Sensor CO2 para El 20% KCD - HP - en - 110602 PDF
Sensor CO2 para El 20% KCD - HP - en - 110602 PDF
Sensor CO2 para El 20% KCD - HP - en - 110602 PDF
Easy application to
Test facilit ies
Cell incubators
CO2 Chambers
Environment controlling system
Environment monitoring system
Ordering Codes
KCD-HP 100x (1% probe) Measurement Sensing Method Dual Wavelength NDIR
1 1000ppm Measurement range options 1%, 10%, 20%
2 2000ppm Accuracy* ±(3%F.S+2%Reading)
• Response time (τ63) < 65 sec
• Measurement time interval 0.75 sec
•
•
A 1% Warm up time < 2 min
Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 1
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
■ Connector input & output signal
Pin No. Name Notes Wire color
1 G+ 24 V DC (+) System Power Red
2 G0 24 V DC (-) Ground Black
3 OUT1 Output 1 (+) 0~10V Yellow
White or Violet
4 OUT2 Output 2 (+) 4~20mA
or Orange
5 RS485A RS485 terminal A Green
6 RS485B RS485 terminal B Blue
1) You should insulate unused output signal line of the cable. It may be
damaged by short.
2. Communication Mode
ASYNC (UART : Universal Asynchronous Receiver Transmitter)
1 Data Frame
Start Stop
Bit D0 D1 D2 D3 D4 D5 D6 D7 Bit
BAUD RATE 38,400bps
Data Bits 8 bit
Parity Bit none
Stop Bit 1 bit
4. Communication protocol
1) Commands for requesting product information and setting status from MASTER to
SLAVE
Command Description
10(0x0A) Transfer measured CO2 value
58(0x3A) Change CO2 module ID (default: 31)
59(0x3B) Change signal output mode of CO2 module(default :current)
Change communication Baud Rate of CO2 module
60(0x3C)
(default 38,400bps)
Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 2
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
2) Commands for data transfer from MASTER to SLAVE
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 System ID
Data block
3 byte 1 6 Block length
length
4 COMMAND char 1 0x0A : data transfer command
5 CheckSum byte 1 0 ~ 255 Sum no.1 to no.4
6 ETX byte 1 0x04 Block End
4) Commands for output type setting changing output status from MASTER to SLAVE
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
Data block
3 byte 1 8 Block length
length
0x3B : output type change
4 COMMAND char 1
command
5
※ Data byte 2 0x0001 : setting to current output
0: 0~10V
1: 4~20mA
6 CheckSum byte 1 0 ~ 255 Sum no.1 ~ no.5
7 ETX byte 1 0x04 Block End
※ Communication data are composed in low byte(0x01)->high byte(0x00) order.
(Little-endian)
Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 3
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
6) Commands for data transfer from SLAVE to MASTER
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
3 Data block length byte 1 26 Block length
System
4 COMMAND char 1 Master transfer command
command
5 System code byte 1 variable
6 Product name string 10 _ “KCD-HP”
7
※ CO2 2 _ Not using this data, it transfer
Measured ‘0x0000’
8
※ Temperature 2 _
value When return value is 0xD5
0x02 ,
Measured
9
※ Humidity 2 _
the formula is
Unsigned (0x02*0x100)+0xD5
value =0x02D5(16)=725(10)
integer
Ex) 1%module→725ppm
10
※ Measured VOC
2 _
10%module→7,250ppm(0.72
5%)
value
20%module→72,500ppm
(7.25%)
■ Dimensions
1% : 161mm
Φ43
Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 4
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
Warranty and Instructions
■ Warranty
This product passes our strict quality control and Korea Digital will repair or
replace this item without charge within 1 year after sales except for damage or
break by customer’s mistakes.
■ Instructions
1. Caution: shock and moisture
1) The characters of NDIR optical system may be changed by impacts.
Never drop this sensor and give it heavy impacts.
2) Don’t use it where water drops and condensation can occur, too
2. Keep operating conditions written above. If you do not, it may break down
or have large errors.
3. Don’t use a sensor without a case to block dust and other pollutants in
case of using for a long time.
4. You should insulate unused output signal line of the cable. It may be
damaged by short.