Buck 210 211 Users Manual
Buck 210 211 Users Manual
Buck 210 211 Users Manual
Table of contents
Introduction 3 SECTION 1: INSTALLATION General Specifications 4 Gas Supplies 7 Preparing the Lab 9 Installation - Basic instrument & flame 10 Installation - Graphite furnace 14 Installation - Graphite furnace autosampler 15 SECTION 2: FLAME SETUP & Model 211 Gas Box Operation 2.0: Configuring the Hardware Wavelength, lamp & burner alignment 16 2.1: Running nitrous oxide / Buffers for nitrous oxide 20 2.2: Operation of the 211 automated gas box 21 SECTION 3: FLAME ANALYSIS 3.0: Setting flame analysis parameters [Cntls] key 23 3.1: Setting flame analysis parameters [Smpls] key 24 3.2: Calibrating Standards and Analyzing Samples [Cal] key 25 3.3: Emission mode 28 SECTION 4: FURNACE ANALYSIS 29 4.0: Installing the graphite tube 30 4.1: Aligning the furnace 30 4.2: Aligning the autosampler 30 4.3: Setting furnace analysis parameters [Cntls] key 32 4.4: Setting furnace analysis parameters [Smpls] key 34 4.5: Calibrating Standards and Analyzing Samples [Cal] key 34 SECTION 5: RUNNING COLD VAPOR AND HYDRIDE 36 SECTION 6: EDITING THE LIBRARY/CHANGING LAMP CURRENTS 37 SECTION 7: METHOD OF STANDARD ADDITIONS 38 SECTION 8: FLAME TECHNIQUES 41 SECTION 9: INTERFERENCES 43 SECTION 10: TROUBLESHOOTING, TIPS & TECHNIQUES 45 47 SECTION 11: MAINTENANCE SECTION 12: STANDARD CONDITIONS / ALTERNATE WLS 48 SECTION 13: MATRIX MODIFIERS FOR FURNACE 53 SECTION 14: COMPUTER DATA COLLECTION 54 14.0: Aanalyze data package 54 14.1: Data collection without the Aanalyze package 57
SECTION 1: Installation
GENERAL SPECIFICATIONS Electrical: 110V AC nominal (+10%), 50/60 Hz
220, 240V AC, 50/60 Hz Power Consumption: 50W
Optics:
Detector: model 928; wide range general purpose, 190-930nm Optional Detectors: model 955; UV enhanced, wide-range, 190-930nm model EMI9783B; narrow range furnace/hydride application, 165-600nm Lenses: Supracil - amorphous silica Monochromator: 0.25m Ebert mount Grating: 32nm x 27nm; 600 grooves/mm Wavelength adjustment: 3 digit mechanical, 0 to 1000nm +1 nm Reproducibility: +0.2 nm Resolution: variable slit - 2, 7, and 20
Operating Modes:
Absorbance/Emission: -0.0820 to 3.2000 Concentration: to 5 significant digits Integration Period: 0.5 to 10 seconds Screen Refresh : 0.5 to 1.5 seconds Recorder Output: 1V/ABS (-0.08 to 3.2V) Background Correction: In-line Deuterium Arc Giant Pulse (Self-reversal)
Burner Assembly:
Design: Polyethylene Pre-mix chamber, glass impact bead dispersion Burner Head: Titanium; air-acetylene head - 4" x 0.026" single slot (nitrous oxide head - 2" x 0.019" single slot) Adjustments: Horizontal g
Performance:
Average Noise (at 3): 0.0018 ABS (Cu at 324.7nm, 7 slit, 5 sec. int.) Sensitivity: see specific element chart (Sect. 4) Reproducibility: <+5%
Front View
Fuel Toggle Switch Gas Flow Indicator Titanium Burner Head Brightness Adjustment Knob LCD Screen Wavelength Display
Rear View
Descrete I/O Comm Port Furnace Port Printer Port Furnace & Burner Cut-outs Analog Output 0-1V
Cooling Fan Aux Cord Power Cord Heat Sink Air Input Fuel Input N20 Input Idetntification Label
Right Side
Slit Width Adjustment
Wavelength Adjustment
Power Switch
Left Side
Lamp turrett
Lamp Socket
When N O is removed from the cylinder at this rate the expanding gas cools the diaphragm of the regulator so that sometimes it freezes, causing loss of regulation. It is therefore advisable to use a Ambient Air Heated Regulator (part no. BS303-0204). All lines carrying N O should be free of grease, oil or other organic material, as it is possible for spontaneous combustion to occur. Cylinders of N O should be considered high pressure cylinders and handled with care at all times.
2 2 2
Argon:
Argon is usually obtained in size 1A cylinders containing about 7000 liters (at STP). It is used with a CGA580 two-stage regulator (part no. BS303-0264). Argon is generally used in conjunction with hydrogen as a flame diluent to provide a cool flame, as a purge gas in the analysis of hydride-forming metals, and also as a sheathing gas in graphite furnace work. Consequently the consumption rate varies widely, depending on the particular application. High purity grade (99.995%) Grade Argon is good for most analyses. Argon in the pre-purified grade (99.998%) can also be used but is not recommended for the Graphite Furnace since the grade often contains a small amount of oxygen which may shorten tube life. Argon cylinders are used at high pressure and should be handled with care at all times.
Nitrogen:
Nitrogen is usually obtained in size 1A cylinders containing about 6500 liters (at STP). It is used with a CGA580 two-stage regulator (part no. BS303-02640). Nitrogen is used similarly to argon, and therefore the consumption rates vary widely with the application. The high purity grade (99.9%) or the extra dry grade (99.7%) is suitable for atomic absorption work. Nitrogen cylinders are used at high pressure and should be handled with care at all times.
Note: All regulators & water supplies must be ready to accept a 1/4" swagelock nut for installation. Note: If performing all 3 types of analysis, you only need one tank of any one gas.
6"(12.4 cm)
220 GF
4"heat fins
The laboratory environment should be regulated to provide stable temperature and humidity. The 210VGP should be maintained at temperatures from 10 to 32 C (50 - 90 F), and relative humidity of 30% to 80%.
0 0 0 0
In many applications data handling may be accomplished using an external printer, the Aanalyze software package from an IBM style PC or laptop computer, a laboratory recorder, or all three. These accessories should be located near the instrument for easy access. It is advisable to place the computer and printer on a separate workbench or table on the right hand side of the 210VGP, since the external connections are made on that side of the instrument. This will prevent the cables from having to cross over the gas lines, and make them more accessible.
Locate the exhaust intake manifold directly above the combustion chamber of the 210 VGP as shown below.
Eighteen Inches
Electrical Connections Plug the power cord into a standard outlet and connect the other end to the power adapter on the back of the 210-VGP unit. International clients MUST check to see that the small red voltage selection switch is set appropriately to 220 Volts.
Aux Cord
Power Cord
Heat Sink
Configuring the Hardware: Before powering up for the first time perform this procedure.
Press and hold the [upper case] button and turn power on by pressing the red button on the right side of the 210. When text appears on the screen the [upper case] button can be released. Make any necessary changes to the hardware configuration for: Graphite Furnace : YES if installed Graphite Furnace Autosampler : 48 Cup for new systems Flame Autosampler : YES if installed Printer : YES if installed. Clock : 12 or 24 hour, US or European Date. . Use the [left/right] arrows to make changes in each category and always press [enter] to initiate the change. Press [esc] when finished. This screen never needs to be accessed again.
ACTIVE ANALYSIS Name: Cu-D2-324.7-Lib 3 Pulse: Wide Lamp: Cu buck sci Bknd: On LIBRARY Name: Cs-D2-852.1-Lib 3 Lamp: Cs buck sci Pulse: Wide Bkgnd: Off
Name: Cu-D2-324.7-Lib 3 Lamp: Cu buck sci Wavelength: 324.7 Set the lamp turret to position 1 Set wavelength and slit as specified Then: (1) (2) (3) Press <ALIGN> Fine adjust lamp position and Wavelength for maximum energy Press <AUTOZERO> when done Or Press any other function key
Press: <SELECT> Select next lamp <ESC> Exit <up arrow> or <down arrow> Browse library <1> Enter a library name <2> Load analysis from library <3> Enter a password Password required to modify library
S ANDBY ANAL S T Y IS
Name: Au-Furn3-242.8 Name: Cu-D2-324.7-lib 3 D2 Mode D2 Mode Lamp 3 L amp 2
P ress: < Align> to normalize bargraph < A/Z> to exit and autozero < E C> to exit - - no change S
3) The bargraph indicates the amount of energy from the lamp reaching the detector. 0 on the scale represents the amount of energy you had after pressing align. Very slowly turn the wavelength knob in any direction. If the bar graph goes off screen to the right press [align] to center it. If the bar graph goes off scale to the left turn the wavelength knob in the other direction. Repeat this procedure until you get the HIGHEST reading (bar graph farthest to the right) you can. Pay attention to the numeric value for energy. This runs on a scale from 0 to 6 and you want to get it as high as possible. The bargraph will follow the energy reading (ie: as the bargraph moves to the right the energy value will increase. Most lamps will give an energy between 2.5 and 4.5. If you have trouble aligning the lamp, press [Bkgnd] to turn the background lamp off and try again. Press [Align] when finished. The wavelength is now set.
Align the lamp 4) On the top left of the instrument are 2 knobs for lamp alignment (front knob for horizontal, rear knob for vertical). Rotate either of the two knobs in any direction watching the sample bargraph, again trying to get the most energy you can in a manner similar to the wavelength alignment. 5) When the energy has been maximized, move to the other knob and repeat the procedure. Go back and forth between the two knobs until you cant get any more energy. (NOTE: There will not be nearly the change in energy as there was with the wavelength). The lamp is set when you cant get any more energy. Press [A/Z] to zero and exit to the analysis screen. Align the burner (vertical) vertical adjustment > vertical adjustment > rotation adjustment > right knob (burner moves up & down) left knob (burner moves front to back) middle knob on older instruments or burner bar on newer
6) With the flame and gasses off place a business card or similar surface on top of the burner so that you can see the lamp image on the card. Rotate the vertical adjust knob so that the bottom edge of the light at the focal point is approximately 4mm from the top of the burner (best position for most analysis). Adjust the horizontal if necessary to get the image over the burner head slot, this is only a rough adjustment for the horizontal. (NOTE: Some elements may require different height settings especially when using nitrous oxide, consult the standard conditions section for these instances). Another way to set the vertical position is to lower the burner head until it is clearly not blocking the beam. Perform an autozero by pressing the [a/z] key. Slowly raise the burner head while watching the absorbance display. When the burner head intersects the beam the number will suddenly go positive. As soon as the reaction is noticed, stop and lower the burner about 2 turns of the dial. Align the burner (horizontal) FLASHBACK HAZARD: ALWAYS turn on the air first, and shut it off last. Make sure the drain tube has a loop approximately 8 inches in diameter for proper drainage and to prevent acetylene from flowing into the waste vessel. NEVER use glass or something that can shatter as the waste vessel. 7) With the air and acetylene tanks on and the D2 background lamp off, press [a/z] to zero the display, turn on the air knob at the instrument while aspirating water so that you have a mist coming from the burner head. Turn the horizontal adjust knob in any direction while noticing the absorbance display. The object is to locate the burner head directly under the lamp beam to get the highest absorbance reading you can. 8) Repeat the procedure with the rotation adjust (either the burner head bar or rotation adjust knob, depending on which instrument you have). If you have the burner bar you must physically rotate the head to get the highest absorbance reading. Generally, having the burner head so it looks square in the burner compartment is adequate.
Adjust the nebulizer 10) While still aspirating the water turn the nebulizer adjust counterclockwise until air bubbles out through the capillary into the beaker. Slowly turn it clockwise until the nebulizer starts aspirating the water and watch the absorbance readings. Adjust the nebulizer to get the highest and most quiet absorbance that you can (this will probably be within a turn or two from where the nebulizer started to aspirate the standard. The nebulizer is now set and should not need adjustment for many months. (NOTE: The nebulizer comes factory adjusted and should not need any initial adjustment). 11) When you are finished turn the background corrector back on (if your method requires it) by pressing [bkgnd]. The background corrector is useful below 450 nanometer to reduce or eliminate false absorbance due to dissolved solids or organic material in the sample. Remember: If you have not changed anything you only need to check these alignments if you are having problems or need to change the burner head height for special analyses (see section 3.2). When you change lamps you only need to align the wavelength and the lamp. Igniting the Flame 1) Open the acetylene tank valve and set to 12-13psi. 2) Turn the air compressor on or open air tank valve and set to 50-60psi. 3) Turn on the air valve at the instrument. 4) Turn on the fuel valve at the instrument and adjust flow for 4 on the flowmeter. 5) Light the burner with a long-handled igniter or striker. 6) You are now ready to set up for running your samples.
Shutdown While the flame is lit, turn off the fuel flow making sure the flame shuts down. Aspirate distilled water for a few minutes with the air still on to cool the burner and flush it out. The flame is now shutdown. Shutdown for the night With the flame off, turn off the acetylene at the tank. Turn on the fuel at the instrument, when the flow meter has dropped to 0 the fuel line has been bled. Now turn off the fuel knob and any other gasses at the tanks then turn off the power.
Igniting the Air / Acetylene Flame: The auto gas box has a special drain sensor that is in line with the drain tube. Fill the sensor with water by unscrewing the top until water is seen coming out the bottom hose. If this is not filled with water the flame will not light. Make sure the drain sensor is hanging vertically. Connect the drain line to the spray chamber and make the electrical connection. The other end of the drain line should go to a plastic acid resistant waste container. 1) Set tank air pressure to 60psi and acetylene tank pressure to 13psi. 2) Press and hold the [air] button on the left panel and check to see you have at least a flow of 5 for both the air and acetylene gases. Adjust the acetylene flow up if necessary or the flame will not ignite then release the [air] button. 3) The interlock lights should all be off except for the [N20 interlock] light. If the burner light stays on then either water needs to be added to the drain sensor or the burner head or blowout plug are not properly in place. 4) Press and hold the [on] button for 5-8 seconds until the flame ignites then release the button. If the flame does not ignite, recheck the gas flows and try again. 5) After ignition adjust fuel down so that it is 1-2 units less then the air unless your analysis requires a higher flow. At this point the [on], [air] and [N20] keys are disabled. 6) To shut flame down press the [off] button and gases will shut down and extinguish the flame Igniting the Nitrous Oxide Flame: 1) Remove the 3 nuts holding the air burner head in place then remove the burner head. 2) Make sure the o-ring is in place then install the nitrous burner, tighten the 3 nuts and connect the interlock pin. Be careful not to over-tighten the nuts. 3) Press the [air] button and check for proper ignition flows. 4) Press and hold the [ON] button for 5-8 seconds until the flame lights. Press the [N20] button and release. The flame will 1st switch to a bright yellow flame then to a bluish, tall nitrous oxide flame. The flame should have a red feather on top of the burner about high, if not, adjust the fuel flow for this condition. Do not lower the fuel so much that the red disappears and try to avoid raising so much that the flame becomes white. The more fuel that is introduced the quicker carbon will build up on the burner slot and cause readings to drift. Some elements may require a higher fuel flow for optimum sensitivity. The carbon can be scraped off from time to time with a long handled screwdriver while the Air/Acetylene flame is lit. See section 3.2 Optimizing the Flame for information on adjusting the flame conditions for maximum absorbance.
1) To shut off the nitrous flame (normal), press the [air] button to switch back to an air based flame, then after 10 seconds press the [off] button. 2) To shut off the nitrous flame (emergency), just press [off] and the system will immediately purge all the flammable gases from the system with a puff of compressed air to extinguish the nitrous flame with a gentle popping sound. This is not a flashback but a safe forced shutdown of the flame.
ALL OTHER INSTRUMENT OPERATION & SET UP IS THE SAME AS THE MODEL 210. PLEASE REVIEW THOSE SECTIONS OF THE MANUAL.
Calibration Reporting: The options are [on] or [off]. Prints calibration information during the calibration procedure as well as the calibration summary of results at the end. Printer: The options are [on] or [off]. Activates the printer. Printer Page Eject: The options are [on] or [off]. Ejects the last page when the analysis is completed. This keeps the next report header at the top of the following page.
Sample Replicates: Designates how many times each sample is to be read. Stabilization time: Designates the length of the read delay before the sample is analyzed. Rinse time: Designates how long the autosampler will rinse between samples.
Bkg energy: 3.132 Bkg abs: 0.002 < CNT : Cal c ntls LS>
Note: You should check the [help] menu for a list of special key functions available to you.
