Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Fona XDG: Service and Installation Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 32

FONA XDG

Service and Installation Manual

English
FONA XDG – Service & Installation Manual

FONA XDG Service and Installation Manual


English Edition

Version 150715 July 2015


Code 69 550 70210

Manufactured by FONA S.r.l.


Via Galilei 11 - 20090 Assago (MI) Italy

Distributed by FONA Dental s.r.o.


Stefanikova 7 SK-811 06 Bratislava, Slovakia www.fonadental.com

Table of Contents

4.1 MAINTENANCE OF WALL ADAPTOR ................... 20


1. INTRODUCTION................................................ 4 4.2 MAINTENANCE OF SUPPORT ARM..................... 20
1.1 PURPOSE........................................................ 4 4.3 MAINTENANCE OF SCISSOR ARM ...................... 21
1.2 EQUIPMENT CLASSIFICATION .............................. 4 4.4 MAINTENANCE OF X-RAY HEAD........................ 22
1.3 APPLICABLE STANDARDS ................................... 4 4.5 MAINTENANCE OF BEAM LIMITING DEVICE......... 22
1.4 ENVIRONMENTAL DATA .................................... 4 4.6 MAINTENANCE OF TIMER ................................ 22
1.5 OBLIGATIONS OF THE INSTALLER ......................... 4 4.7 CHECKING FILAMENT PRE-HEATING TIME .......... 22
1.6 WARNING ...................................................... 4 4.8 SERVICE FUNCTIONS....................................... 23
1.7 DEMONSTRATION ............................................ 4 5. MEASUREMENTS ............................................ 24
2. TECHNICAL DATA ............................................. 5
5.1 LINE VOLTAGE .............................................. 24
2.1 SYSTEM SUPPLY ............................................... 5 5.2 ANODE VOLTAGE – KVP ................................. 24
2.2 X-RAY HEAD ASSEMBLIES .................................. 5 5.3 ANODE CURRENT – MA .................................. 24
2.3 BEAM LIMITING DEVICE .................................... 5 5.4 EXPOSURE TIME ............................................ 24
2.4 TIPSET TIMER ................................................. 5 5.5 LEAKAGE RADIATION ...................................... 25
2.5 MECHANICAL SUSPENSION SYSTEM ..................... 5 5.6 EARTH RESISTANCE- ....................................... 25
2.6 WEIGHTS ....................................................... 5 5.7 EARTH LEAKAGE ............................................ 25
3. ASSEMBLY AND INSTALLATION ........................ 6 APPENDIX A SYSTEMS AND COMPONENTS ............ 26
3.1 WALL MOUNTED SYSTEMS ................................ 6 APPENDIX B ICONS................................................. 27
3.2 DIP SWITCH SETTING ...................................... 16
APPENDIX C EXPOSURE TABLE ............................... 28
3.3 LAYOUT POWER BOARD TIMER......................... 17
3.4 MOBILE SYSTEMS .......................................... 18 APPENDIX D ALARM CONDITIONS ......................... 29
3.5 MOUNTING THE TIMER ................................... 19
APPENDIX E IDENTIFICATION LABELS ..................... 30
4. MAINTENANCE............................................... 20
APPENDIX F COOLING CURVES ............................... 31

2/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

The Control Panel


2 3 4
1
1
5
6 7

18

10 11 12
8

16

13 14 15

9 17

Device for emission of ionizing


1 10 Maxillary incisor
radiation on request
Indication of system turned on and
2 11 Maxillary canine or premolar
ready
3 Irradiation 12 Maxillary molar

4 Alarm 13 Mandibular incisor


mAs display, the controlled
5 14 Mandibular canine or premolar
technique factor
Manual decrease of controlled
6 15 Mandibular molar
technique factor
Manual increase of controlled
7 16 Bite-wing premolar
technique factor
8 Patient size adult/large 17 Digital detector in use

9 Patient size child/small 18 Radiation exposure pushbutton

EXPOSURE INDEX
0.32 0.4 0.5 0.63 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0
CDR XIOS F E D
SENSOR SENSOR FILM FILM FILM
CDR Sensor by Shick Technologies Inc., XIOS Sensor by Sirona Dental Systems GmbH

150715 69 550 70210 3/32


FONA XDG – Service & Installation Manual

1. INTRODUCTION
1.1 Purpose
The FONA XDG X-ray Equipment is designed to fulfil the needs for high resolution intra-oral radiography in
the general dental practice. The systems can be configured for wall, mobile solutions.
The Operator’s Manual and the Service and Installation Manuals supplied with the system are integral part
of the product. The original language of the Operator’s Manual is English.
1.2 Equipment Classification
 IEC: FONA XDG is a Class I, type B equipment
 FDA: FONA XDG is a Class II medical device equipment (21 CFR 872-1800).
1.3 Applicable Standards
The FONA XDG system configurations comply with the following standards.
IEC 601-1 General requirements for safety
IEC 601-1-2 Electromagnetic compatibility
IEC 601-1-3 General requirements for radiation protection in diagnostic X-ray equipment
Particular requirements for the safety of high voltage generators of diagnostic
IEC 601-2-7
X-ray generators
Particular requirements for the safety of X-ray source assemblies and X-ray
IEC 601-2-28
tube assemblies for medical diagnosis
21 CFR 1020.30 Diagnostic x-ray systems and their major components
21 CFR 1020.31 Radiographic equipment
93/42/EEC European Directive concerning medical devices (1993)
1.4 Environmental Data
Applicable ranges of temperature, humidity, and atmospheric pressure are reported here below both for
operation and transport conditions.
Ambient Transport & Storage
Temperature from 10 to 40 °C from –20 to +50 °C
Relative Humidity from 30 to 75% 10 to 90%
Pressure from 700 to 1060 hPa from 500 to 1060 hPa

1.5 Obligations of the Installer


Obligations of the Installer are:
 To make sure that the line voltage specified by the Manufacturer of the equipment is available
and within the specified range.
 For safety reasons verify that a proper switch is available to disconnect from line voltage supply when
needed during installation.
 Install and test the equipment with due diligence according to the installation instructions from the
Manufacturer.
 To provide the Operating Instructions to the User.
1.6 Warning
Use the system only after proper assembly and installation as per manufacturer’s instructions.
X-ray equipment produce ionizing radiation that may be harmful if not properly controlled. It is therefore
recommended that the equipment be operated by trained personnel only in accordance with existing law.
Even if compliant to specifications of electromagnetic compatibility, it is recommended not to use the
equipment in presence of external electromagnetic fields, such as those generated by cellular phones,
which might interfere with the electronic circuits of the system.
1.7 Demonstration
In order to use of the system for demonstration purposes radiation emission has to be inhibited by
disconnecting the supply cables to the X-ray head into the wall adaptor or into the timer.
Cables to be disconnected are those leaving the connection block towards the X-ray head (“out”).
Make sure that the disconnected cables are properly insulated to prevent undesired contacts with live
points. This task has to be done by trained personnel only to avoid the risk of electrical shock.

