ESCO CO2 Incubator CCL-User Manual-V.u-feb2021
ESCO CO2 Incubator CCL-User Manual-V.u-feb2021
ESCO CO2 Incubator CCL-User Manual-V.u-feb2021
“Material in this manual is provided for informational purposes only. The contents and the product described
in this manual (including any appendix, addendum, attachment or inclusion), are subject to change without
notice. Esco makes no representations or warranties as to the accuracy of the information contained in this
manual. In no event shall Esco be held liable for any damages, direct or consequential, arising out of or related
to the use of this manual.”
i
Table of Contents
Introductory Page
i Table of Contents
iii Manual Revision History
vi Warranty Terms and Conditions
viii Introduction
viii 1. Products Covered
x 2. Safety Warning
x 3. Limitation of Liability
xi 4. European Union Directives on WEEE and RoHS
xi 5. Symbols
xii Declaration of Conformity
7 Chapter 2 – Installation
7 2.1 General Requirements
7 2.1.1 Location Requirements
8 2.1.2 Environmental Requirements
8 2.1.3 Power Requirements
9 2.2 Installation
9 2.2.1 Unit Leveling
9 2.2.2 Unit Stacking
9 2.2.3 ULPA Filter Installation
11 2.2.4 Shelves System Installation
11 2.2.5 Connecting the Unit to Electrical Power
11 2.2.6 Connecting the Unit to CO₂ and N2 Gas Supply
12 2.2.7 Connecting Alarm Contacts and RS485 Communication Port
12 2.3 Preliminary Cleaning
13 2.4 Filling the Humidity Pan
13 2.5 Filtered Air Exchange
14 2.6 Use of Additional Devices in the CO₂ Incubator
14 2.7 First Run
15 2.8 UV Lamp Functionality
15 2.9 Replacing Old CO₂ Sensor to the Latest IR Sensor
15 2.10 Disclaimer
CO2 Incubator
ii
25 3.2.5 Service
26 3.2.6 Help
26 3.3 Operation of CO₂ Incubator with Peltier Cooling System
26 3.3.1 Cooling Mode
26 3.3.2 Heating Mode
27 3.4 Diagnostic Readout
28 3.5 Alarm
35 Chapter 6 – Troubleshooting
42 Appendix
43 Appendix A: Log Record
CelCulture®
iii
SERIAL # OF 1 STUNIT
REV. NO. DATE RELEASED DESCRIPTION OF CHANGE REFERENCE FOR
IMPLEMENTATION
Add in CCL-50L and CCL-240L volume
D 12 June 2014 - -
size
Remove A series and upgrade CLM-A
E 8 January 2016 ECN-01-CO₂-2016 2016-106827
to CLM-B
F 12 July 2016 Update cooling incubator operational ECN-05-CO₂-2016 2016-110932
Disclaimer: CCL-050 unit has no
G 17 March 2017 drying function in its - -
decontamination cycle
H 10 May 2017 Add in CCL-UV model - -
Add in troubleshooting instructions - 2015-97750
for outer door alignment and use of
stainless steel hinge
7 September 2017 Add in instructions for changing to - 2015-97017
latest ULPA filter design
I Add in instructions for changing to - 2016-114541
latest IR CO 2 sensor
04 October 2017 Updated warranty policy - -
Updated preliminary cleaning - -
06 November 2017
instructions
18 November 2017 Updated device label - -
13 March 2018 Changed water recommendation - -
Updated directives for low voltage - -
16 March 2018
and electromagnetic compatibility
J
Add in information on compliance to - -
20 June 2018 RoHS in Electrical and Electronic
Equipment directives
Added use of filter adaptor for -
upgrading to current ULPA filter STB-CO2-25
K 27 July 2018 design configuration
Updated parameter recovery time for - -
standard models
Updated minimum height clearance - -
of unit from floor
Updated ambient temperature and - -
24 August 2018 humidity requirements
Updated condition for removal of - -
L
water from humidity pan in model
with Peltier cooling system
Updated the minimum distance of - -
27 August 2018 incubator from air conditioner or
other moving current
Change O₂ sensor to Zirconium
09 November 2018 ECN-CO2-18-003 2018-131940
dioxide
M Additional instruction for when a - -
heat-generating device is to be added
in the incubator
CO2 Incubator
iv
CelCulture®
v
CO2 Incubator
vi
Esco products come with a limited warranty. The warranty period will vary depending on the product purchased,
beginning on the date of shipment from any Esco international warehousing location. To determine which
warranty applies to your product, refer to the appendix below.
Esco's limited warranty covers defects in materials and workmanship. Esco's liability under this limited warranty
shall be, at our option, to repair or replace any defective parts of the equipment, provided if proven to the
satisfaction of Esco that these parts were defective at the time of being sold, and that all defective parts shall be
returned, properly identified with a Return Authorization.
This limited warranty covers parts only, and not transportation/insurance charges.
Factory installed, customer specified equipment or accessories are warranted only to the extent guaranteed by
the original manufacturer. The customer agrees that in relation to these products purchased through Esco, our
limited warranty shall not apply and the original manufacturer's warranty shall be the sole warranty in respect
of these products. The customer shall utilize that warranty for the support of such products and in any event
not look to Esco for such warranty support.
