AcT Service Manual
AcT Service Manual
AcT Service Manual
Service Manual
Beckman Coulter, Inc. makes no representation that, upon furnishing this service manual, the holder of the manual will have
the necessary technical capabilities and know-how to properly troubleshoot and repair any of the equipment specified in the
manual. Beckman Coulter, Inc. assumes no liability whatsoever, including consequential and incidental damages, resulting
from improper operation of Beckman Coulter instruments after maintenance of Beckman Coulter instruments has been
performed by persons not employed by Beckman Coulter, Inc. Furthermore, Beckman Coulter, Inc. assumes no liability
whatsoever for any personal injury or property damage resulting from maintenance and/or repair of Beckman Coulter
instruments performed by persons not employed by Beckman Coulter, Inc.
READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED PERSONNEL BEFORE
ATTEMPTING TO OPERATE INSTRUMENT.
Beckman Coulter, Inc. urges its customers to comply with all national health and safety standards such as the use of barrier
protection. This may include, but it is not limited to, protective eyewear, gloves, and suitable laboratory attire when
operating or maintaining this or any other automated laboratory analyzer.
"This Service Manual contains confidential information of Beckman Coulter, Inc. and its receipt or possession does not
convey any rights to reproduce, disclose its contents, or to manufacture, use, or sell anything it may describe.
Reproduction, disclosure, or use without specific written authorization of Beckman Coulter, Inc. is strictly forbidden."
Revision B, 4/97
Software Version 1.04 (8 and 10)
Numerous changes were made to this manual to reflect Software Version 1.04, and to keep it current and
easier to use. A table of specific AC•T-related problems was added. The master parts list and the Index
were updated. Actual text changes, were made on the following pages: Legal Notices, vi through viii,
1.1-14 through 1.1-3, 2.1-1, 2.6-1, 2.6-2, 5.1-1, 7.1-1, 7.1-4 through 7.1-7, 7.2-1, 7.2-1, 7.2-5, 7.2-6,
A.3-1, B.1-1, B.1-2, and Trademarks.
Revision C, 11/97
Released by CN 034130-0069
Software Version 1.03 (diff)
A complete revision was made to this manual to reflect Software Version 1.03. This includes the
AC•T diff™ Analyzer functionality and the differences between the AC•T 8/10 Analyzer and
AC•T diff Analyzer functionality.
Revision D, 4/98
Released by CN 036420-0009
Software Version 1.04 (diff and diff/Veterinary option)
Numerous changes were made to this manual to reflect Software Version 1.04 and to address any Service
Memos released since revision C. Updated sections include: Contents, ABBREVIATIONS, and Index. New
sections include: Heading 4.14, APERTURE VERIFICATION - AC•T diff/Veterinary Option Analyzer and
Appendix C.2, CANON BJC-250 BUBBLE JET PRINTER. Actual text changes were made on the following
pages: 1.1-1 through 1.1-3, 2.2-1, 2.2-3, 2.2-4, 2.3-2, 2.7-1, 2.7-3 through 2.7-6, 2.7-9, 2.7-10, 2.7-12,
2.7-13, 2.7-15, 2.8-1, 3.2-1 through 3.2-4, 4.4-2, 4.4-3, 4.10-1, 7.1-1, 7.1-17, 7.4-1, 8.1-1, 8.1-3, 8.1-4,
8.1-6, 8.1-7, 8.2-3, 8.2-5, 8.2-15 and A.3-1 through A.3-3. A new revision of the AC•T Series Analyzers,
HYDRO-PNEU, Ac.T SERIES engineering diagram, DCN 6323083, was inserted into Chapter 6.
Revision E, 2/99
Released by CN 037530-0075
Software Version 1.00
Complete revision to include hardware and software information for the AC•T diff 2™ analyzer.
Revision F, 12/01
Released by CN 037560-0170
Software Version: 2.00
The material in revision F change pages was updated for AC•T diff analyzer software version 2.00 and for
any hardware changes made since revision E, including relevant information released by the following
service memos: SM1 804, SM 1809, SM 1819, SM 1839, SM 1871, SM 1887, SM 1896, SM 1911, SM 1988,
SM 1994, SM 3013, SM 3021, SM 3023, SM 3044, SM 3048, SM 3087, SM 3091, and SM 3092. This new
information encompasses the hardware changes made to improve reliability - namely, installing a
redesigned sweep-flow assembly, replacing the single-head vacuum pump with a dual-head pump, and
installing a pair of filters in the diluent lines to the reservoir. In addition, since check digits are not used in
the VANTIVE or CARES systems, all check digits have been removed from the part numbers listed in this
manual.
This document applies to the latest software listed and higher versions. When a subsequent software version affects the
information in this document, the changes will be included on minor revision change pages or summarized on a Notice of
Information Update form and will be released by service memo.
PN 4237339F
REVISION STATUS
Note: Information from the following service memos was considered for this revision but not added
because the information is obsolete: SM 1868, SM 1922, SM 1955.
Changes were made on the following pages: 1.1-1, 1.1-3, 1.1-4, 1.1-5, 1.1-6, 1.2-2, 2.1-3, 2.2-2, 2.2-3,
2.2-4, 2.2-5, 2.2-6, 2.2-7, 2.2-8, 2.3-9, 2.3-10, 2.3-11, 2.3-12, 2.3-13, 2.3-14, 2.3-15, 2.3-16, 2.3-17, 2.6-1
through 2.6-4, 2.8-2, 2.8-8 and 2.8-9, 2.8-12 and 2.8-13, 2.8-14 and 2.8-15, 2.8-16 through 2.8-18, 2.9-1,
3.2-5, 3.2-6, 4.4-1, 4.4-2, 4.4-4, 4.5-4, 4.6-1 through 4.6-4, 4.7-1 through 4.7-4, 4.17-6, 4.15-3, 4.16-1,
4.20-2, 5.1-1, 6.1-1, 7.1-1, 7.1-6, 7.1-16, 7.1-17, 7.1-18, 7.1-22, 7.1-23, 7.2-8, 7.4-1, 8.1-1, 8.1-2, 8.1-3,
8.1-4, 8.1-6, 8.1-6, 8.1-7, 8.1-9, 8.1-9, 8.1-10, 8.1-11, 8.1-14, 8.2-1, 8.2-3, 8.2-4, 8.2-5, 8.2-9, 8.2-11,
8.2-12, 8.2-13, 8.2-14, 8.2-15, 8.2-21, 8.2-24, 8.2-25, 8.2-26, 8.2-27, 8.2-28, 8.2-29, 8.2-30, A.2-1 through
A.2-18, A.3-1 through A.3-4, A.4-1, A.4-7, A.4-10, A.4-11, A.5-2, A.5-6, A.5-8, A.5-9, A.5-10, A.5-11,
A.5-12, and TRADEMARKS page. In addition Headings 4.21, 4.22, and C.3 were added.
PN 4237339F
CONTENTS
LEGAL NOTICES
1 INTRODUCTION, 1.1-1
2.2 DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS, 2.2-1
Traverse Module, 2.2-1
Components, 2.2-1
Diluent Delivery System, 2.2-2
Lytic Reagent Delivery System, 2.2-4
Pneumatic System, 2.2-5
Vacuum Supply, 2.2-6
Probe Wipe, 2.2-6
Count, 2.2-7
Pressure Supply, 2.2-7
Aperture Sensor System, 2.2-8
Hemoglobin Photometer Assembly, 2.2-9
Waste and Cleaner System, 2.2-9
Waste System, 2.2-9
Cleaner System, 2.2-10
Peristaltic Pumps, 2.2-11
PN 4237339F
CONTENTS
2.6 ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers, 2.6-1
Power Supply Filtering and Distribution, 2.6-1
CPU and Memory, 2.6-2
Communication, 2.6-2
Display, 2.6-2
Solenoid Control, 2.6-2
Stepper Motor Control, 2.6-3
D/A Conversion, 2.6-3
Data Acquisition, 2.6-3
Jumpers, 2.6-4
vi PN 4237339F
CONTENTS
PN 4237339F
CONTENTS
4.12 LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS, 4.12-1
Purpose, 4.12-1
Tools/Supplies Needed, 4.12-1
Procedure, 4.12-1
Verification, 4.12-3
4.13 SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2
ANALYZERS, 4.13-1
Tools/Supplies Needed, 4.13-1
Procedure, 4.13-1
Verification, 4.13-2
PN 4237339F
CONTENTS
6 SCHEMATICS, 6.1-1
7 TROUBLESHOOTING, 7.1-1
PN 4237339F
CONTENTS
ABBREVIATIONS, ABBREVIATIONS-1
INDEX, INDEX-1
TRADEMARKS
PN 4237339F
CONTENTS
ILLUSTRATIONS
1.2-1 Warning and Information Labels, 1.2-2
2.1-1 User Interfaces on the AC•T 8/10 and AC•T diff Analyzers, 2.1-2
PN 4237339F
CONTENTS
3.2-4 Canon BJC-250 Bubble Jet Printer - Power/Resume Controls and Indicators, 3.2-4
4.2-1 Top Cover and Rear Panel Removal, AC•T 8/10 and AC•T diff Analyzers, 4.2-1
4.3-1 Power Supply Module Replacement, 4.3-1
4.4-1 Removing the Software (Flash Memory) Card, 4.4-1
4.4-2 Disconnecting the Analyzer Card, 4.4-2
4.4-3 Disengaging the Analyzer Card from the Mounting Posts, 4.4-3
4.5-1 Disconnecting the Cables to the Display Assembly, 4.5-1
4.5-2 Removing and Disassembling the Display Assembly, 4.5-2
4.6-1 Pneumatic Pump Cover and Bracket Removal AC•T 8/10 and AC•T diff
Analyzers, 4.6-2
4.6-2 KNF Dual-Head Pump with Muffler and Choke Installed, 4.6-3
4.7-3 KNF Dual-Head Pump with Muffler and Choke Installed, 4.7-4
4.7-5 Pneumatic Compartment Securing Screws, 4.7-6
4.8-1 Traverse Module Disassembling/Replacement, 4.8-2
4.9-1 Disconnecting the Traverse Module Flex Cable, 4.9-2
4.9-2 Disconnecting the Horizontal Motor Cable, 4.9-2
4.9-3 Disconnecting the Traverse Module Tubing, 4.9-3
4.9-4 Removing the Traverse Module, 4.9-3
4.9-5 Removing the Probe-Wipe Housing from the Traverse Housing
4.9-6 Removing the Probe from the Traverse Housing, 4.9-4
4.9-7 Removing the Belt Clamp, 4.9-5
4.9-8 Removing the Belt, 4.9-5
4.9-9 Removing the Traverse Housing, 4.9-6
4.9-10 Removing the Vertical Motor, 4.9-6 , 4.9-4
4.9-11 Removing the Traverse Interconnect Card, 4.9-7
4.9-12 Removing the Vertical Rack, 4.9-8
4.9-13 Removing the Horizontal Motor, 4.9-8
4.9-14 Removing a Horizontal Sensor, 4.9-9
4.9-15 Probe Installation, 4.9-12
4.10-1 Removing the Aperture/Bath Assembly Shield, 4.10-1
4.10-2 Removing the Hgb Photometer Assembly, 4.10-2
4.10-3 Aperture Sensor System Components, 4.10-3
4.11-1 Removing the Aperture/Bath Assembly Shield, 4.11-1
4.11-2 Removing the Hgb Photometer Assembly, 4.11-2
4.11-3 Disassembling the Hgb Photometer Assembly, 4.11-3
4.12-1 Latex Calibration Screen - AC•T diff Analyzer, 4.12-1
4.12-3 Latex Results Screen, 4.12-2
4.15-1 Hgb Photometer Shipping Posts, 4.15-3
4.15-2 Securing the Traverse Assembly for Shipping, 4.15-4
4.15-3 Preparing FMI Pumps for Shipment, 4.15-5
4.17-1 Removing the Cap-Pierce Module, 4.17-2
4.17-2 Removing the Cap-Pierce Motor, 4.17-3
4.17-3 Front Door Safety Latch Removal, 4.17-4
4.17-4 Disassembling the Cap-Pierce Door Assembly, 4.17-5
PN 4237339F
CONTENTS
PN 4237339F
CONTENTS
TABLES
1.1-1 AC•T Analyzer Customer Documents, 1.1-1
2.1-1 AC•T Series Analyzers - System Specifics, 2.1-3
2.4-1 Switching Supply Specifications , 2.4-1
2.4-2 Power Supply Module - Comparison of Unique Features, 2.4-3
2.5-1 Stepper Motors, AC•T 8/10 Analyzer - Associated Chips, 2.5-4
2.5-2 Solenoids Controlled by Solenoid Driver Circuitry, AC•T 8/10 Analyzer, 2.5-5
2.6-1 Stepper Motor Function Summary, 2.6-3
2.7-1 Touch Screen Specifications, 2.7-2
2.7-2 Touch Screen Connections to Display Controller Driver Card, 2.7-2
2.7-3 LCD Connector CN1, 2.7-3
2.7-4 LCD Connector CN2, 2.7-3
2.7-5 Display Controller Driver Card - J61 Pinout, 2.7-4
2.7-6 Display Controller Driver Card Connectors, 2.7-5
2.7-7 Display Controller Driver Card Jumpers, 2.7-6
2.7-8 Display Controller Driver Card Test Points, 2.7-6
2.8-1 Powerup Cycle - AC•T 8/10 and AC•T diff Analyzers, All Software Versions, 2.8-3
2.8-2 Powerup Cycle - AC•T diff 2 Analyzer, 2.8-4
2.8-3 Whole-Blood Aspirate Cycle - AC•T 8 Analyzer, Software Version 1.03, 2.8-6
2.8-4 Whole-Blood Aspirate Cycle - AC•T 8 and AC•T 10 Analyzers,
Software Version 1.04 and 2.00, 2.8-8
2.8-5 Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Version 1.03, 2.8-10
2.8-6 Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Versions 1.04, 1.05,
1.06, and 2.00, 2.8-12
2.8-7 Whole-Blood Aspirate Cycle - AC•T diff Analyzer with Veterinary Option,
Software Versions 1.04, 1.05, 1.06, and 2.00, 2.8-14
2.8-8 Closed-Vial Aspirate Cycle - AC•T diff 2 Analyzer, Software Version 1.00 and
2.00, 2.8-16
3.1-1 AC•T Analyzer Installation Documents, 3.1-1
3.2-1 Canon BJC-250 Bubble Jet Printer - Available Printer Information/Tests, 3.2-4
4.12-1 Latex Adjustment Criteria, 4.12-2
4.13-1 Target Acceptance Criteria, 4.13-1
5.1-1 Recommended Maintenance Schedule - AC•T 8/10 and AC•T diff Analyzers, 5.1-1
5.1-2 Recommended Maintenance Schedule - AC•T diff 2 Analyzers, 5.1-1
7.1-1 Solenoid Drivers and Related Components, 7.1-7
7.1-2 RBC Clog Test Results, 7.1-17
7.1-3 WBC Clog Test Results, 7.1-17
7.1-4 Motors - Function, Action and Description, AC•T 8/10 and AC•T diff
Analyzers, 7.1-18
7.1-5 Motors - Function, Action and Description, AC•T diff 2 Analyzer, 7.1-19
7.2-1 Warning (Non-Fatal) Error Messages (Icons), 7.2-1
7.2-2 Fatal Error Messages (Numbers), 7.2-2
7.2-3 Allowable Voltage Ranges - AC•T 8/10 Analyzer, 7.2-5
7.2-4 Allowable Voltage Ranges - AC•T diff Analyzer, 7.2-5
7.2-5 Allowable Voltage Ranges - AC•T diff 2 Analyzer, 7.2-5
7.4-1 Specific AC•T 8/10 and AC•T diff Analyzer Problems/Solutions, 7.4-1
8.1-1 Components, 8.1-1
8.1-2 Fasteners, Fittings, and Miscellaneous Hardware, 8.1-8
8.1-3 Optional Items and Supplies, 8.1-14
PN 4237339F
CONTENTS
8.2-1 Triple Syringe Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-1), 8.2-1
8.2-2 Front Panel Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-2), 8.2-3
8.2-3 Front Panel Components, AC•T diff 2 Analyzers (See Figure 8.2-3), 8.2-5
8.2-4 Front Door Components (See Figure 8.2-4), 8.2-7
8.2-5 Right-Side Compartment, Reservoir Components, AC•T 8/10 and AC•T diff
Analyzers (See Figure 8.2-5), 8.2-9
8.2-6 Right-Side Compartment, Transducer Components, AC•T 8/10 and AC•T diff
Analyzers (See Figure 8.2-6), 8.2-11
8.2-7 Right-Side Compartment, Reservoir and Reagent Pump Components, AC•T diff 2
Analyzers (See Figure 8.2-7), 8.2-13
8.2-8 Right-Side Compartment, Vacuum Pump Components, AC•T diff 2 Analyzers
(See Figure 8.2-8), 8.2-15
8.2-9 Rear View Showing Components (See Figure 8.2-9), 8.2-17
8.2-10 Traverse Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-10), 8.2-19
8.2-11 Traverse Module Components, AC•T diff 2 Analyzers (See Figure 8.2-11), 8.2-21
8.2-12 Cap-Pierce Module Components, AC•T diff 2 Analyzers (See
Figure 8.2-12), 8.2-23
8.2-13 Left Side and Top Views (See Figure 8.2-13), 8.2-25
8.2-14 Aperture/Bath Assembly Components on Mounting Plate, AC•T 8/10 and AC•T
diff Analyzers (See Figure 8.2-14), 8.2-27
8.2-15 Aperture/Bath Assembly Components on Mounting Plate, AC•T diff 2 Analyzers
(See Figure 8.2-15), 8.2-29
8.2-16 Aperture/Bath Assembly Shield and Mounting Plate (See Figure 8.2-16), 8.2-30
A.1-1 Analyzer Card - AC•T 8/10 Analyzer, Voltage Limits, A.1-1
A.1-2 Analyzer Card - AC•T diff Analyzer, Voltage Limits, A.1-1
A.1-3 Analyzer Card - AC•T diff 2 Analyzer, Voltage Limits, A.1-1
A.1-4 Target Acceptance Criteria, A.1-1
A.1-5 Latex Adjustment Criteria, A.1-2
A.1-6 Power Supply Module Output, Connector P5 - AC•T 8/10 Analyzer, A.1-2
A.1-7 Power Supply Module Output, Connector P5 - AC•T diff and AC•T diff 2
Analyzers, A.1-2
A.2-1 Analyzer Card Connectors - AC•T 8/10 Analyzers (See Figure A.2-1), A.2-2
A.2-2 Analyzer Card Test Points - AC•T 8/10 Analyzers (See Figure A.2-1), A.2-4
A.2-3 Analyzer Card Connectors - AC•T diff Analyzer (See Figure A.2-2), A.2-6
A.2-4 Analyzer Card Jumpers - AC•T diff Analyzer (See Figure A.2-2), A.2-7
A.2-5 Analyzer Card Test Points - AC•T diff Analyzer (See Figure A.2-2), A.2-8
A.2-6 Analyzer Card Connectors - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-10
A.2-7 Analyzer Card Jumpers - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-12
A.2-8 Analyzer Card Test Points - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-12
A.2-9 Universal AC•T diff/AC•T diff 2Analyzer Card Connectors (See
Figure A.2-4), A.2-14
A.2-10 Universal AC•T diff/AC•T diff 2 Analyzer Card Jumper Settings (See
Figure A.2-4), A.2-17
A.2-11 Universal AC•T diff/AC•T diff 2Analyzer Card Test Points (See
Figure A.2-4), A.2-18
A.2-12 Display Controller Driver Card Connectors, A.2-20
A.2-13 Display Controller Driver Card - J61 Pinouts, A.2-21
PN 4237339F
CONTENTS
PN 4237339F
CONTENTS 1
1 INTRODUCTION, 1.1-1
ILLUSTRATIONS
1.2-1 Warning and Information Labels, 1.2-2
TABLES
1.1-1 AC•T Analyzer Customer Documents, 1.1-1
PN 4237339F
1INTRODUCTION 1
1.1 MANUAL DESCRIPTION
Scope
This manual provides the reference information and procedures needed for servicing and
maintaining COULTER AC•T™ analyzers (hereafter referred to as the AC•T analyzer or the
instrument). It is available both online and in hard copy. The online manual is released on the
Service Resource Kit CD-ROM, PN 6417471.
This manual is to be used in conjunction with the AC•T analyzer customer documents and
does not contain information and procedures already covered in those documents.
Table 1.1-1 lists the customer documents available.
PN 4237339F
INTRODUCTION
MANUAL DESCRIPTION
AC•T diff analyzer Installation and Training Guide for the AC•T Pak™ Reagent English 4237421
Chinese 4237444
French 4237432
German 4237436
Italian 4237440
Spanish 4237428
Installation and Training Guide for the AC•T Tainer™ Reagent English 4237417
Chinese 4237445
French 4237433
German 4237437
Italian 4237441
Japanese 4237449
Spanish 4237429
Operating Summary English 4237420
Chinese 4237446
French 4237434
German 4237438
Italian 4237442
Japanese 4237450
Spanish 4237430
Operator's Guide English 4237416
Chinese 4237447
French 4237435
German 4237439
Italian 4237443
Japanese 4237451
Spanish 4237431
Reference Manual English 4237422
Veterinary Applications Software Options Operator’s Guide English 4237400
PN 4237339F
INTRODUCTION
MANUAL DESCRIPTION 1
Table 1.1-1 AC•T Analyzer Customer Documents (Continued)
AC•T diff 2 analyzer Installation and Training Guide for the diff AC•T Pak™ Reagent and the English 4237517
diff AC•T Tainer™ Reagent French 4237553
German 4237544
Italian 4237561
Japanese 4237556
Spanish 4237550
Portuguese 4237547
Operating Summary English 4237516
French 4237551
German 4237557
Italian 4237562
Japanese 4237554
Spanish 4237548
Portuguese 4237545
Operator's Guide English 4237495
French 4237552
German 4237543
Italian 4237563
Japanese 4237555
Spanish 4237549
Portuguese 4237546
Reference Manual English 4237515
Printers AC•T diff Analyzer Canon® Bubble Jet™ Printer User’s Guide English 4237479
AC•T diff Analyzer Graphics Printer User’s Guide English 4237343
AC•T Series Analyzer Roll Printer User’s Guide English 4237381
AC•T Series Analyzer Ticket Printer User's Guide English 4237380
Notification of Updates
Any service memo that affects the information in this manual will include either change pages
or a Notice of Information Update form for this manual. A Notice of Information Update form
will summarize the changes and list the specific headings, figures, and tables affected.
PN 4237339F
INTRODUCTION
MANUAL DESCRIPTION
Intended Audience
To use this manual effectively, you need the following:
Organization
The material in this manual is organized into eight chapters and three appendices. To make it
easier to access the information:
Chapter 6, SCHEMATICS - A list of the engineering schematics needed for servicing the
AC•T analyzer.
PN 4237339F
INTRODUCTION
MANUAL DESCRIPTION 1
Chapter 7, TROUBLESHOOTING - An in-depth troubleshooting table.
Chapter 8, PARTS LISTS - The master parts list followed by the illustrated parts list.
Numbering Format
Each chapter of this manual is further divided into topics that are numbered sequentially,
beginning at one. The numbering format for the topic heading, which is called the primary
heading, is chapter number, decimal point, topic number. For example, the primary heading
number for the fourth topic covered in Chapter 2 is 2.4.
The page, figure, and table numbers are tied directly to the primary heading number. For
example, Heading 2.4 begins on page 2.4-1, the first figure under Heading 2.4 is Figure 2.4-1
and the first table under Heading 2.4 is Table 2.4-1.
Special Headings
Throughout this manual, WARNING, CAUTION, IMPORTANT, ATTENTION, and Note
headings are used to indicate potentially hazardous situations and important or helpful
information.
WARNING
A WARNING indicates a situation or procedure that, if ignored, can cause serious personal
injury. The word WARNING is in bold-faced text in the printed manual and is red in the
electronic manual.
CAUTION
A CAUTION indicates a situation or procedure that, if ignored, can cause damage to the
instrument.The word CAUTION is in bold-faced text in the printed manual and is red in the
electronic manual.
IMPORTANT
An IMPORTANT indicates a situation or procedure that, if ignored, can result in erroneous
test results.The word IMPORTANT is in bold-faced text in the printed manual and is red in
the electronic manual.
PN 4237339F
INTRODUCTION
MANUAL DESCRIPTION
ATTENTION
An ATTENTION contains information that is critical for the successful completion of a
procedure and/or operation of the instrument.The word ATTENTION is in bold-faced text in
the printed manual and is red in the online manual.
Note
A Note contains information that is important to remember or helpful in performing a
procedure.
Conventions
1. The AC•T analyzer uses icons exclusively. In this document however, text equivalents
(appearing in bold) are substituted for icons. Refer to Figures A.3-1, A.3-2, A.3-3, or
A.3-4 for the icons (along with their text equivalents) used to access menus and screens.
2. Select menu item tt sub-menu item indicates the software options you have to select, as
well as the order in which you should select them. For example:
When the Powerup cycle is completed, select Main icon tt Diagnostics icon tt
Voltages/Sensors icon.
3. In the electronic manual, links to additional information are in blue and are underlined.
To access the linked information, select the blue, underlined text, or in the case of a
graphic, the blue box.
4. A reference to AC•T Series analyzers includes the AC•T 8 analyzer, the AC•T 10
analyzer, the AC•T diff analyzer, the AC•T diff with Veterinary Option analyzer, and the
AC•T diff 2 analyzer.
5. The AC•T 8 analyzer and AC•T 10 analyzer are two different products. However, since
these analyzers are physically identical with only slight differences in their software, the
product names are combined in this document. As a result, a reference to the AC•T 8/10
analyzer means the stated information or procedure applies to both the AC•T 8 analyzer
and AC•T 10 analyzer.
PN 4237339F
INTRODUCTION
SAFETY PRECAUTIONS 1
1.2 SAFETY PRECAUTIONS
Electronic
WARNING Risk of personal injury. Contacting exposed electronic components while the instrument is
attached to power can cause personal injury from electric shock. Power down completely before removing
covers to access electronic components.
WARNING Risk of personal injury or damage to electronic components. While performing maintenance or
service on the instrument, rings and other metal jewelry can become caught in the instrument. To avoid
personal injury or damage to the instrument, remove rings and other metal jewelry before performing
maintenance or service on the electronic components of the instrument.
CAUTION Risk of damage to electronic components. If the power is ON while removing or replacing printed
circuit cards and components, the instrument could be damaged. To prevent damage to electronic
components, always be sure power is OFF before removing or replacing printed circuit cards and
components.
CAUTION Risk of damage to electronic components. Electrostatic discharge (ESD) can damage add-in
circuit cards and other electronic components. If there is a possibility of ESD damage with a procedure,
then perform that procedure at an ESD workstation, or wear an antistatic wrist strap attached to a metal part
of the chassis connected to an earth ground.
Biological
WARNING Risk of personal injury or contamination. If you do not properly shield yourself while servicing
the instrument with the doors open, you may become injured or contaminated. To prevent possible injury or
biological contamination, you must wear appropriate safety glasses, a lab coat, and gloves when servicing
the instrument with the doors open.
Use care when working with pathogenic materials. Means must be available to decontaminate
the instrument, provide ventilation, and to dispose of waste liquid. Refer to the following
publications for further guidance on decontamination:
Troubleshooting
Bring the following Warning to the customer’s attention before advising that customer to
perform any service, maintenance or troubleshooting procedures on the AC•T analyzer. Also,
make sure customers are aware of the Warning and information labels shown in Figure 1.2-1.
PN 4237339F
INTRODUCTION
SAFETY PRECAUTIONS
WARNING Risk of personal injury or contamination. If you do not properly shield yourself while performing
service, maintenance, and troubleshooting procedures, residual fluids in the instrument could injure or
contaminate you. Beckman Coulter recommends that you wear barrier protection, such as appropriate
safety glasses, a lab coat, and gloves throughout the performance of service, maintenance, and
troubleshooting procedures to avoid contact with cleaners and residual fluids in the instrument.
2 1
MOD ASSY S/N
NO. Ac*T NO. xxxxxx xxxxxx
240/120 50/60 0.5/1.0 120 ELECTRICAL POWER
VOLTS HZ AMPS WATTS
SUPPLY AC INPUTS
MANUFACTURED BY COULTER CORPORATION
11800 S.W. 147 AVENUE, MIAMI, FL 33196-2500 U.S.A. 90 - 132v / 198 - 264v
PATENTS ISSUED AND / OR PENDING ETL LISTED
ETECTRICAL EQUIPMENT
FOR LABORATORY USE;
PART 1: GENERAL
LISTED R REQUIREMENTS
26785 CONFORMS TO UL Std 3101-1
ETL TESTING LABORATORIES, INC.
M AN UF AC TU R ED BY COULTER CORPORATION
11 80 0 S .W . 1 47 AV EN U E, MIA M I, F L 331 96 -25 00 U .S.A.
WARNING
TO AVOID ELECTRIC CONNECT ONLY
SHOCK DISCONNECT TO A PROPERLY
J1
POWER CORD PRIOR TO EARTH GROUNDED
ON
F1 F2
AC INPUT
ELECTRICAL POWER
SUPPLY AC INPUTS
REMOVING OR REPLACING OUTLET .
OFF
FUSE 3AG
90 - 132v / 198 - 264v
FUSE .
1 . 5A 250V
!
WARNING
TO AVOID ELECTRIC
SHOCK DISCONNECT
CONNECT ONLY
TO A PROPERLY
REPLACE FUSE ONLY
WITH THE TYPE AND
POWER CORD PRIOR TO EARTH GROUNDED
REMOVING OR REPLACING OUTLET .
ELECTRICAL POWER
SUPPLY AC INPUTS
MOD
NO.
VOLTS
ASSY
NO.
HZ
M AN UF AC TU RE D B Y COULTER CORPORATION
AMPS
S/N
WATTS
90 - 132v / 198 - 264v
11 80 0 S .W . 14 7 A VE NU E , MIAM I, F L 331 96 -25 00 U.S.A .
WARNING
FUSE 3AG TO AVOID ELECTRIC CONNECT ONLY
2. 5A 250V SHOCK DISCONNECT TO A PROPERLY
J1
POWER CORD PRIOR TO EARTH GROUNDED
ON
F1 F2
AC INPUT
ELECTRICAL POWER
REMOVING OR REPLACING OUTLET .
SUPPLY AC INPUTS
OFF
FUSE 3AG
90 - 132v / 198 - 264v
FUSE .
LEAKAGE CURRENT COMPLIES 2 . 5A 250V
FUSE .
REPLACE FUSE ONLY
WITH THE TYPE AND
EARTH GROUNDED
REMOVING OR REPLACING OUTLET .
WITH THE TYPE AND
DATE
INSP BY
DATE
RATING SPECIFIED .
RATING SPECIFIED .
PN 4237339F
CONTENTS 2
2 INSTRUMENT DESCRIPTION, 2.1-1
2.2 DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS, 2.2-1
Traverse Module, 2.2-1
Components, 2.2-1
Diluent Delivery System, 2.2-2
Lytic Reagent Delivery System, 2.2-4
Pneumatic System, 2.2-5
Vacuum Supply, 2.2-6
Probe Wipe, 2.2-6
Count, 2.2-7
Pressure Supply, 2.2-7
Aperture Sensor System, 2.2-8
Hemoglobin Photometer Assembly, 2.2-9
Waste and Cleaner System, 2.2-9
Waste System, 2.2-9
Cleaner System, 2.2-10
Peristaltic Pumps, 2.2-11
Waste System, 2.3-16
Cleaner System, 2.3-17
FMI Pump Operation, 2.3-17
Pump Components, 2.3-18
Crossover Position from Output to Intake, 2.3-18
Intake Position, 2.3-19
Crossover Position from Intake to Output, 2.3-19
Additional Points about the FMI Pumps, 2.3-20
Output Position, 2.3-20
2.4 POWER SUPPLY MODULE, 2.4-1
Components, 2.4-1
Module Differentiation, 2.4-2
, 2.4-2
Power Harness and Connectors
Fuses, 2.4-3
Appearance, 2.4-3
Comparison, 2.4-3
2.6 ANALYZER CARDS - AC••T T diff and AC••TT diff 2 Analyzers
Analyzers,, 2.6-1
Power Supply Filtering and Distribution
Distribution,, 2.6-1
CPU and Memory
Memory,, 2.6-2
Communication,, 2.6-2
Communication
Display,, 2.6-2
Display
Solenoid Control
Control,, 2.6-2
Stepper Motor Control
Control,, 2.6-3
D/A Conversion
Conversion,, 2.6-3
Data Acquisition
Acquisition,, 2.6-3
Jumpers,, 2.6-4
Jumpers
PN 4237339F
CONTENTS
2
2.7 DISPLAY ASSEMBLY
ASSEMBLY,, 2.7-1
Main Components
Components,, 2.7-1
Touch Screen
Screen,, 2.7-2
LCD,, 2.7-2
LCD
Display Controller Driver Card, 2.7-3
Function, 2.7-3
Differences Between the Old and New Versions of the Display Controller
Driver Card, 2.7-5
2.8 SOFTWARE STRUCTURE, 2.8-1
Overview, 2.8-1
Menu Trees, 2.8-1
Instrument Cycles, 2.8-1
Powerup Cycle, 2.8-2
Aspirate Cycle, 2.8-5
2.9 INSTRUMENT STATES, 2.9-1
ILLUSTRATIONS
2.1-1 User Interfaces on the AC•T 8/10 and AC•T diff Analyzers, 2.1-2
2.1-2 User Interfaces on the AC•T diff 2 Analyzer, 2.1-2
2.2-1 Traverse Module - Components, 2.2-1
2.2-2 Diluent Delivery System in AC•T 8/10 and AC•T diff Analyzers - Components and
Flow, 2.2-3
2.2-3 Lytic Reagent Delivery System in AC•T 8/10 and AC•T diff Analyzers-
Components and Flow, 2.2-4
2.2-4 Pneumatic System in AC•T 8/10 and AC•T diff Analyzers - Vacuum Components
and Flow, 2.2-5
2.2-5 Vacuum Chamber (VC1) System in AC•T 8/10 and AC•T diff Analyzer-
Components and Flow, 2.2-7
2.2-6 Pneumatic System in AC•T 8/10 and AC•T diff Analyzers- Pressure Components
and Flow, 2.2-8
2.2-7 Waste/Cleaner System in AC•T 8/10 and AC•T diff Analyzers - Components and
Flow, 2.2-10
2.2-8 Peristaltic Pump Assembly, 2.2-11
2.3-1 Traverse Module - Main Components, 2.3-2
2.3-2 Vertical Positioning of the Probe - Components, 2.3-3
2.3-3 Cap-Pierce Module - Main Components, 2.3-4
2.3-4 Starting the Cycle (right side view), 2.3-5
2.3-5 Vent Pierce (right side view), 2.3-6
2.3-6 Aspiration Pierce - Typical (right side view), 2.3-6
2.3-7 Bottom Sense Function - Components
(right side view), 2.3-7
2.3-8 Aspiration Pierce Using Bottom Sensing (right side view), 2.3-8
2.3-9 Diluent Delivery System in the AC•T diff 2 Analyzer - Components and
Flow, 2.3-9
2.3-10 Lytic Reagent Delivery System in the AC•T diff 2 Analyzer - Components and
Flow, 2.3-11
2.3-11 Pneumatic System in the AC•T diff 2 Analyzer - Vacuum Components and
Flow, 2.3-12
2.3-12 Vacuum Chamber (VC1) System in the AC•T diff 2 Analyzer - Components and
Flow, 2.3-13
2.3-13 Pneumatic System in the AC•T diff 2 Analyzer - Pressure Components and
Flow, 2.3-14
PN 4237339F
CONTENTS
2.3-14 Waste and Cleaner System in the AC•T diff 2 Analyzer - Components and
Flow, 2.3-16
2.3-15 FMI Pump Components, 2.3-18
2.3-16 Crossover from Output to Intake, 2.3-18
2.3-17 Intake Position, 2.3-19
2.3-18 Crossover from Intake to Output, 2.3-19
2.3-19 Output Position, 2.3-20
2.4-1 Power Supply Module - Main Components, 2.4-2
2.7-1 Display Assembly Components, 2.7-1
TABLES
2.1-1 AC•T Series Analyzers - System Specifics, 2.1-3
2.4-1 Switching Supply Specifications , 2.4-1
2.4-2 Power Supply Module - Comparison of Unique Features, 2.4-3
2.5-1 Stepper Motors, AC•T 8/10 Analyzer - Associated Chips, 2.5-4
2.5-2 Solenoids Controlled by Solenoid Driver Circuitry, AC•T 8/10 Analyzer, 2.5-5
2.6-1 Stepper Motor Function Summary, 2.6-3
2.7-1 Touch Screen Specifications, 2.7-2
2.7-2 Touch Screen Connections to Display Controller Driver Card, 2.7-2
2.7-3 LCD Connector CN1, 2.7-3
2.7-4 LCD Connector CN2, 2.7-3
2.7-5 Display Controller Driver Card - J61 Pinout, 2.7-4
2.7-6 Display Controller Driver Card Connectors, 2.7-5
2.7-7 Display Controller Driver Card Jumpers, 2.7-6
2.7-8 Display Controller Driver Card Test Points, 2.7-6
2.8-1 Powerup Cycle - AC•T 8/10 and AC•T diff Analyzers, All Software Versions, 2.8-3
2.8-2 Powerup Cycle - AC•T diff 2 Analyzer, 2.8-4
2.8-3 Whole-Blood Aspirate Cycle - AC•T 8 Analyzer, Software Version 1.03, 2.8-6
2.8-4 Whole-Blood Aspirate Cycle - AC•T 8 and AC•T 10 Analyzers,
Software Version 1.04 and 2.00, 2.8-8
2.8-5 Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Version 1.03, 2.8-10
2.8-6 Whole-Blood Aspirate Cycle - AC•T diff Analyzer,
Software Versions 1.04, 1.05, 1.06, and 2.00, 2.8-12
2.8-7 Whole-Blood Aspirate Cycle - AC•T diff Analyzer with Veterinary Option,
Software Versions 1.04, 1.05, 1.06, and 2.00, 2.8-14
2.8-8 Closed-Vial Aspirate Cycle - AC•T diff 2 Analyzer, Software Version 1.00 and
2.00, 2.8-16
PN 4237339F
2INSTRUMENT DESCRIPTION 2
2.1 SYSTEM OVERVIEW
The AC•T analyzer product line consists of four small, uncomplicated, low cost, hematology
analyzers: the AC•T 8 analyzer, the AC•T 10 analyzer, the AC•T diff analyzer, and the
AC•T diff 2 analyzer. The AC•T 8/10 and the AC•T diff analyzers use the same diluter but
different electronics and software. The AC•T diff and AC•T diff 2 analyzers use very similar
electronics and software, but different diluters. All four models accept open-vial whole-blood
or prediluted samples. The AC•T diff 2 analyzer also has a Cap-Pierce mode. Table 2.1-1
provides a quick reference to instrument specific information concerning each AC•T system.
