Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Sensor CO2 para El 20% KCD - HP - en - 110602 PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

KCD-HP

CO2 Sensor Probe (1%, 10%, 20%)


for OEM Applications
Our CO2 gas sensors get a small deviat ion
unlike NDIR Single type.
So they keep long term stability. KCD-HP100x
Excellent stability and accuracy
- through test ing and calibrat ion with
sophist icated process and techniques

Easy application to
 Test facilit ies
 Cell incubators
 CO2 Chambers
 Environment controlling system
 Environment monitoring system

NDIR type uses opt ical property to


measuring CO2 gas. We make up for a
controller not to be affected by a shock and
KCD-HP200x, 300X
a wave(vibration).
But please consult with our engineers, if you
use it under harsh environments (like [Figures]
construction sites).  Top : 1% Sensor probe
 Bottom left : 10% / 20% Sensor probe
 Bottom right : Holding bracket(Optional)

Ordering Codes
KCD-HP 100x (1% probe)  Measurement Sensing Method Dual Wavelength NDIR
1 1000ppm Measurement range options 1%, 10%, 20%
2 2000ppm Accuracy* ±(3%F.S+2%Reading)
• Response time (τ63) < 65 sec
• Measurement time interval 0.75 sec


A 1% Warm up time < 2 min

KCD-HP 200x (10% probe)


 General
Storage temperature -40~70 ℃
Weight < 250g
2 2%
5 5%  Operating Temperature 5 ~45 ℃
7 7% Conditions Humidity 0~95%RH (Non-condensing)
A 10%  Electrical Power supply (rectifiered) 16~28VDC
Power consumption 70mA average
KCD-HP 300x (20% probe)
 Outputs Analog Outputs 0~5VDC, 0 ~10VDC, 4~20mA
2 12%
Communication Output RS485
5 15%
A 20%  Dimensions 1% Probe Φ43 x 161mm
10% /20% Probe Φ43x 91mm
Cable 1.0m , 6 wires

℃ , intermediate value of detected ranges.


* Under conditions of calibration facilities of production factory, @25
 The recommended calibration interval is one year.
 Specifications and images may change without prior notice.

Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 1
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
■ Connector input & output signal
Pin No. Name Notes Wire color
1 G+ 24 V DC (+) System Power Red
2 G0 24 V DC (-) Ground Black
3 OUT1 Output 1 (+) 0~10V Yellow
White or Violet
4 OUT2 Output 2 (+) 4~20mA
or Orange
5 RS485A RS485 terminal A Green
6 RS485B RS485 terminal B Blue
1) You should insulate unused output signal line of the cable. It may be
damaged by short.

2) Output mode can be set as RS485 communication.


When you change output way (current or voltage), you should first select
‘Current’ or ‘Voltage’ with communication.
Then you can get output on corresponding Pin.
Refer to ‘Setting communication’
(0x0000 : current output 0x0100 : voltage output)

■ RS485 Communication protocol


1. Communication Connector
Pin No. Name Notes
5 RS485 A RS-485 connection terminal A
6 RS485 B RS-485 connection terminal B
(2) WAFER : No.5, 6

2. Communication Mode
ASYNC (UART : Universal Asynchronous Receiver Transmitter)

3. Communication data type

1 Data Frame

Start Stop
Bit D0 D1 D2 D3 D4 D5 D6 D7 Bit
BAUD RATE 38,400bps
Data Bits 8 bit
Parity Bit none
Stop Bit 1 bit

4. Communication protocol
1) Commands for requesting product information and setting status from MASTER to
SLAVE
Command Description
10(0x0A) Transfer measured CO2 value
58(0x3A) Change CO2 module ID (default: 31)
59(0x3B) Change signal output mode of CO2 module(default :current)
Change communication Baud Rate of CO2 module
60(0x3C)
(default 38,400bps)

Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 2
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
2) Commands for data transfer from MASTER to SLAVE
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 System ID
Data block
3 byte 1 6 Block length
length
4 COMMAND char 1 0x0A : data transfer command
5 CheckSum byte 1 0 ~ 255 Sum no.1 to no.4
6 ETX byte 1 0x04 Block End

