Yokogawa EJA110E General Specifications

Download as pdf or txt
Download as pdf or txt
You are on page 1of 17

<<Contents>> <<Index>>

General EJA110E
Differential Pressure Transmitter

[Style: S2]

The high performance differential pressure transmitter


EJA110E features single crystal silicon resonant
sensor and is suitable to measure liquid, gas, or

pressure. EJA110E outputs a 4 to 20 mA DC signal


corresponding to the measured differential pressure.
Its accurate and stable sensor can also measure the

communications. Other key features include quick


response, remote set-up using communications and
self-diagnostics. F

requirement.

Refer to GS 01C31T02-01EN for Fieldbus


communication type and GS 01C31T04-01EN for
PROFIBUS PA communication type for the items
(includes terminal-based linearity, hysteresis, and
repeatability)

Measurement inH2O mbar mmH2O


kPa ±0.055% of Span
Span/Range (/D1) (/D3) (/D4) Reference
Span 0.5 to 5 2.0 to 20 5 to 5050 to 500 accuracy X > span ±(0.005+0.02 URL/span)% of Span
F* -500 to X 2 kPa (8 inH2O)
Range -5 to 5 -20 to 20 -50 to 50
500 URL
5 kPa (20 inH2O)
Span 0.5 to 10 2.0 to 40 5 to 100 50 to 1000 (upper range limit)
L* -100 to -1000 to
Range -10 to 10 -40 to 40
100 1000
100 to
Span 1 to 100 4 to 400 10 to 1000 Reference ±0.055% of Span
10000
M accuracy X > span ±(0.005+0.0025 URL/span)% of Span
-100 to -400 to -1000 to -10000 to
Range X 5 kPa (20 inH2O)
100 400 1000 10000
0.05 to 5 URL
Span 5 to 500 20 to 2000 50 to 5000 100 kPa (400 inH2O)
kgf/cm2 (upper range limit)
H
-500 to -2000 to -5000 to -5 to 5
Range
500 2000 5000 kgf/cm2
H
0.14 to 14 20 to 2000 1.4 to 140 1.4 to 140
Span Reference ±0.055% of Span
MPa psi bar kgf/cm2
V accuracy X > span ±(0.005+0.01 URL/span)% of Span
-0.5 to 14 -71 to -5 to 140 -5 to 140
Range
MPa 2000 bar kgf/cm2 X 100 kPa (400 inH2O)
*: F capsule is applicable for wetted parts material URL
500 kPa (2000 inH2O)
code S. (upper range limit)
L capsule is applicable for wetted parts material
code other than S.
V
Reference ±0.055% of Span
Zero-based calibrated span, linear output, wetted accuracy X > span ±(0.005+0.005 URL/span)% of Span
parts material code S and silicone oil, unless X 1.4 MPa (200 psi)
otherwise mentioned. URL
For Fieldbus and PROFIBUS PA communication 14 MPa (2000 psi)
(upper range limit)
types, use calibrated range instead of span in the

Yokogawa Electric Corporation GS 01C31B01-01EN


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright June 2012
Tel.: 81-422-52-5690 Fax.: 81-422-52-2018 21st Edition Mar. 2020
<<Contents>> <<Index>> 2

Reference ±0.04% of Span


accuracy X > span ±(0.002+0.0019 URL/span)% of Span 50% or Greater Same as reference accuracy
X 5 kPa (20 inH2O) Reference accuracy × 50
50% to Dropout point
Square root output (%)
URL
100 kPa (400 inH2O)
(upper range limit)

H
Reference ±0.04% of Span
F ±(0.08% Span + 0.18% URL)
accuracy X > span ±(0.005+0.0049 URL/span)% of Span M ±(0.07% Span + 0.02% URL)
X 70 kPa (280 inH2O) H ±(0.07% Span + 0.015% URL)
URL V ±(0.07% Span + 0.03% URL)
500 kPa (2000 inH2O)
(upper range limit)

V
Reference ±0.04% of Span F, M, H and V capsules
accuracy X > span ±(0.005+0.0013 URL/span)% of Span ±0.1% of span
X 500 kPa (2000 inH2O)
URL
14 MPa (2000 psi)
(upper range limit)
F ±(0.04% span+0.208% URL)
M, H, V ±0.028% URL

±0.20% of Span @1:1 to 5:1 Rangedown Overpressure condition: up to maximum working


Total probable error, known as a measure of the total pressure
performance of the transmitters under the condition of M, H and V capsules, except for gold-plated
diaphragm.
±0.03% of URL
Total Probable Error = ± E12 + E22 + E32

E1: Reference Accuracy of Calibrated Span M, H and V capsules


E2: Ambient Temperature Effects per 28°C change ±0.1% of URL per 10 years
E3: Static Span Effects per 6.9 MPa change F capsule
±0.2% of URL per one year

