Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Installation Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 52

installation manual

2005-05/10010941_1/en
Table of Contents

1 INTRODUCTION ........................................................................................................ 1
2 PRE-INSTALLATION INFORMATION ....................................................................... 2
2.1 X-ray unit installation information ................................................................................................. 2
2.2 Installation dimensions ................................................................................................................ 4
2.3 X-ray unit main parts .................................................................................................................... 5
2.4 Calibration tools ........................................................................................................................... 6

3 INSTALLING THE X-RAY .......................................................................................... 7


3.1 Attaching the X-ray to a wall ........................................................................................................ 7
3.2 Installing the exposure switch .................................................................................................... 11
3.3 Planmeca Proline XC pan/ceph X-ray - Attaching the Cephalostat ........................................... 12

4 CHECKING THE ALIGNMENT ................................................................................ 13


4.1 Operating the X-ray ................................................................................................................... 13
4.2 Planmeca Proline XC panoramic X-ray - checking the alignment of the radiation beam .......... 17
4.3 Planmeca Proline XC pan/ceph X-ray - checking the alignment of the radiation beam ............ 21
4.4 Planmeca Proline XC pan/ceph X-ray - checking the position of the soft tissue filter ............... 31
4.5 Planmeca Proline XC pan/ceph X-ray - checking the alignment of the head support ............... 33
4.6 Checking the alignment of the patient positioning mechanism .................................................. 34
4.7 Checking the alignment of the patient positioning lights ............................................................ 38

5 SETTING CUSTOMER PREFERENCES ................................................................. 40


5.1 Setting the clock ........................................................................................................................ 40
5.2 Checking and changing the tone of the exposure warning signal ............................................. 42
5.3 Adjusting the starting speed of the vertical carriage .................................................................. 43

Installation manual TOC 1


Table of Contents

Planmeca pursues a policy of continual product development. Although every


effort is made to produce up-to-date product documentation this publication
should not be regarded as an infallible guide to current specifications. We re-
serve the right to make changes without prior notice.

COPYRIGHT PLANMECA 2005-05


PUBLICATION PART NUMBER 10010941 revision 1

TOC 2 Installation manual


INTRODUCTION

1 INTRODUCTION

Planmeca manufactures two versions of the Planmeca


Proline XC X-ray:
-the Planmeca Proline XC Panoramic X-ray
-the Planmeca Proline XC Pan/ Ceph X-ray.
This guide contains all the information required to set up
and calibrate both versions of the Planmeca Proline XC
X-ray.
NOTE You may also need the Planmeca Proline XC X-
ray unit’s technical manual (publication number
10010942) during installation. The alignment in-
structions are given in technical manual.
Protect yourself from radiation when you are checking
the radiation beam alignment.
WARNING FAILURE TO INSTALL THE X-RAY IN AN AP-
PROVED LOCATION MAY BE DANGEROUS TO
BOTH PATIENT AND OPERATOR.

CAUTION IT IS VERY IMPORTANT THAT THE ROOM IN


WHICH THE X-RAY IS INSTALLED AND THE PO-
SITION FROM WHICH THE USER OPERATES
THE EQUIPMENT ARE CORRECTLY SHIELDED
FROM RADIATION. SINCE RADIATION SAFETY
REQUIREMENTS VARY FROM COUNTRY TO
COUNTRY AND STATE TO STATE IT IS THE RE-
SPONSIBILITY OF THE INSTALLER TO ENSURE
THAT ALL LOCAL SAFETY REGULATIONS ARE
MET.

CAUTION The exposure switch Autoprint and Admark film


marking systems are connected to the X-ray
units Planet connectors. Do NOT connect any
other devices to the Planet connectors.

The display values shown in this guide are only examples


and should not be interpreted as recommended values
unless otherwise stated.

The manufacturer, assembler, and importer are responsible for the safety, re-
liability and performance of the unit only if:

- installation, calibration, modification and repairs are carried out by qualified


authorized personnel

- electrical installations are carried out according to the appropriate require-


ments such as IEC364

- equipment is used according to the operating instructions

Installation manual Planmeca Proline XC X-ray unit 1


PRE-INSTALLATION INFORMATION

2 PRE-INSTALLATION INFORMATION

WARNING Do not connect items which are not specified as


part of the system.

NOTE The connection of additional equipment to a mul-


tiple portable socket-outlet must only be possi-
ble by using a tool or be supplied via separating
transformer.

