Heat Value Gas Chromatograph Model: HGC303 User's Manual
Heat Value Gas Chromatograph Model: HGC303 User's Manual
Heat Value Gas Chromatograph Model: HGC303 User's Manual
User’s Manual
HGC000001000P
NOTICE
Safety symbols
Be sure to correctly operate the model HGC303 while strictly observing the safety
precautions provided in this manual-especially the Warnings and Cautions indicated
by the symbols as shown below.
The descriptions of the Warning and Caution signs used in this manual are as follows.
m WARNING
The Warning sign means that serious personal injury, even death, could result if the
instructions given are not strictly observed.
m CAUTION
The Caution sign means that light personal injury and/or equipment damage could
result if the instructions given are not correctly observed.
0344
II 2 GD
m WARNING
NEVER open the terminal box cover while the model HGC303 is energized in a
hazardous location.
m CAUTION
Use the model HGC303 only in an ambient temperature of -10 to 50°C (14 to 122°F)
m CAUTION
Take precautions to prevent corrosion, deformation or damage to the housing or
terminal box cover.
m CAUTION
See that all conduits are properly sealed. Otherwise, the model HGC303 cannot with-
stand the pressure that can result from explosion of an explosive gas inside the housing.
Also, the model HGC303 cannot prevent the explosion of any external explosive gas.
~Note The apparatus has been certified to comply with EN 61241-0, EN 61241-
1 (dust ignition protection).
1.2 The apparatus enclosure must be kept closed in the hazardous areas when the
apparatus is energized because the internal circuit of the apparatus is capable of
igniting the explosive atmosphere. (Never connect any hand-held communicator to
the apparatus terminals by opening the cover, except while no explosive
atmosphere is present.)
1.3 It is required to connect the external earthing terminal of the apparatus to the
equipotential bonding system which includes protective conductors, metal
conduits, metal cable sheaths, steel wire armouring and metallic parts of structures,
but does not include the neutral conductors of the power systems.
2. Cable systems
2.1 Thermoplastic sheathed cables, thermosetting sheathed cables, or elastomeric
sheathed cables can be selected for fixed wiring in the hazardous areas.
2.2 Flameproof cable entry devices (cable glands) certified to comply with EN 60079-
1 and appropriate to the type of cable employed, must be used for the connection of
cables to the apparatus.
3. Conduit systems
For conduit systems, relevant national standards or codes of practice are followed
prior to the following recommendations.
3.1 Screwed heavy gauge steel, solid drawn or seam welded conduit, or flexible
conduit for protection of cables in explosive atmospheres (see ISO 10807) can
be selected for fixed wiring in the hazardous areas.
3.2 Conduit must be threaded for connection to permit the full engagement of five
threads.
3.3 Either conduit entry devices or sealing devices such as stopping boxes are
provided at the wall of the apparatus enclosure to limit the pressure piling effect
and to prevent hot gases from entering the conduit system from the enclosure
containing a source of ignition. Each type of both the devices must be certified
to comply with EN 60079-1.
3.4 The stopping boxes, if used, are filled with a compound which does not shrink
or setting and is impervious to, and unaffected by, chemicals found in the
hazardous area. The depth of the compound in the stopping box is at least equal
to the internal diameter of the conduit, but in no case less than 10 mm.
3.5 When the conduit contains three or more non-seathed single or multi-core
cables, the total cross-sectional areas of cables, including insulation, are not more
than 40% of the cross-sectional area of the conduit.
m CAUTION
• Install the apparatus only in hazardous (classified) locations for which the
apparatus has been approved.
• Seal each conduit entering the apparatus enclosure within 18 in.(457 mm)
from the enclosure.
• Do not open the apparatus enclosure when an explosive atmosphere is
present.
EMC caution
1. Electromagnetic environment
The model HGC303 is intended to be used in an industrial electromagnetic
environment.
2. Electromagnetic immunity conditions
Specification: During test, SCV deviation is less than or equal to 1MJ/m3.
Chapter 1 : Introduction
1-1 : Definition of terms ........................................................................................ 1-1
1-2 : General ........................................................................................................ 1-3
1-3 : Model HGC303 measuring system .............................................................. 1-4
1-4 : Model No...................................................................................................... 1-5
1-5 : Model HGC303 Structure............................................................................. 1-6
1-6 : Fieldbus communication system.................................................................. 1-7
Chapter 2 : Installation
2-1 : Unpacking and storing ................................................................................. 2-1
2-2 : Installing the model HFA100........................................................................ 2-3
2-3 : HGM Installation .......................................................................................... 2-3
2-3-1 : Computer system requirements ..................................................... 2-3
2-3-2 : HGM software installation............................................................... 2-3
2-3-3 : .NET framework 4.0 installation...................................................... 2-6
2-3-4 : Microsoft Visual C++ runtime installation ....................................... 2-8
2-3-5 : Setting the folder access rights ...................................................... 2-11
2-4 : Fieldbus installation ..................................................................................... 2-15
2-4-1 : Fieldbus requirements .................................................................... 2-15
2-4-2 : Fieldbus wiring................................................................................ 2-18
2-5 : Model HGC303 installation .......................................................................... 2-19
2-5-1 : Installation site................................................................................ 2-19
2-5-2 : Model HGC303 dimensions............................................................ 2-20
2-5-3 : Model HGC303 installation example .............................................. 2-21
2-5-4 : Model HGC303 piping .................................................................... 2-22
2-5-5 : Model HGC303 wiring .................................................................... 2-24
Chapter 3 : Operation
3-1 : Starting up the model HGC303.................................................................... 3-1
3-1-1 : Secondary pressure and flow set ................................................... 3-1
3-1-2 : Piping leak check............................................................................ 3-1
3-1-3 : Power on ........................................................................................ 3-2
3-1-4 : Model HGC303 leak check............................................................. 3-4
3-2 : Stopping the model HGC303 ....................................................................... 3-5
3-3 : HGM operation............................................................................................. 3-6
3-3-1 : HGM connection with model HFA100 and HDM303 ...................... 3-7
3-3-2 : Starting up the HGM with model HFA100 ...................................... 3-8
3-3-3 : HGM Main menu ............................................................................ 3-13
3-3-4 : Set up HGM.................................................................................... 3-14
Table of Contents
Chapter 4 : Maintenance
4-1 : Checking and changing the carrier gas ....................................................... 4-1
4-2 : Checking and changing the filters in model HGC303 .................................. 4-1
4-3 : Periodical check........................................................................................... 4-1
Chapter 5 : Troubleshooting
5-1 : Connection with PC ..................................................................................... 5-1
5-2 : HGC status on HGM .................................................................................... 5-2
Appendix
SP (Set Point)
The set value of each variable.
PV (Process variable)
The present value of each variable.
SCV, GCV
Superior Calorific Value, Gross Calorific Value
These parameters are same value of different name.
ICV, NCV
Inferior Calorific Value, Net Calorific Value
These parameters are same value of different name.
