Multi Tron Manual
Multi Tron Manual
Multi Tron Manual
Operating Manual
2020-08-20
We bring life to your laboratory.
The best Multitron
ever
Doc-Nr. 77408 - EN
V.02.00 - Original
info@infors-ht.com
service@infors-ht.com
Multitron - Operating Manual
Table of Contents
Table of Contents
4 Options .................................................................................. 35
4.1 Cooling ......................................................................... 35
4.1.1 Internal Cooling ............................................. 35
4.1.2 External Cooling ............................................ 37
4.1.3 Operating the Cooling Unit ............................ 38
4.1.4 Specifications and Technical Data .................. 38
4.2 Direct Steam Humidification .......................................... 39
4.2.1 Setup and Function ........................................ 39
4.2.2 Connection Conditions .................................. 40
4.2.3 Operating the Direct Steam Humidification .... 41
4.2.4 Specifications and Technical Data .................. 41
4.3 Pressure Elimination Unit for Direct Steam
Humidification............................................................... 42
4.3.1 Setup and Function ........................................ 42
4.3.2 Cleaning the Pressure Elimination Unit ........... 43
4.3.3 Autoclaving the Pressure Elimination Unit ...... 45
4.3.4 Dimensions .................................................... 47
4.3.5 Specifications and Technical Data .................. 47
4.4 CO2 Control .................................................................. 48
4.4.1 Setup and Function ........................................ 48
4.4.2 Connection Conditions .................................. 49
4.4.3 Operating the CO2 Control ............................ 49
4.4.4 Specifications and Technical Data .................. 49
4.5 UV Decontamination ..................................................... 50
4.5.1 Setup and Function ........................................ 50
4.5.2 Operating the UV Decontamination ............... 50
4.5.3 Specifications and Technical Data .................. 50
4.6 Removable Shelf............................................................ 51
Table of Contents
5 Accessories ............................................................................ 57
5.1 Trays ............................................................................. 57
5.1.1 Universal Table Tray ....................................... 57
5.1.2 Pre-Fitted Trays .............................................. 58
5.1.3 Tray with Sticky Stuff ..................................... 63
5.2 Clamps and Other Holders ............................................ 65
5.2.1 Clamps .......................................................... 66
5.2.2 Test Tube Holders .......................................... 67
5.3 Box for Microtitre Plates ................................................ 68
5.4 eve® .............................................................................. 71
6 Installation ............................................................................ 72
6.1 Operating Conditions at the Installation Location .......... 72
6.2 Requirements for the Mains Connection ....................... 73
6.3 Minimum Distances to the Device ................................. 74
7 Operation .............................................................................. 76
7.1 Switching on the Device ................................................ 76
7.2 Loading the Device ........................................................ 76
7.2.1 Opening the Door .......................................... 76
7.2.2 Inserting and Removing the Tray .................... 77
7.2.3 Fitting the Holders ......................................... 80
7.2.4 Tips & Tricks for Loading the Tray .................. 82
7.3 Overview about the Display und Controls ...................... 84
7.3.1 Display Area................................................... 85
7.3.2 Signs and Messages Regarding the Timer
Function ........................................................ 86
7.3.3 Error, Warning, and Alarm Signs .................... 87
7.3.4 Operating Panel ............................................. 88
Table of Contents
Table of Contents
Table of Contents
General Information
1 General Information
This manual enables the safe and efficient handling of the device.
All the information and instructions in this operating manual comply
with the current standards, legal regulations, the latest technological
and scientific developments and the knowledge gained from the
manufacturer’s many years of experience in this field.
General Information
DANGER
The signal word “DANGER” indicates a dangerous situation that
will lead to severe or even fatal injuries if not avoided.
WARNING
The signal word “WARNING” indicates a potentially dangerous sit-
uation that may result in severe or even fatal injuries if not
avoided.
CAUTION
The signal word “CAUTION” indicates a potentially dangerous sit-
uation that may result in minor injuries if not avoided.
ATTENTION
The word “ATTENTION” on a blue bar indicates a situation that
may result in significant damage to property if not avoided.
INFORMATION
Texts located below a grey bar bearing the notice “INFOR-
MATION” provide useful tips and recommendations for ensuring
efficient, fault-free operation of the device.
General Information
Manufacturer name
Designation = Category of device
Type = Device type (name)
S/N = Serial number
Year = Year of manufacture
Mains = Nominal voltage and frequency
Current = Current consumption
Manufacturer address
CE marking
Our Customer Service is at your disposal for technical advice and spe-
cialist enquiries. For contact information, see page 2.
Due to their familiarity with the potential applications of the device,
the Customer Service team is able to provide information on whether
the device can be used for a specific application or modified to han-
dle the planned process.
Furthermore, our colleagues are always interested in new information
and experiences resulting from user’s applications for the device that
may be valuable for the continued development of our products.
This section also refers to areas that are the responsibility of the pro-
vider due to certain risks arising from particular applications for which
the device is used deliberately and with full awareness of the associ-
ated risks.
WARNING
The device is designed and constructed exclusively for the intended
use described above.
Intended use also includes following all the instructions in this man-
ual, especially those relating to:
The installation site
Use of suitable cultivation vessels
User qualifications
Permissible parameter setpoints
Correct operation and maintenance
Any use of the device outside the scope of the intended use as de-
scribed above shall be deemed misuse. This also applies to applica-
tions for which the device is not designed, especially the following:
The device is not protected against explosions. Use and manu-
facture of explosive gases as well as operating the device in the
Ex area are therefore not permitted.
The device is not designed to sufficiently protect its users if path-
ogenic organisms escape uncontrollably. Cultivation of patho-
genic organisms of risk categories 3 and 4 is therefore not per-
mitted.
To use the device for special applications not covered by conven-
tional, intended use, the device must be modified and certified ac-
cordingly by the manufacturer.
ATTENTION
Use of unsuitable or defective cultivation vessels can lead to glass
breakage and therefore damage to property.
To avoid the vessels coming out of the clamps at very high speeds,
they might have to be secured using cable ties underneath the
springs or some other suitable measure.
INFORMATION
For trays with Sticky Stuff, special provisions apply in relation to
maximum permitted speeds. These must be observed to prevent
cultivation vessels from detaching.
Due to the complexity of the device and the potential risks arising
from its operation, the device may only be used by qualified, special-
ist personnel.
2.3.1 Provider
The term “provider” applies to all persons who are responsible for
making the device and the necessary infrastructure available. These
persons may also be included in the group of people known as “us-
ers”, though this is not always the case.
Irrespective of whether a provider is a member of the company’s
board of management or a supervisor, they bear a special level of
responsibility with regard to the processes and the qualification and
safety of the users.
2.3.2 User
General
The term “user” applies to all persons who come into contact with
the device in any way and perform work on or with it. This primarily
applies to the following activities, which can be performed by the
manufacturer’s own specialists or a variety of other persons (it is not
always possible to distinguish clearly between the different types of
person):
Qualified personnel
On account of their specific education, training and – in many cases –
experience, the qualified personnel required for this work are able to
recognise risks and respond accordingly to potential hazards.
The qualified personnel (either internal or external) who cannot be
categorised under the separate “operators” group are made up of
the following groups of persons:
2.3.3 Operator
The “operators” are a specific sub-group of users distinguished by
the fact that they work with the device. They are the true target audi-
ence for this operating manual.
Qualified technicians
Only technicians who have been trained for working in a biotechnol-
ogy laboratory can be considered for the role of operator. These in-
clude:
Technicians in training
Persons in this group who are undergoing training or apprenticeships
are only permitted to use the device under supervision and in accord-
ance with the instructions of a trained and qualified technician.
The term “unauthorised persons” applies to all persons who can ac-
cess the work area but are not qualified to use the device in accord-
ance with the aforementioned requirements.
Unauthorised persons are not permitted to operate the device or use
it in any other way.
The device is used for industrial and scientific purposes. As such, the
provider of the device is individually liable with regard to the legal
requirements relating to occupational health and safety in a biotech-
nology laboratory. In particular:
The provider is responsible for ensuring that the work and envi-
ronmental regulations applicable in a biotechnology laboratory
are observed.
The provider must ensure that the device remains in safe and
proper working condition throughout its entire term of use.
The provider must ensure that all safety equipment is fully func-
tional and is not disabled.
The provider must ensure that the device is only worked on by
qualified users, and that said users receive sufficient training.
The provider must ensure that the protective equipment re-
quired for working with the device is provided and worn.
The provider must ensure that this operating manual remains in
the immediate vicinity of the device throughout its entire term
of use.
This section covers general hazards and residual risks that are always
present when using the device in accordance with normal, intended
use.
The following notices are general in nature. As such, with a few ex-
ceptions they are not repeated in the remaining sections.
This section covers particular hazards and residual risks that may arise
when using the device for special applications in accordance with
normal, intended use.
Since the use of the device for such applications is deliberate, it is the
responsibility of the operators and the provider to ensure that all per-
sonnel are protected from potential damage to health. The provider
is responsible for ensuring that the appropriate protective equipment
for such applications is provided, and that the necessary infrastruc-
ture is in place.
