DM3068 UserGuide
DM3068 UserGuide
DM3068 UserGuide
User’s Guide
Feb. 2014
Copyright
© 2010 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UGC06105-1110
Notices
RIGOL products are protected by patent law in and outside of P.R.C.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding material.
RIGOL shall not be liable for losses caused by either incidental or
consequential in connection with the furnishing, use or performance of this
manual as well as any information contained.
Any part of this document is forbidden to copy or photocopy or rearrange
without prior written approval of RIGOL.
Product Certification
RIGOL guarantees this product conforms to the national and industrial standards
in China as well as the ISO9001:2008 standard and the ISO14001:2004 standard.
Other international standard conformance certification is in progress.
Contact Us
If you have any problem or requirement when using our products or this manual,
please contact RIGOL Technologies, Inc.
E-mail: service@rigol.com
Website: www.rigol.com
Safety Requirement
Please review the following safety precautions carefully before putting the
instrument into operation so as to avoid any personal injuries or damages to the
instrument and any product connected to it. To prevent potential hazards, please
use the instrument only specified by this manual.
Electrostatic Prevention.
Operate in an electrostatic discharge protective area environment to avoid
damages induced by static discharges. Always ground both the internal and
external conductors of the cable to release static before connecting.
Handling Safety.
Please handle with care during transportation to avoid damages to keys, knob and
interfaces as well as other parts on the panels.
The disturbance tests of all models conform to the P/F values of A based
on the standard of EN 61326: 1997+A1+A2+A3 instead of P/F values of
B.
Note:
In order to prevent the fuse from blowing out and protect the
multimeter, please use the current input terminals according to the
following requirements:
1) Do not connect the 10 A and LO Sense/200 mA input terminals into
the current measuring circuit at the same time.
Note: voltages above 300 VAC may be measured only in circuits that are isolated
from mains. However, a transient overvoltage is also present in such circuits.
DM3068 was designed to safely withstand occasional transient overvoltage up to
2500 Vpk. Do not use this device to measure circuits whose transient overvoltage
may exceed this level.
WARNING
Warning statements indicate the conditions or practices that could result in
injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in
damage to this product or other property.
CAT I (1000V)
IEC Measurement Category I. The maximum voltage can be measured by
HI-LO terminal is 1000 Vpk.
CAT II (300V)
IEC Measurement Category II. Inputs may be connected to mains (up to
300 VAC) in the case of overvoltage in Category II.
General Care:
Do not store or leave the instrument in where the instrument will be exposed to
direct sunlight for long periods of time.
Cleaning:
Clean the instrument regularly according to its operating conditions. To clean the
exterior surface, perform the following steps:
1. Disconnect the instrument from all power sources.
2. Clean the loose dust on the outside of the instrument with a lint- free cloth
(with a mild detergent and water). When clean the LCD, take care to avoid
scarifying it.
CAUTION
To avoid damages to the instrument, do not expose them to liquids which
are corrosive.
WARNING
To avoid injury resulting from short circuit, make sure the instrument is
completely dry before reconnecting into a power source.
Environmental Considerations
The following symbol indicates that this product complies with the applicable
European Union requirements according to Directives 2002/96/EC on waste electrical
and electronic equipment (WEEE) and batteries.
DM3068 Overview
Main Features:
Real 6½ digits reading resolution.
Minimum integration time: 0.006 PLC.
Dual display: synchronously display two kinds of characteristics of the same
signal.
Switch between “Normal” and “Preset” operation modes. The “Preset” mode
can quickly recall the preset configurations.
DC voltage measurement range: -1050 V to 1050 V.
DC current measurement range: -10.5 A to 10.5 A.
AC voltage measurement range: True-RMS, 0 V to 787.5 V.
AC current measurement range: True-RMS, 0 A to 10.5 A.
Resistance measurement range: 0 Ω to 110 M Ω; support 2-wire (2WR) and
4-wire (4WR) resistance measurements.
Capacitance measurement range: 0 F to 110 mF.
Frequency measurement range: 3 Hz to 1 MHz.
Continuity and Diode tests.
User-defined any sensor measurements and three types of temperature
sensor measurements: TC, RTD and THERM.
Plenty of math operations such as STA (MAX, MIN, AVG, ALL), P/F, dBm, dB,
REL as well as real-time tendency graph and histogram display functions.
Enable to store data and configuration in USB storage device.
Plenty of interfaces: USB Device, USB Host, GPIB, RS232 and LAN.
Support the command sets of RIGOL DM3068, Agilent 34401A (including
some extensions) and Fluke 45.
Two power management modes: enable or disable the front panel power
switch.
Enable to store 10 groups of system configurations and 5 groups of sensor
configurations in internal memory and recall them when necessary.
Configuration clone: back up all the instrument configurations in a USB
storage device or clone the configurations to other DM3068.
Provide English/Chinese menus and built-in help system.
Powerful remote control and any sensor editing software.
Document Overview
Chapter 1 Quick Start
This chapter guides users to quicky get familiar with the front and rear panels, user
interface and measurement connections of the multimeter.
Chapter 4 Troubleshooting
This chapter lists the failures that might occur when using the multimeter as well
as their solutions.
Chapter 6 Specifications
This chapter lists the specifications and characteristics of the multimeter.
Chapter 7 Appendix
This chapter provides the accessories information as well as the service
information of the multimeter.
Operation step:
“” is used to indicate operation steps. For example, “ REL” means
pressing and then pressing REL.
Contents
DM3068 Overview.................................................................................. IX
Document Overview............................................................................... XI
Index ....................................................................................................... 1
This chapter guides users to quicky get familar with the front and rear panles, user
interface and measurement connections of the multimeter.
General Inspection
Handle Adjustment
Appearance and Dimensions
Front Panel
Rear Panel
User Interface
First-use of the Multimeter
Measurement Connections
To Use the Built-in Help System
To Use the Rack Mount Kit
General Inspection
The consigner or carrier shall be liable for the damage to the instrument
resulting from shipment. RIGOL would not be responsible for free
maintenance/rework or replacement of the unit.
Handle Adjustment
To adjust the handle position of the multimeter, grip the handle by both sides and
pull outward, then rotate the handle to a desired position, see figure below.
231.6
107.0
290.5
Front Panel
1. USB Host
USB storage device can be connected via this interface. By using this interface,
the system configuration or measurement data can be stored into USB
storage device and be recalled when required.
2. LCD
The instrument provides a 256 x 64 monochrome LCD that can display the
current function menus, measurement parameter settings, system status,
prompt messages and so on.
5. Power Button
Press the button to turn on or off the multimeter. Users can enable or disable
this key as required.
Press System Cfg Switch and select “ON” or “OFF”.
Provide auto, single, external and level trigger; enable to set the
reading hold function, number of samples per trigger, delay time
before reading and edge of the trigger input signal as well as the
trigger output parameter.
Provide help information for common operations and the use method
of the built-in help. The multimeter allows users to quickly recall the
help information about any front panel keys and menu softkeys.
9. Range/Direction Keys
Enable auto range.
Rear Panel
Note: the large current fuse stands inside the multimeter and is not allowed to
be replaced by users themselves. If a new fuse is required, please contact
RIGOL.
2. LAN
Through this interface, the multimeter can be connected to the network for
remote control. The multimeter conforms to LXI-C standards; it can be used in
connection with other standard devices to quickly build a testing system and
easily realize LAN based system integration.
3. GPIB
Conform to IEEE-488.2 standards.
4. Power Socket
The multimeter accepts two types of AC supplies. Please use the power cord
provided in the accessories to connect the multimeter to the AC power
through this socket.
Note: a proper voltage scale must be first selected (through the Voltage
Selector) before power connection.
5. VMC
The mutlimeter outputs a low-true pulse from the [VM Comp] connector after
every measurement when VM output is enabled ( VMC ON).
