Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Siemens Basic 100 Mobile X-Ray - Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 67

Basic 100-15

100-30
Mobile Radiographic Unit

Service Manual

VIA ALDO MORO 5/7 Š I-24020 SCANZOROSCIATE Š BERGAMO Š ITALY


+39 035 66.81.63 Š FAX +39 035 66.81.66

Doc. 6101
Service Manual

Index

Section 0

Doc. 7760 Page 1 of 4


Service Manual
INDEX

General Index

Section 0 - Index (doc. 7760) Valid from 6th June 2004

Page
Section 1 - General description (doc. 7761)
1 General description 2
1.1 Applications and use 2
2 Composition 2
2.1 Mobile radiographic unit 2
2.2 Control panel 4
2.3 Collimator 6
3 Technical data 7
3.1 Classification of the apparatus 7
3.2 Technical characteristics 7
3.2.1 Dimensions and weigths 9
3.2.2 kV – mA Relationship 10
3.2.3 RX exposure time 12

Section 2 - Installation (doc. 7762)


1 Introduction 2
2 Unpacking 3
2.1 Packing with cellophane 3
2.2 Packing with carton 5
2.3 Packing with wood crate 6
3 Electrical connections 7
4 Acceptance test 7
4.1 Checking cables and connectors 7
4.2 Mechanical checks 8
4.3 Electrical checks 8

Section 0

Page 2 of 4 Doc. 7760


Service Manual
INDEX

Page
Section 3 – Maintenance (doc. 7763)
1 Routine maintenance 2
1.1 General recommendations 2
1.2 Frequent checks and inspections 3
1.3 General checks and inspections 3
1.4 Cleaning and disinfection 4
2 Special maintenance 5
2.1 Troubleshooting 5
2.2 Replacement of components 6
2.3 List potentiometers 7
2.4 List leds 8
2.5 List fuses 9
3 Spare parts 10

Section 4 – X-ray settings (doc. 7764)


1 Introduction 2
1.1 Dip switch 3
2 Rotating anode stator power supply adjustment 3
3 Set-up procedure 4
3.1 Service menu 5
4 kV adjustment 6
4.1 Adjustment “set-up” voltage 6
4.2 Adjustment of maximum kV setting safety 6
4.3 Checking maximum kV setting safety 7
5 Adjustment of the filament current 8
5.1 Adjustment of filament current 9
6 mAs adjustment 11
7 Adjustment of the battery voltage 12

Section 0

Doc. 7760 Page 3 of 4


Service Manual
INDEX

8 Checking the radiographic parameter 13


8.1 Checking kV and mAs during radiography 13
9 Centering the collimator 15
10 Spring Adjustment 16

Section 5 – Drawings (doc. 7765)

Annex 1 – Test Sheet

Annex 2 – 115Vac Power supply (optional)

Section 0

Page 4 of 4 Doc. 7760


Service Manual

General
Description

Section 1

Doc. 7761 Page 1 of 14


Service Manual
GENERAL DESCRIPTION

1 GENERAL DESCRIPTION
1.1 APPLICATIONS AND USE
The equipment is a MOBILE RADIOGRAPHIC UNIT for radiography on X-ray film, that may be
used in different places and situations: operating theatre, orthopedics, intensive care, emergency
room.

2 COMPOSITION
2.1 MOBILE RADIOGRAPHIC UNIT

18
3

4
15
5

14
8

10

19

13 11

Section 1

Page 2 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

12
2
4

16
9

MOBILE RADIOGRAPHIC UNIT


1 Transport handle with brake 11 Pivoting wheel (front whell)
2 Control panel 12 Power supply cable holder
3 Handle for tilting (optional) 13 Wheels (main whell)
4 X-ray tube head 14 Support for Tilting
5 Lateral goniometer 15 Pantograph arm lock
6 Front goniometer 16 Cassette holder
7 X-ray tube head positioning handle …
8 Collimator 18 X-ray control pushbutton
9 Adjustment of collimator diaphragms 19 Pedal for Stationary brake
10 Rail for filters and accessories

Section 1

Doc. 7761 Page 3 of 14


Service Manual
GENERAL DESCRIPTION

2.2 CONTROL PANEL

14 13 12 11

10

1 2 3 4 5 6 7

Section 1

Page 4 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

7 Unit ON
kV decrease
1 kV
kV increase
8 Unit OFF

9 Data storage in anatomical


mAs decrease technique
2 mAs
mAs increase
10 Potter-Bucky selection

3 HP/LP (1) Power selection 11 X-Ray exposure

12 X-Ray ready
4 Prog N° of exams memorized in Technique
Anatomical
13 Alarm signalling

5 Anatomical Technique Selection 14 Display for Dates and Messages

6 Collimator light activation

(1) A change of the Power level involves a change of the used focal spot:
• High Power (HP) – Big Focus
• Low Power (LP) – Small Focus
In the following paragraphs the abbreviations HP and LP will be used.

Section 1

Doc. 7761 Page 5 of 14


Service Manual
GENERAL DESCRIPTION

2.3 COLLIMATOR

50

51 54

52 52

53

COLLIMATOR
50 Collimator light on 53 Retractile meter
51 Adjustment of transverse diaphragm 54 Adjustment of longitudinal diaphragm
52 Rail for filters and accessories

Section 1

Page 6 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

3 TECHNICAL DATA

3.1 CLASSIFICATION OF THE APPARATUS


CLASSIFICATION – EN 60601 1 § 5
¯ Type of protection against short circuit: CLASS I
¯ Degree of protection against direct and indirect contact: TYPE B
¯ Use conditions: CONTINUOUS WORKING WITH INTERMITTENT LOAD
¯ Unit not to be used in the presence of an inflammable anaesthetic mixture with air or
nitrous oxide

CLASSIFICATION – 93/42/EEC DIRECTIVE


¯ In according with Annex IX: CLASS II b

3.2 TECHNICAL CHARACTERISTICS

Electrical characteristics
SINGLE PHASE VOLTAGE 230 Vac ± 10%, 16 A (Optional: 115 Vac ± 10%)
FREQUENCY 50/60 Hz
STAND-BY WORKING 1 A (115 Vac: 2.5 A)
MAX ABSORBED CURRENT
RADIOGRAPHY WORKING 12 A (115 Vac: 23 A)
LINE COMPENSATION Automatic
LINE RESISTANCE < 2.5 Ω
Radiological characteristics
15 kW 30 kW
LP (Low Power) HP (High Power) LP (Low Power) HP (High Power)
MAX POWER 7.5 kW 15 kW 7.5 kW 30 kW
MAX CURRENT IN RADIOGRAPHY 150 mA 375 mA 150 mA 425 mA
3 ms ÷ 1.3 s 1 ms ÷ 0.6 s 3 ms ÷ 1.3 s 1 ms ÷ 0.5 s
EXPOSURE TIME
Selected by the processor according to the mAs
WORKING FREQUENCY 100 kHz
RANGE KV 40 ÷ 125 (Step of 1 kV)
RANGE MAS 0.5 ÷ 200 in 25 values
A.T. PILOTAGE Inverter driven by IGBT
RIPPLE ≤ 3% at Max Power
TOTAL FILTRATION > 2.7 mmAl
RISING TIME ≤ 1 ms

Section 1

Doc. 7761 Page 7 of 14


Service Manual
GENERAL DESCRIPTION

X-Ray Tube Head


TYPE OF ANODE Rotation with speed 3000 RPM

0.6 mm
FOCAL SPOTS
1.3 mm
All the other information relevant to the X-Ray Tube Head and to the X-Ray Tube can be found in the X-Ray Tube Head
Technical Data Sheet
Collimator (optional)
SHUTTERS TO MULTIPLE PLANS Parallels and perpendicular with manual movement
All the other information relevant to the Collimator can be found in the relative Technical Data Sheet
Dosimeter (optional)
MODEL Kermax-plus VacuDap 2000 with printer optional
ACTIVE AREA 146 x 146 mm2 147 x 147 mm2
MINIMAL DOSE RESOLUTION 1 mGycm2 1 mGycm2
MAXIMAL MEASURABLE DOSE 9999.9999 mGycm2 9999.9999 mGycm2
Operating modes and functionality
INTERFACE USER Polycarbonate flat keyboard with alphanumeric LCD display for all the
operative parameters and messages of possible anomalous conditions –
administrated by a microprocessor.
Two-points techniques (kV-mAs)
OPERATING MODES RADIOGRAPHY
40 programmable anatomic technique (20 for LP and 20 for HP)
X-RAY CONTROL Distance control with double – click and extensible cable (≥4m)
Filament current
mAmin and mAmax
Maximum exposure time
Temperature maximum X-ray tube head
SAFETY
Count thermal units X-ray tube head
Max kV, min kV, max ∆kV, max I
Anode rotation
Microprocessor self – test

