XY-6S-10 English Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

OXYGEN CONCENTRATOR

Product Manual

Model: XY-6S-10

Manual

Please read this instruction manual carefully before using this product.
The company reserves the right to interpret the instructions
The appearance of the product is subject to the actual product
TABLE OF CONTENTS
Foreword······································
·····
···
·· 2
···
Important safety precautions·······························
·3
Related usage symbols·································
··7
Product packing List ··································
····
····
·8
Product operation ··································
····
8
Product introduction ··································
···10
Structure and composition·······························
··· 11
···
Scope of application ·································
···
·· ·11
···
Contraindications···································
···
···
·11
Product use environment requirements··························
··12
Transportation and storage conditions···························
··12
Additional explanations·································
·· 12
···
Oxygen Concentration Value and Flow Function Charts··············13
Product Specifications·································
···
···14
Product part name and function description························
16
Product Cleaning and Maintenance Instructions····················19
Product Warranty···································
···
····
··20
Fault and repair checklist·······························
···
···
·21
Product return step requirements····························
····
·23
Period of use / Date of manufacture···························
···
·24
Environmental Protection·······························
····
··24
·
List of components··································
·····
··25
Product Warranty Card/Qualification Certificate·····················27

1
FOREWORD

Respected user:
Thank you for purchasing our product! I hope that you will become a satisfied user of our
products.
This manual describes the product's precautions, operating procedures, basic functions
and technical parameters, as well as basic troubleshooting, returning service instructions, etc.,
will make you familiar with the product and the operation of the product.
In order to ensure the effective use of this machine, please be sure to read the instructions
in this manual carefully before use. Before you operate the machine, please confirm that you
have read and understood the basic operation of this product, please pay special attention to
all the "Important Safety Notices".
If the product needs to be returned to the factory for repair, please refer to the “Product
Return Step Requirements” on the following page, and return the product according to the
required content.
Please note that some of the illustrations in this manual may not be exactly the same as
what you see on the machine. Please refer to the actual product.
For the random components such as the humidification bottle, oxygen pipe/oxygen mask,
filter, etc. mentioned in this manual, the products that meet the specifications of the supporting
products should be selected.
If you have any questions or other inquiries, please contact the supplier or the
manufacturer's after-sales service.

Version: V0 Date of preparation: July 2019

2
IMPORTANT SAFETY PRECAUTIONS

DANGER (The following items may cause serious injury or death if they violate
the operation)

◆ Measures to reduce the risk of burns, electric shock, fire or human injury.Avoid using an
oxygen concentrator when taking a bath.If you need to use it, please follow the doctor's
instructions, and the oxygen concentrator must be used in another room 2.5 meters
away.
The recommended oxygen pipe should be no more than 11 meters long and cannot be
folded.
Do not place or store the oxygen concentrator in a location where it is easy to drip into
water or other liquids.
Oxygen has a combustion-supporting effect, and the oxygen concentrator should be kept
away from open flames and fire sources when working.When there are objects with high
temperature, sparks or open flame within 1.6 meters, please do not use the oxygen
concentrator. Do not look after the machine after the oxygen concentrator is energized.
◆ Do not smoke while inhaling oxygen. It is strictly forbidden to smoke or ignite an open
flame near oxygen inhalers.
◆ Before cleaning the dust of the oxygen concentrator casing, the power plug must be
removed to prevent electric shock.
◆ When the machine is running, do not open the outer casing and inner casing of the
machine at will, to prevent injury caused by touching the running parts.

Warning (The following items must be strictly enforced, otherwise it may cause
serious consequences)
◆ In order to prevent the oxygen concentrator from malfunctioning or encountering power
outages, oxygen users (such as those in need of oxygen and critically ill patients) need to
be equipped with other spare oxygen supply devices (such as oxygen cylinders, oxygen
bags, etc.).
◆ This product cannot be used to maintain any life. It is recommended that patients with