6. Ext Trig
Trigger the multimeter by connecting a trigger pulse through the [Ext Trig]
connector. Note the external trigger source must be selected ( Source
Ext).
7. USB Device
Connect the PC through this interface. You can use SCPI commands or PC
software to control DM3068 remotely.
8. RS232
Connect the PC through this interface. You can use SCPI commands or PC
software to control DM3068 remotely. You can also use this interface to output
the P/F test result.
9. Power Fuse
The multimeter is already installed with a power fuse before leaving factory. To
replace a new one, please:
1) Turn off the multimeter and remove the power cord.
2) Press down the block tongue using a straight screwdriver (in the direction of
the dotted arrow in the figure below) and pull out of the fuse seat.
3) Select a proper voltage scale.
4) Replace a specified fuse.
5) Reinstall the fuse seat into the slot.
WARNING
In order to avoid electric shock or fire, please just use the specified fuse
and make sure the fuse holder is not short-circuited.
User Interface
STA
MAX
Math MIN
AVG
P/F
dBm
dB
Measurement parameter LXI: LXI connection
Range REL on Mode succeeds
Rmt: Remote Control
Main Display Local: Local Control
Function : USB Storage
Run Vice Display Reading
Indication Vice Display Function
Main Display Operation Menus
Reading
Start your multimeter according to the following steps when using it for the first
time.
3. Start-up Process
1) Normal start: the multimeter executes self-test and then enters the user
interface.
2) Start with a USB storage device containing update file inserted: the
multimeter directly updates if an update file is detected and then starts.
Measurement Connections
DM3068 is designed with many measurement functions. After selecting the desired
measurement function, please connect the signal (device) under test to the
multimeter according to the method below. Do not discretionarily switch the
measurement function when measuring as it may cause damage to the multimeter.
For example, when the test leads are connected to the related current terminals,
AC voltage measurement should not be taken.
+ +
DCV ACV
- -
- DCI/ACI +
+ DCI/ACI -
+ +
2WR 4WR
- -
Capacitance Measurement
+
Capacitance
-
+ +
Open or Diode
Closed Circuit -
-
+ +
AC Signal Sensor
- -
+
Sensor
-
+ Sensor -
Note*: before setting the parameters for DCI sensor when using it, please connect it according
to the figure above and press mA.
The built-in help system of DM3068 provides using help for all the front panel keys
and menu softkeys as well as the using method of the built-in help. Press to
enter the following interface.
Help Topics
1. The method of the test leads.
2. Common measurement.
3. Math.
4. Sensor.
5. Store and recall.
6. Utility.
7. I/O interface.
8. Online help.
9. To change the power fuse.
10. Support.
DM3068 can be mounted into a standard 19-inch rack cabinet. Before the
installation, please remove the package and cushioning material of the multimeter.
Installation Tool
Installation Space
Installation Procedures
1. Remove the handle: grip the handle by both sides and pull outward; then, pull
upward.
2. Install left and right plates: aim the detents of right and left plates at the
openings on support board and insert respectively, then fix them using eight
M4 screws.
3. Place the instrument: aim the parallels at the corresponding openings and
then put the multimeter onto the support board.
4. Fix the instrument: fasten or fit the instrument tightly into the support board
using two fixed fingers in connection with four M4 screws.
5. Mount the front panel: aim the front panel opening at the front of instrument
and fix them using four M4 screws.
6. Load into machine cabinet: mount the rack with instrument onto a standard
19-inch machine cabinet using four M6 screws and four M6 square nuts,
respectively.
7. Post-installation note: the rack holds a height of 3U. Holes pointed out by the
arrows are the installation holes and should be aligned with when installing.
This chapter introduces how to use the functions of the multimeter from the front
panel.
DM3068 provides auto and manual range selecting modes. In auto mode, the
multimeter selects a proper range automatically according to the input signal; in
manual mode, you can use the front panel key or menu key to set the range. The
auto mode can bring a lot of convenience for users while the manual mode
provides higher reading precision.
Note:
“OVER LOAD” will be displayed when the input signal exceeds the current
range.
By default, the range is set to Auto at power-on or after a reset.
Auto mode is recommended if you are not sure about the measurement range
in order to protect the instrument and obtain accurate data.
In FREQ/PERIOD measurements, the multimeter uses different gate times to
include all input signals within 3 Hz and 1 MHz.
The range of CONT measurement is fixed at 2 kΩ; the range of DIODE
measurement is related to the current source currently selected.
1. In DCV, DCI and OHM measurements, press the left and right direction keys at
the front panel ( ) to set the reading resolution to 51/2 or 61/2 digits. The
resolution affects the integration time and vice versa.
2. In ACV and ACI measurements, the resolution is fixed at 61/2 digits. Press the
left and right direction keys to set the filter type.
To Measure DC Voltage
Operating Steps:
2. Make connection
Connect the test leads with the measured signal by referring to
“Measurement Connections”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
To Measure AC Voltage
Operating Steps:
2. Make connection
Connect the test lead with the measured signal by referring to
“Measurement Connections”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
To Measure DC Current
In order to obtain more accurate measurement results, DM3068 deals with small
and large input currents separately. The multimeter will use small current
measurement mode for current within 200 mA and use large current measurement
mode for current that is equal to or higher than 2 A.
Operating Steps:
2. Make connection
Connect the test lead with the measured signal by referring to
“Measurement Connections”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
To Measure AC Current
In order to obtain more accurate measurement results, DM3068 deals with small
and large input currents separately. The multimeter will use small current
measurement mode for current within 200 mA and use large current measurement
mode for current that is equal to or higher than 2 A.
Operating Steps:
2. Make connection
Connect the test lead with the measured signal by referring to
“Measurement Connections”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
To Measure Resistance
Operating Steps:
2. Make connection
Connect the test lead with the measured resistance by referring to
“Measurement Connections”.
Tip
Both ends of the measured resistance should be placed far away from your
hands and desks that can conduct electricity; otherwise, the measurement
result might be inaccurate. The greater the measured resistance is, the
greater the affect will be.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
To Measure Capacitance
Range: 2 nF, 20 nF, 200 nF, 2 μF, 20 μF , 200 μF, 2 mF, 20 mF, 100 mF
Max Resolution: 1 pF (in 2 nF range)
Input Protection: a 1000 V input protection is available in all ranges. 10%
overrange is available on all ranges.
Operating Steps:
2. Make connection
Connect the test lead with the measured capacitance by referring to
“Measurement Connections”.
Tip
Please short contact the two feet of an electrolytic capacitor by using a test
lead before measuring the electrolytic capacitor.
pre-specified value of REL from the actual measurement result and displays
the result. For details, please refer to “Math Operations”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
To Measure Continuity
Operating Steps:
2. Make connection
Connect the test lead with the measured circuit by referring to
“Measurement Connections”.
To Measure Diode
Operating Steps:
2. Make connection
Connect the test lead with the measured diode by referring to
“Measurement Connections”.
Remark*: this function is added since the 01.01.00.01.07.00 version but only the performance
of the 1 mA current source is guaranteed.
To obtain the measured signal’s frequency and period, directly press or use
the secondary measurement function during voltage or current measurment.
Operating Steps:
2. Make connection
Connect the test lead with the measured signal by referring to
“Measurement Connections”.
Press Info and use the up and down direction keys to view the
measurement item, measurement range, number of values, maximum,
minimum, average and standard deviation.
Press List and use the up and down direction keys or the menu softkeys
to view all the measurement results before the last update in table form.
Press HistoG to view the histogram of the average (AVG) and standard
deviation (SDEV).
Press Update to update the “Records”, “Maximum”, “Minimum”, “Average”
and “SDEV” under Info and the display in List and HistoG.
Press Save to enter the store and recall interface; at this point, is
illuminated. You can save the current measurement data into internal
memory or external USB storage device. For details, please refer to the
introduction in “Save and Recall”.
