses on Paramagnetic Oxygen Analyser.
1. Which among the following gases have diamagnetic property (ability to be repelled by
magnetic fields)?
a) Oxygen
b) Nitrogen
c) Nitrogen dioxide
d) Nitric oxide
View Answer
Answer: b
Explanation: Nitrogen has diamagnetic property. Other components that have diamagnetic
property are carbon, boron, etc.
2. Which of the following gases have paramagnetic property (ability to get attracted to magnetic
field)?
a) Nitric oxide
b) Hydrogen
c) Helium
d) Nitrogen
View Answer
Answer: a
Explanation: Nitric oxide has paramagnetic property. Nitrogen dioxide and oxygen are other
gases that have paramagnetic property.
3. The force produced during operation of Paramagnetic oxygen analyzer is proportional to which
of the following?
a) Magnetic susceptibility of sphere
b) Magnetic susceptibility of surrounding gas
c) Difference between magnetic susceptibility of sphere and magnetic field strength
d) Difference between magnetic susceptibility of sphere and that of surrounding gas
View Answer
Answer: d
Explanation: The paramagnetic property of oxygen is used in this method for the analysis of
oxygen. The force produced during operation of Paramagnetic oxygen analyzer is proportional to
difference between magnetic susceptibility of sphere and that of surrounding gas.
4. In Paramagnetic oxygen analyzer, the expression for magnitude of force produced can be
expressed as, F = C (K-KO), where K= magnetic susceptibility of surrounding gas and KO=
magnetic susceptibility of sphere. C denotes which of the following?
a) Magnetic field strength
b) Magnetic gradient
c) Function of magnetic field strength and gradient
d) Magnetic flux
View Answer
Answer: c
Explanation: In the expression for magnitude of force, C denotes the function of magnetic field
strength and gradient. The forces are thus a measure of magnetic susceptibility and hence of
oxygen.
5. In Paramagnetic oxygen analyzer, the electrostatic force that is exerted should have which of
the following features?
a) It should be greater than the magnetic field and must be in the same direction
b) It should be equal to the magnetic field and must be in the same direction
c) It should be greater than the magnetic field and must be in the opposite direction
d) It should be equal to the magnetic field and must be in the opposite direction
View Answer
Answer: d
Explanation: The electrostatic force that is exerted should be equal to the magnetic field and
must be in the opposite direction. It is exerted by two charged vanes.
6. Which of the following is true about the two charged vanes that produce electric field in
Paramagnetic oxygen analyzer?
a) Both the vanes should be at higher potential than the test body
b) Both the vanes should have the same potential as the test body
c) One vane should be at higher potential than the test body and the other vane should be at
lower potential
d) Both the vanes should be at lower potential than the test body
View Answer
Answer: c
Explanation: In Paramagnetic oxygen analyzer, the two charged vanes should be such that one
vane should be at higher potential than the test body and the other vane should be at lower
potential. This electric field produces a force.
7. In Paramagnetic oxygen analyzer, when no oxygen is present the magnetic force balances
torque of the fibre.
a) True
b) False
View Answer
Answer: a
Explanation: In Paramagnetic oxygen analyzer, when no oxygen is present the magnetic force
balances torque of the fibre. If oxygen is present,it would displace the dumb-bell sphere.
8. To improve paramagnetic oxygen analyzer, the quartz suspension must be replaced with
which of the following?
a) Platinum-iridium
b) Phosphor-bronze
c) Gold-palladium
d) Antimony
View Answer
Answer: a
Explanation: To improve paramagnetic oxygen analyzer, the quartz suspension must be replaced
with Platinum-iridium. It is more robust.
9. In the Beckman instrument, the flow rate recommended when the sample enters through a
porous diffusion disc is which of the following?
a) 100-200 cc/min
b) 50-250 cc/min
c) 40-60 cc/min
d) 200-300 cc/min
View Answer
Answer: b
Explanation: In the Beckman instrument, the flow rate recommended when the sample enters
through a porous diffusion disc is 50-250 cc/min. paramagnetic oxygen analyzer was first
described by Pauling.
10. In the Beckman instrument, the flow rate recommended when the sample enters directly for
rapid response is which of the following?
a) 100-200 cc/min
b) 50-250 cc/min
c) 40-60 cc/min
d) 200-300 cc/min
View Answer
Answer: c
Explanation: In the Beckman instrument, the flow rate recommended when the sample enters
directly for rapid response is 40-60 cc/min. paramagnetic oxygen analyzer was first described by
Pauling.
