Cardio Vascular
Cardio Vascular
Cardio Vascular
Unit II
Cardio Vascular and Nervous
System
irL '*o
o
\
Right atrium Aortic valve
s such as
nunicate
rlti-level Mitral valve
rbody to
Eide the
is given
Left ventricle
ssential
heblood Myocardium
rs called Tricuspid valve
::,::r@
;rn:firi;N
ntc,,g
Ef,M
]F{
mltril fi
usm!fir
-O^ ?!mEm
COZ Lung ::rrmdm
_.-
Semilunar
Location of tJ-l
sinus node Left atrium
valve
lh*:ery
Right atrium Aortic valve SStrifM?
Tricuspid Jt -l-a'4
valve Mitral valve
Right ventricle
it€ :mm
Left ventricle
Legs
distributed to the capillaries in the tissues, where it gives up its oxygen and chemical compounds,
takes up carbon dioxide and products of combustion.
The blood returns to the heart along different routes from different parts of the body. It usualil,
passes from the venous side of the capillaries directly via the venous system to either the superior
vena cava or the inferior vena cava, both of which empty into the right atrium. The heart itself is
supplied by two small but highly important arteries, the coronary arteries. They branch from the
aorta just above the heart. If they are blocked by coronary thrombosis, myocardial infarction
follows, often leading to a fatal situation.
The heart rate is partly controlled by autonomic nervous system and partlyby harmone action.
These control the heart pump's speed, efficiency and the fluid flow pattern through the system.
The circulatory systern is the transport system of the body by which food, oxygen, water and
other essentials are transported to the tissue cells and their waste products are transported awav.
This happens through a diffusion process in which nourishment from the blood cell diffuses
lllllI
through the capillary wall into interstitial fluid. Similarly, carbon dioxide and some waste
produ-cts from the interstitial fluid diffuses through the capillary wall into theblood cell.
The condition of the cardiovascular system is examined by haemodynamic measurements and
by recording the electrical activitv of the heart muscle (electrocardiography) and listening to the
heart sounds (phonocardiography). For assessing the performance of the heart as a PumP/
measurement of the cardiac output (amount of blood pumped by the heart per unit time), blood
pressure, blood flow, rate and blood volume are made at various locations throughout the
circulatory system.
The respiratory system in the human body (Fig. 1.3) is a pneumatic system in which-an air Pump
(diaphragm) altemately creates negative and positive pressures in a sealed chamber (thoracic
cavi?y) ,id .urr", air to be sucked into and forced out of a pair of elastic bags (lungs). The lungs
are connected to the outside environment through a passage way comprising nasal cavities,
Pharynx
(throat)
Larynx
Trachea
windpipe
(air passage)
Bronchiole
(smallest air
passage)
mrPounds,
Pleura
p.It usually
hesuperior
eart itself is
dr from the
I infarction Alveoli '
(brarrch from bronchiole
where exchange occurs)
ureaction.
Diaphragm
Esystem.
,water and
rhdaway.
dl diffus€s > Fig.1.3 TheRespiratory system
Biomedical Signal Processing
.. ]} $"-! ILECTROCARDIOGRAPH
records the eiectrical activity of the hear.
Tire electrr:cardiograph (ECG) is an instrument, which
signals from the heart characteristically prececle the
normal mechanical function and
Electrical
ECC provides valuable information
monitoring of these ,ignut, has great clinical significance'
ii-r" po"t"t of an inactive pari (infarction) or an
alrout a r.vide range of Jardiac dis"orders such as
""
Electrocardiographs are used in catheteriza-
mrlargement 1cur,lluc hifeitrophy) of the heart muscle'
ancl for routine diagnostic applications in cardiology'
tion lal:oratoriur, .o.oirlry
"rie.rnits by the heart can be best characterized by vector quantities'
although the eiectric field generated
quantities' i'e' a voltage difference of m\'
it is generally convenient to directly measure only scalar
useful frequency range is usualh'
*rderbetween the gi'en points of inemay' TIr" diagnostically -d'$
accepteci as (').05 to i 50 liz (Golden et al 1'973).The
implifier and writing part should faithfullr' :le st]-
reprociuce signals in this range. A good low frequen.y
i"'po"'" is essential to ensure stability of ot"erat
of several factors like isolati.n between a
the basetrine. High frequency"rerpo:nr" is a compromise Ela
(myographic potentials) and limitations
use{ul ECG signrl rroJ1 othJ, signals of biologiial origin -nes a
and friction' The interference of non-
of tire tlirect writing p"r, ,u"ori"rs due to riass, inertia il€agJ
amplifiers' which are capable oi
biologicai origin .ur, iu handled by usingmodern differential :eccrd
order of L00-120 dB with 5 kO unbalance
providing excellent."i"J.".rp"uiryillcrvrRRof the i6dt
In addition to this, under specially adverse
in ihe leads is a desirable feature of ECG machines.
circumstances, it becomes necessary to include a
notch filter tuned to 50 Hz to reject hum due l*ld
from the changes of the contact ei*tn
to power mains. rrre inJaUility oithe baseline, originating
impedance, demands the applicationof the automaticbaseline
stabilizing circuit' A minimum of anl'I
speecls ir rru."rru,.v (25 and 50 mm Per sec)
for ECG recording' ntct
twt paper
irllf*
-{ss
S.l.lBlockDiagramDescriptionofanElectrocardiosraph ',rfut
machine' The potentials picked up
Figure 5.1 shows the block diagram of an electrocardiograph
bi, ihe patient electrodes are tafe'r tc the leacl selector
u*it.h- In the lead selector, the etrectrodes are
Biomedical Recorders 155
:elected twoby two according to the lead program. Bymeans of capacitive coupling, the signal is
connected symmetrically to the long-tail pair differential preamplifier. The preamplifier is usually
a three or four stage differential amplifier having a sufficiently large negative current feedback,
irom the end stage to the first stage, which gives a stabilizing effect. The amplified output signal is
picked up single-ended and is given to the power amplifier. The power amplifier is generally, of
the push-pull differentical type. The base of one input transistor of this amplifier is driven by the
preamplified unsymmetrical signal. The base of the other transistor is driven by the feeclback
signal resulting from the pen position and connected via frequency selective network The output
of the power amplifier is single-ended and is fed to the pen rr.otor, rvhich deflects the writing arm
on the paper. A direct rt'riting recorder is usually adequate since the ECG signal of interest has
limited bandwidth. Frequency selective network is an R-C network, u,hich provides necessary
damping of the pen motor and is preset by the manufacturer. The auxiliary circuits provide a 1 mV
calibration signal and automatic blocking of the amplifier during a change in the position of tire
lead switch. It may include a speed control circuit for the chart drive motor.
