15ppm Bilge Alarm: Features
15ppm Bilge Alarm: Features
15ppm Bilge Alarm: Features
■Features
● This product adopts an automatic operation method using sample water pressure which is matched to the operation of an oil
separator or pump.
● The automatic washing function of the detection unit can reduce troublesome manual washing work.
● An error code indication provided at occurrence of a device fault and a stain indication of the detection unit facilitate
maintenance and inspection. The product has a high-maintainability structure.
● Operation records such as trouble alarms can be saved in a memory card and confirmed as required.
● For alarms, a double safety method is used. When the oil concentration is 15 ppm or more, and When
sample water is defective or the instrument is defective, a alarm is output.
● Replacement of detecting part (sensor) separately sold can be accepted. as the completion of every five
Advanced device recognized
throughout the world.
This product conforms
to MEPC.107(49)
15ppm Bilge
Alarm
Product Specifications
2012-06 Rev.1.03 (T)
Alteration / correction hysteretic
-
Specifications
Water-proof grade IP = 54
Coating color 7.5BG 7/2 (Since this color is manufacturer standard, it cannot be changed.)
-
■ Outline dimensions ・ Material etc
251
116 1.0 4-Φ10 122
1
15ppm Bilge Alarm 注意
CAUTION
260
1.2
これらの注意を怠ると、機器の損
9 に注意して下さい。
If these attention are neglected,there is a fear
to cause damage,electric leakage,etc,of apparatus besides an obstacle in terms of functions may be found,so pay full attention.
CANADA NO,
USCG No.
EC
15ppm Bilge Alarm
型 式
MOD E L
製造-番 号
FOCAS-2000
2 SERIAL NO.
製造年月
20
3
DATE
7
4
165 4.0
TOKYO JAPAN
5
6
150 or more 116
4-Φ10
4-M8
Maximum
85 or
26
300 or more
100 or more
10
-
Internal Arrangement Drawing and the Terminal Base Arrangement
-
Spare parts model:FOCAS-2000 item
oil concentration detector FOCAS-2000 Hull No.
Spare Parts Table Case No.
2
No Part name Part No. Sketch Material Work- backup mass(kg) remarks
ing
Normal
Meltdown
1 Fuse for three-way Valve 2000-020 Glass tube 2 2 0.00075
type
2A
Fast-blow
Ceramic
2 Fuse for the equipment 2000-010 tube 1 1 0.00075 type
3.15A
0.01
5 Washing brush 2000-030 1
RC1/4×1
7 Φ10 joint 2000-050 C3604 3 0.05 RC1/8×2
DATE TITLE
CHECK DESIGN DRAWIN MAR. 09. 2012 FOCAS-2000
APPOINT
kikuchi Okumura Okumura Spare Parts Table
SCALE
DRAWING NO
8960-01
-
■ Pipe connections
※ 22
2m or less Overall lengthScrewed joint
(RC1/4 φ8 )
Within 1.5m
Do not install Bilge Alarm
than 2m down from the
Drain outlet
Bilge Separator.
head
Bilge Alarm
Hopper
Height of Drainage
Sample water inlet
Fresh
Be sure to set Bilge Alarm above the outlet water inletSeparator.
level of Bilge
(Equivalent to Mineral water, at 0.03 ~ 0.6MPa )
Allow the pipe upstand around Bilge Separator outlet to provide Sample water for testing.
(In accordance with MEPC.107(49) 6.1.1. )
Restrictions on piping
5. The length from the sample water outlet to the end of piping should be within 2m and the fall should be
within 1.5m. Be sure to provide hoppers.
6. Piping length between Bilge Separator and Bilge Alarm should be within 3m. Bilge Alarm should be set above
the Outlet height of Bilge Separator. Do not install Bilge Alarm lower than 2m below the top of Bilge
Separator. Operation pressure of Bilge Alarm is 0.03MPa –0.3MPa. In consideration of recommended pressure
for Bilge Separator, adjust the operation pressure.
Caution:
Be sure to follow the above instructions to run the oil concentration detector normally.