Technical specifications

Measurement of relative humidity


Measurement range 0…100%RH
HD2101.1 HD2101.2
Resolution 0.1%RH
Accuracy ±0.1%RH
Drift after 1 year 0.1%RH/year
Measurement of temperature
Pt100 measurement range -200…+650 °C
Pt1000 measurement range -200…+650 °C
Resolution 0.1°C
Accuracy ±0.1°C
Drift after 1 year 0.1°C/year
Measuring unit °C - °F - %RH - g/kg -
g/m3 - hPa - J/g

Measured values storage - model HD2101.2


Type 2000 pages containing
19 samples each
Quantity Total of 38000 samples
Storage interval 1,5,10,15,30 s;
1,2,5,10,15,20,30 min;
1 hour

Security of stored data Unlimited, independent


of battery charge conditions

Power
Batteries 4 1.5V type AA batteries
Autonomy 200 hours with 1800 mAh
alkaline batteries
Power absorbed with instrument off 20μA
Mains 12Vdc / 1000 mA output
mains adapter
Serial interface RS232C
HD2101.1, HD2101.2
HYGRO- THERMOMETER Type RS232C electrically isolated
HANDHELD - COMMUNICATION / DATA LOGGING Baud rate Can be set from 1200
to 38400 baud
HD2101.1 and HD2101.2 are portable instruments with LCD display. They Data bit 8
measure relative humidity and temperature with combined probes.
Temperature only is measured by Pt100 or Pt1000 immersion, penetration air Parity None
or contact probes. Stop bit 1
When the humidity/temperature combined probe is connected, the instrument
calculates and displays the absolute humidity, the dew point, the partial vapour Flow Control Xon/Xoff
pressure, the wet bulb temperature, the mixing ratio, the enthalpy and the Serial cable length Max 15m
comfort indices.
The probes are fitted with an automatic detection module, with the factory Print interval Immediate or selectable among
calibration data already stored inside. 1,5,10,15,30 s;
1,2,5,10,15,20,30 min;
The instrument HD2101.2 is a data logger. It stores up to 38,000 samples which 1 hour
can be transferred into a PC connected to the serial ports RS232C and USB 2.0 USB interface - model HD2101.2
or into a portable printer.
Type 1.1 - 2.0 electrically isolated
The storing interval, printing, and baud rate can be configured using the menu.

The Max, Min and Avg function calculate the maximum, minimum or average
values.
Other functions include: the relative measurement REL, the HOLD function and
the automatic turning off (excludable).

The instruments have IP66 protection degree.

HD2101.2 CP23

7
ORDERING CODES
Connections
Input module for the probes 8-pole male DIN45326 connector HD2101.1: The kit is composed of the instrument HD2101.1, 4 1.5V alkaline
Serial interface 8-pole Mini-Din connector batteries, operating manual, case and DeltaLog9 software downloadable
from Delta OHM website. Probes and cable must be ordered separately.
USB Interface Mini USB type B
Mains adapter 2-pole connector (positive at centre) HD2101.2: The kit is composed of the HD2101.2 datalogger, 4 1.5V alkaline
batteries, operating manual, case, USB cable CP23 and DeltaLog9 software
Operating conditions downloadable from Delta OHM website. The probes and cable must be
Operating temperature -5…50 °C ordered separately.
Storage temperature -25…65°C HD2110CSNM: 8-pole connection cable Mini-Din - Sub D 9-pole female for
Working relative humidity 0…90%RH without condensation RS232C.
Protection degree IP66
C.206: Cable for instruments series HD21..1 to connect directly to the USB
Instrument Technical Characteristics Input of a PC.
Dimensions
185x90x40mm SWD10: Stabilized 230Vac/12Vdc-1000 mA mains adapter.
(Length x Width x Height)
Weight 470g (complete with batteries) HD40.1: Portable, serial input, 24 column thermal printer, 58mm paper width.
Materials ABS, rubber It uses the cable HD2110 CSNM (optional).

