Instruction Manual for PT650M Series 2004 Version 1/9

MODEL PT650M SERIES 2. Mounting the Product


2.1 Dimensions for Cutting Panel
INSTRUCTION MANUAL

45 - 0
+0.6
Caution
(1) Do not apply a voltage or current exceeding the maximum allowable
value;otherwise,it may damagd the equipment.
(2) Use a power voltage within the operation range; otherwise, it may Cut the panel for mounting to the following dimensions.
result in a fire,electrical shock,or malfunction. 2.2 Mounting the Product to the Panel
(3) The contents of this manual are subject to change without notice. To mount the PT650M to the panel,remove its fittings and insert it through the
(4) Although the contents of this manual have been prepared with extra hole in the front of the panel.From the back of the panel,fix the product to the
care,if you have any questions,or find errors or missing information, panel with the fittings.
contact the sales agent from which you purchased the product or CHI
MEI Electronics Co.,Ltd.
(5) After reading this manual thoroughly,keep it in a convenient place
for future reference.

1. Before Using the Product


Thank you for purchasing the PT650M series.This manual should be passed on
to the person who operates the product.Examine the product for damage
caused by transportation or any other defects. If you find any damage or
defects,
contact the sales agent from which you purchased the product or CHI MEI
Electronics Co., Ltd.
Caution
(1) Mount the product to a panel that is strong enough to hold the
1.1 Model Codes product, If the panel is not strong enough or the product is not fixed
The model lineup of the PT650M series is shown below.Check that the model tightly,it may fall down and cause injury.
(2) The PT650M does not have a power switch,and will thus be
code and specifications of your product match those you specified when
immediately ready for operation upon connecting it to a power
ordering. supply.
(3) If the product is installed inside other equipment,provide sufficient
PT650M- X X – X X heat dissipation to ensure that the temperature inside the equipment

Display unit
01: Single display
02: Multi display

Power unit
1: 100 to 240VAC±10%
2: 9 to 60VDC

Output unit
0: None
1: Comparison
2: Analog
3: RS-232C
4: RS-485
5: Comparison and analog
6: Comparison, analog, and RS-232C
7: Comparison, analog, and RS-485
Instruction Manual for PT650M Series 2004 Version 2/9
3. Terminal Arrangement
Comparison output Serial communication Analog output
3.6 Serial Communication
No. Name Description
RS-232C: transmission;
20 RXD(+)
RS-485: Non-reverse output
RS-232C: reception;
21 TXD(-)
RS-485: Reverse output
22 NC Do not connect this terminal.
Common terminal for
23 SG
communica- tions