This function enables you to easily convert a measured physical quantity (such as
pressure, flow rate and temperature) into a quantity that can be easily measured
(such as voltage, current or resistance). By pre-inputting a response curve, the
multimeter converts and amends the data according to the internal arithmetic and
displays the physical quantity measured on the screen. Users can edit and modify
the display unit of the measured physical quantity at will.
DM3068 supports user-defined (DCV, DCI, 2WR, 4WR, FREQ) and temperature
(TC, RTD, THERM) sensor measurements. Press to enter the following
interface.
The following part introduces how to create a sensor configuration file in details.
For DCV, DCI, 2WR, 4WR or FREQ sensor, you can define a fitting curve. Since the
temperature sensor (TC, RTD, THERM) converts the electrical signal (such as
voltage or resistance) to temperature using the software conversion program
inside the instrument, you can directly choose a desired temperature sensor.
User-defined Sensor
File Name
Input Area
Character
Selecting Area
Press the up or down direction key to place the cursor in the File Name Input
Area or Character Selecting Area.
When the File Name Input Area is selected, the Character Selecting Area is
hidden automatically and you can select the cursor position to be edited by
pressing the right and left direction keys (or press and hold the key).
In Character Selecting Area, choose the desired character using the right
and left direction keys. At this point, the character in the File Name Input
Area changes accordingly (after choosing the desired character, select the
File Name Input Area and press the right direction key to confirm the
character selected and move the cursor to the next position).
Uppercase letter, lowercase letter and number can be use for defining the name.
A: input uppercase letters (A-Z).
a: input lowercase letters (a-z).
1: input numbers (0-9).
Press Meas and enter the data of the physical quantity measured using the
direction keys. Here, as the sensor type is 2WR, the unit is Ω.
Press Corrsp and enter the data of the corresponding physical quantity using
the direction keys. The unit to be used is the one (℃) you have set in step 3.
Press Done to finish the input of a pair of data and enter the following interface.
The icon indicates the arithmetic for the data segment behind this data and
before the next arithmetic mark. The default arithmetic for the first data
segment is Line and you can modify it using Edit.
Start Mark of the Current Data Segment
Arithmetic of the
Current Data Segment
Press to return to the upper-level menu and create the second group of
data in the same way. From the second group of data on, you can use the
segment function to set the arithmetic for different data segment.
Press SEG ON Arith and select Line or Curve arithmetic type.
Tip
For some sensor, the corresponding relation curve might both contain data
segment with good linearity (Line arithmetic is recommended) and data
segment with poor linearity (Curve arithmetic is recommended); in this case,
you need to use the segment function to define different arithmetic for
different data segment.
Press to finish the sensor property setting and returns to the User
interface. Then, press Done to enter the following interface.
Press Save, the multimeter enters the store and recall interface and the
backlight of goes on.
When finished, press Done to save the current sensor configuration into
the internal memory (C:\) or external USB storage device (A:\) with the
specified filename.
If you suspect the stored any sensor configuration files or data, press Edit to
make a modification.
Temperature sensor
Type: choose a desired temperature sensor type from TC, RTD and THERM
and set the related parameters.
Unit: choose the unit of the measured temperature from ℃, ℉ and K. The
conversion relations between these units are:
℉ = (9/5) * ℃ + 32
K ≈℃ + 273.15
Done: save or apply the current temperature sensor configurations.
Press to return to the upper-level menu.
For the temperature sensor measurement, connect the test lead and device under
test by referring to the connection method in “Measurement Connections”.
TC
The DM3068 provides two cold junction compensation modes: internal cold
junction temperature & user-defined cold junction temperature.
To use the TC sensor, you should select the type of the thermocouple and the
compensation mode of the cold junction.
Press TC Type to select the desired thermocouple type and press .
Press TC JUNC, select INT or SIM to input correct cold junction temperature
(with -273 and 999) and then press .
RTD
To use the RTD sensor, set the measurement parameters according to the
following steps.
THERM
To use the THERM sensor, you need to set the resistance parameter of the
multimeter and select the connection mode.
1. Set the resistance parameter
Press THERM Type and select a desired value from 2.2K (2.2 kΩ), 3K (3 kΩ),
5K (5 kΩ), 10K (10 kΩ) and 30K (30 kΩ).
Preset Mode
Press in any mode to enter the following interface, from which you can
choose a preset configuration from up to 10 groups. Note the configurations in this
interface correspond to the 10 groups of system configurations in the internal
memory ( Type Sys Setting).
Press Setn (n=1 to 10); if an effective configuration is stored in the storage unit
corresponding to this key, the beeper sounds and the instrument recalls the
configuration stored directly; if the storage unit selected is empty, the instrument
returns to the previous test state directly.
Besides, you can recall or save preset configurations using function. For
The secondary function key is used to enable the dual display mode, quickly
store the current instrument settings in co-ordination with and quickly enter
the REL setting interface.
Note:
The multimeter measures the two parameters at the same time and the
main and vice displays update respectively.
If the measurement from the main display uses math operation (STA
<MAX, MIN, AVG, ALL>, P/F, dBm, dB), turning on the vice display will
automatically disable the math operation in use, while the vice display
presents the measurement value of the vice display function.
The math operation (MAX, MIN, AVG, dBm, dB, REL) only works with the
main display when dual display function is used. Turning on P/F will
automatically disable the dual display.
If the measurement on the main display uses math operation (REL),
turning on the vice display will not change the display result on the main
display and the vice display presents the measurement value of the vice
display function.
Measurement Configuration
The parameters for different measurement function differ, see table below.
To set the measurement parameter, select a measurement function and then press
. For more details, please refer to the following content.
Integration Time
Integration time is the period during which the multimeter’s analog-to-digital (A/D)
converter samples the input signal for a measurement. The longer the integration
time is, the slower the measurement speed will be and the higher the resolution
will be; the shorter the integration time is, the faster the measurement will be and
the lower the resolution will be. The integration time applies to DCV, DCI, 2WR and
4WR measurements.
NPLC: choose a desired integration time and then press to return to the
upp-level menu.
APER: set a desired integration time using the direction keys and then press
Done.
DC Impedance
Auto Zero
Auto zero (Auto Zero) applies to DCV, DCI, 2WR and 4WR measurements.
ON: the multimeter internally disconnects the input signal and measured
circuit after each measurement, and takes a zero reading. It then subtracts
the zero reading from the preceding reading (displaying the difference
between the measurement value and zero value during the measurement), in
order to reduce the impact of offset voltage from input circuit on
measurement result.
Once: the multimeter takes a zero reading and then disables the auto zero.
The multimeter will subtract this zero reading from the following
measurement value.
OFF: disable the auto zero function. Whatever, the multimeter automatically
takes a zero reading once the function, range or integration time is changed.
The multimeter will subtract this zero reading from the following
measurement value.
Offset Compensation
ON: the multimeter adds two different currents onto the measured resistance
and measures the variation of the voltage between both ends of the
resistance under the two different currents, and then divides the voltage
variation by the current variation to obtain the value of the measured
resistance.
OFF: the multimeter does not execute offset compensation.
Tip
The offset compensation and auto zero are mutually exclusive. The used auto zero
function will be disabled automatically once you turn on the offset compensation,
and vice versa.
AC Filter
AC filter applies to ACV, ACI and FREQ/PERIOD measurements. It can optimize the
low-frequency accuracy and minimize the AC settling time. DM3068 provides three
types of AC filters (slow, medium and fast). The AC filter to be used is determined
by the input signal frequency.
Press Slow, Mid or Fast to select 3 Hz, 20 Hz or 200 Hz as the lowest frequency of
the AC filter. The default is 20 Hz: MEDIUM.
