Jwi Jps Js User Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 13

Chapter

1、 SUMMARY ------------------------------------- 1
2、 INTRODUCTION ------------------------------------- 1
3、 PRECAUTIONS ------------------------------------- 2
4、 CONNECTION OF L/C AND JWI-586 ------------------------------------- 2
5、 KEY BOARD ------------------------------------- 3
6、 SETTING OF CAPACITY AND RESOLUTION ------------------------------------- 4
7、 CONNECTION OF THE INPUT/OUTPUT DELICACY ------------------------------------- 5
8、 FUNCTION SETTING ------------------------------------- 5
9、 OPERATION ------------------------------------- 9
10、 WEIGHING UNIT CONVERSION ------------------------------------- 10
11、 OUTPUT OPERATION ------------------------------------- 10
12、 SPECIFICATION ------------------------------------- 10
13、 APPENDIX 1 ------------------------------------- 11
ERROR MESSAGE ------------------------------------- 11
14、 APPENDIX 2 ------------------------------------- 11
RS-232 FORMAT ------------------------------------- 11
PRINT FORMAT ------------------------------------- 12
RS-232 CONNECTOR ------------------------------------- 12
15、 APPENDIX 3 ------------------------------------- 13
ASSEMBLY ------------------------------------- 13

1、Summary
The JWI-586/JPS/JS series is a multi-function weighing indicator that offer flexibility
together with good features! Ideal weighing indicator for clients that want
stability ,good quality together with a competitive cost .

2、INTRODUCTION:
Thank you for your purchase of this high precision electronic weighing indicator /
scale of our company. The main features of the Weighing Indicator / scale is :-
1. The microprocessor in this indicator features
A: Automatic zero point tracking function.
B: Selection of the ON/OFF of the zero tracking
C: Selection of the baud rate of the RS232 interface
D: The beep(alarm) function can be activated when the weighing reach the HI –LO limits .Multiple
choices for checking functions .
E: Tare function
F: Pre-tare function
G: Automatic tare function
H: The weight accumulation function
2. Easy operating and water-resistant membrane keypad
3. Easy to read backlight HTN- LCD display.
4. Simple counting function is applicable.
5. Tare range is unlimited.
6. Bi-direction RS-232 signal output are applicable.
7. AC power or rechargeable battery is applicable. The battery’s checking system can
check automatically whether or not the battery is recharged.
8. Fast, stable and accurate display reading with simple operation.

1
3、PRECAUTION:
1. Full charge the battery after unpacking the scale.
Recharge the battery: When battery symbol appears on the LCD display, charge the battery with AC
power cord plug in, the indicator of charge will light up in red, when it becomes green means charge
completed. (It takes about 8 hours to full charge the battery.)
2. Install the equipment on a level and stable surface.
3. Do not install the equipment near the air conditioning or a vibrating machine.
4. Install the equipment in an environment with steady temperature(-5℃~40℃),
prevent from rapid temperature changes.
5. Independent AC outlet for this equipment is recommended, check the voltage before
plug in.
6. Warm up the equipment for 15 minutes before use.

4、CONNECTION of L/C and JWI-586:


E+ 1 5
LOAD-CELL
E- 2 4
S+ CONN
S-

3
Shield

PIN SIGNAL
1 E+
LOAD CELL 2 E-
CONNECTION 3 S+
4 S-
5 SHIELD

The solution of the LOAD CELL when its output is too high :
LOAD-CELL
E+
E- 追加電阻
S+
S- 50k~500k ohm

The solution of the LOAD CELL when its output is too low :
LOAD-CELL
E+
E- 追加電阻
S+ 50k~500k ohm
S-

2
5、KEYBOARD:

JWI-586 WEIGHING INDICATOR

CHARGER

ESC ENTER +
GROSS ON
SET PRINT TARE ZERO OFF
NET

MC MR M+ P.T. UNIT

ITEM OPERATION FUNCTION

ON/OFF Power switch


ON/OFF

ZERO Reset to zero


ZERO

TARE Tare
TARE

G.W./N.W. Change G.W & N.W


G.W./N.W.

PRINT Print
PRINT

SET Change the key and checking function setting


SET

CHANGE THE UNIT Change the unit


SET ZERO

PRE-TARE Per- tare


SET TARE

ACCUMULATION Accumulation
SET G.W./N.W.