Transport and storage conditions


MAXIMAL TEMPERATURE –10°C ÷ 55°C
RECOMMENDED TEMPERATURE 0°C ÷ 40°C
RELATIVE HUMIDITY 20% ÷ 90%
ATMOSPHERIC PRESSURE 500 hPa ÷ 1060 hPa
Operating conditions
TEMPERATURE 10°C ÷ 40°C
RELATIVE HUMIDITY 30% ÷ 75%
ATMOSPHERIC PRESSURE 700 hPa ÷ 1060 hPa

Section 1

Page 8 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

Mechanical Characteristics
WIDTH 734 mm
LENGHT 1259 mm
HEIGHT 1467 mm
MIN SOURCE-FLOOR DISTANCE 489 mm
MAX SOURCE-FLOOR DISTANCE 1997 mm
MAX RANGE 1091 mm
FOCUS MAXIMUM HEIGHT, WITH RANGE 1000 MM 1656 mm
PIVOTING FRONT WHEEL Ø 75 360°
MAXIMUM DIFFERENCE IN LEVEL WHICH CAN BE OVERCOME WITH TILTING 25 mm
BACK WHEELS DIAMETER Ø 200
MINIMUM DEFLECTING RAY 977 mm
WEIGHT 191 kg (115 Vac: 205 kg)
MOVEMENT Manual
CASSETTE HOLDER 35X43 4

3.2.1 DIMENSIONS AND WEIGTHS

Drawing N°: 7742

Section 1

Doc. 7761 Page 9 of 14


Service Manual
GENERAL DESCRIPTION

3.2.2 KV – mA RELATIONSHIP

kV - mA Relationship in Radiography mode -


7.5kW

160,00

140,00
mA Radiography

120,00

100,00

80,00

60,00

40,00

20,00
35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125
kV

kV - mA Relationship in Radiography mode -


15kW
420,00

370,00

320,00
mA Radiography

270,00

220,00

170,00

120,00

70,00

20,00
35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125
kV

Section 1

Page 10 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

kV - mA Relationship in Radiography mode -


30kW
470,00

420,00

370,00
mA Radiography

320,00

270,00

220,00

170,00

120,00

70,00

20,00
35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125
kV

Section 1

Doc. 7761 Page 11 of 14


Service Manual
GENERAL DESCRIPTION

3.2.3 RX EXPOSURE TIME


In the following tables the exposure times (s) associated to kV and mAs are reported. Since
the apparatus is a device functioning with a two point technique (kV and mAs) we remind that the
values indicated in the table are merely theoretical and can undergo a variation in relation to the
tolerance of the mA.

TABLE 1 – 7.5KW POWER


mAs
0,5 1 1,3 1,6 2 2,5 3,2 4 5 6,3 8 10 13 16 20 25 32 40 50 63 80 100 130 160 200
kV
40 0,003 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,042 0,053 0,066 0,086 0,106 0,133 0,166 0,213 0,266 0,333 0,420 0,533 0,666 0,866 1,066 1,333
(150mA)
50 0,003 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,042 0,053 0,066 0,086 0,106 0,133 0,166 0,213 0,266 0,333 0,420 0,533 0,666
(150mA)
60 0,004 0,008 0,010 0,012 0,016 0,020 0,025 0,032 0,040 0,050 0,064 0,080 0,104 0,128 0,160 0,200 0,256 0,320 0,400 0,504 0,640
(125mA)
70 0,004 0,009 0,012 0,014 0,018 0,023 0,029 0,037 0,046 0,058 0,074 0,093 0,121 0,149 0,186 0,233 0,299 0,373 0,467 0,588
(107mA)
80 0,005 0,010 0,013 0,017 0,021 0,026 0,034 0,043 0,053 0,067 0,086 0,107 0,139 0,172 0,215 0,268 0,344 0,430 0,537 0,677
(93mA)
90 0,006 0,012 0,015 0,019 0,024 0,030 0,038 0,048 0,060 0,075 0,096 0,120 0,156 0,192 0,240 0,301 0,385 0,481 0,602
(83mA)
100 0,006 0,013 0,017 0,021 0,026 0,033 0,042 0,053 0,066 0,084 0,106 0,133 0,173 0,213 0,266 0,333 0,426 0,533 0,666
(75mA)
110 0,007 0,014 0,019 0,023 0,029 0,036 0,047 0,058 0,073 0,092 0,117 0,147 0,191 0,235 0,294 0,367 0,470 0,588
(68mA)
120 0,008 0,016 0,020 0,025 0,032 0,040 0,051 0,064 0,080 0,101 0,129 0,161 0,209 0,258 0,322 0,403 0,516 0,645
(62mA)
125 0,008 0,017 0,022 0,027 0,034 0,042 0,054 0,068 0,085 0,107 0,136 0,169 0,220 0,271 0,339 0,424 0,542 0,678
(59mA)

TABLE 2 – 15KW POWER


mAs
0,5 1 1,3 1,6 2 2,5 3,2 4 5 6,3 8 10 13 16 20 25 32 40 50 63 80 100 130 160 200
kV
40 0,001 0,002 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,034 0,042 0,053 0,066 0,085 0,106 0,133 0,168 0,213 0,266 0,346 0,426 0,533
(375mA)
50 0,001 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,043 0,053 0,066 0,083 0,106 0,133 0,166 0,210 0,266 0,333 0,433 0,533
(300mA)
60 0,002 0,004 0,005 0,006 0,008 0,010 0,012 0,016 0,020 0,025 0,032 0,040 0,052 0,064 0,080 0,100 0,128 0,160 0,200 0,252 0,320 0,400 0,520
(250mA)
70 0,002 0,004 0,006 0,007 0,009 0,011 0,014 0,018 0,023 0,029 0,037 0,046 0,060 0,074 0,093 0,116 0,149 0,186 0,233 0,294 0,373 0,467 0,607
(214mA)
80 0,002 0,005 0,006 0,008 0,010 0,013 0,017 0,021 0,026 0,033 0,042 0,053 0,069 0,085 0,106 0,133 0,171 0,213 0,267 0,336 0,427 0,534
(187mA)
90 0,003 0,006 0,007 0,009 0,012 0,015 0,019 0,024 0,030 0,037 0,048 0,060 0,078 0,096 0,120 0,150 0,192 0,240 0,301 0,379 0,481 0,602
(166mA)
100 0,003 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,042 0,053 0,066 0,086 0,106 0,133 0,166 0,213 0,266 0,333 0,420 0,533
(150mA)
110 0,003 0,007 0,009 0,011 0,014 0,018 0,023 0,029 0,036 0,046 0,058 0,073 0,095 0,117 0,147 0,183 0,235 0,294 0,367 0,463 0,588
(136mA)
120 0,004 0,008 0,010 0,012 0,016 0,020 0,025 0,032 0,040 0,050 0,064 0,080 0,104 0,128 0,160 0,200 0,256 0,320 0,400 0,504
(125mA)
125 0,004 0,008 0,011 0,013 0,016 0,021 0,026 0,033 0,041 0,052 0,066 0,082 0,107 0,132 0,165 0,207 0,264 0,331 0,413 0,521
(121mA)

