12. SAFETY INSTRUCTIONS FOR THE INSTALLER 3-5
13. OPENING THE TERMINAL 3-7
13.1 Opening the outer casing 3-7
13.2 Main parts of the terminal 3-8
13.3 Removal of the power supplier 3-9
13.4 Removal of the converter 3-10
13.5 Removal of the CPU 3-12
13.6 Removal of the display 3-13
13.7 Replacing the keypad 3-14
13.8 Replacing the lithium battery 3-15
13.9 Installing the MPP board 3-17
13.10 Installing optional boards 3-19
13.11 Installing the optional 4 I/O board 3-19
13.12 Installing the optional serial ports board 3-20
13.13 Installing the optional BCD parallel 5V-calculator board 3-20
13.14 Installing the optional BCD parallel 24V-source board 3-21
14. PERSONALIZING METROLOGICAL PARAMETERS 3-23
14.1 Calibration button 3-23
14.2 Description of the main menus and metrological parameters 3-25
14.2.1 Setup Menu/Scale/Configurations/Metrological 3-26
14.2.2 Setup Menu/ANALOGUE scale/Analogue scale parameters 3-29
14.2.3 Setup Menu/DIGITAL scale/Digital scale parameters 3-29
14.2.4 Setup Menu/Scale/Calibration/Execute 3-29
14.2.5 Setup Menu/Scale/Calibration/Display data 3-33
14.2.6 Setup Menu/Scale/Calibration/Correction 3-33
14.2.7 Setup Menu/DIGITAL scale/Calibration/Angle calibr. 3-35
14.2.8 Setup Menu/DIGITAL scale/Test 3-39
14.2.9 Setup Menu/DIGITAL scale/Test/Terminal data storage 3-40
14.2.10 Setup Menu/DIGITAL scale/Test/Terminal data reinstatement 3-40
14.3 Repeater scale 3-41
14.3.1 Setup Menu/Scale/Repeater scale parameters/Serial port/String
3-41
15. SOFTWARE UPGRADE 3-43
15.1 Setup Menu/Upgrade 3-43
15.1.1 Setup Menu/Upgrade/Serial line 3-44
15.1.2 Upgrade via upgrade board 3-46
16. MAINTENANCE 3-49
16.1 Setup Menu/Maintenance 3-49
17. TROUBLESHOOTING 3-53
17.1 CPU and converter faults 3-54
17.2 Management of errors by peripheral devices 3-57
3-3
D400
3-4
D400
CAUTION
The information and instructions contained in this section are
addressed exclusively to installation personnel who must
have specialised knowledge in the fields of electrical and
electronic engineering and programming.
A number of the operations described herein require the
removal of legal seals and the opening of the terminal
casing.
Furthermore, certain operations can be only performed when
the terminal is connected to the electrical power supply.
In this condition, some of the exposed electrical components
will be live and consequently there is a risk of electric shock.
Any dismantling or opening of the terminal by the user or any
other unauthorised person shall immediately invalidate the
warranty and release the Manufacturer from all liability for
any personal injury or damage sustained.
Also consult the "Safety instructions" " chapter of the user
manual.
3-5
D400
3-6
D400
Key
1. casing screws
2. casing and seal screw
3. plastic casing
Figure 13.1 - Opening the terminal (log0106.jpg)
✔ Unplug the mains lead connector, the platform scale connector and
any other connections;
✔ Remove the six screws indicated in fig. 13.1 on page 3-7 (points 1
and 2. One of the screws is not visible in the figure);
✔ Remove the plastic casing.
3-7
D400
Key
1. Plastic casing
2. Slots for optional boards
3. Power supplier
4. Display
5. MPP
6. CPU
7. Converter for analogue terminal
The converter in digital terminals has no cover ( fig. 13.5 on page 3-11
)
Figure 13.2 - Main parts of the terminal (log0107.jpg)
3-8
D400
Key
1. screws
Figure 13.3 - Removal of the power supplier (log0108.jpg)
3-9
D400
Key
1. screws
2. connector on CPU
3. connector on converter (obscured in photo)
Figure 13.4 - Analogue terminal converter demounting (log0111.jpg)
3-10
D400
Key
1. screws
2. connector on CPU
3. connector on converter (obscured in photo)
Figure 13.5 - Digital terminal converter demounting (citi0202.jpg)
3-11
D400
Key
1. power supplier connector
2. display connector
3. keypad connector
4. optional boards
5. converter connector
6. screws
7. screw posts on D-type connectors
Figure 13.6 - Removal of the CPU (log0112.jpg)
3-12
D400
To remove the CPU board, proceed as follows:
✔ remove any optional boards from slots 1 and 2 ( fig. 13.10 on page
3-17 );
✔ disconnect the connectors (points 1, 2, 3, 5 of fig. 13.6 on page 3-12
);
✔ unscrew the screws (points 6 of fig. 13.6 on page 3-12 ) and screw
posts (points 7 of fig. 13.6 on page 3-12 ).