ALL EXPRESS AND IMPLIED WARRANTIES FOR THE PRODUCT, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTIES AND CONDITIONS OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED
IN TIME TO THE TERM OF THIS LIMITED WARRANTY. NO WARRANTIES, WHETHER EXPRESS OR IMPLIED, WILL
APPLY AFTER THE LIMITED WARRANTY PERIOD HAS EXPIRED. ESCO DOES NOT ACCEPT LIABILITY BEYOND THE
REMEDIES PROVIDED FOR IN THIS LIMITED WARRANTY OR FOR SPECIAL, INDIRECT, CONSEQUENTIAL OR
INCIDENTAL DAMAGES, INCLUDING, WITHOUT LIMITATION, ANY LIABILITY FOR THIRD-PARTY CLAIMS AGAINST
YOU FOR DAMAGES, FOR PRODUCTS NOT BEING AVAILABLE FOR USE, OR FOR LOST WORK. ESCO'S LIABILITY
WILL BE NO MORE THAN THE AMOUNT YOU PAID FOR THE PRODUCT THAT IS THE SUBJECT OF A CLAIM. THIS IS
THE MAXIMUM AMOUNT FOR WHICH ESCO IS RESPONSIBLE.
CelCulture®
vii
These Terms and Conditions shall be governed by and construed in accordance with the laws of Singapore and
shall be subject to the exclusive jurisdiction of the courts of Singapore.
Note: The warranty periods may vary by country. Contact your local distributor for specific warranty details.
For international distributors, warranty period starts two months from the date the equipment is shipped from
Esco facility. This allows shipping time so the warranty will go into effect at approximately the same time the
equipment is delivered to the user. The warranty protection extends to any subsequent owner during the
warranty period. Distributors who stock Esco equipment are allowed an additional four months for delivery and
installation, providing the product is registered with Esco. User can register their products online at
www.escoglobal.com/warranty or complete the warranty registration form include with each product.
Policy updated on 1st January 2015 (This limited warranty policy applies to products purchased on and after 1 st
January 2015)
CO2 Incubator
viii
Introduction
1. Products Covered
CelCulture CO₂ Incubator (Stainless Steel Chamber with Curved Door)
Model Size Sensor Control Filter Decontamination Electrical Rating
CCL-050B-8 50 L IR CO₂ N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-050B-8-SS 50 L IR CO₂ N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-050B-9 50 L IR CO₂ N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-050B-9-SS 50 L IR CO₂ N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-050T-8 50 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-050T-8-SS 50 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-050T-9 50 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-050T-9-SS 50 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170B-8 170 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-8-NF 170 L IR CO₂ N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-8-SS 170 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-9 170 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170B-9-NF 170 L IR CO₂ N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170B-9-SS 170 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-8 170 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-8-NF 170 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-8-SS 170 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-9 170 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-9-NF 170 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-9-SS 170 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-8 240 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-8-NF 240 L IR CO₂ N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-8-SS 240 L IR CO₂ N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-9 240 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-9-NF 240 L IR CO₂ N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-9-SS 240 L IR CO₂ N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-8 240 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-8-NF 240 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-8-SS 240 L IR CO₂/O2 N/A 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-9 240 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-9-NF 240 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-9-SS 240 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CelCulture®
ix
CCL-050T-9-IVF 50 L IR CO₂/O2 N/A 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170B-8-IVF 170 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-9-IVF 170 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-8-IVF 170 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-9-IVF 170 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-8-IVF 240 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-9-IVF 240 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-8-IVF 240 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-9-IVF 240 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CelCulture CO₂ Incubator with UV Lamp (Stainless Steel Chamber with Curved Door)
Model Size Sensor Control Filter Decontamination Electrical Rating
CCL-170B-8-UV 170 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-9-UV 170 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-8-UV 170 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-9-UV 170 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-8-UV 240 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-9-UV 240 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-8-UV 240 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-9-UV 240 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CelCulture CO₂ Incubator with UV Lamp (Stainless Steel Chamber with Flat Door)
Model Size Sensor Control Filter Decontamination Electrical Rating
CCL-170B-8-FD-UV 170 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170B-9-FD-UV 170 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-170T-8-FD-UV 170 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-170T-9-FD-UV 170 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240B-8-FD-UV 240 L IR CO₂ ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240B-9-FD-UV 240 L IR CO₂ ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CCL-240T-8-FD-UV 240 L IR CO₂/O2 ULPA 90oC Moist Heat 220 - 240 V, AC, 50/60 Hz, 1Φ
CCL-240T-9-FD-UV 240 L IR CO₂/O2 ULPA 90oC Moist Heat 110 - 130 V, AC, 50/60 Hz, 1Φ
CO2 Incubator
x
2. Safety Warning
• Anyone working with, on or around this equipment should read this manual. Failure to read,
understand and follow the instructions given in this documentation may result in damage to the unit,
injury to operating personnel and/or poor equipment performance.
• The use of any hazardous materials in this equipment must be monitored by an industrial hygienist,
safety officer or some other suitably qualified individual.
• Before you proceed, you should thoroughly understand the installation procedures and take note of
the environmental / electrical requirements.
• In this manual, important safety related points will be marked with the symbol .
• If the equipment is used in a manner not specified by this manual, the protection provided by this
equipment may be impaired.
