Im01c22d01-01en (17) - Eja310, Eja430
Im01c22d01-01en (17) - Eja310, Eja430
Im01c22d01-01en (17) - Eja310, Eja430
IM 01C22D01-01E
17th Edition
CONTENTS
CONTENTS
ii IM 01C22D01-01E
CONTENTS
REVISION RECORD
iii IM 01C22D01-01E
1. INTRODUCTION
1. INTRODUCTION
Thank you for purchasing the DPharp electronic • The following safety symbol marks are used in this
pressure transmitter. manual:
The DPharp Pressure Transmitters are precisely
calibrated at the factory before shipment. To ensure WARNING
correct and efficient use of the instrument, please read
this manual thoroughly and fully understand how to Indicates a potentially hazardous situation which,
operate the instrument before operating it. if not avoided, could result in death or serious
injury.
Regarding This Manual
• This manual should be passed on to the end user.
• The contents of this manual are subject to change
CAUTION
without prior notice.
Indicates a potentially hazardous situation which,
• All rights reserved. No part of this manual may be
if not avoided, may result in minor or moderate
reproduced in any form without Yokogawa’s written
injury. It may also be used to alert against
permission.
unsafe practices.
• Yokogawa makes no warranty of any kind with
regard to this manual, including, but not limited to,
implied warranty of merchantability and fitness for a
particular purpose. IMPORTANT
• If any question arises or errors are found, or if any Indicates that operating the hardware or software
information is missing from this manual, please in this manner may damage it or lead to system
inform the nearest Yokogawa sales office. failure.
• The specifications covered by this manual are
limited to those for the standard type under the
specified model number break-down and do not
cover custom-made instruments. NOTE
• Please note that changes in the specifications, Draws attention to information essential for
construction, or component parts of the instrument understanding the operation and features.
may not immediately be reflected in this manual at
the time of change, provided that postponement of
revisions will not cause difficulty to the user from a Direct current
functional or performance standpoint.
• Yokogawa assumes no responsibilities for this 1.1 For Safe Use of Product
product except as stated in the warranty.
For the protection and safety of the operator and the
• If the customer or any third party is harmed by the instrument or the system including the instrument,
use of this product, Yokogawa assumes no responsi- please be sure to follow the instructions on safety
bility for any such harm owing to any defects in the described in this manual when handling this instru-
product which were not predictable, or for any ment. In case the instrument is handled in contradiction
indirect damages. to these instructions, Yokogawa does not guarantee
safety. Please give your attention to the followings.
1-1 IM 01C22D01-01E
1. INTRODUCTION
1-2 IM 01C22D01-01E
1. INTRODUCTION
1-3 IM 01C22D01-01E
1. INTRODUCTION
SK PL
CZ
SLO
LT H
BG
LV
EST RO
1-4 IM 01C22D01-01E
2. HANDLING CAUTIONS
2. HANDLING CAUTIONS
This chapter describes important cautions regarding
how to handle the transmitter. Read carefully before
using the transmitter.
The EJA Series pressure transmitters are thoroughly : Refer to USER'S MANUAL
transmitter is delivered, visually check them to make Figure 2.2 Name Plate Example of TIIS Flameproof Type
sure that no damage occurred during shipment.
Also check that all transmitter mounting hardware 2.2 Unpacking
shown in Figure 2.1 is included. If the transmitter was When moving the transmitter to the installation site,
ordered without the mounting bracket or without the keep it in its original packaging. Then, unpack the
process connector, the transmitter mounting hardware transmitter there to avoid damage on the way.
is not included. After checking the transmitter, repack
it in the way it was delivered until installation.
2.3 Storage
Bolt
Process connector The following precautions must be observed when
Process connector storing the instrument, especially for a long period.
Gasket
U-bolt (a) Select a storage area which meets the following
conditions:
• It is not exposed to rain or water.
• It suffers minimum vibration and shock.
• It has an ambient temperature and relative
humidity within the following ranges.
Mounting bracket Ambient temperature:
(L type)
–40 to 85°C without integral indicator
U-bolt nut Transmitter mounting bolt
–30 to 80°C with integral indicator
Relative humidity:
5% to 100% R.H. (at 40°C)
Preferred temperature and humidity:
approx. 25°C and 65% R.H.
(b) When storing the transmitter, repack it as nearly
as possible to the way it was packed when
delivered from the factory.
(c) If storing a transmitter that has been used,
Mounting bracket thoroughly clean the chambers inside the cover
(Flat type)
F0201.EPS flanges, so that no measured fluid remains in it.
Figure 2.1 Transmitter Mounting Hardware Also make sure before storing that the pressure-
detector and transmitter section are securely
2.1 Model and Specifications mounted.
Check
2.4 Selecting the Installation
The model name and specifications are indicated on the
name plate attached to the case. If the reverse operat- Location
ing mode was ordered (reverse signal), ‘REVERSE’ The transmitter is designed to withstand severe
will be inscribed in field *1. environmental conditions. However, to ensure stable
and accurate operation for years, observe the follow-
ing precautions when selecting an installation
location.
2-1 IM 01C22D01-01E
2. HANDLING CAUTIONS
2-2 IM 01C22D01-01E
2. HANDLING CAUTIONS
2-3 IM 01C22D01-01E
2. HANDLING CAUTIONS
2-4 IM 01C22D01-01E
2. HANDLING CAUTIONS
P.1-2.”) [Nonincendive]
Note 1. Model EJA Series differential, gauge, and Hazardous Location Nonhazardous Location
absolute pressure transmitters with Class I, II, Division 2,
optional code /CS1 are applicable for use Groups A, B, C, D, E, F, G
in hazardous locations Class III General
EJA Series Pressure Purpose
Certificate: 1053843 Equipment
Transmitters
• Applicable Standard: C22.2 No.0, No.0.4, No.25, + +
No.30, No.94, No.142, No.157, No.213 Supply – –
• Intrinsically Safe for Class I, Division 1, Groups A, Not Use
B, C & D. Class II, Division 1, Groups E, F & G Safety Barrier
F0204-2.EPS
2-5 IM 01C22D01-01E
2. HANDLING CAUTIONS
2-6 IM 01C22D01-01E
2. HANDLING CAUTIONS
2-7 IM 01C22D01-01E
2. HANDLING CAUTIONS
optional code /KS2 for potentially explo- *1: In any safety barriers used the output current must be limited
by a resistor “R” such that Imaxout-Uz/R.
sive atmospheres:
• No. KEMA 02ATEX1030 X
• Applicable Standard: EN50014:1997, b. CENELEC ATEX (KEMA) Flameproof Type
EN50020:1994, EN50284:1999 Caution for CENELEC ATEX (KEMA) flameproof
• Type of Protection and Marking code: type.
EEx ia IIC T4 Note 1. Model EJA Series differential, gauge, and
• Temperature Class: T4 absolute pressure transmitters with
• Enclosure: IP67 optional code /KF2 for potentially explo-
• Process Temperature: 120°C max. sive atmospheres:
• Ambient Temperature: –40 to 60°C • No. KEMA 02ATEX2148
Note 2. Electrical Data • Applicable Standard: EN50014:1997,
• In type of explosion protection intrinsic safety EEx EN50018:2000
ia IIC only for connection to a certified intrinsically • Type of Protection and Marking Code: EEx d IIC
safe circuit with following maximum values: T6···T4
Ui = 30 V • Temperature Class: T6, T5, and T4
Ii = 165 mA • Enclosure: IP67
Pi = 0.9 W • Maximum Process Temperature:
Effective internal capacitance; Ci = 22.5 nF 85°C (T6), 100°C (T5), and 120°C (T4)
Effective internal inductance; Li = 730 µH • Ambient Temperature: T4 and T6; –40 to 75°C,
Note 3. Installation T5; –40 to 80°C
• All wiring shall comply with local installation Note 2. Electrical Data
requirements. (Refer to the installation diagram) • Supply voltage: 42 V dc max.
Note 4. Maintenance and Repair • Output signal: 4 to 20 mA
• The instrument modification or parts replacement Note 3. Installation
by other than authorized representative of • All wiring shall comply with local installation
Yokogawa Electric Corporation is prohibited and requirement.
will void KEMA Intrinsically safe Certification. • The cable entry devices shall be of a certified
Note 5. Special Conditions for Safe Use flameproof type, suitable for the conditions of use.
• In the case where the enclosure of the Pressure Note 4. Operation
Transmitter is made of aluminium, if it is mounted • Keep the “CAUTION” label to the transmitter.
in an area where the use of category 1 G apparatus CAUTION: AFTER DE-ENERGIZING,
is required, it must be installed such, that, even in DELAY 10 MINUTES BEFORE
the event of rare incidents, ignition sources due to OPENING. WHEN THE AMBIENT
impact and friction sparks are excluded. TEMP.70°C, USE HEAT-RESISTING
CABLES90°C.
• Take care not to generate mechanical sparking
when accessing to the instrument and peripheral
devices in a hazardous location.
Note 5. Maintenance and Repair
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void KEMA Flameproof Certification.