Display 2 rows 4½ digits plus symbols For all suitable probes, see from pag.9 onwards
Visible area: 52x42mm
Time Accessories

Date and time In real time HD75: Saturated solution at 75.4%RH@20 °C for calibration of relative humidity
Accuracy 1min/month max drift probes, fixing adapter M24x1.5, M12x1.
HD33: Saturated solution at 33.0%RH@20 °C for calibration of relative humidity
probes, fixing adapter M24x1.5, M12x1.
HD11: Saturated solution at 11.0%RH@20 °C for calibration of relative humidity
probes, fixing adapter M24x1.5, M12x1.

Protection for humidity probes Ø 26, thread M24x1,5


P1: Technopolymer and 34μm stainless steel grid protection. Operating
temperature: -40…80 °C.
P2: Technopolymer and 20μm sintered PE protection Operating temperature:
-40…80 °C.
P3: 20μm sintered bronze protection Operating temperature: -40…150 °C.
P4: 20μm sintered PE protection. Operating temperature: -40…80 °C.

Protection for humidity probes Ø 14, thread M12x1


P6: 10μm sintered stainless steel protection. Operating temperature: -40…180 °C.
P7: 20μm PTFE protection. Operating temperature: -40…150 °C.
P8: PBT and 10μm stainless steel grid protection. Operating temperature:
-40…120 °C.

A The portable data loggers HD2101.2 are equipped with HID (Human
Interface Device) type USB port with mini USB connector.

For the connection to a PC with the CP23 cable it is not necessary to load
USB drivers.

B For the connection of the models HD2101.1 to the USB port of a PC, the
C.206 USB/serial converter is necessary. The converter is supplied with its
own drivers which must be installed before the connection of the converter
to the PC.

C The port with the Mini-DIN connector is a serial port type RS232C. The serial
port RS232C of a PC or the printer HD40.1 can be connected by the cable
HD2110CSNM.

8
COMBINED DEW POINT AND TEMPERATURE PROBES WITH SICRAM MDODULE
RANGE
CODE SENSORS DIMENSIONS
RH - TEMP
170

Ø 26
HP472ACR

0...100% RH
-20 °C...+80 °C 130 120

Ø 14
HP473ACR

130 215

Ø 14
HP474ACR

0...100% RH
-40 °C...+150 °C 110 560
RH
Ø 14
HP475ACR Pt100

480

0...100% RH
HP475AC1R Ø 14
-40 °C...+180 °C

110 520

0...100% RH 18x4
HP477DCR
-40 °C...+100 °C

130

0...100% RH Ø 14
HP478ACR
-40 °C...+150 °C

0…100%RH For the technical specifications of these probes please


HP480 / HP481 Pt100
-40…+60 °C see page 12-13

PROBES COMMON CHARACTERISTICS


Relative humidity
Sensor Capacitive
Temperature drift @ 20 °C Max 0.02%RH/°C
Response time %RH at constant temperature 10 sec (10÷80%RH; air speed=2m/s) at constant temperature
Temperature with sensor Pt100
Temperature drift @20 °C 0.003%/°C

Accuracy
±1.5% RH (0...90%RH) ±2.0% RH (90...100%) @ T=15...35°C
%RH
±(1.5 + 1.5% of the displayed value)% RH in the remaining temperature range
Temperature ± 0.3°C

PROBES PROTECTION
P1 P2 P3 P4 P6 P7 P8
Operating
Temperature
-40…80 °C -40…80 °C -40…150 °C. -40…80 °C -40…180 °C -40…150 °C -40…120 °C

Technopolymer and Technopolymer and 10μm sintered PBT and 10μm


20μm sintered bronze 20μm sintered PE 20μm PTFE
Material 34μm stainless steel 20μm sintered PE
protection protection
stainless steel
protection
stainless steel grid
grid protection protection protection protection

View

Technical
suitable for probes Ø 26 - thread M 24x1,5 suitable for probes Ø 14 - thread M 12x1
Spec.
9
Notes and warnings:

