ADV-PLUS Operator Manual - SO8816
ADV-PLUS Operator Manual - SO8816
ADV-PLUS Operator Manual - SO8816
Operator Manual
Model: PT-SR-24A-ADVPLUS
Serial Number: 040419
Help Consolidated Help You
Always have this sterilizer’s model and serial number (printed on the first page of this
document) available when calling Consolidated or when calling your Consolidated
representative for service or parts.
CHLORIDES are known to damage the stainless steel material and cause
cracks and pin holes thereby leading to steam leaks.
Copyright © 2012:
Consolidated Sterilizer Systems
76 Ashford Street
P.O. Box 297
Boston, MA 02134-0003 U.S.A.
Tel: (617) 782-6072 Fax: (617) 787-5865
1. INTRODUCTION 6
2.1 Description 7
2.2 Printer 7
2.3 Alarms 8
2.4 Features 8
2.6 Options 8
3.1 Manager 9
3.1.1 Manager Password 10
3.1.2 Cycle Cnt Reset 10
3.1.3 Print Options 11
3.1.4 Vacuum Leak Test (Hi-Vacuum Units only) 11
3.1.5 Time/Date 12
3.1.6 Auto Jacket Times 13
3.1.7 English/Metric Units 13
3.1.8 Jacket Blowdown (if ordered) 14
3.1.9 Passwords Used/Not Used 14
3.1.10 Operator Passwords 14
3.1.11 Gasket Operations (for EZ-Glide™ Vertical Door only) 15
3.2 Service 16
6. ERROR MESSAGES/WARNINGS 34
Touch Screen
On/Off Switch
Printer
2.1 Description
The Advantage PLUS has been designed to bring a high degree of flexibility to sterilizers. A
touch screen is used for system status displays and operator prompts. This enables the
operator to easily control and view all important sterilization parameters and current options
from a single panel.
The control unit contains a battery backup system which saves all machine settings, user
preferences, cycle settings, as well as the date and time in the event of a power loss. For
convenience, the Advantage PLUS allows users to store up to twelve preset cycles, complete
with all necessary parameters in memory. This allows users to easily switch between cycles
without reentering critical values, thereby eliminating problems associated with erroneous
sterilization parameters.
2.2 Printer
The printer is used to generate a hard copy record of each cycle. All pertinent information
about the cycle is printed, including time and date. The unit prints the current time,
temperature, and pressure or vacuum at user selected intervals. Status messages are also
printed as the sterilizer moves from one cycle phase to another. Refer to the table of contents
in this manual for information on servicing and changing the printer paper.
2.4 Features
The optimum jacket temperature for each cycle is set by the factory but can be altered by
service personnel. An automatic jacket control program turns jacket steam on and off at
preset times without operator initiation. Automatic jacket control for each day of the week may
be programmed differently. There is also a jacket control key on the touch screen Main Menu
or Cycle Select Menu.
The print interval is fully programmable and the printer may be disabled if not needed.
Each Consolidated sterilizer is equipped with an adjustable wastewater cooling valve. The
purpose of this valve is to bring the temperature of the waste to 140°F or lower before it
leaves the sterilizer. This valve is completely mechanical. It operates independently of the
sterilizer control system and power supply. If the temperature of the drain is too high or too
low, the valve can be adjusted by turning the screw on its top face.
https://consteril.com/products/smart-options/cloud-enabled-autoclaves/
Consolidated autoclaves can be configured to include remote troubleshooting with live factory
support; ideal for facilities that cannot afford any down-time or that are located in remote
areas where service assistance is hours or days away.
2.6 Options
This sterilizer can be equipped with WaterEco® water savings options to reduce the amount
of water consumed by the sterilizer.
A flood monitor switch is an option, which is designed to detect flood conditions and display
warnings to the user on the Touch Screen (see Section 6.2).
3.1 Manager
To access the Manager Menu, press the MANAGER key in the lower left hand corner of the
main menu. The sterilizer will ask for the manager password. If known, enter the password
and press the enter key . The factory-configured manager password is set to 10.