Hardware
The instrument has an onboard switching Power Supply module rated at 50/60 Hz with a
voltage range of 90 to 264 Vac. The electronics of the instrument are housed on one circuit
card, the Analyzer card, although different models use different Analyzer cards. The Analyzer
cards for the AC•T 8/10 analyzers use through-hole card technology. The Analyzer cards for
the AC•T diff and AC•T diff 2 analyzers use surface-mount technology (SMT).
Fluidic movement on the AC•T 8/10 and AC•T diff analyzers is controlled by the Triple
Syringe module using one motor to drive three different volume syringes and two peristaltic
pumps. Fluidic movement on the AC•T diff 2 analyzer is controlled by FMI metering pumps
or piston/valve pumps.
Three external electrical connectors are provided, one for the waste sensor, one for a parallel
Printer, and one for ASTM host communication.
Software
The instruments use 8 bit microprocessors as their main controllers. Software is stored and
runs directly from Flash Memory cards. The AC•T 8/10 analyzer has a 512 KB Flash Memory
card while the AC•T diff and AC•T diff 2 analyzers use a 1 MB Flash Memory card. Both
models use a 128 KB, non-volatile RAM chip to store persistent data such as the calibration
factors and the user-configuration settings. The Analyzer card for the AC•T 8/10 analyzer has
no additional memory, so the unused portion of non-volatile RAM is used as program runtime
memory. The Analyzer cards for the AC•T diff and AC•T diff 2 analyzers have 256 KB RAM
that is used during program execution.
You also interact with the instrument using the touch screen (Figures 2.1-1 and 2.1-2). To
communicate commands to the instrument, the touch screen displays a series of screens
containing icons and numbers for your use. Screens are available for initiating sample runs,
displaying sample results, establishing instrument settings, and performing various
maintenance routines.
PN 4237339F
INSTRUMENT DESCRIPTION
SYSTEM OVERVIEW
Figure 2.1-1 User Interfaces on the AC•T 8/10 and AC•T diff Analyzers
Touch
screen
Aspirate
switch 7339011E
PN 4237339F
INSTRUMENT DESCRIPTION
SYSTEM OVERVIEW 2
Table 2.1-1 AC•T Series Analyzers - System Specifics
AC•T 8 Analyzer AC•T 10 Analyzer AC•T diff Analyzer AC•T diff 2 Analyzer
Operating Modes Whole-Blood Aspirate Whole-Blood Aspirate Whole-Blood Aspirate Closed-Vial Whole-Blood
Predilute Predilute Predilute Open-Vial Whole-Blood
Predilute
Parameters WBC WBC WBC WBC
RBC RBC RBC RBC
Hgb Hgb Hgb Hgb
Hct Hct Hct Hct
MCV MCV MCV MCV
MCH MCH MCH MCH
MCHC MCHC MCHC MCHC
Plt Plt RDW RDW
LY% Plt Plt
LY# MPV MPV
Pct* Pct*
PDW* PDW*
LY% LY%
MO% MO%
GR% GR%
LY# LY#
MO# MO#
GR# GR#
Software See Figure A.3-2
Menu Trees (Figure A.3-3 for system
with Veterinary Option)
Diluter Operation See Heading 2.2
Power Supply See Heading 2.4
Analyzer Card See Heading 2.6
Display Assembly See Heading 2.7
Powerup Cycle See Table 2.8-1
Aspirate Cycle Software Version 1.03,
see Table 2.8-5
Software Versions 1.04,
1.05, 1.06, or 2.00, see
Table 2.8-6
Veterinary Option for
Software Version 1.04,
1.05, 1.06, or 2.00, see
Table 2.8-7
* Pct and PDW are not FDA approved parameters; therefore, these parameter labels and results are not routinely
displayed in the United States. The PDW result, however, is used as an internal check in the platelet
algorithm.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
2.2 DILUTER SYSTEM - AC•T 8/10 AND AC•T diff ™ ANALYZERS
The AC•T 8/10 and AC•T diff analyzers use different Analyzer cards and software, but the
same diluter. This diluter uses two peristaltic pumps and three syringes to move liquids.
Location diagrams for the fluidic components identified in the following descriptions are
under Heading A.4, AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND
FUNCTIONS.
Traverse Module
The Traverse module acts as a transport system for the probe. Two stepper motors provide the
power that allows the Traverse module to move the probe either vertically or horizontally as
needed. Two dual-channel opto sensors control the positioning of the probe so that it is
properly aligned, as needed, at the aspiration station, the WBC bath, or the RBC bath.
Components
The Traverse module (Figure 2.2-1) consists of:
r The probe
r The traverse housing
r The probe-wipe housing
r Two stepper motors
r The Traverse Interconnect card
r Two dual-channel opto sensors
Vertical
stepper
Traverse motor
housing
Horizontal
stepper
motor
Probe-wipe
housing
Traverse
Interconnect
Probe
card 7339140E
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS
For horizontal travel, the probe has three positions: the aspirate station, over the WBC bath,
and over the RBC bath. The probe also must position itself at three heights: the top position
while travelling, the thief position (aspiration for the RBC dilution while in the WBC bath),
and the bottom position for sample aspiration.
Two stepper motors move the probe. One motor moves it vertically and the other moves it
horizontally. A direct drive gear and rack system is employed for both motors. Two
dual-channel opto sensors, one associated with each motor, detect the probe positions. Using
these sensors with binary encoding allows detection of three possible positions in each
direction with one sensor.
When a function such as aspiration is complete, the probe moves into the traverse housing
where the probe-wipe housing surrounds the tip of the probe and a flow of diluent cleanses
the outside of the probe and high vacuum dries it.
Peristaltic pump PM2 moves diluent from the reagent source to the diluent reservoir via two
hydraulic filters, FLS1 and FLS3; a solenoid, LV10; and two hydrophilic filters, FLS4 and
FLS5. See Figure 2.2-2 for a simplified diagram of the diluent system.
Note: For details concerning the operation of peristaltic pumps, see Peristaltic Pumps at the
end of this section.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
Figure 2.2-2 Diluent Delivery System in AC•T 8/10 and AC•T diff Analyzers - Components and Flow
FLS1
1 2
Diluent 1 2
FLS3
PM2 1 2 1 LV10
1
2 3 2
Probe-wipe 1
housing 2 3
3
1 1
LV7 LV11 Diluent
1 3 2 FLS5
Probe reservoir FS1 2 FLS4 2
3 2
2 3 1 3 2 1 1
3 2 LV12
Sweep
WBC RBC 1 flow
bath bath
FLS1 and FLS3 filter any large particles from the diluent source as well as from peristaltic
pump tubing wear. They are directional and fluid should enter the blue (striped) side of the
filter. LV10 normally connects PM2 output to the reservoir, but when energized, connects
PM2 to the top port of the probe-wipe housing. This is the diluent supply for cleansing the
probe. FLS4 and FLS5 remove smaller contaminating particles, such as some bacteria. (Filters
that are small enough to eliminate all possible bacteria would restrict diluent flow too much.)
Two filters are used in parallel to facilitate initial priming.
LV12 controls the reservoir diluent usage. In its normal state LV12 connects the reservoir to
the 5-mL diluent syringe. This provides diluent while the syringe is being filled, and also
allows delivery of diluent back to the reservoir when only lytic reagent dispense or aspirate
functions are required. A separate port on the diluent reservoir provides sweep-flow diluent
to the RBC aperture through the sweep-flow coil and a check valve.
LV7 and LV11 control diluent output from the syringes. When LV11 is in its normal state, the
diluent syringe dispenses through the aspirate probe. When energized, LV11 routes diluent
from the syringe to LV7, which routes diluent to lower-side ports on the baths. Diluent is
routed to these ports to rinse the baths and to prefill each bath prior to making the dilution.
The remainder of the diluent needed to make the dilution is dispensed through the probe
with the whole-blood sample. In its normal state LV7 is connected to the WBC bath. When
energized, LV7 switches to the RBC bath.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS
Figure 2.2-3 Lytic Reagent Delivery System in AC•T 8/10 and AC•T diff Analyzers- Components and Flow
W B C
F S 2 b a th
1 2
3 2 L V 9
L y tic
re a g e n t
1
3 2 L V 4
L y s e s y r in g e
(1 m L )
1
A T M
3 2 L V 5
P n e u m a tic
p u m p
7 3 3 9 0 4 1 F
Two features of the AC•T analyzer’s lytic reagent delivery system distinguish it from previous
Beckman Coulter hematology analyzers. Using a syringe allows the system software to control
the amount of lytic reagent dispensed. The AC•T analyzer takes advantage of this versatility,
using different amounts of lytic reagent during different cycles. A normal Aspirate cycle uses
415 µL of lytic reagent, the same amount that is dispensed during a Lyse Dispense Verification
cycle and a Prime cycle. The Predilute cycle uses 313 µL while the Lyse Prime cycle (which
assumes a lytic reagent dry state) uses the full 1-mL syringe stroke several times.
The other difference in the AC•T analyzer’s lytic reagent delivery system results from the use
of the Syringe module. Since one motor moves three syringes simultaneously, there are
instances when diluent is being dispensed and lytic reagent delivery is not desired. At these
times, lytic reagent is pushed back toward the reagent container.
An optical liquid sensor, FS2, in the lytic reagent line monitors the lytic reagent supply as it
enters the instrument. Lytic reagent in the sensor transmits light, but air in the sensor
prevents light transmission. The presence of air gaps in the sensor triggers a lytic reagent low
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
warning. A software filter is applied to the data received from the sensor to ensure that
microbubbles, inconsequential in their affect on the lytic reagent delivery volume, do not
trigger the warning.
Pneumatic System
The pneumatic pump is located beneath the on-board reagent compartment. Currently, three
models of a 24-Vdc pump are in use: the Brailsford brushless vacuum pump; the ASF
Thomas brush-type, diaphragm vacuum pump; and the KNF dual-head, brushless-type
pump. The earliest AC•T analyzers used the Brailsford pump, more recent analyzers
(including the AC•T diff 2 analyzer), used the ASF Thomas pump, and the latest use a KNF
dual-head pump. These pumps satisfy the three pneumatic supply requirements:
r Raw vacuum from the pump is used unregulated as a high-vacuum source for the probe-
wipe function.
r Raw (high) vacuum from the pump is regulated to 6-in. Hg for counting.
r Pressure from the pump (exhaust pressure from the single-head pumps) is used to create
bubbles for mixing the WBC and RBC dilutions in their respective baths.
Note: An upgrade to a KNF dual-head pump is required for instruments used at altitudes of
1524 m (5,000 ft) or above.
Five pneumatic solenoids, LV1 through LV5, control this vacuum system. These solenoids are
mounted together on a manifold and are available as an assembly. The vacuum system is
illustrated in Figure 2.2-4.
Figure 2.2-4 Pneumatic System in AC•T 8/10 and AC•T diff Analyzers - Vacuum Components and Flow
A T M
P n e u m a tic
p u m p 1 L V 1
V a c u u m
3 2 L V 2 s e n s o r
F B 1
1
C K 1
A T M
2 1 0
1
V C 1
7 3 3 9 0 4 3 F
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS
Vacuum Supply
LV2 controls the vacuum level applied to the vacuum chamber, VC1. High vacuum from the
pneumatic pump is applied directly to the normally closed port, labelled port 3
(Figure 2.2-4). The normally open port, port 2, is connected to the vacuum being routed
through the vacuum regulator. This configuration makes high vacuum and regulated low
vacuum available at a common port, port 1 which is connected to VC1, LV1, and the vacuum
sensor.
At rest, when the pneumatics are on but the analyzer is not cycling, the high vacuum supply
being regulated to 6-in. Hg by the vacuum regulator is the vacuum available inside the
vacuum chamber, VC1. Regulated vacuum, monitored by the vacuum sensor, is used during
the count portion of the cycle to draw the WBC and RBC dilutions through their respective
apertures and to pull sweep-flow diluent behind the RBC aperture.
The high vacuum applied to VC1 when LV2 is energized is used to scavenge waste during the
probe-wipe function. Solenoid LV1 is used to vent VC1 to atmosphere when it is draining and
during high vacuum/low vacuum transitions.
On the single-head pneumatic pump, LV5 switches the exhaust to atmosphere when vacuum
is being used.
Probe Wipe
The probe-wipe function involves two tasks, probe washing with diluent and probe drying.
Both tasks require high vacuum. To supply the high vacuum needed to accomplish these
tasks, LV8 connects the high vacuum supply in VC1 to the bottom port of the probe-wipe
housing (Figure 2.2-5).
After aspiration, while the probe is moving up into the probe-wipe housing, LV10 is energized
to open the pathway from the diluent source to the top port of the probe-wipe housing and
peristaltic pump PM2 begins pumping diluent from the source container to the top
probe-wipe port. The diluent is forced in a spiral action down to the bottom probe-wipe port,
and with high vacuum assistance, is drawn into VC1. The probe-wipe waste path is split into
two lines for attachment to two outside fittings on VC1.
It is important to connect these tubings to the two outer fittings that allow the waste to enter
behind the splash guards inside VC1. This is necessary to reduce flow rate at each fitting
which reduces splashing and cross-talk in VC1. Once the outside of the probe is washed, PM2
is turned off and LV10 is de-energized which stops the supply of wash diluent. The high
vacuum being applied to the bottom probe-wipe housing port dries the probe before LV8
de-energizes. When the probe is moved up from inside the baths, the vacuum is reapplied to
the VC1 chamber without diluent to dry the probe.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
Figure 2.2-5 Vacuum Chamber (VC1) System in AC•T 8/10 and AC•T diff Analyzer- Components and Flow
WBC RBC
aperture aperture
1 LV17 LV16 1
2
Diluent
Probe-wipe 2 2
housing 3
1 LV8
Probe
Vacuum
X
5 3 6 2 41
VC1
7339045E
Count
Regulated vacuum (6-in. Hg) is used for counting. The WBC aperture is connected to VC1
through LV17 and the RBC aperture through LV16 (Figure 2.2-5). During the count portion
of the cycle, the regulated vacuum available in VC1 is applied to these apertures when valves
LV16 and LV17 are energized. Once the pathway is open, the regulated vacuum in VC1 is
applied to the apertures and the WBC dilution inside the WBC bath is pulled through the
WBC aperture and the RBC dilution inside the RBC bath is pulled through the RBC aperture.
Additionally, on the RBC side, regulated vacuum also pulls diluent through the sweep-flow
line. Two streams of droplets should be entering VC1 during the count portion of the cycle.
Note: Regulated vacuum, 6-in. Hg, low vacuum, count vacuum, and aperture vacuum are
synonymous terms and are, therefore, interchangeable.
Pressure Supply
Exhaust from a single-head pneumatic pump or pressure from a dual-head pneumatic pump
is used as the mixing bubble air supply. These air bubbles pass through the sample dilution
inside each bath and mix the dilution to ensure the particle suspension is uniform throughout
the bath. LV5 switches the pressure between atmosphere and the mixing-bubble path
(Figure 2.2-6). The alternate energizing and de-energizing of LV5 breaks up the air flow and
creates the individual bubbles.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS
Figure 2.2-6 Pneumatic System in AC•T 8/10 and AC•T diff Analyzers- Pressure Components and Flow
W B C R B C
b a th b a th
L y tic
re a g e n t
C V 2
C V 1
C V 3 3 2 L V 3
3 2 L V 4
1 A T M
3 2 L V 5
P n e u m a tic
p u m p 1
7 3 3 9 0 4 4 F
Solenoids LV3 and LV4 control the path for the mixing bubbles. LV4 switches the air path
between the WBC bath lytic reagent (side) port and LV3. LV3 switches the mixing bubbles
between the bottom ports of the RBC and WBC baths. The alternate energizing and
de-energizing of LV5 provides the bubble rate.
Check valves are used in all three mixing bubble paths to ensure that liquid from the bath
does not migrate into the pneumatic system. At the completion of mixing, an air gap remains.
This air gap electrically isolates the reagent systems from the dilution in the bath that is about
to be counted.
r An RBC and a WBC bath and aperture assembly for cell count and size
r A photometer for Hgb measurement
r The vacuum chamber (VC1)
r A sweep-flow tubing spool
r Two fluidic solenoid valves opening the vacuum pathway to each bath
These components are mounted on a plate that is electrically isolated from the chassis and is
enclosed with a shield attached to it. A single ground connection between this assembly and
the preamp shield on the Analyzer card eliminates undesirable ground loops.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
The aperture and bath are one piece. The assembly clips (without fasteners) onto a bath
support. The electrical connection to the bath assembly is via one coaxial cable plugged
directly into the preamp on the Analyzer card. The internal wire is attached to the internal
rear chamber electrode and the coaxial shield is connected to the external bath electrode.
Since most of the assembly is enclosed within the isolated shield/plate assembly, there is little
need to shield both electrode wires individually, although a portion of the coaxial cable inside
the instrument’s chassis is not shielded from noise coming from within the chassis.
Note: Two versions of the aperture/bath assembly are currently in use. The newer version has
an extra tubing port at the top of the bath that is not used by the AC•T 8/10 or AC•T diff
analyzer, but is used by the AC•T diff 2 analyzer. If this bath is used, the newer molded
version of the bath clamp must also be used, since it has a notch to accommodate the extra
fitting.
CAUTION The length of the screw used to mount the Hgb photometer assembly is critical. A screw that is
too long will punch a hole in the WBC bath aperture assembly. Use the correct length screw to mount the
Hgb photometer assembly.
A single screw mounts the Hgb photometer assembly to the WBC bath while two additional
screws and posts are used for shipping support. The assembly is composed of two pieces, a
housing including the photodiode and preamp, and the removable Hgb lamp. A coaxial cable
connects the photodiode preamp to the Analyzer card. A four-wire cable connects the lamp to
the Analyzer card. On the AC•T 8/10 analyzer, two wires are used for the lamp’s voltage and
two wires are used to sense the voltage applied to the lamp. Lamp sensing is used to keep the
lamp’s voltage constant at approximately 3.18 Vdc. On the AC•T diff analyzer, sensing is not
used by the Analyzer card and the supply voltage is 3.6 Vdc.
Waste System
At the completion of a sample run, the remaining sample dilutions, the counted sample, and
the material washed from the outside of the probe must be cleared from the system in
preparation for the next sample. This means the two aperture baths and the vacuum chamber,
VC1, must be drained and rinsed then the accumulated waste expelled from the analyzer.
Peristaltic pump PM1 drains these chambers and LV14 and LV15 control the flow of waste
(Figure 2.2-7).
To effectively remove waste from the system, an individual drain line is attached to the
bottom of each chamber (the WBC bath, the RBC bath, and VC1). Each drain line is
controlled by a small Bio-Chem pinch valve. LV14 normally connects the waste system to the
bottom port of the RBC bath and when energized, switches the waste system to the bottom
port of the WBC bath. LV15 is normally connected to LV14 for bath drainage and when
energized, connects peristaltic pump PM1 to the bottom port of VC1. When PM1 is rotated
clockwise, the selected chamber is drained. LV1 must also be energized to apply atmospheric
pressure when draining VC1.
Note: For details concerning the operation of peristaltic pumps, see Peristaltic Pumps at the
end of this section.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS
Figure 2.2-7 Waste/Cleaner System in AC•T 8/10 and AC•T diff Analyzers - Components and Flow
FS3
1 2 CV4
Cleaner
2 1
Waste
PM1 CV5
FLS2
LV13
WBC RBC
bath bath
VC1
LV14
CV6
LV15
7339042E
Cleaner System
To limit the number of components used in the instrument, peristaltic pump PM1 does
double duty (Figure 2.2-7). Normally PM1 is connected to the waste system, but when LV13
is energized, PM1 is connected to the AC•T Rinse™ shutdown diluent (cleaner). PM1 is
rotated counterclockwise and AC•T Rinse shutdown diluent is brought into the system and
fills both counting baths through their respective waste ports. An inline optical fluid sensor
(FS3) is placed in the tubing from the reagent source to ensure reagent is present.
Note: For details concerning the operation of peristaltic pumps, see Peristaltic Pumps at the
end of this section.
A parallel check valve/filter combination is used to ensure that waste is not pushed back into
the system. When draining waste from the system, the waste passes through CV4. When the
AC•T Rinse shutdown diluent is brought into the system, it passes through CV5 and
hydraulic particle filter FLS2. FLS2 captures debris from the waste system and from the
peristaltic pump tubing wear before debris can enter the baths. FLS2 is directional and fluid
should enter the blue (striped) side of the filter.
PN 4237339F
INSTRUMENT DESCRIPTION
DILUTER SYSTEM - AC•T 8/10 AND AC•T diff™ ANALYZERS 2
Peristaltic Pumps
A peristaltic pump consists of a spool with an attached motor (Figure 2.2-8). The spool is a
cylindrical device which has a rim or ridge at each end and an axial hole for inserting the
motor’s shaft. Flexible tubing is wound around the spool and attached to a connector. When
the motor is turned on, the motor’s rotating shaft rotates the spool. Fluid is forced along by
waves of contraction produced mechanically on the flexible tubing. The direction of flow is
determined by the clockwise or counterclockwise rotation of the motor’s shaft.
Motor
Spool
Flexible tubing
7339131E
Two peristaltic pumps are used on the AC•T 8/10 and AC•T diff analyzers. PM1 is used to
either move shutdown diluent (cleaner) into the analyzer or waste out of the analyzer while
PM2 is used to move diluent from the reagent source into the diluent reservoir.
PN 4237339F
INSTRUMENT DESCRIPTION
POWER SUPPLY MODULE 2
2.4 POWER SUPPLY MODULE
Components
The Power Supply module, a single module mounted on the rear of the AC•T analyzer
(Figure 2.4-1), provides power for the instrument. The Power Supply modules used in the
AC•T Series analyzers vary slightly and, therefore, require three separate part numbers.
The main component of the Power Supply module is a switching supply. See Table 2.4-1 for
the specifications.
PN 4237339F
INSTRUMENT DESCRIPTION
POWER SUPPLY MODULE
Other components comprising the Power Supply module are (Figure 2.4-1):
r Neutral and line 1.5 A, SLO-BLO fuses, AC•T 8/10 and AC•T diff analyzers
Connector
P5
Shielding
case
Power
line cord
connector
Power ON/OFF
switch
Fuse
holders 7339013C
Module Differentiation
The location of connector P5 is shown in Figure 2.4-1. The nine pinouts for the P5 connector
on the AC•T 8/10 analyzer are described in Table A.1-6. The 12 pinouts for P5 connector on
the AC•T diff and AC•T diff 2 analyzers are described in Table A.1-7.
PN 4237339F
INSTRUMENT DESCRIPTION
POWER SUPPLY MODULE 2
The pinouts for P5 were increased to allow for greater +24 Vdc supply current draw by the
AC•T diff and AC•T diff 2 analyzers; two wires are used for the +24 Vdc instead of one. The
need for two different harnesses arose due to instrument design changes that exceeded the
current rating for a single wire. The sense line for the +5 Vdc supply is also passed through
the connector to the card in the 12-pin version.
A 12-pin to 9-pin converter harness is available to allow the newer, 12-pin Power Supply
module for the AC•T diff analyzer to be used on an older AC•T 8/10 analyzer. The 12-pin
Power Supply module for the AC•T diff 2 analyzer is not interchangeable with the Power
Supply modules used on the AC•T 8/10 and AC•T diff analyzers.
Fuses
The AC•T 8/10 and AC•T diff Power Supply module circuitry uses two 1.5 A SLO-BLO fuses;
whereas, the Power Supply module for the AC•T diff 2 analyzer uses two 2.5 A SLO-BLO
fuses.
Appearance
The Power Supply module’s exterior cover matches the paint scheme used on the back of the
analyzer. The AC•T diff 2 analyzer uses a lighter paint scheme than that used on the
AC•T 8/10 and AC•T diff analyzers covers.
Comparison
The major components used in these three Power Supply modules are identical. The
modifications requiring them to have separate part numbers are minimal. Table 2.4-2
provides a quick reference of the features that make the Power Supply modules differ enough
to require a separate part number.
PN 4237339F
INSTRUMENT DESCRIPTION
ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers 2
2.6 ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers
The Analyzer card is the main electronic printed circuit (PC) card in the instrument. It
controls all the input/output functions and data analysis. Refer to Figure 4.4-2 for the
location of the Analyzer card.
Currently the AC•T 8/10, the AC•T diff and the AC•T diff 2 analyzers use different Analyzer
cards. The Analyzer card for the AC•T 8/10 analyzer is described in Heading 2.5.
A major difference among these Analyzer cards is the additional editor circuitry needed by the
AC•T diff and AC•T diff 2 analyzers for creating histograms. The use of surface-mount
technology (SMT) for mounting components allows more circuitry on the same size card so
the overall size of the Analyzer card has not changed. The numbering for its connectors also
remains the same. One obvious difference is the number of mounting holes on the cards. The
Analyzer cards for the AC•T 8/10 and AC•T diff analyzers use seven mounting holes; the
Analyzer card for the AC•T diff 2 analyzer has an eighth mounting hole in its lower right
corner (Figure 4.4-3).
Although the Analyzer card for the AC•T diff 2 analyzer is very similar to the Analyzer card
for the AC•T diff analyzer, it has more driver components to handle the additional stepper
motors, dc motors, and sensors used by the AC•T diff 2 analyzer. Refer to Figures A.2-2 and
A.2-3 for component locations.
New AC•T diff and AC•T diff 2 analyzers are shipped with a Universal AC•T diff/AC•T diff 2
Analyzer card. This card provides better motor drive current and stepping control for the
peristaltic pumps in the AC•T diff analyzer and the FMI sample pump in the AC•T diff 2
analyzer, and is required if these pumps are wired for half-winding. Jumpers are used on the
Universal AC•T diff/AC•T diff 2 Analyzer card to configure the card for the instrument model
and for the type of peristaltic or FMI sample pump (full or half winding) currently in use on
the instrument. Refer to Figure A.2-4 for the jumper locations and to Table A.2-10 for the
jumper configurations.
Additional voltages are also created on the Analyzer card. Chip U8, a 7805CT, 5 Vdc regulator
produces a clean +5 Vdc supply from the +15 Vdc source. The +5 Vdc supply is free of
switching noise and is used for the more sensitive analog circuits. A +15 Vdc precision
reference supply is needed by the vacuum transducer. It is created with U7. Finally, the Red
PN 4237339F
INSTRUMENT DESCRIPTION
ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers
and White Amplifier and Editor circuits use ±10 Vdc references. The ±10 Vdc references are
generated on the Red and White Editor hybrid circuits, U65 and U73, respectively.
The full memory address capability of U116 is used. Chips U111 and U112, are 256 KB of
static RAM, and are used for temporary storage. Another 128 KB of non-volatile battery
backed static RAM, U109, are used to hold user setup information and stored patient and
control results. A 1 MB, Flash Memory card stores the program. Only 640 MB of the 1 MB are
used since that is all that the microprocessor can address.
Communication
Communication to Printer and host computer is provided by U3, a TL16C451 device. This is
an industry-standard device used on many computer motherboards. It provides both the
parallel Centronics™ Printer ouput via a female DB25 connector and RS-232 serial output
using a male DB9 connector. The serial output is capable of baud rates up to 19,200 KB.
Internal buffering and external resistors provide some ESD protection from the peripherals
connected to these ports.
Display
The control of the touch screen and LCD is handled by chip U60, a 74ABT16543 device. U60
is a bidirectional byte-wide data port and byte-wide control port. One bit in the control map
determines the direction of data flow. The data lines pass through RN7, a resistor network of
330-Ohm resistors. This provides ESD protection and matches impedance for the long ribbon
cable used.
Solenoid Control
Solenoid and dc motor control is provided using two 82C55 I/O expanders, chips U99 and
U100. Up to twenty-two dc drivers are selected using U99 and U100. Other lines on U100 are
used to control data acquisition and system modes of operation. U99 is also used to monitor
system and reagent status.
L603C drivers (U102, U103, and U104) drive solenoids and small dc motors, while
RF1K49156 drivers (U124, U125, and U54 on the Analyzer card for the AC•T diff 2 analyzer)
drive larger dc motors requiring higher current.
On the Analyzer card for the AC•T diff analyzer, U102 drives solenoids 1 through 8, U103
drives solenoids 9 through 16, and U104 drives solenoids 17 through 20. U124 drives
connector J12/LV23 and provides power to the vacuum pump. U125 drives connector
J11/LV22 which is currently a spare.
The Analyzer card for the AC•T diff 2 analyzer uses the same components to drive solenoids,
but it must drive five dc motors instead of one, creating a greater requirement for high current
drivers. Connector J36/LV6 uses higher current driver chip U54 instead of the lower current
solenoid driver on U102. The unused solenoid output on U102, formerly used by J36/LV6, is
tied to J31/LV1, doubling the power capability of J31/LV1. A similar scheme is used for
PN 4237339F
INSTRUMENT DESCRIPTION
ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers 2
solenoid connectors LV18/J48 and LV20/J50. Since U104 has eight drivers and only four are
used in the Analyzer card for the AC•T diff analyzer, the extra driver outputs are tied to these
connectors, doubling their power output capability.
All the Analyzer cards have the circuitry to control six motors, but only the Analyzer card for
the AC•T diff 2 analyzer uses six motor drivers. The Analyzer card for the AC•T diff analyzer,
like the one for the AC•T 8/10 analyzer, only has five components mounted.
Output from the microcontroller passes through L6506 and TPIC0298 devices. TPIC0298 is a
power stage and L6506 a current modulation controller. Table 2.6-1 summarizes the devices
and the motor functions they control.
Analyzer Card for AC•T diff Analyzer Analyzer Card for AC•T diff 2 Analyzer
D/A Conversion
Digital to analog conversion is necessary for several functions. This conversion is
accomplished by U43, an AD7228 device. U43 converts digital information supplied by the
CPU into a white cell, red cell, and platelet count-threshold voltage, the contrast adjustment
voltage for the LCD and the voltage pulses used by the diagnostics pulse test.
Data Acquisition
The data acquisition scheme used by the AC•T diff analyzer is a replica of that used by the
COULTER MD II™ Series Analyzer (MD II analyzer). The data acquisition module consists of
Red and White Preamplifier circuits with Dc Restorer circuits, Platelet and White Gain and
Offset Stages, Red, Platelet and White Peak Detector hybrids, Red And White Pulse Editor
hybrids, delay lines, red and white editor state machines, a Platelet Processor circuit and three
a/d interfaces providing three DMA channels. The hybrid daughter cards are the same parts
used by the MD II analyzer and the state machines. U22 and U33 are the same EPLD with the
PN 4237339F
INSTRUMENT DESCRIPTION
ANALYZER CARDS - AC•T diff and AC•T diff 2 Analyzers
same program used on the MD II analyzer. The gain is set during latex calibration using an
independent gain control stage for the red, platelet, and white channels.
The remainder of the data acquisition module is comprised of two ADG506A multiplexer
chips, U35 and U42. U35 and U42 route the system voltages, vacuum voltage, Hgb lamp and
photodetector amplifier output, temperature, red and white aperture voltages and the
26-percentile voltages. The selected voltage is fed to an a/d converter, AD7572 chip (U36)
for processing.
Jumpers
All of the jumpers on the Analyzer card for the AC•T diff analyzer and most of the jumpers on
the Analyzer card for the AC•T diff 2 analyzer are used for sub-assembly testing. Two of the
jumpers on the AC•T diff 2 analyzer are used for system configuration. Refer to Table A.2-7,
Analyzer Card Jumpers - AC•T diff 2 Analyzer (See Figure A.2-3) for the correct
jumper settings.
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY 2
2.7 DISPLAY ASSEMBLY
Main Components
The Display assembly for the AC•T analyzer is mounted on the front door. It is composed of
three separate electronic components fastened together: the touch screen, the LCD, and the
Display Controller Driver card (Figure 2.7-1). The touch screen and LCD are purchased
components while the Display Controller Driver card is designed in-house.
Chassis
ground
strap
P61
Power on
LED cable
Touch screen
Daughter X1 to X4
cable
card
J66
LCD
J65
J63
J64
Display Controller
7339017E
Driver card
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY
Touch Screen
An 8 row x 8 column, digital-resistive touch screen provides system input. Table 2.7-1 lists
the general specifications for the touch screen.
Item Specification
Actuation force 40 to 80 g
Mechanical life 6 million cycles
Contact resistance (open) 20 MΩ
Contact resistance (closed) 10 to 30 KΩ
Operating voltage 5V
Operating current 5 mA
Flex cables connect the rows and columns (displayed as a grid) to the Display Controller
Driver card at connector J63. To connect the touch screen to the Display Controller Driver
card, locate the grid on the back of the touch screen and follow the pinouts on Table 2.7-2.
The grid is displayed in rows and columns. The columns correspond to the “front” flex cable
and the rows to the “back” flex cable.
Columns: Rows:
Pin Front Flex Cable Destination Pin (J63) Back Flex Cable Destination Pin (J63)
1 4 9 1 1
2 3 10 2 2
3 2 11 3 3
4 1 12 4 4
5 5 13 5 5
6 6 14 6 6
7 7 15 7 7
8 8 16 8 8
LCD
The liquid crystal display (LCD) is a 320 x 240 dot, graphics-capable, supertwist LCD with an
effective viewing area of 103 x 80 mm. There are two connectors on the LCD, CN1 and CN2.
CN1 connects directly to the Display Controller Driver card, J62, without a cable, while CN2
is a three-pin connector at the end of two wires (black and red) soldered directly to the LCD.
This connector plugs into CN2 on the high voltage backlight daughter card mounted to the
Display Controller Driver card. Table 2.7-3 and 2.7-4 respectively, describe the pinouts for
these connectors.
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY 2
Table 2.7-3 LCD Connector CN1
Function
The Display Controller Driver card is an interface between the front door components and the
Analyzer card. The Display Controller Driver card:
r Receives power from the Analyzer card and distributes the power to the display, the
power on LED (J64/P64), and the reagent management card reader.
r Receives display data from the Analyzer card and uses the display data to control the
display.
r Receives input data from the touch screen and forwards it to the Analyzer card.
r Provides a two-way communication path between the reagent management card reader
and the Analyzer card.
Major components of the Display Controller Driver card include:
r A buzzer (BZ1) that provides sound for touch screen keypress response.
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY
The interface between the Display controller Driver card and the Analyzer card is through
connector J61. The pins and functions of J61 are detailed in Table 2.7-5.
.
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY 2
Differences Between the Old and New Versions of the Display Controller Driver Card
Two versions of the Display Controller Driver card are currently being used in AC•T Series
analyzers: the old card with the standard through-hole design and the new card with
surface-mount design. Currently, both cards will work in any AC•T Series analyzer.
The old through-hole card design was used on the AC•T 8/10 analyzers and on the original
AC•T diff analyzers. Unlike the old version of the Display Controller Driver card, the new
version of the Display Controller Driver card allows the use of reagent management cards
from a different manufacturer. For this reason, when you replace the Display Controller
Driver card on an AC•T diff or AC•T diff 2 analyzer, you should always use the new version of
the Display Controller Driver card for the replacement. This way, if the decision is made to
change the vendor for the reagent management card, the AC•T diff and AC•T diff 2 analyzers
will be ready for the change.
Note: The new card is designed as a drop-in replacement and will work in any AC•T Series
analyzer; however, if a new Display Controller Driver card is used to replace an old card in an
AC•T 8 analyzer, the analyzer will still be unable to use reagent management cards from a
different manufacturer. Current software for the AC•T 8 analyzer does not support this
capability and there are no plans to add this capability to its program.
Table 2.7-6 lists the connectors used on the two versions of the Display Controller Driver
card.
Most of these connectors have identical locations on the two versions of this card. Because
some connectors are not in the same location, the two versions of the Display Controller
Driver card require different electromagnetic shields for the display assembly. FRUs that
include both the new Display Controller Driver card and the appropriate shield are available,
one for the AC•T 8/10 and AC•T diff analyzers and another for the AC•T diff 2 analyzer. Refer
to the master parts list, Table 8.1-1, or the illustrated parts list, Table 8.2-4, for the part
numbers.
Note: The AC•T diff 2 analyzer uses the same display assembly and the same Display
Controller Driver card as the AC•T diff analyzer. The shield is different because the front door
designs are different.
PN 4237339F
INSTRUMENT DESCRIPTION
DISPLAY ASSEMBLY
Table 2.7-7 lists the jumpers used on the old through-hole design of the Display Controller
Driver card; the new surface-mount design does not need jumpers. Table 2.7-8 lists the test
points available on the two versions of the Display Controller Driver card.
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE 2
2.8 SOFTWARE STRUCTURE
Overview
The AC•T analyzer software can be divided into two sections, the low-level operating system
and the diluter control. The operating system software provides instrument I/O, system error
checks, data analysis, and individual diluter subroutines or functions. The diluter control
software is responsible for all diluter activity. The operating system software provides diluter
functions that energize solenoids, drive motors, and check sensors in the diluter. A sequence
of these functions makes up a diluter table. In effect, these diluter functions become a
high-level programming language and the program using this language is called a diluter
table. Instrument cycles are diluter tables, such as the whole-blood (aspirate) diluter table or
the shutdown diluter table.