TCD
Thermal Conductivity Detector
URV
Upper Range Value
LRV
Lower Range Value
Total (Raw)
Total of raw concentration
Component name
C6+: Hexane and heavier gas
C3H8: Propane
i-C4H10: i-Butane
n-C4H10: n-Butane
neo-C5H12: neo-Pentane
i-C5H12: i-Pentane
n-C5H12: n-Pentane
N2: Nitrogen
CH4: Methane
CO2: Carbon dioxide
C2H6: Ethane
1-2 : General
The model HGC303 is a gas chromatograph designed to analyze natural gas and is able
to transmit a process variable via a Fieldbus signal.
One can easily adjust configuration data and monitor values such as the heat value by
using the HGM.
The heat value monitoring system, which can be controlled from both the model
HGC303 and HGM, will substantially minimize time, cost and maintenance.
This chapter first describes the measuring system and structure of the model HGC303.
After that, the characteristics and the specifications of Fieldbus are described in detail.
First time users of the model HGC303 should read this chapter carefully and thor-
oughly.
~Note *: For satisfying TestSafe Flameproof Certification or IP65, six 1/4 NPT
male connectors are required.
m CAUTION
A block valve is a kind of air actuator valve. It is used mainly for the protection of
the TCD and columns.
It works as sample shut-off valve when the pressure of the carrier gas or air supply
is lower than approximately 294 kPa.
Azbil Corporation recommends that it should be installed.
The Heat Value Gas Chromatograph measures process gases (N2, CO2, C1~C6+) that
are mainly contained in natural gas, calculates heat value, density, Wobbe index and
compressibility factor, and converts them into a Fieldbus signal in the field and trans-
mits the signal to receivers.
Parameters can be remotely set, adjusted, and self-diagnosed with the HGM.
HGC303-I II
I Conduit entry 1/2 NPT female 1
Gas connection 1/4 NPT female Calculation method
II Explosion-protection ISSeP/ATEX flameproof E ISO
NEPSI flameproof N
TestSafe flameproof S
JIS flameproof * + J
Ordinary type + H
FM flameproof F GPA
CSA flameproof C
Terminal housing
Terminal cover
Manifold
Oven cover
HGC000004000D
This chapter guides you through the procedures for installing of your hardware and
software.
Package items
(1) Model HGC303
(2) Mounting bracket set
(3) Wrench for seal plug
(4) Tag number plate set
(5) CD-ROM including HGM software
and user's manual
(6) EC declaration of conformity and
safety instructions *
m CAUTION
Exposing the model HGC303 to the atmosphere might cause deterioration of the
column. Therefore, the model HGC303 has been packed and shipped in a protective
bag with a desiccant. Install and operate the model HGC303 immediately after
breaking the seal.
Inquires
If you have any questions regarding the specifications of your model HGC303, contact
one of the Azbil Corporation products service offices or contact your nearest Azbil
Corporation representative.
When making an inquiry, make sure to provide the model number and product number
of your model HGC303.
Model HGC303 that has been used should be stored by following procedures below.
Step Action
~Note This manual is for the use of HFA100 version 3.0 or later. For combina-
tions of HFA100 versions and corresponding PC software versions, see
“Software Compatibility” in the Appendix.
(10)Installation is complete once the message below appears on your screen. Click
[OK] button.
Check the license acceptance check box and click the [Install] button.
(8) Installation will start. It may take a few minutes until completion.
(9) Installation is complete once the message below appears on your screen. Click
[Finish] button.
Check the license acceptance check box and click the [Install] button.
(9) Installation is complete once the message below appears on your screen. Click
[Finish] button.
~Note Executing the program without this setting will cause a virtual folder prob-
lem caused by Windows User Account Control (UAC). Refer to trouble-
shooting “5-1: Connection with PC”.
(3) HGM Properties, select [Security] tab and click [Edit] button.
(5) Permissions for HGM, check Allow Full control in Permissions for Users, then
click [Apply] button and click [OK] button.
Cable
Various types cables are usable for fieldbus.
Type A is the preferred fieldbus cable.
Azbil Corporation recommends type A as the fieldbus cable to use.
The table below describes the type of cable and its maximum length, which is speci-
fied in the IEC 1158-2/ISA S50.02 Physical Layer Standard.
A Shielded, twisted pair #18 AWG (0.8 mm2) 1900m (6232 ft.)
Shield foil
Shield wire
Shield net
Jacket
Terminators
A terminator is an impedance matching module used near or at the end of a transmis-
sion line. Terminators prevent distortion and signal loss.
A terminator shall be located at both ends of a trunk cable, connected from one signal
conductor to the other.
A trunk is the longest cable path between any two devices on the fieldbus network.
ONLY TWO terminators are required per fieldbus segment.
No connection shall be made between the terminator and cable shield.
The terminator impedance value shall be 100 ohm +/- 2% over a frequency range of
7.8 kHz to 39 kHz.
The model HGC303 and model HDM303 have a terminator at the fieldbus connection
port therefore an additional terminator is not required.
Terminal blocks
The terminal blocks can be the same as those used for 4-20mA.
Connectors
D-sub 9P connector is as specified for standard fieldbus connectors in the IEC/ISA
Standard.
6 Data +
7 Data -
Signal wire
A Fieldbus signal is transmitted via 2-wire isolated signal lines.
Please keep in mind that the Fieldbus signal has polarity, positive (+) and negative (-).
All of the (+) terminals must be connected to each other and similarly, all of the (-) ter-
minals must be connected each other.
An important aspect of fieldbus is that neither of the signal wires are grounded.
Shielding
The preferred type of cable for fieldbus is a shielded cable.
Assemble a lugged shield wire connected to the metallized shield of each cable.
Connect all shield wires together to the terminal block.
In addition, connect the overall shield to the ground at one point in instruments room
to protect against field noise.
Do not ground the shield at multiple points.
Termination
A terminator shall be connected at both ends of the signal wire pair, at the field device
end and the host device end.
Connect the terminator between signal (+) and (-).
ONLY TWO terminators are needed per fieldbus segment.
Never connect a terminator between the signal (+ or -) and cable shield.
Heat shrink sleeve
Jacket
Overall shield
Shield wire
Wire terminal
- A location with minimal exposure to corrosive gases or dust and with good air cir-
culation.
97 (3.8)
244 (9.6)
82 (3.2)
100 (3.9)
Mounting bracket
HGC000008000D
HGC000009000P
~Note *: Remove the vent plug then connect the fitting and pipe when IP65 is
required or when HGC model No. is 'HGC303-1S'.
m WARNING
Purge the carrier gas line before performing any piping, and then verify that there is
no dust remaining in the piping.
Release the gas from the vent line to the air through the header.
There is a possibility that back-pressure from vent line has a lot of influence.
Prepare the carrier gas and valve operating gas as specified in the table below.
Secondary supply
Gas type Purity
pressure
400 ± 50 kPa
Carrier gas Helium 99.99% or higher
(58 ± 7 psi)
Either internal grounding (earthing) terminal (A) or external grounding (earthing) ter-
minal (B) can be used.
At least one grounding (earthing) terminal connection is recommended.
m WARNING
Only a 24V DC supply may be used to operate the model HGC303.
m CAUTION
Confirm that the supply voltage is within 24VDC+/-15% (20.4~27.4V) at the HGC
terminal.
m CAUTION
HGC requires the current of 4A minimum on 24VDC as the power supply.
m CAUTION
Use a power supply which has overcurrent protection capability for this product.