±0.17% of Span @1:1 Rangedown


±0.33% of Span @ 5:1 Rangedown
Total accuracy is a comprehensive measure of
transmitter total performance, covering all major Less than 0.1% of URL when tested per the
factors in actual installation, that cause errors in
measurement. with high vibration level (10-60 Hz, 0.21 mm
As a standard measure, YOKOGAWA uses this to displacement/60-2000 Hz 3 g)
evaluate transmitter performance.
Less than ±0.1% of URL when tested per the
Total Accuracy = ± E12 + E22 + (E3 + E4)2 + E52 application or pipeline with low vibration level (10-60
Hz 0.15mm displacement /60-500 Hz 2g)
E1: Reference Accuracy of Calibrated Span
E2: Ambient Temperature Effects per 28°C change
E3: Static Span Effects per 6.9 MPa change Rotation in diaphragm plane has no effect. Tilting
E4: Static Zero Effects per 6.9 MPa change up to 90 degree will cause zero shift up to 0.4 kPa
E5: Overpressure Effects upto overpressure 16MPa (1.6 inH2O) which can be corrected by the zero
adjustment.
Not only a day-to-day changes in temperature can
affect the measurement and lead to unnoticed errors; 90 ms for Wetted Parts material code S and L,
except for Measurement span code F. 150 ms for
Wetted Parts Material Code H, M, T, A, B, D and W or
events, and other phenomena can have the similar Measurement span code F.
result. Total Accuracy factors in such changes and
errors and provides much comprehensive and dead time of 45 ms (nominal)
practical determination of how a transmitter will
perform under actual plant operation.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN June 1, 2017-00
<<Contents>> <<Index>> 3

Zero can be fully elevated or suppressed, within the


lower and upper range limits of the capsule.

External zero is continuously adjustable with


Upper Range Value and Lower Range Value of the 0.01% incremental resolution of span. Re-range
static pressure can be set in the range between 0 can be done locally using the digital indicator with
and Maximum Working Pressure(MWP). The upper rangesetting switch.
range value must be greater than the lower range
value. Minimum setting span is 0.5 MPa(73 psi). 5-digit numerical display, 6-digit unit display and bar
Measuring either the pressure of high pressure side graph.
or low pressure side is user-selectable.
four of the following variables periodically.;
Absolute Pressure Measured differential pressure, differential pressure
1 MPa or higher: ±0.5% of span in %, scaled differential pressure, measured static
Less than 1 MPa: ±0.5%×(1 MPa/span) of span
Gauge Pressure Reference
Gauge pressure reference is 1013 hPa (1 atm)
Note: Gauge pressure variable is based on the above

the change of atomospheric pressure.


adjustment screw and push button (Integral indicator
code E) offers easy and quick setup for parameters
of Loop test, Tag number, Unit, LRV, URV, Damping,
Output mode (linear/square root), Display out 1, and
Re-range by applying actual pressure (LRV/URV)
and Device Information.

Two wire 4 to 20 mA DC output with digital 69 MPa (10,000 psi) for wetted parts material S,
communications, linear or square root programmable. except for Measurement span F.
BRAIN or HART FSK protocol are superimposed on 47 MPa (6,800 psi) for wetted parts material other
the 4 to 20 mA signal. than S or Measurement span F.
Output range: 3.6 mA to 21.6 mA
Output limits conforming to NAMUR NE43 can be
pre-set by option code C2 or C3.
and over-range error for differential pressure, static
pressure and capsule temperature.
Three or four wire low power 1 to 5 V DC output with differential pressure and static pressure is also
HART, linear or square root programmable. available.
HART protocol are superimposed on the 1 to 5 V DC
signal.
Output range: 0.9 V to 5.4 V DC
for 4 to 20 mA output.

EJA-E series transmitters except Fieldbus,


Analog output status at CPU failure and hardware PROFIBUS PA and 1-5V DC with HART (Low Power)
error;
Up-scale: 110%, 21.6 mA DC or more (standard) the following standards;
IEC 61508: 2010;
Functional Safety of Electrical/electronic/
programmable electronic related systems; SIL 2
capability for single transmitter use, SIL 3 capability
Analog output status at CPU failure and hardware for dual transmitter use.
error; Reliability Data different depending on hardware and
Up-scale: 110%, 5.4 V DC or more (standard) software revision.
For details, refer to Functional Safety Data Sheet.
(Document number: TI 01C25A05-01EN or TI
01C25A05-21EN for option code SLT )
0.00 to 100.00 s by software and added to response The document can be downloaded from the website
time. of Yokogawa.
(Website address: https://www.yokogawa.com/
damping is set to less than 0.5 s, communication
may occasionally be unavailble during the
operation, especially while output changes
dynamically. The default setting of damping
ensures stable communication.

Differential pressure: 45 ms
Static pressure: 360 ms

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 4

0 to 100% RH
10.5 to 32 V DC for lightning protector
(option code /A.)
10.5 to 30 V DC for intrinsically safe, type n, or
16 MPa (2300 psi) nonincendive.
25 MPa (3600 psi) for option code /HG Minimum voltage limited at 16.6 V DC for digital
communications, BRAIN and HART
See graph below

Power supply :
100(14.5) Atmospheric
pressure Power Consumption :
Working
0.96 mA to 3 mA, 27 mW
pressure
kPa abs
(psi abs)
Applicable range
10(1.4)

2.7(0.38) Note that with three-wire connection, the cable length


may affect the measurement accuracy of the output
signal.
1(0.14)
-40 0 40 80 120
(-40) (32) (104) (176) (248) (Approval codes may affect electrical requirements.)
Process temperature °C (°F)
F01E.ai
Up to 2 km (1.25 miles) when using CEV
polyethylene-insulated PVC-sheathed cables.
Communication distance varies depending on type of
cable used.

