Maintenance Manual FOR Turbine Flow Meter: Shanghai Cowell Machinery Co., LTD

Download as pdf or txt
Download as pdf or txt
You are on page 1of 29

MAINTENANCE MANUAL

FOR
TURBINE FLOW METER

Liquid Turbine Flow Meter

SHANGHAI COWELL MACHINERY CO., LTD.


Turbine Flowmeter Instruction2

Content

Ⅰ、 Generality

Ⅱ、 Principle of Operation

Ⅲ、 Product Feature

Ⅳ、 Basic Parameter / Technical Specification

Ⅴ、 Product Category

Ⅵ、 Types Choice

Ⅶ、 Installation Size

Ⅷ、 Caution in Installation

Ⅸ、 Connection

Ⅹ、 Adjustment and Application

Ⅺ、 Caution in Using

Ⅻ、 Problem & Solution

ⅩⅢ、Transportation & Storage

ⅩⅣ、Cautions in Unpacking

ⅩⅤ、Necessary Knowledge on Order

Turbine Flowmeter Instruction2


Turbine Flowmeter Instruction3

Ⅰ、Generality
Turbine Flow Meter (Abbr. TUF) is a main type of Impeller Flowmeters

also including the Anemoscope and Water meter. TUF is made up of

Sensor and Conversion-Show. The Sensor reacts to the average velocity

of fluid with multi-blades rotor so as to speculating the flow value and

the accumulative flow value. The velocity (or circles) of rotor can be

picked up by the way of mechanism, electromagnetic induction,

photoelectricity, before displaying and transmitting the records by

reading device.

It is said that America announced the first TUF patent early in 1886.

The patent in 1914 recorded that the TUF flow value is relevant to

frequency. The first developed TUF in 1938 is applied to measuring the

fuel flow in the aircraft. It is eventurelly achieved to use in the

industry until the end of the world war two, since it is urgent for the jet

engine and liquid jet fuel to demand high accuracy, quick responses flow

measurement instrument. Nowadays, it can be extensively used in the

fields of oil,chemical,defence,science,measuring, etc..

LWGY series Turbine Flowmeters draw the leading technology

integrating with advanced design to produce the new generation of

turbine flowmeter with the features of simple structure, light weight, high

accuracy, good repeatability, flexible reaction, convenient

installation/maintenance/application etc.. It is widely applied to

Turbine Flowmeter Instruction3


Turbine Flowmeter Instruction4

measuring the liquid of which kinematic viscosity is under 5*10 -6 ㎡/s

and have no impurify of fibre,grain etc.,and no corrosive interaction with

the stainless steel 1Cr18Ni9Ti,2Cr13,and A12O3, and hard alloy in seal

pipes. The liquid of kinematic above 5*10-6 ㎡/s can be measured after

real liquid calibration of flowmeter. It can be used in value control,

siren when excess, if co-ordination with special display instrument. So it

is the ideal instrument of measuring flow value and saving energy.

Ⅱ、 Principle of Operation

As the measured liquid flows through the sensor, the drived vane begins

to turn, which velocity is in direct proportion to average flow one in the

pipe. The turn of vane periodically changes the magnetic resistance

value of magnetoelastic transducer. Magnetic flux in the magnetic test

coil happens to change cyclically with it to produce periodic induced

voltage, it is the pulse signal, that will be sent to the display to show after

amplified by magnifier.

Flow rate equation of Turbine Flowmeter includes both practical and

theoretical one:

(1) Practical equation:

Qv=f/k

Qm= Qv vρ

Turbine Flowmeter Instruction4


Turbine Flowmeter Instruction5

Qv refers to volume flow rate, (unit: m3/s)

Qm refers to mass flow rate, (unit ㎏/s)

f : refer to output signal frequency (unit Hz)

k : refer to the Flowmeter factor, (unit P/m3).

The related curve of flowmeter factor and flow rate is in the graph

(Diagram: Turbine flowmeter characteristic curve). As your seeing, the

factor curve can be divided into two parts of linearity and non-linearity.

The linear part accounts for two-thirds of the entire curve which feature

is related to the structure, size of sensors, and fluid viscosity. The

feature in non-linearity part is influenced by friction force from bearing,

the viscosity resistance of liquid. When flow rate is below the lower

limit of sensor, the instrument factor are quickly increasing with it. The

value of pressure loss and the flow rate are similar to be square relations.