If the provider intends to use the device for such purposes, he must
check its suitability for the planned application with the responsible
local authorities.
Overheating shut-down
The heating of the device is protected against overheating by a tem-
perature limiter. This triggers as soon as the temperature on the heat-
ing element exceeds the maximum allowed value and immediately
turns off the heating.
Door monitoring
The position of the door is monitored electronically. If the door is
opened, the shaker drive is stopped immediately. As soon as the door
is fully closed again, the shaker drive restarts automatically.
Position
On the housing of the optional direct steam humidification on the
rear of the device.
Meaning
Observe the instructions in the operating manual when connecting
the direct steam humidification to ensure compliance with the re-
quired water quality. For information regarding the required water
quality, see chapter 12.5.3 "Direct Steam Humidification", page 148.
WARNING
Illegible or missing warning symbols on the device will lead to the
user being exposed to risks that the warning symbols in question
were designed to make him or her aware of.
Important notice
If the return shipment is not accompanied by a signed and complete
declaration of decontamination or it is not affixed to the outer pack-
aging, the shipment will be returned unopened to the sender at their
expense (see also T&C).
Exterior
4 3
5
10 9 8 7 6 5
Interior
1 2
9
5
8 7
1 Finned heating element and axial fans (in back 6 Limit stop table pull-out
wall, not visible) 7 Door heating
2 CO2 and humidity sensor (optional) 8 Stop bar
3 Interior lighting (not visible) 9 Table
4 Pt100 temperature sensor
5 Slide rail
Brief description
The Multitron incubation shaker is used to cultivate micro-organisms
in a laboratory environment. The basic version of the device is fitted
with a shaker drive and a heater. Depending on the version of the
device, it will come with either a 3 mm, 25 mm or 50 mm throw. A
version with an adjustable throw (12.5, 19, 25 and 50 mm) is also
available.
The device can be operated either locally, using the operating panel
on the front of the device, or remotely, using a computer connected
to the device via an Ethernet interface.
Stacking devices
If necessary, up to three units can be stacked to save space. The de-
vices can also be stacked subsequently but stacking must be per-
formed by the manufacturer’s qualified expert personnel or persons
authorised by the manufacturer.
If stacked devices are operated, you have to keep in mind that the
maximum permissible rotation speed is limited for the top units in the
stack.
Overview of options
To be able to adapt the device to different application situations, the
following options are available:
The standard features of the device include the shaking function (pa-
rameter RPM) and temperature control (parameter Temp).
1
2
The counterweight used to balance the mass is fitted under the table.
Depending on the design, the deflection of the circular movement is
either 3, 25 or 50 mm or 12,5, 19, 25 or 50 mm for devices with an
adjustable throw. The table moves in clockwise circles.
Table
The table is used to hold the 85 x 47 cm (type M) tray, various ver-
sions of which are available. Bars on the sides, two stops and two
tapered plugs ensure that the tray is positioned correctly.
1 2 3 4
1 3 4
Adjustable throw
If the device features an adjustable throw, there are four possible po-
sitions for the eccentric throw:
12.5 mm
19 mm
25 mm
50 mm
The counterweights are also adjustable to reduce vibrations during
operation.
INFORMATION
The throw may only be adjusted by personnel that has received
appropriate training and authorisations. If in doubt, contact IN-
FORS HT (see contact details on page 2).
Operation
The Rotation speed parameter is operated via the operating panel.
Beside the actual value, the symbol RPM and the unit min-1 appear on
the display.
For an exact description of how to set the parameters, see chapter
7.4 "Adjusting, Activating and Deactivating Parameters", page 90.
INFORMATION
The permissible load weights depend on the position of the device
in the stack, the throw and the shaking speed (see chapter 7.2.4
"Tips & Tricks for Loading the Tray", page 82).
1 Finned heating element and radial fans (in back 2 Pt100 temperature sensor
wall, not visible)
INFORMATION
The temperature sensor serves exclusively to measure the tempera-
ture of the air in the incubation chamber. The temperature sensor
must not be removed from the holder and immersed in the liquid
in the cultivation vessel.
Operation
The Temperature parameter is operated using the operating panel.
Beside the actual value, the symbol Temp and unit °C appear on the
display.
For an exact description of how to set the parameters, see chapter
7.4 "Adjusting, Activating and Deactivating Parameters", page 90.
The mains connection is located on the right side of the device. Three
different versions of the device are available for different mains volt-
ages:
230 V 50 Hz
230 V 60 Hz
115 V 60 Hz
Prior to connecting the device, make sure that the connecting values
of the device match those of the local power supply. The mains con-
nection must be easily accessible at all times so that the device can be
disconnected from the power supply quickly in case of an emergency.
On the right side of the casing there is a socket (stereo jack, 3.5 mm)
for connecting the device to an alarm system. It is labelled ALARM
EXTERN. The socket is designed for a maximum of 34 V AC/DC, 1 A.
The alarm relay is permanently activated as long as the alarm is active
(not toggling).
Allocation of contacts
1 2 3
1 COM (common)
2 NC (normally closed)
3 NO (normally open)
The Ethernet interface can be used to integrate the device into a net-
work and thus control it using a remote computer. To control the de-
vice using a remote computer, eve® the bioprocess platform software
is required. If the device is to be controlled using other software, a
special communication protocol is required. For further information
contact INFORS HT.
The Ethernet interface makes it possible to send and receive data. For
example, setpoints can be sent to the device or values measured by
the sensors can be sent to eve®. If the device is connected to eve® via
Ethernet or if the device is controlled using eve®, this is displayed on
the operating panel.
For more information on the use of the Ethernet interface, see chap-
ter 7.6 "Using eve® to Operate the Device", page 109.
3.4 Openings
INFORMATION
In case of large fill volumes, we recommend installing the dis-
charge hose to avoid the bearings coming into contact with liquid
if a flask breaks.
If the plug (1) is directed inwards and the bevelled parts of the
plate (2) point outwards, the plate seals the opening.
In contrast, if the plate is mounted in such a way that the bev-
elled parts face the device, this results in a ventilation slot.
INFORMATION
2
The device is always delivered with the ventilation opening open. If
a significantly higher gas consumption is detected when using the
CO2 option, the ventilation opening can be closed if necessary.
Please note that closing the ventilation opening can lead to a lack
of oxygen in the culture and therefore to low growth.
The device has air vents on the right-hand side and the rear of the
casing. When setting up the device, you must ensure that these air
vents remain unobstructed (keeping a minimum distance of 80 mm
to walls or objects on either side) in order to ensure that the exhaust
air can be extracted without obstruction and none of the compo-
nents overheat.
The device features two LED spots to illuminate the incubation cham-
ber. These are located in the door area on the inside of the casing.
In the settings menu (OPTION key) the interior lighting can be com-
pletely deactivated if necessary.
3.6 Frames
Low base, 13 cm
Individual units and stacked devices can be fitted with a 13 cm base.
The base has an adjustable foot that can be used to level the device.
High base, 41 cm
Individual units and devices stacked in pairs can be fitted with a
41 cm base. The base has an adjustable foot that can be used to level
the device.
The interior of the high base can be used as a storage space, e.g. for
trays that are currently not in use. Alternatively, a suitable cooling
unit is available as an option for the high base.
INFORMATION
In case of an emergency shut down via the power switch, all pa-
rameter setpoints are stored and the device restarts immediately
when it is switched on via the power switch.
All device functions can be controlled directly via the operating panel
on the front of the unit. The operating panel is divided into display
and operating sections:
The identification plates for identifying the device are located on the
right side of the casing, directly above the mains connection and on
the front of the door.
For information on the data provided on the identification plate see
chapter 1.3 "Device Identification (Standard Identification Plate)",
page 9.
At the back wall inside the incubation chamber, there is a sticker that
identifies the throw of the device.
Options
4 Options
To adapt the device to the specific usage scenarios, the device can be
enhanced with various options. The options can be ordered at the
same time as the device or can be fitted at a later stage. For detailed
information on the options as well as advice, contact our customer
service (for contact information, see page 2).
4.1 Cooling
For processes that require low temperatures, the device can be fitted
with internal cooling or connected to an external cooling system.
3
2
Options
The device can be equipped with an internal cooling unit. The cooling
unit can either be placed in a bracket on the incubation shaker or in
the high base (900 W version only).
The version with 1800 W cooling capacity is used when very low
temperatures have to be reached. As two independent compressors
are installed in this version, it is also ensured that cooling can con-
tinue if one compressor fails (redundancy).
The cooling unit is an autonomous system and has its own power
supply and mains switch. This is located on the right-hand side of the
bracket or base. The cooling unit is a 2-circuit system with a coolant
tank. Prior to commissioning, it must be filled with coolant by the
manufacturer’s qualified expert personnel or persons authorised by
the manufacturer.
The heat exchanger is located behind the rear wall in the incubation
chamber.
Options
Connection conditions
1 2
The connections for an external cooling system are on the top of the
device. The connections are labelled COOLANT IN (inlet) and COOL-
ANT OUT (outlet). If two or three units with external cooling are
stacked, the lines are connected internally and only the top unit has
connections on the top.