Caution
(1) Use 12 to 28AWG wire for the power,input(except for range 26),external
control,and comparison output connectors.
(2) Tighten the screws for the power,input(except for range 26), external
Input(theshape
Input(the shapedepends
dependson
onthe
theunit)
unit) Externalcontrol
External control Power
Power connector
control,and comparison output connectors to a torque of 0.5 to 0.6 Nm.
(3) Use 16 to 28AWG wire for the analog output connector.
3.1 Power (4) Tighten the screws of analog output connector to a torque of 0.22 to
0.25Nm.
No. Name Description
Power terminal without polarity for
10 POWER 4. Components and their Functions
both DC and AC
Power terminal without polarity for The front panel design of the PT650M series of unit meters
11 POWER
both DC and AC differs depending on the display unit selected.The names and
3.2 External Controls functions of each unit are as shown below.
4.1 Multi-display Unit
No. Name Description
Control for hold function.Enabled
6 HOLD when short-circuited or at the same
potential as COM.
Control for digital zero function.
7 DZ Enabled when short-circuited or at
the same potential as COM.
Control for peak hold function. No. Main Functions
8 PH Enabled when short-circuited or at Name During measurement During parameter
the same potential as COM. setup
Common for all external control Main display Indicates the measured value. Indicates information
9 COM (1) on the parameter to be
terminals.
set.
Indicates the result of judgment
3.3 Strain Gauge Input Unit(Load cell) HI and turns on if the measured
value>HI judgment value
No. Name Description Indicates the result of judgment
1 +SIG Positive input terminal Judgment and turns on if LO judgment value
(2) GO
2 -SIG Negative input terminal indicators ≦the measured value≦HI
3 +EXC Power output for sensor(positive) judgment value.
Indicates the result of judgment
4 -EXC Power output for sensor(negative)
LO and turns on if the measured
Common terminal(grounding term-
5 COM value<LO judgment value.
inal for input circuit)
Turns on if “digital zero backup”is
ME
on
Function Turns on if “peak hold/valley
(3) PH
3.4 Comparison Output indicators hold/peak-valley hold”is on
DZ Turns on if“digital zero”is on
No. Name Decription RE Turns on if RS-232C or RS485
LO output terminal (b contact) Indicates the HI side judgment value.
12 LO-b
(N.C.) Indicates the item in the
13 LO-c Common terminal for LO output (4) Sub-display 1 maximum/minimum/(maximum-
LO output terminal (a contact) minimum)/input value monitoring
14 LO-a mode.
(N.O.)
15 GO-c Common terminal for GO output Indicates the LO side judgment value. Indicates the item to
GO output terminal (a contact) be set
16 GO-a Indicates information on the item in the
(N.O.) (5) Sub-display 2
HI output terminal (b contact) maximum/minimum/ (maximum-
17 HI-b minimum)/input value monitoring
(N.C.)
mode.
18 HI-c Common terminal for HI output
Pressing the Enter and Mode keys Returns to the
3.5 Analog 19 HI-a HI output terminal (a contact) (N.O.)
together changes to the parameter measurement mode.
Output setting mode.
Pressing the Enter and Increment keys
No. Name Description together changes to the
24 COM Common terminal for analog output maximum/minimum/ (maximum-
(6) Enter key
25 A-OUT Current output terminal(4 to 20mA) minimum)/input value monitoring
mode.
Voltage output terminal(1 to 5V, 0 to
26 V-OUT Switches from the maximum/minimum/
1V, 0 to 10V)
(maximum-minimum)/input value
monitoring mode to the comparative
judgment reading mode.
Instruction Manual for PT650M Series 2004 Version 3/9
Pressing the Mode and Enter keys Selects the item to Switches from the
together changes to the parameter be set. maximum/minimum /(maximum-
setting mode. minimum)/input value monitoring
Pressing the Mode and Shift keys mode to the comparative judgment
(7) Mode key together changes to the shift function reading mode.
setup mode.
Pressing the Mode and Increment keys Pressing the Mode and Enter Selects the item to
together turns on/off the “Digital zero” keys together changes to the be set
indicator parameter setting mode.
Pressing the Shift and Enter keys Changes the digit Pressing the Mode and Shift
together changes to the parameter to be set.
keys together changes to the
checking mode.(Comparator data can be (7) Mode key
shift function setup mode.
set.)
Pressing the Shift and Mode keys Pressing the Mode and
together changes to the shift function incremental keys together
(8) Shift key turns on/off the “Digital
setup mode.
Selects from items in the zero”indicator.
maximum/minimum/(maximum- Pressing the Shift and Enter Changes the digit to
minimum)/input value monitoring keys together changes to the be set.
mode.(Hold down the key for about one parameter checking mode.
second.) (Comparator data can be set.)
Pressing the Increment and Mode keys Changes the value Pressing the Shift and Mode
together tums on/off the “Digital or content of a keys together changes to the
zero”indicator selected digit shift function setup mode.
(Increments the Holding down the Shift key
value) (8) Shift key for about one second moves
Pressing the Increment and Enter keys
to the HI judgment value
(9) Increment key changes to the maximum/ minimum/
indicator.
(maximum-minimum)/input value
monitoring mode. Selects from items in the
Resets the maximum/ minimum/ maximum/ minimum/
(maximum-minimum)/input value (maximum-minimum)/input
monitoring mode.(Hold down the key value monitoring mode.(Hold
for about one second.) down the key for about one
second. )
4.2 Single Display Unit Pressing the Increment and Changes the value or
Mode keys together turns on/ content of a selected
off the “Digital digit.(Increments the
zero”indicatoe value.)
Holding down the Increment
key for about one second
moves to the LO judgment
value indicator.
Increment Pressing the Increment and
(9)
key Enter keys changes to the
No. Main Functions maximum/ minimum/
Name During measurement During (maximum-minimum)/input
parameter setup value monitoring mode.
Indicates the measured value. Indicates informa- Resets the maximum/
tion on the para-
minimum/ (maximum-
meter to be set.
(1) Main display minimum)/input value
Indicates information on the item in the
monitoring mode.(Hold down
maximum/ minimum/ (maximum-
minimum)/input value monitoring mode. the key for about one second.)
Indicates the result of judgment and turns 5. Parameter Setup
HI on if the measured value>HI judgment
value.
5.1 Differences between Display Units
Indicates the result of judgment and turns
Judgment
(2) indicators GO
on if LO judgment value≦the measured
value≦HI judgment value.
Indicates the result of judgment and turns
LO on if the measured value<LO judgment
value.
PH
Turns on if “peak hold/valley hold/peak- Setup contents Parameter name
valley hold”is on.
Turns on if “digital zero”is on 5.1.1 Multi-display Unit
DZ Flashes when linearization data output
values are set.
Function
(3) Indicators ME Turns on if “digital zero backup”is on.
Turns on if remote control is being
performed through RS-232C or RS485
RE interface.
Flashes when linearization data input
values are set. Setup contents and parameter name
Pressing the Enter and Mode keys together Returns to the Note: Pressing the Mode key displays the next parameter.
(6) Enter key changes to the parameter setting mode. measurement 5.1.2 Single Display Unit
mode. Note1: Pressing the Mode key with the parameter name shown changes the
Pressing the Enter and Increment keys
together changes to the display to the parameter information indication.If there is no key
maximum/minimum/(maximum- operation for about one second when the parameter name is shown,the
minimum)/input value monitoring mode. display automatically changes to the parameter information
indicati on(however,this change does not automatically occur for
parameters PH/S-HI/FSC,etc.,right after COND/COM/MET is
indicated).
Note2: Pressing the Mode key when the parameter information indication is
Instruction Manual for PT650M Series 2004 Version 4/9
shown results in the next parameter being displayed. information indication shown,the display returns to the parameter name
Note3: If there is no key operation for about 8 seconds with the parameter indication.
5.2 Moving to the Parameter Setup Mode