3-1-3 : Power on
Supply the power to operate the model HGC303 system according to the following
action.
Step Action
~Note After turning on the power, allow 2 hours for the device to warm up.
The carrier gas pressure SP and oven temperature SP have already been factory set in
the model HGC303, therefore, the user doesn't have to worry about setting this data.
Carrier gas pressure SP: less than 300 kPa (43.5 psi)
(SP differs with each model HGC303)
Oven temperature SP: 58°C (136.4°F)
Analyzing cycle: 300 sec.
~Note When the power is supplied to the model HGC303, a model HGC303
status error will appear on HGM monitoring system (oven temperature
error message etc.
The model HGC303 status will automatically return to normal once the
oven temperature reaches 58°C (136.4°F).
~Note *: Recommend supplying the standard gas if it is the first time set-up after
delivery or a long-period storage.
~Note +: If the output value from HGC seems strange after several cycles sup-
plying the process gas, try to do followings:
1. Check the process/standard gas supplies properly and the vent line is not
blocked. If there are problems, rectify them and check the output value
again.
2. Run the HGM program, and make it “on-line”, then start “User's mode”.
3. Check whether the peaks are small or normal, the peak shapes are strange
or not by chromatogram.
4. If the phenomena in section above are observed, stop the process gas and
quite the HGM program.
5. Connect blind plugs or shut the vent lines, then connect Helium gas cylin-
der at the 'INLET' port of HGC.
6. Charge Helium gas at 400kPa (58psi) to 'INLET' then leave it for about one
hour.
7. Return the connection normal and supply process/standard gas for checking
again.
(A)
(B)
HGC000011000D
Step Action
1 Check that the valve operating gas is being supplied.
2 Check the carrier gas has a secondary pressure (A) of 400 ± 50 kPa (58 ± 7
psi).
3 Verify that the carrier gas line valve off and observe the rate of fall in the
indicated primary pressure (B).
4 Leak evaluation procedure.
After introducing the carrier gas into the model HGC303, a normal condi-
tion is confirmed by a rate of fall of less than 1500 kPa (217 psi) per every 5
minutes.If more than 1500 kPa (217 psi) is observed, immediately contact
an Azbil Corporation products service office or the nearest distributor.
If the carrier gas is being used for valve operating gas at the same time, the
carrier gas consumption will be doubled. (less than 3000 kPa (435 psi) per 5
minutes)
m CAUTION
Verify that there is no leak from all connections.
m CAUTION
Do not leave the model HGC303 in the sampling system without plugs or seals at
the connections to vent.
Introduction
The functions of the HGM are described in this chapter.
The HGM is a calibration, configuration and maintenance tool for the model HGC303.
Analysis statuses, process variables and a chromatogram are displayed on its screen,
and information is stored in a database to facilitate routine management and tuning.
~Note 1 There is a possibility that this software will not function properly if
another application software is used at the same time.
Functions
(1) Monitoring heat value, chromatogram and carrier gas pressure / oven temperature
control
(2) Data save (load)
(3) User report
(4) Calibration
(5) Self-diagnostics
(6) Hold model HGC303 outputs to host control system
Refer to the model HDM303 user’s manual regarding the details of each part of the
model HDM303.
(1) Make sure that both the model HGC303 and the model HDM303 are running nor-
mally.
(2) Prepare a personal computer, which has the HGM installed.
(3) Verify that font size is [Small font] and the display resolution 1024 Å~ 768 pixels.
(4) Connect the model HFA100 along the FB line. (Refer to Figure 3-2.)
(5) Connect the USB cable to the USB port of your PC.
(6) Make sure that the model HFA100 installation is correct.
(7) Run the HGM port set program [ComSetHGM.exe].
All program> HGM> HGM Port Setting
(8) Port setting combo box will appear, click [], and select COM port for use, and
click [OK] button.
These settings will be stored in the settings file in the program folder.
(10)Driver program will start, and please wait for periodical running begin.
~Note When the driver program doesn't operate properly, please Refer to
troubleshooting “5-1: Connection with PC”.
[hfadrv2.exe]
[hgmXXX.exe]
[hgmXXX.exe]
[hfadrv2.exe]
[hgmXXX.exe]
[hgmXXX.exe]
(OFFLINE) (ONLINE)
Display Description
Display Description
Analyzer Status Analyzer Status shows whether the HGM is online or not.
The HGM is online if [HGC] is shown.
Refresh The latest update information for communication is dis-
played.
Maintenance mode Authorized service personnel use only.
product key
Data Saving Interval HV1, CV1 and SV1 files are stored onto your PC accord-
ing to the set data saving interval.
Auto Saving The HGM automatically saves files according to the set
auto saving interval
Calculation mode The HGM can calculate heat values using either [ISO] or
[GPA] calculation method.
Follow the procedures given below in order for the HGM to communicate with the
model HGC303.
Online OK OK NA OK OK
Offline OK NA OK OK NA
NA: not available
~Note For details on [GPA mode] selected in Calculation mode, refer to “3-5
: GPA mode” on page 3-55.
Data save
The last 4000 items of data are automatically stored in the RAM of your PC at each
data saving interval.
You can also save data by using the save function (See Table 3-12 or Table 3-14).
The data are saved as text files (.hv1 or.cv1 or.sv1) in C:\program files\hgm\data
(default) folder.
Text files (.hv1 and. cv1) are saved at the same time with the save function, which is
described in Table 3-12 No.3.
Editing data
If you want to edit saved data, open a saved file using to following procedure.
You can edit data using software such as Microsoft Excel™.
1 Start Microsoft Excel™
2 Select [Open]
3 Select the directory where the saved file is stored.
(Default directory C:\program files\hgm\data)
4 Select [All files] in “Files of type”.
5 Select a saved file, then click [open].
6 Follow the messages that come up on screen. (Click [Comma] at “delimiters”.)
HGM000011000S
Example:
Auto saving interval:1day (Selection: min. 1day, max. 10 days)
Data saving interval:5 min. and 5 sec. (Selection: min. 5 min. and 5 sec., max 60 min.
and 60 sec.)
(1) HGM data saving starts at 2001/07/25 19:00.
(This function starts after checking the box in [Setup HGM] then clicking the [OK]
button.)
(2) Analysis data and chromatograms (2001/07/25 19:00-2001/07/25 23:59) is saved
at 2001/07/26 0:00.
Saved file names: 010725as.hv1, 010725as.hv2, 010725as.cv1, 010725as.sv1,
010725as.sv2, 20010725HHMMas.cg1.
“as” stands for auto saving.
(3) Analysis data and chromatograms (2001/07/26 0:00-2001/07/26 23:59) are saved
at 2001/07/27 0:00.
This screen is divided into three graphs. On the right hand side is the measurement
data.
Screen Description
Top (blue) This graph shows heat value and the total of raw concentration
Center (white) Chromatogram
Bottom (red) This graph shows carrier gas pressure and oven temperature
Right panel Process gas analysis data
Indication panel
Data is updated every 5 minutes.
Table 3-11 Description of the indication panel
HGM000016000S
Figure 3-10 Trend graph of SCV and the total concentration (Raw)
Table 3-12 Trend graph of SCV and total raw concentration description
Chromatogram
Chromatogram is updated every 5 minutes.