4/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

2. TECHNICAL DATA
2.1 System Supply
Line Voltage 110-120 V (from 99 V to 132 V in sub-ranges depending on THA mounted)
220-240 V (from 198 V to 264 V in sub-ranges depending on THA mounted)
Line Power 400 W
Line Voltage Range 108 - 132 V for type 93 256 01300, 207 - 253 V for type 93 256 01700,
216 - 264 V for type 93 256 01800
Line Fuse Slow Blow: 6.3 A at 110-120 V, 4 A at 220-240 V, second fuse for 2 phases or cord
Line Frequency 50/60 Hz  1 Hz
Line Resistance  0.4 Ohm at 110-120 V,  0.8 Ohm at 220-240 V
Voltage regulation for max. line current: 2.1% at 110-120 V, 1.4% V at 220-240 V
2.2 X-ray Head Assemblies
120 V for type 93 256 01300, 230 V for type 93 256 01700,
Nominal Line Voltage
240 V for type 93 256 01800
6 A at 120 V for type 93 256 01300, 4 A at 230 V for type 93 256 01700,
Nominal Line Current
4 A at 240 V for type 93 256 01800
70 kVp ± 8% at nominal line voltage
Anode Voltage
66 kVp ± 8% at nominal line voltage – 10%
(peak tube potential)
74 kVp ± 8% at nominal line voltage+ 10%
3.5 mA ± 10% at nominal line voltage
Anode Current
3.0 mA ± 10% at nominal line voltage – 10%
(tube current)
4.0 mA ± 10% at nominal line voltage + 10%
Nominal Power 0.2 kW at 70 kVp, 3.5 mA, 0.1 s
X-ray Insert CF4G070
Anode Tungsten, angle 16° to the tube axis
Focal Spot 0.4 (EN 60336:1995-04)
Focal spot mark Dot embossed on plastic covers of tube-head
Inherent Filtration > 2.5 mm Al/70kVp (EN 60522: 1999)
Duty Cycle 1/15
Radiation Leakage < 0.1 mGy/h a 1 m (< 11.5 mR/h a 1 m)
2.3 Beam Limiting Device
Focus skin distance 21 cm (8.27”)
Beam Limiting Circular radiation field 6 cm (2.35”) diameter /Rectangular 3.2x4.2 cm
Device (1.26”x1.65”)
Maximum 10% eccentricity
2.4 TipSet Timer
Supply Voltage 110-120 for type 93 354 60200, 220-240 for type 93 354 60100
Time-current in mAs: from 0.21 to 11.2 mAs
0.21 0.28 0.35 0.42 0.56 0.70
Exposure factor
0.88 1.12 1.40 1.75 2.20 2.80
3.50 4.40 5.60 7.00 8.75 11.2
Precision ± 0.04 mAs o 10% (whichever the greater) supplied at nominal line voltage
Automatic setting through tooth type selection and patient build, for use with
Exposure factors
traditional film or digital sensor, or manual setting with plus and minus keys.
Irradiation signal Yellow light on hand-switch and on control panel plus acoustic buzzer
Hand-switch Hand-switch with 3 m coiled cord, with remote mounting optional kit
Overall size Width: 15 cm /6”, Height: 24 cm /9”½, Depth: 9 cm /3”½
2.5 Mechanical Suspension System
Wall Adaptor 12 cm /4.7” width, 24 cm /9.4” height, 9 cm /3.5” depth
Short (S): 30 cm /11.8”, Medium (M): 60 cm /23.6”,
Arm Length
Long (L): 80 cm /31.5”, Extra Long (XL) 100 cm /39,4”
Useful Reach Arm S: 138 cm /54.3”, M: 168 cm /66.1”, L: 188 cm /74”, XL: 208 cm /81.9”
Mobile Stand Size 78 x 92 cm (30” ¾ x 36” ¼), Height: 186 cm /73” ¼” with scissor arm
2.6 Weights
Timer 1.7 kg / 3.7 lb Support Arm Arm S: 2.8 kg /6.2 lb , M: 4.0 kg /8.8
X-ray Head 6.6 kg / 14.5 lb lb, L: 4.8 kg /10.6 lb,
Circular BLD 0.1 kg / 0.22 lb XL: 5.4 kg /11,9 lb
Rectangular BLD 0.2 kg / 0.44 lb Wall Adaptor 1.3 kg / 2.9 lb
Scissor Arm 11.7 kg / 25.8 lb Mobile Stand 29.4 kg / 64.8 lb

150715 69 550 70210 5/32


FONA XDG – Service & Installation Manual

3. ASSEMBLY AND INSTALLATION


3.1 Wall Mounted Systems
3.1.1 Installation Options
The timer box can be mounted together with the X-ray head on the right of the wall adaptor or can be
placed remote either inside or outside the examination room. The release hand-switch can be directly
connected to the timer box or placed remote making use of specific mounting kit including 10 m cable. A
door contact and/or a second exposure button can be added via the internal socket on the timer board.
3.1.2 Unpacking
Unpack the components of the system and check the following:
T1 T2 T3
A Each item is in good conditions and was not damaged during
transportation.
B All the items for the desired system configuration are available.
C The line voltage on the labels of timer and X-ray head are
corresponding to the local line voltage.
3.1.3 Structural Requirements
The wall adaptor has to be mounted in a convenient position on left or
right side of the chair or on back wall (head of the patient).
The maximum useful reach is of 208 cm (81.9”) from the wall when using
an 100 cm (39.4”) support arm.
The wall adaptor can be mounted with 2, 4, or 6 bolts, depending on wall
quality.
A Two bolts only (top and bottom central holes – T2, B2) are used when
there is a solid slim column (e.g. iron mounting) with weak sides (e.g.
wooden wall). Considering the requested safety factor, the top bolt
has to withstand a load of 8878 N, comprehensive of a safety factor,
i.e. about 2013 lbs or 906 kg. Proper screw to be selected for a solid B3
connection to the wall. Classes ISO 8.8 (M 8, M 8x1, M 8x1.25) or SAE B1 B2
Grade 5 (5/16” 18UNC, 5/16” 24 UNF) are recommended.
B Four bolts, two on top sides (T1, T3) and two on bottom sides (B1,
B3) is the regular mounting for solid (concrete) wall, but also on large metal plate. Considering the
requested safety factor, each bolt at top has to withstand a load of 4439 N, comprehensive of a safety
factor, i.e. about 1007 lbs or 453 kg.
Proper expansion screw to be selected for a solid connection to the concrete wall; the permissible load
of each screw has to be greater than 308 lbs (about 140 kg).
 On solid concrete use heavy duty metal anchors.
 On hollow bricks use injection chemical fixing.
C Six bolts, three on top (T1, T2, T3) and three at bottom (B1, B2, B3) are required when the wall is not
solid enough and the load has to be distributed on more points. Considering the requested safety
factor, each bolt at top has to withstand a load of 2959 N, comprehensive of a safety factor, i.e. about
671 lbs or 302 kg.
In case the wall is not in condition to withstand the indicated load, corrective actions can be evaluated by
adoption of reinforcing plates:
D Large plate to fit vertical supports at 16” distance, with 4 mounting holes and one cable opening for
the wall mount in the middle.
E In case of a thin (wooden) wall not solid enough, the use of a steel counter plate 2 mm thick can be
the solution. Make sure that the wall is solid enough to carry the load.
F The use of two reinforcing steel plates of about 4 times the surface of the wall adaptor, one by each
side of the wall, can help when a single plate looks not adequate; additional bolts have to be used to
hold together the two plates.
Make sure that the wall is solid enough to carry the load.
Recommended Bolts
Diameter Class Core Section mm2
M 8X1.25 ISO 8.8 36.6
M 8X1 ISO 8.8 39.2
5/16” – 18 UNC SAE- Grade 5 33.8
5/16 –24 UNF SAE- Grade 5 37.41