Menu Trees
Most I/O functions of the operating system software are controlled by the user. This
interaction between the user and the instrument is called the user interface. A touch screen
and LCD provide the physical user interface, while menus and menu items provide the
software user interface. Menus and menu items are displayed on the screen as graphic icons.
Pressing an icon selects that function or menu. Icons are used in place of text for language
independence. Refer to the following figures for a graphic representation of the available
icons and text equivalent for each icon:
Instrument Cycles
Since an instrument cycle is a sequence of diluter functions that makes up a diluter table,
running an instrument cycle is simply running the appropriate diluter table. These diluter
tables, with appropriate parameters such as duration or rate, are stored in their own files. This
method allows changes to an instrument cycle without changing the main program.
Ten instrument cycles provide diluter control in the AC•T 8/10 and AC•T diff analyzers:
Powerup, Whole-Blood Aspirate, Predilute, Prime, Wakeup Prime, Startup, Shutdown, Clean
Baths, Non-Labile Control Mode (AC•T Tron™ control), and the Prepare-to-Ship cycle.
The AC•T diff 2 analyzer has a Closed-Vial Aspirate cycle and an Open-Vial Aspirate cycle in
place of the Whole-Blood Aspirate cycle. The primary aspirate or sample table for the
AC•T 8/10 and AC•T diff analyzers is the Whole-Blood Aspirate cycle; the primary aspirate
table for the AC•T diff 2 analyzer is the Closed-Vial Aspirate cycle.
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE
The following tables summarize the Powerup and the primary Aspirate cycles:
r Powerup cycles:
t Table 2.8-1, Powerup Cycle - AC•T 8/10 and AC•T diff Analyzers, All Software
Versions
r Aspirate cycles:
t Table 2.8-5, Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Version
1.03
t Table 2.8-6, Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Versions
1.04, 1.05, 1.06, and 2.00
t Table 2.8-7, Whole-Blood Aspirate Cycle - AC•T diff Analyzer with Veterinary
Option, Software Versions 1.04, 1.05, 1.06, and 2.00
Realize that changes in software can result in changes to these cycles. Also times are
approximate because they can vary. For example:
r The time to analyze the data depends on the shape of the histogram.
r The time to fill the reservoir up to the sensor depends on the starting level of the reagent
and on the state of the filters and the peristaltic pump tubing. This is even more
pronounced in the Powerup cycle when the reagent level can be anywhere.
r The position of the syringe, and therefore, the time it takes to home the syringe at the
beginning of the Powerup cycle, depends on the state of the instrument when the power
was turned off and what was done to the instrument while the power was off. The first
movement in the Powerup cycle is to home the syringe. If the syringe is already at home,
no time is taken. If the syringe is at the bottom of stroke, up to 2.2 seconds is used. This
time affects the rest of the Powerup cycle. For the Powerup cycles described in Tables
2.8-1 and 2.8-2, step duration is given rather than time.
Powerup Cycle
The Powerup cycle is a diluter cycle that is executed once the main program is up and
running. When power is applied, the software first checks for valid CMOS data, then displays
the copyright screen, does a checksum of the Flash Memory card, checks for clock
movement, setting a default if necessary, and finishes by doing a DVM check and a waste level
check. This activity takes 25 to 30 seconds. When this activity is complete, the Powerup cycle
begins. This Powerup cycle may also be executed when leaving the Diagnostics menu or the
Diluter Functions menu, or when entering into the Verify Predilute function.
Tables 2.8-1 and 2.8-2 describe the Powerup cycle for the current models and software
versions of the AC•T Series analyzers.
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE 2
Table 2.8-1 Powerup Cycle - AC•T 8/10 and AC•T diff Analyzers, All Software Versions
Duration Activity
2.0 Energize LV1, de-energize LV2 through LV17 (when power is applied, some drivers may be on)
0.025 Energize LV1, turn on vacuum pump, energize LV11 and LV7
2.2 to 3.1 Home syringe while draining RBC bath and moving probe up
0.0 to 32.4 Fill diluent reservoir
2.27 Fill diluent syringe (3.0 mL) while draining RBC bath
0.025 Energize LV12
2.45 Dispense diluent syringe into RBC bath, de-energize LV12
9.5 Repeat last 3 steps 2 times
0.0 Energize LV14
0.025 Wait
2.24 Fill diluent syringe (3.0 mL), drain WBC bath
0.025 Energize LV12, energize LV7
2.45 Dispense 3.1 mL from diluent syringe to WBC bath
0.0 De-energize LV12, energize LV7
9.5 Repeat last 5 steps 2 times
2.0 Fill diluent syringe (5.0 mL)
0.0 Check diluent sensor for empty (fluid not at sensor) state
2.9 Dispense diluent syringe to sensor (5.0 mL to reservoir)
2.24 Fill diluent syringe (3.0 mL), drain WBC bath, send probe to aspirate position, check lytic reagent sensor
0.025 De-energize LV14, energize LV12, de-energize LV7, energize LV8
2.45 Dispense 3.0 mL from diluent syringe to WBC bath, fill diluent reservoir
0.0 De-energize LV12, LV8, LV11
BEGIN CLEANUP ROUTINE (performed only if power was turned OFF in midcycle)
0.0 Energize LV11, LV7
0.025 Delay
2.43 Fill diluent syringe with 3.0 mL while draining RBC bath
3.525 Energize LV12, dispense 3.0 mL from diluent syringe to RBC bath while filling diluent reservoir,
de-energize LV12
14.6 Repeat last 3 steps 2 times
0.025 Energize LV14, LV8
2.24 Fill syringe with 3.0 mL while draining WBC bath
0.025 De-energize LV8, energize LV12, de-energize LV7
1.55 Move probe to WBC bath, dispense 0.2 mL from diluent syringe to WBC bath
0.025 De-energize LV11
4.48 Move probe down into WBC bath, dispense (to sensor, about 3.0 mL) from diluent syringe through probe
to WBC bath
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE
Table 2.8-1 Powerup Cycle - AC•T 8/10 and AC•T diff Analyzers, All Software Versions (Continued)
Duration Activity
0.0 De-energize LV12, energize LV11, LV7
0.025 Delay
2.24 Fill diluent syringe with 3.0 mL while draining WBC bath
0.025 Energize LV12, de-energize LV7
2.45 Dispense 3.0 mL from diluent syringe to bottom of WBC bath while filling diluent reservoir
0.025 Energize LV7, de-energize LV12
10 Repeat last 5 steps 2 times
0.025 De-energize LV14, LV7, LV11, energize LV8, LV2, LV10
1.85 Fill diluent reservoir without stopping at sensor, move probe up
1.0 De-energize LV10
1.55 De-energize LV2, LV8, move probe to aspirate position
END CLEANUP ROUTINE (resume normal Powerup cycle at next step)
0.025 Energize LV1, LV15, LV7, LV11
0.9 Fill diluent syringe with 1.8 mL while draining VC1
0.025 De-energize LV7, LV11
1.3 to 3.15 Fill diluent reservoir while further draining VC1
0.025 De-energize LV1, energize LV8
1.3 Aspirate 12.0 µL, move probe down, sound a warning beep, drain VC1
0.025 De-energize LV8, turn vacuum pump off, energize LV1
0.0 to 5.4 Fill diluent reservoir to sensor (tries 5 fills of 10.8 seconds each)
1.127 Overfill diluent reservoir with 2.5 mL (without regard to sensor)
0.0 Check diluent sensor for fluid (system expects fluid at sensor)
0.0 De-energize LV15, LV1
Aspirate Cycle
Pressing the aspirate switch or closing the cap-pierce door initiates the Aspirate cycle. The
main program turns on the vacuum pump and then executes the Aspirate cycle table for the
mode selected.
Tables 2.8-3, 2.8-4, 2.8-5, 2.8-6, 2.8-7, and 2.8-8 describe the primary Aspirate cycles for the
current models and software versions of the AC•T Series analyzers. The primary Aspirate
cycles are described because they are the cycles most frequently used and because they use all
the instrument’s components except the AC•T Rinse shutdown diluent pump and solenoid.
When running repeated Aspirate cycles, the syringe and FMI pump positions and the diluent
level are known at the beginning of the cycle, so timing is more accurate than it is for the
Powerup cycle. For this reason, the first column (Timing) in the Aspirate cycle tables gives a
time relative to the beginning of the cycle.
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE
Table 2.8-5 Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Version 1.03
Time Activity
0 Energize LV2 (high vacuum), aspirate 12.0 µL
0.32 Energize LV10, LV14, LV8, check for 6 in. Hg at VC1, sound probe warning, move up and wipe probe,
drain WBC bath, overfill diluent reservoir
2.32 De-energize LV14, LV10, aspirate 12.0 µL (air gap)
2.64 Energize LV12, LV11
2.7 Drain RBC bath, move probe to WBC bath, rinse WBC bath with 2.6 mL
4.03 De-energize LV8, LV2, LV12, energize LV14, LV7, take “Hgb Blank2” reading
4.08 Drain WBC bath, fill diluent reservoir with 2.5 mL and diluent syringe with 3.250 mL
5.64 Energize LV12, de-energize LV14, prefill RBC bath with 240.0 µL, de-energize LV7
5.84 Prefill WBC bath with 1.5 mL, move probe into WBC bath, partial drain RBC bath, fill reservoir
6.59 Take “Hgb Blank1” reading, de-energize LV11, dispense 980.0 µL with sample from probe to WBC bath,
drain RBC bath, fill reservoir
7.92 Energize LV3, LV11, LV7, send mix bubbles through lytic reagent port (Energize/de-energize LV5 4 times)
9.32 0.6 mL rinse to RBC bath (prefill), drain RBC bath
10.11 De-energize LV3, LV12, fill diluent syringe with 3.0 mL, fill reservoir
11.34 De-energize LV11, energize LV15, LV1
11.36 Drain VC1, aspirate 100.0 µL from WBC bath, fill diluent reservoir, check lytic reagent sensor
12.96 Energize LV8, de-energize LV15, LV1, energize LV2, LV12, LV11
12.99 Prefill RBC bath with 2.0 mL, fill reservoir, move probe up
13.8 De-energize LV8, drain RBC bath, fill reservoir, move probe to RBC bath
16.0 Energize LV8, move probe into RBC bath with vacuum dry, dispense 695.0 µL diluent from diluent
syringe to bottom of RBC bath
16.55 De-energize LV8, LV2, LV7, energize LV9, de-energize LV11, energize LV3
16.58 Dispense 415.0 µL lytic reagent into WBC bath (while dispensing 100.0 µL sample and 2.075 mL diluent
into RBC bath), mix with 3 bubbles through bottom port, fill diluent reservoir
19.85 De-energize LV3, delay 0.25 seconds, de-energize LV9, LV12, energize LV4, mix WBC bath with 4 bubbles
through lytic reagent port
20.9 De-energize LV4, mix RBC bath through bottom port with 2 bubbles
21.4 Energize LV16, LV17, delay 1 second, acquire data
Note: The time for acquiring count data depends on the number of extended counts. When Plt data is
insufficient, sensing is extended up to eight additional 3.0 second counts in an attempt to acquire
sufficient data. Because the time for acquiring data varies, the start time for the next sequence of events
also varies.
36.4 De-energize LV17, LV16, fill diluent syringe with 3.25 mL, drain RBC bath, move probe into RBC bath
37.88 Energize LV12, LV1, LV15, dispense 0.2 mL diluent from probe
38.11 Energize LV7, LV11
38.13 Home syringe, dispense 3.4 mL diluent to RBC bath, drain VC1, fill diluent reservoir, take two Hgb
Sample readings 0.5 seconds apart
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE 2
Table 2.8-5 Whole-Blood Aspirate Cycle - AC•T diff Analyzer, Software Version 1.03 (Continued)
Time Activity
39.78 De-energize LV1, energize LV2, LV14, de-energize LV12, LV15, energize LV8
39.81 Fill syringe with 3.6 mL, drain WBC bath, move probe up, analyze CBC data
Note: The time needed to analyze data depends on the amount of data accumulated during data analysis
and the Plt histogram data. A normal blood takes about 6.0 seconds, but can vary from 4.0 to 8.0
seconds. Because the time needed to analyze this data varies, the start time for the next sequence of
events also varies.
46.2 De-energize LV8, LV2, energize LV12, de-energize LV7, energize LV15, LV1
46.23 Dispense diluent syringe to WBC bath, drain VC1, perform histogram analysis, flagging, clog detection
48.06 De-energize LV12, LV15, LV1, energize LV11, LV7, fill diluent syringe with 1.8 mL, fill diluent reservoir,
move probe to aspirate position, show results to screen
48.3 De-energize LV11, LV7, LV14, energize LV8, LV2, increment cycle counter
48.33 Energize LV3, LV5 (for 200.0 ms), aspirate 12 µL, fill diluent reservoir (do not look at sensor), sound
probe warning, move syringe down
48.84 De-energize LV3, LV8, LV2, LV7, vacuum pump, energize LV1, LV15, fill diluent reservoir to sensor, print
results
Note: The time needed to print results depends on the Printer used and increases significantly when
automatic host transmission is performed. Because the time needed to print varies, the start time for the
next sequence of events also varies.
52.84 Reservoir overfill, drain VC1, zap aperture
54.46 De-energize LV15, LV1
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE
Time Activity
0 Energize LV2 (high vacuum), aspirate 12.0 µL
0.32 Energize LV10, LV14, LV8, check for 6 in. Hg at VC1, sound probe warning, move up and wipe probe,
drain WBC bath, overfill diluent reservoir
2.5 De-energize LV14, LV10, aspirate 12.0 µL (air gap)
2.82 Energize LV12, LV11
2.88 Drain RBC bath, move probe to WBC bath, rinse WBC bath with 2.6 mL
4.3 De-energize LV8, LV2, LV12, energize LV14, LV7, take “Hgb Blank2” reading
4.9 Drain WBC bath, fill diluent reservoir with 2.5 mL and diluent syringe with 3.250 mL
6.6 Energize LV12, de-energize LV14, prefill RBC bath with 240.0 µL, de-energize LV7
6.8 Prefill WBC bath with 1.5 mL, move probe into WBC bath, partial drain RBC bath, fill reservoir
7.92 Take “Hgb Blank1” reading, de-energize LV11, dispense 980.0 µL with sample from probe to WBC bath,
drain RBC bath, fill reservoir
9.25 Energize LV3, LV11, LV7, send mix bubbles through lytic reagent port (energize/de-energize LV5 4 times)
10.65 0.6 mL rinse to RBC bath (prefill), partial drain RBC bath
11.44 De-energize LV3, LV12, fill diluent syringe with 3.0 mL, fill reservoir, drain RBC bath
12.78 De-energize LV11, energize LV15, LV1
12.8 Drain VC1, aspirate 100.0 µL from WBC bath, fill diluent reservoir, check lytic reagent sensor
14.4 Energize LV8, de-energize LV15, LV1, energize LV2, LV12, LV11
14.43 Rinse RBC bath with 2.0 mL, fill reservoir, move probe up
15.32 De-energize LV8, drain RBC bath, fill reservoir, move probe to RBC bath
17.71 Energize LV8, move probe into RBC bath with vacuum dry, dispense 695.0 µL diluent from diluent
syringe to bottom of RBC bath
18.27 De-energize LV8, LV2, LV7, energize LV9, de-energize LV11, energize LV3
18.29 Dispense 415.0 µL lytic reagent into WBC bath (while dispensing 100.0 µL sample and 2.075 mL diluent
into RBC bath), mix with 3 bubbles through bottom port, fill diluent reservoir
21.56 De-energize LV3, delay 0.25 seconds, de-energize LV9, LV12, energize LV4, mix WBC bath with 4 bubbles
through lytic reagent port
22.61 De-energize LV4, mix RBC bath through bottom port with 2 bubbles
23.11 Energize LV16, LV17, move probe up, delay 1 second, acquire data
Note: The time for acquiring count data depends on the number of extended counts. When Plt data is
insufficient, sensing is extended up to eight additional 3.0 second counts in an attempt to acquire
sufficient data. Because the time for acquiring data varies, the start time for the next sequence of events
also varies.
38.11 De-energize LV17, LV16, fill diluent syringe with 3.25 mL, drain RBC bath, move probe into RBC bath
39.72 Energize LV12, LV1, LV15, dispense 0.2 mL diluent from probe
39.95 Energize LV7, LV11
PN 4237339F
INSTRUMENT DESCRIPTION
SOFTWARE STRUCTURE 2
Table 2.8-6 Whole-Blood Aspirate Cycle - AC•T diff Analyzer,
Software Versions 1.04, 1.05, 1.06, and 2.00
Time Activity
39.97 Home syringe, dispense 3.3 mL diluent to RBC bath, drain VC1, fill diluent reservoir, take two Hgb
Sample readings 0.5 seconds apart
41.77 De-energize LV1, energize LV2, LV14, de-energize LV12, LV15, energize LV8
41.8 Fill syringe with 3.6 mL, drain WBC bath, move probe up, analyze CBC data
Note: The time needed to analyze data depends on the amount of data accumulated during data analysis
and the Plt histogram data. A normal blood takes about 6.0 seconds, but can vary from 4.0 to 8.0
seconds. Because the time needed to analyze this data varies, the start time for the next sequence of
events also varies.
48.0 De-energize LV8, LV2, energize LV12, de-energize LV7, energize LV15, LV1
48.03 Dispense diluent syringe to WBC bath, drain VC1, perform histogram analysis, flagging, clog detection
50.02 De-energize LV12, LV15, LV1, energize LV11, LV7, fill diluent syringe with 1.8 mL, fill diluent reservoir,
move probe to aspirate position, show results to screen
51.39 De-energize LV11, LV7, LV14, energize LV8, LV2, LV3, LV5 for 200.0 ms, increment cycle counter
51.42 Aspirate 12 µL, fill diluent reservoir (do not look at sensor), sound probe warning, move syringe down
51.95 De-energize LV3, LV8, LV2, LV7, turn vacuum pump off, energize LV1, LV15, fill diluent reservoir to
sensor, print results if Autoprint ON
Note: The time needed to print results depends on the Printer used and increases significantly when
automatic host transmission is performed. Because the time needed to print varies, the start time for the
next sequence of events also varies.
56.32 Reservoir overfill, drain VC1, zap aperture
58.08 De-energize LV15, LV1
PN 4237339F
INSTRUMENT DESCRIPTION
INSTRUMENT STATES 2
2.9 INSTRUMENT STATES
To save time and reagents whenever power is interrupted, the AC•T Series analyzers record
the time of the last sample run and the state of the instrument, or more specifically the
diluter, as power is lost. The six instrument states are: busy, ready, sleep, exception, error, and
ready-to-ship.
r Busy state - Any time the diluter is in the middle of a cycle performing some operation,
the exact fluidic condition is not known and this is considered the “busy” state.
r Ready state - When the instrument is not performing a cycle and is waiting for user
input.
r Sleep state - The instrument is put to sleep after two hours without running a sample. To
save energy, improve reliability, and force a Prime cycle the Hgb lamp and pneumatic
pump are turned off. If the system is in the Sample Results screen, the probe is retracted.
If the first request, while the instrument is asleep, is to go to the Sample Results screen,
the probe retracts immediately, the system awakens, then the probe returns to the
aspirate position. Retracting the probe prevents using sample aspiration to awaken the
instrument from the sleep state. This is necessary because system vacuum takes time to
build up to sufficient levels to perform a probe wash and is also used to force a Prime
cycle after two hours.
Note: If the screen is not touched for 15 minutes, the screen backlighting is turned off
(the screen goes blank). This is not considered the “sleep state.” Touch the screen to turn
it back on.
r Exception state - When the instrument identifies a situation requiring the user to take
some action before the instrument can continue (for example, a reagent is empty).
r Error state - When the instrument identifies a fatal error condition.
r Ready-to-Ship state - When the “prepare-to-ship” sequence of operations has been
performed.
The state of the instrument at the time it was powered down is used during powerup to
determine what actions to take.
r If the instrument was in a ready state when powered down, a Powerup cycle (home all
hardware, perform one drain and rinse, extend the probe and check the lytic reagent
level) is performed.
r If the instrument was in a ready state when powered down and over two hours have
elapsed since the last Sample cycle, a Startup cycle is automatically performed after the
Powerup cycle is completed.
r If the instrument was in any other state when powered down, an extended Cleanup
cycle is performed during the Powerup cycle, then a Startup cycle (including a
background count) is performed after the Powerup cycle.
PN 4237339F
CONTENTS 3
3 INSTALLATION PROCEDURE, 3.1-1
ILLUSTRATIONS
3.2-1 Epson TM-290P Slip Printer - DIP Switch, SW1, Location, 3.2-1
3.2-2 Connecting Optional Printers, 3.2-2
3.2-3 Epson TM-290P Slip Printer - Location of Controls and Indicators, 3.2-3
3.2-4 Canon BJC-250 Bubble Jet Printer - Power/Resume Controls and Indicators, 3.2-4
TABLES
3.1-1 AC•T Analyzer Installation Documents, 3.1-1
3.2-1 Canon BJC-250 Bubble Jet Printer - Available Printer Information/Tests, 3.2-4
PN 4237339F
Changed with the DEMO VERSION of CAD-KAS PDF-Editor (http://www.cadkas.com).
3INSTALLATION PROCEDURE 3
PART A: INSTRUMENT INSTALLATION
PN 4237339F
INSTALLATION PROCEDURES
OPTIONAL PRINTERS 3
PART B: UPGRADE AND OPTION INSTALLATION
1. The Printers have different power supply connectors, requiring the use of different
power supplies. That means that if you replace one model with the other you must also
replace the power supply.
2. If the Printer is operating but no ticket is in place, the Epson TM-290P Slip Printer
signals the instrument which initiates a printer warning, but the Epson TM-295 Slip
Printer does not signal the instrument. It continues printing as if a ticket is in place. If
the customer does not notice the missing ticket and cycles the next specimen, the data
from the first sample is lost.
For complete information on how to use this Printer, see the COULTER AC•T Series Analyzer
Ticket Printer User’s Guide. For Printer specifications, see Appendix C, OPTIONAL
PRINTERS.
Figure 3.2-1 Epson TM-290P Slip Printer - DIP Switch, SW1, Location
SW 1
ON
1 2 3 4 5 6 7 8 9 10
7303003A
PN 4237339F
INSTALLATION PROCEDURES
OPTIONAL PRINTERS
CAUTION Risk of damage to equipment. Turning the instrument’s power ON before connecting the Printer
to the instrument could damage the instrument. Ensure the instrument’s power is OFF before connecting
the Printer.
TICKET
PRINTER
AC POWER (OPTIONAL)
LINE CORD
TO POWER
SOURCE
7339056C
Verification
1. Perform a Printer self-test:
a. Ensure the instrument’s power is on and the Printer’s power line cord is plugged in.
b. Press the RELEASE key on the Printer (Figure 3.2-3).
c. Insert a ticket in the Printer.
d. Unplug the Printer’s power line cord.
e. Press and hold down the FORWARD key (Figure 3.2-3). While holding the FORWARD
key, plug in the Printer’s power line cord to initiate the self-test.
f. Press the RESET button to stop the self-test (Figure 3.2-3). The self-test stops after
printing a specific number of lines.
2. Print the calibration factors to verify that the Printer and the instrument are working
together properly.
PN 4237339F
OPTIONAL PRINTERS
OPTIONAL PRINTERS 3
Figure 3.2-3 Epson TM-290P Slip Printer - Location of Controls and Indicators
POWER
RELEASE
RESET PAPER OUT
FORWARD
REVERSE
RELEASE
7303007A
To achieve complete printing results, the AC•T diff and AC•T diff 2 analyzers need to be set
up for graphic printing.
r To set up the AC•T diff analyzer properly, refer to Selecting the Printer Type in Chapter 1
of the COULTER AC•T diff Analyzer Installation and Training Guide.
r To set up the AC•T diff 2 analyzer properly, refer to Selecting the Printer Type in
Chapter 1 of the COULTER AC•T diff 2 Analyzer Installation and Training Guide.
For complete information regarding this Printer, refer to the COULTER AC•T diff Analyzer
Canon® Bubble Jet™ Printer User’s Guide.
1. Unpack the Printer removing all items from the box. Ensure you have the following
components:
r Printer
r 24-V Printer Power supply
r Power line cord for the power supply
r Standard Centronics® data cable
r Black ink cartridge
PN 4237339F
OPTIONAL PRINTERS
OPTIONAL PRINTERS
2. Set the Printer on a flat, stable surface. Do not use an angled Printer stand.
CAUTION Risk of damage to equipment. Turning the instrument’s power ON before connecting the Printer
to the instrument could damage the instrument. Ensure the instrument’s power is OFF before connecting
the Printer.
Verification
1. Perform a Printer self-test:
a. Turn the Printer off.
b. Ensure that there is paper in the Printer.
c. To perform the specific tests or gather the information shown in Table 3.2-1, press
and hold the POWER button (Figure 3.2-4) until you hear the number of audible
beeps required, then release the POWER button.
Table 3.2-1 Canon BJC-250 Bubble Jet Printer - Available Printer Information/Tests
Figure 3.2-4 Canon BJC-250 Bubble Jet Printer - Power/Resume Controls and Indicators
BUTTONS
POWER RESUME
POWER
LIGHT
7479001A
PN 4237339F
OPTIONAL PRINTERS
OPTIONAL PRINTERS 3
Ithaca Series 90 Printer
The Ithaca Series 90 Printer (also called the OKIPOS Series 90 Printer in Germany) is
supported by the AC•T diff and AC•T diff 2 analyzers’ software, but it is not available with the
instrument and does not have a Beckman Coulter part number. This Printer must be obtained
from the manufacturer or a third party distributor.
The Ithaca Series 90 Printer is a 40 column, fast, dot matrix Printer. It is intended as a “point
of sale” Printer and hence is very reliable. This Printer is capable of printing graphics from an
Espson graphic driver, just like the Citizen GSX-190 and Canon Bubble Jet BJC-250 Printers.
The Model 93 Printer was used in developing software support. When the instrument is setup
for this Printer, histograms print vertically, rather than horizontally. In addition to its graphic
capability, this Printer can print tickets (slips), and has a journal feature. The journal feature
allows one or two backup copies when used as a Roll Printer. The backup copy is
automatically spooled in a roll at the back of the Printer, similar to many cash register
systems. The other Series 90 models do not have all the printing features, and consequently
do not cost as much. This might be a consideration if all features will not be used.
When using the Ithaca Series 90 Printer as a Ticket Printer, you need to consider three main
differences between it and the Epson Printers.
1. Settings are not available at the Ithaca Series 90 Printer to select the Ticket mode; the
instrument must send a command to put this Printer into the Ticket mode.
Consequently the instrument has to be programmed specifically for this Printer and this
mode.
When the proper command is sent, the Ithaca Series 90 Printer accepts a ticket, but as
soon as it prints the ticket it automatically defaults to the Roll Paper mode.
2. Standard in-house tickets do not work well with the Ithaca Series 90 Printer; third party
tickets are strongly recommended.
All previous Ticket Printers used advance rollers on the underside of the ticket. The
Ithaca Series 90 Printer has its ticket advance rollers on the top of the ticket. The
combination of hard back copy and very smooth (slippery) carbon paper causes the front
copies of standard Beckman Coulter tickets to bunch up.
3. The Ithaca Series 90 Printer has a standard parallel printer output, but does not use the
standard DB25 pin to Centronics 36 pin printer cable. It uses a straight DB25 pin to
DB25 pin cable.
PN 4237339F
OPTIONAL PRINTERS
OPTIONAL PRINTERS
3. Connect the end of the Printer cable labeled P2 (DB-25 connector) to the back of the
instrument (next to the Printer symbol) and secure the connection by tightening the two
screws.
4. Press the Printer cover release button and lift the cover.
ATTENTION: For steps 5 through 12, refer to the supplied Quick Start Guide.
5. Lift out the image drum cartridge. Do not touch the drum’s green surface or remove the
sponge at this time.
6. Without turning the drum cartridge over, remove the protective paper from the image
drum.
7. Reinstall the cartridge in the Printer.
8. Remove the sponge.
9. Unwrap the toner cartridge and shake the cartridge to distribute the toner.
10. Peel the tape off of the bottom of the toner cartridge.
11. Grasp the lock lever and place the left side of the toner cartridge under the end tab.
12. Lower the right side of the toner well, aligning the lock lever slot with the ridge in the
drum cartridge.
13. When the cartridge is in place, push the lock lever forward until it stops.
14. Lower the Printer cover and press it firmly to close.
15. Remove the protective film from the front operator panel.
16. Slide out the paper tray.
17. Fan a stack of paper and place it in the tray under the tabs.
18. Place the tray in the printer and push closed until the tray locks in place.
19. Connect the power cord to the Printer, then plug the cable into a grounded ac outlet.
ATTENTION: DO NOT press the User Option switch while turning the Printer ON, as that will
reset the Printer back to the factory-default settings.
20. Turn the Printer ON. The Printer takes about 10 seconds to warm up. At that time, the
“Ready Light” on the operator panel lights, indicating the Printer is ready to receive data.
Verification
1. Print a Demo and Menu page:
a. Turn the Printer ON.
b. Set the Printer to off line (the green ready light is off).
c. Press the User Option switch for 2 seconds, then release it.
If done correctly, the ready light begins to flash and the Printer prints a Demo page
followed by a list of current menu settings.
2. Compare the menu settings to the settings in the User column in Table C.3-2. If any
setting needs to be changed, reconfigure the Printer. Refer to Configuring the Printer
procedure under Heading C.3, OKIPAGE 14e LED PRINTER.
PN 4237339F
CONTENTS 4
4 SERVICE AND REPAIR PROCEDURES, 4.1-1
PN 4237339F
CONTENTS
4
4.10 APERTURE/BATH ASSEMBLY REPLACEMENT, 4.10-1
Purpose, 4.10-1
Tools/Supplies Needed, 4.10-1
Removal, 4.10-1
Installation, 4.10-3
Verification, 4.10-5
4.12 LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS, 4.12-1
Purpose, 4.12-1
Tools/Supplies Needed, 4.12-1
Procedure, 4.12-1
Verification, 4.12-3
4.13 SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2
ANALYZERS, 4.13-1
Tools/Supplies Needed, 4.13-1
Procedure, 4.13-1
Verification, 4.13-2
PN 4237339F
CONTENTS
4
ILLUSTRATIONS
4.2-1 Top Cover and Rear Panel Removal, AC•T 8/10 and AC•T diff Analyzers, 4.2-1
4.2-2 Opening the AC•T diff 2 Analyzer’s Front Door, 4.2-3
4.2-3 Top and Rear Center Cover Removal, AC•T diff 2 Analyzer, 4.2-4
4.3-1 Power Supply Module Replacement, 4.3-1
4.4-1 Removing the Software (Flash Memory) Card, 4.4-1
4.4-2 Disconnecting the Analyzer Card, 4.4-2
4.4-3 Disengaging the Analyzer Card from the Mounting Posts, 4.4-3
4.5-1 Disconnecting the Cables to the Display Assembly, 4.5-1
4.5-2 Removing and Disassembling the Display Assembly, 4.5-2
4.6-1 Pneumatic Pump Cover and Bracket Removal AC•T 8/10 and AC•T diff
Analyzers, 4.6-2
4.6-2 KNF Dual-Head Pump with Muffler and Choke Installed, 4.6-3
4.7-1 ASF Thomas Pneumatic Pump Replacement - AC•T diff 2 Analyzer, 4.7-2
4.7-2 KNF Dual-Head Pneumatic Pump Replacement - AC•T diff 2 Analyzer, 4.7-3
4.7-3 KNF Dual-Head Pump with Muffler and Choke Installed, 4.7-4
4.7-4 Locations of CV1, CV2, and CV 3 on the AC•T diff 2 Analyzer, 4.7-5
4.7-5 Pneumatic Compartment Securing Screws, 4.7-6
4.8-1 Traverse Module Disassembling/Replacement, 4.8-2
4.9-1 Disconnecting the Traverse Module Flex Cable, 4.9-2
4.9-2 Disconnecting the Horizontal Motor Cable, 4.9-2
4.9-3 Disconnecting the Traverse Module Tubing, 4.9-3
4.9-4 Removing the Traverse Module, 4.9-3
4.9-5 Removing the Probe-Wipe Housing from the Traverse Housing, 4.9-4
4.9-6 Removing the Probe from the Traverse Housing, 4.9-4
4.9-7 Removing the Belt Clamp, 4.9-5
4.9-8 Removing the Belt, 4.9-5
4.9-9 Removing the Traverse Housing, 4.9-6
4.9-10 Removing the Vertical Motor, 4.9-6
4.9-11 Removing the Traverse Interconnect Card, 4.9-7
4.9-12 Removing the Vertical Rack, 4.9-8
4.9-13 Removing the Horizontal Motor, 4.9-8
4.9-14 Removing a Horizontal Sensor, 4.9-9
4.9-15 Probe Installation, 4.9-12
4.10-1 Removing the Aperture/Bath Assembly Shield, 4.10-1
4.10-2 Removing the Hgb Photometer Assembly, 4.10-2
4.10-3 Aperture Sensor System Components, 4.10-3
4.11-1 Removing the Aperture/Bath Assembly Shield, 4.11-1
4.11-2 Removing the Hgb Photometer Assembly, 4.11-2
4.11-3 Disassembling the Hgb Photometer Assembly, 4.11-3
4.12-1 Latex Calibration Screen - AC•T diff Analyzer, 4.12-1
4.12-2 Latex Calibration Screen - AC•T diff 2 Analyzer, 4.12-1
4.12-3 Latex Results Screen, 4.12-2
4.15-1 Hgb Photometer Shipping Posts, 4.15-3
4.15-2 Securing the Traverse Assembly for Shipping, 4.15-4
4.15-3 Preparing FMI Pumps for Shipment, 4.15-5
4.17-1 Removing the Cap-Pierce Module, 4.17-2
4.17-2 Removing the Cap-Pierce Motor, 4.17-3
4.17-3 Front Door Safety Latch Removal, 4.17-4
PN 4237339F
CONTENTS
TABLES
4.12-1 Latex Adjustment Criteria, 4.12-2
4.13-1 Target Acceptance Criteria, 4.13-1
C.3-1 OKIPAGE14e LED Printer - Specifications, C.3-1
C.3-2 Approved Menu Settings Printed Using the Front Panel, C.3-2
C.3-3 Approved Menu Settings Printed from the Installation Software Print Menu, C.3-3
C.3-4 OKIPAGE 14e LED Printer - Operator Panel Lights - Reference, C.3-4
PN 4237339F
4SERVICE AND REPAIR PROCEDURES 4
4.1 GENERAL GUIDELINES
Field Service Engineers can perform the procedures in this section using their standard tool
kit. Any special tools, supplies or equipment required are identified at the beginning of the
applicable procedure under the Tools/Supplies Needed heading.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS 4
4.2 OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS
Purpose
Use the procedures in this section for Removing the AC•T 8/10 and AC•T diff Analyzers’ Top
Cover and Rear Panel, Opening the AC•T diff 2 Analyzer’s Front Door, or Removing the
AC•T diff 2 Analyzer’s Top and Rear Center Cover to access the instrument’s components.
Removing the AC•T 8/10 and AC•T diff Analyzers’ Top Cover and Rear Panel
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B Small-blade screwdriver
Figure 4.2-1 Top Cover and Rear Panel Removal, AC•T 8/10 and AC•T diff Analyzers
TOP COVER
WASTE CONNECTOR
HOST CONNECTOR
PRINTER CONNECTOR
LYTIC REAGENT
FITTING (2)
WASTE
FITTING
REAR
PANEL
7339012A
PN 4237339F
SERVICE AND REPAIR PROCEDURES
OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS
3. Remove the Power Supply module as directed under Heading 4.3, POWER SUPPLY
MODULE REPLACEMENT.
4. Remove the input/output cables attached to the Analyzer card’s rear connectors
(Figure 4.2-1):
r Top (WASTE) connector - round BNC connector used by the waste-full sensor
r Middle (HOST) connector - 9-pin DB connector used for transmission to a Host
computer
r Bottom (PRINTER) connector - 25-pin DB connector used for a parallel Centronics
Printer.
WARNING Risk of personal injury or contamination. Waste can include biohazardous materials. Handle and
dispose of waste according to acceptable laboratory standards.
5. Disconnect the reagent tubing from the rear-panel bulkhead Luer fittings (Figure 4.2-1):
r Waste tubing - red-striped, attaches at the lower left
r Lytic reagent tubing - translucent EVA, connects to the inner connector labeled 2
r Diluent tubing - clear-coiled tubing, connects to the outer connector labeled 1.
6. Remove the rear panel. The rear panel is secured with eight Phillips-head screws, seven
on the back of the unit and one on the lower right (reagent) side that is used to fasten the
vacuum pump cover (Figure 4.2-1).
Note: Hinged, side-access doors are attached to the rear panel and do not have to be
removed from the rear panel.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS 4
5. From the back of the instrument (Figure 4.2-1), install the top cover:
a. Hook the top cover under the top edge of the front panel.
b. Reinstall the three Phillips-head screws across the back of the instrument.
6. Plug the power line cord into an appropriate wall outlet and turn the instrument on (I).
PN 4237339F
SERVICE AND REPAIR PROCEDURES
POWER SUPPLY MODULE REPLACEMENT 4
4.3 POWER SUPPLY MODULE REPLACEMENT
Purpose
Use this procedure to replace the Power Supply module in any AC•T Series analyzer.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
Removal
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. To access the Power Supply module (Figure 4.3-1), remove the top cover as directed in
the procedure for Removing the AC•T 8/10 and AC•T diff Analyzers’ Top Cover and
Rear Panel or for Removing the AC•T diff 2 Analyzer’s Top and Rear Center Cover under
Heading 4.2.
3. Remove the four Phillips-head screws attaching the Power Supply module to the
instrument (Figure 4.3-1).
4. Pull the Power Supply module out just until you have access to the wiring, and unplug
the two Power Supply module connectors (Figure 4.3-1):
r P5 (9 or 12-pin output connector) - from J5 on the Analyzer card
r P79 (1-pin chassis ground connector) - attaches to a green/yellow chassis ground
wire with connector J79.