2-8 IM 01C22D01-01E
2. HANDLING CAUTIONS
c. CENELEC ATEX (KEMA) Intrinsically Safe • The instrument modification or parts replacement
Type/CENELEC ATEX (KEMA) Flameproof by other than authorized representative of
Type/CENELEC ATEX Type n
Yokogawa Electric Corporation is prohibited and
Model EJA Series pressure transmitters with will void Type of Protection “n”.
optional code /KU2 can be selected the type of
[Installation Diagram]
protection CENELEC ATEX (KEMA) Intrinsi-
cally Safe, Flameproof or CENELEC ATEX Hazardous Location Nonhazardous Location
(Zone 2 only)
Type n for use in hazardous locations. Transmitter
Note 1. For the installation of this transmitter,
+ +
once a particular type of protection is Supply Power Supply
– –
selected, any other type of protection
cannot be used. The installation must be
in accordance with the description about
the type of protection in this user’s F0209.EPS
Ratings of the Power Supply as follows;
manual. Maximum Voltage: 30 V
Note 2. In order to avoid confusion, unnecessary
marking is crossed out on the label other
than the selected type of protection when • CENELEC ATEX Type of Protection “Dust”
the transmitter is installed. • Applicable Standard: EN50281-1-1:1997
• Type of Protection and Marking Code: II 1D
• CENELEC ATEX Type of Protection “n” • Maximum surface temperature:
T65°C (Tamb.: 40°C), T85°C (Tamb.: 60°C), and
WARNING T105°C (Tamb.: 80°C)
Note 1. Installation instructions
When using a power supply not having a non-
The cable entry devices and blanking
incendive circuit, please pay attention not to
elements shall be of a certificated type
ignite in the surrounding flammable atmosphere.
providing a level of ingress protection of
In such a case, we recommend using wiring
at least IP6x, suitable for the conditions
metal conduit in order to prevent the ignition.
of use and correctly installed.
• Applicable Standard: EN60079-15
• Referential Standard: IEC60079-0, IEC60079-11 (2) Electrical Connection
• Type of Protection and Marking Code: The type of electrical connection is stamped near
Ex nL IIC T4 the electrical connection port according to the
• Temperature Class: T4 following marking.
• Enclosure: IP67
• Process Temperature: 120°C max.
• Ambient Temperature: –40 to 60°C
Note 1. Electrical Data
Ui = 30 V
Effective internal capacitance; Ci = 22.5 nF
Effective internal inductance; Li = 730 µH
Note 2. Installation
Location of the marking
• All wiring shall comply with local installation
requirements. (refer to the installation diagram) F0200.EPS
2-9 IM 01C22D01-01E
2. HANDLING CAUTIONS
(4) Operation
Tag plate for intrinsically safe type
WARNING
12A819857 132
2-10 IM 01C22D01-01E
2. HANDLING CAUTIONS
When the fill fluid near the sensor part moves b. TIIS Intrinsically Safe Type
from within, the instrument outputs a failure
The model EJA Series pressure transmitter with optional
signal either high or low of the specific signal. In
code /JS3, which has obtained certification according to
that case, generate the alarm to identify that the
technical criteria for explosionprotected construction of
failure signal is output since the event may
electric machinery and equipment (Standards
invalidate the flameproof approval.
Notification No.556 from the Japanese Ministry of
If the optional integral indicator is equipped, the
Labor) conforming to IEC standards, is designed for
indicator identifies the alarm on its display.
hazardous areas where explosive or inflammable gases
Therefore, no other alarm generation is neces-
or vapors may be present. (This allows installation in
sary.
Division 0, 1 and 2 areas)
Hazardous Location Nonhazardous Location
To preserve the safety of flameproof equipment requires
great care during mounting,wiring, and piping. Safety
4 to 20 mA DC 1 to 5 V DC
requirements also place restrictions on maintenance and
Power DCS repair activities. Users absolutely must read “Installation
Transmitter Display
Supply
and Operating Precautions for TIIS Intrinsically Safe
F0210.EPS
Equipment” at the end of this manual.
Figure 2.3 Example of using DCS (Distributed Control
System)
2-11 IM 01C22D01-01E
2. HANDLING CAUTIONS
CAUTION
2.10 EMC Conformity Standards
EN61326-1 Class A, Table 2 (For use in industrial
For using a safety-barrier with a pressure
locations)
transmitter, the safety-barrier must be certified as
a safety-barrier itself. EN61326-2-3
A safety-barrier must be used under the
following condition.
CAUTION
(1) Condition of the current and voltage limits
This instrument is a Class A product, and it is
Maximum output voltage(Uo) ≤ 28V
designed for use in the industrial environment.
Maximum output current(Io) ≤ 94.3mA
Please use this instrument in the industrial
Maximum output power (Po) ≤ 0.66W
environment only.
(2) Category and Group
Category ia
Group II C NOTE
2-12 IM 01C22D01-01E
3. COMPONENT NAMES
3. COMPONENT NAMES
Vertical impulse piping type
Process connection
Pressure-detector section
CPU assembly
Integral
indicator (Note 1) Conduit
Mounting screw connection Zero-
adjustment
screw
Transmitter section
Cover
Setting pin (CN4) flange
Range-setting
switch (Note 1)
(See Subsection 7.6)
Open to atmosphere
Amplifier Cover (∅5mm)
Setting pin (CN4) Burn - out Output at Process connection
position (Note 2) direction burn - out Process connector (Note 1)
H
110% or
HIGH higher
L
H
-5% or
LOW lower (Note3)
L
F0301.EPS
3-1 IM 01C22D01-01E
4. INSTALLATION
4. INSTALLATION
Mounting bracket
IMPORTANT
U-bolt
50 mm(2-inch) pipe
• When welding piping during construction, take Horizontal pipe mounting
care not to allow welding currents to flow
through the transmitter.
• Do not step on this instrument after installation.
• For Model EJA430, the atmospheric opening is
Transmitter
located on the low pressure side cover flange. mounting bolt
The opening must not face upward. See
Subsection 10.4, “Dimensions, ” for the location
of the opening. U-bolt nut
Mounting bracket
50 mm(2-inch) pipe U-bolt F0401.EPS
U-bolt nut
U-bolt
50 mm(2-inch) pipe
F0402.EPS
4-1 IM 01C22D01-01E
4. INSTALLATION
Bolt
Rotate 90° or 180° segments
Conduit connection
Transmitter section
∗1
∗2
Drain/vent plug Horizontal impulse piping type
Conduit connection
Zero-adjustment screw
Pressure-detector section
4-2 IM 01C22D01-01E
4. INSTALLATION
IMPORTANT
F0405.EPS
4-3 IM 01C22D01-01E
5. INSTALLING IMPULSE PIPING
Figure 5.1.1 “H” and “L” Symbols on a Capsule Assembly Figure 5.1.2 Process Pressure Tap Angle (For Horizontal
Piping)
(2) Tightening the Process Connector Mount- (2) Position of Process Pressure Taps and
ing Bolts Transmitter
After connecting the impulse piping, tighten the If condensate (or gas) accumulates in the impulse
process connector mounting bolts uniformly. piping, it should be removed periodically by opening
the drain (or vent) plugs. However, this will generate a
transient disturbance in the pressure measurement, and
therefore it is necessary to position the taps and route
the impulse piping so that any extraneous liquid or gas
generated in the leadlines returns naturally to the
process piping.
• If the process fluid is a gas, then as a rule the
transmitter must be located higher than the process
pressure taps.
• If the process fluid is a liquid or steam, then as a
rule the transmitter must be located lower than the
process pressure taps.
5-1 IM 01C22D01-01E
5. INSTALLING IMPULSE PIPING
Union or
flange
Tee
Tee Tee
Union or flange
Tap valve
5-2 IM 01C22D01-01E
6. WIRING
6. WIRING
6-1 IM 01C22D01-01E
6. WIRING
(Note) Use a check meter whose internal resistance is 10 Ω or less. Figure 6.4.1a Connection between Transmitter and
Distributor
Power supply
+
(2) Intrinsically Safe Type
–
For intrinsically safe type, a safety barrier must be
included in the loop.
Check meter
Transmitter terminal box F0604.EPS
Hazardous Location Nonhazardous Location
Figure 6.3.4 Check Meter Connection
Transmitter terminal box
Distributor
(Power supply unit)
6.4 Wiring
Receiver
instrument
+
CAUTION –
For the intrinsically safe equipment and flame- Safety barrier F0606.EPS
proof equipment, wiring materials and wiring
work for these equipment including peripherals Figure 6.4.1b Connection between Transmitter and
Distributor
are strictly restricted. Users absolutely must read
“Installation and Operating Precautions for TIIS
Intrinsically Safe Equipment” and “Installation
and Operating Precautions for TIIS Flameproof
Equipment” at the end of this manual prior to the
work.
6-2 IM 01C22D01-01E
6. WIRING
6.4.2 Wiring Installation • Measure the cable outer diameter in two directions to
within 1 mm.
(1) General-use Type and Intrinsically Safe • Calculate the average of the two diameters, and use
Type packing with an internal diameter nearest to this value
Make cable wiring using metallic conduit or water- (see Table 6.4.2).
proof glands.
Table 6.4.2 Flameproof Packings and Applicable Cable
• Apply a non-hardening sealant to the terminal box Outer Diameters
connection port and to the threads on the flexible
Wiring Port Applicable
metal conduit for waterproofing. Optional Thread Cable OD Identifying Part
Code Diameter (mm) Mark Number
G11 8 to 10 16 8-10
G 1/2 G9601AM
G12 10.1 to 12 16 10-12
T0601.EPS
Flexible metal conduit • Mounting flameproof packing adapter body to conduit
Apply a non-hardening connection (see Figure 6.4.2c)
sealant to the threads for 1) Screw the flameproof packing adapter into the
Wiring metal waterproofing.
conduit terminal box until the O-ring touches the wiring
Tee port (at least 6 full turns), and firmly tighten the
Drain plug
lock nut.
F0607.EPS
2) Insert the cable through the union cover, the union
Figure 6.4.2a Typical Wiring Using Flexible Metal Conduit coupling, the clamp nut, the clamp ring, the gland,
the washer, the rubber packing, and the packing
box, in that order.
(2) Flameproof Type (TIIS) 3) Insert the end of the cable into the terminal box.
Wire cables through a flameproof packing adapter, or 4) Tighten the union cover to grip the cable. When
using a flameproof metal conduit. tightening the union cover, tighten approximately
one turn past the point where the cable will no
䊏 Wiring cable through flameproof packing adapter
longer move up and down.
for only TIIS flameproof type (see Figure 6.4.2b).