Pressing the ENTER key saves the data and returns to the measurement mode.
(Data are backed up with EEPROM even when the power is turned off.)

5.3 Data Lists and Default Settings


Input unit
Default Equipped Output unit number
Indication Name number
value as
17 0 1 2 3 4 5 6 7
Condition data
PVH Peak hold setup PH O
RANG Measurement range setup *1 x
AVG Number of averaging 1 O
MAV Number of moving averaging setup OFF O
S.UD Step wide setup 1 O
BLNK Indication blank setup OFF O
UNIT Unit setup C x x x x O O x O O
BAUD Baud rate setup 9600 x x x O O x O O
DATA Data length setup 7 x x x O O x O O
P.BIT Parity bit setup E x x x O O x O O
S.BIT Stop bit setup 2 x x x O O x O O
T- Delimiter setup CR.LF x x x x O x x O
ADR Equipment ID setup 00 x x O x x O O O
A.OUT Analog output setup OFF
B.UP Digital zero backup setup OFF O
LINE Linearization setup CLR O
I.SEL Input selection OC x
TR T Tracking zeroing time setup 00 O
TR V Tracking zeroing width setup*2 01 O
SNSR Sensor power setup 10 O
PON Power-on delay setup OFF O
PRO Protect setup OFF O
U-NO. Unit number Indication setup ON O
Comparator data
S-HI HI side judgment value setup 1000 x O x x x O O O
S-LO LO side judgment value setup 500 x O x x x O O O
H-HI HI side hysteresis setup 0 x O x x x O O O
H-LO LO side hysteresis setup 0 x O x x x O O O
Scaling data
FSC Full scale Indication value setup *1 x
FIN Full scale input value setup *1 x
OFS Offset indication value setup *1 x
OIN Offset input value setup *1 x
PS Pre-scaling value setup 1 x
PPR Frequency division setup 1 x
DLHI Digital limiter HI value setup 9999 O
DLLO Digital limiter LO value setup -9999 O
AOHI Analog output HI indication setup 9999 x x O x x O O O
AOLO Analog output OL indication setup 0 x x O x x O O O
DEP Decimal point position setup None O
Linearization data *5 O
Calibration data
ZERO Zero input value *6 0 O
SPIN Span input value *6 2000 O
SPAN Span indication 9000 O
Shift data
SHF Shift data setup 0 O
*1: Each value in the lower part of a cell in the columns on the right is the default value. *2: Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
*3: 5000 for 1 V range and 2000 for 2 A range. *4: 1000 for 1 V range and 400 for 2 A range.
*5: Linearization data are not set up for the default values. *6: This value is not indicated if calibration is done using an actual load.
Instruction Manual for PT650M Series 2004 Version 5/9
The shaded parts show the parameters that must be set for each unit.