3
aerobic therapy follow the instructions of your physician when using this machine for flow
and oxygen inhalation time.
◆ If the patient develops or exhibits an uncomfortable or abnormal reaction while inhaling
oxygen, stop using the product immediately and contact the device supplier or doctor.
◆ Patients with severe illnesses need to be equipped with additional pointing devices when
using this product, or need to be administered separately. Please consult a doctor before
use and follow the doctor's advice.If you have an adverse reaction, please let your doctor
know immediately.
◆ If the product is used in places where the sea level is more than 2000 meters, the oxygen
concentration output will be lower than 90% at the rated maximum flow rate.
◆ Do not share the same oxygen breathing tube or oxygen mask for many people, so as to
avoid cross-infection of viruses or bacteria between users.
◆ Wet bottles should be used in accordance with the specifications of the supporting
products. Do not replace them at will, otherwise it may cause harm to oxygen or oxygen.
◆ Do not press the oxygen connection tube under the bed cover or the seat cushion during
oxygen absorption. Do not sit on the oxygen connection tube to prevent oxygen from
passing through the connection tube properly, resulting in the inability to absorb oxygen.
◆ When the product is operated by no oxygen, the oxygen generated will help
combustion.When no one is breathing oxygen, it is recommended to turn off the power
switch and disconnect the plug.
◆ When using this product, pay attention to whether the power cord is too long to interfere
with the passage of people. It is recommended that the machine be placed next to the
nearest power outlet.
◆ This product cannot be used with non-designated humidification bottles or drug delivery
accessories to avoid affecting product performance.

Be careful (please note the following to prevent damage to the product)

◇ Do not open the casing and chassis of the machine for maintenance. If the user finds any

4
quality problems or finds an alarm, etc., please do not disassemble and repair it. Please
contact the dealer or manufacturer of this product immediately.
◇ The oxygen concentrator should be placed in an environment free of dust, smoke, and
corrosion and poisonous gases.Do not use this product under strong magnetic field
conditions.
◇ The air inlet of the oxygen concentrator should be located in a well ventilated area, and
the air source inlet should be in the place with the least amount of pollutants
(contaminants / including: burning exhaust gas, anesthesia other discharge systems,
vents and vacuum exhaust ports, etc.).
◇ Please use a separate (wall mounted) AC power outlet to use this product. Do not
connect the power strip to the board.Do not use an additional power outlet with other
appliances.
◇ This machine is strictly prohibited from contact with oil and grease. If you need to connect
other pipes, valves and joints as needed, this type of connection should be cleaned
before installation to ensure clean and oil-free (fat).During the connection process, be
careful to keep the parts clean. The connected whole machine must not touch any
flammable oily liquid.
◇ When using the machine, ensure that the bottom exhaust is unblocked. Do not place the
back on the wall at least 30 cm away from the wall. Otherwise, the machine will be
overheated and the casing will be damaged.
◇ The machine can not be turned on and off frequently, and then turned on after turning off.
The interval should not be less than 5 minutes, so as not to affect the life of the
compressor. The manufacturer recommends that the oxygen concentrator should run no
less than 30 minutes each time.
◇ Do not use excessive force when rotating the flow adjustment knob, otherwise the valve
plug may be damaged.When the flow adjustment knob is turned on to the highest, but the
flow indication is zero, shut down immediately and check the fault.
◇ Before receiving the product, you need to cut off the cable tie at the bottom of the
machine and turn it on again. Otherwise, the machine will be damaged.

5
Note (you must pay special attention to the information)

◇ When the machine is turned on, the machine maintains the rated flow rate for 10 minutes
and the oxygen concentration reaches 90% or more.
◇ The oxygen inhalation tube and oxygen mask that are in contact with the patient need to
be cleaned, disinfected and sterilized when used repeatedly.
◇ Among the components of the oxygen concentrator, the humidification bottle needs to be
cleaned every 2 to 3 days.The air intake filter is recommended to be replaced after 1500
hours of use.If the dust or smoke of the product is large, reduce the interval between
replacement parts.In order to ensure the use of the effect, it is recommended to replace
the above parts early.
◇ Diluted water or cold boiled water can be used for the humidification bottle. Do not inject
untreated tap water directly.The water in the humidification bottle should be replaced
once every 2~3 days. It is recommended to replace it every day in summer. If the product
is not used for several days after use, please pour off the water and dry the wet bottle for
use.
◇ Oxygen pipe or oxygen mask should be selected for the specifications of the matching
products. If you use other types of oxygen or oxygen mask, please consult a professional
doctor or supplier of this product.
◇ Do not discard the humidification bottle, oxygen pipe or oxygen mask, and the matching
parts, and send them to a nearby medical waste disposal facility for disposal.
◇ Please use the humidification bottle that is compatible with this machine and fix it on the
body as required. Please do not use it under the state of disassembly.
◇ When using a humidified bottle, pay attention to the water at the highest and lowest water
level of the bottle.
◇ When the unit is scrapped, please contact your local supplier or manufacturer.
◇ Please refer to the "Related Symbols" for instructions on the symbols in the machine,
package and manual.
◇ When you receive the product to open the packaging equipment, please refer to the
“Unpacking Inspection Package and Product Packing List” to check the accessories that
are delivered randomly, and follow the “Product Operation Procedure”.