11. Which of the following happens to the magnetic susceptibility of gases when temperature
changes?
a) It increases with increase in temperature
b) It decreases with increase in temperature
c) It remains constant with increase in temperature
d) It remains constant with decrease in temperature
View Answer
Answer: b
Explanation: Magnetic susceptibility of gases decreases with increase in temperature. Magnetic
susceptibility of oxygen decreases with increase in temperature.
12. If the oxygen content is not greater than 21%, which of the following gases must be used to
set the span point?
a) Oxygen
b) Dry air
c) Nitrogen
d) Hydrogen
View Answer
Answer: b
Explanation: If the oxygen content is not greater than 21%, dry air is used to set the span point.
This is done for calibration purposes.
13. If the oxygen content is greater than 21%, which of the following gases must be used to set
the span point?
a) Oxygen
b) Dry air
c) Nitrogen
d) Hydrogen
View Answer
Answer: a
Explanation: If the oxygen content is greater than 21%, oxygen is used to set the span point. This
is done for calibration purposes.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Magnetic Wind Instruments.
1. Which of the following does not occur in Hartman and Braun oxygen analyzer based on
magnetic wind?
a) Oxygen is attracted to magnetic field
b) Oxygen becomes less paramagnetic due to rise in temperature
c) Resistance change in a resistance connected to the Wheatstone bridge occur
d) Wheatstone Bridge becomes balanced due to temperature change
View Answer
Answer: d
Explanation: In Hartman and Braun oxygen analyzer, the magnetic wind cools the heated
winding. As a result, the bridge becomes unbalanced.
2. Which of the following is not a cause of error that could occur in Hartman and Braun oxygen
analyzer based on magnetic wind?
a) The filament temperature could be affected by changes in thermal conductivity of the carrier
gas
b) Hydrocarbons in sample stream may react on the heated filaments
c) Gravitation chimney flow effect lead to errors when cross tube is horizontal
d) Combustible gases may react on the heated filaments
View Answer
Answer: c
Explanation: Error occurs when cross tube is vertical. When cross tube is horizontal, errors due
to gravitation chimney flow effect is avoided.
3. Hydrocarbons need to be removed from the sample gas to avoid errors in Hartman and Braun
oxygen analyzer.
a) True
b) False
View Answer
Answer: a
Explanation: Hydrocarbons need to be removed from the sample gas to avoid errors. This is
because they may react on the heated filaments and lead to change in temperature and
resistance.
4. Magnetic wind analyzers are also known as which of the following?
a) Thermo-magnetic analyzers
b) Electro-magnetic analyzers
c) Thermo-analytical analyzers
d) Hot wire analyzers
View Answer
Answer: a
Explanation: Magnetic wind analyzers are also known as Thermo-magnetic analyzers. In Europe,
it is known as Magnetic wind analyzers. In United States, it is known as Thermo-magnetic
analyzers.
5. In OXYMAT-M oxygen analyser which of the following is usually used as reference gas?
a) Hydrogen
b) Nitrogen
c) Helium
d) Ambient air
View Answer
Answer: d
Explanation: In OXYMAT-M oxygen analyser, ambient air is usually used as reference gas. Its
measuring procedure is based on pressure difference which develops between two gases having
different oxygen concentration in the magnetic field.
6. Which of the following is used in magnetic wind instruments to determine the amount of
oxygen?
a) Change in resistance
b) Change in capacitance
c) Change in inductance
d) Change in concentration
View Answer
Answer: a
Explanation: In magnetic wind instruments to determine the amount of oxygen, change in
temperature leads to a change in resistance. This change in resistance is measured to determine
the amount of oxygen.
7. Which of the following concepts are not used in magnetic wind instruments?
a) Change in resistance due to cooling by magnetic wind
b) Displacement of membrane by magnetised oxygen
c) Difference in pressure difference between two gases having varying concentration of oxygen
d) By measuring temperature using thermometer
View Answer
Answer: d
Explanation: Temperature is not measured in magnetic wind instruments. The cooling effect of
magnetic wind causes a change in resistance which is measured.