NOTE The multiple portable socket-outlets shall not be


placed on the floor.

When using a multiple portable socket-outlet, do


not connect any other equipment to this than the
ones specified as part of the system.

In cases where additional equipment, in same or


especially in separate locations, are connected
there is always a danger of too high leakage cur-
rents in the system.

2.1 X-ray unit installation information

1. The room in which the X-ray is to be installed and the


position from where the user is to operate the equipment
MUST be correctly shielded from radiation. Since radia-
tion safety requirements vary from country to country and
state to state it is the responsibility of the installer to
ensure that all safety regulations are met.
2. Make sure that there is enough room to install the X-
ray correctly and that the wall will support the X-ray. The
bracket that secures the X-ray to the wall must be able to
resist a pull-out force of 160 kg (353 pounds). NOTE. If
the wall will not support the X-ray the X-ray must be
mounted on the free-standing base, part number 50021.
3. Make sure that the power requirements of the X-ray
are the same as the power that is to be used. The X-ray
is designed to operate using either 100-132, or 180-240
V~ dedicated power supply with a tolerance of ±10%.
The X-ray power requirements are printed on a label
which is attached to the mains cable and also printed on
a label located at the base of the support column. If the
voltage is likely to fluctuate by more than ±10% you will
need to install a power supply regulator to ensure that the
unit operates correctly.
4. Make sure that there is an earthed power outlet within
2.5 meters (8 feet) of where the X-ray is to be installed.
5. Make sure that you have access to an X-ray film pro-
cessor that is in working order and is ready for use.

2 Planmeca Proline XC X-ray unit Installation manual


PRE-INSTALLATION INFORMATION

6. Make sure that you have the correct chemicals for pro-
cessing the X-ray film being used. Make sure that the
solutions are fresh.

7. Make sure that the darkroom is fitted with a safe light


that is suitable for the X-ray film being used.

8. Make sure that the X-ray film will fit the cassette and
that it can be used with the intensifying screen being
used.

Installation manual Planmeca Proline XC X-ray unit 3


PRE-INSTALLATION INFORMATION

2.2 Installation dimensions


W(with ceph.) 2230 (88")
1140 (45")

W(with autoprint) 1630 (64")


540 (21") 1090 (43")

W(standard Pan) 1530 (60")


440 (17") 730 (29") 360 (14")
620 (24 1/2")

D 1475 (58") (With transversal


930 (36 1/2")

D 1035 (40 1/2")

slicing system)
830 (32 1/2")
2050 - 2150 (80 1/2 - 84 1/2")

2200 (86 1/2")

NOTE: If the unit is to be used


with the free-standing base
(part number 50021) the total
height will be 2230mm (88in)

4 Planmeca Proline XC X-ray unit Installation manual


PRE-INSTALLATION INFORMATION

2.3 X-ray unit main parts

Tube head assembly


Column Cassette carriage Rotating unit
assembly

Vertical carriage

Cephalostat
(optional)

Lower shelf

Admark film-marking
system (optional)
XC_Film_main.eps

Graphic user interface

Autoprint film-marking
system (optional)

E
FIC
OF

+ M
=

- M
, Ü TE
R

) EN
0 Å
(
Ä
9 P
´
Ö ?
8 O /
L >.
&
7 I
K <,

Exposure switch
^
6 U
%
J
5 Y M
$
H
4 T N
#
G
3 R B
"
F
2 E V
!
D
1 W C
S
EA
R
Q X
CL A
DE
L
:
Z
;
IFT
SH

Installation manual Planmeca Proline XC X-ray unit 5


PRE-INSTALLATION INFORMATION

2.4 Calibration tools

You will need the following PLANMECA calibration tools


to check the alignment of the X-ray:

Beam alignment tool


Alignment ruler (Part number 50972)
(Part number 50973)

Used to check the position of the Used to check the position of


patient positioning mechanism the panoramic X-ray beam

Frankfort plane alignment tool


Ball phantom
(Part number 50977)
D
EN
1
97

(Part number 50971)


.50
No

D
CA EN
ME 97
1
AN .50
PL
No

E T
IN AR
Y-L IN
E ST
X-L 7 CA
97 ME
.50 AN
NO PL

OY T

E
IN
AR
CA

-L
E ST
ME

Y
IN
X-L
AN
PL

Used to check the position of the patient


positioning mechanism and the positioning lights Used with the ball phantom to check
the position of the Frankfort plane light