ACCUMULATION DISPLAY Accumulation show


SET PRINT

CANCEL THE ACCUMULATION Cancel the accumulation


SET SET

+ : digit+1 ◄: digit displacement

▲: Function selection ENTER: go into the setup

ESC: leave the setup : press the key once

3
6、SETTING OF THE CAPACITY AND THE RESOLUTION:
1.Turn Power on while holding down PRINT key and the scale goes into the function
setting with POS2 shown on the screen.
2. Press PRINT key and then SET key to set the maximum capacity
3. Press G.W./N.W. key in sequence to set the capacity as diagram shown below ; when
FrEE appears on the screen ,press PRINT key and 000000 will shows on the
screen where you can set the maximum capacity according to what you need; pressing
◄ key enables the digit to move leftwards and + key enables the digit to increase;
press SET key to move to the next setting which is the setting of the resolution.
4. Press G.W./N.W. key to set the resolution [1.2.5.10.20.50] in sequence then press
SET key to set the next one that’s the location of the decimal fraction.
5. Press G.W./N.W. key to set the location of the decimal fraction in sequence. Then
press ZERO key and POS2 will shows on the screen. In the end press ZERO
key and this enables the indicator to return to the normal operation mode.
CAPACITY RESOLUTION
1Min. of 2 Min. of 5 Min. of 10 Min. of 20 Min. of 50 Min. of
Max
graduation graduation graduation graduation graduation graduation
300 1/300 ------------ ------------ ------------ ------------ ------------
400 1/400 ------------ ------------ ------------ ------------ ------------
500 1/500 ------------ ------------ ------------ ------------ ------------
600 1/600 1/300 ------------ ------------ ------------ ------------
800 1/800 1/400 ------------ ------------ ------------ ------------
1,000 1/1000 1/500 ------------ ------------ ------------ ------------
1,200 1/1200 1/600 ------------ ------------ ------------ ------------
1,500 1/1500 1/800 1/300 ------------ ------------ ------------
2,000 1/2000 1/1000 1/400 ------------ ------------ ------------
2,500 1/2500 1/1250 1/500 ------------ ------------ ------------
3,000 1/3000 1/1500 1/600 1/300 ------------ ------------
4,000 1/4000 1/2000 1/800 1/400 ------------ ------------
5,000 1/5000 1/2500 1/1000 1/500 ------------ ------------
6,000 1/6000 1/3000 1/1200 1/600 1/300 ------------
8,000 1/8000 1/4000 1/1600 1/800 1/400 ------------
10,000 1/10000 1/5000 1/2000 1/1000 1/500 ------------
12,000 1/12000 1/6000 1/2400 1/1200 1/600 ------------
15,000 1/15000 1/7500 1/3000 1/1500 1/750 1/300
20,000 ------------ 1/10000 1/4000 1/2000 1/1000 1/400
25,000 ------------ 1/12500 1/5000 1/2500 1/1250 1/500
30,000 ------------ 1/15000 1/6000 1/3000 1/1500 1/600
40,000 ------------ ------------ 1/8000 1/4000 1/2000 1/800
50,000 ------------ ------------ 1/10000 1/5000 1/2500 1/1000
60,000 ------------ ------------ 1/12000 1/6000 1/3000 1/1200
75,000 ------------ ------------ 1/15000 1/7500 1/3750 1/1500
80,000 ------------ ------------ ------------ 1/8000 1/4000 1/1600
100,000 ------------ ------------ ------------ 1/10000 1/5000 1/2000
120,000 ------------ ------------ ------------ 1/12000 1/6000 1/2400
150,000 ------------ ------------ ------------ 1/15000 1/7500 1/3000
200,000 ------------ ------------ ------------ ------------ 1/10000 1/4000
250,000 ------------ ------------ ------------ ------------ 1/12500 1/5000
300,000 ------------ ------------ ------------ ------------ 1/15000 1/6000
400,000 ------------ ------------ ------------ ------------ ------------ 1/8000
500,000 ------------ ------------ ------------ ------------ ------------ 1/10000
600,000 ------------ ------------ ------------ ------------ ------------ 1/12000
700,000 ------------ ------------ ------------ ------------ ------------ 1/14000
750,000 ------------ ------------ ------------ ------------ ------------ 1/15000

4
7、CONNECTION of the INPUT/OUTPUT DELICACY:
The loadcell input sensitivity is 0.13u V/D or more . The input sensitivity indicates the
variation in the loadcell output voltage required to change the display one point on the
display.; when designing a weighing system, the loadcell ’s output voltage must accord
with the input V/D of the weighing display so all must accord with the following formula .
In order to achieve a system with a stable performance , it should be designed so that
sensitivity is as great as possible.