8. How are the hydrocarbons in the samples removed in magnetic wind instruments?
a) By using a cold trap
b) By using a filter
c) By using a semipermeable membrane
d) By using magnetic effect
View Answer
Answer: a
Explanation: The hydrocarbons in the samples are removed in magnetic wind instruments by
using a cold trap. It is removed to avoid combustion.
9. Which of the following bridges are used in magnetic wind instruments?
a) Wheatstone bridge
b) Kelvins bridge
c) Andersons bridge
d) Scherings bridge
View Answer
Answer: a
Explanation: Wheatstone bridge is generally used in magnetic wind instruments. The change in
resistance due to cooling effect is measured using Wheatstone bridge.
10. The extremely high inertia of the thin membrane in the oxygen sensor allows it to rapidly
attain equilibrium.
a) True
b) False
View Answer
Answer: b
Explanation: The extremely low inertia of the thin membrane in the oxygen sensor allows it to
attain equilibrium. It reaches equilibrium state in 0.25 to 0.5 seconds.
This set of Analytical Instrumentation Interview Questions and Answers focuses on Introduction
to UV Visible Spectrometers.
1. Beer Lamberts law gives the relation between which of the following?
a) Reflected radiation and concentration
b) Scattered radiation and concentration
c) Energy absorption and concentration
d) Energy absorption and reflected radiation
View Answer
Answer: c
Explanation: Beer Lamberts law give the relation between Energy absorption and Concentration.
It was proposed by Beer and Lambert.
2. In which of the following ways, absorption is related to transmittance?
a) Absorption is the logarithm of transmittance
b) Absorption is the reciprocal of transmittance
c) Absorption is the negative logarithm of transmittance
d) Absorption is a multiple of transmittance
View Answer
Answer: c
Explanation: Transmittance is the ratio of the radiant power transmitted by a sample to the
radiant power incident on the sample. Absorption is the negative logarithm of transmittance.
3. Which of the following is not a limitation of Beer Lamberts law, which gives the relation
between absorption, thickness and concentration?
a) Concentration must be lower
b) Radiation must have higher bandwidth
c) Radiation source must be monochromatic
d) Does not consider factors other than thickness and concentration that affect absorbance
View Answer
Answer: b
Explanation: The law is derived assuming that the radiation is monochromatic. So, if bandwidth
increases it will create deviation.
4. Beers law states that the intensity of light decreases with respect to ___________
a) Concentration
b) Distance
c) Composition
d) Volume
View Answer
Answer: a
Explanation: Beers law states that the intensity of light decreases with the concentration of the
medium. It was stated by Beer.
5. Lamberts law states that the intensity of light decreases with respect to __________
a) Concentration
b) Distance
c) Composition
d) Volume
View Answer
Answer: b
Explanation: Lamberts law states that the intensity of light decreases with respect to
concentration of the medium. It was stated by Lambert.
6. The representation of Beer Lamberts law is given as A = abc. If b represents distance, c
represents concentration and A represents absorption, what does a represent?
a) Intensity
b) Transmittance
c) Absorptivity
d) Admittance
View Answer
Answer: c
Explanation: a represents the absorption constant. It is also known as absorptivity.
7. Which of the following is not true about Absorption spectroscopy?
a) It involves transmission
b) Scattering is kept minimum
c) Reflection is kept maximum
d) Intensity of radiation leaving the substance is an indication of concentration
View Answer
Answer: c
Explanation: In Absorption spectroscopy, reflection must also be kept minimum along with
scattering. Amount of absorption depends on the number of molecules in the material.
8. Transmittance is given as T = P/Po. If Po is the power incident on the sample, what does P
represent?
a) Radiant power transmitted by the sample
b) Radiant power absorbed by the sample
c) Sum of powers absorbed and scattered
d) Sum of powers transmitted and reflected
View Answer
Answer: a
Explanation: P represents radiant power transmitted by the sample. Transmittance is the ratio of
radiant power transmitted by the sample to the radiant power that is incident on it.
9. What is the unit of absorbance which can be derived from Beer Lamberts law?
a) L mol-1 cm-1
b) L gm-1 cm-1
c) Cm
d) No unit
View Answer
Answer: d
Explanation: Absorbance has no unit. The units of absorptivity, distance and concentration
cancel each other. Hence, absorption has no unit.