3. Limitation of Liability
• The disposal and / or emission of substances used in connection with this equipment may be governed
by various local regulations. Familiarization and compliance with any such regulations are the sole
responsibility of the users. Esco’s liability is limited with respect to user compliance with such
regulations.
CelCulture®
xi
• Esco Voyager® is a PC-based software package developed for the remote monitoring, data logging, and
programming / device configuration of some models of Esco controlled environment laboratory
equipment.
If customers want to use their own PC-based software to control Esco equipment, Esco may, at its sole
discretion, after a non-disclosure agreement is executed, provide the Esco communication protocol to
the customer. Under no circumstances shall Esco be responsible for integration work, or providing
additional technical support, unless a special agreement is entered into separately.
Recommended method of disposal is according to The Federal, State and Local Governmental
regulations.
5. Symbols
The information and instructions in this manual may be accompanied by symbols shown in the table below. They
draw the reader's attention to important information on the use, safety and maintenance of incubator.
CO2 Incubator
xii
Declaration of Conformity
In accordance to EN ISO/IEC 17050-1:2010
has been certified independently to comply with the requirement of the following Harmonized Standard:
Low Voltage : EN 61010-1:2010
EMC : EN 61326-1:2013
More information may be obtained from Esco’s authorized distributors located within the European Union. A
list of these parties and their contact information is available on request from Esco.
____________________
Lim Lay Yew
Director, Esco
CelCulture®
1
Esco CelCulture® CO₂ Incubator is designed for high-quality incubation performance through the precise control
of heating and gas injections combined with multiple contamination control and ergonomic features appropriate
for safe incubation of many types of cells, tissues, and similar.
The external body of CelCulture® is electro-galvanized steel coated with silver-ion impregnated ISOCIDE™
antimicrobial powder coating which kills 99.9% of surface contaminants within 24 hours of exposure. The
incubator chamber is stainless steel type 304 with rounded corners for easy cleaning. Its perforated shelves,
shelves system, plenums and humidity pan are also made of shiny stainless steel. Interior components are easily
removed without using any tools. The inner door with tight gasket seal is heated to prevent condensation.
Heater foils directly mounted on the outer surfaces of the inner chamber helps maintain a relatively uniform
temperature with fast recovery time after door opening. A high-density insulation completely protects the
heating elements. It is then surrounded by an air jacket which separates the chamber from the varying ambient
parameters. The unit is equipped with a drift-free, non-dispersive infrared sensor which has temperature and
pressure compensating feature for longer-lasting reliability. For models with suppressed O₂ control, the unit is
equipped with a highly accurate zirconia O₂ sensor with non-depleting technology for longer lasting reliability.
Forced convection design accelerates humidification via humidity pan to prevent desiccation of cell samples. An
ULPA filtration system, which is 99.999% efficient at 0.1 µ to 0.2 µ particulates, is connected to the blower to
ensure proper circulation of an ISO Class 5 clean air standard without affecting cell growth negatively. It is
attached to a blower fan to ensure proper circulation of clean air and equal distribution of heat and gas at any
point in the incubator chamber.
Given the precise measurement and control of temperature and CO₂ level, as well as the multiple contamination
control methods of CelCulture® CO₂ incubator, it is especially suitable for growing mammalian cells at 37°C and
in the presence of 5% CO₂.
1.2 Labels
CO2 Incubator
2
10
11
CelCulture®
3
16
CO2 Incubator
4
11
12
13
14
15
8
16
9
10
17
CelCulture®
5
1
7
2
8
9
3
10
4
11
5
1. Power switch
2. Cooling fan*
3. Power inlet
4. RS485 communication port
5. Analog output (optional)
6. Alarm contacts
7. CO₂ gas inlet #1
8. CO₂ gas inlet #2 (for optional CO₂ backup system)
9. N2 gas inlet #1 (for suppressed O2 control units)
10. N2 gas inlet #2 (for optional N2 backup system in units
with suppressed O2 control)
11. Access ports**
CO2 Incubator
6
1
7
2
8
9
3
10
4
11
5
12
1. Power switch
2. Cooling fan
3. Power inlet
4. RS485 communication port
5. Analog output (optional)
6. Alarm contacts
7. CO₂ gas inlet #1
8. CO₂ gas inlet #2 (for optional CO₂ backup system)
9. N2 gas inlet #1 (for suppressed O2 control units)
10. N2 gas inlet #2 (for optional N2 backup system in units
with suppressed O2 control)
11. Access port
12. Peltier cooling module
CelCulture®
7
Chapter 2 – Installation
2.1 General Requirements
• The spacing between the unit and the side wall should be at least 100 mm (3.9”).
• The spacing at the back of the equipment should be at least 110 mm (4.3”) to allow connections to the
rear panel of the equipment.
CO2 Incubator
8
• The spacing from the floor must be ≥148 mm (5.8”). Never place equipment directly on the floor. It is
recommended to use the optionally available floor stand to elevate the unit from the floor.
Normally, units are rarely moved once they are in their ideal positions, but should there be a
need to relocate or repackage the unit, here are some considerations:
• It is recommended that risk assessment is carried out before the unit is moved.
• Before moving the unit, remember to decontaminate the incubator.
• Before moving the unit, remember to secure all moving parts (e.g. glass door) and remove
the water in the humidity pan to avoid spillage.