6
RELATED USAGE SYMBOL DESCRIPTION

OFF/ON Off / on 220v50Hz/110v60Hz AC 220V/110V

No smoking Date of manufacture

Safety instructions Serial number

Precautions Reference instructions

Class II equipment No open flames

This item cannot be


Type b application part
discarded after use.

Avoid rain
Place up
Keep dry

Fragile items Stacking layer limit

UNPACKING INSPECTION PACKAGE AND PRODUCT PACKING LIST

7
When you receive the product, please open and check the package carefully.This product
is equipped with an upper and lower foam cover. If the cover is damaged, please check the
product immediately. Then, check the product packing list to see if there are missing
accessories.

Product packing List

Serial Product Name / Accessories Name Quantity unit


number
1 oxygen concentrator 1 station

2 humidification bottle 1 One

3 humidification bottle connection tube 1 article

4 Intake filter 1 One

5 Nasal tube 1 article

6 Atomization kit (for atomizer models only) 1 set

7 Product manual (including warranty card, 1 this


certificate)

PRODUCT USE STEPS

1、 When using the product, the back should be at least 30 cm away from the wall. At the
same time, the foam board and other things must be removed, and the bottom must be
kept free of debris to keep the air circulation of the machine and the normal heat
dissipation of the machine.(The new machine is used for the first time, please loosen or
cut off the fixing strap at the bottom of the machine)
2、 Remove the humidification bottle, unscrew the bottle cap counterclockwise, and inst
all cold water or distilled water. The water level must be as indicated. Do not Abo
ve/ below the height/low indicator line of the bottle. Close the bottle and screw the
bottle clockwise. As shown in Figure 1;

8
Maximum water level (maximum)

Minimum water level (minimum)

Figure 1

2. Put the humidification bottle into the fixed seat on the back of the machine, connect one
end of the wet bottle connection tube to the humidification bottle inlet port, and the other end
to the oxygen outlet of the oxygen concentrator, as shown in Figure 2.

Figure 2

3、 Plug the AC power cord of the machine into the indoor power socket, turn on the
power, and turn on the power switch on the operation panel to enter the standby
mode. Press the “Oxygen” button on the operation panel of the machine, and the
machine will start with a slight nitrogen discharge and slight jitter. .
4、 Adjust the flow meter according to the required oxygen output.The flow adjustment
knob is turned counterclockwise to increase the flow rate, and clockwise to decre

9
ase the flow rate.(Oxygen concentration is 90% at rated flow rate)

5、 For the oxygen concentrator, we have equipped a monitoring device that outputs o
xygen concentration. Each time the machine is turned on, the power light on the d
isplay lights up normally. After a few minutes, if the oxygen concentration of the o
utput does not meet the standard requirements (at rated flow rate) ), yellow hypoxi

c warning light on the display Will light up.

6、 Insert the inlet end of the nasal oxygen tube (the shape of the bell mouth) into th
e oxygen outlet of the humidification bottle, as shown in Figure 3. Then place the
oxygen suction tube on the ear of the oxygen inhaler and insert the nasal plug on
the oxygen inhalation tube. Oxygen is absorbed in the nostrils of oxygen, see Fig
ure 4.

Figure 3 Figure 4

7、 When the machine is running, press the “Sleep” button, the machine will enter into
sleep state with the current set flow rate and timing time, and the display will be
off, so as to avoid the blue light that affects the user's sleep. Press the “Sleep” b
utton again,the machine display is on and the status is back to normal.

8、 When the machine is running, the machine running time can be set through the o
peration panel “Time+” and “Time-” keys. The single operation timing time is accu
mulated or decremented by 15 minutes.
10
9、 When the machine is running, if an alarm sounds, check if the power connection i
s loose or if the external power supply has been interrupted.

10. When the oxygen is finished, press the “Oxygen” button on the operation panel an
d turn the power switch after three minutes.Follow the prompts to clean the oxygen pip
e or oxygen mask.When not in use frequently, tie the AC power cord to the strap and
place the unit.

