9907X0043 - 3 Operating Instructions
9907X0043 - 3 Operating Instructions
9907X0043 - 3 Operating Instructions
77-0115
OPERATING INSTRUCTIONS
Digital Point Load Test Apparatus
77-0115
In the interests of improving and updating its equipment, ELE reserves the right to alter specifications to equipment at any time.
ELE International 2015
1 9907X0043 Issue 3
Contents
Section Page
1 Introduction 3
2 Specification 4
4 Operation 5
5 Calculations 7
7 Maintenance 12
8 References 13
Appendix 1 : Bibliography 14
A. Operating Manual 16
B. Installation Manual 30
WEEE Directive
Prologue
The index s is determined from the formula P/D² where D is the distance between the point
load platens at the moment of failure and P the force required causing the failure.
Considerable work has been conducted on this method of strength determination but only
recently is this being related to the field study of specimens. The following is an excerpt from
an article ‘Logging the Mechanical Character of Rock’ by Dr J A Franklin, Mr E Broch and Mr
G Walton in Transactions of the Institute of Mining and Metallurgy volume 80, which
adequately introduces the technique.
‘The Strength Log –
Traditionally, rock strength is either estimated in the field by use of a hammer or in the
laboratory with sophisticated apparatus. The former method is at best qualitative, whereas
the latter requires careful specimen preparation. There is an obvious need for a device to
test quickly and reliably, unprepared core in the field.
A portable point load tester has been developed at Imperial College, London, as an aid to
core logging. The apparatus consists of a small hydraulic pump and ram, with a loading
frame of maximum rigidity easily adjustable to test core of different sizes. Core is loaded
between pointed platens of standard dimensions. The two quantities measured in the test are
the distance, D, between platen contact points, which is read from a graduated scale
incorporated in the load frame, and the force, P, required to break the specimen, which is
read from a calibrated readout unit in the hydraulic circuit. The point load strength index, s, is
the ratio P/D².
This type of test is not new, having been used extensively in the USA, Russia and several
European countries, but mainly as a research laboratory tool rather than as a convenient
technique for field index testing. Theoretical consideration of the point load test showed that
it gives a measure of tensile strength, as indeed does the geological hammer. The results
are, however, sufficiently related to other measures, such as unconfined compressive
strength, to allow this test to give an index of strength in a general sense. The point loading
gives several important advantages:
A the specimen fails at much lower loads than in compression, needing a machine
load capacity less than one-tenth of that usually required for compression;
B core can be tested direct from the core box without previous machining – even weak or
broken rock can thus be tested;
C as fracture initiates in the specimen interior, platen contact conditions are of little
importance.
1 Introduction
The ELE Point Load Apparatus comprises a two-column fixed crosshead frame
mounted directly onto a hand-operated hydraulic jack.
Pressure applied by the handpump in the base of the jack extends the piston carrying
the lower of the two conical points. The upper point is fixed to the crosshead and the
frame can accommodate sample sizes up to 101.6 mm (4 in.) in diameter. A scale is
fitted to indicate the distance between contact points (D).
The applied load is indicated directly in kN on the digital readout display unit. The
maximum load achieved during the test is then held and displayed until the readout is
zeroed ready for the next test.
The apparatus is supplied with a heavy-duty face mask for operator protection.
2 Specification
Case dimensions 320 x 715 x 370 mm (width x height x depth)
Overall weight 25 kg
Load capacity 55 kN
Specimen
max. diameter 101.6 mm
Oil Shell Tellus 37. Note: If T 37 is unavailable
then either T 32 or T 46 may be used but do
not mix oils of different specifications.
Working temperature range +5 Deg C to +40 Deg C
Recommended humidity range 30% to 95%
Batteries 4 X AA
Fig. 1
Note the calibration section within the readout is guarded by a password. If for some
reason you need to re-calibrate the unit contact the ELE service department.
8 References
8.1 ISRM Commission on Testing Methods Working Group on Revision of the Point Load
Test Method.
8.2 The Point Load Strength Test for Concrete Cores, P J Robins, Magazine of Concrete
Research, Vol. 32, No. 111, June 1980
Appendix 1
Bibliography
Operating Manual
Contents Page
OPERATING FEATURES :
A. Display Views 18
B. Keyboard Functions 19
C. Operating Functions 20
SET UP :
A. Display
Views Operation
On the 5-digit numerical display the current weight is usually displayed. When the “PEAK”
function is activated, the maximum weight value is displayed. During setup steps, the
displayed messages and values shown depend on the programming function in execution.
Power On
Display test is executed at instrument power on. An ID code is then
displayed representing software and version number. Please remember
to communicate this ID code in case of a service request.
OVERLOAD CONDITION
When the gross weight exceeds the maximum capacity for more
than 9 divisions, this alert appears on the display.
UNDERLOAD CONDITION
When the gross weight is less than -9999, this alert appears on the
display.
Functions Menu
Navigation
Select Value
C. Operating
Functions Autotare
The autotare command has no effect with an unstable weight (timeout 2 sec.). Parameter 0
BND represents the maximum weight value that can be zeroed (positive or negative),
compared to the zero value stored during calibration operations.
Peak Functions
The peak value is always stored, even when not displayed. It is stored both for positive and
negative weight values. The peak value is lost after power off. The peak value is stored at
weight acquisition frequency (see weight filter).
Set Up Menu
During the calibration procedure the message “CAL” blinks on the display.
During the linearization procedure the message “LIN n” blinks on the display: N is the
number of the linearization point you are programming (N range: from 1 to 5).