I. Keep salt solutions in the dark at a temperature of about 20 °C.


HD75, HD33, HD11 II. Salt solutions are effective and can be used as long as there is salt to be
melted as well as liquid inside them. As a rule, in 33% RH solution make sure
that there is some solid salt left, while in 75%RH solution make sure that
there is some liquid left or salt is wet.
III. For better results, the temperature of the probe and that of the saturated
solution must be as close as possible. Do not forget that plastic materials
are bad conductors of heat. Any difference of tenths of degree between the
sensor and the saturated salt solution leads to errors of RH points.
IV. Do not touch the sensitive element with your hands or other objects .
Scratches and dirt alter the instrument measurement and may damage the
sensor.
V. The measurement chamber must be closed, otherwise the equilibrium can-
not be reached.
Thoroughly screw the probe to the bottle.
VI. The check or calibration sequence for Delta OHM instruments or transmit-
ters is always as follows:
first solution: 75% RH
second solution: 33%RH
third solution: 11% RH (if any)
No sequence is compulsory for checking the sensor.
VII. To calibrate or set up the instrument, follow the instruction manual of the
instrument that you are using.

VIII. If you check, set up or calibrate the instrument at a temperature of other


than 20 °C, see the following table to find out the equilibrium relative
humidity reference value of the salt solution corresponding to the working
temperature. In this table, you will find the saturated salt relative humidity
variation when temperature changes.

HD75, HD33, HD11 Equilibrium relative humidity of selected saturated salt solutions from 0 to 100
RH CALIBRATION - SALT SATURATED SOLUTION °C
Magnesium
Temp. °C Lithium Chloride Sodium Chloride
For checking, setting up or calibrating instruments with relative Chloride
humidity sensors 0 11.23 ± 0.54 33.66 ± 0.33 75.51 ± 0.34
5 11.26 ± 0.47 33.60 ± 0.28 75.65 ± 0.27
Before starting 10 11.29 ± 0.41 33.47 ±0.24 75.67 ± 0.22
15 11.30 ± 0.35 33.30 ± 0.21 75.61 ± 0.18
1. Make sure that inside the chamber containing the saturated salt solutions
there are at the same time: 20 11.31 ± 0.31 33.07 ± 0.18 75.47 ± 0.14
• solid salt 25 11.30 ± 0.27 32.78 ± 0.16 75.29 ± 0.12
• liquid solution or wet salt 30 11.28 ± 0.24 32.44 ± 0.14 75.09 ± 0.11
2. The instrument and the saturated solutions to be used are to be kept in an
environment at stable temperature while checking or calibrating them. 35 11.25 ± 0.22 32.05 ± 0.13 74.87 ± 0.12
3. Wait for at least a couple of hours at stable temperature so that the 40 11.21 ± 0.21 31.60 ± 0.13 74.68 ± 0.13
instrument and the salt solutions reach thermal equilibrium with the 45 11.16 ± 0.21 31.10 ± 0.13 74.52 ± 0.16
environment.
4. Unscrew the cap of the first saturated salt solution to be used for checking 50 11.10 ± 0.22 30.54 ± 0.14 74.43 ± 0.19
or calibrating the instrument. Use: 55 11.03 ± 0.23 29.93 ± 0.16 74.41 ± 0.24
• for probes with thread M24X1,5, the bottle threaded hole M24X1,5 directly; 60 10.95 ± 0.26 29.26 ± 0.18 74.50 ± 0.30
• for probes with thread M12X1, the supplied adapter M24X1,5 / M12X1.
5. If there is any liquid inside the measurement chamber, dry it with clean 65 10.86 ± 0.29 28.54 ± 0.21 74.71 ± 0.37
absorbent paper. The uncertainty of the solution or measurement is not 70 10.75 ± 0.33 27.77 ± 0.25 75.06 ± 0.45
influenced by any liquid left inside the measurement chamber. 75 10.64 ± 0.38 26.94 ± 0.29 75.58 ± 0.55
6. Screw the probe to the bottom of the thread; do not touch the sensitive
element with your hands or any other object or liquid. 80 10.51 ± 0.44 26.05 ± 0.34 76.29 ± 0.65
7. The temperature of the salt solution and that of the sensor must be the 85 10.38 ± 0.51 25.11 ± 0.39
same or very close. Once the sensor is inserted, wait for at least 30 minutes. 90 10.23 ± 0.59 24.12 ± 0.46
8. Connect the probe to the instrument or transmitter. Power or turn them on
as per instructions. 95 10.07 ± 0.67 23.07 ± 0.52
9. After 30 minutes, start the calibration procedure for the first calibration 100 9.90 ± 0.77 21.97 ± 0.60
point according to the instruction manual of the specific instrument.
10. Once you have checked, set up or calibrated the first point, take the probe
out of the bottle and put the cap back on the bottle. Make sure you do not
mix it up with that of other saturated solutions.
11. Repeat points 1, 2, 3 and 4 to perform the second calibration point with the
second saturated solution.