PRODUCT INTRODUCTION

Figure 5
A oxygen concentrator is a machine that provides low-flow oxygen therapy at home, in a
nursing home, or at a patient care center.The Portable oxygen concentrator adopts the
principle of pressure swing adsorption to pressurize and transform air. The molecular sieve
separates the nitrogen and oxygen molecules in the air, and the oxygen is retained and the
nitrogen is discharged.When the power is turned on at normal temperature, oxygen that meets
standards can be continuously separated from the air.

Structure and composition

This product consists of oxygen concentrator, flow meter and humidification bottle.

Scope of application

This device is mainly suitable for the preparation of oxygen.

11
Contraindications

no

Product use environment requirements

1. Ambient temperature: 5 ° C ~ 40 ° C;
2. Relative humidity: ≤ 80%;
3. Atmospheric pressure: 86 kpa~106 kpa;
4. The surrounding environment is free of corrosive gases and strong magnetic fields.

Transport and storage conditions

Ambient temperature range: - 20 ° C ~ +55 ° C;


Relative humidity range: ≤ 93%, and no condensation phenomenon;
Atmospheric pressure range: 50 kPa~106 kPa;

Note: When the storage temperature is lower than 5 °C, the equipment should

be placed in the normal working temperature environment for more than four hours
before use.
Supplementary explanation

Classification of electric shock prevention types: Class II equipment;


Classified according to the degree of protection against electric shock: b-type application part;
Classified according to the degree of protection of the incoming liquid: ipx0 equipment;
Classification according to the degree of safety when using flammable anesthetic gas mixed
with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: not flammable
anesthetic gas mixed with air or with oxygen or nitrous oxide Equipment used in the case of
mixed flammable anesthetic gases;
Classified by operating mode: continuous operation;
No signal output and signal input part;
Use voltage: ~220V, 50Hz;
Working system: This machine is a continuous operation system;
12
The oxygen concentrator has no application part that protects against defibrillation discharge
effects;
Oxygen concentrators are non-permanent installation equipment.

OXYGEN CONCENTRATION VALUE AND FLOW FUNCTION CHART


Oxygen concentration value and flow function chart when the nominal pressure of the Portable
oxygen concentrator is zero (see Figure 6)
This function table is for models with model XY-6S-10 rated flow rate of 10L/min.

Figure 6

13
PRODUCT SPECIFICATIONS

model XY-6S-10

Working voltage (V/Hz) ~220 ± 10% / 50 ± 1 / 110 ± 10% / 60 ± 1

Rated power (w) 600

Flow range at nominal


pressures of zero and 7 kPa 0.5~10
(L/min)

Oxygen concentration at the


nominal zero pressure of
the outlet (within the initial When the oxygen flow rate is 0.5~10L/min, the oxygen
start-up within 10 min, the concentration is ≥90%
specified concentration level
is reached)

Maximum recommended
10 L/min
flow rate (rated flow rate)

At the maximum
recommended flow rate, a
back pressure of 7 kPa is ≤0.5 L/min
applied and the flow rate
changes.

Oxygen concentration at the


recommended maximum
flow rate (within the initial
≥90%
start-up within 10 min, the
specified concentration level
is reached)

14
Flow adjustment range Continuously adjustable from 0~10L/min

Stand-alone net weight (Kg) 23

Machine noise dB(A) ≤55

Dimensions (mm) Length 380* wide 330* high 600

Oxygen output pressure 40-60kPa

At rated flow rate, when the oxygen concentration is ≤82% (±


3%), the yellow hypoxic indicator light is displayed. Please
Hypoxic indicator
turn off the machine immediately, use the spare oxygen, and
contact the supplier or manufacturer immediately.

Compressor safety valve release pressure: 250kPa ± 50 kPa


Oxygen outlet temperature: ≤46°C
When the altitude is between 0 and 2000 meters, the oxygen concentration is ≥90%, and the
efficiency is less than 90% from 2001 meters to 4,000 meters.

