Product Information: Rotanode™ E7100X
Product Information: Rotanode™ E7100X
Product Information: Rotanode™ E7100X
ROTANODE™
E7100X
General Data
Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding ......................................................................................................... Center-Grounded
Nominal X-ray Tube Voltage:
Radiographic ...................................................................................................................... 150 kV
Fluoroscopic ....................................................................................................................... 125 kV
Nominal Focal Spot Value:
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s) ..................................................................... See rating charts
180 Hz 60 Hz 50 Hz
Large Focus .................................................................................. 100 kW 59 kW 55.5 kW
Small Focus .................................................................................... 40 kW 24 kW 21.5 kW
Nominal Radiographic Anode Input Power:
180 Hz 60 Hz 50 Hz
Large Focus .................................................................................... 80 kW 48 kW 44 kW
Small Focus .................................................................................... 30 kW 18 kW 17 kW
★The information contained herein is presented only as a guide for the application of our products. No responsibility is assumed by Canon Electron Tubes
& Devices Co., Ltd. (CETD) for any infringements of patents or other rights of the third parties which may result from its use.
No license is granted by implication or otherwise under any patent or patent rights of CETD or others.
★The information contained herein may be changed without prior notice. It is therefore, advisable to contact to CETD before processing with the design of
equipment incorporating this product.
E7100X
Motor Ratings:
Stator: XS-AG
Starting Running
Driven Frequency [Hz] 180 60 180 60
Input Power [W] 2800 800 200 50
Voltage 3) 5) [V] 352 155 80 38
Current 4) [A] 9.6 10.2 3.0 2.2
Min. Speed Up 1) 7) [s] 1.2 0.8 - -
Capacitor [µF] 6 44 6 44
Min. Braking 2) 7) [s] 3.0 (DC 80V)
Note 1) The speed up time from normal speed to high speed is 2/3 times of the specified speed
up time from 0 to high speed, which is described on motor rating table.
2) To be applied for high speed rotation.
3) Applied voltage between common and main terminal.
4) Common current.
5) The every applied voltage must be never exceeded 110% of the above specification.
6) No more than two high speed starts per minute are permissible.
7) The speed-up time is allowed up to 110% of the above specification.
Anode Speed:
180 Hz .......................................................................................................... Minimum 9700 min-1
60 Hz ........................................................................................................... Minimum 3200 min-1
50 Hz ........................................................................................................... Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ........................................................................................................ 9.4 Ω
Common-Auxiliary Winding ................................................................................................ 28.3 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 °C
Mode of Operation ........................................................................................................... Intermittent
Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 496 mm
Maximum Diameter .......................................................................................................... 195 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter .......................................................................................................................... 100 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ................................................................. 1.1 mm Al / 75 kV IEC60522:1999
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor .................................................................................. 150 kV, 3.4 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 24 kg
High Voltage Receptacle ..................... To meet the requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number ............................................................................................. XH-112V
-2-
E7100X
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 °C
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature ............................................................................................................... -20 ~ 70 °C
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-3-
E7100X
125kV
600 150kV
300
40kV
500
400 200
300
200 100
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
125kV 40kV
600 125kV
250
150kV
500 150kV
200
400
150
300
100
200
100 50
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
-4-
E7100X
300 150
200 100
100 50
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
90kV
TUBE CURRENT [mA]
110kV
400 200 100kV
125kV 110kV
125kV
300 150kV 150
150kV
200 100
100 50
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
-5-
E7100X
110kV 90kV
500 125kV 100kV
110kV
400 150kV 200 125kV
150kV
300
200 100
100
0 0
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.02 0.05 0.1 0.3 0.7 0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s] EXPOSURE TIME [s]
50
100
0
0 0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10
0.01 0.03 0.07 0.2 0.5 1 2 3 5 7 10 0.02 0.05 0.1 0.3 0.7
0.02 0.05 0.1 0.3 0.7
EXPOSURE TIME [s]
EXPOSURE TIME [s]
-6-
E7100X
40kV
800 50kV 24
80kV
600 100kV 20
125~150kV
400 16
200 12
Ef
0 8
4.5 4.7 4.9 5.1 5.3 5.5 5.7 5.8
500 16
60kV
400 80kV 14
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]
100kV
300 125~150kV 12
200 10
Ef
50kV
100 8
40kV
0 6
4.0 4.2 4.4 4.6 4.8 5.0 5.2
-7-
E7100X
Thermal Characteristics
1070
1000 COOLING 215W
900
800
HEAT STORAGE [kJ]
700
600
500
400
300
200
100 HEATING
0
0 60 120 180 240 300
TIME [min]
210
200
180 COOLING
710W
160
HEAT STORAGE [kJ]
140
120
450W
100
80 300W
60
200W
40
20
HEATING
0
0 2 4 6 8 10
TIME [min]
-8-
E7100X
Dimensional Outline
Unit: mm
φ174
ANODE
4-M5 9 DEEP
φ195
244
CENTRAL RAY AND REFERENCE AXIS
164
12°
FOCAL SPOT POSITION
16.5
41.27
149
90
496
φ
92
45°
145
4-M6 8 DEEP
S
16.5
252
CATHODE
C
TERMINAL CONNECTIONS
9 8 7 6 5 1 2 3 4 E
FOCAL SPOT
60±1
NC NC NC NC
C1 M A
100
ET
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect termial No.1 and No.5 or No.6
57
in series circuit.
100
φ70
85
B
210 2-M63.5,P=1.25
-1.5mm≦A≦1.5mm
-1.5mm≦B≦1.5mm
EXPLANATION OF SYMBOLS
CATHODE TERMINAL
C : COMMON
L : LARGE FOCUS
S : SMALL FOCUS
TERMINAL CONNECTIONS
C1 : COMMON : CENTRAL X-RAY
M : MAIN WINDING OF THE STATOR ANODE & CATHODE TERMINAL
A : AUX. WINDING OF THE STATOR : IEC60526 TYPE
NC : NON-CONNECTION
ET : EARTH TERMINAL
-9-
Marketing Engineering Group, Sales Department
1385, Shimoishigami, Otawara-shi, Tochigi 324-8550, Japan
Tel: +81-287-26-6666 Fax: +81-287-26-6060
https://etd.canon
・The head office of Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Environmental Management System ISO14001.
・Canon Electron Tubes & Devices Co., Ltd. has been certified to meet all the requirements of Quality Management Systems ISO9001 and ISO13485.
Product scope is referred to the following URL. https://etd.canon/eng/company/quality.htm.
PE-E7100X 2018-11