Short-Circuit Resistance
This function only applies to continuity test. When the measured circuit has a
resistance lower than the short-circuit resistance, the circuit is considered as
connected, and the beeper sounds (if sound is on). The default short-circuit
resistance is 10 Ω and the setting is stored in nonvolatile memory.
interface.
Set a desired value using the direction keys. The range is from 1 Ω to 2000 Ω
Current Source
Current source applies to the diode measurement function. The current selected is
applied to the diode under test and the multimeter measures the voltage drop
between the two ends of the diode to determine the polarity of the diode.
Note: when measuring diode with relatively larger drain current, do not use the
small current source (as the forward voltage drop measured might be too low to
determine the polarity of the diode).
Gate Time
Gate time (also called Aperture Time) applies to FREQ/PERIOD function. It decides
the resolution of low-frequency measurement. The longer the gate time is, the
higher the resolution of the low-frequency measurement is and the slower the
measurement is, and vice versa. For details, please refer to “Specifications”.
The gate time can be set to1 ms, 10 ms, 100 ms or 1 s and the default is 100 ms.
You can select a desired gate time by pressing the corresponding softkey.
Math Operations
DM3068 provides basic math operations (STA, P/F, dBm, dB and REL) for
measurement results and displays the history data through tendency graph and
histogram.
In DCV, ACV, DCI, ACI, 2WR, 4WR, CAP, FREQ/PERIOD or SENSOR measurement,
press to enter the following interface.
Math
The multimeter provides 5 math functions: STA (MAX, MIN, AVG and ALL), P/F,
dBm, dB and REL. Different measurement function allows different math functions
as shown in the table below.
Take the DCV for instance. Press Math to enter the following interface.
Press the related menu to enter the desired measurement setting interface. Note
that you can only use one math operation from SAT, P/F, dBm and dB at a time, but
REL can be used together with any of the four at the same time.
STA
STA calculates the min, max, average and mean square deviation of readings
during the measurement.
After you choose a desired STA function, press ON to enable it and the multimeter
enters the reading interface. The reading updates during the measurement.
When MAX or MIN is selected, the multimeter takes the first reading as the
maximum (or minimum) for displaying. The reading shown on the screen is
always the maximum (or minimum) within the current readings when
measuring more consecutive readings. The measurement interface displays
the maximum (or minimum) reading, current measurement value and number
of samples.
When AVG is selected, the multimeter always displays the average value of all
the current readings. The measurement interface displays the average value,
current measurement value and number of samples.
When All is selected, the measurement interface displays the current
measurement value as well as the maximum, minimum, average and mean
square deviation of the measured readings and number of samples.
P/F
P/F operation prompts the signal testing result (message display and beep)
according to the specified high and low limit parameters and outputs a negative
pulse through the RS232 interface at the rear panel. As shown in the figures below,
the multimeter outputs a negative pulse from the Pin 1 (or Pin 9) if the test
succeeds (or fails).
+5 V
0V
Negative Pulse
Pin 1 Pin 9
Press High or Low and enter a desired value using the direction keys. Note that the
high limit should always be greater than the low limit and the unit is decided by the
current measurement function. The range of the limit is from -110% to +110% of
the maximum range of the selected measurement function. The settings are
stored in volatile memory and will be cleared automatically at power-off.
After you setting the desired parameters, press ON to enable the P/F operation.
The multimeter automatically exits the P/F setting interface and goes to the
measurement main interface as shown in figure below.
If the current reading exceeds the high or low limit, the vice display shows “HI
FAIL” or “LO FAIL” and the beeper sounds (beeper is on: System
Sound).
WARNING
The P/F signal from pins 1 and 9 of RS232 is not compatible with the
handshake signal (Carrier Detect and Ring Indicator) from standard
RS232.
dBm
dBm represents the absolute value of the power. The dBm operation calculates the
power of the reference resistance according to the measured voltage.
Wherein, Reading denotes the measured voltage and RREF denotes the reference
resistance.
Press dBm to enter the following interface, from which you can use the direction
keys to specify a reference resistance within 2 Ω and 8000 Ω. The setting is stored
in volatile memory and will be cleared automatically at power-off.
dB
dB represents the relative value which is used in the REL operation of dBm value.
When enabled, the multimeter calculates the dBm value of the next reading and
subtracts the preset dB from this value and then displays the result.
Wherein, Reading denotes the measured voltage and RREF denotes the reference
resistance.
Press dB to enter the following interface, from which you can use the direction
keys to specify a value within -120 dBm and +120 dBm (the unit is dBm). The
setting is stored in volatile memory and will be cleared automatically at power-off.
Current: calculate the dBm value on the basis of the reference resistance from
the present dBm menu.
Default: restore the dB preset value to 0 dBm.
On: enable dB operation. The multimeter then automatically goes to the
measurement main interface, as shown in the figure below.
REL
The reading displayed on the screen in REL operation is the difference between
measured and preset values.
In the preset value setting interface, you can directly use the direction keys to
specify a preset value. The unit is decided by the current measurement function.
Tendency Graph
The multimeter draws tendency graph of the measured data through which users
can directly observe the variation of the measured data without other
supplementary means. This function is available for DCV, DCI, ACV, ACI, 2WR,
4WR, CAP, FREQ/PERIOD and SENSOR measurements.
Press Trend to enter the following interface. The screen displays the MAX
and MIN values of the measured data, the drawing time of the tendency graph as
well as the real-time curve and tendency graph of the measured data. The
multimeter supports up to 999 hours 59 minutes and 59 seconds drawing time.
When the drawing time exceeds this limit, the timer is reset and starts to time
again (but the waveform drawn is kept). When the real-time curve reaches the
maximum display length, it is compressed and added onto the tendency graph and
the multimeter starts to draw new real-time curve. When the tendency graph
reaches the maximum display length, it is compressed into half to make sure that
the compressed data on the real-time curve can be added onto it continuously to
form cumulative tendency graph.
Tendency curve
Real-time curve
Press any soft menu at this moment to open the operation menu as shown in
figure below.
Press Stop, the multimeter stops drawing and enters the following interface.
Start: enable the multimeter to redraw the graph according to the current
measurement data.
PrtScn: save the current tendency graph under the root directory of an
external USB storage device in BMP format. Note: make sure that a USB
storage device is correctly connected before using this function.
Hide: hide the operation menu.
Histogram
Press HISTO to enter the following interface. The average (AVG), standard
deviation (SDEV) and a histogram of measured data are displayed on the screen
and update accordingly during the measurement.
Press any soft menu at this moment will open the operation menu as shown in
figure below.
Press Stop, the multimeter stops updating histogram and enters the following
interface.
Trigger
DM3068 provides four types of triggers: auto, single, external and level. It reads
one or more specified number of readings (up to 50 000) each time a trigger signal
is received and allows users to set the delay time between the start of trigger and
reading.
Source: choose a trigger source from Auto, Single, Ext, and Level.
Hold: enable or disable the reading hold function.
Set: set the related parameters.
VMC: enable or disable the trigger output.
Auto Trigger:
Press Auto and Done to enable the auto trigger. The backlight of at the front
panel then turns on, and the multimeter continuously takes readings at the fastest
possible speed according to the current configuration. At this point, pressing
will enable the reading hold function; the backlight will flicker and a stable reading
will be captured and held on the screen.
Single Trigger:
Press Single and Done to enable the single trigger. The backlight of at the
front panel then goes on. The multimeter takes one or specified number of
readings every time you press .
External Trigger:
Press Ext and Done to enable the external trigger. The backlights of both and
at the front panel then go off. The multimeter will then receive the trigger
pulse from the [Ext Trig] connector at the rear panel and trigger at the specified
edge of the pulse signal and acquire measured data.
Level Trigger:
This mode applies to DCV/DCI/OHM measurement functions. Press Level and
enter a desired level using the direction keys, then press Done to enable the level
trigger. The backlights of both and at the front panel then go off. The
multimeter will trigger on the positive or negative edge of the input signal when
the input level passes through the specified trigger level and acquire measured
data.