5.4 Information on Each parameter measurement


Defaul mode.(Press the Mode key changes to the next
Indicatio
Name Setup options t parameter).
n
value 5.4.2 Method of Setting Comparator Data
Condition data This section explains comparator data and shows a typical example of setting
PVH Peak hold setup PH(peak hold)/VH(valley hold)PVH(peak-valley hold) PH the HI side judgment value.The same method applies to all other parameters.
RANG Measurement range setup *1 *1
Number of averaging HI side judgment value : 900
AVG 1/2/4/8/10/20/40/80 1
operations setup
HI side hysteresis value : 200
Number of moving average
MAV OFF/2/4/8/16/32 OFF LO side judgment value : 300
operations setup
LO side hysteresis value : 150
S.UD Step width setup 1(1digit)/2(2 digit)/5(5 digit)/0(10 digit) 1
BLNK Indication blank setup OFF/B-3/B-2/ON OFF
UNIT Unit setup C/F C Indicated value HI side judgment value
BAUD Baud rate setup 9600/4800/2400/384(38400)/192(19200) 9600
DATA Data length setup 7(7bit)/8(8bit) 7 HI side hysteresis
P.BIT Parity bit setup E(even number)/O(odd number)/N(none) E value
S.BIT Stop bit setup 2(2bit)/1(1bit) 2
T- Deiimiter setup CR.LF(CR+LF)/CR CR.LF LO side hysteresis
ADR Equipment ID setup 01~99 00 value
OFF/0~1(0~1V)/0~10(0~10V)/1~5(1~5V) LO side judgment value
A.OUT Analog output setup OFF
/4~20(4~20mA)
B.UP Digital zero backup setup OFF/ON OFF
LINE Linearization setup OFF/ON CLR
I.SEL Input selection OC(open collector)/LGC(logic)/MAG(magnet) OC
TR T Tracking zeroing time setup 00~99 00
Tracking zeroing width
TR V 00~99 01 Multi-display unit Single display unit
setup *2
SNSR Sensor power setup 10(10V)/05(5V) 10
PON Power on delay time setup OFF/ON OFF
PRO Protect setup OFF/1~30 OFF
(1)Press the Mode and Enter keys together during
U-NO. Unit number indication setup OFF/ON ON measurement.
Comparator data
S-HI HI side judgment value setup -9999~9999 1000 (2)Press the Shift key a few times to display the
S-LO LO side judgment value setup -9999~9999 500 comparator data menu.
H-HI HI side hysteresis setup -9999~9999 0
(3) Press the Mode key a few times to display the
H-LO LO side hysteresis setup -9999~9999 0 parameter to be set.
Scaling data
Full scale indication value (4)For a single display unit,press the Mode key to
FSC -9999~9999 *1 change to the parameter information indication.
setup
(The display automatically changes to this
FIN Full scale input value setup -9999~9999 *1
indication in about 1 second,except for parameter
OFS Offset indication value setup -9999~9999 *1 S-HI right after COM is indicated. )
OIN Offset input value setup -9999~9999 *1
PS Pre-scaling value setup 0.001~5.000 1 (5)Press the Shift key(change digit) and press the
PPR Frequency division setup 1~100 1 Increment key(change numeric value)to set to 10.
DLHI Digital limiter HI value setup -9999~9999 9999 Note:The decimal point in the selected digit flashes.
DLLO Digital limiter LO value setup -9999~9999 -9999 (6)Press the Enter key to return to the
Analog output HI indication measurement mode(Pressing the Mode key
AOHI -9999~9999 9999
setup changes to the next parameter)
Analog output LO indication
AOLO -9999~9999 0
setup
Decimal point indication posi- None /place of 100/place of 101/ place of 102/place of
DEP
tion setup 103
None Note:The setup conditions are HI side judgment value>LO side judgment
Linearization data *2 value,HI side judgment value≧LO side judgment value+LO side
Caiibration data hysteresis,and LO side judgment value≦HI side judgment value-HI side
ZERO Zero input value *6 -0.300~2.000 0.000 hysteresis.If these conditions are not satisfied,an error indication appears
SPIN Span input value *6 1.000~3.000 2.000 and the display returns to the HI side judgment value setup.
SPAN Span indication 0~9999 9000