ATTENTION
The installed valve regulates the temperature by closing and open-
ing the cycle. Hence it can be necessary to install a bypass from
the inlet to the outlet to protect the circulation pump against dam-
age.
Options
Options
Options
Observe the following points when using the direct steam humidifica-
tion:
The connection for water supply for the direct steam humidification is
located at the bottom of the casing and is designed for hoses with a
diameter of 6 to 7 mm.
ATTENTION
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the device.
Adding detergents, disinfectants or similar chemicals to the water
for the direct steam humidification can damage the device. Only
use water without additions, in accordance to the specification
(see chapter 12.5.3 "Direct Steam Humidification", page 148).
Options
Options
A B
C
I
H D
G E
Options
The hose lines to the steam generators are routed at the rear of the
unit and are fastened with hose brackets. Up to three steam genera-
tors can be connected to one pressure elimination unit via T-pieces.
Options
5. Remove the lid. To do so, loosen and unscrew the four cross-
head screws.
6. Carefully lift the lid upwards.
7. Clean the vessel and other components as required. For infor-
mation on correct cleaning, see the information below.
Detergents
For cleaning use a mild detergent, e.g. dishwashing liquid or neutral
cleaning agent.
Options
ATTENTION
Aggressive cleaning agents, solvents and abrasive cleaning utensils
(hard sponges, brushes) can scratch surfaces, damage the equip-
ment and impair its function. In addition, do not use cleaning
agents containing chlorine, as stainless steel is not resistant to
substances containing chlorine.
Disinfectants
Only use quaternary ammonium compounds for wipe-down disinfec-
tion. As a tried as tested disinfectant, we recommend Fermacidal D2.
An alcohol-based disinfectant (e.g. 70 % ethanol) can be used to
rinse the system.
Materials
If other cleaning or disinfecting agents should be used, it is the re-
sponsibility of the operator to check whether the agent is compatible
with the materials used. The following table gives an overview of the
materials of the components of the pressure elimination unit.
Components Material
Vessel, lid, pipes, level sensor, float valve, mount- Stainless steal
ing bracket
Rapid couplings Polysulfone
Flat gaskets EPDM
Pressure hoses Silicone (VMQ)
CAUTION
If there is no pressure compensation, overpressure may occur in
the vessel. Therefore always keep the hose piece with the filter
open.
ATTENTION
If the vessel is autoclaved empty, no steam can build up. The steri-
lization success is not guaranteed.
Options
Prerequisites:
INFORMATION
The solenoid valve is closed in de-energized state (normaly closed,
NC). It is therefore not mandatory to fit a blind cap to the water
connection.
1. Close the ports of the water connection and the level sensor
with a blind cap each:
a) Ensure that on both ports a flat gasket is placed.
b) Place the blind cap on the flat gasket.
c) Fix the blind cap with the clamp.
2. Close the water inlet line to the steam generators with the sup-
plied hose piece with rapid coupling.
Options
4.3.4 Dimensions
Dimensions Multitron with pressure elimination unit
Options
CAUTION
During normal operation, small amounts of CO2 are constantly dis-
sipated into the environment. Leaking CO2 can lead to discomfort,
loss of consciousness or even death by suffocation, especially in
small rooms.
– The operator is obliged to install a monitoring system which
permanently monitors and analyses the CO2 concentration in
the room.
– If you are using the CO2 control, only operate the device in
well-ventilated areas.
– Comply with local health and safety regulations and regula-
tions regarding the use of CO2 and its limits.
– Regularly check the gas pipelines for leaks.
INFORMATION
Opening the ventilation opening at the back of the device can lead
to significantly higher gas consumption. We therefore recommend
leaving the ventilation opening closed when using the CO2 control
option.
Options
Gas supply
Constant input pressure not exceeding 0.5 bar (± 0.1)
Only use clean, dry, oil and dust free gas.
Hoses
Use only pressure-resistant and intact hoses.
The hose must have an appropriate diameter; an adapter may
be used, if necessary.
Secure hoses using appropriate clamps.
The CO2 control is operated using the CO2 parameter. The top alpha-
numeric display shows the actual value in % with the symbol CO2. To
prevent large amounts of CO2 from escaping, the CO2 control is de-
activated automatically while the door is open.
Options
4.5 UV Decontamination
Options
The removable shelf (approx. 865 mm x 490 mm) allows static incu-
bation in climatic conditions almost identical to the cultivation flasks.
Lower position
If the removable shelf is in the lower position, shake flasks with a vol-
ume of up to 250 millilitre (maximum height: 150 mm) can be placed
on the tray.
Middle position
If the removable shelf is in the middle position, shake flasks with a
volume of up to 1 litre (maximum height: 220 mm) can be placed on
the tray.
Upper position
If the removable shelf is in the upper position, shake flasks with a vol-
ume of up to 3 litres (maximum height: 320 mm) can be placed on
the tray.
ATTENTION
The inserted removable shelf can change the climate conditions in
the incubation chamber and lead to higher tolerances as it hinders
air circulation. To prevent damage to cultures, the climate above
and beneath the removable shelf should be checked frequently
and, if necessary, monitored using mobile sensors.
Options
The device can be equipped with a LED lighting option. This illumi-
nates the surface of the table at an adjustable light intensity. For ex-
ample, the LED lighting option can be used to cultivate phototrophic
cells.
The LED lighting is affixed to the ceiling inside the chamber. The light
generated by the LED lighting is distributed evenly over the surface of
the table.
The LEDs are protected from being hit by a protective cover made of
acrylic glass. The acrylic protective cover has a large opening in the
middle to allow air circulation. The protective cover reduces the
height of the incubation chamber by approx. 30 mm.
The LED lighting unit is operated using the Light parameter. The bot-
tom of the alphanumeric display shows the value in % with the sym-
bol Light.
The light intensity can be regulated from 1 to 100°%.
Restriction
The LED lighting radiates heat when in operation. Hence, the LED
lighting can only be operated together with a cooling unit (internal or
external cooling system). If the device is operated with internal cool-
ing, the temperature is regulated fully automatically. If the device is
connected to an external cooling system, you must ensure that the
cooling is fully functional when operating the LED lighting.
Options
4.8 Darkening
The glass panel of the door can be covered with white adhesive film
on the inside. This is used for darkening as well as to ensure better
light distribution in the incubation chamber when using LED lighting.
The output module is located at the back of the device in its own cas-
ing. The signal range corresponds to 4 up to 20 mA with a precision
of ± 0.05 mA.
Options
Options
WARNING
If an external device that is supplied with mains power is operated
on the table whilst it is operating, there is a risk that the cable
might break or get jammed. This poses a risk of death by electro-
cution.
– Electrical devices may only be used when the shake function
is switched off (e.g. in conjunction with the Temperature pa-
rameter.)
– Electrical devices may only be used in conjunction with the
Humidity parameter if the device is adequately protected
against humidity. Observe the permissible environmental con-
ditions for operating the device.
ATTENTION
To avoid damage to when installing the cables and/or hoses in the
incubation chamber, make sure that they do not rub and do not
get jammed.
Options
INFORMATION
The inner side of the incubation chamber as well as the door can
become slightly discoloured due to the silver in the lacquer. This
discoloration has no negative impact on the antimicrobial charac-
teristics of the coating or the cultivated product.
Accessories
5 Accessories
In the following section, the available accessories for the device are
described. For special applications, please contact our customer ser-
vice for advice (for contact information, see page 2).
5.1 Trays
Different trays are available for the device; these can be purchased
individually or ordered with the device. In addition to the universal
table tray, which can be equipped as desired, various trays with fixed
features are available for different purposes.
Accessories
Accessories
The trays are designed for plates with the dimensions 85.1 x 127
mm.
Accessories
ATTENTION
The spacers are made of plastic and therefore cannot be auto-
claved.
Accessories
1 2
3 2
The trays with clamping assemblies are designed for microtitre and
deep-well plates with dimensions 85.1 mm x 127 mm. To ensure suf-
ficient clamping force can be achieved, plates with a web height of at
least 3 mm must be used.
Accessories
The bars are sheathed in a soft material and arranged on two differ-
ent levels in order to securely hold bottles of different sizes.
The scope of delivery for the tray includes the fixed frame and 8 mov-
able sliding bars.
Accessories
ATTENTION
If condensate forms on the tray or on the cultivation vessels, the
adhesiveness of the Sticky Stuff is no longer guaranteed. As a re-
sult, cultivation vessels can become detached from the Sticky Stuff
and break.
If you are looking for a very flexible option for equipping the
tray, Sticky Stuff offers an ideal alternative.
However, you have to keep in mind that speeds are limited
when using Sticky Stuff (for more information on this, refer to
the table at the end of the chapter).
Use only vessels with a broad flat base. Large Erlenmeyer flasks
(e.g. 3000 mL) adhere more strongly than smaller ones (e.g.
500 mL).
Ensure that the entire bases of the flasks are on the adhesive
matting. They must not extend beyond the edge of the tray.
