Safety notice
To operate the Instrument safely and effectively, be sure to read the following notices first. Operating
the instrument, without following the appointed methods by the manufacturer, may break down the
defensive function of the system, and cause bodily injury or damage the instrument.
➢ Avoid electric shock 1) When the power-on, the unauthorized maintenance personnel should not pen
the instrument.
2) If liquid enters into the instrument or the instrument leakage, please shutoff the power
immediately, and contact with our company’s Customer Service Department or local distributor in due
course. Improper use of the liquid may cause electric shock and result in damaging the instrument. ➢
Defense for biohazard and chemical hazards Improper use of the sample may result in being infected.
Avoid hand touch the sample, reagent and waste directly. When operate the instrument, be sure to wear
gloves to avoid being infected.
If the skin touched the sample, manage it according to operator working standard or
inquiring a doctor to take remedial measures immediately. Use the reagent carefully to avoid hand and
clothes touch directly.
-4-
Vec 3 Diff Operation Manual Edition: A/0
Once hand or clothes touch, flush the touched area with soap and plenty of water immediately.
If the reagent enters into the eyes incautiously, flush with plenty of water at once, and inquire a doctor
for further treatment.
Be sure not to use flammable dangerous materials around the instrument. Reagents, quality control
serum, some materials in the sample are controlled by Pollution Regulations and Standard for Discharge
of Pollutants. Please abide by the local discharge regulation and inquire related reagent manufacturer.
➢ Prevention of fire and explosion
Operation notice
To operate the instrument safely and reliably, be sure to obey the following notices.
• Purpose
Pay attention to the application range of the instrument statement. Make sure your use does not
exceed the application range.
• Limitation of operating environment The instrument should be installed according to the required
installation environment of the manual. Installation and use out of the appointed range may cause
unreliable results, and make the instrument damaged. Contact with our company’s Customer Service
Department or local distributor if you need to change the state of the instrument.
• Limit of Operator Only trained and authorized personnel by the manufacturer can operate the
Instrument.
• Maintenance and Service Be sure to service and maintain the instrument according to this manual
strictly.
-5-
Vec 3 Diff Operation Manual Edition: A/0
1.2 Structure
Host and accessory. 1.2.1 Host
Control the process of sample collection, dilution and analysis, including the following units:
1. Power Supply unit. 2. Central control circuit unit. 3. Dilute unit. 4. Display unit. 5. Thermal recorder.
1.2.2 Accessory
Including the following accessories: 1. PS/2 mouse. 2. PS/2 keyboard. 3. Printer (Optional) 4. Touch pen
5. Power cord. 6. Ground wire.
1.3 Purpose
It is used for detecting the parameters of RBC, WBC, HGB and differential counter. The analyzer is
used for the determination of the following 20 parameters and 3 histograms blood .
-6-
Vec 3 Diff Operation Manual Edition: A/0
1.4 Specifications
• Standard Classification of the Instrument According to anti-electrical shock hazard: Grade II, Pollution 2
According to the defense for deleterious liquid: General Device(Closing device of non-defensive liquid)
According to the recommended disinfection/sterilization method of the manufactory: Disinfectant device
recommended by manufactory.
According to safety degree in condition that using gas mixture of flammable anesthetic gas and air or
gas mixture of oxygen and nitrous oxide: Do not use the equipment in condition that using gas mixture of
flammable anesthetic gas and air or gas mixture of oxygen and nitrous oxide.
According to working status: Continuous Running Equipment.
• Principles of Measurement Blood Cells are counted and sized by Electrical Impedance Method.
Hemoglobin is determined by Colorimetric Method.
