TXR HF Cal
TXR HF Cal
TXR HF Cal
CA-1004R4
Calibration
HF Series Generators
HF Series Generators
Calibration
REVISION HISTORY
This Document is the english original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
CA-1004R4
HF Series Generators
Calibration
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 CALIBRATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 kV Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
CA-1004R4 i
HF Series Generators
Calibration
ii CA-1004R4
HF Series Generators
Calibration
SECTION 1 INTRODUCTION
CA-1004R4 1
HF Series Generators
Calibration
ACCURACY
PARAMETERS
(with 12 BITS HT Controller)
kV ± (3% + 1 kV)
kV ± (3% + 1 kV)
FLUORO mA ± 10%
The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray tube.
2 CA-1004R4
HF Series Generators
Calibration
Note . Test points on the HT Controller Board can also be used to monitor
the kV and mA readings but should not be used to calibrate the
unit. These test points must be checked with scope:
• Verify that dip switch 3024SW2-3 on the ATP Console CPU Board is in
“On” position to permit the service mode.
CA-1004R4 3
HF Series Generators
Calibration
3000SW2-4 Position during operation -- Digital mA Loop Closed Digital mA Loop Open / Filament Current Constant
Note . Only for Generators with Low Speed Starter (LF-RAC) (it does not
apply to Generators with High Speed Starter - LV-DRAC):
• Be sure that X-ray Tubes configured in E02 and E18 Memory Locations
correspond to X-ray Tubes installed (refer to Configuration document).
Note . For RAD and Fluoro Generators, E17 Memory Location (”Fluoro
Filament Setting”) sets maximum patient Entrance Skin Exposure
Dose Rate. (Refer to Section 2.7 “Fluoro Calibration”).
4 CA-1004R4
HF Series Generators
Calibration
The values stored in these Extended Memory Locations only affect to Exposure
Times for techniques below 20 ms. The Memory Locations which affect short
exposure times are:
MEMORY LOCATION
FUNCTION
TUBE-1 TUBE-2
The Generator has been optimized at the factory to produce correct exposures
at the lower times (<20 ms.) Therefore do not change the value factory set for
E13 and E29 Memory Locations and only adjust the value for E15 and E31
Memory Location according to the HV Cables length.
2. Select the E13 (or E29) Memory Location. Value in this Memory Location
adjusts the time delay of the exposure. It is factory set for a value of 17,
18 or 19 (default value is “18”). Only read this value, do not change it.
3. Select the E15 (or E31) Memory Location. Value in this Memory Location
is set in relation to the length in meters of one of the HV Cables
(1 ft = 0.3048 m). Verify the HV Cable length in meters and set the
following value:
4m 27
6m 31
9m 38
12 m 45
14 m 49
16 m 54
5. Exit from calibration mode and record the new values in the Data Book.
6. Repeat the above calibration process for the second tube (memory
locations E29 and E31).
CA-1004R4 5
HF Series Generators
Calibration
2.3 KV LOOP
Extended Memory Location E06 contains the calibration factor for kV Loop.
Each number above or below of the indicated in the E06 memory location
increases or decreases respectively the kV gain value.
3. Select the E06 Memory Location and read the calibration value by
pressing the “AEC Reset” push-button. Enter the value “200” and store
it by pressing the “AEC Reset” push-button.
6 CA-1004R4
HF Series Generators
Calibration
Select 80 kV, 200 mA and set value “344” on Filament Current Number.
Make an exposure. The mAs read on the mAs Meter must be the same
mAs displayed on the calibration screen with a tolerance of ±0.1 mAs
(tolerance of the parameter and mAs Meter).
If the mAs is low, increase the filament number. If the mA is high (or
“Generator Overload” Error is shown), decrease the filament number.
Press the “AEC Reset” push-button before making a new exposure.
Repeat until the mAs read is correct and the mA station is calibrated.
6. Select: 80 kV, 200 mA, 100 ms. Make an exposure and note the kV value
at the end of the exposure.
d. When it is correct, record the new value for E06 Memory Location
in the Data Book.
CA-1004R4 7
HF Series Generators
Calibration
G Remove the HV Cables from the X-ray Tube and connect them to
the HV Bleeder, then connect a short couple of HV Cables from the
HV Bleeder to the X-ray Tube.
Keep pressed the “kV increase” and “Power On” push-buttons again until
code “222” disappears on the Console to confirm Auto-calibration mode
or press only the “Power On” push-button to leave Auto-calibration mode
and return to manual calibration.
5. Read the kVp measured with the HV Bleeder and enter this value on the
kV Display by pressing the “kV Increase” or “kV Decrease” buttons.
7. Read the new value stored in E06 Memory Location pressing the “AEC
Reset” push-button. Record this value in the Data Book.