Proper tightening is important. If it is too tight, a
• Use only flameproof packing adapters approved by
circuit break in the cable may occur; if not tight
Yokogawa.
enough, the flameproof effectiveness will be
• Apply a nonhardening sealant to the terminal box
compromised.
connection port and to the threads on the flameproof
5) Fasten the cable by tightening the clamp nut.
packing adapter for waterproofing.
6) Tighten the lock nut on the union cover.
7) Connect the cable wires to each terminal.
6-3 IM 01C22D01-01E
6. WIRING
;;
Hazardous area F0611.EPS
Apply a non-hardening
Flameproof
sealant to the threads of Figure 6.5 Ground Terminals
heavy-gauge
these fittings for
steel conduit
waterproofing
6.5 Grounding
600
Grounding is always required for the proper operation
of transmitters. Follow the domestic electrical require- External E–10.5
R= Communication
ments as regulated in each country. For a transmitter load 0.0236
applicable range
with built-in lightning protector, grounding should resistance BRAIN and HART
R (Ω)
satisfy Class C requirement (ground resistance of 10⍀
or less).
250
Ground terminals are located on the inside and outside
of the terminal box. Either of these terminals may be
used.
For TIIS flameproof type and intrinsically safe, Figure 6.6 Relationship between Power Supply Voltage
grounding should satisfy Class D requirements and External Load Resistance
6-4 IM 01C22D01-01E
7. OPERATION
7. OPERATION
NOTE
(a) Introduce a process fluid into the impulse piping Drain valve
and then to the transmitter in the following proce-
dure: F0701.EPS
1) Open the tap valve (main valve) to fill the impulse Figure 7.1 Liquid Flow Measurement
piping with process fluid.
2) Gradually open the stop valve to introduce process
fluid into the transmitter pressure-detector section.
3) Confirm that there is no pressure leak in the Confirming that Transmitter is Operating
Properly
impulse piping, transmitter, or other components.
Using the BT200
(b) Venting Gas from the Transmitter Pressure-detector • If the wiring system is faulty, ‘communication error’
Section appears on the display.
Since the piping in the example of Figure 7.1 is • If the transmitter is faulty, ‘SELF CHECK ERROR’
constructed to be self-venting, no venting operation appears on the display.
is required. If it is not possible to make the piping
self-venting, refer to Subsection 7.5 for instructions. PARAM
C60:SELF CHECK
(c) Turn ON power and connect the BT200. ERROR
Open the terminal box cover, and connect the communication error
BT200 to the SUPPLY + and – terminals.
(d) Using the BT200, confirm that the transmitter is DATA DIAG PRNT ESC
operating properly. Check parameter values or
Communication error Self-diagnostic error
change the setpoints as necessary. See Chapter 8 for
(Faulty wiring) (Faulty transmitter)
BT200 operation. F0702.EPS
7-1 IM 01C22D01-01E
7. OPERATION
IMPORTANT
Zero-adjustment
screw
F0704.EPS
7-2 IM 01C22D01-01E
7. OPERATION
7.2.1 When you can obtain Low Range 7.2.2 When you cannot obtain Low Range
Value from actual measured value Value from actual measured value
of 0% (0 kPa, atmospheric pres- of 0%;
sure);
Convert the actual measured value obtained by a digital
For pressure measurement using gauge pressure manometer or a glass gauge into %.
transmitters, follow the step below before zero point
[Example]
adjustment.
1) Close the tap valve (main valve). The measuring range of 50 to 250 kPa; the actual
2) Loosen the fill plug so that the pressure applied to measured value of 130 kPa.
the transmitter is only the head of the seal liquid.
3) Adjust the zero point at this status. 130–50
Actual measured value= x 100=40.0%
4) After the adjustment, close the fill plug and then 250–50
gradually open the tap valve.
Using the Transmitter Zero-Adjustment Screw
Turn the screw to match the output signal to the actual
Using the Transmitter Zero-adjustment measured value in %.
Screw
Before adjusting a screw, check that the parameter Using the BT200
J20: EXT ZERO ADJ displays ENABLE. See Select the parameter J10: ZERO ADJ. Change the
Subsection 8.3.3 (11) for the setting procedure. set point (%) displayed for the parameter to the actual
measured value (%), and press the ENTER key twice.
Use a slotted screwdriver to turn the zero-adjustment
See Subsection 8.3.3 (11) for operation details.
screw. Turn the screw clockwise to increase the output
or counterclockwise to decrease the output. The zero
SET A display at J10
point adjustment can be made with a resolution of J10:ZERO ADJ
–0.0 %
0.01% of the setting range. Since the degree of zero + 000.0
adjustments varies with the screw turning speed, turn
the screw slowly for fine adjustment and quickly for
coarse adjustment. CLR ESC
SET
J10:ZERO ADJ A display when parameter
–0.0 %
+ 000.0 J10 is selected.
Press key
twice for 0% output 4 mA DC.
CLR ESC
F0705.EPS
7-3 IM 01C22D01-01E
7. OPERATION
NOTE
Drain plug
Whenever shutting down the transmitter for a
long period, remove any process fluid from the Drain screw
transmitter pressure-detector section.
When you loosen the drain screw or drain
plug, the accumulated liquid will be expelled
in the direction of the arrow.
F0707.EPS
7-4 IM 01C22D01-01E
7. OPERATION
7.6 Setting the Range Using the • Do not turn off the power to the transmitter
immediately after completion of the change in
Range-setting Switch the LRV and/or HRV setting(s). Note that
With actual pressure being applied to the transmitter, powering off within thirty seconds after setting
the range-setting switch (push-button) located on the will cause a return to the previous settings.
optional integral indicator plate and the external zero- • Changing LRV automatically changes HRV to
adjustment screw allow users to change (re-range) the the following value.
low- and high-limit values for the measurement range HRV = previous HRV + (new LRV – previous LRV)
(LRV and HRV) without using BT200. However, other
changes in the display settings (scale range and • If the range-setting push-button and external
engineering unit) for the integral indicator requires zero-adjustment screw are not touched during a
BT200. range-change operation, the transmitter auto-
matically switches back to the normal operation
Follow the procedure below to change the LRV and mode.
HRV settings.
[Example] Integral indicator
7-5 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
LCD
8.1 BT200 Operation Precau- (21 character × 8 lines)
tions
8.1.1 Connecting the BT200
The transmitter and the BT200 can be connected
either to the BT200 connection hooks in the transmit- Function keys
ter terminal box or to a relaying terminal board.
Movement keys
ENTER key
Distributor
Shift keys
F0803.EPS
F0801.EPS
Figure 8.2.1a BT200 Key Layout
Figure 8.1.1 Connecting the BT200
MENU SCREEN
Messages
8.1.2 Conditions of Communication Line MENU
A:DISPLAY
BATTERY
B:SENSOR TYPE
Menu choices
Cable resistance Rc Screen title
8.2.2 Operating Key Functions Use the function key [F1] CODE to enter symbols.
The following symbols will appear in sequence, one
(1) Alphanumeric Keys and Shift Keys at a time, at the cursor each time you press [F1]
You can use the alphanumeric keys in conjunction CODE:
with the shift keys to enter symbols, as well as
/ . – , + * ) ( ’ & % $ # ” !
alphanumeric keys.
To enter characters next to these symbols, press [ > ]
to move the cursor.
1 –9
HOME SET ADJ ESC Function commands
T0801.EPS
enter the letter shown on that side which the shift key
represents. You must press the shift key before
Function Command List
entering each letter.
Command Function
ADJ Displays the ADJ menu
Left-side letter on the Right-side letter on
alphanumeric key the alphanumeric key CAPS/caps Selects uppercase or lowercase
CODE Selects symbols
F0806 .EPS
CLR Erases input data or deletes all data
Entry Key-in Sequence DATA Updates parameter data
toggles between uppercase and lowercase each time PARM Enters the parameter number setup mode
you press [F2] CAPS. SET Displays the SET menu
SLOT Returns to the slot selection panel
UTIL Calls the utility panel
Entering uppercase Entering lowercase
*COPY Prints out parameters on display
CODE CAPS CLR ESC CODE caps CLR ESC
*FEED Paper feed
*LIST Lists all parameters in the menu
Entry Key-in Sequence *PON/POFF Automatic printout mode on or off
to lower case *PRNT Changes to the print mode
Boy *GO Starts printing
(B) (o) (y) *STOP Cancels printing
F0807.EPS
* Available on BT200-P00 (with printer).
T0804.EPS
8-2 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
PARAM
01:MODEL
EJA430-DA
INITIAL 02:TAG NO. FUNC
DATA YOKOGAWA 1.MENU
SCREEN 03:SELF CHECK (ESC) 2.UPLOAD TO BT200
GOOD 3.DOWNLOAD TO INST
OK 4.PRINT ALL DATA
(SET) (ADJ)
(HOME MENU SCREEN) (SET MENU SCREEN) (ADJ MENU SCREEN)
MENU MENU MENU
A.DISPLAY C.SETTING J.ADJUST
B.SENSOR TYPE D.AUX SET 1 K.TEST
MENU E.AUX SET 2 M.MEMO
SCREEN H:AUTO SET P:RECORD
HOME SET ADJ ESC HOME SET ADJ ESC HOME SET ADJ ESC
PARAM PARAM
A60:SELF CHECK C60:SELF CHECK
GOOD GOOD
8-3 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
PRESS
A40 INPUT Input (indicated as the — –32000 to 32000
value after zeroing)
A60 SELF CHECK Self-diagnostic — GOOD/ERROR, CAP MODULE FAULT, AMP MODULE FAULT,
messages OUT OF RANGE, OUT OF SP RANGE*1, OVER TEMP (CAP),
OVER TEMP (AMP), OVER OUTPUT, OVER DISPLAY,
ILLEGAL LRV, ILLEGAL HRV, ILLEGAL SPAN, and
ZERO ADJ OVER
B SENSOR Sensor type — Menu name
TYPE
B10 MODEL Model+span — 16 uppercase alphanumerics
B11 STYLE NO.Style number —
pressure*6
B60 SELF CHECK Self-diagnostic — Same as A60
messages
C SETTING Setting data — Menu name
C10 TAG. NO. Tag number 16 alphanumerics As specified when ordered.