Accessories
Vessels that are stuck can be removed from the adhesive mat-
ting by using a syringe to inject water under the flask.
Due to their shape (wide base, short neck), Fernbach flasks in
particular can be difficult to remove. If necessary, cover part of
the adhesive matting with the protective foil provided.
The adhesive power can deteriorate over time due to dust and
dirt. To clean and restore full adhesive power, thoroughly wipe
down the surfaces with a brush or a dish sponge and clear wa-
ter with mild detergent (dishwashing liquid). Then let dry over-
night.
Quaternary ammonium compounds are suitable for disin-
fection.
Consider application times and thoroughly rinse with water. If
the adhesive matting is disinfected at regular intervals, it might
need to be replaced sooner.
Accessories
INFORMATION
Due to limited resistance to disinfectants as well as the risk of un-
intentional detaching of flasks, Sticky Stuff is not suitable for culti-
vating pathogenic microorganisms.
INFORMATION
The following guidelines only apply to undamaged, completely dry
adhesive matting and flasks free of grease. If old or dirty adhesive
matting is used, there is a risk that flasks detach even at lower
speeds.
For individual fitting of the universal table tray, various clamps and
holders are available.
Accessories
5.2.1 Clamps
Clamps in various sizes are available for equipping a universal table
tray. These can be ordered individually and mounted on the universal
table tray.
Plastic clamps
The following plastic clamps are available:
Accessories
Test tube holders are used to securely affix different sized test tubes.
Test tube holders can be screwed onto a universal table tray or
placed on Sticky Stuff adhesive matting.
The following test tube holders are available (details and special types
on request):
If necessary, the inserts with the tubes can be tilted by loosening the
black nuts (1). The tilt can be set freely. Then tighten the nuts again.
The entire inner part with the holders for the tubes can be removed
by loosening the black nuts, whereby the u-shaped holder remains
on the tray.
3
When inserting the inner parts, ensure that the feet of the holder
(part on the tray) on both sides rest between the rubber ring (3) and
the Teflon washer (2).
1 2
Test tube holders can also be used on a tray with Sticky Stuff. To do
this, the pre-installed screw must be removed from the base plate.
ATTENTION
The test tube holders adhere extremely strongly to the Sticky Stuff,
so that water has to be sprayed underneath the edge of the
holder with a syringe in order to remove them. This is done to pre-
vent damage to the holder or the tray.
Accessories
Overview
The box for microtitre plates is used for low-vapour cultivation of mi-
crotitre plates and deep well plates. It protects cultures from drafts in
the incubation chamber and also offers a stable container in which
cultures can be transported.
For shaking, the box for microtitre plates can be screwed onto a uni-
versal table tray or placed on the Sticky Stuff adhesive matting.
Structure
1 2
1 Filter
2 Top plate
3 Microtitre insert
6 4 Quickload spacers
5 Container
6 Filter washer
5 4 3
Alternatively, the box for microtitre plates can be placed on the Sticky
Stuff adhesive matting. In this case, all fastening screws must be re-
moved to avoid damaging the adhesive matting.
Accessories
1 2
Inserting/replacing a filter
To insert or replace a sterile filter (3), the filter washer (2) must be re-
moved first. To do so, press the locking button (1) on the filter
washer and lift up the filter washer. Following that, the filter can be
inserted and the filter washer can be put back in its place.
3
Sterilising the box for microtitre plates
The box for microtitre plates can be autoclaved (recommendation: 20
min at 120 °C). To do so, remove the Quickload spacers first and dis-
infect them using commercial disinfectant.
INFORMATION
When working with maximum rotation speeds, it is strongly rec-
ommended to use a special tray with permanently mounted boxes.
Details on request.
Accessories
Technical data
Container with top plate and tray
Description Value
Dimension 465 x 280 mm
Weight 4.23 kg
Filter, paper, round, Ø 185 mm
Screws for microtitre tray M4 x 8 hexalobular
Screws for microtitre box M4 x 12 Phillips head
Material Aluminium, autoclavable
Detergent Mild dishwashing liquid or neutral
cleaning agent
Quickload spacer
Description Value
Dimension 412 x 14 x 5 mm
Quantity 12 units
Material Polyamide 6.6
can be sterilised
Detergent Mild dishwashing liquid or neutral
cleaning agent
Loading capacity
Standard plate size Value
85.1 x 127 mm
Microtitre plates 18 units
Deep well plates height 42 mm 12 units
Accessories
5.4 eve®
Installation
6 Installation
WARNING
Installation and initial commissioning require trained expert per-
sonnel with sufficient experience. Errors during installation may
lead to dangerous situations or significant damage to property.
– Only the manufacturer’s expert personnel or person’s author-
ised by the manufacturer may install and initially commission
the device.
– The manufacturer must also be consulted if the device is sub-
sequently moved.
Operating conditions
Temperature range 10 up to 30 °C
Humidity 10 up to 85 %
Restrictions Do not expose to direct sun-
light
Do not expose to dust
Do not expose to vibrations
Installation
INFORMATION
In order to achieve the desired temperature in the incubation
chamber, the ambient temperature directly at the device is deci-
sive. This can be considerably higher than the temperature at other
positions in the room due to waste heat from the device or other
devices with strong heat radiation in the immediate vicinity!
Installation
Installation
ATTENTION
Failure to comply with the required minimum distances can result
in components of the device overheating and the device being
damaged. Therefore, please note:
– Never cover the air vents on the right side or those on the
rear of the device.
– Never position the device directly on a wall.
Operation
7 Operation
ATTENTION
The shaker drive can start automatically if the device has not been
switched off correctly beforehand. Loose objects in the incubation
room can damage the device and the cultivation vessels.
Procedure 1. Open the door halfway and wait until the table has stopped.
You can hear a click when the door opening is released.
2. Open the door completely. If the door does not unlock straight
away, lift it up slightly to take the load off the locking mecha-
nism.
Operation
The shaker drive and all other parameters will start up again auto-
matically as soon as the door is closed.
INFORMATION
If you wish to stop the shaker drive slowly, e.g. to avoid culture liq-
uid from spilling out, you have to set the Rotation speed parame-
ter to OFF prior to opening the door. In this case, the shaker drive
stops slowly, with the brake force set in the settings menu (see
chapter 7.7.4 "Setting the Brake Force for Stopping the Table",
page 112).
ATTENTION
The door can only be opened fully when there is no load on the
half-open door. If the door is pushed down violently, this can
damage the door mechanism.
Door alarm
To prevent the door from being left open unintentionally, the device
has a door alarm (beeping). This is triggered if the door is left open
for longer than the set time. If no parameter was active prior to
opening the door, the door alarm is not triggered. The time until the
door alarm is triggered can be set in the settings menu (1 to
10 minutes). If necessary, the door alarm can also be deactivated
completely.
When you open the door completely, the ejection levers (1) under the
table lift the tray out of the table’s locking cones (2). The tray is now
free and can be pulled out of the incubation chamber by hand.
1 2
Operation
To remove the tray from the incubation chamber, the knurled screw
1 2 1 has to be unscrewed again. Following that, the tray can be lifted out
of the locking cones and be pulled out of the incubation chamber by
hand.
Operation
3. Push the tray between the guide rails over the front stop bar up
to the limit stop at the back of the incubation chamber. The tray
must audibly click into place behind the front stop bar.
4. Check whether the tray is placed properly behind the front stop
bar.
5. Device with 25 mm or 50 mm throw: Close the device door.
When the door closes, the ejection levers lower and the tray is
lowered into the two cones. The tray is thus securely affixed to
the table.
6. Device with 3 mm throw: Ensure that the tray is correctly
snapped into the two conical plugs on the table.
7. Device with 3 mm throw: Insert the knurled screw and tighten
it by hand. Tighten the knurled screw so that it cannot come
lose during operation.
8. Check whether the tray is securely affixed to the table.
ATTENTION
If the tray is not held in place securely on the table, the tray’s
movement may damage the interior of the device.
– Do not start up the device unless the tray has been inserted
correctly and affixed securely.
– If you are using a device with a 3 mm throw, make sure the
knurled screw is inserted and tightened correctly.
Operation
ATTENTION
The threaded holes of the tray can be damaged if the fastening
screws are not screwed straight into the thread:
– Place the screws vertically on the drill holes.
– Ensure they can be tightened easily.
ATTENTION
All holders are supplied with pre-fitted screws. If a test tube holder
or a box for microtiter plates is placed on the Sticky Stuff adhesive
matting, there is a risk that the adhesive matting is damaged by
protruding screws.
Remove the screws before you place test tube holders or boxes for
microtiter plates on the Sticky Stuff adhesive matting.
Fitting a holder
Clamps, test tubes and boxes for microtiter plates are fixed to the
tray using screws. The device is delivered with pre-fitted screws (2).
The screws are each secured with one flat gasket at the bottom (1) so
that they cannot fall out. Only use the screws provided or screws of
the same size for assembly.
2
To fit a holder, proceed as follows:
Operation
Screw sizes
ATTENTION
If screws that are too long are used to affix the holders, these will
stick out at the bottom of the tray. This has the effect that the tray
can no longer be inserted and affixed correctly.
