Calibration: Technical Publication CA-1044RA
Calibration: Technical Publication CA-1044RA
Calibration: Technical Publication CA-1044RA
CA-1044RA
Calibration
Mobile X-Ray Unit
Mobile X-Ray Unit
Calibration
REVISION HISTORY
This Document is the english original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
CA-1004R2
Mobile X-Ray Unit
Calibration
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 CALIBRATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
CA-1044RA i
Mobile X-Ray Unit
Calibration
ii CA-1044RA
Mobile X-Ray Unit
Calibration
SECTION 1 INTRODUCTION
CA-1044RA 1
Mobile X-Ray Unit
Calibration
ACCURACY
PARAMETERS
(with 12 BITS HT Controller)
kV ± (3% + 1 kV)
The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray tube.
2 CA-1044RA
Mobile X-Ray Unit
Calibration
Note . Test points on the HT Controller Board can also be used to monitor
the kV and mA readings but should not be used to calibrate the
unit. These test points must be checked with scope:
• Verify that dip switch 3024SW2-3 on the ATP Console CPU Board is in
“On” position to permit the service mode.
CA-1044RA 3
Mobile X-Ray Unit
Calibration
3000SW2-4 Position during operation -- Digital mA Loop Closed Digital mA Loop Open
Note . Only for Generators with Low Speed Starter (LF-RAC) (it does not
apply to Generators with High Speed Starter LV-DRAC):
4 CA-1044RA
Mobile X-Ray Unit
Calibration
The values stored in these Extended Memory Locations only affect to Exposure
Times for techniques below 20 ms. The Memory Locations which affect short
exposure times are:
MEMORY LOCATION
FUNCTION
TUBE-1
The Generator has been optimized at the factory to produce correct exposures
at the lower times (<20 ms.) Therefore do not change the value factory set for
E13 and E29 Memory Locations and only adjust the value for E15 Memory
Location according to the HV Cables length.
2. Select the E13 Memory Location. Value in this Memory Location adjusts
the time delay of the exposure. It is factory set for a value of 17, 18 or 19
(default value is “18”). Only read this value, do not change it.
3. Select the E15 Memory Location. Value in this Memory Location is set
in relation to the length in meters of one of the HV Cables
(1 ft = 0.3048 m). Verify the HV Cable length in meters and set the
following value:
4m 27
6m 31
9m 38
12 m 45
14 m 49
16 m 54
5. Exit from calibration mode and record the new values in the Data Book.
CA-1044RA 5
Mobile X-Ray Unit
Calibration
Extended Memory Location E06 contains the calibration factor for Digital kV
Loop Open. Each number above or below of the indicated in the E06 memory
location increases or decreases respectively the kV gain value.
The kV Gain for Digital kV Loop Open can be manually calibrated with a
Non-Invasive kV Meter (recommended procedure) or Auto-calibrated with HV
Bleeder.
3. Select the E06 Memory Location and read the calibration value by
pressing the “AEC Reset” push-button. Enter the value “200” and store
it by pressing the “AEC Reset” push-button.
6 CA-1044RA
Mobile X-Ray Unit
Calibration
Select 80 kV, 200 mA and make an exposure. The mAs read on the mAs
Meter must be the same mAs displayed on the calibration screen with a
tolerance of ±0.1 mAs (tolerance of the parameter and mAs Meter).
If the mAs is low, increase the filament number. If the mA is high (or
“Generator Overload” Error is shown), decrease the filament number.
Press the “AEC Reset” push-button before making a new exposure.
Repeat until the mAs read is correct and the mA station is calibrated.
6. Select: 80 kV, 200 mA, 100 ms. Make an exposure and note the kV value
at the end of the exposure.
d. When it is correct, record the new value for E06 Memory Location
in the Data Book.
CA-1044RA 7
Mobile X-Ray Unit
Calibration
G Remove the HV Cables from the X-ray Tube and connect them to
the HV Bleeder, then connect a short couple of HV Cables from the
HV Bleeder to the X-ray Tube.
Keep pressed the “kV increment” and “Power On” push-buttons again
until code “222” disappears on the Console to confirm Auto-calibration
mode or press only the “Power On” push-button to leave Auto-calibration
mode and return to manual calibration.
5. Read the kVp measured with the HV Bleeder and enter this value on the
kV Display by pressing the “kV Increment” or “kV Decrement” buttons.