• Parameters of Measurement Basic parameters:
-7-
Vec 3 Diff Operation Manual Edition: A/0
-8-
Calculated Parameters:
Sampling Features:
Sample Volume: Venous Blood Mode: 10uL Venous Blood
Capillary Blood Mode: 10uL Capillary Blood Prediluted Mode: 20uL Capillary Blood
Vec 3 Diff Operation Manual Edition: A/0
Reagent Volumes Required for Single Sample:
Diluent: 26mL Lyse: 0.35mL Venous and Capillary Blood Mode:
WBC/HGB 1:285 RBC/PLT 1:21667 WBC/HGB 1:500 RBC/PLT 1:43333 Cell Counting Aperture Size:
WBC: 100μm RBC/PLT: μm Throughput: Up to 60 samples per hour Working mode: Double counting
pathway +independent HGB detection system Display Liquid Crystal Display (LCD), Touch screen,
resolution: 800*600 Language English Chinese Spanish Indicator 20 Parameters and 3 Histograms
Display μ Alarm Indicator Working Status (including Power) Indicator Light Alarm Interface One power
input socket One printer port One RS232/C serial ports One VGA interfaces Two USB interfaces Two
PS/2 interfaces Recorder Rapid Thermal Recorder: Recording width: 35mm
Paper width: 80mm Precision Specifications of the instrument
Parameter Repeatability
WBC
≤ 5 × 109 / L
RBC
≤ 5× 1012/ L
HGB
≤ 4g / L
HCT
≤5%
PLT
≤8× 109/ L
-9-
Vec 3 Diff Operation Manual Edition: A/0
Linear Range
Parameters Measuring range
Measurement of the linear tolerance
WBC
0~5.9×109/L 6.0~99.9×109/L
- 10 -
(±0.3×109)/L ±5%
RBC
0~0.99×1012/L 1.0~9.99×1012/L
(±0.05×1012)/L ±5%
HGB
0~99g/L 100~300g/L
±2.0 g/L ±2%
PLT
0~99×109/L 100~999×109/L
(±8.0×109)/L ±10%
Operating Ambient Temperature: 18°C~35°C Humidity: 10%RH ~90%RH Atmospheric Pressure: 70.0kPa
~106.0kPa Ambient Storage Temperature: -10°C~40°C Humidity: 10%RH ~90%RH Atmospheric
Pressure: 50.0kPa ~106.0kPa This instrument should be placed at a good ventilation area. Electric
Specifications Power Supply: 100~240Vac,50/60Hz
Vec 3 Diff Operation Manual Edition: A/0
Fuse: F3AL250V Dimensions and Weight Dimensions: 410mm×435mm×472mm Weight: 25kg
1.5 Structure
1.5.1 Front Panel
figure 1-1 1.5.2 Rear panel
figure 1-2
1.6 Operation
Designed according to human engineering and transferring information to users by software windows
and mouse, both are convenient for your operating.
- 11 -
Vec 3 Diff Operation Manual Edition: A/0
LYM: 35-90 fL MID: 90-160fL NEUT: 160-450fL 1.7.4 Results and Calculation of
Parameter Measurement
The parameters of the sample blood test can be described in three ways: a) Measured directly, such as
WBC, RBC, PLT, and HGB. b) Derived from histograms, such as LYM%, MID%, NEUT%, MCV, RDW-
SD,
RDW-CV, MPV, PDW, P-LCR. c) Calculated, such as LYM#, MID#, NEUT#, HCT, MCH, MCHC,
PCT. The derivation of the formularies as follows: MCV derives from histograms and the instrument
collecting and classifying the erythrocytes according to its volume. It is determined by measuring the
average volume of individual erythrocytes, and the unit is fL.
RDW represents the volume distribution of the erythrocyte populations, derived from the RBC
histogram. It can be expressed by the coefficient variation of the erythrocyte Volume as RDW-CV and the
unit is %.It can also be expressed in standard deviation of the erythrocyte volume as RDW-SD, the unit is
fL.
MPV is the average volume of individual platelets, derived from the PLT histogram. It represents the
mean volume of the PLT populations and can be expressed in fL.
PDW derives from the PLT histogram. It represents the geometry standard deviation (10 GSD) of the
volume of the PLT populations.
P-LCR derives from the PLT histogram. It represents the ratio of the larger PLT and the unit is %.
• HCT (%) = RBC*MCV/100
• MCH (pg) = 10 × HGB/RBC
• MCHC (g/L) = 100 ×HGB/HCT
• PCT (%) = PLT ×MPV/100
• LYM%=100×AL/(AL+AM+AG)
• MID%=100×AM/(AL+AM+AG)
• NEUT%=100×AG/(AL+AM+AG)
• LYM#=LYM%×WBC/100
• MID=MID%×WBC/100
• NEUT#=NEUT%×WBC/100 AL: Number of cells in LYM area; AM: Number of cells between
lymphocyte and Neutrophil area; AG: Number of cells in NEUT area.
- 15 -
Vec 3 Diff Operation Manual Edition: A/0
Chapter 2 Installation
2.1 Package
If packing damage is discovered after receiving the instrument, or the instrument is badly damaged,
contact with the freight agent immediately to file a claim according to the damage level. At the same time
contact with your supplier to make sure the packing is complete. Then unpacking and installing the
instrument as the following steps.
2.2 Unpack
Take out the instrument and accessories from the packing case carefully. Preserve the packing
material for future transportation or storage.