10
1 2 3 4

1 2 3 4

5 6 7

11
HP480, HP481

Graph 1: accuracy of the dew point measurement (DP)

Typical installation HP481

HP481 is a combined relative humidity and temperature probe suitable for


in-line installation. The probe can be used in pressurized pipes, or in which
vacuum is required. It is equipped with a G ½” threading for the connection
to the system and can be installed in any position. The probe is equipped
with a sintered steel filter.

Connections

The probes can be connected to the pipe in three ways:


A. by using the measuring chamber with a quick coupling (only HP480);
B. by using the measuring chamber with a threaded G 1/4” connection (only
HP480);
C. directly (without measuring chamber) with a threaded G 1/2” connection.

Connection with measuring chamber and quick coupling:

HP480, HP481 - PROBES FOR TEMPERATURE, RELATIVE


HUMIDITY AND DEW POINT MEASUREMENT IN PIPES.

Compressed air is used for several purposes, many of which require


compressed air with a low humidity level, and so comes the need to know
the dew point (DP) of water vapour in the compressed air that circulates
in the system. The HP480 and HP481 probes are designed specifically for To connect with quick coupling, you can use one of the standard 1/4” cou-
this purpose.
plings provided. Other couplings than those supplied can be used, pro-
The use of dew point measurement in order to limit moisture in compressed vided that they have a G 1/4” thread on the side that fits into the probe.
air distribution systems has many advantages:
• prevents corrosion of metal pipes; Connection with measuring chamber and threaded connection:
• in cold areas, prevents the formation of ice inside the pipes leading to
obstruction of the pipes themselves;
• prevents bacterial growth in plants for medical use
• reduces maintenance costs of pneumatic drives, maintaining the proper
lubrication of moving parts;
• improves the quality of products coming into contact with air, for example
in the drying process of granulates.

Typical installation HP480

The probe can be installed in any position. The connection to the com-
pressed air can be achieved with a threaded connection or with a quick
connection.
For the connection by threaded coupling, the connection must have an
The connection allows for quick installation and removal of the probe external G 1/4” thread on the side which will be placed in the probe. The
without stopping the system. There are 3 different couplings supplied: 1/4” connection must be airtight. When installing or removing the probe, it is
Italian, German and American standard. necessary to depressurize the system.
The probe is equipped with a filter made of sintered steel, stainless
steel measuring chamber and control valve of the air flow. Suitable for
measurement of compressed air with dew point up to class 3 according
to standard ISO8573-1.

12
Direct connection (without measuring chamber) only for HP481 and Dimensions (mm) of the probe with measuring chamber, without quick
threaded connection: coupling (only HP480):

For direct connection of the probe, use a fitting with internal G 1/2” thread
on the side which will be placed in the probe. The connection must be Dimensions (mm) of the probe with measuring chamber and quick
airtight. When installing or removing the model HP480, it is necessary coupling (only HP480):
to depressurize the system. Ensure that the probe does not obstruct the
normal flow of air through the distribution line.
In all modes of installation, it is recommended that you place in the plant,
upstream of the sensor, a safety valve to be closed manually in case of
maintenance of the probe.
Periodically check the cleanliness of the sintered filter of the probe, in order
to maintain optimum response characteristics of the probe. The filter can
be washed with a detergent that leaves no traces.
DIMENSIONS ORDERING CODES
Dimensions (mm) of the probe without measuring chamber:
HP480: Interchangeable temperature and relative humidity probe, com-
plete with SICRAM module. Connection cable 2m. Equipped with 15µ
sintered AISI 316 stainless steel filter, measuring chamber, air flow reg-
ulation valve, and three 1/4” quick couplings (standard Italian, German,
and American).
HP481: Interchangeable temperature and relative humidity combined
54
probe, complete with SICRAM module. Connection cable 2 m.
Equipped with 15μ sintered AISI 316 stainless steel filter, G ½” thread-
ing.