10. What is the unit of molar absorptivity or absorptivity which is used to determine absorbance A
in Beer Lamberts formula?
a) L mol-1 cm-1
b) L gm-1 cm-1
c) Cm
d) No unit
View Answer
Answer: a
Explanation: The unit of absorptivity is L mol-1 cm-1. If concentration is represented as gm per litre
it becomes L gm-1 cm-1.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Absorption Instrumentation and Block Diagram Representation.
1. Which is the missing block in the block diagram for Absorption of Radiation Instrument given
below?
a) Filter
b) Reflector
c) Converging lens
d) Detector
View Answer
Answer: d
Explanation: To measure the amount of radiation absorbed and transmitted detector is very
important. It detects the amount of radiation which is then inferred using read out device.
2. Which of the following detectors does not require a battery and is also known as barrier layer
cell?
a) Photomultiplier tube
b) Photovoltaic cell
c) Photoemissive tubes
d) Photo reflector
View Answer
Answer: b
Explanation: Photovoltaic cell does not require a battery for operation. Its working is entirely
different from Photomultiplier tube or Photoemissive tubes.
3. Which of the following detectors is used to detect light intensities which are very weak?
a) Photomultiplier tube
b) Photovoltaic cell
c) Photoemissive tubes
d) Photo reflector
View Answer
Answer: a
Explanation: PMT is used for detection of light intensities which are weak. As the name suggests,
Photomultiplier tube multiplies the incident electrons using dynodes causing an avalanche of
electrons.
4. How is Tungsten Halogen lamp differ from normal Tungsten filament lamp used in absorption
spectroscopy?
a) It has a tungsten filament and is filled with inert gas
b) Iodine is added to normal filling gas
c) Iodine is coated on tungsten filament
d) Iodine is added to inert gas
View Answer
Answer: b
Explanation: In Tungsten filament lamp tungsten filament is enclosed in a bulb of glass filled with
inert gas or vacuum. In Tungsten Halogen lamp iodine is added to normal filling glass.
5. Instead of glass filters, why gelatin filters could not be used for a long period while both are
Absorption filters?
a) Gelatin tends to evaporate and hence they deteriorate
b) Gelatin is affected by humidity in environment
c) They deteriorate due to absorption of heat leading to changes in gelatin
d) Gelation is affected by temperature in environment
View Answer
Answer: c
Explanation: With absorption of heat they deteriorate due to changes in gelatin. Bleaching of dye
occurs.
6. How does continuous wedge filter differ from normal interference filter used in absorption
spectroscopy?
a) It permits continuous selection of different wavelength
b) It allows narrow band of wavelengths to pass
c) It has two semi-transparent layers of silver
d) Space layer is made of a substance having low refractive index
View Answer
Answer: a
Explanation: It allows continuous selection of wavelength by using a spacer film of graded
thickness. Rest of the options are properties of normal interference filters.
7. Which of the following could be used as the layer of dielectric in interference filters used in
Absorption Spectroscopy?
a) Graphite
b) MgF2
c) Fe
d) AgNO3
View Answer
Answer: b
Explanation: MgF2 is used as layer of dielectric in interference filters. Other material that can be
used is ZnS.
8. How can stability of radiation be achieved in incandescent or discharge source used in
Absorption Spectroscopy?
a) Using filters
b) Using monochromators
c) Using slits
d) By controlling the source voltage
View Answer
Answer: d
Explanation: The intensity of radiation in incandescent source is proportional to the lamp source
voltage. Therefore, by controlling the source voltage stability can be achieved.
9. To tolerate high operating temperatures, which of the following has to be done in incandescent
or tungsten filament lamps?
a) Alloys must be used
b) Nitrogen be used instead of inert gas
c) Envelope is fabricated with quartz
d) Envelope is fabricated with copper
View Answer
Answer: c
Explanation: The envelope is fabricated with quartz to allow high operating temperatures.
Tungsten filament and inert gas are generally used and are not modified.
10. Which of the following is not a reason for laser not being generally used as a source of
radiation for UV, Visible Spectroscopy?
a) High cost
b) Limited range of wavelength
c) Less intensity
d) Complex to work with
View Answer
Answer: c
Explanation: Laser has high intensity. It is used in special applications where cost is not a matter
and limited range of wavelength is required.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Single Beam and Double Beam Instruments.