Figure 3.2 shows a typical Manager Screen. Across the top of this screen the current
Chamber Temperature, Chamber Pressure, and Jacket Temperature are displayed. This
information is for chamber status only and is not settable. This screen also contains several
manager functions.
3.2 Service
The Service Menu is password-protected and only accessible by service personnel. To learn
more about the Service Menu refer to the Service Manual.
NOTE: Some cycles described in this section are optional and may not be
included with this particular sterilizer.
BURN HAZARD
The operator MUST wear protective clothing including face shield, thermal
gloves, and proper lab attire, when attempting to load and unload the chamber
materials.
Steam released from the sterilizer chamber can cause serious burns. Stand
away while opening the door.
If water leaks from the front of the sterilizer, DO NOT open the door. Contact
service personnel.
Never try to forcibly open the doors. If they are hard to open, it is because they are electrically
locked, the door lock solenoids have jammed, or the unit is under pressure. First turn the
door’s handles slightly in the “tighten” direction, and then allow the machine to cool for twenty
or thirty minutes before attempting to open the door again. If the problem persists please
contact your CSS service technician.
NOTE: The user can end a cycle at any point by pressing the red ABORT key in the
lower left hand corner of the screen. If the user chooses to abort the cycle, then
the sterilizer will advance to the exhaust phase immediately.
6. After the conditioning phase has completed, the sterilizer will begin to fill the chamber
with steam until the desired temperature has been reached. This chamber
temperature was set in steps 3 and 4. The screen will read “Waiting for Chamber
Temperature”.
7. Once the desired chamber temperature has been reach, the sterilizer will enter the
Sterilization phase. The sterilizer will maintain chamber temperature and pressure for
the amount of time specified in steps 3 and 4. The screen will read “Sterilizing”.
8. Once the sterilize time has completed, the sterilizer will enter the Exhausting/Drying
phase. At this point, the sterilizer will exhaust steam from the chamber, and dry the
load until the specified dry time has elapsed. The screen will read Exhausting/Drying”.
9. For hinged-door units, the end-of-cycle buzzer will sound as soon as the dry time has
elapsed. If the sterilizer is equipped with an EZ-Glide™ Vertical Door, the door will be
released from the gasket by the vacuum system. Once this operation is complete, the
end-of-cycle buzzer will sound. The cycle is now complete.
INJURY HAZARD
If the wrong cycle is selected to sterilize liquids, the containers may burst or crack
during the processing. Additionally, liquids may boil over resulting in loss of media and
spillage within the chamber.
After completion of a Liquids cycle, take care that the fluids are not agitated during
removal from the chamber. Otherwise the containers may burst or crack.
1. From the Main Menu, touch the CYCLE SELECT key on the lower-right hand corner of
the screen. If passwords are enabled, the user will be prompted to enter his or her
password. Enter the operator password and touch enter . If passwords are disabled,
the cycle select screen will appear.
2. Select LIQUIDS from the list of cycle types.
75 25
250 30
500 40
1000 45
1500 50
2000 55
NOTE: The user can end a cycle at any point by pressing the red ABORT key
in the lower left hand corner of the screen. If the user chooses to abort the
cycle, then the sterilizer will advance to the exhaust phase immediately.
6. After the conditioning phase has completed, the sterilizer will begin to fill the chamber
with steam until the desired temperature has been reached. This chamber
temperature was set in steps 3 and 4. The screen will read “Waiting for Chamber
Temperature”.
7. Once the desired chamber temperature has been reach, the sterilizer will enter the
Sterilization phase. The sterilizer will maintain chamber temperature and pressure for
the amount of time specified in steps 3 and 4. The screen will read “Sterilizing”.
8. Once the sterilize time has completed, the sterilizer will enter the Exhausting phase.
At this point, the chamber will very slowly reduce pressure until it reaches 0 psi. This
slow exhaust allows the liquids to gradually cool to prevent boil-over.
9. After the chamber has reached 0 psi, the sterilizer will read “Waiting for Cool Down”
for ten minutes. This phase will allow the liquids to cool further. During this time, DO
NOT attempt to open the door.