o
o
(,)
LlJ
!o
0)
5
roi tne neari.
ftur,:r-cn and Frequency
selective
' ini.- r::'.atlon feedback
lctr):] i Or an network
nca=eteriza-
;liolcc-,.
or quantihes,
erenie of m\' > Fig. $.t Block diagam of an ECG machine
ge is u,<uallr'
dd larthfullr- A 'stand by' mode of operation is generally provided on the electrocardiograph. In this mude,
re stallitr-of the stylus moves in response to input signals, but the paper is stationar.v. This mode allows the
mr beiir-een a operator to adjust the gain and baseline position controls without wasting paper.
d limitations Electrocardiograms are almost invariablv recorded on graph paper with horizontal and vertical
ence Lrf non- lines at 1 mm intervals with a thicker line at 5 mm intervals. Time measurements and heart rate
rc capable of measurements are made horizontally on the electi'ccardiogram. For routine work, the paper
Q unt,alance recording speed is 25 mm/s. Amplitude measurements are made vertically in millivoits. The
iallv adr-erse sensitivity of an electrocardiograph is typically set at 10 mm/mV.
rt hum due
lsolated Preamplifier:lthad been traditional for all electrocardiographs to have the right leg (RL)
I the contact
electrode connected to the chassis, and from there to the ground. This provided a ready path for
minrmumof
any ground seeking current through the patient and presented an electrical hazard. As the
microshock hazard became better understood, particularly when intracardiac catheters are
employed, the necessity of isolating the patient from the ground was stressed. The American Heart
Association guidelines state that the leakage current should notbe greater than 10 microamperes
tls picked up when measured from the patient's leads to the ground or through the main instrument grounding
{eckodes are
155 Handbook of Biomedical I nstrumentat ion
wire with the ground oPen or intact. For this, patient leads would have
ground for all line operated units.
to be isolated from ttre ftF
[n-
Figure 5'2 shows a block diagram of an isolation preamplifier used in
cardiographs. Difference signals obtained from the right arm left arm (LA)
modern electru
and rig6t leg IRL r
n
1RA1,
is pven to a low-pass filter' Filtering is required on thJinput leadsto
C
electrosurgery and radio frequency emissions and sometimes from the 50 kHz
reduce interference caused tn- ht
respiration detection. The filter usually has a cut off frequency higher than
current
"rJf; bfl
filter is needed to achieve a suitable reduction in high frequency s"ignal.
10 kHz. A multistage L-
r u6Pl
P".f
wQr,
qSmJ
RA
LA
LL
RL lsolated power
transformer
The filter circuit is followed by high voltage and over voltage protection
circuits so that the
amplifier can withstand_large voltages during defibrillation. Hoirever, the
price of this protection
is a relatively high amplifier noise level arising from the high series
resistance in each lead.
The lead selector switch is used to derive the required lead configurations
and give it to a
dc-coupled amplifier. A dc level of 1 mV is obtained by dividing down-the
powu, ,rp"ply, which >F
can be given to this amplifier through a push button foicalibratiJn
of the amilifier. Isoiauon of the
patient circuit is obtained using a low capacitance transformer whose primary
winding is driven
from a 100 kHz oscillator. The transforrner second.ary is used to obtain an
of t5 V for operating the devices in the isolated portion of the circuit and
isolated power supply Ttrp
modulator at 100 kHz, which linearly modulates an ECG signal given to it. The
to drive the slzrchronous mflrl
oscillator frequency sEe"TeF
of 100 kHz is chosen as a compromise so that reasonable size transformers (higher
the frequency eefirrgdd
the smaller the transformer) could be used and that the switching
inexpensive transistors and logic circuitry can be utilized. A square
time is not too fast, so that ffrtrh
wave is utilized to minimize tuqmofu
:_l
ii
li
ii
Biomedical Recorders "t57 i
Orfef,
I
eo that the
rlrotection
tlead.
trive it to a Common-mode rejectiofr mpiif ier
p\,u'hich > Fig. 5.S Imptooement in CMKR using ight
nirr of the leg drioe (Courtesy: Hewlett packard,
USA)
gisdriven
rcrsupplv
ldrronous ]heprlence of stray capacitance atthe input of thepreamplifiercausescorunon-mode currents
to flow in LA and RA, resulting in a voltage arop at ihe electrode resistors.
ftreguency An imbalance of the
stray capacitance or the electrode resistors causes a difference signal. This
frrequencv
difference signal can be
almost eliminated, in that the common-mode currents of stray-capacitances
sL so that are not allowed to
flow through the electrode resistors but are neutralized by currents delivered
rminimize to stray capacitances
from the common-mode rejection amplifier. In other words, the potentials
at A, B and C are
158 Handbook of Biomedical lnstrumentation
equalised through an in-phase component of the common-mode voltage, which the amplier
delivers via C, *a C, to LA and RA. As a result, the potentials at A, B and C are kept eou''
independent of the imbalance in the electrode resistors and stray capacitance.
The modern ECG machines with their completely shielded patient cable and lead u'ires a:r';
their high common-mode rejection, are sufficiently resistant to mains interference. However. tllee
could be locations where such interference cannot be eliminated by reapplying the electrods 'ar
moving the cable, instrument or patient. To overcome this problem, some ECG machines har-e ar
additional filter to sharply attenlate a narrow band centred at 50 Hz. The attenuation providei
could be up to 40 dB. In this way, the trace is cleaned up by the substantial reduction of iirc
frequency interference.