• Make sure to transfer all sample cells to another incubator.
• Unit is heavy so please carry out adequate workplace safety assessment before moving
the incubator. Do not lift using the door or door handle.
• Should only manpower is available (i.e. no suitable equipment), it will usually take 6 or
more people to move a unit manually.
• In relocating the unit through an elevated surface, ensure that the unit is held securely,
and should face the ramp side up, as shown below:
For repackaging:
• Place the unit on the pallet.
• Strap the unit body down to the pallet.
• Repackage as necessary. If possible, use original packaging.
• When moving the unit, use material handling equipment and lift the pallet.
CelCulture®
9
power supply must be used, otherwise damage may be caused to the unit or a hazardous situation may
occur.
• It is recommended that the equipment is connected to a dedicated power source with protective
grounding installed.
• There should be unobstructed access to the main power plug. The power plug is the main disconnecting
device on the unit.
• This equipment can sustain a maximum of ±10% nominal voltage fluctuation; otherwise a power
stabilizer is needed.
• The cable length must be shorter than 3 meters.
2.2 Installation
CO2 Incubator
10
Old ULPA filter configurations such as the collared and bracketed mounting types are now phased-out
due to issues encountered. Operators must check their units and change the ULPA filter mounting type
to the latest type (direct mounting) if their unit is still using any of the old types. Units without ULPA
filter (CCL-NF, 170 L and 240 L models only) can also be added with an ULPA filter with the latest
mounting type. Refer to the table below for the conversion kit needed when upgrading to the latest
filter design.
CelCulture®
11
Gasket: 1051440
No need for any kit
Filter: 1250081
CCL-240_-_ 1x Filter (1250081) + 1x Gasket
5170990 (1051440) + 2x Inline filter
(1050771)
3. Direct Mounting Filter: 1250081
Note: The Gasket (1051440)
Type No need for any kit
is just required to order if it
CCL-170_-_ is damage.
1x Filter (1250081) + 2x Inline
5170992
filter (1050771)
Filter: 1250081
Note: The Gasket (1051440)
No need for any kit
is just required to order if it
CCL-240_-_ is damage.
1x Filter (1250081) + 2x Inline
5170992
filter (1050771)
Note: Refer to the serial number tag located on the right side of the unit for electrical specification. Ensure the
unit is connected to the correct power source.
CO2 Incubator
12
If unit is equipped with the optional built-in gas backup system, there will be 2 gas inlets. Each of the inlet should
be connected to individual gas tanks as described above.
Note: Consult your facility safety officer to ensure that the equipment is installed in accordance to your local
regulations and code.
Caution: Avoid contact with CO₂ gas to eyes and skin. CO₂ and N₂ gases can have a suffocating effect in high
concentrations. During normal operations, small amounts of CO₂ and O₂ are released by the incubator to its
surroundings. Ensure that the room is well-ventilated to prevent suffocation of operators.
A set of relay contacts located on the rear of the unit is provided to monitor for
temperature, humidity, O₂ or CO₂ alarms. The terminals provided on the alarm contact
allow for a NO (normally open) output, a NC (normally closed) and COM (common)
connection. In the event of an alarm condition, the NO contact will close, and the NC
contact will open. Once the alarm is cleared, the contacts return to their normal condition.
The RS485 provides serial communication port for PC. It can be daisy chained from
product to product and connected to PC with Esco Voyager® Software.
Note: Please remove the adhesives (tape, covering, etc.) before performing the automated decontamination
cycle. Failure to do so will result to burnt marks or stains in the shelves and chamber.
Note: Metal surfaces can be cleaned using stainless steel cleaning agents. The inner glass door surfaces can be
cleaned using glass cleaners. Never use chlorine-based disinfectants. Do NOT spray disinfectant directly on the
control panel and on the holes for the electrical panel on the top-side of the unit.
CelCulture®
13
• See section 5.1 Cleaning Procedure for more information on cleaning the unit.
For best operation and longevity of the incubator, sterile distilled water should only be used to fill the humidity
pan. Never fill the pan with more than the indicated maximum level.
Note: Chlorinated tap water should not be used as chlorine can deteriorate the stainless steel. Tap
water may also have a high mineral content, which would produce a build-up of scale in the humidity
pan. Water should always be sterilized or treated with a decontaminant, safe for use with stainless
steel as well as safe for the product, prior to being introduced into the humidity pan. If sensitive cells
are being cultured, do not add decontaminant in the water.
Place the pan directly on the incubator floor to ensure optimum humidity and temperature response. The pan
needs to be inserted completely into the back of the incubator so that condensed water can drip back into
the pan and avoid flooding the bottom of the chamber. Ensure that no water is spilled on the bottom of the
chamber as this will potentially lead to condensation.
If water pan is filled with more than the indicated maximum level and / or not pushed to the back surface of the
chamber, condensation will occur.
It is recommended to check the water level and refill the humidity pan once a week.
For 170/240L Suppressed O2 model, ensure that the sparger is fully immersed in water.
When the incubator is turned off and will not be used for a period of time, empty the water pan and
dry it before putting it back inside the unit. Leaving the humidity pan with water when unit is
powered off will cause condensation to form in the chamber which will damage the sensors.
If high humidity level is required, the filtered air exchange can be removed and “Elevated Humidity” can be
activated in the control panel (call Esco or your distributor to activate this function).