3) Commands for setting or changing ID status from MASTER to SLAVE


No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
Data block
3 byte 1 8 Block length
length
0x3A : system setting changing
4 COMMAND char 1
command
5
※ Data byte 2 0x0001 : setting ID to ‘no.1’ Range : 1~31
6 CheckSum byte 1 0 ~ 255 Sum no.1 to no.5
7 ETX byte 1 0x04 Block End
※ Communication data are composed in low byte(0x01)->high byte(0x00) order.
(Little-endian)

4) Commands for output type setting changing output status from MASTER to SLAVE
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
Data block
3 byte 1 8 Block length
length
0x3B : output type change
4 COMMAND char 1
command

5
※ Data byte 2 0x0001 : setting to current output
0: 0~10V
1: 4~20mA
6 CheckSum byte 1 0 ~ 255 Sum no.1 ~ no.5
7 ETX byte 1 0x04 Block End
※ Communication data are composed in low byte(0x01)->high byte(0x00) order.
(Little-endian)

5) Commands for setting or changing transmission speed from MASTER to SLAVE


No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
Data block
3 byte 1 8 Block length
length
0x3C : command for changing
4 COMMAND char 1
transmission speed
5
※ Data byte 2 0x2580 : set as 9,600bps
6 CheckSum byte 1 0 ~ 255 Sum no.1~no.5
7 ETX byte 1 0x04 Block End
※ Communication data are composed in low byte(0x01)->high byte(0x00) order.
(Little-endian)

Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 3
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
6) Commands for data transfer from SLAVE to MASTER
No Data Types Bytes Ranges Notes
1 STX byte 1 0x03 Block Start
2 Slave ID byte 1 1~31 system ID
3 Data block length byte 1 26 Block length
System
4 COMMAND char 1 Master transfer command
command
5 System code byte 1 variable
6 Product name string 10 _ “KCD-HP”
7
※ CO2 2 _ Not using this data, it transfer
Measured ‘0x0000’
8
※ Temperature 2 _
value When return value is 0xD5
0x02 ,
Measured
9
※ Humidity 2 _
the formula is
Unsigned (0x02*0x100)+0xD5
value =0x02D5(16)=725(10)
integer
Ex) 1%module→725ppm

10
※ Measured VOC
2 _
10%module→7,250ppm(0.72
5%)
value
20%module→72,500ppm
(7.25%)

11 Firmware version byte 1 variable


Sum no.1 to no.11
Ex) Sum value :
12 CheckSum byte 1 0 ~ 255
0x01FF → 0xFF
0xFFFF → 0xFF
13 ETX byte 1 0x04 Block End
※ Communication data are composed in low byte(0x01)->high byte(0x00) order.
(Little-endian)

■ Dimensions

1% : 161mm

Φ43

10% /20%: 91mm

Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 4
KCD-HP
CO2 Sensor Probe (1%, 10%, 20%)
for OEM Applications
Warranty and Instructions
■ Warranty
This product passes our strict quality control and Korea Digital will repair or
replace this item without charge within 1 year after sales except for damage or
break by customer’s mistakes.

■ Instructions
1. Caution: shock and moisture
1) The characters of NDIR optical system may be changed by impacts.
Never drop this sensor and give it heavy impacts.
2) Don’t use it where water drops and condensation can occur, too

2. Keep operating conditions written above. If you do not, it may break down
or have large errors.

3. Don’t use a sensor without a case to block dust and other pollutants in
case of using for a long time.

4. You should insulate unused output signal line of the cable. It may be
damaged by short.

5. If output terminal is damaged by short, repair isn’t free of charge whether


within one-year warranty period.

20100901

Korea Digital Co., Ltd #804 Ace twin tower 2, Guro 3-dong, Guro-gu, Seoul, South Korea
Tel: 02-2109-8883 Fax: 02-2109-8878
ideabox@empal.com www.sensecube.com 5