Ifu A500
Ifu A500
Ifu A500
Perseus A500
Contents
Perseus A500
Typographical conventions
Use of terms
Dräger uses the term "accessories" not only for The product "Perseus A500" is also referred to as
accessories in the sense of IEC 60601-1, but also "Perseus".
for consumables, removable parts, and attached
parts.
Illustrations
Further documents
Trademarks
Trademarks owned by Dräger The following web page provides a list of the
countries in which the trademarks are registered:
www.draeger.com/trademarks
Trademark
Perseus®
Trademarks owned by third-party
Infinity® manufacturers
D-Vapor®
Trademark Trademark owner
Drägersorb®
Selectatec® Datex-Ohmeda
MEDIBUS®
ServiceConnect® Trademark used Trademark owner
under license
WaterLock®
BIPAP Respironics
WARNING CAUTION
A WARNING statement provides important A CAUTION statement provides important
information about a potentially hazardous information about a potentially hazardous
situation which, if not avoided, could result in situation which, if not avoided, may result in
death or serious injury. minor or moderate injury to the user or patient or
in damage to the medical device or other
property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
The term "user group" describes the personnel Where product specific knowledge or tools are
responsible who have been assigned by the required, the service activities must be carried out
operating organization to perform a particular task by specialized service personnel. The specialized
on a product. service personnel was trained by Dräger for these
service activities on this product.
User groups
Clinical users
This user group operates the product in
accordance with the intended use.
Users have medical specialist knowledge in the
field of anesthesia. Users have knowledge of
device monitoring and perioperative care.
Reprocessing personnel
This user group carries out the necessary activities
to reprocess the product.
Reprocessing personnel has specialist knowledge
in the reprocessing of medical devices.
Service personnel
This user group installs the product and performs
the service activities.
Service personnel has specialist knowledge in
electrical and mechanical engineering and
experience in the servicing of medical devices.
Patient safety
Patient monitoring
WARNING WARNING
Risk due to device failure Risk due to failure of flow measurement
Device failure can compromise the correct Deposits that were not removed during
therapy functionality of the device. reprocessing can damage the measuring
wires in the flow sensor or cause a fire.
To ensure immediate remedial action in case
– Before inserting the flow sensor check for
of device failure, the device may only be
visible damage, soiling, and particles.
operated under permanent supervision of
Repeat this check regularly.
users. Always have a manual resuscitator
– Replace flow sensors when damaged,
ready.
soiled, or not particle-free.
WARNING
CAUTION
Dräger recommends that the user remains in
Risk of crushing
the vicinity of the anesthesia machine, i.e.
within a distance of up to 4 meters (12 feet). Movable device parts or attached components
This facilitates fast recognition and response may cause crushing due to clamping. Pay special
in the event of an alarm. attention to edges, movable parts, and corners
when working with the following components:
WARNING – Column cover
– Breathing system cover
Risk due to modifications – Drawers
– Extensible writing tray
Modifications to the product may lead to – Swivel arms for mounted devices
malfunctions and unforeseen risks. This may – Accessories such as gas cylinders,
result in injury to the patient or the user or in vaporizers, CLIC absorber, and CLIC adapter
property damage.
Do not modify this product. WARNING
Risk due to electromagnetic fields
WARNING
Although the medical device does not exceed
Risk of fire the applicable limiting values for
The flow sensor can ignite medications or electromagnetic fields, such radiation can
other substances based on highly flammable interfere with the functioning of pacemakers.
substances. All wearers of pacemakers should maintain a
– Do not nebulize medications or other distance of at least 25 cm (10 in) between
substances that are easily flammable or pacemaker and medical device.
spray them into the device.
– Do not use substances containing alcohol.
– Do not allow flammable or explosive
substances to enter the breathing system
or the breathing circuit.
– Do not use cyclopropane or ether.
Essential performance
General
– Supply of the anesthesia workstation with O2:
If the O2 supply (central gas supply or gas
cylinder) fails, an alarm is issued.
– Supply of the patient with adequately
oxygenated breathing gas:
If the breathing gas contains insufficient levels
of O2, an alarm is issued.
– Patients are not supplied with excessively high
anesthetic gas concentrations:
If excessively high anesthetic gas
concentrations are delivered, an alarm is
issued.
– Monitoring of the airway pressure:
Alarms are issued depending on the set alarm
limits.
Gas measurement
– Breathing gas monitoring:
Measurement of the gas composition with ISO
accuracy.
– Monitoring of the breathing gas concentrations:
If a set alarm limit is exceeded or the gas
measurement fails, alarms will be issued.
Application
Indications/Contraindications ...................... 15
Indications ....................................................... 15
Contraindications............................................. 15
Further information on application .............. 16
Environments of use........................................ 16
Intended use
The Perseus anesthesia workstation is intended Anesthesia is achieved through a mixture of pure
for use in anesthetizing adults, children, and oxygen and Air (medical compressed air) or pure
neonates and can be used for automatic and oxygen and nitrous oxide, with the addition of
manual ventilation, pressure-supported volatile anesthetic agents.
spontaneous breathing, and spontaneous
Ventilation is accomplished on the patient through
breathing.
a laryngeal mask airway, a full-face mask, or an
Perseus is equipped with airway monitoring, gas endotracheal tube.
measurement and device monitoring, O2
The integrated breathing system can be used with
insufflation, and an anesthetic gas receiving
partial rebreathing (low-flow or minimum-flow).
system.
A non-rebreathing system such as the Kuhn or
Medec Water System may be used at the external
fresh-gas outlet (optional).
Indications/Contraindications
Environments of use
WARNING
Risk of explosion and fire
This medical device is neither approved nor
certified for use in areas where oxygen
concentrations greater than 25 Vol%,
combustible or explosive gas mixtures are
likely to occur.
WARNING
Risk of device malfunctions and/or patient
injury and user injury
Magnetic fields can negatively influence the
correct functioning of the medical device and
therefore endanger the patient or user.
Do not use the medical device in rooms
where devices for magnetic field applications
are used (e.g., magnetic resonance imaging).
Overview
Hardware ........................................................ 18
Front (trolley version)....................................... 18
Front (ceiling-mounted version)....................... 19
Screen ............................................................. 19
Plug-in connection with Vapor View option for
up to 3 vaporizers ............................................ 20
Breathing system............................................. 21
Side view from left ........................................... 22
Device column ................................................. 23
Rear (trolley version) ....................................... 24
Rear (ceiling-mounted version) ....................... 26
Interfaces......................................................... 27
Gas inlets......................................................... 28
Hanger yokes (optional) for gas cylinders with
pin-index connections...................................... 29
Auxiliary power sockets (trolley version) ......... 30
Gas mixing unit (electronically controlled) ....... 31
Gas mixing unit (mechanically controlled) ....... 32
Functional scope ........................................... 33
Device versions, options, and accessories...... 33
Gas delivery..................................................... 33
Ventilation modes ............................................ 33
Additional operation modes ............................. 33
Monitoring........................................................ 33
Display on the screen ...................................... 34
Protocoling....................................................... 34
Gas supply....................................................... 34
Gas scavenging............................................... 34
Data exchange, interfaces............................... 35
Gas flow plan ................................................. 36
Breathing system............................................. 36
Gas supply (electronically controlled gas
mixer)............................................................... 37
Gas supply (mechanically controlled gas
mixer)............................................................... 38
Hardware
10335
A
M B
L C
K
D
J
I E
H
F
G
A Screen with integrated workplace illumination M Plug-in connector with Vapor View option for up
B Gas mixing unit (electronically controlled gas to 3 vaporizers (the version for 2 vaporizers is
mixer shown here) shown here)
C Work surface
D Drawer or pull-out writing tray with lockable
compartment (optional)
E Central brake
F Castors with central brake
G Footrest
H Trolley
I Anesthetic gas receiving system with flow
indicator
J CO2 absorber
K APL valve
L External fresh-gas outlet (optional)
18909
A
C
D
A Remote control for controlling and positioning D Device column without trolley
the ceiling supply unit (functions dependent on Although the remote control and the safety sensor
the type of the ceiling supply unit) are mounted on Perseus, from an electrotechnical
B Lockable drawer (optional) point of view they are part of the connected ceiling
C Safety sensor supply unit. The safety sensor functions with any
ceiling supply unit.
Screen
windows
B Audio paused key to suppress the alarm
tones of all active alarms for 2 minutes
C Rotary knob with background illumination to
A select and confirm settings
D Key for switching the workplace
illumination on and off and dimming the
illumination intensity in 3 steps (dark, medium,
and bright)
E D
C B E Key for turning the device on or off
10397
B
A Sensor unit
B Illumination unit
Breathing system
A B C D E F G
10370
H
K J
A Water trap with connection for sample line E Bag elbow with circuit plug, e.g., to seal the
B Pressure gauge (optional) for indicating the Y-piece during an automatic test
pressure in the internal breathing system F APL valve
The pressure in the internal breathing system G Expiratory port
is also indicated on the status display, see (G)
on page 31 and (H) on page 32. H Breathing system cover
I Holder, e.g., for breathing bag hose
WARNING
J CO2 absorber
Pressure gauge indication possibly K Anesthetic gas receiving system
inaccurate
Always compare the pressures indicated on
the pressure gauge with those on the status
display. Only use the pressure gauge as the
primary source of information if the pressure
indicator on the status display has failed.
C Collecting channel
D Inspiratory port
10337
C A
Device column
10369
A
F B
E
D
C
A Column cover
B GCX rail
C Recesses for cables which lead e.g., into the
device arms
D Screw for closing the cable channel lid
E Tabs to hold the cable
F Cable channel
On the ceiling-mounted version, there is a
compartment located under the column cover for
stowing accessories, e.g., a cable.
10336
A
H
A A
G C
F
D D
E E
A Mounting rail
B Gas inlets
C Strain relief for compressed gas hose
(available for the trolley version only)
D Connection for optional pole (38 mm)
E Castor with castor brake
F Gas cylinder holders (optional) with
hook-and-loop strap
(available for the trolley version only)
G Interfaces
H Rating plate
Version with hanger yokes for gas cylinders with pin-index connections (optional)
20406
A
H
B
A A
C
G
D D
E E
A Mounting rail
B Gas inlets
C Strain relief for compressed gas hose
(only available for the trolley version shown
here or in combination with the flexibility trolley)
D Connection for optional pole (38 mm)
E Castor with castor brake
F Gas cylinder holders (optional) with hook-and-
loop strap
(available for the trolley version only)
G Interfaces
H Rating plate
18956
A
Interfaces
10367
A B C D E
G F
A Main switch
B Serial interfaces (COM 1 and COM 2)
C USB interface
D Network interface
E Interface for workstation light (optional)
F IEC connector (socket for power cable)
G Potential equalization pin
H Collecting channel
Gas inlets
10338
A
C B
20405
C
10368
A B
C
A Auxiliary power sockets, depending on
equipment variant
B Isolation transformer switch (optional)
C Circuit breakers or fuses, depending on
equipment variant
10339
I J
A
H
G B
F
E D C
A Status display
WARNING
B O2 flowmeter (for O2 insufflation Aux. O2 and
emergency O2 delivery Add. O2) Risk of mix-up due to deviating arrangement
C Current time or time for Auto On In certain countries, the arrangement of the
D O2 switch (for switching between O2 gases on the status display may differ from
insufflation Aux. O2 and emergency O2 the arrangement of the flow tubes depicted
delivery Add. O2) on the screen.
E Outlet for O2 insufflation, e.g., for nasal Always pay attention to the respective
cannula labeling.
F Symbol for programmed Auto On
Explanation of the symbols which may be
G Display of pressure in the internal breathing displayed, see page 277.
system, see page 21
H O2+ key (O2 flush)
I Symbols for mains power supply and power
supply from internal battery
J Symbols for gas supply (O2, Air, N2O) through
central supply and gas cylinders
20305
J K L
A
I
H B
L L L
E D C
A Status display
B O2 flowmeter (for O2 insufflation Aux. O2)
C Outlet for O2 insufflation, e.g., for nasal
cannula
D Flow control valves for fresh gas (O2, Air, N2O)
E Total flow tube for fresh gas
F Display of the set fresh-gas flows
G Symbols for gas supply (O2, Air, N2O) from
central supply and gas cylinders
H Display of pressure in the internal breathing
system, see page 21
I O2+ key (O2 flush)
J Symbols for mains power supply and power
supply from internal battery
K Symbol for programmed Auto On
L Current time or time for Auto On
Explanation of the symbols which may be
displayed, see page 277.
Functional scope
Gas scavenging
Protocoling
Gas scavenging is conducted through the
Among other things, Perseus can save the integrated anesthetic gas receiving system (AGS).
following data in a logbook: The particle filter it contains filters the ambient air.
– Measured values This ensures the display accuracy of the flow
indicator.
– Set values and their changes
– Patient data
– Ventilation modes
– Events (e.g., alarms, confirmed alarms, switch-
on time and switch-off time)
– Test results
– Gas consumption and anesthetic agent
consumption
Serial interfaces
Two serial interfaces, COM 1 and COM 2, are
provided for data communication using the
Dräger MEDIBUS data protocol.
USB interface
After a Dräger USB flash drive is connected, a
USB interface enables, e.g., the following actions:
– Saving screen content as screenshot
– Saving and loading device configurations
– Saving system test results or protocols (partly
optional) as text file
Network interface
If a corresponding service contract has been
obtained, the Dräger RemoteService function can
be executed via a network connection and the
hospital network.
Perseus can be connected to the
Dräger ServiceConnect Gateway or a
DrägerService computer.
If the connected network offers an NTP service,
the time on the device can be synchronized with
the time on the NTP server.
For more information, see "Connections to IT
networks" on page 253.
Breathing system
20412
H
L
J
M
G
N
E I
F O
C B P
A
A Gas supply from gas mixer N Breathing bag
B Inspiratory pressure measurement O CO2 absorber
C Inspiratory valve P Blower module TurboVent 2
D Inspiratory flow sensor
E Expiratory flow sensor
F Patient
G Expiratory pressure measurement
H Patient gas measurement module
I Expiratory valve
J PEEP/Pmax valve
K APL valve
L Anesthetic gas receiving system
M Changeover between automatic ventilation and
Manual / Spontaneous
20413
H I
G
F Add. O2 A
D
C E
Aux. O2
B
20414
G
Aux. O2
H
F A
E
D
C
Operating concept
Screen............................................................. 40
Main screen ..................................................... 40
Therapy bar ..................................................... 40
Dialog windows................................................ 41
Quick setup window......................................... 41
Color concept ................................................ 42
Colors of the control elements......................... 42
Waveforms and parameters ............................ 42
Color coding for anesthetic agents and medical
gases ............................................................... 43
Day and night colors........................................ 43
Selecting and setting..................................... 43
Setting of parameters ...................................... 43
Canceling the setting procedure or the change
procedure ........................................................ 43
Activation of buttons ........................................ 43
Operating the flow control valves .................... 44
Remote control for the ceiling-mounted
version (combination with Dräger ceiling
supply units) .................................................. 44
Releasing the locking brakes........................... 44
Height adjustment............................................ 45
Status display .................................................. 45
Safety sensor ................................................. 45
Screen
A D
10401
Linked therapy controls B
Certain parameters can be linked to other C
parameters. If one parameter is changed, the
linked parameter is also selected and changed.
Among other things, this applies to the adjustment
of ventilation pressures, ventilation times or during
electronically controlled fresh-gas delivery.
Example: The device can be configured so that a A Dialog window title
change to the PEEP setting automatically causes
a change to Pinsp; as a result, the difference B Horizontal tab to open a page
between PEEP and Pinsp and therefore the tidal C Vertical tab to open subordinate structures
volume remain constant.
D Button for closing the dialog window
Linking therapy controls, see page 162.
10402
corresponding message in the message field (D).
Color concept
10342
The availability of functions and settings is
indicated by certain colors of the therapy controls,
of the therapy bar, and in dialog windows.
Color coding for anesthetic agents and Day and night colors
medical gases
There are 3 color schemes that can be selected:
Standardized color coding complying with – Day light
ISO 5359 / ISO 32 / ISO 5360 is used to identify
– Day dark
anesthetic agents and medical gases.
– Night
The colors for O2, Air, and N2O are adapted to
locally applicable standards. Setting the color schemes, see page 108.
Operating the flow control valves Closing the flow control valve
Turn the flow control valve clockwise to the end
The flow control valves of the mechanically stop.
controlled gas mixer and the O2 flowmeter are
operated as follows: In the subsequent chapters of this document, the
following is represented by simplified explanations:
Opening the flow control valve – "Open the flow control valve."
– "Close the flow control valve."
Turn the flow control valve counterclockwise.
18959
of the ceiling supply unit:
– Releasing the locking brakes for repositioning
– Height adjustment A
The keys on the control panel on the media
column of the respective ceiling supply unit can B
also be used for this purpose.
E
WARNING
Risk of operating errors and incorrect use C
Strictly follow the instructions for use for the D
ceiling supply unit.
WARNING
Risk of injury or material damage
When the medical device is docked to the
ceiling supply unit, unintentional movement
of the device may trap persons or objects
and, in the case of persons, cause crushing Releasing the locking brakes
injuries.
– Take particular care when moving the
Function of the keys on double-arm ceiling supply
medical device.
units:
– Prevent accidental activation of the
remote control, e.g., by objects. A Key (green) to release the brake on the ceiling
bearing
B Key (blue) to release the brake on the
intermediate bearing
Safety sensor
19087
instead is kept in a secure place.