When set to elevated humidity mode, the incubator may be able to achieve up to 85%-90% relative humidity
depending on the humidity level that the unit can achieve in default humidity mode (usually +2-3% of value from
default humidity).
CO2 Incubator
14
Use only the sampling port or the rear access port(s) for tubing or cable access. Never route any cable or tubing
through the door opening as it will disrupt the environment of the chamber.
1. Switch ON the unit. A welcome message will appear on the display. Press SET to continue.
2. Keypad operation will appear on the display. Press SET to continue.
3. Set the date, time, temperature set point, %CO₂ set point and %O 2 set point:
a. Set Date. The digit will blink. Use UP/DOWN buttons to select the current date and press SET
to confirm each digit.
b. Set Time. The digit will blink. Use UP/DOWN buttons to select the current time and press SET
to confirm each digit.
Note: To save power, the LCD backlight will automatically turn off from 11:00 pm to 6:00 am daily.
c. Set Temperature Set Point. Use UP/DOWN buttons to select the temperature set point. Press
SET to confirm.
d. Set %CO₂ Set Point. Use UP/DOWN buttons to select the %CO₂ set point. Press SET to confirm.
e. Set %O2 Set Point. Use UP/DOWN buttons to select the %O₂ set point. Press SET to confirm.
(For suppressed O2 model only)
f. Setting the tank reminder alarm (single tank for tank replacement prediction). Set NEW CO₂
TANK. Set the CO₂ gas net weight (refer to the gas catalog from the supplier). The unit is set to
0031Kg by factory default for a single gas tank for single incubator consumption. If 1 gas tank
supplies several incubators, divide gas net weight with the number of the supplied incubators.
Tank reminder alarm will activate to alert the user before gas supply is depleted completely.
4. Incubator set-up is done.
5. The unit will run a memory check.
6. The incubator will do an initial start-up and will continue to the main menu.
The outer door has a magnetic closure, which can be opened by pulling on the rim of the door. The inner glass
door has a tight gasket seal, which prevents the chamber from being contaminated and the chamber
environment from being disturbed.
When the outer door is opened the blower fan, heating, and gas injection will automatically turn off.
Note: Make sure that the door switch will not be overridden or controlled manually or by any means when the
outer door is opened. During normal operations, never leave the door open as it will cause the incubator to
overheat and may pose a fire hazard.
Allow 2 hours for the incubator to stabilize before loading the unit with samples or before proceeding to
calibration.
Note: When loading the unit, ensure proper spacing between the samples to allow uniform air circulation inside
the chamber. Never place any item touching the side walls of the incubator as it may take a longer time to reach
the set temperature. For your safety, never load samples with unknown or potentially harmful composition.
CelCulture®
15
Always use the humidity pan and humidity pan cover to prevent UV light from escaping.
To check whether the UV lamp is lit, activate the test mode by pressing up, up and set button simultaneously.
Activate it by switching the UV to ON mode. Open the outer door and then press the door switch with the inner
glass door closed. The appearance of visible blue light from the front of the humidity pan cover confirms that
the UV lamp is lit.
UV functionality check must be conducted within less than 10 seconds to ensure zero
risk of UV radiation. UV light is harmful to the eyes, so do not light the UV lamp when
the inner door or humidifying pan cover is open.
Note: Replacement of the old sensor with the new one must only be undertaken by qualified service personnel
to prevent damage to the equipment.
2.10 Disclaimer
Each of the Esco units is subjected to an in-depth inspection prior to delivery. Signs of use caused by tests are
possible and have no influence on the performance and function of the units. The performance and safety of
the incubator, while rigorously evaluated at the factory, cannot be guaranteed after transit and installation.
Therefore, on-site testing is always recommended.
CO2 Incubator
16
DECON LED HEAT LED oC/%RH Display MENU Button Main LCD UP Button ALARM LED
DECON Button INJECT LED %CO2/%O2 Display SET Button DOWN Button MUTE Button
MENU Button
• When the main LCD shows normal display, pressing the MENU button will display the menu options.
• Within the menu display, pressing the MENU button will bring back the previous menu level.
SET Button
• Within the menu display, pressing the SET button will confirm a selection or value.
UP/DOWN Buttons
• Within the menu display, pressing the UP and DOWN buttons will scroll the display up and down.
• When an input is required, pressing the UP and DOWN buttons will increase and decrease a value.
MUTE Button & ALARM LED
• MUTE button will mute the audible alarm for a period of time.
• The red ALARM LED will light up whenever an alarm condition is triggered.
HEAT LED
• HEAT LED will light up whenever a heating process is activated.
INJECT LED
• INJECT LED will light up whenever a gas injection process is activated.
o
C/%RH Display
• When oC is displayed, the display indicates the temperature in the chamber.
CelCulture®
17
• When %RH is displayed, the display indicates the relative humidity in the chamber (optional).
%CO₂/%O2 Display
• When %CO₂ is displayed, the display indicates CO₂ concentration level in the chamber
• When %O2 is displayed, the display indicates O2 concentration level in the chamber (for suppressed O 2
model only)
CO2 Incubator
18
NEW PIN
AUDIBLE ALARM
SET DATE
SET TIME
SET PIN
SET TEMP/%CO2
Note: Menu in the darker boxes are shown in special circumstances.
