Dig Multimeter 079
Dig Multimeter 079
Dig Multimeter 079
18. 19. 22. 24, Rachana Industrial Complex, Plot No.71/1B/14, Telco Road, Indrayani Corner,
General Block. MIDC, Bhosari, Pune -411026 Mob 8975896690,
Mail pims pune@gmail com, info@prajyo.com, Web. wwwprajyo.com
Calibration of Electro- Technical, Mechanical, Thermal, Mass, Torque, Power, HV, CT/PT
CC501
CALIBRATION CERTIFICATE
Job No. P1220830-15 Calibration Date:30/08/2022 Issue Date. 02/09/2022 Page No.:I of 2
1. CALIBRATED FOR
Gemini Instratech Lid Calibration Location Site
B-83. MIDC, Ambad Nashik Maharashtra India 422010 Cond. On Receipt Satisfactory
6. CALIBRATION RESULTS:
CALIBRATION RESULTS:AC Voltage@50 Hz Discipline:Group - Electro-Technical: Alternative Voltage
Sr. No. Range Set Value on Average Reading Error Expanded Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)
N Rdg
0.3 @50Hz 0.301 0.001 +0.19
3V 1.5 @50Hz_ 1.497 -0.003 +0.054
3.0 @501Hz 2.996 -0.004 0.054
30 V 30@5011z 29.97 -0.03 +0.054
300V 300 @50Hz 299.6 -0.4 +0.054
1000V 1000@501Hz 998 2 +0.078
CALIBRATION RESULTS:DC Voltage Discipline:Group-Electro-Technical:Direct Voltage
Sr. No. Range Set Value on Average Reading Error Expanded Coverage Factor
Master On UUC Uncertainty (% (k)
Rdg)
3 mV 2.99mV -0.01mV t0.2
300 mV
15mV 15.02mV 0.02mV t0.04
30mV 29.95 mV -0.05mV_ +0.02
300 mV 299.91 mV 0.09 mV +0.004
3V 3V 2.998 V -0.002 V 0.019
30 V 30 V 29.97V -0.03 V +0.019
300 V 300 V 299.5V -0.5V +0.019
8 1000V 1000 998 -2 V +0.058
is
Fgauppa PIPL
Approved By: S.E.Shinde/
Calibrated By: Rahul Dora D.S.Deshpande / R.B.Kasar
Calibration Engineer d Technical Manager
F: CP-T/05
Prajyo Instrument Pvt. Ltd.
PIMS (Formally: Prajyo Instrument &Maintenance Services)
18, 19, 22, 24, Rachana Industrial Complex, Plot No. 71/1B/14, Telco Road, Indrayani Corner
General Block, MIDC, Bhosari, Pune -411026. Mob.: 8975896690,
Mail pimspune@gmail.com,info@prajyo.com,Web: www.prajyo.com
Calibration of: Electro - Technical, Mechanical, Thermal, Mass, Torque, Power, HV, CT/PT
CC
2503
Job No.:P1220830-15 Calibration Date:30/08/2022 Issue Date.:02/09/2022 Page No.:2 of 2
F (ohpau iP
Calibrated By: Rahul Dora
Calibration Engineer Approved By: S.E.Shinde/
D.S.Deshpande/ R.B.Kasar
Technical Manager
End of Cert1ficate---- F: CP-T/05