1. Check the accessories according to the packing list. 2. Check whether there are mechanical damages
on the instrument and accessories IF there is any problem, please recover the package and contact with
your supplier immediately.
3.2 Startup
If external printer is equipped, turn on the printer power r and ensure they are in ready status.
Press the power button on the rear panel of the instrument, then the power lamp on the front panel of
the instrument lights on. The instrument automatically performs initialization program.
After completing initialization, the instrument will access self-test window. It tests the working
conditions of each part and if there is enough diluent, and lyse. Simultaneously prime and clean the
tubing.
After self-test, the instrument will access blood cell analyzer window, as shown in Figure If the system
detects malfunction, the failure information will be displayed on Information Section.
Figure3-1 The button functions are as follows: Info: Enter the
information window of the next sample, input and modify the information. Flush: To get rid of the aperture
clog. Print: Output the data of the current sample to the appointed device. Menu: Enter the main menu
window.
- 19 -
Vec 3 Diff Operation Manual Edition: A/0
3.2.1 Function
In the blood cell analyzer window, press "func" button, enter the “function“ window.
figure 3-2 The button functions are as follows: Sample Type:
Choose the sample mode among Venous Blood Mode, Capillary Blood Mode and Prediluted Mode.
Diluent: Inject the diluent quantificationally by sample needle. This menu uses only in Prediluted Mode.
Histogram: Enter the histogram adjustment window of the current sample. You can manually adjust the
histograms of the current sample.
3.8 Shutdown
Shutdown routine must be performed before turning off the power everyday. During shutdown procedure,
the instrument will perform daily maintenance and clean the tubing automatically. The procedures are:
a) In the blood cell analyzer window, press ”MENU” button. in the window,
Select "Shutdown" button, pops up shutdown window
- 25 -
Vec 3 Diff Operation Manual Edition: A/0
b) If you do not want to shutdown the instrument for the moment, click "Cancel" to
return to MENU window. c) Press "OK", the instrument will perform daily maintenance and clean the
measuring tubing. When the shutdown procedure finished, the screen displays "Turn off the power now",
turn off the power on the rear panel of the instrument. d) Turn off the printer power (if equipped), clean the
workbench and dispose the
waste.
Notice:
Do not turn off the power of instrument directly while performing the shutdown procedure.
- 26 -
Vec 3 Diff Operation Manual Edition: A/0
Chapter 5 Calibration
The instrument has been calibrated strictly at the factory. You may need to perform calibration
procedures when you replace any component that involves the primary measurement characteristics, or
when the shift occurs during controls. Calibration of the instrument is to provide the assurance that the
instrument is providing results with accuracy as design. The purpose of calibration is to ensure the
accuracy of the measurement result meet the requirement of the design all the time. To ensure the
accuracy of the instrument and obtain reliable measurement results, it is necessary to calibrate the
instrument in the following situations:
a) Installation for the first time or re-setup in a new place. b) The instrument is maintained. c) The result of
quality control is abnormal. d) Replace reagent. To ensure the instrument's precision and obtain reliable
measured results, the instrument should be calibrated in these situations: Warning:
Calibrators should be stored and used properly according to the calibrator manual. Warning:
Make sure the instrument is in a normal state before calibration. Warning:
Avoid using the result of the measurement on medical test or clinical before the calibration is finished
accurately. Check the instrument and reagents carefully before calibration. Make sure the instrument in
normal status, and make sure the required sample mode of the measurement.
Chapter 6 Setting
The instrument has accomplished all the settings at the factory. For customer's convenience, most of
the system parameters can be setup by the operator so as to meet various requirements. Notice:
The operations in this chapter will change the running state of the instrument. Please make sure the
necessity of the changing before operating.
6.2 Printout
In the blood cell analyzer window, press “menu” button, in the window ,Select " setting" in the menu,
Select Print". The instrument will enter the print window. As shown in figure6-2;
- 35 -
Vec 3 Diff Operation Manual Edition: A/0
:
figure 6-2. Click the menu box you want to adjust, pops up the
optional window .Click the item you want to set.
Press "OK", the instrument will save the current parameters and return to the blood cell window. Press
"Cancel" button, the system will not save the current parameters and return to the main menu window.
➢ Print select Select Recorder, the data will only be transported to the recorder. Select Serial port, the
data will only be transported to net. Select Recorder and Serial port, the data will be transported to the
recorder and net at the same time.
