ZEISS EVO
Scanning Electron Microscope
ZEISS EVO
Original Manual
© 2021 Without the prior written consent of ZEISS, this document or any part of it must neither be translated nor reproduced or
transmitted in any form or by any means - including electronic or mechanic methods, by photocopying, recording or by any in-
formation or filing system. The right to make backup-copies for archiving purposes shall remain unaffected thereby. Any viola-
tions may be prosecuted as copyright infringements.
The use of general descriptive names, registered names, trademarks, etc. in this document does not imply that such names are
exempt from the relevant intellectual property laws and regulations and therefore free for general use. This shall also apply if
this is not specifically referred to. Software programs shall entirely remain the property of ZEISS. No program or subsequent up-
grade thereof may be disclosed to any third party, copied or reproduced in any other form without the prior written consent of
ZEISS, even if these copies or reproductions are destined for internal use at the customer's only, the only exception being one
single back-up copy for archiving purposes.
ZEISS reserves the right to make modifications to this document without notice.
ZEISS Table of Contents
Table of Contents
1 General Information........................................................................ 6
1.1 Text Conventions and Link Types ........................................................................ 6
1.2 Explanation of Warning Messages and Additional Information............................ 7
1.3 Further Applicable Documents ............................................................................ 8
1.4 Contact............................................................................................................... 8
2 Safety ............................................................................................. 9
2.1 Intended Use ...................................................................................................... 9
2.2 General Safety Information ................................................................................. 10
2.2.1 Requirements for Operators .................................................................. 10
2.2.2 Safe Operating Condition...................................................................... 10
2.3 Prevention of Hazards ......................................................................................... 11
2.3.1 Mechanical Hazards .............................................................................. 11
2.3.2 Electrical Hazards .................................................................................. 11
2.3.3 Hazards generated by Materials and Substances ................................... 12
2.3.4 Hazards generated by Radiation............................................................ 13
2.3.5 Thermal Hazards ................................................................................... 13
2.4 Warning Labels and Lights .................................................................................. 13
2.4.1 Warning Labels ..................................................................................... 14
2.5 Safety Devices and Interlocks .............................................................................. 18
2.5.1 Protective Cover Panels ......................................................................... 18
2.5.2 Main Disconnect Device ........................................................................ 19
2.5.3 Circuit Breaker ...................................................................................... 19
2.5.4 Locking Devices .................................................................................... 20
4 Installation...................................................................................... 55
7 Troubleshooting .............................................................................114
7.1 Chamber.............................................................................................................115
7.1.1 Initializing the Stage.............................................................................. 115
7.1.2 Defining the Post Initialization Position of the Stage.............................. 116
7.1.3 Changing the Joystick TV Angle ............................................................ 116
7.1.4 Resetting the Touch Alarm .................................................................... 117
7.1.5 Checking the Temperature .................................................................... 117
7.1.6 Replacing the Chamber Door Seal ......................................................... 117
7.1.7 Adjusting the Isolation Mounts ............................................................. 119
7.2 Column ...............................................................................................................120
7.2.1 Baking out the Gun Head...................................................................... 120
7.2.2 Replacing Filaments .............................................................................. 121
7.3 Power Circuit ......................................................................................................147
7.3.1 Checking the Position of the Circuit Breakers ........................................ 147
7.4 PC.......................................................................................................................147
7.4.1 Cleaning up the PC ............................................................................... 147
Glossary ..........................................................................................173
Index...............................................................................................176
1 General Information
This Instruction Manual (further called "Manual") is considered to be part of the EVO workstation.
Herein after referred to as Microscope System.
This document contains basic steps and safety information that must be observed during opera-
tion and maintenance. Therefore, the document must be read by the operator prior to commis-
sioning and must always be available at the place of use of the Microscope System.
This document is an essential part of the Microscope System and, if the Microscope System is
resold, the document must remain with the Microscope System or be handed over to the new
owner.
Select Tools > Goto Control Panel > Air- Follow a path in the software.
lock.
See: Text Conventions and Link Types [} 6]. Link to further information for this topic.
DANGER
Type and source of danger
DANGER indicates an imminently hazardous situation which, if not avoided, will result in death
or serious injury.
WARNING
Type and source of danger
WARNING indicates a potentially hazardous situation which, if not avoided, may result in
death or serious injury.
CAUTION
Type and source of danger
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor
or moderate injury.
NOTICE
Type and source of danger
NOTICE indicates a potentially harmful situation which, if not avoided, may result in property
damage.
Info
Provides additional information or explanations to help operator better understand the con-
tents of this manual.
1.4 Contact
If you have any questions or problems, please contact your local ZEISS Sales & Service Partner or
one of the following addresses:
Headquarters
Email: info.microscopy.de@zeiss.com
Email: courses.microscopy.de@zeiss.com
ZEISS Portal
The ZEISS Portal (https://portal.zeiss.com/) offers various services that simplify the daily work with
your ZEISS systems (machines and software). It is constantly improved and extended to meet your
needs and requirements better.
Service Germany
Email: service.microscopy.de@zeiss.com
2 Safety
This chapter contains general requirements for safe working practices. Any person using the Mi-
croscope System or commissioned with installation or maintenance must read and observe these
general safety instructions. Knowledge of basic safety instructions and requirements is a precondi-
tion for safe and fault-free operation. Operational safety of the supplied Microscope System is
only ensured if it is operated according to its intended use.
If any work is associated with residual risks, this is mentioned in the relevant parts of this docu-
ment in a specific note. When components must be handled with special caution, they are
marked with a warning label. These warnings must always be observed.
Info
Not for therapeutic, treatment, or medical diagnostic evidence.
Info
Not all products are available in every country. Contact your local ZEISS representative for
more information.
PC The Zeiss SEM PC is supplied “as is” and is untested with any third-part software. It is not recom-
mended to install any third-part software on the Zeiss SEM PC as this might cause instabilities
and/or break Carl Zeiss software installations.
SmartSEM The SmartSEM software is intended for the operation of scanning electron microscopes (SEMs).
Software The SmartSEM software has to be run exclusively on a personal computer delivered by ZEISS. Any
other application is not allowed.
Regarding the operation of the microscope, the following regulations must be met:
§ Only operate the microscope according to the operating conditions after correct installation
by a ZEISS service representative.
§ The microscope is only to be used by operators who have been trained by a ZEISS service rep-
resentative. Basic operator training and safety instructions will be provided within the scope
of the initial start-up by ZEISS. Make sure that everyone who operates the microscope only
performs the tasks for which he/she is trained.
§ Operators of the microscope must not deviate from the instructions provided in this manual.
§ Only perform preventive maintenance tasks described in this manual. All tasks of mainte-
nance, service, and repair not described in this manual have to be performed by an authorized
ZEISS service representative.
§ The microscope is to be used in a laboratory environment for commercial and scientific pur-
poses only.
Using the microscope for any other purpose is not allowed and can be hazardous.
Improper use of the Microscope System can easily lead to impairment of its function or even dam-
age to the Microscope System. Damage caused by incorrect operation, negligence or unautho-
rized intervention, in particular by removing, modifying or replacing parts of the Microscope Sys-
tem, cannot be held liable by the device manufacturer. Third-party devices or components that
are not expressly approved by ZEISS may not be used.
The Microscope System, components and accessories may only be operated and maintained by
authorized and trained personnel. The Microscope System may only be used in accordance this
document. If the Microscope System is not used as described, the safety of the user may be im-
paired and/or the Microscope System may be damaged.
Any unauthorized intervention or use other than within the scope of the intended use shall void
all rights to warranty claims. The regional regulations on health protection and accident preven-
tion must be observed at all times and during all work on and with the Microscope System.
Training Authorized ZEISS personnel will provide basic training in operating the Microscope System. As
well as information on equipment safety and maintenance work that can be conducted by the op-
erator. The training will be documented by ZEISS and its completion is to be confirmed by the op-
erator.
Special application training is offered for a fee. Current training dates, additional information and
the registration form can be found at https://www.zeiss.com/microscopy/int/service-support/train-
ing-and-education.html.
If circumstances occur which impair safety and cause changes in operating behavior, the Micro-
scope System must be shut down immediately and a ZEISS service representative should be in-
formed. The Microscope System can only be operated after proper installation by a ZEISS service
representative and in compliance with the operating conditions. The Microscope System may only
be operated after correct installation by a ZEISS service representative and if the operating condi-
tions are adhered to.
§ Do not operate the Microscope System until you have completely read and understood the
entire documentation delivered with the Microscope System.
§ Make sure that all protective cover panels are installed and all warning labels are available and
legible.
§ Ensure conditions and take measures to prevent the build up of electrostatic charge on the
workplace.
Hazardous Voltage High voltages are present inside the microscope. Contact may cause burn or electrical shock.
inside the § Ensure proper grounding. For more information, refer to the Installation Requirements docu-
Microscope ment.
§ Only authorized ZEISS service representatives are allowed to service the microscope.
§ Do not try to service the microscope yourself.
High Leakage High leakage currents are present in the microscope. Contact may cause burn or electrical shock.
Currents § Ensure proper grounding. For more information, refer to the Installation Requirements docu-
ment.
§ Do not operate the microscope without the separate ground connection.
Opening the Gun If you open the gun while the mains power is connected to the microscope, then contact may
cause electrical shock.
§ Always disconnected the mains power from the microscope before you open the gun.
Residual Voltage After unplugging the mains plug residual voltage is present at the pins of the plug which may
at the Mains Plug cause electrical shock.
§ After unplugging the mains plug wait at least 5 s before touching the pins of the mains plug.
Stage on Bias If you open the chamber door while the stage is still on bias voltage, then contact may cause elec-
Voltage trical shock.
§ Always switch off the beam deceleration before you open the chamber door.
Malfunction of Magnetic fields present at the ion getter pumps may disturb the function of medical devices. The
Medical Devices magnetic fields are also present if the microscope is switched off.
near Ion Getter If you wear medical implants that are susceptible to magnetic fields (e.g. cardiac pacemakers), do
Pumps the following:
§ Keep a distance of at least 30 cm from the ion getter pumps.
§ Follow the safety instructions provided by the pump manufacturer.
Biological Biological substances may pose a threat to the health of humans and other living organisms.
Substances § Keep a logbook of the biological substances loaded into the microscope and show it to the
ZEISS service representatives before they perform any work on the microscope.
Lack of oxygen Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange and parts
exchange. Inhaling nitrogen may cause unconsciousness.
§ During specimen exchange, keep the chamber door open as short as possible.
§ Do not inhale the air from within the specimen chamber.
§ Ensure that the area around the microscope is sufficiently ventilated.
§ If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of mental
alertness and/or muscular coordination, depression of sensations, emotional instability, fa-
tigue) leave the room immediately and inform the facility’s safety officer.
Concerning the hazards of nitrogen installations and associated safety precautions refer to the
current version of the guideline IGC Doc 44/18: Hazards of Oxygen-Deficient Atmospheres, pub-
lished by EIGA (European Industrial Gases Association).
To download the document:
1. Go to EIGA homepage www.eiga.eu.
2. Select Publications > EIGA Documents.
3. From the list, select Doc. 44/18.
4. Click Download.
Aggressive or Aggressive or toxic chemicals can cause chemical burns.
Toxic Chemicals § When handling aggressive or toxic chemicals, wear suitable protective clothing, gloves, and
eye/face protection.
§ Do not eat, drink, or smoke while handling toxic chemicals.
§ Refer to local safety regulations.
§ Read and follow the instructions in the material safety data sheet of the chemical. The mate-
rial safety data sheet can be obtained from the supplier of the chemicals.
Oil Mist around The rotary pump RV12 may emit oil mist especially at high gas load (e.g. door accidently open).
the Rotary Pump Inhaled oil mist is harmful to health.
§ Regularly service the oil mist filter (refer to pump manual and material safety data sheets).
Recommended is the connection to an exhaust line.
Environmental When disposing of aggressive or toxic chemicals, there is a threat of damage to the environment.
risk: Disposal of § When disposing of waste that has been generated during a service operation (e.g. used rotary
aggressive or toxic pump oil), comply with all national and local safety and environmental protection regulations.
chemicals
X-rays X-rays are generated inside the microscope during operation. This is unavoidable because elec-
trons are accelerated by voltages up to 30 kV.
§ Do not remove any parts around the column and chamber that are essential for radiation pro-
tection.
§ Use genuine ZEISS parts exclusively.
§ Ensure that all local safety and X-ray protection regulations are met.
§ Only authorized ZEISS service representatives are allowed to service the microscope.
An electron microscope is a radiation source in the sense of the German regulations Strahlen-
schutzgesetz (StrlSchG) and Strahlenschutzverordnung (StrlSchV). This also applies if the X-ray ra-
diation generated is evaluated by detectors (WDX, EDX). The operation of an interfering emitter is
subject to the regulations of StrlSchG and StrlSchV.
The design of the scanning electron microscope ensures that the acceleration voltage of 30 kV
and a local dose rate of 1 µSv / h at a distance of 0.1 m from the outer surfaces of the device are
not exceeded. The operation of these devices does not require a permit in Germany according to
StrlSchG and StrlSchV.
Every single device is subjected to a routine test at Carl Zeiss Microscopy GmbH in Oberkochen. A
certificate of this test is enclosed with the device (test certificate). The routine test refers to the
device in the delivered configuration.
The following points must be observed:
§ A notice is attached to the device, which indicates that X-rays are generated inside the micro-
scope.
§ Subsequent changes to the device by the user usually result in a new assessment of the de-
vice's classification with regard to the StrlSchG and StrlSchV. The user is responsible for com-
plying with the limit values prescribed in the StrlSchG and StrlSchV and the necessary check-
ing of the local dose rate after modifications.
Outside Germany, the user of the microscope has to comply with the local regulations of the
country where the microscope is operated.
Contact Radiation For questions regarding radiation protection, contact the ZEISS Radiation Safety Officer, Carl Zeiss
Protection AG, 73447 Oberkochen, Germany
phone: +49 (0) 7364 20 0
Bakeout Parts of the enclosure in the upper range of the column may become hot during bakeout, particu-
larly after a long bakeout cycle.
§ Do not place any combustible objects on the grids of the electron optical column during bake-
out.
§ After the bakeout procedure, let surfaces cool down before working around the column.
§ Only advanced operators are allowed to perform the bakeout procedure.
Appropriate safety labels on the microscope warn operators of hazards. Each safety label is af-
fixed close to the point where a particular hazard exists. Several labels also provide legal informa-
tion.
Front Side of
Microscope
1
2
3
4
Rear Side of
Microscope
6
7
Gun Head
underneath
Cladding
10
11
12
13
14
15
CEE Connector
16
The safety labels always need to be attached to the correct spots at the microscope. If a safety la-
bel is lost or unreadable, it needs to be reordered via the following reorder numbers:
4 At the front of the chamber door. Risk of electrical shock by stage on bias volt-
age
Make sure the beam deceleration is switched off
before opening the chamber door.
Reorder no. 344700-0033-000-05en
13 At the rear of the microscope The Type plate contains a unique serial number,
the EVO model designation with the according
Carl Zeiss Microscopy GmbH part number, the manufacturing year, and the
73447 Oberkochen, Germany
mains voltage for which the EVO is configured,
Serial number either 230 VAC or 120 VAC.
EVO15
354900-9041-000 Manufactured in year 2017 Order no. 354900-9040-000 (EVO10)
1/N/PE 230 VAC
50 - 60 Hz max. 16 A Order no. 354900-9041-000 (EVO15)
Order no. 354900-9042-000 (EVO25)
Made in UK/GB
Due to hazardous voltages and X-rays inside the microscope, the microscope is equipped with
protective cover panels.
Operation of the microscope is only allowed with attached protective cover panels.