5. Remove the Power Supply module.
Chassis ground
connector
J79
P79
Removal/installation
Power ON/OFF screws
7339020A switch
PN 4237339F
SERVICE AND REPAIR PROCEDURES
POWER SUPPLY MODULE REPLACEMENT
Installation
1. Position the new Power Supply module near the back of the instrument.
2. Attach the two Power Supply module connectors (Figure 4.3-1):
r P5 (9 or 12-pin output connector) - from J5 on the Analyzer card
r P79 (1-pin chassis ground connector) - attaches to a green/yellow chassis ground
wire with connector J79.
3. Slide the Power Supply module in place and fasten it using four Phillips-head screws
(Figure 4.3-1).
4. Reinstall the top cover and rear panel as directed in Installation (AC•T 8/10 and
AC•T diff Analyzers) or the top and rear cover as directed in Installation (AC•T diff 2
Analyzer) under Heading 4.2, OPENING OR REMOVING INSTRUMENT DOORS,
PANELS, AND COVERS.
5. Plug the power line cord into an appropriate wall outlet.
Verification
1. Turn the instrument on (l).
2. When the Powerup cycle has completed, select Main icon tt Diagnostics icon tt
Voltages/Sensors icon.
3. Verify that the voltage supplies on the Voltages/Sensors screen are good:
r The Power Supply module supplies these direct voltages: +5 Vdc, ±15 Vdc, and
+24 Vdc.
r The AC•T 8/10 analyzer’s Voltages/Sensors screen does not display +24 Vdc. A quick
check is to verify that the +15 RF voltage is good. It is a 1% precision-regulated
voltage derived from the +24 Vdc supply.
Note: If a more accurate check of the +24 Vdc supply in the AC•T 8/10 analyzer is
needed, check TP5 on the Analyzer card or pin P5-1 with a DVM. TP5 is located
about 2 cm (0.8 in.) above J5. See Figure A.3-1.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
ANALYZER CARD REPLACEMENT 4
4.4 ANALYZER CARD REPLACEMENT
Purpose
Use this procedure to replace the Analyzer card in any of the AC•T Series instruments.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B PC Card Removal tool - AC•T 8/10 analyzer, PN 5450276 (optional)
B Controls or fresh blood specimens with known values (for Verification)
Removal
CAUTION Risk of damage to electronic components. SMT Analyzer cards are very sensitive to electrostatic
discharge (ESD). The surface-mounted components on the top and the bottom of these cards are very
sensitive to mechanical and electronic stresses. When removing, installing, or transporting SMT Analyzer
cards, ensure that you handle them by the edges; and when transporting SMT Analyzer cards, keep them in
their original packaging. The Analyzer cards in the AC•T diff and AC•T diff 2 analyzers are SMT cards.
1. Print out or record all the instrument settings, especially the calibration factors, gains,
and the Clog Detect parameters. They are stored in the non-volatile CMOS memory chip
and must be re-entered when the new Analyzer card is installed.
r For the AC•T 8/10 analyzer - WPWV RM and AVR RM
r For the AC•T diff and AC•T diff 2 analyzers - WBC Target, RBC Target, and AVR
Target.
2. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
3. Open the front door and remove the software (Flash Memory) card (Figure 4.4-1). To
open the front door of the AC•T diff 2 analyzer, see Opening the AC•T diff 2 Analyzer’s
Front Door under Heading 4.2.
7339089E
4. To access the Analyzer card (Figure 4.4-2), remove the top cover and rear panel as
directed in the procedure for Removing the AC•T 8/10 and AC•T diff Analyzers’ Top
Cover and Rear Panel or for Removing the AC•T diff 2 Analyzer’s Top and Rear Center
Cover under Heading 4.2.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
ANALYZER CARD REPLACEMENT
Traverse module
flex cable
Hgb lamp
(J9)
Hgb preamp
connector (J10)
WBC preamp
Preamp shield
5. Disconnect the cable connectors and the preamp ground from the Analyzer card:
Note: If any connectors or wires are not labeled (especially solenoid wires) label them
before disconnecting them.
a. At the top right corner of the Analyzer card (Figure 4.4-2), disconnect the solenoid,
motor, sensor, dc, and pneumatic pump connectors. For the exact locations of these
connectors, refer to Figure A.2-1, Analyzer Card Component Locations - AC•T 8/10
Analyzers; Figure A.2-2, Analyzer Card Component Locations - AC•T diff Analyzer;
or Figure A.2-3, Analyzer Card Component Locations - AC•T diff 2 Analyzer.
b. At the front edge of the card, disconnect the Display assembly ribbon cable and the
Traverse module flex cable (Figure 4.4-2).
c. At the lower front of the card, disconnect the Hgb lamp and detector cables, the
WBC and RBC coaxial aperture bath cables, and the preamp ground lug connector
(Figure 4.4-2).
CAUTION Risk of damage to electronic components. SMT Analyzer cards are very fragile. Flexing the SMT
Analyzer card creates stress on the components. When removing, installing, or transporting SMT Analyzer
cards, handle them by the edges and ensure that you do not flex them.
6. Remove the Analyzer card from its mounting posts (Figure 4.4-3).
Note: The number and type of mounting posts may differ from analyzer to analyzer.
Early AC•T 8/10 analyzers used nylon snap-on mounting posts. The PC Card Removal
tool helps when removing an Analyzer card from nylon mounting posts.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
ANALYZER CARD REPLACEMENT 4
The AC•T diff and AC•T 8/10 analyzers manufactured since the release of the AC•T diff
analyzer are mounted with six, pressure-held metal standoffs and a threaded mounting
post in the center bottom of the card. See Figure 4.4-3 for the location of these seven
mounts.
The Analyzer card on an AC•T diff 2 analyzer is mounted using four, pressure-held metal
standoffs and four threaded mounting posts. See Figure 4.4-3 for the location of these
eight mounts.
a. If the instrument uses threaded mounting posts, unscrew and remove the fasteners
from those posts.
b. Remove the card from the top three mounting posts.
c. Remove the card from the two center posts. The front, center post is located just
below connector J30.
d. Remove the card from the bottom posts.The front bottom post is located just below
the lower front corner of the preamp shield.
Figure 4.4-3 Disengaging the Analyzer Card from the Mounting Posts
Analyzer
card
Snap 0n
mounting
posts
(6)
Threaded
Threaded mounting
mounting posts
post (4)
(1)
7339090E
7. While holding the wires and connectors aside with one hand, slowly remove the card
from the chassis, through the rear of the instrument.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
ANALYZER CARD REPLACEMENT
Installation
1. While holding the wires and connectors aside with one hand, position the Analyzer card
inside the instrument’s chassis.
CAUTION Risk of damage to electronic components. SMT Analyzer card components are very fragile.
Flexing the SMT Analyzer card creates stress on the components. When removing, installing, or
transporting SMT Analyzer cards, handle them by the edges and ensure that you do not flex them. The
Analyzer cards on the AC•T diff and AC•T diff 2 analyzers are SMT cards.
2. Install the Analyzer card onto its mounting posts (Figure 4.4-3):
a. Install the card onto the top three posts. This should line up the two center posts
and the bottom posts.
b. Gently push the center and lower portions of the card until the card is completely
seated on all the mounting posts.
c. If the card has threaded-metal, mounting posts, secure the card with the original
fasteners.
3. Connect the bath assembly ground wire to the Analyzer card’s ground lug connector,
located on or near the preamp shield (Figure 4.4-2). If this connection feels loose, gently
squeeze the wire lug with pliers to make a tighter connection.
Note: Electrical noise problems occur, especially for Plt and WBC counts, if the ground
lug connection is not firm.
4. Connect the RBC and WBC coaxial aperture bath cables, then the coaxial Hgb connector
to J10 and the four-wire Hgb lamp connector, P9, to J9 (Figure 4.4-2).
5. Connect the Traverse module flex cable and the Display assembly ribbon cable
(Figure 4.4-2).
6. Connect the solenoid, motor, sensor, dc, and pneumatic pump cables to the Analyzer
card. Refer to Figures A.2-1, Analyzer Card Component Locations - AC•T 8/10
Analyzers; A.2-2, Analyzer Card Component Locations - AC•T diff Analyzer; or A.2-3,
Analyzer Card Component Locations - AC•T diff 2 Analyzer, for the locations of these
connectors.
7. Ensure all the connectors are plugged in.
8. For the AC•T 8/10 and AC•T diff analyzers only, reinstall the rear panel as directed in
Installation (AC•T 8/10 and AC•T diff Analyzers) under Heading 4.2, OPENING OR
REMOVING INSTRUMENT DOORS, PANELS, AND COVERS.
9. Reinstall the software (Flash Memory) card.
10. Plug the power line cord into an appropriate wall outlet.
Verification
1. Turn the instrument on (I):
r If the time and date warning is issued, it means the Analyzer card has new or cleared
CMOS and default values were written for all the settings. Press the Continue icon.
r If errors occur, troubleshoot the errors.
r If the instrument attains a ready state with no errors, the motors and sensors are
connected properly and have proper power. Go to step 2.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
ANALYZER CARD REPLACEMENT 4
ATTENTION: Check the instrument settings carefully. If this Analyzer card was installed in
another instrument, the settings for that instrument are not automatically written over.
2. When the instrument reaches the “ready” state, select Main icon tt Settings icon and
check or re-enter the customer’s settings recorded in step 1 of Removal.
3. If the instrument did not perform a Startup on Powerup cycle, or failed background, run
a Startup cycle. Rerun as necessary until the backgrounds are acceptable. Depending on
the condition of the instrument before the Analyzer card was replaced, this may take
several attempts.
4. Select Main icon tt Diagnostics icon tt Voltages/Sensors icon. Ensure all the voltages are
present and all the sensors indicate their current state correctly.
ATTENTION: To avoid interference from electrical noise during the Pulse test, ensure the bath
system is completely primed with diluent and the bath shield is in place.
5. Exit the Voltages/Sensors screen and select the Pulse icon.
a. Use the default values.
b. Ensure that values other than 0.0 are obtained for all four parameters.
6. Reinstall the top cover as directed in Installation (AC•T 8/10 and AC•T diff Analyzers)
or the top and rear cover as directed in Installation (AC•T diff 2 Analyzer) under
Heading 4.2, OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND
COVERS.
7. Verify the instrument operation by running controls or fresh blood with known values.
Run the specimens several times to verify precision as well as accuracy.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
DISPLAY ASSEMBLY - DISASSEMBLING/ASSEMBLING 4
4.5 DISPLAY ASSEMBLY - DISASSEMBLING/ASSEMBLING
Purpose
Use the procedure in this section to completely disassemble then reassemble the Display
assembly. When replacing the entire Display assembly, only portions of this procedure are
required.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B Fine, needle-nose pliers
B Small flat-blade screwdriver (AC•T diff 2 analyzer only)
Removal
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. Open the front door. To open the front door of the AC•T diff 2 analyzer, see Opening the
AC•T diff 2 Analyzer’s Front Door under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
3. At the Display assembly, disconnect P61 (Figure 4.5-1), the main display ribbon cable
from the Analyzer card.
4. At the Display assembly, disconnect the reagent management card reader flex cable
(Figure 4.5-1). This flex cable pulls out of connector J66, leaving the entire connector on
the Display Controller Driver card.
5. Disconnect the chassis ground strap attached to the Display assembly (Figure 4.5-1) by
removing a nut at either end of the chassis ground strap.
P61
7339091E
PN 4237339F
SERVICE AND REPAIR PROCEDURES
DISPLAY ASSEMBLY - DISASSEMBLING/ASSEMBLING
6. To remove an AC•T 8/10 or AC•T diff analyzers’ Display assembly from the front door:
a. Beginning with the top two screws, remove the four outer screws securing the
Display assembly to the front door (Figure 4.5-2).
.
CAUTION On the AC•T 8/10 and AC•T diff analyzers, the power on LED connector, P64, is still attached. If
you move the Display assembly completely away from the front door, you could damage the power on
LED’s connection to the Display assembly. Hold the Display assembly near the front door.
b. When the Display assembly is free, hold it near the front door.
c. Disconnect the power on LED connector, P64, from the Display Controller Driver
card (Figure 4.5-2).
Power on
LED cable, P64
LCD
J65
J63
J64
Display Controller
7339092E
Driver card
7. To remove an AC•T diff 2 analyzer’s Display assembly from the front door:
a. Using a small flat-blade screwdriver, remove the power on LED connector, P64,
from the Display Controller Driver card.
b. Beginning with the top two screws, remove the four outer screws securing the
Display assembly to the front door (Figure 4.5-2).
8. If replacing the entire Display assembly, go to Installation to install the new assembly.
Otherwise go to step 9.
9. Remove the four remaining screws attaching the shield to the rest of the Display
assembly and set the shield aside (Figure 4.5-2).
PN 4237339F
SERVICE AND REPAIR PROCEDURES
DISPLAY ASSEMBLY - DISASSEMBLING/ASSEMBLING 4
10. Disassemble the Display assembly:
a. Disconnect the two touch screen flex cables from J63 on the Display Controller
Driver card (Figure 4.5-2).
b. Disconnect the LCD high voltage cable from CN2 on the Display Controller Driver
card.
c. Remove the four Phillips-head screws holding the assembly together (Figure 4.5-2).
r The Phillips-head screws are accessed on the touch screen side of the assembly
and fastened to threaded hex posts on the Display Controller Driver card side.
r Each screw has two spacers, one separating the touch screen and the LCD, the
other separating the LCD and the Display Controller Driver card.
11. Set aside the touch screen.
12. Carefully disconnect the LCD plug from connector J62 on the Display Controller Driver
card (Figure 4.5-2).
13. Carefully separate the LCD from the Display Controller Driver card (Figure 4.5-2).
Installation
1. Assemble the Display assembly (Figure 4.5-2):
a. Plug the LCD cable into J62 on the Display Controller Driver card.
b. Secure the four Phillips-head screws holding the assembly together.
r The Phillips-head screws are accessed on the touch screen side of the assembly
and fastened to threaded hex posts on the Display Controller Driver card side.
r Each screw has two spacers, one separating the touch screen and the LCD, the
other separating the LCD and the Display Controller Driver card.
c. Connect the LCD high voltage cable to CN2 on the Display Controller Driver card.
d. Connect the two touch screen flex cables to J63 on the Display Controller Driver
card.
2. If the Display Controller Driver card (Figure 4.5-2) uses the original through-hole
technology, ensure the jumpers on the daughter card are set as specified in Table A.2-17,
Display Controller Driver Card Jumpers.
Note: Jumpers are eliminated on the SMT version of the Display Controller Driver card.
3. Locate J66 on the Display Controller Driver card (Figure 4.5-2).
J66, the reagent card flex cable connector, is a small white assembly made up of two
parts. The base of the connector plugs into pins on the Display Controller Driver card.
The upper part clips onto the base and slides out about 3 mm (0.01 in.).
4. Slide the upper part of J66 away from the base, leaving the base installed on the Display
Controller Driver card. This is the position J66 must be in to accept the flex cable from
the reagent management card reader. Once the shield is in place, it is difficult to slide this
portion up.
5. On an AC•T 8/10 or AC•T diff analyzer, connect the power on LED connector, P64, to
J64 on the Display Controller Driver card (Figure 4.5-2).
6. Attach the four screws installing the shield to the Display assembly. Do not tighten the
screws.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
DISPLAY ASSEMBLY - DISASSEMBLING/ASSEMBLING
Verification
1. Turn the instrument on (I). Ensure that the green power on LED lights.
2. As the instrument boots up, watch to see if the LCD backlights, and then watch the
screen to see if the Beckman Coulter copyright message is displayed.
3. When the instrument status is ready, move through all the menu items, touching each
one and ensuring you have full touch-screen control.
a. Using all numerals in the Sample Results, ID entry screen, will test a large portion of
screen.
b. While using the touch screen, listen for sounds indicating that a touch screen press
was received.
4. At the Sample Results screen, make sure the reagent management card reader icon is not
displayed in the lower status area of the screen. If the icon is displayed, touch it to
remove it. If the instrument cannot read the reagent management card, it cannot run any
samples.
5. At the front of the instrument, ensure that the Display assembly is visually centered. If it
is not, loosen the four shield screws and/or the four Display assembly screws, realign the
Display assembly until it appears centered and then tighten the appropriate screws.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PNEUMATIC PUMP REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS 4
4.6 PNEUMATIC PUMP REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS
ATTENTION: Chapter 4 contains two procedures for replacing pneumatic pumps.
r Use this procedure to replace a pneumatic pump in an AC•T 8/10 or AC•T diff analyzer.
r Use the procedure under Heading 4.7, PNEUMATIC PUMP REPLACEMENT - AC•T
diff 2 ANALYZER, to replace a pneumatic pump in an AC•T diff 2 analyzer.
Purpose
In the field, three models of pneumatic pumps are currently in use in the AC•T 8/10 and
AC•T diff analyzers:
r The Brailsford pump, which is now obsolete. If this pump becomes defective, you can
replace it with either an ASF Thomas pump or a KNF dual-head pump.
r The ASF Thomas pump.
r The KNF dual-head pump. This pump is used on new AC•T 8/10 or AC•T diff analyzers
and as an upgrade for instruments installed at high altitudes.
Note: The Brailsford and ASF Thomas pumps cannot build the necessary pressure to
produce adequate mixing when installed at high altitudes. For instruments installed at
1524 m (5,000 ft) or above, you must upgrade the pump to a KNF dual-head pump.
The KNF pump has two pump heads driven by the same motor. One head is used for
vacuum. A muffler is connected to its output port. The other head is used for pressure. A
choke is connected to its input port to decrease the air pressure at its output.
The following procedure can be used for any of these pneumatic pumps.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B Flat-blade screwdriver or needle-nose pliers
B Wire cutters
For field upgrades to a KNF dual-head pump, you will also need the following:
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PNEUMATIC PUMP REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS
Removal
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. Open the left side door of the analyzer and place the internal reagents on the bench.
3. Remove the three Phillips-head screws securing the pneumatic pump cover
(Figure 4.6-1).
Note: The cover forms the platform that the internal reagents are placed on.
Figure 4.6-1 Pneumatic Pump Cover and Bracket Removal AC•T 8/10 and AC•T diff Analyzers
Bracket securing
screws
Bracket
Card
securing
Cover screws
4. If you are removing a Brailsford pump, remove the top cover of the analyzer as directed
in Removing the AC•T 8/10 and AC•T diff Analyzers’ Top Cover and Rear Panel under
Heading 4.2, OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND
COVERS.
5. Disconnect the vacuum and pressure output tubes from the pump’s fittings using a
flat-blade screwdriver or needle-nose pliers.
If the pump is a KNF dual-head pump with a choke installed on one of its ports, remove
that tubing/choke assembly also and set it aside.
6. If the pump is a KNF dual-head pump, remove the two Phillips-head screws securing the
Pump Motor Filter card to the pump’s bracket (Figure 4.6-2), and set the card aside.
7. Disconnect the pump’s electrical connector.
r If the pump is a Brailsford pump, disconnect its cable from LV3 on the
Analyzer card.
For the location of the connector, refer to Figure A.2-1, Analyzer Card Component
Locations - AC•T 8/10 Analyzers, Figure A.2-2, Analyzer Card Component
Locations - AC•T diff Analyzer, or Figure A.2-4, Universal AC•T diff/AC•T diff 2
Analyzer Card Component Locations.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PNEUMATIC PUMP REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS 4
r If the pump is an ASF Thomas pump, disconnect the spade lugs from the spade
terminals on the pump.
r If the pump is a KNF dual-head pump, disconnect its cable from the cable on the
Pump Motor Filter card.
8. Remove the two Phillips-head screws holding the pneumatic pump’s bracket in place
(Figure 4.6-1).
Note: If you are replacing a KNF dual-head pump that was installed as a field upgrade,
only one Phillips-head screw is securing the bracket.
9. Remove the pump from the instrument.
Installation
IMPORTANT Using the incorrect choke will result in improper mixing of the dilutions in the bath. Be sure to
use the black (0.010 i.d. orifice) choke for altitudes of 1524 m (5,000 ft) and above, the brown (0.006 i.d.
orifice) choke for lower altitudes.
1. If you are installing a KNF dual-head pump, attach a tubing/choke assembly to the IN
(∧) port on the pump’s pressure head. See Figure 4.6-2.
a. If a tubing/choke assembly is available from the old pump:
1) Remove the assembly from the old pump.
2) Replace the choke in the assembly with a new choke.
3) Install the tubing/choke assembly on the new pump.
b. If a tubing/choke assembly is not available:
1) Place an O-ring over the threaded end of the barb fitting and screw the fitting
onto the appropriate choke.
2) Attach a 5.08 cm (2 in.) piece of the 0.145 i.d. polyurethane tubing to the barb
end of the fitting.
3) Attach the other end of the tubing to the new pump.
Figure 4.6-2 KNF Dual-Head Pump with Muffler and Choke Installed
Grommet
Bracket
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PNEUMATIC PUMP REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS
Verification
1. Plug the power line cord into an appropriate wall outlet.
2. Turn the instrument on (I). Listen for the pneumatic pump to turn on during powerup.
3. Press Main icon tt Diagnostics icon tt Voltages/Sensors icon. Verify that the pneumatic
pump has turned on. If necessary, adjust the low vacuum reading to 6.00.
4. If you have not done so, install the pneumatic pump cover and close up the instrument.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS 4
4.8 TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff
ANALYZERS
Purpose
Use the procedure in this section to completely disassemble/reassemble the Traverse module
in an AC•T 8/10 or AC•T diff analyzer. Replacing individual parts on the Traverse module
requires only portions of the procedure.
Tools/Supplies Needed
B #1 Phillips-head screwdriver
B #2 Phillips-head screwdriver
B Large, flat-blade screwdriver
B DOW CORNING 33 lubricant, PN 1604007-0
Removal
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. Remove the probe-wipe housing (Figure 4.8-1):
a. Manually remove the metal clip.
b. Pull the probe-wipe housing out the bottom of the traverse housing.
c. Remove the tubing from the probe-wipe housing.
3. Remove the aspirate probe from the traverse housing (see Figure 4.8-1):
a. Manually move the probe and vertical rack until the bottom of the rack (where the
probe is press fit into the vertical rack) is visible through the traverse housing
window.
b. Pry the probe out of the rack using a large, flat-blade screwdriver.
c. Pull the probe out of the bottom of the traverse housing.
d. Remove the aspirate tubing from its top fitting.
ATTENTION: Perform the next step only if the entire Traverse module is to be replaced.
4. Disconnect the traverse flex cable from the Traverse Interconnect card.
5. Remove the Traverse module from the instrument as a unit, by removing three
Phillips-head screws attaching the traverse bracket to the front panel (Figure 4.8-1).
6. Remove the horizontal rack and the traverse housing (Figure 4.8-1):
a. Unscrew one (the only) Phillips-head screw on the right end of the horizontal rack.
b. When free, slide the rack right until the left end disengages from the traverse
bracket.
c. Remove the horizontal rack from the traverse bracket.
d. Remove the traverse housing from the horizontal rack.
PN 4237339F
SERVICE/REPAIR PROCEDURES
TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS
TRAVERSE
FRONT BRACKET
PANEL
HORIZONTAL
RACK
TRAVERSE
HOUSING VERTICAL
MOTOR
STAINLESS STEEL
WEAR STRIP
SHORT
OUTSIDE
ASPIRATE
SCREW
VERTICAL TUBING
CLIP
SENSOR FLAG
HORIZONTAL
SENSOR
RIBBON VERTICAL
TUBING SENSOR
CLAMP
STAINLESS
STEEL
WEAR TRAVERSE
STRIP INTERCONNECT
CARD
VERTICAL
RACK
J59
J58
J57
J55
HORIZONTAL FLEX
MOTOR CABLE
CONNEC-
TOR
O-RING
MIDDLE
PROBE TUBING OF RIBBON
WIPE PROBE
LARGER OUTSIDE
HOUSING
TUBING OF RIBBON
7339027C
PN 4237339F
SERVICE/REPAIR PROCEDURES
TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS 4
7. Disconnect the motor and sensor cables from the Traverse Interconnect card
(Figure 4.8-1):
a. P58 (vertical sensor plug) wiring is routed across the top of the housing and plugs
into J58, the second connector from the top.
b. P59 (horizontal sensor plug) wiring is routed up, then across and down with the
other wiring, and plugs into J58, the top connector on the Traverse Interconnect
card.
8. Remove the Traverse Interconnect card. It is press fit into the housing and does not use
any fasteners (Figure 4.8-1).
9. Remove the horizontal and vertical motors and sensors (Figure 4.8-1):
r The motors are fastened to the housing with two Phillips-head screws each.
r The sensors are snap fit onto the traverse housing and can be pulled off by hand.
10. Remove the vertical sensor from the top of the vertical rack. It snaps onto the vertical
rack without fasteners (Figure 4.8-1).
11. Remove the vertical rack by sliding it out the bottom of the traverse housing
(Figure 4.8-1).
Installation
1. Install the vertical rack into the traverse housing (Figure 4.8-1):
r The vertical rack slides in from the bottom of the traverse housing.
r A guide slot in the vertical rack is positioned at the back.
r The slotted end of the vertical rack (flush to end of rack) is positioned up.
r The gear teeth are to the left.
2. Attach the vertical sensor flag (Figure 4.8-1). It snaps onto the top of the vertical rack
and is positioned in the vertical sensor.
3. Install the vertical and horizontal dual-channel position sensors by snapping them into
place on the traverse housing.
ATTENTION: Although the vertical motor is fastened with two Phillips-head screws, it requires
that you use one shorter screw in the outer position. This is necessary because a longer screw
in the outer position would interfere with the vertical sensor.
4. Secure the vertical and horizontal motors with two Phillips-head screws each
(Figure 4.8-1).
5. Install the Traverse Interconnect card. This card is press fit onto the traverse housing
using two inner fingers on the back of the card and two outer fingers on the component
side of the card (Figure 4.8-1).
6. Connect the vertical and horizontal sensors to the Traverse Interconnect card
(Figure 4.8-1):
a. P58 (vertical sensor plug) wiring routes across the top of the housing and plugs into
J58, the second connector from the top.
b. P59 (horizontal sensor plug) wiring routes up, then across and down with the other
wiring, and plugs into J58, the top connector on the Traverse Interconnect card.
PN 4237339F
SERVICE/REPAIR PROCEDURES
TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS
ATTENTION: Dress the sensor and motor wires as efficiently as possible and tie wrap them
together. These wires move with the traverse housing. Loose wires eventually snag and cause
fatal, probe-position errors.
7. Connect the traverse (vertical and horizontal) motors (Figure 4.8-1)
r Vertical and horizontal connector plugs are P55 and P57 respectively.
r Wires are routed with vertical sensor wire.
8. Engage the horizontal rack (Figure 4.8-1).
r The rack teeth are facing up.
r The right end is threaded.
9. Install the traverse housing onto the traverse bracket (Figure 4.8-1):
a. Ensure that the two stainless steel wear strips are properly in place in the traverse
housing.
b. From inside, slide the left end of the horizontal rack through the left end of the
traverse bracket.
c. Line up the bottom travel guides.
d. Lower the traverse housing onto the front guide edge of the traverse bracket just
until the Phillips-head screw can be fastened to the right end of the horizontal rack.
10. Install the aspirate tubing (Figure 4.8-1):
a. Guide the smaller aspirate tubing behind the vertical sensor.
b. Using a small-blade screwdriver, push the aspirate tubing into the molded tubing
clamp just above the vertical sensor. There should be 24 cm (9.5 in.) of tubing
protruding above the clamp.
c. Feed the tubing into the top of the rack and through to the bottom of the traverse
housing.
11. Install the aspirate probe (Figure 4.8-1):
a. Apply a small amount of silicon lubricant onto the probe mounting O-ring.
b. Attach the probe to the aspirate tubing.
c. Push the probe up until it pushes the vertical rack up. Holding the vertical rack with
one hand and the probe with the other hand, push the probe into the vertical rack.
12. Install the probe-wipe housing (Figure 4.8-1):
a. Slide the probe-wipe housing onto the bottom of the probe with the slotted end up
and the fitting at the back.
b. Move the probe-wipe housing and the probe, up into the traverse housing.
c. At the front, fasten the probe-wipe housing in place using the wire clip.
13. Attach the probe-wipe tubing (Figure 4.8-1):
r The middle tube in the tubing ribbon connects to the top fitting.
r The larger outside (non-aspirate) tubing connects to the bottom fitting.
14. Place the tubing ribbon into its retaining clamp at the back of the traverse housing. The
tubing ribbon runs along the front panel from the right, runs behind the traverse
housing, and then loops back to the traverse housing. Slide it down into its retaining
clamp near the split of the three tubings.
PN 4237339F
SERVICE/REPAIR PROCEDURES
TRAVERSE MODULE DISASSEMBLY/REPLACEMENT - AC•T 8/10 AND AC•T diff ANALYZERS 4
Verification
1. Manually move the probe up and down, ensuring that it moves smoothly without
binding and that nothing interferes with its travel.
2. Manually move the horizontal rack to the WBC bath position and back. Ensuring that
the horizontal rack moves smoothly without binding by confirming that:
r The motor and sensor wires do not snag on anything.
r The flex cable flows and bends without difficulty.
r The ribbon tubing has enough play in it for travel to each end of horizontal
movement. If necessary, adjust the traverse and front-panel tubing clamps.
3. Plug the power line cord into an appropriate wall outlet.
4. Turn the instrument on (l). If the instrument does not perform a startup producing
background results, initiate a startup.
5. If no probe position errors are encountered during startup, the instrument is ready to
be used.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
APERTURE/BATH ASSEMBLY REPLACEMENT 4
4.10 APERTURE/BATH ASSEMBLY REPLACEMENT
Purpose
Use this procedure to replace an RBC or WBC aperture/bath assembly that is defective, or that
has an aperture too large for use with the AC•T diff/Veterinary Options analyzer software.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B Specimen with known values (for Verification)
Removal
1. Select Main icon tt Diluter Functions icon tt Drain icon.
2. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
3. Remove the top cover as directed in the procedures for Removing the AC•T 8/10 and
AC•T diff Analyzers’ Top Cover and Rear Panel, or Removing the AC•T diff 2 Analyzer’s
Top and Rear Center Cover, under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
4. Disconnect the aperture (white coaxial) cables plugged into the preamp, a shielded area
at the lower front of the Analyzer card:
r The WBC cable connects through the top hole in the preamp shield.
r The RBC cable connects through the bottom hole in the preamp shield.
5. Open the front door. To open the front door of the AC•T diff 2 analyzer, see Opening the
AC•T diff 2 Analyzer’s Front Door under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
6. Unfasten the two thumb screws securing the aperture/bath assembly shield to the front
panel (Figure 4.10-1) and remove the shield.
Bath
electrical
shield
7339093E
PN 4237339F
SERVICE AND REPAIR PROCEDURES
APERTURE/BATH ASSEMBLY REPLACEMENT
7. Disconnect the three tubes going to the bottom of the aperture/bath assembly being
replaced.
8. On the AC•T diff 2 analyzer, disconnect the diluent tube from the fitting at the top of the
bath.
WARNING Risk of personal injury. The hemoglobin lamp is hot. You might be burned if you handle the
lamp while it is hot. Let the lamp cool before removing the Hgb photometer assembly.
9. If replacing the WBC aperture/bath assembly, remove the Hgb photometer assembly.
a. Remove the Phillips-head screw securing the Hgb photometer assembly to the WBC
bath (Figure 4.10-2).
b. If the two shipping posts are still attached, remove them (Figure 4.10-2).
Note: The shipping posts, which consist of a screw and a spacer, do not have to be
installed unless you are shipping the instrument.
c. Slide the Hgb photometer assembly off the front of the WBC bath.
Hgb photometer
securing screw
Shipping
screws
7339094E
10. Manually remove the aperture/bath assembly from its aperture bath clamp
(Figure 4.10-3). It snaps in and out of the clamp.
11. Disconnect the aperture/bath assembly’s rear-chamber tubing (Figure 4.10-3):
r RBC bath
t Red-striped tubing from LV16 connects to the upper fitting.
t Sweep-flow tubing attaches to the lower fitting.
r WBC bath - Red-striped tubing from LV17 connects to the rear chamber fitting
PN 4237339F
SERVICE AND REPAIR PROCEDURES
APERTURE/BATH ASSEMBLY REPLACEMENT 4
12. Remove the aperture/bath assembly by carefully guiding the coaxial-cable connector of
the aperture cable through the opening in the front panel. A rubber grommet comes with
the cable. Remove the rubber grommet from the old assembly and use it on the
replacement bath.
5
WBC
bath
LV16
LV17
Sweep flow
spool
7339096E
Installation
1. Install the aperture/bath assembly by carefully guiding the coaxial-cable connector of the
aperture cable through the appropriate opening in the front panel and position the
rubber grommet on the front panel.
ATTENTION: If the aperture/bath assembly’s rear-chamber tubing is not reconnected properly,
the tubes may crimp, and impede flow. Correct routing is from rear chamber fitting, up
through the opening in the machined clamp or around the stem of the molded clamp, and
then down to LV16 or LV17. This routing ensures a proper loop that will not crimp tubing.
2. Attach the aperture/bath assembly’s rear-chamber tubing (Figure 4.10-3):
r If the WBC aperture/bath assembly was replaced, connect the red-striped tubing
from LV17 to the bath’s rear-chamber fitting.
r If the RBC aperture/bath assembly was replaced:
t Connect the red-striped tubing from LV16 to the upper fitting of the bath’s rear
chamber.
t Connect the sweep-flow tubing to the lower fitting of the bath’s rear chamber.
3. Snap the aperture/bath assembly into the aperture bath clamp, orienting it so that the
bath’s rear chamber is at the back.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
APERTURE/BATH ASSEMBLY REPLACEMENT
4. If the WBC aperture/bath assembly was replaced, install the Hgb photometer assembly.
a. Slide the Hgb photometer assembly onto the front of the WBC bath.
CAUTION Risk of damage to the WBC bath. When securing the Hgb photometer assembly to the WBC bath,
if you use a screw that is too long, the screw can break through the bath wall. Use the correct length screw
when securing the Hgb photometer assembly to the WBC bath.
b. Secure the Hgb photometer assembly using the appropriate length Phillips-head
screw (Figure 4.10-2).
c. If shipping the instrument, reinstall the two shipping posts (Figure 4.10-2).
5. Attach the tubings to the bath being replaced.
a. For either the RBC or the WBC aperture/bath assemblies, connect the waste
tubing to the bottom fitting of the bath.
b. For an RBC aperture/bath assembly, diluent flow is divided by two Y-fittings into
three paths.
1) Connect the diluent tubing with the small check valve in the line to the top
fitting on the bath.
2) Connect the other two diluent lines to the two lower side fittings of the bath.
c. For a WBC aperture/bath assembly, diluent flow is divided by a Y-fitting into two
paths.
1) Connect the diluent tubing with the small check valve in the line to the top
fitting on the bath.
2) Connect the other diluent line to the lower-left side fitting on the bath.
3) Connect the lytic reagent line to the lower-right side fitting on the bath.
6. Attach the aperture (white coaxial) cables:
a. The WBC cable connects through the top hole in the preamp shield.
b. The RBC cable connects through the bottom hole in the preamp shield.
7. Position the electrode:
a. Manually move the probe to the bath that was replaced.
b. Manually move the probe down into the center of the bath.
c. Shape and position the bath electrode so that the probe does not touch the electrode
when it goes down into the center of the bath. This is necessary because the probe
thief position is lower than the level of the electrode during a normal cycle.
d. Ensure that the electrode does not touch the side of the bath when it is bent off
center.
8. Reinstall the top cover and rear panel as directed in Installation (AC•T 8/10 and
AC•T diff Analyzers), or Installation (AC•T diff 2 Analyzer), under Heading 4.2,
OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS.
9. Plug the power line cord into an appropriate wall outlet.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
APERTURE/BATH ASSEMBLY REPLACEMENT 4
Verification
1. Turn the instrument on (l). Observe the baths during power up and ensure that there are
no leaks and fluid cycling seems correct. If a Startup cycle is performed, disregard the
results.
2. If a Startup cycle was not performed during power up, run a cycle and ensure that the
electrode is not touched by the probe when the probe is down in the bath.
IMPORTANT Electrical interference occurs when the bath’s electrical shield is not covering the
aperture/bath assembly. This interference will cause sample results to be unacceptable. Always ensure that
the bath’s electrical shield is in place before running a sample.
3. Install the aperture/bath assembly shield, being careful not to pinch any tubing, and
close the front door. Refer to Figure 4.10-1.
4. Perform a Startup cycle and ensure that the backgrounds are acceptable.
5. For instruments using the AC•T diff/Veterinary Options analyzer software, verify the
aperture is not too large as directed under Heading 4.16, APERTURE VERIFICATION -
AC•T diff/Veterinary Option ANALYZER.
6. For AC•T diff and AC•T diff 2 analyzers, check the gain settings as directed under
Heading 4.12, LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS.
7. Run a specimen with known values and verify that the results are acceptable.
8. Perform a verification and/or calibration.
Note: Replacing an aperture often requires calibration which should be performed only
by the appropriate personnel according to local regulations and laboratory protocols.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
HGB PHOTOMETER ASSEMBLY - DISASSEMBLY/REPLACEMENT 4
4.11 HGB PHOTOMETER ASSEMBLY - DISASSEMBLY/REPLACEMENT
Purpose
Use this procedure to remove, disassemble and reassemble the Hgb photometer assembly
when the Hgb lamp, clear lens, blue heat filter, or O-ring needs replaced.
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B A control or a fresh blood specimen with a known value (for Verification)
Removal
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. Remove the top cover as directed in the procedures for Removing the AC•T 8/10 and
AC•T diff Analyzers’ Top Cover and Rear Panel, or Removing the AC•T diff 2 Analyzer’s
Top and Rear Center Cover, under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
3. Open the front door. To open the front door of the AC•T diff 2 analyzer, see Opening the
AC•T diff 2 Analyzer’s Front Door under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
4. Unfasten the two thumb screws securing the aperture/bath assembly shield to the front
panel (Figure 4.11-1) and remove the shield.