10. For hinged door units, once the cool down time has elapsed, the end-of-cycle buzzer
will sound. If the sterilizer is equipped with an EZ-Glide™ Vertical Door, the door will
be released from the gasket by the vacuum system. Once this operation is complete,
the end-of-cycle buzzer will sound. The cycle is complete and the door may be
opened.
6. After the conditioning phase has completed, the pulse-vac phase will begin. For each
pulse, the chamber will first pressurize with steam, and then pull vacuum to the
specified amounts in order to remove air from the chamber prior to sterilization.
7. After the pulse-vac phase has completed, the sterilizer will begin to fill the chamber
with steam until the desired temperature has been reached. This chamber
temperature was set in steps 3 and 4. The screen will read Waiting for Chamber
Temperature.
8. Once the desired chamber temperature has been reach, the sterilizer will enter the
Sterilization phase. The sterilizer will maintain chamber temperature and pressure for
the amount of time specified in steps 3 and 4.
9. Once the sterilize time has completed, the sterilizer will enter the Exhausting/Drying
phase. At this point, the sterilizer will exhaust steam from the chamber, and dry the
load until the specified dry time has elapsed. If the sterilizer is equipped with an EZ-
Glide™ Vertical Door, the door will be released from the gasket by the vacuum
system. Once this operation is complete, the end-of-cycle buzzer will sound.
10. The cycle is now complete.
NOTE: The user can end a cycle at any point by pressing the red ABORT key
in the lower left hand corner of the screen. If the user chooses to abort the
cycle, then the sterilizer will advance to the exhaust phase immediately.
7. The sterilizer will now run the specified number of pulses. For each pulse, the
chamber will first pressurize with steam, and then pull vacuum to the specified
amounts in order to remove air from the chamber prior to sterilization.
8. After the pulse-vac phase has completed, the sterilizer will begin to fill the chamber
with steam until the desired temperature has been reached. The screen will read
Waiting for Chamber Temperature.
9. Once the desired chamber temperature has been reach, the sterilizer will enter the
Sterilization phase. The sterilizer will maintain chamber temperature and pressure for
the amount of time specified.
10. Once the sterilize time has completed, the sterilizer will enter the Exhausting phase.
At this point, the sterilizer will exhaust steam from the chamber. If the sterilizer is
equipped with an EZ-Glide™ Vertical Door, the door will be released from the gasket
by the vacuum system. Once this operation is complete, the end-of-cycle buzzer will
sound.
11. The cycle is now complete.
RISK OF INFECTION
Please note that airborne microbial and particulate contamination is likely to be high in
the decontamination area of the sterile processing department. Users MUST wear
protective clothing and equipment when preparing goods for sterilization or when in
the decontamination area.
All packages must be positioned in the chamber to allow free circulation and
penetration of steam, enhance air elimination, and prevent entrapment of air or
water.
Items capable of holding water, such as basins and trays, should be oriented in
the same direction and arranged so that any condensation will drain out. Wicking
material should be used to absorb condensate.
Large or heavy metal items should be placed on the lowest shelf to enable the
condensate to drain out without wetting any other items in the load.
When using non-woven wrap, ensure there are no low spots that might hold
condensate.
Instruments, such as forceps, must be sterilized in the open position.
CHAMBER OVER PRESSURE The chamber has gone over the pressure
set point threshold (>35psi) and the cycle
has been aborted. Possible bad Pressure
Transducer or possibly sticky steam valve.
CHAMBER TEMP HAS NOT REACHED Chamber has taken longer than 60 minutes
SETPOINT to reach sterilization temperature. Check
steam supply, chamber trap.
JACKET TEMP HAS NOT REACHED Jacket has not reached setpoint in 30
SETPOINT minutes. Check jacket steam supply.
A demo of the SteriNet Cloud service and a getting started guide can be found online at:
https://consteril.com/getting-started-with-the-cloud-connection-portal/
2. If you don’t have an account. Click on the “New Users” button. Clicking this button
will take you to the TetraScience Website. Follow the instructions on the screen.
3. If you have an account. Click on the “Login” button. Clicking this button will take you
to the TetraScience Website. Follow the instructions on the screen.