Section 1

Page 12 of 14 Doc. 7761


Service Manual
GENERAL DESCRIPTION

TABLE 3 – 30KW POWER


mAs
0,5 1 1,3 1,6 2 2,5 3,2 4 5 6,3 8 10 13 16 20 25 32 40 50 63 80 100 130 160 200
kV
40 0,001 0,002 0,003 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,018 0,023 0,030 0,037 0,047 0,058 0,075 0,094 0,117 0,148 0,188 0,235 0,305 0,376 0,470
(425mA)
50 0,001 0,002 0,003 0,004 0,005 0,006 0,008 0,010 0,012 0,015 0,020 0,025 0,032 0,040 0,050 0,062 0,080 0,100 0,125 0,157 0,200 0,250 0,325 0,400
(400mA)
60 0,001 0,002 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,034 0,042 0,053 0,066 0,085 0,106 0,133 0,168 0,213 0,266 0,346
(375mA)
70 0,001 0,002 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,017 0,022 0,028 0,036 0,045 0,056 0,070 0,090 0,112 0,140 0,177 0,225 0,281 0,366
(355mA)
80 0,001 0,002 0,003 0,004 0,005 0,007 0,009 0,011 0,014 0,018 0,023 0,029 0,038 0,047 0,059 0,074 0,094 0,118 0,148 0,186 0,237 0,296
(337mA)
90 0,001 0,003 0,004 0,005 0,006 0,007 0,010 0,012 0,015 0,019 0,025 0,031 0,040 0,050 0,062 0,078 0,100 0,125 0,156
(320mA)
100 0,001 0,003 0,004 0,005 0,006 0,008 0,010 0,013 0,016 0,021 0,026 0,033 0,043 0,053 0,066 0,083 0,106 0,133 0,166
(300mA)
110 0,002 0,004 0,006 0,007 0,009 0,012 0,015 0,019 0,024 0,030 0,039 0,049 0,063 0,078 0,098 0,122 0,156 0,196 0,245 0,308 0,392
(204mA)
120 0,003 0,007 0,009 0,011 0,014 0,017 0,022 0,028 0,035 0,044 0,056 0,070 0,092 0,113 0,141 0,177 0,226 0,283 0,354 0,446
(141mA)
125 0,004 0,008 0,010 0,013 0,016 0,020 0,026 0,032 0,040 0,050 0,063 0,079 0,103 0,127 0,159 0,198 0,254 0,317 0,397 0,500
(126mA)

Section 1

Doc. 7761 Page 13 of 14


Service Manual
GENERAL DESCRIPTION

Section 1

Page 14 of 14 Doc. 7761


Service Manual

Installation

Section 2

Doc. 7762 Page 1 of 10


Service Manual
INSTALLATION

1 INTRODUCTION
The installation procedure consists of different operations to be carried out in the order shown in
the flow chart below.

Unpacking

Electrical connection

Acceptance test

The various operations indicated are described in detail in the following paragraphs.
The installation process can only be considered completed after the TEST SHEET regarding the
acceptance tests provided with this manual has been filled in.

NOTE: the radiological system is normally pre-installed and configured in the factory according to
the specific requirements of the customer. Certain adjustments may, however, be necessary when
components are replaced.
The whole machine parameter adjustment procedure is described in detail in this manual (SECTION
4).

Section 2

Page 2 of 10 Doc. 7762


Service Manual
INSTALLATION

2 UNPACKING
The radiological system is packed in a single case containing all the parts of the apparatus. To
unpack the unit, proceed as described on the following page, according to the type of packing:
¯ simple pallet with cellophane – SEE § 2.1
¯ pallet with carton – SEE § 2.2
¯ wood packing crate – SEE § 2.3
If necessary, follow the unpacking procedure backwards to pack the unit again.

2.1 PACKING WITH CELLOPHANE


1. Cut the tape (4) and remove the slide (5) – SEE FIGURE 1
2. remove the protective bag (6)
3. unscrew the four self-threading screws (7) which fix the anterior wheel fixing bracket (8)
4. free the X-ray Tube Head fixing brackets (9) from the tie-rod (14) - SEE FIGURE 2 -
working on nuts (10)
5. unscrew the self-threading screws (11) and remove the X-ray Tube Head fixing brackets
(9)
6. put the slide (5) on the floor, as shown in FIGURE 1
7. unscrew the screws (12) which fix the stand to the packing bracket (13)
8. slowly move the mobile unit towards the slide (5), paying attention to the packing
bracket (13)
9. let the unit move slowly from the slide (5) till its complete positioning on the floor.

Section 2

Doc. 7762 Page 3 of 10


Service Manual
INSTALLATION

Packing position
6

10

9
12

13 11 7 5 Floor position

FIGURE 1 - LATERAL VIEW

11

10

14

FIGURE 2 - VIEW FROM ABOVE

Section 2

Page 4 of 10 Doc. 7762


Service Manual
INSTALLATION

2.2 PACKING WITH CARTON


1. Remove the screws (1) - SEE FIGURE 3 - which fix the packing carton to the crate pallet
(2)
2. cut the band (3)
3. remove the carton packing from its housing; the mobile unit is as shown in FIGURE 1
4. proceed according to the packing with cellophane procedure (§ 2.1).

2
1

FIGURE 3 - PALLET WITH CARTON

Section 2

Doc. 7762 Page 5 of 10


Service Manual
INSTALLATION

2.3 PACKING WITH WOOD CRATE


1. Dismantle the crate cover (1) – SEE FIGURE 4
2. remove the crate walls: be careful not to ruin the wall (5) that has to be used as slide; the
mobile unit is as shown in FIGURE 1

Note: in the wood packing case, the slide (5) is not mounted as shown in FIGURE 1,
because it is substituted by the wall (5), previously mentioned

3. proceed according to the packing with cellophane procedure (§ 2.1).

1 5

FIGURE 4 - WOOD CRATE

Section 2

Page 6 of 10 Doc. 7762


Service Manual
INSTALLATION

3 ELECTRICAL CONNECTIONS
The power supply single-phase alternate voltage and maximum absorbed current values are
reported both on the unit label and on the technical data (SEE SECTION 1 - § 3.2).

Make sure that the power supply socket is approved for the values reported on
! the unit label

4 ACCEPTANCE TEST
The ACCEPTANCE TEST includes all the control and checking operations listed on the TEST
SHEET. These operations must be grouped together under three general categories:
¯ checking cables and connectors
¯ mechanical checks
¯ electrical checks
Each operation is identified by a code which is then used on the TEST SHEET.

4.1 CHECKING CABLES AND CONNECTORS


Visually control each cable and each connector, checking that there are no interruptions and/or
crushing.
A.1 Power supply cable of the unit
A.2 Radiography control pushbutton cable
A.3 Connection cable between unit/X-ray tube head

Section 2

Doc. 7762 Page 7 of 10


Service Manual
INSTALLATION

4.2 MECHANICAL CHECKS


The mechanical checks include the following operations:
¯ MOVEMENTS: all the movements foreseen must be possible without excessive force or
jerks
¯ BRAKES all the movement locking brakes must be operating and easily applied
¯ SOUNDNESS: check that there are no breakages and/or damage which might have
jeopardised operation and safety of the apparatus.

The single checking phases are listed below.


B.1 Unit sliding movement
B.2 Unit parking brake
B.3 Proper Functioning of the antitilting assemblies
B.4 Integrity of the Control panel
B.5 Integrity and fixing of the carters
B.6 Balance of the pantograph arm.

4.3 ELECTRICAL CHECKS


With regard to the electrical part, correct operation of the following must be checked:
¯ SAFETY DEVICES
¯ SIGNALS

NOTE: Remove the carters to carry out the electrical checks.

The procedure is described below:


C.1 POWER SUPPLY: connect the unit to the mains power supply;
C.2 INITIAL TEST: check that the initial automatic test is correct of the unit;
C.3 SOFTWARE: check that the visualization of installed software;
C.4 LOADING: check that the switching on the unit appears the increment of voltage on
display until 350 V. The complete loading of the capacitors battery is ready in about
40/45 sec;

Section 2

Page 8 of 10 Doc. 7762


Service Manual
INSTALLATION

C.5 DISCHARGING: switch the unit off and check the complete discharging of the
capacitors battery by means the switching off the LD3 led mounted on S25 board;

Do not carry out switching the unit On and Off


!