To disconnect the microscope with all its components from the mains supply, unplug the CEE con-
nector (blue for 230 VAC, yellow for 120 VAC).
The circuit breaker (F10) at the rear of the plinth will disconnect the mains power from the elec-
tronics in the plinth in case of an over-current (including the first pre-vacuum pump and PC). The
other circuit breaker (F11) disconnects mains power from the heaters and the second pre-vacuum
pump.
However, some other components will still be connected to mains power.
The vacuum locking device ensures that gun vacuum and system vacuum are better than the re-
quired thresholds.
Several hardware and firmware measures ensure a safe operation of the beam deceleration kit.
Before a stage bias potential can be applied, a vacuum switch is checked, the turbo pump signal is
checked and the CAN bus communication is checked.
5
2
6
3
7
8
Optional A range of options and accessories is available for the microscope. Examples of available options
Components and and accessories are the following:
Accessories
Detectors Detectors for variable pressure applications
Refer to VPSE Detector [} 39], C2D Detector [} 42], and C2DX de-
tector
BSD detectors for high efficiency and angle selective material charac-
terization.
Refer to BSD Detector (HDBSD, BSD1) [} 43] and YAG BSD Detector
Stage accessories Peltier Coolstage for imaging of hydrated specimens with the Ex-
tended Pressure (EP) mode
Refer to Hydrated Specimen [} 97]
Faraday cup
Refer to Measuring the Probe Current [} 81]
Control panel that allows direct access to the most frequently used
functions
Refer to Control Panel
For full details about the available options and accessories, contact your local ZEISS service repre-
sentative, or sales representative.
5
1
2 6
7
3
8
4
Gun Vacuum In the gun head, a high vacuum is maintained by an ion getter pump. The vacuum in the gun
head is displayed as Gun vacuum in the SmartSEM user interface. It should be below
5 × 10⁻⁷ mbar.
Venting The specimen is located in the evacuated specimen chamber. To open the specimen chamber for
specimen exchange, you have to break the vacuum in a controlled manner. This is done by the
Vent command via the SmartSEM user interface or by pressing the Exchange push button on the
optional control panel.
When the Vent command is received, gaseous nitrogen flows into the specimen chamber via the
vent valve 8 . As soon as the pressure equilibrium is obtained, the chamber door can be
opened to change the specimen.
Evacuating In order to continue operation, the Pump command makes the pre-vacuum pump and the turbo
pump evacuate the specimen chamber.
As soon as the vacuum in the specimen chamber is ready for operation, the column chamber
valve opens and the EHT Vac ready message is displayed in the SmartSEM user interface. Gun
and EHT can be switched on.
Vacuum Modes The microscope can be operated in different vacuum modes:
§ High vacuum (HV) mode
§ Variable pressure (VP) mode (optional)
§ EasyVP mode (optional)
§ Extended pressure (EP) mode (optional)
System Vacuum in In HV mode, both gun and chamber are at high vacuum. The 20 μm or 30 μm mid-column aper-
HV Mode tures are used for the beam profile control. No pressure limiting aperture is mounted under the
objective lens.
HV mode provides the means to obtain higher resolution and enables using the SE detector.
HV mode is the regular operation mode for standard applications and is recommended for imag-
ing conductive specimens.
System Vacuum in The configuration of the microscope needs to be changed to enable variable pressure modes (VP,
Variable Pressure EasyVP and EP Mode), refer to Variable Pressure Modes [} 23].
Modes
Variable pressure (VP) modes enable imaging specimens that are non-conducting, strongly
gassing, or hydrated, without the need for vapor deposition or other preparation procedures.
This is made possible by using a differential pumping system. A pressure-limiting aperture (differ-
ential pumping aperture DPA) is mounted under the objective lens, which allows partial pressures
above 10 Pa to be set in the specimen chamber while maintaining high vacuum in the gun area.
Three different pressure-limiting aperture configurations are available:
§ VP mode with 100 μm VP aperture
§ EasyVP mode with 400 μm EasyVP aperture
§ EP mode with 100 μm EP aperture and 500 μm or 1000 μm beamsleeve aperture.
The following table summarizes the hardware, applications and pressure ranges of the different
vacuum modes:
* If your microscope is a VP type instrument with the through-the-lens pumping option, then
the maximum pressure in VP mode can be increased to 400 Pa.
For all charge compensation methods, both, air and water vapor can be introduced into the
chamber as a charge compensating gas.
Purpose VP mode and EasyVP mode allow partial pressures above 10 Pa to be set in the specimen cham-
ber while maintaining high vacuum in the gun area.
22
3
4
5
VP Mode The 100 μm VP aperture is mounted under the objective lens. The 750 μm mid-column aperture is
Configuration used for imaging. The beam first goes through the mid-column aperture and then passes through
the VP aperture before landing on the specimen surface. In VP mode configuration, only the Nor-
mal and Analysis Optibeam modes can be used for imaging. It is possible to switch between HV
mode and VP mode when using this configuration. The maximum pressure that can be achieved
in the VP mode configuration is 400 Pa for EVO W systems, 273 Pa for EVO LaB₆ and EVO HD mi-
croscopes.
In VP mode configuration, focus wobble is not available.
EasyVP Mode The 400 μm EasyVP aperture is mounted under the objective lens. The 20 μm mid-column aper-
Configuration ture is used for beam control. In EasyVP mode configuration, the beam first goes through mid-col-
umn aperture and then passes through the EasyVP aperture before landing on the specimen sur-
face. The maximum pressure that can be achieved in the EasyVP mode configuration is 133 Pa for
all EVO series microscopes.
In the EasyVP mode configuration, Optibeam modes such as Resolution, Depth, and Analysis
modes can be used for imaging. It is possible to switch between HV mode and EasyVP mode and
the beam alignment is optimum in both conditions without adjusting any gun or column parame-
ters.
In this EasyVP mode configuration, focus wobble is available.
Function The residual gas atmosphere in the specimen chamber creates an interaction region of electrons
and residual gas molecules between the objective lens and the specimen. In this region, high-en-
ergy electrons in the primary electron beam hit the residual gas molecules and ionize them. The
ions generated in these collisions contribute to the compensation of negative charge on the speci-
men.
However, another effect of these collisions is to scatter the electron beam. This is called the “skirt
effect”. The electrons that are lost from the primary beam as a result of this effect provide only a
resolution-limited background signal for imaging purposes. Although it is possible to tolerate
these leakage losses at chamber pressures up to a few hundred Pa, it is necessary to carefully se-
lect and control the important factors such as acceleration voltage, chamber pressure, and beam
path. The signal-to-noise ratio in variable pressure mode can also be improved via the noise re-
duction features of SmartSEM. For details refer to the Software Manual SmartSEM.
Fig. 8: Non-conducting specimen imaged with an acceleration voltage of 20 kV and a 30 μm aperture. Left: High vac-
uum (HV) mode, showing strong charging effects. The electron beam is distorted and high-quality imaging is not pos-
sible. Right: VP mode at 21 Pa. Charges are completely compensated, allowing easy imaging of the specimen.
Purpose Extended Pressure (EP) mode is required for imaging hydrated specimens. This will enable the
study of hydrated specimens in their native state with little or no loss of water in the microscope
(the specimen is kept fully hydrated during the pump down). This pressure mode is typically used
with a Peltier coolstage.
3
4
5
1
When the microscope is configured for EP mode, the 750 μm mid-column aperture is used and
the 100 μm EP aperture is mounted under the objective lens in combination with one of the
beamsleeves apertures (500 μm or 1000 μm).
For achievable pressure ranges, refer to Variable Pressure Modes [} 23].
1
2
7
8
Fig. 10: Schematics of the electron optics
Gun A filament serves as gun 1 .The filament is heated by applying the filament current. The heated
filament emits electrons.
EHT The acceleration voltage (UEHT) accelerates the emitted electrons.
Gun Alignment The gun alignment and emission coils 2 center the electron beam.
and Display Coils
Condenser A double condenser system 3 allows the continuous regulation of the probe current.
Apertures The electron beam passes through the currently selected aperture in the mechanical mid-column
aperture changer 4 to ensure that the optimum quality of beam is maintained.
The design of the aperture changer allows for using 4 apertures. Three apertures are fitted as
standard (positions 0–2). The position of each aperture is marked on the mid-column aperture
changer.
1 30 μm HV mode
2 20 μm HV mode
Optional EasyVP mode (with 400 μm EasyVP
aperture in combination with a beamsleeve
aperture)
Stigmator The stigmator coils 5 compensate for astigmatism so that the electron beam becomes rota-
tionally symmetrical.
Deflection System The electron beam is focused by the objective lens 6 onto the specimen 8
while being de-
flected in a point-by-point scan (raster scan) over the specimen surface by the scanning coils 5
.
Signal Detection When the primary electron beam hits the specimen, certain interaction products are released,
which can be recorded by specific detectors, e.g. the BSD detector 7 . For more information
refer to Principle of Signal Detection [} 30].
The specimen A larger field of Smallest probe Largest depth Largest field of
remains in focus view for naviga- diameter for a of field for a view for naviga-
for any change tion and a large chosen probe chosen probe tion and a very
in probe cur- depth of field. current at any current, at any large depth of
rent. working dis- working dis- field.
tance or accel- tance or accel-
eration voltage. eration voltage.
3.3 Detectors
This chapter describes the generation of secondary and backscattered electrons, and describes
how the detector types, that are available for use with the microscope, use these electrons to pro-
vide imaging, topographical, and other information. It also lists some characteristics of each de-
tector type.
For more information about any of the detectors, contact your local ZEISS service representative.
When a primary electron (PE) beam hits a specimen, certain electron beam interaction processes
occur. The interaction products most frequently used for the generation of images in scanning
electron microscopy are secondary electrons (SEs) and backscattered electrons (BSEs).
Specific types of detectors are able to detect the SEs and BSEs. The detector signals can be used
to create images and produce information about the properties of the specimen.
6
7
Fig. 13: Interaction between primary electron beam and specimen
7 Fluorescent X-rays
Primary Electrons Primary Electrons (PEs) are electrons forming the scanning beam before hitting the specimen.
Secondary Secondary electrons are emitted from the topmost layer of the specimen.
Electrons § SE1 Electrons
Electrons emitted at the point of impact between the beam and the specimen are known as
SE1 type electrons. The amount of electrons emitted at the point of impact is related to the
shape of the specimen.
Secondary electron detectors, such as the InLens SE detector, collect SE1 type electrons from
the surface layer of the specimen and are thus ideal for displaying surface structures.
§ SE2 Electrons
The emergence of backscattered electrons from the specimen excites further emission of sec-
ondary electrons. These are known as SE2 type electrons.
Detectors that collect SE2 type electrons are especially suitable where the working distance is
large. Surface detail as the effect of “lateral illumination” emphasizes the topography of the
specimen.
Backscattered All electrons with energy higher than 50 eV are known as backscattered electrons (BSEs). BSEs are
Electrons generated by elastic scattering in a much deeper range of the interaction volume (up to 1 µm) and
carry depth information. The backscatter coefficient increases with increasing atomic number of
the elements within the specimen. This allows the BSE detector to generate atomic number con-
trast, or compositional contrast images.
The beam scans the specimen and initiates particles to be emitted. A detector collects the emis-
sion and produces an electric signal with an amplitude proportional to the number of particles at
any given time.
C2D Detector [} 42] Measures the current that re- Variable pressure
on VP systems only sults from the ionization of Topography and surface
gas by SEs structure in VP mode
For more details, refer to the document Product Specification of the microscope.
3.3.3 SE Detector
Purpose The SE detector is an Everhart-Thornley type detector. It detects SEs as well as BSEs.
Position The SE detector is mounted on the wall of the specimen chamber, and is therefore classified as a
“chamber detector”. Due to its position in the chamber, the SE detector views the specimen later-
ally.
11
3
4
Fig. 15: Schematics of the SE detector
Function Electrons moving to the detector are attracted/repelled by the collector grid and directed to the
scintillator. The electrons gain energy from the scintillator and thus are able to interact with a
phosphor layer, which generates photons (light). The light travels up a light pipe to a photomulti-
plier. The photomultiplier multiplies the flashes of light and outputs a signal that can be used for
imaging.
The collector voltage can be varied in the range between −250 V and +400 V.
A positive collector voltage generates an electrical field in front of the detector, thus directing the
low energy SEs towards the scintillator.
A negative collector voltage generates a field deflecting the low energy SEs so that they cannot
reach the scintillator and do not contribute to the signal. Only high-energy BSEs reach the scintil-
lator contributing to the image generation. This produces a pseudo-backscattered image, which
shows pronounced material contrast, but largely cancels surface properties and topography.
Fig. 16: Comparison of SE detector images using positive and negative collector bias voltage: SE detector using
+300 V collector bias: Good display of surface structures and topography (left), SE detector using –150 V collector
bias: Extremely strong topography, including shadow generation (right)
Info
For all standard applications, the collector bias should be set to +300 V.
Surface images that show enhanced topographical information can also be generated using
BSE detectors, but they do not show the shadows that can be created using the SE detector.
Applications Unlike the InLens SE detector, which can be used only with acceleration voltages up to 20 kV, the
SE detector can be used in the complete high-voltage range.
Fig. 17: Comparison of surface information at different acceleration voltages: Acceleration voltage = 1 kV: Good, sur-
face-sensitive imaging (left); Acceleration voltage = 10 kV: Thin layers are not seen (right)
Fig. 18: Comparison of surface information at different acceleration voltages: Acceleration voltage = 5 kV: Good, sur-
face-sensitive imaging (left); Acceleration voltage = 15 kV: Transparent surface caused by increased penetration
depth (right)
The working distance has a significant effect on the efficiency of the SE detector. Shadowing ef-
fects occur when the working distance is too short. If the specimen is too close to the objective
lens, most of the electrons will be deflected by the field of the electrostatic lens or move to the
objective lens itself. This means they cannot be detected by the SE detector.
Depending on the specimen material and on the specimen geometry, a minimum working dis-
tance of approximately 4 mm should be used. Extreme signal loss is likely to occur if shorter work-
ing distances than this are used. Conversely, the SE detector is very good when used for imaging
at long working distances. This is particularly important for low magnification imaging that is nec-
essary for adjusting the orientation of the specimen holder or locating a specific area on the speci-
men.
Optimal Initial The following settings provide a good field of view for navigating on the specimen at low magni-
Settings fications:
§ Initial working distance in the range of 10 mm to 20 mm.
§ Acceleration voltage of approximately 10 kV.
The imaging conditions can be readjusted for a desired application after identifying the area of in-
terest on the specimen.
Fig. 19: Field of view at different working distances: The smallest possible magnification depends on the working dis-
tance: Field of view at 30 mm WD: Reasonably good signal-to-noise ratio, minimum magnification levels can be set
(left); Field of view at 5 mm WD: Limits exist on the lowest magnification setting possible (right)
Although images produced by the SE detector always include some backscattered electron com-
ponents, most of the signal is generated by the secondary electrons and the fraction of backscat-
tered signal is negligible. The images obtained by the SE detector are therefore primarily sec-
ondary electron images.
Fig. 20: Comparison of material contrast using SE/BSE detectors on a polished specimen: WD = 5 mm using SE detec-
tor: Material contrast is very poor (left); WD = 9 mm using BSE detector: Clear image showing good material contrast
(right)
In the previous figure the SE image taken at 5 mm working distance shows relatively poor mate-
rial contrast. In this example the reduced yield of secondary electrons is caused by the specimen
preparation technique of polishing the specimen surface. The ratio of SE to BSE electrons is there-
fore altered in favor of backscattered electrons.