0.22 µF or less

3.3 mH or less
used. See graph below.

EN 61326-1 Class A, Table2


600
EN 61326-2-3
E-10.5
R=
External 0.0244 Digital
load Communication
resistance range
BRAIN and HART Sound Engineering Practice (for all capsules)
250 With option code /PE3 and /HG (for M, H and V
capsules and wetted parts material code S.)
Category III, Module H, Type of Equipment: Pressure
R
Accessory-Vessel, Type of Fluid: Liquid and Gas,
Group of Fluid: 1 and 2
10.5 16.6 25.2 42
EN 50581
F02E.ai

EN 61010-1, C22.2 No.61010-1

(Anticipated transient overvoltage 330 V)

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 5

FieldMate Versatile Device Management Wizard:


Refer to GS 01R01A01-01E.
BRAIN TERMINAL: Refer to GS 01C00A11-00E
Power Distributor: Refer to GS 01B04T01-02E or
GS 01B04T02-02E

Fluorinated rubber for option code N2 and N3 ; Registered trademark of Yokogawa


Electric Corporation.

Electric Corporation.
B7 carbon steel, 316L SST or 660 SST

®
: Registered trademark of FieldComm
Group.
OUNDATION Fieldbus; Trademark of FieldComm
Group.

Nutzerorganisation e.v., Karlsruhe, Germany.


[for aluminum housing] Other company names and product names used in
Polyester resin powder coating this material are registered trademarks or trademarks
Mint-green paint (Munsell 5.6BG 3.3/2.9 or its of their respective owners.
equivalent)
[for option code /P or /X2]
Epoxy and polyurethane resin solvent coating

IP66/IP67, Type 4X

316 SST

[Installation code 7, 8 and 9]


2.8 kg (6.2 lb) for measurement span code M, H and
V, wetted parts material code S and L without integral
indicator, mounting bracket, and process connector.
3.7 kg (8.2 lb) for measurement span code F without
integral indicator, mounting bracket, and process
connector.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Mar. 25, 2019-00
<<Contents>> <<Index>> 6

Differential pressure transmitter


Output signal 4 to 20 mA DC with digital communication (BRAIN protocol)
4 to 20 mA DC with digital communication (HART 5/HART 7 protocol)*1
Digital communication (FOUNDATION Fieldbus protocol, refer to
GS 01C31T02-01EN)
Digital communication (PROFIBUS PA protocol, refer to
GS 01C31T04-01EN)
Low Power, 1 to 5 V DC with digital communication (HART 7 protocol)
Measurement 0.5 to 5 kPa (2.0 to 20 inH2O) (For Wetted parts material code S)
span (capsule) 0.5 to 10 kPa (2.0 to 40 inH2O)
(For Wetted parts material code M, H, T, A, D, B and W)
1 to 100 kPa (4 to 400 inH2O)
H 5 to 500 kPa (20 to 2000 inH2O)
V 0.14 to 14 MPa (20 to 2000 psi)
Wetted parts
material *2
Process connections
See the table in the next with Rc1/4 female process connector
page for the codes for a with Rc1/2 female process connector
diaphragm seal system. with 1/4 NPT female process connector
with 1/2 NPT female process connector

Bolts and nuts materia B7 carbon steel


G 316L SST
660 SST
Installation Vertical piping, left side high pressure, and process connection downside
Horizontal piping and right side high pressure
Horizontal piping and left side high pressure
Bottom Process Connection, left side high pressure*3
*3

Cast aluminum alloy


Cast aluminum alloy with corrosion resistance properties*5
ASTM CF-8M stainless steel*6
Electrical connection G1/2 female, one electrical connection without blind plugs
1/2 NPT female, two electrical connections without blind plugs
M20 female, two electrical connections without blind plugs
G1/2 female, two electrical connections and a blind plug*7
1/2 NPT female, two electrical connections and a blind plug*7
M20 female, two electrical connections and a blind plug*7
G1/2 female, two electrical connections and a SUS316 blind plug
1/2 NPT female, two electrical connections and a SUS316 blind plug
M20 female, two electrical connections and a SUS316 blind plug
Integral indicator Digital indicator*8
Digital indicator with the range setting switch (push button)*9
None

304 SST or SCS13A 2-inch pipe mounting, L type (for vertical piping)

K 316 SST or SCS14A 2-inch pipe mounting, L type (for vertical piping)
316 SST or SCS14A 2-inch pipe mounting (for bottom process connection type)
None
Optional Codes

*1: HART 5 or HART 7 is selectable. Specify upon ordering.