One LOAD CELL IN USE:


Formula : 0.13 ≤ 【 ( 5000 × B × D ) ÷ A 】
A: LOAD CELL’s rated load capacity.
B: LOAD CELL’s rated output m V/V
D: Minimum divison
5000: LOAD CELL’s excitation voltage in milivolt / the weighing display offers 5V=5000m V
0,2 is the input sensitivity in V.

Multiples LOAD CELL(the number is N) IN USE:


Formula : 0.2 ≤ 【 ( 5000 × B × D ) ÷ ( A × N ) 】
A: LOAD CELL’s rated load capacity.
B: LOAD CELL’s rated output m V/V
D: Minimum divison
5000: LOAD CELL’s excitation voltage in milivolt / the weighing display offers 5V=5000m V
N: LOAD CELL’s number.
0,2 is the input sensitivity in V.

EXAMPLE:
LOAD CELL’s rated load 750kg,
LOAD CELL’s rated output 3m V/V
The scale need: the maximum capacity 300kg ; the minimum of resolution 0.05kg

Estimation as the following:

0.13 ≤【 ( 5000 × 3 × 0.05 ) ÷ 750 】


conclusion: A=750 B=3 D=0.05 N=1
All data comply with the formula and there is therefore no problem with the design of the weighing
application.

When a lever is used, the lever ratio should be taken into consideration.

n = The lever ratio 0.6 ≤ 5000 x B x D ÷ (A x N x n )

8、FUNCTION SETTING:
Turn power on while holding down SET key and the indicator goes into the
function setting with CAL shown on the LCD display. Pressing SET key in
sequence will allow you to move through the different settings .

5
LCD display Function illumination Selection of function setting Detailed

CAL Calibration Remark 1 of page 7

ON/OFFkg Selection of unit ON/OFF Remark 2 of page 7

Init = kg Selection of initial unit Kg/g/Ib/t Remark 3 of page 7


Default Kg
UM OFF Selection of HI / LO memory function ON/OFF Remark 4 of page 7
Default ON
Fil.1 Selection of digital filtering setup 1/2/4/8 Remark 5 of page 7
Default 2
Aut.No Selection of auto-power off NO/5/10/30/60 Remark 6 of page 7
Default 60
Lit.Aut Selection of backlight OFF/ON/AUT Remark 7 of page 7
Default Aut
Zero.0 Selection of zero range display 0/1/2/3/4/5 Remark 8 of page 7
Default 2
BAu.96 Selection of baud rate of RS-232 2400/4800/9600 Remark 9 of page 8
Default 9600
Prt.Pr Selection of print method Pr/St/Co Remark 10 of page 8
Default Co
Peri Remark 11 of page 8

PC Selection of external setting PC/TP/SH/BP-443D Remark11 of page 8


Default PC
Bp.Un Selection of beep of check function Un/In/no/Lo/3b/OFF Remark 12 of page 8
Default LO
Reset Return to the initial setting in the factory Remark 13 of page 8

TrA.ON Function of the zero tracking ON/OFF Remark 14 of page 8


Default ON
A.t.OFF Automatic tare function ON/OFF Remark 15 of page 8
Default OFF
M OFF Selection of checking & memory function ON/OFF Remark 16 of page 8
Default OFF

6
Remark 1 :
When CAL shown on the screen, no object must be on the pan .press TARE key
to perform the zero calibration , zero CAL will shown on the screen. When zero point
calibration is completed the screen will show next 0000.00 kg for the span calibration.
Press ◄ key to select the digit and + key to change the weight. Put the
corresponding weighing mass (number represents the weight of the weighing mass) on
the pan and then press SET key to end the setting. CAL kg shown on the screen
means the scale goes into the SPAN calibration, when beep sounds and with
PASS kg shown on the screen means the completion of span calibration . Then press
ZERO key to leave the setting function and return to the normal operation mode; press
SET key to set the following function continually.

Remark 2 :
ON kg shows on the screen and then press TARE key to set the unit in
sequence . ”ON” means this unit is in use; press G.W/N.W. key and OFF kg shows
on the screen. ”OFF” means this unit isn’t in use. Press G.W/N.W. key can switch “ON”
and “OFF”

Remark 3 :
Init= kg shown on the screen represents the present initial unit. Press TARE key to
select the wanted unit . and the chosen unit will appear on the screen when you power on
next.( pcs and % cannot be the initial unit).

Remark 4 :
UM OFF shows on the screen and press TARE key to select the memory function
of HI / LO ( pcs and % cannot be the initial unit).”NO” or “YES” can be chosen.