Because the SE detector is mounted on the chamber at a certain angle to the specimen, the speci-
men is always viewed laterally. The SE detector, therefore, provides good surface information. All
other detectors (InLens SE and BSE) view the specimen from above, providing only limited infor-
mation about the topography of the specimen. Surfaces tilted towards the detector provide more
surface detail with brighter edges; specimens tilted away from the detector display shadowing ef-
fects and less surface detail.
Fig. 21: Comparison of topographic contrast – SE/BSE detectors on a fracture surface: SE detector: Image with large
depth of field (left); BSE detector: Good material contrast but with limited topographical detail (right)
Some imaging applications require both, compositional and topographical details. The generation
of mixed SE and BSE images is recommended for such applications. The signal mixing option is
available in the Detectors tab in the SEM Controls panel.
Tilting the specimen increases the signal in the SE detector and sometimes improves the topo-
graphical information. Tilting the specimen towards the SE detector also results in a change of the
solid angle in which both the backscattered and secondary electrons are emitted from the speci-
men.
Parameter Description
Acceleration voltage
1 kV to 5 kV Low-voltage applications for the compensation of charges and for
surface-sensitive imaging
5 kV to 20 kV The average voltage range is suitable for many different applications
Working distance
≥ 4 mm If the working distance is too short, shadowing effects occur which
diminish the efficiency of the detector. Below 20 kV, the SEs are ab-
sorbed by the field of the electrostatic lens
6 mm to 12 mm Useful for the average voltage range (5 kV to 20 kV)
12 mm to 30 mm Recommended only for low magnifications and to increase the depth
of field
Collector voltage
0 V to 400 V Variation of the collector voltage at high magnifications to obtain the
mixed signal
Parameter Description
Aperture
Specimen tilt Tilting the specimen towards the detector increases collection effi-
ciency
Purpose The microscope contains a CCD camera (charge-coupled device camera) inside the specimen
chamber. It is referred to as the chamber CCD camera or chamberscope. It allows you to monitor
the position of the specimen stage and particularly the distance between the objective lens and
the specimen holder.
Position The default location of the chamber CCD camera is at the front of the specimen chamber. Other
positions are available optionally.
NOTICE
Risk of collision
Use the chamber CCD camera to monitor the position of the specimen holder during stage
movements. Pay particular attention to the distance between the objective lens and the top of
the specimen. This applies to vertical movements, but also to horizontal movements, because a
thick specimen may collide with the objective lens from the side.
Function The chamber CCD camera has two illumination modes. The chamber can be illuminated either
with white light or with infrared light. Infrared light gives a grayscale image, whereas white light
gives a color image. In standard settings the mode is automatically selected, depending on the
imaging mode and the selected detectors. White light limits the performance of most detectors.
Therefore infrared illumination is a fallback if white light cannot be used. The performance of
diode detectors is negatively affected by infrared light. If a diode detector is selected, then by de-
fault the chamber CCD camera is disabled. The automatic selection of the illumination mode can
be manually overwritten by the user.
Info
It is highly recommended that the CCD illumination control is set to Auto Detect. This avoids
any problems by the user forgetting to switch the illumination back for a different detector.
Fig. 23: Chamber CCD camera disabled as indicated by a pause sign (e.g. if a diode detector is
selected).
11
22
Function The stage can be operated using the dual joystick or using the SmartSEM software.
The focus on the stage is not maintained when the specimen is tilted. This can be compensated
for by using the Compucentric function in SmartSEM.
Purpose The Variable Pressure Secondary Electron (VPSE) detector is a specific type of SE detector for use
in Variable Pressure mode where a standard SE detector cannot be used. The VPSE detector is not
usable and will not operate in HV mode.
The Variable Pressure mode enables analyzing and imaging of non-conducting specimens without
charging artefacts. This is possible, because positively ionized gas molecules stabilize local charg-
ing. Variable Pressure mode can also be used for strongly gassing or moist specimens without any
need for specimen preparation.
1 VPSE detector
The images that are created with the VPSE detector are similar to images that are created with
conventional SE detectors.
Info
If the chamber pressure rises, then the scattering of electrons is increased and the resolution of
the microscope is reduced.
Try to find the optimal chamber pressure for each individual application.
Dwell Time The dwell time is the amount of time that the electron beam stays at one position on the speci-
men before it moves to the next position.
If the dwell time is too short (i.e. the scan rate is too fast), then there is not enough time for an
“ion cascade” to develop and to create the imaging photons. This reduces the efficiency of the
detector.
If the dwell time is too long (i.e. the scan rate is too slow), then the electron beam delivers a large
amount of energy to each individual spot on the specimen. This may result in charging artifacts on
the images.
The optimal dwell time depends on the specimen and needs to be determined by experiment.
Info
To reduce charging effects, use the Frame Averaging function of SmartSEM. Use fast scan
speeds and increase the number of frames (N).
Collector Bias The collector bias corresponds to the voltage that is applied to the VPSE collector. The collector
bias accelerates the secondary electrons from the specimen surface towards the VPSE detector.
Typical VPSE collector bias values are between 50 % and 80 %.
If the collector bias is too low, the efficiency of the detector is reduced.
If the collector bias is too high, the VPSE detector may receive too much signal and may get satu-
rated. In this case, very bright lines are visible on the image in periodic intervals and proper imag-
ing is no longer possible. Whether a given value for the collector bias is too high, depends on the
specimen, the acceleration voltage, the probe current, and the pressure in the specimen chamber.
Fig. 27: Saturation of the VPSE detector. Bright lines become visible in the microscope image
To eliminate the bright lines in the images, either reduce the collector bias or reduce the chamber
pressure. However, the detector efficiency is reduced by either of these adjustments. You need to
find the optimum parameters for imaging different specimens. It is generally better to reduce the
collector bias, because reducing the chamber pressure can cause new charging effects.
Fig. 28: Charge compensation by adjusting chamber pressure and collector bias. Left: Reducing the collector bias
from 82 % to 78 % results in the elimination of banding effects at 40 Pa. Right: Reducing the pressure in the speci-
men chamber from 40 Pa to 20 Pa results in the elimination of banding effects at a collector bias of 79 %.
Parameter Description
Acceleration voltage
Working distance
Aperture
Parameter Description
7.0 μm to 20 μm With these apertures, the probe current is frequently too low to ob-
tain a sufficient signal-to-noise ratio and the required contrast.
Specimen tilt Avoid large angles of tilt, if possible. Slight tilting can improve effi-
ciency.
Purpose The cascade current detector (C2D) is a special type of detector that you can use instead of the SE
detector to create a secondary electron image under variable pressure conditions.
The SE detector cannot be used under variable pressure conditions: The high potential at the scin-
tillator would cause an electrical breakdown with a lightning flash that runs from the scintillator
to the outer body.
Purpose The BSD is a semiconductor backscattered electron detector with exceptionally high sensitivity.
The detector consists of five diode segments, independently configurable. The BSD1 detector of-
fers faster speed and enhanced sensitivity, compared to the HDBSD detector. The BSD1 detector
is required for quantitative applications such as 3DSM.
Fig. 30: BSD detector with five segments in real life (left) and in SmartSEM (right)
In combination with low incident beam energies, the BSD detector can visualize very fine surface
details in high contrast.
Fig. 31: BSE images showing different contrast with different segment configuration of the BSD
detector; material contrast (left) and topography (right)
NOTICE
Motorized specimen stage
Risk of damaging the detector when operating the motorized specimen stage.
4 Retract the detector head completely after you have finished the work with the detector.
Info
Risk of malfunction: The diode segments are sensitive to the light that is used for illumination
in TV mode (infrared and white).
When you use a diode detector, always make sure that the TV illumination is switched off. If
the CCD Mode is set to Auto Detect, then the TV illumination is automatically switched off
when a diode detector is used.
The BSD detector has applications mainly in materials analysis and in the life sciences.
Material analysis:
§ Metallurgical sections
§ Geological sections
§ Complex materials
Purpose The YAG BSD detector is a scintillation type detector for backscattered electrons.
The YAG BSD detector has a high sensitivity and is specifically designed for high speed imaging.
It can be used in HV mode and in VP mode.
Function The YAG BSD detector uses yttrium aluminum garnet (YAG) as a fast scintillation material that is
mechanically and chemically resistant. Since it is a scintillation type detector, the YAG BSD detec-
tor does not have segments with changeable polarity and you can only use this detector for com-
position imaging. In comparison to segment based BSD detectors, the YAG BSD detector has a
faster response time.
Purpose The optional STEM (Scanning Transmission Electron Microscopy) detector is an electron detector
that can be used to detect transmitted and scattered electrons underneath an ultrathin specimen.
The STEM unit consists of a pre-aligned specimen holder and a Bright-field STEM detector, which
is a diode detector for electrons.
Position The pre-aligned holder fits directly on the specimen stage and carries the specimen.
In variable pressure mode, the STEM detector enables you to explore the nanostructure of non-
conducting specimens.
Info
Risk of malfunction: The diode segments are sensitive to the light that is used for illumination
in TV mode (infrared and white).
When you use a diode detector, always make sure that the TV illumination is switched off. If
the CCD Mode is set to Auto Detect, then the TV illumination is automatically switched off
when a diode detector is used.
Function To improve resolution, the STEM unit enables positioning of the thin specimen close to the objec-
tive lens. The collected signals are equivalent to bright field imaging.
The STEM detector is used in cases where the thickness of a specimen is similar to or less than the
dimensions of the interaction volume. The specimen must be mounted on a TEM grid with a thin
carbon-film support (approximately 10 nm thick). Electrons that pass through the target can then
be collected by the detector and used to form an image.
3.5.1.6 CL Detector
Purpose The Cathodoluminescence (CL) detector is an inclined detector that allows efficient visible or ultra-
violet light collection. The CL detector is ideal for use in geology, mineralogy, and materials sci-
ence applications where it can help in internal structural examination of rocks, ceramics, and
semiconductors.
Function The prerequisite for using this detector is that the specimen emits light when interacting with the
primary electron beam. Differences in crystal structure or the presence of impurities in a cathodo-
luminescent material result in variations in the energy gap between the filled valence bands and
the empty conduction bands, and consequently a change in the CL emission.
The light (photons) emitted by the specimen is collected by the CL detector and converted into a
signal for imaging.
The CL detector is fully integrated into the automatic brightness and contrast control of the micro-
scope and can be used simultaneously with any of the detectors without degrading their perfor-
mance.
The detector can be used during energy-dispersive X-ray spectrometer (EDX) measurements and
wavelength-dispersive spectrometer (WDS) measurements at any valid magnification.
Purpose The dual joystick is used for stage control and specimen navigation.
Position The dual joystick is placed on the microscope desk.
1 2 3
Function All axes are deflection-compensated. When the joystick is moved only slightly, the respective axis
moves slowly. Larger movements of the joystick result in a faster movement of the stage.
The X-, Y-, and Z-axes are magnification-compensated. When working at a low magnification, the
stage moves relatively fast. At higher magnifications the stage movement is slower. The stage is
moving with its maximum speed when viewing the specimen with the CCD (Charge Coupled De-
vice) camera.
The different axes can also be moved simultaneously.
Purpose The control panel allows direct access to the most frequently used functions. It integrates a full
sized keyboard, 11 turning knobs, and 8 push buttons.
Info
The control panel facilitates daily routine tasks. All the functions can be applied by using the
mouse and by macro execution also.
6 7 8 9 10 11
Fig. 37: Control panel
1 Stigmator X | Stigmator Y
Shapes the beam roundness by changing the stigmation deflectors.
2 Aperture X | Aperture Y
Adjusts the mid column shift and tilt deflectors for aligning the beam along the column
axis.
3 Scan Rotate
Rotates the scanning pattern 360° continuously.
This turning knob has a push button function to deactivate the scan rotate function and
reset the scan rotation to 0°.
4 Shift X | Shift Y
Shifts the scanned region of the specimen in the X and Y directions.
5 Brightness | Contrast
§ Brightness
Adjusts the image acquisition chain offset for the currently selected detector. Each
configured detector stores its own brightness.
§ Contrast
Adjusts the gain of the currently selected detector.
6 Magnification | Reduced
§ Magnification
Adjusts the magnification of the system.
§ Reduced
Changes the scan field to a reduced area. The size of the area is determined by the
current sub scan area settings.
7 Wobble
Sweeps the acceleration voltage. If the aperture is slightly misaligned, a shift in X and/or
Y direction can be observed.
8 Freeze
Stops the scan and grabs one complete frame at the current imaging conditions.
9 Exchange | Resume
§ Exchange
Starts the pre-defined macro for specimen exchange with the airlock.
§ Resume
Starts the pre-defined macro to finish specimen exchange with the airlock.
10 Camera
Switches to chamber view.
11 Focus | Scan Speed +/−
§ Focus
Changes the focal point of the column by adjusting the magnitude of the objective
lens.
§ Scan Speed +/−
Increases (+) or decreases (−) the scan speed by doubling or halving the beam dwell
time with each click step.
The SmartSEM software graphical user interface (GUI) allows you to monitor and operate most of
the active components of the microscope.
The following screenshot indicates the main elements of the SmartSEM user interface:
1 2 3 4 5
12 11 10 9 8 7 6
Fig. 38: Screen layout of the user interface
1 Title Bar
Displays the name of the user interface and the logged-on user.
2 Menu Bar
Enables you to access SmartSEM features via sub-menus.
3 AVI Toolbar
Contains the controls to set up, record, and playback video sequences of scanned images.
4 Toolbar
Provides quick access to SmartSEM tools.
5 Image Area with Data Zone
Displays image information and acquisition parameters from the microscope.
6 Thumbnails Panel
The following graphical control elements are used in the SmartSEM GUI.
The user access level defines which parameters are displayed for selection purposes, e.g. in the
status window or annotation parameter selection.
SmartSEM distinguishes different user access levels. Depending on the user access level, different
parameters are accessible. User profiles are defined by the administrator.
Access: Menu Bar > Tools > Administrator
Novice Only the items assigned to the novice category are accessible. These
include most frequently used parameters.
Expert Items assigned to the novice and expert category are accessible.
These include parameters useful for advanced operators.
Service All items are accessible, also including infrequently used items and cal-
ibrations.
Additional to the user access levels there are user privileges which are part of the user profile:
Change Image Di- Enables the user to change the location where all images are saved.
rectory
Change User Direc- Enables the user to change the location where all user specific param-
tory eters and configurations are saved.
The SmartSEM Program Suite comprises the EM server, which implements the internal communi-
cation between control software and microscope hardware, plus several programs and utilities.
The main purpose of the SmartSEM Program Suite is to access all necessary microscopy parame-
ters and software features to capture SEM data and optimize image acquisition.
Access: Windows start menu > SmartSEM
Program Description
Release Notes Contains an overview of all SmartSEM versions including new devel-
opments and specific details.
RemCon32 Serial interface for remote operation via RS232, e.g. for EDX
License: REMCON
SampleHolder- Enables you to inspect the dimensions of all possible specimen holders
Gallery as well as to set the dimensions of the custom specimen holders.
Enables you to activate the available specimen holders for SmartSEM.
SEM Drift Correc- Enables you to compensate for the drift of the specimen by using a
tion reference image and by controlling the beam shift.
License: DRIFT-CORR
Slideshow speed Enables you to adjust the slideshow speed for the Windows Photo
setting Viewer.
SmartSEM Admin- Enables you to manage user profiles and configure instruments.
istrator
SmartSEM User Enables you to record important information during individual work-
Accounting ing sessions, e.g. logon/logoff time, number of TIFF files exported etc.
Program Description
Smart Stage Map- Service activities, for ZEISS service representatives only
ping
4 Installation
Installation and commissioning are carried out by authorized ZEISS service representative. The in-
stallation requirements are to be observed and adhered to. After installation or retrofitting, thor-
oughly check that the Microscope System is in a safe operational state, making sure in particular
that all protective covers (e.g. protection against laser radiation) have been installed.