Tip
Pressing in remote mode will switch the multimeter into local mode.
Reading Hold
The multimeter captures stable reading and holds it on the screen when the
reading hold function is enabled.
The sensitivity range determines whether the reading is stable enough and can be
displayed on the screen. This parameter is expressed in the form of reading
percentage based on the current range. The multimeter will not capture and
display new reading until three continuous readings exceed the selected sensitivity
range.
Press On/Off to enable or disable the reading hold function. When it is enabled, the
backlight of at the front panel starts flashing. The settable sensitivity ranges
are 0.01%, 0.1%, 1% and 10% and the default is 0.1%.
For example, if you choose 0.1%, the input signal into the multimeter is 5 V and
the new reading will be displayed if three continuous readings exceed the range
from 4.9975 V to 5.0025 V.
S No.
The multimeter takes the specified number of readings each time a trigger signal is
received.
Press S No. and enter a desired number within 1 and 50 000 using the direction
keys. The default is 1.
Delay:
User can set the delay time between the trigger signal and the every sample
(reading).
Press Delay to enter the following interface.
Auto: the delay time in this mode is decided by a combination of the used
measurement function, range, integration time and AC filter and other factors.
ZERO: the delay time in this mode is 0 s.
Manu: the delay time in this mode can be set to any value within 0 s and
3600 s.
Slope:
This menu allows you to select an edge type for the pulse from the [Ext Trig]
connector at the rear panel when using external (or level) trigger. The multimeter
will trigger on the specified type of edge.
Trigger Output
From this interface, you can enable or disable the trigger output. When it is
enabled, the multimeter outputs a negative pulse from the [VM Comp] connector
after each measurement.
The multimeter allows you to save the system configuration or measurement data
into the internal memory or an external USB storage device and recall it when
required.
Disk: choose a desired storage location. “C:\” indicates the internal memory,
while “A:\” indicates the external USB storage device.
Type: choose a desired file type for storing or reading. Please refer to “Storage
Type”.
Read: read the selected file. This menu is available only when the current disk
contains the specified type of file.
Save: enter the filename input interface. This menu is available only when the
current disk allows the specified type of file.
Erase: clear the selected file. This menu is available only when the current disk
contains the specified type of file.
Storage Type
Internal Storage
External Storage
The external USB storage device (A:\) can save 7 types of files according to its
capability.
To use the external storage, insert an USB storage device into the USB Host
interface at the front panel and wait until an icon appears in the status bar
on the screen.
USB Host
MENU
FUNC
CAUTION
Do not remove the USB storage device under “A:\” disk operations.
To Save a File
When the current storage disk allows specified file type, press Save to enter the
following interface. The length of the file name should be within 9 characters.
Character Selecting
Area
Place the cursor to the File Name Input Area or Character Selecting Area using the
up and down direction keys.
In File Name Input Area, select the character to be edited by pressing the right
and left direction keys (or press and hold the key), while the Character
Selecting Area is hidden.
In Character Selecting Area, choose the desired character by pressing the right
and left direction keys. At this point, the character in the File Name Input Area
changes accordingly (after choosing the desired character, select the File Name
Input Area and press the right direction key to confirm the character selected
and move the cursor to the next position).
Utility
Command Set
DM3068 supports the command sets of RIGOL (DM3068), Agilent 34401A and
Fluke 45.
Press Cmd to enter the following interface, from which you can choose a
desired command set.
For more information about commands, please refer to the Programming Guide of
DM3068.
Interface Configurations
Users can control DM3068 remotely by configuring the remote interfaces (LAN,
GPIB, USB (configuration-free) and RS232).
LAN
1. LAN Status
Display the current network connection status (LAN Status).
Configured: indicate the LAN is connected.
Unlink: indicate the LAN is not connected.
Press Info and use the up and down direction keys to view other information of
network connection.
Tip
When all three IP configuration modes are ON, the priority order of the
configuration mode is DHCP, AutoIP and ManualIP.
At least one IP configuration mode should be ON.
1) DHCP
The DHCP server from the current network assigns network parameters
to the multimeter such as IP address.
Press DHCP to turn on or off the DHCP configuration mode.
2) AutoIP
The multimeter automatically acquires an IP address from 169.254.0.1 to
169.254.255.254 and a subnet mask 255.255.0.0 according to the
current network configurations.
Press AutoIP to turn on or off the AutoIP configuration mode. To enable
this mode, DHCP must be OFF.
3) ManualIP
Users define the network parameters such as IP address, subnet mask
and gateway.
Press Manual to turn on the ManualIP configuration mode and the
multimeter enters the following interface. To enable this mode, neither
DHCP nor AutoIP should be ON.
IPAddr:
The format of IP address is nnn.nnn.nnn.nnn, for example
172.16.3.32. You are recommended to ask your network
administrator for an available IP address.
Press IPAddr and enter the desired IP address using the direction
keys.
Mask:
Gate:
The format of gateway is nnn.nnn.nnn.nnn, for example 172.16.3.1.
You are recommended to ask your network administrator for an
available gateway.
Press Gate and enter the desired gateway using the direction keys.
Each device connected to the GPIB (IEEE-488.2) interface must have a unique
address. Before using this interface, use GPIB cable to connect the multimeter with
the PC.
GPIB
Use the direction keys to set the GPIB address to any integer within 0 and 30. The
default is 7. This setting is stored in nonvolatile memory.
Before any related operations, use RS232 cable to connect the multimeter to a PC
or data terminal equipment (DTE). Then, set the interface parameters such as
baud rate, parity and handshake to match the mutlimeter with the PC or data
terminal equipment, in order to control the multimeter. The measured result can
be transferred into serial receiving device (PC) via this interface.
RS232
6 7 8 9
WARNING
The P/F signal from pins 1 and 9 of RS232 is not compatible with the
handshake signal (Carrier Detect and Ring Indicator) from standard
RS232.
Baud: select the desired baud rate from 1200, 2400, 4800, 9600, 19200,
38400, 57600 and 115200. The default is 9600. This setting is stored in
nonvolatile memory.
Parity: select the desired parity from None, Odd and Even. The default is
None. This setting is stored in nonvolatile memory.
HS: enable or disable the serial hardware handshake function. When it is
disabled, please do not connect DTR/DSR pins into a logic high level.
2. Print
Press Print to enter the following interface, from which you can turn on or off
the output of measured data from serial ports.
System Configuration
DM3068 supports English/Chinese menus and built-in helps. All the messages,
context helps and help topics are displayed in the language selected.
Press System Lang to enter the following interface and choose a desired
language.
1. Bright
Press Bright and set an appropriate brightness within 0 and 32 using the
direction keys. This setting is stored in nonvolatile memory.
2. Contr
Press Contr and set an appropriate contrast within 0 and 32 using the direction
keys. This setting is stored in nonvolatile memory.
3. Invert
Press Invert to switch the screen between normal and invert display. The
default is normal.
You can enable or disable the beeper for front panel key operations and
short-circuit test. The setting is stored in nonvolatile memory.
1. Deciml
Set the display format of the decimal point on the screen to “ ” or “ ”.
2. Separte
Set the display format of the data separator on the screen to “ ”, “None” or
“Space”. For example,
Decimal point:
Separator:
Decimal point:
Separator: Space
1. PwrOn
Select a system configuration to be used at power-on from “Default” and
“Last” (configuration at last power-off).
This setting will be available at the next power-on.
The multimeter will always use DCV as the default measurement at every
power-on no matter Default or Last is selected.
2. Default
Restore the multimeter to factory defaults (see Table 2-8).
3. Switch
Enable or disable the front power button. The default status is OFF.
ON: enable the front power button. After turning on the power switch at
the rear panel, users need to press the power button at the front panel to
turn on the multimeter.
