Trace 1110 SOP 2013 Rev 0.2013
Trace 1110 SOP 2013 Rev 0.2013
Trace 1110 SOP 2013 Rev 0.2013
TRACE 1110
Standard Operating Procedures
PN 827S07134, Revision 0- 2013
TRACE™ GC 1110 - Standard Operating Procedures
September 2013 Edition
© 2013 Thermo Fisher Scientific. All rights reserved.
Printed in India
Disclaimer
Technical Information contained in this publication is for reference
purposes only and is subject to change without notice. Every effort
has been made to supply complete and accurate information;
however, Thermo Fisher Scientific assumes no responsibility and will
not be liable for any errors, omissions, damage, or loss that might
result from any use of this manual or the information contained
therein (even if this information is properly followed and problems
still arise).
Trademarks
TRACE™ GC is a trademark of Thermo Fisher Scientific Inc., and its
subsidiaries. Other brand and product names may be trademarks or
registered trademarks of their respective companies.
Contents
Chapter 10 Checkout Using FPD with S/SL Injector ( Sulphur mode) ........... 1
Scope ................................................................................................ 1
Recommended Initial Operations .................................................... 2
Operating Procedure ........................................................................ 3
Checkout Using FPD with S/SL Injector (Phosphorus mode)......... 5
Symbols
Warning pertaining to a possible safety hazard or instrument damage.
Hot surface. Avoid contact. Surfaces are hot and may cause personal
injury if touched.
Table 1.2
Detector Abbreviation
Flame Ionization Detector FID
Electron Capture Detector ECD
Nitrogen Phosphorus Detector NPD
Flame Photometric Detector FPD
Thermal Conductivity Detector TCD
Safety
Information Precaution for Gases
Before using gases, carefully read the indications of hazard and the
warning reported in the Safety sheet supplied by the manufacturer
with reference to the relevant CAS number (Chemical Abstract
Service).
Using the TRACE The TRACE GC Ultra Document Set (CD-Rom PN 828S08030)
includes all manuals in electronic format, and serves as your library
GC 1110
for information about the TRACE GC 1110 hardware and software.
Document Set
The TRACE GC 1110 Document Set as paper copy is also available
Furthermore, Thermo Fisher Scientific part numbers (PN) for the
paper copy manuals are provided for each book title.
Chapter at a Glance…
• Scope
• Parts Referenced
• Getting Started
Operating Procedures
• Test Column Conditioning
Scope The Standard Operating Procedures (SOP) described in this book are
a series of instructions, operations and test criteria derived from our
quality policy procedures used for final testing of the TRACE GC
1110. The SOPs have been developed to test and verify instrument
complete analytical performance after the installation has been
completed. This will help you as a guideline, to check if your TRACE
GC 1110 continues to perform according to the original checkout
testing specifications carried out in the factory premises. However,
these tests alone cannot define if the instrument is not performing
according to the original specifications. The checkout is carried out
injecting a standard solution into a test column under analytical
conditions set according to the injector(s) and detector(s) hardware
provided with the GC. Before starting the test checkout, refer to the
Parts Referenced and the Analytical Condition required.
Parts Referenced Table 2.1 The SOPs require the following parts:
Column CAUTION !
When performing column conditioning, the column should be
Conditioning
connected only to the injector leaving the column outlet
disconnected to avoid the possibility of contamination of the
detector base body.
Operating
Table 2.2 Test Column Conditioning
Procedure Parameter Setting
Gases Carrier Gas: Helium or Nitrogen
Control Mode : Constant Flow
Injector Mode: Splitless
Column Flow = 3.40 ml/min (for Capillary column )
Column Flow = 20.0 ml/min (for Packed column )
Oven Initial Temperature = 50 °C
Program Initial Time = 1 min.
Ramp 1 = 10 °C/min.
Final Temperature = 250 °C
Final Time = 30 min.
Injector Temperature = according to the injector in use
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions required for packed Injection
• Recommended Initial Operations
Operating Procedures
• FID-PKD Checkout
Scope Use the following procedure to verify proper FID operation with the
Packed column Injector.