1) Press the [cal] button. The screen will display the calibration form with cup 151 as std1- blank and 152-154 as stds 2 to 4. Note 1: The first standard (at the top of the table) must always be the blank. Note 2: The default settings define Std 2 as the max standard, which is necessary to define for emission analysis but may be ignored for other calibrations. Note 3: If the Calibration Form had previously been edited, other data may be present. Pressing [align] here restores the above default sample data. Note 4: You may wish to set up your own cup numbers and identifiers for use on the Cal Form by editing the sample identifiers under the [smpls] menu (see section 3.1) Note 5: A cup number, name and concentration value must be present otherwise an error will result. Press the [down arrow] to highlight 152 then press [sel] button to move the cursor to the conc column and enter in the concentration of the standards you will be analyzing. Type in the value of the standard and press [enter] or [down arrow] to enter the value. If you have less than 3 standards to run you can highlight the cup you want to remove then press and hold the [upper case] button then press [del] and [enter] to remove the line. If you want to add a 5th or 6th standard type in the cup number of 155 or 156 under 154 and press [enter]. The sample identifier will be brought over from the [smpls] page automatically. To enter more standards you can add appropriate identifiers on the [smpls] page. 2) Press the [cntls] button and select the units to report results in. Select the curve type under degree. Linear requires a minimum of 1 standard, 2nd requires 2 standards, etc. For best results, use at least 1 more standard than required. You should be able to obtain a good calibration curve with 2 or more standards using 2nd order fit. Use the option for 3rd or 4th order curves for more standards only if necessary. Select how many standard replicates to run. The rest of this form contains data from the last calibration for reference only. Press [esc] to return to calibration form. 3) Before starting make sure the instrument and flame have been on for at least 5 minutes. The background corrector should always be on for any analysis below 450 nanometer. Press the [start] button and follow the directions at the bottom of the screen. Aspirate the blank and press the [enter] key. Always allow 5 seconds between aspirating a blank/standard/sample, and reading a sample to let the sample equilibrate in the flame. An absorbance reading of 0.00 will fill in. Aspirate the 1st standard and press [enter]. If you selected to run more than 1 replicate the software will prompt you read the standard that number of times then it will report the average under the abs column. Continue for all standards. 4) When finished press the [cal] button to calculate the curve. If you have made any errors either by selecting the wrong degree or the standards were improperly prepared resulting in a curve that could not be fit, it will be reported here. If there are no errors press the [yes] button to calculate the curve then press the [enter] button to look at the curve. Examine the graph. If the calibration curve appears to be falling off the data points, press [esc], then [cntrls] and change the Polynomial Degree to a lesser value. Then press [esc] then [cal] and recalculate the curve and look at the plot. If your absorbances are higher than about 0.600, you will most likely require at least a 2nd order curve with 3 standards for a good fit as most elements are only linear to this value. If for some reason an improper read was made while calibrating you dont have to perform the whole calibration over. You can highlight the cup position and press either the [a/z] or [read] button to update the absorbance reading then recalculate.
5) After youve viewed a satisfactory curve, press [esc] twice to return to active analysis where you can now read your samples. The instrument display will read continuously as you aspirate your samples. For best results, press the [read] button to have the instrument perform an integration of the signal for the time specified in the library. This will provide a more accurate and reproducible reading as the software is taking an integration over several seconds then reporting an average of the result.
Note 1: It is always a good idea to check a blank and a standard before starting your run and every 5 or 10 sample to insure accuracy. It is possible for any atomic absorption instrument to drift over a period of time due to changing conditions. You can reset your curve if drift occurs by aspirating the blank and pressing [a/z] then aspirating your high standard and pressing [rslp] at any time then continuing on with your samples. Note 2: Your high standard should always be slightly higher than your expected results. If you try to run a sample that is significantly higher than your highest standard you will receive a concentration over-range error message. Note 3: When a file has been loaded into a turret position it will stay there until a new file is selected in the lib screen. When a calibration curve is applied, it to will stay there until a new curve is run. Therefore, if you use the same lamp, turret position and standards and have not loaded a new file to the turret position it is possible to recalibrate the instrument from day to day without even going to the calibration form. Simply let the the instrument and flame warm-up for 5 minutes then from the active analysis screen aspirate the blank and press [a/z] then the high standard and press [rslp]. As long as the absorbance values are close to what they were previously the instrument will recalibrate by adjusting the previous curve that was saved and you can run your samples.
Automated Analysis If you will be using an autosampler and you wish to run fully automated press the [start] key and your samples will run according to the protocol you set in the Smpls page starting at the position you specify. You can set up the Sample Order table under the [smpls] key for the autosampler to read, autozero, or reslope any of your standards before, during or after the run (see section 3.1) You can run manually with an autosampler if you turn read key sampling on. In this mode, after pressing [read], [A/Z], or [Rslp] the screen will ask you to enter a cup location. Enter a number then press [enter] twice. After this sample is read you can press the [up arrow] to select the next sample. Manual Analysis 1.) If you do not have an autosampler you can still use the [start] key. The screen will prompt you for a sample position and that identifier will appear according to what you entered in the smpls page and the sample order table as in the Automated Analysis. This allows you to set up a protocol for checking QC standards during the run so the operator is prompted to check them at certain intervals. After setting up the options under the [smpls] key as described in section 3.1, press the [start] key, enter the starting sample #, press [enter] then simply aspirate your sample and press the [enter] key again. If the Auto-increment option on the [cntls] menu is ON, the sample number will automatically increment for the next reading. Pressing the [up arrow] after the read is complete will advance the cup position to the next position in the smpls page. 2.) If you do not need to set up a protocol for sampling, you can start your run using Read-Key Sampling which gives you the option to press [read], [A/Z] or [rslp] at any time during the run while still using sample identifiers. 3.) If you do not need to set up sample IDs or you only have a couple of samples to run you can skip the smpls page (make sure that read key sampling is off if you do this). Just run your standard curve as usual then go to the active analysis screen, aspirate your samples and press [read]. If you do not want to press [read] for each sample you can press [enter] when you are in the active analysis screen and the instrument will read continuously. Be aware that if you read continuously the result will not be integrated, it will only be a snapshot of the concentration at that instant. You must use the [read] key if you want the result integrated at the time you specified.
1.) sample you can press [enter] when you are in the active analysis screen and the instrument will read continuously. Be aware that if you read continuously the result will not be integrated, it will only be a snapshot of the concentration at that instant. You must use the [read] key if you want the result integrated at the time you specified.
SECTION 4: Basic Operating Procedures - Graphite Furnace Technique THE M-220 GRAPHITE FURNACE SYSTEM & AUTOSAMPLER
The M220 graphite furnace utilizes a rapid, frequencyinduced heating technique, coupled with accessible multistep microprocessor control. The M220 graphite furnace uses 50 L capacity graphite cuvettes (p/n BS007-0699), pyrolitically coated furnace cuvettes (p/n BS009-1504), or LVov platform cuvettes (p/n BS-220-2697). Each element in the 210VGP library has a preconfigured furnace method, which is designed for use when analyzing dilute aqueous samples. The use of matrix modifiers is strongly recommended in all graphite furnace work. Consult the standard conditions listing Section 6 for the elements of interest when selecting the appropriate matrix modifiers. The furnace methods should be adjusted when analyzing other complex samples, for instance biological fluids, where the matrix contains very high levels of dissolved or suspended materials (see Section 2: Editing a Method). ADD PHOTO OF OPEN FURNACE HEAD SHOWING ELECTRODE, TUBE, DETENT, WINDOWS IF POSSIBLE.
(2) If held during power-up: Suppresses power-up reset to allow testing of motors individually. C. TUBE: Moves arm to the furnace HOME or CAL position. D. CUP: (1) Moves arm to cup HOME or CAL position. (2) Holding this button down will first move the arm to a cup position, then lift the arm to the SAFE position, move and/or rotate the tray to position 1 and move the arm into the cup. The CUP and TUBE commands do not affect the calibration. E. STEP UP: If arm has been put in either TUBE or CUP position, moves arm one step up (or repetitively if held). The new arm position is stored as the CALIBRATED position. F. STEP DOWN: If arm has been put in either TUBE or CUP position, Moves arm one step down (or repetitively if held). The new arm position is stored as the CALIBRATED position. 1) Press the [HOME/CAL] button. The POWER light will go off and the STATUS will come on to show that youre operating in HOME mode. 2) Put one of the sample cups into position 1. Press and hold the CUP button for 4 seconds. The arm will immediately move to the cups home position, then, after a few seconds, the autosampler arm will move to the SAFE position, the tray will rotate and/or move to line up position 1, then lower the arm back to the cups HOME position. 3) Using the [STEP DOWN] button, lower the arm step by step into the cup until the Sipper Probe Tip is approximately 1/16 of an inch or more off the bottom of the cup. If you go too far, use the [STEP UP] button to move the arm back up a step. This CALIBRATED position is automatically stored into the autosampler memory. The STATUS light will blink until either the other arm position is calibrated or another motion command is given. 4) Press the [TUBE] button to move the arm to the Home position on the Graphite Furnace side. Using the [STEP DOWN] button, lower the Sipper Probe step by step into the Graphite Tube until its tip is approximately 1mm of the bottom of the tube. With the D2 background lamp off, use the included inspection mirror to look in the right-hand side Furnace Window to check the probes depth. 5) During operation, the POWER and STATUS lights will indicate the state of the autosampler. The following table may be used as a guide for autosampler states: POWER off on blinking(fast) blinking(slow) off off blinking STATUS off off off on blinking blinking STATE power off ready Not communicating Performing a command Home Mode Arm partly calibrated Motor Error (press and hold RESET to clear)
Select phase type: atomize Set (END) temperature 2400 Set time 5.0 Insert/add phase Delete phase <ESC> to exit <ENTER> to save
To Delete the selected phase, highlight Delete Phase, and press [enter]. To Insert a phase Before the selected phase; highlight Insert/Add Phase, and press [enter]. To Modify the existing phase, use the [up/down arrow] keys to scroll to Select Phase Type. Use the right arrow key to select the type of phase desired (listed below), and press [enter]. To modify a Time or Temperature setting, highlight the corresponding line and enter the new value. Press [enter] after each modification or the system will remain unchanged.
Initiallize Init-Miniflow Dry Dry-Ramp Ash Ash-Ramp Atomize Atomize-Ramp Burn Burn-Ramp Cool Autozero Data-Start Data-Stop Int-Flow-On Int-Flow-Off Alt-Flow On Alt-Flow-Off End defaults general operation of gas, internal and external only, sets initial temperature defaults gas flow for internal, external and mini-internal flow, sets initial temperature sets time at temperature for dry stage sets ramp time and final temperature for dry stage sets time at temperature for ash stage sets ramp time and final temperature for ash stage sets time at temperature for atomize stage sets ramp time and final temperature for atomize stage sets time at temperature for burn-up stage sets ramp time and final temperature for burn-up stage sets cool time and temperature time-based event (usually 3 seconds before atomization) time-based event (begin just before atomization, time delays atomize step) time-based event (ends data collection) turns internal internal gas on (mini-flow stays on) stops internal gas flow (mini-flow stays on) turns alternate gas flow on (mini-flow stays on) stops alternate gas flow (mini-flow stays on) default command to stop furnace program
To exit from any screen to the previous screen after changes have been made, press [esc]. The new settings will be active the next time you start the furnace. After all changes have been made, press [esc] from the main controls menu to return to the Active Analysis Screen.
Add 1500ppm NH -NO matrix modifier solution to eliminate halogen losses, can use in conjunction with other 4 3 modifier solutions. Can use up to 4000ppm for very high levels of salt. For samples with high levels of biological/organic carbon: Serum, urine, blood, high algae water, foodstuffs: Make prepared solutions to contain 2-5% HNO to help oxidize the organic carbon and reduce carbide losses, 3 can use in conjunction with other matrix modifier solutions. General recommendations: Refractory metals (Ti, V, Cr, Si, Al, Mo, W, Ba) will form carbides unless nitrate (NO3 salt or acid) is present. 0.2% Nitric acid should be added in addition to matrix modifiers. For the hydride metals, the preferred modifier is 1000ppm Mg-NO3 + 200ppm Pd-NO3 since palladium will have the same basic affect as Ni-NO3 and not contaminate the graphite tube with nickel for future analyses. Most of the heavy metals form volatile halides, so be sure that there is no salt left after the ashing step (you can see the smoke coming out the top of the graphite tube). Graphite Tubes: There are three basic types of tubes used in the graphite furnace: Pyro-Coated Tubes are used for most elements. The pyrolytic graphite coating improves the sensitivity of carbide-forming metals Un-coated Tubes are more economical and provide greater reproducibility for certain elements including Lead and Cadmium. Pyro-Platform Tubes provide a flat surface for sample injection. Although these tubes are generally more expensive, they provide greater longevity and possibly better stability. Before calibration, it is best to make sure the best tube for your analysis is installed and that it is in good condition. See Section 11.2 for instructions on changing tubes. Gas Flow Settings Note: The furnace must be activated using the Read or Start key before flows may be adjusted. The Buck 220GF Power Unit has 4 flow meters on the front: 1) Sheath Gas is Argon flow outside the tube. Its purpose is to prevent the Oxygen in the atmosphere from causing the rapid combustion of the Graphite Tube. The Sheath Gas should be set to 1.25 mL/min, or about half way up the scale. 2) Internal Flow is Argon flow inside the tube. The purpose is to remove unwanted compounds from the furnace tube as they are vaporized or ashed. The Internal Flow should be set to 200 uL/min. The internal flow is turned off during Atomize to allow data collection. 3) Mini-Flow is also Argon gas flow in the tube, but the flow remains ON during atomize. Mini-Flow is used to reduce the sensitivity of the Analyte and/or peak duration. The flow is normally set to 50 uL/min to start, then increased to obtain the desired peak. 4) Alternate Internal Flow is the optional flow of another gas (possibly Oxygen or Air) used during the analysis of special types of samples.
Calibration 1) Make sure the argon regulator is set to 40 psi and the valve is open. Check that the cooling water is flowing (about 1.5 liter/min is sufficient). 2) If you would like to use the Auto-peak Timing, make sure that Peak Timing is set to Auto in the Furnace/ Vapor Controls menu, then skip to step 4. 3) If you are setting the peak timing manually using the fixed timing mode, make sure that Peak Timing is set to Fixed in the Furnace/Vapor Controls menu. Read your high standard using the [read] button. After the furnace program ends, Press [esc], then [cntls], then highlight Furnace/Vapor Controls and press [enter]. Use the Replot Last Peak function to examine the peak and select the proper peak delay and duration so that the entire peak is integrated including the tail. Press [esc] then enter the values in for peak delay and duration. You may return to replot last peak to check to see if the white bar, which denotes the area of the curve which will be integrated, covers the entire peak. 4) Press the [cal] button. By default the blank is in cup 41 and stds 1-3 are in cups 42-44 but you can specify any location. The blank must be in the top location on the calibration form. Enter in the concentration values and curve parameters just as with flame (see section 3.3). For best results, run at least 3 replicates for your blank and standard when running a graphite furnace analysis. The absorbance for your blank will be taken into account when the calibration curve is calculated. 5) Press the [start] button. If you have an Autosampler, it will first inject the high standard and run the furnace program. If you are using Auto-Peak Timing, the software will determine the best integration window for the peak then continue on with the blank and all other standards. If you dont have an autosampler , the software will give directions at the bottom of the screen to instruct you when and which standard to pipette by hand. 6) When the standards are done, you can examine the data for inconsistencies. If one of the standards is not consistent, you can read that standard again by using the arrow keys to highlight the standard and press read. The software will automatically run the same # of replicates for the standard. After you are satisfied with the data, press [cal] to calculate then [enter] to view the plot. If the curve is satisfactory, press [esc] 3 times to return to the active analysis screen. Automated Analysis If you will be using an autosampler and you wish to run fully automated press the [start] key, enter the cup number you would like to start with and your samples will run according to the protocol you set in the Smpls page (see Section 3.1). You can run manually with an autosampler if you turn read key sampling on. In this mode, after pressing [read] the screen will ask you to enter a tray location. Enter a number then press [enter]. Manual Analysis 1) If you do not have an autosampler you can still use the [start] key. The screen will prompt you to inject samples according to what you entered in the Smpls page. Using the [start] key, the system will instruct you to run the specified number of replicates for all samples. Simply inject your sample when prompted and press the [enter] key. The software will automatically turn off the internal flow for better manual injections. 2) If you do not need to set up sample IDs or you only have one test sample to run you can skip the Smpls page. Just run your standard curve as usual then go to the Active Analysis screen. With read-key sampling OFF, inject your sample and then press [read]. If you would like a printout turn read key printing on. The software will not turn the flow off for a second injection. 3) If you run samples using the [read] key with read-key sampling ON, you can run any number of replicates for each sample and the software will automatically turn off flow before each injection. Sample Ids will be used, but if they are not necessary you can leave the default S1-S40. Turn read key printing on to print the
3) 4)
5)
6) 7) 8) 9)
1) Maximum Peak Timing; Sets the length of the peak window for Furnace and Vapor methods. Select 14 or 28 second windows using the arrow keys. 2) Background Gain; Preset for best results. Automatically set with Autozero from the align screen. 3) D2 Background; Preset ON for elements which can make use of it. 4) Wavelength, Slit; Designations you can change for libraries using alternate wavelengths. 5) Lamp; Designation you can change for library using a different lamp. Calibration Curves Calibration parameters (options under calibration controls except replicates) can be saved for the active analysis. The points for the calibration curve will not be saved. However, the curve without the points can be viewed for future reference. Saving Libraries You cannot overwrite the master library files that are preprogrammed into the 210VGP. However, you can save this information by creating a new library name and saving the current analysis information under this new name. After making your changes, use the [esc] key to get back to the active analysis screen, then press the [lib] key. Enter the desired password by selecting <3>.If you have not already changed the name under Measurement Controls , select <1> Enter a library name. The current library name will then come up (Example : Pb-Furn3283.2). You can then use the right arrow key to move the cursor to the part of the name youd like to change. Then, using the [del] key, you can erase some or all of the name. After typing the new name, it may look like this example: Pb-Blood-283.2 (in this case the new name describes a library for running lead in blood). To save the changes, select <4> Save Active Analysis, and respond by pressing the [yes] key when prompted. You can delete a Library youve created by selecting <5> Delete Active Analysis. You cannot alter or delete the master library files.
c(x) =
where S(A) and S(B), are the analytical absorbance readings (corrected for the blank) of solutions A and B, respectively. Vs, and cs, should be chosen so that S(A) is roughly twice S(B) on the average. It is best if Vs, is made much less than Vx, and thus cs, is much greater than Cx, to avoid excess dilution of the sample matrix. If a separation or concentration step is used, the additions are best made first and carried through the entire procedure. For the results from this technique to be valid, the following limitations must be taken into consideration: 1. 2. 3. 4. The analytical curve must be linear. The chemical form of the analyte added must respond the same as the analyte in the sample. The interference effect must be constant over the working range of concern. The absorbance data must be corrected for any additive interference.