Isolation amplifiers are available in the modular form. One such amplifier is Model 274ttos
Analog Devices. This amplifier has the patient safety current as 1.2 pA at 115 V ac 60 Hz and ones
a noisJof 5 pV pp. It has u CUnn of 11rdB, differential input impedance of 1012 fl paralleled
rrif
3 pF ancl common mode impedance as 1011Q and a shunt capacitance of 20 pF' It is optimized ro:
signal frequencies in the range of 0.05 to 100 Hz. Metting van Riin ef al (1990) detail out methocs
for high-quality recording of bioelectric events with special reference to ECG.
Two electrodes placed over different areas of the heart and connected to the galvanometer n-ill
pick up the electiical currents resulting from the potential difference between them. For example,
if under one electrode a wave of L mV and under the second electrode a wave of 0.2 mV occur at the
same time, then the two electrodes will record the difference between them, i.e. a wave of 0.8 m\ '
The resulting tracing of voltage difference at any two sites due to electrical activity of the heart ts
called a"LE.LD" (Figs 5.4 (a)-(d)).
Bipolar Leads: In bipolar leads, ECG is recorded by using two electrodes such that the final frace
corresponds to the difference of electrical potentials existing between them. They are calied
standird leads and have been universally adopted. They are sometimes also referred to a-s
Einthoven leads (Fig. 5.4(a)).
In standard lead I, the electrodes are placed on the right and the left arm (RA and LA). In lead tr'
the electrodes are placed on the right arm and the left leg and in lead Iil, they are placed on the
left arm and the left leg. In all lead connections, the difference of potential measured betw'een
two electrodes is always with reference to a third point on the body. This reference point Ls
conventionally taken as the "right leg". The records are, therefore, made by using three eleckode:
at a time, the right leg connection being always present.
In defining the bipolar leads, Einthoven postulated that at any given instant of the cardiac
cycle, the electrical axis of the heart can be represented as a two dimensional vector. The ECG
measured from any of the three basic limb leads is a time-variant single-dimensional comPonent
of the vector. He proposed that the electric field of the heart couldbe rePresented diagrammaticallr'
as a triangle, with the heart ideally located at the centre. The triangle, known as the
"Einthoaat
triangle" ,is shown in Fig. 5.5. The sides of the triangle represent the lines along which the three
projections of the ECG vector are measured. It was shown that the instantaneous voltage measured
il-
Biomedical Recorders
159
tangLtler
Bipolar Limb Leads
Ept equC
rires and
rTtr. tlwe
drrodes clr
shal'ean
prorided
m of line
C-M means'bommon mode,
Zl4frorn
rrd orters
ded rr-ith
tdzed tor
rrethods
&r n-ill
nample,
curat dre Unipolar limb leads
f0-8m\'.
aheart is
ml hace
E celled
cd to as
Llead tr,
ilcr the Lead AVR--
Lead AVF**
Etrt'een
point is
:todes
eardiac
be ECG
lrcnent
aticelly
frprcn (b)
rtfuee > Fig. S.4. Typ,es of lead connections with typical ECG waaefonns (a)
Bud bipotar
limb leads (b) unip orar rimb reaitsibourtesy : Hezabh p
ackard,
rrsA)
160 Handbook of Biomedical Instrumentation
CH positions
LEAD LEAD
o @
roEn
miEtr
irr'l
in eil
XS EE€d
t-@
f;ti-of[
changE
rgiet
:lgher1
LEAD LEAD o-f LtrIiP
m:rrrdc
@ @ mr-nesP
qbtair*
Ieadsar
1D
LEAD LEAD
@ @
{E-}
Fig. 5.4 Types of lead connections with typical EcG waoeforms (c) position
- oi'tt e ihest lead in unipolar precordial lead rccording u) c leads
(C o utt e sY : H ew I ett P a ck ar d, US A)
Biomedical Recorders
161
> Fig. $.S The Einthoaen tiangle for itefining ECG leads
from any one of the three limb lead positions is approximately
equal to the algebraic sum of the
other two or that the vector sum of the projectioni on all
threeiines is equal to zero.
In all the bipolar lead positions, QRS of a normal heart is
such that the R wave is positive and
is greatest in lead II.
unipolat Leads (v Leads);The standard leads record the difference
in electrical potential between
two points on the body produced by the heart's action.
Quite oftery this vottage14,,itt show smaller
changes than either of the potentials and so better ,"r,Jtirrity
an be obtain"iif th" potential of a
single electrode is recorded. Moreover, if the electrode
is placed on the chest close to the heart,
higher potentials can be detected than normally available
at the timbs. This lead to the development
of unipolar leads introduced by wilson in rbg-+. In ,rrrr,j"*"r,t, the electrocardiogram is
recorded between a single exploratory electrode and_this
the centrfl terminal, which has a potential
corresponding to the centre of the body. In practice, the reference
electrode or central terminal is
obtained by a combination of several eiectrodes tied together at one point. Two
types of unipolar
Ieads are employed which are: (i) limb reads, and (ii)
piecordial reads.
(i) Limb leadsrnunipolar limb leads (Fig. 5.4@)), two of the limb leads
are tied together and
recorded with respect to the third hhb. In the lead identified
as AVR, *ru illit arm is
recorded with respect to a reference established by joining
the left arm and leit leg elec-
trodes' [:r the AVL lead, the left arm isrecorded with respecito
the common junction of the
right arm and left leg. In the AVF lead, the left leg is recorded
with respect to the two arm
electrodes fi ed together.
They are also called augmented leads or'averaging leads'.
The resistances inserted
between the electrodes-machine connections ure knowi as
'averaging resistances,.
(i) Precordial leadsThe second type of unipolar lead is a precordial
lead. It employs an explor-
ing electrode to record the potentiat or tne heart action on
the chest at slx different posi
oition tions' These leads are designated by the capital letter 'V'
followed by u ,rfr.ripi,,umeral,
bads which represents the position of the electiode on the
pericardium. rhe posiiions of the
chest leads are shown in Fig. 5.a(c).
