Hot & Cold Chamber: Construction
Hot & Cold Chamber: Construction
Hot & Cold Chamber: Construction
HOT & COLD CHAMBER AN ISO 9001 : 2015 CERTIFIED CO. & CE MARK PRODUCT
Construction :
The chamber is triple wall at back & bottom inner both the
walls are stainless steel 304 quality duly polished with
mirror finish buffing & outer wall is Stainless Steel 304
quality mat finish. This Construction avoids rusting
problems caused due to condensation at -700C Cold
Chamber at the bottom & hot chamber on the top.
Heating :
Air Heaters are provided for heating above ambient temperatures.
De-Humidification :
De-humidification coils are installed below the cooling coils
for lower humidity.
Refrigeration system :
Cooling :
By hermetically sealed Danfoss / Embruco / Bitzer
Make Compressors, for double stage compressor
cooling system & humidity control.
Cooling coil:
Copper cooling coil is sandwiched between the
“PUFF” insulation & the Outer side of Inner working
Chamber to transfer the cooling by conduction
process for ULTRA Low temperature range. A
Application : separate finned cooling coil is fitted behind the baffle
wall below the blower fan for humidity control system.
These equipments are used for climatic & durability tests of electrical
Condensing Unit :
& electronic components, corrosion test on mechanical assemblies, Heavy duty air cooled condensing unit with
materials for simulated tropical & extreme tropical conditions. condensing fan will be provided at bottom side of the
chamber. The complete assembly is fitted on base in
such a way that any component can be replaced very
easily.
Sensors :
2 Nos PT-100 calibrated sensors for temp.-200C
to +200C is provided inside the chamber. All sensors
calibrated with traceable to ERTL / NABL accredited
Laboratories.
HOT & COLD CHAMBER
Controls : Air Circulation :
Touch screen controller, 4.0” panel with 800*480 resolution, user- The chamber is provided with best air circulation to give
friendly, stable & easy use. Programs : Edit up to 100 PTNs (1000 better temperature uniformity this is achieved by using one set of
steps per pattern), each PTN can be executed 999 times, could motor & blower in baffle wall compartment to push treated air in
store logs at an interval from 1 to 60 seconds for 600 days (24 hr working chamber. this system gives better than 1deg c.
operation), logs downloaded by USB port. PLC & SSR control way, temperature uniformity in working chamber, evaporator, heaters &
stable and easy use. PLC automatic calculation. The brightness steam injection system is from back side, behind baffle wall, the
of the LCD panel can adjustable. Programs auto- save function if air circulation blower is also fitted in this compartment only. This
power cut off suddenly. Password set function. Have HOLD & ADV arrangement helps to mix air pass well mixed air inside the
buttons in programs. Micro computer multi function controller, working chamber. Due to above system the temperature in
main components data automatically output including compressors, complete chamber almost remains equal.
solenoids, heating units, etc.
.
resume automatically. Operational Manual for Controller
IQ,PQ,OQ protocols certificate.
.. ELCB, MCB for Compressor, Boiler heater, Air heater & Mains.
TECHNICAL SPECIFICATIONS
Model : HCCH-02 Internal Dimension (W X D X H)(cms) Capacity
HCCH-A 45 x 45 x 50 100
HCCH-B 60 x 60 x 55 200
HCCH-C 55 x 55 x 100 300
HCCH-D 60 x 60 x 110 400
HCCH-E 60 x 60 x 140 500
HCCH-F 70 x 60 x 145 600
HCCH-G 70 x 70 x 145 700
HCCH-H 75 x 75 x 145 800
HCCH-I 80 x 75 x 150 900
HCCH-I 80 x 80 x 155 1000