INSUFFICIENT WALL OR HARDWARE STRENGHT MAY CAUSE THE WALL ADAPTOR TO


PULL OUT FROM THE WALL AND THE FULL SYSTEM TO FALL ON TO THE PATIENT OR
THE OPERATOR CAUSING INJURIES.

6/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.1.4 Electrical Requirements


The power line cable must be connected to the input terminals (IN) of the timer to supply the timer itself
and to make available power for the X-ray head at the output terminals (OUT), upon request by the
operator (via the hand switch). Power to the X-ray head is thus controlled by the timer (acting as a
SWITCH).

INPUT
POWER X-RAY
LINE SWITCH HEAD

IN OUT
ELECTRONIC WALL
TIMER ADAPTOR

The cables (2 poles plus ground) to connect the power line to the timer and the timer to the wall adaptor
are not provided. Cables with large section conductors have minimum electrical resistance and voltage
drop.

Wire Type Cross Section Area Resistance/m m/0.05 Ohm


AWG 14 2.08 mm2 0.00829 Ohm/m 6.0 m
AWG 12 3.31 mm2 0.00521 Ohm/m 9.6 m

The resistance between the ground provision in the connecting block of the timer and the metal body of
the tube housing assembly has not to exceed 0.1 Ohm. Proper section of the yellow green wire to be used
in case remote mounting of the timer.
The wall adaptor is provided with a common bolt for bonding and grounding blocked in place
with a nut which has not to be removed.

The grounding conductors to be connected as indicated below.


 First the conductor to ground provision in the connecting block of the timer (flat washer, ground
provision, flat washer, lock-washer, nut).
 Second the bonding conductor of support arm plus, in case, the one of the cable in the scissor arm
(flat washer, bonding conductors, flat washer, lock-washer, nut).

LOCK-WASHER
COMMON BOLT
FLAT WASHER
NUT
BONDING BONDING CONDUCTOR
CONDUCTOR OF OF SCISSOR ARM
SUPPORT ARM (IN CASE)
NUT FLAT WASHER
GROUND PROVISION
TO TIMER LOCK-WASHER
BLOCKING NUT DO FLAT WASHER
NOT REMOVE FLAT WASHER

METAL BODY OF WALL SUPPORT

150715 69 550 70210 7/32


FONA XDG – Service & Installation Manual

3.1.5 Mounting and Connecting Sequence


Step 1 Mount the Wall Adaptor Step 7 Connect the Hand-Switch
Step 2 Mount the Timer Step 8 Optional Remote Hand-Switch
Step 3 Mount the Support Arm Step 9 Mount and connect X-ray head
Step 4 Mount the Scissor Arm Step 10 Mount the collimator
Step 5 Connect the Wall Adaptor Step 11 Final Tuning and Set-Up
Step 6 Connect the Timer

3.1.6 Mounting the Wall Adaptor


A Take away the plastic cover removing the screws under the
logo label.
B Use the Wall Adaptor plate or a template to mark the holes on
the wall.
Please note that the timer can be mounted close to the Wall
Adaptor on the right side or in a remote position. Cabling
should have been laid out accordingly.
C Make the holes in the wall according to the applicable type of
mounting with two, four bolts, or six bolts, with or without
reinforcing plate.
D Slide the logo-strip out from the plastic cover to access the
screws and remove it from the metal frame.
E Mount the metal frame on the wall using proper heavy-duty
metal anchors and make sure that the power cable enters
from behind.
F Secure it to the wall ensuring it is leveled. Improper leveling
might cause the arm to move and swing out of position.
G Perform wire connection and final set-up complying with
recommended sequence of actions reported in the following.

8/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.1.7 Mounting the Timer


A Use the mounting plate or the template to mark the
holes on the wall.
B Remove the plastic cover after having taken away the
blocking screws behind the logo strip. Pay attention
when disconnecting the flat cable of the control panel
from the control board.
C Drill the four holes in the wall and secure the timer
making sure it is leveled. Cabling should have been laid
out in accordance to mounting requirements.
D Perform wire connection and final set-up complying with
recommended sequence of actions reported in the
following.

3.1.8 Mounting the Support Arm


A Unpack the support arm and check for completeness of parts.
B Do not remove the rope to pull the cable of the Scissor Arm to the Wall Adaptor through the Support
Arm.
C Push back the cylinder of the side friction not to interfere, lightly grease the shaft of the extension arm
and insert it into the wall adaptor.
D Mount the rotation end-stop spacer and close the side friction without tightening.

150715 69 550 70210 9/32


FONA XDG – Service & Installation Manual

3.1.9 Mounting the Scissor Arm

WARNING. THE SPRINGS IN


THE SCISSOR ARM MAY CAUSE
INJURY TO THE INSTALLER AS
WELL AS DAMAGE TO THE ARM
ITSELF IF NOT HANDLED
PROPERLY. DO NOT REMOVE
THE BINDING STRING UNTIL
WHEN NECESSARY

A Light grease the shaft at the base of the arm and make sure
the ring spacer is in place; push back the cylinder of the
friction into the support arm not to interfere during the
insertion of the shaft.
B Secure the trailing rope to the end of the electrical cable of
the Scissor Arm.
C Pull the other end of the trailing rope to drive the electrical
cable of the Scissor Arm through the Support Arm out into
the Wall Adaptor.
D Hold the opening section of the Scissor Arm, carefully
remove the safety string and allow the arm to open slowly,
away from people.
E Perform final set-up complying with recommended sequence
of actions reported in the following.