Operating
Procedure FID- Packed Injector Checkout
Fig. 3.1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for splitless injection
• Recommended Initial Operations
Operating Procedures
• FID-S/SL Checkout for splitless injection.
Scope Use the following procedure to verify proper FID operation with the
Split/ Splitless Injector.
Recommended
Initial Operations Before starting the checkout, the following operations should be
carried out:
1. Check up the Splitless glass liner(208262940) installed in your
injector , it should be clean and free from any deposits or septum
particles if required clean it or replace with new one, with the
appropriate liner seal(208826974).
2. Replace the septum A new septum should be installed properly in
your injector.
3. Connect the required gas lines, Verify the required gas supplies
are properly connected to your GC.
4. Install the test column. The column currently installed should be
carefully removed and replaced with the required test column and
Perform Column Leak Test.
5. Connect your data handling. Verify that your data handling is
properly connected to your GC system.
Operating
Procedure FID-S/SL Checkout in Splitless Mode
Fig. 4.1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for PTV injection
• Recommended Initial Operations
Operating Procedures
• FID-PTV Checkout for PTV injection
Scope Use the following procedure to verify proper FID operation with the
PTV Injector.
Operating
Procedure FID-S/SL Checkout in PTV Mode
5. Activate your Data System and set the parameters required for the
checkout.
6. In the FID Detector menu , scroll to SIGNAL and for press mV
value followed by ENTER key for Auto zero.
7. Perform a blank analysis injecting pure hexane and press START
on the GC to begin the checkout run.
8. Inject the test mixture and press START on the GC to begin the
checkout run. The resulting chromatogram should look like the
one shown in Figure 5.1.
Fig. 5.1.
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for TCD with PKD injection
• Recommended Initial Operations
Operating Procedures
• TCD-PKD Checkout for PKD injection
Scope Use the following procedure to verify proper TCD operation with the
Packed column Injector.
Recommended
Before starting the checkout, the following operations should be
Initial Operations
carried out:
1. Connect/ensure proper gases are connected to your GC.
2. Replace the septum A new septum should be installed properly in
your injector.
3. Connect the required packed column to your GC.
4. Measure suggested flow of carrier gas for column, reference and
Make up gases.
5. Connect your data handling. Verify that your data handling is
properly connected to your GC system.
Operating
Procedure TCD- Packed Injector Checkout
Before beginning, press CONFIG to verify the GC configuration:
1. Use INJ menu and set the Flow and temperature value for the
required test.
2. Use OVEN menu to display the Column Oven Control Table.
Set the oven temperature and the Oven Program required.
3. Use Det menu to Set the required temperature and flow set point.
4. Activate your Data System and set the parameters required for
the checkout.
5. In the TCD Detector menu, scroll to SIGNAL and press mV
value followed by ENTER for Auto zero
6. Perform a blank analysis injecting pure hexane and press START
on the GC to begin the checkout run.
7. Inject the test mixture and press START on the GC to begin the
checkout run. The resulting chromatogram should look like the
one shown in
Fig. 6.1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for TCD with splitless
injection
• Recommended Initial Operations
Operating Procedures
• TCD-S/SL Checkout for splitless injection
Scope Use the following procedure to verify proper TCD operation with the
Packed column Injector.
Operating
Procedure TCD-S/SL Checkout in Splitless Mode
Fig. 7.1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for TCD with S/SL injection
• Recommended Initial Operations
Operating Procedures
• TCD-S/SL Checkout for S/SL injection
Scope Use the following procedure to verify proper ECD operation with the
Split/ Splitless Injector.
1. Use INJ menu and set the pressure value to have the required
carrier gas flow rate.
2. Use OVEN menu to display the Column Oven Control Table. Set
the oven temperature and the Oven Program required.
4. Use Det menu to Set the required temperature and flow setpoint.
5. Activate your Data System and set the parameters required for the
checkout.
8. Inject the test mixture and press START on the GC to begin the
checkout run. The resulting chromatogram should look like the
one shown in Figure 8.1.