Cephalometric beam alignment tool


(Part number 652232, 18 x 24cm
Ceph head support 652233, 8 x 10 in)
alignment tool Soft tissue filter
(Part number 653126) positioning tool
(Part number 653121)

5
4
Positioning light 25mm
Positioning light 0mm
Soft tissue filter 0mm

2
23
52
.6
No

24
8x
t1
6

sta
alo
ph
Ce
ne
roli
0 2P
20
PM
CA
ME
AN
PL

Used to check the position Used to check the position of


Used to check the position of the soft tissue filter the cephalometric X-ray beam
of the head support (Only Pan/ceph x-ray unit) (Only Pan/ceph x-ray unit)
(Only Pan/ceph x-ray unit)

6 Planmeca Proline XC X-ray unit Installation manual


INSTALLING THE X-RAY

3 INSTALLING THE X-RAY

3.1 Attaching the X-ray to a wall

Unpack the unit but DO NOT REMOVE THE PROTEC-


TIVE STYROFOAM PACKAGING UNTIL THE UNIT
HAS BEEN ATTACHED TO THE WALL AND LEV-
ELLED. If you have the carry handles (PLANMECA part
number 50985) attach them to the support column.
Transport the unit to the installation position.

Styrofoam packaging

X-ray unit

Carry handles

Remove the carrying handles and push the base plate


into the base of the column. You may need to tap it in
Base plate with a hammer.

Base of column

Use the wall plate to mark the position of the wall bracket
Wall plate that secures the X-ray to the wall. Either two or four
screws can be used to secure the bracket to the wall. If
the wall is thin you may have to use a backing plate and
screws and nuts (not supplied).
The bracket should be positioned approximately 2090mm
(82in) from the floor. Use a spirit level to ensure that the
wall plate is level. Decide what type and size of screws
you plan to use and then drill securing holes of the appro-
priate size.

2090mm (82in)
from the floor

Installation manual Planmeca Proline XC X-ray unit 7


INSTALLING THE X-RAY

Wrap tape around the top of the X-ray column approxi-


Top of column mately one foot down from the top. This will prevent the
wall bracket from sliding down the column when you
stand the X-ray up. From the top of the column, slide the
clamp plate part of the wall bracket into the grooves at
the back of the column.

Clamp plate

Wall bracket

Lift the X-ray into position against the wall and secure the
Top of column wall bracket to the previously drilled holes with suitable
screws.

Wall bracket

Screws

With a spirit level make sure that the column is vertical. If


Wall bracket you have to make any adjustments move the base of the
column.
Column

Spirit level

8 Planmeca Proline XC X-ray unit Installation manual


INSTALLING THE X-RAY

NOTE If you are installing a Pan/ceph unit you must se-


cure the X-ray column to the floor with the lower
bracket.
To slide the lower bracket into the groove at the
rear of the column, you will have to turn the
bracket sideways slightly to get it into the slots,
and then use the bracket to mark the positions of
Screws
the fixing holes. Drill two 10mm (3/8in) fixing
holes and then secure the lower bracket to the
floor with the screws and plastic inserts.

Lower bracket Bottom of


column
Inserts

When the unit is vertical tighten the column clamping


Top of column
screws that hold the x-ray column to the wall bracket and,
Clamping screws if the lower bracket is used, tighten the column clamping
screws in the lower bracket.

Remove the protective styrofoam packaging and the tape


that you previously wrapped round the top of the column
to stop the wall bracket from sliding down.

Wall bracket

Clamping screws

Bottom of
column
Lower bracket

Installation manual Planmeca Proline XC X-ray unit 9


INSTALLING THE X-RAY

Remove the transportation pin, clamping screw and


packing piece that lock the rotating unit to the vertical car-
riage.
Also remove the counterweight locking screw and clamp
plate from the rear of the column and the two transporta-
tion pins that hold the vertical carriage in position during
shipping. These should be kept with the X-ray in case the
X-ray needs to be moved at some future date.
The long transportation pin that holds the rotating unit in
position is used to check that the unit is correctly aligned.

Clamping screw
Transportation pin

XC_Shipping_pins.eps

Vertical carriage

Transportation pins

Rotating unit

Column

Counterweight
locking screw and
clamp plate

10 Planmeca Proline XC X-ray unit Installation manual


INSTALLING THE X-RAY

3.2 Installing the exposure switch

The exposure switch must be located at least two meters


(7 feet) from the X-ray in a position that satisfies the local
radiation safety regulations regarding to operator safety
and the shielding of X-ray equipment.

