Product Information: Industrial X-Ray Tube E7599
Product Information: Industrial X-Ray Tube E7599
Product Information: Industrial X-Ray Tube E7599
General Data
Electrical:
Circuit:
High Voltage Generator ...................................................................... Self-rectified, Pulse Voltage
(Pulse Width 1ms, Duty Factor 0.16)
Grounding ............................................................................................................ Anode-grounded
Operating Tube Voltage ................................................................................................. 160 ~ 300 kV
Focal Spot .................................................................................................................. 2.5 ± 0.75 mm
Filament Current ......................................................................................................... Maximum 4.0 A
Mechanical:
Dimensions .................................................................................................... See dimensional outline
Overall Length ................................................................................................................... 196 mm
Maximum Diameter ........................................................................................................ φ 132 mm
Target:
Anode Angle ................................................................................................................. 22 degrees
Construction ....................................................................................................................... Tungsten
Inherent Filtration ..................................................................................................... 1.0 mm Beryllium
X-ray Coverage (Angle) ......................................................................... 1) φ 350 mm at SID 600 mm
Weight (Approx.) ........................................................................................................................ 4.1 kg
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the third
parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
E7599
Note 1) 300kV is absolute maximum rating, tube voltage should not be exceeded 300 kV
when line voltage to the generator fluctuates.
2) Anode temperature is judged by the thermo paint around the X-ray port.
(See dimensional outline)
Environmental Limits
-2-
E7599
Filament Characteristics2000-10-10
E7599
1 ms pulse
Filament Chart
Focal Spot: 2.5 mm
10
8
FILAMENT VOLTAGE [V]
0
3.0 3.2 3.4 3.6 3.8 4.0
Emission Characteristics
2000-10-10
E7599
1Emission
ms pulse
Focal Spot: 2.5 mm
300
280
5mA (Average)
TUBE VOLTAGE [kV]
260
240
220
200
180
160
3.2 3.3 3.4 3.5 3.6 3.7
-3-
E7599
Dimensional Outline
Unit mm
22.5 φ116
°
Filament
4-φ8 Holes φ98
on φ30 Circle
φ80
Filament Cathode
φ76
26
11
10
30
8-φ3.5 Holes
4-M3 Bolts
196±5
on φ66 Circle
2-φ3.5 Holes
12
4
A
φ15
74±0.5
A-Section
Anode
M3 6 Deep
31
7
φ46±0.1
2.4 ±0.05
12.0 ±0.2
+0.25 -0.1
Thermo painting (156℃)
3.2 0 φ98-0.2 (At two points)
φ120 ±0.2
φ132±0.2 M8 12 Deep
φ132 ±0.2
O-Ring Seal Surface 30°
±0 .5゜
O-Ring JISB2401G120
6-M5 6 Deep
3
0
°
±
0
.
5
゜
TOSHIBA
E7599
3-M3 6 Deep
Type Name φ
1
1
0
±
0
.
2
:Direction of Main X-ray
-4-
E7599
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE GMBH
HANSAALLEE 181 40549 DÜSSELDORF, GERMANY
PHONE +49 (0) 211-5296-107 FAX +49 (0) 211-5296-402
For Sales & Technical Services, please contact the following representative:
Toshiba Electron Tubes & Devices Co., Ltd. meets Toshiba Electron Tubes & Devices Co., Ltd. meets
the Environmental Management System Standard, internationally recognized Standards for Quality
ISO14001. Management System ISO9001, ISO13485.