7. Read the new value stored in E06 Memory Location pressing the “AEC
Reset” push-button. Record this value in the Data Book.
8 CA-1044RA
Mobile X-Ray Unit
Calibration
Extended Memory Locations E21, E03 and E05 contain the calibration factor
for Digital mA Loop Closed. Each number above or below the indicated in the
memory locations increases or decreases respectively the mA gain value.
Note . Values in E21, E03 and E05 Memory Locations are only related to
the Generator performance (they are not related to the X-ray
Tube(s) installed), so values in these Memory Locations are
factory adjusted. Only perform this procedure if the HT
Controller Board and/or HV Transformer have been replaced in
the unit.
CA-1044RA 9
Mobile X-Ray Unit
Calibration
2. Enter Auto-calibration mode pressing the “Power On” and “kV increment”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increment” and “Power On” push-buttons again
until code “222” disappears on the Console to confirm Auto-calibration
mode or press only the “Power On” push-button to leave Auto-calibration
mode and return to manual calibration.
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
3.00 and 4.00 mAs) in the mAs Display pressing the “mAs increment” or
“mAs decrement” push-buttons.
6. Read the new value stored in E03 memory location pressing the
“AEC Reset” push-button. Record this value in the Data Book.
10 CA-1044RA
Mobile X-Ray Unit
Calibration
2. Enter Auto-calibration mode pressing the “Power On” and “kV increment”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increment” and “Power On” push-buttons again
until code “222” disappears on the Console to confirm Auto-calibration
mode or press only the “Power On” push-button to leave Auto-calibration
mode and return to manual calibration.
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
7.00 and 8.50 mAs) in the mAs Display pressing the “mAs increment” or
“mAs decrement” push-buttons.
6. Read the new value stored in E05 memory location pressing the
“AEC Reset” push-button. Record this value in the Data Book.
CA-1044RA 11
Mobile X-Ray Unit
Calibration
It is recommended to start with the Small Focal Spot (first group) and continue
with the Large Focal Spot (second group).
12 CA-1044RA
Mobile X-Ray Unit
Calibration
Auto-calibration starts with the minimum available mA station for the selected
Focal Spot at 50 kV and follows with the other combinations of mA stations for
the Focal Spot at the kV break points as indicated in the next table.
1. Select one available workstation for the X-ray Tube selected. This
workstation has to be one of the previously configured as “Direct”.
2. Enter calibration mode and select one of the configured mA stations for
the Small Focal Spot.
3. Enter Auto-calibration mode pressing the “Power On” and “kV increment”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increment” and “Power On” push-buttons again
until code “222” disappears on the Console to confirm Auto-calibration
mode or press only the “Power On” push-button to leave Auto-calibration
mode and return to manual calibration.
4. Check that the Heat Units available for the X-ray Tube are 100%
(kV Display shows “H -- --” on the Console).
CA-1044RA 13
Mobile X-Ray Unit
Calibration
If the Heat Units available for the X-ray Tube are less than 40%,
exposures are inhibited momentarily and code “111” will be flashed on the
Console accompanied of an alarm. The alarm will stop when the X-ray
Tube begins to cool and recovers the Heat Units capacity, exposures can
be performed again even though code “111” is shown on the Console.
At this point, it is recommended to wait until the Heat Units available are
closed to the 80% of the X-ray Tube capacity without making any
exposure.
14 CA-1044RA
Mobile X-Ray Unit
Calibration
6. Repeat the same procedure for the Large Focal Spot selecting one of
the configured mA stations for the Large Focal Spot previous to enter in
Auto-calibration mode.
7. After performing both procedures (for Small and Large Focal Spots),
select in calibration mode each combination of the available mA stations
at the kV break points (40, 50, 80 and 120 kV). Press the “AEC Reset”
push-button to read on the kV Display the new value of the Filament
Current Number stored for each combination and record the new values
in the Data Book.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
9. Turn the Generator power OFF and set Dip Switch 3000SW2-4 on the
HT Controller Board in “Off” position (Digital mA Loop Closed).
This procedure describes the Manual calibration of all the Filament Current
Numbers. It has to be also used to calibrate the kV / mA combinations not
performed during Auto-calibration procedure because it has not been
completed. These combinations have the Filament Current Number set to
“344”, so only these combinations have to be manually calibrated as described
in this procedure.