When greater accuracy is required, the method of standard addition SA.2 found on the following page is recommended:
SA.2 Add equal volumes of deionized water and three standards containing different amounts of the test element
to 4 aliquots of the sample. The aliquots must also be of equal volume. Read the absorbance of each solution and plot as shown below. The concentration of the standards is taken as the X value, with the sample assigned the value X=0. When the resulting line is extrapolated back to zero absorbance, the point of intersection with the horizontal axis is the concentration of the unknown.
A b s o r b a n c e
.20 .18 888 .16 .14 .12 .10 000 .08 .06 .04 .02 .00
-2
-1
1 Concentration (ppm)
(courtesy of BUCK SCIENTIFIC, Inc. Applications Department, Norwalk, CT). The method of standard addition is subject to certain limitations, which must be taken into consideration when examining the results. The curve must be within the liner range of the analysis. For the best results, the slope of the curve should be nearly the same as that of the standards alone. The diagram above shows a typical relationship between the sample analysis (upper curve) and the curve of the standard solutions. If the slope of the standard addition curve differs by more than 20% of the standard curve, the results are suspect. In addition, the effect of interferences should not vary with concentration of the analyte or other components in solution. Spectral interferences are not corrected for by this method; use suitable background correction (i.e., deuterium, giant pulse, etc.). Graphing the results enables the analyst to visually determine the validity of the results by checking for linearity, and by comparison with a curve of the standard solutions; however it leads to some uncertainty in determining the concentration of the unknown. For the highest precision, the unknown should be determined by calculation from: [u] = 0.25{(x) - [{4x - (x)}/{4xxy - xy}]} where: [u] is concentration of unknown [y] is an absorbance value for each corresponding concentration, x
There are about 13 elements that form refractory oxides and cannot be dissociated in even the hottest airacetylene flame. It is necessary then to use a nitrous oxide-acetylene flame for these elements. The N-A flame has the advantage of being able to decompose refractory compounds, but suffers from relatively higher noise caused by emission radiation from combustion by-products (CN, CH and NH). These by-products can also cause specific interferences with some elements where the emission spectrum overlaps an absorbing line. Sometimes this type of interference cannot be removed by background correction, making analysis virtually impossible. In the very hot N-A flame, an ionization suppressant must be added to the sample (usually a Potassium or Lanthanum salt) to prevent the analyte from being lost to the analysis through ionization. Elements with characteristic wavelengths near the start of the vacuum UV range show considerable improvement in sensitivity with an argon-hydrogen flame. It is not yet certain what atomization mechanisms occur, however, it is generally agreed that hydrogen has an active role in the process. Because of its very high transparency the ArH flame gives particularly good sensitivity for Arsenic and Selenium. A more sensitive technique, hydride generation, doesnt require any modification of the combustion mixture. In this technique Arsenic and Selenium are converted to the Arsine or Selenine gas and swept into a quartz cell heated by the flame. The hydride technique is used when highest sensitivity for these elements is required. In
A more sensitive technique, hydride generation, doesnt require any modification of the combustion mixture. In this technique Arsenic and Selenium are converted to the Arsine or Selenine gas and swept into a quartz cell heated by the flame. The hydride technique is used when highest sensitivity for these elements is required. In this case the A-A flame is used but only serves as a convenient source of heat. The technique uses sodium borohydride to reduce the As or Se to free metal then delivers it to a Quartz Tee Tube which is warmed by the flame in beam of the Atomic Absorption Spectrophotometer. Buck Scientific offers the Model 1018 Hydride/ Cold Vaporaccessory which provides gas controls, glassware and mounts for this analysis. This system makes use of the Buck 210VGPs peak-picker software to analyze the data. Buck Scientific also offer the Model 420 Continuous Flow Hydride system which allows you to analyze data using flame integration.
Since elemental Mercury has a significant vapor pressure at room temperatures, and is extremely volatile at 500 C, it is best performed using a flame-less technique. The most common is the Cold Vapor technique which reduces the Mercury to free metal in an acid solution using stannous chloride or sodium borohydride then delivered to an absorption cell mounted in the beam of the Atomic Absorption Spectrophotometer. Buck Scientific offers the Model 1018 Hydride/Cold Vapor Accessory for Mercury analysis using the Buck 210VGPs peak-picker software.
SECTION 9: Interferences
There are basically three categories of interferences that can occur in flame atomic absorption work, termed physical, chemical, and spectral. Chemical interference is most often encountered and is caused by lack of absorption of atoms bound in molecular combination in the flame. This occurs when the flame is not hot enough to dissociate the molecule. Phosphates interfere with Magnesium, Calcium and Barium, and are overcome by adding Lanthanum to the solution. Similarly, Silica interferes in the determination of Manganese and can be eliminated by the addition of Calcium. Forming other compounds with the interfering materials may also eliminate chemical interferences. Although matrix modifiers are primarily added to increase the sensitivity of the analysis, they may also be used to eliminate or reduce interferences. Highly refractory metal oxides, especially those of the rare earth metals, do not dissociate at the temperature of an air-acetylene flame. Other metals dissociated into the atomic state often recombine with oxygen in the flame so rapidly that further atomization is not possible. In these cases an alternate combustion mixture is used, most often a nitrous oxide-acetylene flame, to provide greater heat for decomposition. If an element in the atomic state becomes ionized in the flame its absorption spectra will change, effectively removing it from the analysis. The fraction of ionized atoms in the flame increases with increased temperature, and at the heat of a nitrous oxide-acetylene flame nearly all elements are significantly ionized. This type of interference is most pronounced for elements such as Barium, which is readily ionized but requires high temperature excitation for analysis at the usual concentration range. Ionization can generally be controlled by the addition of a large excess (>1,000 mg/L) of an easily ionized element such as K, Na, Li or Cs to the sample. All metals are not equally stable in a digested solution, especially if it contains only nitric acid, and not nitric and hydrochloric acids together. The digestate should be analyzed as soon as possible, with preference given to Antimony, Barium, Molybdenum, Silver and Tin. Precious metals are less likely to plate on to plastic containers than glass when making dilutions. High concentrations of dissolved solids in the sample may result in interferences from physical (non-atomic) absorbance such as light scattering. If background correction is not used, the sample can be re-analyzed at a nearby, non-specific wavelength*. If an absorbance is found at this wavelength, it is due to a physical effect and the sample should be diluted and/or treated by a filtration, digestion or extraction procedure to remove the interference. * All hollow cathode lamps emit not only the line spectra of the element comprising the cathode, but also that of the fill gas and
other incidental impurities; therefore, it is always possible to find an energetic line somewhere near the resonant wavelength of the element of interest which will not respond to the element, but will respond to physical interferences.
Physical Interference can occur with some sample matrices when the instrument is calibrated using aqueous standards. Samples and standards should be monitored for viscosity differences that may alter the aspiration rate. These interferences may be eliminated by diluting the samples or by mixing standards in a similar matrix. Spectral interference can occur when an absorbing wavelength of an element present in the sample but not being determined falls within the width of the absorption line of the element of interest. The results of the determination will then be erroneously high, due to the contribution of the interfering element to the atomic absorption signal. Interference can also occur when resonant energy from another element in a multi-element lamp, or from a metal impurity in the lamp cathode, falls within the band-pass of the slit setting when that other metal is present in the sample. Narrowing the slit width may sometimes reduce this type of interference. Molecular spectra of certain common compounds have broad absorption profiles and can produce a positive interference; that is, the measured absorbance is greater than the actual absorbance of the analyte. The table below illustrates some common molecular absorbance bands: Table 3: Overlapping Spectra of Some Common Analytes (Source: Norris & West; Analytical Chemistry; 1974, V46, p. 1423).
Analyte Aluminum Bismuth Calcium Cadmium Chromium Copper Wavelength 396.15 206.17 422.67 228.80 359.35 217.89 324.75 324.75 327.40 253.65 241.16 213.86 217.00 323.26 285.21 285.21 279.48 403.31 253.65 231.10 352.45 247.64 271.90 338.29 460.73 250.69 308.21 213.86 Overlapping Element Fe I Ge As Hg Ne Sb Fe Eu Fe Hg Pb Zn Sb Sb Fe Hg Fe Ga Co Sb Fe Pb Fe Fe Fe Si Al Fe Wavelength 396.11 206.16 422.66 228.81 359.35 359.35 217.92 324.73 324.75 327.45 253.65 241.17 213.86 217.02 323.25 285.21 285.24 279.47 403.30 253.65 213.15 352.43 247.64 271.90 338.24 460.77 250.69 308.22 213.86
Cobalt Iron Lead Lithium Magnesium Manganese Mercury Nickel Palladium Platinum Silver Strontium Vanadium Zinc
PROBLEM: DRIFT OR FLUCTUATION IN READINGS POSSIBLE CAUSES: 1) Lamp. To determine if it is the lamp turn off the flame and all gasses. Zero the absorbance and watch the display. After warmup drift should be less than 0.001 per minute and noise should be less than +/- 0.002. Most lamps perform much better than this. If this is stable the problem is probably with the burner system. If a lamp drifts or is noisy, selecting a different operating current may help. Lamp energy values for the 210VGP should be 3.000 or greater in a 0.7nm slit and in the upper 2s to low 3s for a 0.2nm slit. 2) Burner system. If drift or noise only occurs during your run then the burner system is in doubt. Check that the drain is working properly. There should be a steady drip or flow. If there is any water buildup in the drain block you will most certainly get a decrease in absorbance. A gurgling sound from the burner is a good indication of this. Make sure the small clear tubing is in the drain elbow to eliminate air bubbles in the elbow. Make sure the end of the drain tube is not submerged in the waste water. Try readjusting the nebulizer. A new nebulizer may be needed. Make sure there are no leaks in the burner, check the o-rings. 3) Thermal drift. If the lab is subject to temperature changes the optical bench of the instrument may shift causing a slight change in energy. To determine if this is the problem peak up the wavelength for maximum energy and zero the absorbance. If after a period of time the absorbance drifts go back and repeak the wavelength. If you can adjust it for more energy then the wavelength has drifted off peak resulting in an absorbance change. 4) Vibrations in the lab can also cause wavelength drift. Make sure any vibrating equipment including air compressors are not near the instrument. 5) Unstable supply gas pressure. Although the instrument has internal regulation, supply pressure change can cause fluctuation of absorbance for any element that is flame sensitive, particularly iron. Air supply: Many failures of the pneumatics or excessive noise in results can be attributed to contaminated air or acetylene. An air filter is a must when using an air compressor for your supply to filter out oil, water & particulates. It should be cleaned on a regular basis. The inside of the plastic bowl should be cleaned with water and soap and the filter element with ethyl alcohol or similar solvent. Refer to the manufacturers instructions for complete information.
PROBLEM: YELLOW / ORANGE FLAME POSSIBLE CAUSES: 1) Acetylene: If you notice your flame becoming orange in color and it is not due to your samples there is acetone coming from the tank, you should shut down when this is noticed. A new tank should sit for at least several hours undisturbed before use to let the acetone settle. Eventually liquid acetone will appear in the flow tube of the acetylene. For this reason do not let tank pressure drop below 75psi. As your tank pressure drops more acetone will be introduced resulting in decreased absorbance signal and increased background levels.MISCELANEOUS: 1) A 0.500 absorbance screen is supplied with the instrument. It should result in .450 to .550 absorbance when inserted into the light path. This indicates the electronics are working properly.2) Your standards, samples and blank should be prepared in the same matrix as your samples so as to avoid erroneous results.
BACKGROUND ERROR The signal from the D2 lamp has changed. Pressing [a/z] should reset the gain circuitry and clear the error. CANT DIRECT REPORT The user is trying to output data but the printer or remote computer has not been turned on in the software. A printer or remote computer must be connected to use the [START] function from the Active Analysis screen. PRINTER ERROR The instrument is trying to print but the printer is not responding. FURNACE NOT INSTALLED: Furnace Accessory OFF on the initialization screen. Turn the Buck 210 off, then hold the uppercase key down and turn it back on to get to the initialization screen. After turning the option ON with the left or tight arrow key, you must press <enter> to lock in the setting. Press <esc> to return to Active Analysis VERY LOW ENERGY or NO ENERGY 1) Check that the lamp turret # and socket # being used match the lamp number at the top of the screen. Only one lamp at a time is electrically modulated so if the incorrect position is selected you will receive a low energy error or lamp current error. The lamp # is changed by pressing the sel button in either the active analysis or lib mode. 2) Pull the lamp out of the turret and reconnect then look at the lamp. The image should be a bright dot in the center, the image should not be fuzzy, diffused or arcing to the glass. 3) While looking at the lamp press the sel key continually from the active analysis screen. When the lamp number at the top of the screen matches the lamp you are looking at you should see a change in brightness of the lamp. This indicates that the lamp is being modulated and that the lamp circuitry is probably functioning correctly. 4) Check that the wavelength and slits are set correctly. 5) Make sure the proper element lamp is installed. 6) Check all lamp positions. It is possible that only 1 position is not working correctly. 7) If the problem only occurs when the background corrector is on then there may be an energy mismatch between the hollow cathode lamp and background lamp. Autozero from the Align screen should correct the gain to offset this mismatch. 8) Turn the wavelength to read 000. At or very near this wavelength the grating is parallel to the slits and the energy should be very high (5.0 or greater). 9) Remove the top right cover of the instrument and check that the wavelength mechanism appears to be operating as you turn the wavelength knob. The Grating arm should be perpendicular to the wavelength micrometer such that the stainless steel ball lines up with it.
1) SECTION 10.3 Replacement Part Assembly Instructions SECTION 10.3.1: D2 lamp installation and alignment
1) Remove the top left cover of the 210 be removing the 2 screws at the back. 2) Before removing the old D2 lamp, peak up the energy and wavelength and lamp alignment on any hollow cathode lamp. Note the energy value from the align screen on a piece of paper. 3) Turn off the power to the 210. 4) Disconnect the cord to the D2 and remove it from its holder. 5) Install the new D2 lamp in the holder and connect the cord. 6) Turn power on to the 210. 7) Press <ALIGN> 8) Align the D2 lamp by sliding the lamp back and forth in the holder and/or loosening the bracket and moving it up and down. . Roughly center the lamp so the image from the hollow cathode lamp is centered through the hole of the D2 lamp. Use the bar graph as you would to align the hollow cathode lamp alone except in this case you need to maximize energy for both the D2 lamp and hollow cathode lamp signals. 9) Fine tune the hollow cathode lamp adjustment. You should not have to make any large adjustment to maximize the HCL. The value should be very close to when the old D2 lamp was there. 10) Make sure everything is tight and re-install the cover.
Wavelength calibration
1) Press the (align) button and slowly move the wavelength back and forth around 000. The highest energy should occur at 000 wavelength, the numerical value for the energy should be above 5.0. If this is not so follow the procedures below: 2) Pull the wavelength knob out to disengage the gears then rotate the digits (holding the gear in place) so that 000 coincides with the highest energy reading then tighten the gear. If the highest energy appears to be below zero than the gear will have to be loosened first so that the wavelength can be turned negative to find the peak. 3) Check the upper end by tuning to the wavelength of the lamp in use and see that the highest energy occurs at the specified wavelength. Make any further adjustments as necessary. 4) Turn off power and replace covers.
SECTION 10.3.4: Removing the electronics modules Make sure the instruments power is turned off before removing any electronic modules since High Voltage may be Present!
FRONT PANEL 1) Remove top right cover by loosening the two screws holding it in back. Lift the back of the cover enough to clear the screws, then pull the cover backward to remove. 2) Remove the 2 screws from underneath the front panel. 3) Remove the front most screw from the right side panel of the 210 with a 3/32" allen wrench (supplied in 210 packing kit).. 4) Remove the screw from the left side of the front panel. To do this the screw can be accessed through a hole in the burner shield. On older 210s the shield may have to be removed by taking out the 3 screws around the perimeter of the shield, once screws are out the shield can be removed. 5) The front panel can now be removed. If the cables need to be disconnected, label the cables before removal to make re-installation easier. 210-0616 SOFTWARE CHIP SET 1) Remove the front panel. You should not have to disconnect any cables to change the software. 2) Look for the EPROMs located in sockets U224 and U225 in the bottom left hand corner of the circuit board. These chips are rectangular and have a small white label reading V3.XX, C&D and V3.XX, E&F. You can remove the chips by carefully prying them up using a flat head screwdriver, as long as it does not have a magnetic tip. 3) Install the new EPROMs in the proper location taking care not to bend any pins in the process. Keep in mind that the notch in the top of the chip should line up with the notch on top of the socket. The socket number on each chip should correspond with the number below the socket on the board. Take care to make sure you have grounded any static charge before installing the chips since they are static sensitive devices. 4) If you need to install a new 210 library chip as well, youll find the library SRAM in socket # U222 near the EPROMs. You can carefully pry it up with a screwdriver like the others. When installing the new library chip, there is a recessed dot in the corner of the chip that should end up on the end of the socket where the notch is. Be careful not to bend any pins and double check that all pins are in the socket. Double check that the top of the chip (the end marked with the recessed dot) is on the notched side of the socket or the library chip will be permanently damaged when you power the instrument up. 5) Turn on the instrument while holding down the upper case key. keep holding the upper case key until the initialization menu appears on the screen. If screen is blank, double check that the chips are installed in the right location, are oriented properly and have no bend pins. 6) On the initialization screen, use the left and right arrow keys to change the furnace, furnace autosampler, flame autosampler, printer and/or clock options to yes and press enter after each change. Change only the options for the accessories you have. 210-2000 SYSTEM PCB 1) Remove Front panel 2) Remove the intensity knob from the front panel. 3) Disconnect the Display Ribbon Cable.
