Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

3qs008 Digital Brief Amatek

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

DV-I, II, III Series, DV2T and

DV3T Viscometers/Rheometers:
Installation Qualification,
Operation Qualification,
and
Performance Qualification
INSTRUMENTATION AND SPECIALTY CONTROLS DIVISION
11 COMMERCE BOULEVARD
Procedure
MIDDLEBORO, MASSACHUSETTS 02346, USA

3Qs008 Revision 7/26/2016 Approved by: Luis Botelho

This guideline for IQ,OQ,PQ is offered as a supplement to the “Installation” section of


this manual. Some customers have particular requirements for validation. It is hoped that
this guideline will assist in meeting those needs. You may also contact Brookfield
AMETEK for a more detailed document, or on-site IQ,OQ,PQ service conducted by a
Brookfield AMETEK representative.
All Brookfield AMETEK instrumentation should be unpacked, assembled and operated
according to the instrument operating instructions. Recommendations for installation,
operation and performance qualification appear below.

IQ: INSTALLATION QUALIFICATION


− Inspect package for shipping damage.
− Confirm that all components have been supplied as detailed in the Operating
Manual and Packing List.
− Assemble lab stand as shown in manual.
− Mount viscometer/rheometer head to assembled lab stand as shown in manual.
− Verify appropriate spindle set (LV #1-#4)(RV/HA/HB #2-#7).
OQ: OPERATION QUALIFICATION
− Level viscometer/rheometer using adjustable feet.
− Remove shipping cap on spindle coupling nut.
− Power on the Viscometer/Rheometer and follow instructions displayed to
complete autozero.
− Perform oscillation check as described on following page.
PQ: PERFORMANCE QUALIFICATION
− Verify calibration using a minimum of one calibrated viscosity standard fluid
according to the procedure described below.
− Centipoise reading should be within the value stated on the fluid ± the
instrument accuracy and the fluid accuracy as defined below.

Comments:_______________________________________________________________________________________

BEL representative:__________________________________ Verified by:____________________________________


Date: 3Qs008 Page 1 of 4
Successful completion of installation, operation and performance procedures as described
above confirm that the Brookfield AMETEK Digital Viscometer/Rheometer is operating
properly and to specification. In the event that any of the above steps cannot be
performed satisfactorily, please contact Brookfield AMETEK or the dealer from
whom you purchased the equipment.

Oscillation check
This check verifies the mechanical condition of the sensing mechanism in the
viscometer/rheometer. Auto-zero the instrument according to directions in the manual.
Do not attach the spindle; the display should show 0.0% torque. The tolerance for this
and all tests is ±0.1%. Gently grasp the viscometer/rheometer’s lower shaft coupling and
lift slightly. Slowly rotate it until the display % reads 10%-15%, then release.

NOTE: For LV units only, the oscillation check should be 0.0% ±0.2%.

DV2T Viscometer and DV3T Rheometer also have an option under the Device Setting
tab, which instructs the user how to perform the Oscillation Checks.

Operational and Calibration Check


Determining Measurement Accuracy
All Brookfield AMETEK instruments are guaranteed to be accurate to within 1% of the
Full Scale Range (FSR) in use. The FSR is simply the maximum viscosity that can be
measured with any spindle/speed combination. Select a specific spindle and rotational
speed. To determine the resulting FSR, consult the manual of the instrument. The %
display will show 100% and the cP display will show the FSR that can be measured
under current choice of spindle and speed. 1% of this value is the measurement accuracy
under these conditions.

Calibration Check for LV (#1-3) and RV, HA, HB (#1-6) Brookfield Spindles

NOTE: The LV #4 (64) and RV, HA, HB #7 (07) spindles have been omitted from this
procedure. Brookfield AMETEK does not recommend the use of these spindles to
perform a calibration check on your instrument. Reasons pertain to the small amount of
spindle surface that makes contact with the viscosity standard, the difficulty of
establishing the immersion mark precisely and the need for precise temperature control at
25°C in the immediate vicinity of the spindle.

In principle, all Brookfield AMETEK rotational viscometer/rheometers are calibrated


torque meters. Brookfield AMETEK offers a variety of viscosity calibration fluids that
enable one to check the calibration and linearity of the instrument over its operating
torque range of 10%-100% torque.

Comments:_______________________________________________________________________________________

BEL representative:__________________________________ Verified by:____________________________________


Date: 3Qs008 Page 2 of 4
Any spindle and any one of several calibration fluids may be used to perform a
calibration check. We recommend using one spindle and one fluid while taking
measurements at multiple speeds. This enables multiple calculations of viscosity
throughout the torque range of the instrument: one at low torque, one at medium torque
and one at high torque. The FSR, and consequently the instrument accuracy in terms of
centipoise, will be different at each speed.

The calibration fluid must be transferred into a 600 mL low form, Griffin beaker (without
creating any bubbles) and, on LV and RV models, the instrument Guard Leg must be
used. See the manual for a description of the use of the Guard Leg. The beaker, fluid,
spindle and Guard Leg must be stabilized at 25.0°C (±0.1°C) before the calibration check
can be performed.
The example below shows typical calibration results using an LV #3 spindle with a
calibration fluid whose actual viscosity is 5,120 cP; the calibration check might look like
the table below. At each speed a different torque is sensed allowing the automatic
computation of viscosity. The allowable error is the sum of 1% of the FSR plus 1% of the
actual value of the calibration fluid. The lower and upper viscosity limits, indicating
acceptable calibration, equal the actual value of the fluid, in this case 5,120 cP; ± the
allowable error.

RPM MIN MAX cP 1% of % cP ALLOWABLE


cP (FSR) FSR Torque READING ERROR

12 1,000 10,000 100 51.2 5120 ± 151


6 2,000 20,000 200 25.5 5100 ± 251
3 4,000 40,000 400 12.7 5080 ± 451

Remember, the total allowable error when conducting a calibration check is the sum of
1% of the FSR plus 1% of the actual value of the calibration fluid itself. The instrument
reading must be within the limits defined by the actual viscosity value of the calibration
fluid ± the allowable error at each speed. The spindle, speeds and fluid you use may be
different from these.

Comments:_______________________________________________________________________________________

BEL representative:__________________________________ Verified by:____________________________________


Date: 3Qs008 Page 3 of 4
Calibration Test Results

A B C D E F G H I
Fluid 1% of Instrument Spindle FSR cP 1% of % Torque Viscosity cP
Value Fluid Model RPM (depress and hold FSR cP
from Value AUTORANGE key)
Certificate cP
cP

Interpretation of Calibration Test Results


The viscosity reading should equal the cP value on the fluid standard to within the
combined accuracies of the Viscometer/Rheometer and the viscosity standard as
calculated below:
Low Limit Reading High Limit Pass or Fail
= A – (B + G) I = A + (B + G)

Any reading outside these limits may indicate a Viscometer/Rheometer problem. Contact
the Brookfield AMETEK technical sales department or your local authorized Brookfield
AMETEK dealer with test results to determine the nature of the problem.

NOTE: A calibration workbook in Microsoft Excel format may be downloaded from


www.brookfieldengineering.com/support/calibration. This workbook contains several
worksheets, which greatly simplify use of the tables above. Instructions for use and
troubleshooting guide are also included.

Brookfield AMETEK recommends repeating this calibration check periodically to insure


Viscosity accuracy and calibration. We suggest every thirty days but your operating
procedures may dictate otherwise.

Comments:_______________________________________________________________________________________

BEL representative:__________________________________ Verified by:____________________________________


Date: 3Qs008 Page 4 of 4

You might also like