Particle-Size Distribution (Gradation) of Fine-Grained Soils Using The Sedimentation (Hydrometer) Analysis
Particle-Size Distribution (Gradation) of Fine-Grained Soils Using The Sedimentation (Hydrometer) Analysis
Particle-Size Distribution (Gradation) of Fine-Grained Soils Using The Sedimentation (Hydrometer) Analysis
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
2
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
then split to reduce the mass while still having sufficient 4.4 The sedimentation test specimen is mixed with a dis-
quantity to meet the minimum mass requirements of Table 1. persing agent and test water. The slurry is allowed to condition
3.2.2 sample, n—material collected without limitation on and is then thoroughly mixed and placed in a cylinder with
the total mass or size range of particles meeting the minimum additional test water. Readings are taken with a hydrometer and
mass requirements provided in Table 1. thermometer over specific time intervals.
3.2.3 sedimentation sample, n—the minus No. 10 (2.0-mm) 4.5 The mass of particles passing specified particle diam-
or finer material separated from the reduced sample using the eters are calculated and recorded. The results produce a
separation sieve which is then used to obtain the sedimentation tabulation of particle size versus percent passing the separation
specimen and a water content specimen. sieve size that can be graphically presented as a gradation
3.2.4 sedimentation specimen, n—the material obtained curve. The plot is typically expressed as percent passing/finer
from the sedimentation sample having a maximum particle- than the separation sieve size versus the log of the particle size
size no greater than the No. 10 (2.0-mm) sieve to be used in the in millimeters. These results can then be combined with the
sedimentation test and in sufficient quantity to satisfy the results of a mechanical sieve analysis to obtain a complete
minimum mass requirements of Table 1. particle size distribution.
3.2.5 separation sieve, n—the No. 10 (2.0-mm) sieve or 5. Significance and Use
finer (Note 4) used to separate the reduced sample to obtain the
material for the sedimentation sample. 5.1 Particle-size distribution (gradation) is a descriptive
term referring to the proportions by dry mass of a soil
NOTE 4—The methodology for using a sieve finer than the No. 10 distributed over specified particle-size ranges. The gradation
(2.0-mm) is not defined in this standard. The methodology used to obtain
a representative sample using a sieve finer than the No. 10 (2.0-mm) is not curve generated using this method yields the distribution of silt
the same as obtaining the representative sample using the No. 10 (2.0-mm) and clay size fractions present in the soil based on size
sieve as presented in this standard. Additional effort or steps are necessary definitions, not mineralogy or Atterberg limit classification.
to make sure the material passing the finer sieve adequately represents the
sample. Such additional effort or steps should be documented if using a 5.2 Unless the sedimentation sample is representative of the
sieve finer than the No. 10 (2.0-mm) sieve to obtain the sedimentation entire sample, the sedimentation results must be combined with
specimen. a sieve analysis to obtain the complete particle size distribu-
tion.
4. Summary of Test Method
5.3 The clay size fraction is material finer than 2 µm. The
4.1 This test method is used to determine the particle-size
clay size fraction is used in combination with the Plasticity
distribution (gradation) of material finer than the No. 200
Index (Test Methods D4318) to compute the activity, which
(75-µm) sieve as a percentage of the mass used in the
provides an indication of the mineralogy of the clay fraction.
sedimentation test.
5.4 The gradation of the silt and clay size fractions is an
4.2 When the source material contains particles larger than
important factor in determining the susceptibility of fine-
the 3⁄8-in. (9.5-mm) sieve, a reduced sample passing the 3⁄8-in.
grained soils to frost action.
(9.5-mm) sieve shall be obtained using techniques presented in
Test Methods D6913 or another standard. This reduced sample 5.5 The gradation of a soil is an indicator of engineering
shall meet the minimum mass requirements in Table 1 for the properties such as hydraulic conductivity, compressibility, and
3⁄8-in. (9.5-mm) sieve. The material is processed using the shear strength. However, soil behavior for engineering and
moist (referee) preparation method unless the material is other purposes is dependent upon many factors, such as
received in the air-dried state. effective stress, mineral type, structure, plasticity, and geologi-
4.3 The reduced sample is separated using the separation cal origin, and cannot be based solely upon gradation.
sieve to obtain a sedimentation sample. The sedimentation 5.6 Some types of soil require special treatment in order to
sample is then split to obtain the appropriate mass for the correctly determine the particle sizes. For example, chemical
sedimentation test specimen and a water content test specimen. cementing agents can bond clay particles together and should
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
3
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
be treated in an effort to remove the cementing agents when accuracy of at least 60.5°C. The thermometric device must be
possible. Hydrogen peroxide and moderate heat can digest capable of being immersed in the suspension and reference
organics. Hydrochloric acid can remove carbonates by washing solutions to a depth ranging between 25 and 80 mm. Full
and Dithionite-Citrate-Bicarbonate extraction can be used to immersion, also known as complete or total immersion
remove iron oxides. Leaching with test water can be used to thermometers, shall not be used. The thermometric device shall
reduce salt concentration. All of these treatments, however, add be standardized by comparison to a nationally or internation-
significant time and effort when performing the sedimentation ally traceable thermometric device and shall include at least
test and are allowable but outside the scope of this test method. one temperature reading within the range of testing. The
5.7 The size limits of the sedimentation test are from about thermometric device shall be standardized at least once every
100 µm to about 0.1 µm. The length of time required to obtain twelve months. The same thermometric device shall be used
a stable initial reading on the hydrometer controls the upper for all measurements.
range of results, and the test duration controls the lower range. 6.5 Timing Device—A clock, stopwatch, digital timer, or
5.8 The shape and density of the grains are important to the comparable device readable to one second or better.
results. Stokes’ Law is assumed to be valid for spherical 6.6 Balance—Balances shall conform to the requirements of
particles even though fine silt- and clay-sized particles are Specification D4753.
more likely to be plate-shaped and have greater mineral 6.6.1 To determine the mass of the specimen, the balance
densities than larger particles. shall have readability without estimation of 0.01 g. The
5.9 High plasticity clays develop structured water layers on capacity of this balance will need to exceed the mass of the
their surfaces. According to Zhang and Lu3 this near surface container plus soil used to contain the soil suspension after the
water can be as dense as 1.4 g/L. This high-density structured completion of the sedimentation test. In general, a balance with
water causes an error in this test method and shifts the particle a minimum capacity of 1,800 g is sufficient.
size distribution curve upwards. Correction for the structured 6.7 Drying Oven—Vented, thermostatically controlled oven
water is beyond the scope of this standard but values of percent capable of maintaining a uniform temperature of 110 6 5°C
passing above 100 % are possible and should not be excluded throughout the drying chamber. These requirements typically
from the report. require the use of a forced-draft oven.
NOTE 5—The quality of the result produced by this standard is
dependent on the competence of the personnel performing it, and the
6.8 Plate—A clean, nonporous, smooth, solid surface that is
suitability of the equipment and facilities used. Agencies that meet the large enough to pile and split about 500 g of material. The
criteria of Practice D3740 are generally considered capable of competent surface shall not be made of any type of paper product.
and objective testing/sampling/inspection/etc. Users of this standard are
cautioned that compliance with Practice D3740 does not in itself assure 6.9 Specimen-Mixing Container—A glass beaker or equiva-
reliable results. Reliable results depend on many factors; Practice D3740 lent inert container with enough capacity to hold the specimen,
provides a means of evaluating some of those factors. the test water, and dispersant. A container of about 250 mL is
sufficient.
6. Apparatus
6.10 Temperature Maintaining Device—Unless otherwise
6.1 Hydrometer—ASTM hydrometer type 151H or 152H. specified by the requesting agency, the standard test tempera-
These hydrometers shall be in general conformance with the ture shall be in the range of 22 6 5°C. In addition, the
requirements in Specification E100 with the exception of temperature of the soil suspension shall not vary more than
dimensional specifications as indicated in Annex A1. 62°C. Normally, this temperature maintenance is accom-
6.2 Sedimentation Cylinder—At least two glass cylinders plished by performing the test in a room with a relatively
having a height of about 457 mm, an inside diameter between constant temperature. If such a room is not available, the
55 and 64 mm, and a capacity of 1,000 mL. The cylinders shall cylinders shall be placed in an automatically temperature
have an indication mark at 1,000 6 5 mL. One cylinder is used controlled insulated chamber or water bath that maintains a
for the soil suspension and the other one can be used as the temperature within the tolerance specified above.
control cylinder or the wash cylinder. The control cylinder 6.11 Soil Suspension Oven-Drying Container
shall have the same amount of dispersant as the soil suspension (Optional)—A container having smooth walls and capable of
cylinder. See Annex A1. holding approximately 1.5-L of the soil suspension. This
6.3 Separation Sieve—No. 10 (2-mm) or finer sieve used to container shall have a tight fitting lid or fit into a desiccator, to
separate the reduced sample. This sieve is subjected to rough prevent moisture gain during cooling of the oven-dried speci-
operation and shall not be used for quantitative grain size men. This is only required when there is insufficient material to
analysis. obtain the initial water content from a companion specimen.
6.4 Thermometric Device—A thermometric device capable 6.12 Dispersion Apparatus—Use one of the following de-
of measuring the temperature range within which the test is vices to disperse the specimen; however for referee testing, the
being performed readable to 0.5°C or better and having an stirring apparatus shall be used.