Remark 5 :
Fil 1 kg shows on the screen and press TARE key to select 1 or 2 or 3 or 4, the
larger number means slower reaction speed but better filtering effect.

Remark 6 :
Aut.NO kg shows on the screen and press TARE key to set the time of power off
automatically. there are 5,10,30,30,60min to be chosen.

Remark 7 :
Lit.Aut shows on the screen and press TARE key to select in sequence the
backlight options. There are ON,OFF, Auto (light up only if the weight is more than 9
times of the resolution) to be chosen..

Remark 8 :
ZERO.0 shows on the screen and press TARE key to set the zero range ; there
are 0~5 classes to be chosen(the bigger, the wider).

7
Remark 9 :
bAu.96 shows on the screen and press TARE key to set the baud rate ; there are
2400,4800 and 9600 to be chosen..

Remark 10 :
Prt.Pr shows on the screen and press TARE key to set the print mode . there are
Pr,St,Co to be chosen.(Pr means data sent when key pressed; St means data sent when
the reading is stable; Co means data sent continuously.

Remark 11 :
Peri then PC show on the screen and press TARE key to set the
external equipment . There are PC,SH-24,BP443D to be chosen.

Remark 12 :
BP.Un shows on the screen and press TARE key to set the checking function .
There are Un,In,no,Lo,3b,OFF to be chosen.
Un – HI Beep sounds (when reading are over the setting of Hi limit).
In –OK Beep sounds( when reading are within or equal to the setting of Hi and Lo limit.)
No – Beep sounds when reading are out of the setting of Hi and Lo limit .
Lo – Beep alert when reading are less than the setting of Lo limit.
3b---three phases checking(beep silent)
OFF—function doesn’t work

Remark 13 :
RESET shows on the screen and press G.W./N.W. key to return to the initial
situation setting from the factory.

Remark 14 :
tr A.ON shows on the screen and press TARE key to set zero tracking function .
There are ON , OFF to be chosen.

Remark 15 :
At. OFF shows on the screen and press TARE key to set the memory of the
Auto –Tare function . There are ON , OFF to be chosen.

Remark 16 :
M OFF shows on the screen and press TARE key to set in circle the memory
function or not. There are ON , OFF to be chosen.

Remark :
After the above setting, press ZERO key can end the setting function and turn into the
normal weighing operation mode automatically; or press SET key to return to the first
step after remark 16.

8
9、OPERATION(NORMAL POWER ON)
Remark:
Single calibration must be processed before the normal power on. The period between
the first SET key and the next key must be within 3 seconds, otherwise only the function
of the SET key is acknowledged
Unit Switch
Press SET key and then ZERO key to switch the weighing unit.

Pre-tare:
Press SET key and then TARE key to turn into the pre-tare mode with XXXXXkg
appearing on the screen .press ◄ key to choose the digit and + key to change the
weight. after the setting, press PRINT key to enable the scale to go into the operation
mode . Pre-tare value appear in the display .

Accumulation:
Press SET key and G.W/N.W key and then XXXXXX
M+ kg shows on the
screen(accumulation cannot process until the scale is stable with load on the pan)

Accumulation display:
Press SET key and PRINT key and then zero ALLXX kg shows on the screen. The
next screen display XXXXXX kg and after a while the scale return to the operation
mode.

Cancel the accumulation:


Press SET key and then SET key to cancel all the accumulation value.

Simple counting:
Press SET key and then ZERO key in sequence and wait for 0 pcs
zero appearing on
the screen. Then press G.W./N.W. key for 3 seconds and S=100pcs shows on the
screen; press G.W./N.W. key to select 25,50,100 as the three kinds of sampling. Put
the samples on the pan and press SET key .counting begins when CAL
appears on the screen.

Percentage:
Press SET key and then ZERO key in sequence and wait for zero 0 % appearing
on the screen Then press G.W./N.W. key for 3 seconds and S=100% shows on the
screen; press G.W./N.W. key to select 25,50,100 as the three kinds of sampling. Put the
samples on the pan and press SET key calculation of percentage begins when CAL
appears on the screen.

9
Checking:
Press SET key for 3 seconds or more and FUN appears on the screen; press ▲
key to switch Hi, Lo , NC and then press PRINT key to go into the checking setting;
press ◄ key to choose the digit and + key to change the weight. Press SET key to
return to the FUN setting after finishing the setting; pressing ZERO key enables the
indicator to the operation mode. Hi value must be bigger than Lo value, or Error9 will
appears. NC means to cancel the checking setting.