Exposure switch holder Remove the cover from the exposure switch holder, note
that there is a ground lead connected to the cover, and
use the body as a template to mark and drill the two

XC_Ins_exp_sw.eps
securing holes. Secure the holder in position and replace
the cover.

Connect the female plug at one end of the exposure


switch cable to the socket on the underside of the expo-
Body sure switch holder.

Cover
Female
plug

Exposure
switch
cable

Connect the male plug at the other end of the exposure


switch cable to one of the sockets at the bottom of the X-
Exposure ray column. It does not matter which socket the cable is
switch connected to.
Bottom of
cable
column
Male
plug

Installation manual Planmeca Proline XC X-ray unit 11


INSTALLING THE X-RAY

3.3 Planmeca Proline XC pan/ceph X-ray - Attaching the Cephalostat

If you are installing a Planmeca Proline XC PAN/CEPH


unit you must attach the cephalostat to the vertical car-
riage.

Lift the cephalostat and carefully guide the cephalostat


arm round the back of the X-ray so that the adaptor is
behind the vertical carriage.
Position the adapter so that it straddles the vertical car-
riage.

Adapter

Cephalostat arm

Cephalostat XC_inst_ceph_CM.eps

Vertical carriage

Secure the cephalostat to the vertical carriage with the


four screws supplied.

12 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

4 CHECKING THE ALIGNMENT

4.1 Operating the X-ray

Switching the unit on


Connect the unit to a dedicated power supply with a volt-
age appropriate to the unit you are installing.
_o

Switch the unit on. The on/off switch is located on the


underside of the vertical carriage. When you switch the
unit on the main display will be shown on the graphic
user interface and the unit will carry out a self-test which
will last a few seconds.
The unit is then ready for use.
The same switch is used to switch the X-ray off.

NOTE The unit incorporates a self-checking feature


that monitors the operation of the unit. If there is
a malfunction or an operating error the unit will
stop working and an error message will appear
on the GUI. The error message must be cleared
from the display before the unit can be used
again. To clear the message, touch the OK field.
Refer to the technical manual for a list of error
messages and what they mean.
NOTE A help message is shown on the GUI if the select-
ed function is not allowed. The help message
disappears when you cancel the action or re-
lease the exposure button.

Installation manual Planmeca Proline XC X-ray unit 13


CHECKING THE ALIGNMENT

Test mode
The test mode enables the X-ray to be operated without
radiation being generated.
Touch the i-field on the Main display. The list of User set-
tings appears.

From the list of User settings select Behavioural prefer-


ences. The display shown below appears.

Touch the Test mode field to enter the test mode (ON).
Exit the test mode by touching the Test mode field again
(OFF).

14 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

Service mode
Various settings and adjustments can only be made in
the service settings mode.
CAUTION WHEN YOU ARE IN THE SERVICE MODE IT IS POSSI-
BLE TO CHANGE SOME OF THE PRESET FACTORY
SETTINGS. DO NOT PLAY WITH THE KEYS UNLESS
YOU KNOW WHAT YOU ARE DOING.

Touch the i-field on the Main display. The list of User


preference settings appears.

By touching the arrow down field below the list you can
see more features. From the list of User settings select
Service settings (password).

Installation manual Planmeca Proline XC X-ray unit 15


CHECKING THE ALIGNMENT

The service settings mode is password protected and the


password is asked when the Service settings mode is
entered for the first time after switching the unit on. The
password is 1701.

Exit the service mode by touching the OK field on the


respective service mode display and on the Service set-
tings display.

16 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

4.2 Planmeca Proline XC panoramic X-ray - checking the alignment of the radia-
tion beam

Remove the inner cover from tube head assembly. Four


screws hold the tube head cover in position, these are
located at the rear of the tube head assembly. The pri-
mary collimator mechanism is located under the tube
head cover.

Cassette drive
assembly Tube head
assembly

Tube head cover

Place the beam alignment tool into the secondary colli-


mator, it is magnetic so it will stay in position. Leave the
cassette in position as it will reduce the amount of scat-
tered radiation.

Beam
alignment
tool
Touch the ready field to drive the X-ray into the exposure
position. The rotating unit will move to ready position.
Collimator The rotating unit and cassette carriage will move to their
respective ready positions. On the exposure switch
holder a ready indicator light will come on and on the
main display the Ready field will be green.