If flow rate surpassed the upper limit, pay attention to preventing from

cavitation. When the turbine flowmeters have similar structure, their

curves have similar feature but have different system errors.

(Diagram: Turbine flowmeter characteristic curve)

The sensor factor can be worked out by calibration instrument, which

may have no consideration of the sensor’s inside fluid mechanism, and

can be confirmed by inputed flow rate and outputed pulse signals of


Turbine Flowmeter Instruction5
Turbine Flowmeter Instruction6

frequency. So we can see the sensor as a black box , that is convenient

for application. But please note that the conversion factor (or

instrument factor) should comply with some conditions which calibration

condition is the reference condition . If it deviate from this condition, the

factor will happen to change. The changes would be determined in terms

of the sensors type, the pipe installation condition,and fluid physical

parameters.

(2) Theoretical flow rate equation:

According moment of momentum theorem can list the equation of

motion impeller.

J dw dt =M1-M2-M3-M4

In the formula,

J: impeller inertia moment;

dw dt: rotational acceleration;


M1 : Liquid-driven torque

M2 : Viscous resistance moment

M3 : Bearing friction moment

M4 : Magnetic moment.

When impeller is rotating according to constant velocity, J dw dt =0,

and M1=M2+M3+M4. Through the analysis in theory and verification

in experiment ,the formula can be deduced that is:

n=Aqv+B- C qv
Turbine Flowmeter Instruction6
Turbine Flowmeter Instruction7

In the formula,

n: refers to impeller rotational speed;

q v: refers to volume flow rate;

A: the factors related to fluid physical properties ( include density,

viscosity etc.), impeller structure parameters (blade angle, impeller

diameter, flow channel cross-sectional area etc.);

B: the factors related to top vane gap, and fluid flow velocity

distribution;

C: the factor related to friction moment.

The scholars domestic and abroad have put forward to many flow

equations in theory, applied to various sensors structures and fluid

working conditions. Until now, the hydrodynamic characteristic of

turbine instrument ones is still unclear, for it has complicated relationship

with fluid physical property, and flow characteristics. For instance,

when there appears to swirling and unsymmetry velocity distribution in

flow field, the hydrodynamic characteristics are very complicated.

So instrument factors can not be deduced by theoretical formula, can be

confirmed by real flow calibration. But theoretical formula has been

significant in practice. It can be used in instruction in the design of sensor

structure parameter and the forecast ,and assessment of instrument

factor changing rule.

Turbine Flowmeter Instruction7


Turbine Flowmeter Instruction8

Ⅲ、 Product Feature:

(1) High accuracy (regular accuracy ± 1%R, ± 0.5%R,

highest accuracy±0.2% R);

(2) Good repeatability (short-term one reaches 0.05%--0.2%), priority

to be used in trade settlement for its extremely high accuracy in the

regular calibration or on-line calibration.

(3) Pulse frequency signal output is applicable to totality calculation

and computer connection with no zero drift and strong

anti-interference capacity.

(4) High frequency (3-4kHz) can be achieved, and has high resolution.

(5) Wide rangeability: medium or large diameter may reach 1:20,and

small diameters are 1:10.

(6) Compact and light weight structure, convenient installation and

maintenance, extensive application ability.

(7) Application to high pressure measurement with its unnecessary

opening hole to be made into high pressure instruments.

(8) Complete tailored version sensors can be designed to different kinds

of types according to users special needs. For instance, low

temperature type, high pressure type, sanitary type, etc.

(9) Insertion type can be made, that is applicable to large normal

diameters measurement for its little pressure loss, low price,

unnecessary stopping flow to take out it, and convenient installation

Turbine Flowmeter Instruction8


Turbine Flowmeter Instruction9

and maintenance.