C.6 SET-UP DATES IN MEMORY: enter on SET-UP (SEE SECTION 4 - § 3) and check that the
SET-UP dates are in according with dates selected (SEE SECTION 4 - § 3 – TAB. 2);
C.7 X-RAY TUBE FORMATION: when not in used the unit for a long time, before to use the
unit normally, carry out the “formation” of the X-ray tube with a serial of exposure
using the following dates (carry out for each value of kV and mAs 10 exposure and
keep a pause of 1 minute between an exposure and another one):
40 kV 100 mAs
50 kV 80 mAs
60 kV 64 mAs
70 kV 64 mAs
80 kV 50 mAs
90 kV 50 mAs
100 kV 40 mAs
110 kV 40 mAs
120 kV 40 mAs

During this formation check with oscilloscope the value of kV [S19/B board Tp11 (-);
Tp3 (+) ] and mA [S19/B board: Tp11 (-); Tp6 (+) ] on inverter. Select the
oscilloscope in the following way:
CHANNEL A: 1 V/DIV ( 1V read is equal to 50 mA)
CHANNEL B: 1 V/DIV ( 1V read is equal to 20 KV)
BASE TIME: 25ms/DIV

Section 2

Doc. 7762 Page 9 of 10


Service Manual
INSTALLATION

In case of kV wave form presents of high voltage discharge or you hear discharge
inside X-ray tube interrupt the sequence, wait few minutes before starting again;
repeat the initial formation checking that the fuse F7 is interrupted (100A) (SEE
ELECTRICAL DRAW);

Warning: presence of 350 Vdc voltage!


!

C.8 EXPOSURE TIME: check that during exposure time the X-ray passage led until the end
of exposure the display shows the real exposure time in second (SEE SECTION 1 - §
3.2.3).

Section 2

Page 10 of 10 Doc. 7762


Service Manual

Maintenance

Section 3

Doc. 7763 Page 1 of 10


Service Manual
MAINTENANCE

1 ROUTINE MAINTENANCE

1.1 GENERAL RECOMMENDATIONS


The radiological system requires regular checks and maintenance. The following
recommendations have the aim of helping the operator to keep the apparatus in good working and
safe conditions during service.
The system contains mechanical parts subject to wear according to use: following prolonged
use, wear on parts may decrease safety during use. For this reason, it is essential for the checking
and maintenance operations indicated below to be carried out consistently to protect the operators
and patients against any damage caused by mechanical breakdowns.
Correct adjustment of the electrical and electronic systems has a direct influence on the
operation of the system, on the quality of the image and on the electrical safety of the system, as
well as on the level of exposure to radiation the operators and patients are subjected to.
The MAINTENANCE PROGRAMME, described in the following paragraphs, consists of controls
and interventions to be carried out by specialised personnel authorised by the manufacturer. All
maintenance operations are the responsibility of the owner of the apparatus.

Should it be necessary to replace components or parts which may in any way


! condition the safety of the machine, only use original spare parts.

Section 3

Page 2 of 10 Doc. 7763


Service Manual
MAINTENANCE

1.2 FREQUENT CHECKS AND INSPECTIONS


The operating personnel must be suitably trained to be able to carry out the daily and weekly
checks indicated in TABLE 1.
The other controls described in this chapter and the interventions described in the following
chapters are reserved for qualified and authorised personnel of the technical assistance service.

TABLE 1

INTERVAL CHECK
DAILY Operation of the signals, displays and LEDs
CHECKS Operation of the stationary brake
Integrity of the warning and danger labels
WEEKLY Absence of oil leaks from the X-ray tube head
CHECKS Absence of unusual noises in the X-ray tube head during X-ray emission
6-MONTHLY Correct operation and the value of the whole earthing circuit
CHECKS Power supply voltage value
Value of the continuous voltages generated inside the system
Fixing and general state (dust and corrosion) of the boards
Centering of the X-ray tube head-collimator assembly

1.3 GENERAL CHECKS AND INSPECTIONS


Every six months, and in any case according to local legislation in force, the procedures
indicated in the ACCEPTANCE TEST REPORT must be carried out completely. The results obtained
are noted down on the TEST SHEET together with the description of any interventions carried out.
The checks listed below must be added to those contained on the Test Sheet:
D.1 Power supply voltage check
D.2 DC power supply check
D.3 Protection earth check
D.4 Internal earth connection check

Section 3

Doc. 7763 Page 3 of 10


Service Manual
MAINTENANCE

1.4 CLEANING AND DISINFECTION


Products with a high content of alcohol, corrosive and/or abrasive detergents or solvents must
not be used to clean the surfaces of the apparatus.
To disinfect the system, only use methods in compliance with the laws in force regarding
disinfection and protection procedures against explosion.
To carry out the cleaning and disinfection operations, take the following precautions:
turn the system off and disconnect the mains power supply cable
make sure that no liquid gets into the apparatus so as to avoid any short-circuits or
corrosion of the electrical and electromechanical parts.

The unit has not to be used in presence of anaesthetic and/or infiammable


disinfectant and cleaning products.
! If, producing explosive gaseous, mixture, are used, make sure that gases are
dispersed before switching on the unit.

Section 3

Page 4 of 10 Doc. 7763


Service Manual
MAINTENANCE

2 SPECIAL MAINTENANCE

2.1 TROUBLESHOOTING
Troubleshooting can be facilitated by the indications given in TABLE 2, which illustrates the
most common incorrect operating conditions, showing their probable causes. In the presence of the
following alarms visualized on the display, the X-ray exposition is disqualified and the console’s
alarm red led light up. In case it is also present after the execution of the suggested intervention,
please contact the service assistance department.

TABLE 2

N° FAULT NOTED LIKELY CAUSE RECOMMENDED INTERVENTION


1 The unit doesn’t switch on (on key LED No power supply on input Check / Replace the S22-F2 or TR2-
switched on) F24V fuse
The unit doesn’t switch on (on key LED No power supply on input Check / Replace the S22-F1 or F1
switched off) fuse (on input)
2 The apparatus works but does not emit Radiography pushbutton fault Check the Radiography pushbutton
rays – No alarm on the unit display cable and connector
3 Alarm: “SUPPLY FAULT” +15V or –15V fault of the S22 Check power supply (+15VA/-
board supply 15VA/+24VA voltage on S22
voltage)
4 Alarm: “FILAMENT FAULT” 400 Hz power supply absent Check / Replace TR2-F130V0 fuse
X-ray tube filament interrupted Replace the X-ray tube head
S15 board fault Replace the board
5 Alarm: “INVERTER FAULT” If the indication appears on stand- Carry out set-up as in Section 4 - §
by: P1 Set-Up on S22 board faulty 4.2
If the indication appears while Switching the unit off, switching it on
carrying out X-ray: faulty on X-ray and repeat the X-ray
power circuit
F2 and F3 fuses blown Replace F2 and F3 fuses
F6 and F7 fuses blown on the Replace F6 and F7 fuses
Power circuit
6 Alarm: “KV FAULT” During a radiograph the effective Switching the unit off, switching it on
kV are less than 85% of those set: and repeat the X-ray
fault on the power circuit
7 Alarm: “mA FAULT” During an exposure the mA value Switching the unit off, switching it on
is lower than the allowed limit and repeat the X-ray
8 Alarm: “THERMIC SAFETY” X-ray tube head overheated Do not carry out any X-rays and wait
for the X-ray tube head to cool down
Breakage of the thermal sensor Replace the thermal sensor located on
the X-ray tube head and/or check the
relevant wiring on S22 board
9 Alarm: “MAN STOP RX” During a radiograph with cassette, Repeat the radiography
the control pushbutton for X-ray
command has been released early
10 Alarm: “X-RAY LACKING” Error in the high voltage generation Switching the unit off, switching it on
circuit and repeat exposure

