Itemp Hart DIN Rail TMT112: Technical Information
Itemp Hart DIN Rail TMT112: Technical Information
Itemp Hart DIN Rail TMT112: Technical Information
TI114R/09/en
51009802
TMT112
Measuring system The iTEMP® HART® DIN rail TMT112 temperature transmitter is a 2-wire transmitter with an analog out-
put. It has measurement input for resistance thermometers (RTD) in 2-, 3- or 4-wire connection,
thermocouples and voltage transmitters. Setting up of the TMT112 is done using the HART® protocol with
hand operating module (DXR275, DXR375) or PC (e.g. configuration software ReadWin® 2000 or FieldCare).
Input
Measured variable Temperature (temperature linear), resistance and voltage.
Measuring range Depending upon the sensor connection and input signal. The transmitter evaluates a number of different
measurement ranges.
Type of input
Minimum
Type Measurement ranges
measurement range
10 to 400 10
Resistance transmitter Resistance
10 to 2000 100
2 Endress+Hauser
TMT112
Output
Output signal Analog 4 to 20 mA, 20 to 4 mA
Current limit 23 mA
Power supply
Electrical connection
T09-TMT112-04-10-xx-ae-000
Endress+Hauser 3
TMT112
Performance characteristics
Response time 1s
! Note!
The accuracy data are typical values and correspond to a standard deviation of ± 3 (normal distribution), i.e.
99.8% of all the measured values achieve the given values or better values.
1. % is related to the adjusted measurement range. The value to be applied is the greater.
Influence of ambient Total temperature drift = input temperature drift + output temperature drift
temperature (temperature
drift) Effect on the accuracy when ambient temperature changes by 1 K (1.8 °F):
4 Endress+Hauser
TMT112
Influence of cold junction Pt100 IEC 60751 Cl. B (internal reference junction for thermocouples TC)
Installation conditions
Installation instructions Orientation
No limit
Environment conditions
Ambient temperature limits 40 to +85 °C (-40 to 185 °F), for Ex-areas see Ex-certification
Condensation Permitted
Electromagnetic Interference immunity and interference emission according to IEC 61326 and NAMUR NE 21
compatibility (EMC)
Mechanical construction
Design, dimensions
T09-TMT112-06-10-xx-en-000
Housing for DIN rail mounting according to IEC 60715; Dimensions in mm (in)
Endress+Hauser 5
TMT112
Terminals Keyed plug-in screw terminals, core size max. 2.5 mm² (16 AWG) solid, or strands with ferrules
Human interface
Display elements A yellow illuminated LED signalizes: Device is operational.
With the PC software ReadWin® 2000 or FieldCare the current measured value can be displayed.
Operating elements At the temperature transmitter no operating elements are available directly. The temperature transmitter will
be configured by remote operation with the PC software ReadWin® 2000 or FieldCare.
Interface
PC interface Commubox FXA191 (RS232) or FXA195 (USB).
Configurable parameters
Sensor type and connection type, engineering units (°C/°F), measurement range, internal/external cold junc-
tion compensation, cable resistance compensation on 2-wire connection, fault conditioning, output signal
(4 to 20/20 to 4 mA), digital filter (damping), offset, measurement point identification + descriptor
(8 + 16 characters), output simulation, customer specific linearisation, min./max. process value indicator func-
tion.
Hazardous area approvals For further details on the available Ex versions (ATEX, CSA, FM, etc.), please contact your nearest End-
ress+Hauser sales organisation. All relevant data for hazardous areas can be found in separate Ex documenta-
tion. If required, please request copies from us or your Endress+Hauser sales organisation.
6 Endress+Hauser
TMT112
Ordering information
Questionnaire
Endress+Hauser 7
TMT112
Product structure This information provides an overview of the order options available. The information is not exhaustive, how-
ever, and may not be fully up to date. More detailed information is available from your local Endress+Hauser
representative.
Approval
A Non-hazardous area
B ATEX II 2(1) G EEx ia IIC T4/T5/T6
C FM IS, Class I, Div. 1+2, Group A, B, C, D
D CSA IS, Class I, Div. 1+2, Group A, B, C, D
E ATEX II3G Ex nA IIC T4/T5/T6
J CSA General Purpose
Configuration connection
A Factory setup Pt100 3-wire 0...100 °C
1 Thermocouple TC
2 RTD, 2-wire
3 RTD, 3-wire
4 RTD, 4-wire
Configuration
A Factory setup Pt100 3-wire 0 to 100 °C
B Measuring range, see additional specification
C TC configuration range, see questionnaire
D RTD configuration range, see questionnaire
Additional option
A Basic version
B Works calibration certificate 6-point
TMT112- Order code
8 Endress+Hauser
TMT112
Accessories
• Commubox FXA191 (RS232) or FXA195 (USB)
Order code: FXA191-... or FXA195-...
• PC-operating software: ReadWin® 2000 or FieldCare
ReadWin® 2000 can be downloaded free of charge from the internet from the following address:
www.endress.com/readwin
• Hand operating module ’HART® Communicator DXR375’, Order code: DXR375-...
Documentation
• Brief operating instructions ’iTEMP® HART® DIN rail TMT112’ (KA193R/09/a3)
• Functional safety manual TMT112 (SD010R/09/en)
• Additional documentation for use in explosion-hazardous areas:
ATEX II 2(1) G Ex ia IIC (XA022R/09/a3)
ATEX II3G Ex nA II (XA055R/09/a3)
• Functional safety manual TMT112 (SD010R/09/en)
Endress+Hauser 9
TMT112
10 Endress+Hauser
TMT112
Endress+Hauser 11
TMT112
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI114R/09/en/13.10
51009802