Info
Additional information and detailed descriptions are available in the further applicable docu-
ments, or ask your ZEISS Sales & Service Partner.
Info
Further information on the software and its operation is available in the software’s online help.
Before energizing the microscope make sure that the following safety warnings have been read
and fully understood by each person who is in the same room as the microscope at any time:
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4 During specimen exchange, keep the chamber door open as short as possible.
4 Do not inhale the air from within the specimen chamber.
4 Ensure that the area around the microscope is sufficiently ventilated.
4 If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
WARNING
Reaction products
Dangerous reaction products can be present in the specimen chamber during or after opera-
tion.
4 Ensure that there is an appropriate exhaust gas line to remove the waste gas of the pre-
vacuum pump and to transmit it to the outside.
4 Wear lint-free gloves when touching the inner parts of the specimen chamber or the speci-
men.
Before energizing the microscope make sure that the following safety warnings have been read
and fully understood by the person operating the microscope:
WARNING
Residual voltage at the mains plug
After unplugging the mains plug residual voltage is present at the pins of the plug which may
cause electrical shock.
4 After unplugging the mains plug wait at least 5 s before touching the pins of the mains
plug.
WARNING
High leakage current
High leakage currents are present in the microscope. Contact may cause burn or electrical
shock.
4 Ensure proper grounding. For more information, refer to the Installation Requirements
document.
4 Do not operate the microscope without the separate ground connection.
WARNING
Hazardous voltages inside the microscope
Contact may cause electrical shock or burn.
Hazardous voltages are present inside the microscope as long as the power cord is plugged in
or as long as the Main Switch of the EMO Circuit is in the ON position.
To completely cut off the microscope from any mains power:
4 with installed EMO Circuit set the Main Switch to the OFF position.
4 without EMO Circuit unplug the power cord by unplugging the CEE connector from the
CEE FEMALE RECEPTACLE of the mains supply.
WARNING
Restart after emergency off
If the reason for the emergency off is not eliminated, it may be dangerous to restart the micro-
scope.
4 If the microscope has been de-energized due to an emergency, ensure that the reason for
the emergency off does not exist anymore.
4 Make sure it is safe to energize the microscope.
Info
If the system was powered off for a longer time period, the ion getter pump might fail to start.
In this case, bake-out the gun head (supervisor user rights required) or contact your local ser-
vice center. Refer to Baking out the Gun Head [} 120].
To start the microscope, you need to use the buttons at the front of the plinth and follow a de-
fined procedure.
Detailed information about using the help system is given in the Online Help directly.
The following keys are shortcut keys and have special meaning.
Shortcut Function
<F3> Closes all windows except the Tool Bar and Status Bar
Shortcut Function
<F4> Step to next Magnification Table entry, or Undo Centre Feature Mag-
nification
<CTRL + ALT + F> Enable/Disable Mouse Following for Movable Crosshairs Marker
Shortcut Function
<CTRL + S> Calls the Export TIFF dialog to save the image
Info
Mobile phones in the microscope room can cause image quality infringements and in worst
case workflow interruptions.
This section describes basic procedures to obtain an image using the SE detector. To simplify the
procedure, the description uses the SEM Controls panel and status bar functions in the SmartSEM
software.
This procedure consists of the following steps:
1. Preparing the Specimen Holder [} 61]
2. Loading the Specimen Chamber [} 63]
3. Locating the Specimen [} 67]
4. Switching on the Gun | LaB₆ Filament [} 68]
5. Switching on the EHT | LaB₆ Filament [} 68]
6. Switching on the Gun and EHT | Tungsten Filament [} 72]
7. Acquiring an Image [} 73]
8. Optimizing the Image [} 76]
9. Saving the Image [} 79]
Stub pliers
Lint-free gloves
WARNING
Biological hazards
Biological substances may pose a threat to the health of humans and other living organisms.
4 Keep a logbook of the biological substances loaded into the microscope and show it to
the ZEISS service representatives before they perform any work on the microscope.
WARNING
Aggressive or toxic chemicals
Aggressive or toxic chemicals can cause chemical burns.
4 When handling aggressive or toxic chemicals, wear suitable protective clothing, gloves,
and eye/face protection.
4 Do not eat, drink, or smoke while handling toxic chemicals.
4 Refer to local safety regulations.
4 Read and follow the instructions in the material safety data sheet of the chemical. The ma-
terial safety data sheet can be obtained from the supplier of the chemicals.
NOTICE
Environmental risk due to disposal of aggressive or toxic chemicals
When disposing of aggressive or toxic chemicals, there is a threat of damage to the environ-
ment.
4 When disposing of waste that has been generated during a service operation (e.g. used ro-
tary pump oil), comply with all national and local safety and environmental protection reg-
ulations.
NOTICE
Contamination caused by fingerprints
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged pumping
times.
4 Always wear lint-free gloves when touching the specimen, specimen holder, or stage.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
2. Alternatively, to activate the chamberscope, click the TV icon in the Toolbar or press the
Camera button on the control panel.
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4 During specimen exchange, keep the chamber door open as short as possible.
4 Do not inhale the air from within the specimen chamber.
4 Ensure that the area around the microscope is sufficiently ventilated.
4 If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4 During specimen exchange, keep the chamber door open as short as possible.
4 Do not inhale the air from within the specimen chamber.
4 Ensure that the area around the microscope is sufficiently ventilated.
4 If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
CAUTION
Risk of electrical shock by stage on bias voltage
If you open the chamber door while the stage is still on bias voltage, then contact may cause
electrical shock.
4 Always switch off the beam deceleration before you open the chamber door.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
CAUTION
Closing the chamber door
Fingers can be pinched when closing the chamber door.
4 Ensure not to get your fingers caught in the chamber door gap.
NOTICE
Driving the stage
While the stage is driven manually, there is a risk of damaging the objective lens and/or the
specimen.
4 Ensure not to hit the objective lens while driving the stage.
4 Monitor the moving stage in TV mode.
4 To stop the moving stage immediately, press F12 or press the Break push button of the
dual joystick panel.
4 Manually lower the stage before you open the chamber door. Alternatively, activate the Z
move on Vent checkbox in the Stage tab of the SEM Controls panel.
NOTICE
Contamination caused by fingerprints
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged pumping
times.
4 Always wear lint-free gloves when touching the specimen, specimen holder, or stage.
3. If a specimen holder is mounted onto the specimen stage, remove it by sliding it out of the
dovetail rails.
4. Install the prepared specimen holder by sliding
it into the dovetail rails.
Make sure that the dovetail is placed in the cor-
rect orientation so that the flat side of the
dovetail of the specimen holder is flush with
the milled edge of the specimen stage.
5. Check the chamber view to ensure the specimen does not hit any components when it is
introduced into the specimen chamber.
6. Carefully close the chamber door.
à The specimen holder and the specimen inside the chamber are visible in the Image
Area.
NOTICE
Driving the stage
While the stage is driven manually, there is a risk of damaging the objective lens and/or the
specimen.
4 Ensure not to hit the objective lens while driving the stage.
4 Monitor the moving stage in TV mode.
4 To stop the moving stage immediately, press F12 or press the Break push button of the
dual joystick panel.
4 Manually lower the stage before you open the chamber door. Alternatively, activate the Z
move on Vent checkbox in the Stage tab of the SEM Controls panel.
Procedure 1. In the Stage Navigation Bar, select Stage Sideview from the upper drop-down list and
Stage Topview from the lower drop-down list.
INFO: To open the Stage Navigation Bar, navigate to View > Toolbars and activate
Stage Navigation Bar (for Widescreen users). Alternatively, you can access the Stage
Navigation Bar via Stage > Navigation.
2. Click Settings.
à The Stage Navigation Settings dialog is displayed.
3. In the Stage Navigation Settings dialog, click Show Holder Gallery.
à The Sample Holder Gallery dialog is displayed.
4. In the Sample Holder Gallery dialog, select the installed specimen holder.
5. Activate the Is Available checkbox.
6. Close the Sample Holder Gallery dialog.
7. Close the Stage Navigation Settings dialog.
8. In the Stage Topview section of the Stage Navigation Bar, spot the stub with the speci-
men you want to observe.
9. To drive the stub directly under the electron beam, double-click the stub.
Procedure 1. In the Stage Navigation Bar, drag the Zoom View slider to the right end, so that the
schematics are zoomed in.
2. In the SEM Controls panel, select the Detectors tab.
3. In the Detectors section, select USB TV1 from the Signal A drop-down list.
à The inside of the specimen chamber is visible in the Image Area.
4. Use the dual joystick to carefully move up the stage so that the stub you are using is in the
center of the upper schematic.
NOTICE Observe the camera image in order not to crash into the pole piece.
à INFO: After loading the specimen, there is the possibility to use the Sample Type Selec-
tion Function to automatically set a number of key parameters. Refer to Using the Sam-
ple Type Selection Function [} 92].
Info
If you operate the microscope with a LaB₆ filament, then you need to switch on or switch off
the gun and the EHT separately.
Prerequisite ü The chamber and the gun head have been evacuated.
Procedure 1. In the right part of the Status Bar, verify whether the gun is switched on or off.
à If or is displayed, the gun is already switched on and you can skip the fol-
lowing steps.
à If is displayed, the gun is switched off.
2. In the SEM Controls panel, select the Vacuum tab.
3. Verify that the EHT Vac ready readout is EHT Vac ready = Yes.
If not, the correct vacuum is not achieved. Check if the Pump procedure has been com-
pleted.
4. In the right part of the Status Bar, click .
à The pop-up menu for vacuum, gun and EHT activation is displayed.
5. Click Beam On.
à The gun runs up.
When you switch on the EHT, the gun starts emitting electrons.
Info
If you operate the microscope with a LaB₆ filament, then you need to switch on or switch off
the gun and the EHT separately.
Prerequisite ü The chamber and the gun head have been evacuated.
ü The gun has been switched on.
Procedure 1. In the SEM Controls panel, select the Vacuum tab.
2. Check that Vac Status = Ready is displayed.
3. In the SEM Controls panel, select the Gun tab.
4. Double-click the Fil I Target readout.
à The Fil I Target window is displayed.
5. In the input field, enter the required filament current in mA and click OK.
INFO: For a LaB₆ filament, you may start with a filament current of approximately
1800 mA.
6. In the SEM Controls panel, select the Gun tab.
7. Double-click the EHT Target readout.
à The EHT Target window is displayed.
8. In the input field, enter the required acceleration voltage in kV and click OK.
9. In the right part of the Status Bar, click .
à The pop-up menu for vacuum, gun and EHT activation is displayed.
10. Click EHT On.
à The EHT runs up to the set voltage.
à In the right part of the Status Bar, the vacuum, gun and EHT status buttons merge to
.
15. Adjust the Filament I target by increasing or decreasing the value using the fine adjust op-
tions, i.e. small arrows: Increase the Filament I target in small steps and check the emission
image until close to the second peak. The signal increases and the emission image becomes
brighter. At first five regions can be seen on the emission image, i.e. the electron emission
from all the five crystal planes of the LaB₆ (pyramid shape filament). As the current is in-
creased the four symmetrical off axis emission areas collapse into a central spot, i.e. the
emission will be mainly from the tip of the filament and not from the other four planes of
the LaB₆ filament. The best performance of the LaB₆ can be achieved when the tip is fully
saturated and at the second peak. For a new LaB₆ filament the filament current value will
be between 1.95 A to 2.0 A. The Filament I target should never be higher than 2.050 A.
a LaB6 is saturated at first peak
The spot is very well defined and bright but
there is clearly some emission from the side
planes. At first peak, does not provide the
means for obtaining the optimal resolution
and the beam intensity decreases quite dra-
matically, as the filament ages.
16. Adjust the Gun Shift and Gun Tilt: Center the emission image, obtain a symmetrical shape
with the brightest point at the middle of the crosshairs (shift for adjusting and checking the
edges and tilt for moving the brightest point of the emission image to the middle.
17. Set the WD=8.5 mm.
18. On the Apertures tab, activate Focus Wobble and Wobble Fast. Find a feature which is
outstanding (preferably round shape). Change the wobbling amplitude to appropriate val-
ues, adjust the contrast, and set to an appropriate magnification.
19. Adjust the Mid-column aperture by fine-adjusting knobs X and Y. Work with both X and
Y knobs, finding the crossovers, until the image is just wobbling in and out of the screen.
When finished with the adjustments, deactivate Focus Wobble.
20. Do a Hysteresis correction by using Shift + F2 keys on the keyboard.
21. Adjust the Mag and Focus. Increase the magnification to the desired values and adjust the
focus (in several steps).
22. Adjust Stigmation in both X and Y directions in several steps.
23. Re-adjust the Focus, when necessary.
à The beam is now aligned and further adjustments of the mid-column apertures or the
fine-adjustment knobs is no longer required (use only the Auto Aperture Align from
now on).
24. Start imaging different specimens and if necessary change/adjust the EHT, I Probe, WD
and repeat with the Mag, Focus, and Stigmation adjustments.
25. On the Aperture Align window, click on Auto Aperture Align after changing parameters
such as EHT, I Probe, WD.
When you switch on the gun and the EHT, the gun starts emitting electrons.
Info
If you operate the microscope with a tungsten filament, then the EHT and the gun are always
switched on or switched off together. You cannot separately switch on or off the EHT or the
gun.
Info
Tungsten filaments have only a limited lifetime of 100–300 operating hours. If you operate the
microscope with a tungsten filament, then do the following:
4 If the microscope is not in operation, then always switch off the EHT and gun to preserve
the filament.
4 Enable Auto Saturation at shutdown.
Prerequisite ü The chamber and the gun head have been evacuated.
Procedure 1. In the SEM Controls panel, select the Vacuum tab.
2. Check that Vac Status = Ready is displayed.
3. In the SEM Controls panel, select the Gun tab.
4. Double-click the Fil I Target readout.
à The Fil I Target window is displayed.
5. In the input field, enter the required filament current in mA and click OK.
INFO: Depending on the exact type of tungsten filament, you may start with the following
values for the filament current:
For standard tungsten filaments (Agar A054), use 2500 mA.
For longlife tungsten filaments (Agar A054L), use 3200 mA.
6. In the SEM Controls panel, select the Gun tab.
7. Double-click the EHT Target readout.
à The EHT Target window is displayed.
8. In the input field, enter the required acceleration voltage in kV and click OK.
9. In the right part of the Status Bar, click .
à The pop-up menu for vacuum, gun and EHT activation is displayed.
10. Click Beam On.
à The gun is switched on and the EHT runs up to the set voltage.
à In the right part of the Status Bar, the vacuum, gun and EHT status buttons merge to
.
Info
The following procedure describes the best way to quickly obtain an image without the con-
trol panel. You can also use the control panel to adjust aperture alignment, magnification/fo-
cus and brightness/contrast.
The following procedure describes how to acquire an image in high vacuum (HV) mode.
For further information on pressure and operating modes, refer to Selecting the Optibeam Opera-
tion Mode [} 81].
Info
You can also use the Automated or Semi-automated Functions to help you with these tasks.
5. In the Apertures tab of the SEM Controls panel, select the 20.00 µm from the Aperture
Size drop-down list.
à The 20 µm aperture is set.
à If the image is noisy, then change the scan speed to 4 or 5, but no higher than 6, as it
slows down the specimen navigation process.
Procedure 1. To move the specimen stage to a suitable location for imaging, use the X and Y controls on
the joystick.
2. In the Toolbar, select the MAGWD icon.
à The Status Bar displays the values for magnification and focus.
3. In the Status Bar, click .
à The Mag window is displayed.
4. In the Mag input field, enter 500.
5. Click OK.
à The magnification is set to Mag = 500 x.