Biomedical Signal Processing
(i.e., a gravitational force) exactly balances the If the valve is open to atmosphere, then thc
force of the atmospheric pressure. Torricelli found pressure on the mercury in the chamber is equal b
that the height of the mercury column that can be the pressure on the column (i.e., I atm). The rner-
supported by atmospheric pressure is approxi- cury in the column will have the same height ar
mately 0.76 m, or 760 mm. Atmospheric pressure, the mercury in the chamber, and this point is des.
then, is frcquently given in units of millimeters of ignated as a pressure of 0 mm Hg. If the valve h
mercury and I atm is 760 mm Hg.t closed, and the pressurc inside of the chamber ir
The pmper unit of pressure, as established by increasedby operating the pump, then the mercury
scientists and adopted by the National Bureau ol in the column will nse an amount pmportiorul b
Standards, is the torr (after Torricelli), in which I the increased pressure.
torr is equal to I mm Hg (l torr : I mm Hg).t
Gauge prcssures are usually given in millime- Example 9-3
ters of mercury above or below atmospheric pres-
The mercury in a manometer, such as the one ir
surc. A manomet€r is any device that measures
Figurc 9-2b, rises to a height of 120 mm Hg. Frnd
gauge pressure, although in commonly accepted
(a) the gauge pressure and (b) the absolute pressurE
jargon, gauge pressures below I atm are said to be
measured onavacuum gauge, and pressures above Solution
I atm are said to be measured onapressure gauge. (a) 120 mm Hg (by definition)
Both instruments are, however, examples of
manometers. Some manometers, including most of (b) 120 mm Hg * 760 mm Hg : 880 mm Hg
the electronic instruments discussed in this chap-
ter, measurc both positive pressures (i.e., above I We use gauge pressure because it is more easily
atm) and negative pressures (i.e., vacuums). referenced at zero and can be easily recalibrated *
Tlte zeru rcference in gauge pressure measure- each use, and the absolute pressure confers no spe-
ments, therefore, is a pressure of atm. Even I cial advantage as to information content. The vari-
though atmospheric pressure varies from one ation of atmospheric prcssure from one location to
place to another, and in the same location over the another (dependent upon mean sea level) and o\E
course of a few hours, zero can be established at the course of a single day makes the use of gauge
each measurement by setting the zero scale with pressures more advantageous,
the manometer open to atmospheric pressure. Fig-
urc 9-2b shows a mercury manometer similar to
those used to make blood pressure measurements. 9-6 Blood pnessure measurcments
The open tube is connected to a mercury reservoir The earliest recorded attempt at the measurcBd
that is fitted with a rubber squeeze-ball pump that of arterial blood pressure was performed in 1773 by
can be used to increase pressure; a valve is used English scientist Stephen Hales, who used an op*
to either open the chamber to atmosphere or close ended tube inserted directly into an artery in tlE
it off. neck of an rnanesthetized horse (presumably ticd
down securely). The tube was long enough thr
blood rose to a height at which the weight of thc
tThe barcmcter
rcading given in English units is usually inches blood exactly balanced the horse's arterial pressura
of mcrcury (in.Hil. 760 mm Hg = 29.92 in. HE
flf According to Hales's observation, the blood
convcntion werc followed, we would quote physiological
prEssucs in torr, but it is common practice in medicine to use
pulsed its way up the tube, attaining a height of
thc unit mm Hg. To avoid confusion, we witl follow this prac- approximately 4 ft on the first pulse but requiring
tice, but kccp in mind that it is no longer accepted outside the an additional 40 or 50 pulses to attain a final
medical world. height of just over 8 ft. After the blood in thc
Physiologicat Pressure and Other Cardiovascular Measurements and Devices 2s7
Cuff pressure
ll0
100
90
tr80
cE70
t60
9so
Zqo
&30
20
l0
0
Artery
Flgure 9.f
cuff placement for the auscultatory method of blood pressure measurement.
This pressure compresses the artery against the usually recorded in the ratio of systolic over
underlying bone, causing an occlusionthat diastolic (i.e., 120/80 mm Hg).
shuts off the flow of blood in the vessel.
The first use of sphygmomanometry for thc
3. The operator then slowly releases (i.e., reduces) measurement of blood pressure was reported by
the pressure in the cuff, shown in Figure 9-5a Korotkoff in 1905, but the technique was not vef,i.
(about 3 mm Hg/s is usually deemed best) and fied for correlation between indirect and dircc
watches the pressure gauge or mercury column. measurements in animals until 1912. It was not un-
When the systolic pressure first exceeds the til 1931 that a similar correlation was establishcd
cuff pressure, the operator begins to hear some for humans-that is, that variations of less than l0
crashing, snapplng sounds in the stethoscope mm Hg existed between direct and indirect m#
that are caused by the first jets of blood pushing ods. More recently it has been shown that indirec Flg
through the occlusion. These sounds, called diastolic pressures are less in error ifthe reading ir Dir
Korotkoff sounds (Figure 9-5D), continue as taken at the point where the Korotkoff sounds dir (4
the cuff pressure diminishes, becoming less appear. Yet most clinicians prefer to use the pdil
loud as the blood flow through the occlusion where the sounds become muffled because tir
becomes smoother. Korotkoff sounds disappear point can be recognized more consistently. Th dk:
or become muffled when the cuff pressure American Heart Association recommended ir tis
drops below the patient's diastolic pressure. 1967 that muffling be tised as the criterion for & cd
To read the blood pressure, the operator notes astolic pressure but that borfi pressures (i.e., muf- eil
both the gauge pressure at the onsel of fling and the cessation of Korotkoff sounds) be ir efi
Korotkoff sounds (systolic) and when the dicated if a significant difference between tlreo be
sounds become muffled (or disappear) exists. This measurement is recorded as a doubb
altogether (diastolic). These pressures are diastolic pressure (i.e.,120/80177 mm Hg). thc
Physiological Pressure and Other Cardiovascular Measurements and Devices 259
b0
a
o
o.
(a)
frc over
r60
llry for the Cuff pressure
;rcported by 120 E
with contractions; they are related to the spatial and temporal organization of gastric
contractions.
