Product Information: Rotanode™ E7254X E7254FX E7254GX Rotating Anode X-Ray Tube Assembly
Product Information: Rotanode™ E7254X E7254FX E7254GX Rotating Anode X-Ray Tube Assembly
Product Information: Rotanode™ E7254X E7254FX E7254GX Rotating Anode X-Ray Tube Assembly
ROTANODE™
E7254X
E7254FX
E7254GX
These tubes have foci 1.2 and 0.6, and are available for a
maximum tube voltage 150kV.
General Data
Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage:
Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Nominal Focal Spot Value:
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s):
180 Hz 60 Hz 50 Hz
Large Focus .................................................................................... 102 kW 60 kW 55 kW
Small Focus ...................................................................................... 40 kW 23 kW 21.5 kW
Nominal Radiographic Anode Input Power:
180 Hz 60 Hz 50 Hz
Large Focus ...................................................................................... 82 kW 47 kW 43 kW
Small Focus ...................................................................................... 32 kW 18 kW 17 kW
★The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Canon Electron Tubes
& Devices Co., Ltd. (CETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of CETD or others.
★The information contained herein may be changed without prior notice. It is therefore, advisable to contact to CETD before processing with the design of
equipment incorporating this product.
E7254X / E7254FX / E7254GX
Motor Ratings:
Stator: XS-RB
Starting Running
Driven Frequency [Hz] 180 60 180 60
Input Power [W] 3500 1520 90 80
Voltage 2) 4) [V] 490 300 165 160
Current 3) [A] 9.0 7.6 1.3 1.1
Min. Speed Up 6) [s] 2.0 1.0 - -
Capacitor [μF] 6 30 6 30
Min. Braking 2) 6) [s] 2.5 s / 300 V (DC)
Note: 1) To be applied for high speed rotation.
2) Applied voltage between common and main terminal.
3) Common current.
4) The every applied voltage must be never exceeded 110% of the above specification.
5) No more than two high speed starts per minute are permissible.
6) The speed-up and braking time are allowed up to 110% of the above specification.
Anode Speed:
180 Hz ............................................................................................................ Minimum 9700 min-1
60 Hz ............................................................................................................ Minimum 3200 min-1
50 Hz ............................................................................................................ Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ...................................................................................................... 20.2 Ω
Common-Auxiliary Winding ................................................................................................ 38.0 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 ℃
Thermal Switch ........................................................................................................ Normally Closed
Open ............................................................................................................................ 75 ~ 85 ℃
Closed .......................................................................................................................... 45 ~ 65 ℃
Mode of Operation ........................................................................................................... Intermittent
Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 463 mm
Maximum Diameter .......................................................................................................... 172 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter .......................................................................................................................... 100 mm
Construction .................................................................... Rhenium-Tungsten faced molybdenum
Filtration:
Permanent Filtration ............................................................. 0.8 mm Al / 75 kV IEC60522:1999
Available Additional Filter combination (0.4 - 1.5 mm) ..................... Maximum 2.3 mm Al / 75 kV
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor ..................................................................................... 150 kV, 5 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 20 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number ................................................................................................ XH-157
-2-
E7254X / E7254FX / E7254GX
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-3-
E7254X / E7254FX / E7254GX
125kV
40kV
800
150kV
TUBE CURRENT [mA]
600
400
200
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
110kV 60kV
400
125kV
TUBE CURRENT [mA]
50kV
150kV
300 40kV
200
100
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
-4-
E7254X / E7254FX / E7254GX
100kV
600 110kV
TUBE CURRENT [mA]
125kV
150kV
400
200
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
100kV
110kV
125kV
200
150kV
100
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
-5-
E7254X / E7254FX / E7254GX
90kV
600 100kV
TUBE CURRENT [mA]
110kV
125kV
400 150kV
200
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
60kV 50kV
70kV
40kV
300 80kV
TUBE CURRENT [mA]
90kV
100kV
110kV
200 125kV
150kV
100
0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7
-6-
E7254X / E7254FX / E7254GX
1000 16
100kV
900 125kV
800 150kV 14
700
600 12
Ef
500
400 10
300 80kV
60kV
200 8
40kV
100
0 6
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.5
FILAMENT CURRENT [A]
500 12
100kV
125kV
400 10
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]
150kV
300 8
Ef
200 6
80kV
100 4
60kV
40kV
0 2
3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2
-7-
E7254X / E7254FX / E7254GX
Thermal Characteristics
950
200W
900
800
HEAT STORAGE [kJ]
700
600
500 COOLING
400
300
200
100 HEATING
0
0 60 120 180 240 300 360
TIME [min]
285
280
1180W
240 COOLING
1000W
HEAT STORAGE [kJ]
200
600W
160
120
80
300W
40
HEATING
0
0 2 4 6 8 10
TIME [min]
-8-
E7254X / E7254FX / E7254GX
9 8 7 6 5 1 2 3 4 E
69
97.1
φ68.8
NC NC NC
56
C1 M A
ET
TEMPERATURE RELAY
70 121 2-M63.5 P1.27 (NORMALLY CLOSED)
Note
1) Make an input-power protection circuit with the terminals No.5 and No.6.