8 CA-1004R4
HF Series Generators
Calibration
Extended Memory Locations E03 and E05 contain the calibration factor for
Digital mA Loop Closed. Each number above or below the indicated in the
memory locations increases or decreases respectively the mA gain value.
Note . Values in E03 and E05 Memory Locations are only related to the
Generator performance (they are not related to the X-ray Tube(s)
installed), so values in these Memory Locations are factory
adjusted. Only perform this procedure if the HT Controller Board
and/or HV Transformer have been replaced in the unit.
CA-1004R4 9
HF Series Generators
Calibration
2. Enter Auto-calibration mode pressing the “Power On” and “kV increase”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increase” and “Power On” push-buttons again until
code “222” disappears on the Console to confirm Auto-calibration mode
or press only the “Power On” push-button to leave Auto-calibration mode
and return to manual calibration.
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
3.00 and 4.00 mAs) in the mAs Display pressing the “mAs increase” or
“mAs decrease” push-buttons.
6. Read the new value stored in E03 memory location pressing the
“AEC Reset” push-button. Record this value in the Data Book.
10 CA-1004R4
HF Series Generators
Calibration
2. Enter Auto-calibration mode pressing the “Power On” and “kV increase”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increase” and “Power On” push-buttons again until
code “222” disappears on the Console to confirm Auto-calibration mode
or press only the “Power On” push-button to leave Auto-calibration mode
and return to manual calibration.
e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.
4. Enter the mAs value read in the mAs Meter (it must be a value between
7.00 and 8.50 mAs) in the mAs Display pressing the “mAs increase” or
“mAs decrease” push-buttons.
6. Read the new value stored in E05 memory location pressing the
“AEC Reset” push-button. Record this value in the Data Book.
CA-1004R4 11
HF Series Generators
Calibration
3. Enter in Calibration mode. Select the E17 Memory Location and store the
value “086”. Press the “AEC Reset” push-button to read the value stored.
Exit calibration mode.
It is recommended to start with the Small Focal Spot (first group) and continue
with the Large Focal Spot (second group).
12 CA-1004R4
HF Series Generators
Calibration
Auto-calibration starts with the minimum available mA station for the selected
Focal Spot at 50 kV and follows with the other combinations of mA stations for
the selected Focal Spot at 80 kV, 120 kV and 40 kV.
1. Select one available workstation for the X-ray Tube selected. This
workstation has to be one of the previously configured as “Direct”.
2. Enter calibration mode and select one of the configured mA stations for
the Small Focal Spot.
3. Enter Auto-calibration mode pressing the “Power On” and “kV increase”
push-buttons. After releasing both push-buttons, code “222” is flashing
on the Console accompanied of an alarm waiting confirmation for
entering in Auto-calibration mode.
Keep pressed the “kV increase” and “Power On” push-buttons again until
code “222” disappears on the Console to confirm Auto-calibration mode
or press only the “Power On” push-button to leave Auto-calibration mode
and return to manual calibration.
4. Check that the Heat Units available for the X-ray Tube are 100%
(kV Display shows “H -- --” on the Console).
CA-1004R4 13
HF Series Generators
Calibration
If the Heat Units available for the X-ray Tube are less than 40%,
exposures are inhibited momentarily and code “111” will be flashed on the
Console accompanied of an alarm. The alarm will stop when the X-ray
Tube begins to cool and recovers the Heat Units capacity, exposures can
be performed again even though code “111” is shown on the Console.
At this point, it is recommended to wait until the Heat Units available are
closed to the 80% of the X-ray Tube capacity without making any
exposure.
14 CA-1004R4
HF Series Generators
Calibration
6. Repeat the same procedure for the Large Focal Spot selecting one of
the configured mA stations for the Large Focal Spot previous to enter in
Auto-calibration mode.
7. After performing both procedures (for Small and Large Focal Spots),
select in calibration mode each combination of the available mA stations
at the kV break points (40, 50, 80 and 120 kV). Press the “AEC Reset”
push-button to read on the kV Display the new value of the Filament
Current Number stored for each combination and record the new values
in the Data Book.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
10. Turn the Generator power OFF and set Dip Switch 3000SW2-4 on the
HT Controller Board in “Off” position (Digital mA Loop Closed).
This procedure describes the Manual calibration of all the Filament Current
Numbers. It has to be also used to calibrate the kV / mA combinations not
performed during Auto-calibration procedure because it has not been
completed. These combinations have the Filament Current Number set to
“344”, so only these combinations have to be manually calibrated as described
in this procedure.
Although the suggested values (Table 2-1) could change depending on the
X-ray tube used, entering those values will approximate accurate calibration
without making excessive exposures.
CA-1004R4 15
HF Series Generators
Calibration
Table 2-1
mA Calibration Numbers Change
Note.-- The mA station values depends on the Generator model. Some models do not contain all the mA stations listed above.