Key
1. screws
2. display connector
Figure 13.7 - Removal of the display (log0113.jpg)
3-13
D400
Key
1. leverage points
2. slot for display connector
3. keypad connector
Figure 13.8 - Replacing the keypad (log0116.jpg)
3-14
D400
✔ Apply leverage using a screwdriver, for example, at points 1.
✔ Unstick the keypad.
✔ Remove all glue residue with alcohol.
✔ Take care not to get alcohol anywhere other than the contact
surfaces and take care not to damage the seal labels. Allow the
contact surface to dry.
✔ Insert the connector in slot 2 and stick the keypad down, positioning it
carefully in the centre of its housing.
DANGER
Take care not to cause short-circuits between the metal
components.
3-15
D400
Key
1. lithium battery
Figure 13.9 - Position of the lithium battery on the CPU board
(log0150.gif)
3-16
D400
Key
1. connectors on MPP board
2. connector on CPU board (the other connector is not visible in the
figure)
Figure 13.10 - Installing the MPP board (log0117.jpg)
✔ Make sure that jumper P1 on board CPU is connected;
✔ Centre board MPP on the connectors.
Consult fig. 13.11 on page 3-18 where the connectors on the CPU
board are more clearly shown.
3-17
D400
Key
1. Jumper P1
2. Connectors for the MPP board
Figure 13.11 - CPU board (log0152.gif)
3-18
D400
Key
1. slot 1
2. slot 2
3. fixing screw
4. connector on CPU
Figure 13.12 - Installing the optional board (log0120.jpg)
3-19
D400
Key
1. jumper P1
2. jumper P2
Figure 13.13 - Personalizing the BCD parallel 5V-calculator board
(log0121.jpg)
Jumpers
P1 on pin 25 i +5V.
Should be connected to the first terminal in IDC systems. It is generally
disconnected. For special applications refer to the specifications.
P2 identifies the board type.
Connected for BCD 5V-calculator output in number of divisions.
Disconnected for BCD 5V multiplied output.
3-20
D400
Personalize the relative parameters (see Personalization chapter of
advanced user manual).
See Options chapter of the user manual for the connections.
Key
1. jumper P1
2. jumper P2
Figure 13.14 - Installing the BCD parallel 24V-source board
(log0121.jpg)
Jumpers
P1 permanently disconnected
P2 permanently disconnected
Personalize the relative parameters (see Personalization chapter of
advanced user manual).
See Options chapter of the user manual for the connections.
3-21
D400
3-22
D400
Key
1. access hole for calibration button
2. seal label area
3. pointed instrument
Figure 14.1 - Accessing the setup procedure by way of the
calibration button (log0123.jpg)
3-23
D400
If you access the setup procedure by pressing the calibration button it is
possible to modify all the terminal parameters.
The correct procedure is as follows:
✔ remove the seals from the terminal (lead or labels) (point 2 fig. 14.1
on page 3-23 );
✔ switch on the terminal;
✔ press the calibration button using a non-metallic pointed instrument
(point 3 fig. 14.1 on page 3-23 ) through the access hole (point 1 of
fig. 14.1 on page 3-23 );
✔ select the menu display language; a few terminal identification
messages will be displayed briefly (program code, version, serial
number) immediately followed by the Setup menu.
On exiting Setup, the display language selected in the personalization
menu will be restored.
WARNING
Access to setup parameters by way of the calibration button
is restricted to specialised personnel only.
Any tampering by unauthorised personnel shall invalidate the
warranty with immediate effect.
3-24
D400
3-25
D400
3-26
D400
3-27
D400
3-28
D400
3-29
D400
3-30
D400
3-31
D400
3-32
D400
3-33
D400
Weight Editor
If the weight is known, the correction can be
entered directly.