C20 PRESS UNIT Measurement range Selected from mmH2O, mmAq, As specified when ordered.
units mmWG, mmHg, Torr, Pa, hPa, kPa,
MPa, mbar, bar, gf/cm2, kgf/cm2,
inH2O, inHg, ftH2O, psi, or atm
C21 LOW RANGE Measurement range, –32000 to 32000(but within As specified when ordered.
lower range value measurement range)
C22 HIGH Measurement range, –32000 to 32000(but within As specified when ordered.
RANGE higher range value measurement range)
C30 AMP Damping time constant Selected from 0.2*2, 0.5, 1.0, 2.0, 2.0 s
DAMPING 4.0, 8.0, 16.0, 32.0, or 64.0 sec.
C40 OUTPUT Output mode and Selected from OUT:LIN; DSP:LIN, As specified when ordered. — —
8-4 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
Applica-
No. Item Description Rewrita- Remarks bility
Default Value
bility F P L
D AUX SET 1 Auxiliary setting data 1 — Menu name
D31 STAT. P. UNIT Static pressure setting Selected from mmH2O, mmAq, As specified when ordered. —
units mmWG, mmHg, Torr, Pa, hPa, kPa, If not specified, MPa.
MPa, mbar, bar, gf/cm2, kgf/cm2,
inH2O, inHg, ftH2O, psi, or atm
D40 REV OUTPUT Output reversal NORMAL/REVERSE If not specified, NORMAL.
D45 H/L SWAP Impulse piping NORMAL/REVERSE*4 NORMAL — —
accessing direction
D52 BURN OUT CPU error —
HIGH/LOW, –5 to 110%*3 HIGH
D53 ERROR OUT Hardware error HOLD/HIGH/LOW, –5 to 110%*3 HIGH
D60 SELF CHECK Self-diagnostic — Same as A60
messages
E AUX SET 2 Auxiliary setting data 2 — Menu name
E10 DFS MODE DFS mode OFF/ON*5 ON —
SELECT sensor
E15 TEMP ZERO Zero shift 10.00*5 0.00 —
conpensation setup
E30 BI DIRE Bidirectional mode OFF/ON OFF — —
MODE
E50 AUTO Auto-recover from OFF/ON ON
RECOVER sensor error
E60 SELF CHECK Self-diagnostic — Same as A60
messages
H AUTO SET Automatic setup — Menu name
H10 AUTO LRV Automatic measure- –32000 to 32000 Displays the same data as
ment range lower C21.
range value setup
H11 AUTO HRV Automatic –32000 to 32000 Displays the same data as
measurement range C22.
higher range value
setup
H60 SELF CHECK Self-diagnostic — Same as A60
messages
J ADJUST Adjustment data — Menu name
J10 ZERO ADJ Automatic zero –5 to 110.0%*3
adjustment
J11 ZERO DEV. Manual zero
adjustment
J15 SPAN ADJ Manual span adjustment –10.00 to 10.00% 0.00%
J20 EXT. ZERO External zero- ENABLE/INHIBIT
ADJ adjustment screw
permission
J30 OUTPUT 4mA 4mA adjustment –10.00 to 10.00% 0.00%
J31 OUTPUT 20mA 20mA adjustment –10.00 to 10.00% 0.00%
J60 SELF CHECK Self-diagnostic — Same as A60
messages
K TEST Tests — Menu name
K10 OUTPUT in % Test output % setting –5 to 110.0%*3 Displays ‘ACTIVE’
while executing
K60 SELF CHECK Self-diagnostic — Same as A60
messages
M MEMO Memo — Menu name
M10 MEMO 1 Memo 8 uppercase alphanumerics
M20 MEMO 2 Memo 8 uppercase alphanumerics
M30 MEMO 3 Memo 8 uppercase alphanumerics
M40 MEMO 4 Memo 8 uppercase alphanumerics
M50 MEMO 5 Memo 8 uppercase alphanumerics
M60 SELF CHECK Self-diagnostic — Same as A60
messages
P RECORD History of the errors —
8-5 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
8-6 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
F0811.EPS
F0810.EPS
8-7 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
SET
C20:PRESS UNIT
The higher range value is changed • Example: Change from 2.0 sec to 4.0 sec.
kPa
C21:LOW RANGE while the span remains constant.
SET
0.5 kPa C30:AMP DAMPING Use the or key to
C22:HIGH RANGE 2.0 sec
30.5 kPa < 2.0 sec > select 4.0 sec.
DATA DIAG PRNT ESC < 4.0 sec >
< 8.0 sec > Press the key twice to
< 16.0 sec >
ESC enter the setting.
Span = Higher range value – Lower range value
SET
C30:AMP DAMPING Press the (OK) key.
F0813.EPS 4.0 sec
FEED NO OK
• Note, however, that changing the higher range value 0.2sec
does not cause the lower range value to change. 0.5sec
1.0sec
Thus, changing the higher range value also changes
2.0sec
the span.
4.0sec
• Calibration range can be specified with range value
8.0sec
specifications up to 5 digits (excluding any decimal 16.0sec
32.0sec
point) for low or high range limits within the range 64.0sec
of –32000 to 32000.
F0815.EPS
Note1: The damping time constant set here is the damping time
constant for the amplifier assembly. The damping time
constant for the entire transmitter is the sum of the values for
the amplifier assembly and for the capsule assembly. For the
capsule assembly damping time constant (fixed), see the
“General Specifications” found at the end of this manual.
(See Chapter 10.)
8-8 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(4) Output Signal Low Cut Mode Setup (5) Change Output Limits
(D10: LOW CUT, D11: LOW CUT MODE) (D15:OUT LIMIT(L), D16:OUT LIMIT(H))
Low cut mode can be used to stabilize the output The range of normal output is preset at factory from
signal near the zero point. The low cut point can be -5.0 to 110.0% unless otherwise specified, and the
set in a range from 0 to 20% of output. (Hysteresis of output is limited with these upper and lower values.
cut point: ±1%) This output range can be changed, for example, to
meet the requirements of NAMUR, etc. within the
Select “ZERO” as the low cut mode.
settable range. Set the lower limit with D15:OUT
LIMIT(L) and upper limit with D16:OUT LIMIT(H).
LOW CUT mode ZERO at 10%
Settable range : -5.0 to 110.0 (%)
Lower limit < Upper limit
50
(6) Integral Indicator Scale Setup
Output(%) The following 5 displays are available for integral
indicators.
10
D20: DISP SELECT Description
and Display and Related parameters
0 10 50
Input(%) NORMAL % Indicates –5 to 110% range
F0816.EPS depending on the Measurement
range (C21, C22).
A10:OUTPUT (%)
• Example: Change the low cut setting range 45.6 %
from 5% to 10%, and the low USER SET Indicates values depending on the
cut mode from LINEAR to ZERO. Engineering range (D22, D23).(Note 1)
Units set using Engineering unit
SET Set “10.” (D21) are not indicated.
D10:LOW CUT
5.0 %
+ 10.0 Press the key twice to
A11:ENGR.OUTPUT
enter the setting. 20.0 M
CLR ESC
USER & % Indicates user set and % alternately
SET Press the (OK) key. in 3 second intervals.
D10:LOW CUT
10.0 %
Next, the [D11: LOW CUT MODE] A10:OUTPUT (%)
45.6 %
setting panel is displayed.
A11:ENGR. OUTPUT
FEED NO OK
20.0 M
SET Use the or key INP PRES Indicates input pressure.(Note 2)
D11:LOW CUT MODE
LINEAR Indication limits –19999 to 19999.
< LINEAR > to select ZERO.
< ZERO >
Press the key twice to A40:INPUT
ESC
enter the setting. 456 kPa
SET Press the (OK) key. PRES & % Indicates input pressure and %
D11:LOW CUT MODE
ZERO alternately in 3 second intervals.
A10:OUTPUT (%)
45.6 %
FEED NO OK
A40:INPUT
PARAM 456 kPa
D10:LOW CUT
10.0 %
D11:LOW CUT MODE (Note 1) Scale range can be specified with range limit specifications up
ZERO to 5 digits (excluding any decimal point) for low or high range
D20:DISP SELECT
NORMAL % limits within the range of –19999 to 19999.
DATA DIAG PRNT ESC
The range with decimals is available to the third decimal place.
(Note 2) It indicates the value after zeroing.
F0817.EPS
T0808.EPS
8-9 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
FEED NO OK
The “%” disappears from the
PARAM
integral indicator display. D21:DISP UNT
M
D22:DISP LRV
– 50M
D23:DISP HRV
F0819.EPS 50M
DATA DIAG PRNT ESC
F0821.EPS
8-10 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(7) Unit Setup for Displayed Temperature (9) Output Status Display/Setup when a CPU
(D30: TEMP UNIT) Failure (D52: BURN OUT)
When the instrument is shipped, the temperature units This parameter displays the status of 4 to 20 mA DC
are set to degC. Follow the procedure below to output if a CPU failure occurs. In case of a failure,
change this setting. Note that changing the unit here communication is disabled.
changes the unit for A20: AMP TEMP (amplifier
Setting of HIGH or LOW is enabled. This is done
temperature) and A21: CAPSULE TEMP (capsule
with the pin (CN4) on the CPU assembly. See
temperature).
Chapter 3 for details.
Standard specifications
• Example: Change the unit for the temperature The parameter is set to HIGH. If a failure, the
display. transmitter outputs the signal of 110% or higher. The
SET Use the or key to parameter D53: ERROR OUT is set to HIGH from
D30:TEMP UNIT
deg C
< deg C > select “deg F.” the factory.
< deg F >
Press the key twice to Optional code/C1
ESC
enter the setting. The parameter is set to LOW. If a failure, output
F0822.EPS which is –5%*1 or lower is generated. The parameter
D53: ERROR OUT is set to LOW from the factory.