Select Laser Jet, the data will be transported to the HP Laser Jet,1018 printer or HP Laser Jet1008
printer
Select Desk Jet, the data will be transported to the HP desk Jet,1668printer or HP Laser Jet
2668printer
Select Network print, the data will be transported to the computer which connect with HP Laser
Jet1008 printer Notice:
There are different formats when selecting different printer settings.
➢ Print caption Can easily printout the headline on the report.
➢ Print type User can choose the type of the printer. Notice:
The selected printer type must be accord with as the used one, otherwise it will not be able to print
normally.
- 36 -
Vec 3 Diff Operation Manual Edition: A/0
6.3 Control
In the blood cell analyzer window, press “menu” button, in the window, Select "setting" in the menu,
Select " Control", enter the control setting window. As shown in figure 6-3.
figure 6-3. Click the menu box you want to adjust, pops up the
optional window .Click the item you want to set.
Press "OK", the instrument will save the current parameters and return to the blood cell window.
Press "Cancel" button, the system will not save the current parameters and return to the main menu
window. ➢ Auto Set "Print" as "ON", the system will print the test report automatically after the sample test
finished.
Set "Dorm" as "OFF", the instrument will keep working status all along. Set "Dorm" as data. When the
time, which the user does not perform any operation, exceeds the setting time, the instrument will access
sleep state. During sleep time, the instrument will turn off part of the circuit power, and take the sample
needle back. Sleep function can greatly save electrical energy and extend the life of the instrument. Press
the left key of the mouse or any key on the keyboard, sleep will be canceled.
Set "Clean" as "OFF", the instrument will not perform auto-cleaning. Set "Clean" as data. When the
running time of the instrument exceeds the setting time, the instrument will perform "Clean" procedure
once. It can ensure the measured tubing of the instrument clean and prevent the possible measurement
malfunctions and maintain the instrument.
➢ Sound Set "Alarm" as "OFF". When the instrument appears alarm, it will not make alarm sound, but
display the alarm information in the Information Section.
Set "Alarm" as "ON". When the instrument appears alarm, it will make alarm sound and display the
alarm information in the Information Section.
- 37 -
Vec 3 Diff Operation Manual Edition: A/0
Set "Prompt" as "OFF". When the sample measurement finished, the instrument makes no sound.
Set "Prompt" as "ON". When the sample measurement finished, the instrument will make a prompt
sound like "di".
➢ Display Set "Para" as "English". In parameter area of the blood cell analyzer, the parameters will be
displayed in abbreviations.
Set "Limits" as "No Display". In parameter area of the blood cell analyzer, the parameters will not
display reference value of the current sample.
Set "Limits" as "Low To High". In parameter area of the blood cell analyzer, the reference values will
be displayed from low to high.
Set "Limits" as "High To Low". In parameter area of the blood cell analyzer, the reference values will
be displayed from high to low.
6.4 Limits
In the blood cell analyzer window, press “menu” button, in the window, Select " setting" in the menu,
Select " Limits", enter the limits setting window. As shown in figure 6-4.The 8 Group reference values are
displayed at the left side of the screen. The dot in front of the reference value represents that this is the
current selected reference group.
The limits of the reference value are displayed on in the limits boxes on the screen. Click the dot in
front of the group to select the group required for setting.
figure 6-4. Click "Default" button, the instrument will select the
default reference values to replace the current reference values. Click the adjustable bar of the limits box
to adjust the upper and lower limit of the reference value.
Function of menu and shortcut keys: Save: Save the reference values of the current group.
- 38 -
Vec 3 Diff Operation Manual Edition: A/0
Print: Print the reference values of the current group. Back: Quit from current window, and return to main
menu window. Notice:
The default reference value may not suitable for the local conditions. User can modify it according to
the local specific circumstances. Caution:
The changing of the reference value will cause variation of abnormal prompt on hematology target.
Please make sure the necessity of the variation.
Chapter 7 Review
The instrument can automatically store the results after each sample analysis. The operator can
review, search, modify, print, and delete the data.
Chapter 8 Service
This chapter gives the description of various service functions provided by the instrument. These
functions will bring user convenient and pleasure. In the blood cell analyzer window, press ”menu” button,
in the window, Select " Service" in the menu.
figure 8-1
8.1 Flush
This function is used for solving general probe clogs. The instrument makes a fixed pressure and
voltage act on the probe to get rid of the clogs.
8.3 Cleaning
This function is used for the routine cleaning of the probe, sample needle, metering tube
measurement tubing.