6.12.1 Stirring Apparatus (Referee)—A mechanically oper-
3
ated stirring device in which a suitably mounted electric motor
Zhang, C and Lu, N. “What is the Range of Soil Water Density? Critical
Reviews with a Unified Model,” Reviews of Geophysics, Vol 56, No. 3, July 2019, turns a vertical shaft at a minimum speed of 10,000 rpm
pp. 532–562. without load. The shaft shall be equipped with a replaceable
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
4
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
stirring paddle made of metal, as shown in Fig. 1. The shaft inserted on top of and in the center of the larger disk. This
shall be of such length that the stirring paddle will operate device shall be used for referee testing.
between 19.0 mm and 37.5 mm above the bottom of the
dispersion cup. 6.14 Agitation of the Soil Slurry—Any of the following
6.12.1.1 Dispersion Cup—A special dispersion cup con- items may be used to break up the soil aggregations as
forming to either of the designs shown in Fig. 2 shall be described in 11.1:
provided to hold the sample while it is being dispersed. The 6.14.1 Erlenmeyer Flask—A glass flask having a capacity
cup shall contain two sets of three long baffle rods and two sets between 250 mL and 500 mL.
of three short baffle rods rigidly mounted to the interior sides 6.14.2 Dispersion Shaker—A platform, wrist action or simi-
of the cup. This cup is used with the stirring apparatus. lar type shaker having a gyratory, orbital, reciprocating, or
6.12.2 Air Jet Dispersion Device (Optional)—A tube type or similar motion to assist in the dispersion process by continu-
other comparable device that uses compressed air to disperse ously agitating the soaking soil.
the slurry (Note 6). The device requires an air source capable 6.14.3 Ultrasonic Water Bath—The ultrasonic water bath
of providing up to 0.0024 m3/s of air to operate the device, must be large enough to hold a beaker or flask containing the
such that pressures between about 70 and 170 kPa can be soil slurry to be agitated for use in the sedimentation test. The
achieved. The device shall be fitted with a pressure gauge on water level in the bath should be equal to or higher than the
the line between the device and the air source. Water may water level in the specimen container.
condense in the lines when not in use and this water must be
removed. There are two ways to remove the water: use of a 6.15 Desiccator (Optional)—A desiccant containing device
water trap or purging the lines before use. If a water trap is of suitable size used to prevent moisture gain during cooling of
used, it shall be installed on the air line in such a manner to the oven-dried specimen.
prevent condensed water from entering the slurry. This device
6.16 Mortar and Rubber-Covered Pestle (Optional)—
shall not be used in referee testing.
Apparatus suitable for breaking up aggregations of air-dried
NOTE 6—Use of this device or others, causes differing amounts of soil particles without breaking individual particles.
dispersion and should be used with caution. Information on how to
appropriately use this device should be obtained from the manufacturer. 6.17 Miscellaneous Items—Items such as a wash/rinse
0.0024 m3/s is equivalent to 5 ft3/min. The device typically needs at least bottle (squirt bottle), rubber scraper, spatula, and stirring rod
0.0009 m3/s (2 ft3/min) to operate and therefore, some small air compres-
sors are not capable of supplying sufficient air to operate the device.
may be useful.
6.13 Agitator (Optional/Referee)—A hand-held device to 7. Reagents and Materials
mix the soil suspension in the sedimentation cylinder prior to
testing, as shown in Fig. 3. The agitator must not have any type 7.1 Sodium Hexametaphosphate (NaPO3)6—Also referred
of metal, such as a screw head, protruding from the bottom of to as sodium metaphosphate is the dispersion agent (defloccu-
the disk. To aid in strengthening the connection between the lant) required to prevent the fine particles in suspension from
rod and the disk, a smaller disk (about 25 mm or less) having coalescing or flocculating (Note 7). Consult the Safety Data
similar thickness and material as the larger disk, may be Sheet (SDS) for specific information regarding this chemical.
Dimensions are provided to give a sense of scale and will vary depending on manufacturer specifications.
FIG. 1 Typical Detail of Stirring Paddles
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
5
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
Dimensions are provided to give a sense of scale and will vary depending on manufacturer specifications.
FIG. 2 Typical Geometry of the Dispersion Cup
Dimensions are provided to give a sense of scale and will vary depending on manufacturer specifications.
FIG. 3 Detail of Agitator
7.2 Isopropyl Alcohol—Also referred to as isopropanol al- 7.3 Test Water—Distilled or demineralized water is the only
cohol or rubbing alcohol is used as a foam inhibitor. Commer- permissible test fluid. The use of tap water is not permitted.
cially available in concentrations ranging from 70 % to 99 %.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
6
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
NOTE 7—Fine-grained soils requiring the use of a dispersant are those aggregations without destroying the individual particles. Addi-
that do not readily slake in water, such as some highly plastic clays and tional guidance for splitting material to obtain a representative
most tropical soils. Typically, 5 grams per test of dispersant is used to
prevent flocculation and is added directly to the soil, such that the
specimen using a splitter, quartering, or moist stockpile sam-
concentration will equal 5 g/L in the final soil suspension volume. The pling is given in Test Methods D6913, Annex A2.
chemical formula for the dispersant shown above is approximate.
NOTE 8—Air drying causes irreversible changes to some clay particles
8. Sampling that cause permanent flocculation and decreases the fine fraction.4
8.1 General—This test method does not address, in any 9.2 Moist Preparation (Referee)—This preparation method
detail, procurement of the sample. It is assumed the sample is shall be used for materials not received in the air-dried state.
obtained using appropriate methods and is representative of the This method is especially important for any soil whose
material under evaluation. However, the testing agency shall properties are altered due to drying such as, most organic soils,
preserve all samples in accordance with Practice D4220/ many highly plastic fine-grained soils, tropical soils and soils
D4220M, Group B, except if the as-received sample does not containing halloysite. The material is thoroughly mixed to
meet those requirements. In which case, the water content of obtain a uniform reduced sample. Intact fine-grained samples
the material does not have to be maintained. should be chopped/reduced into small pieces, less than ap-
proximately 10 mm, and mixed to make uniform. There is no
8.2 Where data from this test are to be used for correlation need to process the fine-grained materials through a sieve.
with other laboratory or field test data, use the same material as
used for these tests where possible and as much is practical. 9.3 Air-Dried Preparation—This preparation method shall
only be used if the sample is received in the dried condition. In
8.3 The sample can be from a variety of sources and contain order to obtain a uniform reduced sample, the sample is
a wide range of particle sizes. Typically, samples for particle- blended at room temperature.
size analysis are obtained from the following sources: large
bags or buckets, small bags, jar samples, tube samples, or 9.4 If the reduced sample contains particles which are larger
specimens from other tests, such as consolidation, hydraulic than the No. 10 (2.0-mm) sieve, the material shall be separated
conductivity or strength tests. In some cases, such as compac- using a No. 10 (2.0-mm) or finer sieve. Process the entire
tion testing, prior testing may have caused a reduction in reduced sample over the No. 10 (2.0-mm) or finer sieve using
particle sizes; therefore, it may be required to obtain a sample a rubber scraper and, if needed, test water to aid in working the
of the original material, the degraded sample, or both. Test material through the sieve. Check that the material retained on
Methods D6913, Section 9, gives additional information re- the sieve does not contain aggregations of finer particles. Any
garding sampling from the different sources. aggregations should be broken and passed through the sieve. It
8.3.1 Preserve the sample at its original moisture condition is not necessary that the separation be totally complete but the
unless excluded above, and at no time shall the sample be material passing the sieve shall be representative of the passing
allowed to undergo undesirable temperature changes such as fraction. The material passing the separation sieve is termed the
freezing or heating. sedimentation sample and shall meet the minimum mass
requirement of Table 1. Record the separation sieve size that is
8.4 When the sample contains particles larger than the 3⁄8-in. used to separate the sample.
(9.5-mm) sieve, it shall be processed to obtain the reduced
sample. If particle-size separation is necessary, process the 9.5 Estimate the maximum amount of moist mass that can
sample to meet this requirement using the separation proce- be used for the sedimentation test using the following equation
dures provided in Test Methods D6913. (Note 9):
8.4.1 The reduced sample shall have a maximum particle
size that passes through the 3⁄8-in. (9.5-mm) or smaller sieve.
M mest 5 H c 3 S D S S DD
100
%est
3 1 1
w cest
100
(1)
8.4.2 The mass of the reduced sample shall meet or exceed
where:
the mass requirements given in Table 1.