Although the suggested values (Table 2-1) could change depending on the
X-ray tube used, entering those values will approximate accurate calibration
without making excessive exposures.
CA-1044RA 15
Mobile X-Ray Unit
Calibration
Table 2-1
mA Calibration Numbers Change
Note.-- Some Generator models do not contain all the mA stations listed above.
16 CA-1044RA
Mobile X-Ray Unit
Calibration
The kV Display shows again the kV when the “AEC Reset” push-button is
released. Note that in calibration mode, only the mA stations and kV (at the
break points) can be selected.
2. Turn the Generator ON and select one available workstation for the X-ray
Tube selected. This workstation has to be one of the previously
configured as “Direct”.
3. Check that the Heat Units available for the X-ray Tube are 100%
(kV Display shows “H -- --” on the Console).
Exit calibration mode and select the mA station and Exposure Time of
each kV / mA combination not allowed. Increase or decrease the kV
value as required to determine the kV value allowed nearest to the kV
value of the combination. Note the kV value allowed in the respective cell
of Table 2-1.
CA-1044RA 17
Mobile X-Ray Unit
Calibration
6. Make an exposure. The mAs read on the mAs Meter must be the mAs
displayed on the Console ± 0.2 mAs (tolerance of the parameter and
mAs Meter). If the mAs read is close to zero, increase the filament
number in big steps (a.e. increase values in 40). As the mAs read is close
to the mAs displayed on the Console, increase (or reduce) the filament
number in smaller steps (a.e. increase value in 30, 20, 10, ...).
Note . Press the “AEC Reset” push-button to store the new data (filament
number) before selecting the next kV or mA stations.
7. Select the next mA station at 80 kV. Before making any exposure, enter
as filament number the value calibrated for the previous mA station
increased in 10.
Press the “AEC Reset” push-button to read on the kV Display the new
value of the Filament Current Number stored for each mA station at
80 kV. Record the new values in the Data Book.
18 CA-1044RA
Mobile X-Ray Unit
Calibration
CA-1044RA 19
Mobile X-Ray Unit
Calibration
13. Select the mA station, Exposure Time and kV value noted for each
kV / mA combination not allowed (refer to step-4.). Check calibration at
these kV / mA combinations by making exposures. If needed, enter in
calibration mode and correct the Filament Current Number of the
respective kV / mA combination not allowed.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
20 CA-1044RA
Mobile X-Ray Unit
Calibration
Extended Memory Location E06 contains the calibration factor for Digital kV
Loop Open. Each number above or below of the indicated in the E06 Memory
Location increases or decreases respectively the kV gain value.
The kV Gain for Digital kV Loop Open can be manually calibrated with a
Non-Invasive kV Meter (recommended procedure) or Auto-calibrated with HV
Bleeder.
3. Select the E06 Memory Location and read the calibration value by
pressing the “Read” button. Enter the value “200” and store it by pressing
the “Store” button.
CA-1044RA 21
Mobile X-Ray Unit
Calibration
Select 80 kV, 200 mA, Large Focus and make an exposure. The mAs
read on the mAs Meter must be the same mAs displayed on the
calibration screen with a tolerance of ±0.1 mAs (tolerance of the
parameter and mAs Meter).
If the mAs is low, increase the filament number. If the mA is high (or
“Generator Overload” Error is shown), decrease the filament number.
Press the “Store” button before making a new exposure. Repeat until the
mAs read is correct and the mA station is calibrated.
22 CA-1044RA
Mobile X-Ray Unit
Calibration
6. Enter in Manual Mode of the Application and select: 80 kV, 200 mA,
100 ms. Make an exposure and note the kV value at the end of the
exposure.
d. When it is correct, record the new value for E06 Memory Location
in the Data Book.
CA-1044RA 23
Mobile X-Ray Unit
Calibration
G Remove the HV Cables from the X-ray Tube and connect them to
the HV Bleeder, then connect a short couple of HV Cables from the
HV Bleeder to the X-ray Tube.
24 CA-1044RA
Mobile X-Ray Unit
Calibration
4. Read the kVp measured with the HV Bleeder and enter this value on the
kV Display by pressing the “kV Increase” or “kV Decrease” buttons.
CA-1044RA 25
Mobile X-Ray Unit
Calibration
Extended Memory Locations E03 and E05 contain the calibration factor for
Digital mA Loop Closed. Each number above or below the indicated in the
Memory Locations increases or decreases respectively the mA gain value.