Section 3

Doc. 7763 Page 5 of 10


Service Manual
MAINTENANCE

N° FAULT NOTED LIKELY CAUSE RECOMMENDED INTERVENTION


11 Alarm: “MAX TIME” The radiography time has exceeded Check that the mA are correct
the concurred limit Check the mA reading on board S22-
Tp1 or S19/B-Tp6
12 Alarm: “STARTER FAULT” During anode preparation it was Check the fuses on board S21A
not put into rotation Check the stator winding of the X-ray
tube head
Fault on board S21/A Replace board S21/A
Fault on board S20 Replace board S20
13 Alarm: “SWITCH OFF FOR 1 MIN” Capacitors bank still loads With the unit off wait 1 minute before
switching it on
14 Alarm: “BATTERY FAULT” Power Circuit Fault Contact technical service
15 Alarm: “OVERVOLTAGE BATTERY” Battery Circuit Fault Contact technical service
16 Alarm: “X-RAY COMMAND ACTIVE” The operator has pressed the Release the radiography pushbutton
radiography command before the and wait until the system is ready
system had finished the initial
control stage
17 Alarm: “X-RAY TUBE TOO HOT” It is not possible to begin exposure Wait for the X-ray tube to cool down.
since the remaining thermal units
available are too few
18 Alarm: “WAIT CONNECTION” The keyboard does not Switching the unit off, switching it on
communicate with the unit and repeat exposure
19 Alarm: “POTTER FAULT” Potter Fault Check the F4 - F5 fuses of the Potter
(only with Potter installed) power supply and the K3 relè

2.2 REPLACEMENT OF COMPONENTS


Replacement of a component can mean the need to carry out checks and adjustments in order to
reset correct operation of the radiological unit. TABLE 3 shows typical situations where an
adjustment is necessary.

TABLE 3
COMPONENT REPLACED ADJUSTMENT AND CHECKS REFERENCE
15 BOARD Verify mA set up in SET UP SECTION 4 - § 5
S19/B BOARD Check of kV e mA SECTION 4 - § 5.1 and 8
S20 BOARD Check Run voltage
SECTION 4 - § 2
Check start 800 ms
S22 BOARD Check max Set kV (P1) SECTION 4 - § 4.3
Check max exposure time (P2) SECTION 4 - § 4.1 and 4.2
S25 BOARD Check max voltage of battery SECTION 4 - § 7
S28 BOARD Check LCD Display contrast (P1)

Section 3

Page 6 of 10 Doc. 7763


Service Manual
MAINTENANCE

Before effecting the monoblock replacement make the following operations:


for safety purposes add some counterweight on the front part of the unit in order
to avoid the possible tilting of the same.
fasten the pantograph arm to the base of the unit (using some ropes) to avoid the
!
arm making not required movements, due to the loss of weight when the X-ray
tube head is removed.
Consult in any case Mechanical Assembly and Dismantling of the present section.

2.3 LIST POTENTIOMETERS

BOARD POTENTIOMETER VALUE NAME NOTE


S15 P1 5K MANUAL SET I FIL (ONLY FOR TEST) Only factory adjustment
P2 5K MAX I FG Only factory adjustment
P3 5K MAX I FP Only factory adjustment
S17 P1 10K FEEDBACK KV+ Only factory adjustment
P2 10K Only factory adjustment
P3 10K FEEDBACK KV- Only factory adjustment
P4 10K Only factory adjustment
S19/B P1 2K MAX CURRENT Only factory adjustment
P2 2K FREQUENCY SET Only factory adjustment
P3 2K MA FLUORO OFFSET ADJ. N.U.
R66 FEEDBACK MA ADJ Only factory adjustment
S20 P1 1K RUN VOLTAGE Only factory adjustment
P2 1M START 800MS Only factory adjustment
S22 P1 2K MAX SET KV Only factory adjustment
P2 100K MAX TIME RAD Only factory adjustment
S25 P1 5K MAX V CAPACITORS BANK Only factory adjustment
S28 P1 20K LCD CONTRAST Only factory adjustment

Section 3

Doc. 7763 Page 7 of 10


Service Manual
MAINTENANCE

2.4 LIST LEDS

BOARD LED COLOR NAME NOTE


S15 LD1 YELLOW +24V
LD2 GREEN +15VA
LD3 RED SAFETY MAX I FIL
S19/B LD1 YELLOW COM-RX
LD2 RED ∆ KV MAX
LD3 RED KV > 110%
LD4 RED KV MIN
LD5 RED I MAX
LD6 YELLOW COM. 1
LD7 YELLOW COM. 2
LD8 YELLOW PRE-RX
LD9 GREEN +15V
LD10 GREEN -15V
S20 LD1 YELLOW STARTER ON
LD2 GREEN STARTER OK
LD3 YELLOW FLUORO PREP (N.U.)
S22 DL1 YELLOW N.U.
DL2 YELLOW N.U.
DL3 YELLOW N.U.
DL4 YELLOW N.U.
DL5 YELLOW N.U.
DL6 YELLOW N.U.
DL7 YELLOW N.U.
DL8 YELLOW N.U.
DL9 YELLOW STARTER OK
DL10 YELLOW PREP K1A
DL11 YELLOW RAD
DL12 YELLOW KV 85%
DL13 YELLOW POTTER OK
DL14 GREEN THERMIC
DL15 YELLOW KV OK
DL16 GREEN FIL OK
DL17 YELLOW PRE RX
DL18 YELLOW RX
DL19 YELLOW PREP K1A
DL20 YELLOW START POTTER
DL21 GREEN +24VA
DL22 GREEN +24VA
DL23 YELLOW ON/OFF
DL25 GREEN +24VA
DL26 YELLOW CAPACITORS BANK CHANGE
DL27 YELLOW CAPACITORS BANK CHARGE
DL28 YELLOW OUT_WD
DL29 YELLOW PREP STARTER
DL30 YELLOW FP
DL31 YELLOW CORR MA

Section 3

Page 8 of 10 Doc. 7763


Service Manual
MAINTENANCE

2.5 LIST FUSES

BOARD NAME TYPE NOTE


LOADING BATTERY F2 16A
Input power supply protection
F3 16A
INVERTER F6 63A
Protection Inverter
F7 63A
TR1 TRANSFORMER F1 200MA-T Protection TR1 Transformer
TR2 TRANSFORMER F230V 2A-T
F115V 3.15A-T
F135V 1A-T Protection TR2 Transformer
F24V 2A-T
F12V 10A-T
POTTER BUCKY F4V 2A-T Protection Potter Bucky (Optional)
F5V 2A-T
AUTOTRANSFORMER F10 2A-T Protection 115Vac power supply (Optional)
S21/A F1 10A-T Input power supply protection S21/A Board
F2 10A-T
F3 500MA-T Protection TR1 Transformer S21/A Board
S22 F1 1A-T
Protection Bridge Rectifier (24Vac)
F2 2A-T
S25 F1 80mA-T Input power supply protection S25 Board
S29 F1 200mA-T
Protection Bridge Rectifier Group 1 (24Vac)
F2 200mA-T
F3 200mA-T
Protection Bridge Rectifier Group 2 (24Vac)
F4 200mA-T
F5 200mA-T
Protection Bridge Rectifier Group 3 (24Vac)
F6 200mA-T
F7 200mA-T
Protection Bridge Rectifier Group 4 (24Vac)
F8 200mA-T
F9 315mA-T Input power supply protection S29 Board

Section 3

Doc. 7763 Page 9 of 10


Service Manual
MAINTENANCE

3 SPARE PARTS
In case of replacing of parts that can be negative for the safety of the machine, use only original
parts. For the list of the spare parts and relative codes to make reference to the SECTION 5 –
EXPLODED DRAWINGS.

NOTE: When you require spare parts ,it is necessary to tell the code of the piece and the serial
number of the unit too.

Section 3

Page 10 of 10 Doc. 7763


Service Manual

X-ray settings

Section 4

Doc. 7764 Page 1 of 18


Service Manual
X-RAY SETTINGS

1 INTRODUCTION
The radiological system is fully regulated in the factory: it is therefore only necessary to install it in
order to use it.
All the information contained in this section and SECTION 3 of this manual must be used when
extraordinary maintenance is necessary.