If you have to replace lost fastening screws, you must use screws
that comply with the specification below.
Steel clamps
Plastic clamps
Operation
The working volume should not exceed 1/3 of the total volume
of the vessel.
If any of the vessels protrude over the edge of the tray, they
could collide with the inner wall, resulting in the device being
damaged or vessels breaking. As such, always ensure that the
cultivation vessels do not protrude when positioning them on
the tray.
Operation
Stacked devices
With devices stacked in three, very low loads (< 10 kg per unit) and
high rotation speeds (> 300 min-1) at the same time can result in vi-
brations occur. In this case it is recommended to place additional cul-
tivation vessels filled with water on the tray. This ensures a smooth
run. With a load of more than 10 kg per unit, no significant vibra-
tions occur and the device can be operated over the entire rotation
speed range without any limitations.
WARNING
If the loading mass of the table is too high or too low or the load
is distributed unevenly, high rotation speeds can cause vibrations.
This can result in the device moving uncontrollably.
– Do not operate the device without tray and loading.
– In case of strong vibrations, reduce the rotation speed and
check the loading weights or the distribution of the load.
Operation
1 7
2
8
4 9
10
6
4 Lower display: Parameter setpoints and actual 9 Lower operating panel: For setting the
values (RPM, humidity); time (timer 2) parameters (RPM, humidity), time (timer 2)
5 Display area for the timer function 10 Further operating keys (selection, TIMER and
OPTION)
6 Display for the ethernet function when the de-
vice is operated via ethernet Parameters in gray are only available when the
according option is installed.
Operation
Operation
In the upper display area the sign Time 1 lights up, when within the
timer function (see below) the setpoint value for the duration of the
first phase (Phase 1) can be entered.
First, the message OFF appears, if the timer is not activated.
In a similar way but not at the same time, the according information
for the second phase (Phase 2, Time 2) is shown in the lower display
area.
When entering a time between one minute (00:01) and 23 hours and
59 minutes (23:59) the units h and min light up.
When entering a time between 24 hours or one day (01:00) and the
maximum of 9 days and 23 hours (09:23) the units days and h light
up.
The sign Timer indicates that either within the timer function the set-
points for the duration (Time 1 and/or Time 2) can be entered or that
at least one timer has been activated.
The symbol for the cycle mode lights up in addition to the sign Timer
as soon as two durations (Time 1 and Time 2) have been entered and
both timers are activated. Then the device alters between the param-
eter setpoints of the first and the second phase in an endlessly re-
peated cycle.
The signs Phase 1 and Phase 2 indicate in which phase the device is
operating if both timers are active with different parameter settings.
/
They also indicate for which phase (Time 1 or Time 2) the parameter
setpoints can be entered or the actual values of the parameters are
shown on the alphanumeric display.
Operation
Between the upper and the lower alphanumeric display area there is
a separate field with various signs associated with error, warning, and
alarm messages.
The big red general symbol for warnings lights up in connection with
error messages and alarms to emphasise the importance of the mes-
sage.
It always flashes up when the setpoints are changed to avoid mis-
takes. The warning symbol also lights up, if somebody tries to enter a
setpoint value or tapping the ON/OFF key without tapping the SET
key first. In addition, it also appears if someone tries to enter an inva-
lid setpoint value.
The sign ALARM appears in connection with other signs like ERROR,
High or Low and it indicates that there is a problem with the device.
In addition, there will be an acoustic alarm.
The sign RESTARTED indicates that the device had been disconnected
from the power supply and did restart automatically on the basis of
the saved parameter setpoints.
Operation
The various keys on the operating panel serve to select and set the
parameters, timers and other functions and modes. On the panel
there are selection keys in orange (parameters, operator mode) and
in blue (timer function) as well as white keys for the setting.
The white keys for the setting exist twice because they are arranged
in an upper and a lower group within the operating key area which
corresponds with the according display area.
INFORMATION
The sensor controlled soft-touch keys are behind a glass cover and
they only need to be touched very gently.
INFORMATION
Tapping an arbitrary key on the operating panel turns on the light
inside the incubation chamber. 20 seconds after tapping the last
key it is automatically turned off.
With the TIMER key the timer function is selected. Afterwards, the
two timers can be programmed using the keys in the upper or lower
operating area.
On both display areas the symbols Time 1 (upper area) und Time 2
(lower area) are displayed
For detailed information see chapter 7.5 "Timer Function", page 95.
By holding the OPTION key for at least 2 seconds the operator mode
is selected. It allows to change the setup of the device.
The OPTION key is also used to select the functions within the opera-
tor mode.
For detailed information see chapter 7.7 "Settings Menu (Option
Function)", page 111.
Operation
With the Plus key the values for the parameters or the time are in-
creased.
By tapping the key once, the value is increased with the smallest in-
crement for the particular parameter. By holding the key, the incre-
ment increases in ever bigger steps.
With the Minus key the values for the parameters or the time are re-
duced.
By tapping the key once, the value is reduced with the smallest incre-
ment for the particular parameter. By holding the key, the increment
increases in ever bigger steps.
The ON/OFF key serves to turn the parameters and timers on or off.
Operation
Has the device been turned on via the power switch and is the initiali-
sation process terminated, the device is ready for getting pro-
grammed.
INFORMATION
Which temperatures can effectively be reached depends on various
factors, like the ambient temperature, the ventilation or the tem-
perature of the other device in a stack.
Operation
INFORMATION
The highest rotation speed allowed depends on the throw, the po-
sition of the device in a stack and the load on the tray (mass).
Humidity (Humid)
The parameter humidity (Humid) does not belong to the standard
scope of delivery and therefore it is only available if the according op-
tion is installed.
When entering the setpoint value for the relative humidity it is shown
in percent with an accuracy of one decimal place.
If the parameter is not installed, the display remains empty.
The sign High or Low lights up, if the current value of a parameter is
above respectively below its setpoint value.
After a certain time span or when a critical value is reached an alarm
will be set off, because most probably an error or defect hinders the
device to reach or maintain the setpoint value in question.
The possible setpoints for the parameters rotation speed and temper-
ature can be limited in the settings menu. If an attempt is made to
enter a higher value, the message bloc appears.
Operation
1 2 3
Procedure 1. Select the desired pair of parameters with the Select key (1).
The alphanumeric display field shows the current values of the
selected parameter pair.
2. To activate programming, tap the SET key (2) in the upper or
the lower setting key area, depending on which parameter set-
point should be set.
The alphanumeric display field shows last set setpoint of the pa-
rameter. The Setpoint sign indicates, that the parameter setpoint
can be set.
INFORMATION
The Setpoint sign lights up for 10 seconds. If you do not make an
entry in that time, you have to tap the SET key again.
3. Use the Plus or Minus key (3) to set the desired setpoint of the
selected parameter.
INFORMATION
The entered value is saved automatically. A separate confirmation
is not necessary. If the parameter is already activated, the changes
take effect immediately.
Operation
1 2 3
INFORMATION
Steps 1 and 2 can be omitted, if the parameter is turned on di-
rectly after entering the setpoint value.
Procedure 1. Select the desired pair of parameters with the Select key (1).
The message OFF indicates, that the parameter is switched off.
2. Tap the SET key (2) in the upper or the lower setting key area,
depending on which parameter should be switched on.
On the alphanumeric display the setpoint value of the parameter
is shown. The sign Setpoint indicates, that the parameter can be
switched on.
3. To switch on the parameter, tap the ON/OFF key (3) in the ac-
cording operating key area.
The parameter is instantly turned on.
The sign Setpoint disappears. On the alphanumeric display the
message on appears for a very short time, then the current value
of the parameter is shown.
INFORMATION
When the parameter rotation speed (RPM) is activated, the shaker
table starts to move instantly. It is therefore suggested to turn it on
last.
Operation
INFORMATION
It takes a while for the parameters to reach the entered setpoint
value. The sign High or Low lights up, if the current value of a pa-
rameter is above respectively below its setpoint value.
Operation
7.5.1 Overview
The timer function allows you to programme defined sequences (e.g.
changing parameter setpoints after a defined time period or delaying
the start of the cultivation process). The timer can be programmed in
two different modes:
INFORMATION
More complex sequences and procedures (e.g. daylight curves) can
be realised with the bioprocess platform software eve®.
The signs Time 1 and Time 2 indicate, that the duration of phase 1
(Time 1) respectively phase 2 (Time 2) can be entered.
The signs Phase 1 and Phase 2 indicate which phase is active or for
which phase the parameter setpoints can be entered:
The sign is blinking: The phase is active.
The sign is lit up: The parameter setpoints of the corresponding
phase can be entered.
If the cyclic change between phase 1 and phase 2 is active, the sign
cycle lights up.
Operation
Application examples:
Delayed start of the cultivation process
Induction of a protein expression
Slow down or stop the cultivation process after the given time
period
INFORMATION
To programme a one-time change from phase 1 to phase 2, the
timer for the second phase has to be switched off (Time 2 = off). If
the timer for the second phase is switched on (Time 2 = on), there
will be a cyclic change between the two phases (see the next ex-
ample).