(8) Operation Mode Setup
*1 : When optional code /F1 is specified, substitute the value with
(D40: REV OUTPUT) –5 with –2.5.
This parameter allows the direction of the 4 to 20
mA output to be reversed with respect to input.
• Example: Standard specifications
Follow the procedure below to make this change.
D52: BURN OUT pin (CN4) position: H
HIGH
• Example: Change 4 to 20 mA output to 20 to
4 mA output. • Example: Optional code/C1
SET Use the or key D52: BURN OUT pin (CN4) position: L
D40:REV OUTPUT
NORMAL LOW
< NORMAL > to select REVERSE.
< REVERSE>
Press the key twice to F0824.EPS
ESC
enter the setting.
(10) Output Status Setup when a Hardware
F0823.EPS
Error Occurs (D53: ERROR OUT)
This parameter allows the setting of the output status
when a hardware error occurs. The following three
selections are available.
(a) HOLD; Outputs the last value held before the
error occurred.
(b) HIGH; Outputs an output of 110% when an error
has occurred.
(c) LOW; Outputs an output of –5%*1 when an error
has occurred.
Note: A hardware error means CAP MODULE FAULT of Er.01 or
AMP MODULE FAULT of Er. 02 which are shown in 8.5.2
“Errors and Countermeasures.”)
*1 : When optional code /F1 is specified, substitute the value with
–5 with –2.5.
8-11 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(11) Range Change while Applying Actual Note that changing the higher range value does not
Inputs (H10: AUTO LRV, H11: AUTO cause the lower range value to change but does
HRV)
change the span.
This feature allows the lower and higher range values
to be set up automatically with the actual input
applied. If the lower and higher range values are set, • Example 2: When the higher range value is to
C21: LOW RANGE and C22: HIGH RANGE are be changed to 10 kPa with the
changed at this same time. present setting of 0 to 30 kPa, take
the following action with an input
Follow the procedure in the figure below. pressure of 10 kPa applied.
The measurement span is determined by the higher
SET
and lower range values. Changing the lower range H11:AUTO HRV
Press the key twice.
30 kPa
value results in the higher range value changing auto- + 30 The higher range value is changed
matically, keeping the span constant. to 10 kPa.
ESC
8-12 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(a) Follow the procedure below when setting the (b)-2 Follow the procedure below to use J11: ZERO
present output to 0% (4 mA). DEV.
SET
J10:ZERO ADJ Zero adjustment is completed. SET Set the correction value, 1.50.
J11:ZERO DEV.
0.0 % 2.50 %
Press the (OK) key. 1.50 Press the key twice.
FEED NO OK ECS
A10:OUTPUT (%) Output is 0%. A10:OUTPUT (%) The output is changed to 40%.
0.0 % 40.0 %
F0830.EPS
F0828.EPS
(b) In tank level measurement, if the actual level When the zero point is adjusted, the displayed value
cannot be brought to zero for zero adjustment, of A40 is as follows.
then the output can be adjusted to correspond to
the actual level obtained using another measuring [Example]
instrument such as a glass gauge. When the zero point is shifted by 20 kPa for the
actual pressure, the parameter of A40 indicates 50
[Example] kPa.
Measurement range: 50 to 250 kPa, Actual value:
130 kPa.
70 kPa
Actual value 50 kPa
Actual = –Measurement range lower range value
x 100
value(%) Measurement range higher range value
–Measurement range lower range value Zero
20 kPa adjustment
Zero adjustment
130–50 volume(J11)
0 kPa 0 kPa
= x 100=40.0%
250–50 Actual differential Input value
pressure/pressure of A40 F0839.EPS
A10:OUTPUT (%)
Present output is 41.0%.
41.0 %
CLR ESC
F0829.EPS
8-13 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(c) Zero Point Adjustment Using the External Zero You can manually perform the trimming procedure
Adjustment Screw by using J15: SPAN ADJ.
• Enabling/inhibiting of zero point adjustment using • Span adjustment value
the external zero-adjustment screw on the transmitter The span adjustment value is calculated as follows.
(J20: EXT ZERO ADJ)
Follow the procedure below to enable or inhibit zero Span adjustment value (%) = P1 A40 100
P1
point adjustment from the zero-adjustment screw on
P1: Actual differential pressure/pressure value
the transmitter.
A40: Input (indicated as the value after zeroing)
This is set to “ENABLE” when the instrument is
shipped. Measurement pressure
ESC
enter the setting.
F0831.EPS
SET
J15:SPAN ADJ
Set 0.35.
0.15 %
(13) Span Adjustment - 0.35 Press key twice.
Each DPharp EJA series transmitter is factory
DEL CLR ESC
characterized according to the specification. Mount-
ing position effects or zero shifts caused by static SET
J15:SPAN ADJ
Press the (OK) key.
-0.35 %
pressure are typically compensated by a zero adjust-
ment.
FEED NO OK
A span adjustment is a function to correct the slope
error from a zero point in characterizing 100% point Note: Enter 0.00 to J15: SPAN ADJ to reset the
(HRV). This function can be used when span drifts span adjustment to the initial value at the
may be caused or characterization to the specific shipment.
pressure standard is required.
F0847.EPS
8-14 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
(14) Test Output Setup (K10: OUTPUT X%) 8.4 Displaying Data Using the
This feature can be used to output a fixed current BT200
from 3.2 mA (–5%) to 21.6 mA (110%) for loop
checks. 8.4.1 Displaying Measured Data
Note: When optional code /F1 is specified, output range is from
The BT200 can be used to display measured data.
3.6mA(–2.5%) to 21.6mA(110%).
The measured data is updated automatically every 7
• Example: Output 12 mA (50%) fixed current. seconds. In addition, the display can be updated to
the present data value at any time by pressing the
SET Set “50.0%.”
K10:OUTPUT X % (DATA) key. For parameters associated with the
0.0 %
+ 050.0 Press the key twice to display of measured data, see Subsection 8.3.1,
output a fixed current at 50%. “Parameter Summary.”
ESC
F0832.EPS
HOME SET ADJ ESC
A20:AMP TEMP
tion cable is disconnected during test output,
it is held for approximately 10 minutes.
2. Press the (OK) key to release test output F0834.EPS
immediately.
8.4.2 Display Transmitter Model and
Specifications
(15) User Memo Fields (M: MEMO) The BT200 can be used to display the model and
This feature provides 5 user memo fields, each specifications of the transmitter.
holding up to 8 alphanumeric characters. Up to 5
items such as inspection date, inspector, and other
• Example: View transmitter model name.
information can be saved in these fields.
MENU Press .
A:DISPLAY
B:SENSOR TYPE
• Example: Save an inspection date of January
30, 1995.
HOME SET ADJ ESC
PARAM Set “95.1.30” in the order of year,
M10:MEMO 1
F0833.EPS
F0835.EPS
8-15 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
The following four areas can be checked. HOME SET ADJ ESC
Since communications will be The history of up to four errors can be stored. When the 5th
unsuccessful if there is a problem error has occurred, it is stored in “P10”. The error stored in
communication error
“P13” will be deleted, and then, the error in “P12” will be
in the connection to the BT200, the
copied to “P13”. In this sequence, the history of the most
ESC display at the left will appear.
previously occurred error will be removed from memory.
Recheck the connection. “GOOD” will be displayed if there was no previous error.
Press the (OK) key.
SET Select P10: ERROR REC1 and
P10:ERROR REC 1
ERROR press the key to display
< ERROR >
< ILLEGAL LRV > the error message.
• Example 2: Setting entry errors < ILLEGAL HRV >
ESC
PARAM The initial data panel shows the
01:MODEL
EJA430-DA <(a) SETUP PANEL>
02:TAG NO. result of current transmitter
YOKOGAWA
03:SELF CHECK diagnostics. For the details of the messages listed below, see Table
ERROR
OK 8.5.1 Error Message Summary.
8-16 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
NOTE
F0838.EPS
8-17 IM 01C22D01-01E
8. BRAIN TERMINAL BT200 OPERATION
Integral
Output Operation
Indicator BT200 Display Cause Countermeasure
during Error
Display
None GOOD
---- ERROR
Er. 01 CAP MODULE Capsule problem.*1 Outputs the signal Replace the capsule
FAULT (Hold, High, or Low) when error keeps
set with parameter appearing even after
D53. restart.*2
Er. 02 AMP MODULE Amplifier problem. Outputs the signal Replace amplifier.
FAULT (Hold, High, or Low)
set with parameter
D53.
Er. 03 OUT OF RANGE Input is outside Outputs high range Check input.
measurement range limit value or low
limit of capsule. range limit value.
Er. 04 OUT OF SP Static pressure Displays present Check line pressure
RANGE exceeds specified output. (static pressure).
range.*3
Er. 05 OVER TEMP Capsule temperature Displays present Use heat insulation or
(CAP) is outside range output. make lagging to keep
(–50 to 130°C). temperature within
range.
Er. 06 OVER TEMP Amplifier temperature Displays present Use heat insulation or
(AMP) is outside range output. make lagging to keep
(–50 to 95°C). temperature within
range.
Er. 07 OVER OUTPUT Output is outside high Outputs high or low Check input and range
or low range limit range limit value. setting, and change
value. them as needed.
Er. 08 OVER DISPLAY Displayed value is Displays high or low Check input and
outside high or low range limit value. display conditions and
range limit value. modify them as
needed.
Er. 09 ILLEGAL LRV LRV is outside setting Holds output Check LRV and
range. immediately before modify as needed.
error occurrence.
Er. 10 ILLEGAL HRV HRV is outside setting Holds output Check HRV and
range. immediately before modify as needed.
error occurrence.
Er. 11 ILLEGAL SPAN SPAN is outside Holds output Check SPAN and
setting range. immediately before change as needed.
error occurrence.
Er. 12 ZERO ADJ OVER Zero adjustment is too Displays present Readjust zero point.
large. output.