10/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.1.10 Connecting the Wall Adaptor with Timer at side

MAINS LINE
INPUT TO
TIMER

YELLOW-GREEN SUPPLY CABLE


BONDING OUTPUT TO
CONDUCTOR OF TUBE HEAD
SUPPORT ARM

SUPPLY CABLE
FROM TIMER
TO TUBE HEAD

YELLOW-GREEN
GROUND PROVISION
TO TIMER
YELLOW-GREEN YELLOW-GREEN
GROUND PROVISION GROUND PROVISION
FROM MAINS ON TIMER

A Pass the conductor of the grounding provision (already connected to the common bolt) into the timer
for connection to the central ground point in the connecting block (protective earthing).
B Connect the grounding provision of the extension arm to the common bolt on the wall plate (flat
washer, ground provision, lock-washer, nut).
C Pass through the supply cable of the arm into the timer for connection to the OUT positions (supply
cable output to X-ray head).
3.1.11 Connecting the Wall Adaptor with Timer remote
A In case the connection to the mains is done starting from the wall adaptor a 5 position connection
block has to be mounted. This approach is recommended allowing for a lower ground resistance.
a The ground wire of the mains has to be connected in first position on the common bolt and
blocked with a nut.
b The mains line goes to the remote timer and back through the connection block to the X-ray head.
c The other grounding provisions for support arm and of cable returning from the timer to be
connected to the common bolt and blocked with a second nut.
d The ground resistance to consider is actually the one of the cable in the scissor arm.

B In case the connection to the mains is done through the timer box, the ground resistance to consider
includes the wire from the timer to the wall adaptor.
a Prepare the terminations of the cable for remote connection of timer with tips to fit the connecting
blocks in timer and connecting blocks plus the common bolt for grounding in wall adaptor.
Make sure the copper section of the yellow green wire is large enough (see section 0,
Electrical Requirements, at page 7).

YELLOW-GREEN TO YELLOW-GREEN
COMMON BOLT ON
WALL ADAPTOR

WALL TIMER
ADAPTOR SIDE
SIDE

150715 69 550 70210 11/32


FONA XDG – Service & Installation Manual
C Remove from the common bolt for bonding and grounding on the wall adaptor the existing cable for
ground provision.

BOLT FOR
BONDING AND
GROUNDING

SUPPLY CABLE
FROM REMOTE
SUPPLY TIMER TO TUBE
CABLE HEAD
TO TUBE
HEAD

YELLOW-GREEN
BONDING CONDUCTOR
SUPPORT ARM YELLOW-GREEN
GROUND PROVISION
YELLOW-GREEN FROM/TO TIMER
BONDING CONDUCTOR
SCISSOR ARM
D Connect the cable for remote connection of the timer by placing the ground wire as first conductor on
the common bolt for bonding and grounding; block with washers and nut.
E Place the phase and neutral wires to the connecting block.
F Connect to the common bolt for bonding and grounding the bonding conductor of the extension arm
and the one of the cable of the arm.
G Connect the phase and neutral conductors of the supply cable in the arm to the connecting block.
See section 0,
H Electrical Requirements, at page 7 to determine cable length and maximum distance.

CONNECTIONS IN THE REMOTE TIMER

CONNECTING BLOCK
IN THE TIMER

MAINS SUPPLY INPUT TO TIMER

SUPPLY CABLE FROM REMOTE


TIMER TO TUBE HEAD

12/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.1.12 Connecting the Timer


FUSE F4
NEUTRAL

GROUND

LIVE ZERO OHM


RESITOR
HAND-SWITCH
CONNECTION
JUMPERS
FUSE F2

HAND-SWITCH
SOCKET TST2

IN
HAND-SWITCH
OUT SOCKET TST1

FUSE F1

A Turn-off the voltage supply line. Connect the three wires from the voltage supply to the terminal
block (IN on the left) but do not connect any outgoing wire (OUT on the right) to Wall Adaptor.
B Make sure that the “live” conductor is the “hot” one:
a Connect an AC voltmeter or a test light between the block terminal “live in” and “ground”.
b Turn line voltage supply “ON”. If full line voltage is measured (test lamp lights) the wiring is
correct. If not turn line voltage supply “OFF”, reverse “live in” and “neutral in” wires and repeat the
measurement; eventually the full line voltage should be read between “line” and “ground” (test
lamp lights).
c Make sure no voltage is read between “neutral in” and “ground”; if not check line voltage
distribution.
C Test the timer for full functionality keeping the output cables disconnected not to supply the load.
D Connect the load.
a Turn OFF the line voltage supply line,
b Connect the three wires of additional cable outgoing to the Wall Adaptor, Arm, Tube Housing
Assembly. Live and neutral wires to the arm (X-ray head) can be interchanged.
Fuses of FONA XDG Timer
Ref. 115 V 230 V and 240 V
F1 6.3 AT 5x20 4 AT 5x20
F2 6.3 AT 5x20 4 AT 5x20
F3 315 mAT 5x20 315 mAT 5x20
F4 80 mAT 5x20 50 mAT 5x20
3.1.13 Connecting the Hand-Switch
A The hand-switch is provided with a 3 m coiled cord to be plugged one side in the hand-switch, the
other side in the lower left corner of the timer’s board, either externally (connector TST1) or internally
to the timer (connector TST2). Internal connection is for remote hand-switch or for additional control
safety switches.

150715 69 550 70210 13/32


FONA XDG – Service & Installation Manual
B Jumpers are available to activate properly the hand-switch connection:
a Position TST1 to enable external socketTST1,
b Position TST2 to enable the internal socket TST2,
c Position TST1/2 f to enable the internal and the external sockets.
3.1.14 Optional Remote Hand-Switch
The hand-switch can be remotely mounted by making use of the optional kit made of a remote box (wall
holder) and connection cable 10 m long.
The connection of the wires is here indicated.

The wires of the switch are number 3 and 4


(those in the middle of the connector).

3.1.15 Mounting and Connecting the X-ray Head


A Remove the handgrip pipe
HANDGRIP from the arm and place it onto
PIPE the tube-head shaft of the
coupling yoke.
B Lubricate the shaft, insert it
into the arm passing the
cables into the coupling yoke,
then hold it in place with the
retaining fork.
C Attach the connectors of the
supply and grounding wires
and push them into the side
opening of the coupling yoke.
D Pull down the O-ring and
push-up the handgrip pipe,
then fix it in place with the
screw at bottom.
E Verify smooth movement of
the tube head around its
vertical and horizontal axis.
RETAINING Tune side friction if necessary
FORK referring to section 4.4,
Maintenance of X-ray Head, at
page 22 for details.
F Mount the cover on the yoke opening eventually.

COUPLING
YOKE

14/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.1.16 Beam Limiting Device


A Mount the beam limiting device on the tube-head and lock it close with a clock-wise rotation.

