Instruction for Use

IFU/01, Rev 0, Date 12.10.2022

Otoscope

Intended for visual examination of the outer ear canal and tympanic membrane (eardrum) by direct viewing
through the ear opening.

CAUTION: • If you suspect that liquid or moisture could have entered the
handle, it must not be charged under any circumstances. This
• If you do not plan to use the device for a long time or if you
travel with it, remove the rechargeable batteries from thecould lead to a life-threatening electric shock, especially in the
handle. case of plug-in handles.

• New batteries should be inserted once the light intensity ofTothe
extend the service life of the battery, the battery should
instrument becomes weaker, even on a full charge. only be charged once the light intensity of the instrument has
• To obtain the best possible light output we recommend always weaker.
become
fitting high quality batteries
Instruction for Use
IFU/01, Rev 0, Date 12.10.2022

Otoscope

with a moist cloth until visually clean. Wipe disinfection as


specified by the disinfectant manufacturer. Only disinfectants
CLEANING AND DISINFECTION with proven efficacy should be used, taking into account the
Cleaning and disinfection of the medical devices serves tonational requirements. After disinfection, wipe the instrument
protect the patient, the user and third parties and to preserve thedown with a moist cloth to remove possible disinfectant
value of the medical devices. Due to the product design and theresidues. The components that come into contact with the skin
materials used, no defined limit can be specified for thecan be rubbed down with alcohol or a suitable disinfectant.
maximum number of reprocessing cycles that can be carriedPlease Note!
out. The life span of the medical devices is determined by their
• Never immerse the instrument heads and handles in
function and by gentle handling of the devices. Defective
products must undergo the reprocessing procedure described liquids! Take care to ensure that no liquids get inside
before being returned for repair. the casing!
Cleaning and Disinfection • This item is not approved for automated
The instrument heads and handles can be cleaned externally reprocessing and sterilization.

These instructions provided above have been validated by the medical device manufacturer as being CAPABLE of preparing a
medical device for re-use. It remains the responsibility of the processor to ensure that the processing is actually performed
using equipment, materials and personnel in the processing facility to achieve the desired result. This requires validation and
routine monitoring of the process. Likewise, any deviation by the processor from the instructions provided should be properly
evaluated for effectiveness and potential adverse consequences.
Symbols
Symbols may be used on some package labeling for easy identification.

TECHNICAL SUPPORT:
 In case of defective and malfunctioning products, please inform to Asha Surgical Industries at 20,
NEAR JAROUT CHOWNK, VILLAGE MANDAUR AMBALA CITY, HARYANA-134003, India.
 In case of adverse incident, inform competent authority of Member state in which the user or
patient is established.
 Dispose these products as per Hospital Norms or Country Regulations.