1. Which of the following statements is false about single beam absorption instruments?
a) Tungsten bulb is used as source
b) Beam splitter is used to get parallel beam
c) Test tube is used as sample holder
d) Photovoltaic cell as detector
View Answer
Answer: b
Explanation: Single beam instruments make use of one beam. Therefore, beam splitters are not
required in single beam instruments.
2. Which of the following statement is false about double beam absorption instruments?
a) It is similar to single beam instruments except two beams are present
b) Tungsten bulb is used as source
c) Reference beam must have higher intensity than sample beam
d) Both the beams after they pass through respective samples are compared
View Answer
Answer: c
Explanation: Reference beam cannot have higher intensity than sample beam. The beam is split
into two beams of equal intensity.
3. Which of the following is not an application of colorimeter?
a) Paints
b) Inks
c) Cosmetics
d) Composition detection
View Answer
Answer: d
Explanation: Colorimeter is not used to determine composition. Its application is paints, dyes,
inks, cosmetics and plastics.
4. In photometers, the readings of the specimen are initially obtained in the form of which of the
following parameters?
a) Transmittance
b) Absorption
c) Wavelengths
d) Volume
View Answer
Answer: a
Explanation: In photometers, the reading is initially obtained in the form of transmittance as some
radiation is absorbed by the sample and the rest of the beam is transmitted. This transmitted
beam is measured by the detector.
5. Colorimeters are used in applications where great accuracy is required.
a) True
b) False
View Answer
Answer: b
Explanation: Colorimeters are used in applications where great accuracy is not required. They
are also known as photometers.
6. In the diagram of single beam photometer given below, identify the component that is not
marked.
a) Monochromator
b) Absorption filter
c) Sample holder
d) Interference filter
View Answer
Answer: c
Explanation: In single beam spectrophotometer, the beam passed through the sample which is
held in the sample holder. The transmitted beam is measured by the detector.
7. Colorimeters are used to determine concentration of solutions.
a) True
b) False
View Answer
Answer: a
Explanation: Colorimetry is the science of colour measurement. It is used to determine
concentrations of solutions.
8. Which of the following is the purpose of balance indicator in double beam photometer or
colorimeter?
a) Selects particular wavelength
b) Splits the wavelength selected into two equal beams
c) Detects and indicates the amount of light falling on it
d) Indicates the difference between the output of two photometers
View Answer
Answer: d
Explanation: It compares the output of the two photometers obtained using two beams. It
indicates the output.
9. Which of the following is the purpose of beam splitter in double beam photometer or
colorimeter?
a) Splits beam into two equal intensity beams
b) Splits beam in such a way that sample beam has higher intensity
c) Splits beam in such a way that reference beam has higher intensity
d) Merge two equal intensity beams into single beam
View Answer
Answer: a
Explanation: Beam splitter splits beam into two equal intensity beams. One beam passes through
the sample and other through the reference.
10. Which of the following is a source used in spectroscopy?
a) LASER
b) Tube light
c) Sodium vapour lamp
d) Tungsten lamp
View Answer
Answer: d
Explanation: Tungsten lamp is the source used in spectroscopy. It is the source used in UV,
Visible spectroscopy.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Instrumentation of IR.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Fourier Transform Infra-red Spectrometers.
1. Which of the following is not true about Fourier Transform Infrared (FTIR) spectrometer?
a) It is of non-dispersive type
b) It is useful where repetitive analysis is required
c) Size has been reduced over the years
d) Size has increased over the years
View Answer
Answer: d
Explanation: FTIR is of non-dispersive type of instruments and is used for repetitive analysis.
Initially, it was bulky and cost was high and hence, it was limited to special application. Now, the
size has been reduced.
2. In the most widely used beam splitter, a thin film of ________ is sandwiched between two
plates of low refractive index solid. Fill the blank with the suitable option.
a) Mylar
b) Silicon carbide
c) Ferrous oxide
d) Silver chloride
View Answer
Answer: a
Explanation: Generally, mylar is the thin film used in beam splitters. Beam splitter has to pass
50% of the radiation and reflect 50% of the radiation.
