Guide for replacing the cover of GC85A and GC70 table cover to safety cover.
Model SW Version
Parts/Files Required
MI81-08381A / A/S-UPGRADE KIT : GC85, GC70 IR TABLE
1
Procedure
Table 1. GC85A/GC70 parts
IR Sensor ASSY
MI97-03894A
ASSY-TABLE-IR-SENSOR-RX 1EA
2
Procedure
1. Disassemble the current cover
1-1 Move the table top frame to the end towards the transversal.
1-2 Unscrew three bolts(Figure 1,TH,+,M4,L10) and remove the cover.
1-3 Repeat 1-1 and 1-2 on 4 areas.
1 2
Figure 4. Replace Cable Veyor Bracket to Prevent Interference
3
Procedure
2-1 Turn the system off.
2-2 Unscrew 4 bolts of the bracket that attached on main body then remove foot
switch connector in the bracket.
2-3 Unscrew 5 bolts that fix the bracket to the cable veyor. (Figure 6)
(2 bolts from the inside, 2 bolts from the outside, 1 bolt from the side)
4
Procedure
NOTE: When assembling the new bracket, if you cannot put the foot switch cable
into the jack because the cable is too short, change the foot switch cable wiring
after opening the Main-board.
3-2 After removing the bolts (FH,+,M4,L6,1EA) marked in figure 10A, assemble the
bracket (MI61-04724A) with new bolts (FH,+,M4,L8,1EA) as in figure 10B.
M/B
4
Procedure
3-3 Wiring Operation
1) Put TRANS L(SERVICE) Cable (MI39-02236A) into MD2D board.
DIO1
ELU
2) Put the black line into the A6 pin of DIO1 connector, and A7 pin into the white line.
6
Procedure
3) Put the TRANS L(SERVICE) CABLE into the Bracket. Put the connecter into the
bracket with the 1 label on the left side as shown in Figure 13.
4)Wire the TRANS L(SERVICE) CABLE as shown in Figure 14 and fix it with cable ties.
7
Procedure
6) Wire IR-1, 2, 3, 4 CABLEs as in Figure 16, and fix them with CABLE TIEs.
IR-2
IR-1
IR-4
IR-3
8
Procedure
4. Sensor assembly
4-1 Assemble a sensor and two bosses on top cover. (SEMS,M4,L10,2EA)
9
Procedure
5. Assemble the new safety cover.
5-1 Assemble Cover Table Top. (MI63-02709A, MI63-02710A)
NOTE: When assembling the top cover, make sure that the connected harness does
not get stuck or come out.
5-2 Put the bolt-cap sheet(MI63-02384A) on top of the tighten bolts. (Figure 22)
10
Procedure
5-3 Attach the Safety Labels on each Cover Table Top as shown in figure 23.
11
Procedure
6. Software upgrade
- Press the Upgrade > button to upgrade the package.
12
Procedure
7. Setting
7-1 Set Up
- Set the values of the device option and the FT sensor menu.
13
Procedure
7-2 Set up > Device options > Ceiling Movement Method setting
1) Direct mode : when changing the Position, move to step 1.
2) Semi – Indirect mode (New feature) : when changing the Position, if the moving
stroke is short or if you can only change the SID, move to step 1. If not, move to step 3.
3) Indirect mode (Default) : when changing the Position, move to step 3.
14
Procedure
7-3 Set up > Device options > Operation Method for Auto Movement Key setting.
1) Single Press : move with a single press of the Auto position key.
2) Dead-Man (Default) : it only moves while pressing the Auto position key.
15
Procedure
7-4 Set up > Device options > Speed of Stand Fast Key setting
1) Fast (Default) : the key of the Stand panel moves fast.
2) Mid : the Fast key of Stand panel moves in the medium speed.
3) Slow : the Fast key of Stand panel moves Slow.
4) Off : the Fast key of Stand panel does not move.
16
Procedure
7-5 Set up > FT Sensor > FT Threshold (0.1N) setting
1) Set the threshold of power on the FT sensor to produce an Error.
2) Default setting is 200 and it can be adjusted within the scope from 100 to 300. As it
gets closer to 100 the sensitivity increases, generating errors, and as it gets closer to
300 it gets duller.
17
Procedure
7-6 Mechanical Specification > Table IR Sensor Installed setting
1) Not Installed (Default) : IR sensor is not installed.
2) Installed : IR sensor is installed.
18
Procedure
8. Operation check
8-1 IR sensor operation check
- Go to Set up > Diagnosis to check IR sensor operation condition.
19
Procedure
8-2 FT sensor threshold operation check
- Check if it generates an error when applying force surpassing the threshold value on
the THU.
19
Action Notification Report
TITLE : GC85A, GC70 Table Safety Application Guide (IR Sensor) with Safety Patch
NO : DGRMULTI-B18002M
Contact Address
Site Information
Country / Hospital
Hospital Address
System model
□□□□□□□□□□□□□□□
System SN
Action
Confirm
------------------------------------- -------------------------------------
Service Engineer Name Service Engineer Signature
21