3.1.17 Final Tuning and Set-Up


A Remove the line voltage fuse(s) from the timer (Fuse 1 and Fuse 2).
B Tune friction screw in Wall Adaptor for rotation of Support Arm (see section 4.1, Maintenance of Wall
Adaptor at page 20 for details).
C Tune friction screw in Support Arm for rotation of Scissor Arm around it vertical axis (see section 4.2,
Maintenance of Support Arm at page 20).
D The friction points in the scissor arm and those for the rotation of the X-ray head around its horizontal
axis are tuned in factory. Should you need to repeat the procedures refer to sections 4.3, Maintenance
of Scissor Arm, at page 21 and 4.4, Maintenance of X-ray Head, at page 22, respectively.
E Make sure the Timer switch is in the “OFF” position.
F Set the dip switches SW2 as indicated in the following to define:
a Enable/disable function for correction of exposure in presence
of line voltage fluctuations.
b The nominal line voltage of the X-ray head (see label).
c Set dip switch 4 to the anode current set in the X-ray head
label, which is 3.5 mA for FONA XDG.
d Set the pre-heating time according to type of X-ray insert used
in the X-ray head assembly; the measured value is marked on
the X-ray head.
e Select the technique factor on display as mAs (always mAs for
the timer of FONA XDG).
G Put back the line voltage fuse(s).
H Put back all plastic covers and logo strips.
I Connect the system to the line voltage by switching the general line
voltage switch on and/or plug the power cord into the wall socket.
J Switch the timer “ON”.
The system is now ready for functional checks.

150715 69 550 70210 15/32


FONA XDG – Service & Installation Manual

3.2 Dip Switch Setting

DIP 4 DIP 8
ANODE CURRENT TECHNIQUE
FACTOR
DISPLAYED

DIP 1
TECHNIQUE DIPS 2-3 DIPS 5-6-7
FACTOR TUBE HEAD PRE HEATING
CORRECTION TYPE TIME

DIP 1 DIPS 5 – 6 – 7
Technique Factor Correction Pre Heating Time

Disabled 100 ms

Enabled 120 ms

DIPS 2- 3
140 ms
Tube Head Type
110 or 220
160 ms
VAC
115 or 230
180 ms
VAC
120 or 240
200 ms
VAC
127 or 250
220 ms
VAC
DIP 4
240 ms
Tube Head Anode Current
DIP 8
3.5 mA
Technique Factor Displayed

7 mA s

mAs

16/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.3 Layout Power Board Timer

POWER
SUPPLY
SET-UP KEYBOARD 5 V LED
MICROSWITCHES CONNECTOR

BACK UP TIMER
TEST POINT & LED
MICROPROCESSOR
WATCH DOG LED
POWER
SUPPLY
24 V LED

5VA TRANSFORMER
PRIMARY 115/230 V
SECONDARY 18 V
FUSE 4
T50mA @230V
T80mA @115V
FUSE 3
T 315 mA

JUMPER HANDSWITCH
FUSE 2 JUMPERS

FUSE 1 & FUSE 2


T4A @230V
T6.3A @115V

INPUT
MAINS LINE OUTPUT
NEUTRAL LINE TO MAINS
& LIVE TUBEHEAD SWITCH

HANDSWITCH
SOCKETS

150715 69 550 70210 17/32


FONA XDG – Service & Installation Manual

3.4 Mobile Systems


3.4.1 Unpacking
Unpack the components of the system and check the following:
A Each item is in good conditions and was not damaged during transportation.
B All the items for the desired system configuration are available.
C The line voltage on the labels of timer and X-ray head correspond to the existing local
line voltage.
3.4.2 Room Preparation
Make sure that a wall socket is available close to the dental chair in reach of the mobile X-
ray equipment.
Such a socket has to be provided with connection to the protective grounding.
The line voltage cable can come without plug to be added according to the local standard.
3.4.3 Mounting and Connecting Sequence
Step 1 Assemble the Mobile Stand
Step 2 Mount the Scissor Arm
Step 3 Mount the Scissor Arm
Step 4 Mount the Timer
Step 5 Connect the Line Voltage Cable
Step 6 Connect the Timer
Step 7 Activate second Fuse
Step 8 Mount and connect X-ray head
Step 9 Mount the collimator
Step 10 Final Tuning and Set-Up

3.4.4 Assembling the Mobile


Stand
A Mount the legs below the central base.
B Mount rear wheels and front wheels with brakes.
C Attach the pole with timer plate.
D Complete the stand with cable holder and
handles

3.4.5 Mounting the Arm


A Loosen the friction and push back the cylinder of
the same not to interfere during shaft insertion.
B Insert the arm. Refer to section 3.1.9 at page 10
for details.

18/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

3.5 Mounting the Timer


The timer is blocked in place with four bolts.
3.5.1 Connecting the Line
voltage cable
A The line voltage cable for the mobile has
to be completed with local plug.
B The power supply cord bonding
conductor has to be blocked (first) to the
common bolt for bonding and grounding
using lock-washer and nut.
C The grounding wire for the timer is then
placed on the same common bolt and
blocked (second) with lock- washer and
nut.
D Connect the grounding wire for the timer
to the ground provision on the terminal
block of the timer and the two line wires
of the power cord to the input points of
the terminal block of the timer.

3.5.2 Connecting the Timer


For the connection of the output cable to the
timer please refer to section 3.1.11 at page
13.

3.5.3 Activation of Fuse 2


Units provided with supply cord and
plug must be equipped with fuses
on both line phases. The activation POWER
on the second fuse is done by cutting the SUPPLY
Jumper Fuse 2 (zero Ohm resistor) close to CORD
Fuse F2. See position on layout in section 0 BONDING
at page 17.
CONDUCTOR

3.5.4 Mounting and Connecting the X-ray Head


Refer to section 3.1.15 at page 14.

3.5.5 Mounting the Beam Limiting Device


Refer to section 3.1.16 at page 15.

3.5.6 Final Tuning and Set-Up


Refer to section 3.1.17 at page 15.

150715 69 550 70210 19/32


FONA XDG – Service & Installation Manual

4. MAINTENANCE
Here in the following the list of recommended maintenance actions to keep the FONA XDG system in
compliance with the Performance Standard.
It is the responsibility of the User to maintain the equipment in compliance with the standard. Failure of
the User to properly maintain the equipment may relieve the Manufacturer, or its Agent, from responsibility
for any injury, damage or non compliance which may result.
Maintenance for the FONA XDG system to be done regularly by a service technician at least once every 24
months, with regular checks performed by the operator every year.
Any defect or malfunction should be corrected immediately by qualified personnel with adequate training.

WARNING.

 Any defective item affecting A safe use must be repaired or replaced


 Only original certified components and spare parts must be used for repairs or replacements.
Correction of damages to the identification labels to be handled with the Manufacturer. Any defect or
non-compliance must be reported promptly to the Manufacturer or to its Local Agent.
 Always disconnect the system from the line voltage supply using the general line voltage switch for the
room where the equipment is located before performing any maintenance activity
 Avoid using liquid or spray detergents which may enter into the equipment and cause corrosion.
 Avoid the use of solvents or corrosive detergents which can damage painted surfaces and plastic
covers.
4.1 Maintenance of Wall Adaptor

POSITION OF 21
DISK SPRINGS IN
FRICTION SCREW FRICTION
7 GROUPS OF 3 SCREW

A Remove the cover and verify that the mounting is closely


connected to the wall and stays firm and steady during the
movements of the system. BOLT FOR
B Verify for proper connection of grounding wires to the common BONDING AND
bolt.
C Check friction during rotation of Extension Arm and adjust if
GROUNDING
needed.
D Verify that the technical label with identification data outside the
plastic cover at bottom is in place and is readable.