Bath
electrical
shield
7339093E
5. Disconnect the Hgb lamp cable, P9, and the Hgb preamp coaxial cable, P10, from the
Analyzer card connectors, J9 and J10, respectively. For the locations of the connectors,
see Figures A.2-1, Analyzer Card Component Locations - AC•T 8/10 Analyzers, A.2-2,
Analyzer Card Component Locations - AC•T diff Analyzer, or A.2-3, Analyzer Card
Component Locations - AC•T diff 2 Analyzer.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
HGB PHOTOMETER ASSEMBLY - DISASSEMBLY/REPLACEMENT
WARNING Risk of personal injury. The hemoglobin lamp is hot. You might be burned if you handle the
lamp while it is hot. Let the lamp cool before removing the Hgb photometer assembly.
6. Remove the Phillips-head screw fastening the Hgb photometer assembly to the WBC
bath (Figure 4.11-2).
7. If the two shipping posts are still attached, remove them. (Figure 4.11-2).
Note: The shipping posts, which consist of a screw and a spacer, do not have to be
installed unless you are shipping the instrument.
Shipping
posts
Hgb photometer
securing screw
Shipping
screws
7339094E
8. Remove the Hgb photometer from the WBC bath, and pull the Hgb preamp and the Hgb
lamp cables through the front panel.
Note: It is often easier to remove the rubber grommet from the panel with the cables, and
then take the grommet off the cables.
9. Disassemble the Hgb photometer (Figure 4.11-3) by removing the following
components from the Hgb detector housing:
Note: Disassemble carefully. These components are small and can be easily lost.
a. Remove the lamp assembly by pushing the assembly into the housing and twisting
counterclockwise.
b. Remove the clear lens.
c. Remove the blue heat filter.
d. Remove the compression O-ring.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
HGB PHOTOMETER ASSEMBLY - DISASSEMBLY/REPLACEMENT 4
Figure 4.11-3 Disassembling the Hgb Photometer Assembly
O-ring
Lens
Heat
filter
Lamp
7339095E holder/heatsink Lamp
10. Remove the lamp from its heatsink housing by loosening the Allen setscrew and pulling
the lamp out. Refer to Figure 4.11-3.
Installation
1. Insert the lamp into the back end of the lamp housing and the heatsink until the back
end of the lamp is flush with the back end of the heatsink. Tighten the lamp in place with
the Allen setscrew (Figure 4.11-3).
2. Assemble the photometer (Figure 4.11-3) by inserting the following into the Hgb
detector housing:
a. Insert the clear lens.
b. Insert the blue heat filter.
c. Insert the compression O-ring.
d. Push the lamp housing into the Hgb detector housing and twist counterclockwise.
Note: With a fresh O-ring, the lamp housing pins usually rest in the center notch
and provide good tension.
CAUTION Risk of damage to the WBC bath. When securing the Hgb photometer assembly to the WBC bath,
if you use a screw that is too long, the screw can break through the bath wall. If you use a screw that is too
short, you can cause damage to the plastic threads. Use the correct length screw when securing the Hgb
photometer assembly to the WBC bath.
3. Slide the Hgb photometer assembly onto the WBC bath. Tighten in place using the
appropriate length Phillips-head screw (Figure 4.11-2).
4. If shipping the instrument, install the two shipping posts (Figure 4.11-2).
5. Dress the Hgb preamp and the Hgb lamp cables to the rear of bath. Place a rubber
grommet over both cables if they were removed from the panel.
6. Feed both cables through the panel. Install the grommet if it is not in place.
7. Connect cables, P9 and P10, to the Analyzer card connectors, J9 and J10, respectively.
For the locations of the connectors, see Figures A.2-1, Analyzer Card Component
Locations - AC•T 8/10 Analyzers, A.2-2, Analyzer Card Component Locations - AC•T
diff Analyzer, or A.2-3, Analyzer Card Component Locations - AC•T diff 2 Analyzer.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
HGB PHOTOMETER ASSEMBLY - DISASSEMBLY/REPLACEMENT
Verification
1. Turn the instrument on (I).
2. When the instrument reaches the “ready” state, select Main icon tt Diagnostics icon tt
Voltages/Sensors icon.
3. Change the Preamp and/or Hgb Gain adjustment. Verify that the Hgb output moves
up/down as you move the gain up/down. Set the Hgb output to3700 ±100 using either
the Preamp Gain or the Hgb Gain controls.
Note: The Analyzer cards used by the AC•T diff and AC•T diff 2 analyzers have only one
amplifier, the Hgb Gain adjustment, for setting the Hgb output.
4. Select the Save and Exit icon.
5. Run a control or a fresh blood specimen with a known value and verify that the Hgb
parameter is recovered properly.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS 4
4.12 LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS
ATTENTION: On instruments using AC•T diff/Veterinary Option analyzer software, you must
verify the apertures are not too large before adjusting the gains. Do the procedure under
Heading 4.16, APERTURE VERIFICATION - AC•T diff/Veterinary Option ANALYZER,
before doing this gain adjustment procedure.
Purpose
Use this procedure to adjust the gains of the amplifiers for the pulse signals received from the
apertures. This procedure compensates for variations in aperture size that could affect
size-sensitive parameters derived from histograms like the WBC differential and MCV.
Note: On instruments using both AC•T diff analyzer and AC•T diff/Veterinary Option
analyzer software, you can set the gains using either software and no additional adjustments
are needed to use the other software. The higher gains needed for the various animal species
are accomplished by the software based on the gains set in the LATEX GAIN ADJUSTMENT -
AC•T diff AND AC•T diff 2 ANALYZERS procedure for normal laboratory use.
Tools/Supplies Needed
B Latex particles, PN 6857371-8
Procedure
1. Ensure that the apertures are clean. If protein buildup is suspected, bleach the apertures
as directed under Clean the Baths, in the Special Procedures and Troubleshooting
manual.
2. Select Main icon tt Diagnostics icon tt Latex Calibration icon. An intermediate screen
(Figure 4.12-1 for an AC•T diff analyzer or
displayed indicating the next action to perform.
7339053E
PN 4237339F
SERVICE AND REPAIR PROCEDURES
LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS
3. The top numerical entry box is highlighted showing the particle size (default = 94.3 fL).
Enter the particle size of the latex particles you are using.
Note: The Save icon is not needed. The Save icon is used for the manual entry of the gain
values below the particle size.
4. Select Continue icon. An icon is displayed indicating that the instrument is ready to
aspirate the latex particles.
5. Aspirate the latex particles:
a. Present the latex particles to the aspirate probe (open-vial position if using an
AC•T diff 2 analyzer.)
b. Press the aspirate switch as you would to run a sample.
6. When the latex cycle is completed, a screen similar to Figure 4.12-3 is displayed. Verify
that the WBC and RBC counts and the corresponding CVs displayed on the screen are
within the limits shown on Table 4.12-1.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
LATEX GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS 4
7. Check the measured values for WMCV and RMCV:
r If the measured values are within ±1.5 fL of the latex particle size entered in step 3,
no adjustment is necessary.
r If the measured values are not within ±1.5 fL of the latex particle size entered in
step 3, adjust the gains and verify that they are within the limits in Table 4.12-1. If
the gains are outside the limits, troubleshoot and correct the problem.
8. Select WIA, RIA or Both to select the parameter to change.
r Select Save and Exit icon to save the new gains and exit the program.
r Select Exit icon to exit the program without saving any changes.
Verification
Verify instrument calibration and calibrate if necessary.
Note: MCV and RDW are most affected by a change in RIA gain. The WBC differential is most
affected by a change in WIA gain.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2 ANALYZERS 4
4.13 SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2
ANALYZERS
Tools/Supplies Needed
B Ten or five normal, fresh-blood specimens
B COULTER 4C® PLUS Normal cell control (preferred)
Procedure
ATTENTION: This procedure is more accurate if performed at the normal operating temperature
of the instrument. During sample analysis, the Target values are scaled according to the
temperature of the reagent in the diluent reservoir.
1. Ensure the apertures are clean. If protein buildup is suspected, bleach the apertures as
directed under Clean (Bleach) the Baths, in the Operator’s Guide, Section 6, Service and
Maintenance.
2. Run the first specimen or the 4C PLUS Normal cell control.
3. Print the Sample Details screen or from the Sample Details screen, record the WBC, RBC,
AVR Measured, WBC Measured, and RBC Measured values.
4. Repeat step 3 above until you have ten values.
Note: If using 4C PLUS Normal cell control, cycle the control ten times. If using five
normal fresh blood specimens, cycle each blood twice. If using ten normal fresh blood
specimens, cycle each blood once.
5. Calculate the Mean value of the ten runs for WBC, RBC, AVR Measured, WBC Measured,
and RBC Measured.
r If you used normal fresh blood specimens, the Mean values for AVR Measured,
WBC Measured, and RBC Measured are the new Target values.
r If you used 4C PLUS Normal cell control, multiply the Mean values for AVR
Measured, WBC Measured, and RBC Measured by the values below to obtain the
new Target values:
t Mean value of AVR Measured x 0.98 = new Target value
t Mean value of WBC Measured x 1.05 = new Target value
t Mean value of RBC Measured x 1.02 = new Target value.
6. Calculate the CV for the WBC and RBC parameters.
7. Ensure that the results from steps 5 and 6 meet the criteria in Table 4.13-1.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2 ANALYZERS
Verification
Run controls and/or blood specimens and verify there are no clog detect problems.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
VERIFY INSTRUMENT OPERATION IN PREDILUTE MODE 4
4.14 VERIFY INSTRUMENT OPERATION IN PREDILUTE MODE
Purpose
The purpose of the Verify Predilute function is to aid you in troubleshooting the Predilute
mode. This function allows you to make prediluted samples more quickly and consistently
than manual methods. It also provides a means of determining if a Predilute mode problem is
in the instrument or in sample preparation.
DO NOT use this method for normal sample analysis. If enough whole blood is available in
the specimen tube for the instrument to make a prediluted sample, then enough blood is
available for the instrument to analyze the blood in the Whole-Blood mode and that is how
the sample should be analyzed.
Tools/Supplies Needed
B Empty sample container
Procedure
1. Select Main icon tt Diagnostics icon tt Verify Predilute icon. The screen displays the next
action to perform.
2. Present a whole blood sample to the aspirate probe.
Note: If operating an AC•T diff 2 analyzer, the Verify Predilute function uses the
Open-Vial mode.
3. Press the aspirate switch.
a. Whole blood is aspirated.
r If operating an AC•T 8/10 or AC•T diff analyzer, 20 µL of whole-blood is
aspirated.
r If operating an AC•T diff 2 analyzer, 50 µL of whole-blood is aspirated.
b. The probe is wiped clean.
c. The probe is presented after being wiped.
4. Present an empty sample container to the aspirate probe.
5. Press the aspirate switch. Diluent is dispensed into the sample container along with the
whole blood aspirated earlier.
r If operating an AC•T 8/10 or AC•T diff analyzer, 1580 µL (1.58 mL) of diluent is
dispensed into the sample container along with the 20 µL of whole blood.
Note: The sample produced has the same dilution as a prediluted fingerstick sample
prepared using a 20 µL pipette with the instrument dispensed volume of diluent.
r If operating an AC•T diff 2 analyzer, 4000 µL (4.0 mL) of diluent is dispensed into
the sample container along with the 50 µL of whole blood. The pulsing action of the
AC•T diff 2 analyzer is normal.
Note: The sample produced has the same dilution as a prediluted fingerstick sample
prepared using a 50 µL pipette with the instrument dispensed volume of diluent.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
VERIFY INSTRUMENT OPERATION IN PREDILUTE MODE
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PREPARING THE INSTRUMENT FOR LONG-TERM SHUTDOWN OR SHIPMENT 4
4.15 PREPARING THE INSTRUMENT FOR LONG-TERM SHUTDOWN OR SHIPMENT
Tools/Supplies Needed
B Empty sample container
B Bleach solution (one part distilled water and one part bleach from the bleach container)
B Distilled water
B 5-mL to 20-mL syringe with attached tubing, optional
Note: In step 2, a silicone tubing (approximately six to eight inch long) works best. In
step 13, a polyurethane tubing with an internal diameter large enough to fit securely, but
snugly over the pump fitting will maximize the pressure application.
Procedure
1. Remove the pickup tubes from their reagents and place them into an empty container.
Leave the waste line in the waste receptacle.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PREPARING THE INSTRUMENT FOR LONG-TERM SHUTDOWN OR SHIPMENT
c. A screen opens displaying 3 (indicating the third stage of preparation) and an icon
that refers the operator to the manual for instructions if they do not know how to
proceed.
8. Remove the cleaning agent pickup tube from the bleach container and place it into
distilled water.
9. Press 3:
a. The instrument cycles the distilled water through the system.
b. After cycling, a screen opens displaying 4 (indicating the fourth stage of
preparation) and an icon that refers the operator to the manual for instructions if
they do not know how to proceed.
10. Remove all the pickup tubes from the distilled water and place them into the empty
sample container.
11. Press 4 and the instrument clears the distilled water from the system.
12. When the system has been mostly cleared of fluids, the final screen displays a power
on/off icon.
WARNING Risk of personal injury or contamination. Waste can include biohazardous materials. Handle and
dispose of waste according to acceptable laboratory standards.
14. Turn the instrument off (O), unplug the power line cord from the wall outlet and
disconnect the waste tubing from the instrument. The instrument is now in long-term
shutdown. If the instrument is to be packed for shipment, go to Preparation for
Shipment.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PREPARING THE INSTRUMENT FOR LONG-TERM SHUTDOWN OR SHIPMENT 4
Preparation for Shipment
Tools/Supplies Needed
B #2 Phillips-head screwdriver
B Two shipping posts, PN 2852302, and two shipping screws, PN 2852303, for the Hgb
photometer assembly, if not already installed
B One 3.5-in. shipping screw, PN 2851323, for the Traverse module
B Pump cushion, PN 8024587, with attached caution tag, PN 4276819
Procedure
1. Open the front door. To open the front door of an AC•T diff 2 analyzer, see Opening the
AC•T diff 2 Analyzer’s Front Door under Heading 4.2, OPENING OR REMOVING
INSTRUMENT DOORS, PANELS, AND COVERS.
2. If not installed, install the two Hgb photometer assembly shipping posts and screws
(Figure 4.15-1).
Shipping
posts
Hgb photometer
securing screw
Shipping
screws
7339094E
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PREPARING THE INSTRUMENT FOR LONG-TERM SHUTDOWN OR SHIPMENT
3. Secure the Traverse module with a 3.5-in. Phillips-head shipping screw to prevent
movement during shipment (
4. Close the front door.
r If shipping an A C•T 8/10 or AC•T diff analyzer, pack the instrument in its original
box, if available.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
INSTRUMENT DECONTAMINATION 4
4.21 INSTRUMENT DECONTAMINATION
Purpose
Use this procedure to decontaminate the instrument’s diluent delivery system whenever you
suspect contamination is the cause of elevated background counts.
ATTENTION: Throughout this decontamination procedure, handle the pickup tubes carefully to
avoid recontaminating the system. To avoid contamination in the future, always place the
pickup tubes on a known clean surface or material when changing the reagents and stress this
action to the operator.
Tools/Supplies Needed
B One diluent pickup tube, PN 6805685 for the AC•T Pak reagent, or PN 6805771 for the
AC•T Tainer reagent
B One diluent reservoir with float, PN 6806261
B Two liters 5% acetic acid (white distilled vinegar)
B One empty reagent container (for waste)
B Two liter distilled water
B Two 20 mL syringes
B One fresh reagent supply (Pak or Tainer)
Procedure
WARNING Failure to adhere to the following procedure could mix reagents with acetic acid, exposing you
and your customer to hazardous chemicals. Follow the procedure exactly as written.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
INSTRUMENT DECONTAMINATION
WARNING Mixing reagents with acetic acid, could expose you and your customer to hazardous chemicals.
Do not allow waste containing acetic acid to be routed into a waste container containing lytic reagent, nor
into a sink or open drain. Remove any waste container containing lytic reagent from the instrument.
13. Remove the instrument waste line from its waste container or drain (whichever is
applicable) and route it to an empty (lytic reagent free) waste container.
14. Go to Decontaminating with Acetic Acid.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
INSTRUMENT DECONTAMINATION 4
10. Do six Startup cycles to completely flush out the distilled water. (On the Main screen,
press the Startup icon.)
11. Turn the instrument off (O).
12. Manually drain the diluent reservoir with the 20 mL syringe.
13. Remove the waste container used to collect the acetic acid solution and reinstall the
original waste container or return the waste line to the sink drain (whichever is
applicable).
14. Go to Installing New Reagents and Checking the Instrument.
PN 4237339F
SERVICE AND REPAIR PROCEDURES
PREAMP DUMMY LOAD TEST 4
4.22 PREAMP DUMMY LOAD TEST
Purpose
Use this procedure as a troubleshooting tool to help determine if problems affecting the test
results are in the pneumatic/hydraulic system or in the electronic system.
Tools/Supplies Needed
B Two Coax Load Cable Assemblies (dummy loads), PN 6029105
Procedure
1. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
2. To access the Analyzer card (Figure 4.4-2), remove the top cover and rear panel as
directed in the procedure for Removing the AC•T 8/10 and AC•T diff Analyzers’ Top
Cover and Rear Panel or for Removing the AC•T diff 2 Analyzer’s Top and Rear Center
Cover under Heading 4.2.
3. On the Analyzer card, disconnect the RBC and WBC aperture cables from J7 and J8,
respectively.
For the exact locations of these connectors, refer to Figure A.2-1, Analyzer Card
Component Locations - AC•T 8/10 Analyzers; Figure A.2-2, Analyzer Card Component
Locations - AC•T diff Analyzer; or Figure A.2-3, Analyzer Card Component Locations -
AC•T diff 2 Analyzer.
4. Connect the Coax Load Cable Assemblies to J7 and J8 on the Analyzer card.
5. Reinstall the top cover and rear panel as directed in Installation (AC•T 8/10 and
AC•T diff Analyzers) or Installation (AC•T diff 2 Analyzer) under Heading 4.2,
OPENING OR REMOVING INSTRUMENT DOORS, PANELS, AND COVERS.
6. Plug the power line cord into the ac wall outlet and turn the instrument on (I).
7. Cycle a sample of air and verify the directly measured parameter results are zero and the
histograms on the AC•T diff and AC•T diff 2 analyzers have no populations or “spikes”.
r If the test passed, the problem is probably in the pneumatic/hydraulic system.
r If the test failed, the problem is probably in the electronic system, possibly a
malfunction of the Analyzer card or the Power Supply module.
8. Turn the instrument off (O) and unplug the power line cord from the wall outlet.
9. Remove the top cover and rear panel to access the Analyzer card.
10. On the Analyzer card:
a. Disconnect the Coax Load Cable Assemblies from J7 and J8.
b. Reconnect the RBC and WBC aperture cables to J7 and J8, respectively.
11. Troubleshoot and correct the problem. When you are finished:
a. Reinstall the top cover and rear panel.
b. Plug the power line cord into the ac wall outlet and turn the instrument on (I).
PN 4237339F
CONTENTS 5
5 MAINTENANCE PROCEDURES, 5.1-1
TABLES
5.1-1 Recommended Maintenance Schedule - AC•T 8/10 and AC•T diff Analyzers, 5.1-1
5.1-2 Recommended Maintenance Schedule - AC•T diff 2 Analyzers, 5.1-1
PN 4237339F
5MAINTENANCE PROCEDURES 5
5.1 RECOMMENDED MAINTENANCE SCHEDULE
Perform maintenance for the AC•T 8/10 and AC•T diff analyzers according to the schedule in
Table 5.1-1
Table 5.1-1 Recommended Maintenance Schedule - AC•T 8/10 and AC•T diff Analyzers
PN 4237339F
CONTENTS 6
6 SCHEMATICS, 6.1-1
PN 4237339F
6SCHEMATICS 6
6.1 SCHEMATICS REQUIRED
The following is a list of the engineering schematics you need for troubleshooting the AC•T
Series instruments. Two tubing lists (production tubing instructions) are also included.
Name PN
Pneumatic/Hydraulic, AC•T 8/10 and AC•T diff Analyzers 6323083
HYDRO-PNEU, Ac.T SERIES
PN 4237339F
Changed with the DEMO VERSION of CAD-KAS PDF-Editor (http://www.cadkas.com).
CONTENTS 7
7 TROUBLESHOOTING, 7.1-1
PN 4237339F
CONTENTS
ILLUSTRATIONS
7.1-1 Diagnostics Menu, AC•T 8/10 Analyzer, 7.1-1
7.1-2 Diagnostics Menu, AC•T diff/Veterinary Option Analyzer, 7.1-1
7.1-3 Diagnostics Menu, AC•T diff and AC•T diff 2 Analyzers, 7.1-1
7.1-4 Voltages/Sensors Screen - AC•T 8/10 Analyzer, 7.1-2
7.1-5 Voltages/Sensors Screen - AC•T diff Analyzer, 7.1-2
7.1-6 Voltages/Sensors Screen - AC•T diff 2 Analyzer, 7.1-2
7.1-7 Solenoid Screen, AC•T diff 2 Analyzer, 7.1-6
7.1-8 Sample Details Screen Example - AC•T 8/10 Analyzer, 7.1-9
7.1-9 Sample Details Screen Example - AC•T diff Analyzer, 7.1-13
7.1-10 Hgb Results Analysis, 7.1-14
7.1-11 Motors Screen, AC•T 8/10 and AC•T diff Analyzers, 7.1-18
7.1-12 Motors Screen, AC•T diff 2 Analyzer, 7.1-18
7.1-13 Latex Calibration Screen - AC•T diff Analyzer, 7.1-23
7.1-14 Latex Calibration Screen - AC•T diff 2 Analyzer, 7.1-23
TABLES
7.1-1 Solenoid Drivers and Related Components, 7.1-7
7.1-2 RBC Clog Test Results, 7.1-17
7.1-3 WBC Clog Test Results, 7.1-17
7.1-4 Motors - Function, Action and Description, AC•T 8/10 and AC•T diff
Analyzers, 7.1-18
7.1-5 Motors - Function, Action and Description, AC•T diff 2 Analyzer, 7.1-19
7.2-1 Warning (Non-Fatal) Error Messages (Icons), 7.2-1
7.2-2 Fatal Error Messages (Numbers), 7.2-2
7.2-3 Allowable Voltage Ranges - AC•T 8/10 Analyzer, 7.2-5
7.2-4 Allowable Voltage Ranges - AC•T diff Analyzer, 7.2-5
7.2-5 Allowable Voltage Ranges - AC•T diff 2 Analyzer, 7.2-5
7.4-1 Specific AC•T 8/10 and AC•T diff Analyzer Problems/Solutions, 7.4-1
PN 4237339F
7TROUBLESHOOTING 7
7.1 DIAGNOSTICS MENU
The Diagnostics menu (
Figure 7.1-3 for the AC•T diff and AC•T diff 2
analyzers) provides information and manual control of many instrument functions that are
needed to adjust or troubleshoot the instrument.
7515007A
Individual control of instrument components is possible from the Solenoids screen and the
Motors screen. If any instrument functions are used, the instrument state is unknown. To
ensure that the instrument is ready to run a cycle, the Powerup cycle is automatically
performed when leaving the Diagnostics menu.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Voltages/Sensors Screen
The Voltages/Sensors screen ( for the AC•T 8/10 analyzer, Figure 7.1-5 for the
AC•T diff analyzer, and for the AC•T diff 2 analyzer) provides information about
the instrument state that can be used for troubleshooting. Voltages, sensors, Hgb amplifier
gain, and vacuum are displayed on this screen. The reagent temperature in the diluent
reservoir is also indicated. This value is not currently used by the AC•T 8/10 analyzer’s
software but it is used by the clog detection algorithm on the AC•T diff and AC•T diff 2
analyzer. A description of each screen item by its screen name follows
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
Hgb
This value represents the Hgb amplifier output or “Amp Out.” It is given in A/D units rather
than voltage. For a rough conversion to mV, consider that 5.0 V is represented by 4095 A/D
units, the highest possible output from the A/D chip. While in the Voltages/Sensors screen,
you can adjust and save the Hgb value. Adjust the gain so that the output with clean diluent
in the WBC bath is 3700 ±100. This represents the Hgb blank reading. Although the gain can
be set manually while in the Voltages/Sensors screen, it is adjusted automatically during the
Startup cycle when it is outside the range of 2867 to 4090 A/D units. This ensures that the
Hgb blank level is checked at least once a day when the instrument is in use and is adjusted
automatically when required.
Note: The AC•T diff and AC•T diff 2 analyzers tends to give slightly lower readings. For
example, it is common to get readings in the 9 to 10 V range for RAV, while the AC•T 8/10
analyzer rarely gives a reading <10.0 V.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Vac
Vac is the count or low vacuum reading given in inches of mercury. The instrument has no
indicator for the unregulated pump vacuum.
To adjust the low vacuum, you watch the low vacuum reading as you turn the regulator
adjustment knob.
r On the AC•T 8/10 and AC•T diff analyzers, the regulator adjustment knob is located on
the upper right front panel behind the Display assembly.
The vacuum regulator is a mechanical device, so the adjustment does not have to be saved
before leaving the Voltages/Sensors screen.
The vacuum transducer used to obtain a vacuum reading on the AC•T 8/10 and AC•T diff
analyzers has a small range of measurement for maximum resolution. Any vacuum under the
minimum reading shows the minimum reading and any vacuum over the maximum reading
shows the maximum reading. Current hardware gives a minimum reading of 5.36 in. Hg and
a maximum reading of 6.5 in. Hg. These numbers may vary slightly with different hardware
implementations, but should not change significantly.
Temp
Temp is a temperature reading in degrees Celsius. A temperature sensor is housed in the
lower portion of the diluent reservoir to measure the diluent temperature. The diluent
temperature should be very close to the temperature at the aperture during sample analysis.
Voltages
The AC•T series analyzers monitor several of the system voltages. The Voltages/Sensors
screen is an active display providing a full DVM function for these system voltages.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
r The +5 V, +15 V, -15 V and +24 V are supplied directly from the Power Supply module to
the Analyzer card. All other voltage supplies are created on the Analyzer card from these
voltages.
r The 15 RF voltage is a precision-regulated voltage used as a reference voltage by the A/D
converter for vacuum transducer translation. On the AC•T diff 2 analyzer, this voltage is
labeled VACRF because the voltage used for reference may be 8 V or 15 V, depending on
the vacuum transducer used.
r The 200 Vdc supply on the AC•T diff and the AC•T diff 2 analyzers provide a lower
resolution to the voltage display than the AC•T 8/10 analyzer does. The lower resolution
allows a display of the full voltage range from 0 V to about 242.5 V, instead of the narrow
80 V to 220 V range displayed on the AC•T 8/10 analyzer.
r Separate ±10 V RBC and WBC circuit supplies are created and indicated on this screen.
r Hgb lamp is the Hgb lamp supply voltage reading for the Analyzer card. On the AC•T diff
and AC•T diff 2 analyzers, the Hgb lamp supply voltage should be 3.6 V. The Hgb lamp
supply on the AC•T 8/10 analyzer is 3.18 V.
Exit Icons
The two icons in the lower right corner are the exit icons. The icon on the left, resembling a
file folder, is the Save and Exit icon. When you press the Save and Exit icon the current preamp
gain and Hgb gain settings are saved and then the screen is exited. The icon on the right, the
return arrow icon, is the Exit icon and pressing it returns you to the Diagnostics menu without
saving the gain settings.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Solenoids Screen
The Solenoids screen is used to energize up to six solenoids simultaneously. Solenoids are
designated by their number. A solenoid that is energized is shown with inverse characters. See
Figure 7.1-7.
On the AC•T 8/10 and AC•T diff analyzers, numbered keys represent solenoid drivers 1
through 17 and 23. On the AC•T diff 2 analyzer, numbered keys represent solenoid drivers 1
through 23.
7339116E
The key labeled 1 .. 23 briefly energizes each solenoid driver in sequence. Solenoids that are
energized stay on until that solenoid number is pressed again or the Diagnostics menu, not
the Solenoids screen, is exited. This makes it possible to exit the Solenoids screen and move
to another Diagnostics menu item, such as the Motors screen, while keeping up to six
solenoids energized.
Although the numbers on the Solenoids screen represent solenoid drivers, and most drivers
are connected to solenoids of the same number, several solenoid drivers drive dc motors.
Table 7.1-1 summarizes the solenoid drivers and their related components.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
Table 7.1-1 Solenoid Drivers and Related Components
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Using blood collected in a tube eliminates many sample problems such as tissue debris. Using
the instrument to make dilutions, rather than capillary fingerstick pipettes, provides
consistent and accurate dilutions from sample to sample. Refer to Heading 4.12, LATEX
GAIN ADJUSTMENT - AC•T diff AND AC•T diff 2 ANALYZERS.
When the Verify Predilute function is entered, a cleanup cycle is performed to ensure that the
instrument is ready to make a dilution. Software Version 1.03 for the AC•T 8 analyzer allows
only one dilution to be made. Software Version 1.04 for the AC•T 8/10 analyzer and the
software for the AC•T diff and AC•T diff 2 analyzers allow many dilutions to be made before
exiting the Verify Predilute function.
Sample Results
IMPORTANT Reporting results from the Sample Details screen can cause inaccurate patient diagnosis. The
Sample Results screen and printout suppresses results when a problem is detected during data
accumulation. Results suppressed on the Sample Results screen and printout are presented on the Sample
Details screen as an aid in troubleshooting problems and are not for patient diagnosis.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
Figure 7.1-9 Sample Details Screen Example - AC•T diff Analyzer
When a clog is detected, sample results are suppressed on the Sample Results screen. Instead
of the measured value, Xs (XXXXX) are displayed, printed, and transmitted. To troubleshoot
a clog detection, or a problem generating a clog detect, it is useful to have the measured value
of the parameter or parameters that were suppressed. This information is reported on the
Sample Details screen for WBC, RBC, Hgb, MCV, and Plt.
Hgb Readings
Hgb measurements are obtained by measuring the amount of 525-nm light transmitted
through a sample and comparing that result to the amount of 525-nm light transmitted
through a blank solution. To ensure the integrity of the readings, the instrument takes two
Hgb blank readings and two sample readings are taken. These four readings are displayed on
the Sample Details screen. The instrument verifies the integrity of the readings by using the
process shown in Figure 7.1-10.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Is Non-fatal
Blank1 or Yes
warning:
Blank2 < 2048 Hgb VOLTAGE
A/D units FAILURE
?
No
Is
Non-fatal
Blank1 or Yes warning:
Blank2 > 4090
Hgb VOLTAGE
A/D units
FAILURE
?
No
Calculate
Hgb using
both blank
readings
No
No
Use Hgb 1
as result 7339050C
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
When troubleshooting Hgb problems on the AC•T diff or AC•T diff 2 analyzers, consider the
following:
r On the AC•T diff analyzer, Hgb Blank2 is taken on the WBC bath rinse. After the aspirate
switch is pressed, the WBC bath is drained and rinsed. The Hgb Blank2 reading is taken
on this rinse.
r On both the AC•T diff and AC•T diff 2 analyzers, Hgb Blank1 is taken on the 1.5 mL
sample prefill that enters the WBC bath through the lower port This is the first phase of
the sample dilution, the rest of the diluent is dispensed through the probe with the
sample.
r Hgb Sample1 is taken one second before the WBC sample is drained from the WBC bath.
r Hgb Sample2 is taken 1/2 second after the Hgb Sample1.
WBC/RBC Measured
These parameters are 26-percentile values representing pulse width. A 26-percentile value is
determined by the Editor circuit for each one-second count period. Twelve values are
recorded during sample data accumulation, and these twelve values are averaged to produce
the reported measured value. The measured value is then compared to the Target value to
determine if this clog test passes.
The limits for the measured values generating clogs are as follows:
WBC Target value × 0.88 ≤ WBC Measured ≤ WBC Target value × 1.12
RBC Target value × 0.87 ≤ RBC Measured ≤ RBC Target value × 1.15
RBC CV
The CV is calculated using the 12 measured, 26-percentile values determined during sample
data collection. This test fails if the calculated CV is not within the following range:
When sample data accumulation is complete, there are 12 ratio values for WBC and RBC. The
CV of these values is calculated and used for clog detection. The idea behind this criteria is
based on consistency. The ratio of good pulses may vary for a particular sample because of cell
size, shape, elasticity, or concentration. It may even vary with the concentration of other cells
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
or interfering particles or substances. Once a sample dilution is made however, whatever the
ratio is, it should be consistent through the 12 count periods.
WBC/RBC Voteout
When there is a single aperture count voteout, WBC or RBC, the count from that aperture is
compared to the average count of the other two apertures. If this count is >7.0% higher than
the average of the other two apertures, the voteout test fails.
RAV CV
An RBC Aperture Voltage (RAV) reading is taken for each one-second count period. The CV is
calculated for the 12 readings. If the resulting CV is >5.0%, this clog test fails.
AVR Measured
The 12 readings for RBC Aperture Voltage (RAV) and WBC Aperture Voltage (WAV) are
averaged. The RAV average is then divided by the WAV average to obtain the Aperture Voltage
Ratio (AVR) Measured. This test fails if:
AVR (average) > AVR Target × 1.09 ; this condition generates an RBC failure.
AVR (average) < AVR Target × 0.97; this condition generates a WBC failure.
WBC/RBC/AVR Target
WBC and RBC Target are target 26-percentile values used for comparison with the WBC and
RBC Measured values during sample analysis. AVR Target is the target Aperture Voltage Ratio
used for comparison with the AVR Measured clog detection test. They are determined by
averaging the value obtained when running several normal samples. See Heading 4.13,
SETTING CLOG DETECTION TARGET VALUES - AC•T diff AND AC•T diff 2
ANALYZERS, for the procedure to determine these values.
Note: The numeric keypad and Save and Exit icon are used to enter the Target values.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
When clog tests are performed, flagging is accomplished according to Table 7.1-2.
Motors Screen
The Motors screen allows you to individually operate the stepper motors used by the AC•T
series analyzers. Beginning with the AC•T 8/10 analyzer, Software Version 1.04, and in all the
AC•T diff and AC•T diff 2 analyzers’ software, the Motors screen also displays the state of the
sensors associated with the stepper motors. See Figures 7.1-11 and 7.1-12.
Pressing a Motors key performs the diluter table default action for the selected motor. On the
AC•T 8/10 and AC•T diff analyzers, 15 different motor actions are required during
instrument operation, so five motor functions are available on the Motors screen
(Table 7.1-4). The AC•T diff analyzer uses 24 different motor functions (Table 7.1-5).
Many of the motor functions appear similar. For example in Table 7.1-4, M0 - ASF operates
the aspirate syringe to aspirate a sample and M2-DSF operates the diluent syringe to recharge
the syringe, but both functions move the syringe pistons down.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Table 7.1-4 Motors - Function, Action and Description, AC•T 8/10 and AC•T diff Analyzers
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
When using motor functions, you are in control of the instrument. Normal system safeguards
used by system cycles are bypassed, and must be observed. Most notable, solenoid functions
are separate from motor functions. During an instrument cycle, the proper solenoids are
energized before doing a motor function. Entering this screen without first energizing the
appropriate solenoids from the Solenoids screen may enable fluidic pathways that are not
desirable.
For example, the function M15 - CF (Table 7.1-4) rotates PM1, the waste peristaltic pump,
counterclockwise, against the waste flow.
r During a normal Shutdown cycle, LV13 is energized before calling function M15 - CF,
changing the hydraulic path from the waste container to the shutdown diluent container.
Then, without energizing any other solenoid, performing M15 - CF fills the RBC bath
with shutdown diluent.
r Performing the M15 - CF function from the Motors screen does not energize LV13, so
the RBC bath fills with waste from the waste system. Furthermore, since draining waste
is also a separate function, the RBC bath overflows if you repeatedly perform this
function without draining the RBC bath. To duplicate the Shutdown cycle, you must
energize the appropriate solenoids from the Solenoids screen first, before entering the
Motors screen.
CAUTION Risk of damage to mechanical components. Using fill functions repeatedly can drive the syringe
pistons down and out of the glass syringe barrels. Subsequent upward movement of the syringe without
manually placing the pistons back into the glass barrels may cause the syringe barrels to break. Always
open the right side door and observe the syringe assembly when using syringe motor functions. If the
pistons are pulled out of the syringe barrels, manually guide them back in while using the thumbscrew at
the bottom of the syringe assembly.
CAUTION Motor functions M10-PDHC and M20-DCHC leave holding current applied to the motor after
motion has stopped. Doing motor functions M10-PDHC or M20-DCHC without doing the opposite function
within 1 minute can damage the driver circuit on the Analyzer card or the stepper motors. Do motor
function M9-PUP within 1 minute of doing M10-PDHC. Do motor function M21-DOPN within 1 minute of
doing M20-DCHC.
If you are using probe functions, direction of travel is determined by the needs of the system
during normal instrument cycles. Some functions may generate error messages if the position
of the probe before initiating the function is ignored. For example, during a normal
instrument cycle the probe always goes to the RBC bath position from the WBC bath, never
from the aspirate position. The direction of travel is left. If you initiate function M8-PRB
while at the aspirate position, the probe moves left. Since the probe is already at its left-most
position, and the RBC bath position is to the right, you drive the traverse housing against the
left mechanical stop, never get to the RBC bath, and generate fatal error 11.
On the AC•T 8/10 and AC•T diff analyzers, the syringe sensor is the simplest. It is a single
sensor that displays ON when the syringe is at the top position and displays OFF when the
syringe is not at the top position.
The horizontal and vertical traverse sensors are actually double sensors where the binary state
is used to determine a horizontal or vertical position. The Motors screen displays the probe
position rather than the individual sensor state.
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
On the AC•T 8/10 and AC•T diff analyzers:
When the probe is at one of the indicated positions, that position is ON, otherwise it is OFF.
This means that all six positions can indicate OFF but no more than one horizontal and one
vertical position should indicate ON at one time.