Note . Values in E03 and E05 Memory Locations are only related to the
Generator performance (they are not related to the X-ray Tube(s)
installed), so values in these Memory Locations are factory
adjusted. Only perform this procedure if the HT Controller Board
and/or HV Transformer have been replaced in the unit.
26 CA-1044RA
Mobile X-Ray Unit
Calibration
CA-1044RA 27
Mobile X-Ray Unit
Calibration
28 CA-1044RA
Mobile X-Ray Unit
Calibration
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
3.00 and 4.00 mAs) in the mAs Display pressing the “mAs increase” or
“mAs decrease” buttons.
CA-1044RA 29
Mobile X-Ray Unit
Calibration
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
7.00 and 8.50 mAs) in the mAs Display pressing the “mAs increase” or
“mAs decrease” buttons.
30 CA-1044RA
Mobile X-Ray Unit
Calibration
It is recommended to start with the Small Focal Spot (first group) and continue
with the Large Focal Spot (second group).
CA-1044RA 31
Mobile X-Ray Unit
Calibration
Auto-calibration starts with the minimum available mA station for the selected
Focal Spot at 50 kV and follows with the other combinations of mA stations for
the Focal Spot at the kV break points as indicated in the next table.
3. Select the Small Focal Spot by pressing the “Small” button and then
press the “OK” button.
32 CA-1044RA
Mobile X-Ray Unit
Calibration
CA-1044RA 33
Mobile X-Ray Unit
Calibration
6. Check that the Heat Units capacity available for the X-ray Tube is 100%
or nearly (HU 0% displayed on the screen).
If the Heat Units value displayed is more than 60%, exposures are
inhibited momentarily and message “Tube too hot” is shown on the
screen. This message will disappear and exposures can be made again
when the X-ray Tube begins to cool and recovers the Heat Units capacity.
34 CA-1044RA
Mobile X-Ray Unit
Calibration
Select the Large Focal Spot by pressing the “Large” button and then
press the “OK” button.
CA-1044RA 35
Mobile X-Ray Unit
Calibration
9. After performing both procedures (for Small and Large Focal Spots),
enter in “Manual Calibration” mode and select each combination of the
available mA stations for each Focal Spot at the kV break points (40, 50,
80 and 120 kV). Press the “Read” button to read on the calibration panel
the new value of the Filament Current Number stored for each
combination and record the new values in the Data Book.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
11. Perform the Auto-calibration procedure for the second X-ray Tube.
12. Turn the Generator power OFF and set Dip Switch 3000SW2-4 on the
HT Controller Board in “Off” position (Digital mA Loop Closed).
This procedure describes the Manual calibration of all the Filament Current
Numbers. It has to be also used to calibrate the kV / mA combinations not
performed during Auto-calibration procedure because it has not been
completed. These combinations have the Filament Current Number set to
“344”, so only these combinations have to be manually calibrated as described
in this procedure. If Auto-calibration for one of the Focal Spots has been
successful, it is only required to perform the manual calibration of the mA station
do not calibrate for the other Focal Spot.
Although the suggested values (Table 2-1) could change depending on the
X-ray tube used, entering those values will approximate accurate calibration
without making excessive exposures.
In “Manual Calibration” mode, the Filament Current Numbers are shown on the
calibration panel by pressing the “Read” button after selecting the respective
kV / mA combination. The value can be changed by pressing the “Increase” or
“Decrease” buttons of the calibration panel and stored by pressing the “Store”
button.
Note that in calibration mode, only the mA stations and kV (at the break points)
can be selected.
36 CA-1044RA
Mobile X-Ray Unit
Calibration
Table 2-2
mA Calibration Numbers Change
Note.-- The mA station values depends on the Generator model. Some models do not contain all the mA stations listed above.
CA-1044RA 37
Mobile X-Ray Unit
Calibration
4. Check that the Heat Units available for the X-ray Tube is 100% or nearly
(HU 0% displayed on the screen).
6. Exit calibration mode. Select the mA station and Exposure Time of each
kV / mA combination not allowed. Increase or decrease the kV value as
required to determine the kV value allowed nearest to the kV value of the
combination. Note the kV value allowed in the respective cell of
Table 2-1.