WARNING
A B
Risk of injury
If the cable connection between Perseus and
the ceiling supply unit is not established, the
safety sensor will be inoperative.
Before using Perseus, make sure that the C
cable connection has been established.
45125
Perseus can be used with ceiling supply units from
other manufacturers. In this case, a special kit (see
list of accessories) is required. The ceiling supply
unit must be prepared before Perseus can be
mounted.
A B Dräger will provide the manufacturer of the ceiling
supply unit with the document "Perseus A500
anesthesia workstation - ceiling-mounted version -
interface description for ceiling supply units". The
instructions in this document must be followed
during construction.
C
Make sure that the keyswitch on the ceiling The ceiling-mounted version of Perseus can be
supply unit is in position (A). Contact service mounted on a flexibility trolley and can then be
personnel if necessary. used as a trolley version.
When Perseus is mounted on the flexibility trolley,
WARNING the remote control is inoperative and the safety
sensor is disabled.
Risk of injury
Information on attaching accessories to Perseus
If the ceiling supply unit has not been set up on the flexibility trolley can be found in the chapter
for operation with Perseus or the keyswitch is "Mounting of accessories".
in an incorrect setting position, the safety
sensor will be inoperative.
Only operate Perseus with ceiling supply
units that have been set up for Perseus and
have the keyswitch in the correct setting
position. If necessary, contact service
personnel for clarification and specialized
service personnel if the keyswitch is faulty.
Mounting of accessories
Depending on the arm length, the following Multiple attached P1 + P2 + P... <8500 mm*kg
maximum weights are permissible: components: (<738 in*lbs)
23537
Depending on the position of the mounting rail, the
following weights are permissible:
B2 Position Maximum weight
On the side of the work- 10 kg
ing surface (22 lbs)
At the top on the rear of 2.5 kg
B1 the device (5.5 lbs)
WARNING
A Risk of tipping over
If the weight of the mounted accessories
exceeds the permissible maximum weight,
Example of a permissible configuration: the medical device may tip over.
Observe the maximum weight per arm.
Maximum Arm weight Weight of the
length attached com-
ponent WARNING
400 mm 4 kg 5 kg Risk of tipping over
(15.7 in) (8.8 lbs) (11 lbs)
If the weight of the accessories is unevenly
300 mm 3 kg 12 kg distributed about the device, the medical
(11.8 in) (6.6 lbs) (26.5 lbs) device may tip over.
Distribute the weight evenly.
P1= 400 mm * (4 kg + 5 kg) = 3600 mm*kg
(P1=15.7 in * (8.8 lbs + 11 lbs) = 310.9 in*lbs)
P2= 300 mm * (3 kg + 12 kg) = 4500 mm*kg Special features of the ceiling-mounted
(P2= 11.8 in * (6.6 lbs + 26.5 lbs) = 390.6 in*lbs)
version
3600 mm*kg + 4500 mm*kg = 8100 mm*kg
(310.9 in*lbs + 390.6 in*lbs = 701.5 in*lbs)
WARNING
8100 mm*kg < 8500 mm*kg
(701.5 in*lbs < 738 in*lbs) Risk of tipping over
If a combination of arms on one device side If the ceiling-mounted version has been
exceeds the value of 8500 mm*kg (738 in*lbs), placed on the flexibility trolley or the service
tipping stability as per IEC 60601-1 is no longer cart, the information for fitting the
given. Check the tilting stability. accessories still applies as for the ceiling-
mounted version.
– Pay attention to the maximum weight of
the accessories.
– Pay attention to the distribution of the
weight.
CAUTION WARNING
Risk of injury or material damage Risk of injury or material damage
If an accessory is fitted to Perseus which projects If Perseus is mounted on a flexibility trolley
below the lowest point of the anesthesia and additional devices have been attached, it
workstation, the accessory may strike objects or is possible that the maximum load capacity
persons when the ceiling supply unit is lowered. will be exceeded when Perseus is coupled to
the ceiling supply unit again.
Only fit accessories that maintain a distance of at
– Observe the maximum load capacity of the
least 12 cm (4.7 in) from the floor when the
ceiling supply unit.
anesthesia workstation is lowered to its lowest
– Use only those combinations of patient
point.
monitoring and accessories that are
permissible for the ceiling-mounted
The total weight of all accessory parts that are version of Perseus.
fastened to standard rails or are placed on the – Observe the maximum weights and tilting
shelves must not exceed 20 kg (44 lbs). moments.
Additionally, all product labels regarding the
maximum load that are attached to standard rails,
shelves, and ceiling supply units must be
observed.
CAUTION
Risk of damage to the device
If the maximum load capacity of the ceiling
supply unit is exceeded, the load cannot be
supported.
Refer to the instructions for use of the ceiling
supply unit and observe the maximum load
capacity.
WARNING
Risk of injury or material damage
If Perseus is not properly mounted on the
flexibility trolley or is being dismounted from
it, the trolley may tip over or Perseus may fall
off.
The mechanical coupling and uncoupling of
Perseus and the flexibility trolley must only
be performed by specialized service
personnel.
NOTE
20095
Prior to initial commissioning, the breathing
system and blower module TurboVent 2 must be
reprocessed. Carry out the reprocessing as
described in the reprocessing instructions
supplied with the product.
WARNING
3 Pass the power cable around the lower lug (A)
Risk due to incorrect mains voltage or for strain relief.
missing protective ground
WARNING
If the device is connected to a power socket
with incorrect mains voltage or a power Risk of electrical overload
socket without a protective ground, an
electric shock may occur. If the device is connected to additional power
socket strips, this may lead to increased
Connect the device only to power sockets leakage current. The leakage current may
with correct mains voltage and a protective exceed the permissible values.
ground. – Do not connect the device to additional
power socket strips.
NOTE – Do not connect additional power socket
strips to the auxiliary power sockets under
The mains plug must be readily accessible so the column covering.
that the power supply to Perseus can be
interrupted quickly in the event of a device 4 Plug the power cable into the mains power
failure. socket on the wall.
5 Check the displays for mains voltage and
1 Connect the power cable to the device and
battery on the status display.
secure it.
6 Set the main switch to position I, see page 27.
2 Lay the power cable in the cable conduit on the
right side of the device.
Charging the battery Make sure that the maximum power consumption
of additional devices does not exceed permissible
The internal batteries are automatically charged by values.
the mains power supply.
WARNING
WARNING Risk of fire
Risk of device malfunction Sources of ignition such as electrosurgical
If the batteries are not sufficiently charged, it equipment, laser surgical equipment, sparks,
may not be possible to maintain operation for damaged cables, or damaged plugs can
long enough if the mains power supply fails. cause fires in combination with oxygen or
nitrous oxide. As a result, the patient or user
Before first operation or after storage, charge may be at risk.
the batteries for at least 8 hours. – Maintain a distance of at least 200 mm
(7.9 in) between electrical connections
and components which conduct oxygen
WARNING and nitrous oxide.
Risk due to reduced power supply from the – Cables and connections must be
internal battery sufficiently insulated and must not be
damaged. Check cables for damage daily.
Batteries are wear parts. The capacity of the – Do not allow the device to come into
battery diminishes with the period of use. contact with sources of ignition.
Check the functional capability of the battery
by performing regular inspections. WARNING
Risk of electric shock
WARNING
The connection of devices to auxiliary power
Risk of patient injury sockets can lead to an increased leakage
current. If the protective ground of one of
If Perseus is operated or connected to the these devices fails, the leakage current may
mains power supply at ambient temperatures rise above the permissible values. Only
above 35 °C (95 °F), the battery cannot be connect devices with the permission of the
charged properly. The power supply out of respective manufacturer. Have the leakage
the battery may be limited. current checked by service personnel.
Do not expose the device to temperatures If the permissible value is exceeded, use a
above 35 °C (95 °F) on a permanent basis. mains power socket on a wall instead of the
auxiliary power socket of the device.
WARNING
Establishing potential equalization
Risk of electric shock and of device Differences in electrical potential between devices
malfunction can be reduced by potential equalization.
Any connected devices or device Potential equalization does not replace the
combinations not complying with the protective ground connection.
requirements in these instructions for use
may compromise correct functioning of the During operation, the potential equalization
medical device. connectors must be readily accessible and must
– Do not connect high-frequency surgery be removable without tools.
equipment to the auxiliary power sockets
of the anesthesia machine. Connecting the potential equalization cable
– Before using the medical device, refer to
and strictly comply with the instructions 1 Connect the potential equalization cable to the
for use of all connected devices and potential equalization pin on the device.
device combinations. 2 Connect the potential equalization cable to a
potential equalization connector of the hospital
WARNING (e.g., wall, ceiling supply unit, operating table).
WARNING
Risk of electric shock
Connecting devices to the serial data
interfaces can lead to an increased leakage
current. If the protective ground connector of
one of these devices fails, the patient leakage
current may rise above the permissible
values.
– Have the leakage current checked by
service personnel.
– If the permissible value is exceeded,
disconnect the devices from the serial
interfaces.
– Do not touch the connectors of the
interfaces and the patient simultaneously.
WARNING
Risk due to impermissible data connections
The impermissible use of data interfaces can
result in new hazards.
– Only make connections to data interfaces
with permission from the responsible
organization (IT representative and the
hospital equipment officer).
– Follow the information in the chapter
"Connections to IT networks".
Intrahospital transport
When parking, always engage the brakes (central 1 Visually check the medical device for damage,
brake for front castors, individual wheel brakes at particularly the hoses and cables.
rear), especially on inclined surfaces. 2 Any damage must be repaired by service
personnel before using the device.
WARNING WARNING
Risk due to gas supply failure Risk due to failure of central gas supply
All gas supplies (central gas supply, gas If the central gas supply fails, additional
cylinders) must be correctly connected since devices such as such devices which are
otherwise the backup system (gas cylinders) connected to the gas outlets on the device,
will not be available if gas supply fails. will no longer be supplied with gas. The
– Make sure that all compressed gas hoses device will be supplied from the gas
are correctly connected to the rear side of cylinders.
the device.
When using such additional devices, always
– After connecting the gas supply, check for
monitor the gas supply independently of the
correct function.
main device.
– Even when the anesthesia machine is
connected to the central gas supply, the
gas cylinders should remain at the device WARNING
with valves closed as backup.
Danger to the patient and user
WARNING The device may be damaged if the strain relief
of the compressed gas hoses is not used
Risk of device malfunction correctly.
Gas supply (central gas supply or gas Use the strain relief of the compressed gas
cylinders): To avoid damage to devices hoses correctly.
connected to the gas supply, only use
medical gases. In particular, observe the
national and international standards WARNING
regarding the use of medical gases.
Risk of fire
Ignition sources in combination with oxygen
can lead to fires.
Do not position oxygen sources in the vicinity
of ignition sources, e.g., electrical
connectors.
10343
Pressure reducers have an internal relief
valve. If a defect occurs, gas can escape into
the ambient air.
Do not block or cover the relief valve.
WARNING
B
Risk of damage to the device
D
When connecting the pressure reducers,
ensure that they do not protrude beyond the
device.
A
4 Tightly screw the pressure reducers (D) to the
gas cylinder valves. The connections must fit
C C each other directly; do not use transition
pieces.
20421
H I A the connections (M).
B 3 Two fixing positions are possible for the gas
cylinder holder (K). Adjust the position of the
C gas cylinder holder to the size of the gas
cylinder in use. Contact service personnel to do
D this.
4 Remove the protection cap from the head of
the cylinder.
E When changing gas cylinders:
G
F 5 Remove the old sealing washer (D).
6 Insert a new sealing washer (D) at the cylinder
Rear of device: holder (I).
7 Make sure that both pin-index pins (A) are
M present below the gas inlet (B).
20415
10386
Depending on the configuration, Perseus can be C
operated with vaporizers which have a Dräger C
Auto Exclusion plug-in adapter or a Selectatec
plug-in adapter. Dräger recommends using only
vaporizers that have been tested and are listed in
the list of accessories.
WARNING
B B
Risk due to incorrect anesthetic agent
delivery
If the vaporizer is filled with the wrong
anesthetic agent or if it is not filled
sufficiently, incorrect anesthetic gas 1 Set the vaporizer on the plug-in connector
concentrations or concentrations that are too evenly and firmly.
low can occur as a result. 2 Turn the locking lever (A) clockwise. The lever
– Follow the instructions for use for the is in the locked position when it is pointing left.
vaporizer.
– Compare the color coding and labeling on 3 Check the vaporizer filling level through the
the vaporizer used with the anesthetic sight glass (B). If necessary, fill the vaporizer.
agent bottle and the anesthetic agent 4 Turn the control dial to position 0; the key (C)
indicated on the screen. latches into place.
5 Check the locking:
The vaporizers used must conform to the Set the control dial on a vaporizer to a position
ISO 8835-4 or ISO 80601-2-13 standard. If the other than 0 and make sure the other vaporizer
internal patient gas measurement module fails, an remains locked in its 0 position. Repeat the test
independent gas measurement system complying with the other vaporizer.
with ISO 80601-2-55 must be used.
6 Set both control dials to position 0.
CAUTION
10354
Risk of ambient air contamination
If the anesthetic gas receiving system is not
connected to the disposal system, contamination
of the ambient air with anesthetic gas may result. A
Ensure the following:
– The anesthetic gas receiving system is
correctly connected to the disposal system.
– The anesthetic gas scavenging system is
functioning properly. Check the flow indicator.
B
Perseus is equipped with an integrated anesthetic
gas receiving system (AGS).
10353
the terminal unit of the scavenging system.
As an option, the integrated anesthetic gas
receiving system can be operated in combination
with a control valve. Observe the assembly
instructions for the control valve.
Assembling the breathing system 2 The following parts must be free from deposits:
– Incident flow meshes
WARNING – Non-return valves (red and blue)
If necessary, remove deposits on the valve
Risk of insufficient anesthetic gas plates of the non-return valves with a soft cloth.
concentrations
If the incident flow meshes are damaged,
If the component connections of the contact DrägerService.
breathing system are not leak-tight enough,
ambient air may be added to the anesthetic
gas mixture. Installing the TurboVent 2 blower module
Make sure that all components of the
10540
breathing system are connected tightly.
10394
10391
B
F
E
B
Mounting the upper part of the breathing Inserting the flow sensors and the ports
system housing
WARNING
10357
I Risk of fire
High temperatures arise in the flow sensors,
for example, in operation or while calibrating
during a system test. Due to the high
temperatures, residual vapors of flammable
G disinfectants (e.g., alcohols) and residues
I that were not removed during reprocessing
I may ignite.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
H – Before inserting the flow sensor, check for
visible damage and soiling, such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
10358
1 Place the upper part (G) of the breathing L
system on the lower part (H).
2 Pay attention to the correct alignment of the
three levers (I), see illustration above.
3 Turn the 3 levers (I) by about 120° clockwise.
K
J
10366
The CLIC adapter allows the following disposable
CO2 absorbers to be used:
– Infinity ID CLIC Absorber 800+
– CLIC Absorber 800+
– Infinity ID CLIC Absorber Free B
– CLIC Absorber Free
10392
10364
C
Mounting the flexible arm (optional) or 1 Attach the attachment piece of the arm (B) to
the bag elbow the connection piece (A) on the breathing
system and screw down tightly with the two
knurled screws. Check that the arm is fixed
The breathing bag can be mounted either on the securely!
flexible arm or, using the bag elbow and a
breathing hose, mounted directly on the breathing 2 Attach the elbow (C) to the end of the flexible
system. arm.
3 Align the flexible arm so that collisions with
other mounted assemblies are prevented.
Mounting the flexible arm
10347
C Attach the bag elbow (D) directly to the
connection piece on the breathing system.
A
B
Safety information
WARNING NOTE
Risk due to leakage in coaxial breathing This device is made without natural rubber latex.
circuit
To minimize the risk of exposure to latex, use
Leakage in the inner hose of a coaxial breathing bags and breathing hoses that are not
breathing circuit may result in CO2 made with natural rubber latex.
rebreathing or inadequate gas exchange. The
device can only detect such leakage if a
separate test with a coaxial test adapter is
performed.
– Use a coaxial test adapter to check the
inner hose for leakage, see page 125.
Next, perform a leakage test on the entire
breathing circuit.
– Monitor the measured gas concentrations
during ventilation.
WARNING WARNING
Risk due to particles and dust Risk due to occluded components in the
breathing circuit
In order to protect the patient from particles
and dust, a filter must be used between the If filters, hoses, or endotracheal tubes are
inspiratory limb of the breathing system and occluded and sample gas is being taken from
the patient. between the patient and the occluded
component, the sample gas flow may lead
Use a patient-side filter or a filter at the
immediately to negative pressure in the
inspiratory port.
lungs.
Or
Filter at inspiratory port and expiratory port, connector
Lateral connectors for the sample line support
for sample line at Y-piece:
CO2 measurement and help to flush the dead
space between Y-piece and hose adapter.
If using an expiratory filter is not possible
(e.g., due to an intrinsic PEEP due to air trap-
ping), hygienically reprocess the device after
use with this patient.
Carry out the reprocessing as described in the
reprocessing instructions supplied with the
product.
When fine pored filters (A) are used, do not connect the sample line
between the tube and the filter:
A
When no filters, but only an HME (B) is in use, set the alarm limits for
MV low and Paw high to suitable values:
10345
10346
A
WARNING
Risk due to pinched breathing bag
If the breathing bag is pinched, excessive
airway pressures or a lack of fresh gas may
result.