Technical specifications HP480 HP481


Relative humidity
Sensor capacitive
Measuring range 0…100%RH
Accuracy ± 1,5%RH (0..90%RH), ± 2%RH (remaining field)
(@ T = 15…35 °C)
Accuracy ± (1,5 + 1,5% of the measured value)%RH
(@ T = -40…+60 °C)
Long term stability < 1%RH/year
Temperature
Sensor Pt100
Measuring range -40…+60 °C
Accuracy ± 0,25 °C
Dew point

Sensor Parameter calculated from the measurement of temperature and relative humidity

Measuring range -40…+60 °C DP

Accuracy ± 2 °C DP (-40…0 °C DP)


(@ T = 20 °C) ± 1 °C DP (0…+20 °C DP)

Accuracy see graph 1


(@ T = -40…+60 °C)
General features
Connection G 1/2" or G 1/4" or quick coupling G 1/2"
Regulation of the air flow From 0,2 to 3 l/min
Cable length 2m 2m (other lenghts on request)
Filter Sintered 15µ AISI 316 steel
Material of the measuring chamber AISI 304 stainless steel
Operating temperature of the probe -40…+80 °C
Operating pressure of the probe 0...16 bar -1...16 bar
Protection degreee IP65
Compatibility with ethylene oxide The maximum allowed concentration in continuous operation which causes a deviation within 2% is 3 ppm
(C2H4O)

13
TEMPERATURE PROBES – RESISTANCE THERMOMETERS

Delta OHM offers a wide choice of Platinum resistance thermometers with resistance equal to 100 Ω at 0 °C and temperature coefficient α as defined by
the IEC 60751 standard: Pt100, R0=100 Ω, α= 3.851∙10-3 °C-1.

For particular applications, probes with Pt1000 sensor or with a thermistor sensor are available. The response time τ0.63 indicated for each probe is the
response time of the sensor to a temperature variation, with a variation of the measured signal corresponding to the 63% of the total variation. The
response times are referred:

• in water at 100 °C for immersion probes;

• to the contact with a metal surface at 200 °C for surface probes;

• to an air temperature of 100 °C for air probes.

The IEC 60751:2008 standard defines the tolerance classes of the resistance thermometers as summarized in the following table:

Temperature Range
Tolerance class WIRE WOUND sensor THIN FILM sensor Tolerance [°C]

classe AA (1/3 DIN) from -50 °C to 250 °C from 0 °C to 150 °C ±(0.1+0.0017∙|t|)

classe A from -100 °C to 450 °C from -30 °C to 300 °C ±(0.15+0.002∙|t|)

classe B from -196 °C to 600 °C from -50 °C to 500 °C ±(0.3+0.005∙|t|)

classe C from -196 °C to 600 °C from -50 °C to 600 °C ±(0.6+0.01∙|t|)

On request, the probes can be assembled with a compatible connector chosen from TP471 and TP47.

The TP471 connector developed by Delta OHM contains an electronic module (SICRAM) that allows the probe error to be adjusted. During the Quality
Control, the probes provided with this module are individually checked in our laboratories, linearizing the characteristic and allowing more stringent
accuracy over the entire working range.

The following graph shows the Delta OHM SICRAM module probe TP472I typical error values obtained from the calibrations performed in our ISO17025
calibration laboratory. The graph highlights the effectiveness of the linearization performed on the probes.