When the Pulse icon is pressed, a screen is presented to allow entry of a minimum and
maximum voltage Values can range from 0 to 255. The minimum and maximum defaults are
1 and 255, respectively.
Latex Calibration
With the introduction of histograms and the differential parameters, greater control of
amplifier gain is required. The amplifiers are adjustable and a latex calibration function is
provided for gain adjustment. See Heading 4.12, LATEX GAIN ADJUSTMENT - AC•T diff
AND AC•T diff 2 ANALYZERS, for the procedure to adjust the amplifier gain using latex
particles.
Selecting Main icon tt Diagnostics icon tt Latex Calibration icon displays the Latex Calibration
screen (Figure 7.1-13) for an AC•T diff analyzer or for an AC•T diff 2
analyzer).
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU 7
Figure 7.1-13 Latex Calibration Screen -
AC•T diff Analyzer
7339053E
The top item on the Latex Calibration screen is the latex particle size. If the instrument’s
power has been off since the last time the latex particle size was used, this value defaults to a
particle size of 94.3 fL.
The next three fields on the Latex Calibration screen display values representing the amplifier
gains for WBC, RBC, and Plt gain. If you know what the gains should be, you can enter them
here without performing the Latex Gain Adjustment procedure. Note that the Plt gain should
be set to the same value as the RBC gain for normal operation.
During the automatic adjustment procedure, prompts to change gains only mention WBC
and RBC gains. Any requested change to the RBC automatically produces the same change in
the Plt value. If however, you manually enter the RBC gain, the Plt gain is not automatically
entered. You must manually enter the same value that was entered for the RBC gain.
The Latex Calibration screen also displays the following icons at its bottom:
PN 4237339F
TROUBLESHOOTING
DIAGNOSTICS MENU
Cycle Counter
The instrument’s cycle counter is incremented during each aspirate cycle, whether Whole
Blood, Predilute, Control or Reproducibility mode. On the AC•T 8/10 analyzer the cycle
counter icon appears on the Time/Date Setup screen. On the AC•T diff and AC•T diff 2
analyzers, the cycle counter icon appears on the Startup screen. The Startup screen is a more
practical location for customer use but requires running the Startup cycle to get the cycle
count. For service use, the cycle counter icon appears on the Diagnostics menu on the
AC•T diff analyzer, Software Version 1.04 and later, and on the AC•T diff 2 analyzer.
Prepare to Ship
This function is used to prepare an instrument for shipment or for long-term shutdown.
There are four stages to preparing an instrument for shipment. By the end of the procedure,
the instrument has been decontaminated, flushed with distilled water and dried. For
procedures, refer to Heading 4.15, PREPARING THE INSTRUMENT FOR LONG-TERM
SHUTDOWN OR SHIPMENT.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES 7
7.2 TROUBLESHOOTING ERROR MESSAGES
The system displays two types of error messages, warning (non-fatal) and fatal error
messages.
Warning messages indicate an instrument condition that could affect operation, such as low
reagents or a Printer that is off line. A warning message is indicated by the "yield sign"
warning icon, a message specific icon, and a continuation icon. Pressing the Continue icon
allows the user to continue operation and correct the condition where applicable. Table 7.2-1
shows the icons that identify warning (non-fatal) errors.
Transmission incomplete
Hgb voltage failure - Hgb Blank or Sample reading outside range 2048 to 4095 A/D
units
Channelizing buffer overflow error - usually indicates an overflow in one of the Plt
channels
Timekeeper failure
Waste full
Diluent empty
Patient storage data corrupt, AC•T diff and AC•T diff 2 analyzers
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES
Fatal error messages (Table 7.2-2) are indicated by an Error icon and an Error Message icon or
number. Fatal error messages do not have a Continue icon available. The instrument must be
turned OFF then ON to continue operation. The Error icon is displayed until the fatal error is
resolved.
10 All Probe did not reach WBC position. For details, see Probe Did Not
Reach Position Errors 6 through 11 and 19.
11 All Probe did not reach RBC position. For details, see Probe Did Not
Reach Position Errors 6 through 11 and 19.
12 AC•T 8/10 and AC•T diff analyzers Syringe did not reach up position. For details, see Syringe Did Not
Reach/Leave Up Position Errors 12 and 13, AC•T 8/10 and AC•T diff
Analyzers.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES 7
Table 7.2-2 Fatal Error Messages (Numbers) (Continued)
AC•T 8/10 and AC•T diff analyzers Diluent level error during Powerup table execution. For details, see
Diluent Level Error During Powerup.
All Vacuum <6 in. Hg before probe wipe. For details, see Vacuum Error.
Service Bypass
An error that occurs during powerup creates a loop and consequently the instrument cannot
be used. For troubleshooting a powerup problem, a “service bypass” function is provided to
get past the powerup fatal error message. To initiate the service bypass:
This procedure bypasses most normal powerup operations and displays the Main screen.
r All cycles and menus, including the Diagnostic menu are active.
r When running cycles, such as the Aspirate cycle, any error encountered is reported as a
fatal error and forces a power off.
r When using the Solenoids screen or the Motors screen, any error encountered is reported
as a warning, using the error number as the message icon. Pressing the Continue icon
returns control to the Solenoids or the Motors screen.
r System states are ignored and normal instrument operation could be affected.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES
When the instrument is again ready for normal operation, always power up without using the
aspirate switch. If the instrument is ON and was turned ON using the aspirate switch, you
must turn the instrument OFF, and then back ON again without using the aspirate switch.
PCMCIA Error 1
Error 1 is generated when the system fails to verify a checksum of the PCMCIA code. All
program code is contained in Flash Memory. Major problems with reading the Flash Memory
or corruption of Flash Memory data prevents the checksum verification from occurring or the
display from being used. As a result, this error should be very rare, but if it occurs, the most
likely problem is the Flash Memory (software) card itself.
DVM Error 3
A DVM check is made during powerup and during a menu-selected startup. System voltages
are checked to see if they are within allowable ranges. If any of these voltages are outside their
allowable range (shown in Table 7.2-3 for the AC•T 8/10 analyzer, in Table 7.2-4 for the
AC•T diff analyzer, and in Table 7.2-5 for the AC•T diff 2 analyzer), error 3 is displayed on
the screen. The A/D converter has an input range of 0 to 10 Vdc, so voltages are converted to
this range before presentation to the A/D converter. The test points given in the table are
before the conversion, representing the actual voltage.
The POWER GOOD/POWER FAIL signal from the Power Supply module is also monitored.
A POWER GOOD/POWER FAIL signal also triggers error 3.
If a DVM failure occurs, there may be a problem with the Power Supply module or the
Analyzer card, or poor line voltage may be triggering the POWER GOOD/POWER FAIL
signal.
Table 7.2-4 Allowable Voltage Ranges - AC•T diff Analyzer
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES 7
Probe Did Not Reach Position Errors 6 through 11 and 19
The Traverse module is responsible for moving the aspirate probe to different horizontal and
vertical positions.
r The AC•T 8/10 and AC•T diff analyzers use two dual-channel optical sensors to identify
position, one for vertical and one for horizontal movement. Vertical positions are
determined by holes in a flag that snaps onto the top of the vertical rack and rides
through the vertical sensor. Horizontal positions are determined by cutouts in the front
portion of the main traverse mounting bracket.
During system operation, the probe is sent enough steps to go slightly beyond the sensor read
position. If the probe does not get to the position, horizontal or vertical, that it is sent to, a
corresponding error message is generated. See Table 7.2-2, Fatal Error Messages (Numbers),
for specific messages.
Mechanical and electrical problems that can cause errors 6 through 11 and 19 include:
r Obstruction or resistance to probe movement that causes the stepper motor to miss
steps.
r Obstructions in the optical sensor path that cause incorrect sensor reads.
r Poor electrical connections to the motors and sensors, the responsibility of the Traverse
Interconnect card and the flex cable.
r Problems with the Power Supply module, since the DVM check is made only during
powerup. This is especially true on the AC•T 8/10 analyzer, since the DVM does not
check the +24 Vdc supply.
r Failure of any of the following components: motors, sensors, Traverse Interconnect card,
flex cable, and Analyzer card.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES
Syringe Did Not Reach/Leave Up Position Errors 12 and 13, AC•T 8/10 and AC•T diff Analyzers
The syringe assembly only uses one sensor and it is at the top position. At various times, the
syringe is homed, sent to the top position. When the syringe is not detected at the home
position after it is sent there, a Syringe Did Not Reach Up Position (error 12) message is
generated.
Syringe movement is complex as many up/down movements can occur before the syringe is
sent back to the home position. It is not practical to track this movement and determine
exactly when the syringe should be back at the home position. Instead, the syringe is sent to
the home sensor after a period of activity to start once again from a known position. Though
movement is not tracked precisely, the syringe should never be at the home sensor after a
downward move. If this is detected, a Syringe Did Not Leave Up Position (error 13) message
is generated.
r A problem with the motor and lead screw of the syringe mechanism
r Dislodged syringe pistons
r A damaged, obstructed, or out of position syringe sensor
r Deteriorating piston seals that result in too much friction
r Broken syringe bodies
r Plugs in hydraulic tubing or solenoid valves that cause syringe deadheading.
Since all the components involved in error 16 are located on the Analyzer card, any time
error 16 is generated there is a problem on the Analyzer card.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING ERROR MESSAGES 7
Steps Missing (Syringe Motor) Error 17, AC•T 8/10 and AC•T diff Analyzers
Error 17 indicates something is wrong with syringe movement. The syringe assembly has one
sensor, located at the top of stroke. The syringe is sent to this sensor near the beginning of a
cycle so that it has a known starting point. Because there are three syringes for one motor,
many upward and downward moves are made before the syringe may be sent back to the
sensor. The system keeps track of the number of steps for each move. When the syringe is
finally sent back to the sensor, the number of downward steps should equal the number of
upward steps, leaving a difference of 0. Error 17 is generated if the difference is not near 0. A
margin of error is given, since normal stopping and directional changes can produce small
step errors.
Note that when using the Motors screen, a default number is used to generate error 17 when
the syringe is sent to the sensor using the M1 motor function. If you are not at the correct
position (within the limits for error) before this function is used, error 17 is generated. This
does not however, mean that there is a problem with the syringe.
During the Powerup cycle, the diluent level sensor is checked twice. It is checked after a large
amount of diluent is used by the system, which should indicate that the diluent level is low,
not at the sensor. It is checked again after the reservoir is intentionally overfilled, which
should indicate that diluent is at sensor. If the sensor reading is not what it should be at either
check, a fatal diluent error is generated.
This routine is used to check the operation of the diluent sensor, not to check the diluent
level, but a problem with the diluent reagent delivery system can cause the fatal version of the
diluent error to occur. For instance, if there is a leak in the diluent input that does not allow
the reservoir to fill, the sensor never sees diluent on the second check.
The most common problem with the diluent level sensor is bubbles, a film, or a droplet of
diluent developing on the surface of the sensor.
Chronic diluent errors during powerup can also be caused by clogged diluent filters.
Vacuum Error
More than 6-in. Hg is needed to do a probe wipe. The vacuum is checked before probe wipe
and if the vacuum is less than 6-in. Hg this error is generated. This check ensures there is no
vacuum leak and that the vacuum can build to this level quickly.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING CLOG DETECTION PROBLEMS 7
7.3 TROUBLESHOOTING CLOG DETECTION PROBLEMS
Purpose
This section provides an in-depth discussion of the parameters used for clog detection and
why they are used. A greater knowledge of these parameters and what can cause them to vary
should make troubleshooting clog detection problems simpler.
Description
The basic parameter used for clog detection is pulse width. This is an important parameter
because it represents the time it takes a particle or cell to travel through a sensitive area of the
aperture. Since the instrument applies a constant fixed vacuum to draw cells through a fixed
aperture, the time of travel or pulse width should be constant. The time of travel and pulse
width do vary some, however, because not all particles take a straight path through the
aperture. A cell passing straight along the center axis of the aperture achieves the shortest
time and the narrowest pulse width. A cell approaching the aperture from the side takes an
angular path through the sensing zone and aperture and consequently takes longer to clear
the sensing zone and produces a wider pulse width. A wider pulse width means that the cell
was in the sensing zone for a longer time and that it took longer to travel through the
sensing zone.
The 26-percentile value is based on readings taken at the ½ height width and ¾ height width.
To understand a ½ height width, first consider a triangular pulse. If you were to draw a
horizontal line at ½ the height of the triangle, and measure the length between the points
where the line intersects the two vertical sides of the triangle, this measurement would be the
½ height width of the triangle or the pulse width.
The reason this method of measuring the pulse width is better than the method used by the
AC•T 8/10 analyzer is that this value is independent of height. This can be illustrated by
considering another triangle with exactly the same base width, but with a different height.
This triangle represents a particle of a different size, but taking exactly the same path and time
through the aperture. Measuring the ½ height width gives the same value as was obtained for
the first triangle. Measuring the width at a fixed threshold gives a wider pulse width for a
taller triangle and a narrower pulse width for a shorter triangle.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING CLOG DETECTION PROBLEMS
After measuring the pulse widths, the instrument must determine a value representing all the
pulses passing through the Editor circuit during a count period. The first value to consider is
an average, but an average value might not be the best choice. If all the pulse widths during a
count period were plotted, with the vertical axis representing the number of pulses and the
horizontal axis representing increasing width, the resulting distribution would be similar to a
normal Plt distribution curve. It would show a fast rise on the left of the peak, and a more
gradual drop on the right of the peak. The average width would be on the right, but fairly
close to the peak. The average width would represent the 50 percentile, since 50% of the
pulses would be on the left of this point, and 50% would be on the right. The 26 percentile is
a point on the curve where 26% of the pulses are on the right of that point and 74% are on the
left. The 26 percentile point is well to the right of the peak, where the curve is approaching
the baseline.
Keeping this normal distribution in mind, consider what happens if a clog occurs. Pulse
widths increase because there is no optimal path through the center axis of the aperture.
Pulses no longer fall on the left end of the curve and the drop on the right side of the peak is
more gradual as pulses are added. The average, and possibly the peak, move slowly to the
right. The 26 percentile also moves to the right, since more pulses were being plotted on the
right end of the curve. The 26 percentile is more sensitive to a change in wider pulse widths,
which is what a clog produces, than the 50 percentile or a 74 percentile.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING CLOG DETECTION PROBLEMS 7
When a cell is fixed, as cells are for controls and calibrators, some of the elasticity is lost.
Consequently, the cell travels through the aperture faster, creating a narrower pulse
width. This is why there is a difference in the 26-percentile value between fresh bloods
and controls. This is also why sickle cells produce a narrow pulse width. Sickle cells have
rigid cell walls which means they do not have the same elasticity as normal cells and
travel through the aperture much faster.
The primary purpose of the Editor circuit in deriving the WBC and RBC Ratios is to screen
the pulse train for well-shaped pulses and channelize only well-shaped pulses for
constructing histograms. The aperture signal is applied to the editor and the rising edge of a
pulse triggers the processing of that pulse. No other pulses are processed until processing is
completed for the previous pulse. This means that only a percentage of pulses are even looked
at by the editor. This is further complicated for the RBC editor because Plts are taken from the
same signal, and Plt pulse processing is given a higher priority because of the smaller number
of Plts generally available. If the Plt count is high, the number of pulses looked at by the RBC
editor decreases. Pulses processed and rejected by the editor are not counted, so the
WBC/RBC Ratio is a ratio of the channelized pulses (well-shaped pulses) to the raw count
from the counter (total pulses at the aperture).
The ratio of well-shaped pulses may be very low, but still acceptable. For this reason, the
actual ratio is not used for RBC clog detection, and the ratio has a wide limit for WBC clog
detection. More useful information is gained from the CV of the ratio values. A ratio is
calculated for each count period yielding 12 readings for RBC and WBC. A CV of the 12
values is calculated and limits placed on the result. The assumption is that the ratio should
not change. Samples with ratios of 75% and 20% should remain at 75% and 20% for all 12
count-periods. Factors such as vacuum, cell concentration, and physical cell characteristics
may determine the ratio, but remain constant for count duration.
A clogged aperture affects flow patterns and in turn affects the ratio. Cells with an irregular
path through the aperture are rejected by the editor, decreasing the ratio. If the clog is severe,
the pulse rate is reduced and may allow a higher percentage of cells to be processed,
increasing the ratio.
Changes in flow rate caused by temperature or vacuum have no affect until they cause the
pulses to be so wide that they get rejected by the editor. This triggers vacuum alarms and the
26-percentile check.
The Ratio CV is affected when a temporary clog is present long enough to significantly
change one or more count periods. If a sample yields a low ratio, the Ratio CV calculation is
higher for mathematical reasons alone.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING CLOG DETECTION PROBLEMS
WAV/RAV/AVR
WAV and RAV are the voltages measured across the WBC and RBC apertures, respectively.
AVR is the ratio of the RAV to WAV, which should be constant for a given set of apertures.
Since the voltage across the aperture is directly proportional to resistance at the aperture, this
is the most obvious form of clog detection. This is also why the voltage reading for the WBC
aperture is about half the voltage reading for the RBC aperture. Resistance is lower for the
larger WBC aperture, so voltage is lower. An obstruction in the aperture causes an increase in
resistance which causes an increase in voltage. A cracked aperture is larger, with less
resistance, so the voltage across the aperture decreases.
The WAV, RAV and AVR are affected directly by aperture size, temperature of the reagents,
resistive properties of the reagents (electrolytes), and any bubbles in the aperture’s path. The
instrument software compensates to some degree for temperature changes and using the AVR
Ratio also minimizes the affects of temperature.
PN 4237339F
TROUBLESHOOTING
TROUBLESHOOTING AC•T 8/10 AND AC•T diff ANALYZER SPECIFIC PROBLEMS 7
7.4 TROUBLESHOOTING AC•T 8/10 AND AC•T diff ANALYZER SPECIFIC PROBLEMS
The causes of several problems on the AC•T 8/10 or AC•T diff analyzers may not be
immediately obvious. Table 7.4-1 describes some of these problems and their
causes/solutions, including solutions that may be different from those used for other
Beckman Coulter instruments.
Table 7.4-1 Specific AC•T 8/10 and AC•T diff Analyzer Problems/Solutions
PN 4237339F
CONTENTS 8
8 PARTS LISTS, 8.1-1
ILLUSTRATIONS
8.2-1 Triple Syringe Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Table 8.2-1), 8.2-1
8.2-2 Front Panel Components, AC•T 8/10 and AC•T diff Analyzers (See
Table 8.2-2), 8.2-2
8.2-3 Front Panel Components, AC•T diff 2 Analyzers (See Table 8.2-3), 8.2-4
8.2-4 Front Door Components (See Table 8.2-4), 8.2-6
8.2-5 Right-Side Compartment, Reservoir Components, AC•T 8/10 and AC•T diff
Analyzers (See Table 8.2-5), 8.2-8
8.2-6 Right-Side Compartment, Transducer Components, AC•T 8/10 and AC•T diff
Analyzers (See Table 8.2-6), 8.2-10
8.2-7 Right-Side Compartment, Reservoir and Reagent Pump Components, AC•T diff 2
Analyzers (See Table 8.2-7), 8.2-12
8.2-8 Right-Side Compartment, Pneumatic Pump Components, AC•T diff 2 Analyzers
(See Table 8.2-8), 8.2-14
8.2-9 Rear View Showing Components (See Table 8.2-9), 8.2-16
8.2-10 Traverse Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Table 8.2-10), 8.2-18
8.2-11 Traverse Module Components, AC•T diff 2 Analyzers (See Table 8.2-11), 8.2-20
8.2-12 Cap-Pierce Module Components, AC•T diff 2 Analyzers (See
Table 8.2-12), 8.2-22
8.2-13 Left Side and Top Views (See Table 8.2-13), 8.2-24
8.2-14 Aperture/Bath Assembly Components on Mounting Plate, AC•T 8/10 and AC•T
diff Analyzers (See Table 8.2-14), 8.2-26
8.2-15 Aperture/Bath Assembly Components on Mounting Plate, AC•T diff 2 Analyzers
(See Table 8.2-15), 8.2-28
8.2-16 Aperture/Bath Assembly Shield and Mounting Plate (See Table 8.2-16), 8.2-30
TABLES
8.1-1 Components, 8.1-1
8.1-2 Fasteners, Fittings, and Miscellaneous Hardware, 8.1-8
8.1-3 Optional Items and Supplies, 8.1-14
8.2-1 Triple Syringe Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-1), 8.2-1
8.2-2 Front Panel Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-2), 8.2-3
8.2-3 Front Panel Components, AC•T diff 2 Analyzers (See Figure 8.2-3), 8.2-5
8.2-4 Front Door Components (See Figure 8.2-4), 8.2-7
8.2-5 Right-Side Compartment, Reservoir Components, AC•T 8/10 and AC•T diff
Analyzers (See Figure 8.2-5), 8.2-9
PN 4237339F
CONTENTS
8.2-6 Right-Side Compartment, Transducer Components, AC•T 8/10 and AC•T diff
Analyzers(See Figure 8.2-6), 8.2-11
8.2-7 Right-Side Compartment, Reservoir and Reagent Pump Components, AC•T diff 2
Analyzers (See Figure 8.2-7), 8.2-13
8.2-8 Right-Side Compartment, Vacuum Pump Components, AC•T diff 2 Analyzers
(See Figure 8.2-8), 8.2-15
8.2-9 Rear View Showing Components (See Figure 8.2-9), 8.2-17
8.2-10 Traverse Module Components, AC•T 8/10 and AC•T diff Analyzers (See
Figure 8.2-10), 8.2-19
8.2-11 Traverse Module Components, AC•T diff 2 Analyzers (See Figure 8.2-11), 8.2-21
8.2-12 Cap-Pierce Module Components, AC•T diff 2 Analyzers (See
Figure 8.2-12), 8.2-23
8.2-13 Left Side and Top Views (See Figure 8.2-13), 8.2-25
8.2-14 Aperture/Bath Assembly Components on Mounting Plate, AC•T 8/10 and AC•T
diff Analyzers (See Figure 8.2-14), 8.2-27
8.2-15 Aperture/Bath Assembly Components on Mounting Plate, AC•T diff 2 Analyzers
(See Figure 8.2-15), 8.2-29
8.2-16 Aperture/Bath Assembly Shield and Mounting Plate (See Figure 8.2-16), 8.2-30
PN 4237339F
8PARTS LISTS 8
8.1 MASTER PARTS LIST
The parts are listed in alphabetic order within the following categories:
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
Card, Analyzer, Universal (for AC•T diff and AC•T diff 2 analyzers only) 6707111 8.2-13 4
Card, Display Controller Driver, SMT 6706703 8.2-4 15
Card, Display Controller Driver, through hole 6706342 8.2-4 15
Card, Pump Motor Filter 6029207 8.2-8 21
8.2-13 12
Card, Transient Suppressor, Power Supply module 6705663 8.2-9 11
Card, Traverse Interconnect, AC•T 8/10 and AC•T diff analyzers 6706334 8.2-10 4
Card, Traverse Interconnect, AC•T diff 2 analyzer 6706753 8.2-11 3
Chamber, vacuum isolator and waste assembly, molded 6805032 8.2-14 14
8.2-15 14
Choke, 0.010 orifice, black 6213011 8.2-13 11
Choke. 0.006 orifice, brown 6213009 8.2-13 11
Cover, cap-pierce door, AC•T diff 2 analyzer 1024537 8.2-12 15
Cover, front, AC•T diff 2 analyzer 6806393 8.2-4 6
Cover, front, with partial label, AC•T 8/10 and AC•T diff analyzers 6805673 8.2-4 6
Cover, pneumatic pump compartment, AC•T diff 2 analyzer, ASF Thomas pump 6806364 8.2-8 1
Cover, pneumatic pump compartment, AC•T diff 2 analyzer, KNF dual-head pump 6807605 8.2-8 26
Display, graphic LCD screen 4837341 8.2-4 19
Display, LCD and keypad assembly with SMT, AC•T 8/10 and AC•T diff analyzers 7000580 8.2-4 5
Display, LCD and keypad assembly with SMT, AC•T diff 2 analyzer 7000623 8.2-4 5
Door, left side with hinge, no latch, AC•T 8/10 and AC•T diff analyzers 6805316 8.2-13 15
Door, right side with hinge, no latch, AC•T 8/10 and AC•T diff analyzers 6805672 8.2-6 20
Fan, 24 Vdc, mounts on rear of panel of the AC•T diff 2 analyzer 6806535
Fan, filter and cage assembly, for rear panel of the AC•T diff 2 analyzer 2603078
Filter, fluid barrier, 0.45 micron, green 6232803 8.2-6 22
8.2-8 18
Filter, for 24 Vdc fan, AC•T diff 2 analyzer 2603081
Filter, waste particle, basket type, 149 micron mesh, double-barb, AC•T diff 2 analyzer 6233045 8.2-7 22
Note: Position filter with the arrow pointing towards the waste pump.
Filter, water, hydrophilc, 5 µ, with male Luer fitting, AC•T 8/10 and AC•T diff analyzers 6233052 8.2-6 1
Filter, water, hydrophilic, 0.45 µ, replacement kit (two filters in kit), AC•T 8/10 and 6915577
AC•T diff analyzers
Note: These filters are mounted on the inside surface of the right-side door.
Filter, water, hydrophilic, 0.45 µ, replacement kit (two filters in kit), AC•T diff 2 analyzer 6915526 8.2-7 10
Note: These filters are mounted inside the right compartment on the pneumatic
pump cover.
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-1 Components (Continued)
Flag, vertical position, for optical sensor, AC•T 8/10 and AC•T diff analyzers 1023357 8.2-10 11
Foam trap, glassware and stopper assembly 6806764 8.2-7 20
Fuse, 1.5 A SLO BLO, F1 and F2 on Power Supply module, AC•T 8/10 and AC•T diff 5102029 8.2-9 14
analyzers
Fuse, 2.5 A SLO BLO, F1 and F2 on Power Supply module, AC•T diff 2 analyzer 5120184 8.2-9 14
Gear, cap-pierce door, with setscrew, AC•T diff 2 analyzer 2527842 8.2-12 10
Gear, traverse belt tensioner, AC•T diff 2 analyzer 2527743 8.2-11 21
Guard, cap-pierce probe, AC•T diff 2 analyzer 1024587 8.2-12 22
Holder, tube, AC•T diff 2 analyzer 6806615 8.2-12 2
Holder, tube, door assembly, AC•T diff 2 analyzer 6806616 8.2-12 9
Housing, cap pierce, AC•T diff 2 analyzer 6806481 8.2-12 1
Housing, lamp holder and heatsink, molded 6805691 8.2-14 3
8.2-15 3
Housing, probe-wipe, AC•T 8/10 and AC•T diff analyzers 6859716 8.2-10 14
Housing, probe-wipe, AC•T diff 2 analyzer 6807013 8.2-11 12
Housing, traverse guide, AC•T diff 2 analyzer 6706321 8.2-11 10
Housing, traverse, molded, AC•T 8/10 and AC•T diff analyzers 6805479 8.2-10 17
Housing, traverse, molded, AC•T diff 2 analyzer 6806502 8.2-11 15
Insulation, vacuum pump cover, side, AC•T diff 2 analyzer 1024433 8.2-8 15
Insulation, vacuum pump cover, top and bottom, AC•T diff 2 analyzer 1024434 8.2-8 14
Insulation, vacuum pump cover, top, AC•T diff 2 analyzer 1024432 8.2-8 16
Insulation, vacuum pump, bracket side, AC•T diff 2 analyzer 1024435 8.2-8 8
Insulator, display fish paper, AC•T 8/10 or AC•T diff analyzer with original through-hole 1023501
Display Controller Driver card
Insulator, display fish paper, AC•T analyzer with SMT Display Controller Driver card 1024137
Keyboard, touch screen 2016703 8.2-4 20
Knob, vacuum regulator 9908180 8.2-2 2
8.2-8 5
Lamp, Hgb Photometer assembly 6805344 8.2-14 2
8.2-15 2
LED, power indicator assembly, AC•T 8/10 and AC•T diff analyzers 6805394 8.2-4 23
LED, power indicator assembly, AC•T diff 2 analyzer 6806620 8.2-4 23
Lens, Hgb lamp 1023237 8.2-14 6
8.2-15 6
Lens, Hgb lamp, heat filter, blue 6102189 8.2-14 5
8.2-15 5
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-1 Components (Continued)
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
Shield, plastic, splash guard, AC•T 8/10 and AC•T diff analyzers 6805734 8.2-10 5
Software, AC•T 10 analyzer, Version 1.04 6706655 8.2-2 1
Software, AC•T 8 analyzer, Version 1.04 6706654 8.2-2 1
Software, AC•T diff 2 analyzer, Spanish Version 1.00 6915345 8.2-3 1
Software, AC•T diff 2 analyzer, Version 1.00 6915288 8.2-3 1
Software, AC•T diff analyzer, English, Version 1.06, card only 6915355 8.2-2 1
Software, AC•T diff analyzer, English, Version 1.06, upgrade kit, mailable 6915356 8.2-2 1
Software, AC•T diff analyzer, Spanish, Version 1.06, card only 6915347 8.2-2 1
Software, AC•T diff/Veterinary Option analyzer, Version 1.06, kit with manual 6915366 8.2-2 1
Software, AC•T diff/Veterinary Option analyzer, Version 1.06, kit without manual 6915367 8.2-2 1
Spool, peristaltic pump, AC•T 8/10 and AC•T diff analyzers 6859650 8.2-5 3
Spool, peristaltic pump, new molded version, AC•T 8/10 and AC•T diff analyzers 6706686 8.2-5 3
Spool, sweep-flow, assembly (includes 13 ft. of EVA tubing, PN 3202284, and two 6807701 8.2-14 17
flangeless fittings, PN 6233080) 8.2-15 20
Switch, aspirate, with cable and connector 6805396 8.2-13 5
Switch, Power Supply module and instrument ON/OFF 5120214 8.2-9 10
Syringe, assembly, 1 mL, without barrel boot seal 6707095 8.2-1 5
Syringe, glass barrel assembly, with boot and plunger, 250 µL 6706808 8.2-1 4
Syringe, glass barrel assembly, with boot and plunger, 5 mL 6706809 8.2-1 3
Syringe, glass barrel assembly, with boot and plunger, without boot seal, 250 µL 6707087 8.2-1 4
Syringe, glass barrel with Luer fitting, 1 mL (replaces PN 2527682) 6706652 8.2-1 6
Syringe, glass barrel with Luer fitting, 250 µL (replaces PN 2527683) 6706653 8.2-1 13
Syringe, glass barrel with Luer fitting, 5 mL (replaces PN 2527680) 6706651 8.2-1 7
Syringe, piston, 1 mL 2527677 8.2-1 5
Syringe, piston, 250 µL 2527678 8.2-1 11
Syringe, piston, 5 mL 2527679 8.2-1 8
Traverse, probe traverse module, AC•T 8/10 and AC•T diff analyzers 7000558 8.2-2 8
Tubing, 0.082 i.d., polyurethane, clear 3202036
Tubing, 0.082 i.d., polyurethane, green striped 3202208
Tubing, 0.082 i.d., polyurethane, red striped 3202205 8.2-8 7
Tubing, 0.082 i.d., polyurethane, blue striped 3202209
Tubing, 0.082 i.d., polyurethane, yellow striped 3202207
Tubing, 0.145 i.d., polyurethane, clear 3202039 8.2-14 19
8.2-15 22
Tubing, coiled reagent 1023593
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-1 Components (Continued)
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-2 Fasteners, Fittings, and Miscellaneous Hardware (Continued)
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-2 Fasteners, Fittings, and Miscellaneous Hardware (Continued)
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
PN 4237339F
PARTS LISTS
MASTER PARTS LIST 8
Table 8.1-2 Fasteners, Fittings, and Miscellaneous Hardware (Continued)
PN 4237339F
PARTS LISTS
MASTER PARTS LIST
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
8.2 ILLUSTRATED PARTS
Figure 8.2-1 Triple Syringe Module Components, AC•T 8/10 and AC•T diff Analyzers (See Table 8.2-1)
1 3 7
6
4
8
5
12 10
11
7339008E
Table 8.2-1 Triple Syringe Module Components, AC•T 8/10 and AC•T diff Analyzers (See Figure 8.2-1)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-2 Front Panel Components, AC•T 8/10 and AC•T diff Analyzers (See Table 8.2-2)
15
14
13
12
11 2
9
10
8
3
6 5
7339006C
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-2 Front Panel Components, AC•T 8/10 and AC•T diff Analyzers (See Figure 8.2-2)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
23
1
22
4 5
21 2
20
19
18
13
17
14
12
15
16
9
10
8
11
7339002C
7
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-4 Front Door Components (See Figure 8.2-4)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-5 Right-Side Compartment, Reservoir Components, AC•T 8/10 and AC•T diff Analyzers
(See Table 8.2-5)
14 1
3
7
8
2
13
12
6
RIGHT SIDE VIEW
11
7
8
9
10 7339003E
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-5 Right-Side Compartment, Reservoir Components, AC•T 8/10 and AC•T diff Analyzers
(See Figure 8.2-5)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-6 Right-Side Compartment, Transducer Components, AC•T 8/10 and AC•T diff Analyzers
(See Table 8.2-6)
5 2
1
21
5
19
7
17 8
16
20
15 10
18
14
13 12
11
7339004E
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-6 Right-Side Compartment, Transducer Components, AC•T 8/10 and AC•T diff Analyzers
(See Figure 8.2-6)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
6
3
5
2
3
4
17
3
4
11 12 7
8
16
15
10
14
13 10
7339009C
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-9 Rear View Showing Components (See Figure 8.2-9)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-10 Traverse Module Components, AC•T 8/10 and AC•T diff Analyzers (See Table 8.2-10)
17
8
1
2
12
9
16
11
10
10
4
2
9
15
7
5
13
14 6
7339007C
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-10 Traverse Module Components, AC•T 8/10 and AC•T diff Analyzers (See Figure 8.2-10)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-13 Left Side and Top Views (See Table 8.2-13)
Original 1
Ac •T 8/10 &
17
Ac •T diff analyzers 18 19
Ac •T diff2
analyzer 2
3 4
16 15 9 8
14
13
or
10
7339046F
12 11
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-13 Left Side and Top Views (See Figure 8.2-13)
6707111 Card, Analyzer, Universal (for AC•T diff and AC•T diff 2 analyzers only)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-14 Aperture/Bath Assembly Components on Mounting Plate, AC•T 8/10 and AC•T diff
Analyzers (See Table 8.2-14)
2 7
2 6
2 5
9 1
2 4
8 1 0
7 6
2 5 5
4
2 3 3
2
2 5
2 2
1 1
2 1
1 2
1 3
1 4
8
1 5
2 0 1 9 1 8
1 6
1 7
7 3 3 9 0 0 1 F
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS 8
Table 8.2-14 Aperture/Bath Assembly Components on Mounting Plate, AC•T 8/10 and AC•T diff Analyzers
(See Figure 8.2-14)
PN 4237339F
PARTS LISTS
ILLUSTRATED PARTS
Figure 8.2-16 Aperture/Bath Assembly Shield and Mounting Plate (See Table 8.2-16)
4
2
1
7339047A
Table 8.2-16 Aperture/Bath Assembly Shield and Mounting Plate (See Figure 8.2-16)
PN 4237339F
CONTENTS 9
A QUICK REFERENCE INFORMATION, A.1-1
A.4 AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS, A.4-1
Summary, A.4-1
Solenoids, A.4-1
Main Analyzer Components, A.4-2
Components Located Behind the Front Door of an AC•T 8/10 or AC•T diff
Analyzer, A.4-2
Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff
Analyzer, A.4-5
PN 4237339F
CONTENTS
ILLUSTRATIONS
A.2-1 Analyzer Card Component Locations - AC•T 8/10 Analyzers, A.2-1
A.2-2 Analyzer Card Component Locations - AC•T diff Analyzer, A.2-5
A.2-3 Analyzer Card Component Locations - AC•T diff 2 Analyzer, A.2-9
A.2-4 Universal AC•T diff/AC•T diff 2 Analyzer Card Component Locations, A.2-13
A.2-5 Display Controller Driver Card Component Locations - (Through-Hole
Design), A.2-19
A.2-6 Display Controller Driver Card Component Locations - (Surface-Mount
Design), A.2-20
A.3-1 Software Menu Tree - AC•T 8/10 Analyzer, A.3-1
A.3-2 Software Menu Tree - AC•T diff Analyzer, A.3-2
A.3-3 Software Menu Tree - AC•T diff/Veterinary Option Analyzer, A.3-3
A.3-4 Software Menu Tree - AC•T diff 2 Analyzer, A.3-4
A.4-1 View of an AC•T 8/10 or AC•T diff Analyzer with the Front Door Open, A.4-2
A.4-2 View Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer, A.4-5
A.4-3 View Inside the Left Compartment of an AC•T 8/10 or AC•T diff Analyzer, A.4-10
A.5-1 View of an AC•T diff 2 Analyzer with the Front Door Open, A.5-3
A.5-2 View Inside the Right Compartment of an AC•T diff 2 Analyzer, A.5-6
A.5-3 View Inside the Pneumatic Pump Compartment of an AC•T diff 2
Analyzer, A.5-10
PN 4237339F
CONTENTS
9
A.5-4 View Inside the Left Compartment of an AC•T diff 2 Analyzer, A.5-13
C.1-1 Epson TM-290P Slip Printer - Location of Controls and Indicators, C.1-1
C.2-1 Canon BJC-250 Bubble Jet Printer - Power/Resume Controls and
Indicators, C.2-1
C.2-2 Canon BJC-250 Bubble Jet Printer - Cartridge Lock and Paper Thickness
Levers, C.2-2
C.3-1 OKIPAGE 14e LED Printer Front Operator Panel, C.3-4
TABLES
A.1-1 Analyzer Card - AC•T 8/10 Analyzer, Voltage Limits, A.1-1
A.1-2 Analyzer Card - AC•T diff Analyzer, Voltage Limits, A.1-1
A.1-3 Analyzer Card - AC•T diff 2 Analyzer, Voltage Limits, A.1-1
A.1-4 Target Acceptance Criteria, A.1-1
A.1-5 Latex Adjustment Criteria, A.1-2
A.1-6 Power Supply Module Output, Connector P5 - AC•T 8/10 Analyzer, A.1-2
A.1-7 Power Supply Module Output, Connector P5 - AC•T diff and AC•T diff 2
Analyzers, A.1-2
A.2-1 Analyzer Card Connectors - AC•T 8/10 Analyzers (See Figure A.2-1), A.2-2
A.2-2 Analyzer Card Test Points - AC•T 8/10 Analyzers (See Figure A.2-1), A.2-4
A.2-3 Analyzer Card Connectors - AC•T diff Analyzer (See Figure A.2-2), A.2-6
A.2-4 Analyzer Card Jumpers - AC•T diff Analyzer (See Figure A.2-2), A.2-7
A.2-5 Analyzer Card Test Points - AC•T diff Analyzer (See Figure A.2-2), A.2-8
A.2-6 Analyzer Card Connectors - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-10
A.2-7 Analyzer Card Jumpers - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-12
A.2-8 Analyzer Card Test Points - AC•T diff 2 Analyzer (See Figure A.2-3), A.2-12
A.2-9 Universal AC•T diff/AC•T diff 2Analyzer Card Connectors (See
Figure A.2-4), A.2-14
A.2-10 Universal AC•T diff/AC•T diff 2 Analyzer Card Jumper Settings (See
Figure A.2-4), A.2-17
A.2-11 Universal AC•T diff/AC•T diff 2Analyzer Card Test Points (See
Figure A.2-4), A.2-18
A.2-12 Display Controller Driver Card Connectors, A.2-20
A.2-13 Display Controller Driver Card - J61 Pinouts, A.2-21
A.2-14 Display Controller Driver Card, J62 Pinouts, A.2-22
A.2-15 Display Controller Driver Card, J63 Pinouts, A.2-22
A.2-16 Display Controller Driver Card, J65 Pinouts, A.2-22
A.2-17 Display Controller Driver Card Jumpers, A.2-23
A.2-18 Display Controller Driver Card Test Points, A.2-23
A.4-1 Solenoid Functions - AC•T 8/10 or AC•T diff Analyzers, A.4-1
A.4-2 Components behind the Front Door of an AC•T 8/10 or AC•T diff Analyzer, A.4-3
A.4-3 Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff
Analyzer, A.4-6
A.4-4 Components Inside the Left Compartment of an AC•T 8/10 or AC•T diff
Analyzer, A.4-10
A.5-1 Solenoid Functions - AC•T diff 2 Analyzers, A.5-1
A.5-2 Components behind the Front Door of an AC•T diff 2 Analyzer, A.5-3
A.5-3 Components Inside the Right Compartment of an AC•T diff 2 Analyzer, A.5-6
A.5-4 Components Inside the Pneumatic Pump Compartment of an AC•T diff 2
Analyzer, A.5-11
PN 4237339F
CONTENTS
A.5-5 Components Inside the Left Compartment of an AC•T diff 2 Analyzer, A.5-13
B.1-1 Warning (Non-Fatal) Error Messages (Icons), B.1-1
B.1-2 Fatal Error Messages (Numbers), B.1-1
C.1-1 Epson TM-290P Slip Printer - Specifications, C.1-1
C.1-2 Epson TM-290P Slip Printer - Function of Controls and Indicators, C.1-2
C.1-3 Epson TM-290P Slip Printer - DIP Switch Settings for SW1, C.1-2
C.1-4 Epson TM-290P Slip Printer - International Character Sets - DIP Switch Settings
for SW1-2 through SW1-4, C.1-2
C.2-1 Canon BJC-250 Bubble Jet Printer - Specifications, C.2-1
C.2-2 Canon BJC-250 Bubble Jet Printer - Function/Status of Controls and
Indicators, C.2-2
C.3-1 OKIPAGE14e LED Printer - Specifications, C.3-1
C.3-2 Approved Menu Settings Printed Using the Front Panel, C.3-2
C.3-3 Approved Menu Settings Printed from the Installation Software Print Menu, C.3-3
C.3-4 OKIPAGE 14e LED Printer - Operator Panel Lights - Reference, C.3-4
PN 4237339F
AQUICK REFERENCE INFORMATION A
A.1 TOLERANCES AND LIMITS
Analyzer Card Voltage Limits
PN 4237339F
QUICK REFERENCE INFORMATION
TOLERANCES AND LIMITS
Table A.1-7 Power Supply Module Output, Connector P5 - AC•T diff and AC•T diff 2 Analyzers
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
A.2 CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Component Locations
Figure A.2-2 Analyzer Card Component Locations - AC•T diff Analyzer
LV1 thru 20
LV1 2 3 4 5 6 7 8 9 10
M1
S1
X9 S9
M2
J30
S10
S2
S3
J14
J4
J6 S8
S13
TP3 TP4
J9 X8
TP26 J3
TP25
J5
J2
TP5 W APT
J8
TP6 X5
TP9
TP8 R APT X4
TP10
J7
X3 U7
X2 X6 X7 J10 E2 E3 E4 E5 E6 E7 E8 E9 E10 7339051F
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Connectors
Table A.2-3 Analyzer Card Connectors - AC•T diff Analyzer (See Figure A.2-2)
Reference
Designator Card
DCN 6323022 Label Plug Location Description
J2 J2 P2 Lower rear Parallel Printer, rear panel interface, 25 conductor
J3 J3 P3 Center rear Serial out, rear panel interface, 9 conductor
J4 J4 P4 Center rear Waste sensor, rear panel interface, BNC connector
J5 J5 P5 Lower rear Power in, from Power Supply module, 9 conductor
J6 J6 P6 Top center Cover interlock, currently just jumpered, 2 conductor
J7 J7 P7-R Lower front Red aperture cable, coaxial
J8 J8 P8-W Lower front White aperture cable, coaxial
J9 J9 P9 Lower front Hgb lamp, 4 conductor
J10 J10 P10 Lower front Hgb sensor (preamp), coaxial
J11 LV22 Unused Top rear Spare, solenoid 22, dc motor, 2 conductor
J12 LV23 WM23 Top rear Solenoid 23, vacuum pump, 2 conductor
J13 S13 WM13 Center rear Sensor 13, diluent reservoir thermistor sensor,
3 conductor
J14 J14 P14 Center front Traverse module, flex-connect cable, 20 conductor
J15 S1 WM1 Top rear Diluent reservoir sensor, 4 conductor
J16 S12 Unused Top rear Spare sensor for M6, 4 conductor
J17 S9 WM9 Top rear Syringe motor sensor, 4 conductor
J18 S10 WM10 Center rear Sensor 10, vacuum transducer, 3 conductor
J19 S2 WM2 Center rear Sensor 2, lytic reagent optical-fluid sensor, 3 conductor
J20 S3 WM3 Center rear Sensor 3, AC•T Rinse, optical-fluid sensor, 3 conductor
J21 S14 Unused Center rear Spare, sensor 14, 3 conductor
J22 S15 Unused Center rear Spare, sensor 15, 3 conductor
J23 S16 Unused Center rear Spare, sensor 16, 3 conductor
J24 M5 WM5 Top rear Motor 5, syringe-assembly motor, 5 conductor
J25 M6 Unused Top rear Spare, motor 6, spare sensor motor, 5 conductor
J26 M1 WM1 Top rear Motor 1, waste peristaltic pump motor, 5 conductor
J27 M2 WM2 Top rear Motor 2, diluent peristaltic pump motor, 5 conductor
J28 S8 WM8 Center rear Sensor 8, aspirate switch, 3 conductor
J29 J29 P29 Top front Flash Memory card, 68 conductor
J30 J30 P30 Center front Display assembly ribbon cable, 26 conductor
J31 LV1 WM1 Top rear, upper row Solenoid 1, 2 conductor
J32 LV2 WM2 Top rear, upper row Solenoid 2, 2 conductor
J33 LV3 WM3 Top rear, upper row Solenoid 3, 2 conductor
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Table A.2-3 Analyzer Card Connectors - AC•T diff Analyzer (See Figure A.2-2) (Continued)
Reference
Designator Card
DCN 6323022 Label Plug Location Description
J34 LV4 WM4 Top rear, upper row Solenoid 4, 2 conductor
J35 LV5 WM5 Top rear, upper row Solenoid 5, 2 conductor
J36 LV6 WM6 Top rear, upper row Spare, solenoid 6, 2 conductor
J37 LV7 WM7 Top rear, upper row Solenoid 7, 2 conductor
J38 LV8 WM8 Top rear, upper row Solenoid 8, 2 conductor
J39 LV9 WM9 Top rear, upper row Solenoid 9, 2 conductor
J40 LV10 WM10 Top rear, upper row Solenoid 10, 2 conductor
J41 LV11 WM11 Top rear, lower row Solenoid 11, 2 conductor
J42 LV12 WM12 Top rear, lower row Solenoid 12, 2 conductor
J43 LV13 WM13 Top rear, lower row Solenoid 13, 2 conductor
J44 LV14 WM14 Top rear, lower row Solenoid 14, 2 conductor
J45 LV15 WM15 Top rear, lower row Solenoid 15, 2 conductor
J46 LV16 WM16 Top rear, lower row Solenoid 16, 2 conductor
J47 LV17 WM17 Top rear, lower row Solenoid 17, 2 conductor
J48 LV18 Unused Top rear, lower row Spare, solenoid 18, 2 conductor
J49 LV19 Unused Top rear, lower row Spare, solenoid 19, 2 conductor
J50 LV20 Unused Top rear, lower row Spare, solenoid 20, 2 conductor
Jumpers
X5
Table A.2-4 Analyzer Card Jumpers - AC•T diff Analyzer (See Figure A.2-2)
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Test Points
Table A.2-5 Analyzer Card Test Points - AC•T diff Analyzer (See Figure A.2-2)
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Analyzer Card - Universal AC•T diff/AC•T diff 2 Analyzers
r For component locations, refer to Figure A.2-4.