16 CA-1004R4
HF Series Generators
Calibration
The kV Display shows again the kV when the “AEC Reset” push-button is
released. Note that in calibration mode, only the mA stations and kV (at the
break points) can be selected.
2. Turn the Generator ON and select one available workstation for the X-ray
Tube selected. This workstation has to be one of the previously
configured as “Direct”.
3. Check that the Heat Units available for the X-ray Tube are 100%
(kV Display shows “H -- --” on the Console).
CA-1004R4 17
HF Series Generators
Calibration
Exit calibration mode and select the mA station and Exposure Time of
each kV / mA combination not allowed. Increase or decrease the kV
value as required to determine the kV value allowed nearest to the kV
value of the combination. Note the kV value allowed in the respective cell
of Table 2-1.
6. Make an exposure. The mAs read on the mAs Meter must be the mAs
displayed on the Console ± 0.1 mAs (tolerance of the parameter and
mAs Meter). If the mAs read is close to zero, increase the filament
number in big steps (a.e. increase values in 40). As the mAs read is close
to the mAs displayed on the Console, increase (or reduce) the filament
number in smaller steps (a.e. increase value in 30, 20, 10, ...).
Note . Press the “AEC Reset” push-button to store the new data (filament
number) before selecting the next kV or mA stations.
18 CA-1004R4
HF Series Generators
Calibration
7. Select the next mA station at 80 kV. Before making any exposure, enter
as filament number the value calibrated for the previous mA station
increased in 10.
Press the “AEC Reset” push-button to read on the kV Display the new
value of the Filament Current Number stored for each mA station at
80 kV. Record the new values in the Data Book.
CA-1004R4 19
HF Series Generators
Calibration
13. Select the mA station, Exposure Time and kV value noted for each
kV / mA combination not allowed (refer to step-4.). Check calibration at
these kV / mA combinations by making exposures. If needed, enter in
calibration mode and correct the Filament Current Number of the
respective kV / mA combination not allowed.
Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.
16. Repeat the above calibration process for the second tube.
20 CA-1004R4
HF Series Generators
Calibration
This section describes the adjustments needed to calibrate the AEC according
to the customer input. Therefore, AEC exposures will be made during the
calibration process in order to insure AEC functionality.
The Optical Density is controlled by the values stored in the respective memory
locations. These values are influenced by film speed, screen speed, dark room
procedures and customer requirements.
Note . For AEC calibration, use the same Film and Cassettes used by the
customer. AEC calibration must be performed using the Medium
Film/Screen speed combination. The Medium Film/Screen speed
has to be double of the Slow and half of the Fast (a.e. 200--Slow,
400--Medium, 800--Fast).
The following table indicates the Memory Locations related to AEC Calibration.
CA-1004R4 21
HF Series Generators
Calibration
Illustration 2-1
Automatic Exposure Control
DAC
Jumper
COMPARATOR
Adjustment V ref
(Potentiomer) Reference
V out
AEC STOP
to cut the exposure
Ramp
ATP Console CPU (from Ion Chamber)
Data Bus
Microprocessor
EXP ORDER
STRT DR
Make sure the automatic processor works correctly, and the concentration and
temperature of the solutions comply with manufacturer specifications.
Obtain a sensitometric curve to determine gamma (γ) of the film and the solution
quality. The procedure normally requires a sensitometer, but if it is not available
proceed as follows:
2. Develop and measure the Density (d) of each, d(f1) and d(f2).
22 CA-1004R4
HF Series Generators
Calibration
Note . Some Ion Chambers does not provide balance potentiometers for
Three Field Detectors. Continue calibration process in Section
2.6.3 “Optical Density Adjustment”.
1. Set the Master Gain potentiometer of the Ion Chambers to mid range
(refer to Ion Chamber documentation).
2. Set the following jumper on the AEC Control Board in position A: JP3 for
Board A3012--x1/x2/x5 or JP2 for Board A3012--x6/x7/x9.
3. Set SID at the Focal Distance of the Grid installed in the Table Bucky
(usually 100 cm) or in the Vertical Bucky Stand (usually 150 cm).
4. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
6. Enter in calibration mode and verify that AEC Calibration number in E04,
E09, E20 and E23 Memory Locations is 70. The range for AEC
calibration numbers are from 20 to 120. Then exit from calibration mode.
G RAD: 70 kV, 200 mA Large Focus (or the first mA station for Large
Focus if 200 mA station is set for Small Focus) and 1 second
back-up time.
8. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary, change the Copper thickness or mA station and make the
exposure again. (Last exposure parameters are displayed by pressing
the “AEC Reset” button of the Console.)
CA-1004R4 23
HF Series Generators
Calibration
10. Deselect “Left Area” and select “Right Area”. Make an exposure without
film in the cassette and record the Exposure Time. If this time is not equal
to the “Center Area” time (± 7%), adjust “Right Area” balance
potentiometer to increase or decrease Exposure Time. Repeat process
until both Exposure Times for “Center Area” and “Right Area” are equal.