1) Remove all screws holding the board to the panel. 2) When installing the new system PCB, be careful to align the connector for the keyboard before screwing down. 210-2002 KEYBOARD 1) Remove System PCB 2) Remove screws keyboard to the front panel 3) Remove the standoffs holding down the keyboard while making note of their location. 998-1046 LIQUID CRYSTAL DISPLAY 1) Remove the System PCB 2) Remove screws holding the display to the front panel 3) Transfer the backlight power cable extension to the new display and run past the bottom of the system PCB before re-assembly. 210-1006 POWER SUPPLY 1) 2) 3) 4) Remove front panel but you do not have to disconnect any ribbon cables. Unplug the power connectors from the Power Control PCB. Remove the screw from the bracket on the front of the power supply Remove 2-4 screws holding the power supply to the 210 chassis in the rear
210-2006 POWER CONTROL PCB ASSEMBY (LAMP POWER SUPPLY) 1) Remove the Front Panel. Only the large 36pin ribbon cable going to the Power Control PCB needs to be disconnected. 2) Unplug the 3 power supply plugs from the right side of the Power Control PCB. 3) Remove the 3 or 4 screws holding the finned heatsink to the rear of the 210. 4) Slowly pull the supply outward to expose the aluminum cover but do not pull all the way out. Remove the screws holding this cover on. Again, label all cables before removal of supply to make things easier. Once all cables are removed the supply can be pulled out. DETECTOR (PMT OR PHOTOMULTIPLIER TUBE) 1) Remove the top right cover 2) At the rear left of the optical base is a small black cover held in place with 2 screws, remove this cover. 3) CAREFULLY!!! remove the detector from the socket by wiggling back and forth while at the same time pulling up (try to grab the detector by its base so as not to break the glass envelope). Once removed, put in a safe place. The detector is very expensive so exercise extreme caution. 210-2004 PREAMP ASSEMBLY 1) Remove the Front Panel Assembly. 2) Remove the PMT 3) Remove the 2 screws on either side of the hole where the detector was. You will need to hold the nuts underneath the optical base to prevent them from turning. The preamp is enclosed in the small metal case. Once the screws are removed the pre-amp assembly can be taken out.
211-0621 FLAME CONTROL PCB 1) Remove left side top cover by loosening the two screws holding it down in the back. Then lift the back up to clear the screw heads, slide the panel back then lift the front up to expose the PC Board. 2) Disconnect the 4 cables to the board making note of there location. 3) Unscrew the transistor from the cover first, then the 4 screws holding down the board. 211-0623 FLAME SENSOR PCB 1) Remove the burner assembly including the X-Y Table by removing the two screw in the front of the burner compartment. 2) Remove the 5 screws securing the burner shield and slide it out far enough to access the flame sensor PCB 3) Be careful with the white wire, which is the low current 500V feed for the flame sensor which does not go down immediately after power shutdown due to capacitors on the line. 4) Disconnect the two plugs to the board and unscrew the two screws to remove the board. 5) When installing the new board, make sure the small hole in the black flame sensor cover lines up with the hole in the shield. If not, slide the sensor up or down in its holder to align. 220GF REPLACEMENT PARTS IGBT PCB ASSEMBLY, PCB MAIN 1) To find out if the IGBT assembly is bad we first need to check theyre drivers. When the IGBTs burn out they almost always burn out theyre drivers and/or regulators as well as possibly damaging some lands on the Printed Circuit Board due to excessive current drawn through these circuits. As a result, we recommended that the PCB be replaced along with the IGBTs. With the IGBTs (plugs P8 & P9) disconnected you can power up the bad board to see what other components are bad. DO NOT TAKE MEASUREMENTS ON THE CIRCUIT BOARD WITH THE IGBTS CONNECTED BECAUSE THERE COULD BE LETHAL VOLTAGES ON THE BOARD WHEN THE IGBTS ARE CONNECTED. Also and accidental shorts on the IGBT driver section of the board could blow the IGBTs if they are connected. With the IGBTs (P8 & P9) disconnected you can power up the furnace board and make the following measurements with a DC voltmeter to determine what components are blown: a. Measure the output voltage of each LM78L15 Regulator (U216, U221, U218 ) to see if they are putting out 15 volts. If any are putting out the proper voltage then that regulator and its associated driver chip are probably both O.K. b. If any are not, remove the associated TC1426CPA driver chip and check the voltage again. If the output of the regulator is O.K. now, just the driver chip is blown. If the output of the regulator is still not approximately 15volts, then both the regulator and driver chip are bad. c. If any of the regulator or driver chips are blown then the IGBT PCB assembly (part # 220-1090) must be replaced. The main PCB (# 220-0021) must be repaired, however we recommend replacement since the board cannot be tested under similar load without risking another set of IGBTs. The IGBT are high speed, high voltage switching devices that control the duty cycle controlling the temperature of the furnace. It is often impossible to explain their demise since the driver circuitry is damaged when they blow. Good grounding is important since the circuit breaker will not trip if all circuits are not properly grounded. Before connecting a new IGBT PCB Assembly to a repaired main PCB, compete the Furnace Printed Circuit Board Test as outlined on the next page. 2.Disconnect 220VAC Furnace Power Cord.
3. To make it easier to work on the furnace, rotate the instrument so you can access the Furnace Control Box from the rear. To accomplish a complete electronics test you will need to access the instruments software so rotating the instrument clockwise 90 degrees and supporting the right side with a tall stool or cart is a good way to accomplish this without disconnecting water or gas lines. 4. Remove the 6 screws on the top part of the back cover on the Furnace Control Box and fold the top half down, exposing the Circuit Board. Remove the screw holding the regulator (U222) to the heat sink shield, then remove the 4 screws holding down the shield. Clip the wire tie holding the IGBT wires down to the back cover if applicable. Carefully disconnect the other 3 or 4 plugs across the top of the circuit board, then lift up the top half of the back cover and remove the 6 screws on the bottom half of the back cover to remove the back cover with the circuit board attached. 5. Disconnect the incoming water line from the outside of the IGBT heat sink and the outgoing from the inside. Remove the 4 IGBT mounting screws on its heat sink and remove the IGBT PCB assembly with the sink from the power unit box such that you can access the terminal block. 6. Loosen the screws on the terminal block and remove the wires one by one, making note of their location for re-assembly. Check all connections to the IGBTs for corrosion and make sure there are no stray strands that could short to another terminal. A bad or erroneously wired connection on one of the IGBT terminals can cause them to blow again. Reconnect the transformer and capacitor according to the following wiring diagram:
4 Black Wire, 170 V from Capacitor 3 Black Wire, Transformer Lead 2 Black Wire, Transformer Lead 1 Red Wire, +170 V from Capacitor
If your unit has a long ground wire running behind the transformer up to a ground terminal, disconnect it. Use the short ground wire on the new assembly instead and use the included hardware to tie it to the existing hole in the shelf. 7. Feed the control cables through the hole and mount the assembly. Connect the water connections to the heat sink 8. Attach the back panel with the NEW PCB attached. Carefully connect the 4 plugs to the new circuit board making sure all pins are inside the receptacles. DO NOT RECONNECT IGBTS (P8 & P9) AND TRY TO RUN THE FURNACE WITH THE OLD CIRCUIT BOARD OR YOU WILL BURN THEM OUT AGAIN. If and plugs are misplaced the IGBTs may blow again. 9. Attach the Shield, Heat Sink to the standoffs and secure the Regulator to the shield. Be sure to connect the ground wire to the shield. Close the Furnace box up and test with low power first.
The following test procedure is used to make sure the main PCB is working properly to avoid damage to the IGBTs: 1. Physically inspect the circuit board with a magnifier looking for bad solder connections, short circuits or burnt lands especially if there has been a problem with the power part of the circuitry previously. The IGBTs are sensitive to short circuits and poor contacts even if they are only momentary. 2. The printed circuit board tests MUST be performed with the IGBTs disconnected since there is LETHAL VOLTAGES on the board with IGBTs connected and the power on as well as the potential to damage the IGBTs while taking measurements on the board. With the power off, make sure the IGBTs (P8 & P9) are disconnected, then power up the furnace. 3. Using an oscilloscope with a trigger lead, connect the trigger to the top (closest to driver chip U217) of 10k resistor R230, and its ground to the bottom of C217. Set the time/div to .1 ms. Connect a 10x probe to pin 1 on the P8 header (right most pin) and its ground to the bottom of C217. Set the volts/div to .5v. You will see the waveform on the scope until you activate the furnace from the 210VGP. 4. On the 210VGP, load a furnace library into one of the lamp positions. Under Controls select furnace cleanup cycle. Set the cleanup time for 99sec. Highlight Run Cleanup Cycle and press enter. If the furnace responds with Furnace Not Responding or repeats the message Furnace Initializing for more than one minute, then there is a communication problem. Turn the furnace off, wait a few seconds, then turn it on and try again. If it is still not communicating, check that the furnace communications cable is properly installed, and check that the outputs of regulators U214, U215, and U222 are +12vdc, -12vdc, and 5vdc respectively. You can also check the handshaking signals from the 210 on J1, using the schematic as a guide. 5. When the furnace responds, on the scope youll first see the baseline with a line of small dots above it representing the small duty cycle for furnace initialization. Then you should see the still wave form of the burn duty cycle which looks something like this: ___ ___
_______
___________
_______
If you get this waveform, then this IGBT Driver is operational. Try the same test on P9, pin 1 (leftmost pin), but youll have to move the trigger to R231 with ground C220 since the IGBT drivers are completely Isolated. Both P8 pin 1 and P9 pin 1 must have the same waveform to properly fire the IGBTs. 6. The center driver on the PCB drives one gate in each IGBT package. To test, hook up the trigger to R233 and ground to C223 like above. Compare the waveforms on P8, pin 7 and P9, pin 7. The waveforms should be the same as the other driver outputs but the two should be out of phase as shown in the diagram below: ___ P9, pin7: _______ ___________ _______ ___
___
___
___________
___________
7.
Both these Waveforms need to be present and out of phase to ensure the IGBTs are driven properly. Keep in mind that you may need to restart the furnace cleanup cycle on the 210VGP if you need more time to complete the tests. This test does not ensure the IGBTs are 100% safe to connect since the test cannot be made under a proper load. However, if you complete this test, inspect the wiring to ensure it agrees with the wiring diagram after servicing and check for corroded contacts or bad solder connections you can be confident that you should not have problems Press <esc> on the 210 to end the cleanup cycle when you are done. TURN THE POWER TO THE FURNACE OFF before connecting the IGBTs to P8 & P9 and closing the back of the furnace.
PROCEDURE FOR INSTALLING SOFTWARE IN THE BUCK 220GF 1.)Make sure the power is shut off and unplug the furnace box. Remove the top seven screws on the back of the 220GF furnace box and let the top half of the back down exposing the circuit board. Remove metal shield covering the circuit board if applicable. DO NOT disconnect any wires and be careful not to damage any of the wiring. 2.) Look for the socket labeled U110 (this number will be upside down as you are looking on the board). The chip in U110 is a 512K EPROM in a slightly larger socket so that we can expand to a bigger capacity EPROM for future upgrades if necessary. Before removing this chip, note its orientation in the socket. The left side of the chip ( the side with no notch ) is all the way in the left side of the socket. the right side of the chip(notched) falls two pins short of filling the socket in each row. The notch in the chip is still on the same side as the notch in the socket. The new chip is the same size and is therefore installed in the same manner. 3.) After youve examined the position of the chip, you can remove it by carefully prying it up with a flathead screw driver. Please do not discard the chip because it can be reprogrammed for future updates. Please put it in the packaging the new chip came in and return it to us. 4.) Install the chip marked FU110, V1.20 into the socket marked U110 toward the left side of the socket as described in step 2. Make sure the notch on the chip is on the same side as the notch on the socket and be careful not to bend any pins. 5.) In order to test this new software, you must have V3.54 software installed in the Buck 210VGP, have a furnace library loaded (Pb Furn3 for example) for a lamp, and have your water and argon gas connected to the furnace. If youve never run the Buck 220GF furnace before, read the instruction manual before running. 6.) With the read key sampling and read key printing options set to off under the controls menu, press the read button to start the furnace. If ,after a delay , the instrument responds with furnace not responding, check to make sure the chip as well as the furnace cable are properly installed. Then press esc twice on the 210VGP, turn the furnace power off, wait a few seconds and turn the furnace back on before trying again. If the furnace responds with a furnace ready message, you know the software is working . 7.) Press ESC to stop the furnace program.
PROBLEM: FLAME KEEPS SHUTTING DOWN: 1.) The Burners vertical adjustment is too high or low for the Flame Sensor to work properly. 2.) There is some other obstruction in between the flame sensor and the burner head. 3.) Tubing for Air or Fuel supply damaged, restricting flow
When aligning the autosampler the pipette tip should be aligned so that the tip delivers the sample about 1 to 2 mm from the bottom of the tube. The tip should be touching the sample as it pulls out of the tube. The inside of the contact rings where the tube touches should be kept clean of deposits. This can be done with a q-tip and some methanol, or a small burnishing tool for more stubborn deposits. Poor contact will cause the furnace not to heat even though it runs through the program. Tube life depending on sample matrix should be about 100-500 cycles. If tube burning is noticed then either the argon is contaminated with air, there is a leak, or there is a severe draft near the instrument (a draft can displace the argon in the furnace area). If poor precision is noticed using the sampler then pipette by hand and see if it clears up. If you get inconsistent results from the same solution, the temperatures may be set too low, or you may need some matrix modifier to stabilize the species in the tube. If your baseline does not return to zero after each run, you are getting a build up of residue in the tube. This should be burned up using the burn out feature, or by doing some blank runs with 10 % nitric acid to oxidize any residual organics or 10% hydrochloric acid to volatilize any metals. Common elements like Na, K, Mg, Ca, Zn and Cu represent contamination and may show poor reproducibility on graphite furnace. Also since they are very sensitive elements by flame there is not much need to use the furnace. Here are some solutions for problems you may encounter while running the 220GF Graphite Furnace:
PROBLEM: FURNACE WILL NOT HEAT UP AS BUCK 210 RUNS THE PROGRAM
1.) If the furnace sounds as if it would when heating up, the tube is not making good contact with the electrode. If the tube is old, replace it and twist the tube around a few times in each electrode before clamping the electrodes together. If necessary, tighten the set screw on the detent a quarter turn to increase tension on the tube. 2.) After a couple of years you may need to replace the electrodes in the furnace head to ensure proper contact. A extra pair of electrodes are included with the instrument. 3.) If the furnace does not make any sound while going through the furnace cycle, a qualified service technician to assess the power circuitry as described in Section 10.3.4.
PROBLEM: SHORT TUBE LIFE 1.) The presence of oxygen in the tube will cause it to deteriorate rapidly. If there is any amount of oxygen in the Argon gas, or if air from the buildings ventilation system is blowing on the furnace the graphite tubes will not last long. 2.) The tube will deteriorate more rapidly at higher temperatures. Since some elements require Atomization temps as high as 2900 degrees, tubes wont last as long when running these elements. You can, however, reduce the time at this temperature without sacrificing a lot of sensitivity by using 2-4 Atomize steps where the second is 2900 degrees for 1 sec, for example, followed by more short steps at decreasing temperatures. 3.) The chemical composition of your samples can cause the graphite tubes to deteriorate faster. The simplest way to reduce this effect is by making use of the low detection limits of the system as best you can and dilute your samples as much as you can. Changing Furnace Transformer Taps There are two transformers in the 220GF Power Unit Box, both with taps for 208 and 240 VAC. A qualified electrical technician can easily change the taps if necessary using the following instructions. 1. Unplug the Power Unit from the 208 or 240 VAC outlet. 2. Remove the 6 screws on the upper half of the power unit box and fold down the top half of the back panel. Be careful not to disturb any of the wiring to the circuit board. 3. The small transformer is located on the bottom left-hand corner of the top compartment. Move the red wire to the appropriate quick disconnect tab. 4. Install the screws back in to the top half of the back panel then remove the six on the bottom. 5. The large transformer provides the large step up in amperage needed to heat the graphite tube. Connect the black transformer wire to the appropriate tab using the 10-32 pan head screw.
SECTION 10.6: Troubleshooting The 220AS Graphite Furnace Autosampler Motor Errors
While working with the Buck 220AS you may encounter motor errors from time to time. This is when the Autosampler stops and both lights are blinking. Normally this condition may be cleared up with a hard reset (performed by pressing and holding the reset button for 5 seconds). This resets the motor position with respect to motor phase. Just pressing and releasing the <reset> button will not clear the error but merely set the mechanisms back to home position. If the motor errors occur repeatedly, youll need to assess which mechanism is at fault by watching to see what the last motion was before the error [arm movement, tray moving left/right (translation), tray rotation or syringe]. Repeated motor errors may also be caused by inconsistent line voltage. If the line voltage is consistantly higher or lower the normal (120 or 240 VAC), you may change the line voltage selector on the power module using the following procedure: Changing Line Voltage: 1.) 2.) 3.) 4.) Unplug the power cord, Remove the fuse holder with a small blade screwdriver Using needle nose pliers carefully pull out the small circuit board. Rotate the board and plastic indicator for the best voltage (preferably + or 7 volts from your actual line voltage) and re-insert the board. 5.) Re-install the fuse holder. In the following troubleshooting guide, we will address each mechanism (as described above) separately: PROBLEM : ARM MOTOR ERROR 1.) Make sure the Autosampler is properly aligned (see section 4.2) such that the probe does not come in contact with the cup or the furnace. 2.) Inspect the mechanism by removing the necessary covers to see if the motion is restricted, or if any belts or pulleys are loose. You must not be able to move the arm with respect to its pulley and chopper or the set screws must be tightened. PROBLEM : SYRINGE MOTOR ERROR 1.) Check the coupler between the syringe and motor. Make sure the drive screw is screwed into the coupler as far as it goes and that its set screw is tight. 2.) Make sure the syringe plunger is into the coupler all the way and that its set screw is secured PROBLEM : TRANSLATION MOTOR ERROR 1.) Make sure the tray assembly slides smoothly back and forth when the power is turned off.