Mmest = estimated maximum moist mass, nearest 1 g,
9. Preparation of the Test Specimen Hc = hydrometer capacity, g (either 45 for 151H or 55 for
9.1 Specimen Procurement—This standard presents two 152H),
preparation methods to obtain the sedimentation specimen %est = estimated percentage of material passing the No.
from the reduced sample: moist and air-dried. In these prepa- 200 (75-µm) sieve, nearest 1 %, and
wcest = estimated water content, nearest 1 %.
ration methods, moist and air-dried refers to the condition of
NOTE 9—The mass of the sedimentation test specimen should be
the material or sample as it is being reduced to an appropriate selected based on the amount of material that will be in suspension at the
particle size and mass. The test shall not be performed on time of the first reading. Using the maximum capacity of the hydrometer
oven-dried material. The moist preparation method shall be provides the highest resolution but is not a requirement. Exceeding the
used for referee testing and for samples not received in the capacity will make it impossible to obtain the initial readings. The
capacity of the 152H hydrometer is about 55 g of dry soil and the capacity
air-dried state. The air-dried method shall only be used on of the 151H is about 45 g. The wet mass should be adjusted to account for
materials received in the air-dried state (Note 8). Since some
fine-grained, air-dried soils aggregate, a mortar and rubber
covered pestle is used to break up aggregations. Care must be 4
Sridharan, A., Jose, B.T., and Abraham, B.M., Technical Note on “Determina-
taken to avoid disintegration or reduction of individual par- tion of Clay Size Fraction of Marine Clays,” Geotechnical Testing Journal,
ticles. Use only enough force as necessary to break up the GTJODJ, Vol. 14, No. 1, March 1991, pp. 103-107.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
7
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
the water content and the fraction of particles smaller than the No. 200 first 5 to 8 minutes of the test can be difficult. During that time it is
(75-µm) sieve. For example, if the water content is estimated at 20 %, acceptable to read the hydrometer to the nearest 1⁄2 division.
using a 151H hydrometer, and an estimated percent passing the No. 200
(75-µm) sieve of 95 %, the estimated moist mass is 57 g. 10.1.2 To insert the hydrometer correctly do the following:
First, make sure the stem is dry. Then, gently hold it by the
9.6 If the sedimentation sample contains sufficient material,
stem with one or both hands and slowly lower it to the depth
then split or quarter the sedimentation sample into at least two
slightly below the point at which it just floats. The hydrometer
portions: one for the water content determination and one for
should then rise slightly and become stable. This insertion
the sedimentation test. The water content specimen shall
process should take between 5 to 15 seconds.
contain 50 6 10 g of material. The sedimentation specimen
should be less than the limit computed in 9.5. 10.2 Temperature-Density Correction—During a test, the
9.7 If the sedimentation sample has limited material, do not suspension fluid density changes, therefore calculations for the
obtain a water content specimen. Obtain the dry mass, Md, of quantity of particles in suspension must account for fluid
the sedimentation specimen at the end of the sedimentation test density changes due to temperature, presence of dispersant, and
as discussed in 11.12 and calculated in 12.1.2. Reduce the mass the meniscus rise. The temperature-density correction is shift-
to be less than the limit computed in 9.5. ing the hydrometer scale, which is factory set for distilled
water at 20°C. There are two ways to determine this correction:
NOTE 10—If there is enough material after splitting/quartering to obtain take companion measurements in a control cylinder filled with
the necessary masses for both the sedimentation and water content the reference solution during the testing or generate a reusable,
specimens, the sedimentation sample is considered to have sufficient
material. If there is not enough material to obtain both the necessary calibration relationship. Both options require the use of a
masses of the specimens, the sedimentation sample is considered to have reference solution composed of test water and the same amount
limited material. of dispersant used in the sedimentation test cylinder. The
9.8 Record the mass of the moist soil, Mm, used for the meniscus correction is automatically accounted for in the
sedimentation test to the nearest 0.01 g. temperature-density correction for both options by consistently
reading the hydrometer at the top of the meniscus as described
9.9 Place the sedimentation specimen in the specimen-
in 10.3.
mixing container and record the identification of the specimen-
mixing container. 10.2.1 Reference Solution—The reference solution shall be
prepared with the same amount of dispersant as the soil
9.10 If sufficient material is available, immediately use the suspension. Fill a control cylinder to the 1,000 mL mark with
other specimen for determination of the water content in a mixture of test water and the same amount of dispersant used
accordance with Test Methods D2216, and record the water in the soil suspension cylinder. The test water and dispersant
content, wc, to the nearest 0.1 %. shall be well mixed such that no visible crystals can be seen,
10. Verification/Preparation of Apparatus and the reference solution shall be agitated to make sure the
dispersant is adequately mixed throughout the control cylinder.
10.1 Hydrometer—Check and record the dimensions of the
10.2.1.1 Companion Measurements—Use a control cylinder
151H or 152H hydrometers as presented in Annex A1 in
filled with the reference solution in conjunction with the soil
accordance with the interval listed in the Annex. The hydrom-
suspension cylinder to obtain the correction. Position the
eter shall be free of cracks and chips, which can compromise
control cylinder in the same temperature controlled location as
the integrity of the hydrometer. The body and stem of a
the test cylinders so that all cylinders are at or near the same
hydrometer seldom change over time, unless they have been
temperature. For each hydrometer and temperature reading
subjected to corrosive materials or have been damaged, that is,
taken in the soil suspension cylinder, take a corresponding
chipped or cracked. They only need to be checked before initial
hydrometer and temperature reading in the control cylinder.
use or after damage has been suspected or seen. Since it is
However, it is permissible for one control cylinder reading to
possible for the paper scale inside the hydrometer stem to slide
be used in conjunction with readings in the soil suspension for
down, the hydrometer reading in the test water, shall be
up to 30 minute intervals. Record the elapsed time and the
checked and documented according to the interval presented in
hydrometer and temperature readings of the control cylinder on
Annex A1 to make sure the scale has remained in its proper
the data sheet(s) only when measured (Note 12).
place. If the scale has moved, the hydrometer shall be replaced.
10.1.1 Hydrometer Readings—Hydrometer readings are NOTE 12—Readings taken in one control cylinder may be used for
taken to the nearest 1⁄4 division (Note 11). Reading the multiple test cylinders. Therefore, if the hydrometer and temperature
hydrometer can be difficult. A properly placed hydrometer readings in the control cylinder are only taken once during the first 30
minutes of the test, only that reading is recorded on each of the data
should neither bob nor rotate appreciably when released in the sheet(s) to which it applies. These measured readings are not to be written
soil suspension. It is important for the stem to be dry and clean in for the other times during the first 30 minutes when a measurement was
when inserting it into the soil suspension. If the stem is wet not actually determined.
above the reading point it will add mass to the hydrometer 10.2.2 Calibration Relationship—This option generates a
causing the reading to be too low. If the stem is not clean, general calibration relationship between the hydrometer read-
variations in the meniscus rise will result. In this application, ing of a control cylinder filled with the reference solution and
the hydrometers are always read at the top of the meniscus for the temperature, which eliminates the need for companion
the reason stated in 10.3. measurements during the test. A temperature-based general
NOTE 11—Reading the hydrometer to the nearest 1⁄4 division during the calibration relationship is required and established for each
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
8
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
hydrometer. A sedimentation cylinder is filled to the 1,000 mL 1606 setting the scale to read mass of solids in solution for a particle
mark with test water with the same amount of dispersant used specific gravity of 2.65.
in the soil suspension. Be sure the solution is thoroughly 10.3 Meniscus Correction—Hydrometers are designed to be
mixed, and the temperature is allowed to equilibrate. Then read at the fluid surface, however, the fluid is wetting to the
insert the hydrometer as described in 10.1.2. Take the reading glass such that the soil suspension will rise up the hydrometer
at the top of the meniscus and record this reading and the stem making a reading at the fluid surface impossible at times.
temperature of the solution. Rinse the hydrometer well with It is common practice to read the hydrometer at the top of the
test water between readings and dry it prior to taking the next meniscus and it must be taken there even when it is possible to
reading. Increase/Decrease the temperature of the cylinder, see through the soil suspension. The meniscus rise has a small
allow it to come to temperature equilibrium and repeat the impact on the effective depth determination since the reading is
measurement process. Take at least five different hydrometer taken slightly above the surface of the suspension. This rise
and temperature readings within the range of temperatures causes a change to the third significant digit in the computed
expected during the sedimentation test. Calculate the constant particle size. The meniscus correction is performed before
A or B as discussed below. initial use of the hydrometer and after damage has been
10.2.2.1 The 151H hydrometer measures the specific grav- suspected or noticed.
ity of the fluid relative to distilled water at 20°C. The 10.3.1 The meniscus correction, Cm, as shown in Fig. 4, is
calibration measurements are used to compute the constant A in determined by inserting the hydrometer with a clean and dry
the following equation. The standard deviation of the five stem and without bobbing into the test water with the proper
computed “A” values shall be less than 0.0005. The average amount of dispersant. The reading at the top of the meniscus
value of A is used when computing the temperature-density and the reading where the plane of water surface intersects the
correction. stem are recorded. The difference between these two readings
(top of meniscus minus plane of water surface) is the meniscus
A n 5 R 151,n 1 ~ 7.784 3 1026 3 T n ! 1 ~ 4.959 3 1026 3 T 2n ! (2) correction, Cm. Therefore, the meniscus correction is a nega-
where: tive value and has units of the hydrometer scale. In accordance
A = average specific gravity shift (151H hydrometer), with Test Method E126, the plane of water surface reading
nearest 0.0001, shall be obtained using the following guidance. Observe a
R151,n = 151H specific gravity hydrometer at reading, n, in point slightly below the plane of the water surface and raise the
reference solution, readable to 0.00025 or better, line of vision until this surface, seen as an ellipse, becomes a
T = temperature at reading, n, readable to 0.5°C or straight line. The point where this line cuts the hydrometer
better, and scale is the hydrometer reading. Holding a white card behind
n = subscript indicating the reading number during the cylinder just below the water level will improve the
calibration. visibility of the surface. The hydrometer readings shall be
recorded to the nearest 1⁄4 division.