PRODUCT AND COMPONENT NAME AND FUNCTION DESCRIPTION

* Description of the following parts*

15
1. Oxygen outlet: Connect to the humidification bottle or nasal oxygen tube with a
humidification bottle connection tube
2. Atomizing port: use of atomizing function (only for atomizing machine type): Add
appropriate amount of liquid to the atomizing cup (according to the doctor's advice), do
not exceed the maximum scale line.Before starting the machine, unscrew the atomizing
port plug and connect the connecting pipe to the atomizing port of the main machine and
the air inlet of the atomizing cup.Turn on the power, adjust the flow rate of the oxygen flow
meter to 1.5L/min, and follow the doctor's instructions for atomization treatment.After the
atomization is finished, close the oxygen concentrator, screw the atomization port plug
into the atomization port and tighten.To continue to breathe oxygen, turn the oxygen
concentrator back on.
3. Oxygen button: machine start / standby mode
4. Sleep button: When the machine is started, the display goes out;
5. Timing +: Timing time increment setting
6. Timing -: Timing time decrement setting
7. Current overload protector: protect the whole circuit
8. Switch: whole machine power switch
9. Display:

Figure 8
9.1. Power indicator light--The machine is in standby mode when it is in the flashing state, and

16
it is in the working state when it is on.
9.2. Sleep Breathing Light--The machine is in sleep mode when the light is in the state, the
display is off, sleep breathing
Open
9.3. Low-oxygen indicator light--The machine oxygen concentration is lower than 82% when it
is on; (When the actual flow exceeds the designed flow of the machine, the oxygen
concentration may be lower than 82%)
9.4. Fault indicator light--The machine parts are faulty, please contact the after-sales service;
9.5. Accumulated time--the cumulative use time of the machine;
9.6. Timing time--the working time of the machine, the unit is min (minutes);
9.7. Oxygen concentration--Display the current oxygen concentration value in L/min (L/min);
9.8. Oxygen flow rate--Display the current flow rate value in units of l (liter);

Remarks: The oxygen concentrator has the following fault detection functions: low oxygen

alarm and component failure alarm;

When the oxygen machine works normally for 10 minutes, the oxygen concentration detection
(low oxygen alarm) function of the oxygen machine starts the real-time detection of the
working state of the oxygen machine. When the oxygen concentration is ≤82% (±3%), the
yellow hypoxia alarm icon lights up. (When the actual flow exceeds the rated flow, the oxygen
concentration may be less than 82%), please shut down immediately when the low oxygen
alarm is encountered; when the machine parts are abnormal and cannot work normally, the
red fault alarm icon lights up. And continue to alarm, the whole machine stops running, please
shut down immediately;
10. Oxygen flow meter (refer to Figure 9)

17
Adjustment
knob

Flow ball

XY-6S -10 model


Figure 9

Flow meter adjustment knob: Oxygen flow meter switch knob to regulate and control the
output oxygen flow
The level of the flow ball in the oxygen flow meter corresponds to the flowmeter scale,
indicating the flow rate (L/min)

Note: Check the oxygen flow meter and confirm that the flow ball is centered under the
tick mark specified by your oxygen flow meter (the maximum flow rate of the XY-6 model
oxygen machine is 3L/min, please do not let the flow ball be higher than 3L, XY-6S model
oxygen machine maximum flow rate For 5L/min, please do not let the flow ball be higher than
5L).

The flow in the oxygen prescription is very important. Do not increase or decrease the
reference flow of oxygen provided by your doctor.

If the meter's regulator valve rotates clockwise, the flow will decrease (and eventually the
oxygen flow will be turned off).If it is rotated counterclockwise, the flow will increase.

11. Intake filter cover: The intake filter can be replaced after the cover is removed:
12. Machine air inlet: machine oxygen source and machine cooling air source, ensure that
the air port is 30cm away from debris or wall during normal use, so as not to affect the
18
normal air intake of the machine;
13. Product nameplate/factory number label: product performance label, label product serial
number.
14. AC power cord.

MACHINE USE PRECAUTIONS

 When the oxygen concentrator is working, there must be no other items (including
walls) within 30 cm of the back of the machine; objects with high temperature, sparks
or open flames should not be present within 1.6 m.
 Smoking is prohibited during oxygen inhalation. When working in an oxygen
concentrator, it is strictly forbidden to smoke next to or near it.
 Do not use in dust, smoke, corrosion, poisonous or polluting gases and strong
magnetic fields.
 The oxygen concentrator is strictly prohibited from coming into contact with flammable
oils (including lubricants/oils).
 The air intake filter is recommended to be replaced after 1500 hours of use.
 Please do not use a mobile socket or share the same socket with other appliances.
 If you feel uncomfortable or abnormal, stop using the product immediately.
 Be sure to read the instructions in the instruction manual carefully before use.
 Unauthorized personnel must not disassemble the oxygen concentrator casing.