PROBLEM : ROTATION MOTOR ERROR 1.) Make sure the tray screw is tight, securing the tray Here are some other situations you may encounter with simple solutions: PROBLEM: NO SENSITIVITY, NO INJECTION 1.) Autosampler is not aligned properly to pick up sample in the cup. 2.) The rinse water bottle is empty. You will have to press <RESET> several times to purge the system after refilling the bottle. PROBLEM: SIPPER PROBE TIP DAMAGED The teflon sipper tube can be pulled through the probe, then cut off at a 45 degree angle with about 3/8 inch (9.5mm) of teflon tube protruding from the end.
999-2202 Burner head slot cleaner 990-1536 Burner head O-ring [2] 990-1333 Burner head retaining nuts [3] 990-1531 O-ring for X-Y Table [2] 210-5087 Spray Chamber only 210-6070 Spray Chamber Only, 211 210-0552 Spray chamber [complete] - except burner head 210-0542 Burner head Assembly [air/acetylene], 210 210-0541 Burner head Assembly [nitrous oxide], 210 210-0632 Burner head Assembly [air/acetylene], 211 210-0631 Burner head Assembly [nitrous oxide], 211 210-0538 Collar, Burner head 210-0618 Collar, Burner head, 211 200-0088 Impact bead 999-1549 Spray chamber drain elbow 200-1090 Blow out plug + O-ring 999-3083 O-ring for blow out plug [2] 200-0056 Nebulizer - complete [stainless steel] 200-0080 Nebulizer - needle assy only 210-0551 Nebulizer - body only 200-0060 Nebulizer - complete [corrosion resistant] 210-0554 X-Y Table Assembly 210-0606 Chimeny Assembly 210-6007 Sample Tray 210-0051 Glass shield, burner compartment, for old style 210 BS186-0450 0.5 Absorbance screen BS30040 Spares kit - includes nebulizer, impact bead, blow out plug & O-ring, capillary tubing [10 feet], nebulizer cleaning wire [10 pieces], burner slot cleaner, retaining nuts [3] ELECTRONICS SECTION 210-0540 Front panel complete and tested 210-1006 Computer style main power supply 210-2006 Power control assembly (includes chassis) 210-2000 System pcb PART # 210-2002 990-1072 998-1038 DESCRIPTION Keyboard Power Cord Fuses, 100mA for HCL [3]
Fuses, 200mA for D2 [3] Software with library chip Software with library chip for 210 with Furnace Keycap Set (complete) Liquid crystal display Photomultiplier -R928 Lamp socket end piece Lamp socket cap
PNEUMATICS 999-1532 999-1534 999-1555 210-1086 210-0607 OPTICAL 210-1052 200-1053 200-1052 210-5026 210-1116 999-2032 999-3419 TOOLS 999-3127 999-3123 999-1635 999-3124 Hex wrench 3/16 for burner head Hex wrench 5/32 for burner block Hex Driver, 1/4 for Furnace Power Cable 1/16 Allen Wrench D2 lamp Grating Monochromator mirror Lenses Counter, Wavelength Micrometer for wavelength Monocover strap Fuel toggle switch Fuel adjustment valve Regulator [internal air] Solenoid [air/nitrous oxide] Flowmeter [new]-for 210 sn 606 or greater
GRAPHITE FURNACE SUPPLIES 220-0021 Furnace PCB 220-1090 IGBT PCB Assembly BS290-1820 Non-pyro coated graphite tubes [10 pack] BS007-0699 Non-pyro coated graphite tubes [50 pack] BS013-5653 Pyro coated graphite tubes [10 pack] PART # DESCRIPTION BS009-1504 BS220-2697 204-1044 204-1045 990-1537 Pyro coated graphite tubes [50 pack] Platform graphite tubes [10 pack] Left Contact Ring Right Contact Ring O-ring for water connections [2]
O-ring for windows [2] Furnace Communucations Cable Variable Pipettor 10-100uL Cooling Water install kit (ball valve, needle valve ,tube)
FURNACE AUTOSAMPLER BS008-7056 BS011-9079 220-3056 990-1636 990-1072 220-0339 220-0340 220-0341 220-0342 220-3117 220-3118 220-3089 220-3076 Polyethylene sample cups for furnace autosampler Polystyrene sample cups for furnace autosampler Polystyrene 8.5mL Sample Cups (pkg 50) 3/32 Allen Wrench Power Cord Communication cable Adjustable Feet (2) Mounting Hadware Kit Sipper probe 4' of rinse tubing Stiff tube for rinse bottle Rinse bottle Inspection mirror
F G H K
4.0% Phosphate (as ammonium) 1.0% Magnesium nitrate 2.0% Nickel nitrate 0.2% Palladium (as nitrate)
ACCESSORIES BS303-0106 Acetylene Pressure Regulator BS303-0263 Air Pressure Regulator BS303-0264 Argon Pressure Regulator BS303-0265 Hydrogen Pressure Regulator BS303-0204 Nitrous Oxide Pressure Regulator R06-221 Pressure Regulator with Gauge BS303-0407 Hood and Vent Kit, Stainless Steel BS303-0417 Hood and Vent Kit, Stainless Steel (220V) BS303-0313 Oil-less Air Compressor BS303-0314 Oil-less Air Compressor (220V) PART # DESCRIPTION BS303-0229 BS057-2471 6103A BS30055 999-1514 999-1594 990-1855 Air Filter Replacement Cartridge for Filter Flash Arrestor Atomic Absorption Spectrometry by S.J. Haswell Fitting, 1/4 Tube to 1/4 NPT Male Fitting, 1/4 Tube to 1/4 NPT Female Black Flexible Tubing (per foot) with nuts & Ferrules
Blue Flexible Tubing (per foot) with nuts & Ferrules Red Flexible Tubing (per foot) with nuts & Ferrules Nebulizer Capillary Tubing Nebulizer Cleaning Wire Flame Ignitor
TABLE 1: Flame Atomic Absorption Concentration Ranges TABLE 1: Flame Atomic Absorption Concentration Ranges
Wl (Nm) 309.3 217.6 193.7 553.6 234.9 222.8 249.7 228.9 422.7 852.1 357.9 240.7 324.8 421.2 400.8 459.4 368.4 287.4 265.1 242.8 286.6 410.4 303.9 264.0 248.3 550.1 Slit (Nm) 0.7 0.2 0.7 0.7 0.7 0.7 0.7 0.7 0.7 0.2 0.7 0.2 0.7 0.2 0.2 0.2 0.2 0.7 0.2 0.7 0.2 0.2 0.7 0.2 0.2 0.2 Detec Limit (mg/L) 2.00 0.30 0.25 0.50 0.04 0.10 0.01 0.05 0.04 0.05 0.005 0.20 0.05 Sens Check (mg/L) 25 12.5 22.5 10 0.75 10 300 0.75 2 5 2 3.5 2 22.5 15 15 425 30 50 7.5 225 20.0 17.5 250 2.5 1250 Linear Range (mg/L) 50.00 20.00 25.00 25.00 4.00 25.00 450 2.00 5.00 7.50 5.00 5.00 5.00 33.75 22.50 22.50 637.5 45.00 75.00 10.00 337.5 30.00 26.25 375 5.00 1875 Flame Type Color N-A, rich/red A-A, lean/blue A-A, lean/blue N-A, rich/red N-A, rich/red A-A, lean/blue N-A, rich/red-wh A-A, lean/blue N-A, rich/red A-A, lean/blue A-A, rich/yellow A-A, lean/blue A-A, lean/blue N-A, rich/red N-A, rich/red N-A, rich/red N-A, rich/red A-A, lean/blue N-A, rich/red A-A, lean/blue N-A, rich/red N-A, rich/red A-A, lean/blue A-A, rich/yellow A-A, lean/blue N-A, rich/red
Metal Aluminum (Al) Antimony (Sb) Arsenic (As) Barium (Ba) Beryllium (Be) Bismuth (Bi) Boron (B) Cadmium (Cd) Calcium (Ca) Cesium (Cs) Chromium (Cr) Cobalt (Co) Copper (Cu) Dysprosium (Dy) Erbium (Er) Europium (Eu) Gadolinium (Gd) Gallium (Ga) Germanium (Ge) Gold (Au) Hafnium (Hf) Holmium (Ho) Indium (In) Iridium (Ir) Iron (Fe) Lanthanum (La)
TABLE 1: Continued
Metal Wl (Nm) Slit (Nm) Detec Limit (mg/L) Sens Check (mg/L) Linear Range (mg/L) Flame Type Color
Lead (Pb) 283.3 0.7 0.08 10 20.00 A-A, lean/blue Lithium (Li) 670.8 0.7 1 1.50 A-A, lean/blue Lutetium (Lu) 336.0 0.2 125 187.5 N-A, rich/red Magnesium (Mg) 285.2 0.7 0.005 0.015 1.50 A-A, lean/blue Manganese (Mn) 279.5 0.7 0.03 1.25 2.50 A-A, lean/blue Mercury (Hg) 253.7 0.7 ~ 5.0 A-A, lean/blue Molybdenum (Mo) 313.3 0.7 0.80 15 20.00 N-A, rich/red Neodymium (Nd) 492.4 0.2 175 262.5 N-A, rich/red Nickel (Ni) 232.0 0.2 0.05 3.5 4.00 A-A, lean/blue Niobium (Nb) 334.4 0.2 350 525 N-A, rich/red Osmium (Os) 290.9 0.2 22.5 33.75 N-A, rich/red Palladium (Pd) 244.8 0.2 0.15 5 10.00 A-A, lean/blue Phosphorus (P) 213.6 0.2 7000 10500 N-A, rich/red Platinum (Pt) 265.9 0.2 0.80 50 20.00 A-A, lean/blue Potassium (K) 766.5 0.7 0.01 1 3.00 A-A, lean/blue Praseodymium (Pr) 495.1 0.2 1000 1500 N-A, rich/red Rhenium (Re) 346.0 0.2 325 487.5 N-A, rich/red Rhodium (Rh) 343.5 0.2 4.5 6.75 A-A, lean/blue Rubidium (Rb) 780.0 0.7 25 37.5 A-A, lean/blue Ruthenium (Ru) 349.9 0.2 15 22.5 A-A, lean/blue Samarium (Sm) 429.7 0.2 150 225 N-A, rich/red Scandium (Sc) 391.2 0.2 7.5 11.25 N-A, rich/red Selenium (Se) 196.0 0.2 0.50 15 25.00 Ar-H Silicon (Si) 251.6 0.2 1.00 50 50.00 N-A, rich/red Silver (Ag) 328.1 0.7 0.02 1.25 3.00 A-A, lean/blue Sodium (Na) 589.0 0.2 0.005 0.25 2.00 A-A, lean/blue Strontium (Sr) 460.7 0.2 2.5 3.75 N-A, rich/red Tantalum (Ta) 271.5 0.2 275 412.5 N-A, rich/red Technetium (Tc) 261.4 0.2 50 75 A-A, rich yellow Tellurium (Te) 214.3 0.7 10 15 A-A, lean/blue Terbium (Tb) 432.6 0.2 150 225 N-A, rich/red Thallium (Tl) 276.8 0.7 0.40 15 25.00 A-A, lean/blue Thulium (Tm) 371.8 0.2 10 15 N-A, rich/red Tin (Sn) 286.3 0.7 1.00 75 25.00 N-A, rich/red Titanium (Ti) 364.3 0.2 1.00 40 25.00 N-A, rich/red Tungsten (W) 400.9 0.2 0.5 225 337.5 N-A, rich/red Uranium (U) 351.5 0.2 2750 4125 N-A, rich/red Vanadium (V) 318.4 0.2 0.40 45 75.00 N-A, rich/red Ytterbium (Yb) 398.8 0.2 2.5 3.75 N-A, rich/red Yttrium (Y) 410.2 0.2 40 60 N-A, rich/red Zinc (Zn) 213.9 0.7 0.005 0.50 2.50 A-A, lean/blue Zirconium (Zr) 360.1 0.2 150 225 N-A, rich/red NOTE: The notations refer to preferred technique where FAAS is generally not suitable: hg means hydride technique; cv means cold vapor technique
Cobalt
Copper
Dysprosium
Erbium
Europium
396.2 308.2 206.8 231.2 189.0 197.2 350.1 none 222.8 306.8 206.2 227.7 none 326.1 239.9 455.5 359.4 360.5 425.4 427.5 429.0 242.5 241.2 252.1 243.6 304.4 352.7 346.6 327.4 216.5 222.6 249.2 224.4 404.6 418.7 419.5 416.8 386.3 415.1 389.3 408.8 381.0 390.5 462.7 466.2 321.1 321.3 311.1
0.7 0.7 0.2 0.2 0.7 0.7 0.2 0.2 0.7 0.2 0.2 0.7 0.7 2.0 0.7 0.7 0.7 0.7 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2
1.1 1.6 1.5 2.1 0.8 2.0 16.0 2.4 3.7 8.6 14.0 435 120 85 1.7 2.2 3.0 3.8 4.5 1.2 1.8 2.0 2.9 12 22 30 2.0 6.0 15 72 157 1.1 1.2 1.6 6.8 2.7 2.7 5.0 7.0 8.4 20 1.3 1.5 12 15 15
Gadolinium
Gallium
Germanium
Gold Hafnium
Holmium
Indium
Iridium
Iron
407.9 378.3 405.8 405.4 371.4 419.1 367.4 404.5 394.6 294.4 417.2 250.0 245.0 272.0 259.2 271.0 275.5 269.1 267.6 312.3 307.3 289.8 296.5 405.4 416.3 417.3 404.1 410.9 412.7 422.7 325.6 410.5 451.1 256.0 208.9 266.5 237.3 285.0 250.3 254.4 351.4 248.8 302.1 252.7 372.0 373.7
0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.7 0.7 0.7 0.7 0.2 0.2 0.2 0.2 0.7 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.7 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2
1.0 1.1 1.2 1.3 1.7 2.7 2.9 3.2 6.5 1.0 1.4 9.0 9.6 20 2.2 2.4 2.6 3.8 1.8 900
1.3 1.7 4.2 5.2 9.8 11 24 1.0 2.9 3.1 12 0.3 1.2 1.3 1.4 1.7 2.1 8.6 1.7 3.7 4.6 5.7 10
TABLE 2: Continued
Wl Lanthanum 418.7 495.0 357.4 365.0 392.8 217.0 261.4 368.4 323.3 331.2 337.7 356.8 298.9 451.9 202.6 279.8 280.1 403.1 317.0 379.8 319.4 386.4 390.3 315.8 320.9 ? ? ? 231.1 352.5 341.5 305.1 346.2 358.0 334.9 408.0 335.8 412.4 357.6 305.9 263.7 301.8 330.2 247.6 276.3 340.5 214.9 Slit 0.2 0.2 0.2 0.2 0.2 0.7 0.7 0.7 0.7 0.2 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.2 0.7 0.7 0.7 0.7 0.7 0.7 0.2
Rs
1.6 1.7 4.0 4.0 4.0 0.4 10 25 235 1.8 2.0 2.1 9.2 11 24 1.3 1.9 9.5 1.6 1.8 2.0 2.5 3.3 4.0 8.7 Platinum
Wl 306.5 283.0 293.0 273.4 Potassium 769.9 Praseodymium 513.3 473.7 492.5 502.7 Rhenium 346.5 345.2 Rhodium 369.2 339.7 350.2 365.8 Rubidium 794.8 420.2 Ruthenium 372.8 379.9 392.6 Samarium 476.0 511.7 520.1 472.8 Scandium 390.8 402.4 402.0 405.5 327.0 408.2 327.4 Selenium 204.0 206.3 207.5 Silicon 250.7 252.8 252.4 221.7 221.1 Silver 338.3 Sodium 589.6 330.2 Strontium none
Slit 0.7 0.2 0.7 0.2 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 2.0 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.2 0.2 0.7 0.2 2.0
Rs
2.1 3.4 3.7 4.1 2.3 1.4 2.2 2.2 2.5 1.7 2.4 1.7 2.5 3.7 6.0 2.1 120 1.4 2.2 11 1.4 1.4 1.6 2.0 1.0 1.4 1.8 2.7 3.2 7.0 12 3.0 11 35 2.8 3.2 3.7 4.3 8.0 1.9 1.0 185
Lead
Lithium Lutetium
Magnesium Manganese
Molybdenum
Neodymium
Nickel
Niobium
Osmium
Palladium
Phosphorus
0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2
1.5 3.3 3.5 4.5 6.6 1.1 1.2 1.4 1.5 1.9 2.5 1.6 1.8 3.2 3.6 1.0 2.7 3.0 2.0
TABLE 2: Continued
Wl 260.8 265.7 293.4 255.9 265.3 269.8 275.8 260.9 429.7 426.2 318.2 423.8 363.6 317.3 225.9 238.6 431.9 390.1 406.2 433.8 410.5 377.6 238.0 258.0 410.6 374.4 409.4 418.8 420.4 375.2 436.0 341.0 224.6 235.5 270.6 303.4 254.7 219.9 300.9 Slit 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.7 0.2 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0. 0.2 0.2 0.7 0.7 0.2 0.7 0.2 0.7
Rs
2.1 2.5 2.5 2.5 2.7 2.7 3.1 4.1 6.5 8.1 10 11 11 100 15 50 1.2 1.6 1.8 2.0 3.6 2.7 6.7 24 1.4 1.6 1.7 1.9 3.0 5.7 9.3 14 0.5 0.8 2.0 2.8 4.4 4.7 5.9 Tungsten
Tantalum
Technetium
Uranium Vanadium
Tellurium Terbium
Ytterbium
Yttrium
Thallium
Thulium
Zinc Zirconium
Wl 255.1 294.4 268.1 272.4 294.7 283.1 289.6 287.9 430.2 358.5 356.7 306.6 306.0 305.6 320.2 390.2 346.4 246.4 267.2 407.7 412.8 414.3 362.1 307.6 354.8 303.0 301.2 298.5 362.4
Slit 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.7 0.2 0.2 0.2 0.2 0.2
Rs
0.5 0.7 0.7 0.7 0.7 1.0 1.4 2.4 7.2 0.3 0.5 2.4 2.4 3.0 6.4 6.5 3.5 7.5 40 1.1 1.2 1.4 2.0 4700 1.5 1.5 1.7 1.7 1.9
Tin
Titanium
Sample Preparation of Glass and Ceramic Materials for Atomic Absorption Analysis
The analysis of silicate-based glass and ceramics has always been a moderately difficult job, since the components that comprise the glass are usually not stable together in solution. For example, most typical acidic digestions that solubilize silicon will cause calcium fluoride to precipitate. If no fluoride is present, the silicon will not be fully decomposed, and if the solution is heated, boron can be lost, etc. There are several industry standard procedures that are used for specific elements, thereby requiring at least two (2) preparations to be used for the complete, reproducible analysis of each sample. If only certain components are important, then perhaps a single preparation can be employed. Listed below are some methods developed at Buck Scientific for use on the Model 210VGP Atomic Absorption Spectrophotometers. (1) Fuse 0.2 1.0 grams of powdered glass with 5x the amount of a mixed lithium carbonate/zinc oxide flux (2:1) for 10-15 minutes at 900 950C in an Ni or Pt crucible. Cool and dissolve the melt in hydrochloric acid (1:1) with warming. The resultant solution is quantitatively diluted, filtered to remove insoluble silica, and analyzed for Ca, Mg, Al, B, Na, K, Fe, Cu, and Mn. If Cr, V, Ti or Mo are to be determined, the flux should contain ~1/ 2 part sodium peroxide, or the melt can be dissolved in 25% hydrochloric acid and 10% hydrogen peroxide with satisfactory results.