*2:

to plant facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken diaphragm

hypochlorite, and high-temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the
wetted parts material.
*3: Only applicable for Wetted parts material code S.
*4: Not applicable for measurement span code F.
*5: Not applicable for electrical connection code 0, 5, 7, 9 and A.
*6: Not applicable for electrical connection code 0, 5, 7 and 9.
*7: Material of a blind plug; aluminum alloy for code 5 and 9, and SUS304 for code 7.
*8: Not applicable for output signal code G.
*9: Not applicable for output signal code F.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN June 1, 2017-00
<<Contents>> <<Index>> 7

# Hastelloy C-276 *2 (Diaphragm)


ASTM CF-8M *1 316 SST
F316L SST, 316L SST (Others)
# Hastelloy C-276 *2 (Diaphragm)
ASTM CF-3M *7 316L SST
F316L SST, 316L SST (Others)
# ASTM CF-8M *1 Hastelloy C-276 *2 316 SST
# ASTM CF-8M *1 Monel 316 SST
ASTM CF-8M *1 Tantalum 316 SST
# Hastelloy C-276 equivalent *3 Hastelloy C-276 *2 Hastelloy C-276 *2
Hastelloy C-276 equivalent *3 Tantalum Hastelloy C-276 *2
# Monel equivalent *4 Monel Monel
# Super Duplex SST equivalent *5 Hastelloy C-276 *2 Super Duplex SST *6
*1: Cast version of 316 SST. Equivalent to SCS14A.
*2: Hastelloy C-276 or ASTM N10276.
*3: Indicated material is equivalent to ASTM CW-12MW.
*4: Indicated material is equivalent to ASTM M35-2.
*5: Indicated material is equivalent to ASTM A995 Grade5A.
*6: ASTM S32750 or EN 10272 1.4410.
*7: Cast version of 316L SST. Equivalent to SCS16A.

Please refer to the latest standards for details. Selected materials also conform to NACE MR0103.

The table below shows the codes dedicated for the combination with a diaphragm seal system. They are only
available when the transmitter is ordered in combination with a diaphragm seal system. Please also refer to GS
01C25W01-01EN.

With C80F , C81F , C82F or C70S With C80F , C81F , C82F or C70S
diaphragm seal diaphragm seal
With C80F , C82F or C70S diaphragm seal
With C80F , C82F or C70S diaphragm seal
With C80F , C82F or C70S diaphragm seal
With C80F , C82F or C70S diaphragm seal
With C80F , C81F or C82F diaphragm seal With C80F , C81F or C82F diaphragm seal
G
for high vacuum use for high vacuum use
With C20F or C30S direct mount seal
With C20F or C30S direct mount seal
With C20F or C30S direct mount seal With C80F or C70S diaphragm seal
With C80F compensation capillary system
With C20F direct mount seal
diaphragm seal
C80F , C81F , C82F , C20F , C70S and C30S

direct mount seal, C70SW or C70SE remote mount hygienic diaphragm seal, and C30SW or C30SE direct mount hygienic seal
respectively.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Sep. 27, 2018-00
<<Contents>> <<Index>> 8

For other agency approvals and marine approvals, please refer to GS 01C25A20-01EN.

Factory Mutual FM Explosionproof Approval *1


(FM) Applicable Standard: FM3600, FM3615, FM3810, NEMA 250, ANSI/UL 61010-1, ANSI/UL 61010-2-30
Explosionproof for Class I, Division 1, Groups B, C and D, Dust-ignitionproof for Class II/III, Division 1,
Groups E, F and G, in Hazardous locations, indoors and outdoors (Enclosure: Type 4X)

Temperature class: T6, Amb. Temp.: –40 to 60°C (–40 to 140°F)


FM Intrinsically safe Approval *1 *3
Applicable Standard: FM 3600, FM 3610, FM 3611, FM 3810, ANSI/ISA-60079-0, ANSI/ISA-60079-11,
ANSI/ISA-61010-1, NEMA 250

Enclosure: Type 4X, Temp. Class: T4, Amb. Temp.: –60 to 60°C (–75 to 140°F)
Intrinsically Safe Apparatus Parameters
[Groups A, B, C, D, E, F and G] Vmax=30 V, Imax=200 mA, Pmax=1 W, Ci=6 nF, Li=0 µH
[Groups C, D, E, F and G] Vmax=30 V, Imax=225 mA, Pmax=1 W, Ci=6 nF, Li=0 µH
Combined FF1 and FS1 *1 *3
ATEX ATEX Flameproof Approval *1
Applicable Standard: EN IEC 60079-0, EN 60079-1, EN 60079-31

II 2G, 2D Ex db IIC T6...T4 Gb, Ex tb IIIC T85°C Db


Degree of protection: IP66/IP67
Amb. Temp. (Tamb) for gas-proof :
T4; –50 to 75°C (–58 to 167°F), T5; –50 to 80°C (–58 to 176°F), T6; –50 to 75°C (–58 to 167°F)
Process Temp. for gas-proof (Tp):
T4; –50 to 120°C (–58 to 248°F), T5; –50 to 100°C (–58 to 212°F), T6; –50 to 85°C (–58 to 185°F)
Max. surface Temp. for dust-proof: T85°C (Tamb: –30 to 75°C, Tp: –30 to 85°C) *2
ATEX Intrinsically safe Approval *1 *3
Applicable Standard: EN 60079-0, EN 60079-11