When attaching and aligning, pay attention to
the following:
–The breathing bag is visible.
–The breathing bag is not pinched.
–The breathing bag can inflate freely.
Risk due to changed hose lengths If necessary, take into consideration additional
parts such as water traps or additional hoses. The
Changed hose lengths can change resistance specifications for the resistance of the breathing
and compliance. Especially for neonates, this system can be found on page 249. The
may cause increased or decreased ventilation specifications for all other accessories can be
volumes. found in the respective instructions for use.
For neonates in particular, do not use flex hoses. In these instructions for use, the specifications for
the resistance in the inspiratory limb are regarded
as negative values. The resistance values given in
the instructions for use for the accessories must
therefore be subtracted from the inspiratory
resistance of the breathing system.
Example of calculation: Adult hose with filter on Y-piece, no filters on the ports.
10388
(optional)
WARNING
Risk of excessively high airway pressure
1 Tightly screw the sample line to the elbow of
Without a pressure-relief valve or breathing the non-rebreathing system and water trap (A).
bag, airway pressure may become too high. For elbows without port for sample line:
Only connect breathing systems with Fit the T-piece with T-piece filter directly to
breathing bags or pressure-relief valves the elbow and tightly screw the sample line
which comply with ISO 8835-2. to the T-piece filter. The part numbers for
the T-piece and the T-piece filter are listed
Strictly follow the instructions for use for the non- in the list of accessories.
rebreathing system and the transfer hose. Or
To prevent contaminating the ambient air, connect If necessary, use the gas sampling port of
the gas outlet of the breathing system with the inlet the filter on the Y-piece.
on the AGS. Use transfer hose with overpressure Ensure the correct course of the sample
valve complying with ISO 8835-3. line. Do not use adapters.
2 Connect the fresh-gas hose of the non-
WARNING rebreathing system to the external fresh-gas
outlet (B).
Risk of faulty gas delivery
O2 and CO2 and any anesthetic gases must
also be monitored for non-rebreathing
systems.
The sample line must be connected to the
elbow and the water trap on the anesthesia
machine.
10389
C
CAUTION
Risk due to open AGS inlet nozzle
After using a non-rebreathing system, reinsert
the sealing plug into the inlet nozzle to prevent
contamination of the ambient air with anesthetic
gases.
10348
Use of soda lime over prolonged periods can B
increase the inspiratory CO2 values.
Check the color of the soda lime regularly
and exchange as needed, especially if the
inspiratory CO2 value increases A
unexpectedly.
10349
Risk of insufficient ventilation
If the CO2 absorber is not correctly locked
into place, system leakage may occur.
After mounting and replacing, make sure the
C
CO2 absorber is firmly locked into place.
Filling and mounting 3 Insert a new disposable dust filter (E). Only use
dust filters from the list of accessories. Only
use undamaged filters.
10350
4 Insert CO2 absorber (F) on the breathing
system from below and then rotate it
counterclockwise as far as it will go.
Strictly observe the instructions for use for
Drägersorb 800 Plus or Drägersorb Free soda
D lime.
Water trap
WARNING
E
Risk of infection
The water trap may contain infectious fluid.
F – Proceed carefully when emptying and take
protective measures if necessary.
– Follow the infection prevention policies
and reprocessing regulations of the
healthcare facility.
WARNING WARNING
Risk of incorrect measured gas values Risk due to leakage from improperly
connected or damaged sample line
Blocked water traps or blocked sample lines
prevent correct gas measurement. As a Connect the sample line correctly, otherwise
result, incorrect gas values may be displayed. faulty gas measurements may result.
– Use only Dräger sample lines.
– Replace the O-rings of the water trap
10344
mount every 2 years.
CAUTION
Risk of misleading data
Silicon can get into the measuring cuvette and
disrupt measurement.
Do not spray the O-rings of the water trap holder
with silicon spray.
WARNING
Risk of fire
Silicon or aerosol residues in the water trap
can cause fires.
Screw on the sample line at the Y-piece, HME
Do not spray the O-rings of the water trap filter, or hose adapter and at the water trap.
holder with silicon spray. Ensure the correct course of the sample line.
Do not use adapters.
Getting started
Turning on Perseus
10340
and assembled to be ready for operation in
accordance with the reprocessing instructions (see
page 46).
CAUTION
Risk of device malfunction
Condensation may form when the device is
moved from a cold storage location to a warm
environment.
A
To prevent condensate formation and resulting
failures of electrical components, do not turn on
the device after abrupt temperature changes for
1 to 2 hours. 1 Connect the device to the mains power supply.
2 Set the main switch to position I.
WARNING
3 Turn on Perseus: Press the (A) button.
Risk of explosion and fire
If the battery charge is sufficient, Perseus will also
Do not set the device into operation if oxygen start without the power plug being plugged in as
leakage is suspected in the medical device or long as the main switch is not set to position 0.
its vicinity.
When the starting procedure is completed, the
Stop all oxygen supplies and contact service Standby page is displayed.
personnel.
WARNING
Risk due to unbraked device
An unbraked device may accidentally move
during operation.
Prevent this by actuating the central brake or
the castor brakes and check their function.
WARNING
Risk of device malfunction
Some safety systems are only checked
during start-up.
Restart Perseus at least once per month to
maintain proper functionality.
On the Standby page, the readiness for operation If the system is not fully operational, the most
resulting from the system test is indicated by color important irregularities (C) are displayed and a
(A). specific test (D) is recommended as a remedy
where possible.
10410
Color Meaning
Green System is fully operational.
Yellow System is operational with limita-
tions.
There are functional restrictions.
Take further measures to ensure
patient safety (e.g., external moni-
toring).
Red System is not operational.
Call service personnel.
Gray The system has not been tested.
Emergency start-up
10376
20341
D C
A
B A
C B
2 Set O2 switch (C) to the Add. O2 position. 2 Open the flow control valve (A) and set the
3 Open the flow control valve (B) and set the desired O2 flow. If required, press the O2+ key
desired O2 flow. If necessary, press the (C) to fill the breathing bag quickly.
O2+ (D) key to quickly fill the breathing bag. 3 Monitor the set flow on the total flow tube (B).
4 Monitor the set flow on the O2 flowmeter (A). 4 Set the anesthetic gas concentration at the
5 Set the anesthetic gas concentration at the vaporizer.
vaporizer. 5 Manually ventilate the patient.
6 Manually ventilate the patient. 6 Switch on the device.
7 Switch on the device. 7 As soon as the Standby page is displayed,
8 As soon as the Standby page is displayed, start the therapy, see page 87.
start the therapy, see page 87.
9 Set the O2 switch (C) to Aux. O2 to stop the
increased fresh-gas flow.
10 Close the flow control valve (B).
Operation
Safety information
WARNING CAUTION
Risk of strangulation Risk of injury
Negligent placement of hoses, cables, sample Looking directly into the LEDs of the workplace
lines, and similar device components can illumination and of the illumination unit on the
endanger the patient. vaporizer plug-in connection can cause damage
to the retina.
Use particular caution when establishing
connections to the patient. Do not look directly into the LEDs. Make sure that
the patient is not dazzled by the LEDs.
CAUTION
CAUTION
Risk due to falsified gas measurement values
and failure of the gas measurement Risk due to unsuitable illumination
Aerosols can damage the patient gas If illumination without neutral colors is used
measurement module and the membrane of the during the medical examination of the patient,
water trap. Propellants can falsify the gas this may result in, e.g., misinterpretation of skin
measurement values. coloring.
– Do not use the device's workplace light during
Do not use medication nebulizers.
examinations.
– For examinations, use an examination light
WARNING conforming to IEC 60601-2-41.
Risk of faulty gas delivery
O2 and CO2 and anesthetic gases must be
monitored every time breathing systems are
used.
Ensure adequate gas monitoring.
WARNING
Risk of misinterpretation
Misdiagnosis or misinterpretation of the
measured values or other parameters can
endanger the patient.
Do not make therapeutic decisions based
solely on individual measured values and
monitoring parameters. Therapeutic
decisions must be made solely by the user.
WARNING WARNING
Risk of misinterpretation Risk of infection
Measured gas values and waveforms such as Disease-causing germs can be transmitted
the CO2 waveform are determined on the between patients via the circuit plug. If a used
basis of the composition of the sample gas. Y-piece or filter is fitted to the circuit plug,
The composition of the sample gas is and then later a reprocessed component is
affected by many factors and their fitted (e.g., during a leakage test), the new
interactions, especially in patients with low component may become contaminated.
body weight. This may result in biased
Do not fit hoses that have already been used
measurements or waveforms.
and that have attached filters or Y-pieces to
The following factors affect the sample gas the circuit plug. Instead, hang them over the
measurement: handle on the left-hand side of the device.
– Dead space
– Patient resistance, patient compliance WARNING
– Type of surgical procedure Risk due to occluded components in the
breathing circuit
– Gas sampling site
– Breathing circuit, filter, sample line, tube If filters, hoses, or endotracheal tubes are
occluded and sample gas is being taken from
– Ventilation settings and the resulting between the patient and the occluded
ventilation component, the sample gas flow may lead
– Leakage immediately to negative pressure in the
lungs.
– Spontaneous breathing
– Cardiogenic oscillations Ensure the following when ventilating
pediatric patients and neonates:
Observe the following: – When fine pored filters are used, do not
– Do not make therapeutic decisions based connect the sample line between the tube
solely on individual measured values or and the filter.
parameters. – When no filters but only an HME is in use,
set the alarm limits for MV low and Paw
– If possible, minimize the effects of the high to values that are suitable for this
factors described above, e.g., take the patient so that an incipient occlusion can
sample gas from a gas sampling site close be detected.
to the patient, minimize leakage, adjust the
ventilation settings.
CAUTION
CAUTION Danger from an unintentional therapy start
Danger due to inaccurate gas measurement If Perseus is in the Standby mode and the flow
values control valves are open, ventilation starts in the
Man / Spon mode.
Due to the technical characteristics of gas
measurements, the measured gas values might Take care that this does not happen
be inaccurate at high respiratory rates and at unintentionally.
certain I:E ratios.
Observe the technical data.
WARNING
Risk of fire
Components such as power supply units that
heat up are unable to cool down in enclosed
storage locations and may cause a fire.
Do not keep components that heat up in the
drawers or under the column cover.
10461
Start
Select patient and check data
A B
A B C
Select mode
Standby
C
There are 2 possibilities for starting the therapy:
There are two possibilities for loading patient data:
Quick start with Manual / Spontaneous mode
A Defining a new case
To perform a quick start with the displayed
fresh-gas settings (A), touch the Quick start B Continuing a case
Man/Spon button (B). Depending on the selection, the patient data in
Or area (C), e.g., weight or age, are filled in
automatically.
Normal start with customized settings
1 Touch the Start... button (C). Defining a new case
2 Adjust the patient data and ventilation settings.
Select to start the therapy for a new patient:
Depending on the patient category needed,
touch the New adult, New pediatric, or New
neonate button (A).
The ventilation parameters and alarm limits are set
to the configured start settings, see page 150. The
set value for Ti is set automatically based on RR
so that the I:E ratio set in the start settings are the
result, see 158.
10463
start settings. A
D C B E
Checking the patient data
1 Select a ventilation mode (A) as needed.
The following ventilation modes are available:
WARNING
– Man / Spon
Risk due to incorrect settings – PSV (optional)
Different standard alarm limits or therapy – PC
settings might be configured for medical – APRV (optional)
devices within the same area. The user must
– VC - AF
observe the following:
– Make sure that the set values are suitable – VC
for the patient. The following operation modes are also available:
– Make sure that the alarm system is neither – Ext. FGO (optional)
rendered useless by setting extreme
values for the alarm limits nor deactivated – Pause
by switching off the alarms. For additional information about the ventilation
– Check the start settings for alarms and modes, see page 258.
alarm settings each time the ventilation
2 Set the fresh-gas delivery.
mode is changed.
– Only switch off alarms if the safety of the Perseus is equipped with a minimum O2 delivery
patient is not affected. that ensures that a minimum quantity of oxygen is
delivered, see page 268.
CAUTION Electronically controlled gas mixer:
Select the carrier gas (B).
Risk due to incorrect setting for patient age Set the O2 concentration (D) and fresh-gas
Incorrectly setting the patient age can lead to flow (C).
inappropriate xMAC values and therefore to an To prevent the soda lime from drying out at an
inappropriate anesthetic gas delivery. increased rate, the electronically controlled gas
Always set patient age correctly. mixer limits the setting range for the fresh-gas
flow FG flow. This restriction depends on the
set tidal volume VT and the set respiratory rate
10462
RR.
A
Only use the total flow tube as the primary Electronically con- Mechanically con-
source of information in the following cases: trolled gas mixer trolled gas mixer
– The status display has failed.
– The values on the status display do not In the Standby Open the flow con-
match the values on the total flow tube. mode, touch Quick trol valves for fresh
start Man/Spon. gas.
Use the flow control valves to adjust the
fresh-gas flows. Opening the start dialog in Standby mode
Use the total flow tube to check the total
flow set, see page 32. Touch the screen in the monitoring area.
Or
CAUTION
Squeeze the breathing bag.
Risk of patient injury
Unsuitable soda lime can result in disintegration Starting when time is limited
products from the anesthesia gases.
When time is limited, it is possible to bypass the
Use suitable soda lime such as Drägersorb Free. adjustment of the patient data and the ventilation
settings. Start the therapy as follows:
WARNING 1 Touch the screen.
Risk of patient injury 2 Check the displayed start values.
Under certain conditions, acetone can 3 To start therapy, press the rotary knob.
accumulate in the body in patients under 4 Adjust the patient data and ventilation settings
anesthesia. As a result, the patient could be as soon as possible.
put at risk.
Do not perform low-flow anesthesia on
patients with ketoacidosis or patients under
the influence of alcohol.
CAUTION
Risk of patient injury
The use of minimum-flow or low-flow settings can
lead to accumulation of metabolic by-products in
the breathing system.
When using minimum-flow setings or low-flow
settings follow the recommendations of
professional societies (e.g., regular flushing of
the breathing system).
10380
The pressure limitation set with the APL valve only
takes effect during manual ventilation or
spontaneous breathing.
WARNING
Risk of excessively high airway pressures
If the ventilator fails, the device switches into
the Man / Spon ventilation mode.
The APL valve should also be set to a In the Manual / Spontaneous mode, lifting the
pressure limitation value suitable for the valve relieves pressure from the breathing system.
patient when using automatic ventilation
modes. If the ventilator fails, ventilate the
patient manually. Spontaneous breathing
10379
(Man) and spontaneous breathing (Spont) is
made at the APL pressure limitation valve, see
page 18.
Manual ventilation
10378
10403
A In ventilation modes in which breaths are to
be triggered only by the patient (e.g., PSV),
B C adverse settings or sensor failure can lead to
insufficient ventilation.
1 In the therapy bar, touch the tab (A) of the new Set the respiratory rate to a suitable value so
ventilation mode. that a minimum ventilation of the patient is
When the ventilation mode is changed, the start maintained.
settings are adopted from the parameters of the
previous ventilation mode and the patient data, or WARNING
they are sensibly derived.
Risk of insufficient ventilation
In addition, the alarm settings are adjusted to
appropriate values, see page 143. The display of the spontaneous minute
volume MVspon shows the volume that
2 Adjust the therapy as needed with the therapy
results from the patient's own breathing and
controls (B) or buttons (C).
machine support. If machine support is
3 Activate the ventilation mode by pressing the triggered by small tidal volumes of the
rotary knob. A signal tone is emitted when the patient, a large part of MVspon is achieved by
mode is changed. machine support and not by the patient's own
breathing. In this case, MVspon shows a high
value although the actual spontaneous
Synchronizing the breaths minute volume is very low.
Do not base therapy decisions solely on the
Turning on the synchronization activates the set value displayed for MVspon.
pressure support, for example, see page 258.
10464
1 Turn the synchronization on or off with the For tidal volumes below 20 mL or when using
buttons Sync. on (A) or Sync. off (B). unblocked tubes:
2 As needed, the expanded therapy bar can be Use pressure-controlled ventilation, see table
displayed using the More buttons (C) or (D); "Ventilation modes and effective parameters"
here, additional parameters (Trigger, ∆Psupp, starting on page 260.
etc.) can be adapted to the patient.
Using non-rebreathing systems When ventilation modes are changed, the CBM
mode remains active. Changing to the Standby
Only available with the option External fresh-gas mode deactivates the CBM mode.
outlet. Deactivating the CBM mode activates the apnea
Prerequisite: The non-rebreathing system is alarms.
connected, see page 74.
Activating
CAUTION
1 Open the Alarms dialog window.
Risk of gas contamination
2 Touch the Settings tab (A).
The extracted sample gas is also returned to the
internal breathing system during operation with Alarms
10465
an external fresh-gas outlet if the Perseus A
sample gas measurement is used.
Use breathing circuit to close the breathing
system or set the APL valve to spontaneous
breathing. Flush the breathing system each time
patients or anesthetic gas are changed!
B C
Conducting the fresh gas to the external outlet
1 Start Ext. FGO operating mode.
2 Adjust the fresh-gas delivery. Set vaporizer as
required.
3 For Cardiac bypass mode (CBM), touch the
Strictly observe the instructions for use of the non- button On (B).
rebreathing system.