(2) Fuse 0.2 1.0 grams of powdered glass with ~7x the amount of either sodium hydroxide or potassium hydroxide for 10 30 minutes at 400 500C in an Ni or Pt crucible. The melt is cooled, dissolved in water, and promptly analyzed for Si, B, Mo, Ti, V, Zn, halogens, P and Al. (3) Moisten 0.1 0.5 grams of powdered glass with 1ml of alcohol in a Pt or PtFe dish, add ~5ml water, 1ml perchloric acid, 1ml sulfuric acid, and ~10ml hydrofluoric acid. Digest on a steam bath or hot plate until fumes are evolved. Add 1ml sulfuric acid, 10ml hydrofluoric acid, and 210ml methyl alcohol. Take to near dryness (sulfuric acid minimizes undesirable reactions from perchloric acid). Add 520ml of hydrochloric acid, dilute, and analyze. This procedure allows the accurate determination of almost all elements except Si and B. Adding 0.1 0.2% lanthanum or strontium will improve the response for Al, Ca and Mg. The Buck 210VGP series can accurately determine these major and minor elements in ranges from 0.005% to over 20%, depending on the dilution used, with typical precision of 1.5% or better if using bracketed standards. The entire procedure, from preparation of sample to final analysis, can be standardized to a simple format so that a novice lab technician can perform the test easily. Analyst: Gerald J. De Menna
SIC: 321, 322, 3252, 326
NOTE: All values listed below are in Weight Percent (%wt) in the original glass sample; [K] = known assay, [M] = measured data.
Soda-Lime (green) Element [#2 prep] Si Al Zn B [#3 prep] Na K Zn Ca Mg Mn Ba Al Pb Cu Fe [K] 34.2 0.17 0.18 1.42 [M] 34.0 0.15 0.19 1.38
Borosilicate (hard, Hi-B) [K] 37.8 1.21 .011 3.93 [M] 37.2 1.24 .009 3.90 [K] 30.6 0.09 1.11 0.05
10.0 0.49 0.16 6.15 2.35 .015 0.17 0.15 <0.01 0.08 4.08
10.3 0.48 0.18 5.97 2.40 .016 0.15 0.14 <0.01 0.09 3.95
2.97 .012 .006 .007 .008 .004 4.77 1.22 <0.01 .002 .024
3.08 .011 .007 .009 .008 .005 4.61 1.20 <0.01 .002 .022
4.22 6.97 1.07 0.28 0.05 0.11 1.38 0.08 16.3 0.16 0.41
4.25 7.03 1.04# 0.30 0.04 0.12 1.42 0.07* 16.1 0.15 0.42
The above data shows the powerful flexibility and stability of the Buck 210VGP system for the wide-ranging requirements of the glass and ceramics industry. The excellent correlation between the known and measured values ranges from 0.6% to 3.0% (average RSD = 1.6%) demonstrates the precision of the instrument. The fact that two different preparations gave superb reproducibility on two different elements, Zn (#) and Al (*), reveals the high accuracy of the chemistry.
NOTE: All values listed below are in Weight Percent (%wt) in the original glass sample; [K] = known assay, [M] = measured data.
Soda-Lime (green) Element [#2 prep] Si Al Zn B [#3 prep] Na K Zn Ca Mg Mn Ba Al Pb Cu Fe [K] 34.2 0.17 0.18 1.42 [M] 34.0 0.15 0.19 1.38
Borosilicate (hard, Hi-B) [K] 37.8 1.21 .011 3.93 [M] 37.2 1.24 .009 3.90 [K] 30.6 0.09 1.11 0.05
10.0 0.49 0.16 6.15 2.35 .015 0.17 0.15 <0.01 0.08 4.08
10.3 0.48 0.18 5.97 2.40 .016 0.15 0.14 <0.01 0.09 3.95
2.97 .012 .006 .007 .008 .004 4.77 1.22 <0.01 .002 .024
3.08 .011 .007 .009 .008 .005 4.61 1.20 <0.01 .002 .022
4.22 6.97 1.07 0.28 0.05 0.11 1.38 0.08 16.3 0.16 0.41
4.25 7.03 1.04# 0.30 0.04 0.12 1.42 0.07* 16.1 0.15 0.42
The above data shows the powerful flexibility and stability of the Buck 210VGP system for the wide-ranging requirements of the glass and ceramics industry. The excellent correlation between the known and measured values ranges from 0.6% to 3.0% (average RSD = 1.6%) demonstrates the precision of the instrument. The fact that two different preparations gave superb reproducibility on two different elements, Zn (#) and Al (*), reveals the high accuracy of the chemistry.
AA3002: Determination of Trace Lead, Poly-Phenols and Tannins in Wines Using a Single Analytical Instrument The manufacturing of wine and other fermented beverages has typically been a fairly simple process. Historically, the expertise of the sage wine-master was used to gauge the quality of the product. Today, many environmental concerns, health factors, and government rules, prompt the producers of fine wines to monitor several important components in the grape, the fermentation vats, and in the final wine. Typically, several different analytical instruments have been used to perform these tests including Flame Atomic Absorption Spectrophotometers (FAAS), UV Spectrophotometers (UV), and Colorimeters (VIS). The 210VGP Atomic Absorption system from Buck Scientific, Inc. employs a high performance monochromator and a unique cell-holding attachment that allows it to be used as three analytical tools in one: FAAS, UV and VIS. This means that three of the more critical tests can be performed with a single device. The FAAS system normally uses a Lead Hollow Cathode Lamp (HCL) to measure the Lead signal at 283nm, while the UV test for PolyPhenols at 280nm is best done with a D2 (Deuterium) HCL. The VIS determination of Tannins at 520nm uses a Cr (Chromium) HCL and the Buck #5711 cell holder. Tannins can also be measured at 760nm using a Cs (Cesium) HCL. Other test procedures for measurements at 420nm and 620nm can be done with a Rb (Rubidium) HCL. Examples of actual samples are listed on the back. Analyst: Gerald J. DeMenna
SIC: 2084
Napa Varietal Must 8.7 Napa Varietal First Ferment 6.3 Medocino County Sauvignon Blanc 0.93 LaCour Pavillion Bordeaux 1.4
768
2,570
227
982
169
64
403
1,328
AA3003: Determination of Major Components and Trace Contaminants in Assorted Plating Baths by Atomic Absorption Analysis
Precious metal, non-ferrous, and ferrous plating operations cover almost every facet of modern life: automotive parts, kitchen & bathroom fixtures, electrical parts, electronic components, furniture, tools, plumbing, construction supplies, die castings, jewelry, etc. Good manufacturing practices, and lately, government regulations for environmental safety, have prompted many plating operations to perform more stringent tests on their products. This determines the composition of the major elements in the bath, and also verifies the cleanliness of the bath (i.e. no toxic or hazardous metals). Checking the effluent and wastewater from a production operation is also required under most state guidelines. In the past, simple titrimetric or colorimetric tests were sufficient for the determination of major components, but lacked the speed for good quality control methodology, and sensitivity for trace analysis. Atomic Absorption Spectrophotometry has been an accurate and precise analytical tool for many laboratory environments for over 50 years. The general purpose system lacks the spectral interferences of emission spectroscopy techniques and is significantly more sensitive and accurate than methods such as X-ray fluorescence. The Buck 210VGP AA system is substantially less expensive than almost any other analytical tool, while keeping its unique flexibility and versatility.
Lab Requirements: 5.0ml & 10.0ml Class-A Pipets, 100ml & 500ml Class-A volumetric flasks, distilled water, 110V AC, and air & acetylene tanks. Procedure: (1) Pipet 5.0ml of bath into a 100ml flask and dilute with water. Dilution = 1:20 / Use to determine trace elements (~200 ppb) (2) Pipet 10.0ml of 1:20 into a 100ml flask and dilute with water. Dilution = 1:200 / Use to analyze minor elements (0.1% - 0.0005%) (3) Pipet 10.0ml of 1:200 into a 500ml flask and dilute with water. Dilution = 1:10,000 / Use to analyze major elements (10% - 0.1%) Calibration: A 1.0 ppm standard and matrix blank for trace elements. A 10 to 50 ppm standard and a water blank for major / minor elements. Analysis: Insert element lamp, peak in wavelength, optimize energy, set blank zero, calibrate on standard, aspirate sample and read concentration. The Buck 210VGP Series can accurately determine most elements from 100 ppb to over 10%, depending on the dilution, with typical precision levels of 1.5% or better if using bracketed standards. The entire procedure can be standardized to a simple, easily followed format for a lab technician.
SIC: 3471
Analytical Results for Various Plating Baths Samples: Preparation: Calibration: Instrument: Conditions: Three (3) baths from an electronics application, a die-cast factory and a chromefinishing fixtures plant. Three serial dilutions, as previously described. Buck Certified Atomic Absorption Standards. The Buck 210VGP Atomic Absorption Spectrophotometer. Air / Acetylene flame, Integrate mode, Normal parameters.
NOTE All values listed below are in Volume Percent (%v) in the original bath sample unless otherwise noted.
2.3%
1.08%
0.69%
(Traces in PPM, mg/L) Cd Pb Mn Co 79 ppm 228 ppm 24 ppm <5 ppm 2.8% 1.7% 3.9% 554 ppm 130 ppm 472 ppm 326 ppm 1.5% 1.9% 0.7% 27 ppm 115 ppm 2.1% 83 ppm 471 ppm 2.2% 1.8% 2.3% 5.3%
The above data shows the powerful flexibility and stability of the Buck 210VGP system for the wide-ranging requirements of the plating industry. The excellent correlation between major / minor values ranges from 0.7% to 3.2% (average RSD = 1.3%), exemplifying the high precision of the instrument. The overall accuracy of the various trace metals supports the interference-free quality of the data. Together, the combinations provide for an unmatched pair in economy and performance.
AA3004: Analysis of Whole Blood for Trace Lead (Pb) by Graphite Furnace AAS
Toxic effects from heavy metals (Pb, As, Se, Cd, Bi, Sn, Tl, and Hg) have been characterized in medical literature for decades. Originally, complex extractions were performed to isolate the poisonous components from the complex biological matrices: blood, serum, urine and tissues. Recently, optical emission and atomic absorption spectroscopy techniques were developed to determine these toxic elements. As analytical instrumentation performance improved, real limits of detection were also improved. The advent of Inductively-Coupled Plasma (ICP) and Graphite Furnace AAS (GFAAS) brought the limits of analysis into the parts-per-billion (PPB) range for many matrices. This capability has been invaluable in screening blood and urine samples from pediatric and chronically ill patients to track the exposure and elimination of lead. The determination of trace lead in blood samples is very important, particularly in third world and urban environments, where environmental pollution, lead-based paints, gasolines, and poor plumbing systems increase the exposure of lead in various forms. The standard clinical procedure for the measurement of lead in blood is very simple and is based on the use of highly sensitive GFAAS instrumentation. The procedures listed below and the data were developed on and obtained from the Buck 210VGP Atomic Absorption System with the Model 220 Graphite Furnace option. Sample / Standard Preparation: There are several procedures for the quantitative and reproducible preparation of Whole Blood samples prior to atomic analysis. Below is the most popular technique: (1) Collect Blood and preserve with Heparin, EDTA, or Oxalate. (2) Dilute 100L of sample with 400L of Diluent (contains 0.25% Triton X-100, 2000 ppm ammonium dihydrogen phosphate, and 750 ppm magnesium nitrate). These matrix modifiers serve to aid hemolysis, sequester the Pb, and increase atomization efficiency. (3) Make 0, 0.01, 0.05, 0.1, 0.25, and 0.5 mg/L lead standards in diluent. (4) Analyze 20L loads with the following GFAAS conditions: 283.3nm wavelength, 7A slit, D2 background correction, Peak Height, Argon purge (~50ml/min), Auto-Zero off, grooved furnace tube. Dry: Ambient to 125C in 15 second ramp, 5 second hold. Ash: 125C to 600C in 45 second ramp, 20 second hold. Atomize: 600C to 2400C in fast ramp or step, 5 second hold. This technique provides a well-matched calibration equivalent to 40, 200, 400, 1000, and 2000g/L (ppb) lead in the original blood sample. The precision ranges from ~3% relative at the upper limit to ~8% at the low end, with a method detection limit of 25g/L in blood calculated from the blank variance taken at 2-sigma.
SIC: 801, 8062, 8071
The superb precision and accuracy shown by the above data are typical of the quality optical components and engineering design of the Buck AA systems. The industrially rugged nature of this design also allows the instrument to be used in a wide range of environments, from Hospital R&D Centers, to Clinical Testing Laboratories with much ease and confidence.
AA3005: Determination of Wear Metals and Additives (Soaps) in Lubricating Oils by Atomic Absorption
Lubricating & hydraulic fluids are crucial to the proper functioning of many types of machines: automobiles, farm equipment, heavy-duty motors, airplanes, and military hardware. Most lubes are simple hydro-carbon based oils or greases that have certain additives in them to prevent wear and abrasion of components (piston rings, bearings, etc.) or oxidation degradation of the oil. These additives are typically soaps made from an organic acid and a metal (usually barium, zinc, calcium, or magnesium). As the oil ages, these materials are continually lost and need to be monitored so that they can be replaced to provide the proper level of protection. Wear metals are formed in lube oils under the harsh conditions of temperature and pressure that occur in heavy machinery. Under continuous heat and pressure, the surface of the metal piece becomes slightly oxidized, forms salts with the degradation products of the oil, and becomes soluble in the oil. The friction of motion in machinery also causes micro-fine particles to actually shear off the surface and become suspended in the oil. Determining what these wear metals are and their respective levels is very important to the proper maintenance and operation of the equipment. Copper & tin can come from bearing wear, lead & antimony come from Babbit metal bushings, chromium & molybdenum from piston rings and seals, nickel & iron from crankshaft wear and engine body damage, aluminum & silicon from catalyst residues, and sodium & silicon from bad air filters and from water / coolant leaks. Careful monitoring of these materials can be used in preventative maintenance of large, expensive engines, minimizing potential repair costs. One of the best technologies for this type of determination is Atomic Absorption Spectrometry. There are several techniques for the analysis of metals in petroleum matrices by AAS. The simplest is dilution of the oil with a solvent like MIBK or kerosene, and analysis using organo-metallic standards. This solution burns well in an air / acetylene flame and gives very good detection limits. Another calls for the dry ashing of the sample, where the organics are burned off to leave an inorganic residue (ash). The ash is dissolved in acids and analyzed as an aqueous solution. However, significant errors can occur in this method because many metals are easily volatilized (sublimation, distillation), giving low assays. Recently, the advent of microwave digestion technology enables the chemist to decompose the oil with acids under pressure. This results in a clear, aqueous solution that can be analyzed using simple water-based standards. For any of these methods, the Buck 210VGP system and an optional MEGA-1200 Microwave Digestion system are ideally suited for rapid and economical determinations of wear metals and additives in all types of hydrocarbons and lubricant media.