*1: This error code appears at a capsule problem or when an illegal overpressure is applied to the
pressure sensor.
*2: If the normal pressure is regained, the Er.01 will disappear according to the setting of the parameter
of E50: AUTO RECOVER. When the E50: AUTO RECOVER is set to ON(defalut setting), the Er.01
will disappear automatically. When the E50: AUTO RECOVER is set to OFF, restart the transmitter
to cancel Er.01. If no error code appears then, perform necessary adjustment such as zero-adjust-
ment to continue the operation. If the error code still exists, replace the capsule assembly.
*3: For Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is
not a measured value.
T0811 .EPS
8-18 IM 01C22D01-01E
9. MAINTENANCE
9. MAINTENANCE
9-1 IM 01C22D01-01E
9. MAINTENANCE
Note: The above table contains the instruments capable of performing calibration to the 0.2% level. Since special maintenance and management
procedures involving traceability of each instrument to higher-level standards are required for calibration to the 0.1% level, there are
difficulties in calibration to this level in the field. For calibration to the 0.1% level, contact Yokogawa representatives from which the
instrument was purchased or the nearest Yokogawa office.
F0901.EPS
9-2 IM 01C22D01-01E
9. MAINTENANCE
bly
CAUTION
This section describes procedures for disassembly and
reassembly for maintenance and component replace-
Cautions for TIIS Flameproof Type Transmit-
ment.
ters
Always turn OFF power and shut off and release Users are prohibited by law from modifying the
pressures before disassembly. Use proper tools for all construction of a flameproof type transmitter.
operations. Table 9.4.1 shows the tools required. This would invalidate the agency approval and
Table 9.4.1 Tools for Disassembly and Reassembly the transmitter’s use in such rated area.
Thus the user is prohibited from using a flame-
Tool Quantity Remarks
proof type transmitter with its integral indicator
Phillips screwdriver 1 JIS B4633, No. 2
Slotted screwdriver 1
removed, or from adding an integral indicator to
Allen wrenches 2 JIS B4648 a transmitter. If such modification is absolutely
One each, nominal 3 and required, contact Yokogawa.
5 mm Allen wrenches
Wrench 1 Width across flats, 17 mm
Torque wrench 1 This subsection describes the procedure for replacing
Adjustable wrench 1 an integral indicator. (See Figure 9.4.2)
Socket wrench 1 Width across flats, 16 mm
Socket driver 1 Width across flats, 5.5 mm Removing the Integral Indicator
Tweezers 1
T0902.EPS
1) Remove the cover.
2) Supporting the integral indicator by hand, loosen its
two mounting screws.
CAUTION
3) Dismount the LCD board assembly from the CPU
Precautions for CENELEC, IECEx, and TIIS assembly.
Flameproof Type Transmitters When doing this, carefully pull the LCD board
• Flameproof type transmitters must be, as a assembly straight forward so as not to damage the
rule, removed to a non-hazardous area for connector pins between it and the CPU assembly.
maintenance and be disassembled and reas-
sembled to the original state. For details, see
“Installation and Operating Precautions for TIIS
Flameproof Equipment” later in this manual.
• On the flameproof type transmitters the two
covers are locked, each by an Allen head bolt
(shrouding bolt). When a shrouding bolt is
driven clockwise by an Allen wrench, it is going
in and cover lock is released, and then the
cover can be opened.
When a cover is closed it should be locked by a
shrouding bolt without fail. Tighten the shroud-
ing bolt to a torque of 0.7 N·m.
Shrouding Bolt
Shrouding Bolt
9-3 IM 01C22D01-01E
9. MAINTENANCE
Attaching the Integral Indicator 9.4.2 Replacing the CPU Board Assembly
Integral indicator can be installed in the following three
This subsection describes the procedure for replacing
directions.
the CPU assembly. (See Figure 9.4.2)
9-4 IM 01C22D01-01E
9. MAINTENANCE
9-5 IM 01C22D01-01E
9. MAINTENANCE
Bolt
Process connector
YES Does problem exist in
Process connector gasket
receiving instrument?
NO
Inspect receiver.
Check/correct
F0905.EPS Check transmitter.
environmental conditions.
Figure 9.4.4 Removing and Mounting the Process Operating conditions
Connector
Check/correct operating
conditions.
F0906.EPS
9-6 IM 01C22D01-01E
9. MAINTENANCE
Connect BRAIN TERMINAL and check self-diagnostics. NO Refer to error message summary in
Subsection 8.5.2 to take actions.
Is power supply NO
polarity correct? Are valves opened or NO
closed correctly?
Are power
NO
supply voltage and load
resistance correct? YES
Is there any pressure leak?
Is there
continuity through the NO
transmitter loop wiring?
Do the loop numbers
match?
9-7 IM 01C22D01-01E
9. MAINTENANCE
Is impulse piping NO
connected correctly?
Is transmitter
installed where there is YES
marked variation in
temperature?
Were appropriate NO
instruments used for
calibration?
NO
Is output adjusted correctly?
9-8 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10-1 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10-2 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10-3 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10-5 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10-6 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
Long vent Total vent plug Length: 112 mm (standard, 32 mm), Material: SUS316 U
Fast response Update time: 0.125 sec or less, see GS for the response time F1
Failure alarm down-scale *1 Output status at CPU failure and hardware error is –5%, 3.2 mA or less. C1
Failure alarm down-scale: output status at CPU failure and
C2
Output signal limits: hardware error is –5%, 3.2 mA or less.
NAMUR NE43 compliant *1
3.8 mA to 20.5 mA Failure alarm up-scale: output status at CPU failure and
C3
hardware error is 110%, 21.6 mA or more.
Stainless steel Amplifier housing material; SCS14A stainless steel (equivalent to SUS316 cast stainless
E1
amplifier housing steel or ASTM CF-8M)
Gold-plate Gold-plated diaphragm A1
Stainless steel tag plate SUS304 tag plate wired onto transmitter N4
130 Pa abs {1 mmHg abs} Minimum input pressure: 130 Pa abs{1 mmHg abs} at range calibrating testing S1
* 1: Applicable for Output signal code D and E. The hardware error indicates faulty amplifier or
capsule. When combining with Optional code F1, output status for down-scale is –2.5%,
3.6 mA DC or less.
10-7 IM 01C22D01-01E
10. GENERAL SPECIFICATIONS
10.4 Dimensions
䊉 Model EJA310 and EJA430 [Style: S2]
(1.81)
46
(5.83)
(3.07)
148
ø78
234(9.21)
Terminal
(4.02)
side
102
Internal
indicator
Ground (Optional)
terminal Conduit
connection Open to
Mounting bracket
( 2.09)
(Optional) Terminal
ø78
External indicator
conduit connection side
(7.76)
197
Blind plug
(5.75)
(6.38)
(Optional)
146
Ground
162
terminal
Shrouding bolt
(Note 4) High Low
pressure pressure
side (Note 1) side
Process
connection Open to
Process Vent plugs atmosphere (Note 3)
27 (1.06)
(4.88)
46 connector (ø5mm)
124
F1003.EPS
Note 1: When INSTALLATION CODE ‘2’, ‘3’ or ‘8’ is selected, high and low pressure side on above
figure are reversed.
(i. e. High pressure side is on the left side.)
Note 2: When INSTALLATION CODE ‘3’ or ‘7’ is selected, process connetion and mounting bracket on
above figure are reversed.
Note 3: Applicable for EJA430.
Note 4: Applicable only for ATEX, IECEx, and TIIS Flameproof type.
10-8 IM 01C22D01-01E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
1. General words, they are not safe products involved with factors such
as chemical reactions, geographical changes or the like other
The following describes precautions on electrical apparatus than affected by electric energy from the equipment itself.
of intrinsically safe construction (hereinafter referred to as
intrinsically safe apparatus).
Following the Labor Safety and Health Laws of Japan, an 2. Electrical Apparatus of Intrinsic
intrinsically safe apparatus must undergo type tests in order Safety Type of Explosion-
to be certified by the Technical Institute of Industrial Safety, Protected Construction
Inc. These tests are required to satisfy either the technical
criteria for electrical machinery and equipment in compliance The intrinsic safety type of explosion-protected construction
with explosionproof standards involving inflammable gases is a method of protection applicable to a circuit or part of a
or vapors and for machinery and equipment having circuit in which, under prescribed test conditions, no spark or
explosionproof performance (standards notification no. 556 thermal effect, whether produced normally or accidentally, is
from the Japanese Ministry of Labor) (hereinafter referred to capable of causing a prescribed explosive gas to ignite. In
as technical criteria), in conformity with IEC Standards, or other words, electrical apparatus of this construction is
the “Recommended Practice for Explosion-Protected intended to suppress electrical energy thereby preventing
Electrical Installations in General Industries,” published in ignition of a given explosive gas atmosphere even though
1979. Such a certified apparatus can be used in hazardous spark or high thermal effect occurs in the electric circuitry.
locations where inflammable gases or vapors may be present. Intrinsically safe electrical apparatus generally comprise
Certified apparatus includes a certification label and an intrinsically safe apparatus installed in a hazardous location
equipment nameplate with the specifications necessary for and a safety barrier (associated apparatus), installed in a non-
explosion requirements as well as precautions on explosion hazardous location, aimed at preventing electrical energy
protection. Please confirm these precautionary items and use from flowing into the electric circuitry of intrinsically safe
them to meet specification requirements. apparatus.
For electrical wiring and maintenance servicing, please refer However, battery-operated, portable intrinsically safe
to “Internal Wiring Rules” in the Electrical Installation apparatus or the like may be used alone.
Technical Standards as well as “USER’S GUIDELINES for
Electrical Installations for Explosive Gas Atmospheres in
General Industry,” published in 1994.
3. Terminology
(1) Intrinsically safe apparatus: Electrical apparatus in which
all the circuits are intrinsically safe circuits.