Deactivating
Activating or deactivating the CBM For Cardiac bypass mode (CBM), touch the
mode button Off (C).
Or
The CBM mode allows patient monitoring without In the main menu bar, touch the Exit CBM
unnecessary alarms during extracorporeal button.
oxygenation of the patient by a heart-lung
machine.
Properties of CBM mode:
– All gas concentrations are measured
independent of the respiratory phase.
– The CO2 apnea and pressure apnea alarms
are inactive.
The CBM mode can be used in all active
ventilation modes.
Activating
Start Monitoring operating mode.
Deactivating
After Standby or change to ventilation mode.
For more information, see page 162.
Maneuvers (optional)
Perseus has various maneuvers for lung Inspiration hold, expiration hold
recruitment. During a maneuver, various data
concerning the lung mechanics are displayed so Perseus provides functions whereby a breath can
that the user can assess the progress of the be initiated or extended, or expiration can be
maneuver. extended.
The reminder function (option) reminds the user to This can be useful in situations where the lungs of
perform a maneuver. A reminder is issued after the the patient are not supposed to move, e.g., during
first change to a ventilation mode with medium or use of imaging techniques.
high breathing support and also at settable
intervals after the ending of a maneuver. The Procedures
31824
Consider recruitment message is displayed in
the waveforms for flow and pressure.
The use of lung recruitment maneuvers is the sole
responsibility of the user.
Dräger recommends that the patient's
hemodynamics always be monitored while the
maneuvers are being performed.
Available maneuvers
– Insp./Exp. hold C
– One-step recruitment D
– Multi-step recruitment
1 Open the Procedures dialog window. Manual inspiration/inspiration hold
2 Touch the tab for the required maneuver. This maneuver is available in the volume-
controlled modes, the pressure-controlled modes,
Procedures
31825
31825
patient category
31826
that the set values for
both PEEP max and
Pinsp max can be
reached.
K L M
U Breaths@Max Number of breaths at the
Pinsp max pressure level
V Breaths/Step Number of breaths after
T U V W which the next pressure
N
step begins
Q O
S R W Reminder Time after which a
P reminder for a new
maneuver is issued
K Trends of the measured values of pressure,
compliance, or tidal volume 3 Touch the Start button (N) and confirm.
L Predicted trends of inspiratory pressure and During the maneuver, the inspiratory pressure is
expiratory pressure during the maneuver automatically increased in steps to the set
("what if..." function) maximum Pinsp max value and the expiratory
pressure is increased to the maximum PEEP max
The exact progress of the maneuver is influ-
value. Both pressures are held at the highest
enced by the following values:
pressure level for a certain number of breaths
– Parameters entered in this dialog Breaths@Max and then reduced again in steps.
– Values set for RR and PEEP in the therapy The level of the pressure rise and the pressure
bar decrease is dependent on the selected patient
category. The appropriate settings are made in the
M Values of pressure, compliance, or tidal vol- system setup.
ume
The Paw high alarm limit is checked at the start of
1 Using the Pinsp max (S) and PEEP max (R) the maneuver. If the alarm limit is too low, it is
therapy controls, set the pressures that are to changed to a value 5 hPa above the set pressure.
be reached during the maneuver. After the maneuver, the alarm limit is reset to its
original value.
Maneuver
Insp./Exp. hold
Alarm One-step recruit- Multi-step recruit-
Man. insp./Insp. ment ment
Exp. hold
hold
Pressure alarms:
Alarm limit is
Airway pressure Alarm limit is adjusted
--- --- adjusted to
high to Pinsp max + 5
Pressure + 5
When set to Auto, the
alarm limit is automat-
ically adjusted to half
When set to Auto,
way between PEEP
the alarm limit is
Airway press. con- max and Pinsp max.
tinuously high Stopped1) --- automatically
Example:
adjusted to
PEEP max = 20 hPa
Pressure +3
Pinsp max = 40 hPa
Adjusted alarm limit =
30 hPa
Airway pressure
not achieved Stopped1) Stopped1) Stopped1) ---
Volume alarms:
Inspiratory tidal vol- Suppressed in VC Suppressed in VC Suppressed in VC
---
ume high and VC - AF and VC - AF and VC - AF
Minute volume low Displayed no sooner than 90 seconds after the start of the maneuver.
Apnea alarms:
Apnea (no pres-
sure)
Apnea (no flow) Stopped1) ---
Apnea (no CO2)
Apnea
1) Existing alarms are preserved. No new alarms will be issued.
12203
in et The values specified in the table apply to a patient
age of 40 years.
O2
1 MAC corresponds to:
N2O (in 100 % O2)
Halothane 0.77 Vol%
Iso Enflurane 1.7 Vol%
Isoflurane 1.15 Vol%
Desflurane 6.0 Vol%
Sevoflurane 2.1 Vol%
N2O 105 Vol%
xMAC Age years
The age-corrected MAC values are calculated
The MAC value is a simple navigation aid for according to the equation of W.W. Mapleson
anesthetic agent delivery. (British Journal of Anaesthesia 1996, pp. 179-
185).
Perseus displays the measured inspiratory and
expiratory values for O2, N2O, and anesthetic The equation applies to patients older than 1 year.
gases, and the xMAC in the monitoring area.
Measured nitrous oxide concentration or MACage corrected = MAC1) x 10(–0.00269 x (age –40))
anesthetic gas concentration is only displayed
when it is not zero.
The xMAC is the MAC multiple calculated from the
current expiratory measured values and the age-
dependent MAC values. If no respiratory phases
are detected, expiratory values and xMAC cannot
be displayed.
The integrated MAC algorithm is based on the
MAC values shown in the following table. These
values are guiding values only. The binding values
are specified on the package information leaflet of
the anesthetic agent.
1) 40 years
Perseus offers the option of displaying predictive Using the anesthetic gas prediction
concentrations of anesthetic agent in the breathing
gas. This requires the Vapor View option and Prerequisite: The Agent prediction parameter
Dräger Vapor 3000 or D-Vapor 3000. field must be configured in the monitoring area and
respiratory phases must be detected.
The anesthetic gas prediction is appropriate for
patients with the following data only:
10453
Iso prediction xMAC
Height: 150 to 200 cm
(59.1 to 78.7 in) E
Weight: 40 to 140 kg
(88 to 308 lbs) C D
Age: 18 to 90 years
WARNING
10452
FiO2 prediction
Risk of incorrect therapy settings D
The anesthetic gas prediction is based on
mathematical models and does not provide
individual patient values. A B
Do not make therapy decisions based solely
C
on the displayed anesthetic gas prediction.
The following is displayed:
The anesthetic gas concentration of the currently A Section for the trend
delivered anesthetic agent (or of the measured
primary anesthetic agent if delivery is switched off) B Section for the prediction
is displayed as a color diagram. The C Current O2 fresh-gas concentration
corresponding expiratory concentration is shown
D Prediction curve ("what if..." function)
by the color coding for the anesthetic agent.
The previously measured concentrations are WARNING
displayed in the left-hand section (A) and the
predicted concentrations in the right-hand section Risk due to incorrect O2 prediction
(B). The predicted concentrations are dependent Therapy decisions should not be made solely
on the vaporizer setting, the set fresh-gas flow, on the basis of a displayed O2 prediction.
and on various measurements.
Electronically controlled gas mixer only: The inspiratory O2 concentration is displayed as a
During the setting procedure, 2 dotted lines (D) are line.
displayed for the fresh-gas flow (the so-called The previously measured concentration is
"what if..." function). When the value is altered with displayed in the left-hand section (A) and the
the therapy control, the lines move and show a predicted concentration in the right-hand section
preview of the predicted concentration. (B). The predicted concentration is dependent on
the set O2 flow and the measured patient values.
10451
bag for sufficient filling. Stopwatch
A
10512
Surplus
Econometer Timer
20070
The bar graph indicates the qualitative fresh-gas
supply.
The volumeter can be used to monitor and assess The low-flow wizard displays bar graphs for the
the ventilation during spontaneous breathing, required fresh-gas flow and the total flow. Both bar
manual ventilation, or mechanical ventilation. graphs are to the same scale. The required fresh-
gas flow is calculated on the basis of balancing the
gas volumes.
10450
Volumeter
B
20152
Required FG flow
F VT A
C
Volume
Total flow
D Efficient
E
An evaluation of the total flow is displayed below
The bar graph indicates the inspiratory and the bar graph:
expiratory tidal volume.
At the end of the inspiration, the delivered tidal Evaluation Color Meaning
volume is displayed as a bar (A). Too high Yellow The fresh-gas flow is
At the end of the expiration, the difference between possibly too high. If the
the inspiratory and the expiratory tidal volume (F) fresh-gas flow can be
is displayed. reduced, both fresh gas
and anesthetic agent will
The expiratory tidal volume is displayed next to the be saved.
bar graph (C).
Efficient Green No action is necessary.
Too low Red The fresh-gas flow is too
Using the volumeter (minute volume low.
measurement)
Check the fresh-gas
1 Touch the Start button (B) to start. flow.
The bar graph displays the individual
Check the position of
measured breaths in segments (E). The
the breathing bag.
summed volume (D) is displayed next to the
bar graph. Refill bag Red The fresh-gas flow is too
2 Touch the Stop button (B) to stop. low.
3 To reset the volumeter and time display to zero, Refill the breathing
touch the Reset button (B). bag immediately,
e.g., with O2 flush.
The volumeter stops automatically after
60 seconds. The measured values are displayed An insufficient fresh-gas flow may trigger the
for 4 minutes and then deleted. Fresh gas low or leakage or Emergency air inlet
activated alarms, for example.
Views
10475
Perseus offers two view settings for the monitoring
area:
– Standard view A
10474
A B
D E F G
Changing with the View button
1 2 3
In addition to the standard view, the four additional Touch the View button.
parameter fields (D), (E), (F), and (G) are 1 2 3
View button.
1 2 3
Two buttons are available for changing the current The screen displays the first view .
view:
– Views... button
1 2 3
– View button
The monitoring area can be customized during The following loops are available:
operation: – Pressure-Volume loop
1 Touch a waveform or parameter field. The – Flow-Volume loop
Quick setup window opens.
10476
10402
Quick setup
B C
A B D
1 Open the Views dialog window.
2 For Content (A), select the desired content
from the list. 2 Touch the Loops button.
For a list of the possible screen content, see The following is displayed:
page 173. A Current loop and the last 5 loops
3 For Scale (B), select the desired scale. B Parameters:
– Cdyn
WARNING
– R
Risk of insufficient monitoring – TC
National and medical regulations require The area (E) can be configured so that the
certain parameters to be displayed. Pressure-Volume loop is displayed.
Always consider the relevant regulations
31735
Displaying or deleting reference loops Adjusting the sweep speed and the
To save a reference loop, touch the Save ref. scale
button (C).
Or 1 Open the System setup dialog window.
To delete the reference loop, touch the Delete 2 Touch the Screen layout > Waveforms
ref. button (D). tab (A).
10485
Quick setup window. A
B
Displaying mini-trends
C A
Mini-trends (B) can be displayed for the C C
waveforms (A).
C C
10477
B A
Setting the sweep speed
Touch button (B). Set the sweep speed.
Changing the color scheme and the 3 To change the color scheme, touch one of the
screen brightness buttons at (B).
4 Set the screen brightness at (C).
1 Open the System setup dialog window.
2 Touch the Screen layout > General tab (A).
System setup
10479
A
B A
During operation, measured values for ventilation, The logbook can save up to a maximum of 20000
gases, and the device are available in tabular entries. Logbook data are displayed in table form.
overviews. 1 Open the Trends/Data dialog window.
1 Open the Trends/Data dialog window. 2 Touch the Logbook tab (A).
2 Touch the Values tab (A).
Trends/Data
10480
B
E
The entries in the logbook cannot be deleted and Moving the cursor
are retained even after the device has been
switched off and on again or following a power The exact measured values for a specific point in
supply failure. When the storage limit is reached, time can be displayed numerically in area (F). To
the oldest entries are overwritten. do so, move the cursor.
Use the rotary knob to move the cursor (G).
Creating entries and associated settings, see
page 115. Or
Touch the corresponding area on the screen.
Trends
Displaying installed options
Trends are displayed in the form of a graphic or a
table. Listing of the additionally installed software
1 Open the Trends/Data dialog window. options.
2 Touch the Graphical trends (A) or Tabular 1 Open the System setup dialog window.
trends tab (B). 2 Touch the Licenses/Options tab.
The following illustration shows the graphical
trend:
Displaying an overview of the
Trends/Data
10481
10482
breathing sound volume
A
WARNING B
C A
Risk of operation error
The acoustic alarm signals might not be
heard if functions such as "Breathing sound"
are used or when operating in a noisy
environment.
Always set the alarm tone to be sufficiently
loud.
Setting via the Alarms dialog window Activating or deactivating CO2 alarms
In the Alarms dialog window, the alarm limits can
be set either manually or automatically. The CO2 monitoring (affects the alarms for inCO2,
etCO2, and CO2 apnea) can be activated or
1 Open the Alarms dialog window. deactivated.
2 Touch the Limits tab (C). Deactivation is indicated in the header bar and in
Alarms
the parameter field by the symbol.
10483
C 1 Open the Alarms dialog window.
2 Touch the Settings tab (A).
Alarms
10465
A
D
B
E
F F F F G
Manual setting
1 Set the upper alarm limits (D).
2 Set the lower alarm limits (E).
3 For CO2 alarms, touch the button (B):
On: Alarms are activated.
Automatic setting Off: Alarms are deactivated
Alarm limits can be adapted to current measured Or
or setting values. Use the CO2 alarms off button in the main
1 Touch the Autoset limits button in the main menu bar to activate or deactivate the alarms.
menu bar. This button is only visible in the following
The Autoset all button (G) is selected. ventilation modes:
2 Confirm to adapt the alarm limits for all – Manual / Spontaneous
parameters. – External fresh-gas outlet
Or – Pause
To adjust the alarm limits for an individual
parameter, touch one of the buttons Autoset The alarm system is immediately activated when
(F) and confirm. the CO2 monitoring is activated.
[0$& LQVS[0$&
H[S[0$&
upwards and downwards so that no alarm is
issued during this time. Similarly, the value is not
limited to 1.0 during this time.
ORZHUDODUPOLPLW
A BC DE F GH
Patient data can be changed during operation. 2 Make the desired changes.
1 Open the Patient dialog window. Changes influence, e.g., therapy suggestions, as
is recognizable by the position of the arrow at
Patient
the therapy controls.
10484
The current therapy settings remain unaffected.
When patient category is changed, age, weight,
and height are adapted as needed so they remain
within the described limits, see page 157.
Exporting data
Prerequisite: USB flash drive is connected to the In the Standby mode, the following data can be
USB interface. exported to a USB flash drive:
During a saving process, the button is displayed as – System test results
activated (dark green). – Logbook
The data are stored in the "Draeger\ExportData" Selection from the following time periods is
directory. possible:
– Last case
– Today
Exporting the screen contents – All
– Alarm history
The screen contents can be exported to a USB
flash drive as a screenshot. – Trends
Touch the Export screenshot button in the
main menu bar.
The screenshot will be saved as a ".png" file.
1 Open the Trends/Data dialog window. 3 Touch the desired button (B).
2 Touch the Export tab (A). The data will be saved as a ".txt" file.
Trends/Data
10496
A
B
B
B
Other settings
Switching the breathing system warmer When switching to Standby mode, the warmer is
on or off reset to the value configured in System setup.
CAUTION
The breathing system warmer should only be
switched off in special situations (e.g., for Risk due to faulty or switched-off breathing
intentional reduction of the body temperature of system warmer
the patient).
Increased condensation or accumulation of water
1 Open the System setup dialog window. in the breathing system and the hoses may
2 Touch the Therapy tab (A). occur.
System setup
Increase fresh-gas flow as required. Remove
10513
CAUTION
Risk of inaccurate measured values
The accuracy of the flow measurement may be
compromised if the breathing system warmer is
switched off.
When the breathing system warmer is switched
off, do not make therapy decisions solely on the
3 To switch the warmer on or off, touch the
basis of the displayed values for flow and
corresponding button (B).
volume.
10486
A
A
10359
– Prevent leakage, e.g., at endotracheal
tubes, laryngeal mask airways, face
masks, Y-piece, breathing system
including hoses, filters, and breathing
bag, at the external fresh-gas outlet, and
at the outlet for O2 insufflation. B
– Use only intact and leak-free hoses at the C
outlet for O2 insufflation. A
– Before beginning laser surgery or
electrosurgery, flush with sufficient air
(<25 % O2), and flush beneath the surgical Mechanically controlled gas mixer:
drapes as well.
– Close the flow control valve on the O2
20364
Prerequisite:
– The appropriate accessory is connected to the
outlet for O2 insufflation (A).
– With electronically controlled gas mixer:
The O2 switch is horizontal in the Aux. O2
position (C).
Beginning O2 insufflation
Open the flow control valve (B) of the
O2 flowmeter.
Ending O2 insufflation
Close the flow control valve (B) of the
O2 flowmeter.
Change of patient
WARNING
Risk due to residual concentrations of
anesthetic agent
Even after flushing the breathing system and
breathing circuit, residual concentrations of
anesthetic agent might not be sufficiently
reduced.
For patients suspected of having malignant
hyperthermia, do not rely solely on the
flushing function.
CAUTION
Risk of impaired ventilation due to faulty drying
function
The breathing system and breathing hoses may
still retain traces of moisture even after the
flushing and drying functions are performed.