10.2.2.2 The 152H hydrometer measures the mass of par-
ticles (specific gravity of 2.65) in a suspension of distilled 10.4 Effective Depth—The effective depth, also referred to
water at 20°C. The temperature-density correction provides the as “true depth,” is used in the calculation of the particle fall
offset mass reading for the hydrometer for a specific tempera- distance for each reading. The effective depth is defined as the
ture and dispersant concentration. The calibration measure- distance from the center of (volume) buoyancy of the hydrom-
ments are used to compute the constant B in the following eter to the surface of the soil suspension. The equation to
equation (Note 13). The standard deviation of the five com- calculate the effective depth when the hydrometer is inserted
puted “B” values shall be less than 0.5 g/L. The average value and removed between readings requires certain dimensional
of B is used when computing the temperature-density correc- measurements in order to do the calculation. Refer to Fig. 5.
tion. 10.4.1 Determine and record the volume of the hydrometer
bulb, Vhb, to the nearest 1 cm3 using the procedure given in
B n 5 R 152,n 1 ~ 1.248 3 1022 3 T n ! 1 ~ 7.950 3 1023 3 T 2n ! (3)
where:
B = average mass reading shift (152H hydrometer),
nearest 0.1,
R152,n = mass in reference solution hydrometer at reading, n,
readable to 0.25 g/L or better,
T = temperature at reading, n, readable to 0.5°C or
better, and
n = subscript indicating the reading number during
calibration.
NOTE 13—The equations relating the hydrometer readings to tempera-
ture are based on the same water density-temperature expression as used
in Test Methods D854. For the 151H specific gravity hydrometer, the
constants in the water density-temperature equation from Test Methods
D854 are divided by the density of water at 20°C (0.99821 g/mL). For the
152H mass in suspension hydrometer, the constants are multiplied by FIG. 4 Meniscus Correction Diagram
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
9
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
well as the minimum hydrometer reading, Hr1. Record both
values to the nearest 0.1 cm. Refer to Fig. 5.
10.4.4 Measure and record the inside cross-sectional area of
the sedimentation cylinder, Ac, to the nearest 0.1 cm2 using the
procedure given in A1.4.2.1.
10.5 Sedimentation Cylinder—Check and record the dimen-
sions of the sedimentation cylinders as presented in Annex A1
in accordance with the interval presented in Annex A1.
10.6 Sieves—See Practice E11 for the verification require-
ments of the sieves used in this test method.
10.7 Miscellaneous—The ancillary equipment used in con-
junction with these test methods shall be calibrated/verified/
checked according the intervals listed in Practice D3740 and
performed in accordance with their applicable standards.
11. Procedure
11.1 Add 5.0 6 0.1 g of sodium hexametaphosphate to the
sedimentation specimen in the specimen-mixing container
obtained in 9.9 or dissolve this amount of dispersant in about
100 mL of test water and add it to the sedimentation specimen.
Record to the nearest 0.01 g the actual amount of dispersant,
Mdisp, placed in the container or dissolved in the test water. If
added directly to the sedimentation specimen, first add at least
100 mL of test water to the specimen and then the dispersant to
form a slurry of milkshake consistency. The amount of test
water to be added during this step should be sufficient enough
only to facilitate the process of breaking apart the soil
aggregations. Mix the contents with a spatula or similar device
until all of the soil aggregations are broken-up (Note 14).
NOTE 14—If hand mixing is not efficient, use a 250 to 500 mL
Erlenmeyer flask along with either a wrist or orbital dispersion (mechani-
cal) shaker to vigorously agitate the soil slurry in a minimum of 150 mL
of test water for a few hours or until all the soil aggregations are
broken-up. Be aware the dispersion cup has a capacity of about 400 mL
and the stirring apparatus is only efficient in dispersing soil aggregations
that will pass between the baffle rods. An ultrasonic water bath, along with
the flask, may also be used to agitate the slurry.
11.2 Prior to the overnight conditioning period, disperse the
slurry using either the stirring apparatus or an air jet device.
11.2.1 If using the stirring apparatus, transfer the slurry to
the dispersion cup. Use a wash/rinse bottle filled with test
water to aid in the transfer and make sure all of the slurry has
been transferred to the dispersion cup. Add additional test
water as necessary such that the cup is about half full, then use
the stirring apparatus to blend the soil for about 1 minute (Note
15).
NOTE 15—Subsection 11.2.1.1 provides the referee procedure to dis-
perse the soil. However, the stirring apparatus is known to create fines by
breaking fragments off weak grains. Alternative methods of dispersion
FIG. 5 Effective Depth Dimensional Measurements should be considered when testing friable materials.
11.2.1.1 Transfer all the dispersed slurry into the sedimen-
tation cylinder. Use the wash/rinse bottle filled with test water
Annex A1.3.4. The bulb includes everything from the bottom as needed to make sure all of the slurry is transferred to the
tip up to the base of the stem. cylinder. Add test water to the sedimentation cylinder to bring
10.4.2 Determine and mark the center of (volume) the bottom of the meniscus of the slurry to within 61 mm of
buoyancy, cb, using the procedure given in A1.3.5. the 1,000 mL mark.
10.4.3 Measure the distance between the center of (volume) 11.2.2 If using the tube type air jet device, transfer the slurry
buoyancy, cb, and the maximum hydrometer reading Hr2, as to the sedimentation cylinder. Use a wash/rinse bottle filled
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
10
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
with test water to aid in the transfer and make sure all of the is less critical with respect to the calculations than the addition of water
slurry has been transferred to the cylinder. Add additional test (~5 mL) to the cylinder.
water as necessary to bring the volume to no more than 250 mL NOTE 17—The number of turns during this minute should be approxi-
mately 60 counting the turn upside down and back as two turns. Any soil
in the sedimentation cylinder. remaining in the bottom of the cylinder during the first few turns should
11.2.2.1 Before placing the device into the cylinder, slowly be loosened by vigorously shaking of the cylinder while it is inverted.
allow air to flow until the gauge reads about 7 kPa. This initial 11.4 If using a companion measurement to obtain the
pressure is needed to prevent the slurry from entering the air temperature-density correction, prepare the reference solution
jets when the device is inserted into the cylinder and to remove in a control cylinder by dissolving the same amount (5.0 6 0.1
any water that has condensed in the lines. Then, slowly lower g) of sodium hexametaphosphate as used in the soil suspension
the device into the slurry. Make sure the rubber stopper is cylinder in test water. Add test water to the sedimentation
securely in place at the top of the cylinder to prevent the slurry cylinder to bring the bottom of the meniscus of the solution to
from being ejected from the cylinder. 61 mm of the 1,000 mL mark.
11.2.2.2 For clayey soils increase the pressure to about 170
kPa and for sandy soils increase the pressure to about 70 kPa. 11.5 Cover the cylinder(s) to prevent evaporation and allow
Once the pressure is reached, disperse the slurry for five the cylinder(s) to sit overnight either in a temperature con-
minutes. At the end of five minutes, reduce the pressure to trolled insulated chamber or water bath or in an area of
about 7 kPa and lift the air jets out of the slurry and wash any relatively constant temperature. This conditioning period al-
slurry remaining on the device back into the cylinder. Once the lows the temperature to equilibrate and for the specimen to
device is washed off, turn off the air flow to the device and add deflocculate (Note 18). The soil slurry is now referred to as a
test water to the sedimentation cylinder to bring the bottom of soil suspension.
the meniscus of the slurry to within 61 mm of the 1,000 mL 11.6 At the end of the conditioning period, check the
mark. suspension for indications of flocculation (Note 19). If the
11.3 Mix the slurry using the agitator (referee) or the tipping suspension has indications of flocculation, then the test is
method (Note 16). The agitator device is the preferred/referee invalid and the suspension should be discarded since it has
method for mixing. Check for the presence of foam on top of been contaminated with dispersant. A new specimen would
the slurry after mixing. If a significant amount of foam is need to be treated to prevent flocculation. Such treatment is
present, it may be necessary to reduce the foam using isopropyl beyond the scope of this test method.
alcohol just prior to the start of testing (See 11.7.1). NOTE 18—Strong interparticle bonding can occur in suspensions having
11.3.1 To use the agitator, insert the paddle to the lower 1⁄4 high salt concentrations. The dispersant will not be effective in breaking
of the cylinder. Stroke the agitator at a rate of about one cycle these bonds. In such circumstances the salt should be leached from the soil
per second over a distance of several centimeters to dislodge before proceeding with the sedimentation test. This leaching results in a
considerable increase in effort and difficulty to process the material and is
any material stuck to the bottom of the cylinder. After the not discussed in this standard.
material is dislodged, the agitator should be rapidly moved NOTE 19—Flocculation of clay particles can be difficult to detect. Minor
downwards until the paddle almost comes into contact with the flocculation will shift the particle-size distribution (gradation) towards a
bottom of the cylinder and then upwards with a slower motion. coarser fraction. Moderate flocculation will cause a plateau in the curve as
The downward stroke should take about one to two seconds the sedimentation process stops. Major flocculation will completely halt
sedimentation at an early stage. Only major flocculation is visually
while the upward stroke should take about two to three detectable showing a lack of material collected at the bottom of the
seconds. As this process is repeated, the elevation of the cylinder, a surface layer of clear fluid, or horizontal cracks in the
starting position of the agitation cycle is raised until the bottom suspension.
of the agitator is significantly above the mid-height of the soil 11.7 Once the suspension has temperature equilibrated and
suspension. Keep the paddle submerged at all times during deflocculated, the suspension shall once again be mixed to
mixing. Mix for about one minute or until the suspension create a uniform suspension. Repeat the mixing procedure as
appears uniform. described in 11.3. Upon completion of agitation, make sure the
11.3.2 To use the tipping method, first place a rubber cylinder is on a stable surface and in a location where it will not
stopper in the open end of the cylinder or use the palm of the be subjected to any jarring or disturbance and immediately start
hand to cover the opening. Then turn the cylinder upside down the timer as directed in either 11.7.2 or 11.7.3. If using a
and back for a period of one minute to complete the agitation temperature controlled water bath, immediately place the
of the slurry (Note 17). Using the cylinder tipping method is cylinder back into the water bath after agitation.
not very efficient and tipping for more than one minute will 11.7.1 If significant foam develops on top of the suspension
typically be required when testing highly plastic clays. This after mixing, immediately add up to three drops of isopropyl
method of agitation may leave some soil aggregations. Use of alcohol to the suspension to reduce or eliminate the foam.
the agitator is greatly encouraged.