3. To adjust the magnification, hold down the left mouse button and drag the mouse within
the Image Area in left/right direction.
à The current magnification is indicated in the Status Bar.
4. To adjust the focus, change the working distance. Hold down the mouse wheel and drag
the mouse within the Image Area in left/right direction.
à Alternatively, you can press <Ctrl+F> to use the fine autofocus or you can press <Ctrl
+Shift+F> to use the coarse autofocus.
à The current working distance is indicated in the Status Bar.
5. Adjust contrast and brightness again.
Once you have generated an initial image, you can adjust various parameters to optimize the im-
age.
Info
The following procedure describes how to manually improve the image quality. You can also
use the Auto Aperture Alignment Function.
Info
The following procedure describes the best way to quickly optimize the image without the
control panel. You can also use the control panel to adjust aperture alignment, magnification/
focus and brightness/contrast.
Procedure 1. Step by step, raise the magnification to about twice the desired final magnification (e.g.
50,000 x) and focus in between.
Use the Coarse mode or the Fine mode of adjustment, as appropriate. To toggle between
Fine and Coarse mode, in the Status Bar, click or .
2. To adjust the magnification and the focus, hold down the left mouse button or the mouse
wheel, respectively, and drag the mouse within the Image Area.
If you want to move the field of view at high magnifications, use the Beam Shift function instead
of moving the stage.
Procedure 1. In the Panel Configuration Bar, double-click Beam Shift.
2. To shift the beam, in the Beam Shift navigation
box, use the scroll bars or the red marker.
Prerequisite ü Adjusting the size and position of the small frame (reduced raster) requires the license RE-
DUCED.
Procedure 1. In the Toolbar, click the REDUCE icon.
à A small scan frame is displayed. This frame defines the specimen area to be scanned by
the electron beam.
à The image outside the scan frame is frozen.
2. To change the position of the scan frame, click on the green border line and use the mouse
to drag and drop the frame.
3. To change the size of the scan frame, click on the small blue squares on the green border
line and drag them to the desired size.
4. Focus the image in the reduced raster.
Info
If the VP 100 μm aperture is fitted under the objective lens, focus wobble is not available.
Procedure 1. In the Toolbar, from the Faster/Slower drop-down list, select Scan Speed = 7.
Alternatively, in the SEM Controls panel, select the Scanning tab, and from the Scan Speed
drop-down list, select Scan Speed = 7.
à The scan speed is set to Scan Speed = 7.
2. Bring the image into focus.
6. To deactivate the reduced raster, in the Toolbar, click the REDUCE icon.
Procedure 1. Once a steady contrast and brightness level is reached, click the PHOTO icon in the Tool-
bar.
INFO: This runs a macro that automatically changes the scan speed to 8, the noise reduc-
tion to line integration with 4 lines and freezes the image at the end of the frame.
à A red dot at the right bottom of the image area indicates that the image is frozen.
2. From the Menu Bar, select File > Save Image.
à The Export TIFF dialog is displayed.
Automatic gun align is used to automatically align the shift and/or the tilt of the beam, to the cen-
ter of the column.
Two automatic gun align functions are available:
§ Standard Align uses the normal imaging mode to align the gun.
§ Advanced Align uses the emission image to align the gun.
Procedure 1. From the Panel Configuration Bar, select Gun Alignment.
à The Gun Alignment dialog is displayed.
2. To automatically align the gun with the help of the emission image, click Advanced Align.
3. To automatically align the gun with the help of the normal imaging mode, click Standard
Align.
Auto aperture align is used to improve the quality of the images. When the aperture align is se-
lected the software automatically aligns the beam with respect to the EasyVP aperture.
Info
The maximum achievable probe current depends on the currently selected EHT and the in-
stalled aperture configuration.
Measuring the probe current using the Faraday cup ensures that the current displayed in the soft-
ware equals the incident probe current. The Faraday cup consists of a strongly absorbing material
with a cavity covered by a small aperture. If the beam is focused in this cavity, no secondary elec-
trons and no backscattered electrons leave the Faraday cup.
EVO can be operated in different Optibeam operation modes, depending on the type of applica-
tion. Based on the requirements for probe current, working distance, etc., Optibeam determines
the optimum lens settings to achieve the best performance from the column.
The following modes are accessible in HV mode (when no aperture is fitted) or when the EasyVP
aperture is fitted:
§ Analysis mode: Specimen remains in focus for any change in probe current.
§ Field mode: A large field of view for navigation with a long depth of field.
§ Resolution mode: Smallest probe diameter for a chosen probe current at any working dis-
tance or acceleration voltage.
§ Depth mode: Largest depth of field or a chosen probe current at any working distance or ac-
celeration voltage.
§ Fisheye mode: Extreme field of view for navigation and a very large depth of field.
If the fixed aperture is fitted, some Optibeam modes may not be available. The fixed apertures are
apertures, which are screwed into the column of the microscope and which you need to distin-
guish from the click-stop apertures.
Prerequisite ü Fisheye mode requires the optional SmartSEM software license FISHEYE.
Procedure 1. In the SEM Controls panel, select the Apertures tab.
2. Use the drop-down list and the checkboxes to
select the operation mode.
You need to select an appropriate detector depending on the application and the pressure mode.
In addition to the standard SE detector, several optional detectors are available.
SE HV
C2D VP
C2DX VP, EP
BSD HV, VP
STEM HV, VP
CL HV, VP
The variable pressure mode can either be normal VP or extended EP. This depends on the fac-
tory configuration. If EP is enabled the VP is replaced by EP.
For information on special set-up procedures for the detectors, refer to:
§ Setting up the SE Detector [} 83]
§ Setting up the C2D Detector [} 83]
§ Setting up the C2DX Detector [} 84]
§ Setting up the BSD Detector [} 85]
§ Setting up the YAG BSD Detector [} 87]
Procedure 1. Select the Detectors tab of the SEM Controls panel.
2. Select the detector from the Signal A drop-
down list.
Procedure 1. In the Detectors tab of the SEM Controls panel, select SE2 from the Signal A drop-down
list.
2. Use the Collector Bias scroll bar to adjust the
collector bias.
INFO: The default value is 300 V.
Info
The C2D detector requires VP pressures of at least 20 Pa to acquire a good signal.
A good starting point is EHT = 20 kV, I Probe = 500 pA, Chamber pressure = 30 Pa,
WD = 10 mm, and scan speed = 5.
NOTICE
Inserting the C2DX detector
When you manually insert the detector, there is a risk to damage the C2DX detector.
4 Use the chamberscope image to observe if there is enough space between the objective
lens and the specimen.
4 If there is not enough space between the objective lens and the specimen, then lower the
stage position before you insert the detector.
4 Insert the C2DX detector carefully and observe the moving C2DX detector via the cham-
berscope.
4 If the BSD detector is mounted on the lens, then ensure that the C2DX can be inserted
without touching the BSD. This is best done with the chamber vented so that it can be ob-
served directly.
Info
The C2DX detector requires VP pressures of at least 20 Pa to acquire a good signal.
A good starting point is EHT = 20 kV, I Probe = 500 pA, Chamber pressure = 30 Pa,
WD = 8.5 mm, and scan speed = 5.
8. Click a segment symbol to toggle its status between on (white), inverted (black), and off
(gray).
9. To confirm the settings, click Apply.
10. To select compositional mode, click BSD: COMPO.
11. To select topography mode, click BSD: TOPO.
INFO: The topography mode has a default setting that can be changed by the user. To
change the default setting, select the desired segment combination and click BSD: Set
Topo.
12. Activate BSD Auto Range. Optimize the image by adjusting brightness and contrast.
INFO: The BSD detector has four amplification ranges (BSD Gain: Low, Medium, High, Very
High). Activating BSD Auto Range automatically switches the ranges, depending on the
contrast setting of the detector. Alternatively, the BSD Gain can be manually selected from
the drop-down list when BSD Auto Range is not activated. The suitable BSD Gain depends
on the signal strength (acceleration voltage, probe current, WD and sample material com-
position).
INFO: Risk of malfunction: the diode segments are sensitive to the light used for illumina-
tion in TV mode (infrared and white). When using a diode detector, always ensure that the
TV illumination is switched off. If the CCD Mode is set to Auto Detect, the TV illumination
is automatically switched off when a diode detector is used.
NOTICE
Inserting the YAG BSD detector
When you manually insert the detector, there is a risk to damage the YAG BSD detector.
4 Use the chamberscope image to observe if there is enough space between the objective
lens and the specimen.
4 If there is not enough space between the objective lens and the specimen, then lower the
stage position before you insert the detector.
4 Insert the YAG BSD detector carefully and observe the moving YAG BSD detector via the
chamberscope.
Beam deceleration is mostly used when imaging specimens with BSE detectors. By means of stage
biasing, a negative potential is applied to the specimen. This reduces the electron beam’s landing
energy, which enables you to obtain more specimen surface detail, especially at low magnifica-
tions. At high magnifications, you can obtain better image resolutions. The effectiveness strongly
depends on the specimen type. For metallic specimens, image improvements can easily be
achieved.
Beam deceleration can only be applied in HV mode. A potential in the range of 0 to −5 kV is used.
If the EHT voltage is set to 5 kV and a stage bias potential of −4 kV is applied, the landing energy
of the primary electrons is reduced to 1 kV.
The beam deceleration kit is mounted through a plate on the chamber door. A specially designed
specimen holder is generally used. This is to isolate the specimen holder, i.e. the bias is only ap-
plied to the specimen.
Fig. 41: Beam deceleration kit (left) and 9-stub beam deceleration specimen holder (middle and right)
CAUTION
Risk of electrical shock by stage on bias voltage
If you open the chamber door while the stage is still on bias voltage, then contact may cause
electrical shock.
4 Always switch off the beam deceleration before you open the chamber door.
NOTICE
Touch alarm disabled
When beam deceleration or stage bias is activated, the stage touch alarm is automatically de-
activated. A warning is displayed to inform you that the touch alarm is not enabled. This
means that any stage collisions will not be detected.
4 Move the stage carefully.
Prerequisite ü The stage has been initialized before fitting the beam deceleration specimen holder.
ü HV mode is active.
ü The beam deceleration specimen holder is fitted.
Procedure 1. In the Stage Navigation Bar, click Settings.
INFO: To open the Stage Navigation Bar, navigate to View > Toolbars and activate
Stage Navigation Bar (for Widescreen users). Alternatively, you can access the Stage
Navigation Bar via Stage > Navigation.
à The Stage Navigation Settings dialog is displayed.
2. Activate the Safe Navigation checkbox.
3. In the Stage Navigation Settings dialog, click Sample Holder Gallery.
11. Use the Beam deceleration volts scroll bar to adjust the stage bias potential until you are
satisfied with the results.
Stage biasing can also be used for applying low voltages in the range of +20 V to −20 V to speci-
mens. This low-voltage specimen biasing is normally used in combination with the SE detector
and enables you to achieve an improvement in the image contrast.
Low-voltage specimen biasing can be applied under HV and VP conditions.
Info
Low-voltage biasing works with any standard, conducting, dovetail mounted specimen holder.
It does not work with lens-mounted specimen holders and other non-dovetail mountings.
If the Beam deceleration carousel checkbox is fitted and activated in SmartSEM, only high-
voltage beam deceleration can be used and low-voltage biasing is disabled.
CAUTION
Risk of electrical shock by stage on bias voltage
If you open the chamber door while the stage is still on bias voltage, then contact may cause
electrical shock.
4 Always switch off the beam deceleration before you open the chamber door.
NOTICE
Touch alarm disabled
When beam deceleration or stage bias is activated, the stage touch alarm is automatically de-
activated. A warning is displayed to inform you that the touch alarm is not enabled. This
means that any stage collisions will not be detected.
4 Move the stage carefully.
11. Use the Stage bias Low Volts scroll bar to adjust the voltage until you are satisfied with
the result.
Filament § Long Fil. Life § Fil I set to first § Long Fil. Life
activated peak for qualita- deactivated
tive analysis
Pressure-limiting – – –
aperture
Scanning parameters Cycle time = 20 s to Cycle time ≥ 20 s for Cycle time ≥ 1.3 min
reduce noise X-ray mapping to reduce noise
The sample type selection function enables the user to obtain an image of any specimen quickly,
i.e. a reference image, without putting any effort in selecting the operating parameters (vacuum
mode, acceleration voltage, probe current, and detector). The quality of the initial image can sub-
sequently be improved, if necessary, by modifying the imaging parameters.
Procedure 1. From the Panel Configuration Bar, select Sample Type Selection.
à The Sample Type Selection menu is dis-
played.
WD 5 mm 8.5 mm
To use the optional VP mode or EasyVP mode, you need to install the VP or the EasyVP pressure-
limiting aperture under the objective lens.
Additionally, you need to select the mid-column aperture according to the pressure mode and the
pressure-limiting aperture and adjust some settings in SmartSEM.
Prerequisite ü The microscope is equipped with the VP mode or EasyVP mode option.
Procedure 1. In the SEM Controls panel, select the Vacuum tab.
2. To vent the specimen chamber, click Vent.
3. If a lens-mounted BSD is fitted, remove the BSD
and place it in the parked position.
6. Use the VP aperture tool to either install the VP aperture or the EasyVP aperture.
INFO: This should be finger-tight only.
7. Refit the lens-installed BSD detector.
8. In the SEM Controls panel, select the Apertures tab.
9. Click Select Aperture and select the appropri-
ate aperture.
10. To evacuate the specimen chamber, click Pump in the Vacuum tab.
11. Turn the knob of the mid-column aperture
changer, to select the appropriate aperture:
Select the 20 μm aperture (position 1) for
EasyVP mode.
Select the 750 μm aperture (position 3) for VP
mode.
12. In the Apertures tab, match the aperture size from the Aperture Size drop-down list to
the selected aperture.
13. In the Vacuum tab, click Go To VP.
14. To adjust the working pressure, use the VP Target scroll bar.
5.7.2.2 Aligning the Gun and Mid-column Aperture for EasyVP Mode
To achieve an optimal alignment of the beam through the 20 μm mid-column aperture, you need
to perform an alignment procedure.
Info
You can refer to Using the Automatic Gun Alignment Functions [} 79] and Using the Auto
Aperture Alignment Function [} 80] to help you with these tasks.
Info
Once the alignment is optimized, do not change the position of the X and Y micrometer
gauges on the mid-column aperture changer.
6. Click Emission.
7. Adjust the Emission Zoom to 35.
8. To zero the aperture alignment, select Aperture Align from the Panel Configuration
Bar.
à The Aperture Align window is displayed.
9. To set the Aperture Align value to zero, click the 0 button.
10. Make sure that the filament is saturated.
11. Set Brightness and Contrast to 50 %.
12. Adjust Contrast to make the outer region of the emission image visible around the brighter,
central part of the emission image.
13. Adjust the Brightness and Emission Zoom to improve the visibility.
14. Adjust the emission image with Gun Shift and Gun Tilt so that the bright part is at the cen-
ter of the image.
Observation of hydrated (moist) specimen at low kV (start at 1 kV) slows down the dehydration of
moist specimens that leads to structural collapse.
Optimum imaging of moist specimens, requires an EVO microscope with the options Peltier cool-
stage and Extended Pressure (EP) mode. This setup enables you to study hydrated specimen in
their native state with little or no loss of water. For EP mode, the 100 μm EP aperture is installed
as well as the 500 or 1000 μm beamsleeve.
The imaging in EP mode consists of the following steps:
§ Installing or Deinstalling the EP and Beemsleeve Apertures [} 98]
§ Mounting the Peltier Coolstage [} 101]
§ Activating the Peltier Coolstage [} 105]
§ Purging the Chamber [} 106]
§ Setting Initial Operating Parameters [} 106]
§ Adjusting Operating Parameters [} 107]
Recommended The following parameters are recommended for hydrated specimens. The values given are only
Parameters suggestions and the operator may find that different values will give better information for the
types of specimens that are being investigated.