2) Do not connect terminal No.1 and No.5 or 6 in series circuit.
φ134
ANODE
CENTRAL RAY AND
REFERENCE AXIS
10
4-M6 8 DEEP
12°
41.27
φ
287
148
463
89
92
.1
4-M5 9 DEEP
10
φ172 C
CATHODE
S L
EXPLANATION OF SYMBOLS
CATHODE TERMINAL
C : COMMON
B
L : LARGE FOCUS
S : SMALL FOCUS
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS -1.5mm≦B≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE
-9-
E7254X / E7254FX / E7254GX
121 70
TERMINAL CONNECTIONS
φ68.8
PRESSURE DETECTIVE RELAY
(NORMALLY CLOSED)
1 2 3 4 E
69
9 8 7 6 5
2-M63.5 P1.27
NC NC NC
C1 M A
96
FOCAL SPOT
56
ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
11
Note
1) Make an input-power protection circuit with the terminals No.5 and No.6.
2) Do not connect terminal No.1 and No.5 or 6 in series circuit.
φ134
4-M6 8 DEEP
CENTRAL RAY AND REFERENCE AXIS
10
12°
41.27
φ
287
463
148
89
92
.1
45°
4-M5 9 DEEP
C φ172
10
S L
EXPLANATION OF SYMBOLS
CATHODE TERMINAL
C : COMMON
B
L : LARGE FOCUS
S : SMALL FOCUS
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS -1.5mm≦B≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE
- 10 -
E7254X / E7254FX / E7254GX
121
9 8 7 6 5 1 2 3 4 E
NC NC NC
C1 M A
ET
70
97.1
TEMPERATURE RELAY
(NORMALLY CLOSED)
56
Note
1) Make an input-power protection circuit with the terminals No.5 and No.6.
2) Do not connect terminal No.1 and No.5 or 6 in series circuit. FOCAL SPOT
11
CENTRAL RAY AND REFERENCE AXIS
φ134
ANODE
4-M6 8 DEEP
10
12°
41.27
φ
463
148
287
89
92
.1
4-M5 9 DEEP
φ172
C
10
CATHODE
L S
EXPLANATION OF SYMBOLS
CATHODE TERMINAL
C : COMMON
B
L : LARGE FOCUS
S : SMALL FOCUS
-1.5mm≦A≦1.5mm
TERMINAL CONNECTIONS -1.5mm≦B≦1.5mm
C1 : COMMON
M : MAIN WINDING OF THE STATOR
A : AUX. WINDING OF THE STATOR : CENTRAL X-RAY
NC : NON-CONNECTION ANODE & CATHODE TERMINAL
ET : EARTH TERMINAL : IEC60526 TYPE
- 11 -
Marketing Engineering Group, Sales Department
1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
https://etd.canon
・The head office of Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
・Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. https://etd.canon/eng/company/quality.htm.
PE-E7254X/FX/GX 2018-11