Check the breathing system and breathing hoses
for condensate.
Tests
10411
Yellow A non-critical fault has been detected.
Test results Available tests
The device can be operated with
restricted function.
A Red A serious fault has been detected.
Operation is not possible or is forbid-
den.
Gray Function not tested.
CAUTION WARNING
Risk due to device malfunction and/or patient Risk of patient injury
injury
During the system test, the system is
If the system test is canceled, it is possible that pressurized.
some malfunctions might not be detected. Thus,
To prevent patient injury, do not perform the
more attention is required during operation.
system test on the medical device if a patient
Perform the system test every day. If the system is connected.
test is canceled during execution, perform it
again as soon as possible.
Special characteristics of Dräger coaxial 3 If all components are ready for operation, touch
hoses the button (B). The automatic test will start.
If a component is not ready for operation, touch
the button (C).
When using coaxial breathing hoses, always use The walk-through mode will start.
the leakage assistant to check the inner hose of
the coaxial breathing circuit for leakage prior to the
31887
system test or leakage test, see page 125. System test
System test
H
D
A G
After the test, the final test result is displayed on Leakage assistant
the Standby page, see page 81.
Prerequisite when vaporizers are connected:
Test interruption due to irregularities – Vaporizer is vertically level and securely
mounted on the plug-in adapter.
If an irregularity is detected during the automatic
test, the following occurs: – Filling inlet is closed.
– The test is interrupted. In this test, a continuous pressure is generated
– An acoustic signal sounds. and the current leakage value is displayed.
10514
displayed. Leakage assistant
Test details
A
10510
I
If the leakage value changes as a result of
J changes made to the device (e.g., loosening or
readjusting hose connections), this can help locate
Remedying the cause:
the cause of the leakage.
1 Remedy the cause of the interruption.
If there is leakage, the leakage assistant (K) The displayed leakage value may differ from the
can be used to support troubleshooting. value that was determined in the leakage test. The
reasons for this are the different measuring
2 Touch the Repeat button (I) and repeat the test methods and different pneumatic ranges.
of the component.
1 Touch the button for the test (Leakage
Accepting the irregularity: assistant).
Touch the Accept button (J) and continue the 2 Follow the instructions (A).
test.
Accepted irregularities prevent the total result of Checking a vaporizer for leakage
the system test from indicating "fully operational"
and are protocoled in the logbook. Prerequisites:
– Gas mixer:
– With electronically controlled gas mixer:
The flow control valve of the O2 flowmeter
is closed.
– With mechanically controlled gas mixer:
All flow control valves are closed.
– The vaporizers are closed.
– The breathing circuit is correctly connected.
– The APL valve is set to 30.
31878
processed, using as an example a device with an
electronic gas mixer and factory defaults.
The scope of the checklist can vary due to differing
system settings.
The instructions on the screen take precedence.
Prerequisites
31875
31879
A
5 Set the O2 switch to Aux. O2.
1 Connect the hoses (A).
2 Seal the Y-piece (B).
31876
20 30 40
31861
filling level has fallen below the refill mark.
31864
1 The locking lever points left, indicating the
vaporizer is locked.
3 The control dial is set to position 0 and the key
31863
is locked in placed.
31862
31868
WARNING
Risk due to impermissible gas supply
pressure
Impermissible gas supply pressures may
cause incorrect gas composition.
Check the supply pressures of the central gas
supply and of the gas cylinders before
operation.
WARNING
Risk due to impermissible gas supply
Using oxygen supplies with less than 100 %
O2 may cause incorrect gas composition. 1 Open the gas cylinder valves slowly.
Check that the displayed pressures are
When using O2, only use 100 % O2. sufficient.
When using pressure reducers without
WARNING electronic pressure measurement, read the
pressure from the pressure gauge.
Danger when using O2 concentrators
2 Close the gas cylinder valves.
The following effects may occur: On devices that are equipped with Advanced
– Discrepancies between the set value and Cylinder Support, the gas cylinder valves can
the actual value for fresh-gas flow and O2 remain open during operation. These devices are
concentration in the fresh gas identified by an appropriate label near the gas
– Inaccurate measured values for volume, inlets, see page 28.
anesthetic agent consumption,
econometer (optional), and low-flow
wizard
– Inaccuracy in the FiO2 prediction
– Inaccurate volume delivery in volume-
controlled ventilation modes
– Accumulation of argon in low-flow
operation and minimal-flow operation
Do not use any O2 concentrators.
31865
electronically controlled gas mixer)
Prerequisite: Y-piece is sealed.
10375
D C
A B
3 Close the flow control valve (C).
4 Reset O2 switch (D) to the Aux. O2 position.
1 Set O2 switch (A) to the Add. O2 position. Checking the flow control valves (mechanically
2 Open the flow control valve (B) of the O2 controlled gas mixer)
flowmeter and set the desired O2 flow.
42690
31866
Breathing circuit
31899
Prerequisite:
– Breathing system is complete and locked.
– Breathing system cover is in place.
31900
E
G F Valves
31901
2 Extendable hoses (F) are extended to the
foreseen application length.
NOTE
Do not change the length of flex hoses after the
test is done.
Components
31866
Loudspeakers
31907
Touch the button and wait for 2 acoustic
signals.
NOTE
If the acoustic signals are not emitted, contact
service personnel.
Bronchial suction
The breathing bag fills and the inflow of gas is
audible.
31905
31902
20 30 40
31906
range").
When the flow indicator is floating in range (C)
("restricted range"), certain fresh-gas flow rates
should not be exceeded, see "Anesthetic gas
receiving system" in chapter "Technical data".
This can prevent contamination of the ambient
air.
Accessories
Soda lime
1 Make sure there is a manual resuscitator on
the device.
31922
2 Check the functional integrity of the manual
resuscitator.
Sample line
WARNING
Risk due to soda lime drying out
31923
The soda lime loses moisture. If the moisture
level falls below the minimum moisture level,
the following adverse reactions occur
regardless of the type of soda lime and the
inhalational anesthetic agent used:
Decreased CO2 absorption, increased
generation of heat in the CO2 absorber
resulting in increased breathing gas
temperature, formation of CO, absorption
and/or degradation of the inhalational
anesthetic agent.
– Do not use unnecessarily high fresh-gas
flows.
– Only use the O2 flush if necessary.
– With electronically controlled gas mixer:
Only use the emergency O2 delivery if
necessary.
– With mechanically controlled gas mixer:
Do not leave the flow control valves open
unnecessarily long. Connect the sample line.
NOTE
Only perform leakage tests with CLIC absorber
locked into place because this affects the system
compliance values.
Water trap
31921
Ending operation
To store Perseus
1 Set the main switch to position 0.
2 Disconnect from mains power supply if
necessary.
Alarms
Displaying alarms
In the event of an alarm, the system displays the Depending on the overall alarm situation, it is
relevant alarm message in the alarm message possible that the 10-tone sequence for the high
field (A). The parameter field (B) for the alarm- alarm priority is only issued as a 5-tone sequence
generating parameter flashes. due to the coinciding occurrence of alarms.
Regardless of the set alarm volume, the No O2
A
10490
B B
B
B
B
B
Alarm priorities
Response to alarms
Displaying information on alarms 3 Refer to the information under Cause (C) and
Remedy (D) to remedy the error.
1 Touch the alarm in the header bar. A list of all possible alarms can be found in chapter
Or "Troubleshooting", see page 191
Open the Alarms dialog window and touch the
Current alarms tab (A).
Alarms
10488
A
B
C D
The alarm tone can be suppressed for a maximum Use "ALARM RESET" Alarm priority is
of 2 minutes. to downgrade alarm changed to low.
priority.
Use "ALARM RESET" Alarm is reset.
10362
to acknowledge alarm.
10495
A
Or
Press the Audio paused key (A) below the Option 2:
screen, see page 19.
Alarms
10488
The symbol and the remaining time for the
suppressed alarm tone are displayed in the header B
bar.
After the alarm suppression has expired, the alarm
tone immediately resumes if the cause of the
alarm still exists.
While the alarm tone is suppressed, only those
alarms with a higher alarm priority or internal
priority rating are issued acoustically, see
page 191.
C
Reactivating the alarm tone
Option 1 Option 2
Press the Audio paused key again.
Touch the ALARM In the Alarms > Cur-
RESET button (A) in the rent alarms dialog
Downgrading and acknowledging alarm header bar and confirm. window (B), touch the
Reset all button (C) and
messages confirm.
Some alarms can be downgraded to low priority or All the alarms displayed All alarms will be down-
can be cleared completely. The relevant alarms in the alarm message graded or reset.
can be identified in the table "Alarm – Cause – field will be downgraded
Remedy" on page 191 by the following remedial or reset.
messages:
After an alarm is downgraded, the respective
priority rating is maintained , see table "Alarm –
Cause – Remedy" on page 191.
10492
1 Open the Alarms dialog window.
2 Touch the Alarm history tab (A).
Alarms B
10489
A
When the ventilation mode is changed, the alarm Some modes can be configured whether or not the
settings are also adapted. settings are adopted.
Depending on the mode, alarm settings can either However, the settings can be adjusted at any time
be adopted or set to Off. during operation.
Activating the alarms related to volume – No sooner than 60 seconds after an Apnea
(no flow) or Apnea (no pressure) alarm.
The upper alarm limit for MV is deactivated at the In volume-controlled ventilation modes, the alarm
start of the ventilation, but can be set during the limit for the inspiratory tidal volume is automatically
ventilation. The upper alarm limit for VTi is set to 130 % of the set tidal volume.
automatically set to 130 % of the set VT in volume-
controlled ventilation modes. In pressure-
controlled ventilation modes, it is deactivated at
the start of the ventilation, but can be set during
Resetting the Apnea (no CO2) alarm
the ventilation.
When changing to a ventilation mode with higher
The MV low alarm is delayed in certain cases and respiratory support, the Apnea (no CO2) alarm is
is issued as follows: reset. If the apnea situation persists, an alarm
– No sooner than 90 seconds after a case starts, appears after the time specified in the table "Alarm
delay and alarm escalation".
– No sooner than 60 seconds after changing to a
mode with greater respiratory support, see
page 259.
Priority
Alarm
Low Medium High
inCO2 high
etCO2 high After two successive
etCO2 low --- respiratory phases ---
FiO2 high and 15 seconds
Inspiratory N2O high
After two successive
inAgent low respiratory phases --- ---
and 15 seconds
After two successive
respiratory phases
and 15 seconds
FiO2 low --- --- or
after 30 seconds if no
respiratory phases are
detected
Priority
Alarm
Low Medium High
After two successive
respiratory phases
and 15 seconds
inAgent high --- or >165 seconds later
after 30 seconds if no
respiratory phases are
detected
longer than
30 seconds:
insp. MAC
≥
insp. MAC 3
≥ and
Inspiratory xMAC high --- 3 exp. MAC
for more than ≥
180 seconds 2.5
or
insp. MAC
≥
5
xMAC low 0 to 60 s >60 s ---
at the latest after
20 seconds
(for RR
≥
6) 15 seconds later
or (for RR
Apnea (no CO2)
Apnea (no flow)
At the latest after ≥
--- 35 seconds 6)
Apnea (no pressure)
(for RR <6) or
Apnea
or 30 seconds later
at the latest after (for RR <6)
65 seconds for the
Pause, Manual /
Spontaneous and Ext.
FGO modes
At the latest after 20 seconds
(15 seconds for RRmin
Apnea Ventilation ≥ ---
4)
(configurable, see page 155)
No CO2 detected >60 s --- ---
Inspiratory tidal volume high After 3 successive
--- ---
Tidal volume not achieved breaths
Priority
Alarm
Low Medium High
> 15 seconds above
Airway press. continuously high --- --- the manually or auto-
matically set limit
Pmean < –2
Airway pressure negative --- --- or
Paw < –10
Airway pressure
>(PEEP +5 hPa)
PEEP/CPAP high --- ---
during more than 10
consecutive breaths
>15 seconds (for RR
≥
6)
Airway pressure not achieved --- ---
or
>30 s
(for RR <6 or PSV)
If a minute volume of
>50 % of the sug-
Cardiac bypass mode still gested value is mea-
--- ---
active? sured after CBM
mode has been acti-
vated for >60 seconds
after 30 s
or
in the event of the
additional alarm
"Apnea (no flow)" or
Breathing bag almost
Fresh gas low or leakage Breathing bag empty "Apnea (no pressure)"
empty
or
in the event of the
additional alarm
"Emergency air inlet
activated"
O2 measurement not available
N2O measurement not available
Agent measurement not avail-
able
O2 measurement temporarily
inaccurate
CO2 sensor accuracy low >20 s --- ---
Agent measurement temporar-
ily inaccurate
N2O measurement temporarily
inaccurate
Measured gas concentrations
temporarily inaccurate
If no breaths have been detected even after the CO2 waveform, and the symbol "Alarm
leaving the Standby or Pause modes, the temporarily inactive" is displayed beside the
breathing gas is monitored with regard to an O2 relevant parameters.
concentration that is too low or an anesthetic gas
Once 2 breaths have been detected, the message
concentration that is too high, regardless of the
and the symbol disappear. Only then, the
respiratory phase. At the same time, the message
respiratory-phase-dependent O2, CO2, N2O, and
Waiting for respiratory phases is displayed in
anesthetic gas alarms are completely activated.
Combined alarms
Suppressed alarms
Configuration
Device settings
changed manually:
Touch field (A).
Or
Set the date and time in the system setup.
The source for the time synchronization can also
be set in the system setup. For more information,
see page 164.
Standby
Setting range
Headline/
Description
Parameter
Color scheme Day light; Day dark; Night Set the color scheme, see page 43
Screen brightness 10 to 100 Set the screen brightness.
80
Setting range
Headline/
Description
Parameter
Rename views 1 Standard; 2 Expert; 3 Expert Specify names for screen layouts.
Touch the button with the respec-
tive screen layout, enter the new
name on the screen keyboard, and
confirm with the button or with
the rotary knob.
Number of wave- 3; 4
forms (view 3)
Default view 1 Standard; 2 Expert; 3 Expert Specify the standard view.
Save as system Current view Save current screen layout.
defaults (only available during operation)
All views Save all screen layouts.
Setting range
Headline/
Description
Parameter
Sweep speed 6.25; 12.5; 25 6.25; 12.5; 25 6.25; 12.5; 25 Specify sweep speed.
[mm/s]
VT scale Auto; Auto; Auto; Specify scale for volumeter.
[mL] 0 to 10; 0 to 10; 0 to 10;
0 to 50; 0 to 50; 0 to 50;
0 to 150; 0 to 150; 0 to 150;
0 to 500; 0 to 500; 0 to 500;
0 to 1000; 0 to 1000; 0 to 1000;
0 to 2000 0 to 2000 0 to 2000
Flow scale Auto Auto Auto Specify scale for flow waveform.
[L/min] -5 to 5 -5 to 5 -5 to 5
-10 to 10; -10 to 10; -10 to 10;
-30 to 30; -30 to 30; -30 to 30;
-60 to 60; -60 to 60; -60 to 60;
-120 to 120 -120 to 120 -120 to 120
O2 scale Auto; Specify scale for O2 waveform.
[%] 0 to 100; 15 to 35; 25 to 45; 35 to 55;
45 to 65; 55 to 75; 65 to 85; 75 to 95;
85 to 105
CO2 scale [%]; [kPa]: Auto; 0 to 6; 0 to 12 Specify scale for CO2 waveform.
[mmHg]: Auto; 0 to 50; 0 to 100
Paw scale Auto; -5 to 20; -7.5 to 30; -10 to 40; -20 to 80 Specify scale for the Paw wave-
[mbar]; [hPa]; form.
[cmH2O]
Agent prediction Full range; 0 to 2 xMAC; 0 to 1 xMAC Specify scale for anesthetic gas
scale prediction.
Flow-volume loop ISO standard; Dräger Specify coordinate axes for
flow/volume loop display.
Setting range
Headline/
Description
Parameter
CO2; Default color; color palette with 7 additional Specify parameter colors.
Paw; colors
Flow, volume
O2; Default color; ISO color
Agent
Setting range
Headline/
Description
Parameter
Alarm limits for ventilation and gases for each patient category
FiO2 [%] 19 to 99; 19 to 99; 19 to 99; Inspiratory oxygen fraction
Off Off Off
90
FiO2 [%] 18 to 98 18 to 98 18 to 98
20 20 20
etCO2 Expiratory CO2 concentration
[%]; [kPa] 0.1 to 9.8; 0.1 to 9.8; 0.1 to 9.8;
Off Off Off
7.0 7.0 7.0
[mmHg] 1 to 75; 1 to 75; 1 to 75;
Off Off Off
53 53 53
etCO2
[%]; [kPa] Off; Off; Off;
0.0 to 9.7; 0.0 to 9.7; 0.0 to 9.7;
[mmHg] Off; Off; Off;
0 to 74 0 to 74 0 to 74
inCO2 Inspiratory CO2 concentration
[%]; [kPa] 0.1 to 1.4; 0.1 to 1.4; 0.1 to 1.4;
Off Off Off
1.0 1.0 1.0
[mmHg] 1 to 10; 1 to 10; 1 to 10;
Off Off Off
8 8 8
Paw high 5 to 99 5 to 99 5 to 99 Airway pressure
[mbar]; [hPa]; 40 25 20 If the airway pressure is above the
[cmH2O] set value for Paw low for longer
than 15 s, the alarm Airway press.