9
1
2
10
3
4
5
6
11 13
7
8 HGM000017000S
12
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data.
1 7
2
3
4
6 8
HGM000019000S
11
Figure 3-13 Trend graph of carrier pressure and oven temp. control
This data is displayed according to the set data saving interval (Refer to “3-3-4 : Set up
HGM” on page 3-14).
Default interval: 5 sec.
Table 3-14 Description of trend graph of carrier gas pressure and oven temperature control
No. Display Description
1 URV (Oven Temp.) Upper Range Value for oven temperature.
Click the [URV] button to change the URV value
Default value: 62 deg.C.
2 Load (Offline) Recall saved data. File name extension:.sv1
3 Save (Online) The latest data is saved.
Default directory is “C:\Program files\hgm\data”.*
4 Print Verify that your printer is connected and working properly.
5 30min This indicates the time range of the horizontal axis. Select a time
(Time range select) range from pull-down menu:30 min., 60 min., 3 hours, 6 hours,
12 hours, 1 day, 2 days, 3 days, 6 days, 12 days
Default: 30min.
6 LRV (Oven Temp.) Lower Range Value for oven temperature
Default value: 54 deg. C.
Click the [LRV] button to change the LRV value
7 URV (Carr Press.) Upper Range Value for carrier pressure
Default value: 350kPa
Click the [URV] button to change the URV value
8 LRV (Carr Press.) Lower Range Value for carrier pressure.
Default value: 150 kPa
Click the [LRV] button to change the LRV value
9 Black circle Black indicates oven temperature.
10 Red circle Red indicates carrier pressure.
11 Time Online: Date and time of the latest data
Offline: Date and time of when the data was saved.
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data.
3-3-7 : Report
To create a report, click the [report] icon in the right panel of User's mode.
The following entry form for process gas data will appear.
HGM000020000S
To print out a report, click the [Preview] button then [print out].
~Note When you want to save report data, save either the report or the
chromatogram.
The chromatogram file (.cg1) includes the report data.
Section Description
Data items Display of user input data and saved file name
Gas Analysis by model PV1 -11 outputs data from model HGC303
HGC303
HGC Outputs Configura- Display of model HGC303 configuration data (PV12 -
tion data 20)
Heatvalue Calculation by All heat value data is calculated by the HGM and is dis-
HGM played.
HGC Status Model HGC303 status is displayed (online mode only)
-% DEV RF limit
-PV High/Low Alarm Configuration
-Output setting in Auto Calibration
-Normalization method
-Multi-stream function setting
Display Description
Default value
PV1-11 Low 0, High 100
PV12-15, 19 and 20 Low 0, High 10000
If some values exceed the High or Low Limit, HGC status button on HGM user's
mode screen changes red and blinks. (Click this button and get the error detail.)
~Note To complete the limit value change, close the PV High/Low Alarm
Configuration, then click the [Download to HGC] button on the con-
figuration mode screen.
Time 55 6:00 05 10 15 20 25 30 35 40
~Note HGC also transmits the error status “Total (raw) error” with New RF of
PV1 to 11 to HDM, if “Calibration data (New RF)” is selected.
~Note HDM LCD display of PV1 to 11 may overflow (ex: 9999.999) during output-
ting New RF.
~Note This setting is available for application of HGC (version 3.3 or later) and
HGM (version 4.83 or later) combination.
When click the [Load] button, configuration mode loads the setting from [Setup
HGM] screen.
The setting on the [Setup HGM] or on the [Configuration mode] is different. Each set-
ting is used as follows.
-Setup HGM: Used for the internal arithmetic of HGM
-Configuration mode: Used for the internal arithmetic of HGC
Different normalization method can be selected between HGC and HGM. Selecting
the same method is highly recommended without appropriate grounds.
~Note If the setting is changed, the color of the modified cell background
changes from blue to red. And that color returns from red to blue after
downloading to HGC.
Step Action
1 Click on [Return Menu]
2 Click on [Quit]
3 Wait until the hfadrv2 command screen closes.
3-4 : Calibration
The device has already been calibrated at the factory but we recommend that you
recalibrate it with your own calibration gas to ensure the accuracy of analysis in the
following cases:
1 When the model HGC303 is newly installed.
2 When an unused model HGC303 is started up again.
3 At a calibration interval decided by the user.
(Recommended calibration cycle is every 6 months.)
4 After the model HGC303 has been repaired or its parts have been replaced.
Manual calibration
Manual calibration procedure is as follows:
Manual calibration
Online HGM User’s mode
Change the inlet gas
from sample gas to calibration gas
Wait for output stabilization
Click the [calibration] button
in HGM User’s mode
(Refer to “3-3-6 : Main displays of HGM” on page 3-22)
Perform calibration
(Refer to “3-4-6 : Calibration methods” on page 3-53)
Calibration complete
Change the inlet gas
from calibration gas to sample gas
Waiting for output stabilization
Manual calibration procedure complete
Auto calibration
The auto calibration function will perform manual calibrations automatically at user
defined cycle, which is set by operator.
The setting is activated once the [start] button in auto calibration box is pressed.
(Figure 3-17)
(It’s not always necessary that the HGM is communicating with the model HGC303
after these settings have been made.)
2001/09/14 2001/09/15
18:00 20:00 0:00 20:00 2001/09/16
0:00 20:00 2001/09/17
0:00
Program start
No action
Calibration
Calibration mechanism
20:00 20:30 21:00
1/2 1/2
Calibration
HGC000016000D
m CAUTION
(1) Verify that the current time is correct in the [HGC time adjusting] box. [Current]
means the present time, which is being transmitted from the model HGC303.
(2) One measurement data will be used as calibration data. (not average data)
Therefore, set the enough hold time.
(3) Auto calibration will start on the day after the setting date.
(4) Model HGC303 holds the process variables to host control system during [Hold
time].
The model HGC303 discerns whether a new calibration should be performed or not by
using an RF (response factor) during auto calibration.
RF (Response Factor)
An RF is a correction factor used to discern whether a new calibration is correct as
compared with last valid calibration. (For each component)
The model HGC303 uses peak height for RF calculation.
Equation:
PH
RFn = -----------n-
Cal n
Where:
RF n = Response factor for component “n“
% DEV RF Limit
The model HGC303 automatically calculates a “RF% DEV” and decides whether the
value is smaller than the “% DEV RF Limit” or not.
The user can manually change the “% DEV RF Limit” in Configuration mode.
(Refer to “3-5-5 : Configuration mode” on page 3-58)
Signal to
No Calibration OLD RF NEW RF Judgement model
HDM303
m CAUTION
(1) For manual calibration and semi auto calibration, the RF will be the new RF
even if the |RF% DEV| > “%DEV RF Limit”. So the judgement is always OK.
(Example No.4 and 7)
(2) Only the checked components in calibration box are recognized and judged.
Refer to No.11 on “Table 3-20 Calibration factor setting” on page 3-50 for
details on the check box of each component.
1
18
2
3
4
5
6 7
19
15
14
17
9
20
21
22
16
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data
3-4-5 : Report
Click the [Preview] button and the following preview screen will appear.
