BA-88A Service Manual
BA-88A Service Manual
BA-88A Service Manual
1. EQUIPMENT CONFIGURATION................................................................................................................ 4
ASTM E1381-91 ” Specification for Low-Level Protocol to Transfer Messages Between Clinical Laboratory
Instruments and Computer Systems ”
ASTM E1394-91 ” Standard Specification for Transferring Information Between Clinical Instruments and
Computer Systems ”
ASTM E1238 ” Standard Specification For Transferring Clinical Observations Between Independent Computer
Systems ”
LIS
Connection
ASTM 1381
or HPRIM1.2
Follow the codification specified in the reference documentation, which is complemented by the
following :
3.1.1. QUERY :
If kryptor is running in Query mode , the requests are formatted as follow :
HEADER
H|\^&|||KRYPTOR^AUTOMATE KRYPTOR|||||LIS||P|1|199709010810<CR >
QUERY
Q|1|^123456||ALL||||||||O<CR>
nd
Field 12.3, 2 component : sample ID read on the carousel.
TERMINATOR :
L|1|F<CR>
Only one request record may be outgoing at a time. The receiver of the request must give a
response to the query . Kryptor will wait for this response before sending the next message.
HEADER :
H|\^&|||LIS|||||KRYPTOR||P|1|19970901103100<CR >
PATIENT :
P|1|REF_PATIENT|||NAME of the PATIENT^First name of the patient|||M<CR >
Patient ID, patient name ( truncated to 13 characters ) and comments are downloaded to the
KRYPTOR, but these fields are not returned with the results upload.
The other fields ( field 8.8 : date of birth, field 8.9 : sex...) are not used in this configuration.
TEST REQUEST :
Batch mode :
O|1|01201^01^10||^^^CEA^^1\^^^AFP^^2^19970512V26.1\^^^LH|R||||||A||||||||||||||O<CR >
If a response exists for the requested sample , the answer is the same as a request message in batch
download mode except for field 9.26 :
O|1|01201||^^^CEA^^1\^^^AFP^^2^19970512V26.1\^^^LH|R||||||A||||||||||||||Q<CR >
O|1|01201|||||||||||||||||||||||Z<CR >
TERMINATOR :
L|1|F<CR >
HEADER
H|\^&|||KRYPTOR^ KRYPTOR ANALYSER|||||LIS||P|1|199709011410<CR >
PATIENT :
P|1|||||||U<CR >
In this configuration, the PID et PATIENT NAME are not sent back (uploaded).
ORDER :
O|1|03104^01^04||^^^NSP^^5|R||||||A||||||||01^04||||||F<CR >
The unique test code is sent back.
Cassette and position are returned in 9.3 and 9.20 fields.
COMMENT (optional) :
C|1|I|Comment about this order<CR>
Comment as it was manually entered on the analyser.
RESULT :
R|1|^^^NSP^^5^^F|8.123|||L||F|||19970901133025|19970901140910|517017000187<CR >
A comment segment detailing Kryptor error codes follows this result segment
Field 10.12 : date and time on which the test has been started
Field 10.13 : date and time on which the test has been performed
Field 10.14 : Kit Id
COMMENT (optional) :
C|1|I|33\39<CR>
TERMINATOR :
L|1|F<CR >
Several samples for several patients may be included in the same ASTM message.
Query
<ENQ>
<ACK>
Response to Query :
<ENQ>
<ACK>
<EOT>
New Query
<ENQ>
<ACK>
Response to Query :
<ENQ>
<ACK>
<EOT>
<EOT>
<ENQ>
<ACK>
<STX>4O|1|02315000^01^10||^^^CEA^^1^19970512V46.2\^^^FSH^^^19970104V450|R||||||A||||||||||||||O<CR >
<ETX>xx<CR ><LF >
<ACK>
<EOT>
Results upload :
<ENQ>
<ACK>
<STX>3O|1|02315000^01^10||^^^CEA^^1|R||||||A||||||||01^10||||||F<CR ><ETX>xx<CR
><LF >
<ACK>
Contention :
The instrument system has priority to transmit information. When the LIS sends an <ENQ> to begin a
transmission and receives <ENQ> for response, KRYPTOR takes priority.
The LIS system will have to stop trying to transmit, and prepare to receive by sending <ACK>
Retries :
3 retries maximum after receiving a NAK. After 3 retries, the message upload is aborted , and the
sender ends the communication through sending a ” EOT ”.
In batch mode, the number of orders in the same communication frame is limited to 10
(between ENQ and EOT).