OFF: disable the front power button. Users can directly start the
multimeter by turning on the power switch at the rear panel. However,
you can still turn off the multimeter by pressing the power button at the
front panel.
Measurement Parameters
*Short-circuit Resistance 10 Ω
AC Filter Mid
DC Input Impedance 10 MΩ
Reading Resolution 6½
Measurement Function DCV
Range Auto
Math Operations
Math Status OFF
Math Item STA
Math Register All registers cleared
dBm Reference Resistance 600 Ω
Trigger Parameters
*Delay Auto
*Number of Samples 1
*Holding Range 0.1%
*Trigger Source Auto
System
*Sound ON
*Separator ,
Simplified Chinese in China and English
*Language
in other area
Normal (black text on white
*Display Mode
background)
Error Queue Cleared
Saved Instrument Parameters and Data No change
Remote Interfaces
*GPIB Address 7
*Interface USB
*Baud Rate 9600
*Parity Check None (8 data bit)
*HS (Handshake) OFF
Remark*: Parameters marked with data * in the above table are stored in nonvolatile memory.
Use the left and right direction keys to move the cursor, use the up and down
direction keys to modify the value and press Done to finish the system clock
setting.
DM3068 can automatically detect the system hardware and display the results in
error list.
Slftst: press this softkey and then press Run to execute the self-test.
Info: provide information about the multimeter such as the model and serial
number.
Error: list the newest error information (up to 21 items) from the error queue.
This chapter introduces how to control the multimeter remotely using either virtual
panel based on web page or interface commands.
Web Control
Command Control
Web Control
DM3068 confirms to LXI-C standard and can be remotely controlled via web page.
Please take the following steps.
1) Press “Network Status” to enter the interface as shown in the figure below
to view the setting information about TCP/IP and network hardware.
Set the IP via this interface, press “Apply” to enable the modification and
press “Cancel” to disable this setting. To restore to factory setting, press
“LAN Configuration Initialize”.
Explanation:
The password input box is displayed and the correct user name and
password are required when using the “Network Settings” function if the
multimeter has had a password set on it (the default user name and
password are empty).
3) Press “Web Control” to enter the instrument web page control interface.
At this point, you can configure the related parameters and acquire the
measurement result.
4) Press “Help” to enter the interface as shown in the figure below to acquire
the instrument information and network setting information.
5) Press “Security” to enter the interface as shown in the figure below. Input
the old password and then input the new password twice; press “Apply”
and a message is displayed indicating that the password setting succeeds;
the password modification of the network setting of network control
function succeeds.
Command Control
User-Defined Programming
Users can program and control the multimeter using standard SCPI (Standard
Commands for Programmable Instruments) commands. For more details about
commands and programming, please refer to the Programming Guide of DM3068.
DM3068 uses USB, LAN, RS232 or GPIB interface to communicate with PC. The
next part will introduce how to control the instrument remotely using Ultra Sigma
through various interfaces in details. To obtain the software and detailed operation
methods, please contact RIGOL salesman or technical support.
5. Communication Test
Right click the source name
“DM3068 (USB0::0x1AB1::0x0C94::DM3A020100823::INSTR)” and then
select “SCPI Panel Control” to open the remote command control panel, from
which you can send commands and read data as shown in the figure below.
Turn on the Ultra Sigma and click . Click “Search” in the pop-up
window and the software searches for the instrument resources currently
connected to the LAN and the resources found are displayed at the right of the
window. Select the desired resource name and click “OK” to add it as shown in
the figure below.
5. Communication Test
Right click the source name “DM3068 (TCPIP::172.16.3.32::INSTR)” and then
select “SCPI Panel Control” to open the remote command control panel, from
which you can send commands and read data as shown in the figure below.
5. Communication Test
Right click the source name “DM3068 (GPIB0::25::INSTR)” and then select
“SCPI Panel Control” to open the remote command control panel, from which
you can send commands and read data as shown in the figure below.
1. Instrument Connection
Connect the multimeter to PC using RS232 cable.
succeeds, the searched devices are shown in resource list at the right side of
the panel; otherwise, follow the guides to solve the problem.
5. Communication Test
Right click the source name “DM3068 (ASRL1::INSTR)” and then choose “SCPI
Panel Control” to open the remote command control panel, from which you
can send commands and read data as shown in the figure below.
Chapter 4 Troubleshooting
The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still, please contact RIGOL and provide your device
information ( T/C Info).
This chapter introduces how to eliminate the potential error during your
measurement in order to obtain accurate measurement result.
Rs
HI
Ideal
Vs Ri Meter
LO
To reduce the effect of loading errors and noise interference, you can set the input
resistance of the multimeter to “>10GΩ” for 200 mV, 2 V and 20 V ranges, and
“10MΩ” for the 200 V and 1000 V ranges.
True-RMS AC Measurements
The AC measurement of DM3068 has true RMS response. The average thermal
power of the resistor over a certain period is directly proportional to the square of
the true RMS of the voltage of the resistor over this period of time, regardless of
the waveform shape. DM3068 can accurately measure the true RMS voltage or
current if the energy of the voltage or current waveform outside the effective
bandwidth of the multimeter can be ignored. The effective AC voltage bandwidth
of DM3068 is 800 kHz and the effective AC current bandwidth is 100 kHz.
The multimeter's ACV and ACI functions measure the “AC coupled” true RMS value,
namely the RMS value of the AC component (the DC component is filtered out) of
the input signal. As shown in table 5-1, since the sine, triangle and square (50%
duty cycle) waves do not have DC offset, their AC true RMS values equal the
AC+DC true RMS values.
Table 5-1 True- RMS AC Measurements of Sine, Triangle and Square waves
Waveform Crest Factor (C.F.) AC RMS AC+DC RMS
V
V V
0
2
2 2
V
V V
0
3
3 3
V V
1
0
t
-V
T
C.F . C.F .
The AC coupled true RMS measurement is especially useful for measuring small AC
signals having DC offset such as the AC ripple from DC power supplies. However, in
some cases you need to measure the AC+DC true RMS value. First you can
measure the DC and AC components of the signal using DCV and ACV
measurements, and then calculate the AC+DC true RMS value according to the
formula below. In order to obtain the best AC rejection, select 6.5 digits when
measuring DC voltage.
RMS ( AC + DC ) = AC 2 + DC 2
A common misconception is that "since a multimeter can measure the true RMS of
signal, its sine wave accuracy specifications apply to all waveforms." Actually, the
waveform shape of the input signal can affect the measurement accuracy. A
common way to describe the signal wave shape is “crest factor”. Crest factor is the
ratio of the peak value to RMS value of a waveform.
In general, the greater the crest factor is, the greater the energy contained in high
frequency harmonics will be. All multimeters have errors that are relevant to the
crest factor. Crest factor errors of DM3068 are listed in “AC Characteristics” in
chapter 6.
You can estimate the measurement error due to signal crest factor according to
the equation below.
- C.F.2 × F
Bandwidth Error = ×100% (% reading)
4π × BW
C.F . : signal crest factor
F : pulse fundamental wave frequency
BW : multimeter’s effective bandwidth
Example:
Calculate the approximate measurement error for a pulse train with a crest factor
of 2 and a fundamental frequency of 20 kHz. Assume that the multimeter’s 1-year
accuracy is ± (0.05%× reading + 0.03%×range).
- Rs
Loading Error (%) = × 100%
Rs + 1MΩ
F : input frequency
Cin : input capacitance (100 pF) plus test lead capacitance
Chapter 6 Specifications
DC Characteristics
100.0000 MΩ 200 nA || 10 MΩ 0.300 + 0.010 0.800 + 0.010 0.800 + 0.010 0.1500 + 0.0002
Diode Test 2.0000 V[7] 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
Continuity Test 2000.0 Ω 1 mA 0.002 + 0.010 0.008 + 0.020 0.010 + 0.020 0.0010 + 0.0020
[1] Specifications are for 90-minute warm-up and 100 PLC integration time. When the integration time is <100 PLC, add the appropriate “RMS Noise
Adder” listed in the following table.