11. Check that the three scan Areas have equal Exposure Time (± 7%):
G “Center Area” with “Left Area”, “Center Area” with “Right Area”,
and “Left Area” with “Right Area”.
Note . The Film Optical Density must be measured always on the same
point for all the X-ray Films developed during this procedure.
Illustration 2-2
How to measure the Film Optical Density
A K
USEFUL BEAM
DENSITOMETER FILM
24 CA-1004R4
HF Series Generators
Calibration
1. Set the Detector Gain switches (or gain potentiometer) of the Ion
Chamber to mid range (refer to Ion Chamber documentation). (If the Ion
Chamber is provided with the Generator, this adjustment is Factory set.)
The following tables indicate the switch position for the “Vacutec” Ion
Chamber (Factory set).
Positive ON
OUTPUT SIGNAL
Negative OFF
Positive or Negative polarity of the ramp signal is selected with a switch at the Ramp Module.
OUTPUT SIGNAL The Ramp Module is a 9-pin Sub-D connector plugged to the Ion Chamber cable.
Positive polarity of the ramp signal is factory set.
NORMAL FACTORY SELECTION: 1 V / µGy (10 V≃ 10 µGy) - Positive OFF OFF ON OFF
CA-1004R4 25
HF Series Generators
Calibration
2. Set the following jumper on the AEC Control Board in position A: JP3 for
Board A3012--x1/x2/x5 or JP2 for Board A3012--x6/x7/x9.
3. Set SID at the Focal Distance of the Grid installed in the Table Bucky
(usually 100 cm) or in the Vertical Bucky Stand (usually 150 cm).
4. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
6. Enter in calibration mode and verify that AEC Calibration number in E04,
E09, E20 and E23 Memory Locations is 70. The useful range for AEC
calibration numbers is from 20 to 120. Then exit from calibration mode.
G RAD: 70 kV, 200 mA Large Focus (or the first mA station for Large
Focus if 200 mA station is set for Small Focus) and 1 second
back-up time.
8. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary, change the Copper thickness (or if needed change the mA
station) and make the exposure again. (Last exposure parameters are
displayed by pressing the “AEC Reset” button of the Console.)
26 CA-1004R4
HF Series Generators
Calibration
Note . The AEC accuracy is better in all the useful range as the AEC
calibration number is closer to 70.
11. Repeat the above steps for all the Ion Chambers installed with the
Generator.
CA-1004R4 27
HF Series Generators
Calibration
1. Set the following jumper on the AEC Control Board in position B: JP3 for
Board A3012--x1/x2/x5 or JP2 for Board A3012--x6/x7/x9.
2. Set to mid range, the following potentiometer on the AEC Control Board:
R10 for Board A3012--x1/x2/x5 or R22 for Board A3012--x6/x7/x9.
Note . The density adjustment of ALL the Ion Chambers in the system are
affected by this potentiometer.
3. Set SID at the Focal Distance of the Grid installed in the Table Bucky
(usually 100 cm) or in the Vertical Bucky Stand (usually 150 cm).
4. Collimate the X-ray beam so that it completely covers all three fields but
does not extend beyond limits of the phantom.
6. Enter in calibration mode and verify that AEC Calibration number in E04,
E09, E20 and E23 Memory Locations is 70. The useful range for AEC
calibration numbers is from 20 to 120. Then exit from calibration mode.
G RAD: 70 kV, 200 mA Large Focus (or the first mA station for Large
Focus if 200 mA station is set for Small Focus) and 1 second
back-up time.
8. Make an exposure without film in the cassette and note the Exposure
Time displayed on the Console, it should be approximately 100 ms. If
necessary for that, change the Copper thickness (or if needed change
the mA station) and make the exposure again. (Last exposure
parameters are displayed by pressing the “AEC Reset” button of the
Console.) Take note of the final Copper thickness and mA station
selected.
28 CA-1004R4
HF Series Generators
Calibration
10. Insert a cassette with the Medium Film/Screen combination used by the
customer. Make an exposure with each Ion Chamber (use the final
Copper thickness and the selected mA station noted before), develop the
film and take note of the Optical Density measured for each Ion Chamber.
11. Take the maximum and minimum value of the O. Densities measured
before, sum them and divide the result by two. Subtract this value from
density 1.0 (or the customer preference O. Density) to obtain the O.
Density difference.
O.D. max + O.D. min
O.D. difference = O.D. customer preference --
2
Calculate the new O. Density by adding the O. D. difference (with its sign
+ or --) to the value of the O. Density measured for the Ion Chamber closer
to 1.0 (or the customer preference O. Density).