To meet intrinsically safe requirements, equipment that can (2) Associated apparatus: Electrical apparatus in which there
be termed an “intrinsically safe apparatus” must: are both intrinsically safe circuits and non-intrinsically
safe circuits that can affect the safety of intrinsically safe
(1) be certified by the Technical Institute of Industrial
circuits.
Safety, Inc. in accordance with the Labor Safety and
(3) Safety barrier: A specific type of associated apparatus,
Health Laws of Japan and have the appropriate mark of
which consists mainly of safety barrier elements, and
certification labeled on its case, and
serves to limit the flow of excessive electrical energy,
(2) be used in compliance with the specifications marked on
which is capable of causing ignition of a given explosive
its certification label, equipment nameplate and
gas or vapour of a non-intrinsically safe circuit into
precautionary information furnished.
concerned intrinsically safe circuits.
Note: Intrinsically safe apparatus satisfy their performance under (4) Apparatus of category “ia”: Intrinsically safe electrical
specific conditions. They are not always absolutely safe under apparatus and associated apparatus which are incapable
every operational and environmental condition. In other of causing ignition of a given explosive gas or vapour
with the appropriate safety factors such as:
1 EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
— when up to two countable faults are applied and, in (2) For pressure transmitters, pH transmitters, temperature
detectors and the like, safety barriers that can be
addition,
combined are already specified. Other safety barriers
— when non-countable faults produce an onerous cannot be used.
condition.
(5) Apparatus of category “ib”: Intrinsically safe electrical Note 2: Testing Intrinsically Safe System
An assembly (as a system) in which intrinsically safe
apparatus and associated apparatus which are incapable
apparatus and safety barriers are combined is assessed to
of causing ignition of a given explosive gas or vapour, ensure that its safety requirements are satisfied. A tested and
with the appropriate safety factors such as: certified system incorporates a certification number
— when up to one countable fault is applied and, in (intrinsically safe apparatus and safety barriers have the same
certification number).
addition,
— when non-countable faults produce an onerous Note 3: Impossible Combinations of Apparatus Certified Under
condition. Different Standards
(6) Safety rating: A rating to be designated to intrinsically Intrinsically safe apparatus certified under technical criteria
and safety barriers certified under the “Recommended
safe apparatus as well as associated apparatus and is the
Practice for Explosion-Protected Electrical Installations in
maximum rating allowable for maintaining intrinsic General Industries” (1979) and vice versa cannot be
safety of concerned intrinsically safe circuits. combined even if their combination requirements are
satisfied.
2 EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS INTRINSICALLY SAFE EQUIPMENT
intrinsically safe apparatus may be used at an ambient Installations for Explosive Gas Atmospheres in General
temperature up to 60°C. So, specifications should be checked Industry” issued in 1994 by the Japanese Ministry of Labour,
before installing intrinsically safe apparatus. the Research Institute of Industrial Safety.
In intrinsically safe construction, safety shall be maintained (2) Maintenance and Inspection
as an intrinsically safe system involving intrinsically safe (a) Visual inspection
apparatus and safety barriers connected thereto, and electrical Visually inspect the external connections of intrinsically
wiring (through intrinsically safe circuits) interconnected safe apparatus and safety barriers, and cables for damage
between them. In other words, even when safety or corrosion as well as other mechanical and structural
requirements are maintained individually by intrinsically safe defects.
apparatus and safety barriers, they shall not be affected by (b) Adjustments
electrical or magnetic energy caused by electrical wiring. Zero, span and sensitivity adjustments shall be made
To make electrical wiring for intrinsically safe circuits, you with applicable adjusting potentiometers and mechanical
must: adjustment screws.
These maintenance adjustments shall be made in a non-
(a) refer to the equipment configuration diagram and make hazardous location.
electrical wiring properly;
(b) prevent intrinsically safe wiring from being contacted
with non-intrinsically safe wiring, and separate the CAUTION
intrinsically safe circuit from other electrical circuits;
(c) prevent intrinsically safe wiring from being If intrinsically safe apparatus and safety barriers
electrostatically and magnetically affected by non- require maintenance service and checking, a
intrinsically safe wiring; gas detector shall be used to ensure that there
(d) reduce wiring inductance and capacitance produced is no explosive gas in the location (mainte-
between the intrinsically safe apparatus and safety nance servicing shall be conducted in a non-
barrier where possible, and use a shorter cable between hazardous location).
the intrinsically safe apparatus and safety barrier than
specified if the maximum permissible inductance of the
cable is specified as operating conditions; (3) Repair
(e) conform to conditions of installation such as wiring Intrinsically safe apparatus and safety barriers shall be
method, earthing or the like, if any; and repaired by manufacturers.
(f) protect the outer sheath of cables from damage with
appropriate measures. (4) Prohibition of modifications and specifica-
tion changes
Do not attempt to make modifications or change specifica-
tions which may affect safety.
7. Maintenance and Inspection of
Intrinsically Safe Apparatus
and Safety Barriers
Maintenance and inspection of intrinsically safe apparatus
and safety barriers shall be limited to within the instructions
described in applicable instruction manuals. If other than this
is required, contact the manufacturers. For more information,
refer to the “USER’S GUIDELINES for Electrical
3 EX-A03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
To meet flameproof requirements, equipment that can be (4) Path length of joint surface
termed “flameproof” must: On a joint surface, the length of the shortest path through
(1) Be certified by a Japanese public authority in accordance which flame flows from the inside to outside of the
with the Labour Safety and Health Laws of Japan and flameproof enclosure. This definition cannot be applied to
have a certification label in an appropriate location on its threaded joints.
case, and
(5) Gaps between joint surfaces
(2) Be used in compliance with the specifications marked on
its certification label, equipment nameplate and The physical distance between two mating surfaces, or
precautionary information furnished. differences in diameters if the mating surfaces are cylindrical.
2. Electrical Apparatus of Flame- Note: The permissible sizes of gaps between joint surfaces, the path
length of a joint surface and the number of joint threads are
proof Type of Explosion- determined by such factors as the enclosure’s internal
Protected Construction volume, joint and mating surface construction, and the
explosion classification of the specified gases and vapours.
Electrical apparatus which is of flameproof construction is
subjected to a type test and certified by the Japanese Ministry
of Labour aiming at preventing explosion caused by electrical
apparatus in a factory or any location where inflammable
gases or vapours may be present. The flameproof
1 EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
2 EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
CAUTION
3 EX-B03E
Customer DPharp EJA Series
Maintenance Transmitter Section
Parts List
5
2 4
11
A 10
12
13 1
2 2
1
14
6
7-1 7-2
8 9
5
1 Bellow 2 Cover
F9341RA Cast-aluminum alloy
F9341RJ SCS14A stainless steel
2 F9341JP 2 O-ring
3 Below 1 Case Assembly (Note 1)
F9341AR SCS14A stainless steel for 1/2 NPT (two electrical connections)
4 – 1 Name Plate
5 Bellow 4 Screw
F9900RG For cast-aluminum alloy case assembly
F9900RR For SCS14A stainless steel case assembly
F9342AD For BRAIN protocol version TIIS Intrinsically safe type with /F1(Optional code /JS3 and /F1)
F9342AF For BRAIN protocol version (Optional code /F1)
F9342AM For HART protocol version with write protection switch (Optional code /F1)
7-2 F9342BF For FOUNDATION Fieldbus protocol
F9342BG For FOUNDATION Fieldbus protocol with PID/LM function (Optional code /LC1)
8 F9900RP 2 Cap Screw
9 Y9612YU 2 Screw
10 Below 1 Plug
F9340NW For Pg13.5
F9340NX For M20
G9330DP For G1/2
G9612EB For 1/2 NPT
11 Bellow 1 Cover Assembly
F9341FM Cast-aluminum alloy
F9341FJ SCS14A stainless steel
12 Below 1 LCD Board Assembly For integral indicator
F9342BL Without range-setting switch
F9342BM With range-setting switch
13 F9342MK 2 Mounting Screw
14 F9300PB 2 Label
Note 1 : Applicable for BRAIN and HART Protocol version (Output signal code D and E).
For FOUNDATION Fieldbus protocol version (Output signal code F), contact Yokogawa local office.