5.4.3 Method of Setting Scaling Data


5.4.1 Method of Setting Condition Data This section explains comparator data and shows a typical example of setting
This section shows a typical example of setting the peak hold parameter.The the full scale indication parameter.The same method applies to all other
same methld applies to other parameter. parammeters.

Multi-display unit Single display unit Input voltage :0~10V


Indication : 0~5.000

(1)Press the Mode and Enter keys together


during measurement. FSC: 5000
FIN: 9999
(2)Press the Mode key to change to the peak OFS: 0
hold setup mode. OIN: 0
DLHI: 3000
(3)For a single display unit,press the Mode key DLLO: -2000
to change to the parameter information DEP: The place of 103 lit
indication.(The display automatically changes to
this indication in about 1 second,except right
after COND is indicated.)
Note:For the Digital limiter,values larger than the DLHI setponit are not
indicated even if signals greater that the value set in the DLHI para-
(4) Press the Increment key a few times to set to
meter are input (for DLLO parameter,values smaller than the DLLO
Valley Hold.
setpoint are not indicated.)
(5) Press the Enter key to return to
Instruction Manual for PT650M Series 2004 Version 6/9

Multi-display unit Single display 5.4.4.2 Equivalent Calibration


unit Equivalent calibration means that calibration is carried out according to the
ratings(specifications) of such a sensor as a load cell.It is not necessary to
connect the sensor or to apply pressure to the sensor.
(1)Press the Mode and Enter keys together
during measurement. Multi-display Single display
unit unit

(2) Press the Shift key a few times to change


to the scaling data menu. (1)Press the Shift, Increment, and Enter keys together
during measurement.
(3) Press the Mode key a few times to display
the parameter to be set. (2)Press the Increment key to select the equivalent
(4) For a single display unit,press the Mode calibration mode.
key to change to the parameter information
indication.(The display automatically changes (3) Press the Mode key to move to the equivalent
to this indication in about 1 second, except for calibration node.
parameter FSC right after MET is indicated.)

(5) Press the Shift key (change digit)and press (4) Press the Shift key to display the zero-input setup
mode.
the Increment key(change numeric value)to set Note:For a single display unit,the unit automatically
to10. returns to ZERO indication if there is no key opera-
Note:The decimal point in the selected digit tion for about 8 second,In such a case,press the Mode
flashes. key to return to the numerical value indication
(5)Press the Shift key (change digit) and the
Increment key(change numeric value) to set 0.004.
(6)Press the Enter key to return to the Note:The decimal point in the selected digit flashes.
measurement mode( Pressing the Mode key
changes to the next parameter.)
(6)Press the Mode key to change to the span input
value setup mode.