For this Ion Chamber, adjust the potentiometer (turning it clockwise will
decrease the O. Density) R10 for Board A3012--x1/x2/x5 or R22 for
Board A3012--x6/x7/x9 (at least one turn). Make another exposure,
develop the film and check the Optical Density. Repeat this action
(adjusting the potentiometer and making exposures) until the new
Optical Density (calculated before) for this Ion Chamber is obtained.
12. The Optical Density for each Ion Chamber has to be adjusted to 1.0 (or
the customer preference O. Density).
Enter in calibration mode, select the AEC calibration number for the
respective Memory Location (E04, E09, E20 or E23). Change the AEC
calibration number: the Optical Density increases / decreases when the
calibration number is increased / decreased. Then exit calibration mode.
Make another exposure, develop the film and check the Optical Density.
Repeat this action (changing the respective Memory Location and
making exposures) until the desired Optical Density is obtained.
Note . The AEC accuracy is better in all the useful range as the AEC
calibration number is closer to 70.
13. Repeat step-12. until the desired Optical Density for each Ion Chamber
is obtained.
14. Record all the values for the Memory Locations in the Data Book.
CA-1004R4 29
HF Series Generators
Calibration
2.6.4 kV COMPENSATION
Perform the following steps to determine whether or not AEC Tracking of Optical
Density must be adjusted.
1. Enter in calibration mode and verify that AEC Tracking number in E08,
E10 and E24 Memory Locations is “0”. Then exit from calibration mode
and Service mode.
G RAD: 55 kV, 200 mA Large Focus (or the first mA station for Large
Focus if 200 mA station is set for Small Focus) and 1 second
back-up time.
3. Make an exposure without film in the cassette and check that the
Exposure Time is lower than 1 second. If necessary for that, change the
Copper thickness (or if needed change the mA station) and make the
exposure again. Take note of the final Copper thickness and mA station
selected for 55 kV.
4. Select 90 kV. Make an exposure without film in the cassette and check
that the Exposure Time is higher than 20 ms. If necessary for that, change
the Copper thickness (or if needed change the mA station) and make the
exposure again. Take note of the final Copper thickness and mA station
selected for 90 kV.
5. Select 110 kV. Make an exposure without film in the cassette and check
that the Exposure Time is higher than 20 ms. If necessary for that, change
the Copper thickness (or if needed change the mA station) and make the
exposure again. Take note of the final Copper thickness and mA station
selected for 110 kV.
30 CA-1004R4
HF Series Generators
Calibration
7. If the variation value is not ±0.2 of the Optical Density (±20% of Image
Gray Level / Dose Level with CR or DR) calculate the new value for the
AEC Tracking number in each Memory Location in the following manner:
b. Values for AEC Tracking range is from --10 to +10. Determine the
correct value for the needed AEC Tracking change.
AEC-kV
TRACKING +10 +9 +8 +7 +6 +5 +4 +3 +2 +1 0 --1 --2 --3 --4 --5 --6 --7 --8 --9 --10
CURVE
VALUE TO ENTER
IN THE MEMORY 10 9 8 7 6 5 4 3 2 1 0 255 254 253 252 251 250 249 248 247 246
LOCATION
c. Enter the new value and store it by pressing the “AEC Reset”
push-button. Exit calibration mode and repeat this process from
step-6. until the desired density compensation is obtained.
CA-1004R4 31
HF Series Generators
Calibration
9. Record all the values for the Memory Locations in the Data Book.
10. Repeat the above steps for all the Ion Chambers installed with the
Generator.
11. Record all the values for the Memory Locations in the Data Book.
Illustration 2-3
AEC kV Tracking Curve
250
200
DAC OUTPUT in byte (8 Bit) (U3-2 of AEC Control Board)
150
100
70 10
8
50 6
4
2
0
--2
--4
0
--6
--8
--10
--50
40 50 60 70 80 90 100 110 120 130 140 150
kV
32 CA-1004R4
HF Series Generators
Calibration
1. Enter in calibration mode. Select E23 Memory Location and set a value
of “127”. Then exit from calibration mode and Service mode.
G RAD: 70 kV, 200 mA Large Focus (or the first mA station for Large
Focus if 200 mA station is set for Small Focus) and 1 second
back-up time.
CA-1004R4 33
HF Series Generators
Calibration
Note . If the photo tube voltage required is too low, decrease the AEC
calibration number in the E23 Memory Location and repeat the
process.
34 CA-1004R4
HF Series Generators
Calibration
The variation percentage of the density scale is factory set at 25%. This
variation can be set according to the customer preferences by changing the
value stored in E12 Memory Location. This value applies to both tubes.
Depending on the Console model, the range of the scale is “from --2 to +2” or
“from --4 to +4” (optional ).
The following tables show some examples for the variation percentage of the
density scale with reference to the value stored in E12 Memory Location.
DENSITY SCALE FROM --2 TO +2 . VARIATION OVER NORMAL OPTICAL DENSITY (N)
E12 VALUE
--2 --1 0 +1 +2
NOTE: If the value stored in E12 Memory Location is 0, 25 or 255, the variation percentage is 25%.