WD 8.5 mm 5 mm
To use the optional EP mode, you need to install the EP and beamsleeve pressure-limiting aper-
tures under the objective lens.
Additionally, you need to select the 750 μm mid-column aperture and adjust some settings in
SmartSEM.
Info
The 500 µm beamsleeve allows EP pressures up to 3000 Pa, but only 0.5 mm maximum field
of view.
The 1000 µm beamsleeve allows EP pressures up to 1000 Pa, and 1 mm maximum field of
view.
You need to activate the stage and the EP option in the software and set the working pressure
before you can use the EP mode.
NOTICE
Rotating the Peltier coolstage
The Peltier coolstage can be damaged when stage rotation is used.
4 Only move the stage in X and Y direction.
2. To evacuate the system, click Pump in the Vacuum tab of the SEM Controls panel.
3. Wait until Vac Status = Ready and EHT Vac ready = Yes are displayed.
This can take some time.
4. From the Menu Bar, select Stage > Navigation.
à The Stage Navigation panel is displayed.
5. Click Settings.
à The Stage Navigation Settings dialog is displayed.
6. In the Stage Navigation Settings dialog, click Show Holder Gallery.
à The Sample Holder Gallery dialog is displayed.
7. In the Sample Holder Gallery dialog, click CoolStage MK3.
8. Activate the Is Available checkbox.
13. Select the USB connection and activate the Humidity Option checkbox.
14. In the Vacuum tab of the SEM Controls panel, select Go to EP.
15. Use the EP Target scroll bar to adjust the working pressure to 10 Pa.
16. Adjust the other working parameters. For information on working parameters for different
specimens, refer to Hydrated Specimen [} 97].
You need to purge the chamber to remove air from the water bottle and to fill the chamber with
pure water vapor.
Procedure 1. From the Panel Configuration Bar, select Extended Pressure.
à The Extended Pressure window is displayed.
2. Activate the Peltier checkbox.
3. Click Purge Settings.
à The Purge Control window is displayed.
4. Check that EP Gas = Air is selected.
5. If the water kit has recently been fitted, filled with water, or has not been used for a while,
set Purge cycles to 10.
6. If the system has recently been used in the wet mode, set Purge cycles to 3.
7. Set Purge Max to 1000 Pa.
8. Set Purge Min to 100 Pa.
9. Click EP Gas = Air to toggle the vacuum status to EP Gas = Water vapour.
10. To start purging, click Manual Purge H2O.
à The pressure in the chamber varies between the set maximum and minimum levels. Bub-
bles may be seen in the water bottle. All air is eventually removed from the water bottle.
à The process can take a while to be completed. The progress is displayed in the status bar.
Procedure 1. In the Extended Pressure window, set the Peltier Target to 1 °C.
2. Adjust Humidity Target to a suitable value.
3. Adjust the position of the green cross on the phase diagram to change the environment of
the specimen between water vapor, water, or ice.
4. Switch on the beam.
5. Use the stage navigation to position the specimen under the beam.
INFO: Use the Stage Navigation Bar or the chamberscope to monitor stage movement.
Procedure 1. Increase the pressure to 550 Pa and keep the temperature constant at 1 °C.
2. Increase the pressure in steps of 20 Pa and review the image.
à The water droplets start to appear around 600 Pa.
3. Once the water droplets are clearly visible on the stub, make a note of the pressure.
INFO: The rate of water condensation from the chamber on the stub and the evaporation
of the droplets from the stub to the chamber should be at equilibrium.
4. In the Purge Control window, set Purge Min to the pressure value obtained in the previ-
ous step.
5. Set Purge Max to a pressure value that is around 100 Pa higher than the Purge Min
value.
6. Adjust the Purge Cycle to 3.
7. Vent the chamber without changing any of the parameters.
8. Remove the plain stub and dry the area underneath the stub on the Peltier coolstage.
9. Carefully install the specimen on a new stub, place and tighten the stub on the Peltier cool-
stage.
10. Use a pipette for putting a few droplets of distilled water on the Peltier coolstage holder.
INFO: Do not put any water droplets on the stub or directly on the specimen as the water
droplets can cover the specimen surface and prevent viewing the areas of interest.
11. Move the Peltier coolstage slightly to the right or left of the chamber before closing the
chamber door.
This prevents any droplets splattering up on the C2D, BSD, or the objective lens area during
the pump down process.
12. Close the chamber door and pump down the system.
13. Switch on the beam.
14. Use the stage navigation to position the specimen under the beam.
INFO: Use the Stage Navigation Bar or the chamberscope to monitor the stage move-
ment.
15. If necessary, increase the pressure, but in very small steps.
INFO: If you increase the pressure in large steps, water can accumulate on the specimen
and prevent imaging.
16. If necessary, adjust the humidity level to achieve 100 % humidity.
INFO: This changes both the Peltier temperature and the chamber pressure.
17. If necessary, increase the EHT.
INFO: This increases the signal strength but can damage the specimen.
18. If necessary, reduce the spot size.
INFO: This provides higher resolution but means loss of signal.
19. If necessary, reduce the WD.
INFO: This increases the signal strength but reduces the field of view, which means that a
montage of specimen images may be required.
To obtain the maximum lifetime of a LaB₆ emitter, switch off the EHT when the microscope is not
in use and switch the gun to STANDBY mode when the microscope is not used for longer peri-
ods of time. In STANDBY mode, the filament continues to be heated, and the vacuum in the
electron column and in the specimen chamber are maintained. It is recommended that you do not
turn off the gun during its service lifetime.
The STANDBY mode is also the recommended mode to store the microscope. In this case, acti-
vate the Partial Vent on Standby checkbox in the Vacuum tab of the SEM Controls panel. This
can help to prevent oil vapors from penetrating into the specimen chamber during the period of
storage.
To further extend the service life time, you can activate the Long Fil. Life checkbox in the Gun
tab of the SEM Controls panel. This reduces the filament current to “first peak” conditions,
which provides satisfactory operating conditions for general microscopy. If optimum performance
is required, the Long Fil. Life checkbox should be deactivated.
Info
If you operate the microscope with a tungsten filament, then the EHT and the gun are always
switched on or switched off together. You cannot separately switch on or off the EHT or the
gun.
Info
The tungsten filaments have only a limited lifetime of 100–300 operating hours. If you operate
the microscope with a tungsten filament, then do the following:
4 If the microscope is not in operation, then always switch off the EHT and gun to preserve
the filament.
7. Only when interrupting work for longer periods: Press the OFF button.
à The microscope switches to OFF mode.
à The electron column is partially vented.
à Computer, electronic components, and vacuum system are switched off.
à A 24 V auxiliary voltage is still present to
restart the microscope.
Procedure 1. From the Menu Bar, select File > Log Off.
à A system message is displayed.
2. Click Yes.
3. Close the EM Server.
à A system message is displayed.
4. Click Yes.
INFO: The EM server remains active.
WARNING
Residual voltage at the mains plug
After unplugging the mains plug residual voltage is present at the pins of the plug which may
cause electrical shock.
4 After unplugging the mains plug wait at least 5 s before touching the pins of the mains
plug.
Prerequisite ü The microscope is in Standby mode or in OFF mode, refer to Finishing the Work Session | LaB₆
Filament [} 108] or Finishing the Work Session | Tungsten Filament [} 108].
Procedure 1. If the microscope is not yet in OFF mode, press the Off button at the front of the plinth.
à When all subsystems are fully deactivated,
the Off button lights up red permanently.
2. Close and lock the main shut-off valves at the installation site.
3. If the EMO circuit is not installed, unplug the
power cord by unplugging the CEE connector
from the CEE FEMALE RECEPTACLE of the
mains supply.
9 The microscope is completely cut off from the electrical main supply.
WARNING
Residual voltage at the mains plug
After unplugging the mains plug residual voltage is present at the pins of the plug which may
cause electrical shock.
4 After unplugging the mains plug wait at least 5 s before touching the pins of the mains
plug.
NOTICE
Components in the high voltage circuitry
When the microscope, especially the gun, is fully on, an abrupt shutdown of all electrical sup-
plies may damage some components in the high voltage circuitry, mainly the cathode.
4 Use the emergency off only in an emergency situation with personnel injury.
Info
Additional information and detailed descriptions are available in the further applicable docu-
ments, or ask your ZEISS Sales & Service Partner.
DANGER
Electric shock due to live parts
When the Microscope System is still switched on, coming in contact with live parts can lead to
electric shock or burn.
4 Switch off Microscope System prior to opening or cleaning.
4 Disconnect live parts from the power supply.
Info
Keep track of maintenance work and contact the ZEISS service representative in time.
A list of ZEISS locations and authorized service partners can be found at:
http://www.zeiss.com/microscopy
The change of consumables and chemicals has to be performed by a ZEISS service representative
at mandatory intervals.
The times scheduled are designed for the maximum equipment performance level (i.e. 24 h per
day of permanent operation).
Interval Component/Part
As required Filament
NOTICE
Functional impairment due to dirt and moisture
Dirt, dust and moisture can impair the Microscope System’s functionality and can cause short-
circuits.
4 The ventilation slots must be unobstructed at all times.
4 Perform regular maintenance and cleaning according to the instructions in this document
and according to the instructions in the applicable documents.
4 Make sure that no cleaning liquid or moisture gets inside the Microscope System.
4 In case of damage, the affected parts of the Microscope System must be taken out of op-
eration.
If required, use a moistened lint-free cloth to clean the outside of the microscope.
7 Troubleshooting
The following table provides information about solving common problems.
Info
If you cannot solve the problem or if you are unsure about a certain technical difficulty, con-
tact your local ZEISS service representative.
Chamber does not Isolation mounts Refer to Adjusting the Isolation Mounts
move freely have lost air [} 119]
Gun is switched off Gun has been Refer to Baking out the Gun Head [} 120]
automatically switched off for
safety reasons since
gun vacuum is worse
than the pressure
threshold
Image is bad at low Working distance is Reduce the working distance to a maximum
EHT (e.g. 1 kV) too long of 7 mm
Microscope is dead Circuit breaker is Refer to Checking the Position of the Circuit
tripped (lower posi- Breakers [} 147]
tion)
After a power fail- Stage needs to be Refer to Initializing the Stage [} 115]
ure, the stored stage initialized
position cannot be
approached correctly
SE image is noisy Scintillator is used up Contact your local ZEISS service representa-
tive to have the scintillator replaced
Stage does not move Stage needs to be Refer to Initializing the Stage [} 115]
initialized
Vac ready = OK is System vacuum is Check the chamber door seal for cleanliness.
not displayed after bad due to a vac- If required, refer to Replacing the Chamber
specimen exchange uum leak at the Door Seal [} 117]
chamber door
Vac ready = OK is Gas ballast at rotary Deactivate gas ballast at the pre-vacuum
displayed very late pump or scroll pump pump
after specimen ex- is activated
change
If you use a LaB₆ fila- The pumping capac- Refer to Baking out the Gun Head [} 120]
ment and gun vac- ity of the ion getter
uum is worse than pump decreases in
1 × 10⁻⁶ mbar the course of time,
thus deteriorating
the gun vacuum
Tab. 6: Troubleshooting
7.1 Chamber
If a stored stage position cannot be approached or if the stage does not move or does not move
accurately, the stage needs to be initialized.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
Prerequisite ü The specimen chamber has been evacuated, refer to Loading the Specimen Chamber [} 63].
ü Requires the Stage Initialise privilege.
ü If there are any large specimens inside the chamber, remove them before initializing.
Procedure 1. From the Menu Bar, select Stage > Stage Initialise.
à The Initialise Stage window is displayed.
2. Confirm via Yes.
à The stage initialization process takes a few minutes.
à INFO: If initialization of the stage does not solve the stage problem, contact your local
ZEISS service representative.
You can configure the position to which the stage drives after the initialization procedure. Other-
wise, the stage drives to the center position.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
In TV mode (chamberscope), the dual joystick and stage may appear to move in opposite direc-
tions. This is because the selected CCD camera is installed at a certain angle relative to the stage.
Thus, the camera shows a side-inverted view. To remedy this, you need to change the joystick TV
angle setting in the software.
Info
If you are working with two CCD cameras: The joystick TV angle can only be set for one CCD
camera. When selecting the other CCD camera, you have to change the setting.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
To prevent damage, a touch alarm is integrated in the microscope. If the specimen or the speci-
men holder touches the chamber walls, the detectors, or the objective lens, the stage is stopped
immediately. An audible warning sounds and an on-screen message is displayed.
Possible reasons for replacing the chamber door seal are the following:
§ Chamber door does not close tightly
§ Bad chamber vacuum
This procedure consists of the following steps:
1. Venting the Specimen Chamber [} 64]
2. Replacing the O-ring [} 118]
3. Evacuating the Specimen Chamber [} 66]
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4 During specimen exchange, keep the chamber door open as short as possible.
4 Do not inhale the air from within the specimen chamber.
4 Ensure that the area around the microscope is sufficiently ventilated.
4 If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
CAUTION
Moving the specimen stage
Fingers can be trapped by the moving specimen stage.
4 Always close the chamber door before moving the specimen stage.
4 To remove parts fallen into or near to the stage use a tool (e.g. tweezers) instead of your
fingers.
CAUTION
Closing the chamber door
Fingers can be pinched when closing the chamber door.
4 Ensure not to get your fingers caught in the chamber door gap.
NOTICE
Contamination caused by fingerprints
Contamination caused by fingerprints can lead to vacuum deterioration or prolonged pumping
times.
4 Always wear lint-free gloves when touching the specimen, specimen holder, or stage.
3. Inspect the groove that holds the O-ring and remove any contamination.
4. Insert the new chamber door O-ring.
5. Close the chamber door.
Parts and Tools Isolation mounts setting tool (350600-1476-000), delivered with the microscope
CAUTION
Crushing fingers between table and microscope chamber
The gap between table and microscope chamber is designed to allow the specimen chamber
to swing freely. The size of the gap varies during operation and can cause crushing injuries.
4 Do not place your fingers between table and microscope chamber.
NOTICE
Isolation mounts
The isolation mounts are specified only up to 4.5 bar. If you pump up the isolation mounts
above 4.5 bar, then the isolation mounts could be destroyed.
4 When pumping up the isolation mounts, avoid pressures above 4.5 bar.
Procedure 1. To check if the isolation mounts need to be adjusted, give the column a gentle nudge:
If the column is able to move freely, then skip the subsequent steps; no further action is re-
quired.
If the column is unable to move freely, then adjust the isolation mounts as described in the
following steps.
2. To adjust the isolation mounts, remove the
caps from the inlet valves at the back of the
plinth.
7.2 Column
The gun vacuum deteriorates with time. For an optimum use of the LaB₆ filament it is essential
that the gun vacuum is always in the 10⁻⁷ mbar region. If the pressure rises above this value, it is
strongly recommended to bake out the gun head.
This procedure consists of the following steps:
1. Switching off the Gun | LaB₆ Filament [} 120] or Switching off the Gun and EHT | Tungsten Fil-
ament [} 120]
2. Starting the Bakeout [} 121]
3. Switching on the Gun | LaB₆ Filament [} 68] or Switching on the Gun and EHT | Tungsten Fila-
ment [} 72]
Info
If you operate the microscope with a LaB₆ filament, then you need to switch on or switch off
the gun and the EHT separately.
Info
If you operate the microscope with a tungsten filament, then the EHT and the gun are always
switched on or switched off together. You cannot separately switch on or off the EHT or the
gun.