(7)For a single display unit,press the Mode key to


The following explains the frequency measuring unit.(The same method
display the parameter information indication.
applies to the full scale indication parameter.)
(8) Press the Shift key (change digit) and the
Increment key (change numeric value) to set 1.002.
Indication : 0.0~500.0
Output : 4~20mA (9)Press the Mode key to change to the span
indicating value setup mode.

(10)For a single display unit,press the Mode key to


HOHI: 5000 display the parameter information indication.
AOLO: 0
(11) Press the Shift key (change digit) and the
Increment key (change numeric value) to set 2000.
Note:The decimal point in the selected digit flashes.

Note:For analog output scaling,set the indication value for an output current (12)Press the Mode key to return to the measurement
of 20 mA in the AOHI parameter and set the indication value for an mode.
output current of 4 mA in the AOLO parameter(for 4~20mA output).

5.4.4 Method of Setting Calibration Data


5.4.4.1 Actual Load Calibration
Actual load calibration means that calibration is carried out by applying
actually measured pressure to a sensor such as a load cell connected to the
meter.
Multi-display Single display unit
unit

(1)Press the Shift,Increment,and Enter keys


together during measurement

(2)Press the Mode key to change to the actual load


calibration mode..

(3)Press the Mode key while applying pressure that


sill cause the display to show zero.

(4)For a single display unit,press the Mode key to


change to the parameter information indication.

(5)Press the Shift key(change digit) and Increment


key(change numeric value)to set 8000.
Note:The decimal point in the selected digit flashes.

(6)Press the Mode key to return to the


measurement mode.
Instruction Manual for PT650M Series 2004 Version 7/9
Increment key for about 1 second to zero the indication at that moment.
Note: Operation with the control terminals takes priority over operation with
the front panel keys. The Digital Zero function is disabled if the
control terminals are made to go through the off-on-off
sequence with the function enabled by means of the front panel keys.
5.4.5 Method of Setting Linearization Data 6.3 Peak Hold Function
The linearization function meas a function that changes the slope of straight The Peak Hold functon retains one of the maximum(peak hold)/minimum
lines in the relationship between the input and indication by correcting the (valley hold)/maximum-minimum(peak-valley hold) value and provides
relations at arbitrary points.Linearization data are set using the input value output for that value. Selection from these values is made using the condition
(indicated value before correction) and the output value (indicated value after data. The peak hold function is enabled by shortcircuiting the PH and COM
correction) at each arbitrary point. terminals or setting both terminals to the same voltage level.

Multi-display unit Single display 7. Output Function


unit 7.1 Comparison Output function
The PT650M series of unit meters is designed so that the two judgment values
HI and LO can be set for the measured (indicated) value to provide the results
(1) Press the Mode and Enter keys during of judgment as relay contact output.(This function is effective when the meter
measurement.
is equipped with a comparison output unit.) For details on the contact ratings
(2) Press the Shift key a few times to move to and other specifications,refer to the section “Output Specifications.”
the linearization data menu. 7.2 Analog Output Function
The PT650M series of unit meters can output an analog signal for an indicated
(3) Press the Mode key to move to the setup value (when the meter is equipped with an analog output unit). There are four
for the number of data to be corrected.. output ranges, 0 to1V/0 to10V/1 to 5V/4 to 20mA,from which a selection can
be made using the condition data. In addition,the analog output of the
(4) Press the Shift key(change digit) and the
Increment key (change numeric value) and PT650M series allows for arbitrary output scaling.This scaling can be
then press the Mode key to after the number achieved by setting the indication value for an output of the maximum scale
of data to be corrected has been set. value (20mA for 4 to 20mA output range) in the AOHI parameter of the
(5) For a single display unit,press the Mode
scaling data.
key to change to the parameter information
indication.
Note:For a single display unit,the unit
7.3 RS-485 Interface Function
automatically returns to N-01 indication if The PT650M series can be equipped with an RS-485 interface (when the
there is no key operation for about 8 meter is provided with an RS-485 unit). For details on the RS-485 function,
seconds.In this case,press the Mode key to see the separate manual on communication functions.
return to the numeric value indication.
(6)Press the Shift key (change digit) and 7.4 RS-232C Interface Function
Increment key(change numeric value)and then The PT650M series can be equipped with an RS-232C interface (when the
press the Mode key after the N-01 input value
has been set. meter is provided with an RS-232C unit). For details on the RS-232C
(7)Press the Shift key (change digit) and
function,see the separate manual on communication functions.
Increment key(change numeric value) and 8. Specifications and External Dimensions
then press the Mode key after the N-01 output
value has been set. 8.1 Input Specifications
(8)Press the Mode key to display the setup for Load cell Zero adjusting Highest
output values of the data to be corrected.. Span adjusting range Accuracy
power range resolution
Note: For a single display unit,the unit is
automatically returned to N-02 indication if 5V 0.5μV/digit
there is no key operation for about 8 second. -0.3 to 2mV/V 1~3mV/V ±(0.1%FS+2digit)
In this case,press the Mode key to return to
10V 1μV/digit
the numeric value indication. Input circuit: Single ended type
(9)Press the Shift key (change digit) and Operating system: △Σconversion
Increment key (change numeris value) and Maximum sampling rate: 12.5 times per second
then press the Mode key after the N-02 input
value has been set. Noise rejection ratio: NMR(normal mode rejection)50dB or more (50 or 60Hz)