Paw low Auto; Auto; Auto; continuously high is issued. If the
[mbar]; [hPa]; 3 to 97 3 to 97 3 to 97 setting Auto is selected, the lower
[cmH2O] alarm limit is automatically
adapted to the set value for PEEP.
inSev 0.10 to 9.95 0.10 to 9.95 0.10 to 9.95 Sevoflurane
[%]; [kPa] 4.40 5.10 6.70
Setting range
Headline/
Description
Parameter
Alarm limits for ventilation and gases for each patient category
inDes 0.1 to 20.0 0.1 to 20.0 0.1 to 20.0 Desflurane
[%]; [kPa] 12.5 14.5 19.0
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Deactivate the alarm limit when activating Man/Spon, CPAP, The CPAP and CPAP / PSV
CPAP/PSV with ∆Psupp < 5, or Pause? modes are optional.
FiO2 high Yes; No Yes; No Yes; No Specifies the alarm behavior when
changing to a different ventilation
MV low Yes; No
mode.
MV high Yes; No These settings apply only to a
change to a ventilation mode with
xMAC low Yes; No
lower or no breathing support (see
etCO2 low Yes; No page 259).
The alarm behavior at the start of
etCO2 high Yes; No
the therapy is defined by the con-
inCO2 high Yes; No figuration in the vertical tab Alarm
limits.
inAgent low Yes; No
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO2 high Yes; No Specify alarm behavior in CBM
mode.
MV low Yes; No
MV high Yes; No
inAgent low Yes; No
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Default ventilation Buttons with available ventilation modes Specify the default ventilation
mode Man/Spon mode at start of therapy.
VT and RR start settings Specify the tidal volume and respi-
ratory rate.
Based on Patient category; Ideal body weight
Selected: Specify tidal volume and respira-
[Patient category] tory rate based on patient cate-
gory.
VT 20 to 2000 20 to 2000 20 to 2000
[mL] 500 150 50
RR 3 to 100 3 to 100 3 to 100
[1/min] 12 20 30
Selected: Specify the tidal volume and respi-
[Ideal body weight] ratory rate based on ideal body
weight.
VT 20 to 2000
Set the tidal volume and the respi-
[mL] 100 kg (220 lbs): 700
ratory rate for the supporting
75 kg (165 lbs): 520
points 5; 15; 75; 100 kg (11; 33;
15 kg (33 lbs): 110
165; 220 lbs).
5 kg (11 lbs): 35
For calculated values for ideal
RR 3 to 100 body weight values that lie
[1/min] 100 kg (220 lbs): 10 between these four supporting
75 kg (165 lbs): 12 points, the start settings for tidal
15 kg (33 lbs): 26 volume and respiratory rate are
5 kg (11 lbs): 32 interpolated linearly. For ideal
weight values lying outside these
supporting points, calculation pro-
ceeds with the values of the high-
est or lowest supporting point.
The start settings for VT and RR influence the start The "offset" value corresponds to the respective
values of the alarm limits for MV high, MV low, and offset setting for automatic alarm adjustment. The
VTi high: "offset" value can be set in the System setup >
Alarms > vertical tab Autoset limits.
MV high = VT x RR x (1 + offset);
minimal: 2.0 L/min The following applies in volume-controlled modes:
Setting range
Headline/
Description
Parameter
Start settings for ventilation Specify start settings for the venti-
lation.
Pmax 7 to 80 7 to 80 7 to 80
[mbar]; [hPa]; 40 30 25
[cmH2O]
Pinsp 3 to 80 3 to 80 3 to 80
[mbar]; [hPa]; 15 15 15
[cmH2O]
∆Psupp Off; Off; Off;
[mbar]; [hPa]; 1 to 80 1 to 80 1 to 80
[cmH2O] 10 10 10
Insp. term. 5 to 80 5 to 80 5 to 80
[%PIF] 25 25 25
PEEP Off; Off; Off;
[mbar]; [hPa]; 2 to 35 2 to 35 2 to 35
[cmH2O] 3 3 3
Slope 0 to 2 0 to 2 0 to 2
[s] 0.2 0.2 0.2
RRmin Off; Off; Off;
[1/min] 3 to 25 3 to 25 3 to 25
6 10 15
I:E 1:50 to 50:1 1:50 to 50:1 1:50 to 50:1
1:2.0 1:2.0 1:1.0
%Tplat 0 to 60 0 to 60 0 to 60
[%] 20 20 20
Trigger 0.3 to 15 0.3 to 15 0.3 to 15
[L/min] 4.0 2.0 1.0
Sync. On; Off On; Off On; Off
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
General settings
Default maneuver One-step recruitment; Multi-step recruitment; Defines the default maneuver.
Insp./Exp. hold
Reminder Off, 10 to 180 Off, 10 to 180 Off, 10 to 180 If the value is not set to Variable , a
[min] reminder for a maneuver is given
after the first switch to a controlled
ventilation mode. Defines the time
after which, following the end of a
One-step recruitment or Multi-step
recruitment maneuver, a reminder
for a further maneuver is given.
Layout
Displayed parame- Cdyn; VT Defines which additional parame-
ter ter is displayed in the One-step
recruitment and Multi-step recruit-
ment dialogs.
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Setting range
Headline/
Description
Parameter
Start settings for fresh gas Select start settings for the fresh-
gas delivery.
FG O2 21 to 100 21 to 100 21 to 100 Set the O2 concentration.
[%] 100 100 100
FG flow 0.20 to 15.00 0.20 to 15.00 0.20 to 15.00 Set the flow.
[L/min] 2.00 2.00 2.00
Minimal O2 flow Off, 50 to 300 Off, 50 to 300 Off, 50 to 300 Set the minimum O2 flow that is
(carrier gas: Air) Off Off Off delivered when Air is used as car-
[mL/min] rier gas.
Do not set this value too small;
recommended is, e.g., 200 for
adults, 100 for pediatric patients,
and 50 for neonates.
Minimal O2 flow 50 to 300 50 to 300 50 to 300 Set the minimal O2 flow that is
(carrier gas: N2O) 200 200 200 delivered when N2O is used as
[mL/min] carrier gas.
Do not set this value too small;
recommended is, e.g., 200 for
adults, 100 for pediatric patients,
and 50 for neonates.
Carrier gas Air; N2O Air; N2O Air; N2O Set the carrier gas.
Setting range
Headline/
Description
Parameter
Default selection for Continue case; Load patient data: The button that
"Start" dialog New adult; New pediatric; New neonate is selected here is preselected the
first time the Start dialog window is
opened or after changes to the
system setup.
Weight Specify start settings for:
[kg] 30 to 300 5 to 50 0.4 to 10 – Patient weight
80 25.0 3.0
[lbs] 67 to 661 12 to 110 0.9 to 22 – Patient height
176 55 6.6 – Patient age
Height If the Height parameter is config-
[cm] 120 to 300 50 to 300 Off; 20 to 80 ured to Off for the New neonate
185 100 Off patient category, the correspond-
[in] 48 to 118 20 to 118 Off; 8 to 31 ing setting control will not be dis-
73 39 Off played in the Start dialog window.
Age [years, Neo: 12 to 130 0 to 16 0 to 24
months] 32 8 6
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
General
Always use walk- On; Off When this function is switched on,
through mode system tests and leakage tests will
always be executed in walk-
through mode.
Sample line is con- On; Off Specify whether the sample line is
nected during test connected to the Y-piece or to the
filter on the Y-piece during the
automatic tests.
When On is configured, the device
can automatically check whether,
e.g., breathing hoses or central
supply hoses are incorrectly con-
nected.
When Off is configured, additional
checks are required during the
system test.
Test gas supply
Central O2 supply On; Off Select which gas supplies are
tested during the automatic sys-
Central Air supply On; Off
tem test.
Central N2O supply On; Off If Auto On is configured, only
check the gas supplies whose
O2 cylinder On; Off
actual availability is guaranteed
Air cylinder On; Off after the automatic start-up. Set-
ting both O2 sources to Off - in
N2O cylinder On; Off
order to make the system test suc-
ceed - is not possible.
Verify O2 delivery On; Off Select whether the system test will
check that the O2 supply is actu-
ally delivering oxygen.
The Sample line is connected
during test parameter must also be
set to On for this.
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
A logbook entry with measured values is created Create additional logbook entries
with measured values.
Every 1 min; 2 min; 5 min; 10 min; 15 min Create entries regularly.
For all high-priority On; Off Create entries for alarms.
alarms
For all medium-pri- On; Off
ority alarms
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Setting range
Headline/
Description
Parameter
Certain tabs in the System setup dialog window Further information is displayed in Standby >
have a button for resetting the respective start System setup > System status:
settings to the factory settings.
System setup
31778
Resetting the changes in a dialog window
1 Open the corresponding tab.
2 Touch the Factory defaults button and
confirm.
A
C A
D
B
The following software options require an Loading the activation code from a USB
activation code to be entered, followed by flash drive
activation:
– Pressure Support Prerequisite: A USB flash drive with valid licenses
– APRV is connected to the USB interface.
– FiO2 prediction Touch the Load from USB button (B).
– Econometer trend The activation codes are read and displayed in the
list (C).
– Low-flow wizard
– Breathing sound
– Lung recruitment Entering the activation code
– Lung recruitment reminder
1 Touch the Enter code button (D).
– Inspiration hold / Expiration hold
2 Enter the activation code and confirm with OK.
– Logbook export
The license is displayed in the list (C).
Trial licenses for these options are time-limited.
An activation code is linked with the serial number
of the respective device and cannot be transferred. Activating the licensed software option
The activation codes can either be loaded from a
USB flash drive or entered manually.
The licensed software options must be activated
Open page System setup > before they become available.
Licenses/Options > Licenses/Options (A).
1 Select the corresponding license from the
System setup list (C).
10501
D B
10505
Waveforms and associated parameter
fields
Volume MV, VT, RR
etCO2
Volume
10456
CO2
10454
10506
CO2
10502
O2
CO2 in/et, RR
O2
10457
CO2
10503
O2, ∆
Paw
O2
10507
Paw
10455
Primary agent
Paw (3)
Iso
10458
Volume-controlled modes:
Parameters PIP, Pplat, PEEP
All other modes:
Parameters PIP, Pmean, PEEP "Empty"
Paw
10504
10508
10416
Compliance trend CO2
et in
20035
Cdyn PEEP
RR
Paw
10419
Compliance Paw
PIP
33361
Compliance
Cdyn mean
Paw (3)
Volume-controlled modes:
Parameters PIP, Pplat, PEEP
etCO2
All other modes:
Parameters PIP, Pmean, PEEP
10420
CO2
10417
et Paw
PIP Pmean
PEEP
CO2 in/et
Paw (4)
10421
Paw
PIP Pplat
Pmean PEEP
10499
20036
MV ∆VT VTi
Mand Spon
VT
Paw
Volume
10418
10498
Volume
MV VT
VT MV
RR
RR
MV×CO2 trend O2
WARNING
10460
Risk of incorrect therapy settings
O2
The numerical value displayed for the
MV×CO2 parameter is not sufficiently
accurate to enable therapy decisions to be
made.
Do not make therapy decisions based solely
on the displayed numerical value. Only the
trend curve can be used for therapeutic O2, ∆
decisions.
10422
MV×CO2 O2
20033
in ∆
WARNING
10427
Econometer
Risk of incorrect therapy settings
The numerical value displayed for the O2
Deficit Efficient Surplus
uptake parameter is not sufficiently accurate
to enable therapy decisions to be made.
Do not make therapy decisions based solely
on the displayed numerical value. Only the
trend curve can be used for therapeutic
decisions. Low-flow wizard (optional)
20152
Required FG flow
20034
O2 uptake
Total flow
Efficient
10424
10423
N2O
Iso
RR Gas supply
10426
Gas supply
10449
RR
O2 Air N2O
13718
10452
Volumeter FiO2 prediction
Start
VT
Volume
10451
10425
Stopwatch Vaporizer sett.
Start
Iso Sev Des
Refill
Timer Empty
Timer
20070
10509
A B
Econometer trend
31702
WARNING
Risk of mix-up
In some countries, the representation and
order of the virtual flow tubes on the screen
may differ from that illustrated here.
Pay attention to the labeling of the virtual flow
tubes.
WARNING
Paw
Insufficient fresh-gas supply
The indication on the virtual flow tubes is
intended only as additional information.
Do not use the virtual flow tubes alone when
making therapeutic decisions.
10511
Surplus
Efficient
Deficit
O2 N2O Econometer
Flow tubes (mechanically controlled gas mixer) When the Agent prediction or PV loop parameter
field is displayed, the flow tubes will be displayed
WARNING at reduced size:
Risk of mix-up
20788
In some countries, the representation and
order of the virtual flow tubes on the screen
may differ from that illustrated here.
Pay attention to the labeling of the virtual flow
tubes.
O2 Air N2O
20367
O2 Air N2O
Troubleshooting
Leakage
Leakages may result in the system not being – The O-rings on the inspiratory port or
operational or being operational with limitations expiratory port are damaged, soiled, or
only. missing.
– The flow sensors are incorrectly installed or
WARNING damaged. The rear O-ring is missing.
Risk due to leaks – The upper part of the breathing system housing
is incorrectly fitted or damaged.
As a result of leakage, ambient air may get
into the breathing gas and cause a reduction – The valves or seals of the breathing system are
in the depth of anesthesia, or ambient air may damaged.
become enriched with anesthetic gas and – The circuit plug is scratched or damaged.
consequently put the user at risk.
– The filling or emptying connections on the
Perform the leakage test before using the vaporizer are leaking or are open. The
device. Reduce leakage to a minimum. vaporizer is incorrectly fitted. The O-ring is
missing or damaged. The control dial is not set
WARNING to the 0 position.
– The CO2 absorber or the CLIC adapter is not Breathing Disconnect the breathing hoses.
securely screwed to the breathing system. hoses Connect the inspiratory port and
expiratory port with a hose that
– The APL valve is not correctly fitted to the is known to be without leakages.
breathing system or is not set to 30 hPa Connect the breathing bag
(cmH2O). directly to the breathing system.
– The breathing bag, the breathing hoses, the Y- Vaporizers Remove the vaporizers.
piece, or the microbial filter is incorrectly fitted
or damaged. 1 Perform a leakage test, see page 123, and use
– The flexible arm for the breathing bag the leakage assistant if necessary.
(optional) is incorrectly fitted to the breathing 2 Contact service personnel if the leakages
system. The sealing ring is soiled or damaged. cannot be localized.
– The water trap is not connected.
– The sample line is not connected, is kinked, or
is leaking.
– The connections for the sample line are
damaged.
A B
A failure of the central gas supply can result in Changing an empty gas cylinder
simultaneous device malfunctions on all systems
connected to it. 1 Close the valve of the empty gas cylinder.
Perseus signals an alarm if the gas supply for the 2 Completely use up or completely vent any gas
gases O2, Air, or N2O (optional) fails. remaining in the pressure reducer and in the
hose between Perseus and the gas cylinder.
WARNING If there is no patient connected, venting can be
performed as follows:
Risk of contaminating the gas supply
– Disconnect the central O2 supply.
When the central gas supply is connected, – Open the flow control valve of the O2
the smallest internal leakage can cause flowmeter. Wait until gas is no longer
contamination of the supply gases. flowing.
If the central gas supply fails during – Close the flow control valve of the O2
operation, disconnect the hoses for the failed flowmeter again.
gas from the central supply.
3 Unscrew the pressure reducer from the gas
cylinder valve.
Open the corresponding gas cylinder.
4 Replace the gas cylinder with a full gas
Restore central gas supply. cylinder.
Electronically controlled gas mixer only: If the 5 Connect the pressure reducer to the new filled
central gas supply for a gas fails and there is no gas cylinder, see page 58.
sufficiently filled gas cylinder connected (see
page 31 "Status display"), a substitute gas is used: 6 Open the valve of the filled gas cylinder.
WARNING
Risk of patient recovering consciousness
If the gas supply fails completely, further
operation takes place through gas supply of
the anesthesia machine with ambient air.
Anesthetic agents are no longer delivered
and the inspiratory anesthetic gas
concentration in the breathing gas decreases.
Monitor the gas mixture carefully and use
intravenous anesthetic agents if need be.
CAUTION
Risk of increased anesthetic gas concentration in
the ambient air
If the breathing bag is not connected, expiratory
anesthetic gases may escape from the breathing
system.
Ensure adequate circulation of the ambient air.
If the fresh-gas delivery has failed, the emergency In the event of a fault, figures and instructions
O2 delivery can be used to deliver oxygen and showing how to start the emergency O2 delivery
anesthetic agent. The current ventilation mode and are displayed in areas (A) and (B).
the fresh-gas deficiency detection remain active.
The emergency O2 delivery is started as folows:
WARNING 1 Set the O2 switch (A) upwards to the Add. O2
position. (Pay attention to the figure on the
Risk of patient injury screen.)
If the gas mixer fails, no fresh gas is The Internal FG flow failure alarm will then be
delivered. automatically downgraded.
2 Open the flow control valve on the O2
Check vaporizer setting. Supply the patient
flowmeter and set the desired flow. This O2
with O2. Use emergency O2 delivery.
flow flows through the vaporizer.
Internal FG flow failure 3 Check vaporizer setting.