HGC Calibration
Manual calibration
Table 3-22 Manual calibration procedure
Step Action
Select [Start]
Click [Refresh]
Each RF is calculated by HGM
RF%DEV>%DEV RF limit
RF error Click [Close]
Component
message will
Data table
appear
RF%DEV<=%DEV RF limit
HGC is calibrated
Data save
End
Both the [Refresh] and the [Calibration Start] buttons are active when in manual
calibration mode.
m CAUTION
Perform calibration after verifying that the carrier pressure and oven temperature
are both stable.
Do not input zero into any box.
Do not calibrate without calibration gas.
Auto calibration
Table 3-23 Operating auto calibration procedure
Step Action
If you need to cancel auto calibration, click [Stop] from the Auto calibration setting
panel.
Semi-auto calibration
Table 3-24 Operating semi auto calibration procedure
Step Action
(4) Configure the other values such as “Data Saving interval”, and then click the [OK]
button. The exchanging procedure is then completed.
~Note The file (.sv2) includes date/time, oven temperature (unit: °F) and
carrier gas pressure (unit: psi).
This screen is divided into three graphs with measurement data on the right.
Screen Description
Top (blue) Graph to monitor heat value and the total of raw concentration
Center (white) Chromatogram
Bottom (red) Graph to monitor carrier gas pressure and oven temperature
Right panel Process gas analysis data
The differences between ISO and GPA mode in User's mode are given below.
Table 3-29 The difference between ISO and GPA mode in user’s mode
Item ISO GPA
Top (Blue) Left vertical axis Unit: MJ/m3
Unit: BTU/CF or
BTU/lbm
Center (white) Chromatogram ISO and GPA use the same display.
Bottom (Red) Left vertical axis Unit: degree C Unit: degree F
Right vertical axis Unit: kPa Unit: psi
Right measure- Data Box (Rt, PH etc.) ISO and GPA use the same display.
ment data Refer to the next
Calculated Value Real or Ideal CV
page.
Oven temperature Unit: degree C Unit: degree F
Carrier gas pressure Unit: kPa Unit: psi
Other function ISO and GPA use the same function.
Indication panel
The data displayed in this panel is updated every 5 minutes.
Table 3-30 Indication panel description (GPA)
No. Panel Description
1
1 Date time Current date and time
2 Status Communication status appears when the 2
HGM is communicating with model
HGC303.
3 Data box Click a button to select a data type.
Default: Retention time (sec.)
4 Calculated Value Select between for BTU (dry), BTU (sat), 3
Density, or GPM
Default: BTU (dry)
Refer to description of calculated value on
page 3-69.
5 Oven Temp. and Oven temperature and carrier gas pressure
Carr. Pres.
6 Chromatogram The last 300 chromatograms are stored in
RAM. Save data as required. Select [pre-
vious XX] or [latest] to view the chro-
matogram.
If [previous XX] is selected, the auto
reload function stops. XX: 01-299
Return to [latest] to monitor the latest
chromatogram.
Auto reload function starts again.
7 [Field work] The model HGC303 holds outputs to the
host control system during field mainte-
nance.
Click the [Field work] button then [ON], 4
to set the holding time to [24hrs]. [Field
work] button blinks while performing
5
fieldwork.
8 [Calibration] Click the [Calibration] button to perform
calibration. The [Calibration] button 6
blinks during auto calibration.
Refer to “3-4 : Calibration” on page 3-44 8
9 [Report] Click the [Report] button to make a 9
report. 7
Refer to “3-3-7 : Report” on page 3-28. 10
10 HGC Status Green indicates that the model HGC303 is 11
functioning normally.
If the color changes to red, click this but-
ton to view the error message.
Refer to “Chapter 5 : Troubleshooting” on
page 5-1.
11 Return Menu Exit from User's mode
Return to Main Menu
Total (raw) Total raw concentration Total (raw) Total raw concentration
Real GHV Real Gross Heating Value (dry) Real GHV Real Gross Heating Value (sat)
Real NHV Real Net Heating Value (dry) Real NHV Real Net Heating Value (sat)
GHM Gross Heating Value per unit mass (dry) Z Compressibility factor (sat)
NHM Net Heating Value per unit mass (dry)
Z Compressibility factor (dry)
Density Others
Refer to the appendix for the calculation formula for each value.
9 11 10
1 7
2
3
4
6 8
12
Figure 3-26 BTU trend graph and the total of raw concentration
Table 3-31 Trend graph of BTU and Total raw concentration description
11 Disp. data Select a data type for heat value. (Gross/Dry, Gross/Sat, Net/Dry, Net/
Sat, GHM Dry, NHM Dry)
12 Time Online: Date and time of the latest data (data is reloaded every 5 min.)
Offline: Date and time of when the data was saved.
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data.
Chromatogram
1
2
3 10
4
5
6 11
7
8
13 12
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data.
1 7
2
3
4
8
6
11
Figure 3-28 Trend graph of carrier pressure and oven temp. control
Table 3-33 Trend graph of Carrier gas pressure and Oven temperature control
* In case of 64bit type operation system, data are saved in c:\Program files (x86)\hgm\data.
hfadrv2 stoped at “ClearVCR” HFA fieldbus Check the FB cable connection with
connection error HFA100.
HDM Check the FB cable connection with
connection error HDM303.
Confirm HDM303 powered on.
hfadrv2 couldn’t find “HGC303” HGC Check the FB cable connection with
Connection error HGC303.
Confirm that power is supplied to HGC303.
8
x 9 = 100 – y i + y 10 + y 11
i = 1
8 N2 3 2.95 3
11 C2H6 3 2.95 3
Decide which normalization method should be selected, and configure it to the model
HGC303 at the configuration mode. In addition, don't forget to setup it at the [Setup
HGM] screen for HGM.
Formulas
This section shows formulas for heat value calculations.
1) Real dry heating value per unit volume
2) Real saturated (sat) heating value per unit volume
3) Dry heating value per unit mass
4) Compressibility factor (dry)
5) Compressibility factor (sat)
6) Real Relative density, gas (dry)
7) Relative density, liquid
8) Gas density
9) Liquid density
10) Wobbe index (dry)
11) Wobbe index (sat)
12) Gallons per thousand cubic feet (GPM)
13) Reid vapor pressure
14) Weight fraction
15) Liquid volume fraction
16) Specific Heat K
n id
id xi GHV i
GHV - = -------------------------------------------
i =1
RealGHV dry = ---------------- -
Z dry Z dry
2) Real saturated (sat) heating value per unit volume: HV (sat) [BTU/CF]
Real GHV (sat), Real NHV (sat) are calculated using the following formula.
ex.