10、UNIT CONVERSION
1 Ib=453.59237g 1 t=1000kg=1000000g
1 jin=500g
1 HK-liang=37.799375g 1HK-jin=16HK-liang
1 TW-liang=37.49995g 1TW-jin=16TW-liang

11、OUTPUT OPERATION
OP-01 : RS-232
OP-02 : Large LED display

12、SPECIFICATION:
TYPE JWI-586
Input sensitivity 0.13u V/DIV
Input the range of the voltage -0.5m V to 16.5 mV
LOAD CELL Excitation voltage 5V DC ±5%
LOAD CELL driving capability Drive Up to max. 8 loadcells (350 ohm ,120m A)
Non-linearity 0.006% of full scale
Input the resistance 10M ohm or more
A/D converting ∆-∑
A/D internal resolution 700000 count
A/D conversion rate 8 times / set
Out display resolution 15000 count
LCD digit 6 digit
KEY amount set, print, g. w./n. w. , tare, zero, on/off
Power standard 110V or 220V & 6V 4A Recharge BAT

10
13、Appendix 1:

1、Error message
Message Problem operation
Err2 Initial zero point over+10%
-Err2 Initial zero point over-10%
Err Over/low A/D resolution range
Over Accumulation exceeds 99 times
Err4 EEPROM Chksum error
Err5 Over load (max: capacity +9e)
Err6 Wrong weight value keyed in when performing calibration
Err7 Setting of capacity and resolution incorrect; (resolution≥15000 or
capacity and resolution can’t be divided directly)
Err8 Setting of operation unit and initial unit incorrect; (only % and pcs are
settled as operation)
Err9 Set LOW value ≥ Hi value
Battery symbol Low battery alarm

2、Trouble shooting:
Where Possible Error message Trouble shooting
Power on Err2,-Err2 Check and remove the object from weighing
pan or malfunction of LOAD CELL
Power on Err3,-Err3 Check if A/D or LOAD CELL malfunction
Power on Err4 Beep alarm. switch power off and power on
again, or make the calibration
Power on Battery symbol Charge the battery
Normal weighing mode Err5 Check if weighing object over the capacity
+9e
calibration Err6 Automatic calibration or change value

14、Appendix 2:
RS-232 OUTPUT FORMAT
Baud Rate : 2400、4800、9600
Data Bit : 8
Parity : N ( None )
Stop Bit : 1
Code : ASCII
Bit Format :
LSB MSB

0 1 2 3 4 5 6 7 8

Start Stop
Parity
Bit Bit

11
Data Format:
1、 kg
G/N . W . : +/- k g CR LF
Weight
Example
N.W. : +2.2352kg
G.W. : +2.2352kg

G=GROSS N=NET

RS-232 INPUT FORMAT


Baud Rate: 2400 4800 9600
Data Bit : 8
Parity : N(None)
Stop Bit : 1
Code : ASCⅡ
Command on computer:
‘R’=READ ‘Z’=ZERO ‘T’=TARE

Printing Format (example : kg)


SH-24 BP-443D
T.W.:+0.0000kg T.W.:+0.0000kg
N.W.:+1.6025kg N.W.:+1.6025kg
G.W.:+1.6025kg G.W.:+1.6025kg

RS-232 Connector

1
14
2
TXD1 SCALE to PC
15
3
RXD PC to SCALE
16
4
17
5
18
6
19
TXD2 SCALE to PC
7
20
8
21
9
22
10
23
11
24
12
25
13

CONNECTOR DB25

12
15、APPENDIX 3:
ASSEMBLY

INTRODUCTION OF PARTS
(1)stem base (2)stem (3)bottom bracket (4)upper bracket
(5)display (6)screw (7)screw (8)plastic knob
(9)slot (10)channel (11)bolt

n Drill the Load Cell thread on the stem base(1) through the stem(2),insert
the stem (2)in the hole of the screw base and lock the two screw(6)
tightly.
n Drill the Load Cell thread through the bottom bracket (3), insert the
bottom bracket (3) in the stem(2) and lock the screw(7) tightly.
n Drill the Load Cell thread through the upper bracket (4),jointing the
display(5), and insert the upper bracket (4) in the slot (9) to the slot’s half
deepness【the thread should be outside the upper bracket】.
n Lay the Load Cell thread in the channel(10),assembling the bottom
bracket (3) and the upper bracket (4) then inserting the upper bracket (4)
in the slot (9) entirely, In the end lock the bottom bracket (3) and the
upper brackets(4) via the plastic knob(8) and then insert the bolt in.
n Use the plastic knob(8) to adjust the display to the optimum situation and
then lock close to fix up.

13