II 1G, 2D Ex ia IIC T4 Ga, Ex ia IIIC T85°C T100°C T120°C Db


Degree of protection: IP66/IP67
Amb. Temp. (Tamb) for EPL Ga: –50 to 60°C (–58 to 140°F)
Maximum Process Temp. (Tp) for EPL Ga:120°C
Electrical data: Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=27.6 nF, Li=0 µH
Amb. Temp. for EPL Db: –30 to 60°C *2
Max. surface Temp. for EPL Db: T85°C (Tp: 80°C), T100°C (Tp: 100°C), T120°C (Tp: 120°C)
Combined KF22, KS21 and ATEX Intrinsically safe Ex ic *1 *3
[ATEX Intrinsically safe Ex ic]
Applicable Standard: EN 60079-0, EN 60079-11
II 3G Ex ic IIC T4 Gc, Amb. Temp.: –30 to 60°C (–22 to 140°F) *2

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Mar. 30, 2020-00
<<Contents>> <<Index>> 9

Canadian CSA Explosionproof Approval *1


Standards
Association Applicable Standard: C22.2 No. 25, C22.2 No. 30, CAN/CSA-C22.2 No. 94,
(CSA) CAN/CSA-C22.2 No. 61010-1, CAN/CSA-C22.2 No. 61010-2-030, CAN/CSA-C22.2 No. 60079-0,
CAN/CSA-C22.2 No. 60079-1, CAN/CSA-C22.2 No. 60529
Explosion-proof for Class I, Groups B, C and D.
Dustignition-proof for Class II/III, Groups E, F and G.

Temp. Code: T6...T4


Ex d IIC T6...T4 Enclosure: IP66/IP67
Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
–50 to 75°C(–58 to 167°F) for T6 *2

No additional sealing required


Primary seal failure annunciation: at the zero adjustment screw
CSA Intrinsically safe Approval *1 *3

[For Division System]


Applicable Standard: C22.2 No.0, C22.2 No.94, C22.2 No.157, C22.2 No.213, C22.2 No.61010-1,
C22.2 No.61010-2-030

Enclosure: Type 4X, Temp. Code: T4 Amb. Temp.: –50 to 60°C(–58 to 140°F) *2
Electrical Parameters: [Intrinsically Safe] Vmax=30V, Imax=200mA, Pmax=0.9W, Ci=10nF, Li=0 µH
[Nonincendive] Vmax=30V, Ci=10nF, Li=0 µH
[For Zone System]
Applicable Standard: CAN/CSA-C22.2 60079-0, CAN/CSA-E60079-11, CAN/CSA-E60079-15,
CAN/CSA-C22.2 No.60529
Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66/IP67
Amb. Temp.: –50 to 60°C(–58 to 140°F) *2, Max. Process Temp.: 120°C(248°F)
Electrical Parameters: [Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0 µH
[Ex nL] Ui=30V, Ci=10nF, Li=0 µH

No additional sealing required


Primary seal failure annunciation: at the zero adjustment screw
Combined CF1 and CS1 *1 *3
IECEx IECEx Flameproof Approval *1
Applicable Standard: IEC 60079-0, IEC60079-1

Flameproof for Zone 1, Ex d IIC T6...T4 Gb Enclosure: IP66/IP67


Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
–50 to 75°C(–58 to 167°F) for T6
IECEx Intrinsically safe and Flameproof Approval *1*3
Intrinsically safe Ex ia

Applicable Standard: IEC 60079-0, IEC 60079-11


Ex ia IIC T4 Ga Enclosure: IP66/IP67
Amb. Temp.: –50 to 60 °C(–58 to 140 °F), Max. Process Temp.: 120 °C(248 °F)

Intrinsically safe Ex ic

Applicable Standard: IEC 60079-0, IEC 60079-11


Ex ic IIC T4 Gc IP code: IP66
Amb. Temp.: –30 to 60°C(–22 to 140°F) *2, Max. Process Temp.: 120°C(248°F)

Flameproof

Applicable Standard: IEC 60079-0, IEC60079-1


Flameproof for Zone 1, Ex d IIC T6...T4 Gb Enclosure: IP66/IP67
Max.Process Temp.: T4;120°C(248°F), T5;100°C(212°F), T6; 85°C(185°F)
Amb.Temp.: –50 to 75°C(–58 to 167°F) for T4, –50 to 80°C(–58 to 176°F) for T5,
–50 to 75°C(–58 to 167°F) for T6
*1: Applicable for Electrical connection code 2, 4, 7, 9, C and D.

