3qs008 Digital Brief Amatek
3qs008 Digital Brief Amatek
3qs008 Digital Brief Amatek
DV3T Viscometers/Rheometers:
Installation Qualification,
Operation Qualification,
and
Performance Qualification
INSTRUMENTATION AND SPECIALTY CONTROLS DIVISION
11 COMMERCE BOULEVARD
Procedure
MIDDLEBORO, MASSACHUSETTS 02346, USA
Comments:_______________________________________________________________________________________
Oscillation check
This check verifies the mechanical condition of the sensing mechanism in the
viscometer/rheometer. Auto-zero the instrument according to directions in the manual.
Do not attach the spindle; the display should show 0.0% torque. The tolerance for this
and all tests is ±0.1%. Gently grasp the viscometer/rheometer’s lower shaft coupling and
lift slightly. Slowly rotate it until the display % reads 10%-15%, then release.
NOTE: For LV units only, the oscillation check should be 0.0% ±0.2%.
DV2T Viscometer and DV3T Rheometer also have an option under the Device Setting
tab, which instructs the user how to perform the Oscillation Checks.
Calibration Check for LV (#1-3) and RV, HA, HB (#1-6) Brookfield Spindles
NOTE: The LV #4 (64) and RV, HA, HB #7 (07) spindles have been omitted from this
procedure. Brookfield AMETEK does not recommend the use of these spindles to
perform a calibration check on your instrument. Reasons pertain to the small amount of
spindle surface that makes contact with the viscosity standard, the difficulty of
establishing the immersion mark precisely and the need for precise temperature control at
25°C in the immediate vicinity of the spindle.
Comments:_______________________________________________________________________________________
The calibration fluid must be transferred into a 600 mL low form, Griffin beaker (without
creating any bubbles) and, on LV and RV models, the instrument Guard Leg must be
used. See the manual for a description of the use of the Guard Leg. The beaker, fluid,
spindle and Guard Leg must be stabilized at 25.0°C (±0.1°C) before the calibration check
can be performed.
The example below shows typical calibration results using an LV #3 spindle with a
calibration fluid whose actual viscosity is 5,120 cP; the calibration check might look like
the table below. At each speed a different torque is sensed allowing the automatic
computation of viscosity. The allowable error is the sum of 1% of the FSR plus 1% of the
actual value of the calibration fluid. The lower and upper viscosity limits, indicating
acceptable calibration, equal the actual value of the fluid, in this case 5,120 cP; ± the
allowable error.
Remember, the total allowable error when conducting a calibration check is the sum of
1% of the FSR plus 1% of the actual value of the calibration fluid itself. The instrument
reading must be within the limits defined by the actual viscosity value of the calibration
fluid ± the allowable error at each speed. The spindle, speeds and fluid you use may be
different from these.
Comments:_______________________________________________________________________________________
A B C D E F G H I
Fluid 1% of Instrument Spindle FSR cP 1% of % Torque Viscosity cP
Value Fluid Model RPM (depress and hold FSR cP
from Value AUTORANGE key)
Certificate cP
cP
Any reading outside these limits may indicate a Viscometer/Rheometer problem. Contact
the Brookfield AMETEK technical sales department or your local authorized Brookfield
AMETEK dealer with test results to determine the nature of the problem.
Comments:_______________________________________________________________________________________