Product cleaning and maintenance instructions

1、 Clean the whole machine casing: 1~2 times a month, please wipe the outside of the
fuselage with a clean soft towel with a little detergent, and then dry it with a dry towel.

2、 Cleaning the humidification bottle: The humidification bottle can be cleaned separately
with detergent and hot water, or mixed with white vinegar and water in a ratio of 1:3 as a
bactericide, and the separated humidification bottle is placed in the solution for 30
minutes. Then wipe the humidification bottle.

3、 Replacing the air intake filter: Intake filter cotton is an important part of the cleaning
oxygen machine air source. In order to ensure that the machine continues to provide
19
fresh and sufficient oxygen, it is recommended to replace the air intake filter every 1500
hours.

Note: replacement of the oxygen pipe


In view of the fact that the oxygen pipe is a one-time use product, the company does not
recommend reusing the oxygen pipe to prevent the disease from causing disease.Each
bag has a separate oxygen tube that can be purchased at the dealer.The company is not
responsible for the adverse consequences arising from repeated use.

PRODUCT WARRANTY

Within 12 months from the date of purchase, it is a quality guarantee period. If the product
is not artificial and cannot operate normally, Oxygen should repair or replace the parts for the
user without charge.
1、 During the warranty period, under normal use conditions, due to the quality problems
caused by the purchased product itself and the warranty conditions, please return the
completed warranty card to the company for repair, or carry it. The warranty card and the
purchase ticket are repaired free of charge at the point of sale of the purchased product.If
the warranty card is lost, the warranty period of the product you purchased will be based
on the results of the database query, or the date of purchase on the 30th day of the
fuselage code.

2、 Machines that fail due to unauthorized modifications or additional functions are not
accepted.

3、 Save the warranty card and the purchase ticket as the warranty certificate for the product,
please keep it in a safe place and lose it.

The following conditions are not free of charge:

1、 The product is out of warranty;


2、 No warranty certificate;
3、 Failure caused by operation in accordance with the instructions;

20
4、 Damage caused by the dismantling of maintenance personnel of the company;
5、 Malfunction, scratch or breakage caused by movement;
6、 The user's own failure to repair, disassemble, and assemble the product;
7、 Normal damage to consumables and consumables;
8、 Failure and damage caused by force majeure (such as fire, flood, earthquake, etc.).

Fault and repair checklist

To avoid electric shock, do not open the case by yourself.Opening the enclosure can
only be carried out by a specialist authorized by the oxygen concentrator manufacturer.

The [Fault and Repair Checklist] attached below will help you analyze and properly repair
the malfunction of the oxygen concentrator.If the suggested procedure does not help, use an
alternate oxygen concentrator and notify the supplier of the oxygen concentrator for
repair.Please do not try to try any other repairs.

Malfunction
Possible cause Maintenance method
symptoms
When the switch The power cord
Check the connection between the power cord and the wall
is turned on, the plug is not properly
outlet. Check the connection between the power cord and the
machine does not plugged into the
wall outlet after the 220v voltage checker is installed.
run, the power power outlet.
indicator does not
light, the alarm The power outlet is Check the home circuit breaker and restart if necessary.If the
sounds, and the not powered. same problem occurs again, use another AC outlet.
alarm light is on.

When the switch Check the air filter cotton and if it is dirty, clean it according to
The intake filter is
is turned on, after item 3 of the “Product Cleaning and Maintenance
clogged.
the machine is Instructions”.
running for 1 The exhaust port Check the base exhaust port and make sure the exhaust port
minute, the power is blocked. is not blocked.

21
indicator light is The oxygen tube,
Separate the oxygen tube, catheter or mask and if the flow is
on, the red alarm catheter, mask or
restored, it needs to be cleaned or replaced.
light is on, and an oxygen connection
Take the oxygen pipe out of the oxygen outlet. If the flow rate
alarm sound may tube is blocked or
rises, replace the oxygen pipe and use it.
be heard. defective.

When the switch The cable tie that


is turned on, the fastens the
Turn off the machine, unplug the power cord, place the side of
power indicator compressor at the
the machine on the ground, cut the tie that fastens the
lights up and a bottom of the
compressor, and pull it out.
low frequency machine is not cut
vibration is heard. and pulled out.