Comparison between the classes defined in IEC 60751 standard and the typical error of the Delta OHM TP472I SICRAM probe
1,0

0,8

0,6

0,4

0,2
Tolerance [°C]

0,0 TP472I probe typical error

-0,2

-0,4

-0,6

-0,8

-1,0
-200 -100 0 100 200 300 400 500
Temperature [°C]

14
Tolerance as a function of temperature. The temperature range refers to the platinum wire wound probes.

Temperature [°C]

Tolerance [°C] -196 -100 -50 0 100 250 300 350 450 500 600

class AA --- ± 0.27 ± 0.19 ± 0.10 ± 0.27 ± 0.53 ± 0.61 ± 0.70 --- --- ---

class A --- ± 0.35 ± 0.25 ± 0.15 ± 0.35 ± 0.65 ± 0.75 ± 0.85 ± 1.05 --- ---

class B ± 1.28 ± 0.80 ± 0.55 ± 0.30 ± 0.80 ± 1.55 ± 1.80 ± 2.05 ± 2.55 ± 2.80 ± 3.30

class C ± 2.56 ± 1.60 ± 1.10 ± 0.60 ± 1.60 ± 3.10 ± 3.60 ± 4.10 ± 5.10 ± 5.60 ± 6.60

accuracy TP472I ± 0.30 ± 0.30 ± 0.20 ± 0.10 ± 0.20 ± 0.20 ± 0.30 ± 0.30 ± 0.30 ± 0.30 ---

By means of the calibration, the purchased instrument can be metrologically


characterized, determining the systematic error of the thermometer and
ensuring at the same time the traceability to international standards. Delta
OHM Laboratories are able to provide this service by issuing calibration reports
according to ISO 9001 or ACCREDIA LAT certificates in compliance with ISO/IEC
17025 standard, recognized internationally through ILAC MRA agreements.

15
Pt100 PROBES WITH TP471 SICRAM MODULE
CODE T (°C) ACCURACY USE τ0.63 DIMENSIONS
1 2 1 2 1 2 1 2
ON ON ON ON

Tc = K Tc = J Tc = T Tc = N

±0.1 °C (@ 0 °C)
-196 Ø3
TP472I ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3s 300

+500
±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-50
TP472I.O ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3s Ø3

+300 230

±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-50
TP473P.I ±0.2 °C (-50 °C ≤ t ≤ 250 °C)
+400
±0.3 °C (t < -50 °C; t > 250 °C)
5s Ø4
150

±0.1 °C (@ 0 °C)
-50
TP473P.O ±0.2 °C (-50 °C ≤ t ≤ 250 °C)
+300
±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-50
TP474C.O ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 5s Ø5

+300 230

±0.3 °C (t < -50 °C; t > 250 °C)

-50 ±0.1 °C (@ 0 °C)


TP475A.O 12s Ø4

+250 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 230

±0.1 °C (@ 0 °C)
-50
TP472I.5 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3s Ø6

+400 500

±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-50
TP472I.10 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3s Ø6

+400 1000

±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-70
TP49A.I ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3,5s
+250
±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-70
TP49AC.I ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 5,5s
+250
±0.3 °C (t < -50 °C; t > 250 °C

-70 ±0.1 °C (@ 0 °C)


TP49AP.I 4s
+250 ±0.2 °C (-50 °C ≤ t ≤ 250 °C)

-50 ±0.1 °C (@ 0 °C) Ø3


TP87.O 3s
+200 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 70

16
Pt100 PROBES WITH TP471 SICRAM MODULE
CODE T (°C) ACCURACY USE τ0.63 DIMENSIONS

Contact probe
for solar panels,
-40 ±0.1 °C (@ 0 °C)
TP878.O with SICRAM
+85 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) module.
Cable L = 2 m
60s
Contact probe
for solar panels,
-40 ±0.1 °C (@ 0 °C)
TP878.1.O with SICRAM
+85 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) module.
Cable L = 5 m

Penetration
probe for
-20 ±0.1 °C (@ 0 °C) compost,
TP879.O 60s with SICRAM
+120 ±0.2 °C (-50 °C ≤ t ≤ 250 °C)
module.
Cable L = 5 m
Mignon head, cable length = 2m