DENSITY SCALE FROM --4 TO +4 (OPTIONAL) . VARIATION OVER NORMAL OPTICAL DENSITY (N)
E12
VALUE --4 --3 --2 --1 0 +1 +2 +3 +4
NOTE: With scale from “--4 to +4” the useful range for the value stored in E12 Memory Location is from 1 to 25 .
Record the value for E12 Memory Location in the Data Book.
CA-1004R4 35
HF Series Generators
Calibration
This generator uses Pulsed Fluoro at fixed or variable rate. This technique is
a series of short exposures at the TV frame rate (fixed rate) or at the selected
PPS (variable rate).
• E17 is used to store data that controls the Maximum Skin Dose Radiation
at the following maximum levels (Regulation limits) (1 Rad = 8.7 mGy).
Note that in practice, the rejection limits for entrance exposure rate must
be somewhat less than the maximum specified due to Dosimeter
calibration accuracy. (Refer to Table 2-2.)
Table 2-2
Rejection Limits Based on Meter Calibration Accuracy
REJECTIONS LIMITS
METER CALIBRATION
ACCURACY FOR 5 R/min (43.5 mGy/min) MAXIMUM FOR 10 R/min (87 mGy/min) MAXIMUM
36 CA-1004R4
HF Series Generators
Calibration
• E19 is used to set maximum Fluoro kV. This value is determined by the
type of TV camera and type of images desired. For general fluoroscopic
use with a conventional TV system, 120 kV is recommended. A lower
maximum setting will produce more contrast on the TV system but less
penetration for large patients.
• E25, E26, E27 are used to calibrate the Fluoro mA Display. These values
will be shown on the Fluoro mA Display during Fluoro exposures
whenever maximum PPS are selected.
Note . The Fluoro mA Display values entered into the Extended Memory
Locations E25, E26 and E27 are also used to calculate Heat Units.
It is important to enter accurate values.
Make sure that the Small Filament of the X-ray tube is properly
warmed-up (at less 15 minutes).
4. Select E19 Memory Location and set the maximum Fluoro kV at120 kV
or more if it is possible.
CA-1004R4 37
HF Series Generators
Calibration
Illustration 2-4
Fluoro Entrance Exposure Rate Test Set-up
DIAGNOSTIC SOURCE
ASSEMBLY
PROBE
SET AS CLOSE
AS POSSIBLE
USEFUL BEAM
30 cm
(12”)
IMAGE SYSTEM
OVERTABLE SOURCE
IMAGE SYSTEM
SPOT FILM DEVICE
PROBE
TABLE TOP
DIAGNOSTIC
SOURCE
ASSEMBLY
UNDERTABLE SOURCE
38 CA-1004R4
HF Series Generators
Calibration
6. Select the maximum PPS and Non-ABC mode. Make a Fluoro exposure
at maximum kV that will be used in the system and measure the dose
applied, it should not be over the Rejection Limits for 10 R/min (87
mGy/min) or 5 R/min (43.5 mGy/min) (refer to Table 2-2.).
In case that the value is not acquired, change the Fluoro Filament Setting
stored in E17 as required until it is correctly calibrated to that dose. Be
in mind that radiation increases or decreases in accordance to value in
E17 is increased or decreased.
7. The mA values displayed during Fluoro exposures are stored in E25, E26
and E27 Memory Locations. The method used to obtain the Fluoro mA
values is to measure the average mA using a mA meter in Fluoro.
Select the maximum PPS and Non-ABC mode. Make the following
Fluoro exposures:
10. Select E19 Memory Location if it is required to reduce the value for the
maximum Fluoro kV in the installation.
11. Select the E25, E26 and E27 Memory Locations and store the respective
mA values noted before multiplied by 10 (a.e. if the mA value noted is
“3.2”, store the value “32” in the respective Memory Location).
CA-1004R4 39
HF Series Generators
Calibration
The closed-loop ABC System can monitor the Image Tube Output Brightness
through two ways: Photomultiplier Tube or TV Camera.
Brightness error at the comparator output is sent as “kV Up” and “kV Down” to
the Generator where is used to drive the Fluoro kV control. The closed-loop
operation requires more or less brightness thru “kV Up” and “kV Down” demand
signals respectively. Patient Entrance Dose is automatically varied so that
optimum constant Image Tube Output Brightness is maintained.