External (cutaneous) electrodes can record the signal known as the electrogas-
trogram (EGG). Chen et al. [38] used the following procedures to record cutaneous
EGG signals. With the subject in the supine position and remaining motionless,
the stomach was localized by using a 5 M H a ultrasound transducer array, and the
orientation of the distal stomach was marked on the abdominal surface. Three active
electrodes were placed on the abdomen along the antral axis of the stomach with
an inter-electrode spacing of 3.5 cm. A common reference electrode was placed
6 cm away in the upper right quadrant. Three bipolar signals were obtained from the
three active electrodes in relation to the common reference electrode. The signals
were amplified and filtered to the bandwidth of 0.02 - 0.3 Hz with 6 dB/octave
transition bands, and sampled at 2 Ha.
The surface EGG is believed to reflect the overall electrical activity of the stomach,
including the electrical control activity and the electrical response activity. Chen et
al. [38] indicated that gastric dysrhythmia or arrhythmia may be detected via analysis
of the EGG. Other researchers suggest that the diagnostic potential of the signal has
not yet been established [35,36]. Accurate and reliable measurement of the electrical
activity of the stomach requires implantation of electrodes within the stomach [39],
which limits its practical applicability.
near the right sternal border. The pulmonary area lies at the left parasternal line in
the second or third left-intercostal space [23].
A normal cardiac cycle contains two major sounds - the first heart sound (Sl)
and the second heart sound (S2). Figure 1.24 shows a normal PCG signal, along with
the ECG and carotid pulse tracings. S1 occurs at the onset of ventricular contraction,
and corresponds in timing to the QRS complex in the ECG signal.
-2
I
0.1
I
0.2
I
0.3
Y 'I
0.4
I
0.5
I
0.6
I
0.7
I
0.8
I
0.9
I
E
u0
:
I
2 0
0
-1 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Time in seconds
Figure 1.24 Three-channel simultaneous record of the PCG,ECG,and carotid pulse signals
of a normal male adult.
The initial vibrations in S1 occur when the first myocardial contractions in the
ventricles move blood toward the atria, sealing the atrio-ventricular (AV - mitral
and tricuspid) valves (see Figure 1.25). The second component of S 1 begins with
abrupt tension of the closed AV valves, decelerating the blood. Next, the semilunar
(aortic and pulmonary) valves open and the blood is ejected out of the ventricles.
The third component of S1 may be caused by oscillation of blood between the root
of the aorta and the ventricular walls. This is followed by the fourth component of
S 1, which may be due to vibrations caused by turbulence in the ejected blood flowing
rapidly through the ascending aorta and the pulmonary artery.
Following the systolic pause in the PCG of a normal cardiac cycle, the second
sound S2 is caused by the closure of the semilunar valves. While the primary
vibrations occur in the arteries due to deceleration of blood, the ventricles and atria
also vibrate, due to transmission of vibrations through the blood, valves, and the
valve rings. S2 has two components, one due to closure of the aortic valve (A2)
36 INTRODUCTION TO BIOMEDICAL SIGNALS
Figure 1.25 Schematic representation of the genesis of heart sounds. Only the left portion
of the heart is illustrated as it is the major source of the heart sounds. The corresponding
events in the right portion also contribute to the sounds. The atria do not contribute much to
the heart sounds. Reproduced with permission from R.F. Rushmer, Cardiovascular Dynamics,
4th edition, @W.B.Saunders, Philadelphia, PA, 1976.
EXAMPLES OF BIOMEDICAL SIGNALS 37
and another due to closure of the pulmonary valve (P2). The aortic valve normally
closes before the pulmonary valve, and hence A2 precedes P2 by a few milliseconds.
Pathologic conditions could cause this gap to widen, or may also reverse the order
of occurrence of A2 and P2. The A2 - P2 gap is also widened in normal subjects
during inspiration. (Note: The PCG signal in Figure 1.24 does not show the A2 and
P2 components separately.)
In some cases a third heart sound (S3) may be heard, corresponding to sudden
termination of the ventricular rapid-filling phase. Because the ventricles are filled
with blood and their walls are relaxed during this part of diastole, the vibrations of
S3 are of very low frequency. In late diastole, a fourth heart sound (S4) may be
heard sometimes, caused by atrial contractions displacing blood into the distended
ventricles. In addition to these sounds, valvular clicks and snaps are occasionally
heard.
Heart murmurs: The intervals between S 1 and S2, and S2 and S 1 of the next
cycle (corresponding to ventricular systole and diastole, respectively) are normally
silent. Murmurs, which are caused by certain cardiovascular defects and diseases,
may occur in these intervals. Murmurs are high-frequency, noise-like sounds that
arise when the velocity of blood becomes high as it flows through an irregularity
(such as a constriction or a baffle). Typical conditions in the cardiovascular system
that cause turbulence in blood flow are valvular stenosis and insufficiency. A valve is
said to be stenosed when, due to the deposition of calcium or other reasons, the valve
leaflets are stiffened and do not open completely, and thereby cause an obstruction or
baffle in the path of the blood being ejected. A valve is said to be insufficient when it
cannot close effectively and causes reverse leakage or regurgitation of blood through
a narrow opening.
Systolic murmurs (SM) are caused by conditions such as ventricular septal defect
(VSD - essentially a hole in the wall between the left ventricle and the right ven-
tricle), aortic stenosis (AS), pulmonary stenosis (PS), mitral insufficiency (MI), and
tricuspid insufficiency (TI). Semilunar valvular stenosis (aortic stenosis, pulmonary
stenosis) causes an obstruction in the path of blood being ejected during systole. AV
valvular insufficiency (mitral insufficiency, tricuspid insufficiency)causes regurgita-
tion of blood to the atria during ventricular contraction.
Diastolic murmurs (DM) are caused by conditions such as aortic or pulmonary
insufficiency (AI, PI), and mitral or tricuspid stenosis (MS, PS). Other conditions
causing murmurs are atrial septal defect (ASD), patent ductus arteriosus (PDA), as
well as certain physiological or functional conditions that cause increased cardiac
output or blood velocity.