11.7.2 If using the agitator, start the timer when the paddle
NOTE 16—When using the tipping method, there will likely be some of the agitator breaks the surface of the soil suspension. After
soil particles remaining on the rubber stopper or hand and on the sides of removing the agitator allow the “free” liquid on it to drain back
the sedimentation cylinder above the 1,000 mL mark. Be careful not to into the sedimentation cylinder.
lose much material by scraping the material adhering to the rubber stopper
or hand off onto the rim of the cylinder. Do not rinse these particles back 11.7.3 If using the tilting method, start the timer after
into the cylinder. The minor loss of mass (~0.02 g or less, if care is taken) completion of the last inversion.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
11
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
11.8 Hydrometer readings shall be taken at elapsed times of content taken in 9.10, then the dry mass is obtained from the
approximately, 1, 2, 4, 8, 15, 30, 60, 240, and 1440 minutes. soil suspension used for the sedimentation test.
Additional readings can be taken to better define the particle- 11.12.1 After the last hydrometer reading is obtained, trans-
size distribution (gradation) relationship. fer all of the soil suspension to the oven drying container as
11.8.1 About 15 to 20 seconds before a reading is required, described in 6.11. To facilitate the complete removal of all of
gently lower the hydrometer into the sedimentation cylinder as the suspension, agitate the cylinder to distribute the soil as
described in 10.1.2. At the prescribed elapsed times and when described in 11.3 (Note 21). Pour the agitated suspension into
the hydrometer is stable, read and record the hydrometer the container and record the identification of the container. If
reading, rm, to the nearest 1⁄4 division, and record the elapsed using the container as the tare to calculate the dry mass, also
time, tm. record the mass of the container prior to adding the suspension
11.8.2 When removing the hydrometer from the suspension, to the nearest 0.01 g. Use a wash/rinse bottle to aid in
the removal process should take about 5 to 10 seconds and be transferring the slurry.
with a steady motion. Generally, there will be a drop of NOTE 21—It is not necessary to duplicate the exact requirements of 11.3
suspension at the tip of the hydrometer. Touch the tip to the in order to sufficiently distribute the soil. Use only as many turns (tipping
inside lip of the cylinder and allow the drop to flow back into method) or strokes (agitator method) as needed to dislodge the material
the cylinder. from the bottom of the cylinder. The purpose of this agitation is to reduce
the amount of additional water needed to remove all of the suspension
11.8.3 With a spinning motion, place the hydrometer into a from the cylinder. Additional test water may be added to thoroughly clean
wash cylinder filled with test water to clean off the hydrometer the cylinder and remove all of the suspension.
(Note 20). Once the hydrometer is clean, remove and dry it off 11.12.2 Dry the suspension to a constant mass in the drying
prior to the next reading or the reading in the control cylinder oven at 110 6 5°C. Usually constant mass is achieved after 48
if using companion measurements. hours when using a forced-draft type oven. If there is any
NOTE 20—While it is ideal to rinse the hydrometer off in a wash uncertainty if the specimen has thoroughly dried, it is neces-
cylinder that can be easily cleaned, it is permissible to rinse the sary to perform the constant mass test after an additional six
hydrometer in another suitable container filled with test water providing hours in the drying oven, to verify it has indeed completely
there is ample room for the hydrometer to spin. dried before proceeding. Cover the container while cooling in
11.9 Immediately after taking a hydrometer reading, gently preparation of making the mass measurement (Note 22).
insert the thermometric device into the soil suspension and NOTE 22—Determining the dry mass after the test adds additional time
record the temperature, Tm, to the nearest 0.5°C or better. Do and consideration. Because the oven must remove large amounts of water,
not allow the thermometric device to create disturbance in the the drying time takes longer and the constant mass check interval is
suspension. After removing the thermometric device, cover the extended. The large surface area of the specimen in the container allows
cylinder to prevent evaporation when the time between the the material to potentially absorb more moisture from the air as it cools.
readings exceeds five minutes. 11.12.3 Remove the container from the drying oven and
11.9.1 The temperature reading taken at the start of the allow it to cool in a desiccator or a tightly covered/sealed
sedimentation test may be used for the initial series of time container. After the container has cooled, determine and record
readings up to 30 minutes. The temperature does not need to be the dry mass of the soil plus dispersant, Mdd, to the nearest 0.01
measured more frequently than 30 minutes and shall be g.
recorded at the elapsed times of actual measurements. 11.12.4 After recording the dry mass of soil plus dispersant,
Mdd, cover the specimen with tap water and allow the specimen
11.10 When using a companion measurement to determine to soak. During soaking, gently stir the specimen to facilitate
the temperature-density correction, the hydrometer, rd,m, and the separation of particles.
temperature, Tm, readings of the control cylinder shall be
11.12.5 Pour the soaked material over the No. 200 (75-µm)
measured and recorded in accordance with 10.2.1.1.
sieve (Note 23). Take care in transferring the soil suspension
11.11 If the dry mass, Md, of the sedimentation specimen from the container to the wash sieve so as not to lose material.
will be determined from a companion specimen, the soil Make sure there is no remaining material in the container and
suspension is now ready to be washed over the No. 200 that the wash water is running clear before transferring the
(75-µm) sieve. material retained on the sieve into the oven drying container.
11.11.1 Remix the suspension to dislodge the sediment at Record the identification of the container if a different con-
the base of the cylinder and then pour the soil suspension over tainer is used and proceed to 11.13.
the No. 200 (75-µm) sieve. Take care in transferring the soil NOTE 23—Test Methods D1140 provides information on the washing
suspension from the cylinder to the wash sieve so as not to lose technique to use.
material. Make sure there is no remaining material in the
11.13 Dry the material retained on the No. 200 (75-µm)
cylinder and that the wash water is running clear before
sieve to a constant mass in the drying oven at 110 6 5°C.
transferring the material retained on the sieve into the oven
Usually, constant mass is achieved overnight (~12-16 hours)
drying container. Record the identification of the container and
when using a forced-draft type oven. If there is any uncertainty
proceed to 11.13.
if the specimen has thoroughly dried, perform the constant
11.12 If the dry mass, Md, of the sedimentation specimen mass test after an additional two hours in the drying oven, to
was not determined using the companion specimen water verify it has indeed completely dried before proceeding.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
12
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
11.14 Remove the container from the drying oven and allow where:
it to cool in a desiccator or cover the container with a tight rd,m = 152H g/L hydrometer offset at reading, m, nearest 0.1
fitting lid. After the container has cooled, determine and record g/L,
the dry mass of the soil retained on the No. 200 (75-µm) sieve, B = average mass shift (152H hydrometer), nearest 0.1
Mdr, to the nearest 0.01g. During the washing process, the g/L,
dispersant has been removed and the resulting dry mass will T = temperature at reading, m, readable to 0.5°C or better,
only include the particles retained on the sieve. and
m = subscript indicating the reading number during the
12. Calculations sedimentation test.
12.1 Calculate the dry mass, Md, of the sedimentation 12.3 Temperature-Density Correction: Companion
specimen using either of the following methods. Measurement—When using the companion measurement to
12.1.1 Dry Mass Using Moist Mass and Water Content— obtain the temperature-density correction, use the recorded
Using the water content, wc, of the companion specimen hydrometer offset reading taken in the control cylinder, rd,m,
determined in 9.6 and the initial moist mass, Mm, of the that corresponds to the hydrometer used during the test.
sedimentation specimen, calculate the dry mass of the sedi- 12.4 Mass Percent Finer—For each hydrometer reading
mentation specimen based as follows: taken in the soil suspension, compute the mass of material still
Mm in suspension as a percentage of the sedimentation specimen
Md 5 (4)
11 S D
wc
100
using the appropriate equation for the type of hydrometer used
during the test.
where: 12.4.1 For each 151H hydrometer reading, calculate and
record the mass percent finer using the following equation:
Md = mass of dry soil, nearest 0.01 g,
Mm = mass of moist soil, nearest 0.01 g, and
wc = water content, nearest 0.1 %.