±0.1 °C (@ 0 °C)
-50
TP880/300.I ±0.2 °C (-50 °C ≤ t ≤ 250 °C)
+450
±0.3 °C (t < -50 °C; t > 250 °C)

60s
Mignon head, cable length = 2m

±0.1 °C (@ 0 °C)
-50
TP880/600.I ±0.2 °C (-50 °C ≤ t ≤ 250 °C)
+450
±0.3 °C (t < -50 °C; t > 250 °C)

±0.1 °C (@ 0 °C)
-110
TP35.5AF.5S ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 3s
+180
±0.3 °C (t < -50 °C; t > 250 °C)
Cable L = 5 m. Shield in Inox + PTFE

Globe-thermometer probe for


measurement of radiant heat with
0150mm. Accuracy according to ISO 7243
TP875.I
ISO 7726.
Pt100 sensor, 4-wire cable L=2 m.
Supplied with SICRAM module.

-30 ±0.1 °C (@ 0 °C)


+120 ±0.2 °C (-50 °C ≤ t ≤ 250 °C) 15'
Globe-thermometer probe for
measurement of radiant heat with 0
50mm. Accuracy according to ISO 7243
TP876.I ISO 7726.
Pt100 sensor, 4-wire cable L=2 m.
Supplied with SICRAM module.

17
Pt100/Pt1000 PROBES WITH TP47 CONNECTOR WITHOUT SICRAM MODULE
CODE T (°C) CLASS USE τ0.63 DIMENSIONS

TP47.100.O
(Pt100)
-50 +250
TP47.1000.O
(Pt1000)
Class A 3s

TP87.100.O
(Pt100)
-50 +200
TP87.1000.O
(Pt1000)

Pt100 PROBES ENDING WITH FREE WIRES


Globe-thermometer probe for measurement of radiant
heat with 0150mm. Accuracy according to
TP875.1.I ISO 7243 ISO 7726.
Pt100 sensor, 4-wire cable L=2 m .
-30
+120 Class A 15s
Globe-thermometer probe for measurement of radiant
heat with 050mm. Accuracy according to
TP876.1.I ISO 7243 - ISO 7726.
Pt100 sensor, 4-wire cable L=2 m.

Contact probe for


solar panels
-40 4-wire
TP878.1SS.O Class A 60s
+85 cable L = 5 m

Penetration
probe for
-20 compost
TP879.1.O Class A 60s
+120 4-wire
cable L = 5 m

-40
TP32MT.1P.I Class A 40s
+100

-50
TP32MT.1P.2 +250 Class A 40s

-40
TP32MT.2.I +100 Class A 60s

-110
TP35.5AF.5 Class A 3s
+180
Cable L = 5 m. Shield in Inox + PTFE

18
TEMPERATURE PROBES FOR INDUSTRIAL USE
CODE T (°C) CLASS USE τ0.63 DIMENSIONS

HD882/EK -40 Not 5s


(KTY81) +150 applicable

HD882/ E/100 -50


Class A 5s
(Pt100) +300

HD882/GK -50 Not 5s


Environmental
(KTY81) +100 applicable

HD882/G100 -50 5s
Class A Environmental
(Pt100) +100

0
HD882/L104 Process 7s
+250 Class A
(Pt100) Thread

HD882/L106 0
Process
(Pt100) +250 Class A 15s
Thread

Process Thread -
HD882M100/600 -50
Class A Miniature 15s
(Pt100) +450 Head

HD882DM100/600
-50 Process Thread -
(Pt100) Class A 15s
+450 DIN B Head

HD882M100/300 Process Thread - 15s


-40
(Pt100) Class A Miniature
+100 Head

HD882DM100/300 -50 15s


Process Thread -
(Pt100) +250 Class A
DIN B Head

CONNECTORS

Connector without SICRAM module. It can be connected to


TP47 4-wire Pt100 probes (and 3-wire with some instruments) or
2-wire Pt1000 probes.

Connector with SICRAM electronic module for the


connection of resistance thermometers and the correction
of the characteristic of the sensor.
TP471 It can be connected to 3-wire or 4-wire Pt100Ω platinum
temperature probes.
assembling and calibration only in Delta OHM