Sample: Preparation:
30W Engine Oil; abused & oxidized, 0.1 10 micron particulates. (1) 10 grams diluted with 3:1 MIBK / Xylene solvent to 100ml (2) 10 grams dry ashed @ 650C, dissolved in acids to 100ml (3) 10 grams microwave digested in HNO3 / HCIO4 / H2O2 to 100ml Buck Certified AA Standards; Aqueous and Organo-Metallic, at 0.5 ppm and 5 ppm levels in appropriate matrices. Buck 210VGP Atomic Absorption Spectrophotometer. Air or N2O / Acetylene flame, normal parameters.
Values are mg / kg (ppm) in the original sample unless otherwise noted. AVG / sd = Average & deviation of 3 preps; D.L. = 3-sigma detectability.
Method (1) Wavelength (nm) AVG / sd 2.7 / 0.4 240 32 / 2.1 324 248 589 283 251 0.25 Method (2) D.L. D.L. 1.9 / 0.8 0.08 115 / 8.5 0.65 24 / 4.7 0.22 47 / 0.9 11 / 1.3 0.03 0.43 Method (3) AVG / sd 2.5 / 0.1 103 / 23 0.12 12 / 0.1 44 / 2.4 5.5 / 0.4 0.17 8.1 / 0.5 0.27 0.25
D.L. 0.01 122 / 6.5 0.20 30 / 1.6 0.03 0.01 0.05 0.02 0.17 0.12 0.15 0.31
357
0.50 12 / 0.5
4.9 / 0.8 0.46 232 8.4 / 0.7 18 / 0.9 0.65 5.4 / 1.1 224 205 / 4.2 0.77
0.34 0.25
214
0.51
0.13
The above data shows the powerful flexibility and stability of the Buck 210VGP system for the wideranging requirements of the wear metals industry. The high precision of the major and trace data exemplifies the accuracy of the instrument. The overall high sensitivity of the various trace metals supports the interference-free quality of the data. Together, the combination provides for an unmatched pair in economy and performance.
AA3006: Determination of Major Electrolytes, Minor Minerals, and Trace Heavy Metals in Physiological Fluids by Flame and Graphite Furnace Atomic Absorption Spectroscopy
Accurate measurements of specific metals in various biological systems is extremely important. These metals can be classified into several groups: the high level major electrolytes (Na, K, Ca, Mg), which are crucial to normal primary physiological processes like cellular activity and metabolism; the minor minerals (Fe, Mn, Zn, Co, Cu), which are present in lower levels to act as metabolic agents and enzyme catalysts; and the trace heavy metals, or toxics, that should not be present in a healthy organism (Ag, Al, As, Se, Pb, Cr, Cd, Hg, Ni, TI, Sr, Ba and V). These last elements are either known to be detrimental to the body, or have a currently unknown activity that is purported to be harmful. Such a listing of elements is prone to rapid changes. Some vitamin metabolic processes and enzymatic reactions appear to need very small amounts of poisonous metals like Se and Cr. Metals like Al and Sn do not have any well-documented harmful effects, but are believed to be related to Alzheimers Disease and kidney dysfunction, respectively. Even certain odd metals like Mo and Au, not normally found in the food system, are found to disturb digestion and skeletal processes. Other elements like Li, Pt, and La are being used in drug products and require monitoring in the body. Obviously, there is a need for an analytical system to be able to differentiate and quantitate these 20+ elements easily and reproducibly. The most economical instrument suitable for this work is the Buck 210VGP Atomic Absorption Spectrophotometer. The Buck 210, based on a unique Stable-Beam technology that gives superior sensitivity and stability over older doublebeam systems, is extremely simple to install and operate. The various excitation sources (air, nitrous oxide, hydride, and graphite furnace) allows analyses to be made in a multitude of matrices (tissue digests, serum, plasma, whole blood, packed RBCs, urine, amnion, CSF, etc.) for metals from the parts-per-billion to the percentage level. Most clinical environments prefer to mg% or mEq/L for the major electrolytes and in mg/dL (or g/dL) for the toxic elements. Chemical considerations are quite simple for handling physiological matrices, particularly if employing the Buck 210AA with its perfectly balanced in-line deuterium lamp background corrector. This feature permits the analyst to use a single, simple aqueous standard for calibration of many different sample matrices, which would have required individual matrixmatched standards for each type of sample. The enclosed data shows a simplified, standardized technique for the preparation of various body fluids, and the conditions for the precise determination of a range of metals.
SIC: 8071
The above data shows the powerful flexibility and stability of the Buck 210VGP systems for the wide-ranging requirements of the Clinical Laboratory. The high precision of the Major and Trace data exemplifies the stability of the instrument. The overall high sensitivity of the various Trace metals supports the interference-free quality of the data. The result is an unmatched system in economy and performance.
AA3007: Determination of the Environmental TCLP Metals in Waste-Waters, Solid Wastes, and Soils by Flame Atomic Absorption Spectrophotometry
Hazardous wastes can originate from many sources industrial processes, product formulations, synthetic operations, virtually any manufacturing business. The nature and composition of the wastes can vary so much that one industrys discharge can serve as anothers raw materials. It is the tremendous transfer of these materials their storage, transportation, spillage and leakage into the environment which has caused great concern. Regulation of these discharged solid and liquid materials is a crucial part of the U.S. EPA and the State-level Departments of Environmental Protection programs for defining the degree of hazard of a particular material. The basic methods describe procedures to leach any soluble heavy metals from samples using acid(s) (to simulate acid rain, for example). While there is a lot of printed material available from these Government agencies, the basic protocols for the collection and preparation of this type of sample is referenced in the Standard Methods of SW-846 as part of the Resource Conservation and Recovery Act (RCRA). Principle The Toxic Characteristic Leach Procedure (TCLP) is a specific protocol to extract and decompose any complex or bound metals in a waste-water, solid waste or soil sample. This makes a solution suitable for atomic analysis by Flame Atomic Absorption Spectroscopy (FAAS) or Inductively Coupled Plasma Optical Emission Spectrocopy (ICP-OES). Flame AAS is an inexpensive, efficient method that has been time tested for many decades. ICP-OES is a newer technique that is faster, but significantly more expensive than FAAS. The sample preparation range from boiling water samples with acid, to shaking a solid sample in an acid slurry for 18 hours, to digesting at controlled temperatures in a fluoro-polymer microwave oven pressure vessel. The protocols for microwave oven assisted leachings are defined in the U.S. EPA Methods 3015 (liquids) and 3051 (solids). Practice An assortment of samples is prepared according to the appropriate methodology, and the resulting solutions are ready for instrumental analysis. The Buck Model 210VGP Flame AA system with the Nitrous Oxide assembly and the Model-420 Hydride unit are set up for running the metals listed in the TCLP program. Accuracy is ensured and detection limits are maximized by using the unique in-line Deuterium Background Corrector and the wavelengthindependent Variable Giant Pulse Corrector (standard parts of the 210VGP) to eliminate any interferences in the sample measurements.
NOTE Values are mg/L (ppm) in the original sample // D.L. = 3-sigma detectability
Element Ag Cd Pb Ni Cr Al Ba Be As Se Sn Hg
Wavelength (nm) 328 228 283 232 357 394 553 234 193 196 286 253
Waste-D.L. 0.04 0.02 0.15 0.10 0.08 0.65 0.35 0.02 0.001 0.002 0.005 0.00001
Water 0.09 0.23 0.72 2.48 1.34 5.63 4.52 0.03 0.28 0.16 0.08 0.0008
Soil 0.36 0.69 1.82 2.17 0.61 (100) 2.43 <0.02 0.13 0.05 0.17 0.0012
Sludge 6.97 1.21 8.03 5.85 3.53 (100) 0.54 0.39 0.74 0.83 1.49 0.0185
The above data shows the powerful flexibility and stability of the Buck 210VGP systems for the wide-ranging requirements of the Environmental industry. The overall high sensitivity of the various Trace metals supports the interference-free quality of the data. The combination of components provides for an unmatched system in economy and performance.
AA3008: Indirect Determination of Gold Purity by Measurement of Impurities with Flame Atomic Absorption Spectroscopy
The determination of the purity of scrap and refined gold is a very important procedure for the financial markets, the jewelry business and the electronics industry. Over the years, the classical fire assay technique has been used for gold determination but this is a very laborintensive procedure, and prone to many operator errors. Hundreds of books have been written with special steps and secret protocols to follow for accurate and reproducible fire assays; but it still is a subjective method and is sometimes used more for semi-quantitative analyses for those labs without experienced asssayers. Trying to determine the exact concentration of gold instrumentally can be done directly by measuring the gold itself, but preparing an accurate dilution of a sample is very difficult unless the volumetrics are calibrated, the water is thermostatted and the balance is certified. Slight changes in temperature, or vessel cleanliness or calibration can result in percentage errors in the final data. The error of the measuring instrument (Gravimetric balance, Arc / Spark, Flame-AA, Plasma Emission, etc) can also affect the final value by defining a limiting precision; for example: D-C Plasma has a systematic error of 1.5%; so the best data for gold off a DCP instrument will have a built-in error, or variability, of 1.5%. A better, and more useful procedure, is determining the total impurities in the sample and subracting this from 100% to get the gold by difference. Many large companies (Engelhard, Texas Instruments, Krementz) use this procedure to assay their incoming gold. Reporting errors can be reduced to less than 0.1% in some case for a more accurate purity determination of the gold material. Atomic Absorption Spectroscopy is a very acceptable technique for gold assays, and is a very simple technique for technicians and chemical operators to learn. Sample preparation and calibration are easily performed according to a standardized procedure, and the operation of Buck AA instruments takes only a few hours to master. Performing these analyses is actually simple on the Buck Model 210VGP Atomic Absorption system. The high optical throughput and stability of our Stable-Beam design provide excellent sensitivity, superb precision and, with the proprietary Background Correction modes on the 210, complete freedom from interferences. The unique single-mirror monochromator used in the optical system is the key to the systems small size and high performance. It is the incorporation of the in-line Deuterium Background Correction that will minimize or eliminate most types of spectral or spectrochemical interferences without any loss in data quality. The data shown on this appnote is a clear statement to the stability and energy of the Buck AA instruments. Analyst: Gerald J. DeMenna
SIC: 1041, 367, 3911, 6011
Analysis of Gold Purity by Measurement of Impurities Samples: Preparation: Standards: Results: [1] Smelted scrap (~75% Au) [2] Refined ore (~99% Au) 1.0gm in 25ml Aq. Regia, fumed with HCl, dilute to 100gm. A Hydrochloric Acid Blank (0), and a 1 ppm (g / gm) Standard. Values are ppm (g / gm) in the original solid sample, with RSD calculations on triplicate runs (using 1:100 dilution factor).
Element Silver Cadmium Chromium Copper Iron Nickel Lead Tin Zinc
Wavelength (nm) 328 228 357 324 248 232 283 286 214
Flame Type Air even Air lean Air rich Air even Air lean Air lean Air even N2O rich Air even
Sample #1 / RSD 5220 ppm / 1.3% 1740 ppm / 2.1% 120 ppm / 2.6% 51850 ppm / 0.8% 690 ppm / 3.0% 82700 ppm / 1.2% 21500 ppm / 1.6% 60700 ppm / 3.2% 10350 ppm / 0.9% 234,870 (23.49%) 76.51% Au
Sample #2 / RSD 3.8 ppm / 2.5% <0.9 ppm / <2.5 ppm / 15 ppm / 1.8% <3.0 ppm / 33 ppm / 1.7% 12 ppm / 2.4% 265 ppm / 4.7% 21 ppm / 1.3% 365 (0.036%) 99.964% Au
AA3009: Analytical Methodology for the Characterization of Steels and Iron Alloys by Atomic Absorption Analysis
The analysis of irons and steels has always been a very important job in the industrial and commercial environment. The grading of steels and iron alloys is critical to characterizing their respective strengths to corrosion and stress. There are also many protocols for the several hundred types of steels available. Fe is always the primary component, with elements such as Ni and Cr as major consstituents, and Mn, Zn, B, P, Cu, V, Ti, Co, Pb, Mo, Si, and Sn as essential minor elements. Much of the standard methods of analyses have been documented and issued by agencies like the ASTM (American Society of Testing and Materials), and the ISO (International Standards Organization). Some of the procedures listed here are taken from the ASTM Standards, volume 03.05 (Chemical Analysis of Metals); methods E350 (Low Alloy Steels), E351 (Cast Irons), E352 (Tool and High Alloy Steels, and E353 (Stainless Steels). All these standards employ Flame Atomic Absorption Spectrophotometry (FAAS). Preparation of the samples and calibration of the instrumentation is actually quite easy using the Buck Scientific Model 210VGP AA system. The unique components allow interference-free measurements at high levels of sensitivity and precision. Aspiration of the sample by the high efficiency nebulizer provides tremendous sample throughput, and the proprietary inline Deuterium (D2) Lamp Background Corrector gives the highest energy of any commercially available system. These features combined with the single-mirror, high-resolution monochromator and low noise electronics create an integrated analytical tool of remarkable power. An example of a universal sample preparation technique is listed below. This procedure gives complete recovery of all components with no losses or chemical interferences, and can be used for Certified Standards from ASTM, NIST / NBS, BCS, IUPAC, or other regulatory agencies. Two basic dilutions cover the range of concentrations from 0.001% up to 25.0%. Preparation of Irons and Steels for Flame AAS Analyses: Aqua Nasty [1] Weigh 1.0gm sample chips into a tared 4oz HDPE Nalgene bottle. [2] Add 10ml HCl, 2ml HNO3, 1ml HF, and 1gm K2S2O8. Heat in water bath or in microwave oven for 5 10 minutes. [3] Dilute to 100grams final weight, or cool and dilute to 100ml volume. This is the 1% solution (1:100 dilution factor) for minor elements. [4] Weigh or pipet 1gm or 1ml of this to 100ml to make 100 ppm solution (1:10,000 dilution factor) for Major components. The Buck 210VGP AA can accurately determine these major and minor elements, depending on the dilution used, with typical precision levels of 1% or better if using bracketted standards. The entire procedure can determine over 40 elements per hour for high turnover and high quality results.
NOTE All values listed below are in Weight Percent (%wt) in the original Steel sample; [K] = Known assay, [M] = measured data.
Element
[M]
18.0 13.0 1.75 0.25 0.89 0.23 .008 2.10 0.45 0.68 0.22 .029 .036 -
[M]
0.70 0.20 0.19 4.80 0.24 .080 0.47 0.21 0.12 .060 .029 .030 .039 -
Cr Ni Mn Co Al Cu Ag Mo Ti Si V Pb Zn Sn Mg Bi As
2.45 1.17 0.79 0.54 0.35 0.67 .011 0.48 0.17 2.86 0.15 0.17 0.88 .013 .097 .046 0.12
The above data shows the powerful flexibility and stability of the Buck 210 system for the wide-ranging requirements of the Iron and Steel industry. The excellent correlation between the Known and Measured values ranges from 0.6% to 3.0% (average RSD = 1.6%), exemplifying the high precision of the instrument. The fact that a single preparation gives superb reproducibility on three (3) widely varying metals, with a range of difficult elements, reveals the accuracy of the chemistry. The above components provide for an unmatched system in economy and performance.
AA3011: Evaluation of Mineral Supplements for Content and Purity by Flame / Graphite Furnace AAS
Doctors, dieticians and health officials recommend a balanced diet that provides essential minerals for proper nutrition in the human body. The accurate analysis of metals in vitamin / mineral preparations and food supplements is very important. These metals can be classified into several groups the major electrolytes (Na, K, Ca, Mg), crucial to normal primary physiological processes like cellular activity and metabolism; the minor minerals (Fe, Mn, Zn, Cu), present in lower levels to act as metabolic agents and enzyme catalysts; and the micro (trace) minerals, in low levels (Se, Cr, Mo), for less defined reactions. There is a need for an analytical tool to quantitate these elements easily and reproducibly. Ferrous gluconate is a source of iron, but can be contaminated with manganese, titanium and vanadium. Dolomite is an excellent source of calcium and magnesium, but is often contaminated with lead and silver. Zinc oxide provides an essential form of easily absorbed zinc, though the amount of cadmium and arsenic found can be a problem. Examining some of the common over-the-counter and prescription formulations, one can see that the potential contaminants present in some of these higher concentration minerals is as important as the accurate determination of these essential minerals themselves. For this type of low-level analysis in such a complex matrix (high organics and alkali elements), graphite furnace atomic absorption spectroscopy (GFAAS) with deuterium background correction (D2) is one of the few acceptable techniques for error-free determination of trace metals. GFAAS is also the preferred technique for measuring the desired micro-minerals. The electrolyte metals and minor minerals are best done by simple flame AA (FAAS). Both flame and furnace techniques can suffer from tremendous interferences when low level elements are determined in dirty or complex matrices. The alkali elements (Na, K, Mg, Ca) and the refractory metals (Al, Si) can create background effects that will add signficant inaccuracies to an analysis. The presence of organic materials in the sample matrix can cause the formation of smoke particulates in a graphite furnace analysis and cyanogen bands in a flame method; which also contribute to analytical errors. The nature of these interferences can be compensated for by the use of an efficient D2 Background Correction system; and the unique design of the Buck 210VGP AA system provides the maximum energy throughput for maximum sensitivity and accuracy.