4.2 Maintenance of Support Arm

POSITION OF 15 FRICTION
DISK SPRINGS IN SCREW
FRICTION SCREW
5 GROUPS OF 3

A Slide partially out the extension arm from wall mount, inspect for
wear of the joint and lubricate for smooth rotation.
B Check friction during rotation of Arm and adjust relevant screw if
needed.
C Verify that the technical label with identification data is in place and is readable.

20/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

4.3 Maintenance of Scissor Arm


FRICTION SPRING
POINT 3 TUNING
POINT

FRICTION
POINT 2

SPRING FRICTION
TUNING POINT 4
POINT

A Balance the arms (2 sections) and tune


frictions (4 points) if needed in normal
conditions with X-ray head mounted.
B To balance the arm, for each arm’s
section:
LABEL a Remove plastic covers at the joints.
b Reduce the friction at the 2 friction
points by loosening the socket set
screw (allen) into the female friction
pin which is blocking male and female
pins together, then loosen them to
reduce friction.
c WARNING. Trying to loosen the
friction pins without having loosen the
FRICTION socket set screw might lead to
breaking the friction pins.
POINT 1 d Tune the springs using an allen key
for even balancing with arm section in
intermediate position. UNLOCK
e Tighten the two friction pins for little
resistance to the movement, and
SOCKET SET SCREW
eventually tighten the socket set WITH ALLEN KEY
screw (allen) into the female friction
pin to block male and female friction pins together.

ALLEN KEY
C Specifically for the arm, at least ALLEN KEY
every 24 months in case of
normal use or after every 5000
cycles performed, have the
service technician inspect for
wear of pins and levers.
a Remove plastic covers.
b Check for wearing out of pin and levers, starting from the joint at the
base of the arm.
c Make sure no part is loose and no vibrations are occurring during movements.
d Apply silicon spray lubricant if needed.
e Replace the worn out parts or the arm itself if they appear damaged.
D To replace defective pins or levers
a have the X-ray head disconnected with arm fully open to avoid injuries for sudden opening.
b Loosen completely the two springs of the arm before extracting pins or removing levers.
E Mount back the X-ray head if removed and secure it with the retaining fork.
F Slide partially out the Scissor Arm from the Support Arm, inspect for wear of the vertical joint and
lubricate for smooth rotation.
G Verify that the label with identification data is in place and is readable.

150715 69 550 70210 21/32


FONA XDG – Service & Installation Manual

4.4 Maintenance of X-ray Head

HOLDING
POSITION OF 8 SCREW
DISK SPRINGS IN
BLOCKED
FRICTION SCREW
4 GROUPS OF 2
A Inspect for damage/wear to the X-ray tube-
head and support system.
B Inspect for oil leakage. Replace tube-head if FRICTION
necessary. SCREWS
C Check for position stability. If necessary remove
the cap on the yoke side, tune the three friction
screws (with disk springs) to be moderately tight for smooth rotation.
The fourth one without disk springs must be blocked in place
D Check that electrical cables are correctly twisted and not damaged or worn down.
E Verify that the technical label with identification data is in place and is readable.
F Dismount the tube-head from the arm and lubricate the pin for smooth rotation.
Warning: remove the tube-head with the arm completely open to avoid the risk of injuries.
G Check the retaining clip and replace if worn down. Do not mount the tube-head until all checks of the
suspension system are completed.
4.5 Maintenance of Beam Limiting Device
A Verify that the collimator is properly mounted and firmly closed.
B Verify that the internal cone is not damaged. Replace the collimator if necessary.
C Verify that the technical label with identification data is in place and is readable.
4.6 Maintenance of Timer
A Inspect the panel for damages on the surface.
B Inspect the cable of the exposure switch for wear.
C Check that all lights (except the yellow one) are working when the system is switched on.
D Perform a functional check with verification of a) setting of dip switches, b) yellow light, c) buzzer,
d) alarm A07 for “Exposure push button pressed at power on”, e) alarm A08 for “Exposure stopped by
the operator”.
E Verify that the fuses mounted comply with those indicated on the label and listed below.
F Verify that the technical label with identification data is in place and is readable.

Fuses of FONA XDG Timer


Ref. 115 V 230 V and 240 V
F1 6.3 AT 5x20 4 AT 5x20
F2 6.3 AT 5x20 4 AT 5x20
F3 315 mAT 5x20 315 mAT 5x20
F4 80 mAT 5x20 50 mAT 5x20
4.7 Checking Filament Pre-Heating Time
During the life of the system the wearing down of the filament may result in a different pre-heating time
(PHT). The value set at installation needs therefore to be verified and in case corrected.
A At a fixed distance measure the X-ray dose for exposure of 2.0 s (nominal irradiation of 100 pulses at
50 Hz, 120 at 60 Hz).
B At same distance measure the dose for exposure of 0.2 s (nominal irradiation of 10 pulses at 50 Hz, 12
at 60 Hz).
C If dose at 0.2 s is 1/10 of dose at 2 s, the value set for PHT is fine.
D If dose at 0.2 s is lower than 1/10 of the dose at 2 s, PHT has to be increased. Proceed at steps of 20
ms until fine.
E If dose at 0.2 s is higher than 1/10 of the dose at 2 s, PHT has to be reduced. Proceed at steps of 20
ms until fine.

22/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

4.8 Service Functions


A READING EXPOSURE COUNTER. Enter by switching ON the
unit while pressing at the same time for 2 s, the three keys
plus , minu s , and upper incisor . The cumulative
number of exposures is of 9 digits shown on the display in
groups of three digits, from the most to the least significant,
selectable using the plus , minus keys. Each group is
identified by the dot at bottom, on the left for the most
significant group, in centre for the intermediate group, at the

right for the least significant one. Press the sensor key to
exit (E).

B READING ENERGY COUNTER. Enter by switching ON the unit


while pressing at the same time for 2 s, the three keys plus

, minus , and upper molar . The cumulative value


of the energy imparted in Joule (J) is of 9 digits shown on the
display in groups of three digits, from the most to the least
significant, selectable using the plus , minus keys.
Each group is identified by the dot at bottom, on the left for
the most significant group, in centre for the intermediate
group, at the right for the least significant one.

Press the computer key to exit (E).

C DIGITAL VOLT METER. This function allows the user to display


the line voltage level and the frequency, alternating.
a Enter this function by switching ON the unit while
pressing at the same time

the , the , and the sensor keys.


b Exit by switching the system OFF.