Info
Tungsten filaments have only a limited lifetime of 100–300 operating hours. If you operate the
microscope with a tungsten filament, then do the following:
4 If the microscope is not in operation, then always switch off the EHT and gun to preserve
the filament.
4 Enable Auto Saturation at shutdown.
Info
You cannot work with the microscope while the bakeout procedure runs.
NOTICE
Hot surfaces during bakeout
Parts of the enclosure in the upper range of the column may become hot during bakeout, par-
ticularly after a long bakeout cycle.
4 Do not place any combustible objects on the grids of the electron optical column during
bakeout.
4 After the bakeout procedure, let surfaces cool down before working around the column.
4 Only advanced operators are allowed to perform the bakeout procedure.
Prerequisite ü Requires the Supervisor privilege and the user level Service.
ü Only advanced operators are allowed to perform the bakeout procedure.
Procedure 1. In the Panel Configuration Bar, double-click Bakeout.
à The Bakeout dialog is displayed.
2. If the Full service bakeout checkbox is available, deactivate the Full service bakeout
checkbox.
INFO: Full service bakeout includes column heating that may lead to column misalignment.
3. From the Bakeout drop-down list, select a bakeout cycle.
For 2 hours heating / 1 hours cooling, select Quick.
For 8 hours heating / 2 hours cooling, select Overnight.
For 40 hours heating / 3 hours cooling, select Weekend.
For a cycle defined by the operator, select User.
INFO: You can switch to Standby mode while the bake-out procedure is running. Before
you switch to Standby mode, ensure that the Partial Vent on Standby checkbox is deacti-
vated in the SEM Controls panel.
4. To start the bakeout procedure, click Bakeout Start.
5. After bakeout, allow the column to cool down.
INFO: The pressure changes due to residual heat in the gun area. A cooling period is neces-
sary to allow the vacuum to improve. The cooling period depends on the room tempera-
ture. The gun monitor monitors the pressure improvement as the cooling time progresses.
The filament is the electron source of the microscope. To extract electrons from the filament, it is
heated up and a high voltage is applied. During operation, the filament slowly degrades. The fila-
ment must be replaced if it is used up or damaged.
Info
To display the operating hours of the filament, select View > SEM Status > Select > Fila-
ment Age.
Info
ZEISS encourages all customers to keep a record on the filament exchanges. Such a record
simplifies maintenance.
With the EVO microscope, you can use the following types of filaments:
§ Tungsten filaments (pre-aligned)
§ Tungsten filaments (not pre-aligned)
§ LaB₆ filaments
The procedure for replacing the filament is similar for all types of filaments. For details and differ-
ences, refer to the lists for the different types of filaments after the overview figure.
Venting the Gun and the Specimen Chamber
Only LaB₆:
Fig. 44: Overview of the procedure for replacing the different types of filament
The preparations for filament exchange are the same for all filament types.
This procedure consists of the following steps:
1. Venting the Gun and the Specimen Chamber [} 124]
2. De-energizing the Microscope [} 125]
3. Disassembling the Gun [} 126]
4. Disassembling the Firing Unit [} 127]
5. Cleaning the Firing Unit [} 128]
6. Cleaning the Anode [} 130]
In order to replace the filament, you need to disassemble the gun. Before you can disassemble the
gun, you need to prepare the microscope: You need to switch off the gun and to vent the micro-
scope. In SmartSEM, you need to select the filament type that you want to install.
6. Vent the specimen chamber. Refer to Venting the Specimen Chamber [} 64].
This procedure completely cuts off the microscope from the electrical main supply.
WARNING
Residual voltage at the mains plug
After unplugging the mains plug residual voltage is present at the pins of the plug which may
cause electrical shock.
4 After unplugging the mains plug wait at least 5 s before touching the pins of the mains
plug.
Prerequisite ü The microscope is in Standby mode or in OFF mode, refer to Finishing the Work Session | LaB₆
Filament [} 108] or Finishing the Work Session | Tungsten Filament [} 108].
Procedure 1. If the microscope is not yet in OFF mode, press the Off button at the front of the plinth.
à When all subsystems are fully deactivated,
the Off button lights up red permanently.
2. Close and lock the main shut-off valves at the installation site.
3. If the EMO circuit is not installed, unplug the
power cord by unplugging the CEE connector
from the CEE FEMALE RECEPTACLE of the
mains supply.
9 The microscope is completely cut off from the electrical main supply.
You need to disassemble the gun so that you can dismount the filament holder and replace the
filament.
WARNING
Malfunction of medical devices near ion getter pumps
Magnetic fields present at the ion getter pumps may disturb the function of medical devices.
The magnetic fields are also present if the microscope is switched off.
If you wear medical implants that are susceptible to magnetic fields (e.g. cardiac pacemakers),
do the following:
4 Keep a distance of at least 30 cm from the ion getter pumps.
4 Follow the safety instructions provided by the pump manufacturer.
CAUTION
Hot firing unit
During operation, the firing unit gets hot. If you touch the hot firing unit, you may burn your-
self.
4 After you switch off the microscope, wait for 15 minutes before you touch any parts of
the firing unit.
WARNING
Risk of electrical shock
If you open the gun while the mains power is connected to the microscope, then contact may
cause electrical shock.
4 Always disconnected the mains power from the microscope before you open the gun.
2. Wait at least fifteen minutes for the firing unit to cool down.
Before you can replace the filament, you need to disassemble the firing unit, which contains the
filament holder.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
Over time, the filament deposits material on the firing unit and the anode. This can damage the
aperture and reduce the filament life time. You need to remove these deposits during every fila-
ment exchange.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
Over time, the filament deposits material on the firing unit and the anode. This can damage the
aperture and reduce the filament life time. You need to remove these deposits during every fila-
ment exchange.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
When you unpack a new filament holder, you need to use compressed air to clean it from any
dust particles.
7.2.2.3.1 Disassembling the Filament Holder | Tungsten (Not Pre-Aligned) and LaB₆
Before you can install a new filament to the filament holder, you first need to disassemble the fila-
ment holder.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
Parts and Tools Tungsten filament (not pre-aligned): Pack of 10 filaments (350010-2079-000)
Lint-free gloves
Adjust key (350061-1904-000)
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
7.2.2.3.4 Installing the Filament to the Filament Holder | Tungsten (Not Pre-Aligned) and LaB₆
After you have disassembled the filament holder and removed the old filament, you can install a
new filament.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
To assemble the firing unit, you need to reinstall the filament holder.
If you do not use a pre-aligned filament, you also need to install the tensator spring washer.
The firing units for LaB₆ and tungsten filaments are different. If you want to replace a LaB₆ fila-
ment with a tungsten filament, you have to replace the firing unit as well.
Info
Without further ado, you cannot install a firing unit for LaB₆ filaments to a microscope that is
designed for tungsten filaments. This can only be achieved by an upgrade of the microscope at
the factory.
At the customer’s location, this upgrade cannot be performed.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
Parts and Tools EVO firing unit for tungsten filaments (350071-2832-000) including a Wehnelt assembly and a
box of ten standard tungsten filaments. Only used when replacing LaB₆ with tungsten fila-
ments.
Lint-free gloves
Can of compressed air
Adjust key (350061-1904-000)
Hex key, 1.5 mm
Procedure 1. If the filament is not pre-aligned, insert the tensator spring into the firing unit.
INFO: The tensator spring washer is not designed to be used in combination with pre-
aligned filaments. If you install the tensator spring washer in combination with a pre-
aligned filament, then the filament height is incorrect. If you install a pre-aligned filament,
then do not install the tensator spring washer.
2. Align the slot at the side of the filament holder
with the pin inside the firing unit.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
3. Slightly shake the firing unit to check that the brass retaining washer is mounted correctly.
4. If the brass retaining washer makes any sound when shaking, then tighten the brass retain-
ing washer a bit more.
When you install the brass retaining washer, you need to screw it in with the correct number of
turns. The number of turns affects the filament distance and the possible applications according
to the following table:
Tab. 7: Tightening the brass retaining washer with the correct number of turns.
You need to check the correct filament distance and the centering of the filament with a stereo
microscope or with a hand lens.
Info
If the filament is not accurately centered or if the filament distance is not correctly adjusted,
then this may negatively affect the filament lifetime and the performance.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
4. If the filament is not centered, then turn the grub screws to center the filament.
INFO: The grub screws have pointed ends.
5. If the tip of the filament is not flush with the top of the firing unit, then turn the brass re-
taining washer to adjust the tip of filament distance.
6. Once the filament is flush with the top of the firing unit, turn the brass retaining washer in
anticlockwise direction.
Choose the number of turns according to the table above.
9 The filament distance is correctly adjusted.
The tip of the LaB₆ filament needs to be located at a distance of 0.375 mm below the top of the
firing unit (Wehnelt cap).
To adjust the filament tip to this position, you need to install and tighten the brass retaining
washer. You also need to adjust the grub screws to center the filament.
You need to check the correct filament distance and the centering of the filament with a stereo
microscope or with a hand lens.
Info
If the filament is not accurately centered or if the filament distance is not correctly adjusted,
then this may negatively affect the filament lifetime and the performance.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
4. Once the filament is flush with the top of the firing unit, turn the brass retaining washer by
270° in anticlockwise direction.
INFO: Now the filament tip is correctly aligned and is located approximately 0.375 mm be-
low the top of the firing unit.
5. Tighten or untighten the grub screws until the filament is centered.
INFO: The grub screws have pointed ends.
Use a stereo microscope or a hand lens to check wether the filament centered.
Before you can reinstall the gun to the microscope, you need to install the firing unit to the gun.
NOTICE
Contamination by dust particles or fingerprints
Dust particles or skin grease can cause contaminations, which lead to flashovers or to bad vac-
uum.
4 Wear lint-free gloves throughout the whole procedure of replacing the filament.
3. If your microscope is equipped with a column safety switch, check that the safety switch is
correctly lined up when you close the gun.
Before you can use the filament for SEM imaging, you need to evacuate the microscope and start
up the filament.
3. Wait until the messages Vac Status = Ready and EHT Vac ready = Yes are displayed.
INFO: It may take some time until the vacuum is established and the messages are dis-
played.
4. In the Gun tab of the SEM Controls panel, deactivate Long Fil. Life.
5. Switch on the electron beam.
6. Set the following values:
EHT = 10.00 kV
Spot Size = 500
Fil I Target = 1950 mA
To achieve an optimal alignment of the beam, you need to align the emission image to the center
of the image area.
NOTICE
Persisting electrical problems
Tripped circuit breakers or blown fuses may be a hint for an electrical problem in the micro-
scope.
4 If a circuit breaker keeps tripping or a fuse keeps blowing, de-energize the microscope
completely and contact your ZEISS service representative for assistance.
7.4 PC
It is important to periodically clean up the PC. This shortens the boot sequence and loading times.
If necessary, refer to standard Windows manuals for instructions to do this.
Procedure 1. Backup the database to the server or to other storage.
2. Delete the temporary files.
3. Check for adequate free space on the hard drives.
4. Backup the log file (EMServer.log).
5. Erase the original log files in the LOG folders.
6. Check that Windows updates are applied and that service packs are applied.
INFO: Each service pack includes all the patches since the last major release.
8.1 Decommissioning
If the Microscope System is not used for an extended period such as several months, it should be
shut down completely and secured against unauthorized access. Complete decommissioning of
the Microscope System should be executed by your ZEISS service representative.
WARNING
Malfunction of medical devices near ion getter pumps
Magnetic fields present at the ion getter pumps may disturb the function of medical devices.
The magnetic fields are also present if the microscope is switched off.
If you wear medical implants that are susceptible to magnetic fields (e.g. cardiac pacemakers),
do the following:
4 Keep a distance of at least 30 cm from the ion getter pumps.
4 Follow the safety instructions provided by the pump manufacturer.
WARNING
Biological hazards
Biological substances may pose a threat to the health of humans and other living organisms.
4 Keep a logbook of the biological substances loaded into the microscope and show it to
the ZEISS service representatives before they perform any work on the microscope.
WARNING
High leakage current
High leakage currents are present in the microscope. Contact may cause burn or electrical
shock.
4 Ensure proper grounding. For more information, refer to the Installation Requirements
document.
4 Do not operate the microscope without the separate ground connection.
WARNING
Radiation hazard due to X-rays
X-rays are generated inside the microscope during operation. This is unavoidable because elec-
trons are accelerated by voltages up to 30 kV.
4 Do not remove any parts around the column and chamber that are essential for radiation
protection.
4 Use genuine ZEISS parts exclusively.
4 Ensure that all local safety and X-ray protection regulations are met.
4 Only authorized ZEISS service representatives are allowed to service the microscope.
WARNING
Reaction products
Dangerous reaction products can be present in the specimen chamber during or after opera-
tion.
4 Ensure that there is an appropriate exhaust gas line to remove the waste gas of the pre-
vacuum pump and to transmit it to the outside.
4 Wear lint-free gloves when touching the inner parts of the specimen chamber or the speci-
men.
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4 During specimen exchange, keep the chamber door open as short as possible.
4 Do not inhale the air from within the specimen chamber.
4 Ensure that the area around the microscope is sufficiently ventilated.
4 If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
WARNING
Tilting hazard when removing the microscope from the crate
When removing the microscope from the wooden crate, it can tilt and crush a person.
4 Use a forklift to remove the microscope from the wooden crate.
WARNING
Crushing hazard when lowering the microscope
The microscope and its components are heavy. When the load is lowered during transport and
positioning, body parts can be crushed.
4 Maintain a safe distance.
4 Do not walk or place your hands or feet under the load while it is being lowered.
4 Wear safety shoes and gloves.
NOTICE
Damage during transport
Sensitive components of the microscope can get damaged during transport.
4 The microscope may only be transported in air-suspended vehicles.
4 Moving parts must be secured during transport to prevent them from slipping or tipping
over.
4 Install shock/tilt watches.
4 Avoid rocking the crates back and forth.
4 Devices for transporting the microscope must be rated to handle its full weight and dimen-
sions. Note the weight information on the package and on the shipping document.
4 Check that none of the items has been damaged during shipment.
4 Otherwise contact your local ZEISS service representative.
§ Please check the location requirements for door and hallway widths.
§ Check the entrance to the building and to the final site for suitable ramps and compliant ele-
vators that can match the weights of the Microscope System where necessary.
§ Some components, such as the tables, are large, heavy or bulky and may require extra assis-
tance to get the units into the allocated site.
Maximum shock § Do not drop or bump the boxes during movement or storage. Any acceleration shall be
resistance < 10 g.
§ Evaluate packaging shock and tilting sensors on delivery and after internal transport.
Packaging The microscope is delivered in two containers:
§ Reusable wooden crate
Dimensions and weight of crate:
1480 × 1480 × 2070 mm³ (W × D × H), appr. 878 kg
§ Cardboard box on Euro pallet
1400 × 1220 × 1350 mm³ (W × D × H), appr. 340 kg
Dimensions of plinth un-crated:
784 × 1015 × 1780 mm³ (W × D × H, depth is 1215 mm with PSU on transit clamp fitted to
the rear of plinth), appr. 650 kg
Check that none of the items has been damaged during shipment.
Guidelines for Due to the heavy weight of microscopes, a forklift has to be used to remove the microscopes
Unpacking from the wooden crate:
§ The forklift used must have a sufficient load capacity.
§ Refer to the weights of the microscope stated in this chapter.
Conditions during The packed microscope has to be stored in a dry place.
Storage and
Transport
Allowable Allowable temperature during on-site storage and transport:
temperature § Between −10 °C and +70 °C
Info
24 hours before installation of the Microscope System it is required that the boxes be at
recommended room temperature to avoid ingress of humidity, which is very harmful to optical
paths, and to ensure effective stability of the Microscope System during installation and test-
ing.