MORE HELP
CelCulture®
19
3.2.1 Settings
3.2.2 Calibration
Place the reference instrument or its measuring probe on the center of work zone. Route the connecting cable
through the access port located at the rear of work zone, and highly to seal the access port tightly once cables
are placed there. Close doors and wait until the value displayed in the reference instrument is stabilized. Record
the measured value.
Enter MENU|CALIBRATION and select the responding options. When asked, enter the measured value. Press
SET to confirm.
When calibrating the zirconia O₂ sensor: If the difference value between the incubator’s display reading and
the reading from the reference instrument is ≥1.0%, it is recommended to do resetting of the O₂ sensor following
the instructions below.
CO2 Incubator
20
Resetting the O₂ sensor: When resetting the O₂ sensor, prepare an electronic gas analyzer and the magnetic
rear access port plug then follow below procedure:
By default magnetic rear access port plug for By default magnetic rear access port plug for 50L
170/240L is installed directly to the unit from the is NOT installed directly to the unit from factory.
factory. This plug can be found in retrofit kit box.
For 50 L models:
1. Disconnect CO2 and N2 gas injection from the unit. No gas allowed during doing reset calibration.
2. Open the outer door and the inner glass door to allow the temperature in the chamber to be
similar to the ambient temperature (e.g. ambient temperature is 25°C, then recommended
incubator temperature on display should be around 24°C to 26°C).
3. Once temperature has stabilized, remove the original plug and install the rear access port plug
with magnet to the left rear access port hole inside the chamber for 10 seconds then check the %O₂
actual display on the incubator control panel which should have changed to 17.7% to 18%.
CelCulture®
21
4. Remove the rear access port plug with magnet and install back the original plug without magnet to
the left rear access port hole and close the chamber doors. Set the %CO₂ and %O₂ to 5.0% and
allow gas to be injected.
5. Allow the unit to stabilize for 1 hour, then measure the %CO₂/%O₂ concentration by inserting
around 10cm of electronic gas analyzer tubing through the sampling port on the glass door of the
chamber. Turn on the pump of the gas analyzer and wait for 5 minutes when the O₂ reading has
stabilized.
6. Check if there is a difference between the readings from the gas analyzer and from the incubator
actual display. Set the O₂ offset (MENU > CALIBRATION > %O₂ SENSOR > CALIBRATE) when needed.
1. Disconnect CO2 and N2 gas injection from the unit. No gas allowed during doing reset calibration.
2. Open the outer door and the inner glass door to allow the temperature in the chamber to be similar to
the ambient temperature (e.g. ambient temperature is 25°C, then recommended incubator
temperature on display should be around 24°C to 26°C).
3. Once temperature has stabilized, install the rear access port plug with magnet from left side to the
right rear access port hole inside the chamber for 10 seconds. Check the %O₂ actual display on the
incubator control panel which should have changed to around 17.7 % - 18%.
4. Remove the rear access port plug with magnet and install back to the left rear access port hole and
close the chamber doors. Set the %CO₂ and %O₂ to 5.0% and allow gas to be injected.
CO2 Incubator
22
5. Allow the unit to stabilize for 1 hour, then measure the %CO₂/%O₂ concentration by inserting around
10cm of electronic gas analyzer tubing through the sampling port on the glass door of the chamber.
Turn on the pump of the gas analyzer and wait for 5 minutes when the O₂ reading has stabilized.
6. Check if there is a difference between the readings from the gas analyzer and from the incubator actual
display. Set the O₂ offset (MENU > CALIBRATION > %O₂ SENSOR > CALIBRATE) when needed.
To reset the temperature, CO2, and RH sensors to its default calibration, select DEFAULT CALIBRATION. Press
SET button to confirm.
3.2.2.2 Analog Output Calibration (for units with optional analog output)
A set of relay contacts are provided at the rear of the incubator to allow the unit to output analog signals
representing the temperature, %CO₂, %O2 (for suppressed O2 model), and %RH (for units with optional humidity
display) depending on the options available in your incubator. This allows the chamber to be connected to an
in-house data acquisition or alarm system.
The analog data output can be set to operate in either DC (0-5V) or current (4-20mA) mode. The factory default
setting is voltage. The voltage of the analog output can be calibrated using a calibrated digital multimeter.
Place the multimeter’s measuring probe on the responding + and - terminals of the analog output located on
the rear panel and record the measured value.
Enter MENU|SERVICE and select the responding options. When asked, enter the measured value. Press SET to
confirm.
3.2.3 Admin
CelCulture®
23
3.2.3.5 New CO₂ Tank (for unit without the optional gas backup switch system)
The CO₂ tank depletion reminder automatically calculates how much CO₂ gas is left in the tank. It alerts the user
one week before the gas is depleted which gives the user some buffer time to order new tanks.
MENU ADMIN NEW CO2 TANK XXXX
CO2 Incubator
24
optional %RH display), the relay will make or break to signal to a remote device that the incubator is in a faulty
condition.
TEMP ALARM RELAY ENABLE | DISABLE
SPANISH
FRENCH
ITALIAN
CelCulture®
25
3.2.3.23 Set RS-485 Address (for unit with optional Voyager® monitoring software)
User can set the RS-485 address for PC connection.