BAUD RATE
Select a value: 9600, 19200, 38400, 57600 e 115200 bit / sec.
COMMUNICATION PROTOCOL
DATA FORMAT
Select combination for parity / data bits number / stop bits number.
COMMUNICATION ADDRESS
Address used for master/slave protocols (from 0 to 99).
H. Password Access Menu
<csum> = string data control sum. It is calculated executing the exclusive OR (XOR) of all
characters from STX (or from <addr>) to ETX, excluding the latter; the XOR result is
resolved into 2 characters considering the 4 highest bit (first character) and the for lowest
bit (second character) separately; the 2 resulting characters are then coded in ASCII; (e.g.:
XOR = 5Dh; <csum> = "5Dh" i.e. 35h and 44h).
AVAILABLE COMMANDS:
Current net weight request.
Current gross weight request.
Current peak value request.
Autotare command.
Peak reset command.
The unit connected to the instrument (usually a personal computer) is the master-unit and is
the only unit that can start a communication procedure.
The communication procedure must always start with the transmission of a string by the
master-unit, followed by a response string from the interested slave-unit.
The quotes enclose the constant characters (attention to capital / lowercase characters); the
symbols < and > enclose variable numerical fields.
Current Net Weight Request
Autotare Command
The quotes enclose the constant characters (attention to uppercase / lowercase characters);
the symbols < and > enclose variable numerical fields.
STX (start of text) = 0x02h, ETX (end of text) = 0x03h, EOT (end of transmission) = 0x04h,
ACK (acknowledge) = 0x06h, NAK (No acknowledge) = 0x15h.
<Addr> = Serial communication address + 0x80h (e.g.: address 2: <Addr> = 0x82h (130 in
decimal format)).
<status> = 1 byte, see following table (bit = 1 if condition is TRUE).
<gross>, <net>, <peak> = 8 ASCII bytes, weight value is right hand justified (without non-
significant zeroes, with decimal point and negative sign).
If overload condition: “^^^^^^^^”.
If underload condition: “------------- ”.
If weight reading error “ O-L “.
<csum> = string data control sum. It is calculated executing the exclusive OR (XOR) of all
characters from STX (or from <addr>) to ETX, excluding the latter; the XOR result is
resolved into 2 characters considering the 4 highest bit (first character) and the for lowest
bit (second character) separately; the 2 resulting characters are then coded in ASCII;
(e.g.: XOR = 5Dh; <csum> = "5Dh" i.e. 35h and 44h).
Appendix 2 : B
Installation Manual
Contents Page
MAIN FEATURES :
A. Technical Features 32
B. Symbols 33
D. Warnings 33
CONNECTIONS :
E. Power Supply 34
I. Summary of Wiring 36
MAIN FEATURES :
A. Technical Features
Load cells: max 2 350 Ohm parallel load cells (or 4 700 Ohm load cells).
Load cells power supply 3.3Vdc
Linearity < 0.01% of full scale
Temperature drift < 0.001% of full scale / C°
A/D converter resolution 24 bit
Input signal range From -3.9 mV/V to +3.9 mV/V
Digital filter From 0.25 Hz to 3 Hz selectable
Weight decimal digits from 0 to 3 decimal digits
Zero and full scale calibration Executable via keyboard
Load cells cable fail check Always working
Further information.
D. Warnings
The power supply cable must be channelled separately from other power
supply cables with different voltages, from the load cells cables and from
the I/O cables.
Pin 1 on J5 + Alim.
Pin 2 on J5 GND
The load cell cable must not be channelled with other cables (i.e. Outputs
connected to remote switches or power supply cables), but must follows its
own route.
Any cable extension must be carefully shielded, the colour code must be
respected and a cable of the same type as the one provided from the
manufacturer must be used. The extensions on the cables must be
soldered, otherwise a supportive terminal block or a junction box must be
used.
The load cell cable shouldn’t have more conductors than those effectively
used (4 or 6). In case of a cable with more conductors, connect those
unused wires to the –power supply (terminal block 2).
A maximum of 4 x 350-ohm load cells can be connected to the instrument in parallel. The
load cell excitation voltage is 3.3 Vdc and is protected against a temporary short circuit .
The instrument’s input signal range requires the use of load cells with sensitivity from 1 mV/V to 3.9
mV/V. The load cell wires must be connected to terminals 1 ...4 of 4 pins connector (or soldered).
Connect the load cell’s cable shield to the load cell body, or to GND.
NUM. J6
1 + Excitation
2 - Excitation
3 + Signal
4 - Signal
G. RS485 Serial Wiring
A bipolar shielded cable must be used and the shield must be connected to
GND.
The cable must not be channelled with other cables (i.e. outputs connected
to remote switches or power supply cables), but must follow its own route.
NUM. J5
1 + ALIM. 6 - 13 Vdc
2 GND
3 CAN H
4 CAN L
5 RS485 +
6 RS485 -
A bipolar shielded cable must be used and the shield must be connected to
GND.
The cable must not be channelled with other cables (i.e. outputs connected
to remote switches or power supply cables), but must follow its own route.
NUM. J5
1 + ALIM. 6 - 13 Vdc
2 GND
3 CAN H
4 CAN L
5 RS485 +
6 RS485 -
I. Summary of Wiring
NUM. J6
1 + Excitation
2 - Excitation
3 + Signal
4 - Signal
NUM. J5
1 + ALIM. 6 - 13 Vdc
2 GND
3 CAN H
4 CAN L
5 RS485 +
6 RS485 -
DIRECTIVE ON WASTE ELECTRICAL & ELECTRONIC EQUIPMENT (WEEE)