32032
Ventilator failure
WARNING
Risk due to malfunction of the inspiratory
flow measurement
If the inspiratory flow sensor malfunctions,
the device automatically switches to
pressure-controlled ventilation. If there is an
additional malfunction of the inspiratory
pressure sensor, the system automatically
switches to the non-synchronized PC - CMV
mode.
In either case, check the ventilation settings
and adjust as needed.
The screen does not respond to operation. It has 2 Use emergency O2 delivery, see page 183.
failed or the screen display is faulty. 3 Check vaporizer setting.
1 Switch Perseus to Standby mode: 4 Manually ventilate the patient.
Press the button and confirm with the rotary
knob. 5 Ensure corresponding substitute monitoring.
Complete failure
WARNING
Risk of device malfunction
If the breathing bag does not fill with fresh
gas, the patient cannot be sufficiently
ventilated.
– Check the oxygen supply and, if
necessary, open the gas cylinder valves.
– If fresh gas still is not delivered or manual
ventilation is not possible, close the flow
control valve of the O2 flowmeter.
– Disconnect the patient from the device
and use a replacement device!
A
A
Support request
If the device is configured for remote maintenance, 1 In the Standby mode, touch the Tests...
device information can be sent to Dräger in the button.
event of a problem. Proceed as follows to send a 2 Touch the Request support button.
support request to Dräger:
Alarm messages are displayed in hierarchal form In order to classify the alarms within an alarm
in the alarm message field of the header bar, see priority, internal priority numbers are given in the
page 140. table below. The most critical alarm is given the
number 255. Lower numbers indicate a lower
Different background colors indicate the priority
alarm priority.
levels of the alarms.
The following table lists the alarm messages in
In the Current alarms table and Alarm history
alphabetical order. If an alarm occurs, the table
table, the priority of the alarm messages is also
helps to quickly identify causes and remedies. The
indicated by exclamation marks.
possible causes and remedial measures should be
High !!! Red looked through in the order they are listed until the
alarm is resolved.
Medium !! Yellow
Some alarms appear in this table several times
Low ! Cyan with different priorities because their priority may
change under certain conditions, see page 144.
Service
Inspection....................................................... 217
Remote service ............................................... 217
Safety checks .................................................. 217
Maintenance................................................... 219
Safety information
WARNING WARNING
Risk due to inappropriately reprocessed Risk if maintenance is not performed properly
products
If the device is connected to the power supply
The product may be contaminated with or the gas supply during maintenance, there
infectious agents. is a risk of personal injury and property
damage.
Before maintenance and before returning the
product for repair, reprocess the product. Before performing maintenance, disconnect
Carry out the reprocessing as described in all electrical connections from the power
the reprocessing instructions supplied with supply and all gas connections from the gas
the product. supply.
WARNING WARNING
Risk if service is not performed regularly Risk when the housing is being opened
Wear and material fatigue of the components Under the housing, there are live electrical
may lead to device failure and malfunctions. components, which may cause an electric
shock.
Perform service at the specified intervals.
The housing may only be opened by those
WARNING user groups that are assigned to that
particular measure.
Risk if service is not performed properly
Personal injury and property damage may WARNING
occur if service is not performed properly.
Risk of patient injury
Service must be performed by those user
If maintenance activities are carried out
groups that are assigned to the particular
during ventilation, the patient will be put at
measure.
risk.
Only carry out maintenance activities when
there is no patient connected to the device.
Concept Definition
Service All measures (inspection, mainte-
nance, repair) intended to maintain or
restore the functional integrity of a
product
Inspection Measures intended to determine and
assess the current state of a product
Mainte- Regular specified measures intended
nance to maintain the functional integrity of a
product
Repair Measures intended to restore the
functional integrity of a product after a
failure
Inspection
Maintenance
Cylinder pressure
reducer, type MK:
– Sealing Every 3 years
– Sieve insert Every 3 years
– Sintered metal filter Every 3 years
on the cylinder con-
nection
– Diaphragm unit Every 6 years Replace Specialized service
personnel
– Valve cone Every 6 years
– Blow-off valve Every 6 years
– Safety valve spring Every 6 years
Repair
Disposal
Safety information
WARNING
Risk due to inappropriately reprocessed
products
The product may be contaminated with
infectious agents.
Before disposal, reprocess the product. Carry
out the reprocessing as described in the
reprocessing instructions supplied with the
product.
The disposal of electrical and electronic devices is Dräger will organize the return of the device.
subject to special guidelines. This device must be Additional information is available at
disposed of in accordance with national www.draeger.com (search term: WEEE).
regulations. In countries of the European Union,
Disposing of accessories
When disposing of the following accessory parts, Dispose on the following articles according to
observe the hospital hygiene regulations and the hospital hygiene regulations:
respective instructions for use: – Sample line
– Flow sensors – Disposable dust filter
– Breathing hoses – AGS
– Filter, HME, HMEF
– Breathing bag
– Masks
– Water trap
– CLIC absorber, Infinity ID CLIC absorber
– Soda lime
Technical data
General information
Ambient conditions
During operation
Temperature 10 to 40 °C
(50 to 104 °F)
Ambient pressure 620 to 1060 hPa
(9.0 to 15.3 psi)
Relative humidity 20 to 95 %, non-condensing
CO2 concentration 300 to 1000 ppm
Height Up to 4000 m (13123 ft)
During storage and transport
Temperature
Device without battery –20 °C to 60 °C
(–4 °F to 140 °F)
Battery –15 °C to 40 °C
(5 °F to 104 °F)
Fresh-gas delivery
Ventilator
Ventilator (continued)
Ventilator (continued)
Inspiratory flow Flow for adults Minimum 1 L/min, maximum >180 L/min
Results from the VT / Pinsp and Ti settings
Inspiratory flow Flow for children and neo- Minimum 1 L/min, maximum 60 L/min
nates Results from the VT / Pinsp and Ti settings
Pressure-related settings:
Inspiratory pressure Pinsp PEEP +1 to 80 hPa (cmH2O);
±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Pressure limitation Pmax PEEP +5 to 80 hPa (cmH2O);
±10 % of the set value or ±3 hPa (cmH2O) (the
larger value applies)
Lower pressure level Plow in APRV mode Off, 2 to 35 hPa (cmH2O);
±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Upper pressure level Phigh in APRV mode Plow +1 to 80 hPa (cmH2O);
±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Pressure amplitude above PEEP: ∆Psupp
In the CPAP/PSV and PC - BIPAP / PS Off, 1 to (80 - PEEP) hPa (cmH2O),
modes ±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
In the VC - SIMV / PS and VC - SIMV / Off, 1 to (Pmax - PEEP) hPa (cmH2O),
PS / AutoFlow modes ±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Positive end-expiratory pressure PEEP Off, 2 to 35 hPa (cmH2O);
±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Continuous positive pressure CPAP Off, 2 to 35 hPa (cmH2O);
±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Inspiratory pressure Pressure in the One- 3 to 80 hPa (cmH2O) and greater than PEEP;;
step recruitment maneuver ±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Highest inspiratory pressure Pinsp max in 15 to 80 hPa (cmH2O);
the Multi-step recruitment maneuver ±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Highest PEEP pressure PEEP max in the 2 to 35 hPa (cmH2O);
Multi-step recruitment maneuver ±10 % of the set value or ±2 hPa (cmH2O) (the
larger value applies)
Pressure amplitude above PEEP: ∆Pres- 5 to 30 hPa (cmH2O);
sure in the Multi-step recruitment maneuver ±10 % of the set value or ±2 hPa (cmH2O)
(the larger value applies)
Ventilator (continued)
Breathing system
Suction flow
Normal range 32 to 50 L/min
At lower end of restricted range 14 L/min
Maximum fresh-gas flow to prevent contaminat-
ing ambient air
For external breathing systems (normal 16 L/min
range)
For external breathing systems (restricted 5 L/min
range)
For internal breathing systems (restricted 7 L/min
range)
Low-flow wizard
Range 0 to 8 L/min
Accuracy ±25 % of measured value or ±100 mL/min (the
larger value applies)
Respiratory rate
Total RR
Spontaneous RRspon
Mandatory RRmand
Range 0 to 150 /min
Accuracy ±1 /min or 10 % (the larger value applies)
Resolution of displayed value 1 /min
T0...90 <45 s (RR ≥6 /min)
<105 s (RR <6 /min)
Dynamic compliance
Compliance Cdyn
Mean compliance Cdyn mean
Range 0 to 200 mL/hPa (mL/cmH2O)
Accuracy ±30 % or ±3 mL/hPa (mL/cmH2O) (the larger
value applies)
If during automatic ventilation there is still an
inspiratory flow at the end of inspiration, the
device will determine how much volume could
not be applied. This volume will then be
included with the normal volume in the compli-
ance calculation.
With increasing spontaneous breathing activity,
the compliance values can be distorted. In this
case the measurement accuracy may be
reduced.
Resolution of displayed value 0.1 mL/hPa (mL/cmH2O)
Resistance R
Range 0 to 100 hPa / (L/s) (cmH2O / (L/s))
Accuracy ±3 hPa / (L/s) (cmH2O / (L/s)) or ±30 % of the
measured value (the larger value applies)
(With increasing spontaneous breathing activity,
the R values can be severely distorted. Hence
maintaining measurement accuracy cannot be
guaranteed with spontaneous breathing).
Resolution of displayed value 1 hPa / (L/s) (cmH2O / (L/s))
Elastance E
Range 0.005 to 10 hPa/mL (cmH2O/mL)
Resolution of displayed value 0.001 hPa/mL (cmH2O/mL)
Gas measurement Sidestream gas measurement (The sample gas
is fed back into the breathing system and
included in calculations for measurements and
delivery);
All values are measured under ATPS condi-
tions;
Sample gas flow standardized to STPD condi-
tions.
The measurement is corrected for ambient pres-
sure.
Due to the T10...90 time and the sampling rate,
the accuracies of the measured values for O2,
CO2, N2O, and anesthetic agent may deviate at
respiratory rates of 75 /min or higher and an I:E
ratio of 1:2. The influence of respiratory rate and
the I:E ratio on the accuracy has been verified in
a simulated breathing system using a rectangu-
lar waveform for the gas concentration.
End-tidal measured values are calculated for
each breath from the local maxima and minima
of the real-time measurements during expira-
tion.
Sample gas flow 200 mL/min ±10 %
Maximum time until emptying of the water 41 h (sample gas under BTPS conditions at
trap is necessary 23 °C ambient temperature)
System response time The system response time results from the typi-
cal delay and the gas type specific T10...90 time.
Sensor sampling rate <50 ms
Time after switch-on until the specified accu- <500 s
racy is attained
Time until CO2 measured values are dis- <90 s
played with reduced accuracy
Typical delays <5 s
Cross sensitivity None with respect to alcohol (<3000 ppm in
blood),
Acetone (<1000 ppm), methane, water vapor,
NO and CO
Detection Automatic
Primary gas At the latest at 0.3 Vol%
Secondary gas At the latest at 0.4 Vol% or 0.1 xMAC (the larger
value applies)
With a desflurane concentration greater than
4 Vol%, mixture detection occurs at the latest
when the concentration of the second anes-
thetic gas rises above 10 % of the desflurane
concentration.
The secondary gas becomes the primary gas
when the expiratory xMAC value is more than
0.2 xMAC above that of the primary gas.
Minimum displayed concentration The specified detection thresholds refer to rising
anesthetic gas concentrations (e.g., at the start
of an operation). If the anesthetic gas concen-
tration falls, a concentration of down to
0.05 Vol% will be measured, based on the last
anesthetic agent detected. Below this concen-
tration, a value of 0 Vol% will be displayed.
xMAC Based on patient age, anesthetic agent, and
nitrous oxide concentration
(the xMAC is corrected for ambient pressure)
Range 0 to 9.9
Accuracy Corresponds to that of the gas measurement
Resolution of displayed value 0.1
Measurement of supply pressures
Central supply
Range 0 to 9.8 kPa x 100
0 to 140 psi
0 to 0.98 MPa
Accuracy ±4 % or ±0.2 kPa x 100 (the larger value
applies)
±4 % or ±3 psi (the larger value applies)
±4 % or ±0.02 MPa (the larger value applies)
Resolution of displayed value 0.1 kPa x 100
1 psi
0.01 MPa
Gas cylinders
Range 0 to 250 kPa x 100
0 to 3600 psi
0 to 25 MPa
Anesthetic agent
Range 0 to 999.9 mL liquid
Accuracy ±25 %
Resolution of displayed value 0.1 mL
Waveforms O2 concentration
Primary anesthetic agent concentration
CO2 concentration
Airway pressure
Volume (only for loops)
Flow
Sweep speed 6.25; 12.5; 25 mm/s
Scale
Airway pressure Paw (t) –20 to 80 hPa (cmH2O)
Flow (t) –120 to 120 L/min
Volume (t) 0 to 2000 mL
O2 (t) 0 to 100 Vol%
CO2 (t) 0 to 100 mmHg (0 to 12 Vol%, 0 to 12 kPa)
Anesthetic agent (t)
Halothane 0 to 5 Vol% (kPa)
Enflurane 0 to 6 Vol% (kPa)
Isoflurane 0 to 5 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
Loops Pressure-Volume
Flow-Volume
Operating characteristics
RFID system
Operating frequency 13.56 MHz ± 50 ppm (broadband)
Transmitter power ≤42 dBμA/m (200 mW ± 1 dB)
Modulation ASK (Amplitude-Shift Keying)
Electromagnetic compatibility Tested in compliance with IEC 60601-1-2
Protection classes
Device I, in compliance with IEC 60601-1
Applied parts (connections for breathing TYPE BF
hoses)
Penetration of liquids IP20 according to IEC 60529,
meets IEC 60601-2-13
Classification in compliance with Directive II b
93/42/EEC, Annex IX
UMDNS Code Universal Medical Device 10-134
Nomenclature System – nomenclature system
for medical devices
Use of latex Perseus is not made with natural rubber latex.
Device outputs
Pin assignment
Pin 1 n/c
Pin 2 RXD
Pin 3 TXD
Pins 4, 6 Pins 4 and 6 are connected internally
Pin 5 SHLD-GND
Pins 7, 8 Pin 7 and Pin 8 are connected internally
Pin 9 n/c
Housing SHLD-GND
USB interface Only connect USB storage media that do not
have their own power supply. Do not connect
any charging cables.
Mains power sockets Observe the maximum current for each power
socket, the total current for all power sockets,
and the total permissible leakage current.
Mains power sockets have no battery back-up
and are independent of the switch-on state of
the device.
Power sockets (version without isolation
transformer)
Power socket type IEC
Power sockets with protective ground 2 (maximum 3.15 A per power socket)
Power sockets without protective ground 2 (maximum 1 A per power socket)
Fuse type Fuses, T3,15AH250V IEC 60127-2/V and
T1AH250V IEC 60127-2/V, size 5 mm * 20 mm
Power sockets
(version with isolation transformer)
Power socket type Country-specific
Fuse type Electronic automatic circuit breakers: 4 x 3 A,
slow-blow, for voltages up to 250 V and break-
ing capacity 2000 A.
And 1 x 8 A, slow-blow, for voltages up to 250 V
and breaking capacity 2000 A, in each case
conforming to IEC 60934
Power sockets with protective ground 4 (maximum 3 A per power socket)
Total current Max. 8 A
Main switch for isolation transformer and all
power sockets
Relevant standards
Diagrams
10574
3UHVVXUH>K3DFP+2@
ದ
ದ
ದ
ದ
ದ
ದ
ದ
ದ
ದ
ದ
)ORZ>/PLQ@
Device combinations
EMC declaration
This device was tested for electromagnetic This device may be used only in environments
compatibility using accessories from the list of specified in section "Environments of use" on
accessories. Other accessories may only be used page 16.
if they do not compromise the electromagnetic
compatibility. The use of non-compliant Emissions Compliance
accessories may result in increased Radiated emis- Class A, group 1 (30 MHz to
electromagnetic emissions or decreased sions 1 GHz)
electromagnetic immunity of the device.
Conducted emis- Class A, Group 1 (150 kHz to
This device may be used in the direct vicinity of sions 30 MHz)
other devices only if Dräger has approved this
device arrangement. If no approval has been given
by Dräger, it must be ensured that this device NOTE
functions correctly in the desired arrangement
The emissions characteristics of this equipment
before use. The instructions for use for the other
make it suitable for use in industrial areas and
devices must be followed.
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required), this equipment
might not offer adequate protection to radio-
frequency communication services.
The user might need to take mitigation
measures, such as relocating or re-orienting the
equipment.
This medical device is equipped with an RFID Changes and modifications that have not been
module (radio frequency identification) to enable expressly approved by Dräger may result in the
wireless communication with Infinity ID user no longer being permitted to operate the
accessories. device.
This medical device has been designed and Dräger hereby declares that this medical device,
manufactured to comply with emission limit values including its radio equipment, is in compliance with
for high-frequency energy. These limit values are Directive 2014/53/EU.
incorporated in international safety standards such
The complete EU Declaration of Conformity can be
as IEC 60601-1-2 and standards for radio
viewed at the following internet address:
equipment such as EN 300330 and have been
http://www.draeger.com/doc-radio
defined by regulatory authorities.
The RFID system of this medical device complies
with Part 15 of the FCC regulations and the
license-free RSS regulations of Industry Canada.
Operation is subject to the following 2 conditions:
1 This medical device does not cause any
harmful interference.
2 The medical device is not liable to damage
caused by the reception of interference,
including interference causing undesired
operating conditions.