r For details concerning a specific connector, refer to Table A.2-10.
r For details concerning a specific jumper, refer to Table A.2-10.
Note: The jumpers on the Universal AC•T diff/AC•T diff 2 Analyzer card configure the
card for the model of the instrument and for the type of peristaltic or FMI sample pump
(full or half winding) currently in use on the instrument. Table A.2-11 lists both the wire
colors for pump motor connectors J26, J27, and J24 to help you determine if the pump is
full or half winding, and the corresponding jumper settings.
r For details concerning a specific test point, refer to Table A.2-11.
Component Locations
Figure A.2-4 Universal AC•T diff/AC•T diff 2 Analyzer Card Component Locations
LV1 2 3
CB8 X1 X10
4 5 6 7 8 9 10 LV22 LV23
LV11 12 13 14 15 16 17 18 19 20 LV23 M5
J29 TP32
X12
M6
TP44
X11
TP35
TP33
M1
X9 S1
J30 M2
X1 S9
S10A
S2
M3 TP36
TP34
J6
TP3 TP45 TP2 TP1
J4
J14 TP15 TP20 S19
J9 J3
X4 TP24 TP37 S3
TP38 TP42
TP17 TP41 S8
TP26
J5 TP23 S18
TP43
TP5,6,9 S13
W APT TP11 TP31
J8 TP27 J2
TP16
TP12 TP29
TP8,10 TP30
TP39
X8
TP21 TP13
X3 TP19
TP28
R APT TP18 TP14
X2 TP22
TP4
TP40
J7 TP7
U7
X6 X7 J10 E2 E3 E4 E5 E6 E7 E8 E9 E10
7339146F
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Connectors
Table A.2-9 Universal AC•T diff/AC•T diff 2Analyzer Card Connectors (See Figure A.2-4)
Reference
Designator Card
DCN 6323409 Label Plug Location Description
J2 J2 P2 Lower rear Parallel Printer, rear panel interface, 25 conductor
J3 J3 P3 Center rear Serial out, rear panel interface, 9 conductor
J4 J4 P4 Center rear Waste sensor, rear panel interface, BNC connector
J5 J5 P5 Lower rear Power in, 200 Vdc from Power Supply module, 9 conductor
J6 J6 P6 Top center Cover interlock, currently just jumped, 2 conductor
J7 J7 P7-R Lower front Red aperture cable, coaxial
J8 J8 P8-W Lower front White aperture cable, coaxial
J9 J9 P9 Lower front Hgb lamp, 4 conductor
J10 J10 P10 Lower front Hgb sensor (preamp), coaxial
J11 LV22 Unused Top rear Spare, solenoid 22, dc motor, 2 conductor
J12 LV23 WM23 Top rear Vacuum pump, 2 conductor
J13 S13 WM13 Center rear Sensor 13, diluent reservoir thermistor sensor, 3 conductor
J14 J14 P14 Center front Traverse module, flex-connect cable, 20 conductor
J15 S1 WM1 Top rear Diluent reservoir sensor, float, 4 conductor
J16 S12 Unused Top rear Spare sensor for M6, 4 conductor
J17 S9 WM9 Top rear AC•T diff analyzer: Syringe motor sensor, 4 conductor
AC•T diff 2 analyzer: Sensor 9, sample pump home sensor,
4 conductor
J18A S10A WM10 Center rear Sensor 10A, vacuum transducer, 3 conductor
J19 S2 WM2 Center rear Sensor 2, lytic reagent optical-fluid sensor, 3 conductor
J20 S3 WM3 Center rear Sensor 3, AC•T Rinse, optical-fluid sensor, 3 conductor
J21 S14 WM14 Center rear AC•T diff analyzer: Not used.
AC•T diff 2 analyzer: Sensor 14, open-vial horizontal position
sensor, 3 conductor
J22 S15 WM15 Center rear AC•T diff analyzer: Not used.
AC•T diff 2 analyzer: Sensor 15, closed-vial horizontal
position sensor, 3 conductor
J23 S16 WM16 Center rear AC•T diff analyzer: Not used.
AC•T diff 2 analyzer: Sensor 16, RBC bath horizontal position
sensor, 3 conductor
J24 M5 WM5 Top rear AC•T diff analyzer: Motor 5, syringe-assembly motor,
5 conductor
AC•T diff 2 analyzer: Motor 5, sample FMI pump,
5 conductor
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Table A.2-9 Universal AC•T diff/AC•T diff 2Analyzer Card Connectors (See Figure A.2-4) (Continued)
Reference
Designator Card
DCN 6323409 Label Plug Location Description
J25 M6 WM25 Top rear AC•T diff analyzer: Not used.
AC•T diff 2 analyzer: Motor 6, closed-vial station door motor,
5 conductor
J26 M1 WM1 Top rear AC•T diff analyzer: Motor 1, waste peristaltic pump motor,
5 conductor
AC•T diff 2 analyzer: Motor 1, lyse FMI pump motor,
5 conductor
J27 M2 WM2 Top rear AC•T diff analyzer: Motor 2, diluent peristaltic pump motor,
5 conductor
AC•T diff 2 analyzer: Motor 2, diluent FMI pump motor,
5 conductor
J28 S8 WM8 Center rear AC•T diff analyzer: Sensor 8, aspirate switch, 3 conductor
AC•T diff 2 analyzer: Sensor 8, open-vial station aspirate
switch, 3 conductor
J29 J29 P29 Top front Flash Memory card, 68 conductor
J30 J30 P30 Center front Display assembly ribbon cable, 26 conductor
J31 LV1 WM1 Top rear, upper row AC•T diff analyzer: Solenoid 1, 2 conductor
AC•T diff 2 analyzer: Rinse pump
J32 LV2 WM2 Top rear, upper row Solenoid 2, 2 conductor
J33 LV3 WM3 Top rear, upper row Solenoid 3, 2 conductor
J34 LV4 WM4 Top rear, upper row Solenoid 4, 2 conductor
J35 LV5 WM5 Top rear, upper row Solenoid 5, 2 conductor
J36 LV6 WM6 Top rear, upper row AC•T diff analyzer: Not used.
AC•T diff 2 analyzer: Waste pump
J37 LV7 WM7 Top rear, upper row Solenoid 7, 2 conductor
J38 LV8 WM8 Top rear, upper row Solenoid 8, 2 conductor
J39 LV9 WM9 Top rear, upper row Spare, solenoid 9, 2 conductor
J40 LV10 WM10 Top rear, upper row Solenoid 10, 2 conductor
J41 LV11 WM11 Top rear, lower row Solenoid 11, 2 conductor
J42 LV12 WM12 Top rear, lower row Solenoid 12, 2 conductor
J43 LV13 WM13 Top rear, lower row Solenoid 13, 2 conductor
J44 LV14 WM14 Top rear, lower row Spare, solenoid 14, 2 conductor
J45 LV15 WM15 Top rear, lower row Solenoid 15, 2 conductor
J46 LV16 WM16 Top rear, lower row Solenoid 16, 2 conductor
J47 LV17 WM17 Top rear, lower row Solenoid 17, 2 conductor
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Table A.2-9 Universal AC•T diff/AC•T diff 2Analyzer Card Connectors (See Figure A.2-4) (Continued)
Reference
Designator Card
DCN 6323409 Label Plug Location Description
J48 LV18 Unused Top rear, lower row AC•T diff analyzer: Not used
AC•T diff 2 analyzer: Waste/cleaner select
J49 LV19 WM19 Top rear, lower row AC•T diff analyzer: Not used
AC•T diff 2 analyzer: Diluent reservoir pump
J50 LV20 Unused Top rear, lower row AC•T diff analyzer: Not used
AC•T diff 2 analyzer: Fan, back panel
J57 M3 WM3 Center front Motor 3, traverse horizontal motor
J80 S24 WM24 Top center Sensor 24, specimen tube bottom seeking sensor
J81 S17 WM17 Top center Sensor 17, WBC bath horizontal position sensor
J82 S18 WM18 Center rear Sensor 18, lyse pump home sensor
J83 S19 WM19 Center rear Sensor 19, diluent pump home sensor
J98 S25 WM25 Top center Sensor 25, closed-vial aspirate sensor
(door-closed position sensor)
Jumpers
The jumpers on the Universal AC•T diff/AC•T diff 2 Analyzer card configure the card for the
model of the instrument and for the type of peristaltic or FMI sample pump (full or half
winding) currently in use on the instrument. Table A.2-11 lists both the wire colors for pump
motor connectors J26, J27, and J24 to help you determine if the pump is full or half winding,
and the corresponding jumper settings.
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Table A.2-10 Universal AC•T diff/AC•T diff 2 Analyzer Card Jumper Settings (See Figure A.2-4)
X2 Always OFF Always OFF Always OFF Always OFF Factory use only.
X7 Always OFF Always OFF Always OFF Always OFF Factory use only.
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Test Points
Table A.2-11 Universal AC•T diff/AC•T diff 2Analyzer Card Test Points (See Figure A.2-4)
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Display Controller Driver Card
r For component locations on the circuit card using through-hole design, refer to
Figure A.2-5.
r For component locations on the circuit card using surface-mount design, refer to
Figure A.2-6.
r For details concerning a specific connector, refer to Table A.2-12.
r For J61 pinout details, refer to Table A.2-13.
r For J62 pinout details, refer to Table A.2-14.
r For J63 pinout details, refer to Table A.2-15.
r For J65 pinout details, refer to Table A.2-16.
r For details concerning a specific jumper on the circuit card using through-hole design,
refer to Table A.2-17.
Note: The circuit card using surface-mount design has no jumpers.
r For details concerning a specific test point, refer to Table A.2-18.
Component Locations
Figure A.2-5 Display Controller Driver Card Component Locations - (Through-Hole Design)
X4
J66
X1 TP1
X3
J65
X2
TP3
J62
J63
TP4
J61
J64
TP2
7339133E
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Figure A.2-6 Display Controller Driver Card Component Locations - (Surface-Mount Design)
J67
J65
J66
J62
J61
TP1
TP3
J63
TP4
J64
TP2
7339132E
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
.
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS
Columns: Rows:
Pin Front Flex Cable Destination Pin (J63) Back Flex Cable Destination Pin (J63)
1 4 9 1 1
2 3 10 2 2
3 2 11 3 3
4 1 12 4 4
5 5 13 5 5
6 6 14 6 6
7 7 15 7 7
8 8 16 8 8
PN 4237339F
QUICK REFERENCE INFORMATION
CIRCUIT CARD CONNECTORS, JUMPERS, AND TEST POINTS A
Jumpers
Test Points
PN 4237339F
QUICK REFERENCE INFORMATION
SOFTWARE MENU TREES
TRANSMISSION VERIFY
PREDILUTE CONTROL
DISPENSE INFO
LYTIC REAGENT
CALIBRATION SAMPLE
FACTORS DETAILS ASSAY VALUES
ZAP AND RANGES
APERTURES
PRINTERS/ MOTORS
PROFILES CALIBRATION
RINSE + MIX TARGET VALUES
LABORATORY PULSE
ID REPRODUCIBILITY
DRY PRIME
DILUENT RUN
LATEX
PRINT SETUP CALIBRATION
REPORT SWEEP FLOW CARRYOVER
RUN
PREPARE TO
CLEAN BATHS SHIP 7339058F
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
A.4 AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
Summary
Most Diluter functions are accomplished by fluidic components that are interconnected by
tubing and controlled by timed solenoid signals. This section briefly describes the functions
of these fluidic components and shows their locations.
Solenoids
Solenoids are electronic valves that are energized or de-energized to perform a function or to
activate other components in the Diluter. Table A.4-1 identifies the function of each solenoid
used in the AC•T 8/10 and AC•T diff analyzers. Solenoids LV7, LV8, LV9, LV10, LV11, and
LV12 are attached to and control the operation of ASCO® Angar™ valves. Solenoids LV13,
LV14, and LV15 are attached to and control the operation of small Bio-Chem™ pinch valves.
Use the Pneumatic/Hydraulic Schematic or the figures referenced in Table A.4-1 to locate
these solenoids. (For the part number and location of the Pneumatic/Hydraulic Schematics
schematic file, see Chapter 6.)
Solenoid Figure
Number Function Reference
LV1 VIC vent select A.4-2, 7
LV2 High vacuum enable A.4-2, 8
LV3 WBC air mix select A.4-2, 9
LV4 Lytic reagent air mix select A.4-2, 10
LV5 Air mix enable A.4-2, 11
LV6 Spare
LV7* RBC bath diluent select A.4-2, 3
LV8* Probe-wipe waste select A.4-2, 4
LV9* Lytic reagent dispense enable A.4-2, 1
LV10* Probe-wipe diluent select A.4-2, 2
LV11* Bath diluent select A.4-2, 5
LV12* Syringe diluent select A.4-2, 6
LV13† Cleaner select A.4-2, 14
LV14† WBC bath drain select A.4-2, 21
LV15† VC1 drain select A.4-2, 20
LV16 RBC count A.4-1, 8
LV17 WBC count A.4-1, 7
LV23 Pneumatic pump motor A.4-3, 4
Note: Solenoids LV18 through LV22 are not used on the AC•T 8/10 and AC•T diff analyzers.
* Solenoid is attached to and controls the operation of an Angar valve.
† Solenoid is attached to and controls the operation of a small Bio-Chem pinch valve.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
r Table A.4-2 describes the main components behind the front door. Use Figure A.4-1 to
locate these components.
r Table A.4-3 describes the main components inside the right compartment. Use
Figure A.4-2 to locate these components.
r Table A.4-4 describes the main components inside the left compartment. Use
Figure A.4-3 to locate these components.
Components Located Behind the Front Door of an AC•T 8/10 or AC•T diff Analyzer
Figure A.4-1 View of an AC•T 8/10 or AC•T diff Analyzer with the Front Door Open
1 2
1. Software card slot
3. Vacuum chamber
13
4. Hgb assembly VAC ADJ
5. Sweep-flow spool
6. WBC bath
7. VL17
8. VL16
9. RBC bath RBC WBC
12
10. Aspirate switch
7339123E 10 9 8 7 6 5 4
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
Table A.4-2 Components behind the Front Door of an AC•T 8/10 or AC•T diff Analyzer
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-1, 1 Software card slot Opening for inserting the metallic software card. This
software card must be firmly seated into the track of the
slot; otherwise, the card could fall inside the analyzer.
A.4-1, 2 Vacuum adjust knob Knob used to the adjust the vacuum regulator (RG1) which VAC RGL TR
regulates the low (6-in. Hg) vacuum used to pull the
dilution through the WBC and RBC aperture and
sweep-flow diluent behind the RBC aperture.
A.4-1, 3 Vacuum chamber Distributes low vacuum (6-in. Hg) to pull dilution through VC1 or
the aperture in both baths and sweep-flow diluent through VAC ISLR
the sweep-flow line behind the RBC aperture. Also CHMBR
provides the high vacuum needed for scavenging the
sweep-flow line behind the RBC bath, and for cleansing
and drying the probe inside the probe-wipe housing. Also
acts as a holding chamber for the waste that accumulates
during these count, scavenge, and cleaning functions. May
also be referred to as a vacuum isolator chamber.
A.4-1, 4 Hgb assembly Hgb lamp and preamp provide the hemoglobin blank and
sample readings needed to calculate the Hgb result.
A.4-1, 5 Sweep-flow spool Tubing wound around a spool that acts as a choke to CK2
regulate diluent delivery through the sweep-flow line
behind the RBC aperture.
A.4-1, 6 WBC bath Holds the WBC dilution for mixing and for collecting WBC WBC
and Hgb data, including differential data if the analyzer is
an AC•T diff analyzer. May also be referred to as the WBC
aperture/bath assembly.
A.4-1, 7 LV17 Angar two-way solenoid valve used to apply the low LV17
vacuum (6-in. Hg) in the vacuum chamber (VC1) to the
WBC aperture for accumulating count and size data.
De-energized - the pathway from VC1 to the WBC aperture
is closed.
Energized - the pathway from VC1 to the WBC aperture is
open. As low vacuum is applied to the WBC aperture,
dilution is pulled from the WBC bath through the WBC
aperture via LV17 and enters the VC1 as a stream of
droplets.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
Table A.4-2 Components behind the Front Door of an AC•T 8/10 or AC•T diff Analyzer (Continued)
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-1, 8 LV16 Angar two-way solenoid valve used to apply the low LV16
vacuum (6-in. Hg) in the vacuum chamber (VC1) to the
RBC aperture for accumulating count and size data.
De-energized - the pathway from VC1 to the RBC aperture
is closed.
Energized - the pathway from VC1 to the RBC aperture is
open. As low vacuum is applied to the RBC aperture,
dilution is pulled from the RBC bath through the RBC
aperture via LV16 and enters the VC1 as a stream of
droplets. Vacuum applied to the RBC aperture is
simultaneously applied to the sweep-flow system.
A.4-1, 9 RBC bath Holds the RBC dilution for mixing and for collecting RBC RBC
and Plt data. May also be referred to as the RBC
aperture/bath assembly.
A.4-1, 10 Aspirate switch When pressed by the operator, initiates a cycle in the
selected mode of operation - either the Whole-Blood mode
or Predilute mode.
A.4-1, 11 Probe Input pathway for sample aspiration in the Whole-Blood ASP PROBE
and Predilute modes. Also the input pathway for aspirating
a portion of the WBC dilution from the WBC bath for
transfer to the RBC bath. May also be referred to as the
aspirate probe.
A.4-1, 12 Probe-wipe housing Housing has an internal spiral configuration that allows
diluent entering the top port to clean the outside of the
probe as the diluent spirals downward and exits through
the bottom port to the vacuum chamber (VC1). May also
be referred to as the probe-wipe block.
A.4-1, 13 Traverse module Encases and protects the probe. Controls horizontal and
vertical positioning of the probe using stepper motors,
direct drive gear systems, and dual-channel opto sensors.
Also cleanses the probe after aspiration. Figure 2.2-1
illustrates the main components.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer
Figure A.4-2 View Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer
1. LV9 1 2 3 4 5 6
2. LV10
3. LV7 23
4. LV8
5. LV11
7
6. LV12 8
7. LV1 9
8. LV2 10
9. LV3 11
10. LV4
11. LV5 12
12. Waste/rinse pump (PM1)
13. Diluent pump (PM2)
14. LV13
15. Diluent reservoir 13
16. Aspiration syringe (250 µL) 22
17. Diluent syringe (5 mL)
18. Syringe module
19. Lytic reagent syringe (1 mL) 14
20. LV15
21. LV14 21 20 19 18 17 16 15 7339124E
22. Rinse (cleaner) optical sensor
23. Vacuum transducer
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
Table A.4-3 Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-2, 1 LV9 Angar three-way solenoid valve used to control lytic LV9
reagent usage.
De-energized - provides a bidirectional pathway between
the lytic reagent source and the 1-mL lytic reagent syringe.
Energized - provides a pathway between the 1-mL lyse
syringe and the left-side port of the WBC bath.
A.4-2, 2 LV10 Angar three-way solenoid valve used to control diluent LV10
movement.
De-energized - provides a pathway from the diluent source
to the reservoir for diluent being pumped by peristaltic
pump PM2.
Energized - provides a pathway from the diluent source to
the top port of the probe-wipe housing for diluent being
pumped by peristaltic pump PM2.
A.4-2, 3 LV7 Angar three-way solenoid valve used to route diluent LV7
dispensed by the sample syringe via LV11 to the lower-side
ports of the baths.
De-energized - routes diluent to the right-side port of the
WBC bath to prefill the bath.
Energized - routes diluent through a Y-connector (FY8) to
both the left- and right-side ports of the RBC bath to prefill
the bath.
A.4-2, 4 LV8 Angar two-way solenoid valve used to connect the vacuum LV8
chamber (VC1) to the bottom port of the probe-wipe
housing.
De-energized - the pathway from the bottom port of the
probe-wipe housing to VC1 is closed.
Energized - the pathway from the bottom port of the
probe-wipe housing to the VC1 is open to allow high
vacuum to either pull diluent through the housing to clean
the outside of the probe or function as a source of air for
drying the outside of the probe.
A.4-2, 5 LV11 Angar three-way solenoid valve used to control diluent LV11
output from the sample syringe.
De-energized - top of the probe is connected to the sample
syringe for aspiration or diluent dispense.
Energized - routes diluent from the sample syringe to LV7
for bath prefill.
A.4-2, 6 LV12 Angar three-way solenoid valve used to control the routing LV12
of diluent. De-energized - provides a bidirectional pathway
between the diluent reservoir and the 5-mL diluent syringe.
Energized - provides a unidirectional pathway from the
5-mL diluent syringe to the 250-µL sample syringe and
LV11.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
Table A.4-3 Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer (Continued)
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-2, 7 LV1 Two-way solenoid valve used to control venting of the LV1
vacuum chamber (VC1).
De-energized - atmospheric vent closed.
Energized - vents the top of VC1 to atmosphere when the
chamber is draining or during a high vacuum/low vacuum
transition.
A.4-2, 8 LV2 Three-way solenoid valve used to control the vacuum level LV2
going through port 1 to the vacuum chamber (VC1).
De-energized - low vacuum (6-in. Hg) is available at LV2,
port 1.
Energized - high vacuum is available at LV2, port 1.
A.4-2, 9 LV3 Three-way solenoid valve used to switch the mixing LV3
bubbles between the bottom port of the RBC and WBC
baths.
De-energized - supplies mixing bubbles to the RBC bath.
Energized - supplies mixing bubbles to the WBC bath.
A.4-2, 10 LV4 Three-way solenoid valve used to switch pressure between LV4
the WBC bath and LV3.
De-energized - pressure is routed to LV3.
Energized - pressure is routed to the WBC bath lytic reagent
port (left-side port).
A.4-2, 11 LV5 Three-way solenoid valve used to switch the pressure from LV5
the pneumatic pump between atmosphere and LV4.
De-energized - pressure from the pneumatic pump is
vented to atmosphere (necessary for proper operation of a
single-head pump when vacuum is being used in the
system).
Energized - pressure from pneumatic pump routed to LV4
for creating mixing bubbles.
A.4-2, 12 Waste/rinse pump This peristaltic pump functions as either a waste pump or a PM1
rinse pump, depending on its rotation.
r As a waste pump, PM1 rotates clockwise and drains the
WBC bath, the RBC bath, and the vacuum chamber
(VC1).
r As a rinse pump, PM1 rotates counterclockwise and
pumps AC•T Rinse shutdown diluent (cleaner) into the
system filling both baths through their respective waste
ports.
A.4-2, 13 Diluent pump Peristaltic pump used to pull diluent from the diluent PM2
container into the diluent reservoir.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
Table A.4-3 Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer (Continued)
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-2, 14 LV13 Small Bio-Chem solenoid pinch valve directs the flow of LV13
liquid being moved by peristaltic pump PM1.
De-energized - peristaltic pump PM1 is connected to the
waste system for draining the WBC bath, the RBC bath, and
the vacuum chamber (VC1).
Energized - peristaltic pump PM1 is connected to the AC•T
Rinse shutdown diluent (cleaner) supply for input of
cleaning agent during the Shutdown routine.
A.4-2, 15 Diluent reservoir Distribution chamber for supplying diluent to the diluent RSD
syringe. Diluent flow between the reservoir and diluent
syringe is bidirectional to allow diluent flow back into the
reservoir when the Triple Syringe module is energized but
diluent supplied by the diluent syringe is not required. Also
supplies diluent through the sweep-flow line behind the
RBC aperture.
A.4-2, 16 Aspiration syringe (250 µL) Syringe is capable of aspirating different volumes SAMPLE
depending on the mode of operation: SRNG
r 12 µL of whole-blood is aspirated when using the
Whole-Blood mode of operation.
r 20 µL of whole-blood is aspirated when using the
Predilute mode of operation.
May also be referred to as the sample or aspirate syringe.
A.4-2, 17 Diluent syringe (5 mL) Syringe is capable of dispensing different volumes as DILUENT
required in the cycle for prefilling the baths and then SRNG
providing the proper volume of diluent for making the final
WBC and RBC dilutions. The diluent syringe also provides
the diluent needed for rinsing the WBC and RBC baths.
A.4-2, 18 Syringe module Module used to control fluidic movement. From right to
left, consists of a 250-µL sample syringe, a 5-mL diluent
syringe, and 1-mL lytic reagent syringe. One motor
simultaneously drives all three volume syringes. May also
be referred to as the Triple Syringe module.
A.4-2, 19 Lytic reagent syringe (1 mL) Delivers lytic reagent to the WBC bath to lyse RBCs and LYSE SRNG
react with the hemoglobin. Syringe is capable of delivering
different volumes of lytic reagent depending on the cycle:
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
Table A.4-3 Components Inside the Right Compartment of an AC•T 8/10 or AC•T diff Analyzer (Continued)
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-2, 20 LV15 Small Bio-Chem solenoid pinch valve used to route the flow LV15
of waste or shutdown reagent (cleaner).
De-energized - connects peristaltic pump PM1 to LV14.
Depending on the state of LV13, either drains the bath or
fills the bath with shutdown diluent (cleaner) as specified
by the de-energized or energized state of LV14.
Energized - connects peristaltic pump PM1 to the bottom
port of the vacuum chamber (VC1) for draining.
A.4-2, 21 LV14 Small Bio-Chem solenoid pinch valve used to control the LV14
flow of waste from the baths or the flow of shutdown
diluent (cleaner) to the baths.
De-energized - depending on the state of LV13 and LV15,
connects either the waste system or the shutdown diluent
(cleaner) system to the bottom port of the RBC bath.
Energized - depending on the state of LV13 and LV15,
connects either the waste system or the shutdown diluent
(cleaner) system to the bottom port of the WBC bath.
A.4-2, 22 Rinse (cleaner) optical sensor An inline fluid sensor placed in the tubing from the cleaning FS3
reagent source to ensure cleaning reagent is present.
A.4-2, 23 Vacuum transducer Monitors the low vacuum supply in VC1, prior to energizing VAC XDCR
count solenoids VL16 and VL17, to verify the proper level
of low vacuum (6-in. Hg) is available for collecting count
data. May also be referred to as the vacuum sensor.
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS
Components Inside the Left Compartment of an AC•T 8/10 or AC•T diff Analyzer
Figure A.4-3 View Inside the Left Compartment of an AC•T 8/10 or AC•T diff Analyzer
1 . C o n n e c tio n fo r r e a g e n t p ic k u p tu b e 1 1
2 . C o n n e c tio n fo r r e a g e n t p ic k u p tu b e 2
3 . C o n n e c tio n fo r r e a g e n t p ic k u p tu b e 3
4 . P n e u m a tic p u m p
7 3 3 9 1 2 5 F
Table A.4-4 Components Inside the Left Compartment of an AC•T 8/10 or AC•T diff Analyzer
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-3, 1 Connection for reagent Connects the diluent reagent supply to the analyzer.
pickup tube 1
r Reagent 1 of the AC•T Tainer or diff AC•T Tainer reagents.
r Reagent 1 of the AC•T Pak reagents.
A.4-3, 2 Connection for reagent Connects the lytic reagent supply to the analyzer.
pickup tube 2
r Reagent 2 of the AC•T Tainer or diff AC•T Tainer reagents.
r Reagent 2 of the AC•T Pak reagents.
A.4-3, 3 Connection for reagent Connects the cleaning agent supply to the analyzer.
pickup tube 3
r Reagent 3 of the AC•T Tainer or diff AC•T Tainer reagents.
r Use this connection for the AC•T Rinse shutdown diluent
supply.
PN 4237339F
QUICK REFERENCE INFORMATION
AC•T 8/10 and AC•T diff COMPONENT LOCATIONS AND FUNCTIONS A
Table A.4-4 Components Inside the Left Compartment of an AC•T 8/10 or AC•T diff Analyzer (Continued)
Reference
Figure Designator
Reference Component Function DCN 6323083
A.4-3, 4 Pneumatic pump Located underneath the panel floor of the on-board reagent PM3
compartment.
All three types of 24-Vdc pneumatic pumps currently in use, the
Brailsford brushless-type vacuum pump, the ASF Thomas
brush-type, diaphragm vacuum pump, and the KNF
brushless-type, dual-head pump, provide the vacuum and
pressure needed by the system. For details, see Pneumatic
System under Heading 2.2, DILUTER SYSTEM - AC•T 8/10 AND
AC•T diff™ ANALYZERS.