D TEST CONTROL PANEL. Function i to test lights, buttons, and


segments of numerical display on the control panel.
a Enter by switching the system “ON” while pressing at the

same time the , the , and the adult keys.


b Press the or keys to check the segments of the
numerical display.
c Press the X-ray pushbutton for the “rAY” message.
d Press each button on the panel to turn ON and OFF the
corresponding light.
e At the last button pressed the system exit this function.

E FIRMWARE VERSION. This function returns the code of the


version of firmware loaded.
a Enter this function by switching the system “ON” while
pressing at the same time

the , the , and the adult child keys.


b Exit by switching the system OFF.

5.

150715 69 550 70210 23/32


FONA XDG – Service & Installation Manual

5. MEASUREMENTS
All measurements to be done with the equipment supplied at nominal line voltage with specified line
resistance. The tolerances on the measured values have to take into account the precision of each
measurement instrument. Measurements have to be done by trained personnel only to avoid the risk of
electrical shock.
5.1 Line Voltage
The Line voltage can be measured using a Volt Meter for Alternate Current within the proper range. A
service function can be activated on the timer to display the current line voltage in V and frequency in Hz
(see section 4.8 of this Manual).
5.2 Anode Voltage – KVp
The kVp level is the actual peak value of the Anode voltage which stabilizes once the preheating time of
the filament has elapsed (in about 0.2 s) and the high voltage transformer is actually loaded. The kVp level
can be measured with a non-invasive kVp meter placed in front of the BLD following the instructions in the
user manual of the instrument. A correct measure can be done with an exposure time of 500 ms or more
having introduced a delay of the reading instrument of about 300 ms to allow the voltage level to stabilize
after the pre-heating time has elapsed.
5.3 Anode Current – mA
WARNING. To prevent high voltage shock make sure
the system is disconnected from the power supply
when connections to the measurement points are
performed. Electric discharge might occur in case of improper
operation. A1
The Anode current in mA is the actual average value of the
tube current which raises when the filament has warmed-up,
after the pre-heating time has elapsed.
Remove the back cover of the tube-head, disconnect the
jumper A1-A2 and replace it with a resistor of 1000 Ohm.
The anode current can be read with a DC voltmeter connected A2
to points A1 and A2, with the 1 V corresponding to 1 mA.
After the measurement disconnect the 1000 Ohm resistor and
put back the jumper.
5.4 Exposure Time
Here below the timing indications in s (R10 scale) and line pulses at 50 and 60 Hz for each value of
exposure factor in mAs for tubehead at 3.5 mA.
mAs Pulses Pulses mAs Pulses Pulses
@3.5 mA s 50 Hz 60 Hz @3.5 mA s @50 Hz @60 Hz
0.21 0,06 3 4 1.75 0,50 25 30
0.28 0,08 4 5 2.20 0,63 31 38
0.35 0,10 5 6 2.80 0,80 40 48
0.42 0,12 6 7 3.50 1,00 50 60
0.56 0,16 8 10 4.40 1,25 62 75
0.70 0,20 10 12 5.60 1,60 80 96
0.88 0,25 12 15 7.00 2,00 100 120
1.12 0,32 16 19 8.75 2,50 125 150
1.40 0,40 20 24 11.2 3,20 160 192
The actual exposure (loading) time is “determined as the time interval between the instant when the tube
current first rises above the 25% of its maximum value and the instant when it finally falls below the same
value”.
In order to assure the requested exposure time (irradiation), the pre-heating time of the insert used, which
has been set-up at installation, is taken into consideration. The actual switch-on time (SOT) of the X-ray
head assembly is therefore the sum of the filament pre-heating time (PHT) and of the requested exposure
time (RET).
SOT = PHT + RET
The measurement of the exposure time can be done with kVp & Time meters with the ability to trigger at
75% of maximum kV level.

24/32 69 550 70210 150715


FONA XDG – Service & Installation Manual
Exposure time can also be measured after having connected a resistor of 1000 Ohm between points A1
and A2 on the small circuit board on the back of the X-ray head, by monitoring the voltage across it (tube
current) with an oscilloscope, and counting the number of pulses in the time interval.
When the function to correct the effects of line fluctuations on the emitted X-ray dose is activated, the
exposure time is corrected in order to assure dose consistency. When the actual line voltage is lower than
the nominal line voltage, the actual exposure time is increased. Conversely, when the actual line voltage is
higher than the nominal line voltage, the actual exposure time is decreased.
Do disable this correction function (dip switch number 1 of SW2 moved to OFF position) to verify exposure
time accuracy.
5.5 Leakage Radiation
The value of leakage radiation of the X-ray tube assembly is measured at 1 m distance taking into account
the energy provided in 1 hour. In case the measurement is performed at a distance different from 1 m, the
value has to be properly corrected.
The applicable technique factors for FONA XDG are of 70 kV and 3.5 mA. The point of maximum leakage
has to be considered, excluding the direction of the primary beam. The duty cycle factor has also to be
taken into account for the computation of the actual dose rate. In case for example a measure is done at
50 cm distance from the source with an exposure of 1 s, the following computations have to be performed:
 A duty cycle of 1/15 means maximum of 1 s exposure every 15 s, i.e. 240 s per hour.
 Dose at 100 cm is computed taking into account the inverse square law, i.e. at twice the distance the
dose intensity is reduced to one quarter. D1h100cm = D1s50cm * 240 * (50/100)2 = D1s50cm * 60
where: D1s50cm is the dose measured at 50 cm with 1 s exposure
and D1h100cm is the dose computed at 1 m distance for the energy of 1 hour.
5.6 Earth Resistance
The measurement of the earth resistance to be done after having disconnected the system for the power
supply line. Using a voltage generator of not more than 6 V, 50/60 Hz, a current of 25 A is applied from 5 s
to 10 s through the earth terminal in the connecting block of the timer to any metal part which can be in
touch to the patient/operator (applied part). The voltage drop is then measured and the earth resistance
computed. The system complies for values up to 0,1  (extended to 0,2  in case of mobile system).
5.7 Earth Leakage
Connect the Timer to the measuring circuit as
shown in the figure (R1=10 k  5%, R2=1
k  1%, C1=0,015 F 5%)
A Set the input supply voltage (Vp) equal to
110% of the nominal supply voltage.
B Connect a voltmeter (Vx with 1 MOhm input
impedance) to the indicated points
C Switch the Timer “ON”
D Make two measures in Normal Conditions
(NC with S1 “ON”)
 S5 “ON”: live and neutral conductors in
normal condition.
 S5 “OFF”: swap of live and neutral
conductors.
E Make two measures in Single Fault
Condition (SFC with S1 “OFF”).
 S5 “ON”: live and neutral conductors in normal condition.
 S5 “OFF”: swap of live and neutral conductors.
Please note that 1V equals to 1mA. Values of leakage currents have to be within the limits of table below.