Ⅳ、 Basic Parameters / Technical Specification

1. Technical Specification:
4,6,10,15,20,25,32,40 (tread connection)
Nominal Diameter(mm)
15,20,25,32,40 (tread and flange connection)
and Connection method
50,65,80,100,125,150,200 ( flange connection)
Regular accuracy ±1%R, ±0.5%R,
Accuracy Class
Highest accuracy ±0.2% R
Measurement Range Rate 1:10,1:15,1:20
Instrument material 304 stainless steel; 316L stainless steel; etc.
Medium Temperature(℃) -20 ~ +120 ℃
Temperature:-10~+55℃,
Ambient Conditions Relative Humidity: 5%~90%
Atmosphere Pressure:86~106Kpa
Sensor: pulse frequency signal,
low level≤0.8V
high level≥8V.
Signal Output
Transmitter: current signal
4~20mA DC
two wires
Sensor: +12V DC, +24V DC (option)
Supply Power Transducer: +24V DC
Scene display type meter: 3.2V Lithium cell
Signal Transmission Line STVPV 3×0.3 (three wires), 2×0.3(two wires)
Transmission Distance ≤1000m
Signal Line Interface Internal thread M20×1.5
Explode-proof Class ExdIIBT6
Protection Class IP65

Turbine Flowmeter Instruction9


Turbine Flowmeter Instruction10

2. Measurement range and Working pressure for liquid


Nominal Regular Expanding Regular tolerance Special tolerance
Diameter Flow rate flow rate pressure(Mpa) pressure(Mpa)
(mm) (m3/h) (m3/h) (flange connection)
DN4 0.04—0.25 0.04—0.4 6.3 12, 16, 25

DN6 0.1—0.6 0.06—0.6 6.3 12, 16, 25

DN10 0.2—1.2 0.15—1.5 6.3 12, 16, 25

DN15 0.6—6 0.4—8 6.3, 2.5(flange) 4.0, 6.3, 12, 16, 25

DN20 0.8—8 0.45—9 6.3, 2.5(flange) 4.0, 6.3, 12, 16, 25

DN25 1—10 0.5—10 6.3, 2.5(flange) 4.0, 6.3, 12, 16, 25

DN32 1.5—15 0.8—15 6.3, 2.5(flange) 4.0, 6.3, 12, 16, 25

DN40 2—20 1—20 6.3, 2.5(flange) 4.0, 6.3, 12, 16, 25

DN50 4—40 2—40 2.5 4.0, 6.3, 12, 16, 25

DN65 7—70 4—70 2.5 4.0, 6.3, 12, 16, 25

DN80 10—100 5—100 2.5 4.0, 6.3, 12, 16, 25

DN100 20—200 10—200 2.5 4.0, 6.3, 12, 16, 25

DN125 25—250 13—250 1.6 2.5, 4.0, 6.3, 12, 16

DN150 30—300 15—300 1.6 2.5, 4.0, 6.3, 12, 16

DN200 80--800 40—800 1.6 2.5, 4.0, 6.3, 12, 16

Turbine Flowmeter Instruction10


Turbine Flowmeter Instruction11

3. Measurement range and Working pressure for gas


Model Diameter Flow Rate Initial Flow Rate Tolerance
(mm) (m3/h) (m3/h) pressure(Mpa)
(flange connection)
LWQ-25A 0.7—7 0.6 4.0 Flange or Thread

LWQ-25B 25 (1”) 1.5—15 1.0 4.0 Flange or Thread

LWQ-25C 3—30 2.0 4.0 Flange or Thread

LWQ-40A 4—40 2.5 4.0 Flange or Thread


40 (1.5”)
LWQ-40B 8—80 3 4.0 Flange or Thread

LWQ-50A 10—100 3.5 4.0 Flange


50 (2”)
LWQ-50B 15—150 4 4.0 Flange

LWQ-80 15—300 4 1.6 Flange


80 (3”)

LWQ-100 20—400 5 1.6 Flange


100 (4”)

LWQ-150 50—1000 8 1.6 Flange


150 (6”)

LWQ-200 100—2000 20 1.6 Flange


200 (8”)

LWQ-250 150—3000 30 1.6 Flange


250 (10”)

LWQ-300 200—4000 40 1.6 Flange


300 (12”)

Ⅴ、 Product Category

1. LWGY series can be divided into two categories by function:

 Turbine flow sensor / transmitter

 Intelligent integration Turbine Flowmeter

2. Function illustration:

 Turbine flow sensor/ transmitter

This kind of products have no scene display function,only produce

Turbine Flowmeter Instruction11


Turbine Flowmeter Instruction12

signals to transmit output to far distance. The flow signals can be

divided into pulse or current (4-20ma) signal. This instrument has

low price, high assemble, small size, so can be applicable to match

second displayer,PLC,DCS so on computer control system to use.