Connections to IT networks
Examples of subsequent changes to the network: The LAN interface allows synchronization with an
NTP (Network Time Protocol) server using the
– Changing the network configuration NTP protocol.
– Removing devices from the network
– Adding new devices to the network Required characteristics
– Performing upgrades or updates on devices
The LAN must ensure the connection between the
that are connected to the network
device and the following destinations:
– Service Connect Gateway or DrägerService
Computer
– NTP server
Connections between Perseus and the DrägerService computer or the NTP server
Function Protocol Perseus Direction Remote port Remote Connection
port partner
SNMP V3 UDP 161 >1023 SCG
TCP SNMP V3 UDP >1023 162 SCG
(trap)
FTP TCP >1023 21 SCG New, estab-
(command) lished
FTP TCP >1023 21 SCG Established
(command)
FTP TCP >1023 >1023 SCG New, estab-
(data) lished
FTP TCP >1023 >1023 SCG Established
(data)
SNTP UDP >1023 123 NTP server
DHCP UDP 67 67 DHCP server
Required characteristics
The RS232 interface is a point-to-point connection.
A connected device must prevent access by
unauthorized users to the data that are sent over
the RS232 interface and must itself be protected
from infections by malware and computer viruses.
Open-source software
Principles of operation
Additional parame-
Base parameters ters
Group Tab Ventilation mode
(normal therapy bar) (expanded therapy
bar)
Manual ventila- Man/Spon Manual / Spontaneous CPAP1)
tion / Sponta-
neous
breathing
Pressure-sup- PSV1) CPAP / PSV Trigger Insp. term.
ported ventila- ∆Psupp
tion RRmin
PEEP
Slope
Pressure-con- PC PC - CMV Pinsp Slope
trolled ventila- RR
tion PEEP
Ti
Sync. off
PC - BIPAP Pinsp Trigger
∆Psupp1) = Off Slope
RR
PEEP
Ti
Sync. on
PC - BIPAP / PS1) Pinsp Trigger
∆Psupp1) >0 Insp. term.1)
RR Slope
PEEP
Ti
Sync. on
APRV2) PC - APRV Phigh
Thigh
Slope
Plow
Tlow
Additional parame-
Base parameters ters
Group Tab Ventilation mode
(normal therapy bar) (expanded therapy
bar)
Volume-con- VC - AF VC - CMV / AutoFlow Pmax Slope
trolled ventila- VT
tion RR
PEEP
Ti
Sync. off
VC - SIMV / AutoFlow Pmax Trigger
VT ∆Psupp1) = Off
RR Slope
PEEP
Ti
Sync. on
VC - SIMV / PS / Auto- Pmax Trigger
Flow1) VT ∆Psupp1) >0
RR
Insp. term.1)
PEEP
Slope
Ti
Sync. on
VC VC - CMV Pmax %Tplat
VT
RR
PEEP
Ti
Sync. off
VC - SIMV Pmax Trigger
VT ∆Psupp1) = Off
RR %Tplat
PEEP
Ti
Sync. on
VC - SIMV / PS1) Pmax Trigger
VT ∆Psupp1) >0
RR
Insp. term.1)
PEEP
Slope
Ti
%Tplat
Sync. on
1) Requires software option PSV
2) Requires software option APRV
Pressure-supported ventilation
PSV (optional)
10563
Paw ∆Psupp
PEEP
Slope 7LPH
1/RRmin
Flow
Trigger
Insp. term.
7LPH
SDWLHQWWULJJHUHGEUHDWK PDFKLQHWULJJHUHGEUHDWK
Pressure-controlled ventilation
PC
10562
Paw Pinsp 7ULJJHUZLQGRZ
LQVS
PEEP
∆Psupp
Ti
1/RR
Flow
Trigger Trigger
Insp. term.
7LPH
PDFKLQHWULJJHUHGEUHDWK SDWLHQWWULJJHUHGEUHDWK
APRV (optional)
10561
Paw Phigh
Plow
Slope 7LPH
Thigh Tlow
Flow
7LPH
PC - APRV
– Pressure-controlled
– Time-controlled
– Machine-triggered
– Spontaneous breathing at continuous positive
airway pressure with short pressure releases
In PC - APRV, the patient's spontaneous breathing
occurs at the upper pressure level Phigh. This
pressure level is maintained for the duration of
Thigh.
The number of pressure releases is determined by
the Thigh and Tlow settings. The releases are
time-controlled and are not triggered by the
patient. The duration is determined by Tlow.
Volume-controlled ventilation
VC
10558
7ULJJHUZLQGRZ
LQVS
Paw
Pmax
∆Psupp
PEEP
%Tplat 7LPH
Slope
Ti
1/RR
Flow
Trigger Trigger
Insp. term.
7LPH
PDFKLQHWULJJHUHGEUHDWK SDWLHQWWULJJHUHGEUHDWK
VC - CMV – Machine-triggered
– Volume-controlled – Constant inspiratory flow
– Pressure-limited
– Time-controlled
VC - SIMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered or patient-triggered
– Constant inspiratory flow
– Synchronized inspiration
In VC - SIMV, the patient can breathe
spontaneously at any time, while the number of
mandatory breaths is predefined. When
synchronization is switched on, the breaths are
adapted to the spontaneous breathing of the
patient. If inspiratory effort by the patient is
detected during the inspiratory trigger window, a
patient-triggered breath will be initiated.
VC - SIMV / PS
This mode is similar to VC - SIMV, except that the
patient's spontaneous breathing at the PEEP level
during the expiratory phase is pressure-supported
with ∆Psupp when outside the trigger window.
VC - AF
10560
Paw 7ULJJHUZLQGRZ
LQVS
PEEP
Pmax
∆Psupp
7LPH
Slope Slope
Ti
1/RR
Flow
Trigger Trigger
Insp. term.
7LPH
PDFKLQHWULJJHUHGEUHDWK SDWLHQWWULJJHUHGEUHDWK
With AutoFlow, the set tidal volume VT is applied If no automatic ventilation has previously taken
for all mandatory volume-controlled breaths with place, a volume-controlled test breath with
the lowest required pressure. The patient can constant inspiratory flow is performed when
breathe spontaneously throughout the entire starting a ventilation mode with AutoFlow in order
respiratory cycle, during both inspiration and to estimate the lung parameters. The inspiratory
expiration. The pressure patterns and flow patterns pressure required at the start is determined from
of the mechanical inspiratory breaths correspond this test breath. Each additional breath-related
to those of pressure-controlled ventilation. readjustment of the inspiratory pressure is limited
to ±3 hPa (cmH2O). The pressure difference
Due to the patient's spontaneous breathing efforts
(inspiratory pressure - PEEP) is at least 5 hPa
or compliance changes in the lungs, the tidal
(cmH2O) and the upper inspiratory pressure limit is
volume in an individual breath may deviate from
set by Pmax. If the set value for VT is reduced, the
the set tidal volume VT. However, averaged over
inspiratory pressure will be reduced by a greater
time a tidal volume corresponding to the set
amount if necessary.
volume VT is applied.
Minimum O2 delivery
Influence of patient category In the Neo patient category, the ideal patient
weight is equal to the entered patient weight.
– Alarm limits and start settings for therapy The calculated ideal patient weight affects:
– Volumeter scale – Start settings for tidal volume VT
– Flow measurement and software algorithms to – Start settings for respiratory rate RR
suppress artifacts
– Start settings for VT and MV alarm limits
– Maximum duration of a pressure-supported
breath – Flow trigger
VT and RR are only dependent on the ideal patient
weight when the Ideal body weight function is
Influence of the ideal patient weight and selected in System setup > Therapy, see
patient height page 157.
Changing the patient weight during an automatic
Ideal patient weight describes the portion of the ventilation has no effect on the current ventilation
body that is relevant to setting the ventilation settings.
parameters (patient body weight minus assumed
excess fat).
In the Adult and Ped patient categories, the ideal Influence of the patient age
patient weight is calculated from the entered
patient height. During operation, the set age influences:
– Calculation of MAC value
CAUTION WARNING
Risk of incorrect values for resistance and Risk of inappropriate operating life
compliance
Exchange monitoring only considers the
If an unused Infinity ID accessory is in the absolute period of use and not the current
immediate vicinity of the device, values may be condition of the Infinity ID accessory and
transferred inadvertently from this accessory. therefore does not excuse the user from
periodic checks of the accessory.
Do not keep unused Infinity ID accessories in the
vicinity of the device. The exchange interval which can be set for
the exchange monitoring does not represent
CAUTION a guarantee for the maximum period of use of
the accessory.
Risk of inappropriate compliance values
When accessories are added to an Infinity ID Anti-interchange security
breathing circuit, the values for compliance and
leakage may deviate from those saved on the
WARNING
breathing circuit.
Always perform the leakage test before starting Risk of interchanged or incorrect breathing
therapy in order to determine the actual values hoses
for compliance and resistance. If the test cannot The Infinity ID function to prevent
be performed because the patient is already interchanging the breathing hoses does not
connected, particular attention is required during excuse the user from checking the
ventilation. accessories.
The Infinity ID function to prevent
CAUTION interchanging does not represent a guarantee
Malfunctions of the Infinity ID functionality that the hoses are correctly connected.
Particular EMC situations or defects in Infinity ID When Infinity ID breathing hoses and Infinity ID
components can cause permanent alarms. breathing bags are used, the incorrect connection
To prevent distracting the user under these of breathing hoses and the breathing bag is
circumstances, contact service personnel to detected and reported. Hoses that are incorrectly
deactivate the Infinity ID alarms. connected with the breathing system trigger an
automatic alarm.
High Depending on the overall alarm situation, this tone sequence Yes
may be played as a 5-tone sequence due to the timing of the
individual alarms.
Medium Yes
Low No
When Signal
Therapy start or change of venti-
lation mode
Timeout
20427
provided with a combination lock. When delivered,
the combination on the lock is set to 3333.
20428
20429
20432
2 Turn the knob to position C. 2 Turn the key counterclockwise to the final
position stop.
3 Set the new combination.
20433
20430
4 Turn the knob to CLOSE. 3 Turn the knob to OPEN. The digit wheels show
the current combination.
The new combination is stored.
Annex
Abbreviations................................................. 275
Symbols.......................................................... 277
Abbreviations
Symbols
Gas inlet
Product labels
Group
Group
View1)
1 2 3
Group
Group
Group
Group
Password
Cut-out from the instructions for use Perseus To prevent unauthorized alteration, the start
A500 Software 2.0n settings for Perseus A500 are protected by the
following configuration password:
0000
Index
Gas scavenging M
Ensuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Gas supply MAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Checking. . . . . . . . . . . . . . . . . . . . . . . . . . . 128 Main screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Establishing . . . . . . . . . . . . . . . . . . . . . . . . . 57 Main switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Mains power supply
Establishing . . . . . . . . . . . . . . . . . . . . . . . . . 52
H Mains power supply failure . . . . . . . . . . . . . . . 182
Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 219
Hose configurations . . . . . . . . . . . . . . . . . . . . . . 70 Maneuvers (optional). . . . . . . . . . . . . . . . . . . . . 95
Hose set and filters Manual ventilation . . . . . . . . . . . . . . . . . . . 90, 260
Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Measured values
Viewing . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . 282
I Minimum O2 delivery . . . . . . . . . . . . . . . . . . . . 268
Illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Minimum O2 flow . . . . . . . . . . . . . . . . . . . . . . . 161
Importing configurations . . . . . . . . . . . . . . . . . . 171 Mini-trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Minute volume measurement . . . . . . . . . . . . . 104
Infinity ID. . . . . . . . . . . . . . . . . . . . . . . . . . . 35, 169 Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Accessory exchange interval monitoring . . 270
Accessory replacement intervals . . . . . . . . 169 N
Accessory support . . . . . . . . . . . . . . . . . . . . 35
Activating the function. . . . . . . . . . . . . . . . . 169 N2O
InfinityID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269 Operation without N2O. . . . . . . . . . . . . . . . 163
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217 Name
Inspiratory port . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Names for screen layouts . . . . . . . . . . . . . . . . 150
IT networks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253 Network
Configuring . . . . . . . . . . . . . . . . . . . . . . . . 168
Network interface . . . . . . . . . . . . . . . . . . . . . . . 27
L New case
Language Defining . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164 Non-rebreathing system
Latex. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Connecting. . . . . . . . . . . . . . . . . . . . . . . . . . 74
Leakage Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Causes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Current values . . . . . . . . . . . . . . . . . . . . . . 121 O
Locating . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Leakage assistant. . . . . . . . . . . . . . . . . . . . . . . 122 O2 flowmeter . . . . . . . . . . . . . . . . . . . . . . . . 31, 32
Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 O2 flush
Linked therapy controls . . . . . . . . . . . . . . . . . . . 41 Checking . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Logbook O2 insufflation
Creating entries . . . . . . . . . . . . . . . . . 115, 167 Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Viewing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 O2 switch . . . . . . . . . . . . . . . . . . . . . . . . . . . 31, 37
Loops O2 waveform scale
Configuring . . . . . . . . . . . . . . . . . . . . . . . . . 151 Specifying . . . . . . . . . . . . . . . . . . . . . . . . . 151
Using. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 O2+ . . . . . . . . . . . . . . . . . . . . . . . . . 31, 32, 37, 38
Low-flow wizard . . . . . . . . . . . . . . . . . . . . . . . . 104 Checking . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Lung recruitment. . . . . . . . . . . . . . . . . . . . . . . . . 95 Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Open-source software . . . . . . . . . . . . . . . . . . . 256
Operation
Safety information . . . . . . . . . . . . . . . . . . . . 85
P Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Parameter fields . . . . . . . . . . . . . . . . . . . . . . . . 173 Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Parameters Saving the layout . . . . . . . . . . . . . . . . . . . . 150
Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Setting brightness . . . . . . . . . . . . . . . 108, 150
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285 Screen brightness
Patient category . . . . . . . . . . . . . . . . . . . . . . . . 269 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Patient data Screenshot
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Exporting . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Checking. . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9, 169
Influence on device behavior . . . . . . . . . . . 269 Service personnel . . . . . . . . . . . . . . . . . . . . . . . . 7
Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Setting procedure
Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Canceling. . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Potential equalization Settings
Establishing . . . . . . . . . . . . . . . . . . . . . . . . . 54 Resetting to starting settings . . . . . . . . . . . 115
Potential equalization pin . . . . . . . . . . . . . . . . . . 27 Soda lime
Power sockets . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Checking . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Power supply failure . . . . . . . . . . . . . . . . . . . . . 182 Software options
Prediction Activating . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Anesthetic gas concentration . . . . . . . . . . . 101 Displaying . . . . . . . . . . . . . . . . . . . . . . . . . 109
FiO2 concentration . . . . . . . . . . . . . . . . . . . 102 Sound volume
Pressure-controlled ventilation . . . . . . . . . 260, 263 Setting alarm tones . . . . . . . . . . . . . . . . . . 110
Pressure-supported ventilation. . . . . . . . . 260, 262 Setting breathing sound. . . . . . . . . . . . . . . 110
Spontaneous breathing . . . . . . . . . . . . . . . 90, 260
Q Start dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Opening . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Quick setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Start settings . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Gas delivery. . . . . . . . . . . . . . . . . . . . . . . . 161
R Patient category. . . . . . . . . . . . . . . . . . . . . 162
Resetting . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Rating plate . . . . . . . . . . . . . . . . . . . . . . . . . 24, 25 Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . 157
Rear Start values . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Ceiling-mounted version . . . . . . . . . . . . . . . . 26 Stopwatch
Trolley version . . . . . . . . . . . . . . . . . . . . . . . 24 Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Recruitment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Remote control . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277
Remote service . . . . . . . . . . . . . . . . . . . . . . . . 217 Synchronized ventilation . . . . . . . . . . . . . . . . . . 92
Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220 System test
Reprocessing personnel. . . . . . . . . . . . . . . . . . . . 7 Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Configuring . . . . . . . . . . . . . . . . . . . . . . . . 166
Respiratory support . . . . . . . . . . . . . . . . . . . . . 259 Test types . . . . . . . . . . . . . . . . . . . . . . . . . 122
Rotary knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
T
S
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . 223
Safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . 217 Test results . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Safety sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Testing the system. . . . . . . . . . . . . . . . . . . . . . 121
Sample line Therapy
Connecting . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Ending . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Scale Setting and starting . . . . . . . . . . . . . . . . . . . 88
Adjusting . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Therapy controls . . . . . . . . . . . . . . . . . . . . . 40, 42
Time
Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
V
Valves
Inserting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Vapor View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Vaporizer
Checking. . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Illumination . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Using. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Ventilation mode
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Ventilation modes . . . . . . . . . . . . . . . . . . . . 33, 258
Ventilation parameters
Linked setting . . . . . . . . . . . . . . . . . . . . 41, 162
Ventilator failure . . . . . . . . . . . . . . . . . . . . . . . . 186
View
Available views . . . . . . . . . . . . . . . . . . . . . . 105
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Customizing . . . . . . . . . . . . . . . . . . . . . . . . 106
Volume
Setting the breathing sound volume . . 153, 167
Manufacturer
9510595 – 6500.510 en
Á9510595fÈ
© Drägerwerk AG & Co. KGaA
Edition: 3 – 2020-01
(Edition: 1 – 2018-10)
Dräger reserves the right to make modifications
to the medical device without prior notice.
ision]_0000048349_No.2402