[2] 10% overrange on all ranges except DCV 1000 V and DCI 10 A ranges.
[3] Relative to calibration standards.
[4] For each additional volt over ± 500 V, add 0.03 mV error.
[5] For continuous current > 7A DC or 7A AC rms, 30 seconds ON and 30 seconds OFF.
[6] Specifications are for 4–wire resistance measurement or 2–wire resistance measurement using REL operation. Without REL operation, add 0.2 Ω
additional error in 2-wire resistance measurement.
[7] Accuracy specifications for the voltage measured at the input terminal only. 1 mA test current is typical. Variation in the current source will cause some
variation in the voltage drop across a diode junction.
[1] Typical value. Resolution is defined as the RMS noise of DCV 20 V range (set auto zero to “Once”).
[2] Normal mode rejection ratio for power frequency ±0.1%. For power-line frequency ±1%, subtract 20 dB. For ±3%, subtract 30 dB.
[3] Maximum rate for DCV, DCI, 2-wire resistance and 4-wire resistance functions.
[4] The basic DC accuracy specifications include RMS noise at 100 PLC. For <100 PLC, add “RMS Noise Adder” to the basic DC accuracy specifications.
Measuring Characteristics
DC Voltage
Input Impedance 200 mV, 2 V, 20 V ranges: 10 MΩ or >10 GΩ
(For these ranges, input beyond ±26 V are clamped through 106 kΩ)
200 V and 1000 V ranges: 10 MΩ±1%
Input Protection 1000 V
Input Offset Current 50 pA, at 25℃, typical
CMRR (common mode 140 dB for 1 kΩ unbalanced resistance in LO lead, ±500 VDC peak maximum.
rejection ratio)
Resistance
Measurement Method 4-wire or 2-wire resistance
Current source referenced to LO input
Open-circuit Voltage Limited to <10 V
Max. Lead Resistance 10% of range per lead for 200 Ω and 2 kΩ ranges, 1 kΩ per lead on all other ranges
(4-wire)
Input Protection 1000 V on all ranges
Offset Compensation Available on 200 Ω, 2 kΩ and 20 kΩ ranges.
DC Current
Shunt Resistor 100 Ω for 200 uA, 2 mA
1 Ω for 20 mA , 200 mA
0.01 Ω for 2 A, 10 A
Input Protection Externally accessible 500 mA, 250 V high speed fuse at the rear panel for 200 uA, 2 mA, 20 mA and 200 mA ranges.
Internal 10 A, 250 V high speed fuse for 2 A and 10 A ranges.
Continuity/Diode Test
Response Time 300 samples/sec, with audible tone
Continuity Threshold Adjustable from 1 Ω to 2000 Ω
Auto Zero OFF Operation (typical value)
Following instrument warm-up at the environment temperature ±1℃ and <5 minutes, add 0.0001 % range + 2 uV error for DCV and 2 mΩ error for
resistance.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics and input signal changes. The default measurement delay can
ensure the correctness of the first reading for most measurements.
Measurement Considerations
Teflon or other high-impedance, low-dielectric absorption wire insulation is recommended for these measurements.
AC Characteristics
20 kHz-50 kHz 0.10 + 0.04 0.12 + 0.05 0.15 + 0.05 0.012 + 0.005
50 kHz-100 kHz 0.55 + 0.08 0.60 + 0.08 0.60 + 0.08 0.060 + 0.008
100 kHz-300 kHz 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.20 + 0.02
750.000 V[5] 3 Hz-5 Hz 1.00 + 0.02 1.00 + 0.03 1.00 + 0.03 0.100 + 0.003
5 Hz-10 Hz 0.35 + 0.02 0.35 + 0.03 0.35 + 0.03 0.035 + 0.003
10 Hz-20 kHz 0.04 + 0.02 0.07 + 0.03 0.08 + 0.03 0.008 + 0.003
20 kHz-50 kHz 0.10 + 0.04 0.12 + 0.05 0.15 + 0.05 0.012 + 0.005
50 kHz-100 kHz 0.55 + 0.08 0.60 + 0.08 0.60 + 0.08 0.060 + 0.008
100 kHz-300 kHz 4.0 + 0.50 4.0 + 0.50 4.0 + 0.50 0.20 + 0.02
True RMS AC 200.0000 uA 3 Hz-5 Hz 1.10 + 0.06 1.10 + 0.06 1.10 + 0.06 0.200 + 0.006
Current [8] 5 Hz-10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.100 + 0.006
10 Hz-5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
5 kHz-10 kHz 0.35 + 0.70 0.35 + 0.70 0.35 + 0.70 0.030 + 0.006
2.000000 mA 3 Hz-5 Hz 1.00 + 0.04 1.00 + 0.04 1.00 + 0.04 0.100 + 0.006
5 Hz-10 Hz 0.30 + 0.04 0.30 + 0.04 0.30 + 0.04 0.035 + 0.006
10 Hz-5 kHz 0.12 + 0.04 0.12 + 0.04 0.12 + 0.04 0.015 + 0.006
5 kHz-10 kHz 0.20 + 0.25 0.20 + 0.25 0.20 + 0.25 0.030 + 0.006
20.00000 mA 3 Hz-5 Hz 1.10 + 0.06 1.10 + 0.06 1.10 + 0.06 0.200 + 0.006
5 Hz-10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.100 + 0.006
10 Hz-5 kHz 0.15 + 0.06 0.15 + 0.06 0.15+ 0.06 0.015 + 0.006
5 kHz-10 kHz 0.35 + 0.70 0.35 + 0.70 0.35 + 0.70 0.030 + 0.006
200.0000 mA 3 Hz-5 Hz 1.00 + 0.04 1.00 + 0.04 1.00 + 0.04 0.100 + 0.006
5 Hz-10 Hz 0.30 + 0.04 0.30 + 0.04 0.30 + 0.04 0.035 + 0.006
10 Hz-5 kHz 0.10 + 0.04 0.10 + 0.04 0.10 + 0.04 0.015 + 0.006
5 kHz-10 kHz 0.20 + 0.25 0.20 + 0.25 0.20 + 0.25 0.030 + 0.006
2.000000 A 3 Hz-5 Hz 1.10 + 0.06 1.10 + 0.06 1.10 + 0.06 0.100 + 0.006
5 Hz-10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.035 + 0.006
10 Hz-5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
5 kHz-10 kHz 0.35 + 0.70 0.35 + 0.70 0.35 + 0.70 0.030 + 0.006
10.00000 A[6] 3 Hz-5 Hz 1.10 + 0.08 1.10 + 0.10 1.10 + 0.10 0.100 + 0.008
5 Hz-10 Hz 0.35 + 0.08 0.35 + 0.10 0.35 + 0.10 0.035 + 0.008
10 Hz-5 kHz 0.15 + 0.08 0.15 + 0.10 0.15 + 0.10 0.015 + 0.008
Additional Low Frequency Errors (% of reading) Additional Crest Factor Errors (non-sine wave) [7]
Frequency AC Filter Crest Factor Error (% of reading)
Slow Medium Fast
10 Hz-20 Hz 0 0.74 -- 1-2 0.05
20 Hz-40 Hz 0 0.22 -- 2-3 0.2
40 Hz-100 Hz 0 0.06 0.73 3-4 0.4
100 Hz- 200 Hz 0 0.01 0.22 4-5 0.5
200 Hz-1 kHz 0 0 0.18
>1 kHz 0 0 0
[1] Specifications are for 90-minute warm-up, slow ac filter and sinewave input.