Fig 8.1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for TCD with S/SL injection
• Recommended Initial Operations
Operating Procedures
• NPD-S/SL Checkout for S/SL injection
Scope Use the following procedure to verify proper NPD operation with the
Split/ Splitless Injector.
Operating
Procedure NPD-S/SL Checkout in Splitless Mode
1. Use INJ menu and set the pressure value to have the required
carrier gas flow rate.
2. Use OVEN menu to display the Column Oven Control Table. Set
the oven temperature and the Oven Program required.
3. Use Split menu to display the appropriate Split/Splitless Control
Table.
4. Use Det menu to Set the required temperature and flow set point.
Activate your Data System and set the parameters required for the
checkout.
12. In the NPD Detector menu , scroll to SIGNAL and press mV
value followed by ENTER for Auto zero.
5. Perform a blank analysis injecting pure iso-octane and press
START on the GC to begin the checkout run.
6. Inject the test mixture and press START on the GC to begin the
checkout run. The resulting chromatogram should look like the
one shown in Figure 9.1.
Fig. 9.1
Selectivity :
Azobenzene to Octadecane should be > 2000 : 1
Methyl Parathion to Octadecane should be > 20000 : 1
Chapter at a Glance…
• Scope
• Parts Referenced
• Analytical Conditions Required for FPD with S/SL injection
• Recommended Initial Operations
Operating Procedures
• FPD-S/SL Checkout for S/SL injection
Scope Use the following procedure to verify proper FPD operation with the
Split/ Splitless Injector.
Parameters Setting
Injector Inj Temperature = 230 °C
(Capillary ) Control Mode : Constant Flow
Carrier Gas: Nitrogen
Injector Mode: Splitless
Inlet Press = 3.40 ml/min (Constant Flow )
Oven Program Initial Temperature = 60 °C
Initial Time = 1 minute
Ramp 1 = 20 °C/minute
Final Temperature = 200 °C
Final Time = 1 minute
Max Temperature = 300 °C
Split Operating Mode = Splitless
Splitless Time = 0.4 minutes
Split Flow = 60 ml/min
Septum Purge = 2 ml/min
Injected Volume 1 µl + needle of Test Mixture
Detector Det Temperature = 250 °C
Detector Range = 10^1
Hydrogen = 110 ml/min
Air-1 = 60 ml/min
Air-2 =160 ml/min
FPD Supply PMTVoltage = 500-600 volt
FPD PMT voltage = 5.50 (dial set approx,550 V)
Aquisition Chrom-Card, Frequency = 10 Hz
Frequency
Operating
FPD-S/SL Checkout in Splitless Mode
Procedure
Before beginning, press CONFIG to verify the GC configuration:
1. Use INJ menu and set the pressure value to have the required
carrier gas flow rate.
2. Use OVEN menu to display the Column Oven Control Table. Set
the oven temperature and the Oven Program required.
3. Use Split menu to display the appropriate Split/Splitless Control
Table.
4. Use Det menu to Set the required temperature and flow setpoint.
5. Activate your Data System and set the parameters required for the
checkout.
6. In the FPD Detector menu scroll to SIGNAL and press mV value
followed by ENTER for Auto zero.
7. Perform a blank analysis injecting pure iso-octane and press
START on the GC to begin the checkout run.
8. Inject the test mixture and press START on the GC to begin the
checkout run. The resulting chromatogram should look like the
one shown in Figure 10.1.
9. Typical Chromatograph of SOP Sample of 10 ppm Methyl
Parathion in Hexane FPD-S mode ( with Sulphur filter 393 nm )
Fig. 10.1
While on Iris-32 Area of Methyl Parathion Peak should be greater than 200
counts (mV).
@ with Sample Qty of Injection 1.0 microlitre cold needle split less
injection .
Checkout Using
Shut off the FPD flame gases ,Cool down the FPD detector Replace
FPD with S/SL the Sulphur interference filter (P/N-525249984 ) by Phosphorous
Injector filter (P/N-525249985 ) and perform the FPD Phosphorus mode
(Phosphorus checkout for details please refer to service manual for detail
mode) procedure for changing filters in FPD detector
Fig 10.2
@ with Sample Qty of Injection 1.0 microlitre cold needle split less