KB169 G KIM Host Lab Instructions Page 11 /17
4. HPRIM 1.2 transfer, Configuration 2
Follow the codification specified in the reference documentation, which is complemented by the
following :
HEADER :
H|^~\&|||LIS2 ||ORM|||KRYPTOR ||P|H1.2|199709011213<CR>
field 7.5 : ” LIS2 ” Name of the host
field 7.10 : ” KRYPTOR ” Name of the receiver
field 7.12 :”P” Processing ID = (P)roduction
field 7.13 : ” H1.2 ” Version
field 7.14 : YYYYMMDDHHMM Date and time of the request
PATIENT :
P|1|PID123|||SMITH^Arnold|||U<CR>
Field 8.3 : PID, taken into account and sent back, if specified
Field 8.5 : Additional patient identification, sent back, if specified.
Field 8.6 : Patient Name, sent back if specified
Field 8.8 : date of birth, taken into account but not returned
Field 8.9 : sex, taken into account but not returned
TEST REQUEST :
OBR|1|01280^REQUEST_NUMBER||02020~01170~03680|R||||||N||||1||||01^05<CR>
The dilution cannot be specified and the default dilution (1) is used.
Field 9.16 : specifies the sample type (serum, plasma, urine, etc.) - optional
TERMINATOR :
L|1|F<CR >
HEADER
H|^~\&|||KRYPTOR^KRYPTOR ANALYSER||ORU|||LIS2||P|H1.2|199709011455<CR>
PATIENT :
P|1|PID123|||SMITH^Arnold|||U<CR>
The patient data : PID, patient name are sent back.
ORDER :
OBR|1|01201^REQUEST_NUMBER||02020|R|||||| N||||||||01^05<CR>
The unique analyte code is uploaded.
Cassette and position are sent back ( non operationnel with version 2.03 )
RESULT :
OBX|1|NM|02020||0.63|||>|||F||19970901151000|19970901152010|C16034123432<CR>
Field 10.15 : date on which the test has been achieved on the analyser.
Field 11.4 : Error flags given by KRYPTOR, separated by a repeat delimiter ( the max length is 20
characters )
TERMINATOR :
L|1|F<CR >
Download :
H|~&\^|||LIS2||ORM|||KRYPTOR|||H1.2|199709011711
P|1|12345678|||DURIEUX~Jocelyne~F~~Mlle||19581105|F|||||||||||||||||||||R.A.S
OBR|1|01000106~Request number 1||04960|||||||N||||1~0~0
P|2|123456-0023456|||MARCELIN~Jean-Pierre~~~MR||19560613|M
OBR|1|01000107~Request number 2||04960|||||||N||||1~0~0
L|1|F
Result upload :
All results are given as displayed on Kryptor result list interface. Unit can be modified in Kryptor
analyte preferences.
Analyte names can be converted, by using the KLIS.INI file.
Code Meaning
0 System Warning : Low Laser Power
1 System Error : Reaction Plate Position
2 System Warning :Reagent Cooler Out Of Range
3 System Warning : Incubator Out Of Range
4 System Error : Missed Reading
5 System Error : Unknown
6 System Error : Insufficient Sample Volume
7 System Error : Insufficient Reagent Volume
8 System Error : Insufficient Diluant Volume
9 System Warning : Late T0
10 System Warning : Late T1
11 System Warning : Late TE
12 System Error : Pipetting
13 System Error : Clot Detected
14 System Error : Pipetting in Sample Area
15 System Error : Pipetting in Dilution Area
16 System Error : Pipetting in Reagent Area
17 System Error : Pipetting in Wash Area
18 System Error : Pipetting in Reaction Area
19 System Warning : Heated Tip Out Of Range
20 System Error : Interrupted
21 System Warning : Not Performed
22 Calibration : Removed by User
24 Check Results
25 Math Error
26 Reflex Test Launched
27 Rerun by Reflex testing
28 Generated by Reflex testing
30 Data Error : Unknown Error
31 Data Error : Ratio
32 Data Error : Response
33 Data Warning : Abnormal
34 Data Warning : Out Of Range
35 Data Warning : Detection Limit
36 Data Warning : Above Max Range
37 Data Warning : Standard Warning
38 Data Warning : Calibrator Warning
39 Data Warning : Below Normal
40 Data Warning : Above Normal
41 Data Warning : Imposed
42 Concentration not consistent with dilution used
43 Data Warning : Control Warning
44 Data Warning : Abnormal Kinetics
45 Data Warning : Internal flag
46 Data Warning : Internal flag
47 Data Warning : 25 mUI hCG
48 Data Warning: Control out of 2 SD range
51 Inconsistent incubation time
52 System Error: Missed Flashes
69 Pre-incubating in process
70 Error during phase 2 of pre-incubating