id
1 – xw GHV
Real GHV (sat) = --------------------------------------------
Z sat
C6+ * 5276.5 4889 20875 19342 0.0898 95.96 3.31309 0.67991 5.6685 22.616 3.019 1.0576
C3H8 2516.1 2315 21654 19922 0.0347 44.0956 1.5225 0.50719 4.2285 36.391 188.62 1.1316
i-C4H10 3251.9 3000 21232 19590 0.0441 58.1222 2.0068 0.56283 4.6925 30.637 72.644 1.0969
n-C4H10 3262.3 3011 21300 19658 0.047 58.1222 2.0068 0.5842 4.8706 31.801 51.567 1.0947
neo-C5H12 3993.9 3691.4 20958 19371 0.05774 72.15 2.4911 0.5967 4.975 26.11 35.9 1.0775
i-C5H12 4000.9 3699 21044 19456 0.0576 72.1488 2.4911 0.62514 5.212 27.414 20.474 1.0775
n-C5H12 4008.7 3707 21085 19481 0.0606 72.1488 2.4911 0.63071 5.2584 27.658 15.576 1.0764
N2 0 0 0 0 0.00442 28.0134 0.9672 0.80687 6.7271 91.128 0 1.3996
CH4 1010 909.4 23892 21511 0.0116 16.0425 0.5539 0.3 2.5 59.138 5000 1.3073
CO2 0 0 0 0 0.0195 44.0095 1.5195 0.81716 6.8129 58.746 0 1.2929
C2H6 1769.7 1619 22334 20429 0.0238 30.069 1.0382 0.35628 2.9704 37.488 800 1.1932
n-C6H14 4755.9 4404 20943 19393 0.0776 86.1754 2.9754 0.66406 5.5364 24.38 4.961 1.0635
n-C7H16 5502.6 5100 20839 19315 0.0951 100.2019 3.4597 0.68823 5.7379 21.73 1.619 1.0544
n-C8H18 6249 5796 20760 19256 0.1128 114.2285 3.944 0.70655 5.8907 19.57 0.5349 1.0476
Water 50.31 0 1059.8 0 0.0651 18.0153 0.62202 1 8.3372 175.62 0.9505 1.3295
Air 0 0 0 0 0.00537 28.9625 1 0.87586 7.3022 95.678 0 1.4002
*C6/C7/C8 = 47/35/17
GPA2145-09 constants are used for HGC (version 3.4 or later) and HGM (version 4.85
or later).
GPA2145-03 constants are used for HGC (version 3.0 or later) and HGM (version 4.60
or later).
GPA2145-00 constants are used for HGC (version earlier than 3.0) and HGM (version
earlier than 4.60).
ISO calculation
ISO calculation of Calorific value, density, relative density, Wobbe index and
compression factor will be calculated using PV1-PV11(normalized) and Helium
(normalized).
ISO calculation of Calorific value, density, relative density, Wobbe index and
compression factor will be calculated using PV1 - PV11 (normalized) and Helium
(normalized).
Formulas(ISO6976:1995)
This section shows formulas for heat value calculations of ISO6976:1995.
1) Compression factor
2) Superior calorific value
3) Inferior calorific value
4) Density
5) Relative density
6) Wobbe Index
4) Density (kg/m3)
Ideal gas density, Density (ideal)
= [sum (xj × Mj) + xhelium × Mhelium] × p2 / (R × (t2 + 273.15))
Real gas density, Density (real) = Density (ideal) / Zmix
where, xj = mole fraction for component j
xhelium = mol fraction for Helium
Mj = molar mass for component j (ISO6976 Table 1)
Mhelium = molar mass for Helium (ISO6976 Table 1)
p2 = metering reference pressure = 101.325 (kPa)
R = molar gas constant = 8.31451(J / (mol × K))
t2 = metering reference temperature (°C)
5) Relative density
Ideal gas relative density, Relative density (ideal)
= [sum (xj × Mj) + xhelium × Mhelium] / Mair
Real gas relative density, Relative density (real)
= Relative density (ideal) × Zair / Zmix
where, xj = mole fraction for component j
xhelium = mol fraction for Helium
Mj = molar mass for component j (ISO6976 Table 1)
Mhelium = molar mass for Helium (ISO6976 Table 1)
Mair = molar mass for Air (ISO6976 Table 1)
Zair = compression factor for Air at (t2, p2) (ISO6976 Table 1)
Formulas(ISO6976:2016)
This section shows formulas for heat value calculations of ISO6976:2016.
1) Compression factor
2) Calorific value on a molar basis
Gross calorific value
Net calorific value
3) Calorific value on a volume basis
Ideal gas gross calorific value
Ideal gas net calorific value
Real gas gross calorific value
Real gas net calorific value
4) Associated properties
Ideal gas relative density
Ideal gas density
Ideal gas gross Wobbe index
Ideal gas net Wobbe index
Real gas relative density
Real gas density
Real gas gross Wobbe index
Real gas net Wobbe index
1) Compression factor Z
Compression factor Z = 1 - (p2/p0) * [ sum(xj * sj) + (xhelium * sjhelium) ]^2
where HcG0j ideal gas gross molar basis calorific value of component j
(Table 3)
4) Associated properties
Ideal gas relative density G0 = M / Mair
ISO6976
title: Natural gas - Calculation of calorific values, density, relative density and Wobbe
indices from composition
ISO6976 specifies calculation formulas and physical properties.
Following table shows ISO6976 edition and our product software relationship.
HGC000013000D
Drawings
Software Compatibility
PC S/W Version
HGM HDS
HFA100 Version
FEATURES
Small size for easy field installation Digital communication for system integration
The model HGC303 has a compact design thus facilitating The model HGC303 is capable of supporting MODBUS
field installation. In addition, the device can be mounted protocols with optional analog output capabilities and has
with a sampling system in the field. been tested with leading flow computers.
• Small compact packaging PC monitoring and online diagnostics
• No analyzer house is required The Heat Value Gas Chromatograph Monitor (HGM) is a
• Flameproof certified PC-based software that allows the user to view all data and
diagnostic information from a laptop computer.
Pre-engineered analysis and calculations for
Natural Gas Metering
The model HGC303 has pre-engineered analysis and cal- ADVANTAGES
culations for Natural Gas Metering so that no additional Simple to start-up and easy to maintain
programming or application work is required.
A huge amount of time and cost in the analyzer system start
• Easy to set up straight out of the box up phase can be saved with the model HGC303’s unique
• Analysis of 11 components and pre-configured value packaging and pre-engineered functions.
calculations The unit’s easy-to-maintain design contributes to time and
• Analysis and calculations based on international standards cost savings and it can be repaired without the need for
analyzer expertise.