MENU ADMIN SET RS-485 ADDRESS 1 ≤ X ≤ 32
3.2.5 Service
CO2 Incubator
26
3.2.6 Help
This menu section will show brief explanation on the basic control of the unit.
3.3 Operation of CO₂ Incubator with Peltier Cooling System (applicable for CCL-P
series)
Note: The incubator is working according to the room temperature. Stable room/ambient temperature is
therefore required. When the temperature set point is equal to or below the ambient temperature, put empty
humidity pan in the chamber or if it has water, drain first and put back the dried pan inside the chamber.
CelCulture®
27
CO2 Incubator
28
3.5 Alarm
The CelCulture® CO₂ Incubator alarm system is shown in the table below. When an alarm is active, the error message appears
in the display of the controller. Pressing MUTE button will disable the audible alarm for the ring back period. However, the
visual alarm continues until the incubator returns to a normal condition.
CelCulture®
29
Cleaning instruction for Cu model: Copper (Cu) chamber interior, after a period of time, will have discoloration.
Use vinegar, baking soda or lemon soap to clean the chamber surface. About 1 tablespoon of salt per 1 cup of
vinegar can be used to remove tarnishes. If there is a patina (greenish stain), use smooth/fine sandpaper to
remove the stain.
CO2 Incubator
30
CelCulture®
31
Gently unhook the back of the plenum and pull it out from the chamber.
CO2 Incubator
32
Unplug the rear access port plug and remove the inlet filter.
Plug the rubber cover at the left and right access port.
b. Remove the shelf system. See section 2.2.4 Shelves System Installation.
Items:
1. Back Pilaster
2. Front Pilaster
3. Railing
4. Tray or Shelf
Procedure:
1. Remove the tray or shelf.
2. Remove the railing by removing the front end
upward from the front pilaster and moving the
back end backward from the back pilaster.
3. The pilaster can be removed by unscrewing
the thumb screws holding the pilasters.
c. Remove the sparger and tubing (for suppressed O₂ model) and the humidity pan.
Note: When cleaning and disinfecting, always observe the safety instructions and hygiene regulations.
a. Thoroughly remove dirt and residues from the surfaces of the work space using the mild soap
solution and a sponge, or spray an appropriate disinfectant onto work space surfaces and all
dismantled parts. Allow disinfectant to react as specified by manufacturer.
Note: Do not spray disinfectant directly onto sensors, on the control panel, and on surfaces near the
electrical panel to prevent damage to electrical parts. Use a cloth soaked with disinfectant to wipe the
control panel and the exterior body.
CelCulture®
33
b. Rinse the soap solution or disinfectant twice using the distilled water in a wash bottle. If the
part is hard to rinse with a wash bottle, use a wet sponge to remove the soap.
c. Dry all rinsed part with a clean lint-free cloth or paper towel.
d. Wipe the interior components and exterior surfaces with 70% ethanol then allow to dry.
7. Re-assemble the unit by reversing step 5. Ensure that the equipment has completely dried before
commencing normal operation.
Note: Surface of a CO₂ incubator with copper chamber (CCL-Cu) is prone to dirt & fingerprint marks (once touched with
hands) and has a tendency to give a greenish appearance when oxidation occurs (upon contact with carbon dioxide).
Clean regularly with lemon soap, vinegar or baking soda. About 1 tablespoon of salt per 1 cup of vinegar can be also
used to remove tarnishes. If the surface or the humidity pan turned green, use sand paper. Contact your Local Distributor
for the available passivation solution to use.
5.2 Decontamination
Decontamination in CelCulture® CO₂ incubator uses 90°C moist heat to kill organisms that commonly
contaminate workspace including all installed components. The entire program run of the decontamination
routine takes approximately 15 hours (overnight process). The chamber is cool and dry at the end of the cycle
(applicable to CCL-170L/240L units only). No further wipe down is necessary.
Note: CCL-50L unit does not have a decon pump to dry the chamber after the decontamination cycle. Further wipe down
of the chamber is required for CCL-50L models since condensation will normally occur after the cycle.
Note: The incubator must not be used to decontaminate objects other than the incubator and its
components.
Back
• Keep the ULPA filter inside the incubator during decontamination (for CCL-170L/CCL-240L units only).
• CCL-50L unit has no top duct, no ULPA filter installed, and has a small blower fan for forced convection.
No need to remove blower during decontamination.
CO2 Incubator
34
• Replace water from the humidity pan with distilled water according to the following volumes:
• Once the decontamination is done, fill in the humidity pan (400 ml water for CCL-050L and 800 ml for
CCL-170/240L,) and reinstall the rear access plug (CCL-170/240L).
CelCulture®
35
Chapter 6 - Troubleshooting
Under no circumstances should the unit be disassembled, repaired, or converted by unqualified people. Failure
to follow these instructions may cause personal injury and/or loss of property due to malfunctions, electric
shocks or fire.
Should the unit malfunction, turn off the incubator and disconnect the power supply. Continuing to operate the
unit when it shows signs of malfunctioning may cause electric shocks or fire.
The following troubleshooting guidelines address some of the basic problems that can be solved by the user or
operator. However, if the problem persists, please contact your local distributor.
CO2 Incubator
36
Air leakage through Inner door • Check installation of the gasket to ensure that it is properly secured to the unit and that there
gasket. is no gap.