*3: Not applicable for output signal code Q.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Mar. 30, 2020-00
<<Contents>> <<Index>> 10

High accuracy type *1*20*28 High accuracy


High pressure-proof structure*28 Maximum working pressure for differential pressure measurement: 25MPa *18 HG
Painting Color change *2

Coating change Anti-corrosion coating*3


316 SST exterior parts 316 SST zero-adjustment screw and setscrews*4
Fluoro-rubber O-ring
Lightning protector Transmitter power supply voltage: 10.5 to 32 V DC (10.5 to 30 V DC for intrinsically safe type.)
Allowable current: Max. 6000 A (1×40 µs), Repeating 1000 A (1×40 µs) 100 times
Applicable Standards: IEC 61000-4-4, IEC 61000-4-5
Oil-prohibited use*5*28 Degrease cleansing treatment

Oil-prohibited use with Degrease cleansing and dehydrating treatment


dehydrating treatment*5*28

*28

Calibration units*6 P calibration (psi unit)


bar calibration (bar unit) (See Table for Span and Range Limits.)
M calibration (kgf/cm2 unit)
Plug option*23*24*28 Long vent*7: Total length: 119 mm (standard: 34 mm); Total length when combining with option
code K1, K2, K5, and K6: 130 mm. Material: 316 SST U1
Without vent and drain plugs
Gold-plated capsule gasket*8*28 Gold-plated 316L SST capsule gasket. Without drain and vent plugs. GS
Gold-plated diaphragm *9 Surface of isolating diaphragms are gold plated, effective *29

for hydrogen permeation. *28

Output limits and failure


operation*10
Failure alarm down-scale: Output status at CPU
NAMUR NE43 Compliant
Output signal limits:
3.8 mA to 20.5 mA*21 Failure alarm up-scale: Output status at CPU
failure and hardware error is 110%, 21.6 mA or more.
Body option*11*28 Right side high pressure, without drain and vent plugs
Terminal N1 and Process connection, based on IEC61518 with female thread on both sides of cover
Side

L H
F03E.ai

Wired tag plate 316 SST tag plate wired onto transmitter
*12 Software damping, Descriptor,
Message
Software damping
European Pressure PED 2014/68/EU
Equipment Directive*19*28 Category III, Module H, Type of Equipment: Pressure Accessory-Vessel,
Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2

*13*28 *14
*15
*14*30
*15*25

Vent screw, Capsule gasket*14*22*24

Diaphragm, Capsule body, Vent and Drain plug, Vent screw, Capsule gasket*15*22*24
Pressure test/ Test Pressure: 16 MPa(2300 psi) Nitrogen Gas*17
*16*28
Test Pressure: 25 MPa(3600 psi) for option code HG Retention time: one minute
Parameter list*26 List of setting and adjustment parameters
Functional safety(SIL)*27

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 11
*1: Applicable for capsule code M, H and V when combined with wetted parts material code S. Not applicable for /A2.

*5: Applicable for Wetted parts material code S, M, H and T.

D3, and D4.


*7: Applicable for vertical impulse piping type (Installation code 7) and Wetted parts material code S, H, M and T.
*8: Applicable for wetted parts material code S; process connection code 0 and 5; and installation code 8 and 9.
Not applicable for option code U1, N2, N3 and M11. No PTFE is used for wetted parts.
*9: Applicable for wetted parts material code S or L. /A2 is not applicable with FM approval.

*11: Applicable for wetted parts material code S, M, H and T; process connection codes 3, 4, and 5; installation code 9; and
mounting bracket code N. Process connection faces on the other side of zero adjustment screw.

*14: Applicable for process connections codes 0 and 5.


*15: Applicable for process connections codes 1, 2, 3, and 4.

*17: Dry nitrogen gas is used for oil-prohibited use (option codes K1, K2, K5, and K6).
*18: Applicable for Measurement span code M, H, or V; and wetted parts material code S or L. The URL of static pressure range
is also extended to 25 MPa.
*19: Applicable for measurement span code M, H and V and wetted parts material code S. If compliance with category III is
needed, specify this option code with HG.
*20: Not applicable for output signal code Q.
*21: The 1 to 5 V voltage output corresponding to 4 to 20 mA current output is applied to output signal code Q which is non-
compliant to NAMUR NE43.
*22: Not applicable with plug option code UN.
*23: Not applicable for installation code -U.
*24: Not applicable with option code N1, N2, N3 and GS.
*25: Applicable for option code UN and N1.
*26: Applicable for output signal code D and J.
*27: Not applicable for output signal code F, G, Q, process connections code for diaphragm seal system.
*28: Not applicable with process connections code for diaphragm seal system B, C, D, E, F, G, Q, R, P and T.
*29: Not applicable with process connections code for diaphragm seal system B, G, P, and T.
*30: Applicable for option code UN, N1, and GS.

The table below shows the codes dedicated for the combination with a diaphragm seal system. They are only
available when the transmitter is ordered in combination with a diaphragm seal system. Please also refer to GS
01C25W01-01EN

Degrease cleansing treatment

Degrease cleansing and dehydrating treatment

*1

*2

*1

*2

Capsule body*1

Capsule body*2

*1

*2

*1: Applicable with process connections code for diaphragm seal system C, D, Q, and R.
*2: Applicable with process connections code for diaphragm seal system E and F.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Mar. 30, 2020-00
<<Contents>> <<Index>> 12

Unit: mm (approx.inch)

Electrical connection
for code 5, 9, A, and D.
Process connector downside

54 (0.24)
(Installation code 7)

6
242(9.53) 110(4.33)

(2.13)
175(6.89)
12 39
97 129(5.08) 2-inch pipe
(3.82) (0.47) (1.54) (O.D. 60.5 mm)
Mounting bracket External indicator Process connector upside
(L-type,optional) Conduit connection
(2.05)