In the case of any intervention, it is essential for the adjustment operations to be


! carried out in the sequence indicated.

In TABLE 1 indicates the potentiometers which do not require any further


adjustments since they have already been regulated and sealed in the factory.
!

TABLE 1
BOARD POTENTIOMETER DESCRIPTION
S15 P1 Manual Setting of the filament current
P2 Maximum current of large focus
P3 Maximum current of small focus
S17 (X-ray tube head) P1
Return of the kV+
P2
P3
Return of the kV-
P4
S19/B P1 Maximum current
P2 Frequency setting
S22 P1 Maximum kV setting
P2 Maximum exposure time
S25 P1 Maximum battery voltage

NOTE : Jumper S19/B-J2A and J2B ON: safety kVmax set up for rotating anode
Jumper S19/B-J2A and J2B OFF: safety kVmax set up for stationary anode

Section 4

Page 2 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

1.1 DIP SWITCH


Here follows the function of the dip-switch S22-SW1 pins:

1 – Only factory adjustment 6 – Only factory adjustment


2 – Only factory adjustment 7 – Only factory adjustment
3 – ON – It doesn’t display the name when switched on 8 – Only factory adjustment
4 – Only factory adjustment
5 – ON – Small Focus when switched on

2 ROTATING ANODE STATOR POWER SUPPLY


ADJUSTMENT
Connect an AC voltmeter (true RMS if possible) with full-scale set
fs: 300V
at 300 V between the TB3-2 and TB3-3 test points of board S21/A
(power supply of the main winding);
TB3-2 TB3-3

connect the probe of an oscilloscope between IC2-8 (+) and Dz1


Anode (-) of board S20 and prepare the instrument to read a time
of 800 ms;
take the Radiography Command pushbutton.

For maximum operator safety, at this point it is advisable to temporarily disable


! the X-ray command signal by disconnecting the wire coming from S22-X4-3.

only press the first click of the Radiography Command


pushbutton (preparation pushbutton);
check that the stator is supplied with a voltage of about 230 Vac
for approximately 0,8 seconds. If necessary, adjust this time
interval using potentiometer P2 of board S20;
check that once the 0,8 seconds have passed the stator is supplied
with a voltage of about 70 Vac. If necessary, adjust this value by
using poten-tiometer P1 of board S20.

Section 4

Doc. 7764 Page 3 of 18


Service Manual
X-RAY SETTINGS

3 SET-UP PROCEDURE
To access the modality of SET-UP procedure to follow what underneath reported:
press the keys of the console indicated to side at the same time

the wording "SET-UP PROCEDURE ?" will appear on the display;


YES
to confirm selection, press the key indicated at the side marked
with "YES", whereas to cancel the procedure, press the key
NO
marked "NO".

NOTE: In the Set-up procedure, the functions for editing the content of the fields shown on the
display have been assigned to the keys as indicated below.

YES +
kV mAs

NO -

TABLE 2 – MINIMUM, MAXIMUM AND OF DEFAULT VALUES OF SET-UP ADJUSTMENT

FUNCTION VALUE
DIRECTORY FILE MINIMUM MAXIMUM DEFAULT

KV40 300 600 380


kV ADJUSTMENT
kV120 1000 1500 1190
150 mA (40 kV) 1000 2000 1600
SMALL FIL. ADJ 7.5 kW
75 mA (100 kV) 900 2000 1360

375 mA (40 kV) 1000 2000 1690


15 kW
150 mA (100 kV) 900 2000 1450
LARGE FIL. ADJ
425 mA (40 kV) 1000 2000 1750
30 kW
300 mA (100 kV) 900 2000 1560

MAS ADJ kV 050 (50 mAs) 250 1000 530

BATTERY V = 350

Section 4

Page 4 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

3.1 SERVICE MENU


When the unit is off, pushing both the keys underneath you enter the SERVICE MENU of the unit
which subsequently visualizes the following data:
name of the device and software version
chosen anatomical program
X-ray tube model
total number of radiographic exposures
total mAs accumulated.

kV mAs

Section 4

Doc. 7764 Page 5 of 18


Service Manual
X-RAY SETTINGS

4 KV ADJUSTMENT

4.1 ADJUSTMENT “SET-UP” VOLTAGE


The SET KV voltage is supplied by the microcomputer and must be adjusted according to the
relationship kV/20 (i.e.: 1V = 20kV).
First of all, enter the SET-UP procedure, following the instructions
indicated in the previous paragraph;
On board S19/B connect the probes of a Vdc digital multimeter
between CP1-10 (+) and TP11 (-);
kV ADJ a 40 kV
in Set-up select "kV ADJ at 40 kV";
2,00 V
adjust the value until a reading of 2,00 V is obtained on the
+ -
voltmeter;
CP1-10 TP11

kV ADJ a 120 kV
in SET-UP select "kV ADJ at 120 kV";
6,00 V
adjust the value until a reading of 6,00 V is obtained on the
+ -
voltmeter;
CP1-10 TP11
exit the SET-UP procedure by using the key;
then select 100 kV, 80 kV and 60 kV, checking that the voltage
read on the voltmeter is Vread = (KvSET/20) ± 25 mV.

4.2 ADJUSTMENT OF MAXIMUM KV SETTING SAFETY


The maximum parameter value of SET KV under operating conditions must be 6 V,
corresponding to 120 kV. The MAX SET KV safety device intervention is normally set to 130 KV.
connect the probes of a Vdc digital voltmeter between test points
TP10 (+) and TP9 (-) of board S22;
turn the system on and use potentiometer P1 until a reading of 6,5 6,5 V

V ± 0,1 V is obtained. + -

TP10 TP9

Section 4

Page 6 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

4.3 CHECKING MAXIMUM KV SETTING SAFETY


Connect the voltmeter to test points TP10 (+) and TP9 (-) of board
S22
lower the MAX SET KV safety intervention to 100 kV adjusting 5V
potentiometer P1 until a reading of 5 V ± 0,1 V is obtained on the + -

voltmeter; TP10 TP9

checking that by setting the kV to a higher value than 100 kV the


"KV FAULT" alarm goes off;
6,5 V
put the MAX SET KV safety intervention back to 130 kV using
+ -
potentiometer P1 again until a reading of 6.5 kV ± 0,1 V is
TP10 TP9
obtained.

Section 4

Doc. 7764 Page 7 of 18


Service Manual
X-RAY SETTINGS

5 ADJUSTMENT OF THE FILAMENT CURRENT


The following tables indicate the mA anode values associated with the kV and the typical filament
currents determined on a sample X-ray tube head for 30 kW, 15 kW and 7.5 kW power.

TABLE 3 RADIOGRAPHY – 30 KW

Set kV kV mA IFilament
S22: TP10 – TP9 on display S19/B: TP6 - TP11 S15: CP6-2
[V] [kV] [1V=10mA] [mA]

2 40 425 870

2,5 50 400 817

3 60 375 822

3,5 70 355 803

4 80 337 814

4,5 90 320 806

5 100 300 795

5,5 110 204 769

6 120 141 753

Pre-switching: 466

TABLE 4 RADIOGRAPHY – 15 KW

Set kV kV mA IFilament
S22: TP10 – TP9 on display S19/B: TP6 - TP11 S15: CP6-2
[V] [kV] [1V=10mA] [mA]

2 40 375 799

2,5 50 300 752

3 60 250 733

3,5 70 214 723

4 80 187 713

4,5 90 166 703

5 100 150 693

5,5 110 136 684

6 120 125 673

Pre-switching: 448

Section 4

Page 8 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

TABLE 5 RADIOGRAPHY – 7.5 KW

Set kV kV mA IFilament
S22: TP10 – TP9 on display S19/B: TP6 - TP11 S15: CP6-2
[V] [kV] [1V=10mA] [mA]

2 40 150 826

2,5 50 150 777

3 60 125 757

3,5 70 107 747

4 80 93 737

4,5 90 83 727

5 100 75 716

5,5 110 68 706

6 120 62 696

Pre-switching: 471

5.1 ADJUSTMENT OF FILAMENT CURRENT


NOTE: by Filament Current the current which runs through the primary winding of the
transformer supplying the power to the filament itself is intended (Coolidge Transformer).
MEASUREMENTS
• To measure the Filament Current, connect an ammeter with true fs: 1 A

RMS reading in series with terminal CP6-2 of board S15. Set


the full-scale to 1A in AC. CP6

• to measure the anode mA in RADIOGRAPHY, connect the probe 1V=50mA

of an oscilloscope between test points TP6 (+) and TP11 (-) of + -

board S19/B to display the wave form of the mA (reading TP6 TP11

correspondence is 1V = 50 mA) and between test points TP3


1V=20kV
(+) and TP11 (-) of board S19/B to display the wave form of + -

the kV (reading correspondence is 1V = 20 kV).