The value entered must lie within the range:
- (Capacity/100) < Weight < + (Capacity/100)
Gravity Editor
place of calibration
On selecting Campogalliano (headquarters
of the Manufacturer) the gravitational
acceleration value of Campogalliano (9.80552
m/s) is entered automatically), otherwise it is
possible to enter a gravitational acceleration
value within the range (9.60000 to 10.00000)
m/s.
place of use
As the gravitational acceleration of the place
of calibration is known, in order to calculate
the weight correction value it is also necessary
to know the gravitational acceleration of the
place of use. You can enter a value within the
range (9.60000 to 10.00000) m/s.
Alternatively, you can enter the Coordinates
of the place of use: latitude (0 to 90)° and
altitude (-1000 to +9000) m.
On the basis of these data, the program will
calculate the weight correction and enter the
value obtained in the parameter Weight .
3-34
D400
3-35
D400
3-36
D400
3-37
D400
3-38
D400
Consult the Personalization chapter of the advanced user manual for all
the other parameters in this menu.
3-39
D400
3-40
D400
3-41
D400
Consult the Personalization chapter of the advanced user manual for the
other parameters in this menu.
Refer to chapters 15 and 16 of this section for maintenance and
software updates.
3-42
D400
In the first case, all the parameters including sampling and the archives,
remain unchanged.
In the second case, the parameters, metrological data, any prints,
transmissions and personalized messages do not change. However,
remember that if the new program involves substantial changes to the
operating mode or further parameters, the old configuration could be
wrongly interpreted. This is why it is advisable to print the report before
and after the update in order to make sure that the parameters are
correct.
The data in archives common to both programs remain unchanged.
The logo of the terminal can be loaded and cancelled via Dialogic. It is
not changed by initialization or by an update.
CAUTION
If the CPU board is replaced it may also be necessary to
update the software on the new board so that it matches that
of the old board.
Make sure that the installed software and board are
compatible with each other.
Do not interrupt the updating procedure as this could impair
the result of the downloading operation.
More detailed information is available on request from the
Manufacturer.
3-43
D400
There are two possible methods for upgrading the firmware, which are
described in the following paragraphs ( par. 15.1.1 on page 3-44 and
par. 15.1.2 on page 3-46 ).
You will need a file containing the new software (generally the name of
this file consists of the 6 digits of the program code plus the extension
.a37). Locate this file on the PC. Use a program that contains the kermit
binary file transfer protocol (e.g. Windows Hyperterminal).
Configure the program on the PC as follows:
✔ 38400,8,N,1;
✔ no port control.
3-44
D400
Follow the menu path:
The terminal will set on hold and give the .waiting message,
cancelling the program it contains..
Transmit the file with the new software version from the PC by means of
the Kermit procedure.
On receiving the file, the terminal displays the message .loading.
On completion of the transfer procedure the terminal displays
.correctly terminated: switch off. (transfer completed
successfully; switch off the terminal).
Should any error messages be displayed notify Assistance Service.
Switch the terminal off and then on again before using the new software.
3-45
D400
CAUTION
This operation exposes some parts of the internal circuitry.
Be careful not to cause accidental short-circuits!
The terminal detects the presence of program on the upgrade board and
displays the message .Upgrade software from board . The code
of the program to be installed is also displayed.
The subsequent messages indicate the successful deletion of the old
program, the loading and testing of the new program.
Should any error messages be displayed notify Assistance Service.
✔ Switch off the terminal;
✔ remove the upgrade board;
✔ replace the jumper P1 on the CPU board.
3-46
D400
Key
1. Jumper P1
2. Connectors for upgrade board
Figure 15.2 - CPU board (log0152.gif)
3-47
D400
Key
1. connectors on the upgrade board
2. connector on the CPU board (the other connector is not visible in the
figure)
Figure 15.3 - Installing the upgrade board (log0117.jpg)
3-48
D400
16. MAINTENANCE
WARNING
Accessing these menu options could delete all the data in
the terminal memory!
3-49
D400
3-50
D400
3-51
D400
3-52
D400
17. TROUBLESHOOTING
This chapter contains descriptions of the error messages that may be
displayed on the terminal along with possible remedies. See also
chapter Troubleshooting of the user manual.
3-53
D400
3-54
D400
3-55
D400
3-56
D400
where:
3-57
SOCIETÀ COOPERATIVA
BILANCIAI