(optional) (Installation code 6)


52

Vent/Drain
Integral indicator plugs*6 Process
connection
(optional)
97(3.82)
223(8.78)

67(2.64)
ø78(3.07)

138(5.43)*2
95(3.74)

41
67(2.64)

223(8.78)
41

97(3.82)
Process
Conduit connector
connection (optional)
Process connection 54
Zero adjustment High Low

(2.05)
pressure (2.13) pressure

52
Ground terminal side*1 side*1

F04E.ai

Conduit
External indicator 89(3.50) 67 connection 110(4.33)
Conduit connection (2.64) Zero
adjustment 12 39
(optional) 95(3.74) Integral indicator
(0.47) (1.54)
(optional)
54
ø78(3.07)

(2.76)

(2.13)
ø70

6
175(6.89)
145(5.71)

129(5.08)

(0.24)
Vent plug, Ground
Drain plug*4*6 terminal
Vent plug*4*6
Electrical connection
for code 5, 9, A, and D. Vent plug*4*6
Drain plug*4*6
124(4.88)

41 Drain plug*4*6
54(2.13)
(1.61) Process Low
High
connection pressure
(1.63)

115(4.53) pressure
41.3

side*1 side*1
Process connector
(1.85)
47

Mounting bracket (optional) 2-inch pipe


(Flat-type,optional) (O.D. 60.5 mm)
130(5.12)*3 F05E.ai

(i.e. High pressure side is on the right side.)

*5: When electrical connection code 7 or C is selected, a blind plug is protruded up to 8 mm (0.31 inch) from the conduit connection.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 13
Unit: mm (approx.inch)

Electrical connection
for code 5, 9, A, and D.
Process connector downside

(0.24)
(Installation code 7)

6
256(10.10)
194(7.64) 110(4.33)

(2.13)
54
97 143(5.63) 9*4 39
2-inch pipe
(3.82) (0.35) (1.54)
(O.D. 60.5 mm)
Mounting bracket
(L-type,optional) External indicator Vent/Drain plugs*6 Process connector upside
Conduit connection
(2.05)

(Installation code 6)
(optional)
52

Process connection
Integral indicator
(optional)
102(4.02)
234(9.21)

72(2.83)

(1.81)
ø78(3.07)

148(5.83)*2

46
(3.74)

(2.76)
ø70
95
(1.81)

234(9.21)
72(2.83)
46

102(4.02)
Process
Conduit connector
connection (optional) 54
Process connection
Zero adjustment High (2.13) Low

(2.05)
Ground terminal pressure pressure

52
side*1 side*1

F06E.ai

External indicator Conduit connection


Conduit connection 94(3.70) 72(2.83) 110(4.33)
(optional) Zero
95(3.75) adjustment 9*4 39
Integral indicator
54 (optional) (0.35) (1.54)
ø78(3.07)

(2.13)
6
(2.76)
ø70

(0.24)
159(6.26)

143(5.63)
194(7.64)

Ground terminal
Electrical connection
for code 5, 9, A, and D. Vent plug*6
Vent plug*6
Drain plug*6
124(4.88)

46 Process connection Drain plug*6


(1.81) 54
High Low
125(4.92) Process connector pressure (2.13) pressure
(optional) side*1 side*1
(1.63)

2-inch pipe
41.3

(1.85)
47

(O.D. 60.5 mm)


Mounting bracket
(Flat-type,optional)

133*3
(5.24) F07E.ai

(i.e. High pressure side is on the right side.)

*4: 15 mm (0.59 inch) for right side high pressure.


*5: When electrical connection code 7 or C is selected, a blind plug is protruded up to 8 mm (0.31 inch) from the conduit connection.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 14
Unit: mm (approx.inch)

58 67 110(4.33)
(2.28) (2.64) Zero
External indicator
Conduit connection Conduit connection adjustment 12 39
95(3.74)
(optional) Integral indicator (0.47) (1.54)
54 (optional)

ø78(3.07)
(2.13)
6
145(5.71)

(0.24)

129(5.08)
175(6.89)
Ground terminal

Electrical connection
for code 5, 9, A, and D.
Vent plug
Drain plug
41 54 Drain plug
High Low
Vent plug (1.61) (2.13)
(1.63)

pressure pressure
41.3

115(4.53) side 130*1 side


Process connection
Drain plug (5.12)
Process connector
(optional)
F08E.ai

External indicator 63 72 Zero 110(4.33)


Conduit connection adjustment
Conduit connection (2.48) (2.83)
9 39
(optional) 95(3.74) Integral indicator
(0.35) (1.54)
54 (optional)
ø78(3.07)

(2.13)
6
(0.24)
159(6.26)

143(5.63)
194(7.64)

Ground terminal

Electrical connection
for code 5, 9, A, and D.
Vent plug
Drain plug
46 Drain plug
54
(1.63)

Vent plug (1.81)


41.3

High (2.13) Low


Process connection pressure pressure
125(4.92) side 133(5.24)*1 side
Drain plug Process connector (optional) F09E.ai

*1: When Option code K1, K2, K5, or K6 is selected, add 30 mm (1.18 inch) to the value.
*2: When electrical connection code 7 or C is selected, a blind plug is protruded up to 8 mm (0.31 inch) from the conduit connection.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 15
Unit: mm (approx.inch)

54
(2.13) 95(3.74) 188(7.40)
6
Integral indicator
(0.24) 73 110(4.33)
(optional) Zero
(2.87)
adjustment 12 39
Conduit connection (0.47) (1.54)
Electrical connection

ø78(3.07)
for code 5, 9, A, and D.

