Vr6100 Measurement
Vr6100 Measurement
Vr6100 Measurement
■Measurement overview
The block memory is the memory which is exclusive for the timer measurement to store the data that is
automatically stored during the timer measurement. B mark indicates on the memory mode display of the LCD
screen. When executes the timer measurement with settings of timer ON and real time octave display OFF
respectively, the calculation values and measurement conditions of 3 axes or 1 selected axis is stored
automatically after completing each measurement. This also applies to the real time octave measurement
(optional function), but in this function, the measurement conditions are stored in the real time memory. The
followings are an explanation of the measurement procedure without using the real time function. In addition,
timer function is with saving electricity function.
■Block memory
The block memory is divided into 10 blocks. Enable to select the data to be stored from three axes or one axis.
Maximum 480 combined data for three axes and maximum 1440 combined data for one axis can be stored per
1 block. The recalled data can be displayed in various formats of calculated values by pressing the panel
switch [L Leq LE], [AXIS] or [LIST].
-1-
■Storing data to block memory
The block memory is interlocked with the timer measurement to storage data as follows.
<Caution>
The data cannot be stored in the block memory with the panel switch operation.
The timer measurement which is executed in the real time octave display cannot be stored in the
block memory.
When the data are stored up to block 9 by the previous timer measurement, only the timer
measurement is executed and the measurement data are not stored, unless the previous block
memory data are cleared.
When the data are not stored in the block, mark "----" is indicated on the LCD screen of memory
address.
-2-
■Saving electricity mode function
-3-
■Measurement procedure
Set the timer measurement OFF and real time octave display OFF respectively.
Press the panel switch [MENU], and switch to the menu mode. Menu mode is displayed by switching the
menu screen.
At first, press the panel switch [<] or [>] (ADDRESS), and flash setting item on the main menu [4.DATA
MEMORY] and press the panel switch [ENTER] to switch to (4.DATA MEMORY) menu.
Press the panel switch [<] or [>] (ADDRESS) to flash setting item [2.B AXIS], and press the panel switch
[ENTER] to flash the setting value of the store axis selection for the block memory. Press the panel switch
[<] or [>] (ADDRESS) to flash the X/Y/Z or 3 axes (XYZ) to be set, and press the panel switch [ENTER] to
return the flashing cursor to the setting item [2.B AXIS].
Press the panel switch [<] or [>] (ADDRESS) to flash the setting item [OK], and press the panel switch
[ENTER]. If the panel switch [ENTER] is pressed when [OK] is flashing, the setting conditions are
confirmed and return to the menu screen. Also, at this time, when the panel switch [MENU] is pressed, the
measurement screen is switched to the screen which is directly after the setting condition is confirmed. In
addition, if the condition setting is not confirmed, press the panel switch [ENTER] to flash [CANCEL].
<Caution>
The axis selections are managed in block unit. Therefore, selected axis is available for the new block
which is being stored.
The setting value cannot be changed during the calculation, internal calibration signalling output or
memory recall.
-4-
2. Setting the timer function
Start the timer function after the selection of store axis.
① Press panel switch [<] or [>] (ADDRESS) to move the flashing cursor to the setting item "minute".
② Press panel switch [<] or [>] (ADDRESS) to change the "minute".
③ Press panel switch [<] or [>] (ADDRESS) to move the flashing cursor to the setting item "second".
④ Press panel switch [<] or [>] (ADDRESS) to change the "second".
⑤ In the end, press the panel switch [ENTER] to conform the adjustment of hour, minute and second.
-5-
Setting the measurement time
[2.MEAS.T]
-6-
Setting the measurement period [3.PERIOD] or total measurement time [4.TOTAL]
Set measurement period [3. PERIOD] and total measurement time [4. TOTAL] respectively with the
same procedure explained in the paragraph of “Setting of measurement time [2.MEAS.T]”.
The figure in the paragraph of “Confirmation of all setting conditions” is the setting examples when
the measurement period [3. PERIOD] is 8 hours 00 minute 00 second (008:00:00), and all measurement
time is 24 hours 00 minute 00 second (0024:00:00) respectively.
-7-
3. Timer measurement procedure
After setting the timer measurement condition, timer measurement executes as the following procedure.
<Caution>
Automatically switch to the saving electricity mode when the waiting time to start the measurement
after pressing the panel switch [START] is more than 30 seconds, and no key operation for 10
seconds.
When the measurement period setting is 30 seconds or more, and the panel switch [PRINT] is
pressed for 0.5 seconds or more right after starting the timer measurement, the measurement results
can be outputted from the printer immediately after each measurement. Also, at this time, RS mark is
flashed.
- END -