40 CA-1004R4
HF Series Generators
Calibration
Illustration 2-5
ABC System for Photomultiplier Tube
PT SPLY
(Photomultiplier Voltage supply)
MONITOR
VIDEO
U12
R13 U2
TV
I/V
PHOTOMULTIPLIER
CONVERTER
LIGHT
PEAK SAMPLE
U10 DETECT. &
PHOTOMULTIPLIER
BOARD (A3168--xx)
PT INPUT HOLD U8
(Brightness signal)
AMPLIFIER
KV UP & KV DWN
PATIENT
X-RAY TABLE
HV HT CONTROL PCB INTERFACE
TRANSFORMER CONTROL PCB
X-RAY TUBE
POWER MODULE
CA-1004R4 41
HF Series Generators
Calibration
Window reference could be adjusted first to set mid-way the brightness level
(+5 VDC), and second to increase or decrease the range of response and
sensitivity of the kV control to input variations (brightness changes). If oscillation
occurs during ABC fluoro exposure, increase the dead zone by adjusting
Resistors R11 and R14 on the AEC Control Board. (Refer to Illustration 2-6 for
ABC waveforms).
Illustration 2-6
ABC Waveforms in AEC Control Board
Brightness level
(Gain adj.)
ABC IN
(U8--8)
UP PT reference (R14)
High Logic
--kV Down (J1--19) Low
Logic
Command to drive
the fluoro kVp DOWN
--kV UP (J1--6)
Command to drive
the fluoro kVp UP
1. Be sure that the Video System and the Image Intensifier are powered and
operating correctly.
42 CA-1004R4
HF Series Generators
Calibration
6. Calculate the value of the optimum radiation (that will give optimum
brightness) usually is 2μR/frame at 9” FOV (for more information refer to
Image System documentation).
Examples:
For 25 frame/second optimum radiation is 3 mR/min.
2 μR/frame x 25 frame/s = 50 μR/s.
50 μR/s x 60 s/min = 3000 μR/min = 3 mR/min.
For 30 frame/second optimum radiation is 3.6 mR/min.
2 μR/frame x 30 frame/s = 60 μR/s.
60 μR/s x 60 s/min = 3600 μR/min = 3.6 mR/min.
CA-1004R4 43
HF Series Generators
Calibration
10. Make a Fluoro exposure and check that the kV does not change.
11. Stop the Fluoro exposure and select 40 kV. Make a Fluoro exposure and
check that the kV value goes to 70 kV (or the kV obtained in step-7.)
±2 kV without System problems.
12. Stop the Fluoro exposure and select 100 kV. Make a Fluoro exposure
and check that the kV value goes to 70 kV (or the kV obtained in step-7.)
±2 kV without System problems.
44 CA-1004R4
HF Series Generators
Calibration
The ABC can be performed with an Analogic Signal Output (ABC OUT)
proportional to the brightness or with the Composite Video Signal Output from
the TV Camera.
The “ABC OUT” signal from the TV Camera is compatible with the Generator
when the range is from 0 VDC (dark image) to 10 VDC (bright image) and the
Optimum Brightness is achieved at around 6 VDC.
When a TV Camera without a direct “ABC OUT” signal is used, the Composite
Video signal (which amplitude is proportional to the image tube output
brightness) is sent to an RF Adaptation Board where it is transformed into an
“ABC OUT” analogic signal. In this case, the brightness level is taken from a
rectangular window from the center of the raster. (Refer to Illustration 2-7).
This analogic signal in the AEC Control Board (“PT Input” in Board
A3012--x1/x2/x5 and “ABC OUT” in Board A3012--x6/x7/x9) is filtered, and
compared to a window reference. Brightness error at the comparator output is
sent as “kV Up” and “kV Down” to the Generator where is used to drive the fluoro
kV control. The closed-loop operation requires more or less brightness thru “kV
Up” and “kV Down” demand signals respectively. Patient entrance dose is
automatically varied so that constant image tube output brightness is
maintained.
Illustration 2-7
ABC System for TV Camera
MONITOR
CAMERA
TV
R5 -- C11 U8
WINDOW ADJ.
R11 -- R14
KV UP & KV DWN
PATIENT
X-RAY TABLE
HV HT CONTROL PCB
TRANSFORMER
X-RAY TUBE
POWER MODULE
CA-1004R4 45
HF Series Generators
Calibration
The optimum brightness level in ABC mode is set by adjusting the gain at R27
on the RF Adaptation Board. The “ABC OUT” signal requires an input range
from 0 to +10 VDC (the stabilized value will be between 5 and 7 VDC).
Window reference could be adjusted first to set mid-way the brightness level (+6
VDC), and second to increase or decrease the range of response and sensitivity
of the kV control to input variations (brightness changes). If oscillation occurs
during ABC fluoro exposure, increase the dead zone by adjusting R11 and R14
on the AEC Control Board. (Refer to Illustration 2-8 for ABC waveforms)
Illustration 2-8
ABC Waveforms in AEC Control Board
Brightness level
(Gain adj.)