Operation
Application example:
Simulation of day and night
Operation
Operation
INFORMATION
Possibly the message OFF for the second timer (Time 2) appears on
the lower alphanumeric display. Then the TIMER key needs to be
tapped again to get to the first timer (Time 1).
2. To set the time, tap the SET key in the upper setting key area.
The sign Setpoint indicates, that the time can be entered.
3. Use the Plus or the Minus key in the upper setting key area to
set the time for the first phase (Time 1).
The time set is shown either in hours (h) and minutes (min) or in
days (days) and hours (h).
Operation
4. Tap the ON/OFF key in the upper setting key area to activate
the timer (Time 1).
On the upper alphanumeric display the message on appears for
a very short time and then the time entered is shown
In the separate timer display area the sign Timer is lit up and the
sign Phase 1 lights up for the first time.
INFORMATION
It is not possible to set the parameters for the first phase as long as
the timer for the first phase has not been activated.
Entering the setpoints for the first phase and switching on the
parameters
As soon as the timer for the first phase (Time 1) has been switched
on, the different parameters for phase 1 can be selected using the
Select key and afterwards be switched on.
INFORMATION
In this phase of the programming process the sign Setpoint re-
mains lit up and does not disappear after 10 seconds.
6. Tap the Plus or Minus key in the corresponding setting key area
to set the desired setpoint values and switch the parameters on
using the ON/OFF key.
INFORMATION
With the Select key the parameters and the time entered for the
first phase can be selected and changed over and over again.
Operation
8. Make sure the second timer is turned off. If not, use the
ON/OFF key in the lower setting key area to turn the timer
(Time 2) off.
INFORMATION
If the timer should be used to stop the cultivation process after a
defined time, all parameters for phase 2 can be switched off. This
causes the parameters to switch off after the set time for phase 1
has elapsed and therefore to stop the cultivation process.
Operation
10. Tap the Plus or Minus key in the corresponding setting key area
to set the desired setpoint values and switch the parameters on
using the ON/OFF key.
12. Tap the ON/OFF key in the lower setting key area to start the
timer function.
On the alphanumeric displays the messages Strt (upper) and run
(lower) appear. In the lower display area the white sign Setpoint
is lit up. The display then changes to showing the remaining
time of the first phase on the upper and the message run on the
lower alphanumeric display.
In the separate timer display area the sign Timer is lit up. The
blue sign Phase 1 is blinking.
Operation
13. Tap the Select key to quit the timer programming mode.
If the Select key is not used, the display changes to the normal
operating mode automatically after 60 seconds.
The displays now show the current values of the integral param-
eters temperature and rotation speed. If other parameters are
installed, the display shows the current values of each pair of pa-
rameters for 20 seconds.
The cultivating process now runs with the parameter setpoint values
for the first phase, until the time entered for the first timer (Time 1)
has elapsed.
After the time for the first phase has lapsed, the message End ap-
pears on the lower alphanumeric display. This message alternates
every 20 seconds with the current values of the parameters for the
second phase. In addition, there is an acoustic alarm every 60 sec-
onds.
The cultivating process now runs with the setpoint values of the sec-
ond phase until the device is stopped manually by turning the param-
eters off. If the alarm is confirmed by tapping the TIMER key, the
acoustic alarm stops, the message End does not appear anymore.
INFORMATION
The timer function can at any time be deactivated separately or by
turning off the first timer. The cultivating process then continues
with the last active parameter setpoint values.
Operation
Programming the timer for a cyclic change works the same way as
programming the timer for a single change. The only difference is
that for a cyclic change the second timer (Time 2) is switched on as
well.
As soon as the second timer is switched on, the symbol cycle lights
up in the separate timer display area.
INFORMATION
It is not possible to activate the timer for the second phase as long
as the timer for the first phase has not been activated.
After starting the timer, the cultivating process runs with the alternat-
ing setpoint values of the first and the second phase until the device
is stopped manually by turning the parameters off. The symbol of the
currently active phase (Phase 1 or Phase 2) is blinking.
INFORMATION
The timer function can at any time be deactivated separately or by
turning off the first timer. When the first timer is turned off, also
the second timer is deactivated and the cultivating process contin-
ues with the last active parameter setpoint values.
Operation
3. If required, tap the Plus or the Minus key in the upper setting
key area to set the time for the first phase (Time 1).
INFORMATION
Do not tap the ON/OFF key after changing the timer setting, as
this would turn off the timer immediately.
If this happens in the cycle mode with the timer for the first phase,
also the timer for the second phase is deactivated. And if acci-
dentally the second timer is turned off, it cannot be activated
again.
4. Tap the TIMER key to get to the time settings for the second
phase (Time 2).
On the lower alphanumeric display the setpoint value last en-
tered for the time interval of the second phase (Time 2) is
shown. The sign Setpoint indicates, that the time can be en-
tered.
5. If required, tap the Plus or the Minus key in the lower setting
key area to set the time for the second phase (Time 2).
6. Tap the TIMER key again to quit the programming mode of the
timer.
On the alphanumeric display the remaining time of the active
phase and the message run appear.
7. Tap the Select key return to the normal operating mode of the
display.
Operation
Procedure 1. Select the desired pair of parameters by tapping the Select key
– if necessary several times.
In the two alphanumeric display areas the setpoint values of the
selected pair of parameters are shown.
2. Tap the SET key in the upper or lower operation key area.
The sign Setpoint indicates, that a setpoint can be set.
3. Tap the Plus or Minus key in the according setting key area to
set the new setpoint value.
The new setpoint value for the parameter appears on the ac-
cording alphanumeric display.
Operation
2. Use the Select key to select the desired parameters within the
phase chosen.
In the two alphanumeric display areas the setpoint values of the
selected pair of parameters are shown.
The sign Setpoint indicates, that a setpoint can be set.
3. Tap the Plus or Minus key in the according setting key area to
set the new setpoint value.
The new setpoint value for the parameter appears on the ac-
cording alphanumeric display.
4. Use the Select key to get to the next pair of parameters or quit
the parameter setting mode with the TIMER key.
5. Tap the Select key to return to the normal operation mode of
the display.
Operation
2. Tap the ON/OFF key in the lower operating key are to turn the
timer function off.
On the alphanumeric displays the messages Strt (upper) and OFF
(lower) appear. In the lower display area the white sign Setpoint
is lit up.
In the separate timer display area the sign Timer and maybe the
symbol cycle are lit up.
INFORMATION
The timer function is not deactivated by opening the door or by an
interruption in the power supply.
Operation
INFORMATION
For detailed information on how to use eve® to operate the device,
refer to the documentation supplied with eve®.
The device comes with the option of being operated from an external
computer via the Ethernet interface. This requires use of the eve®
platform software for bioprocesses. The Ethernet interface can be
used to send or receive data. For example, setpoints can be sent to
the device or parameters can be activated or deactivated. In addition,
the actual values of the parameters and the values measured on the
sensors as well as alarm and error messages can be transferred to
eve® and thus logged.
INFORMATION
To use the Ethernet interface with other tools, a special communi-
cation protocol is required. For more information, contact INFORS
HT.
Points to be observed
The following points must be observed if you want to use eve® to
control the device:
Operation
Requirements
To ensure the device can be operated via the Ethernet interface, the
function must be activated in the settings menu. Here, you can
choose between the options on (read only) and Auto (read and write)
(see chapter 7.7.9 "Setting up Data Exchange via Ethernet", page
115).
Operation
The settings menu enables you to adjust the main device settings. To
open the device’s settings menu, proceed as follows:
On the top display, the OPt message appears with the symbols
Setpoint and High. The Settings menu is thus open and you can
make the desired settings.
Tapping the OPTION key again lets you switch to the next func-
tion. If no key is tapped for 60 seconds, the display returns to
normal operating mode. If all functions are clicked through, the
device also returns to normal operating mode.
INFORMATION
The limits for the allowed setpoints of the parameters can only be
set within a few factory-set values.
Operation
Default value: 65 °C
Operation
INFORMATION
When opening the door of the device, the table is stopped with
the maximum brake force, regardless which brake force is set.
If the key pad lock has been activated, you can then tap the OPTION
key followed by the SET key to enter a PIN (number between 0 and
9999). Once the PIN has been entered, this must be confirmed using
the ON/OFF key (the message PIN OK appears on the top display).
INFORMATION
By entering the number “1756” (Super PIN), the PIN is deactivated
and set to zero.
INFORMATION
When the key pad lock is active, this information can be read out
via the Ethernet interface. In this way, when the device is remotely
controlled (e.g. using eve®), it can be determined and logged
whether the key pad lock is active or not.
The key tone (clicking sound when tapping a key) can be activated or
deactivated. To do so, tap the SET key (the sign Setpoint appears)
and then use the ON/OFF key to activate or deactivate the key tone.
Operation
Operation
You can set whether and in which direction data can be sent via the
Ethernet interface. To do so, tap the SET key (the writing Setpoint
appears) and then use the ON/OFF key to select the desired setting.