TP3 TP11

Section 4

Doc. 7764 Page 9 of 18


Service Manual
X-RAY SETTINGS

PROCEDURE (example with Small focus LP 7,5 kW)


• activate the SET-UP procedure;
• select "SMALL FIL ADJ";
• adjust the "DAC" value set until a filament current, displayed on
the digital voltmeter, equal to the one indicated in TABLE 5 for
40 kV (826) is obtained – Typical DAC 1600 – FIGURE 1;
• give the X-ray command and check that the anode mA are 150.
If necessary, adjust the SET-UP until the desired value is
obtained;
• proceed with SET-UP;
• adjust the set value of the "DAC" until a filament current,
displayed on the digital voltmeter, equal to the one indicated in
TABLE 5 for 100 kV (716) is obtained – Typical DAC 1360 –
FIGURE 2;
• give the X-ray command and check that the anode mA are 75;
if necessary, adjust the SET-UP until the desired value is
obtained;
• proceed with SET-UP by selecting "LARGE FIL ADJ";
• Repeat the same operations described for "SMALL FIL ADJ"
with 40 kV and 100 kV. For the anode and filament current mA
values, refer to TABLE 3 (30 kW) and TABLE 4 (15 kW)
according to the power of the machine.
• Exit SET-UP.

826 mA

150mA (40kV) 1600

FIGURE 1 – DAC (40 KV) ADJUSTMENT

Section 4

Page 10 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

716 mA

75mA (100kV) 1360

FIGURE 2 – DAC (100 KV) ADJUSTMENT

6 mAS ADJUSTMENT
MEASUREMENTS
• Remove the bridge between terminals CP3-1 and CP3-3 of
board S19/B
• connect a mAsmeter between the above-mentioned terminals.
PROCEDURE
• activate the SET-UP procedure
• select "mAS ADJ" and give the X-ray command
• check that the mAs value on the instrument is really the one set 50 mAs

(50 mAs – Typical DAC 0530); if necessary, adjust the SET-UP + -

until the desired value is obtained CP3-1 CP3-3

• exit SET-UP.

Section 4

Doc. 7764 Page 11 of 18


Service Manual
X-RAY SETTINGS

7 ADJUSTMENT OF THE BATTERY VOLTAGE


MEASUREMENTS
• To measure the battery voltage, connect a voltmeter with true 350 V

RMS reading between terminals CP2-1 (+) and CP2-3 (-) of + -

board S25. CP2-1 CP2-3

Warning: presence of 350 Vdc voltage


!
PROCEDURE
• activate the SET-UP procedure;
• select "V BATTERI ADJ";
• Adjust the Set-up until the point in which the battery voltage
measure value, reaches the battery voltage set value (tip. 350V);
• exit SET-UP.

Section 4

Page 12 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

8 CHECKING THE RADIOGRAPHIC PARAMETER


After having completed the operations indicated in the previous paragraphs (§ 2, § 4, § 5), it is
necessary to check the values of the signals proportional to the kV and to the mA which come from
the INVERTER. The tolerance for the kV signal is ± 5% whereas for the mA it is ± 15%.

NOTE: If the kV value is not correct, it is necessary to repeat the SET kV adjustment.
NOTE: An oscilloscope with two channels is required for the checks below.

8.1 CHECKING KV AND mAs DURING RADIOGRAPHY


Connect the probe of the oscilloscope as indicated below: Ch. A

CHANNEL A: TP3 (+) – TP11 (-) of board S19/B (REAL kV); + -

TIME BASE: 10 ms TP3 TP11

If necessary, connect the probe of the oscilloscope as indicated Ch. B


below: + -

CHANNEL B: TP6 (+) – TP11 (-) of board S19/B (mA); TP6 TP11

TIME BASE: 10 ms
set the minimum value of the kV (40 kV – 425 mA)
give the X-RAY command and check that the values read on the
oscilloscope correspond with those indicated in TABLE 6; (FIGURE
3 shows an example of a 100 kV wave form)
repeat the previous operations, setting the kV values to 80 kV and
120 kV both for 15 kW and 7.5 kW powers.

TABLE 6 – 30KW

kV mAs mA Time (s)

Acceptance range
Acceptance range
Set S19: TP3 - TP11 Set Theoretical Theoretical
(±10%)
[1V = 20 kV]

40 38 ÷ 42 16 14.5 ÷ 17.5 425 0.037

80 76 ÷ 84 16 14.5 ÷ 17.5 337 0.047

120 114 ÷ 126 16 14.5 ÷ 17.5 141 0.113

Section 4

Doc. 7764 Page 13 of 18


Service Manual
X-RAY SETTINGS

TABLE 7 – 15KW

kV mAs mA Time (s)

Acceptance range
Acceptance range
Set S19: TP3 - TP11 Set Theoretical Theoretical
(±10%)
[1V = 20 kV]

40 38 ÷ 42 16 14.5 ÷ 17.5 375 0.042

80 76 ÷ 84 16 14.5 ÷ 17.5 187 0.085

120 114 ÷ 126 16 14.5 ÷ 17.5 125 0.128

TABLE 8 – 7.5KW

kV mAs mA Time (s)

Acceptance range
Acceptance range
Set S19: TP3 - TP11 Set Theoretical Theoretical
(±10%)
[1V = 20 kV]

40 38 ÷ 42 8 7.2 ÷ 8.8 150 0.053

80 76 ÷ 84 8 7.2 ÷ 8.8 93 0.086

120 114 ÷ 126 8 7.2 ÷ 8.8 62 0.129

FIGURE 3 – ACQUISITION OF THE KV WAVE FORMS DURING RADIOGRAPHY [100 KV – 16 MAS – 30 KW]

Section 4

Page 14 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

FIGURE 4 – ACQUISITION OF THE mA WAVE FORMS DURING RADIOGRAPHY [100 KV – 16 MAS – 30 KW]

9 CENTERING THE COLLIMATOR


For any necessary centering of the collimator, refer to the TECHNICAL MANUAL of the collimator
itself, enclosed with this manual.

Section 4

Doc. 7764 Page 15 of 18


Service Manual
X-RAY SETTINGS

10 SPRING ADJUSTMENT
The BALANCED SUSPENSION system of the mobile unit allows flowing movement of the ARM
vertically and, therefore, of the X-RAY UNIT. This system has been studied and made to reduce the
need for maintenance to a minimum.
As time passes and due to the continual stresses, the balancing system may lose a very small
amount of its efficiency. It is usually sufficient to adjust the BOOM PIN (3) which acts directly on
the CLUTCH DISKS (2); disassemble the arm carters for this operation (SEE 3 – MECHANICAL
ASSEMBLY AND DISMANTLING).

FIGURE 5 - EXPLODED DRAWING OF THE PANTOGRAPH ARM

Section 4

Page 16 of 18 Doc. 7764


Service Manual
X-RAY SETTINGS

If this operation is not sufficient to restore the optimum balanced condition, it is necessary to carry
out an adjustment on the thrust of SPRING (6) as described below:
remove the front casing of the UNIT
loosen the BOOM PIN (3) using a hexagonal spanner, taking
care of the sudden and unwanted movement of the ARM and
X-RAY UNIT
this brings the Arm in the highest possible position
with a second person’s help let the arm down till the NUT (8)
is accessible in the WINDOW (7)
with a spanner (12 mm), screw up or unscrew NUT (8) 4

through the WINDOW (7) on the SHAFT WITH SPRING (4).