38 CA-1044RA
Mobile X-Ray Unit
Calibration
8. Make an exposure. The mAs read on the mAs Meter must be the same
mAs displayed on the calibration screen with a tolerance of ± 0.1 mAs
(tolerance of the parameter and mAs Meter). If the mAs read is close to
cero, increase the filament number in big steps (a.e. increase values in
40). As the mAs read is close to the mAs displayed on the Console,
increase (or reduce) the filament number in smaller steps (a.e. increase
value in 30, 20, 10, ...).
Note . Press the “Store” button to store the new data (filament number)
before selecting the next kV or mA stations.
9. Select the next mA station at 80 kV. Before making any exposure, enter
as filament number the value calibrated for the previous mA station
increased in 10.
CA-1044RA 39
Mobile X-Ray Unit
Calibration
Press the “Read” button to read on the calibration panel the new value
of the Filament Current Number stored for each mA station at 80 kV.
Record the new values in the Data Book.
40 CA-1044RA
Mobile X-Ray Unit
Calibration
12. Check calibration at all break points (making exposures) and correct any
calibration points as needed.
14. Exit calibration mode. Select the mA station, Exposure Time and kV
value noted for each kV / mA combination not allowed (refer to step-6.).
Check calibration at these kV / mA combinations by making exposures.
If needed, enter in calibration mode and correct the Filament Current
Number of the respective kV / mA combination not allowed.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
17. If required, perform the calibration procedure for the second X-ray Tube.
CA-1044RA 41
Mobile X-Ray Unit
Calibration
This section describes the adjustments needed to calibrate the AEC according
to the customer input. Therefore, AEC exposures should be made during the
calibration process in order to insure AEC functionality.
AEC Film Density is controlled by the values stored in the respective memory
locations. These values are influenced by film speed, screen speed, dark room
procedures and customer requirements.
Note . For AEC calibration, use the same Film and Cassettes used by the
customer.
Illustration 2-1
AEC Function Waveform
EXP ORDER
STRT DR
Voltage Reference
( calibration data setting )
IC1
EXP Start
AEC STOP
(AEC CONTROL PCB)
EXP Stop
42 CA-1044RA
Mobile X-Ray Unit
Calibration
Make sure the automatic processor works correctly, and the concentration and
temperature of the solutions comply with manufacturer specifications.
Obtain a sensitometric curve to determine gamma (γ) of the film and the solution
quality. The procedure normally requires a sensitometer, but if it is not available
proceed as follows:
2. Develop and measure the Density (d) of each, d(f1) and d(f2).
d(f2) − d(f1)
γ= mAs(f2)
log 10 mAs(f1)
1. Set the detector gain potentiometer of the Ion Chamber to mid range
(refer to Ion Chamber documentation).
Note . Some Ion Chambers are not provided with detector gain and
balance potentiometers. Continue calibration process in Section
2.5.3 “Density Gain Adjustment”.
2. Set the following jumper on the AEC Control Board in position A: JP3 for
Board A3012--05.
4. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
6. Enter in calibration mode and verify that AEC Calibration number in E04,
Memory Location is 70. The useful range for AEC calibration numbers
is from 20 to 120. Then exit from calibration mode.
CA-1044RA 43
Mobile X-Ray Unit
Calibration
8. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary for that, change the Copper thickness or mA station and make
the exposure again.
10. Deselect “Left Area” and select “Right Area”. Make an exposure without
film in the cassette and record the Exposure Time. If this time is not equal
to the “Center Area” time (± 7%), adjust “Right Area” balance
potentiometer by increasing or decreasing Exposure Time. Repeat
process until both Exposure Times for “Center Area” and “Right Area” are
equal.
11. Check that the three scan Areas have equal Exposure Time (± 7%):
G “Center Area” with “Left Area”, “Center Area” with “Right Area”,
and “Left Area” with “Right Area”.
1. Set the detector gain potentiometer (or switches) of the Ion Chamber to
mid range (refer to Ion Chamber documentation).
2. Set the following jumper on the AEC Control Board in position A: JP3 for
Board A3012--05 or JP2 for Board A3012--06.
4. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
6. Enter in calibration mode and verify that AEC Calibration number in E04,
Memory Locations is 70. The useful range for AEC calibration numbers
is from 20 to 120. Then exit from calibration mode.
44 CA-1044RA
Mobile X-Ray Unit
Calibration
8. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary for that, change the Copper thickness or mA station and make
the exposure again.
10. To change the density, adjust the Master Gain (potentiometer) of the Ion
Chamber by increasing or decreasing Exposure Time. Repeat this
process until obtain the desired density (or customer preference).