According to different signal outputs, its can be divided into

LWGY-N and LWGY-A types.

 LWGY—N sensor

12--24V DC power supply, three wires pulse outputs,

high level≥8V, low level≤0.8V, signal transmission

distance≤1000M.

 LWGY—A transmitter

24V DC power supply, two wires current (4—20mA) signal

output, signal transmission distance≤1000M.

 Intelligent integration turbine flowmeter

It adopts an advanced super-low power consumption single-chip

microprocessor technology to make up of new intelligent flowmeter

with turbine flow sensor and accumulative calculation displayer

integration. It has many obvious advantages which are double-row

LCD display at the scene, compact structure, direct and clear

reading, high reliability, anti-interference from outside power,

anti-thunder attack, and low cost ,etc.

It has the instrument factors’ three points rectified, non-linear

Turbine Flowmeter Instruction12


Turbine Flowmeter Instruction13

intelligently compensated, and revision at the scene.

High clear LCD display simultaneously shows both instant flow rate

(four valid figures) and accumulative flow rate (eight valid figures,

and accumulative flow rate (eight valid figures with reset). All valid

data can be kept for ten years. This kind of turbine flowmeters all

are explosion-proof products, and the explosion-proof class is

ExdIIB6.

This type of turbine flowmeters can be divided into type LWGY—B

and LWGY—C in terms of supply power and the remote signal

transmitting methods.

LWGY—B type: supply power 3.2V10AH(Lithium battery) can

continuously run more than four years, but no signal output.

LWGY—C type: supply power 24V DC outside, output normal

two wires current signal (4-20 m A) , and can add RS485 or HART

communication according to different scene demand.

Turbine Flowmeter Instruction13


Turbine Flowmeter Instruction14

Ⅵ、Type Choice
Model
Explanation
LWGY- □/ □/ □/ □/ □/ □/ □
4 4mm, normal flow range0.04-0.25m3/h,wide flow range0.04-0.4m3/h
6 6mm, normal flow range0.1-0.6m3/h,wide flow range0.06-0.6m3/h
10 10mm, normal flow range0.2-1.2m3/h,wide flow range0.15-1.5m3/h
15 15mm normal flow range0.6-6m3/h,wide flow range0.4-8m3/h
20 20mm normal flow range0.8-8m3/h,wide flow range0.4-8m3/h
25 25mm normal flow range1-10m3/h,wide flow range0.5-10m3/h
DN 32 32mm normal flow range1.5-15m3/h,wide flow range0.8-15m3/h
(mm) 40 40mm normal flow range2-20m3/h,wide flow range1-20m3/h
50 50mm normal flow range4-40m3/h,wide flow range2-40m3/h
65 65mm normal flow range7-70m3/h,wide flow range4-70m3/h
80 80mm normal flow range10-100m3/h,wide flow range5-100m3/h
100 100mm normal flow range20-200m3/h,wide flow range10-200m3/h
125 125mm normal flow range25-250m3/h,wide flow range13-250m3/h
150 150mm normal flow range30-300m3/h,wide flow range15-300m3/h
200 200mm normal flow range80-800m3/h,wide flow range40-800m3/h
N Basic type, +12Vsupply power, pulse output, high level≥l8V, low level≤0.8V
A 4—20mA two wires current output, remote transmitting type.
Type B Battery supply power, scene display type.
C scene display/4—20m A two wires current output
C1 Scene display/ RS485 communication protocol
C2 Scene display /HART communication protocol
Accuracyc 05 Accuracy class 0.5
lass 10 Accuracy class 1.0
Measurement range W Wide flow range turbine
mark S Standard measurement range turbine
S 304 Stainless steel
Materials
L 316(L) Stainless steel
N No mark, non-explosion-proof
Explosion-proof
E Explosion-proof(ExdIIBT6)
N Normal (reference to picture before)
Pressure class
H(x) High pressure (reference to picture before)

Note: DN15—DN40 need thread connection regularly,but can be made into flange
connection through adding the “FL” to the nominal diameter at its end.