B-1
FUNCTIONAL SPECIFICATIONS Analyzer outputs
PV1 C6+ (sum of C6+)(mol%)
Principle of measurement
PV2 C3H8 (propane)(mol%)
Gas chromatography PV3 i-C4H10 (i-butane)(mol%)
Measured gas streams PV4 n-C4H10 (n-butane)(mol%)
1 PV5 neo-C5H12 (neo-pentane)(mol%)
PV6 i-C5H12 (i-pentane)(mol%)
Analyzed components PV7 n-C5H12 (n-pentane)(mol%)
11 PV8 N2 (nitrogen)(mol%)
Analysis time PV9 CH4 (methane)(mol%)
300 sec. PV10 CO2 (carbon dioxide)(mol%)
PV11 C2H6 (ethane)(mol%)
Detector PV12 SCV (real) (MJ/m3) [default], (kJ/m3), (kWh/m3)
Micro TCD (Thermal Conductivity Detector) or SCV (ideal) (MJ/m3), (kJ/m3), (kWh/m3)
Chromatographic method or ICV (real) (MJ/m3), (kJ/m3), (kWh/m3)
ISO 6974, part 4 or ICV (ideal) (MJ/m3), (kJ/m3), (kWh/m3)
PV13 Density (real) (kg/m3) [default]
Heat value calculation method or Density (ideal) (kg/m3)
ISO 6976 or Relative density (real)
or Relative density (ideal)
Gas to be analyzed
PV14 Wobbe index (real) (MJ/m3) [default], (kJ/m3), (kWh/m3)
Natural gas
or Wobbe index (ideal) (MJ/m3), (kJ/m3), (kWh/m3)
Component measuring ranges and minimum PV15 Compressibility factor Zmix
detection PV16 Total of raw concentrations(mol%)
Components Ranges Minimum detection PV17 Oven temperature(degree C)
(mol%) (mol%) PV18 Carrier gas pressure(kPa)
Sum of C6+ 0-0.3 0.01 PV19 ICV (real) (MJ/m3) [default], (kJ/m3), (kWh/m3)
C3H8 (propane) 0-3 0.05 or ICV (ideal) (MJ/m3), (kJ/m3), (kWh/m3)
i-C4H10 (i-butane) 0-1 0.01 PV20 Relative density (real) [default],
Relative density (ideal) or Helium normalized conc. (mol%)
n-C4H10 (n-butane) 0-1 0.01
neo-C5H12 0-0.5 0.01 Auto-calibration
(neo-pentane)
External solenoid valve and HDM contact are required.
i-C5H12 (i-pentane) 0-0.5 0.01
n-C5H12 (n-pentane) 0-0.5 0.01 Normalization of concentrations
N2 (nitrogen) 0-20 0.1
On-line diagnostics
CH4 (methane) 50-100 -
CO2 (carbon dioxide) 0-10 0.05 Hazardous area certification
C2H6 (ethane) 0-15 0.05 ISSeP/ATEX certifications: II 2 GD EEx d IIC T6
IP65
B-2
Process Gas
Temperature
-10°C to +50°C
Flow rate
50 ± 20ml/min
Dust and mist
None
Moisture
Less than 2000 ppm
Coexisting components limit
H2 < 0.1 mol%
He < 0.1 mol%
Oxygen < 0.1 mol%
H2S (dry) < 0.1 mol%
Ambient temperature limits
-10°C to +50°C
-40°C to +70°C for storage and transportation
Ambient humidity Range
0-95%RH
CE marking
Electromagnetic compatibility (EMC):2014/30/EU
Equipment explosive atmospheres (ATEX): 94/9/EC
PERFORMANCE SPECIFICATIONS
Repeatability of analysis
± 0.05% CV
PHYSICAL SPECIFICATIONS
Color
Metallic light green, silver
Material
Body
Cast aluminum
Oven
Cast aluminum
Wet-parts
304 Stainless steel, polyimide
Sensor
Pt, glass, gold
Dimensions
W: 100 mm × D: 115 mm × H: 244 mm
Weight
3.5kg
B-3
COMMUNICATIONS
The model HGC303 communicates to a PC for configuration, maintenance and data transmission. PC and HGC bus connec-
tions are provided as standard equipment.
A specific Windows-based model HGC303 software, the HGC Monitor HGM, enables convenient model HGC303 instrument
control in a user-friendly environment. Retransmission of data to the central system can be performed via the Internet. Mod-
bus communication is also available for networking with, for example, a flow computer or SCADA system.
B-4
MODEL SELECTION
Heat Value Gas Chromatograph
HGC303 - I II
DIMENSIONS
[Unit: mm]
B-5
B-6
Heat Value Gas Chromatograph
GPA standard version
Model HGC303
OVERVIEW
The model HGC303 is the world smallest gas chromato-
graph, which is capable of analyzing 11 different compo-
nents of natural gas and digitally publishing the derived
parameters, such as calorific value, Wobbe-index, or density.
By using our leading sensing & control technology, we de-
signed the model HGC303 specifically for custody transfer,
quality control, and other natural gas applications to meet
the demands of the expanding natural gas market.
The model HGC303’s size, weight, cost and other great
features gives the user the benefits of functionality, flexibility
and economy from the production site to the pipe line sta-
tion to the gas distributor and onto the end user’s station.
FEATURES
Small size for easy field installation Digital communication for system integration
The model HGC303 has a compact design thus facilitating The model HGC303 is capable of supporting MODBUS
field installation. In addition, the device can be mounted protocols with optional analog output capabilities and has
with a sampling system in the field. been tested with leading flow computers.
• Small compact packaging PC monitoring and online diagnostics
• No analyzer house is required The Heat Value Gas Chromatograph Monitor (HGM) is a
• Flameproof certified PC-based software that allows the user to view all data and
diagnostic information from a laptop computer.
Pre-engineered analysis and calculations for
Natural Gas Metering
The model HGC303 has pre-engineered analysis and cal- ADVANTAGES
culations for Natural Gas Metering so that no additional Simple to start-up and easy to maintain
programming or application work is required.
A huge amount of time and cost in the analyzer system start
• Easy to set up straight out of the box up phase can be saved with the model HGC303’s unique
• Analysis of 11 components and pre-configured value packaging and pre-engineered functions.
calculations The unit’s easy-to-maintain design contributes to time and
• Analysis and calculations based on international standards cost savings and it can be repaired without the need for
analyzer expertise.
B-7
FUNCTIONAL SPECIFICATIONS Analyzer outputs
PV1 C6+ (sum of C6+) (mol%)
Principle of measurement
PV2 C3H8 (propane) (mol%)
Gas chromatography PV3 i-C4H10 (i-butane) (mol%)
Measured gas streams PV4 n-C4H10 (n-butane) (mol%)
1 PV5 neo-C5H12 (neo-pentane) (mol%)
PV6 i-C5H12 (i-pentane) (mol%)
Analyzed components PV7 n-C5H12 (n-pentane) (mol%)
11 PV8 N2 (nitrogen) (mol%)
Analysis time PV9 CH4 (methane) (mol%)
300 sec. PV10 CO2 (carbon dioxide) (mol%)
PV11 C2H6 (ethane) (mol%)
Detector PV12 Real Gross HV (dry) (BTU/CF) [default]
Micro TCD (Thermal Conductivity Detector) or Real Gross HV (sat) (BTU/CF)
Chromatographic method or Ideal Gross HV (dry) (BTU/CF)
ISO 6974, part 4 or Ideal Gross HV (sat) (BTU/CF)
or Gross HV (dry) (BTU/lbm)
Heat value calculation method PV13 Gas Density (lb/1000CF) [default]
GPA2172 or Real Relative Density (dry gas)
or Real Relative Density (sat gas)
Gas to be analyzed
or Ideal Relative Density (dry gas)
Natural gas
or Ideal Relative Density (sat gas)
Component measuring ranges and minimum PV14 Real Wobbe index (dry) (BTU/CF) [default]
detection or Real Wobbe index (sat) (BTU/CF)
or Ideal Wobbe index (dry) (BTU/CF)
Components Ranges Minimum detection
or Ideal Wobbe index (sat) (BTU/CF)
(mol%) (mol%)
PV15 Compressibility factor Z (dry gas)
Sum of C6+ 0-0.3 0.01
PV16 Total of raw concentrations (mol%)
C3H8 (propane) 0-3 0.05
PV17 Oven temperature (degree C)
i-C4H10 (i-butane) 0-1 0.01
PV18 Carrier gas pressure (kPa)
n-C4H10 (n-butane) 0-1 0.01
PV19 Real Net HV (dry) (BTU/CF) [default]
neo-C5H12 0-0.5 0.01
or Real Net HV (sat) (BTU/CF)
(neo-pentane)
or Ideal Net HV (dry) (BTU/CF)
i-C5H12 (i-pentane) 0-0.5 0.01 or Ideal Net HV (sat) (BTU/CF)
n-C5H12 (n-pentane) 0-0.5 0.01 or Net HV (dry) (BTU/lbm)
N2 (nitrogen) 0-20 0.1 PV20 Real Relative Density (dry gas) [default]
CH4 (methane) 50-100 - or Real Relative Density (sat gas)
CO2 (carbon dioxide) 0-10 0.05 or Ideal Relative Density (dry gas)
C2H6 (ethane) 0-15 0.05 or Ideal Relative Density (sat gas) or Specific Heat K
Auto-calibration
External solenoid valve and HDM contact are required.