ABC IN
(U8--8)
UP PT reference (R14)
High Logic
--kV Down (J1--19) Low
Logic
Command to drive
the fluoro kVp DOWN
--kV UP (J1--6)
Command to drive
the fluoro kVp UP
46 CA-1004R4
HF Series Generators
Calibration
2.8.2.1 ABC SYSTEM ADJUSTMENT WITH ABC SIGNAL FROM TV CAMERA COMPATIBLE WITH
THE GENERATOR
1. Be sure that the Video System and the Image Intensifier are powered and
operating correctly.
7. Calculate the value of the optimum radiation (that will give optimum
brightness) usually is 2μR/frame at 9” FOV (for more information refer to
Image System documentation).
Examples:
CA-1004R4 47
HF Series Generators
Calibration
10. Adjust the TV Camera to obtain +6 VDC (or the voltage supplied by the
TV Camera as optimum brightness) on TP3 (ABC IN) on the AEC Control
Board (Refer the TV Camera manuals).
12. Select the 70 kV (or the kV obtained in step-8.) and ABC mode.
13. Make a Fluoro exposure and check that the kV does not change.
14. Stop the Fluoro exposure and select 40 kV. Make a Fluoro exposure and
check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±2 kV without System problems.
15. Stop the Fluoro exposure and select 100 kV. Make a Fluoro exposure
and check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±2 kV without System problems.
48 CA-1004R4
HF Series Generators
Calibration
2.8.2.2 ABC SYSTEM ADJUSTMENT WITH ABC SIGNAL FROM TV CAMERA NOT COMPATIBLE
WITH THE GENERATOR
1. Be sure that the Video System and the Image Intensifier are powered and
operating correctly.
4. Set on the RF Adaptation Board, Jumpers JP21 in position “A” and JP20
in position “B” (for negative ABC signal) or in position “C” (for positive
ABC signal).
7. Calculate the value of the optimum radiation (that will give optimum
brightness) usually is 2μR/frame at 9” FOV (for more information refer to
Image System documentation).
Examples:
CA-1004R4 49
HF Series Generators
Calibration
10. Select the 70 kV (or the kV obtained in step-8.) and Non-ABC mode.
11. Adjust R29 (OFFSET) on the RF Adaptation Board to have 0 VDC in TP7.
12. Adjust gain at R27 (Gain) on the RF Adaptation Board to make the “ABC
OUT” signal equal to +6 VDC. Measure “ABC OUT” in TP-7 of RF
Adaptation Board or in TP3 (”ABC IN”) of the AEC Control Board.
14. Select the 70 kV (or the kV obtained in step-8.) and ABC mode.
15. Make a Fluoro exposure and check that the kV does not change.
16. Stop the Fluoro exposure and select 40 kV. Make a Fluoro exposure and
check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±2 kV without System problems.
17. Stop the Fluoro exposure and select 100 kV. Make a Fluoro exposure
and check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±2 kV without System problems.
50 CA-1004R4
HF Series Generators
Calibration
1. Be sure that the Video System and the Image Intensifier are powered and
operating correctly.
4. Set on the RF Adaptation Board, Jumpers JP21 in position “A” and JP20
in position “A”.
7. Calculate the value of the optimum radiation (that will give optimum
brightness) usually is 2μR/frame at 9” FOV (for more information refer to
Image System documentation).
Examples:
CA-1004R4 51
HF Series Generators
Calibration
9. Perform the adjustment for the window test. This window defines the
area of the image where the brightness will be captured for the ABC.
a. Calculate the image area πr2 (clear circle on the monitor). (r=circle
radius).
10. Select the 70 kV (or the kV obtained in step-8.) and Non-ABC mode.
11. Adjust the gain at R27 on the RF Adaptation Board to make the “ABC
OUT” signal equal to +6 VDC. Measure “ABC OUT” in TP7 of RF
Adaptation Board or in TP3 (”ABC IN”) of the AEC Control Board.
52 CA-1004R4
HF Series Generators
Calibration
13. Select the 70 kV (or the kV obtained in step-8.) and ABC mode.
14. Make a Fluoro exposure and check that the kV does not change.
15. Stop the Fluoro exposure and select 40 kV. Make a Fluoro exposure and
check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±4 kV without System problems.
16. Stop the Fluoro exposure and select 100 kV. Make a Fluoro exposure
and check that the kV value goes to 70 kV (or the kV obtained in step-8.)
±4 kV without System problems.
CA-1004R4 53
HF Series Generators
Calibration
Verify that all Configuration and Calibration data have been properly stored in
memory.
1. Enter in calibration mode and check that the values noted for the
“Filament Current Numbers” and “Extended Memory Locations” tables
of the Data Book are the same that the values displayed and stored in
memory. Press the “AEC Reset” button to read the stored values.
3. Turn the Generator OFF and verify position of dip switches on the HT
Controller Board are:
4. Set the Test dip switch 3024SW2-3 on the ATP Console CPU Board in
“Off” position to place the Generator in normal operating mode.
54 CA-1004R4