3. Which of the following is not the function of drive mechanism in Fourier Transform Infrared
Spectrophotometer?
a) Movement of mirror to obtain a satisfactory interferogram
b) Acquire a good interferogram pattern
c) Allow 50% of the beam to pass
d) Keep the speed of the moving mirror constant
View Answer
Answer: c
Explanation: Drive mechanism does the functions specified in the other options. It the function of
the beam splitter to allow 50% of the beam to pass through.
4. Only pyroelectric transducer or pyroelectric crystals are used as detectors in Fourier
Transform Infrared Spectrophotometer (FTIR). What is the main reason for other types of thermal
detectors are not being used in FTIR spectrophotometer?
a) Less accuracy
b) Slower response
c) Less precision
d) Less sensitivity
View Answer
Answer: b
Explanation: Other thermal detectors are not employed due to slower response of the detectors.
Pyroelectric transducers or pyroelectric crystals are chosen for their high speed, accuracy,
precision, sensitivity and resolution.
5. Which of the following is not the advantage of Fourier Transform Spectrometers?
a) Signal to noise ratio is high
b) Information could be obtained on all frequencies
c) Retrieval of data is possible
d) Easy to maintain
View Answer
Answer: d
Explanation: The instrument is not easy to maintain and it is a disadvantage. The other
disadvantage is that the cost is high.
6. Which of the following has to be computed to determine transmittance and absorbance at
various frequencies?
a) Ratio of signal and noise
b) Ratio of sample and reference spectra
c) Sample spectra
d) Reference spectra
View Answer
Answer: b
Explanation: The ratio of sample and reference spectra needs to be computed to determine
transmittance and absorbance. First, reference interferogram is obtained and then sample is
obtained.
7. Computer accepts analog signals directly.
a) True
b) False
View Answer
Answer: b
Explanation: Computer does not accept analog signals directly. An A/D converter is required to
feed the signals to the computer.
8. Which of the following is the reference that is generally used in FTIR interferometer?
a) Air
b) NaCl solution
c) Alcohol
d) Base solution
View Answer
Answer: a
Explanation: Air is generally used as the reference. It is scanned for about 20 to 30 times and the
results are stored in a computer.
9. In Michelson Interferometer, if the reflected and transmitted beams are in phase at the beam
splitter, then maximum intensity will reach the detector.
a) True
b) False
View Answer
Answer: a
Explanation: If the reflected and transmitted beams are in phase at the beam splitter then
maximum intensity will reach the detector. If they are out of phase, then minimum intensity will
reach the detector.
10. The diagram given below is the representation of Fourier transform interferometer. Identify
the missing block in the block diagram.
a) Pyroelectric crystal
b) Display
c) High speed plotter
d) A/D converter
View Answer
Answer: d
Explanation: Computer does not accept analog signals directly. An A/D converter is required to
feed the signals to the computer.
11. In Michelsons interferometer, the frequency of the detector output can be determined by
translating the _________ of movable mirror and the ___________ of monochromatic radiation.
a) Velocity, wavelength
b) Thickness, intensity
c) Length, velocity
d) Angle, intensity
View Answer
Answer: a
Explanation: The frequency can be determined by translating the velocity of the movable mirror
and the wavelength of monochromatic radiation. The reflected beam passes towards the
movable mirror.
12. In Michelsons interferometer, the __________ of the detector output will depend upon the
intensity of incoming radiation.
a) Velocity
b) Frequency
c) Amplitude
d) Phase
View Answer
Answer: c
Explanation: In Michelsons interferometer, the amplitude of the detector output will depend upon
the intensity of incoming radiation. If movable mirror is moved uniformly, the output will be a sine
wave.
13. Why is the computer necessary in Fourier Transform Spectrometer?
a) To display the detector output
b) To process the detector output
c) To determine the amplitude
d) To determine the frequency
View Answer
Answer: b
Explanation: The computer is necessary to process the output of the detector. The Fourier
Transform of output is determined using software using computers.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Atomic Absorption Spectroscopy.
This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on
Flame Emission Photometers.
a) Anode
b) Cathode
c) Grid
d) Dynode
View Answer
Answer: b
Explanation: The cathode is formed of half cylinder. The anode is a metal ring near the centre of
the glass.
15. Which of the following is not an advantage of photovoltaic cell which is used as a detector in
Flame emission photometers?
a) Portable
b) No external supply
c) Robust in construction
d) Does not show fatigue
View Answer
Answer: d
Explanation: It does show fatigue. It shows decrease in response with continued illumination.