Installation manual Planmeca Proline XC X-ray unit 17


CHECKING THE ALIGNMENT

Enter the test mode (See section “Test mode” on page


14) to switch the radiation off.
Press and hold down the exposure button to drive the
rotating unit to a position where you can safely view the
secondary collimator when standing behind the tube
head. Stop pressing the exposure button when the rotat-
ing unit is in a convenient position.

An error code will appear on the display. Clear the error


code from the display by touching the OK field.
Exit the Test mode.

On the main display, touch the kV/mA field.

The Select kV/mA display appears. Select a kilovolt and a


milliampere value high enough to enable the radiation beam
to be seen in a darkened room, for example 62kV and 6mA.
The actual values will depend on how dark the room is.

18 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

Enter the service mode (See section “Service mode” on


page 15). By touching the arrow down field below the list
you can see more features. From the list of Service set-
tings select Primary collimator.

The display shown below appears.

Darken the room sufficiently so that you will be able to


see the image of the radiation beam on the alignment tool
(it is fluorescent and will glow when the radiation beam
strikes it), but not so dark that you cannot see the borders
of the alignment rectangle.

Stand behind the tube head and protect yourself from


radiation. Press and hold down the exposure button.
The image of the radiation beam will appear on the align-
ment tool.

Installation manual Planmeca Proline XC X-ray unit 19


CHECKING THE ALIGNMENT

The radiation beam must appear within the borders of the


rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC panoramic X-ray tech-
nical manual for information on how to align it.

UP

no. 50972
Radiation beam

PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN


NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT

DOWN

When the collimator has been checked, and if necessary


adjusted, exit the service mode, see section “Service
mode” on page 15.

You must now take an exposure to check that the patient


positioning mechanism is correctly aligned, turn to sec-
tion 4.6 “Checking the alignment of the patient positioning
mechanism” on page 34.

20 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

4.3 Planmeca Proline XC pan/ceph X-ray - checking the alignment of the radiation
beam

Panoramic X-ray beam


Remove the inner covers from tube head assembly and
the cassette drive assembly. Four screws hold the tube
head cover in position, these are located at the rear of
the tube head assembly. The screws that hold the cas-
sette drive cover in place are located at the front and
lower side of the cover. The primary collimator mecha-
nism is located under the tube head cover.

Cassette drive
assembly
Tube head
assembly

Cover Tube head cover

Place the beam alignment tool into the secondary colli-


mator, it is magnetic so it will stay in position. Note which
way round it goes. Leave the cassette in position as it will
reduce the amount of scattered radiation.
Beam
alignment
tool

Secondary collimator Primary


collimator

Installation manual Planmeca Proline XC X-ray unit 21


CHECKING THE ALIGNMENT

Touch the ready field to drive the X-ray into the exposure
position. The rotating unit will move to ready position.
The rotating unit and cassette carriage will move to their
respective ready positions. On the exposure switch
holder a ready indicator light will come on and the Ready
field will be green.

Enter the test mode (See section “Test mode” on page


14) to switch the radiation off.
Press and hold down the exposure button to drive the
rotating unit to a position where you can safely view the
secondary collimator when standing behind the tube
head. Stop pressing the exposure button when the rotat-
ing unit is in a convenient position.

An error code will appear on the display. Clear the error


code from the display by touching the OK field.
Exit the Test mode.

On the main display, touch the kV/mA field.

22 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

The Select kV/mA display appears. Select a kilovolt and a


milliampere value high enough to enable the radiation beam
to be seen in a darkened room, for example 62kV and 6mA.
The actual values will depend on how dark the room is.

Enter the service mode (See section “Service mode” on


page 15). By touching the arrow down field below the list
you can see more features. From the list of Service set-
tings select Primary collimator.

Installation manual Planmeca Proline XC X-ray unit 23


CHECKING THE ALIGNMENT

The display shown below appears. The number above


the arrow fields indicates the collimator number and the
collimator is changed with the Next field. Select collimator
number 0.

Darken the room sufficiently so that you will be able to


see the image of the radiation beam on the alignment tool
(it is fluorescent and will glow when the radiation beam
strikes it), but not so dark that you cannot see the borders
of the alignment rectangle.
Stand behind the tube head and protect yourself from
radiation. Press and hold down the exposure button. The
image of the radiation beam from the panoramic collima-
tor will appear on the alignment tool.

