Product Information: Toshiba X-Ray Tube DF-161SBR

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Product Information PI

TOSHIBA X-RAY TUBE


DF-161SBR

Stationary Anode X-ray Tube

 Especially designed for mobile type surgical


X-ray TV apparatus.

 Provided with an insulation cylinder and lead


cylinder.

 This tube has focus 1.6, and 0.5 foci and is


available for maximum tube voltage 125 kV.

 Installed in the same enclosure with the high


voltage transformer.

General Data

Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center grounded
Nominal X-ray Tube Voltage (IEC60613:2010) ....................................................................... 125 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................... 1.6
Small Focus ............................................................................................................................... 0.5
Nominal Anode Input Power (at 1.0s) ..................................................................... See Rating Chart
Large Focus ........................................................................................................................ 4000 W
Small Focus ........................................................................................................................... 700 W
Nominal Radiographic Anode Input Power (IEC60613:2010):
Large Focus ........................................................................................................................ 4100 W
Small Focus ........................................................................................................................... 720 W
Exposure Duty Cycle:
Large Focus ........................................................................................................................... 1:40
Small Focus ........................................................................................................................... 1:20
(Exposure Time : interval time)

★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
DF-161SBR

Mechanical:
Dimensions:
Overall Length ......................................................................................... See dimensional outline
Maximum Diameter ................................................................................. See dimensional outline
Target:
Anode Angle ............................................................................................................... 16 degrees
Material .......................................................................................................................... Tungsten
Inherent Filtration .................................................................................... At least 0.8 mm Al at 50 kV
X-ray Coverage .................................................................................. 430 × 430 mm at SID 909 mm
Weight ....................................................................................................................... Approx. 1100 g
Cooling Method .............................................. Oil immersed (60℃ Max.) and convection oil cooling.
Tube Holding: ..................................................................................... Holding the insulation cylinder..

Absolute Maximum and Minimum Ratings


(At any time, these values must not be exceeded.)

Maximum X-ray Tube Voltage (IEC60613:2010) .................................................................... 125 kV


Between Anode (or Cathode) and Ground ............................................................................. 62.5 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current (IEC60613:2010):
Large Focus .......................................................................................................................... 60 mA
Small Focus ........................................................................................................................... 15 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 4.3 A
Small Focus .......................................................................................................................... 3.1 A
Filament Voltage: (At Maximum Filament Current)
Large Focus ................................................................................................................. 5.0 ~ 6.4 V
Small Focus ................................................................................................................. 3.5 ~ 4.9 V
Filament Frequency Limits ........................................................... DC or AC (Sine Wave) 0 ~ 20 kHz
Thermal Characteristics: (See Anode Thermal Characteristics)
Anode Heat Content ......................................................................................................... 35.5 kJ
Maximum Anode Heat Dissipation Rate ............................................................................. 600 W
Maximum Radiographic Exposure Time ..................................................................................... 10 s
Maximum Fluoroscopic Duration ........................................................................ 600 W - 20 minutes

Environmental Limits

Operating Limits (in dielectric oil):


Oil Temperature ............................................................................................................ 10 ~ 60℃
Oil pressure .............................................................................................................. 70 ~ 140 kPa
Shipping and Storage Limits:
Temperature ................................................................................................................ -40 ~ 70℃
Humidity ........................................................................................................................ 10 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa

-2-
DF-161SBR

Maximum Rating Charts


(Absolute maximum rating charts)

Constant Potential High-Voltage Generator

Nominal Focal Spot Value: 1.6

60 60kV 40kV
70kV
50kV
80kV
50
90kV
TUBE CURRENT [mA]

100kV
40
110kV
125kV
30

20

10

0
0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
EXPOSURE TIME [s]

Refer to IEC60613:2010

Nominal Focal Spot Value: 0.5

15 40kV
50kV

60kV
TUBE CURRENT [mA]

70kV
10
80kV
90kV
100kV
110kV
125kV
5

0
0.1 0.2 0.3 0.5 0.7 1 2 3 5 7 10
EXPOSURE TIME [s]

Refer to IEC60613:2010

-3-
DF-161SBR

Emission & Filament Characteristics


Constant Potential High-Voltage Generator

Nominal Focus Spot Value: 1.6

60 6
40kV
80kV
50 5.5

FILAMENT VOLTAGE [V]


TUBE CURRENT [mA]

40 5
125kV

30 4.5
Ef

20 4

10 3.5

0 3
3.3 3.5 3.7 3.9 4.1 4.3

FILAMENT CURRENT [A]

Note 1) This graph indicates typical characteristics.


Note 2) Refer to IEC60613:2010

Nominal Focus Spot Value: 0.5


15 5

40kV
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]

10 4
80kV

125kV

Ef
5 3

0 2
2.3 2.5 2.7 2.9 3.1

FILAMENT CURRENT [A]

Note 1) This graph indicates typical characteristics.


Note 2) Refer to IEC60613:2010

-4-
DF-161SBR

Anode Thermal Characteristics


40
COOLING
35 600W

30
HEAT STORAGE [kJ]

25

20 300W

15

10
100W
5
HEATING
0
0 60 120 180 240 300 360 420 480 540 600

TIME [s]

-5-
DF-161SBR

Dimensional Outline
Unit: mm

.05
.1 ±0
4-
O4

0.1
8 ±
2
O

M5 10DEEP
O 40

21
26

86 ±1.5
O 22 ±1
O 25 ±1
60

40

176 ±2.5

CENTRAL X-RAY

FOCAL SPOT
70
90

LEAD CYLINDER

O 60 ±1.5 INSULATION CYLINDER


O 64 ±1.5

K
L

7.3)
C ( O1

SHORT PIN L : LARGE FOFUS


S : SMALL FOCUS
S C : COMMON
N K : CATHODE
N : NOT FOR USE

-6-
DF-161SBR

OVERSEAS SUBSIDIARIES AND AFFILIATES

EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

For Sales & Technical Services, please contact the following representative:

・TOSHIBA ELECTRONICS EUROPE GMBH


HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (211) 5296-107 FAX +49 (211) 5296-402

・TOSHIBA AMERICA ELECTRONIC COMPONENTS, INC.


2150 EAST LAKE COOK ROAD, SUITE 310
BUFFALO GROVE, ILLINOIS 60089 USA
PHONE +1 (847) 484-2400 FAX +1 (847) 541-7287

・TOSHIBA ELECTRON DEVICES & MATERIALS (SHANGHAI) CO., LTD. (TEMS)


RM1606, SH-PLAZA,
No.336, XIZANG ROAD (MIDDLE), SHANGHAI, 200001, CHINA
PHONE +86 (21) 6361-0077 FAX +86 (21) 6351-5760

(Product Development Engineering Department)


1385 SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN, 324-8550, JAPAN
PHONE: +81-287-26-6666 FAX: +81-287-26-6060
http://www.toshiba-tetd.co.jp/

・Toshiba Electron Tubes & Devices Co., Ltd. meets the Environmental Management System Standard, ISO14001.
・Toshiba Electron Tubes & Devices Co., Ltd. meets internationally recognized Standards for Quality Management System ISO9001, ISO13485.

You might also like