Important: Make sure you read and understand all of the information
Contents
1 Foreword…………………………………………………………3
2 Symbols…..........…………………………………………………4
3 Safety notice……...............………………………………………5
4 Product introduction…....................................……..…………7
5 Operation condition………….....………………………………7
6 Scope of application…………..........……………………………8
7 Technical parameter………........………..………………………8
8 Structures and Functions……………….....……………………10
9 Operation instructions…………..........................................…11
10 Alarms and safety devices.........………....…………………16
11 Maintenance…...........………………………..……………17
12 Troubleshooting…….........................................……………20
13 Accessories………………………………………………………22
14 Condition for transportation and storage................................23
15 Contact us…………………………………………………………23
1 Foreword
Thank you for purchasing our products, hoping you will be satisfied with
our products.
2 Symbols
The following table is a list of symbols and definitions that used with the Oxygen
Concentrator.
Symbol Description Symbol Description
Warning – Describes a Indicating its conformity with the
hazard or unsafe practice Medical Devices Directive
that can result in severe 93/42/EEC.The 0197 is the
bodily injury or death number of the Notified Body.
Caution – Describes a
hazard or unsafe practice
“ON” (power)
that can result in property
damage
T6.3AL/
Manufacturer Type and rating of fuse
250V
3 Safety notice
:This unit is not a life-support device , and in certain circumstances oxygen therapy can
be hazardous, it is suggested that if any patient who needs oxygen treatment, please follow
doctor’s advice to choose the right flow and period for oxygen before using the oxygen
concentrator.
:In the event of an alarm, you observe your oxygen concentrator not working properly,
or if you feel discomfort, consult your Equipment Provider and /or your physician
immediately.
:This device manufactures high concentration oxygen, which promotes rapid burning.
Keep oxygen concentrator far away from open flames and no smoking around the patient.
:Do not leave a nasal oxygen cannula under bed coverings or chair cushions. If the unit
is turned on without use, the oxygen will help the flammable material get fire .
:Electrical shock hazard. Do not remove covers while the unit is plugged in. Only your
Equipment Provider or a qualified service technician should remove the covers or service
the unit.
:Care should be taken to prevent the unit from getting wet or allowing water to enter
the unit.
:Do not place the oxygen concentrator in surroundings where its airflow is obstructed.
:Always place the concentrator on a hard surface. Never place the concentrator on a
surface such as bed or couch, where the concentrator may tip or fall.
:Ensure the bottom smooth exhaustion during operating, otherwise the oxygen
:5 minutes are needed from oxygen concentrator from warming up to reach regular
the oxygen concentrator is still running. ALWAYS exercise CAUTION with regard to the use
of portable communications equipment in the area around such equipment.
The materials used in the system won’t create environment hazard. The packing materials
of the system are recyclable, and they must be collected and disposed according to the
related regulation in the country or region where the package of the system or its
accessories is opened. The nasal oxygen tube is made of medical PVC, and if it is thrown
away, it could not be bio-degradable, so it will cause the pollution. Any material of the
system, that may cause pollution in the environment, must be collected disposed strictly
complied with the local rules and requirements.
4 Product introduction
OxyTec Series oxygen concentrator is a device that extracts oxygen from atmospheric air. It
will typically be an electrically-powered molecular sieve (artificial zeolite) used to separate
nitrogen from ambient air. It could be applied widely in the hospitals at all different level,
clinics, health centers and family nursing, health care for the old person, mental workers
and students, etc.. The oxygen concentrator can supply 1-2 patients simultaneously, with
steady oxygen flowing out, safe and reliable, low cost, adjustable flow. The key parts of the
concentrator adopt anti-tiring and anti-aging design, and the planned life of the whole
concentrator reaches up to 20,000 hours or 5 years, whichever comes first.
6 Scope of application:
For Medical Use:
Medical oxygen supplied by the concentrator is beneficial to cure the disease or heart and
blood vessel system, chronic pulmonary system, the brain and blood vessel system,
chronic pulmonary tuberculosis, and other oxygen lacking symptoms, etc.
For Health care:
Medical oxygen can be used for athletics and intellectuals and brainworkers, etc. to
eliminate fatigue and also suit for the departments of health care, sanatorium, healthy,
plateau military camps and hotels and other places where need oxygen.
7 Technical Parameters
OxyTec-10 OxyTec-10
Dual Flow
Voltage 220V AC +22v 50 Hz
Average power consumption <550 (W)
Ambient operating 5°C to 40°C
temperature
Flow rate 0.1 to 10.0LPM
Oxygen concentrating 78%-95%
Outlet pressure 10PSI; 20PSI
Weight 27kg 28kg
Dimensions (LxWxH) 365X375X600 mm
Sound level < 50dB (A)
Electromagnetic According EN 60601-1-2
compatibility EMC
Protection class according to Class II; Type B
EN 60601-1
Classification in according Class II a
with EC Directive 93/42/EEC
Annex IX
Figure 1 Figure 2
8.1 Indicating Lamp
8.1.1 Total 8 indicating lamps and their indication for the OxyTec-10 models, which is shown
as Figure 1, are as follows:
a. P.O.: power switch (green lamp)
b. P.F.: power failure(red lamp)
c. H.P.: high pressure(red lamp)
d. L.P.: low pressure(yellow lamp)
e. H.T.: over heated temperature(red lamp)
f. H.O2: oxygen purity is 85%, (blue lamp)(Accuracy:±3%)
g. L.O2.: oxygen purity is < 85%,(red lamp)(Accuracy:±3%)
8.2 Power switch
8.3 Oxygen flow meter
The location of float in the oxygen flow meter shows the outlet oxygen flow (L/min.).