[2] 10% overrange on all ranges except ACV 750 V and ACI 10 A ranges.
[3] Relative to calibration standards.
[4] Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range and <50 kHz, add 0.1% of range additional error. For 50 kHz
to 100 kHz, add 0.13% of range.
[5] ACV 750 range limited to 8x107 Volt-Hz. When the input exceeds 300 V rms, add 0.7 mV error for each additional volt.
[6] For continuous current > DC 7A or AC 7A rms, 30 seconds ON and 30 seconds OFF.
[7] When the frequency is lower than 100 Hz, the specification of slow filter is only for sine wave input.
[8] Specifications are for sinewave input >5% of range. For inputs from 1% to 5% of range, add 0.1% of range additional error. Specifications are typical
values for 200 uA, 2 mA, 2 A and 10 A ranges when frequency is >1 kHz.
Measuring Characteristics
True RMS AC Voltage
Measurement Method AC-coupled True-RMS -- measure the ac component of input with up to 400 V DC bias on any range.
Crest Factor ≤ 5 at full range
Input Impedance 1 MΩ ± 2%, in parallel with <150 pF capacitance on any range
Input Protection 750 V rms on all ranges
AC Filter Bandwidth Slow: 3 Hz – 300 kHz
Medium: 20 Hz – 300 kHz
Fast: 200 Hz – 300 kHz
CMRR (common mode 70 dB, for the 1 kΩ unbalance in LO lead, <60 Hz common mode signal frequency, ±500 VDC peak maximum.
rejection ratio)
True RMS AC Current
Measurement Method Direct coupled to the fuse and shunt; AC-coupled True RMS measurement (measure the AC component).
Crest Factor ≤ 3 at full range
Max. Input DC + AC current peak value <300% of range. Current with DC current component <10 A rms.
Shunt Resistor 100 Ω for 200 uA, 2 mA
1 Ω for 20 mA , 200 mA
0.01 Ω for 2 A, 10 A
Input Protection Externally accessible 500 mA, 250 V high speed fuse at the rear panel for 200 uA, 2 mA, 20 mA and 200 mA ranges.
Internal 10 A, 250 V high speed fuse for 2 A and 10 A ranges.
Settling Time Considerations
The default measurement delay of the multimeter can ensure the correctness of the first readings of most of the measurements. Make sure the RC circuit of
input terminal has been fully settled (about 1 s) before accurate measurement.
Appling >300 Vrms (or >5 Arms) will cause self-heating in signal-conditioning components and the errors are included in the instrument specifications.
Internal temperature changes due to self-heating may cause additional error on lower ac voltage ranges. The additional error will be less than 0.02% of
reading and will generally dissipate within a few minutes.
Measuring Characteristics
Frequency and Period
Measurement Method Reciprocal-counting technique, AC-coupled input using the AC voltage function.
Input Impedance 1 MΩ ± 2%, in parallel with <150 pF capacitance on any range
Input Protection 750 V rms on all ranges
Measurement Considerations
All frequency counters are susceptible to error when measuring low–voltage, low–frequency signals. Shielding inputs from external noise pickup is critical for
minimizing measurement errors.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a dc offset voltage change. The input blocking RC time constant
must be allowed to fully settle (about 1 s) before the most accurate measurements are possible.
Capacitance Characteristics
Measuring Characteristics
Capacitance Measurement
Measurement Method Apply constant current into the capacitance, and measure the voltage changing rate.
Connection Type 2-wire
Measurement Considerations
Since small capacitance measurements are susceptible to error caused by external noise, shielding inputs from external noise pickup is critical for minimizing
measurement errors.
Temperature Characteristics
[1]
Accuracy Specifications
Function Probe Type Type Optimum Range 1 Year Temperature Coefficient
TCAL℃±5℃ 0℃ to (TCAL℃-5℃)
(TCAL℃+5℃) to 50℃
Temperature α=0.00385 -200℃ to 660℃ 0.16℃ 0.01℃
RTD[2]
α=0.00389 -200℃ to 660℃ 0.17℃ 0.01℃
(R0 is within 49 Ω
α=0.00391 -200℃ to 660℃ 0.14℃ 0.01℃
and 2.1 kΩ)
α=0.00392 -200℃ to 660℃ 0.15℃ 0.01℃
2.2 kΩ -40℃ to 150℃ 0.08℃ 0.002℃
3 kΩ -40℃ to 150℃ 0.08℃ 0.002℃
Thermal
5 kΩ -40℃ to 150℃ 0.08℃ 0.002℃
Resistance
10 kΩ -40℃ to 150℃ 0.08℃ 0.002℃
30 kΩ -40℃ to 150℃ 0.08℃ 0.002℃
B 0℃ to 1820℃ 0.76℃ 0.14℃
E -270℃ to 1000℃ 0.5℃ 0.02℃
J -210℃ to 1200℃ 0.5℃ 0.02℃
K -270℃ to 1372℃ 0.5℃ 0.03℃
Thermocouple[3]
N -270℃ to 1300℃ 0.5℃ 0.04℃
R -270℃ to 1768.1℃ 0.5℃ 0.09℃
S -270℃ to 1768.1℃ 0.6℃ 0.11℃
T -270℃ to 400℃ 0.5℃ 0.03℃
[1] Specifications are for 90 minutes warm-up. Probe error excluded.
[2] Specification is for 4WR resistance measurement or 2WR resistance measurement using REL operation for sensor.
[3] Relative to cold junction temperature, accuracy is based on ITS-90. Built-in cold junction temperature refers to the temperature inside the banana jack
and its accuracy is ±2.5 ℃.
Measuring Characteristics
Measurement Considerations
The built-in cold junction temperature tracks the temperature inside the banana jack. The change of temperature in banana jack might cause additional
error. When using the built-in cold junction compensation, connect the sensor terminal of the thermocouple to the banana jack and warm it up for more than
3 minutes to minimize the error.
Measurement Rate
[2] Use the default delay setup rate (the default trigger delay).
[3] The maximum rate available when trigger delay is set to 0.
[4] 20V range, fast filter, 1kHz input.
[5] Measure 20nF capacitance on 200nF range. The measurement period changes with the capacitance under test. The maximum measurement period on
100mF is 4s (typical value).
General Specifications
Display 256×64 LCD, dual display, Chinese/English menu and online help.
Power Supply AC 100 V – 120 V, 45 Hz – 440 Hz
AC 200 V – 240 V, 45 Hz – 66 Hz
Detect the power frequency automatically at power-on, 400 Hz defaults to 50 Hz
Power Consumption 25 VA Max
Working Environment Full accuracy for 0℃ to 50℃
Full accuracy to 80% R.H. at 40℃ Non-coagulation
Storage Temperature -40℃ to 70℃
Operation Altitude Up to 2000m
Safety IEC 61010-1; EN 61010-1; UL 61010-1; CAN/CSA-C22.2 No. 61010-1
Measurement CAT I 1000V/CAT II 300V
Pollution Degree 2
EMC EN 61326-1
Weight About 3.2 kg (without package)
Dimension (height×width×length): 107.0 mm×231.6 mm×290.5 mm
Remote Interface GPIB, 10/100Mbit LAN, USB 2.0 Full Speed Device & Host (support USB storage device), RS-232C
Programming Language SCPI
LXI Compatibility LXI Class C, Version 1.2
Warm-up Time 90 minutes
Chapter 7 Appendix
Note: All the standard or optional accessories can be ordered from you local RIGOL
Office.
Appendix B: Warranty
RIGOL warrants that its products mainframe and accessories will be free from
defects in materials and workmanship within the warranty period.
RIGOL does not provide any other warranty items except the one being provided
by this summary and the warranty statement. The warranty items include but not
being subjected to the hint guarantee items related to tradable characteristic and
any particular purpose. RIGOL will not take any responsibility in cases regarding
to indirect, particular and ensuing damage.
Index