The radiation beam must appear within the borders of the


rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC X-ray unit technical
manual for information on how to align it.

UP
no. 50972

Radiation beam

Panoramic collimator
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT

DOWN

24 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

After the collimator number 0 has been checked, and if


necessary adjusted, select the collimator number 1 with
the arrow fields.
Stand behind the tube head and protect yourself from
radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 1 on
the alignment tool.

The radiation beam from collimator number 1 must


appear within the borders of the rectangle marked on the
alignment tool and fill the lower 80 - 85% of the rectangle.
If the beam is not aligned correctly refer to the x-ray unit’s
technical manual.
Exit the service mode, see “Service mode” on page 15.

Cephalometric X-ray beam


Select the Cephalostat program as follows. Touch the
Prog. field on the main display.

Installation manual Planmeca Proline XC X-ray unit 25


CHECKING THE ALIGNMENT

The Select program type appears. Select Ceph. The


rotating unit will move automatically to the correct posi-
tion for taking cephalometric exposures.

Orientate the cephalometric beam alignment tool so that


Cassette it is vertical and then place it in the cephalometric cas-
holder sette holder. Position it so that it is on the left of the cas-
sette holder.

4 5

Beam
alignment
tool
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm


6

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232

Select the collimator number 4.

26 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

Select exposure values of approximately 62kV, a milliam-


pere value of 6 and an exposure time of about one sec-
ond. The actual values will depend on how dark the room
is.
Stand behind the tube head and protect yourself from
radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 4 will
appear on the alignment tool.

The radiation beam must appear within the borders of the


rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC Pan/Ceph X-ray unit
technical manual for information on how to align it.
Radiation beam

4 5
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm


6

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232 collimator 4

NOTE The soft tissue filter will filter a narrow area of the
left side of the radiation beam.

Cassette Slide the beam alignment tool to the right of the cassette
holder
holder.
6

4 5

Beam
alignment
tool
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232

Installation manual Planmeca Proline XC X-ray unit 27


CHECKING THE ALIGNMENT

Select the collimator number 5.

Stand behind the tube head and protect yourself from


radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 5 will
appear on the alignment tool.

The radiation beam must appear within the borders of the


rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC Pan/Ceph X-ray unit
technical manual for information on how to align it.

Radiation beam
6

4 5
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232


collimator 5

28 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

Turn the beam alignment tool so that it is horizontal.


Cassette
holder

No. 652232
Beam 6

alignment
tool

PLANMECA PM 2002 Proline Cephalostat 18x24


Positioning light 25mm

Positioning light 0mm

Soft tissue filter 0mm


5
4

Select the collimator number 6.

Stand behind the tube head and protect yourself from


radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 6 will
appear on the alignment tool.

Installation manual Planmeca Proline XC X-ray unit 29


CHECKING THE ALIGNMENT

The radiation beam must appear within the borders of the


rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC Pan/Ceph X-ray unit
technical manual for information on how to align it.

Radiation beam

No. 652232
6

PLANMECA PM 2002 Proline Cephalostat 18x24


Positioning light 25mm

Positioning light 0mm

Soft tissue filter 0mm


5
4

collimator 6

NOTE The soft tissue filter will filter a narrow area of the
left side of the radiation beam.
When all the collimators have been checked, and if nec-
essary adjusted, exit the service mode, see section “Ser-
vice mode” on page 15.
You must now check the position of the soft tissue filter.
Turn to section 4.4 “Planmeca Proline XC pan/ceph X-ray
- checking the position of the soft tissue filter” on page
31.

30 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

4.4 Planmeca Proline XC pan/ceph X-ray - checking the position of the soft tissue
filter

Clip the soft tissue filter positioning tool to the soft tissue
filter which is at the end of the cassette carriage. This will
enable you to see the soft tissue filter when you check its
Soft tissue filter position.
positioning tool

Cassette carriage

Soft tissue filter


Drive the soft tissue filter to the zero position by touching
the soft tissue filter arrow keys. Select the collimator
number 4

Orientate the Cephalometric beam alignment tool so that


Cassette it is vertical and then place it in the cephalometric cas-
holder sette holder. Position it so that it is on the left of the cas-
sette holder.

4 5

Beam
alignment
tool
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm


6

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232

Installation manual Planmeca Proline XC X-ray unit 31


CHECKING THE ALIGNMENT

Select exposure values of approximately 62kV, a milliam-


pere value of 6 and an exposure time of about one sec-
ond. The actual values will depend on how dark the room
is.