8.4 Knob of oxygen flow meter switch
It adjusts and controls the outlet oxygen flow.
Do not Switch it over-forced, or else it is easy to damage the valve core. Switch it
counterclockwise to turn on, clockwise to turn off.
8.5 Outlet for Atomization (optional)
8.6 Intake air filter
Once need to replace, refer to 11.2 on page15
8.7 Storage cabinet
It is used for storage oxygen cannula
8.8 LCD display( Liquid crystal display )
a.. It can display some status of during operation of the oxygen concentrator, refer to 7.on
page 9
b. When starting the oxygen concentrator, the LCD screen is lighted, and it will return to
screen saver mode in 15 minutes. But if you press the right key during working, the screen
will be lighted again.
8.9 Timing buttons
The two buttons are used for timing adjustment, and each press of the left button(△) will
increase timing by 10min, the max timing is 40 hours. And each press of the right button(▽)
will decrease timing by 10min. When the right button(▽) is pressed to reduce timing till “0”,
the oxygen concentrator will turn off automatically.
8.10 Humidifier
Humidifier which is used for humidifying oxygen and preventing throat and nasal mucosa
stimulated by dry oxygen and dry hard sputum difficult to spit out.
8.11 Knob of Atomizer switch (optional)
8.12 Rating label
8.13 Appliance inlet
8.14 Holder
9 Operation instructions
:Be certain to place the unit where all sides are at least 30 cm away from walls,
draperies, furniture, or other obstructions. Do not place the unit in a confined area.
:Do not turn on or off frequently. To restart the oxygen concentrator after turning off, no
less than 5 minutes are necessary (namely, exhaust internal gas of the oxygen
concentrator completely, for if air compressor turns on with pressure, its life will be
shortened)
9.1. If used with a humidifier, unscrew the flask from the humidifier in clockwise direction,
pour in proper distilled water or cold boiled water within the scale between the max line and
the min line, then re-connect the top cover to the humidifier bottle, as shown in Figure 5.
The humidifier must comply with the general requirement of the 93/42/EEC European
Directive as appropriate.
Figure 5
9.2. Connect the nasal oxygen canula to the humidifier outlet nozzle or to the concentrator
outlet if a humidifier has not been prescribed. Then set the nasal oxygen cannula over
patient’s ears, insert the nasal oxygen cannula into patient’s nostrils to absorb oxygen; The
nasal oxygen canula should be limited to 20 meters long, in order to ensure that the oxygen
flow rate remains within specification values. The best absorbing time for health care keeps
40-50 minutes per time, absorbing time for medical treatment shall be followed doctor’s
advice, as shown in Figure 6.
Figure 6
9.3. Insert the power plug plug into the electrical outlet of the correct voltage and frequency
as defined in 7(Technical Parameters)on page 8 , and the power connector connected with
the appliance inlet of the oxygen concentrator, then set the I/O power switch to the “I”
position to turn the unit on, at the same time the P.O. lamp will light(8.1.1a and 8.1.2a).
Figure 7
9.4. To set the flow of supplemental oxygen, turn the knob of oxygen flow meter switch left or
right until the ball inside the flowmeter centers on the flow line number recommended
oxygen absorbing flow.(counterclockwise—on, clockwise—off).
Figure 8
Flow value:
Oxytec-8 Dual Flow:10 position flow value from 0.5 ~ 5L/min on flowmeter as shown in
figure 8. The maximum recommended flow: 5L/min.
Oxytec-8 : 16 position flow value from 0.5 ~ 8L/min on flowmeter. The maximum
recommended flow: 8L/min.
In compliance with the ISO8359 standard, the flow supplied is equal to the flow set on the
flowmeter, accurate to within 10% or 200ml/min, whichever is the larger of the two
The variation of the maximum recommended flow does not exceed 10% of the indicated
value when a back pressure of 7kPa is applied to the output of the device. The maximum
Oxygen Concentration:
-at 2L/min: >90%
-at 5L/min: 93%( 3%)(Oxytec-8 Dual flow)
-at 8L/min: 93%( 3%)(Oxytec-8)
:It is very important to select only the prescribed level of oxygen. Change the flow
selection only under the guidance of your physician.
9.5 . When finished the absorbing, set the I/O power switch to the “O” position to turn off the
unit, if there is discontinuous use, please unplug the power plug, as shown in Figure 9
Figure 9
9.6. If the patient needs timing oxygen absorbing, please refer to 8.9 on page 10
:Using distilled water to do the atomization for several seconds after each operation
may lighten the crystallization caused by medical solution.