Turbine Flowmeter Instruction14


Turbine Flowmeter Instruction15

Ⅶ、 Installation Size

LWGY 4~10 Sensor Structure & Installation Diagram

1.Strainer 2.Before Straight Pipe 3.Impeller 4. Preamplifier 5.Body 6. Back Straight Pipe

LWGY 15~40 Sensor Structure & Installation Diagram

1. Body 2. Former Guide Part 3. Impeller 4. Back Guide Part 5. Preamplifier

Turbine Flowmeter Instruction15


Turbine Flowmeter Instruction16

LWGY 50~200 Sensor Structure & Installation Diagram

1.Ball Bearing 2. Former Guide Part 3. O ring 4. Body 5. Preamplifier 6. Impeller


7. Bearing 8. Shaft

Nominal hole
L(mm) G D(mm) d (mm)
diameter(mm) number
4 295 G1/2
6 330 G1/2
10 450 G1/2
15 75 G1 φ 65 φ 14 4
20 80 G1 φ 75 φ 14 4
25 100 G5/4 φ 85 φ 14 4
32 140 G2 φ 100 φ 14 4
40 140 G2 φ 110 φ 18 4
50 150 φ 125 φ 18 4
65 170 φ 145 φ 18 4
80 200 φ 160 φ 18 8
100 220 φ 180 φ 18 8
125 250 φ 210 φ 25 8
150 300 φ 250 φ 25 8
200 360 φ 295 φ 25 12

Turbine Flowmeter Instruction16


Turbine Flowmeter Instruction17

Ⅷ、 Cautions in Installation

(1) The installation site:

Sensor should be installed in the sites where is convenient to maintain,

have no vibration of pipe, no strong electromagnetic interference,

and hot radiation influence. The typical pipe installation system of

turbine flowmeter is following as the picture. The each part of

configuration can be chosen in view of the objects measured ,which

needn’t all. It is sensitive for turbine flowmeter to velocity aberrance

and rotating flow, so entering sensor should be the pipe flow

developed enough and match the necessary straight pipe or rectifier. If

upstream side components of flow resistence are variables, the

pipeline length upstream generally is not less than 20D and the

pipeline length downstream is not less than 5D. If the installation

space does not satisfy these demands, the flow rectifier may be

installed between the component of flow resistence and sensor. The

sensor should be installed outside where avoids the direct sunshine

and rain.
(picture)
Upstream Single Double Double Concentric Open Open Downstream
component 90°angle 90°angle 90°angle reducing whole half side length
types bend bends at the bends at the pipe valve valve
same level different level

L/DN 20 25 40 15 20 50 5

Turbine Flowmeter Instruction17


Turbine Flowmeter Instruction18

(2)The installation demands on connection with pipes:

The horizontally installed sensor demands the pipeline inclination

shouldn’t be visible (generally within 5°), and the vertically installed

one should be same as it.The site needed to run continuouslly should

install the by-pass pipe and reliable cut-off valve. It must be assured that

the by-pass pipe has no leakage when measuring.

Location of sensor in a new pipeline is replaced into a short pipe first.

After the pipeline inside has been cleared , the short pipe can be changed

back into sensor formally. For this step always has been reglected, the

sensor may often be damaged during clearing pipeline.

If the measured fluid includes impurity , the filter should be installed

before sensor of upstream side. To continuous flow liquid should

install two sets of filters which clear impurity in turn, or choose auto

clearing type filer. If the air mixes in the liquid, the eliminator

should be installed in the upstream side. The mouth of filter or

eliminator must be led to safe site.

If the location of sensor is at the lower point of the pipeline, the

drain valve should be fixed after the sensor to discharge the impurity

regularly in order to prevent from dwelling deposit. If the measured

liquid is easy to be aerified , the exit pressure of sensor should be

more than Pmin in order to prevent from air pockets that may

damage the accuracy and live time.

Turbine Flowmeter Instruction18


Turbine Flowmeter Instruction19

Pmin=2⊿P+1.25Pv Pa

Pmin: The lowest pressure, Pa;

⊿P: the pressure loss while the sensor flow rate is the biggest Pa;

Pv : the saturation vapour pressure when the use temperature arrives

at the highest point Pa.