Note: The ASF Thomas vacuum pump diaphragm and motor are
a single assembly.
PN 4237339F
BERROR MESSAGES B
B.1 WARNING AND FATAL ERROR MESSAGE TABLES
Table B.1-1 Warning (Non-Fatal) Error Messages (Icons)
Transmission incomplete
Hgb voltage failure - Hgb Blank or Sample reading outside range 2048 to 4095 A/D units
Channelizing buffer overflow error - usually indicates an overflow in one of the Plt channels
Timekeeper failure
Waste full
Diluent empty
Patient storage data corrupt, AC•T diff and AC•T diff 2 analyzers
PN 4237339F
ERROR MESSAGES
10 All Probe did not reach WBC position. For details, see Probe Did Not
Reach Position Errors 6 through 11 and 19.
11 All Probe did not reach RBC position. For details, see Probe Did Not
Reach Position Errors 6 through 11 and 19.
12 AC•T 8/10 and AC•T diff analyzers Syringe did not reach up position. For details, see Syringe Did Not
Reach/Leave Up Position Errors 12 and 13, AC•T 8/10 and AC•T diff
Analyzers.
13 AC•T 8/10 and AC•T diff analyzers Syringe did not leave up position. For details, see Syringe Did Not
Reach/Leave Up Position Errors 12 and 13, AC•T 8/10 and AC•T diff
Analyzers.
AC•T 8/10 and AC•T diff analyzers Diluent level error during Powerup table execution. For details see
Diluent Level Error During Powerup.
All Vacuum <6 in. Hg before probe wipe. For details. see Vacuum Error.
PN 4237339F
COPTIONAL PRINTERS C
C.1 EPSON TM-290P SLIP PRINTER
Specifications
Table C.1-1 Epson TM-290P Slip Printer - Specifications
Item Specification
Print method Impact dot matrix
Shuttle-type, 7-pin head
Unidirectional printing
Use 0.63 mm column spacing
Use 4.23 mm (1/6 in.) line spacing
Ribbon Exclusive ribbon cassette
Type ERC-27
Purple ink
Life expectancy about 1,500,000 characters
Paper Types:
r Normal (high quality)
r Pressure sensitive.
Carbon copy
Total thickness:
r Single-ply paper (no copy) - 0.09 to 0.25 mm (135 kg paper)
r Copy paper - 0.09 to 0.35 mm.
Maximum copies - 1 original and 2 copies
Interface Uses parallel interface (Centronics compatible)
Figure C.1-1 Epson TM-290P Slip Printer - Location of Controls and Indicators
POWER
RELEASE
RESET PAPER OUT
FORWARD
REVERSE
RELEASE
7303007A
OPTIONAL PRINTERS
Table C.1-2 Epson TM-290P Slip Printer - Function of Controls and Indicators
Control/Indicator Function
RESET button Starts Printer as if power was just turned on
POWER LED (green indicator) Lights when power is on
RELEASE LED (green indicator) Lights when Printer is in release state
Blinks when Printer is in error state
PAPER OUT LED (red indicator) Lights when paper is out
FORWARD key Feeds paper forward
REVERSE key Feeds paper backward
RELEASE key Releases paper
Table C.1-3 Epson TM-290P Slip Printer - DIP Switch Settings for SW1
Positions/Settings
1 2 3 4 5 6 7 8 9 10
OFF ON ON ON OFF OFF OFF OFF OFF OFF
ON = Autofeed Used only for International Not Used
Character Sets (Table C.1-4).
Table C.1-4 Epson TM-290P Slip Printer - International Character Sets - DIP Switch Settings for SW1-2
through SW1-4
Positions/Settings
Item Specification
Print method Bubble jet ink
Ribbon BC-02 black cartridge:
r Life - 500 pages (5% coverage)
r Throughput - 3.1 ppm (standard), 4 ppm (draft)
Paper Weight (auto feeder) - 64 to 105 g/m2 (17 to 28 lb)
Sheet feed capacity (auto feeder):
r 0.4 in.
r 100 sheets
r 75g/m2 (20 lb) paper.
Interface IEEE 1284 with parallel Centronics connector
Noise Level Approximately 43 dB (per ISO 9296)
Operating Temperature 32°F to 95°F (0°C to 35°C)
Power Consumption 20 W (5 W in idle mode)
POWER RESUME
POWER
LIGHT
7479001A
OPTIONAL PRINTERS
CANON BJC-250 BUBBLE JET PRINTER
Figure C.2-2 Canon BJC-250 Bubble Jet Printer - Cartridge Lock and Paper Thickness Levers
Paper thickness
lever
Cartridge lock
(in locked position)
7339060A
Table C.2-2 Canon BJC-250 Bubble Jet Printer - Function/Status of Controls and Indicators
Table C.3-2 Approved Menu Settings Printed Using the Front Panel
Table C.3-4 OKIPAGE 14e LED Printer - Operator Panel Lights - Reference
The following list is a composite of the abbreviations, acronyms and reference designators used in this
manual. When the same abbreviation (or reference designator) is used for more than one word (or type of
component), all meanings relevant to this manual are included, separated by semicolons.
PN 4237339F ABBREVIATIONS-1
ABBREVIATIONS
G mA - milli-amperes
g - grams max - maximum
g/dL - grams per deciliter MB - megabytes
MCH - mean cell hemoglobin
H MCHC - mean cell hemoglobin concentration
Hct - hematocrit MCV - mean corpuscle volume
Hg - mercury MGND - motor ground
Hgb -hemoglobin MHz - megahertz
Hz - hertz mL - milliliters
mm - millimeters
I
i.d. - internal diameter
N
IEEE - Institute of Electrical Engineers nm - nanometers
in. - inches
in. Hg - inches of mercury O
I/O - input/output o.d. - outside diameter
ISO - International Organization for OP - output port
Standardization
P
J P - plug connector
J - receptacle connector PC - printed circuit
PCF - pump correction factor
K PCMCIA - Personal Computer Memory Card
K - kilos International Association
KΩ - kilohms PCV - probe closed vial
KHz - kilohertz PD - persistent data
KNF - Plt - platelet count
PM - pump
L PN - part number
l - long
p-p - peak-to-peak
L - coil
ppm - pages per minute
LCD - liquid crystal display
PUP - probe up position
LED - light emitting diode
LTVAR - long term variance
Q
LV - solenoid Q - transistor
QA - quality assurance
M
MΩ - megohms
M - motor
ABBREVIATIONS
S
S - switch; solenoid
SEMS -
SGND - solenoid ground
SLO-BLO - slow blow
SMT - surface mount technology
SW - switch
T
Temp - temperature
thd - thread
TP - test point
U
U - integrated circuit
V
V - volts
Vac - vacuum; volts alternating current
VAR - variance
ABBREVIATIONS
INDEX
door filters
Cap-Pierce module, disassembly, 4.17-1 blue, location, 8.2-10
Cap-Pierce module, safety latch removal, 4.17-1 blue, maintenance schedule, 5.1-1
door, front blue, part number, 8.2-11
component part numbers, 8.2-7 FLS1. See FLS1
components, illustration, 8.2-6 FLS3. See FLS3
door, front, AC•T diff2 analyzer green, location, 8.2-10
opening, 4.2-3 green, part number, 8.2-11
opening, illustration, 4.2-3 part numbers, 8.1-2
opening, tools/supplies needed, 4.2-3 waste, AC•T diff 2 analyzer, 2.3-17
Flash Memory card
amount of memory, AC•T 8/10 analyzer, 2.1-1
E amount of memory, AC•T diff analyzer, 2.1-1
Editor circuits connector on Analyzer card, AC•T 8/10
voltage used, 2.6-2 analyzer, A.2-2
electrical interference connector on Analyzer card, AC•T diff
troubleshooting, 7.4-1 analyzer, A.2-6
electrode, aperture/bath assembly connector on Analyzer card, AC•T diff 2
positioning, 4.10-4 analyzer, A.2-10, A.2-15
EMC Motor Filter 2 card function, AC•T 8/10 and diff analyzers, 2.1-1
function, AC•T diff 2 analyzer, A.5-12 function, AC•T diff analyzer, 2.6-2
engineering schematics function, AC•T diff 2 analyzer, 2.6-2
included in this manual, 6.1-1 possible cause of PCMCIA (1) error, 7.2-4
Epson printer. See Printer, Epson, C.1-2 removing, illustration, 4.4-1
error icons source of controlling program, 2.5-3
in fatal error messages, 7.2-2 FLS1
in warning error messages, 7.2-1, B.1-1 function in diluent delivery system, AC•T 8/10
error messages, fatal and AC•T diff analyzers, 2.2-3
description, 7.2-2 function in waste and cleaner system, AC•T
table of, 7.2-2, B.1-1 diff 2 analyzer, 2.3-17
error messages, warning FLS2
description, 7.2-1 function in waste and cleaner system, AC•T 8/10
table of, 7.2-1, B.1-1 and AC•T diff analyzers, 2.2-10
error state FLS3
definition, 2.9-1 function in diluent delivery system, AC•T 8/10
exception state and AC•T diff analyzers, 2.2-3
definition, 2.9-1 fluid barrier
exhaust port function in vacuum system, AC•T diff 2
function, AC•T diff 2 analyzer, A.5-12 analyzer, A.5-11
location, illustration, AC•T diff 2 location in vacuum system, illustration, AC•T
analyzer, A.5-10 diff 2 analyzer, A.5-10
external fluidic systems. See pneumatic system; hydraulic
electrical connectors, 2.1-1 systems
FMI pumps
F description, 2.3-17
function, 2.1-1
fans
main components, 2.3-18
part numbers, 8.1-2
main components, illustration, 2.3-18
fatal error messages
FMI pumps, AC•T diff 2 analyzer
See error messages, fatal, 7.2-2
diluent, A.5-6
INDEX
location, illustration, AC•T 8/10 and AC•T diff Lyse Prime cycle
analyzers, A.4-5 amount of lytic reagent dispensed, AC•T 8/10
location, illustration, AC•T diff 2 and AC•T diff analyzers, 2.2-4
analyzer, A.5-10 lyse pump
operational states, AC•T 8/10 and AC•T diff advantages of using an FMI pump, AC•T diff 2
analyzers, A.4-7 analyzer, 2.3-10
operational states, AC•T diff 2 analyzer, A.5-11 component in lytic reagent delivery system,
type of valve, AC•T 8/10 and AC•T diff AC•T diff 2 analyzer, 2.3-11
analyzers, A.4-7 FMI pump, function, AC•T diff 2 analyzer, A.5-7
type of valve, AC•T diff 2 analyzer, A.5-11 function, AC•T diff 2 analyzer, 2.3-10
LV7 location, illustration, AC•T diff 2 analyzer, A.5-6
function, AC•T 8/10 and AC•T diff pump volume, A.5-7
analyzers, A.4-6 See also FMI pumps
function, AC•T diff 2 analyzer, A.5-8 lyse syringe, AC•T 8/10 and AC•T diff analyzers
location, illustration, AC•T 8/10 and AC•T diff function, 2.2-4
analyzers, A.4-5 See also Triple Syringe module
location, illustration, AC•T diff 2 analyzer, A.5-6 Lytic field
operational states, AC•T 8/10 and AC•T diff on Voltages/Sensors screen, AC•T 8/10 analyzer,
analyzers, A.4-6 description, 7.1-5
operational states, AC•T diff 2 analyzer, A.5-8 lytic reagent
type of valve, AC•T 8/10 and AC•T diff amount dispensed for various cycles, AC•T 8/10
analyzers, A.4-6 and AC•T diff analyzers, 2.2-4
type of valve, AC•T diff 2 analyzer, A.5-8 function, 2.2-4, 2.3-10
LV8 lytic reagent delivery system, AC•T 8/10 and AC•T
function, AC•T 8/10 and AC•T diff diff analyzers
analyzers, A.4-6 components and flow, simplified
function, AC•T diff 2 analyzer, A.5-9 schematic, 2.2-4
location, illustration, AC•T 8/10 and AC•T diff description, 2.2-4
analyzers, A.4-5 lytic reagent delivery system, AC•T diff 2 analyzer
location, illustration, AC•T diff 2 analyzer, A.5-6 components and flow, simplified
operational states, AC•T 8/10 and AC•T diff schematic, 2.3-11
analyzers, A.4-6 description, 2.3-10
operational states, AC•T diff 2 analyzer, A.5-9 lytic reagent pump. See lyse pump
type of valve, AC•T 8/10 and AC•T diff lytic reagent syringe, AC•T 8/10 and AC•T diff
analyzers, A.4-6 analyzers
type of valve, AC•T diff 2 analyzer, A.5-9 function, A.4-8
LV9 volume, A.4-8
function, AC•T 8/10 and AC•T diff
analyzers, A.4-6
location, illustration, AC•T 8/10 and AC•T diff M
analyzers, A.4-5 maintenance
operational states, AC•T 8/10 and AC•T diff table showing component, schedule,
analyzers, A.4-6 responsibility, 5.1-1
type of valve, AC•T 8/10 and AC•T diff manifold
analyzers, A.4-6 pneumatic, AC•T 8/10 and AC•T diff
Lyse Dispense Verification cycle analyzers, 2.2-5
amount of lytic reagent dispensed, AC•T 8/10 pneumatic, AC•T diff 2 analyzer, 2.3-11
and AC•T diff analyzers, 2.2-4 manual
abbreviations used in, ABBREVIATIONS-1
conventions, 1.1-6
INDEX
single-head pump, exhaust port, location, vacuum components and flow, simplified
illustration, AC•T diff 2 analyzer, A.5-10 schematic, 2.3-12
vacuum port, location, illustration, AC•T diff 2 Power ON LED
analyzer, A.5-10 circuit, 2.5-6
pneumatic pump compartment power ON/OFF switch
cover removal, illustration of, AC•T diff 2 in Power Supply module, 2.4-2
analyzer, A.5-10 location, illustration, 2.4-2, 4.3-1, 8.2-16
function, AC•T diff 2 analyzer, A.5-8 part number, 8.1-6, 8.2-17
location, illustration, AC•T diff 2 analyzer, A.5-6 Power Supply module
pneumatic pump motor components, 2.4-2
location, illustration, AC•T diff 2 connector P5, AC•T 8/10 analyzer, pin
analyzer, A.5-10 descriptions, A.1-2
pneumatic pump, AC•T 8/10 and AC•T diff connector P5, AC•T diff and AC•T diff 2
analyzers analyzer, pin descriptions, A.1-2
bracket and cover removal, illustration, 4.6-2 connectors, AC•T 8/10 analyzer, 2.4-2
disconnecting for Analyzer card connectors, AC•T diff analyzer and AC•T diff 2
removal/installations, 4.4-2 analyzers, 2.4-2
function in pneumatic system, 2.2-5 description, 2.1-1, 2.4-1
in sleep state, 2.9-1 function, 2.6-1
models in current use, 4.6-1 fuses, AC•T 8/10 and AC•T diff analyzers, 2.4-2
types used, 2.2-5 fuses, AC•T diff 2 analyzer, 2.4-2
pneumatic pump, AC•T diff 2 analyzer Hz rating, 2.1-1
disconnecting for Analyzer card installation, 4.3-2
removal/installations, 4.4-2 main components, illustration, 2.4-2
function in pneumatic system, 2.3-11 part numbers, 8.1-4, 8.2-17
in sleep state, 2.9-1 removal, 4.3-1
location, illustration, ASF Thomas pump, 4.7-2 removal/installation, tools/supplies
location, illustration, KNF dual-head needed, 4.3-1
pump, 4.7-3 removal/installation, verification, 4.3-2
models in current use, 4.7-1 removing to allow top cover and rear panel
replacement, ASF Thomas pump, 4.7-1 removal, AC•T 8/10 and AC•T diff
replacement, ASF Thomas pump, analyzers, 4.2-2
illustration, 4.7-2 replacing, illustration, 4.3-1
replacement, KNF dual-head pump, 4.7-3 schematic, part number, 6.1-1
replacement, KNF dual-head pump, switching supply specifications, table of, 2.4-1
illustration, 4.7-3 voltage range, 2.1-1
type used, 2.3-11 Powerup cycle
pneumatic system, AC•T 8/10 and AC•T diff description, 2.8-2
analyzers error (15), diluent level error during
description, 2.2-5 powerup, 7.2-8
engineering schematic, part number, 6.1-1 initiated when exiting the Diagnostics
pressure components and flow, simplified menu, 7.1-1
schematic, 2.2-8 one of ten instrument cycles, 2.8-1
vacuum components and flow simplified table, duration and activity, AC•T 8/10 and
schematic, 2.2-5 AC•T diff analyzers, 2.8-3
pneumatic system, AC•T diff 2 analyzer table, duration and activity, AC•T diff 2
description, 2.3-11 analyzer, 2.8-4
engineering schematic, part number, 6.1-1 Preamp Gain field
pressure components and flow, simplified on Voltages/Sensors screen, description, AC•T
schematic, 2.3-14 8/10 analyzer, 7.1-3
INDEX
VIC system components and flow, simplified horizontal motor removal, AC•T 8/10 and AC•T
schematic, 2.3-13 diff analyzers, 4.8-1
probe-wipe block horizontal motor removal, AC•T diff 2
function, AC•T diff 2 analyzer, A.5-5 analyzer, 4.9-1
probe-wipe block. See probe-wipe housing horizontal sensor removal, AC•T 8/10 and AC•T
probe-wipe housing diff analyzers, 4.8-1
component in Traverse module, AC•T 8/10 and horizontal sensors removal, AC•T diff 2
AC•T diff analyzers, 2.2-1 analyzer, 4.9-1
function, AC•T 8/10 and AC•T diff latex gain adjustment, AC•T diff and AC•T diff 2
analyzers, A.4-4 analyzers, 4.12-1
function, AC•T diff 2 analyzer, A.5-5 pneumatic pump replacement, AC•T 8/10 and
location, illustration, AC•T 8/10 and AC•T diff AC•T diff analyzers, 4.6-1
analyzers, A.4-2 pneumatic pump, AC•T diff 2 analyzer,
location, illustration, AC•T diff 2 analyzer, A.5-3 replacement, 4.7-1
probe-wipe housing, AC•T 8/10 and AC•T diff Power Supply module replacement, 4.3-1
analyzers preparing instrument for shipment/long-term
location, illustration, 4.8-2 shutdown, 4.15-1
removal, 4.8-1 probe removal, AC•T 8/10 and AC•T diff
probe-wipe housing, AC•T diff 2 analyzer analyzers, 4.8-1
component in Traverse module, 2.3-1 probe removal, AC•T diff 2 analyzer, 4.9-1
location, 2.3-2 probe/tube alignment, AC•T diff 2
removal from traverse housing, analyzer, 4.18-1
illustration, 4.9-4 probe-wipe housing removal, AC•T 8/10 and
procedures AC•T diff analyzers, 4.8-1
Analyzer card replacement, 4.4-1 setting clog detection Target values, AC•T diff
aperture/bath assembly electrode, and AC•T diff 2 analyzers, 4.13-1
positioning, 4.10-4 soaking an FMI pump head, 4.20-1
aperture/bath assembly replacement, 4.10-1 top and rear center panel removal, AC•T diff2
Cap-Pierce module door disassembly, 4.17-1 analyzer, 4.2-4
Cap-Pierce module front door safety latch top cover and rear panel removal, AC•T 8/10 and
removal, 4.17-1 AC•T diff analyzers, 4.2-1
Cap-Pierce module motor removal, 4.17-1 traverse housing removal, AC•T 8/10 and AC•T
Cap-Pierce module optical switch diff analyzers, 4.8-1
removal, 4.17-1 traverse housing removal, AC•T diff 2
Cap-Pierce module removal/disassembly, 4.17-1 analyzer, 4.9-1
decontamination, instrument, 4.21-1 Traverse Interconnect card removal, AC•T 8/10
Display assembly disassembly, 4.5-1 and AC•T diff analyzers, 4.8-1
dummy load test, 4.22-1 Traverse Interconnect card removal, AC•T diff 2
Flash Memory (software) card removal, 4.4-1 analyzer, 4.9-1
FMI pump decrystallization, 4.20-1 Traverse module disassembling/replacement,
front door, AC•T diff2 analyzer, opening, 4.2-3 AC•T 8/10 and AC•T diff analyzers, 4.8-1
general guidelines, 4.1-1 Traverse module disassembly/replacement,
Hgb output adjustment, 4.11-4 AC•T diff 2 analyzer, 4.9-1
Hgb photometer assembly verifying instrument operation in Predilute
disassembly/replacement, 4.11-1 mode, 4.14-1
horizontal drive belt removal, AC•T diff 2 vertical motor removal, AC•T 8/10 and AC•T diff
analyzer, 4.9-1 analyzers, 4.8-1
horizontal drive belt tension adjustment, AC•T vertical motor removal, AC•T diff 2
diff 2 analyzer, 4.9-14 analyzer, 4.9-1
vertical rack removal, AC•T diff 2 analyzer, 4.9-1
INDEX
vertical sensor removal, AC•T 8/10 and AC•T waste. See waste pump
diff analyzers, 4.8-1 waste/rinse, AC•T 8/10 and AC•T diff analyzers
publications volume, A.4-7
on decontamination, 1.2-1 waste/rinse, location, illustration, AC•T 8/10 and
Pulse circuit AC•T diff analyzers, A.4-5
description, 2.5-5
pulse test
AC•T 8/10 analyzer, in troubleshooting, 7.1-22 R
AC•T diff analyzer and AC•T diff 2 analyzers, in RAV
troubleshooting, 7.1-22 conditions that decrease this voltage
pulse width reading, 7.1-3, 7.3-4
factors affecting, 7.3-2 conditions that increase this voltage
measuring, 7.3-1 reading, 7.1-3, 7.3-4
measuring, AC•T 8/10 analyzer, 7.1-10, 7.1-11 typical readings, 7.1-3
measuring, AC•T diff analyzer, 7.1-15 used in clog detection, 7.3-4
varieties of, 7.3-1 RAV field
pumps, 2.3-11 on Voltages/Sensors screen, description, 7.1-3
diaphragm. See diaphragm pumps RAV VAR field
diluent, AC•T 8/10 and AC•T diff analyzers on Sample Details screen, AC•T 8/10
volume, A.4-7 analyzer, 7.1-10
diluent, function, AC•T diff 2 analyzer, A.5-8 RBC
diluent, location, illustration, AC•T 8/10 and erratic results, troubleshooting, 7.4-1
AC•T diff analyzers, A.4-5 high results, troubleshooting, 7.4-1
diluent, location, illustration, AC•T diff 2 low results, troubleshooting, 7.4-1
analyzer, A.5-6 processor circuitry, 2.5-2
diluent. See diluent pump voteout criteria, 7.1-12
FM See FMI pumps RBC aperture
part numbers, 8.1-4 voltage. See RAV
peristaltic. See peristaltic pumps; PM1; PM2 RBC aperture/bath assembly
PM1, location, illustration, AC•T 8/10 and AC•T draining, AC•T 8/10 and AC•T diff
diff analyzers, A.4-5 analyzers, 2.2-9
pneumatic, function, AC•T diff 2 draining, AC•T diff 2 analyzer, 2.3-16
analyzer, A.5-12 function in aperture sensor system, AC•T 8/10
pneumatic, location, illustration, AC•T diff 2 and AC•T diff analyzers, 2.2-8
analyzer, A.5-10 function in aperture sensor system, AC•T diff 2
reservoir fill pump. See reservoir fill pumps analyzer, 2.3-15
reservoir fill, function, AC•T diff 2 function in pneumatic system, AC•T 8/10 and
analyzer, A.5-8 AC•T diff analyzers, 2.2-8
reservoir fill, location, illustration, AC•T diff 2 function in waste and cleaner system, AC•T 8/10
analyzer, A.5-6 and AC•T diff analyzers, 2.2-9
rinse, function, AC•T diff 2 analyzer, A.5-8 function, AC•T diff 2 analyzer, A.5-4
rinse, location, illustration, AC•T diff 2 probe position in Traverse module, AC•T 8/10
analyzer, A.5-6 and AC•T diff analyzers, 2.2-1
rinse. See rinse pump See also aperture/bath assembly
sample. See sample RBC bath
vacuum. See pneumatic pump function, AC•T 8/10 and AC•T diff
waste, function, AC•T diff 2 analyzer, A.5-8 analyzers, A.4-4
waste, location, illustration, AC•T diff 2 function, AC•T diff 2 analyzer, A.5-4
analyzer, A.5-6 location, illustration, AC•T 8/10 and AC•T diff
analyzers, A.4-2
INDEX
setting clog detection Target values, AC•T diff Service Resource Kit CD-ROM
and AC•T diff2 analyzers, 4.13-1 part number, 1.1-1
See also specific field names settings
sample pump Hgb output, 7.1-3
aspiration volumes, AC•T diff 2 analyzer shipment
component in diluent delivery system, AC•T preparing instrument, procedure, 4.15-1
diff 2 analyzer, 2.3-9 Shutdown cycle
function, AC•T diff 2 analyzer, 2.3-10, A.5-6 one of ten instrument cycles, 2.8-1
location, illustration, AC•T diff 2 analyzer, A.5-6 shutdown diluent system. See waste and cleaner
See also FMI pumps system
Sample Results screen shutdown procedure
function in removal/installation of Display long term, 4.15-1
assembly, 4.5-4 signals
in sleep state, 2.9-1 Analyzer card, 2.7-4, A.2-21
Sample1 field APERTURE, 2.5-1
on Sample Details screen, AC•T 8/10 analyzer, Display Controller Driver card, 2.7-4, A.2-21
description, 7.1-12 POWER GOOD/POWER FAIL, 7.2-4
Sample2 field simplified schematics
on Sample Details screen, AC•T 8/10 analyzer, Diluent Delivery System in AC•T 8/10 and AC•T
description, 7.1-12 diff Analyzers, 2.2-3
schematics, engineering Diluent Delivery System in the AC•T diff 2
included in this manual, 6.1-1 Analyzer, 2.3-9
screens Lytic Reagent Delivery System in AC•T 8/10 and
Latex Calibration. See Latex Calibration screen AC•T diff Analyzers-, 2.2-4
Latex Results. See Latex Results screen Lytic Reagent Delivery System in the AC•T diff 2
Motors. See Motors screen Analyzer, 2.3-11
Sample Details. See Sample Details screen Pneumatic System in AC•T 8/10 and AC•T diff
Solenoids. See Solenoids screen Analyzers - Pressure, 2.2-8
touch. See touch screen Pneumatic System in AC•T 8/10 and AC•T diff
Verify Predilute. See Verify Predilute screen Analyzers - Vacuum, 2.2-5
Voltages/Sensors. See Voltages/Sensors screen Pneumatic System in the AC•T diff 2 Analyzer -
sensors Pressure, 2.3-14
diluent level. See diluent level sensor Pneumatic System in the AC•T diff 2 Analyzer -
disconnecting for Analyzer card Vacuum, 2.3-12
removal/installations, 4.4-2 Vacuum Isolator Chamber (VC1) System in
horizontal traverse motor, 7.1-20 AC•T 8/10 and AC•T diff Analyzers, 2.2-7
horizontal. See horizontal sensors Vacuum Isolator Chamber (VC1) System in the
optical liquid. See FS1; FS2; liquid optical AC•T diff 2 Analyzer, 2.3-13
sensors Waste and Cleaner System in AC•T 8/10 and
opto. See opto sensors AC•T diff analyzers, 2.2-10
part numbers, 8.1-5 Waste and Cleaner System in the AC•T diff 2
Sensor Connector circuit, AC•T 8/10 Analyzer, 2.3-16
analyzer, 2.5-6 sleep state
syringe, 7.1-20 condition of Hgb lamp, 2.9-1
vertical traverse motor, 7.1-20 condition of pneumatic pump, 2.9-1
vertical. See vertical sensor definition, 2.9-1
vertical. See vertical sensors probe position, 2.9-1
waste. See waste sensor Sample Results screen determines probe
service bypass function, 7.2-3 position, 2.9-1
INDEX
piston seals, maintenance schedule, 5.1-1 traverse housing, AC•T diff 2 analyzer
system interconnect component in Traverse module, 2.3-1
engineering schematic, part number, 6.1-1 location, 2.3-2
system overview movement, 2.3-2
hardware, 2.1-1 removal, 4.9-1
instruments in the AC•T series, 2.1-1 removal, illustration, 4.9-6
interacting with the AC•T analyzer, 2.1-1 Traverse Interconnect card, AC•T 8/10 and AC•T
software, 2.1-1 diff analyzers
system specifics, table of, 2.1-3 component in Traverse module, 2.2-1
disconnecting to remove Traverse module, 4.8-1
location, illustration, 4.8-2, 8.2-18
T part number, 8.1-2, 8.2-19
Temp field removal, 4.8-1
on Voltages/Sensors screen, description, 7.1-4 Traverse Interconnect card, AC•T diff 2 analyzer
top and rear center panel, AC•T diff2 analyzer location, 2.3-2
installation, 4.2-5 location, illustration, 8.2-20
removal, 4.2-4 part number, 8.1-2, 8.2-21
removal/installation, illustration, 4.2-4 removal, 4.9-1
removal/installation, tools/supplies removal, illustration, 4.9-7
needed, 4.2-4 Traverse module
top cover and rear panel, AC•T 8/10 and AC•T diff function, AC•T 8/10 and AC•T diff
analyzers analyzers, A.4-4
installation, 4.2-2 function, AC•T diff 2 analyzer, A.5-5
removal, 4.2-1 location, illustration, AC•T 8/10 and AC•T diff
removal/installation, illustration, 4.2-1 analyzers, A.4-2
removal/installation, tools/supplies location, illustration, AC•T diff 2 analyzer, A.5-3
needed, 4.2-1 Traverse module, AC•T 8/10 and AC•T diff
touch screen analyzers
component in the Display assembly, 2.7-1 component part numbers, 8.2-19
control, AC•T diff analyzer, 2.6-2 components, 2.2-1
control, AC•T diff 2 analyzer, 2.6-2 connector on Analyzer card, A.2-2
description, 2.7-2 description, 2.2-1
input managed by Display Controller Driver disassembly/replacement verification, 4.8-5
card, 2.7-4 disassembly/replacement, illustration, 4.8-2
location, illustration, 8.2-6 disassembly/replacement, tools/supplies
location, illustration, AC•T 8/10 and AC•T diff needed, 4.8-1
analyzers, 2.1-2 disconnecting for Analyzer card
location, illustration, AC•T diff 2 analyzer, 2.1-2 removal/installation, 4.4-2
specifications, table of, 2.7-2 exploded view, 4.8-2, 8.2-18
See also Display assembly function, 2.2-1
Transient Suppressor card installation, 4.8-3
location, illustration, 8.2-16 removal, 4.8-1
part number, 8.1-2, 8.2-17 Traverse module, AC•T diff analyzers
traverse housing connector on Analyzer card, A.2-6
component in Traverse module, AC•T 8/10 and Traverse module, AC•T diff 2 analyzer
AC•T diff analyzers, 2.2-1 component part numbers, 8.2-21
traverse housing, AC•T 8/10 and AC•T diff components, illustration, 2.3-2
analyzers components, listing of, 2.3-1
location, illustration, 4.8-2 description, 2.3-1
removal, 4.8-1
INDEX
disassembly/replacement procedure V
verification, 4.9-14 Vac field
disassembly/replacement, procedures, 4.9-1 on Voltages/Sensors screen, description, 7.1-4
disassembly/replacement, tools/supplies vacuum
needed, 4.9-1 error, 7.2-3, 7.2-8
disconnecting flex cable, illustration, 4.9-2 minimum and maximum readings, AC•T 8/10
disconnecting for Analyzer card and AC•T diff analyzers, 7.1-4
removal/installation, 4.4-2 regulator, part number, 8.1-5
disconnecting horizontal motor cable, vacuum adjust
illustration, 4.9-2 making a low vacuum adjustment, AC•T diff 2
disconnecting tubing, illustration, 4.9-3 analyzer, 4.7-7
exploded view, 8.2-20 vacuum adjust knob
horizontal drive belt removal, 4.9-1 function, AC•T diff 2 analyzer, A.5-11
horizontal drive belt tension adjustment, 4.9-14 location, illustration, AC•T 8/10 and AC•T diff
horizontal motor removal, 4.9-1 analyzers, A.4-2
horizontal sensors removal, 4.9-1 location, illustration, AC•T diff 2
probe removal, 4.9-1 analyzer, A.5-6, A.5-10
removal, 4.9-1 making a low vacuum adjustment, AC•T diff 2
removal, illustration, 4.9-3 analyzer, A.5-11
traverse housing removal, 4.9-1 vacuum chamber
Traverse Interconnect card removal, 4.9-1 function, AC•T 8/10 and AC•T diff
vertical motor removal, 4.9-1 analyzers, A.4-3
vertical rack removal, 4.9-1 function, AC•T diff 2 analyzer, A.5-4
Traverse module, AC•T diff 2 analyzers location, illustration, AC•T 8/10 and AC•T diff
connector on Analyzer card, A.2-10, A.2-14 analyzers, A.4-2
Triple Syringe module location, illustration, AC•T diff 2 analyzer, A.5-3
exploded view, 8.2-1 See also VC1
function, 2.1-1 vacuum isolator chamber. See vacuum chamber;
function in lytic reagent delivery system, 2.2-4 VC1
See Syringe module vacuum port
troubleshooting function, AC•T diff 2 analyzer, A.5-12
AC•T 8/10 and AC•T diff analyzer specific location, illustration, AC•T diff 2
problems, 7.4-1 analyzer, A.5-10
clog detection problems, 7.3-1 vacuum pump. See pneumatic pump
DVM error (3), 7.2-4 vacuum regulator
error messages, 7.2-1 function, AC•T diff 2 analyzer, A.5-11
service bypass function for powerup location, illustration, AC•T diff 2
errors, 7.2-3 analyzer, A.5-10
using pulse test, AC•T 8/10 analyzer, 7.1-22 vacuum sensor
using pulse test, AC•T diff and AC•T diff 2 function, AC•T diff 2 analyzer, A.5-11
analyzers, 7.1-22 location, illustration, AC•T diff 2
using the dummy load test, 4.22-1 analyzer, A.5-6, A.5-10
using Voltages/Sensors screen, 7.1-2 See also vacuum transducer
verify predilute, 7.1-8 vacuum transducer
tubing function, AC•T 8/10 and AC•T diff
lists, part numbers, 6.1-1 analyzers, A.4-9
maintenance schedule, 5.1-1 location, illustration, AC•T 8/10 and AC•T diff
part numbers, 8.1-6 analyzers, A.4-5
See also vacuum sensor
INDEX
waste and cleaner system, AC•T 8/10 and AC•T diff function in aperture sensor system, AC•T 8/10
analyzers and AC•T diff analyzers, 2.2-8
components and flow, simplified function in aperture sensor system, AC•T diff 2
schematic, 2.2-10 analyzer, 2.3-15
description, 2.2-9 function in pneumatic system, AC•T 8/10 and
waste and cleaner system, AC•T diff 2 analyzer AC•T diff analyzers, 2.2-8
components and flow, simplified function in waste and cleaner system, AC•T 8/10
schematic, 2.3-16 and AC•T diff analyzers, 2.2-9
description, 2.3-16 function, AC•T diff 2 analyzer, A.5-4
Waste field probe position in Traverse module, AC•T 8/10
on Voltages/Sensors screen, description, 7.1-5 and AC•T diff analyzers, 2.2-1
waste pump See also aperture/bath assembly
component in waste and cleaner system, AC•T WBC bath
diff 2 analyzer, 2.3-16 function, AC•T 8/10 and AC•T diff
function, AC•T diff 2 analyzer, A.5-8 analyzers, A.4-3
location, illustration, AC•T 8/10 and AC•T diff function, AC•T diff 2 analyzer, A.5-4
analyzers, A.4-5 location, illustration, AC•T 8/10 and AC•T diff
location, illustration, AC•T diff 2 analyzer, A.5-6 analyzers, A.4-2
waste sensor location, illustration, AC•T diff 2 analyzer, A.5-3
conditions that turn ON its indicator, 7.1-5 WBC Measured field
digital sensor indicator, 7.1-5 on Sample Details screen, AC•T diff analyzer,
waste/rinse pump description, 7.1-15
function, AC•T 8/10 and AC•T diff WBC Ratio CV field
analyzers, A.4-7 on Sample Details screen, AC•T diff analyzer,
location, illustration, AC•T 8/10 and AC•T diff description, 7.1-16
analyzers, A.4-5 Whole-Blood Aspirate cycle. See Aspirate cycle
WAV WPWV AVG field
conditions that decrease this voltage on Sample Details screen, AC•T 8/10 analyzer,
reading, 7.1-3, 7.3-4 description, 7.1-11
conditions that increase this voltage WPWV LTVAR field
reading, 7.1-3, 7.3-4 on Sample Details screen, AC•T 8/10 analyzer,
typical readings, 7.1-3 description, 7.1-11
used in clog detection, 7.3-4 WPWV RM field
WAV field on Sample Details screen, AC•T 8/10 analyzer,
on Voltage/Sensors screen, description, 7.1-3 description, 7.1-11
WAV VAR field WPWV VAR field
on Sample Details screen, AC•T 8/10 analyzer, on Sample Details screen, AC•T 8/10 analyzer,
description, 7.1-10 description, 7.1-10
WBC
erratic results, troubleshooting, 7.4-1
high results, troubleshooting, 7.4-1
low results, troubleshooting, 7.4-1
processor circuitry, 2.5-1
voteout criteria, 7.1-12
WBC aperture
voltage. See WAV
WBC aperture/bath assembly
draining, AC•T 8/10 and AC•T diff
analyzers, 2.2-9
draining, AC•T diff 2 analyzer, 2.3-16
TRADEMARKS
The BECKMAN COULTER logo, 4C, AC•T, AC•T diff, AC•T diff 2, AC•T Pak, AC•T Rinse,
AC•T Tainer, AC•T Tron, COULTER, COULTER COUNTER, diff AC•T Pak, diff AC•T Tainer,
and MD II are trademarks of Beckman Coulter, Inc.
All other trademarks, service marks, products, or services are trademarks or registered
trademarks of their respective holders.