All Rights Reserved, Copyright © 1995, Yokogawa Electric Corporation. CMPL 01C22A01-02E
14th Edition: Oct. 2008(KP)
Yokogawa Electric Corporation
Customer Mode EJA310 and EJA430
Maintenance Absolute and Gauge Pressure Transmitter
Parts List (Pressure-detector Section)
15
11
18
Low pressure side
19
16
1
3 21–1
4
5 3
For EJA310 low pres-
10 6 sure side cover flange
12
22
21–2
23 13 14
All Rights Reserved, Copyright © 1999, Yokogawa Electric Corporation. CMPL 01C22D00-01E
5th Edition: Apr. 2006(YK)
Yokogawa Electric Corporation
2
14
Low pressure side
3
21–1
13 1
11–1
3
12
2
8 22
10
9 For EJA310 low pres-
23 sure side cover flange
18
21–2
20
17 15
1 — 1 Capsule Assembly (see Table 1, Table 2, Table 3 and Table 4 on page 5) (Note 1)
2 F9300AJ 1 O-Ring
3 Below 2 Gasket
F9340GA Teflon-coated SUS316L Stainless Steel (for EJA310 and EJA430 with
F9340GC Teflon-coated SUS316L Stainless Steel (degreased) Wetted Parts Material code S)
F9340GE PTFE Teflon
F9340GF PTFE Teflon (degreased) (for EJA430 with Wetted Parts Material code H, M, T, A, D and B)
4 Below 1 Cover Flange for EJA310 and EJA430 (High Pressure Side) (Note 2) (Note 4)
F9340VA Rc 1/4 SCS14A Stainless Steel (for EJA310 and EJA430 with Wetted Parts Material code S)
F9340VB 1/4 NPT For Horizontal Impulse Piping Type
F9340VC Rc 1/4 SCS14A Stainless Steel (for EJA310 and EJA430 with Wetted Parts Material code S)
F9340VD 1/4 NPT For Vertical Impulse Piping Type
F9340VP Rc 1/4 SCS14A Stainless Steel (for EJA430 with Wetted Parts Material code H, M and T)
F9340VR 1/4 NPT For Horizontal Impulse Piping Type
F9340VN Rc 1/4
SCS14A Stainless Steel (for EJA430 with Wetted Parts Material code H, M and T)
F9340VQ 1/4 NPT For Vertical Impulse Piping Type
F9340VS Rc 1/4 Hastelloy C-276 equivalent (for EJA430 with Wetted Parts Material code A and D)
F9340VT 1/4 NPT For Horizontal Impulse Piping Type
F9340VU Rc 1/4 Hastelloy C-276 equivalent (for EJA430 with Wetted Parts Material code A and D)
F9340VV 1/4 NPT For Vertical Impulse Piping Type
F9340TP Rc 1/4 Monel equivalent (for EJA430 with Wetted Parts Material code B)
F9340TR 1/4 NPT For Horizontal Impulse Piping Type
F9340TN Rc 1/4 Monel equivalent (for EJA430 with Wetted Parts Material code B)
F9340TQ 1 1/4 NPT For Vertical Impulse Piping Type
5 Below Vent Plug
F9340SA R 1/4 SUS316 Stainless Steel (for EJA310 and EJA430 with Wetted Parts
F9340SB 1/4 NPT Material code S, H, M and T)
F9340SE R 1/4
F9340SF 1/4 NPT Hastelloy C-276 (for EJA430 with Wetted Parts Material code A and D)
F9340SS R 1/4
Monel (for EJA430 with Wetted Parts Material code B)
F9340ST 1/4 NPT
6 Below 1 Vent Screw
D0114PB SUS316 Stainless Steel (for EJA310 and EJA430 with Wetted Parts Material code S, H, M and T)
F9340SK Hastelloy C-276 (for EJA430 with Wetted Parts Material code A and D)
F9270HG Monel (for EJA430 with Wetted Parts Material code B)
Apr. 2006
Subject to change without notice. Printed in Japan.
CMPL 01C22D00-01E
3
F9340SW R 1/4
D0117MS 1/4 NPT Monel (for EJA430 with Wetted Parts Material code B)
8 Below 1 Drain/Vent Plug
F9340SC R 1/4 SUS316 Stainless Steel (for EJA310 and EJA430
F9340SD 1/4 NPT with Wetted Parts Material code S, H, M and T)
F9340SG R 1/4
F9340SH 1/4 NPT Hastelloy C-276 (for EJA430 with Wetted Parts Material code A and D)
F9340SU R 1/4
F9340SV 1/4 NPT Monel (for EJA430 with Wetted Parts Material code B)
9 Below 1 Drain/Vent Screw
F9270HE SUS316 Stainless Steel (for EJA310 and EJA430 with Wetted Parts Material code S, H, M and T)
F9340SJ Hastelloy C-276 (for EJA430 with Wetted Parts Material code A and D)
D0117MR Monel (for EJA430 with Wetted Parts Material code B)
10 Below 4 Bolt
F9340AB SCM435 Chrome Molybdenum Steel
F9275EC R 1/4
F9275ED 1/4 NPT
23 F9275EE 1 Needle Assembly (degreased), SUS316 Stainless Steel
(Note 1) In case of degrease cleansing treatment (Optional code/K1 or K5), consult YOKOGAWA local office.
(However, see Table 1, Table 2, Table 3 and Table 4 in case of Optional code/K2 or K6)
(Note 2) In case of degrease cleansing treatment (Optional code/K1, K2, K5 or K6), consult YOKOGAWA local office.
(Note 3) The Plug Qty is 1 for Vertical Impulse Piping Type.
(Note 4) These Cover Flanges shall be applied to the model with the Style Code "S2" which was supplied from Aug. 16, 1996.
Apr. 2006
Subject to change without notice. Printed in Japan.
CMPL 01C22D00-01E
4
EJA430
Table 3. EJA430 Capsule Assembly Part Number (Item 1)
For General-use type, Flameproof type and Intrinsically safe type (Except TIIS Intrinsically safe type)
Installation of High Pressure Capsule Wetted Parts Material Code
Transmitter Side Code S (*1) S (*2) H, A T, D M, B
Horizontal A F9349LA F9352LA F9349LE F9349LJ F9349LN
Right
Impulse B F9349MA F9352MA F9349ME F9349MJ F9349MN
Piping Type A F9349LB F9352LB F9349LF F9349LK F9349LP
Left
B F9349MB F9352MB F9349MF F9349MK F9349MP
Vertical A F9349LC F9352LC F9349LG F9349LL F9349LQ
Right
Impulse B F9349MC F9352MC F9349MG F9349ML F9349MQ
Piping Type A F9349LD F9352LD F9349LH F9349LM F9349LR
Left
B F9349MD F9352MD F9349MH F9349MM F9349MR
Apr. 2006
Subject to change without notice. Printed in Japan.
CMPL 01C22D00-01E
REVISION RECORD
Title: Model EJA310 and EJA430 Absolute Pressure and Gauge Pressure
Transmitter
Manual No.: IM 01C22D01-01E
Edition Date Page Revised Item
9th Sep. 1998 2-14 2.10 • Delete EMC Conformity Standards Tables and move the
section to page 2-14.
2-15 2.10 • Remove Page 2-15.
8-19 8.3.2(8) • Correction made in BURN OUT figure.
11-2 11.1.2 • Add Electrical connection code 7, 8, and 9.
11-6 11.2.1 • Add Wetted parts material code M.
11-7 11.2.2 • Add Wetted parts material code M.
• Add Elactrical connection code 7, 8, and 9.
CMPL CMPL 1C22A1-02E 4th 5th
Page 2 • Add Part No. to Item 3 (For PG13.5 and M20).
• Add Part No. to Item 10 (For 1/2 NPT, Pg13.5, and M20).
CMPL 1C22E1-01E 6th 7th
Page 5 • Add Wetted parts material code M to Table 1 and 2.
10th Feb. 2000 Changed to Electronic File Format.
Revised a book in a new format.
(The location of contents and the associated page numbers may
not coincide with the one in old editions.)
Major Revised Items:
1. Explosion class and option code of JIS flameproof approval.
Explosion class: Ex ds IIC T4(old) to Ex do IIC T4X(new).
Option code: /JF1(old) to /JF3(new)
2. Option code for flameproof packing adapter for JIS flameproof
approval.
Option code: /G1 and /G2(old) to /G11 and /G12(new)
3. Add “Pa” and “hPa” as the unit for calibration range.
4. Part number change for CPU Board Assembly.
11th Sep. 2000 2-8 2.9.4.b • Change ambient temperature limit and add Caution for cable
wiring.
2-9 2.9.5.b Add Caution for /JS3
3-1 3 Add Note for /F1
8-4,8-5 8.3.1 Add parameter E10, E14 and E15. Add Note for /F1.
8-8 8.3.3(3) Add Note for /F1.
REVISION RECORD.EPS
IM 01C22D01-01E
Edition Date Page Revised Item
11th Sep. 2000 8-12 8.3.3(11)(12)
(Continued) Add Note for /F1.
8-15 8.3.3(16) Add Note for /F1.
10-1 10.1 Change the capsule damping time constant.
10-4 10.3 Add /JS3 and delete /JS1.
Change Amb.Temp. for /KF1.
CMPL CMPL 1C22A1-02E 6th 7th
• Add Parts No. to item 7-1 (For /JS3 and /F1).
• Add Parts No. to item 7-2 (For /LC1).
CMPL 1C22A1-02E 7th 8th
• Add Note for Case Assembly.
CMPL 1C22B0-01E 1st 2nd
• Change Parts No. of Capsule Assembly for JIS Intrinsically safe
type.
13th May 2002 1-2 1.1 • Add “1.1 For Safety Using.”
2-7 2.9.4 • Add descriptions based on ATEX directive.
10-4 10.3 • Add Optional code K2.
10-5 • Add Optional code C2 and C3.
• Change Optional code T02 to T12
15th Apr. 2006 1-2 1.1 • Add (e) Explosion Protected Type Instrument and (f) Modification
1-3 1.3 • Add “1.3 ATEX Document”
2-6 2.9.3 • Add “IECEx Certification” and delete “SAA Certification”
2-11 2.9.5 • JIS Certification TIIS Certification
2-12 2.12 • Add Low Voltage Directive
10-5, 10-6 10.3 • Add Certificate numbers and Applicable standards
10-6 • Add option code /SU2 and delete option code /SU1
10-7 • Add option code /PR and /N4
CMPL CMPL 01C22A01-02E 11th 01C22A01-02E 12th
JIS Intrinsically safe type TIIS Intrinsically safe type
CMPL 01C22D00-01E 4th 01C22D00-01E 5th
JIS Intrinsically safe type TIIS Intrinsically safe type
IM 01C22D01-01E
Edition Date Page Revised Item
17th Oct. 2008 2-9 2.9.4 Change explosion protection marking for type n from EEx to Ex.
2-12 2.10 Update EMC conformity standards.
7-1 7.1 Modify layout.
8-4 and 8-5 8.3.1 Add new parameters.
8-6 8.3.2 Add items in table 8.3.1.
8-9 and later 8.3.3 Add (5) Change Output Limit and (13) Span Adjustment.
Re-number the items.
8-18 8.5.2 Modify descriptions and notes for Er.01.
9-1 9.3 Add a note for calibration.
9-5 9.4.3 Add a note for cleaning.
10-3, 10-4 10.2 Add new suffix codes.
10-6, 10-7 10.3 Add Sealing statement for CSA.standards., Add /HC.
10-8 10.4 Correct errors.
CMPL CMPL 1C22A01-02E 13th 14th
Change part No .of item 5 and 8.
REVISION RECORD.EPS
IM 01C22D01-01E