• Check the joining ends of the gasket to make there is no gap. Use silicon sealant to seal the
joins if there is a gap.
• Check gasket profile is not deformed or torn. Otherwise, replace new door gasket.
Defective mainboard • Check if the triac heater is not defective. Measure the voltage (ON/OFF testing under TEST
MODE). Must be <5 V when OFF.
Otherwise, replace the mainboard.
Small fan position too far to O2 • Move closer the small fan to O2 sensor housing
sensor
Problem 4: Humidity level is below standard humidity level (For units with RH display options installed)
Possible Cause Corrective Actions
There is no water or not enough • Ensure there is enough water in the humidity pan. For normal operating temperature,
water in the humidity pan. humidity pan should be half filled.
CelCulture®
37
Problem 12: Condensation occurs on the surface of inner door after decontamination process (For CCL-170
and CCL-240 units)
Possible Cause Corrective Actions
Sampling port on inner door is • Ensure the sampling port is capped.
not installed or capped.
Leakage through magnetic gasket • Check to make sure outer door can close fully and no gap.
on outer door. • If there is gap and cannot be closed properly, check that magnetic strength is not diminished
and PVC gasket profile is not deformed. Otherwise, replace with new magnetic gasket.
Decontamination pump failed. • Contact local distributor.
Filtered air exchange not working • Check whether filtered air exchange is installed.
or not installed. • There is a short tubing with a 1 μm filter attached to the access port located at the back of
the chamber.
• Check that filter is not saturated and tubing is not blocked.
• Remove the ULPA filter and make sure the tube is connected to the hole in the blower intake,
above where the ULPA filter is mounted.
• Check also if 2nd filter is not saturated and tubing is not blocked.
• Loosen the 2 thumb screws at the front of the top duct and lower the top duct carefully.
Check if the tubing is secured at the top of the chamber and tubing is not blocked. Re-secure
any loose tubing connection and close the top duct.
CO2 Incubator
38
Problem 16: Incubator have a severe condensation on glass door and inside chamber during cooling mode
Possible Cause Corrective Actions
Humidity pan inside chamber • Take out or dry out the humidity pan.
Problem 17: Incubator cannot reach temperature 8°C below ambient during cooling mode (CCL-P model)
Possible Cause Corrective Actions
Humidity pan filled with water • Remove humidity pan from the chamber, remove top duct during cooling mode.
Temperature ambient is too high. • Turn on air conditioning.
Room air conditioning is OFF • Set temperature ambient minus 6-8 above set point.
(If incubator set point is 15°C, so the ambient temperature should be 23°C-24°C)
• Go to TEST MODE, choose O2 VALVE2. Press ON and OFF to check the function.
Cooling control malfunction See the respond of temperature display decrease gradually,
• Contact local distributor if the temperature is not responding and decreasing
• Hear the sound of the cooling module running, otherwise contact local distributor.
Defective Peltier cooling module • Contact local distributor.
Problem 18: Incubator have severe condensation during cooling mode (CCL-P model)
Possible Cause Corrective Actions
Humidity pan filled with water • Remove and take out humidity pan when unit in cooling operation mode.
• Open glass door regularly to balance and clip the humidity.
• Contact local distributor if problem persists.
Problem 19: CO₂ is not injected, display shows 0.0% or % CO₂ display not responding
Possible Cause Corrective Actions
Defective CO₂ inject valve
• Change CO₂ valve, otherwise contact local distributor.
Defective valve channel from I-
• Change mainboard, otherwise contact local distributor.
series mainboard
CelCulture®
39
Problem 21: O₂ is not injected, display shows 0.0% or % O₂ display not responding
Possible Cause Corrective Actions
Defective N₂ inject valve
• Change N₂ valve, otherwise contact local distributor.
Defective valve channel from I-
• Change mainboard, otherwise contact local distributor.
series mainboard
The tubing connected to N₂ is
• Do correction for tubing
bent
CO2 Incubator
40
CelCulture®
41
Contamination Control
1) Main body is electro-galvanized steel with ISOCIDE™ antimicrobial coating;
2) 90˚C moist heat OVERNIGHT decontamination cycle (HPA validated);
3) ULPA filter********
Contamination Control Methods
4) 0.2-micron inlet filter for gas inputs;
5) 1-micron air circulation filter
6) Optional copper chamber and shelves
All data recorded is specified for standard models with unloaded chambers and tested under optimum factory setting conditions of
23°C and 60% ambient humidity.
*Results are achieved when tested at 37 0C as set point. Results may vary if set point changes and calibration is needed.
**For temperature not exceeding 37.2°C.
***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.
****For CO₂ level not exceeding 5.2%.
*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.
****** For O₂ level not lower than 4.8%.
******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%.
******** Not available for 50 L models.
CO2 Incubator
APPENDIX
APPENDIX A: LOG RECORD
Thermostatic Unit : __________________________________________________________
Serial Number : __________________________________________________________
Person in-charge : __________________________________________________________
1. This log record should be used by the operator to record any new agent/virus/bacteria/protist that has
been introduced to the unit during its operation, problems encountered, etc.
2. Any decontamination procedure performed by either the user or the technician should be recorded
down as well.
3. Please also record any major maintenance procedure performed by the service technician, for example:
parts replacement, recertification, etc.