8.3 Disposal
The Microscope System and its components must not be disposed of as domestic waste or
through municipal disposal companies. They must be disposed of in accordance with applicable
regulations (WEEE Directive 2012/19/EU). ZEISS has implemented a system for the return and recy-
cling of devices in member states of the European Union that ensures suitable reuse according to
the EU Directives mentioned. The customer is responsible for decontamination.
Info
Detailed information on disposal and recycling is available from your ZEISS Sales & Service Part-
ner.
8.4 Decontamination
A decontamination statement must be submitted before returning any used objects to the ZEISS
location.
If reliable decontamination cannot be guaranteed, the hazard must be marked according to appli-
cable regulations. In general, a well-visible warning sign must be affixed to the article itself and to
the outside of the packaging, together with detailed information on the type of contamination.
B C D E
F
1 2
G
A
5
3
6 4
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes Main Components Length Width (mm) Height Weight (kg)
(mm) (mm)
Pollution degree 2
Parameter Value
Installation cate- II
gory
Exhaust Line If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Nitrogen Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. The nitro-
gen can be taken either from a gas cylinder or from an in-house supply system.
Parameter Requirement
Flow rate Approx. 3 l/min for ventilation of specimen chamber with chamber
door open
Pressure 0.2–3.3 bar
Connection hose 4 mm inside diameter. 10 m are delivered with the microscope.
Instrument con- Quick exchange connector. One is delivered with the microscope.
nection
Compressed Air Compressed air is used to operate several valves and the auto leveling system. The necessary com-
Supply pressed air can be either generated by a compressor (part no. 345596-0000-000) or taken from a
gas cylinder or from an in-house supply system.
Magnification Range: <7x – 1,000,000x referenced to Polaroid 5" × 4.5" image for-
mat
Parameter Description
Image framestore 32768 × 24576 pixel, signal acquisition by pixel, line and frame inte-
gration and averaging, including drift compensated frame averaging
(limitations may apply to averaging mode and maximum scan speed
for large images)
Specimen stage Type: 5-axes motorized Cartesian controlled via the SmartSEM user
interface or operated by a dual joystick control box
Movements:
§ X = 80 mm
§ Y = 100 mm
§ Z = 35 mm
§ T = −10° to 90°
§ R = 360° continuous
INFO: The movements may be reduced by specimen size, operating
conditions, and accessories attached.
Specimen weight:
Standard stage
§ 0.5 kg (full motion)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Large Z stage (optional)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Parameter Description
CL detector (optional):
Cathodoluminiscence (CL) chamber detector (option for non-VP con-
figurations, included in VP option)
SCD (optional)
Specimen current detector
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes Main Components Length Width (mm) Height Weight (kg)
(mm) (mm)
Pollution degree 2
Parameter Value
Parameter Value
Installation cate- II
gory
Exhaust Line If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Nitrogen Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. The nitro-
gen can be taken either from a gas cylinder or from an in-house supply system.
Parameter Requirement
Flow rate Approx. 3 l/min for ventilation of specimen chamber with chamber
door open
Pressure 0.2–3.3 bar
Connection hose 4 mm inside diameter. 10 m are delivered with the microscope.
Instrument con- Quick exchange connector. One is delivered with the microscope.
nection
Compressed Air Compressed air is used to operate several valves and the auto leveling system. The necessary com-
Supply pressed air can be either generated by a compressor (part no. 345596-0000-000) or taken from a
gas cylinder or from an in-house supply system.
Parameter Requirement
Magnification Range: <7x – 1,000,000x referenced to Polaroid 5" × 4.5" image for-
mat
Image framestore 32768 × 24576 pixel, signal acquisition by pixel, line and frame inte-
gration and averaging, including drift compensated frame averaging
(limitations may apply to averaging mode and maximum scan speed
for large images)
Specimen stage Type: 5-axes motorized Cartesian controlled via the SmartSEM user
interface or operated by a dual joystick control box
Movements:
§ X = 125 mm
§ Y = 125 mm
Parameter Description
§ Z = 50 mm
§ T = −10° to 90°
§ R = 360° continuous
INFO: The movements may be reduced by specimen size, operating
conditions, and accessories attached.
Specimen weight:
Standard stage
§ 0.5 kg (full motion)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Large Z stage (optional)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Parameter Description
CL detector (optional):
Cathodoluminiscence (CL) chamber detector (option for non-VP con-
figurations, included in VP option)
SCD (optional)
Specimen current detector
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes Main Components Length Width (mm) Height Weight (kg)
(mm) (mm)
Parameter Value
Pollution degree 2
Parameter Value
Installation cate- II
gory
Exhaust Line If toxic chemicals or biological specimens are used an exhaust line is recommended to remove the
waste gas of the pre-vacuum pump and to transmit it to the outside.
Nitrogen Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. The nitro-
gen can be taken either from a gas cylinder or from an in-house supply system.
Parameter Requirement
Flow rate Approx. 3 l/min for ventilation of specimen chamber with chamber
door open
Pressure 0.2–3.3 bar
Connection hose 4 mm inside diameter. 10 m are delivered with the microscope.
Instrument con- Quick exchange connector. One is delivered with the microscope.
nection
Compressed Air Compressed air is used to operate several valves and the auto leveling system. The necessary com-
Supply pressed air can be either generated by a compressor (part no. 345596-0000-000) or taken from a
gas cylinder or from an in-house supply system.
Magnification Range: <7x – 1,000,000x referenced to Polaroid 5" × 4.5" image for-
mat
Parameter Description
Image framestore 32768 × 24576 pixel, signal acquisition by pixel, line and frame inte-
gration and averaging, including drift compensated frame averaging
(limitations may apply to averaging mode and maximum scan speed
for large images)
Specimen stage Type: 5-axes motorized Cartesian controlled via the SmartSEM user
interface or operated by a dual joystick control box
Movements:
§ X = 130 mm
§ Y = 130 mm
§ Z = 50 mm
80 mm with large stage (option)
§ T = −10° to 90°
§ R = 360° continuous
INFO: The movements may be reduced by specimen size, operating
conditions, and accessories attached.
Specimen weight:
Standard stage
§ 0.5 kg (full motion)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Large Z stage (optional)
§ 2 kg (not tilted)
§ 5 kg (only XY motion)
Parameter Description
CL detector (optional):
Cathodoluminiscence (CL) chamber detector (option for non-VP con-
figurations, included in VP option)
SCD (optional)
Specimen current detector
The Microscope System is in compliance with the requirements of the following regulations and
directives:
The Microscope System and its accessories have been classified as instrument category 9 (labora-
tory equipment or comparable standard). It also comply, as applicable, with the EU-regulations
2011/65/EU (RoHS) and 2012/19/EU (WEEE).
The resulting EMC environment can change through the use / installation of 3rd party compo-
nents (e.g. detectors / plasma cleaners). As soon as the EMC environment of a 3rd party compo-
nent is "domestic environment" or "EMC controlled environment", the entire Microscope System
must be classified as an "EMC controlled environment". Due to the lower immunity requirement,
the user must take suitable protective measures with regard to electromagnetic fields in the instal-
lation room as well as interferences from the mains power supply.
In addition to the European and international guidelines and standards, the 21 CFR §1040.10:
"Performance Standards for light emitting products - laser products" applies for the USA.
The following EMC user notice is for Korea only:
기종별 사용자안내문
European and International Directives / Standards: For more information on ISO, CSA, SEMI certifi-
cates or CE Declarations of Conformity, contact your ZEISS Sales & Service Partner.
ZEISS works according to a certified Environment Management System according to ISO 14001.
The Microscope System was developed, tested and produced in accordance with the valid regula-
tions and guidelines for environmental law of the European Union.
10.1 Consumables
Small pliers –
Stubs –
Tweezers –
Lint-free gloves –
Glossary
aBSD Condenser
Annular Backscattered Electron Detector Device that collects and focuses the elec-
tron beam onto the specimen.
Aperture
Mechanical limitation of an opening ori- D
ented perpendicular to the optical axis, Depth
which filters out electrons whose trajec-
tories (tracks) do not run close to the op- DECT
tical axis.
Digital Enhanced Cordless Telecommuni-
cations
Astigmatism
Lens aberration that distorts the shape of Depth of field
the electron beam, compensated by the
Distance along the optical axis which an
stigmator.
object in the specimen can be moved
while remaining in focus.
AWD
Analytical Working Distance DPA
Differential Pumping Aperture
Backscattered electrons
High-energy electrons that are liberated EBIC
from the specimen surface when the
Electron Beam Induced Current
specimen is hit by the primary electron
beam.
EC
Bakeout European Community
Degassing of surfaces of a vacuum sys-
tem by heating during the pumping EDS
process. Energy Dispersive X-ray Spectroscopy
BSD EDX
Backscattered Electron Detector Energy Dispersive X-ray Spectroscopy
BSE EHT
Backscattered Electron Extra High Tension
C2D EIGA
Cascade Current Detector European Industrial Gases Association
C2DX EM server
Extended Cascade Current Detector A server that implements the internal
communication between control soft-
CAN ware and microscope hardware.
Controller Area Network
EMO
CCD Emergency Off
Charge-Coupled Device
EP
CL Extended Pressure
Cathodoluminescence
Faraday cup R
Small insulated metal container, R-axis (Rotation)
equipped with an aperture where elec-
trons can enter but not escape. Used to RF
measure the specimen current in the mi-
Radio Frequency
croscope.
SCD
Focus wobble
Specimen Current Detector
Function that sweeps the focus of the
objective lens backwards and forward
through the focus on the specimen Scintillator
plane. When the aperture is misaligned a Substance that absorbs electrons and in
lateral shift is observed. response, fluoresces photons while re-
leasing the previously absorbed energy.
GUI
Graphical User Interface SE
Secondary Electron
H
Height Secondary electrons
Low-energy electrons that are emitted
HDBSD from the specimen surface when the
specimen is hit by the primary electron
High Detection Backscattered Electron
beam. Secondary electrons are generated
Detector
by inelastic scattering.
HV
SEM
High Vacuum
Scanning Electron Microscope
IGC
SEMI
Industrial Gases Council
Semiconductor Equipment and Materials
International (SEMI) is a industry associa-
LaB₆ tion comprising companies involved in
Lanthanum hexaboride the electronics design and manufacturing
supply chain.
PC
Personal Computer STEM
Scanning Transmission Electron Micro-
PE scope
Protective Earth (ground)
Stigmator
PE Compensates astigmatism (lens aberra-
tion), so that the electron beam becomes
Primary Electron
rotationally symmetrical.
Penning gauge
T
Device for measuring high vacuum in the
T-axis (Tilt)
vacuum system.
TEM
Pre-vacuum pump
Transmission Electron Microscope
A pump for generating a pre-vacuum.
User
Primary electron beam
Person examining a sample under the mi-
Narrowly bundled beam of accelerated
croscope.
electrons that hit the specimen surface.
Warning label
A label or sign that provides safety-re-
lated information (e.g. safety signs) and
informs about possible risks and hazards.
WD
Working Distance
WDS
Wavelength Dispersive X-ray Spec-
troscopy
WDX
Wavelength Dispersive X-ray Spec-
troscopy
X
X-axis
X-ray
Type of high energy electromagnetic ra-
diation, that is generated during the op-
eration of electron microscopes.
Y
Y-axis
YAG
Yttrium Aluminum Garnet
Z
Z-axis
ZEISS
ZEISS is an internationally leading tech-
nology enterprise operating in the fields
of optics and optoelectronics. Further in-
formation about ZEISS can be found at
www.zeiss.com.
Index
A BSD 43, 85
C2D 42, 83
Acceleration voltage 27 C2DX 85
Accessory 172 CL 46
Air Conditioning and Quality 154, 159, 164 Everhart-Thornley 33
Alignment SE 33, 83
Aperture 80 STEM 45
Gun 79 VPSE 39
Analysis mode 29 YAG BSD 44, 87
Aperture 23, 27, 113 Differential pumping aperture 23
Automated function Disposal 148, 151
Auto aperture alignment 80 DPA 23
Auto gun alignment 79 Dual joystick 47
B E
Backscattered electron 31, 43 EasyVP mode 23
Bakeout 120 EHT 27
Baking out the gun head 120 Electron optical column 27
Beam deceleration 88 Electron Optics 156, 162, 167
BSD detector 43, 85 Emergency shutdown 111
BSE 43 Emitter life 113
Environmental Requirements 156, 161, 167
C EP mode 23, 26
Everhart-Thornley detector 33
C2D detector 42, 83
EVO column 27
C2DX detector 85
Exhaust Line 155, 161, 166
Cathodoluminescence 31, 46
Extended pressure mode 23, 26
CCD camera 37
Chamber door seal 117
Chamberscope 37 F
Checking Faraday cup 81
Position of the circuit breaker 147 Field mode 29
Temperature 117 Filament 27, 122
Chemical 113 Fisheye mode 29
Circuit breaker 147 Focus wobble 76
CL detector 46
Comissioning 56
Compressed Air Supply 156, 161, 167 G
Consumable 113, 171 General Safety Information 10
Consumable and chemical 113 GUI 50
Contamination 151 Gun head, bakeout 120
Control panel 48
Cover panel, protective 18
H
Hazard
D Electrical hazard 11
Decontamination 151 Generated by materials and substances 12
Depth mode 29 Generated by radiation 13
Detection System 158, 163, 169 Mechanical hazard 11
Detector 32, 82 Radiation hazard 13
Thermal hazard 13
Hazards 11
Prevention 11
HD BSD 43
I Requirements
for Operators 10
Image Resetting the touch alarm 117
Acquisition 61, 73 Resolution mode 29
Optimization 76
Saving 79
Initializing the stage 115 S
Installation 55 Safe Operating Condition 10
Intended use 9 Safety 9, 112
devices 18
J interlocks 18
Safety label 14
Joystick TV angle 116 Sample type selection 92
Saving images 79
K Scintillator 113
SE detector 33, 83
Keyboard shortcut 59
Secondary electron 31
Shortcuts 59
L Shutdown
License Emergency 111
REDUCED 77 Finishing the work session 108
Location Requirements 155, 161, 166 Microscope 148
Locking device 20 Signal detection 30
SmartSEM
User interface 50
M SmartSEM program suite 53
Mains Connection 154, 159, 165 SmartSEM user interface 50
Maintenance 112 Software 8, 50
Consumable and chemical 113 Spare part 171
interval 112 Special keys 59
schedule 112 Specimen Chamber 157, 162, 168
work 113 Specimen current 81
Microscope Stage 157, 162, 168
De-energizing 110, 125 Post initialization position 116
Shutdown 148 Stage bias
Switching on 57, 58 High voltage 88
Mid-column aperture 27 Low voltage 89
Stage control 47
Stage initialization 115
N Starting SmartSEM 59
Nitrogen 156, 161, 166 Starting the microscope 58
STEM detector 45
Switching off
O Gun 120
ON/OFF switch 19 Gun and EHT 120
Online help 59 Microscope 108, 110, 125
Operation 56 Switching on
Operation modes 29, 81 EHT 68
Optimizing the image 76 Gun 68
Gun and EHT 72
P Microscope 57, 58
Switching to standby 108
Performance data 153, 159, 164
Pressure-limiting aperture 23
Primary electron 31 T
Probe current 81 Temperature 117
Tool 172
R Touch alarm, reset 117
Training 10
Replacing the chamber door seal 117 Transmitted electron 31
Troubleshooting 114
Type plate 18
U
User access level 52
User interface 50
User privilege 52
V
Vacuum system 22
Variable pressure mode 23
VP mode 23
VPSE detector 39
W
Warning
labels 13
lights 13
Weight and Sizes 153, 159, 164
Y
YAG BSD detector 44, 87
Z
ZEISS
Portal 8
Service agreements 112