2-Wire Programmable Transmitter
2-Wire Programmable Transmitter
2-Wire Programmable Transmitter
5331D
– RTD, TC, Ohm, or mV input
– Extremely high measurement accuracy
– 1.5 kVAC galvanic isolation
– Programmable sensor error value
– For DIN form B sensor head mounting
Application
Applications
• Linearized temperature measurement with Pt100...Pt1000,
Ni100...Ni1000, or TC sensor.
• Conversion of linear resistance variation to a standard analog
current signal, for instance from valves or Ohmic level
sensors.
• Amplification of a bipolar mV signal to a standard 4...20 mA
current signal.
Technical characteristics
• Within a few seconds the user can program PR5331D to
measure temperatures within all ranges defined by the norms.
• The RTD and resistance inputs have cable compensation for
2-, 3- and 4-wire connection.
• Continuous check of vital stored data for safety reasons.
Mounting / installation
• For DIN form B sensor head mounting.
This page is automatically generated on the basis of information provided on www.prelectronics.com and affiliated websites. It is provided to you as a service and for information
purpose only. While we have attempted to maintain the information as accurately as possible, the page may contain errors or omissions for which we disclaim any and all liability
Environmental Conditions Min. measurement range (span).................... 5 mV
Input resistance.............................................. 10 MΩ
Operating temperature................................... -40°C to +85°C
Calibration temperature................................. 20...28°C Output specifications
Relative humidity............................................ < 95% RH (non-cond.)
Protection degree (encl./terminal).................. IP68 / IP00 Current output
Signal range................................................... 4…20 mA
Mechanical specifications Min. signal range............................................ 16 mA
Load (@ current output)................................. ≤ (Vsupply - 7.2) / 0.023 [Ω]
Dimensions.................................................... Ø 44 x 20.2 mm Load stability.................................................. ≤ 0.01% of span / 100 Ω
Weight approx................................................ 50 g Sensor error indication................................... Programmable 3.5…23 mA
Wire size........................................................ 1 x 1.5 mm2 stranded wire NAMUR NE43 Upscale/Downscale............... 23 mA / 3.5 mA
Screw terminal torque.................................... 0.4 Nm
Vibration......................................................... IEC 60068-2-6 Common output specifications
2...25 Hz......................................................... ±1.6 mm Updating time................................................. 440 ms
25...100 Hz..................................................... ±4 g
of span........................................................... = of the presently selected
range
Common specifications
Supply I.S. / Ex marking
Supply voltage............................................... 7.2...30 VDC
ATEX.............................................................. II 1 G Ex ia IIC T4...T6 Ga, II 1
Internal power dissipation.............................. 25 mW…0.7 W D Ex ia IIIC Da, I M1 Ex ia Ma
Isolation voltage IECEx............................................................. Ex ia IIC T4...T6 Ga, Ex ia IIIC
Da, Ex ia I Ma
Isolation voltage, test /
working........................................................... 1.5 kVAC / 50 VAC FM, US........................................................... Cl. I, Div. 1, Gp. A, B, C, D
T4/T6; Cl. I Zone 0, AEx ia IIC
T4/T6; Cl. 1, Div. 2, Gp. A, B,
Response time C, D, T4/T6
Response time (programmable).................... 1...60 s CSA................................................................ Cl. I, Div. 1, Gp. A, B, C, D Ex
Voltage drop................................................... 7.2 VDC ia IIC, Ga
Warm-up time................................................. 5 min. INMETRO....................................................... Ex ia IIC T6...T4 Ga, Ex ia IIIC
Da
Programming................................................. Loop Link
Signal / noise ratio......................................... Min. 60 dB
EEprom error check....................................... < 3.5 s
Observed authority requirements
Accuracy........................................................ Better than 0.05% of selected EMC............................................................... 2014/30/EU
range RoHS.............................................................. 2011/65/EU
Signal dynamics, input................................... 20 bit EAC................................................................ TR-CU 020/2011
Signal dynamics, output................................. 16 bit
Effect of supply voltage change..................... < 0.005% of span / VDC Approvals
EMC immunity influence................................ < ±0.5% of span DNV-GL Marine............................................. Stand. f. Certific. No. 2.4
Extended EMC immunity: NAMUR ATEX 2014/34/EU.......................................... KEMA 06ATEX0062X
NE21, A criterion, burst.................................. < ±1% of span
IECEx............................................................. DEK 13.0035X
FM.................................................................. FM17US0013X
Input specifications CSA................................................................ 1125003
Common input specifications INMETRO....................................................... DEKRA 16.0013 X
Max. offset...................................................... 50% of selected max. value EAC Ex TR-CU 012/2011.............................. RU C-DK.GB08.V.00410
RTD input
RTD type........................................................ Pt100, Ni100, lin. R
Cable resistance per wire.............................. 5 Ω (max.)
Sensor current............................................... Nom. 0.2 mA
Effect of sensor cable resistance
(3-/4-wire)....................................................... < 0.002 Ω / Ω
Sensor error detection................................... Yes
TC input
Thermocouple type........................................ B, E, J, K, L, N, R, S, T, U,
W3, W5, LR
Cold junction compensation
(CJC).............................................................. < ±1.0°C
Sensor error detection................................... Yes
5331D-062719
Voltage input
Measurement range....................................... -12...800 mV