8.5 Park
Manufacturer suggests you do not use this function .If the instrument is not going to be used in 2 week or
longer, use this function to clean and empty the instrument. It is convenient
- 48 -
Vec 3 Diff Operation Manual Edition: A/0
for user to store the instrument. Prepare distilled water before performing the function. .
8.9 Maintenance
User can finish the weekly maintenance according to the prompt information on the screen. Prepare
probe cleaner before performing the function.
8.10 Mechanic
Enter mechanic window. User can complete the detection of the valve, motor, circuit and interface.
It is helpful to the insurance of the malfunction.
figure 8-2 Valve:
"ON" represents the valve is on. "OFF" represents the valve is off. Motor:
"OK" represents that the motor is in normal status. "FAIL" represents the motor is failure. Recorder:
Click this button, the recorder will print a test paper Printer:
- 49 -
Vec 3 Diff Operation Manual Edition: A/0
Click this button, the external print will print a test paper Back
It will return to the service menu and the instrument initialize automatically.
figure 8-3
8.11 Special
The instrument enters special maintenance window. User can adjust the running control data.
This function should be performed in guidance of the engineers of authorized by our company
- 50 -
Vec 3 Diff Operation Manual Edition: A/0
Chapter 9 Maintenance
As other precision instrument, only careful daily service and periodic maintenance can the instrument
have a good working status, and can we get the reliable measurement results and have few malfunctions.
This chapter introduces some preventive methods for service and maintenance. If you want to know
more relative information, please contact the customer service department of our company.
According to the requirements for maintenance of the instrument during the using procedures, we
divide the preventive service and maintenance into the following types: daily, weekly, monthly, yearly and
maintenance according to actual need. Warning:
It is important for the hospital or organization that employs this instrument to carry out a reasonable
maintenance schedule. Neglect of this may result in machine breakdown.
/
Maintenance
“button. b) Operate as screen shows. c) Rotate the bottle covers of rest reagents and store them as the
reagent operation introduction. User should perform efficient action to prevent the material from
deteriorating, mis-eating and misusing. d) Plug the stopples which were pulled out at the first installation
into the corresponding tube connector. e) Pull out diluent connection tube ,, lyse connection tube and
waste connection tube and clean them with distilled water. Dry them in shady place, then packing them
into plastics. f) Pull out the power wire, packing it in plastics after cleaning with neutral detergent. g) Put
the instrument and parts packed in plastics into packing cases.
- 52 -
Vec 3 Diff Operation Manual Edition: A/0
Chapter 10 Troubleshooting
This chapter contains information that is helpful in identifying and resolving instrument problems that
may occur in the operation of the analyzer. If the problems can not be corrected with the aid of this chapter,
the user should contact the Customer Service Department of our company.
Chapter 11 Help
The instrument provides on-line help during operation. It is helpful for user to solve common problems
and indicate the next operation. Besides, the instrument provides help for user to read and search the
information in window. In the blood cell analyzer window, Select "Menu/Help" in the menu, the help
windows will pop up. As shown in figure 11-1.
figure 11-1 This window shows in chapter mode. User can search
help information conveniently. Click the scroll bar up and down, read the information front and back.
Function of the menu and shortcut keys: Content: Select the required contents.
- 55 -
Vec 3 Diff Operation Manual Edition: A/0
Appendix 1 Symbol
- 56 -
Biohazard
High Temperature
Caution
Usb Interface
Vec 3 Diff Operation Manual Edition: A/0
Appendix 2 Specification
2.1 Reagent
Diluent, Lyse,
(±0.3×109)/L ±5%
RBC
0~0.99×1012/L 1.0~9.99×1012/L
(±0.05×1012)/L ±5%
HGB
0~99g/L 100~300g/L
±2.0 g/L ±2%
PLT
0~99×109/L 100~999×109/L
(±8.0×109)/L ±10%
2- 5 Linear Range
2.9 Input/Output
Touch Pen Keyboard(Option) Mouse(Option) Printer(Option) R232 Port VGA Port Internet Port Power
supply:
Voltage Frequency Main Unit 100~240Vac 50/60 Hz Work environment
Temperature: 18°C~35°C Humidity: 10%RH ~90%RH Atmospheric Pressure: 70.0kPa ~106.0kPa
Storage environment Temperature: -10°C~40°C Humidity: 10%RH ~90%RH Atmospheric Pressure:
50.0kPa ~106.0kPa This instrument should be placed at a good ventilation area. Dimensions and Weight
Dimensions: 410mm×435mm×472mm Weight: 25kg Contraindication : None
- 59 -