This operation modifies the pre-load applied to the
COMPRESSION SPRING (6)
to check correct balance, make sure that the ARM can
remain in the horizontal position without any support and
loaded with the clutch (3) 6

on completion of balancing, put the clutch back to a


compression value which, although keeping the arm steadily
immobile, in any position allows easy handling (balancing to
be checked with the assembled arm carters)
on completion of the operations, replace the casings in their
7
original positions.

FIGURE 6 - SHAFT WITH SPRING

Section 4

Doc. 7764 Page 17 of 18


Service Manual
X-RAY SETTINGS

Section 4

Page 18 of 18 Doc. 7764


Service Manual

Drawings

Section 5

Doc. 7765 Page 1 of 2


Service Manual
DRAWINGS

Index
Electrical
¾ 7615 Electronic Group 100
¾ 7616 S15 Board
¾ 7548 S17 Board
¾ 7617 S19/B Board
¾ 7618 S20 Board
¾ 7619 S21/A Board
¾ 7620 S22 Board
¾ 7621 S25 Board
¾ 7622 S28 Board
¾ 7623 S29 Board

Topographic
¾ 630 S14 Board
¾ 1653 S15 Board
¾ 4617 S17 Board
¾ 7822 S19/B Board
¾ 46 S20 Board
¾ 457 S21/A Board
¾ 1401 S22 Board
¾ 1419 S25 Board
¾ 1141 S28 Board
¾ 1420 S29 Board

Exploded view
¾ 7766 Mechnical
¾ 7862 Electrical

Section 5

Page 2 of 2 Doc. 7765


EXPLODED VIEW
MECHANICAL

3 N° DESCRIPTION CODE NOTE

X-RAY TUBE HEAD SUPPORT GROUP 01.02.008.012


Depends on the
1 X-ray tube head 100KHz
customer
2 S17 X-ray tube head Board 01.17.008.002

3 Sup. carter with flat band 01.02.008.002


1
2 14
4 Inf. carter without flat band 01.02.008.003

Light blue goniometer 01.02.014.009


5
Light green goniometer 01.02.010.148

6 Brake disk DI 38.5 DE 62.5 01.02.001.108

Small X-ray tube head handle (grey) 01.02.008.006


7
Small X-ray tube head handle (green) 01.02.025.058

8 Distance ring for rotating group 01.02.025.065

9 Rotating adaptor Ralco mm. 18 49.18.002.001

10 Lamp for centering 12V 100W 49.05.012.001

11 Collimator R105 12V 49.18.006.001

4
KEYBOARD GROUP 01.24.012.015
12
6 Depends on the
12 Control Panel
customer
S28 Keyboard card 01.17.010.002
5 13
S62 Serial keyboard card 01.17.037.001

14 Double-click pushbutton 01.02.003.007

8 13

10 9

Drawing N° 01.02.008.012 - 003507

11

Drawing N° 01.24.012.015 - 003304

Doc. 7766 PAGE 1 OF 1


Annex 1
TEST SHEET

Test Sheet
The present document has to be filled up by the operator during the ordinary and extraordinary maintenance procedures.
The indicated procedures are described in detail on the SECTION 2 - § 4 of the equipment’s service manual.

CODE OPERATION REFERENCES / NOTES RESULT


A.1 Power supply cable of the unit
A.2 Radiography control pushbutton cable
A.3 Connection cable unit/X-ray tube head

B.1 Unit sliding movement


B.2 Unit parking brake
B.3 Integrity of the control panel
B.4 Integrity and fixing of the carters
B.5 Balance of the pantograph arm

C.1 Check power supply


C.2 Check initial test
C.3 Check software
C.4 Check loading capacitors
C.5 Check discharging capacitors

Doc. 7767 Page 1 of 2


Annex 1
TEST SHEET

CODE OPERATION REFERENCES / NOTES RESULT


C.6 Check Set-up dates in memory
C.7 X-ray tube formation button
C.8 Check X-ray exposure time

UNIT MODEL:

UNIT SERIAL NUMBER:


DATE: ____ /____ /_________ OPERATOR:

Doc. 7767 Page 2 of 2


EXPLODED VIEW
ELECTRICAL

S14
S22
TH1

F6

BR1

R1
Drawing N° 01.02.047.018 - 007823
S15 R2

LF1

Fan
Fvent

F12V

L1
N° DESCRIPTION CODE NOTE S19/B
ELECTRONICS GROUP 01.02.047.018
BR1 Graetz bridge Semikron SKB 30/16A1 49.06.018.015 F24V

C1/Starter 30 µF 450 Vdc Power capacitor 49.06.012.035


FAN Fan 80x80x25 230Vac 50Hz 49.06.023.008 S25 F135V

F1- Fvent Fuse 5x20 200mA 250V 49.06.022.025


F2-F3 Fuse 10x38 16A 500V GL 49.06.022.003 F230V

F6 Fuse Brush 63FE (63A 660V) 49.06.022.001


F12V Fuse 5x20 10A 250V 49.06.022.022 F3

F24V-F230V Fuse 5x20 2A 250V 46.06.022.013


F135V Fuse 5x20 1A 250V 49.06.022.011 F2
K1B-K2-K0 BC6-30-01 24Vdc 4kW Contactor 49.06.015.017
K2/A CA6-11M Auxiliary contactor 49.06.015.021 F1
2
L1 2,5 mm Filter inductor 01.02.010.195
LF1 EPCOS B84112-B-B120 Filter 49.06.030.006 SW1

R1-R2 ATE RB 106 25E Resister 49.06.011.080 S20


S14 S14 Filter board 01.17.005.005 TR1

S15 S15 Filament board 01.17.001.020 S21/A


S19/B S19/B Inverter board 01.17.008.036 K1B
S20 S20 Starter board 01.17.006.001 TR3
S21/A S21/A Rotation board 01.17.003.002 K2/A

S22 S22 Micro board 01.17.010.001 K2


S25 S25 Capacitors bank loading board 01.17.005.009
SW1 Int. Aut. Multi 9 C60N C16A 2P 49.05.008.002 C1/Starter
K0 TR2
TH1 Powerblock TT61 N 12 KOF 49.06.021.013
TR1 TM 15/24 Module transformer 01.05.001.017
TR2 300VA Tor. transformer 01.05.001.026 Drawing N° 01.02.047.012 - 007824
TR3 230-250/330 600VA Tor. transformer 01.05.001.016

Doc. 7862 Page 1 of 3


EXPLODED VIEW
ELECTRICAL

Q2 C26

L2

C29

C27 F7

Q1

TR4

S29 TA1

Drawing N° 01.02.047.015 - 007830 Drawing N° 01.02.025.018 - 003302

N° DESCRIPTION CODE NOTE


C26-C27 1500 µF 400Vdc Elect. capacitor 49.06.012.079
C29 (80A) 2 µF 1700 Vdc mkV capacitor 49.06.012.078
F7 Fuse Brush 100 FE (100A 660V) 49.06.022.010
L2 5,1 µH 45112 Tor. inductor 01.05.003.006
Q1-Q2 Tr. IGBT Siemens BSM 300GB 120 DLC 49.06.017.025
S29 S29 Driver board 01.17.008.023
TR4 100kHz Driver tor. transformer 01.05.001.018
TA1 ENCO 10701 1/500 Amper transformer 01.05.004.002

Doc. 7862 Page 2 of 3


EXPLODED VIEW
ELECTRICAL

R4A

R5

R4

C1 C11

R5A

Drawing N° 01.02.047.016 - 007831 Drawing N° 01.02.047.017 - 007832

N° DESCRIPTION CODE NOTE


C1÷C20B V 16000 µF 350Vdc Itelcond Elect. capacitor 49.06.012.076
R4-R4A-R5-R5A IRE RCC 47 Ω Resister 49.06.011.088
Drawing N° 01.02.025.027 - 003259

Doc. 7862 Page 3 of 3

You might also like