Features of heart sounds and murmurs, such as intensity, frequency content, and
timing, are affected by many physical and physiological factors such as the recording
site on the thorax, intervening thoracic structures, left ventricular contractility, posi-
tion of the cardiac valves at the onset of systole, the degree of the defect present, the
heart rate, and blood velocity. For example, S 1 is loud and delayed in mitral stenosis;
right bundle-branch block causes wide splitting of S2; left bundle-branch block re-
sults in reversed splitting of S2; acute myocardial infarction causes a pathologic S3;
and severe mitral regurgitation (MR) leads to an increased S4 [40, 41, 42, 43, 441.
38 INTRODUCTION TO BIOMEDICAL SIGNALS
Although murmurs are noise-like events, their features aid in distinguishing between
different causes. For example, aortic stenosis causes a diamond-shaped midsystolic
murmur, whereas mitral stenosis causes a decrescendo - crescendo type diastolic -
presystolic murmur. Figure 1.26 illustrates the PCG, ECG, and carotid pulse sig-
nals of a patient with aortic stenosis; the PCG displays the typical diamond-shaped
murmur in systole.
Recording PCG signals: PCG signals are normally recorded using piezoelectric
contact sensors that are sensitive to displacement or acceleration at the skin surface.
The PCG signals illustrated in this section were obtained using a Hewlett Packard
-
HP21050A transducer, which has a nominal bandwidth of 0.05 1,000 Hz.The
carotid pulse signals shown in this section were recorded using the HP2 1281A pulse
transducer, which has a nominal bandwidth of 0- 100 Hz.PCG recording is normally
performed in a quiet room,with the patient in the supine position with the head resting
on a pillow. The PCG transducer is placed firmly on the desired position on the chest
using a suction ring and/or a rubber strap.
Use of the ECG and carotid pulse signals in the analysis of PCG signals will be
described in Sections 2.2.1,2.2.2, and 2.3. Segmentation of the PCG based on events
detected in the ECG and carotid pulse signals will be discussed in Section 4.10. A
particular type of synchronized averaging to detect A2 in S2 will be the topic of
Section 4. I 1. Spectral analysis of the PCG and its applications will be presented in
Sections 6.2.1, 6.4.5, 6.6, and 7.10. Parametric modeling and detection of S1 and
S2 will be described in Sections 7.5.2 and 7.9. Modeling of sound generation in
stenosed coronary arteries will be discussed in Section 7.7.1. Adaptive segmentation
of PCG signals with no other reference signal will be explored in Section 8.8.
pharynx, larynx, trachea, bronchi and bronchioles. The passage way bifurcates
to carry air into
each of the lungs wherein it again subdivides several times to irry
air into and out of each of the
many tiny air spaces (alveoli) within the lungs. In the tiny air spaces of the lungs
is a membrane
interface with the hydraulic system of the body through which certain
gases can defuse. oxygen
is taken into the blood from the incoming air and carbon dioxide
is transferred from the blood to
the air under the control of the pneumatic pump. Thus, the blood
circulation forms the link in the
supply of oxygen to the tissues and in the removal of gaseous waste products of
metabolism. The
movement of gases between blood and the alveolar air is basically due to constant
molecular
movement or diffusion from points of higher pressure to points of lower pressure.
. An automatic respiratory control centre in the brain maintains heart pump operation at a speed
that is adequate to supply oxygen and take away carbon dioxide as requlredty
the system. h each
minute, undernormal conditions, about250ml of oxygen are taken upand 256 d
of CO, are given
out by the body and these are the amounts of the two gases, which enter and leave
the blood in the
lungs. Similar exchanges occur in reverse in the tisiues where oxygen is given
up and CO, is
removed. The exact arrrount of CO, expired depends upon the metatolism,
the acid-base balance
and-the pattem of respiration' The exchange of gases takes place in the
alveoli and can be achieved
by the normal 15-20 breaths/min, each one involving abolt soo ml of air.
. The respiratory system variables which are important for assessing the proper functioning of
thesystem are respiratory rate,respiratory air flow, respiratory volu^L u.rd
.or,.entration of CO,
in the expired air. The system also requires measurements to be made of certain
volumes and
capacities such as the tidal volume, vital capacity, residual volume, inspiratory
reserve volume
and expiratory reserve volume. The details of these are given in Chaptei 13.
(Fig. 1.4) made up of the encephalon (brain) and the spinal thu peripheral nervous system
comprises all the nerves and groups of neurons outside the "oid.
brain and the spinal cord.
The brain consists of three parts, namely, the cerebrum, cerebellum and, thebrain
stem.
Cerebrum: The cerebrum consists of two well demarcated hemispheres, right and left and each
hemisphere is sub-divided into two lobes:/rontal lobe and temporaliobein the"lefthemisphere
and
parietal and occipital lobes inthe right hemisphere (Fig. 1.5). The outer layer
of the brain is called the
cerebral cortex. All sensory inputs from various parts of the body eventually
reach the cortex, I
where certain regions relate specifically to certain modalities of sensory inftrmation. me[
Variou-s
areas are responsible for hearing, sight, touch and control of the
voluntai muscles of the bodr,. silur
F undamentals of Medical lnstrumentation
arn'air into
*each of the
lmembrane
use. Oxygen
ttrcblood to tr
-e
lElinkinthe (,
tbolism. The o
I molecular
UJ
rrat a speed
fui.L:reach
Qaregiven Posterior
rHoodinthe nerve
t and COz is roots
hasebalance Cervical
Spinal
rbe achieved spinal
nerve
cord
rrtioning of
ation of CO,
olumes and
:fffe Thoracic
"rOlUme spinal
cord
E
o
o
E
'a
ndinates the
a
he effective,
imuscle are
nloped and Lumbar
information spinal
cord
retof input
us system is Sacral
spinal
,u.rs system cord
L
l
dtand each
bphere and > Fig. 1.4 Centtalneruous system,humanbrain and spinal coril
iscalled the
ttre cortex, The cerebral cortex is also the centre of intellectual functions. The frontal lobes are essential for
cu Various intelligence, constructive imagination and thought. Here, large quantities of information can be
f thebody. stored temporarily and correlated, thus making a basis for higher mental functions.