Evaluation of Mineral Supplements for Content and Purity by Flame / Graphite Furnace AAS
Samples:
Preparation:
1) An enriched Wheat Flour; 2) A Multi-Vitamin & Mineral Capsule; 3) A Dolomitic Limestone [a natural Calcium and Magnesium source]; and 4) A Mineral Supplement Concentrate Dissolve 1gram of sample or 1 capsule in 100ml 5% Nitric Acid. Filter and dilute to 50ml to make a 1:50 dilution, or a 2% solution to be used for trace elements. Dilute 1ml of this to 100ml for a 1:5,000 sample dilution for the electrolytes and major metals. Buck Certified standards at 5 ppm [Flame] and 0.1 ppm with modifiers for [Furnace]. Buck 210VGP AA with the 220 Graphite Furnace and 420 Hydride accessories as noted. Flame for Na, K, Mg, Ca, Cu, Zn, Fe, Mn Hydride for Se and As Furnace for Cr, Mo, Ag, Pb, and Cd Values are noted as Percent, ppm, or ppb in the original sample; or mcg in the capsule.
Element Sodium Potassium Magnesium Calcium Copper Iron Manganese Zinc * Arsenic Selenium Chromium * Lead * Silver * Cadmium Molybdenum
Wavelength 589 nm 766 285 422 324 nm 248 257 214 194 nm 196 357 nm 283 328 228 313
Detection Limit 0.5 ppm 2.0 ppm 0.05 ppm 1.0 ppm 0.05 ppm 0.10 ppm 0.10 ppm 0.05 ppm 10 ppb 25 ppb 50 ppb 25 ppb 10 ppb 5.0 ppb 50 ppb
Flour 2,575 ppm 4,150 ppm 375 ppm 1,500 ppm 21 ppm 315 ppm 46 ppm 185 ppm <10 ppb 69 ppb 3.8 ppm <25 ppb 1.7 ppm <5 ppb 2.4 ppm
Capsule <10 mg 15 mg 20 mg 100 mg 120 mcg 100 mg 1.5 mg 5 mg <5 mcg 7.9 mcg 12.5 mcg <10 mcg <5 mcg <2 mcg 9.3 mcg
Lime 0.83% 0.24% 9.82% 17.2% 0.09% 0.35% 0.18% 1.04% 47.5 ppm 16.3 ppm 0.33% 0.29% 0.12% 665 ppm 0.09%
Conc 255 ppm 810 ppm 175 ppm 75 ppm 550 ppb 1070 ppm 58 ppm 245 ppm <10 ppb 78 ppb 155 ppb <25 ppb <10 ppb <5 ppb 165 ppb
Sample and Standard Preparation procedures for Trace Elements in Lead [1] Preparation of Lead (Pb) Samples: Drill Pb metal to make small pieces. Weigh 2.0 grams of these pieces on a balance. Place the sample in a 400ml Pyrex breaker and add 10ml H2O, 2.5 grams of tartaric acid and 7ml HNO3. Warm on a hot-plate until the sample dissolves. Dilute to 100ml in a volumetric flask with water (H2O). This prepares a 2% sample solution, with a dilution factor of 50. [2] Preparation of Sulfuric Acid (H2SO4) Samples: Using a graduated cylinder, place 75ml of water (H2O) into a 100ml volumetric flask. Slowly add sulfuric acid (H2SO4) to the water, swirling to mix. Solution will bubble up and get hot, do NOT spill it on yourself wear gloves! Add H2SO4 to bring the final volume to 100ml. This prepares a 25% sample solution, with a dilution factor of 4. [3] Preparation of Pure / Waste Water (H20) Samples: Using a digital pipettor or glass pipet, add 1ml nitric acid (HNO3) to 100ml volumetric flask. Add water sample to 100ml mark. There is basically no dilution and the energy relates directly to the sample. [4] Preparation of Multi-element Standards for Calibration: Using a pipet or pipettor, add 100ml of each of the 1000ppm stock concentrate standards (Buck standards) to a 1 Liter volumetric flask containing 250ml water (H2O) and 50ml nitric acid (HNO3). The following groupings will prepare stable 100ppm Stock Standards . (Add water to bring final volume to 1 Liter): A) Bi, Ni, Ag, Zn, Cu, Cd B) Fe, Sb, Sn, As, Al, Ca Dilute the 100ppm Stock Standards into the following Working Standards: 25ml Stock + 1ml HNO3 to 100ml in Volumetric Flask = 10ml 5ml 2ml 1ml 0.5ml
Use these standards for Cu, Ag, Fe, Ca: 0, 0.5, 2, 5ppm. Use these standards for Ni, Zn, Cd: 0, 0.5, 1, 2ppm. Use these standards for Bi, Sb, Sn, As, Al, Pb: 0, 5, 10, 25ppm. Analyst: Gerald J. DeMenna
SIC: 173, 351, 353, 361, 362, 366, 3691, 3692, 3714, 3724, 5013
Instrument: Conditions:
Results:
Lead Sample #19, Lead Sample #21, Sulfuric Acid (~98%) 2% solutions of lead in 5% HNO3 / 2.5% tartric acid; 10% solution of sulfuric acid (1:10 dilution) 0.5 and 2.5 g / ml (ppm) analyte metal standard in 2% high-purity lead matrix, 2% high-purity lead matrix blank; for lead samples. 1.0 g / ml (ppm) analyte metal standard in pure (distilled / deionized) water, pure water blank; for sulfuric acid sample. Buck 210VGP Atomic Absorption Spectrophotometer, Giant Pulse and In-Line D2 Correction, and Model 420 Hydride Generation system. Standard operating conditions for 210 unit; analytical parameters and correction modes as listed per element; air / acetylene flame for Ni, Ag, Zn, Cu, Fe, Cd; nitrous oxide / acetylene for Al, Ca; argon hydrogen for As, Sb, Sn, Bi. Values are weight percent (% w/w) in original sample: Data based on 1:50 Pb dilution and 1:10 H2SO4 dilution: D.L. [detectability] based on 2-sigma statistics for Pb samples.
Wavelength 232nm 328 213 324 223 248 217 224 193 228 309 422 D.L. 0.0015% 0.0003 0.0003 0.0005 0.0008 0.001 0.0005 0.0007 0.0001 0.0005 0.005 0.0005 Lead # 19 <0.0015% 0.0017 0.0009 0.0150 0.0167 <0.001 <0.0005 <0.0007 <0.0001 0.0006 <0.005 0.0006 Lead # 21 0.0018% 0.0018 0.0004 0.0093 0.0184 0.002 0.0012 0.0009 0.0011 0.0009 <0.005 0.0022 H2SO4 <0.0015% 0.0005 0.0027 0.0021 <0.0008 0.005 <0.0005 <0.0007 <0.0001 0.0008 0.006 0.0472
Element Ni Ag Zn Cu Bi Fe Sb Sn As Cd Al Ca
These data show the powerful flexibility and stability of the Buck 210VGP system for the wide-ranging requirements of the manufacturing industry. The overall high sensitivity of the various trace metals supports the interference-free quality of the data. The combination of unique components provides an un-matched system in performance and economy.
This represents a composite protocol based on several State DEP and USEPA Standard Methods for the Collection, Preparation and Storage of Potable Water]. (1) Take a 100.0ml of the Sample, add 20ml concentrate (~65%) Nitric Acid; transfer to a plastic vessel (beaker, wide-mouth bottle); heat in a lab-guide microwave oven at low-medium power for ~10 - 15 minutes or until the volume is reduced back down to ~100ml. Cool and transfer to a Volumetric Flask, re-dilute to 100.0ml final volume. Buffer a portion of the Sodium for Flame-AAS with 1,000ppm Lanthanum (La) and Lithium (Li) for Ca, Mg, and Ba work; Matrix Modify another portion of GFAAS work with 1,000ppm Magnesium Nitrate and 500ppm Palladium Nitrate (per USEPA).
(2)
Calibration:
A linear, 2-point Calibration was made using the appropriate Matrix Blank for the Flame or Furnace and the Standards set within the CAL MAX range for each metal. Buck Model 210/211 AAS with attachments as described below: * Air Flame for Na, Mg, Ca, Fe, Cu, Cr * Furnace for Pb, Cr, Ba, Al, As, Fe * Cold Vapor for Hg. * Nitrous Flame for Ba, Al * Hydride for As
Instrument: Conditions:
Values are g/L (ppb) unless otherwise noted, with [%RSD} on triplicate measurements. Element
Na Mg Ca Fe Cu Ba Al Cr Pb As Hg
Detection Limit
0.005ppm 0.005ppm 0.025ppm 0.05ppm / 2ppb 0.005ppm / 1ppb 0.5ppm / 5ppb 2ppm / 25ppb 0.02ppm / 3ppb 0.8ppb 0.001ppm / 1ppb 0.02ppb / (100ml)
CAL MAX
2.0ppm 3.0ppm 5.0ppm 5ppm / 250ppb 5ppm / 250ppb 25ppm / 500ppb 50ppm / 1000ppb 5ppm / 250ppb 250ppb 1.0ppm / 250ppb 2.0ppb
METHOD
AIR-Flame AIR-Flame AIR-Flame AIR / GF AIR / GF N2O / GF N2O / GF AIR / GF GFAAS HYD / GF 1018-CV
SMP-1 / NY
34ppm [2.1] 1.3ppm [1.9] 0.6ppm [2.2] 0.48ppm [3.1] 0.23ppm [1.6] 375ppb [4.1] 160ppb [2.9] 745ppb [2.3] 2.1ppb [2.0] N/D N/D
SMP-2 / CA
11ppm [1.8] 2.6ppm [2.3] 1.2ppm [2.1] 0.29ppm [2.5] 0.09ppm [1.4] 625ppb [3.5] 215ppb [3.1] 460ppb [2.1] 3.4ppb [2.2] N/D 0.04 [6.7]
SMP-3 / NJ
73ppm [1.5] 1.9ppm [1.3] 1.7ppm [1.8] 0.18ppm [2.2] 0.36ppm [1.5] 170ppb [4.4] 97ppb [2.7] 255ppb [1.7] 1.1ppb [1.6] 8.4ppb [3.4] N/D
The above data shows the powerful flexibility and stability of the Buck 210/211 AA systems for the wide-ranging requirements of the water laboratory. The high precision of the Major and Trace data exemplifies the stability of the 210GF. The overall high sensitivity of the various Trace metals supports the interference-free quality of the data. The result is an unmatched system in economy and performance.
The three modes of operation used on the Buck 210 are Flame Absorption, Flame Emission and Graphite Furnace. The Graphite Furnace methods may also be used for Cold Vapor / Hydride operation. The flam_215.exp file is for use on older instruments which use V2.15 firmware in the Buck 210. Load each method file individually by clicking on its file name then clicking [OK]. In the field labeled New Method Name, enter a name for the method (i.e. Flame Absorbance Method) to identify it in the Analyze Program, Then click [OK]. Repeat the process for any other methods you will be needing. 1) Click on [system] then [configure methods]. Click on the new method name then click [OK]. Click on the [Instrument] button then the [rs-232] button and enter your computer serial port information and a baud rate (the maximum baud rate on the Buck 210 is 38400). Click on the [OK] button 3 times then click on [system] again. Repeat this step for all methods loaded above. Running the software 1) Load the analysis you wish to run into a lamp position on the 210VGP. Leave the instrument in absorbance mode since all concentration values will now be calculated with Analyze/Buck. 2) Turn on your computer and click on the [Analyze] icon with your mouse to start the program. 3) Click on the [run] button (the leftmost button on the tool bar). Select the method for the Buck 210VGP by highlighting its method name. If you wish to work on a worksheet that has already been created, highlight the worksheet number or description to bring this worksheet up, then click on the [OK] button. If you wish to create a new worksheet, click on the [new] button at the bottom of the window. 2) Type a description and operator name for the worksheet. Although this information is not required, it will allow you to identify the worksheet so you can look at it later. The time, date and worksheet numbers are set by the software and cannot be changed by the user. When youre done, click on the [OK] button. The software will setup the header on the worksheet automatically when data is first sent from the Buck 210VGP. 3) If you are running an Absorbance method, you will need to setup sample identifiers for your standards. If you are running a Concentration method, you may skip this step. Click on options then Analyze2 then setup to get to the standard curve setup screen. Click on the AutoCalc on button so the software automatically calculates a concentration value for each absorbance reading. Click on the [setup] button next to Blank Correction. Youll see the default sample identifier for the blank is Blank. Do not change the cell location, but you may change the ID. Click [OK] when finished. Click on the [setup] button for the Standard Curve. Here youll see a table with default sample ID s S1-S20. Youll need to type in values into the concentration column for each of the standards youll be using and make note of the ID. You may change the sample ID to make it easier to differentiate between standards. There is also options to set up concentration ranges, QC standard Checks, etc. on these screens. Consult the Analyze instruction manual for information on these options. When finished click [OK]. NOTE: IF YOU ARE RUNNING WITH VER 2.15 SOFTWARE ON THE 210VGP FOLLOW INSTRUCTIONS UNDER A BELOW, IF YOU ARE RUNNING VER 3.48 OR HIGHER FOLLOW INSTRUCTIONS UNDER B THEN BOTH VERSIONS CONTINUE WITH STEP 6.
) With the communications cable connected from the Buck 210VGPs Comm Port to the computers serial port, Click on the [start] button (the one with a green arrowhead on it) below the Aanalyze softwares menu bar. This function will initialize communication with the instrument, but will not start the analysis on the Buck 210VGP. To start the instrument, press the [print] key on the Buck 210VGP. Use the down arrow key to move the cursor to the [printer] option, then use the right arrow key to turn it off. Use the up arrow key to move the cursor to the [comm port] option, then use the right arrow key to turn it on. Press the up arrow key to highlight [start report] and press [enter]. The instrument will return to the active analysis screen and your ready to run your standards and samples. B) With the communications cable connected from the Buck 210VGPs Comm Port to the computers serial port, Click on the [start] button (the one with a green arrowhead on it) below the Aanalyze softwares menu bar. This function will initialize communication with the instrument, but will not start the analysis on the Buck 210VGP. To start the instrument, press the [cntls] key on the Buck 210VGP, highlight [remote communication] then press [enter]. Use the right arrow key to move the cursor to the remote mode option, then use the [right] arrow key to change to [library/report] then [enter], select proper baud rate using [left/right] arrows then press [enter], then [esc] twice to return to the main screen. Press [start] on the 210 to run your standards & samples. 1) As you read your standards and samples in absorbance mode, a value for absorbance will appear in the Analyze worksheet a few moments after the absorbance value appears on the Buck 210VGPs active analysis screen. When you autozero the Autozero delta (representing the change in absorbance after autozero) will appear. For best results, Autozero on your blank first, read your blank and then your other standards in the order of increasing concentration so it will be easy to identify them when your ready to calibrate. For best results, enter the appropriate sample ID for each blank and standard as they are read so you dont count erroneous readings as standards. Use the value for reading the blank in your calculation and not the autozero delta. You must enter sample IDs exactly as they are written in the set up table (they are case sensitive) and all blanks and standards must have an ID or they will not be considered in the calibration. You may enter the sample IDs at any time. You may also wish to enter sample ID s for you sample as well, just be sure you dont use any sample ID s from the standard curve setup table. 2) If you are not using AutoCalc, youll need to calculate concentration values by clicking on [options] on the menu bar. Select [Analyze2], then [Calculate]. Click on [yes] when you are prompted to update the worksheet. Click on [OK]. If you wish to print the calibration curve, click on the [File] option on the menu bar. Select [print], then [exit to print the calibration curve]. When you return to the worksheet, concentration values for each standard and sample will be displayed on the worksheet. To print out the worksheet, Click on the [print] button (the one with a picture of a printer on it) below the menu bar. When youre done, Click on the [stop] button (marked with a red square on the tool bar). 8) If you have more work to do on this worksheet, Click on the [close] button before exiting the program or opening another worksheet. If youre done working with this particular worksheet, you should save it to the hard drive using the [archive] function. While the worksheet is open, Click [Samples] on the menu bar, then select [Archive]. Enter a label for the worksheet so that you can identify it for future reference. You may also enter an optional operator name and description for the archive file. Click on the [OK] button. The worksheet will automatically be deleted from the method and stored in the archive file. If you do not archive the worksheets the program will use more memory since the worksheets are loaded into RAM with the program.
9) To retrieve a worksheet that has been archived, follow the procedure for starting a new worksheet. After the worksheet has been created, click on [Samples] on the menu bar, then select [Unarchive]. Select the worksheet you wish to retrieve and then click on the [OK] button. 10) If you are done with the worksheet but do not want to save it, you should delete it from memory by clicking on [Worksheet] on the menu bar, then select [Delete worksheet]. If the worksheet you wish to delete is not currently running, youll have to bring it back up to delete it. It is important to either archive or delete worksheets once they are complete because building up too many worksheets may cause the software to exceed the memory requirements of your computer.
7) From the cntls menu highlight report and display controls and press [enter]. Turn read key reporting, read key report header and calibration reporting to on. The printer can be on or off depending if you wish an immediate hard copy or not. Under the cntls menu you can turn read key sampling on if you wish to use the sample identifiers page. 8) Calibrate your standards and run your samples as usual. The information will appear on the computer in the same format as on the printer. NOTE: Once the comm is active, you cannot change any analytical parameters until you finish/exit. 9) When finished go back into the cntls screen and change remote communication controls back to off and return to active analysis. 10) To process the ascii file in Excel, close the Procomm program by pressing esc and alt x and start the Excel program. 11) Select file then open and set files of type to text files/.prn. Browse thru the Procomm directory and select the data file you named above. 12) Excel will prompt you with the text import wizard then follow these steps. Highlight delimited and press next, highlight tabs and press next then select general and finish. 13) Edit the information as needed and save.
END OF MANUAL