Flow control valve should be fixed in the sensor’s downstream ,and

the cut-off valve at the upstream side all should be opened, whose

valves may not produce vibracation and leakage toward outside. To

the flow range that might make the reversed flow should prevent the

fluid’s reversed flow with fixing the check valve. Both sensor and

pipeline should be concentric. The sealed washer is not allowed to

have it protruded into the pipeline. The liquid sensor should not be

fixed at the top of the horizontal pipeline lest the air converging into

the pipe stops in the sensor not to be expelled so as to effect the

measurement.

The sensor’s front and back pipe sections should be supported firmly

so as not to produce vibration. If the fluid is easy to condense, the

measurement of keeping temperature should be taken in the sensor and

its front and back sections of pipeline.

Turbine Flowmeter Instruction19


Turbine Flowmeter Instruction20

Ⅸ、Connection Way

◆ Turbine flow sensor/transmitter:(model LWGY-N, model

LWGY-A )
1. Basic type:

LWGY-□N type Turbine flow meter connection way

LWGY-□A type turbine transmitter connection way

Turbine Flowmeter Instruction20


Turbine Flowmeter Instruction21

2. Anti-explosion type:
LWGY-□N type turbine flow meter sensor connection way:

LWGY-□A type turbine flow transmitter connection way:

Turbine Flowmeter Instruction21


Turbine Flowmeter Instruction22

◆ Intelligent integration turbine flowmeter(model LWGY-C)

Ⅹ、 Adjustment and Application

LWGY-N basic type turbine flow meter:

This sensor has been calibrated and adjusted before sales, so needn’t

examination.

The sensor conbines with displayer: in the first place, checking the

output feature(the pulse frequent range,level,wide etc.) which should

match the entry feature of displayer. The displayer parameters must

set in terms of sensor factors. The sensor power , wire, and

Turbine Flowmeter Instruction22


Turbine Flowmeter Instruction23

resistance must match each other as well.. In addition, the sensor’s

prepositional amplifier must be considered to prevent from

electromagnetic interference, for instance, to take action of rain

proof.

LWGY-A turbine flow transmitter:

This transmitter should be set the flow rate output zero point and

the full range value well according to the customer demand when

purchasing.

When the flowmeter works on and the flow rate output zero point

should be adjust on site, the operation method is doing as the

follows:

Close the valves of flowmeter pipe, confirm there is not flow rate in

pipe; put on the power, the series-connected current meter can

monitor the flowmeter’s output current; slightly adjust the W502

potentiometer on the circuit board to come the output current back to

4m A.

Note: the flowmeter’s full range value couldn’t be adjusted on site

after it works; If need, please return it to factory to complete that in

the standard installation according to your need.

LWGY-B intelligent turbine flowmeter display on site


 The internal parameter set:
(only engineer operation)

01. the panel keys illustration

Turbine Flowmeter Instruction23


Turbine Flowmeter Instruction24

 entry(exit)parameter set menu:in the working state

simultaneously press the key and key F;

 cursor moves towards right:in the parameter set state


press the key

 cursor adds one figure:in the parameter set state


press the key

 parameter menu switch : in the parameter set state


press key F;

 cumulative flow value to clear: in the working state

simultaneously press the key F and key 。

02. internal parameter illustration:

The flow meter program includes three menus,which

shows with three screens that is the three points

modification:the top one is the value of frequency, the

bottom one is the meter factor of this flow point. The

three menus can circularly be switched with the key F

LWGY-C intelligent turbine flow meter display on site with


4-20mA current output

Turbine Flowmeter Instruction24


Turbine Flowmeter Instruction25

01.The panel key illustration:

The turbine flow meter with battery and supply power

on site display

02.Internal parameter illustration:

The meter has four menu programs which displays in four

screens: the first three menus are three points parameters

modification the same as the type o f battery supply on site

display turbine flow meter;the fourth one is the full range value

of current 4-20m A (the 20m A point flow value). These menus

can be circularly switched with the key F

Ⅺ、 Cautions in Using

(1) The switch order putting into running

※The sensor that have not the branch pipe should slightly

open up the half upstream valve, then the downstream valve.

When running for a while through a small rate( about ten

minutes), open the whole upstream valve and the downstream

valve to the normal flowrate.

Turbine Flowmeter Instruction25


Turbine Flowmeter Instruction26

※The sensor with branch should first open the branch pipe

valve, the half upstream valve, the downstream valve, close

the branch valve to small flowrate, and running for a while.