L0 Handbook of Biomedical lnstrumentation
Cerebral cortex
Parietal lobe
Thalamus
Corpus callosum
Hypothalamus
Occipital lobe
Frontal lobe
Pituitary
Temporal lobe
Bnri:
Cerebellum Jer'&:
iectiu
..
a500
,;erati[
Medulla oblongata
r
=ct-il
> frig. 1 .S Cut-azuay section of the human byain :g"*q
rrhich
Each point in the motor centre in the cerebral cortex (Fig. 1.6) corresponds to a the Io
certain bodv
movement. In the anterior part of the parietal lobe lies the terminal station for the nerve pathways temfI
conducting sensation from the opposite half of the body. The sensory centre contains tt.t.rst-
counterparts
of the various areas of the body in different locations of the cortex. 'ih" ,"nro.y inputs
come from Sfirr,
the legs, the torso, arms, hands, fingers, face and throat etc. The amount of surface
allotted to each the Ia
part of the body is in proportion to the number of sensory nerves it contains rather than its
physical size. The visual pathways terminate in the posterior part of the occipital lobe. The rest
actua] tftIei
of andp
the occipital lobes store visual memories, by mear,sbf rrhich we interpret what we
see. Thecr
On the upper side of the temporal lobe, the acoustic pathways terminate making it as a hearing
hrd]-,
centre. This is located just above the ears. Neurons responding to different frequJncies
of sound contr
input are spread across the region, with the higher frequencies located towards the front and iorv
a resu
frequencies to the rear of the ear. The temporal lobes are also of importance for the storage
process h-isLle
inthelong-terrnmemory.
Th
Cerebellum: The cerebellum acts as a physiological microcomputer which intercepts various neurc
sensory and motor nerves to smooth out the muscle motions which could be otherwise jerky. It nainl
also
consists of two hemispheres which regulate the coordination of muscular movementi triurst
elicited br-
the cerebrum. The cerebellum also enables a person to maintain his balance. sirnu
tr: all
11
F undamentals of Medical lnstrumentation
Motor
ciital lobe
> Fig. 1.6 Sites of some actiaity centres inthe cercbral cortex
The basic functional unit of the nervous system is the neuron. A typical neuron consists of a
nucleated cell body and has several processes or branches (Fig. 1.7). The size and distribution of
these branches vary greatly at different sites and in cells with different functions, but the two main
kinds are: the qxone and the dendrite. The dendrites normally conduct impulses toward the cell il
F
body and the axons conduct away from it.
Cell body
I
;
&
F
lmpulse transmission
12-1 Objectives
1. Be able to introduce the biological principles and spinal cord), PNS (nerve pairs), and ANS
underlying the human central nervous system (sympathetic and parasympathetic systems).
(CNS), the peripheral neryous system (PNS), 3. Be able to identify specific areas of the brain
and the autonomic nervous system (ANS). concerned with bodily sensory and motor
2. Be able to describe the structure and function functions.
of the neuron (single nerve cell), CNS (brain
3. Describe nerve impulse conduction through 9. What is the purpose of the three membranes
one neuron and several neurons connected (meninges) covering the brain?
together.
10. Does human behavior control brain function,
4. Describe the basic structure and function of or does brain function control human
the CIVS, PNS, and ANS. behavior?
367
368 CHAPTER 12
NERVOUS SYSTEM
Thc Nervos System ir concerncd with thc INTEGPATXSN and @NTROL
of ell bodilv functions.
It hae ipccielizcd in IR,RITABILITY- cha abi/rty 6nccivc tad4TssljP
and arso i.
"ofi'6ffiii3t
f;7f:fii{lhi#;:ffi*
end troi coVFDNTrrtrc cc'ur*s.
linkcd by en
artlying
or
I
TISSUES end ORGANS
MOTOR Nervc fibrcs c.rrv m.as.qr!
to Ti3suc. end Oroanifrom t'hc
of thc body Brain or- Spinal Conj.
tfu@i./
crll, forkiniccrlt.
Hffi,
CERESR.AT CEEEEEtU'M
conTEX
(dprfutent
(,tlylhatca) (l,bt.ayc/tnated) Autlortoark
cbh)
_..-. AXON... \..
:::.....MYELrN
tpoE ot
NANVIER
-
-. SCHWANN -
onecrtar I '. CELr
ol l,ltfutSE I NUCLEUS
cot{NECTtvz
"---TlssuE sxcerx._
.----tlcrye Endtngs--
S&clct l Smooth-.
,nuaclc
lvlxC mutltpola? ncutrcncc
alrc *t'24&;AP,T
,tmr nnction.
12.2
McNausht & catlander, tttustrated physiotosltUsed
m, ffi;lll"J by permission of
370 CHAPTER 12
involves response to environmental stimuli, while fiber. Figure l2-2 shows a nerve cell
perception relates to the recognition of symbolic skeletal muscle and to smooth muscle.
patterns. These patterns are abskact (not physical) the nerve impulse travels in one dfuc(i
and are composed of linguistic symbols. The abil- from dendrites to nerve endings. The a.rc
ity of the human to think of the physical world in the entire length of the nerve cell, ard
abstract terms accounts for the extraordinary talent sunounded by a myelin sheath (segffi
to manipulate objects, construct houses, and con- lating covering). The neurilemma
trol his or her environment. this sheath and is composed of Sclrxm
Nodes of Ranvier act to speed up the
pulse transmission.
12-4 The neuron (single nerue cell)
The rransrn ission of the nene impukc's
Tlte neuron is the fundamental unit of the nervous a result of biochemicals that travel
system. It is a single cell composed of a cell body synapse (space between nerve cells). The
(soma), several short input projections (dendrites), membrane permeability is the chief reason
and a long propagation channel (axon). The axon pulse transfer across the synaptic junctim
together with its sheath (covering) forms the nerve l2-3 shows the synaptic ultrastructure.