Nm 5 S Gs
Gs 2 1 DS D
V sp
ρ ~ r 2 r d,m ! 3 100
Md c m
(8)
T
nearest 0.0001 (dimensionless),
= temperature at reading, m, readable to 0.5°C or better,
N m 5 0.6226 3 S Gs
Gs 2 1D S D
3
V sp
Md S D
100
~ r m 2 r d,m ! 3 1000 (9)
and where:
m = subscript indicating the reading number during the
0.6226 = correction factor to adjust for particle specific
sedimentation test.
gravity,
12.2.2 For the 152H hydrometer, compute the offset reading rm = hydrometer reading in suspension at reading m,
value for each sedimentation test reading using the following readable to 0.25 g/L, and
equation: rd,m = hydrometer offset reading from reference solution
r d,m 5 B 2 1.248 3 1022 3 T m 2 7.950 3 1023 3 T m2 (7)
at same temperature as reading m, nearest 0.1 g/L.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
13
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
12.5 Effective Depth—This value is used in the calculation Md = initial dry mass of the sedimentation specimen with-
of the particle fall distance for each hydrometer reading. The out dispersant, nearest 0.01 g, and
following equation is used to calculate the travel distance of the Mdr = dry mass retained on the No. 200 (75-µm) sieve,
particles when the hydrometer is inserted immediately before a nearest 0.01 g.
reading and is removed until the next reading.
H m 5 H r2 1 S~ ~
H r1 2 H r2 !
r 2 2 r 1!
3 ~ r 2 2 r m 1 C m! 2
V hb
2A c D S D 13. Report: Test Data Sheet(s)/Form(s)
13.1 The methodology used to specify how data are re-
(10) corded on the test data sheet(s)/form(s), as given below, is
covered in 1.13 and in Practice D6026 (Note 26).
where:
Hm = distance particles fall at reading m when the hydrom- 13.2 Record as a minimum the following general informa-
eter is inserted only for an individual reading, 2 tion (data):
significant digits, cm, 13.2.1 Identification of the material being tested, such as the
Vhb = volume of the hydrometer bulb up to the base of the project identification, boring number, sample number, and
stem, nearest 1 cm3, depth.
Ac = cross-sectional area of the sedimentation cylinder, 13.2.2 Test number, if any, testing dates and the initials of
nearest 0.1 cm2, the person(s) who performed the test.
Hr = distance between the center of (volume) buoyancy and 13.2.3 The sample preparation method used: moist or air-
the minimum (Hr1) and maximum (Hr2) hydrometer dried
readings nearest 0.1 cm, 13.2.4 The specific gravity of the sedimentation specimen
rm = hydrometer reading in suspension at reading m, 151H: and indicate if the value is assumed or measured.
readable to 0.00025 (dimensionless) 152H: readable 13.2.5 The following apparatus identification used during
to 0.25 g/L, the test:
Cm = meniscus correction: 151H nearest 0.25 division (di- 13.2.5.1 Hydrometer type (151H or 152H) and identifica-
mensionless) 152H: nearest 0.25 g/L, tion number.
r = the minimum (r1) and maximum (r2) hydrometer 13.2.5.2 Sedimentation cylinder identification number.
reading (dimensionless or g/L), and 13.2.5.3 Thermometric device identification number.
m = subscript indicating the reading number during the 13.2.5.4 Balance identification number.
sedimentation test. 13.2.5.5 Oven identification number.
12.6 Maximum Particle Diameter in Suspension—For each 13.2.5.6 Wet washing sieve identification number.
hydrometer reading, calculate and record the particle diameter 13.2.6 Method used to disperse the specimen.
of the soil using the following equation (Note 25): 13.2.7 Method used to agitate the specimen.
13.2.8 Description and classification of the soil in accor-
Dm 5 SŒ 18µ
·
Hm
ρ wg ~ G s 2 1 ! t m
D 3 10 (11)
dance with Practice D2488 or when Atterberg limit data are
available, Practice D2487.
where: 13.2.9 Describe any material that was excluded from the
sedimentation specimen.
Dm = particle diameter, two significant digits, mm,
µ = dynamic viscosity of water at 20°C, 0.0100 g/cm-s, 13.2.10 Describe any problems that were encountered.
ρw = mass density of water at 20°C, 1.00 g/cm3, 13.2.11 Indicate any prior testing performed on the speci-
g = acceleration due to gravity, 980.7 cm/s2, men.
Gs = specific gravity of soil, three significant digits 13.3 Record as a minimum the following test specimen
(dimensionless), data:
tm = elapsed (fall) time, two significant digits, s, 13.3.1 The size of the separation sieve.
Hm = particle fall distance, two significant digits, cm, and 13.3.2 The water content of the material passing the No. 10
m = subscript indicating the reading number during the
(2.0 mm) or finer sieve, if determined.
sedimentation test.
13.3.3 Moist mass of the sedimentation specimen.
NOTE 25—Eq 11 assumes the density of water and the viscosity are
constant throughout the sedimentation experiment. Therefore, the viscos-
13.3.4 Dry mass of the sedimentation specimen and indicate
ity and water density values are rounded to 3 significant digits. how obtained: using companion water content or direct mea-
surement after testing.
12.7 Percent Passing the No. 200 (75-µm)—Calculate the 13.3.5 Amount of dispersant used in the test.
percent passing the No. 200 (75-µm) sieve using the following 13.3.6 The percent passing the No. 200 (75-µm) sieve.
equation: 13.3.7 The start time and date of the test.
S
P p 5 100 1 2
M dr
Md D (12)
13.3.8 Indicate if a foam inhibitor was used.
13.3.9 Tabulation of the hydrometer, temperature, elapsed
where: time, offset, and effective depth readings from the sedimenta-
tion test.
Pp = percent passing the No. 200 (75-µm) sieve, nearest 13.3.10 Tabulation of the mass percent finer and the particle
0.1 %, diameter in mm.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
14
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
13.3.11 A graph of the percent passing versus the log of CL—Lean clay, CL, 89 % fines, LL=33, PI=13, gray, soil
particle size in mm (Note 27). had been air dried and pulverized. Local name—Annapolis
NOTE 26—Appendix X1 shows an example data sheet and an example
Clay
of the graphical display of the results of the test. ML—Silt, ML, 99 % fines, LL=27, PI=4, light brown, soil
NOTE 27—It is possible to have percent passing values above 100 % had been air dried and pulverized. Local name—Vicksburg Silt
due to the effects of structured water around clay particles. Therefore, the
test report should include all of the results. 14.5 Repeatability (r)—The difference between repetitive
5 results obtained by the same operator in the same laboratory
14. Precision and Bias
applying the same test method with the same apparatus under
14.1 Precision—These estimates of precision are based on constant operating conditions on identical test material within
the results of the interlaboratory test program conducted by the short intervals of time would, in the long run, in the normal and
ASTM Reference Soils and Testing Program. This program correct operation of the test method, exceed the following
was conducted in preparation for writing D7928—Test Method values only in one case in 20.
for Particle-Size Distribution (Gradation) of Fine-Grained 14.5.1 Repeatability can be interpreted as the maximum
Soils Using Sedimentation (Hydrometer) Analysis. Three dif- difference between two results, obtained under repeatable
ferent soils were used in the hydrometer testing program. A conditions that are accepted as plausible due to random causes
description of these soils is given in 14.4. In the program, some under normal and correct operation of the test method.
laboratories performed three replicate tests per soil type
(triplicate test laboratory), while other laboratories performed a 14.5.2 Repeatability limits are listed in Table 2 and Table 3.
single test per soil type (single-test laboratory). The precision 14.6 Reproducibility (R)—The difference between two
statement for this test method is based solely on the triplicate single and independent results obtained by different operators
laboratory results. Seven laboratories participated in this study. applying the same test method in different laboratories using
Each of the labs reported three replicate test results for each of different apparatus on identical test material would, in the long
the fine-grained soils, at specified times during the test over a run, in the normal and correct operation of the test method,
period of time (up to 1440 minutes). exceed the following values only in one case in 20.
14.2 Every “test result” represents an individual determina- 14.6.1 Reproducibility can be interpreted as the maximum
tion. Except for the use of data from fewer than seven difference between two results, obtained under reproducible
laboratories (in those instances requiring the exclusion of conditions that are accepted as plausible due to random causes
outlier test results), Practice E691 was followed for the design under normal and correct operation of the test method.
and analysis of the data; the details are given in ASTM 14.6.2 Reproducibility limits are listed in Table 2 and Table
Research Reports No. D18-2000. Each measurement is used to 3 below.
compute both a percent passing and a particle diameter. Each
quantity is evaluated as an independent variable. The precision 14.7 The above terms (repeatability limit and reproducibil-
estimates vary with soil type and judgment is required when ity limit) are used as specified in Practice E177.
applying these estimates to a specific soil. 14.8 Any judgment in accordance with statements 14.5 and
14.3 This precision statement was determined through sta- 14.6 will have an approximate 95 % probability of being
tistical examination of all usable test results, from seven correct. The precision statistics in this section must not be
laboratories, on three materials. To judge the equivalency of treated as exact mathematical quantities which are applicable
two test results, it is recommended to choose the material to all circumstances and uses. The limited number of materials
closest in characteristics to the test material. tested may lead to times when differences greater than pre-
14.4 Soil Types—The soils used in the multilaboratory test dicted by the study results will arise, sometimes with consid-
program are described below in accordance with Practice erably greater or smaller frequency than the 95 % probability
D2487. In addition, the local names of the soils are given. limit would imply.
CH—Fat clay, CH, 99 % fines, LL=60, PI=39, grayish 14.9 Bias—There is no accepted reference value for this test
brown, soil had been air dried and pulverized. Local name— method, therefore bias cannot be determined.