Then open the whole upstream valve, close the whole branch

valve( be assure of no leak) , finally adjust the downstream

valve to the needed flowrate.

(2) The low and high temperature fluid starts up

When low temperature fluid flows through the pipe, first the

water should be expelled, then running for fifteen minutes

with a minimum flow, and gradually rise to the normal flow.

When stop flowing , also should gradually reduce to

approaching pipe temperature and ambient temperature.

The high temperature fluid’s running is similar to this low one.

(3) Other notes:

1) Opening and closing the valve should be slow. If adopted

the auto control switch, it is best to use “two open, two close”

way to prevent the fluid against vane wheel to damage it.

2) Check the sensor’s downstream pressure to adopt measures

to prevent cavitation.

3) For the sensor factors could appear to change should

regularly calibrate away from pipe line. If the flow is not

within the allowed range , sensor should be change .

Turbine Flowmeter Instruction26


Turbine Flowmeter Instruction27

4) Cleaning the pipe needs confirm to the standards of used

flow direction,value,pressure,and temperature etc.,otherwise

can make the accuracy fall,even damage.

5) Strengthen the check for sensor in order to assurance of

long time normal working. As finding the unnormal , the

measure should be taken. For instance, hearing the unnormal

voice as monitoring the vane wheel rotation

Ⅻ、 Problem and Solution

problem Possible reason solution


No showing or no total Check: 1)find the problem point
adding when liquid 1)open circuit、 loose with electrical meter or
normally flows. contact (wire power replace this circuit board
wire\fuse\coil\PCB) with spare one.
2)the vane wheel has no 2)clean or replace vane
rotation wheel ,and assure no
rubbing with its
neighboring parts.
The flow showing is 1) filter blocks up 1) clean up the filter
gradually falling. 2) valve in pipe is loose 2) repair or replace the
to the core valve
3) vane wheel has 3) clean the sensor, then
impurity need to calibrate again
Its screen has still flow 1) the cable has no good 1) repair or replace to
showing when liquid has groud wire with the have a good groud
no flow outer interference; wire;
2) the pipe with 2) strengthen the pipe
vibracation to produce line, or install blacket
error signal to prevent from
3) the cutoff valve has vibracation;
leakage with leaking 3) maintain or replace
flow valve
4) internal circuit or 4) gradually check and
component of clear up the
displayer is damaged interference source.
to produce interference

Turbine Flowmeter Instruction27


Turbine Flowmeter Instruction28

The displaying value has 1) The sensor’s internal 1)-4) need first find cause
obvious difference with tunnel wrong; so that use the correct
experience estimation one 2) Sensor’s interior methods;
appears cavitation; 5)replace the magnet
3) The flow inside pipe material
causes problems 6)choose the proper sensor
4) The displayer interior
wrong
5) The effect of
permanent magnet
material is weaker and
weaker
6) The real flow is not
within its normal range

ⅩⅢ、Transportation and Storage

The sensor should be put in the solid wooden box(small diameters

can be put in carton) and cannot be free to wobble in the box.

When carrying, it must be care to put down ,and refuse to load or

unload crustily.

The location of reservation should be confirmed to the conditions as

the following:

1. avoid rain and humidity;

2. avoid mechanical vibration and strike ;

3. temperature range:-20℃--+55℃;

4. relative humidity: not more than 80%;

5. ambient environment does not include corrosive gas.

ⅩⅣ、Cautions in unpacking

When opening box, files and accessory should be complete. The

files in the box include a user manual, a piece of test certificate, and

Turbine Flowmeter Instruction28


Turbine Flowmeter Instruction29

a piece of packing list. The sensor should be observed whether it

happens damage during transportation so that dealing with it well.

Users must protect the certificate from loss otherwise the instrument

factors cannot be set.

ⅩⅤ、Necessary knowledge on order


User should notice that when ordering turbine flowmeter , the proper

model specification should be chosen according to fluid’s nominal

diameter, operating pressure, operating temperature, flow range, the

fluid category and the surrounding condition. The anti-explosion

type sensor should be chosen when having explosion-proof demand

and noticing strictly the explosion-proof classes.

When the display instrument is matched by our company, please

refer to the related instruction to choose your proper model or use

our design of technological engineer for your choosing in terms of

your information offering. The cable using in sending signal you

want should provide the length and specification.