Default value: 0
The device has a door alarm. This is triggered if the door is left open
for too long. To activate or deactivate the door alarm, tap the SET
key (the sign Setpoint appears) and then use the ON/OFF key.
The time after which the alarm is triggered can be set. To do so, tap
the SET key again (the sign Setpoint appears) and then use the Plus
or Minus key to set the desired time period (1-10 minutes).
Operation
INFORMATION
Improper modification of the throw of a device with an adjustable
throw represents a safety risk. The throw may therefore only be
changed by personnel that has received appropriate training and
authorisations. If you have any questions, contact your local IN-
FORS HT representative (see contact details on page 2).
Operation
ATTENTION
Parameters, that have not been switched off prior to switching off
the device, are automatically activated when the device is switched
on. This can lead to damages to the device and cultivation vessels.
Always switch off all parameters before switching off the device.
INFORMATION
The parameter setpoints remain stored for approximately one
month.
Procedure 1. Switch off all parameters. Make sure that, in addition to the ro-
tation speed, parameters that are not visible, such as tempera-
ture, humidity or CO2, are also switched off.
2. Switch off the device by pressing the power switch.
3. If you do not intend to use the device for a long time, discon-
nect the device from the power supply.
Rectifying Faults
8 Rectifying Faults
The following section describes possible reasons for faults and how
to rectify them.
Fault messages
A distinction is made between alarms (ALARM) and error messages
(ERROR):
Rectifying Faults
INFORMATION
An alarm is only triggered if the actual value of the parameter
does not change for a certain period. If there is a fluctuation, the
counter for triggering the alarm is reset.
INFORMATION
If the setpoint for a parameter is limited in the settings menu of
the device and an attempt is made to enter higher value, the bLoc
message appears.
The alarm message RESTARTED appears if the device has switched it-
self back on based on the stored parameters after a power failure.
After power is restored, the device automatically restarts with the
same parameters to avoid damage to the cultures due to an uninten-
tional termination of the process. The alarm message indicates that
the process has been interrupted. However, it is not possible to deter-
mine how long the interruption lasted.
Rectifying Faults
Open
The device door is open. When the door is opened, the device is au-
tomatically stopped using the greatest possible breaking effect.
Error 1, Sensor
The Pt100 sensor is not returning any readings.
Error 1, Fan 1 - 4
One of the fans is blocked or defective.
Rectifying Faults
Error, Heat
The control of the motor has overheated because of excessively
high rotation speed or excess loading. Once the device has cooled
down it can be restarted.
Error, Blocked
The motor or tray is blocked. Remove any foreign objects from the
incubation chamber, if necessary, dismount the table (see chapter
9.2.1 "Cleaning", page 129).
Error, Belt
The transmission belt is torn and has to be replaced.
Error, Control
The rotation speed of the shaker drive is more than 50 min-1 higher
than the maximum possible speed set in the settings menu.
Error, Sensor
The humidity sensor is not returning any readings (humidity sensor
not connected or defective).
Error, H2O
The water supply for humidification is interrupted (empty water
tank or defective hose).
Error 1, Heat
The heating or temperature recording of the steam humidifier is de-
fective.
Error 2, Heat
The heating or temperature recording of the steam humidifier is de-
fective.
Rectifying Faults
Error, Sensor
The CO2 sensor is not returning any readings (CO2 sensor not con-
nected or defective).
Error, Control
The actual CO2 value does not rise (CO2 supply interrupted, pressure
too low or door not closed correctly).
Error, Program
Due to a new program version, problems with EEProm or problems
with one of the internal data buses, EEProm was initialised.
After this fault, it might be necessary to recalibrate temperature
measuring again.
Error, Lock
The message appears when the key pad is locked with a PIN. If the
message appears although the PIN function is not active, the key
pad cannot be initialised. If the problem is caused by one of the in-
ternal data buses, the mainboard or HMI unit must be replaced.
Rectifying Faults
The following tables describe possible faults for which usually no er-
ror message appears on the display and for which no alarm is
sounded, a few exceptions aside. In many cases operators can resolve
these themselves.
General faults
Fault
After activating the power switch, the display and the power switch do not light up.
Possible cause Measures Personnel
Power supply of the device is Check if the plugs are plugged in correctly. Operator
interrupted.
Check the mains connection.
The fuse of the device (230 V Change the fuse respectively reset the thermal pro- Operator
version) is defective or the ther- tection switch (see chapter 8.3, page 127). If the
mal protection switch (115 V fault occurs more than once, contact your local IN-
version) was triggered. FORS HT representative.
Fault
The tray cannot be locked.
Possible cause Measures Personnel
The tray is blocked by foreign Remove foreign objects from the table. Operator
objects.
The tray is bent. Replace the tray. Operator
Fault
The interior lighting does not work.
Possible cause Measures Personnel
Interior lighting is deactivated. Activate interior lighting (see chapter 7.7.7, page Operator
114).
The LED unit of the interior Contact your local INFORS HT representative. INFORS HT service tech-
lighting is defective. nician or licenced dealer
Rectifying Faults
Fault
Strong vibrations occur.
Possible cause Measures Personnel
The load is distributed une- Load the tray in the centre. If possible, do not Operator
venly. place any heavy weights in the corners of the tray.
The rotation speed is too high. Reduce the rotation speed. Operator
The device is not standing Level the table or the device (adjustable foot on Operator
straight. the base).
The substructure is too weak. Place the device on a stable floor. Operator
Fault
The shaker drive accelerates too quickly or runs too fast.
Possible cause Measures Personnel
Fault in measuring the rotation Switch the device off and then on again. Operator
speed.
The motor control is defective. Contact your local INFORS HT representative. INFORS HT service tech-
nician or licenced dealer
Rectifying Faults
Fault
The temperature does not reach the desired setpoint.
Possible cause Measures Personnel
The heating is not switched on. Activate the temperature parameter. Operator
The setpoint is set too low. Raise setpoint. Operator
The ambient temperature does Check the ambient temperature and correct it if Operator
not meet the requirements. necessary (see chapter 6.1, page 72
Door not closed completely. Close door completely. Operator
The heating fans do not work. Contact your local INFORS HT representative. INFORS HT service tech-
nician or licenced dealer
INFORMATION
The re-measuring of the temperature only provides reliable infor-
mation when calibrated measuring tools are used and only if they
are used at points specified by INFORS HT. A measurement at un-
defined positions in the casing will not produce any usable data.
For information about temperature measuring, contact your local
INFORS HT representative or request a quote for calibrating the pa-
rameters.
Rectifying Faults
Fault
CO2 concentration too high (deviation < 1%).
Possible cause Measures Personnel
CO2 input pressure too high. Reduce input pressure. Operator
Fault
CO2 value fluctuates, actual value is not constant.
Possible cause Measures Personnel
Inlet pressure or flow too high. Decrease inlet pressure or flow. Operator
Fault
The humidity parameter was deactivated automatically.
Possible cause Measures Personnel
Transfer error of the humidity Restart software (switch the device off and then on Operator
sensor or temperature sensor. again).
Inlet pressure is too high. Reduce inlet pressure (max. 0.3 bar). Operator
Steam generator is defective. Contact your local INFORS HT representative. INFORS HT service tech-
nician or licenced dealer
Rectifying Faults
INFORMATION
The fuses may only be replaced by fuses of the same rating. For in-
formation concerning the requirements for the fuses see chapter
12.3 "Specifications of the Basic Unit", page 140.
Procedure 1. Turn off the device and pull out the mains plug.
2. Unlock the plug for the fuses by pressing together the two flaps
and pull out the plug at the same time.
3. Remove the defective fuse.
4. Insert a new fuse with the correct number of Amperes.
5. Push the plug as far back in the opening as possible until it
snaps in.
6. Re-establish the power supply to the device.
The provider must return the device or the faulty component part(s)
to the manufacturer if, after consulting the service department of the
local dealer or the manufacturer, on-site diagnosis and/or repair is
not possible.
INFORMATION
When returning the device, the component part or accessory for
repair, it is required for the safety of all parties involved and be-
cause of legal provisions that a lawful declaration of decontamina-
tion is present. Refer to main chapter “Safety and Responsibility”,
chapter “Declaration of Contamination” for details.
WARNING
To prevent life-threatening electric shocks, always switch off the
device and disconnect it from the power supply before carrying
out any maintenance or cleaning.
9.1 Maintenance
The device requires hardly any maintenance. This reduces the running
costs to certain regular checks and cleaning. The following table de-
scribes the maintenance work that is required to ensure optimum,
fault-free operation.
ATTENTION
Insufficient cleaning and disinfection can lead to damage to cul-
tures due to contamination.
9.2.1 Cleaning
Detergent
ATTENTION
Aggressive cleaning agents, solvents and abrasive cleaning utensils
(hard sponges, brushes) can scratch surfaces, damage the device
and impair its function.
Notices on cleaning
To clean the surfaces, use a soft cloth, ideally lint-free. This applies in
particular to the front window. Disinfect using commercially available
disinfectant if necessary.