Stand behind the tube head and protect yourself from


radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 4 will
appear on the alignment tool.

The radiation beam must appear within borders of the


rectangle marked on the beam alignment tool and to the
right of the line marked “Soft tissue filter 0mm”.

4 5 Radiation beam If the left-hand side of the radiation beam does not reach
the line marked “Soft tissue filter 0mm” or extends
passed the line the soft tissue filter is in the wrong posi-
tion it must be repositioned. Refer to the Planmeca Pro-
Positioning light 25mm
Soft tissue filter 0mm

Positioning light 0mm

line XC Pan/Ceph X-ray unit technical manual for


information on how to carry out this task.
6

PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232

“Soft tissue filter 0mm” line


Area of soft tissue filter

32 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

4.5 Planmeca Proline XC pan/ceph X-ray - checking the alignment of the head
support

Rotate the head support to the 90° position and place the
two ear posts into their holders if they are not already
installed.
Embedded in the end of the left (V or L) ear post there is
a metal ball and embedded in the end of the right ear
post (O or R) is a metal ring. The images of these that
appear on an exposed film are used to check the align-
ment of the ear posts.

Place a film in the cephalometric cassette and slide the


Head support cassette into the cassette holder.

Cephalometric cassette

Select exposure values of approximately 70, a milliam-


pere value of 10 and an exposure time of about half a
second.

Protect yourself from radiation and take an exposure.


Remove the film and process the film

On the film you will see the images of a metal ball and a
metal ring. They must be concentric, of if not exactly con-
centric there must be a small gap between the edge of
the ball and the inner edge of the circle.

Ball

Ring
Correct Correct Wrong

If the ball touches the ring the head support must be


adjusted. Refer to the PM 2002 XC Pan/Ceph X-ray unit
technical manual.

Installation manual Planmeca Proline XC X-ray unit 33


CHECKING THE ALIGNMENT

4.6 Checking the alignment of the patient positioning mechanism

NOTE If the PLANMECA AUTOPRINT film marking sys-


tem is to be used with the X-ray you can check
the operation of the AUTOPRINT at the same time
as you check the position of the patient position-
ing mechanism. To do this enter suitable infor-
mation into the AUTOPRINT system using the
AUTOPRINT keyboard and then check that this
information appears correctly on the ball phan-
tom exposure. Refer to the user’s manual sup-
plied with the AUTOPRINT for information on
how to use the system.

Slide the ball phantom into the patient positioning mecha-


nism.
Ball phantom

Set the position of the patient positioning mechanism to 0


with the Layer arrow fields.

Touch the Entry field to move the cassette carriage to the


loading position.

34 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

In the darkroom open the cassette and place a film inside


the cassette and then place a piece of paper the same
Paper size on top, the paper will reduce the radiation that
reaches the film. Close the cassette.
Film

Cassette

Slide the cassette into the cassette holder. Make sure


that you push it all the way in.

The exposure values 60 kV and 2 mA must be selected


to produce a clear ball phantom exposure and it is only
possible to select these values in the Test exposure
mode. Enter the service settings mode. From the list of
service settings select Test exposure.

The display shown below appears. Touch the kV/mA


field.

Installation manual Planmeca Proline XC X-ray unit 35


CHECKING THE ALIGNMENT

Select a kilovolt value of 60 and a milliampere value of 2.


Touch the OK field.

On the Test exposure display touch the OK field. Touch


the ready field to drive the X-ray into the exposure posi-
tion. The rotating unit will move to ready position.

Take an exposure. Hold the exposure button down for


the duration of the exposure.

36 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

Touch the Entry field to return the cassette to the loading


position, remove the cassette and process the film.

120 ± 2mm 120 ± 2mm The images of 23 balls should appear on the film. They
should be round, all the same size, and evenly spaced.
Note that they will not necessarily be at the same vertical
height. The distance from the center of the middle ball to
the center of the tenth ball to the left and the tenth ball to
the right should be 120mm ± 2mm. If this distance is not
correct the patient positioning mechanism must be
adjusted. Refer to PM 2002 XC Panoramic X-ray unit
technical manual for information on how to do this.