Vicksburg Buckshot Clay
15. Keywords
5
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D18-2000. Contact ASTM Customer 15.1 clay; grain-size; hydrometer analysis; particle-size dis-
Service at service@astm.org. tribution (gradation); sedimentation; sieve analysis; silt
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
15
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
TABLE 2 Results for Percent Finer (Nm) by Mass at Specific Time Intervals
Material AverageA Repeatability Reproducibility Repeatability Reproducibility
Percent Finer Standard Deviation Standard Deviation Limit Limit
s0dx̄ Sr SR r R
1 minute reading
High Plasticity Clay 92.40 1.86 2.88 5.20 8.06
Low Plasticity Clay 80.47 1.59 2.95 4.45 8.26
Low Plasticity Silt 86.17 1.83 3.02 5.13 8.45
2 minute reading
High Plasticity Clay 87.45 1.51 2.38 4.24 6.67
Low Plasticity Clay 75.70 1.12 2.30 3.14 6.43
Low Plasticity Silt 74.76 2.66 3.04 7.44 8.51
4 minute reading
High Plasticity Clay 81.56 1.18 1.54 3.31 4.30
Low Plasticity Clay 72.02 0.91 1.79 2.55 5.01
Low Plasticity Silt 57.51 2.15 2.15 6.02 6.02
15 minute reading
High Plasticity Clay 69.09 0.56 1.58 1.56 4.41
Low Plasticity Clay 62.86 0.72 1.85 2.02 5.17
Low Plasticity Silt 27.86 2.10 2.10 5.88 5.88
30 minute reading
High Plasticity Clay 63.26 1.34 1.99 3.75 5.57
Low Plasticity Clay 56.24 0.86 2.06 2.39 5.77
Low Plasticity Silt 19.48 0.89 2.23 2.48 6.24
60 minute reading
High Plasticity Clay 57.77 1.10 1.197 3.07 5.53
Low Plasticity Clay 49.93 0.69 1.95 1.92 5.46
Low Plasticity Silt 13.96 0.48 2.77 1.35 7.77
120 minute reading
High Plasticity Clay 53.32 0.64 1.98 1.80 5.55
Low Plasticity Clay 43.88 0.85 2.33 2.39 6.52
Low Plasticity Silt 11.12 0.99 2.88 2.77 8.06
240 minute reading
High Plasticity Clay 49.72 0.62 2.08 1.75 5.83
Low Plasticity Clay 37.84 0.58 1.97 1.63 5.51
Low Plasticity Silt 10.02 0.50 2.50 1.41 7.000
1440 minute reading
High Plasticity Clay 42.42 0.71 2.15 2.00 6.02
Low Plasticity Clay 27.04 0.77 2.47 2.16 6.90
Low Plasticity Silt 8.30 0.60 2.92 1.68 8.17
A
The average of the laboratories’ calculated averages.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
16
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
TABLE 3 Results for Percent Diameter (Dm) (Values in mm) at Specific Time Intervals
Material AverageA Repeatability Reproducibility Repeatability Reproducibility
Percent Diameter Standard Deviation Standard Deviation Limit Limit
s0dx̄ Sr SR r R
1 minute reading
High Plasticity Clay 0.0375 0.0004 0.0008 0.0011 0.0021
Low Plasticity Clay 0.0404 0.0004 0.0006 0.0011 0.0017
Low Plasticity Silt 0.0377 0.0005 0.001 0.0013 0.0029
2 minute reading
High Plasticity Clay 0.0271 0.0002 0.0007 0.0006 0.0019
Low Plasticity Clay 0.029 0.0002 0.0006 0.0005 0.0018
Low Plasticity Silt 0.0281 0.0004 0.0008 0.0012 0.0021
4 minute reading
High Plasticity Clay 0.0193 0.0001 0.0008 0.0003 0.0023
Low Plasticity Clay 0.0205 0.0001 0.0008 0.0003 0.0024
Low Plasticity Silt 0.0211 0.0003 0.0008 0.0007 0.0023
15 minute reading
High Plasticity Clay 0.0108 0.0001 0.0002 0.0002 0.0006
Low Plasticity Clay 0.0112 0.0001 0.0002 0.0002 0.0006
Low Plasticity Silt 0.0123 0.0002 0.0002 0.0004 0.0006
30 minute reading
High Plasticity Clay 0.0078 0 0.0002 0.0001 0.0005
Low Plasticity Clay 0.0081 0 0.0002 0.0001 0.0005
Low Plasticity Silt 0.009 0.0001 0.0001 0.0002 0.0004
60 minute reading
High Plasticity Clay 0.0056 0 0.0001 0 0.0003
Low Plasticity Clay 0.0059 0 0.0001 0.0001 0.0004
Low Plasticity Silt 0.0065 0 0.0001 0.0001 0.0002
120 minute reading
High Plasticity Clay 0.0041 0 0.0001 0.0001 0.0002
Low Plasticity Clay 0.0043 0 0.0001 0 0.0003
Low Plasticity Silt 0.0046 0 0.0001 0.0001 0.0001
240 minute reading
High Plasticity Clay 0.0029 0 0 0 0.0001
Low Plasticity Clay 0.0031 0 0.0001 0.0001 0.0002
Low Plasticity Silt 0.0033 0 0.0001 0 0.0001
1440 minute reading
High Plasticity Clay 0.0012 0 0 0.0001 0.0001
Low Plasticity Clay 0.0013 0 0 0.0001 0.0001
Low Plasticity Silt 0.0014 0 0.0001 0 0.0001
A
The average of the laboratories’ calculated averages.
ANNEX
(Mandatory Information)
A1.1 General—There are many factors controlling the over- A1.2.1 The total length of the hydrometer,
all accuracy of sedimentation (hydrometer) test results. This HL = 275 - 285 mm.
section covers how the equipment constants used in calculating
A1.2.2 The scale key reference point for the 152H hydrom-
the sedimentation (hydrometer) test results are determined or
eter is 0.0 6 1 g ⁄L and 1.000 6 0.001 specific gravity for the
checked. The equipment tolerances for the hydrometer and
151H hydrometer read at the bottom of the meniscus.
sedimentation cylinder are presented below.
A1.2.2.1 The length from bottom of hydrometer to refer-
A1.2 Hydrometer Standard Dimensions —The 152H and ence reading, Hb = 243 - 247 mm.
151H hydrometers have specified requirements in accordance A1.2.2.2 The length of the quoted nominal scale as stated in
with Specification E100, as shown in Fig. A1.1, and the Specification E100, Hs = 81 - 85 mm.
hydrometer constants or equipment accuracy checks are listed
A1.2.3 The overall length of the hydrometer bulb, HLb =
with the accepted tolerances and the main points summarized
115 - 142 mm.
below. Hydrometer dimensions as stated here have a slightly
larger range of acceptability to reflect current manufacturing A1.2.4 The distance from the bottom of the hydrometer to
practices. These dimensions take precedence over the E100 the maximum diameter of the bulb, Hcb 58 - 71 mm (Note
tolerances. A1.1).
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
17
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
A1.2.5 The maximum diameter of the bulb, reading shall be adjusted by the ratio of the density of water at
Dm = 29 - 32 mm. the calibration temperature to the density of water at the check
temperature. The check temperature shall be measured to the
NOTE A1.1—It is possible for the maximum diameter and the center of
(volume) buoyancy to have different locations on the bulb. When nearest 0.5°C and the density value associated with that
calculating effective depth, the location of the center of (volume) temperature obtained from Table A1.1. The reference reading
buoyancy should be used to calculate Hr1 and Hr2, not the location of the shall be checked and documented prior to initial use, on an
maximum diameter. annual basis, or after damage has been suspected or seen. If the
A1.2.6 The submerged volume of the hydrometer bulb, Vhb hydrometer does not read within the allowable range, it shall
50 - 72 cm3. not be used.
A1.2.7 Stem shall be uniform and extend at least 15 mm A1.3.4 Volume—The volume dimension, Vhb, shall be made
above the top of the graduation and remain cylindrical for at and recorded to the nearest 1 cm3 using either the direct or
least 3 mm below the lowest graduation. The diameter of the indirect method. The direct method is done by inserting the
stem is set to meet the scale requirements. hydrometer into a graduated cylinder filled with test water
having a minimum scale sensitivity of 5 mL per division. Read
A1.3 Checking the Dimensions—The dimensions provided and record the volume in the cylinder prior to inserting the
in A1.2.2 through A1.2.7 shall be checked and documented hydrometer to the nearest 2 mL. Insert the hydrometer into the
prior to initial use or after damage has been suspected or seen. test water just to the base of the stem then read and record the
If any of the standard dimensions do not fall within the volume to the nearest 2 mL. The volume of the hydrometer is
allowable ranges, the hydrometer shall not be used. the difference in the volume readings before and after insertion
A1.3.1 Distance/Length—The length dimensions shall be into the test water. The volume is the average of three
made and recorded to the nearest 0.5 mm. They can be made determinations that are within 2 mL of each other. The indirect
using either, and given in order of preference: a height gauge method is done by making a mass measurement. The volume
(digimatic, dial, or vernier), calipers (digimatic, dial, or of the hydrometer bulb can be measured by placing a partially
vernier), or a ruler and square (the square is used to transcribe filled beaker of room temperature test water on a balance, zero
the measurement from the hydrometer to the ruler). The line of or tare the balance, and then lowering the hydrometer into the
contact between the hydrometer stem and bulb can be estab- water just to the base of the stem. While holding the hydrom-
lished and marked using one of two methods. The first choice eter in place, read and record the mass of displaced water to the
is to measure the diameter of the stem with calipers, increasing nearest 0.1 g. Using an approximate mass density of water
this measurement by about 1 mm and marking the spot where equal to unity (1), the volume of the hydrometer bulb in cm3
the calipers encounter the bulb. The other method is to visually will be equal to the mass reading.