Repeat steps (5)to(9) until all the settings have 8.2 Common Specifications
been made. Display : 7-segment LED display(character height:14.2 mm on main display
and 8 mm on sub-display)
(10)When all the settings have been made,
Polarity indication : Automatically indicated when the calculated result is
press the Enter key to return to the
negative
measurement mode.
Indication range: -9999~9999
Note:The setup conditions are N-1<N-2…N-15<N-16
Over-range alarm: OL or –OL for input signals outside the indication range
6. Control Functions Decimal point : Can be set at an arbitrary digit
6.1 Hold Function External control : HOLD+, PH, DZ(reset for frequency measuring unit)
The Hold function temporarily retains the indication.The hold function is
Operating temperature and humidity range : 0 to 50℃, 35~83%RH
enabled by shortcircuiting the HOLD and COM terminals or setting both
(non-condensing)
terminals to the same voltage level. As a result the display unit retains the
Storage temperature and humidity range : -10 to70℃, 60%RH or less
indication give at that moment.
Power supply : 100 to 240VAC±1 0 % for AC power supply unit
9 to 50VDC for DC power supply unit
6.2 Digital Zero Function
The Digital Zero function zeros the indication given at an arbitrary Power consumption : 5W
timing. External dimensions: 96mm(W) × 48mm(H) × 146.5mm(D)
Thereafter, the function shows the amount of change from the point Note: Depth(D) denotes the maximum value.
of Weight : 450g
zeroing.However, this function serves as an indication resetting function for a Withstand voltage : 2000VAC for 1 min.between power terminals and input
frequency measuring unit.Thus,the Digital Zero function can be used to reset terminal, and between power terminals and each output
the indication when there is no input signal at all.Note that,the on/off control
of terminal(AC power supply)
the Digital Zero function can be achieved by means of terminal control or Withstand voltage : 500VDC for 1 min.between power terminals and input
front terminal, and between power terminals and each output
panel keys.In the case of terminal control, the Digital Zero function is turned
on terminal(DC power supply)
by shortcircuiting the DZ and COM terminals or setting both terminals Withstand voltage : 500VDC for 1 min. between input terminal and each
to output terminal,and between analog output terminal and
the same voltage level. The indication at that moment is zeroed. In
the case communication terminals
of control with the front panel keys, hold down the Mode key and 2000VAC for 1 min between case and each output
press the terminal (common to both AC and DC supply)
Instruction Manual for PT650M Series 2004 Version 8/9
Insulation resistance : 100MΩ between the above terminals when 500V
DC is applied
8.3 Output specifications 8.4 External dimensions
8.3.1 Output for Comparison
Conditions for comparison Judgment result
Indicated value>Upper limit judgment value HI
Lower limit judgment value≦Indicated valeu≦Upper limit judgment value GO
Lower limit judgment value>Indicated value LO