Installation manual Planmeca Proline XC X-ray unit 37


CHECKING THE ALIGNMENT

4.7 Checking the alignment of the patient positioning lights

Place the Frankfort plane alignment tool into position on


Frankfort plane top of the ball phantom.
alignment tool EN
D

No
.50
97
1

7 CA
97 ME
.50 AN
NO PL

Y
AO
T

E
IN
AR
C

-L
E ST
ME

Y
IN
X-L
AN
PL

Touch either of the Layer arrow fields to switch the posi-


tioning lights on and then a second time to drive the
patient positioning mechanism to 0 if it is not already
there.

38 Planmeca Proline XC X-ray unit Installation manual


CHECKING THE ALIGNMENT

The Frankfort plane light beam must be located between


the two black lines on the alignment tool. It must be hori-
zontal.

D
N
E

1
97
Frankfort plane light beam

50
o.
N

A
7

C
97

E
.50

M
N
LA
NO

P
OY

T
E

R
IN
CA

A
E
-L

T
ME

IN

S
Y

-L
AN

X
PL

Midsagittal light beam


Focal layer light beam

The focal layer light beam must be positioned so that it is


on the black reference line on the side of the ball phan-
tom.
The midsagittal plane light beam must strike the black
line on the front of the ball phantom.
If any of the light beams do not coincide with the black
lines or they are crooked or not in focus, they must be
adjusted. Refer to the technical manual.

Installation manual Planmeca Proline XC X-ray unit 39


SETTING CUSTOMER PREFERENCES

5 SETTING CUSTOMER PREFERENCES

5.1 Setting the clock

If the time that appears on the main display is not correct


it must be reset.
Touch the i-field on the Main display. The list of User set-
tings appears.

From the list of User settings select Time&date adjust-


ment. Press either of the arrow fields to set the correct
minutes (0 - 9). The up field will increase the number and
the down field will decrease it.
Each time you press the Next field different clock digits
will start to flash:
- firstly minutes
- secondly hours

40 Planmeca Proline XC X-ray unit Installation manual


SETTING CUSTOMER PREFERENCES

- thirdly days
- fourthly months

- and finally years.

Press either of the arrow fields to change any of the time


settings that are incorrect.
Touch the OK arrow field to exit the clock setting mode
when you have finished setting the time.

Installation manual Planmeca Proline XC X-ray unit 41


SETTING CUSTOMER PREFERENCES

5.2 Checking and changing the tone of the exposure warning signal

When an X-ray exposure is taken you will hear a warning


tone indicating that radiation is being generated. The
pitch of the tone can be altered if required.
Touch the i-field on the Main display. The list of User set-
tings appears.

From the list of User settings select Sound tone. The


display shown below appears.

Touch either of the arrow fields to increase or decrease


the pitch of the tone if required.
Touch the OK field to accept the new or existing tone and
exit the mode.

42 Planmeca Proline XC X-ray unit Installation manual


SETTING CUSTOMER PREFERENCES

5.3 Adjusting the starting speed of the vertical carriage

The user may wish to change the speed at which the ver-
tical carriage starts to move up and down.
Enter the Service Setting mode as describe in section
“Service mode” on page 15.

Upward movement
Select the Motor up adjustment from the Service settings
list.

The display shown below appears.

The current upward starting speed of the vertical carriage


as a percentage of the maximum speed is shown on the
display.

Installation manual Planmeca Proline XC X-ray unit 43


SETTING CUSTOMER PREFERENCES

Touch the right and left arrows to increase or decrease


the speed. The new value will appear on the display.
Touch the “up” arrow field to check the speed. The verti-
cal carriage will move up at the new speed. You can drive
the vertical carriage downwards by touching the “down”
arrow field.
Touch the OK field to accept the new speed.

Downward movement
Select the Motor down adjustment from the Service set-
tings list.

The display shown below appears.

44 Planmeca Proline XC X-ray unit Installation manual


SETTING CUSTOMER PREFERENCES

Touch the right and left arrows to increase or decrease


the speed. The new value will appear on the display.
Touch the “down” arrow field to check the speed. The
vertical carriage will move up at the new speed. You can
drive the vertical carriage downwards by touching the
“up” arrow field.
Touch the OK field to accept the new speed.
Exit the service mode.

Installation manual Planmeca Proline XC X-ray unit 45


www.planmeca.com
PLANMECA OY Asentajankatu 6, 00880 Helsinki, Finland, tel. +358 20 7795 500
fax +358 20 7795 555, e-mail: sales@planmeca.com, www.planmeca.com

You might also like