Flgure 12-3
Synaptic ultrastructure.
The Human Nemous SYstem 3'71
alrangements:
nect to each other in the following
E and PostsynaPtic membranes and one to one, one to many'many to one' or. many to
rI "fr#
wwilz?'r-wz*th-ffi
fr\,qg::t- -"'-.H#
Wffim,ffi
g*{c*flf*-,zr,Ma
hner-tblecll
sy'itl caa.
Thc largcst part of thc human brain is the CEREBRUM - madc up ol I r-:m
2 CEnEBPAL-HEMISPHERES Each of thase is dividcd into LOBEs. rrgrlh
lNlTlATlN6 CENTRES for RECEIVING CENTRES for r[ .{ m&
OUTGO,NG mcstagcs INCOMING information
I qr4!i
mollllt!
rql6{
t. r*xrs
@iruf,m
r2-5 S
c{
lme r.am
ls;irc ,3:
::!9il4.!.&1r|["
rrrxli-lr Fl
jmrm-flu
Laroe uncharted arcas of the Ccrebral Hemispheres IrrrHm!l,(
are'probably conccrncd with MENTAL PROCESSES
such as /nte//tbcncc, Memory, Judgcmcnq
/mcgthatrbn,- Creativc and Consc,ous ThoughA
The surface of thc brein shows many folds or
CONVOLUTIONS. This has Lhe eflect ol -u ::@
incrcasino thc.mounc of GREY MATTER Prcscnt. mcl :ori.
Thc GREY'MATTER, forms thc outcr layer or r=i lmEE
CORTEX. lt contains thecell bodios ot thc lma !ri:t[
NEURONES arranqcd in many intcrconnccting trr4ur:3: fi
layero to form a 3'dimcnsional network.
uef -i]:ri l
,About 90% of all Ncrve Cclls are
in thc Cerebral Gortex. iu:ur;r 1a
r 0a :[]iflir
Flgure 12{ lmro la
The cerebrum. (From McNaught & Callander, lllustrated Physiology. Used by permission ol 7'm-; fr
Churchill Livingstone.) Fnlir:conu
372
The Human Nervous SYstem 313
imuri :€ required to conduct before a triggering tion, and the cerebrum--voluntary movement'
is reached. The AND and OR functions
tM*,nr:,-nl. sensation, and intelligence.
mrr:;r--neC by nerve cells act to control bodily
,lcilt$rJ1a-:;1u on and refl exes. 12-5-1 Brain stem
lrcx action in the human involves many reflex
lmil- {
rervous reflex is an involuntary action re- Tlte brain sfern consists of the medulla (oblon-
mnr:n:.c ;aused by stimulation of an afferent nerve
gata), pons, midbrain, and diencephalon'
ullll::x or receptor. The knee jerk in response to The medulla automatically controls heart rate
iltur ;: .:ri a hammer is one example of a reflex arc, and breathing. (Actually, most essential life sys-
tems are controlled here.) Reflex functions such
as
fum ::,mponents of which are shown in Figure
; : ltrc components of the arc are: coughing, sneezing, and vomiting are associated
witli the medulla. Indeed, one modern definition of
- '€ceptor, which detects change. clinical death is the absence of lower brain EEG
\ '.:" efferent neuron, which conducts the nerve
activity.
i:;ulse from the sensory area to the CNS' The pons is about 2.5 cm long and forms a no-
ticeable bulge on the anterior surface of the brain
*. :enter or synapse, which connects neurons
stem. It Ttrnction.t as a relay station for motor res-
" ,:"eetler. piratory and auditory fibers from the cerebrum and
{ \ :rain processing area. cerebellum. Other impulses from eye movement'
head muscles, and taste sensors also pass through
'. ',:- efferent neuron, which conducts nerve here.
::ulses from the CNS to an organ for The midbrain is a wedge-shaped portion of the
1;f,roPriate response. stem. Midbrain tissues function as a motor relay
:. \: effector or organ, which resPonds to station for fibers passing from the cerebrum to the
:-a:ntain homeostasis. cord and cerebellum. Integration of visual and au-
ditory reflexes, including those concemed with
avoiding objects, also occur here.
'2-5 Structure and function of the T\e diencephalon forms the superior (top) part
central nervous system* of the brain stem. As part of the original forebrain'
*:r brain is defined as "a large sott mass of nerve it develops into the thalamus and hypothalamus'
: .,*e contained within the cranium' lhe en' T\e thalamus receives fibers from the hearing
strucrures of the inner ear and visual system' It
r,'.t!on."t Three major structures compose the
also provides Pathways for somatic sensory sys-
:';.,: (Figures l2-6 through l2-9)'. the brain
All this sensory information eventually
tems-
ini'.F,-automatic vital system control, the cerebel'
reaches the cerebrum, where it is processed'
a--involuntary muscle control and coordina-
to the properties of
The hypothalaLnus responcls
blood passing through nerv'e connections' The en-
"-:c central nervous system is composed of the brain and docrine systen is controlled through nerve re-
i;::a cord, Many excellent drawings and photographs have- sponses, affecting emotional behavior Patterns'
rr by anatomists. Yokochihas excellent pictures of
o,resenred
Other functions controlled via chemical interaction
u::rurl brain tisiue. Frank H. Netter, Ciba Pharmaceutical
picture Presenu- with the pituitary gland we temperature regula-
'*.crcrs, lnc., has a superb l3-minute motion
-;c irom the Department of Anatomy, U'C'L'A' Medical don, water balance, food intake, gastric secretion'
.uLr:l lTeaching Films, Houston, Texas, Clemintioe and Hard- sexual behavior, and sleeping pattems.
n,:r enlitled Guides to Dissection, The Cranial Caviry- As a general arousal mechanism, the reticular
),rtal
-:lvr's
of the Brain.
F'A' Davis Co' activation system (MS) functions with the thala-
Cyclopedic Medical Dictionary' of
;! lidelphia, 1973). mus to Prepare the cerebral cortex (higher parts
l"t4 CHAPTER 12
'.,-At
i ';;;rtii;&
vtt)tdla
dsdotc. \. \