Control system: Micro computer operating system


Judgment value setup range: -9999~+9999
Hysteresis :Can be set in the range of 1 to 999 digits for each judgment value
Operating speed :Depends on the sampling rate.
Output method : Relay contact output (Make and break contacts for HI and LO
and make contacts for GO)
Output rating : 240VAC, 8A(resistive load)and 30VDC,8A(resistive load)

8.3.2 Analog Output


Output type Load resistance Accuracy Ripple
0 to 1V 10kΩ or more
0 to 10V 10kΩ or more ±50mVp-p
±(0.5% of FS)
1 to 5V 10kΩ or more
4 to 20mA 500Ω or less ±25mVp-p
Note:The ripple ratings for the 4~20mA output are when the load resistance of
250Ω and the output current of 20mA are applied.
Conversion system : PWM conversion
Resoluton : Equivalent to 13 bits
Scaling : Digital scaling
Response speed : About 0.5 second

8.3.3 Communicating Function


RS-232C RS-485
Synchronization system Start and stop synchronization

Communication system Full duplex Two-wire half duplex(Polling selecting system)


9. Warranty and After-service
Communication rate 38400bps/19200bps/9600bps/4800bps/2400bps
9.1Warranty
Start bit 1 bit The warranty period shall be one year from the date of delivery. Any failure
Data length 7 bits/8 bits that arises during this period and the cause thereof is judged to be obviously
Even parity/odd parity/non-parity attributable to Zhuhai CHI MEI Electronics Co., Ltd. Shall be remedied at no
Error detection BCC(block,check,and character) cost.
check sum

Stop bit 1 bit/2 bits


9.2 After-service
Character code ASCII code
This product is manufactured,tested,inspected,and then shipped under
Communication control stringent quality control. Should the product fail,however,contact (or send the
No procedure
procedure product to) your vendor or Zhuhai CHI MEI directly. (It is advisable that you
Non-inversion(+) send a memo describing the failure in as much detail as possible along with
Signal name used TXD, RXD, SG
And inversion (-) the product returned.)

Number of connectable
1 1 up to 31 meters
unit

Line length 15m Up to 500m (total)

Delimiter CR+LF/CR
Instruction Manual for PT650M Series 2004 Version 9/9

Appendix 1:
Setting table:
Default
Indication Name
value
Condition data
PVH Peak hold setup PH
RANG Measurement range setup *1
AVG Number of averaging 1
MAV Number of moving averaging setup OFF
S.UD Step side setup 1
BLNK Indication blank setup OFF
UNIT Unit setup C
BAUD Baud rate setup 9600
DATA Data length setup 7
P.BIT Parity bit setup E
S.BIT Stop bit setup 2
T- Delimiter setup CR.LF
ADR Equipment ID setup 00
A.OUT Analog output setup OFF
B.UP Digital zero backup setup OFF
LINE Linearization setup CLR
I.SEL Input aelection OC
TR T Tracking zeroing time setup 00
TR V Tracking zeroing width setup*2 01
SNSR Sensor power setup 10
PON Power-in delay setup OFF
PRO Protect setup OFF
U-NO. Unit number Indication setup ON
Comparator data
S-HI HI side judgment value setup 1000
S-LO LO side judgment value setup 500
H-HI HI side hysteresis setup 0
H-LO LO side hysteresis setup 0
Scaling data
FSC Full scale Indication value setup *1
FIN Full scale input value setup *1
OFS Offset indication value setup *1
OIN Offset input value setup *1
PS Pre-scaling value setup 1
PPR Frequency divistion setup 1
DLHI Digtal limiter HI value setup 9999
DLLO Digtal limiter LO value setup -9999
DEP Decimal point position setup None
Linearization data *5
Calibration data
ZERO Zero input value *6 0
SPIN Span input value *6 2000
SPAN Span indication 9000
Shift data
SHF Shift data setup 0
*1: Each value in the lower part of a cell in the columns on the right is the
default value.
*2: Tracking zero width setup parameter is not indicated if the tracking time is
set to OFF(0).
*5: Linearization data are not set up for the default value.
*6: This value is not indicated if calibration is done using an actual load.