Limit of earth leakage current for X-ray equipment


Type of connection Normal Condition Single Fault Condition
to line voltage supply (S1 “ON”) (S1 “OFF”).
Permanent 5 mA 10 mA
With power cord 2,5 mA 5 mA

150715 69 550 70210 25/32


FONA XDG – Service & Installation Manual

Appendix A
Systems and Components
FONA XDG Systems FONA XDG Systems FONA XDG Systems
at 230 V at 240 V at 120 V
Description Code Code Code
Wall Syst. Arm S 30 cm 97 586 62320 97 566 62320 97 546 62320
Wall Syst. Arm M 60 cm 97 586 62620 97 566 62620 97 546 62620
Wall Syst. Arm L 80 cm 97 586 62820 97 566 62820 97 546 62820
Wall Syst. Arm L 100 cm 97 586 62120 97 566 62120 97 546 62120
Mobile System 97 586 67820 97 566 67820 97 546 67830

C
G H
F

I
A
G

L
E J

D
K

ID Description Code ID Description Code


A Wall adaptor 93 105 11000 X-ray head XDG 120 V 93 256 01300
Support arm S 30 cm 93 150 17100 F X-ray head XDG 230 V 93 256 01700
Support arm M 60 cm 93 150 17200 X-ray head XDG 240 V 93 256 01800
B
Support arm L 80 cm 93 150 17300 TipSet Timer 110-120 V 93 356 60200
G
Support arm L 100 cm 93 150 17400 TipSet Timer 220-240 V 93 356 60100
C Scissor arm 93 150 12010 H 6 cm circular field BLD Grey 91 360 00020
Mobile stand 93 150 20080 I 3x4 cm X-ray field BLD 91 300 00040
D
Mobile stand UL/CSA 93 150 20090 J Coiled cable 76 190 25580
E Wall plate 16” 86 100 11500 K Hand-switch W/O cable 76 190 25590
L Kit remote hand-switch 93 190 00155

26/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

Appendix B
Icons

IEC Type B Compliance to European


Equipment Community Requirements

X-ray On Compliance to Canadian and US


Irradiation Standards

Line voltage supply On - System


Examine Annexed Documentation
Ready

Off (Disconnected from Line voltage


Increase Exposure Time (one step)
Supply)

On (Connected to Line voltage


Decrease Exposure Time (one step)
Supply)

Child – Small Patient ~ Alternate Current

Adult – Large patient Fuse

Upper Incisor Protective Earth

Neutral Point (for equipment


Upper Canine/Premolar N permanent connected to line)

Live Point (for equipment


Upper Molar L permanent connected to line)

Lower Incisor Inherent Filtration

Lower Canine/Premolar  Focal Spot

Fragile,
Lower Molar
Handle With Care

Bite Wing - Interproximal Fear of Humidity

Up
Digital Receptor
Do Not Overturn

Radiography Push Button Stacking Limit

Separate Collection,
Ionizing Radiation
Do Not Dispose

150715 69 550 70210 27/32


FONA XDG – Service & Installation Manual

Appendix C
Exposure Table

Upper
Upper Canine/ Upper
Incisor Premolar Molar

Lower Lower
Large Lower Canine/ Molar or Small
Patient Incisor Premolar Bitewing Patient

0.21 0.28 0.35 0.42  0.32


CDR
0.28 0.35 0.42 0.56  0.40 SENSOR
XIOS
0.32  0.35 0.42 0.56 0.70  0.50 SENSOR
CDR
SENSOR 0.40  0.42 0.56 0.70 0.88  0.63
XIOS
SENSOR 0.50  0.56 0.70 0.88 1.12  0.80
E
0.63  0.70 0.88 1.12 1.40  1.00 FILM
0.80  0.88 1.12 1.40 1.75  1.25
E
FILM 1.00  1.12 1.40 1.75 2.20  1.60
D
1.25  1.40 1.75 2.20 2.80  2.00 FILM
1.60  1.75 2.20 2.80 3.50  2.50
D
FILM 2.00  2.20 2.80 3.50 4.40  3.20
2.50  2.80 3.50 4.40 5.60  4.00
3.20  3.50 4.40 5.60 7.00 EXPOSURE
INDEX
4.00  4.40 5.60 7.00 8.75
5.60 7.00 8.75 11.20
EXPOSURE
INDEX 3.50 4.40 5.60 7.00
4.40 5.60 7.00 8.75
5.60 7.00 8.75 11.20
FONA XDG Exposure Factors in mAs
CDR Sensor: Schick Technologies. Inc. Long Island City, NY, USA
XIOS Sensor: Sirona Dental Systems GmbH, Bensheim, Germany
Film D: Kodak Ultraspeed, Agfa Dentus M2 – Film E: Kodak Ektaspeed Plus - Film F: Kodak Insight

28/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

Appendix D
Alarm Conditions

Timer Alarm Conditions

Code Fault /Error Signal Reset


By acknowledgement
X-ray requested during Green lamp on the panel or when
A 01
cool-down period (System Ready) flashing system has cooled
down
Green lamp (System Ready) Automatically when
Line voltage below lower
A 02 and red lamp (Alarm) line voltage is back in
limit
flashing range
Green lamp (System Ready) Automatically when
Line voltage above upper
A 03 and red lamp (Alarm) line voltage is back in
limit
flashing range
System Ready (green) lamp
Line Frequency By switching system
A 06 and Alarm (red) lamp
Detection Failure off and on again
flashing
Exposure push button Red lamp By acknowledgement
A 07
pressed at power on (Alarm) flashing on the panel
By acknowledgement
Exposure stopped by the Red lamp
A 08 on the panel or after 1
operator (Alarm) flashing
m
Exposure stopped by the Red lamp (Alarm) By switching system
A 09
back-up timer switched on off and on again

Red lamp (Alarm) By switching system


A 10 Back-up relay failure
switched on off and on again

Power switching device Red lamp (Alarm) By switching system


A 11
failure switched on off and on again

Line dips during Red lamp (Alarm) By acknowledgement


A12
exposure switched on on the panel

150715 69 550 70210 29/32


FONA XDG – Service & Installation Manual

Appendix E
Identification Labels

30/32 69 550 70210 150715


FONA XDG – Service & Installation Manual

Appendix F
Cooling Curves

COOLING CURVE OF X-RAY INSERT

kJ

Time (min)

COOLING CURVE OF TUBE HOUSING ASSEMBLY


kJ

Time (min)

150715 69 550 70210 31/32


We reserve the right to make any alterations which may be required due to clinical improvements

FONA XDG Service and Installation Manual – English Edition 150715

FONA Dental s.r.o. FONA S.r.l.


Stefanikova 7 SK-811 06
Bratislava, Slovakia
www.fonadental.com
Via Galilei 11 - 20090 Assago
(MI) Italy *6955070210*

You might also like