109(4.29)
159(6.26)
Extenal indicator
Conduit connection
(optional)
Ground terminal
(2.36)

Mounting bracket*1
60

(optional) Vent plug*4

2-inch pipe Process connector 54


(O.D. 60.5 mm) (optional) High (2.13) Low
pressure pressure
Process connection side 130(5.12)*2 side
129(5.08)
F10E.ai

192(7.56)
95(3.74) Integral indicator
110(4.33)
(optional) Zero
Extenal indicator 9 39
Conduit connection 95(3.74) Conduit connection adjustment
(0.35) (1.54) Vent plug*4
(optional)
ø78(3.07)

54
159(6.26)

(2.13)

143(5.63)
6
238(9.37)

(0.24) Ground terminal

Electrical connection Vent plug*4


79(3.11)

for code 5, 9, A, and D.

2-inch pipe Process connector 54


(O.D. 60.5 mm) (optional) High Low
(2.13)
pressure pressure
Mounting bracket side 133(5.24)*2 side
(optional)
129(5.08)
41.3
(1.63) F11E.ai

*1: A transmitter with SST housing is not applicable for mounting to horizontal 2-inch pipe.

*3: When electrical connection code 7 or C is selected, a blind plug is protruded up to 8 mm (0.31 inch) from the conduit connection.

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 20, 2019-00
<<Contents>> <<Index>> 16

+ 1
Power supply and output terminals
SUPPLY – 2

CHECK + 3
– 2
External indicator (ammeter) terminals*1*2

Ground terminal

*1: When using an external indicator or check meter, the internal

*2: Not available for FOUNDATION Fieldbus and PROFIBUS PA


Terminal 1 Terminal 3 communication types.
Terminal 2

+ 1
Power supply terminals
SUPPLY – 2

VOUT + 3 1 to 5 V DC with HART communication


– 2 terminals
Ground terminal
Three or four wire connection. For four wire connection, both supply
and signal lines use SUPPLY - terminal.
F12E.ai

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN Nov. 7, 2016-00
<<Contents>> <<Index>> 17

Specify the following when ordering


Tag number
2. Calibration range and units Software
damping *1
Output mode
any decimal point) for low or high range limits Calibration range
within the range of -32000 to 32000. When lower range value
reverse range is designated, specify Lower Calibration range
Range Value(LRV) as greater than Upper Range upper range
Value(URV). When square root output mode is value
Selected from mmH2O, mmH2O(68°F),
2) Specify only one unit from the table, ‘Factory mmAq*2, mmWG*2, mmHg, Pa, hPa*2,
Calibration range kPa, MPa, mbar, bar, gf/cm2,
3. Select linear or square root for output mode and unit kgf/cm2, inH2O, inH2O(68°F), inHg, ftH2O,
ftH2O(68°F) or psi. (Only one unit can be
display mode.
linear mode. Designated differential pressure value
4. Display scale and units (for transmitters equipped Display setting
with the integral indicator only)

engineering units scale: Static pressure ‘0 to 16 MPa (25 MPa*3


display range Measuring high pressure side.

*1: To specify these items at factory, option code


point) for low or high range limits within the range or is required.
of -32000 to 32000. Unit display consists of 6-digit, *2: Not available for HART protocol type.
*3: Applicable for option code HG.

be displayed on the unit display.


5. HART PROTOCOL
ASTM JIS
316 SUS316
6. TAG NO (if required) F316 SUSF316
316L SUS316L
BRAIN, 22 characters for HART, and 16 characters
F316L SUSF316L
for /N4 tag) are engraved on the stainless steel tag
304 SUS304
7. SOFTWARE TAG (for HART only. if required) F304 SUSF304
660 SUH660
*1 B7 SNB7
CF-8M SCS14A
alphanumeric capital letters.

*1 (22 characters) in the


EU WEEE (Waste Electrical and Electronic Equipment)
*1: applicable only when HART 7 is selected. Directive is only valid in the EU.
This instrument is intended to be sold and used only
Specifying option code or will allow further as a part of equipment which is excluded from WEEE
Directive, such as large-scale stationary industrial
items and setting range.
[/CA : For HART communication type] therefore, subjected to the exclusion from the scope
1) Descriptor (up to 16 characters) of the WEEE Directive. The instrument should be
2) Message (up to 30 characters) disposed of in accordance with local and national
3) Software damping in second (0.00 to 100.00) legislation/regulations.
[/CB : For BRAIN communication type]
1) Software damping in second (0.00 to 100.00)

All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation GS 01C31B01-01EN April 10, 2018-00
Subject to change without notice.