When it is not possible to get a proper film density adjustment or for a fine
film density adjustment, change the AEC calibration number within a
range from 20 to 120. For that, enter in calibration mode set the new AEC
calibration number for the respecetive Memory Location (E04). The Film
Density increases when the calibration number is increased, and
decreases when the number is decreased. Then exit calibration mode.
11. Repeat the above steps for all the Ion Chambers installed with the
Generator.
1. Set the following jumper on the AEC Control Board in position B: JP3 for
Board A3012--05; to allow Density adjustment with the potentiometer
R10 for Board A3012--05 or R22 for Board A3012--06. Set potentiometer
to mid range.
CA-1044RA 45
Mobile X-Ray Unit
Calibration
3. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
5. Enter in calibration mode and verify that AEC Calibration number in E04,
Memory Locations is 70. The useful range for AEC calibration numbers
is from 20 to 120. Then exit from calibration mode.
7. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary for that, change the Copper thickness or mA station and make
the exposure again. Repeat the above steps for all the Ion Chambers.
9. Calculate the average value of the Ion Chamber densities. Adjust the
potentiometer R10 for Board A3012--05 in order to get a value of 1 for the
density of the Ion Chamber closest to the average value.
10. Repeat step-7 for each Ion Chamber and re-adjust the potentiometer
R10 for Board A3012--05 until all the Ion Chamber densities should be
approximately 0.8 to 1.2 (or the density customer preference).
11. Obtain the desired density by changing the respective AEC calibration
number within a range from 20 to 120. For that, enter in calibration mode
set the new AEC calibration number for the respecetive Memory Location
(E04). The Film Density increases when the calibration number is
increased, and decreases when the number is decreased. Repeat this
step for all the Ion Chambers. Then exit calibration mode.
46 CA-1044RA
Mobile X-Ray Unit
Calibration
2.5.4 kV COMPENSATION
Perform the following steps to determine whether or not AEC Tracking of film
density must be adjusted.
1. Enter in calibration mode and verify that AEC Tracking number in E08,
Memory Locations is “0”. Then exit from calibration mode and Service
mode.
3. Measure the film density obtained with those exposures and check that
the film variation range is the same ±0.2 of the optical density obtained
at 70 kV.
4. If the variation value is not ±0.2 calculate the new value for the AEC
Tracking number in each Memory Location in the following manner:
b. Values for AEC Tracking range is from --10 to +10. Determine the
correct value for the needed AEC Tracking change.
AEC-kV
TRACKING +10 +9 +8 +7 +6 +5 +4 +3 +2 +1 0 --1 --2 --3 --4 --5 --6 --7 --8 --9 --10
CURVE
VALUE TO ENTER
IN THE MEMORY 10 9 8 7 6 5 4 3 2 1 0 255 254 253 252 251 250 249 248 247 246
LOCATION
c. Enter the new value and store it by pressing the “AEC Reset”
push-button. Exit calibration mode and repeat this process until
the desired density compensation is obtained.
CA-1044RA 47
Mobile X-Ray Unit
Calibration
Illustration 2-2
AEC kV Tracking Curve
250
200
DAC OUTPUT in byte (8 Bit) (U3-2 of AEC Control Board)
150
100
70 10
8
50 6
4
2
0
--2
--4
0
--6
--8
--10
--50
kV
48 CA-1044RA
Mobile X-Ray Unit
Calibration
AEC is calibrated with “Density 0” selected (normal film density). Density can
be increased or decreased by selecting “ --2, --1, 0, +1, +2”.
The variation percentage of the density scale is factory set at 25%. This
variation can be set according to the customer preferences by changing the
value stored in E12 Memory Location.
The following table shows some examples for the variation percentage of the
density scale with reference to the value stored in E12 Memory Location.
CA-1044RA 49
Mobile X-Ray Unit
Calibration
Verify that all Configuration and Calibration data have been properly stored in
memory.
1. Enter in calibration mode and check that the values noted for the
“Filament Current Numbers” and “Extended Memory Locations” tables
of the Data Book are the same that the values displayed and stored in
memory.
3. Turn the Generator OFF and verify position of dip switches on the HT
Controller Board are:
4. Set the Test dip switch 3024SW2-3 on the ATP Console CPU Board in
“Off” position to place the Generator in normal operating mode.
50 CA-1044RA