The red fault


alarm indicator The machine part
Contact your supplier.
lights up and the is faulty.
alarm sounds.
The flow meter is Confirm that the flowmeter is adjusted to the specified
not adjusted position.
The yellow Check the air filter cotton and if it is dirty, clean it according to
Intake filter cotton
hypoxia indicator item 3 of the “Product Cleaning and Maintenance
blockage
lights up. Instructions”.
The exhaust port Check the base exhaust port and make sure the exhaust port
is blocked. is not blocked.
If the machine still does not work properly after the above method, or if there are any other problems with
the machine, please contact the product supplier or service point.

* If there is no special explanation, please follow the above instructions.

Product return step requirements

22
Dear Customer:
Hello! First of all, thank you very much for your trust and support to Oxygen. We will
wholeheartedly strive to provide you with satisfactory after-sales service.We apologize for the
abnormal use of your products and promise that they will solve them as soon as possible.
Please follow the steps below to return the products to our company for repair or return:
1、 Complete the product warranty card that came with the end of this manual and return it
with the product.
 Please fill in the reason why the product must be returned to the factory for maintenance.
If the machine received by our after-sales personnel does not have any fault description,
it will take time to re-test and confirm the return to the machine state, which will increase
the time for solving the problem.
 Please indicate your contact details. If necessary, our sales staff may call you to confirm
the machine again.
2、 The description of the accessory parts of the product is as follows:
 For repaired machines, accessories such as oxygen users and humidification bottles are
disposable consumables and do not need to be returned at the same time as the
machine.
 For the returned machine, it is necessary to ensure that the attached parts are sealed and
packaged intact and will not affect the secondary sales, otherwise the price of such
accessories may be deducted when returning.
 If you want to add accessories that come with the purchase, you can also fill in the
instructions on the repair order.
3、 Please remove the product's humidification bottle and do not have water in the bottle.
 If the returned product has a humidified bottle and the water in the bottle is not cleaned,
the water will be injected back into the machine, causing secondary damage to the
product.
 Since such damage occurs as a result of failure to follow the instructions, not the quality
of our products, it should be carried out in accordance with the paid warranty terms that
you need to pay for repairs.
4、 When returning to the factory, please use the matching box of our products.
 The packaging of our products is designed according to the overall rationality of the

23
products to ensure the safety of the transportation process. If you use other packaging,
there may be damage to the transportation of the products.
 This kind of damage occurs, it is not the quality problem of our products, it should be
implemented according to the paid warranty clause that you need to pay for repair; and
in order to ensure the safety when the repair machine is sent back again, our company
will replace the original packaging, the cost will also It is listed in the list you need to pay.
5. (If necessary) The product is sterilized before being issued.
 Since some users have infectious diseases, if your machine is used by such patients,
please wipe the outer casing with alcohol before packaging the machine, or (under the
guidance of staff) for other more professional Disinfection, so as not to infect people
and people who will come back to the machine again.

Please refer to our recommendations after the product manual to return to the machine.
If you have any questions, please call us! Finally, thank you for your support!

Term of use / date of manufacture


1. The period of use: 5 years.
2. Date of manufacture: See product performance label.

Environmental protection
1. Disposal of waste and residue
When the oxygen pipe, humidification bottle and supporting components are used,
do not discard them and send them to a nearby medical waste disposal facility for
disposal.
2. When the machine is scrapped, please contact your local supplier or manufacturer.
3. The disposal of waste and residue should comply with the corresponding national
laws and regulations.

List of components

24
Serial number Component name Quantity

1 Main control board 1 block

2 display board 1 block

3 Oxygen Sensor 1

4 Oxygen system solenoid valve 1

5 compressor 1 set

6 DC fan 1

7 switch 1

8 Overload protector 1

Portable oxygen concentrator

25
Product warranty card (user association)

Customer Name: ________________________________

Phone: ________________________________________

Contact address: ________________________________

Zip code: ______________________________________

Product Name: oxygen concentrator

Model: ________________________________________

Body number: ___________________________________

Dealer: ________________________________________

Address: _______________________________________

Phone: ________________________________________

Zip code: ______________________________________

Purchase date: __________________________________

26
CERTIFICATE
Product Name: Portable oxygen concentrator
Product model: XY-6S-10
The product inspection meets the technical requirements and the
product is qualified.
inspector:
Inspection date:

Shenzhen Hongxinyuan Electronics Co., Ltd.

27