Normalization of concentrations
On-line diagnostics
Hazardous area certification
FM Approvals
Explosionproof for C1 I, Div 1, Gps C & D
Flameproof for Cl I, Zone 1, AEx d IIB
Dust-ignitionproof for Cl II &III, Div 1, Gps E, F&G
hazardous (classified) locations
Seal All Conduits Within 18 in.
NEMA Type 4X
IP 65
B-8
Process Gas
Temperature
14°F to 122°F (-10°C to +50°C)
Flow rate
50 ± 20ml/min
Dust and mist
None
Moisture
2000 ppm or less
Coexisting components limit
H2 < 0.1 mol%
He < 0.1 mol%
Oxygen < 0.1 mol%
H2S (dry) < 0.1 mol%
Ambient temperature limits
14°F to 122°F (-10°C to +50°C)
-40°F to 158°F (-40°C to +70°C) for storage and transportation
Ambient humidity Range
0-95%RH
CE marking
Electromagnetic compatibility (EMC): 2014/30/EU
PERFORMANCE SPECIFICATIONS
Repeatability of analysis
± 0.05% CV
PHYSICAL SPECIFICATIONS
Color
Metallic light green, silver
Material
Body
Cast aluminum
Oven
Cast aluminum
Wet-parts
304 Stainless steel, polyimide
Sensor
Pt, glass, gold
Dimensions
W: 3.9 in. × D: 4.5 in. × H: 9.6 in.
(W: 100 mm × D: 115 mm × H: 244 mm)
Weight
7.7lbs (3.5kg)
B-9
COMMUNICATIONS
The model HGC303 communicates to a PC for configuration, maintenance and data transmission. PC and Fieldbus connec-
tions are provided as standard equipment.
A specific Windows-based model HGC303 software, the HGC Monitor HGM, enables convenient model HGC303 instrument
control in a user-friendly environment. Retransmission of data to the central system can be performed via the Internet. Mod-
bus communication is also available for networking with, for example, a flow computer or SCADA system.
B-10
MODEL SELECTION
Heat Value Gas Chromatograph
HGC303 - I II
DIMENSIONS
[Unit: inch]
B-11
B-12
Terms and Conditions
We would like to express our appreciation for your purchase and use of Azbil Corporation's products. You are required to
acknowledge and agree upon the following terms and conditions for your purchase of Azbil Corporation's products (system
products, field instruments, control valves, and control products), unless otherwise stated in any separate document, including,
Ter
without limitation, estimation sheets, written agreements, catalogs, specifications and instruction manuals.
2. Ascertainment of suitability
s
You are required to ascertain the suitability of Azbil Corporation's product in case of your use of the same with your
machinery, equipment, etc. (hereinafter referred to as “Equipment”) on your own responsibility, taking the following
matters into consideration:
(1) Regulations and standards or laws that your Equipment is to comply with.
(2) Examples of application described in any documents provided by Azbil Corporation are for your reference
purpose only, and you are required to check the functions and safety of your Equipment prior to your use.
(3) Measures to be taken to secure the required level of the reliability and safety of your Equipment in your use a
Although azbil is constantly making efforts to improve the quality and reliability of Azbil Corporation's
products, there exists a possibility that parts and machinery may break down.
You are required to provide your Equipment with safety design such as fool-proof design, *1 and fail-safe
n
design*2 (anti-flame propagation design, etc.), whereby preventing any occurrence of physical injuries, fires,
significant damage, and so forth. Furthermore, fault avoidance, *3 fault tolerance,*4 or the like should be
d
incorporated so that the said Equipment can satisfy the level of reliability and safety required for your use.
S1
(2) For use of specific purposes, such as:
* Nuclear energy/radiation related facilities
[For use outside nuclear energy controlled areas] [For use of Azbil Corporation's Limit Switch For Nuclear
Energy]
* Machinery or equipment for space/sea bottom
* Transportation equipment
[Railway, aircraft, vessels, vehicle equipment, etc.]
* Antidisaster/crime-prevention equipment
* Burning appliances
* Electrothermal equipment
* Amusement facilities
* Facilities/applications associated directly with billing
(3) Supply systems such as electricity/gas/water supply systems, large-scale communication systems, and traffic/air
traffic control systems requiring high reliability
(4) Facilities that are to comply with regulations of governmental/public agencies or specific industries
(5) Machinery or equipment that may affect human lives, human bodies or properties
(6) Other machinery or equipment equivalent to those set forth in items (1) to (5) above which require high reliability
and safety
6. Other precautions
Prior to your use of Azbil Corporation's products, you are required to understand and comply with specifications (e.g.,
conditions and environment of use), precautions, warnings/cautions/notices as set forth in the technical documents
prepared for individual Azbil Corporation's products, such as catalogs, specifications, and instruction manuals to ensure
the quality, reliability, and safety of those products.
7. Changes to specifications
Please note that the descriptions contained in any documents provided by azbil are subject to change without notice for
improvement or for any other reason.
For inquires or information on specifications as you may need to check, please contact our branch offices or sales offices,
or your local sales agents.
9. Scope of services
Prices of Azbil Corporation's products do not include any charges for services such as engineer dispatch service.
Accordingly, a separate fee will be charged in any of the following cases:
(1) Installation, adjustment, guidance, and attendance at a test run
(2) Maintenance, inspection, adjustment, and repair
(3) Technical guidance and technical education
(4) Special test or special inspection of a product under the conditions specified by you
Please note that we cannot provide any services as set forth above in a nuclear energy controlled area (radiation
controlled area) or at a place where the level of exposure to radiation is equivalent to that in a nuclear energy controlled
area.
AAS-511A-014-09
S2
Document Number: CM2-HGC100-2001
Document Name: Heat Value Gas Chromatograph
Model: HGC303
User’s Manual
Date: 1st edition: Apr. 2001
19th edition: May 2017
Issued/Edited by: Azbil Corporation