:If atomizing cannot work then open the cover of the bottle and add clean water in
small amount. Rotate the white ball which lies in the bottle with the gas resource connected
and select the proper angle to gain a better atomization.
a. Open the cover of medical cup, and add atomization remedy that needed, then close the
cover.
b. Connect the joint of atomization nozzle (or mask) with the cover of medical cup, and then
connect the other end of atomizer connection tube with the atomization outlet and turn on
atomizer.
c. Turn on the power of oxygen concentrator, and shut up flow meter, then it is ready for
atomization treatment.
d. Do clean the atomization devices followed by the instructions of the atomization devices.
Figure 10
Figure 11 Figure 12
10 Alarms-Safety devices
10.1 Alarms
a. Power failure alarm: In case of a loss of mains power or when the power cord is not
plugged into the wall outlet, an audible alarm is activated with red indicator on(8.1.1b and
8.1.2b on page 10). The troubleshooting is referred to 12 on page 17.
b. low&high pressure alarm: There is a pressure sensor on the main board to check the
system pressure, when the pressure is lower than 0.1Mpa,there is an audible alarm with
yellow indicator on(8.1.1d and 8.1.2c on page 10) and the oxygen concentrator is still
working; When the pressure is higher than 0.23Mpa,there is an audible alarm with red
indicator on(8.1.1c and 8.1.2d on page10) and the oxygen concentrator is stopped. The
troubleshooting is referred to 12 on page 17.
c. Low oxygen concentration alarm (OCSI): The oxygen concentration will rise to the
normal level in five minutes of operation. When oxygen purity is 85%, the green lamp
(8.1.1f and 8.1.2e on page 10) is on, when oxygen purity is <85%, red lamp is on for
audible alarm, indicating low purity(8.1.1g and 8.1.2f on page10). Refer to the
troubleshooting on page 17. Or call your supplier to service the device
d. Temperature alarm: There is a temperature sensor on the main board to check the
internal temperature, when the temperature is higher than 50°C in the oxygen concentrator,
there is an audible alarm with red indicator on(see H.T. on the lamp) and the oxygen
concentrator will be stopped. The troubleshooting is referred to 12 on page 17.
a. Compressor motor:
Thermal safety is ensured by a thermal switch situated in the motor winding(145±5℃).
b. Safety valve:
This is fitted on the compressor outlet and is calibrated to 2.5 bar(250kPa).
11 Maintenance
:Disconnect the power cord from the electrical outlet before you clean the cabinet.
:Do not operate the concentrator without the filters installed, or while filters are wet.
These actions could permanently damage the concentrator.
NOTE: If legally binding regulations govern the installation, service and/or the operation of
the product, it is the responsibility for the operator to observe and follow these regulations.
NOTE: Modifying the product is not permitted.
11.1. Clean the whole body: In the condition of power off, make a clean for the outside
body by soft towel with little mild household cleaner , and then wipe it up with dry towel,
once or twice per month.
Figure 13
11.2. Clean air filter: It is a critical step for daily maintenance to clean intake air filter, at
least twice a month.
Detail steps: remove the two intake air filters on both sides of the body, clean them with mild
household cleaner and clean it with clean water completely, get ride of extra water and dry
up naturally, finally set back after dry up, as shown in Figure 14.
Air filter
Figure 14
Figure 15
Method II: First, remove the silicone tube connected with oxygen output, take down the
humidification bottle, Rotate the air filter with counterclockwise direction as Figure
15,take down the filter cloth, clean with detergent, wash by clean water, wipe off the
water. and make it to be dry in air naturally ,then install back to the Air filter.
Figure 16
11.4. Clean the humidifier (if the humidifier is prescribed by a physician)
Daily:
- Empty the water from the humidifier.
- Rinse the humidifier flask under running water.
- Fill humidifier up to the mask with distilled water.
Regularly:
Figure 18
12 Troubleshooting
If your concentrator fails to operate properly, please refer to the troubleshooting chart on the
following pages for probable causes and solutions. If problems with the equipment continue,
please contact your Equipment Provider.
NOTE: If the unit has not been used for an extended time period, it needs to operate for
several minutes before power failure alarm can become activated.
4 Too noisy exhaustion 1. The joint of exhaustion 1.Connect the joint well
muffler fallen off 2.Have the muffler replaced
2. Exhaustion muffler broken
5 The oxygen The system pressure is too low. Check every gas circuit
concentrator is connectors with soapy water
working but the L.P. whether there are air leakage.
lamp is light with
audible alarm.
6 The oxygen The temperature in the oxygen 1.Check the fan’s connector on
concentrator is concentrator is too high. the main board whether it is bad
13 Accessories
Your concentrator includes the following components:
• Intake air filter (two pieces, part number:GL-01)
• Secondary filter (one piece, part number:GL-02)
The concentrator comes with two air filters and one secondary filter already installed.
: Please use the parts mentioned in this chapter, if the use of other parts can degrade
minimum safety and performance.
: Please choose the suitable humidifier and Nasal oxygen cannula, they must:
- be oxygen compatible,
- be biocompatible,
- include a means to prevent the propagation of fire and accord with requirements of ISO
8359:1996/Amd.1:2012