Splashing water
Always use a wet cloth when cleaning the base tray; never pour wa-
ter into the tray. Make sure that no water splashes into the bearings.
After cleaning the device, especially the interior and the base tray, dry
it with a cloth.
Dust
Over time, dust can collect in the vents and on other exposed parts. It
can be removed carefully using a vacuum cleaner.
CAUTION
After being folded up, the table must be held in position by hand.
If the table is not secured properly, there is a risk of it folding back
down again. This can cause injuries.
– If possible, have a second person hold the table in place once
it has been folded up. Alternatively, you can place an object
under the table to stop it from folding back down.
– Proceed with caution when folding the table back down.
INFORMATION
If any of the screws are lost or damaged, make sure that they are
only replaced with original screws (M6x12).
9.2.2 Disinfection
Only use quaternary ammonium compounds for wipe-down disinfec-
tion. As a tried as tested disinfectant, we recommend Fermacidal D2.
ATTENTION
Heat (temperatures above 80 °C), aggressive disinfectants such as
chlorine bleach and UVC radiation can damage the device and sig-
nificantly limit the function and useful life of the machine.
The inbound delivery and transport to the assembly location are per-
formed only by INFORS HT employees or by persons authorised by
INFORS HT.
10.1 Transport
WARNING
Improper transport, the use of incorrect auxiliary equipment and
careless handling of the device may lead to injuries and severe
property damage.
10.2 Storage
INFORMATION
When returning the device for disassembly or disposal, it is re-
quired for the safety of all parties involved and because of legal
provisions that a lawful declaration of decontamination is present.
Refer to main chapter “Safety and Responsibility”, chapter “Decla-
ration of Contamination” for details.
11.1 Disassembly
Prior to disassembly:
Switch off the device and lock any isolation switch in the ‘off’
position.
Physically disconnect the main energy supply from the device
and wait for components to fully discharge.
Remove and dispose of all additional consumable items, auxiliary
components and/or spent processing material in an environmen-
tally acceptable manner.
11.2 Disposal
WARNING
Electronic waste, electronic components, lubricants or other auxil-
iary materials/supplies are subject to hazardous waste regulations
and may only be disposed of by registered specialist disposal firms.
Two units stacked with high base frame and the following options:
Low cover (1)
CO2 control (2)
Analog output (3)
Direct steam humidification (4)
1
3
2
4
Three units stacked with low base frame and the following options:
Top cooling (1)
CO2 control (2)
Analog output (3)
Direct steam humidification (4)
2 4
All dimensions in mm
12.2 Connections
1 2
3
4
5
1 Coolant in (hose connector DN10) / max. 4 bar 3 Demineralised water in (hose connector DN06) /
2 Coolant out (hose connector DN10) / pressure- max. 0.3 bar
less 4 Discharge outlet (female thread G1/4‘‘)
5 CO2 in (hose connector DN04) / max. 0.5 bar
All dimensions in mm
Weight
Description Value Unit
Single unit (3 mm throw) 129 kg
Single unit (25 mm throw) 137 kg
Single unit (50 mm throw) 144 kg
Single unit (adjustable throw) 143 kg
Base frame 13 cm 26 kg
Base frame 41 cm 45 kg
Alarm connection
Description Value Unit
Type Stereo jack, 3.5 mm
Relais Rectifier NO / NC
max. 1A 34V AC/DC
Ethernet connection
Description Value Unit
Type RJ45
Data transmission 10/100 Mbps Ethernet
Discharge outlet
Description Value Unit
Hose nozzle ¼ inch
Hose diameter 10 mm
12.3.5 Material
Component Material
Casing Polyurethane (PUR-IHS)
with foam protection
Door PUR-IHS, safety glass
Cover plate temperature control Stainless steel (1.4301-2B)
Table Aluminium, anodised
12.3.6 Emissions
INFORMATION
The specified temperature range refers to the temperature directly
on the device. In case of heat build-up due to insufficient ventila-
tion, the temperatures on the device can be significantly higher
than the room temperature.
Description Value
IP protection type 20 (as per DIN 60529)
ATTENTION
Using the wrong auxiliary materials can result in significant dam-
age to property.
Only use the auxiliary materials in accordance with the table be-
low.
Rotation speeds
INFORMATION
The actual rotation speed that can be achieved depends on differ-
ent factors, such as loading, vessel type (e.g. flasks with baffles) or
holder (e.g. clamps or Sticky Stuff).
The minimum rotation speed for all device variants is 20 min-1. The
maximum rotation speed of a unit depends on the throw and the
position of the unit in the stack:
Individual unit 3 mm 25 mm 50 mm
1000 min-1 400 min-1 350 min-1
Electrical data
Description Type 230 V Type 115 V Unit
Value
Max. power consumption 1
540 540 W
Max. current consumption 4.2 8.4 A
Fuses (5 x 20 mm, slow-blown) 10 - A
Thermal protection switch - 15 A
1
) Power consumption of the compressor
Various
Electrical data
Description Type 230 V Type 115 V Unit
Value
Max. power consumption 1 1080 1080 W
Max. current consumption 8.4 16.8 A
Fuses (5 x 20 mm, slow-blown) 10 - A
Thermal protection switch - 15 A
1
) Power consumption of the compressor
Various
ATTENTION
The installed valve regulates the temperature by closing and open-
ing the cycle. Hence it can be necessary to install a bypass from
the inlet to the outlet to protect the circulation pump against dam-
age.
Casing dimensions
Description Value Unit
Height 340 mm
Depth 80 mm
Width 200 mm
Various
Description Value Unit
Additional mass 3 kg
Water consumption 1 8.3 g/h
Temperature range for use 2
Up to 40 °C
Heating time of the unit 5 min
1
) At an ambient temperature of up to 20 °C and at a temperature of 37 °C and
relative humidity of 85 % in the incubation chamber.
2
) Temperature in the interior at which functioning can be guaranteed
Electrical data
Description Value Unit
Voltage 230 V
Max. power consumption 125 W
Max. current consumption 0,4 A
Bar heater (in door) 24 V
15 W
Door seal heating 24 V
17 W
Floor heating 24 V
17 W
ATTENTION
Using tap water can quickly lead to limescale in the vaporiser of
the humidification unit, which would affect correct functioning.
Ultra pure water (WFI, water for injection purposes) must not be
used either as this is highly corrosive and thus damages the device.
Dimensions
Description Value Unit
Depth 185 mm
Height 337 mm
INFORMATION
For detailed information on the external dimensions, refer to the
dimension drawings (see chapter 4.3.4 "Dimensions", page 47).
Connections
Description Value
Water in Tri-Clamp port DN 8
Inlet pressure: max. 6 bar
Connection data
Description Value Unit
Connection (hose Ø) 3 to 4 mm
Type of connection Legris
Input pressure 0.5 bar
Various
Description Value Unit
Additional mass 0.5 kg
12.5.6 UV Decontamination
Additional mass 4 kg
1
) Temperatur in incubation chamber
Index
14 Index
A B
Index
Index
F use ................................................................. 55
Fans ................................................................... 28 Humidification ................................................... 39
Fault notifications ............................................ 122 Humidity sensor
Faults ......................................................... 88, 120 position .......................................................... 40
CO2 control .................................................. 128 specification ................................................. 151
direct steam humidification .......................... 128 I
general faults ....................................... 122, 125 Identification of the device ................................. 34
shaker drive .................................................. 126 Identification of the throw ................................. 34
temperature control ..................................... 127 Identification plate
Fernbach flasks, requirements ............................ 11
content ............................................................. 9
Flasks ................................................................. 11 position .......................................................... 34
Floor heating ..................................................... 39 Installation ......................................................... 72
Frame Installation location ............................................ 72
base, 13 cm ................................................... 32 Intended use ...................................................... 10
base, 41 cm ................................................... 33 Interior dimensions ........................................... 142
Functional description
Interior lighting
analog output ................................................ 53
configure ...................................................... 116
CO2 control .................................................... 48
functional description ..................................... 32
cooling ........................................................... 35
specification ................................................. 144
direct steam humidification ............................ 39
Intermediate base .............................................. 51
external cooling .............................................. 37
Internal cooling
interior lighting .............................................. 32
pressure elimination unit ................................ 42 specification ................................................. 147
temperature control ....................................... 28 Interrupted power supply ................................. 119
Fuses K
position .......................................................... 18 Key tone activate/deactivate ............................. 115
replace ......................................................... 129 L
type .............................................................. 143 LED lighting
G operate ........................................................... 52
Guidelines ............................................................ 9 specification ................................................. 153
H Lighting
Heating ............................................................ 147 configure ...................................................... 116
Height above sea level ..................................... 117 functional description ..................................... 32
High ................................................... 92, 120, 121 specification ................................................. 144
Loading .............................................................. 83
Holders
Low ................................................... 92, 120, 121
fit ................................................................... 80
overview ......................................................... 65 M
screw sizes ..................................................... 81 M tray ................................................................ 57
Hose pass-through Main switch ....................................................... 33
position .......................................................... 54 Mains connection
Index
Microtitre box P
Index
Index
identification .................................................. 34 V