determine this line of contact and mark it. The visual determi-
A1.3.5 Center of (Volume) Buoyancy—Determine the center
nation may be assisted by feeling the contact point with fingers.
of (volume) buoyancy using the direct or indirect method
Marking can be accomplished by placing lengthwise a thin
described in A1.3.4, except the bulb is inserted until the mass
strip of label paper/marking tape across the line of contact and
or water level change is half of the volume of the bulb as
then marking it with a pencil.
determined in A1.3.4. Mark the hydrometer at the water level
A1.3.2 Scale Length—The quoted scale length shall be and measure the distance from the mark to the tip of the
measured and recorded to the nearest 0.5 mm using either hydrometer.
calipers or a machinist ruler. The error stated in Note A1.1 is
ignored. A1.4 Sedimentation Cylinder Dimensions—The sedimenta-
tion cylinder constants or equipment accuracy checks are listed
A1.3.3 Scale Reference Reading—The 152H hydrometer
with the accepted tolerances and summarized below. The
should read 0.0 6 1 g/L and 1.000 6 0.001 specific gravity for
cylinder shall be checked and documented prior to initial use or
the 151H hydrometer at the bottom of the meniscus when
after damage is suspected or seen.
placed in distilled water that is free of gas bubbles at the
calibration temperature of the hydrometers. At temperatures A1.4.1 The accuracy of key volume mark is 1,000 6 5 mL
other than the calibration temperature of the hydrometer, the in mL or cm3.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
18
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
TABLE A1.1 Density of Water (ρw) Versus Temperature (T)A
T ρw T ρw T ρw T ρw
(°C) (g/ML) (°C) (g/ML) (°C) (g/ML) (°C) (g/ML)
15.0 0.99910 16.0 0.99895 17.0 0.99878 18.0 0.99860
.1 0.99909 .1 0.99893 .1 0.99876 .1 0.99858
.2 0.99907 .2 0.99891 .2 0.99874 .2 0.99856
.3 0.99906 .3 0.99890 .3 0.99872 .3 0.99854
.4 0.99904 .4 0.99888 .4 0.99871 .4 0.99852
.5 0.99902 .5 0.99886 .5 0.99869 .5 0.99850
.6 0.99901 .6 0.99885 .6 0.99867 .6 0.99848
.7 0.99899 .7 0.99883 .7 0.99865 .7 0.99847
.8 0.99898 .8 0.99881 .8 0.99863 .8 0.99845
.9 0.99896 .9 0.99879 .9 0.99862 .9 0.99843
19.0 0.99841 20.0 0.99821 21.0 0.99799 22.0 0.99777
.1 0.99839 .1 0.99819 .1 0.99797 .1 0.99775
.2 0.99837 .2 0.99816 .2 0.99795 .2 0.99773
.3 0.99835 .3 0.99814 .3 0.99793 .3 0.99770
.4 0.99833 .4 0.99812 .4 0.99791 .4 0.99768
.5 0.99831 .5 0.99810 .5 0.99789 .5 0.99766
.6 0.99829 .6 0.99808 .6 0.99786 .6 0.99764
.7 0.99827 .7 0.99806 .7 0.99784 .7 0.99761
.8 0.99825 .8 0.99804 .8 0.99782 .8 0.99759
.9 0.99823 .9 0.99802 .9 0.99780 .9 0.99756
23.0 0.99754 24.0 0.99730 25.0 0.99705 26.0 0.99679
.1 0.99752 .1 0.99727 .1 0.99702 .1 0.99676
.2 0.99749 .2 0.99725 .2 0.99700 .2 0.99673
.3 0.99747 .3 0.99723 .3 0.99697 .3 0.99671
.4 0.99745 .4 0.99720 .4 0.99694 .4 0.99668
.5 0.99742 .5 0.99717 .5 0.99692 .5 0.99665
.6 0.99740 .6 0.99715 .6 0.99689 .6 0.99663
.7 0.99737 .7 0.99712 .7 0.99687 .7 0.99660
.8 0.99735 .8 0.99710 .8 0.99684 .8 0.99657
.9 0.99732 .9 0.99707 .9 0.99681 .9 0.99654
27.0 0.99652 28.0 0.99624 29.0 0.99595 30.0 0.99565
.1 0.99649 .1 0.99621 .1 0.99592 .1 0.99562
.2 0.99646 .2 0.99618 .2 0.99589 .2 0.99559
.3 0.99643 .3 0.99615 .3 0.99586 .3 0.99556
.4 0.99641 .4 0.99612 .4 0.99583 .4 0.99553
.5 0.99638 .5 0.99609 .5 0.99580 .5 0.99550
.6 0.99635 .6 0.99607 .6 0.99577 .6 0.99547
.7 0.99632 .7 0.99604 .7 0.99574 .7 0.99544
.8 0.99629 .8 0.99601 .8 0.99571 .8 0.99541
.9 0.99627 .9 0.99598 .9 0.99568 .9 0.99538
A
Reference: CRC Handbook of Chemistry and Physics, David R. Lide, Editor-In-Chief, 74th Edition, 1993 – 1994.
A1.4.1.1 Volume—Verify the 1,000 mL mark is correct by Repeat this procedure at two more locations on the inside
performing the following steps. Place the cylinder on a balance bottom of the cylinder for a total of three readings. Determine
and zero it. Add distilled water free of air bubbles, having a and record the average distance, D1000, to the nearest 1 mm
known temperature, T, nearest 1°C, to the cylinder until the (Note A1.2). Calculate and record the area of the cylinder, Ac,
balance reads the required mass, Mw. The mass, Mw, in g, (Ac = 1,000 × (10/D1000)) to the nearest 0.1 cm2.
equals 1,000 times the density of water, ρw, at T taken from
Table A1.1. Read the water level at the bottom of the meniscus. NOTE A1.2—For example, if the 1,000 mL mark is determined to be 360
The level should be within about 5 mL or ~2 mm of the 1,000 mm from the inside bottom of the cylinder, the inside area is 27.8 cm2 and
the diameter is then 59.5 mm. For a 63.5 mm inside diameter cylinder, the
mL mark on the cylinder. If the mark is not correct, remark the 1,000 mL mark should be approximately 316 mm from the inside bottom.
cylinder with the correct 1,000 mL line or do not use the
cylinder. If the cylinder is remarked, the correct or incorrect A1.4.3 The inside area’s uniformity above the key volume
line shall be clearly marked as such to prevent misuse of the mark.
incorrect line marking. Record the volume to the nearest 1 cm3. A1.4.3.1 Area Uniformity—Determine and record the uni-
A1.4.2 The inside area, Ac, in cm2. formity of the area above the 1,000 mL. Add 150 6 1 g of
A1.4.2.1 Area—Determine the inside area of the cylinder, bubble free, distilled water to the cylinder filled with 1,000 mL
Ac, by using a commercial-grade meter stick or tape measure of distilled water. Measure the change in height of the water
reading in cm or mm. Insert the meter stick or tape measure surface to 0.1 mm. If the water surface raises within 5 % of the
into the cylinder until it touches the inside bottom of the expected value (150 ⁄ ρw × (10 ⁄Ac)), then the area is consid-
cylinder. Read and record the distance from the inside bottom ered uniform. If this tolerance isn’t met, the cylinder shall not
of the cylinder to the 1,000 mL mark to the nearest 1 mm. be used.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
19
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
APPENDIX
(Nonmandatory Information)
X1.1 General—The data sheets in Fig. X1.1and Fig. X1.2 relationship for the H151 Hydrometer and Fig. X1.2 data are
are provided as examples to assist the user by showing results for the H152 Hydrometer. Figs. X1.4 and X1.5 are examples of
of the calculations performed. Fig. X1.3 is provided as an typical checks of the hydrometer and sedimentation cylinder.
example of how the results of Fig. X1.1 can be displayed Fig. X1.6 is an example of the meniscus correction and
graphically. Fig. X1.1 data are calculated using the calibration determination of the A constant for a 151H hydrometer.
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
20
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
FIG. X1.1 Example Data Sheet Using Calibration Relationship and H151 Hydrometer
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
21
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
FIG. X1.2 Example Data Sheet Using Calibration Relationship and 152H Hydrometer
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
22
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
23
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
24
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
25
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
FIG. X1.6 Example of Constant A and Meniscus Correction Determination for H151 Hydrometer
SUMMARY OF CHANGES
In accordance with Committee D18 policy, this section identifies the location of changes to this standard since
the last edition (2017) that may impact the use of this standard. (May 1, 2021)
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
26
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.
D7928 − 21´1
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
Copyright by ASTM Int'l (all rights reserved); Tue Sep 7 19:20:12 EDT 2021
27
Downloaded/printed by
Jony Gutirrez (Jony Gutirrez) pursuant to License Agreement. No further reproductions authorized.